Drager Babytherm 8000 Service Manual

Dräger Medical, Inc.
Technical Service Bulletin # 3
Product: Babytherm 8000 OC Date: November 30, 1999 Re: Securing the Radiant Heater RH 600 Reference Doc: IDM no. 4 / 10.97 Reason: Unsecured fixing screws of the Radiant Heater RH 600 can loosen
Solution: The 4 fixing screws of the Radiant Heater RH600 shall be secured
with Loctite 221 in the course of servicing.
Devices affected: All Babytherm 8000 OC units having Radiant Heater RH600. When: At next Service Call or Preventive Maintenance. This action was to
be completed by February 28, 1998. Please verify that all units have been modified.
Additional info: This TSB is identical to TSB RH 600W # 1 Cost: Covered by Customer, post warranty. Ordering Info: 79 01 966, screw locking agent, low, Loctite 221 Distribution: Dräger Service Personnel and Authorized Service Organizations for
NCS products.
If you have any questions, please contact Technical Support by phone at 1-800-543-5047 or by fax at 1-215-723-5935
Dräger Medical, Inc. Technical Product Manager
Andreas Lenke
6132.200 Babytherm 8000 WB / OC TSB No. 3 November 1999 1/1