NIR-Online GmbH
Altrottstrasse 31
D-69190 Walldorf
E-mail: info.nir-online@buchi.com
NIR-Online reserves the right to make changes to this manual as required on the basis of future insights,
especially with respect to layout, illustrations and technical detail.
This manual is copyright. Information from it may not be reproduced, distributed, or used for competitive purposes, nor made available to third parties. The manufacture of any component with the aid of
this manual without prior written agreement is also prohibited.
Page 3
NIR-Online GmbHTable of contents
Table of contents
1About this document........................................................................................................... 5
1.1Mark-ups and symbols ..........................................................................................................5
9.2.3 Spare parts specifications.........................................................................................30
ivOperation Manual X-Sential™
Page 5
NIR-Online GmbHAbout this document | 1
1About this document
This operation manual is applicable for all variants of the instrument.
Read this operation manual before operating the instrument and follow the instructions to ensure safe and trouble-free operation.
Keep this operation manual for later use and pass it on to any subsequent user or
owner.
NIR-Online GmbH accepts no liability for damage, faults and malfunctions resulting
from not following this operation manual.
If you have any questions after reading this operation manual:
u Contact NIR-Online GmbH Customer Service.
service.nir-online@buchi.com
1.1Mark-ups and symbols
NOTE
This symbol draws attention to useful and important information.
R This character draws attention to a requirement that must be met before the in-
structions below are carried out.
u This character indicates an instruction that must be carried out by the user.
ð This character indicates the result of a correctly carried out instruction.
Mark-upExplanation
WindowSoftware Windows are marked-up like this.
TabTabs are marked-up like this.
DialogDialogs are marked-up like this.
[Button]Buttons are marked-up like this.
[Field names]Field names are marked-up like this.
[Menu / Menu item]Menus or menu items are marked-up like this.
StatusStatus is marked-up like this.
SignalSignals are marked-up like this.
Operation Manual X-Sential™5/34
Page 6
2 | SafetyNIR-Online GmbH
2Safety
2.1Proper use
The sensor is used to analyze substances and samples in production and in the laboratory. The sensor is intended exclusively for that purpose.
The sensor can be used in laboratories and production facilities for the following operations:
Quality control
Process optimization
Reference measurements
2.2Use other than that intended
The use of the instrument other than described in proper use and specified in technical data is use other than that intended.
The operator is responsible for damages or hazards that are caused by use other
than that intended.
Specially the following uses are not permitted:
Use of the sensor in areas which require explosion-safe devices.
2.3Warning notices in this document
Warning notices warn you of dangers that can occur when handling the device.
There are four danger levels, each identifiable by the signal word used.
Signal wordMeaning
DANGERIndicates a danger with a high level of risk which could result in
death or serious injury if not prevented.
WARNINGIndicates a danger with a medium level of risk which could result in
death or serious injury if not prevented.
CAUTIONIndicates a danger with a low level of risk which could result in mi-
nor or medium-severity injury if not prevented.
NOTICEIndicates a danger that could result in damage to property.
6/34Operation Manual X-Sential™
Page 7
NIR-Online GmbHSafety | 2
1
2.4Warning and directive symbols
The following warning and directive symbols are displayed in this operation manual
or on the instrument.
SymbolMeaning
General warning
Dangerous electrical voltage
Material damage
Read manual
Fig.1: Location of safety notices and warning signs on the instrument
1
General warning
Clean Unit when dust exceeds 5
mm thickness
2.5Staff qualification
Unqualified persons are unable to identify risks and are therefore exposed to greater
dangers.
The device may only be operated by suitably qualified persons.
These operating instructions are aimed at the following target groups:
Operation Manual X-Sential™7/34
Page 8
2 | SafetyNIR-Online GmbH
Users
Users are persons that meet the following criteria:
They have been instructed in the use of the device.
They are familiar with the contents of these operating instructions and the appli-
cable safety regulations and apply them.
They are able on the basis of their training or professional experience to assess
the risks associated with the use of the device.
Operator
The operator is responsible for the following aspects:
The instrument must be correctly installed, commissioned, operated and serviced.
Only suitably qualified staff may be assigned the task of performing the opera-
tions described in these operating instructions.
The staff must comply with the locally applicable requirements and regulations for
safe and hazard-conscious working practices.
Safety-related incidents that occur while operating the instrument are to be re-
ported to the manufacturer.
service.nir-online@buchi.com
NIR-Online service technicians
Service technicians authorized by NIR-Online have attended special training courses
and are authorized by NIR-Online GmbH to carry out special servicing and repair
measures.
2.6Personal protective equipment (production)
Follow the rules regarding personal protective equipment that are applicable at the
installation site.
The operation of the sensor does not require additional protective equipment.
2.7Personal protective equipment (laboratory)
Depending on the application, hazards due to heat and/or corrosive chemicals may
arise.
u Always wear appropriate personal protective equipment such as safety goggles,
protective clothing and gloves.
u Make sure that the personal protective equipment meets the requirements of
the safety data sheets for all chemicals used.
2.8Modifications
Unauthorized modifications may impair safety and lead to accidents.
u Use only genuine NIR-Online accessories, spare parts and consumables.
u Technical modifications to the instrument or accessories should only be carried
out with the prior written approval of NIR-Online GmbH and only by authorized
NIR-Online service technicians.
NIR-Online GmbH accepts no liability whatsoever for damage arising as a result of
unauthorized modifications.
8/34Operation Manual X-Sential™
Page 9
NIR-Online GmbHProduct description | 3
1
2
3
3Product description
3.1Description of function
The sensor is an optical instrument for nondestructive determination of substances
and concentrations in a sample.
A sample absorbs and reflects light across the entire wavelength spectrum according
to its color and chemical composition. The signal reflected by the sample is recorded
and analyzed by a spectrometer.
The sensor uses a lamp to produce near-infrared radiation that interacts with the
molecules of the sample. The interaction between sample and light produces a
characteristic spectrum.
The light reflected from the sample is collected through two sets of fiber optics
that direct the light to the NIR and visible spectrophotometers, respectively. The
visible spectrophotometer consists of a diffraction grating to spatially disperse the
light according to wavelength and a silicon photodiode array containing multiple
elements which measure the light intensity for specific wavelength intervals. The
NIR spectrophtometer consists of a diffraction grating to spatially disperse the
light according to wavelength and an indium-gallium-arsenide photodiode array
containing multiple elements.
The produced measurement results are converted to data sequences.
The data sequences are transferred to a computer via an interface.
A computer program compares the data sequence curve with a calibration model
and in that way determines the chemical composition of the sample.
3.2Configuration
3.2.1Front view
Fig.2: Front View
1Measurement window2Flange
3Heat sink
Operation Manual X-Sential™9/34
Page 10
3 | Product descriptionNIR-Online GmbH
1
2
3
4
5
6
3.2.2Rear view
Fig.3: Rear view
1Journal Button2Power and scanning indicator light
3Type plate (part)4Ground connection
5Power and signal connector6Type plate (part)
3.3Scope of delivery
NOTE
The scope of delivery depends of the configuration of the purchase order.
Accessories are delivered as per the purchase order, order confirmation, and delivery note.
(Equipotential bonding)
10/34Operation Manual X-Sential™
Page 11
NIR-Online GmbHProduct description | 3
5
6
7
8
15
14
1
2
3
4
9
13
12
10
11
3.4Type plate
The type plate identifies the instrument. See Chapter3.2.2 "Rear view", page10
Fig.4: Type plate
1Company name and address2Product option
3Revision number4Product name
5Serial number6Production date
7IP class8ATEX information
9Power consumption
(nominal)
10 Power consumption
(maximum)
11 Ambient temperature12 Current draw
(maximum)
13 Certificates14 Operating voltage
15 Product type
The following product options are possible:
LetterOption
ANIR
DVIS
EGold reflector
(X-One)
GLamp position rev. 1.3.2
NSystem temperature 0 - 80 °C
OHumidity sensor 0-100% RH
SX-Quvette
(Fiber lens (approx. 0 - 2 cm))
Operation Manual X-Sential™11/34
Page 12
3 | Product descriptionNIR-Online GmbH
LetterOption
VExternal button
X2X-Beam 002
(approx. 15 cm measurement distance)
3.5Technical data
3.5.1Sensor
SpecificationsX-Sential
Dimensions (W x D x H) 200 x 200 x 100 mm
Weight5 kg
Max. operating pressure 30 bar at flange
Product temperature
-10 °C to +130 °C
(temperature at flange
with water cooling)
Product temperature
-10 °C to + 70 °C
(temperature at flange
without water cooling)
Vibrations0.2 G at 0.1 - 150 Hz
Wavelength spectrum
900 - 1700 nm; 11100 - 5880 cm
NIR range
Wavelength spectrum
350 - 900 nm; 28500 - 11100 cm
Visible range
DetectorDiode array
Average measurement
20 spectra/s
time
IP CodeIP69 / IPX9K
Type of lampTungsten-halogen dual lamp
Lifetime lamp18000 h (2 x 9000 h)
Minimum clearance on
100 mm
all sides
-1
-1
Connection Voltage85 to 264 VAC
Frequency50/60 Hz
Power consumption30 W
Temperature stabilization
ASDC (Advanced Spectral Drift Control): active temperature control to ±1°C from set system operating temperature. Deviations will lead to automatic white reference
measurement to account for spectral drifts.
12/34Operation Manual X-Sential™
Page 13
NIR-Online GmbHProduct description | 3
3.5.2Installation box
SpecificationsInstallation box
Dimensions (W x D x H)300 x 300 x 167mm
Weight
(excluding cables)
Weight
(inc. cables, 2 x 10 m)
Frequency50/60 Hz
Power consumption30 W
Power supply85 - 264 VAC
3.5.3Ambient conditions
Max. altitude above sea
level
Ambient temperature-10 °C ≤ Tamb ≤ +40 °C
Max. relative air humidity
Storage temperaturemax. 45 °C
3.5.4Materials
ComponentMaterials of construction
6 kg
7.4kg
2500 m
< 90 % non-condensing
CasingAluminium (nickel coated), SS 316L 1.4404 flange
SealsNBR (standard sealing material)
FFKM (optional)
Operation Manual X-Sential™13/34
Page 14
3 | Product descriptionNIR-Online GmbH
3.5.5Software
The sensor is controlled via the SX-Suite software package. It consists of the following components:
NameDescriptionTypical usageUserOccurence
SX-ServerInstrument
driver / usage
of special functions
NameSpecial function DescriptionUserOccurence
SX-ServerConveyor beltOptimized for
MixControl end-
Sample movement detection
Read out instrument status
Setup of instrument hardware
measurement
of moving objects on a conveyor belt
point of mixing
processes
Verify sample
flow
OperatorAs required
NIR adminFor installation
and maintenance
NIR adminAs required
NIR adminAs required
NIR adminAs required
NameDescriptionTypical usageUserOccurence
SX-CenterUser interface
(online/lab
mode)
SX-BackupData backup
scheduler
Recipe/product
and calibration
mangement
View results (table, trend,
charts, reports)
Reference data
management
Automated
backup of measurement data,
results and calibrations
3.5.6Computer system requirements
The system requirements for the computer are as follows:
Operating systemWindows 10 Pro
Cental processing unitIntel Core i5 generation 6600 or later
OperatorDaily workflow
(if not fully automated)
NIR adminDuring installa-
tion
RAMAt least 4 GB
14/34Operation Manual X-Sential™
Page 15
NIR-Online GmbHProduct description | 3
Hard disk spaceAt least 80 GB free disk space
Use a hard disk suitable for continuous operation.
Data backup
Network or external
hard disk
Screen resolutionAt least 1280x1024
LANAt least 1 x 100 Mbit/s LAN
USB 2.0/3.0At least 1 USB connection per sensor and 1x USB per Data-
PCI/PCIe1 slot for Profibus card
SoftwareWord and Microsoft Excel 2003 or later
At least 0.5 GB free disk space
Additional 20 MB per day and sensor
Lab I/O box
(for Profibus connection)
3.5.7Installation site (laboratory)
The installation site has a firm, level surface.
The installation site meets the safety requirements. See Chapter2 "Safety",
page6
The installation site has enough space that cables can be routed safely.
The installation site has no obstacles (e.g. water taps, drains, etc.).
The installation site has an own mains outlet socket for the instrument.
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site allows that the power supply can be disconnected at any time
in an emergency.
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter3.5 "Technical data", page12
3.5.8Installation site (production)
The installation site has a firm, level surface.
The installation site meets the safety requirements. See Chapter2 "Safety",
page6
The installation site has enough space that cables can be routed safely.
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter3.5 "Technical data", page12
The installation site meets the specifications according to the installation point.
See Chapter5.1 "Establishing installation point", page17
The installation site has a own mains outlet socket for the instrument.
The installation site has a sample removal point at a distance of < 1 m.
The installation site allows a direct product measurement.
The installation site has constant product flow.
The layer thickness of the product to be measured is at least 30 mm.
Operation Manual X-Sential™15/34
Page 16
4 | Transport and storageNIR-Online GmbH
4Transport and storage
4.1Transport
NOTICE
Risk of breakage due to incorrect transportation
u Make sure that all parts of the device are safely packed in such a way as to pre-
vent breakage, ideally in the original box.
u Avoid sharp movements during transit.
u After transportation, check the device for damage.
u Damage that has occurred in transit should be reported to the carrier.
u Keep packing for future transportation.
4.2Storage
u Make sure that the ambient conditions are complied with (see Chapter3.5 "Tech-
nical data", page12).
u Wherever possible, store the device in its original packaging.
u After storage, check the device for damage and replace if necessary.
16/34Operation Manual X-Sential™
Page 17
NIR-Online GmbHInstallation | 5
40
60
R6
0
15
°
15°
140
5Installation
5.1Establishing installation point
The fixing points or bolts conform to M6 A2-70/7.3 Nm.
Establish the installation point according to the specified data of the flange.
Fig.5: Dimensions of flange
Operation Manual X-Sential™17/34
Page 18
5 | InstallationNIR-Online GmbH
< 1 m
1
2
3
4
5
5.2Installation point in piping system (example)
Fig.6: Configuration
1Flow restrictor2Bypass
3Analyser4Sample removal point
5Flow restrictor
18/34Operation Manual X-Sential™
Page 19
NIR-Online GmbHInstallation | 5
ProductionServer/Control room
Remote computer
Installation box
with mains adapter
Sensor cables
(power supply,
RS 422)
USB converter
1 sensor per measurement
point
Up to 4 sensors per computer
Process control
system
SX-Client
Display:
- Results
- Charts
Chemometric
software
manual calibration
SX-Plus
Remote access via Ethernet
Frame
grabber card
USB
Optional
interfaces:
-Profibus
-4-20 mA/1-9 V
-TCP/IP
-OPC/XML/SQL
-Modbus
Control center
Product flow
Data cable
(RS422)
5.3Installation (example)
Operation Manual X-Sential™19/34
Page 20
5 | InstallationNIR-Online GmbH
5.4Sensor installation
The following installation positions are possible:
Fig.7: Fixing sensor with screws
Tools required:
Torque wrench, size Torx T30
Tightening torque: 8.4 Nm +-1
Precondition:
R Installation point has been established. See Chapter5.1 "Establishing installation
point", page17.
R The fixing points or bolts conform to M6 A2-70 15 mm
u Fix the sensor to the installation point using the bolts.
20/34Operation Manual X-Sential™
Page 21
NIR-Online GmbHInstallation | 5
5.5Connecting the sensor
Tools required:
Torque wrench, size 7 mm AF
Torque wrench, size Torx T20
NOTE
Make sure that the power is not switched on when connecting the sensor.
NOTE
Loss of performance due to use of unsuitable device cables
Max. cable length between installation box and sensor 10 m.
u Put the cable guard over the sensor ca-
ble.
u Connect the sensor cable to the sensor.
u Secure the connector.
Tightening torque: 2.5 Nm ± 0.5
u Attach the cable guard to the sensor.
Tightening torque: 2 Nm ± 0.5
u Attach the ground cable to the sensor.
Operation Manual X-Sential™21/34
Page 22
5 | InstallationNIR-Online GmbH
5.6Establishing electrical connections
NOTICE
Risk of instrument damage because of not suitable power supply cables.
Not suitable power supply cables can cause bad performance or an instrument damage
u Use only BUCHI power supply cables.
Precondition:
R The electrical installation is as specified on the type plate.
R The electrical installation is equipped with a proper grounding system.
R The electrical installation is equipped with suitable fuses and electrical safety fea-
tures.
R The installation site is as specified in the technical date. See Chapter3.5 "Techni-
cal data", page12
u Connect the power supply cable to the connection on the instrument. See Chap-
ter3.2 "Configuration", page9
u Connect the mains plug to an own mains outlet socket.
22/34Operation Manual X-Sential™
Page 23
NIR-Online GmbHOperation | 6
6Operation
The instrument is operated via the SX-Suite software on a computer. See "SX-Suite
User Manual " and "SX-Plus User Manual ".
6.1Journal button
Pressing the button generates a journal entry.
6.2Entering reference data in the journal
To perform a calibration and continuously check the calibration, reference data is required.
Continuous checking of the calibration is performed according to the requirements
of the production process.
NOTE
The journal entry is identified by date and time.
u Press and hold the journal button for one second.
ð The connected software creates a journal entry.
u Remove the sample at the sample removal point.
u Mark sample with date, time and sensor number.
u Carry out a laboratory analysis.
u Insert the reference data in the journal for creating the calibration model. See
"SX-Suite User Manual " and "SX-Plus User Manual "
Operation Manual X-Sential™23/34
Page 24
7 | Cleaning and servicingNIR-Online GmbH
7Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this section.
Any servicing and repair work which involves opening up the casing may only be
carried out by NIR-Online service technicians.
u Use only genuine NIR-Online consumables and spare parts in order to ensure cor-
rect operation of the device and preserve the warranty.
7.1Notes on servicing
NOTICE
Risk of property damage due to failure to remove dust from heat sink
If dust is not cleaned off the heat sink it may cause the sensor to fail.
u Make sure that the layer of dust is no thicker than 5 mm.
7.2Regular maintenance work
ComponentActionInterval
Casingu Wipe down the casing with a damp
cloth.
Warning symbolsu Check that the warning symbols on the
sensor are legible.
u If they are dirty, clean them.
OpticsNOTICE!Have operation carried out by
NIR-Online service technician
u Replace lamps.
CasingNOTICE!Have operation carried out by
NIR-Online service technician
u Check and replace seals
Weekly
Weekly
Annually
Annually
24/34Operation Manual X-Sential™
Page 25
NIR-Online GmbHTaking out of service and disposal | 8
8Taking out of service and disposal
8.1Disposal
The operator is responsible for proper disposal of the instrument.
u When disposing of equipment observe the local regulations and statutory re-
quirements regarding waste disposal.
u When disposing, observe the disposal regulations of the materials used. Materials
used see Chapter3.5 "Technical data", page12
8.2Returning the instrument
Before returning the instrument, contact the NIR-Online GmbH Service Department.
service.nir-online@buchi.comand ask for an RMA number.
Declares, that the products/Erklärt, dass die Produkte/Déclare par la présente que les produits/Dichiara che i prodotti/Declara que los productos:
Process AnalyserX-Sential
comply with the requirements of the European Directives/denAnforderungen der Europäischen Richtlinien/estconforme auxexigences des directives européennes/soddisfa i requisitidelle normeeuropee /cumple los requerimientos de las Directivas Europeas:
IEC 61326-1:2012
Electrical equipment for measurement, control and laboratory use.EMC requirements:General requirements.
EN 60529 (VDE 470 Teil1):2014-09EN 60529:1991 + A1:2000+ A2:2013
Ingress protection testing –IP69 / IP69k
RoHS2 directive 2011/65/EC
Walldorf‚July1st, 2021
Michael MarkusMichael EckertManaging DirectorProductManagement
9Appendix
9.1Certificates
9.1.1CE certificate
26/34Operation Manual X-Sential™
Page 27
NIR-Online GmbHAppendix | 9
9.2Spare parts and accessories
NOTE
Any modifications of spare parts or assemblies are only allowed with the prior written permission of NIR-Online GmbH.
9.2.1Accessories
Order no.
USB-RS422 interface11060741
Analog interface (DataLabIO)11060742
Profibus card
PCI, High Profile
Profibus card
PCI, Low Profile
Profibus card
PCI Express, High Profile
Profibus card
PCI Express, Low Profile
9.2.2Mounting accessories
Mounting accessories are hardware interfaces between the instrument and the
process. Depending on the setup, specific mounting accessories might be needed for
an implementation into the production facility.
Weld-in Flange
Provides the ability to remove instrument while keeping the
process sealed.
Flange with sapphire window and purge port.
Adapter plate, ø140/106 mm, for wall thick-
ness up to 8,5 mm
11060743
11062589
11063000
11063001
Order no.Image
11060754
Material: Stainless steel DIN 1.4404
(SST316L) / DIN 1.4571 (SST316Ti)
Sealing material FFKM White G74S, FDA com-
pliant 15°C (+59°F) to 260°C (+500°F)
Operating pressure -0.5 to 30 bar. Max. pres-
sure 100 bar short term
Purge port M5 (ø4mm tube adapter needed)
to prevent condensation or detect leakage
High grade sapphire crystal optical lens, pol-
ished for reduced adhesion
Dead volume max. 60 mm³
Operation Manual X-Sential™27/34
Page 28
9 | AppendixNIR-Online GmbH
Order no.Image
Weld-in Flange Pipe
Flange with sapphire window and purge port for installation
in pipes or bended surfaces.
Outer diameter: 140 mm.
Material: Stainless steel DIN 1.4404 (SST316L)
Sealing material: FFKM White G74S
Operating pressure: -0.5 to 30bar. Max. pres-
sure 100 bar short term
Purge port M5 (ø4mm tube adapter needed)
to prevent condensation or detect leakage
The pipe diameter has to be specified upon or-
der
Weld-in Flange Hopper
Flange with sapphire window and purge port for installation
in hopper or bended surfaces with different diameters.
Outer diameter: 140 mm
Material: Stainless steel DIN 1.4404 (SST316L)
Sealing material: FFKM White G74S
11068800
11068801
Operating pressure: -0.5 to 30bar. Max. pres-
sure 100 bar short term
Purge port M5 (ø4mm tube adapter needed)
to prevent condensation or detect leakage
The upper and lower hopper diameter has to
be specified upon order
Weld-in Plate
For instruments in direct contact with the product.
Plate with opening, fitting to instrument flange.
Dimensions: 160 x 241 x 3 mm
Material: DIN 1.4301 (SST304)
Thread bolts M6
Bypass Sampler
For free flowing goods (mealy / grainy).
Bypass with feeder and sampling point.
Pneumatic sampler (min. 5 bar / 72.5 psi wa-
ter or oil free compressed air DIN ISO 8573
Class 1)
11060753
11061670
Screw-conveyor (feeding capacity 1.5 t/h)
Motor (380V/50Hz ATEX A22 0.25 KW)
Requires bypass-installation box and a Data-
Lab IO device
28/34Operation Manual X-Sential™
Page 29
NIR-Online GmbHAppendix | 9
Order no.Image
X-Square
For all free flowing powders and granulates.
The X-Square can be inserted in the product stream or bypass.
Inspection panel (Plexiglas)
Adapted for Jacob pipes Ø150 mm
Stainless steel DIN 1.4301 electro polished
X-Cell DN50, Standard Flange DN50, PL1, 10 bar
For gas, liquid and paste-like products.
The cell can be inserted in the product stream or bypass.
Material DIN 1.4404 (SST316L)
Sealing material: FFKM White G74S
Operating pressure up to 10 bar (145 psi). TÜV
certificate upon request
Measurement slit 26 mm, configurable be-
tween 1 and 15 mm with additional adapter
DN 50 flange (other sizes upon request)
Clearance volume max. 120 mm³ Cells can be
customized with different diameter and
flanges
11061669
11063018
There are various dimensions of the X-Cell available in the
pricelist.
X-Cell 4 Edge DN50
For liquid and paste-like products.
The cell can be inserted in the product stream or bypass.
Material: DIN 1.4404 (SST316L)
Sealing material: FFKM White G74S
Operating pressure: max 3bar
Flange: DN50
Path length: 34mm
Typical use: Wine Applications
11068822
Operation Manual X-Sential™29/34
Page 30
9 | AppendixNIR-Online GmbH
Order no.Image
VARINLINE Sensor Adapter Flange, Type N, 10bar
For opaque products like powder or granules.
In combination with a path length adapter also for transparent liquid, gel or pasty products.
Material DIN 1.4404 (SST316L)
Sealing material FFKM White G74S (FDA com-
pliant), or custom
Operating pressure up to 10 bar (145 psi). TÜV
certificate upon request
Build for DN50 DIN 32676, process connection
type N
Product temp. -14 °C(+5 °F) to 230 °C (+446 °F)
Path length adapter configurable between 0,5
to 42 mm
Path Length Adapter
To measure transparent liquids with the X-Cell.
The reflector reduces the length of the optical path.
Material DIN 1.4404 (SST316L)
11061674
Gap 1 / 2 / 5 / 10 / 15 mm available
Diffuse or polished surface
Water Cooler Flange
Can be used with all instruments, only in com-
bination with X-Cell or Weld-in Flange
Product temperature above 70 °C to 130 °C. A
flow rate of 5 l water per hour at 20 °C is required
40 °C over temp switch for external alarm pur-
pose, NO (Normally Open) circuit
Water connectors for 8/6 mm hose
9.2.3Spare parts specifications
Power supply
NOTICE
Risk of property damage due to incorrectly connected mains adapter
An incorrectly connected mains adapter may cause the sensor to fail.
11060752
u Make sure that the current limiter is set to more than 4.5 A.
u Make sure that the voltage is 12.7 VAC.
30/34Operation Manual X-Sential™
Page 31
NIR-Online GmbHAppendix | 9
Specification
Power supply input voltage: 85 - 264 ± 10% VAC
Rated voltage: 12 VDC
Rated current: ≥ 4.5A
Typical peak-to-peak residual ripple: 50 mV
Max. peak-to-peak residual ripple: 200 mV
Device cables
NOTE
Loss of performance due to use of unsuitable device cables
Max. cable length between installation box and sensor 10 m.
Sensor Cable
Pin assignment on device connector viewed from rear of instrument:
Fig.8: Pin assignment
1PIN 1- blue, ground2PIN 2- red, 12.7 VDC
3PIN 3- green, RxD-4PIN 4- yellow, TxD+
5PIN 5- white, TxD-6PIN 6- brown, RxD+
7PIN 7- not connected8PIN 8- not connected
Operation Manual X-Sential™31/34
Page 32
9 | AppendixNIR-Online GmbH
1
23
4
RS422 Data Cable
Pin assignment on the Moxa viewed from rear of the Moxa: