Buchi NIR-Online X-Sential Operation Manual

Page 1
NIR-Online
®
X-Sential™
Operation Manual
11594282 | A en
Page 2
Product Identification: Operation Manual (Original) X-Sential™ 11594282
Publication date: 08.2021
NIR-Online GmbH Altrottstrasse 31 D-69190 Walldorf E-mail: info.nir-online@buchi.com NIR-Online reserves the right to make changes to this manual as required on the basis of future insights, especially with respect to layout, illustrations and technical detail. This manual is copyright. Information from it may not be reproduced, distributed, or used for competi­tive purposes, nor made available to third parties. The manufacture of any component with the aid of this manual without prior written agreement is also prohibited.
Page 3
NIR-Online GmbH Table of contents
Table of contents
1 About this document........................................................................................................... 5
1.1 Mark-ups and symbols ..........................................................................................................5
2 Safety....................................................................................................................................6
2.1 Proper use .............................................................................................................................6
2.2 Use other than that intended ................................................................................................. 6
2.3 Warning notices in this document.......................................................................................... 6
2.4 Warning and directive symbols.............................................................................................. 7
2.5 Staff qualification ................................................................................................................... 7
2.6 Personal protective equipment (production)..........................................................................8
2.7 Personal protective equipment (laboratory)........................................................................... 8
2.8 Modifications..........................................................................................................................8
3 Product description.............................................................................................................9
3.1 Description of function ...........................................................................................................9
3.2 Configuration ......................................................................................................................... 9
3.2.1 Front view ...................................................................................................................9
3.2.2 Rear view..................................................................................................................10
3.3 Scope of delivery .................................................................................................................10
3.4 Type plate............................................................................................................................11
3.5 Technical data ..................................................................................................................... 12
3.5.1 Sensor ......................................................................................................................12
3.5.2 Installation box..........................................................................................................13
3.5.3 Ambient conditions ...................................................................................................13
3.5.4 Materials ...................................................................................................................13
3.5.5 Software....................................................................................................................14
3.5.6 Computer system requirements................................................................................14
3.5.7 Installation site (laboratory).......................................................................................15
3.5.8 Installation site (production)......................................................................................15
4 Transport and storage ......................................................................................................16
4.1 Transport ............................................................................................................................. 16
4.2 Storage ................................................................................................................................16
5 Installation..........................................................................................................................17
5.1 Establishing installation point ..............................................................................................17
5.2 Installation point in piping system (example).......................................................................18
5.3 Installation (example) ..........................................................................................................19
5.4 Sensor installation ............................................................................................................... 20
5.5 Connecting the sensor.........................................................................................................21
5.6 Establishing electrical connections......................................................................................22
6 Operation............................................................................................................................23
6.1 Journal button......................................................................................................................23
6.2 Entering reference data in the journal .................................................................................23
7 Cleaning and servicing ..................................................................................................... 24
7.1 Notes on servicing ...............................................................................................................24
7.2 Regular maintenance work..................................................................................................24
Operation Manual X-Sential™ iii
Page 4
Table of contents NIR-Online GmbH
8 Taking out of service and disposal.................................................................................. 25
8.1 Disposal...............................................................................................................................25
8.2 Returning the instrument ..................................................................................................... 25
9 Appendix ............................................................................................................................ 26
9.1 Certificates...........................................................................................................................26
9.1.1 CE certificate ............................................................................................................26
9.2 Spare parts and accessories ...............................................................................................27
9.2.1 Accessories ..............................................................................................................27
9.2.2 Mounting accessories ...............................................................................................27
9.2.3 Spare parts specifications.........................................................................................30
iv Operation Manual X-Sential™
Page 5
NIR-Online GmbH About this document | 1
1 About this document
This operation manual is applicable for all variants of the instrument. Read this operation manual before operating the instrument and follow the instruc­tions to ensure safe and trouble-free operation. Keep this operation manual for later use and pass it on to any subsequent user or owner. NIR-Online GmbH accepts no liability for damage, faults and malfunctions resulting from not following this operation manual. If you have any questions after reading this operation manual:
u Contact NIR-Online GmbH Customer Service.
service.nir-online@buchi.com
1.1 Mark-ups and symbols
NOTE
This symbol draws attention to useful and important information.
R This character draws attention to a requirement that must be met before the in-
structions below are carried out.
u This character indicates an instruction that must be carried out by the user.
ð This character indicates the result of a correctly carried out instruction.
Mark-up Explanation
Window Software Windows are marked-up like this.
Tab Tabs are marked-up like this.
Dialog Dialogs are marked-up like this.
[Button] Buttons are marked-up like this.
[Field names] Field names are marked-up like this.
[Menu / Menu item] Menus or menu items are marked-up like this.
Status Status is marked-up like this.
Signal Signals are marked-up like this.
Operation Manual X-Sential™ 5/34
Page 6
2 | Safety NIR-Online GmbH
2 Safety
2.1 Proper use
The sensor is used to analyze substances and samples in production and in the labo­ratory. The sensor is intended exclusively for that purpose. The sensor can be used in laboratories and production facilities for the following op­erations:
Quality control
Process optimization
Reference measurements
2.2 Use other than that intended
The use of the instrument other than described in proper use and specified in tech­nical data is use other than that intended. The operator is responsible for damages or hazards that are caused by use other than that intended. Specially the following uses are not permitted:
Use of the sensor in areas which require explosion-safe devices.
2.3 Warning notices in this document
Warning notices warn you of dangers that can occur when handling the device. There are four danger levels, each identifiable by the signal word used.
Signal word Meaning
DANGER Indicates a danger with a high level of risk which could result in
death or serious injury if not prevented.
WARNING Indicates a danger with a medium level of risk which could result in
death or serious injury if not prevented.
CAUTION Indicates a danger with a low level of risk which could result in mi-
nor or medium-severity injury if not prevented.
NOTICE Indicates a danger that could result in damage to property.
6/34 Operation Manual X-Sential™
Page 7
NIR-Online GmbH Safety | 2
1
2.4 Warning and directive symbols
The following warning and directive symbols are displayed in this operation manual or on the instrument.
Symbol Meaning
General warning
Dangerous electrical voltage
Material damage
Read manual
Fig.1: Location of safety notices and warning signs on the instrument
1
General warning
Clean Unit when dust exceeds 5 mm thickness
2.5 Staff qualification
Unqualified persons are unable to identify risks and are therefore exposed to greater dangers. The device may only be operated by suitably qualified persons. These operating instructions are aimed at the following target groups:
Operation Manual X-Sential™ 7/34
Page 8
2 | Safety NIR-Online GmbH
Users
Users are persons that meet the following criteria:
They have been instructed in the use of the device.
They are familiar with the contents of these operating instructions and the appli-
cable safety regulations and apply them.
They are able on the basis of their training or professional experience to assess
the risks associated with the use of the device.
Operator
The operator is responsible for the following aspects:
The instrument must be correctly installed, commissioned, operated and serviced.
Only suitably qualified staff may be assigned the task of performing the opera-
tions described in these operating instructions.
The staff must comply with the locally applicable requirements and regulations for
safe and hazard-conscious working practices.
Safety-related incidents that occur while operating the instrument are to be re-
ported to the manufacturer. service.nir-online@buchi.com
NIR-Online service technicians
Service technicians authorized by NIR-Online have attended special training courses and are authorized by NIR-Online GmbH to carry out special servicing and repair measures.
2.6 Personal protective equipment (production)
Follow the rules regarding personal protective equipment that are applicable at the installation site. The operation of the sensor does not require additional protective equipment.
2.7 Personal protective equipment (laboratory)
Depending on the application, hazards due to heat and/or corrosive chemicals may arise.
u Always wear appropriate personal protective equipment such as safety goggles,
protective clothing and gloves.
u Make sure that the personal protective equipment meets the requirements of
the safety data sheets for all chemicals used.
2.8 Modifications
Unauthorized modifications may impair safety and lead to accidents.
u Use only genuine NIR-Online accessories, spare parts and consumables.
u Technical modifications to the instrument or accessories should only be carried
out with the prior written approval of NIR-Online GmbH and only by authorized NIR-Online service technicians.
NIR-Online GmbH accepts no liability whatsoever for damage arising as a result of unauthorized modifications.
8/34 Operation Manual X-Sential™
Page 9
NIR-Online GmbH Product description | 3
1
2
3
3 Product description
3.1 Description of function
The sensor is an optical instrument for nondestructive determination of substances and concentrations in a sample. A sample absorbs and reflects light across the entire wavelength spectrum according to its color and chemical composition. The signal reflected by the sample is recorded and analyzed by a spectrometer.
The sensor uses a lamp to produce near-infrared radiation that interacts with the
molecules of the sample. The interaction between sample and light produces a characteristic spectrum.
The light reflected from the sample is collected through two sets of fiber optics
that direct the light to the NIR and visible spectrophotometers, respectively. The visible spectrophotometer consists of a diffraction grating to spatially disperse the light according to wavelength and a silicon photodiode array containing multiple elements which measure the light intensity for specific wavelength intervals. The NIR spectrophtometer consists of a diffraction grating to spatially disperse the light according to wavelength and an indium-gallium-arsenide photodiode array containing multiple elements.
The produced measurement results are converted to data sequences.
The data sequences are transferred to a computer via an interface.
A computer program compares the data sequence curve with a calibration model
and in that way determines the chemical composition of the sample.
3.2 Configuration
3.2.1 Front view
Fig.2: Front View
1 Measurement window 2 Flange
3 Heat sink
Operation Manual X-Sential™ 9/34
Page 10
3 | Product description NIR-Online GmbH
1
2
3
4
5
6
3.2.2 Rear view
Fig.3: Rear view
1 Journal Button 2 Power and scanning indicator light
3 Type plate (part) 4 Ground connection
5 Power and signal connector 6 Type plate (part)
3.3 Scope of delivery
NOTE
The scope of delivery depends of the configuration of the purchase order.
Accessories are delivered as per the purchase order, order confirmation, and deliv­ery note.
(Equipotential bonding)
10/34 Operation Manual X-Sential™
Page 11
NIR-Online GmbH Product description | 3
5
6
7
8
15
14
1
2
3
4
9
13
12
10
11
3.4 Type plate
The type plate identifies the instrument. See Chapter3.2.2 "Rear view", page10
Fig.4: Type plate
1 Company name and address 2 Product option
3 Revision number 4 Product name
5 Serial number 6 Production date
7 IP class 8 ATEX information
9 Power consumption
(nominal)
10 Power consumption
(maximum)
11 Ambient temperature 12 Current draw
(maximum)
13 Certificates 14 Operating voltage
15 Product type
The following product options are possible:
Letter Option
A NIR
D VIS
E Gold reflector
(X-One)
G Lamp position rev. 1.3.2
N System temperature 0 - 80 °C
O Humidity sensor 0-100% RH
S X-Quvette
(Fiber lens (approx. 0 - 2 cm))
Operation Manual X-Sential™ 11/34
Page 12
3 | Product description NIR-Online GmbH
Letter Option
V External button
X2 X-Beam 002
(approx. 15 cm measurement distance)
3.5 Technical data
3.5.1 Sensor
Specifications X-Sential
Dimensions (W x D x H) 200 x 200 x 100 mm
Weight 5 kg
Max. operating pressure 30 bar at flange
Product temperature
-10 °C to +130 °C
(temperature at flange with water cooling)
Product temperature
-10 °C to + 70 °C
(temperature at flange without water cooling)
Vibrations 0.2 G at 0.1 - 150 Hz
Wavelength spectrum
900 - 1700 nm; 11100 - 5880 cm
NIR range
Wavelength spectrum
350 - 900 nm; 28500 - 11100 cm
Visible range
Detector Diode array
Average measurement
20 spectra/s
time
IP Code IP69 / IPX9K
Type of lamp Tungsten-halogen dual lamp
Lifetime lamp 18000 h (2 x 9000 h)
Minimum clearance on
100 mm
all sides
-1
-1
Connection Voltage 85 to 264 VAC
Frequency 50/60 Hz
Power consumption 30 W
Temperature stabiliza­tion
ASDC (Advanced Spectral Drift Control): active tempera­ture control to ±1°C from set system operating tempera­ture. Deviations will lead to automatic white reference measurement to account for spectral drifts.
12/34 Operation Manual X-Sential™
Page 13
NIR-Online GmbH Product description | 3
3.5.2 Installation box
Specifications Installation box
Dimensions (W x D x H) 300 x 300 x 167mm
Weight (excluding cables)
Weight (inc. cables, 2 x 10 m)
Frequency 50/60 Hz
Power consumption 30 W
Power supply 85 - 264 VAC
3.5.3 Ambient conditions
Max. altitude above sea level
Ambient temperature -10 °C ≤ Tamb ≤ +40 °C
Max. relative air humid­ity
Storage temperature max. 45 °C
3.5.4 Materials
Component Materials of construction
6 kg
7.4kg
2500 m
< 90 % non-condensing
Casing Aluminium (nickel coated), SS 316L 1.4404 flange
Seals NBR (standard sealing material)
FFKM (optional)
Operation Manual X-Sential™ 13/34
Page 14
3 | Product description NIR-Online GmbH
3.5.5 Software
The sensor is controlled via the SX-Suite software package. It consists of the follow­ing components:
Name Description Typical usage User Occurence
SX-Server Instrument
driver / usage of special func­tions
Name Special function Description User Occurence
SX-Server Conveyor belt Optimized for
Mix Control end-
Sample move­ment detection
Read out instru­ment status
Setup of instru­ment hardware
measurement of moving ob­jects on a con­veyor belt
point of mixing processes
Verify sample flow
Operator As required
NIR admin For installation
and mainte­nance
NIR admin As required
NIR admin As required
NIR admin As required
Name Description Typical usage User Occurence
SX-Center User interface
(online/lab mode)
SX-Backup Data backup
scheduler
Recipe/product and calibration mangement
View results (ta­ble, trend, charts, reports)
Reference data management
Automated backup of mea­surement data, results and cali­brations
3.5.6 Computer system requirements
The system requirements for the computer are as follows:
Operating system Windows 10 Pro
Cental processing unit Intel Core i5 generation 6600 or later
Operator Daily workflow
(if not fully au­tomated)
NIR admin During installa-
tion
RAM At least 4 GB
14/34 Operation Manual X-Sential™
Page 15
NIR-Online GmbH Product description | 3
Hard disk space At least 80 GB free disk space
Use a hard disk suitable for continuous operation.
Data backup Network or external
hard disk
Screen resolution At least 1280x1024
LAN At least 1 x 100 Mbit/s LAN
USB 2.0/3.0 At least 1 USB connection per sensor and 1x USB per Data-
PCI/PCIe 1 slot for Profibus card
Software Word and Microsoft Excel 2003 or later
At least 0.5 GB free disk space Additional 20 MB per day and sensor
Lab I/O box
(for Profibus connection)
3.5.7 Installation site (laboratory)
The installation site has a firm, level surface.
The installation site meets the safety requirements. See Chapter2 "Safety",
page6
The installation site has enough space that cables can be routed safely.
The installation site has no obstacles (e.g. water taps, drains, etc.).
The installation site has an own mains outlet socket for the instrument.
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site allows that the power supply can be disconnected at any time
in an emergency.
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter3.5 "Technical data", page12
3.5.8 Installation site (production)
The installation site has a firm, level surface.
The installation site meets the safety requirements. See Chapter2 "Safety",
page6
The installation site has enough space that cables can be routed safely.
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter3.5 "Technical data", page12
The installation site meets the specifications according to the installation point.
See Chapter5.1 "Establishing installation point", page17
The installation site has a own mains outlet socket for the instrument.
The installation site has a sample removal point at a distance of < 1 m.
The installation site allows a direct product measurement.
The installation site has constant product flow.
The layer thickness of the product to be measured is at least 30 mm.
Operation Manual X-Sential™ 15/34
Page 16
4 | Transport and storage NIR-Online GmbH
4 Transport and storage
4.1 Transport
NOTICE
Risk of breakage due to incorrect transportation
u Make sure that all parts of the device are safely packed in such a way as to pre-
vent breakage, ideally in the original box.
u Avoid sharp movements during transit.
u After transportation, check the device for damage. u Damage that has occurred in transit should be reported to the carrier.
u Keep packing for future transportation.
4.2 Storage
u Make sure that the ambient conditions are complied with (see Chapter3.5 "Tech-
nical data", page12).
u Wherever possible, store the device in its original packaging.
u After storage, check the device for damage and replace if necessary.
16/34 Operation Manual X-Sential™
Page 17
NIR-Online GmbH Installation | 5
40
60
R6
0
15
°
15°
140
5 Installation
5.1 Establishing installation point
The fixing points or bolts conform to M6 A2-70/7.3 Nm. Establish the installation point according to the specified data of the flange.
Fig.5: Dimensions of flange
Operation Manual X-Sential™ 17/34
Page 18
5 | Installation NIR-Online GmbH
< 1 m
1
2
3
4
5
5.2 Installation point in piping system (example)
Fig.6: Configuration
1 Flow restrictor 2 Bypass
3 Analyser 4 Sample removal point
5 Flow restrictor
18/34 Operation Manual X-Sential™
Page 19
NIR-Online GmbH Installation | 5
ProductionServer/Control room
Remote computer
Installation box
with mains adapter
Sensor cables
(power supply,
RS 422)
USB converter
1 sensor per measurement
point
Up to 4 sensors per computer
Process control
system
SX-Client
Display:
- Results
- Charts
Chemometric
software
manual calibration
SX-Plus
Remote access via Ethernet
Frame
grabber card
USB
Optional
interfaces:
-Profibus
-4-20 mA/1-9 V
-TCP/IP
-OPC/XML/SQL
-Modbus
Control center
Product flow
Data cable
(RS422)
5.3 Installation (example)
Operation Manual X-Sential™ 19/34
Page 20
5 | Installation NIR-Online GmbH
5.4 Sensor installation
The following installation positions are possible:
Fig.7: Fixing sensor with screws
Tools required:
Torque wrench, size Torx T30
Tightening torque: 8.4 Nm +-1
Precondition: R Installation point has been established. See Chapter5.1 "Establishing installation
point", page17.
R The fixing points or bolts conform to M6 A2-70 15 mm
u Fix the sensor to the installation point using the bolts.
20/34 Operation Manual X-Sential™
Page 21
NIR-Online GmbH Installation | 5
5.5 Connecting the sensor
Tools required:
Torque wrench, size 7 mm AF
Torque wrench, size Torx T20
NOTE
Make sure that the power is not switched on when connecting the sensor.
NOTE
Loss of performance due to use of unsuitable device cables
Max. cable length between installation box and sensor 10 m.
u Put the cable guard over the sensor ca-
ble.
u Connect the sensor cable to the sensor. u Secure the connector.
Tightening torque: 2.5 Nm ± 0.5
u Attach the cable guard to the sensor.
Tightening torque: 2 Nm ± 0.5
u Attach the ground cable to the sensor.
Operation Manual X-Sential™ 21/34
Page 22
5 | Installation NIR-Online GmbH
5.6 Establishing electrical connections
NOTICE
Risk of instrument damage because of not suitable power supply cables.
Not suitable power supply cables can cause bad performance or an instrument dam­age
u Use only BUCHI power supply cables.
Precondition:
R The electrical installation is as specified on the type plate.
R The electrical installation is equipped with a proper grounding system.
R The electrical installation is equipped with suitable fuses and electrical safety fea-
tures.
R The installation site is as specified in the technical date. See Chapter3.5 "Techni-
cal data", page12
u Connect the power supply cable to the connection on the instrument. See Chap-
ter3.2 "Configuration", page9
u Connect the mains plug to an own mains outlet socket.
22/34 Operation Manual X-Sential™
Page 23
NIR-Online GmbH Operation | 6
6 Operation
The instrument is operated via the SX-Suite software on a computer. See "SX-Suite User Manual " and "SX-Plus User Manual ".
6.1 Journal button
Pressing the button generates a journal entry.
6.2 Entering reference data in the journal
To perform a calibration and continuously check the calibration, reference data is re­quired. Continuous checking of the calibration is performed according to the requirements of the production process.
NOTE
The journal entry is identified by date and time.
u Press and hold the journal button for one second.
ð The connected software creates a journal entry.
u Remove the sample at the sample removal point. u Mark sample with date, time and sensor number. u Carry out a laboratory analysis.
u Insert the reference data in the journal for creating the calibration model. See
"SX-Suite User Manual " and "SX-Plus User Manual "
Operation Manual X-Sential™ 23/34
Page 24
7 | Cleaning and servicing NIR-Online GmbH
7 Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this sec­tion.
Any servicing and repair work which involves opening up the casing may only be carried out by NIR-Online service technicians.
u Use only genuine NIR-Online consumables and spare parts in order to ensure cor-
rect operation of the device and preserve the warranty.
7.1 Notes on servicing
NOTICE
Risk of property damage due to failure to remove dust from heat sink
If dust is not cleaned off the heat sink it may cause the sensor to fail.
u Make sure that the layer of dust is no thicker than 5 mm.
7.2 Regular maintenance work
Component Action Interval
Casing u Wipe down the casing with a damp
cloth.
Warning symbols u Check that the warning symbols on the
sensor are legible.
u If they are dirty, clean them.
Optics NOTICE!Have operation carried out by
NIR-Online service technician
u Replace lamps.
Casing NOTICE!Have operation carried out by
NIR-Online service technician
u Check and replace seals
Weekly
Weekly
Annually
Annually
24/34 Operation Manual X-Sential™
Page 25
NIR-Online GmbH Taking out of service and disposal | 8
8 Taking out of service and disposal
8.1 Disposal
The operator is responsible for proper disposal of the instrument.
u When disposing of equipment observe the local regulations and statutory re-
quirements regarding waste disposal.
u When disposing, observe the disposal regulations of the materials used. Materials
used see Chapter3.5 "Technical data", page12
8.2 Returning the instrument
Before returning the instrument, contact the NIR-Online GmbH Service Department. service.nir-online@buchi.comand ask for an RMA number.
Operation Manual X-Sential™ 25/34
Page 26
9 | Appendix NIR-Online GmbH
Declaration of conformity Konformitätserkärung Déclaration de conformité Dichiarazione di conformita Declaración de conformidad
Büchi NIR-Online GmbH Altrottstr. 31 69190 Walldorf Germany
Declares, that the products / Erklärt, dass die Produkte / Déclare par la présente que les produits / Dichiara che i prodotti / Declara que los productos:
Process Analyser X-Sential
comply with the requirements of the European Directives / den Anforderungen der Euroischen Richtlinien / est conforme aux exigences des dir ectives européennes / soddisfa i requisiti delle norme europee / cumple los requerimientos de las Directivas Europeas:
IEC 61326-1:2012
Electrical equipment for measurement, control and laboratory use. EMC requirements: General requirements.
EN 60529 (VDE 470 Teil 1):2014-09 EN 60529:1991 + A1:2000 + A2:2013
Ingress protection testing IP69 / IP69k
RoHS 2 directive 2011/65/EC
Walldorf July 1st, 2021
Michael Markus Michael Eckert Managing Director Product Management
9 Appendix
9.1 Certificates
9.1.1 CE certificate
26/34 Operation Manual X-Sential™
Page 27
NIR-Online GmbH Appendix | 9
9.2 Spare parts and accessories
NOTE
Any modifications of spare parts or assemblies are only allowed with the prior writ­ten permission of NIR-Online GmbH.
9.2.1 Accessories
Order no.
USB-RS422 interface 11060741
Analog interface (DataLabIO) 11060742
Profibus card
PCI, High Profile
Profibus card
PCI, Low Profile
Profibus card
PCI Express, High Profile
Profibus card
PCI Express, Low Profile
9.2.2 Mounting accessories
Mounting accessories are hardware interfaces between the instrument and the process. Depending on the setup, specific mounting accessories might be needed for an implementation into the production facility.
Weld-in Flange
Provides the ability to remove instrument while keeping the process sealed.
Flange with sapphire window and purge port.
Adapter plate, ø140/106 mm, for wall thick-
ness up to 8,5 mm
11060743
11062589
11063000
11063001
Order no. Image
11060754
Material: Stainless steel DIN 1.4404
(SST316L) / DIN 1.4571 (SST316Ti)
Sealing material FFKM White G74S, FDA com-
pliant 15°C (+59°F) to 260°C (+500°F)
Operating pressure -0.5 to 30 bar. Max. pres-
sure 100 bar short term
Purge port M5 (ø4mm tube adapter needed)
to prevent condensation or detect leakage
High grade sapphire crystal optical lens, pol-
ished for reduced adhesion
Dead volume max. 60 mm³
Operation Manual X-Sential™ 27/34
Page 28
9 | Appendix NIR-Online GmbH
Order no. Image
Weld-in Flange Pipe
Flange with sapphire window and purge port for installation in pipes or bended surfaces.
Outer diameter: 140 mm.
Material: Stainless steel DIN 1.4404 (SST316L)
Sealing material: FFKM White G74S
Operating pressure: -0.5 to 30bar. Max. pres-
sure 100 bar short term
Purge port M5 (ø4mm tube adapter needed)
to prevent condensation or detect leakage
The pipe diameter has to be specified upon or-
der
Weld-in Flange Hopper
Flange with sapphire window and purge port for installation in hopper or bended surfaces with different diameters.
Outer diameter: 140 mm
Material: Stainless steel DIN 1.4404 (SST316L)
Sealing material: FFKM White G74S
11068800
11068801
Operating pressure: -0.5 to 30bar. Max. pres-
sure 100 bar short term
Purge port M5 (ø4mm tube adapter needed)
to prevent condensation or detect leakage
The upper and lower hopper diameter has to
be specified upon order
Weld-in Plate
For instruments in direct contact with the product.
Plate with opening, fitting to instrument flange.
Dimensions: 160 x 241 x 3 mm
Material: DIN 1.4301 (SST304)
Thread bolts M6
Bypass Sampler
For free flowing goods (mealy / grainy).
Bypass with feeder and sampling point.
Pneumatic sampler (min. 5 bar / 72.5 psi wa-
ter or oil free compressed air DIN ISO 8573 Class 1)
11060753
11061670
Screw-conveyor (feeding capacity 1.5 t/h)
Motor (380V/50Hz ATEX A22 0.25 KW)
Requires bypass-installation box and a Data-
Lab IO device
28/34 Operation Manual X-Sential™
Page 29
NIR-Online GmbH Appendix | 9
Order no. Image
X-Square
For all free flowing powders and granulates.
The X-Square can be inserted in the product stream or by­pass.
Inspection panel (Plexiglas)
Adapted for Jacob pipes Ø150 mm
Stainless steel DIN 1.4301 electro polished
X-Cell DN50, Standard Flange DN50, PL1, 10 bar
For gas, liquid and paste-like products.
The cell can be inserted in the product stream or bypass.
Material DIN 1.4404 (SST316L)
Sealing material: FFKM White G74S
Operating pressure up to 10 bar (145 psi). TÜV
certificate upon request
Measurement slit 26 mm, configurable be-
tween 1 and 15 mm with additional adapter
DN 50 flange (other sizes upon request)
Clearance volume max. 120 mm³ Cells can be
customized with different diameter and flanges
11061669
11063018
There are various dimensions of the X-Cell available in the pricelist.
X-Cell 4 Edge DN50
For liquid and paste-like products.
The cell can be inserted in the product stream or bypass.
Material: DIN 1.4404 (SST316L)
Sealing material: FFKM White G74S
Operating pressure: max 3bar
Flange: DN50
Path length: 34mm
Typical use: Wine Applications
11068822
Operation Manual X-Sential™ 29/34
Page 30
9 | Appendix NIR-Online GmbH
Order no. Image
VARINLINE Sensor Adapter Flange, Type N, 10bar
For opaque products like powder or granules.
In combination with a path length adapter also for transpar­ent liquid, gel or pasty products.
Material DIN 1.4404 (SST316L)
Sealing material FFKM White G74S (FDA com-
pliant), or custom
Operating pressure up to 10 bar (145 psi). TÜV
certificate upon request
Build for DN50 DIN 32676, process connection
type N
Product temp. -14 °C(+5 °F) to 230 °C (+446 °F)
Path length adapter configurable between 0,5
to 42 mm
Path Length Adapter
To measure transparent liquids with the X-Cell.
The reflector reduces the length of the optical path.
Material DIN 1.4404 (SST316L)
11061674
Gap 1 / 2 / 5 / 10 / 15 mm available
Diffuse or polished surface
Water Cooler Flange
Can be used with all instruments, only in com-
bination with X-Cell or Weld-in Flange
Product temperature above 70 °C to 130 °C. A
flow rate of 5 l water per hour at 20 °C is re­quired
40 °C over temp switch for external alarm pur-
pose, NO (Normally Open) circuit
Water connectors for 8/6 mm hose
9.2.3 Spare parts specifications
Power supply
NOTICE
Risk of property damage due to incorrectly connected mains adapter
An incorrectly connected mains adapter may cause the sensor to fail.
11060752
u Make sure that the current limiter is set to more than 4.5 A. u Make sure that the voltage is 12.7 VAC.
30/34 Operation Manual X-Sential™
Page 31
NIR-Online GmbH Appendix | 9
Specification
Power supply input voltage: 85 - 264 ± 10% VAC
Rated voltage: 12 VDC
Rated current: ≥ 4.5A
Typical peak-to-peak residual ripple: 50 mV
Max. peak-to-peak residual ripple: 200 mV
Device cables
NOTE
Loss of performance due to use of unsuitable device cables
Max. cable length between installation box and sensor 10 m.
Sensor Cable
Pin assignment on device connector viewed from rear of instrument:
Fig.8: Pin assignment
1 PIN 1- blue, ground 2 PIN 2- red, 12.7 VDC
3 PIN 3- green, RxD- 4 PIN 4- yellow, TxD+
5 PIN 5- white, TxD- 6 PIN 6- brown, RxD+
7 PIN 7- not connected 8 PIN 8- not connected
Operation Manual X-Sential™ 31/34
Page 32
9 | Appendix NIR-Online GmbH
1
2 3
4
RS422 Data Cable
Pin assignment on the Moxa viewed from rear of the Moxa:
Fig.9: Pin assignment
1 PIN 1 green, TxD- (A) 2 Pin 2 white (from green), TxD+ (B)
3 Pin 3 orange, RxD+ (B) 4 Pin 4 white (from orange). RxD- (A)
When using the supplied Moxa D-Sub 9-pole connector, swap the cables on pin 1 and 2.
32/34 Operation Manual X-Sential™
Page 33
Page 34
BUCHI Affiliates:
Europe
Quality in your hands
Switzerland/Austria
BÜCHI Labor technik AG
CH – 9230 Flawil T +41 71 394 63 63 F +41 71 394 64 64 buchi@buchi.com www.buchi.com
Italy
BUCHI Italia s.r.l.
IT – 20010 Cornaredo (MI) T +39 02 824 50 11 F +39 02 575 12 855 italia@buchi.com www.buchi.com/it-it
NIR-Online
Germany
NIR-Online GmbH
DE – 69190 Walldor f T +49 6227 73 26 60 F +49 6227 73 26 70 nir-online@buchi.com www.nir-online.de
Benelux
BÜCHI Labortechnik GmbH
Branch Office Benelux NL – 3342 GT Hendrik-Ido-Ambacht T +31 78 684 94 29 F +31 78 684 94 30 benelux@buchi.com www.buchi.com/bx-en
Russia
BUCHI Russia/CIS
RU – 127006 Moscow T +7 495 36 36 495 russia@buchi.com www.buchi.com/ru-ru
France
BUCHI Sarl
FR – 91140 Villebon-sur-Yvette T +33 1 56 70 62 50 F +33 1 46 86 00 31 france@buchi.com www.buchi.com/fr-fr
Spain
BUCHI Ibérica S.L.U.
ES – 08960 Barcelona T +34 936 06 8010 iberica@buchi.com www.buchi.com/es-es
America
Brazil
BUCHI Brasil Ltda.
BR – Valinhos SP 13271-200 T +55 19 3849 1201 F +55 19 3849 2907 brasil@buchi.com www.buchi.com/br-pt
Germany
BÜCHI Labortechnik GmbH
DE – 45127 Essen T +800 414 0 414 0 (Toll Free) T +49 201 747 49 0 F +49 201 747 49 20 deutschland@buchi.com www.buchi.com/de-de
United Kingdom
BUCHI UK Ltd.
GB – Suffolk CB8 7SQ T +44 161 633 1000 F +44 161 633 1007 uk@buchi.com www.buchi.com/gb-en
USA/Canada
BUCHI Corporation
US – New Castle, DE 19720 T +1 877 692 8244 (Toll Free) T +1 302 652 3000 F +1 302 652 8777 us-sales@buchi.com www.buchi.com/us-en
Asia
China
BUCHI China
CN – 200233 Shanghai T +86 21 6280 3366 F +86 21 5230 8821 china@buchi.com www.buchi.com/cn-zh
Korea
BUCHI Korea Inc.
KR – Seoul 153-782 T +82 2 6718 7500 F +82 2 6718 7599 korea@buchi.com www.buchi.com/kr-ko
India
BUCHI India Private Ltd.
IN – Mumbai 400 055 T +91 22 667 75400 F +91 22 667 18986 india@buchi.com www.buchi.com/in-en
Malaysia
BUCHI Malaysia Sdn. Bhd.
MY – 47301 Petaling Jaya, Selangor T +60 3 7832 0310 F +60 3 7832 0309 malaysia@buchi.com www.buchi.com/my-en
BUCHI Support Centers:
South East Asia
BUCHI (Thailand) Ltd.
TH-Bangkok 10600 T +66 2 862 08 51 F +66 2 862 08 54 bacc@buchi.com www.buchi.com/th-th
Middle East
BÜCHI Labor technik AG
UAE – Dubai T +971 4 313 2860 F +971 4 313 2861 middleeast@buchi.com www.buchi.com
Indonesia
PT. BUCHI Indonesia
ID – Tangerang 15332 T +62 21 2222 9478 F +62 21 5568 7906 indonesia@buchi.com www.buchi.com/id-in
Singapore
BUCHI Singapore Pte. Ltd.
SG – Singapore 609919 T +65 6565 1175 F +65 6566 7047 singapore@buchi.com www.buchi.com/sg-en
Latin America
BUCHI Latinoamérica S. de R.L. de C.V.
MX – Mexico City T +52 55 9001 5386 latinoamerica@buchi.com www.buchi.com/es-es
Japan
Nihon BUCHI K.K.
JP – Tokyo 110-0008 T +81 3 3821 4777 F +81 3 3821 4555 nihon@buchi.com www.buchi.com/jp-ja
Thailand
BUCHI (Thailand) Ltd.
TH – Bangkok 10600 T +66 2 862 08 51 F +66 2 862 08 54 thailand@buchi.com www.buchi.com/th-th
We are represented by more than 100 distribution partners worldwide. Find your local representative at: www.buchi.com
Loading...