Patient Safety
Attention: take note of accompanying documents!
▲
!
▲
First read the Instructions for Use.
Use of the equipment presupposes regular
checking by specially-trained staff.
Operation
Make sure that the unit has been positioned
safely and is stable and secure.
Prior to use: always ensure the functioning
of the audible and visual alarms during the
automatic check (see page 6). Also check the
staff call and syringe setting features for
possible damage.
Connection to patient is only permissible
when the device is switched on. Interrupt the
connection to change syringes. There is otherwise a danger of incorrect dosage.
Select syringe/catheter suitable for use with
the connection system and medical application.
Make the connection ensuring that the
infusion line is free of kinks so that a flow is
possible.
Replace disposable articles after 24 hours.
It is presupposed that installation in rooms
used for medical purposes will comply with the
appropriate regulations (e.g. VDE 0100, VDE
0107 or IEC publications). Care must also be
taken to observe regional specifications and
national variations.
Do not use in an area endangered by risk of
explosion.
Compare displayed and input values. Use
only when these match.
Other Components
Work in the area of the pressure shut-down
facility, or variations in pressure (e.g. as caused
by change in level), can affect the accuracy of
the device.
Where several infusion lines are connected,
the possibility of their exerting a mutual
influence on each other cannot be excluded.
Cases of possible incompatibility can be found
in the directions on use of the medicament or
appliances in question.
See also VDE 0753, Part 5, "Application Rules
for Parallel Infusion - Possible Application
Methods", or the BBM application directives
for parallel infusion (38910004).
Only combinations of equipment, accessories,
working parts and disposable parts that have
been shown to be compatible shall be used.
The use of disposable parts that have not
been tested or approved might well exert an
influence on the technical data.
Connected analogue and digital components
must verifiably satisfy the EN specifications
(e.g. EN60950 on data processing devices and
EN60601 on medical electrical devices). Anyone
who additionally connects devices to the signal
input or output part is a system configurator
and is thereby responsible for compliance with
the systems standard EN60601-1-1.
Safety Standards
Perfusor compact satisfies all safety standards
for medical electrical devices in terms of the
IEC 601-1 and IEC 601-2 (-24) publications.
Note: IEC 601-1 corresponds to the European
Standard EN60601.
2
Perfusor compact
Contents
Perfusor compact / Overview Page 4
Operation Page 6
Technical Data Page 10
Start-up and Trumpet Curves Page 12
Warranty Page 13
Ordering Page 14
The Perfusor compact is a transportable
infusion syringe pump in accordance with
EN60601-2-24 (draft), Points 2.2.18 and 2.2.23
that is suitable for dispensing liquids in
nutritional and infusion therapy. The medical
specialist must decide on suitability for
application on the basis of the warranted
properties and the technical data.
For further details please refer to these
Instructions for Use.
3
Overview
Handle
Always use the handle when carrying
Display
Shows all important features at a glance: rate,
type of syringe, mains or rechargeable battery,
carriage control
"Alarms" on page 9).
and alarm symbol (see
Operation
For special functions first press the
Syringe holder
Locks in the syringe. To remove, pull and
swing to the left.
The drive unit can be moved by hand after
the locking lever has been opened.
Short instructions for use and Syringe Table
See the underside of the device.
F button.
Battery
Press here to change the battery. Interrupt the
connection to the patient during changing of
the battery. Switch off the device and pull the
cover downwards. Always renew all batteries,
taking care to observe waste disposal regulations.
4
Multi-Function Connector (MFC)
Connection for staff call; ambulance cars (12V)
and interface.
Mains Connection
Connection for the mains supply. In the event of
power failure, automatic switch-over to battery.
Input Correction
Interrupt alarm signal for 2 min.
On/Off. Press for 2 sec to switch off.
Start/Stop Infusion
Decimal Point
Inspection on Delivery
Despite careful packaging the risk of transport
damage cannot be entirely excluded. Upon
delivery please check that nothing is missing.
Do not make use of a damaged device! Contact
the service department.
For special functions (green) first press
Transport/Carriage
A maximum of three devices can be connected
together. Special care is required here if a
patient is already connected. Avoid external
mechanical influence!
Locking Devices Together
Place one device on top of the other. Push down
the connecting part until clicks into place. Lock
by turning the vane until it is vertical.
To disconnect, turn until horizontal and then
neu
push upwards.
Stand Clamp/Attachment on Stand
Attach the Perfusor compact from above,
clicking it into place. To release, press the black
button. For safety purposes attach each device
separately to the stand.
Packaging: Reusable, therefore environmentally
acceptable (returns will be accepted).
Extent of Delivery: Perfusor compact, power
lead, pole clamp, instructions for use,
4 batteries.
F.
5