B Braun Outlook 200 User manual

OUTLOOK”
200
SAFETY
INFUSION
SYSTEM
SERVICE
MANUAL
Rx
BIBRAUN
SHARING
950953
EXPERTISE
Rev B (9/04)
950953
Rev B (9/04)
OUTLOOK™
200
‘THIS
MANUAL
OR
MACHINE-READABLE
©
2004
B.
BRAUN
WORKING
AND
TOGETHER
ITS
AFFILIATES.
MAY
NOT,
IN
FORM
MEDICAL
FOR
EXCELLENCE
WHOLE
INC.
OR
WITHOUT
ALL
SERVICE
IN
PART,
BE
COPIED,
PRIOR
WRITTEN CONSENT
RIGHTS
RESERVED.
IN
SAFETY,
PHOTOCOPIED,
DIOSEGUARD,
MANUAL
REPRODUCED,
OF
B.
BRAUN
DOSESCAN
AND
MEDICAL
OUTLOOK
TRANSLATED,
INC.
ARE
TRADEMARKS
OR
CONVERTED
OF
B.
TO
ANY
BRAUN
ELECTRONIC
MEDICAL
INC.
ase
ンク
B
BRAUN
SHARING
EXPERTISE
950953
Rev B (9/04)
©
OUTLOOKTM
200
SERVICE
MANUAL
5.0
CLEANING
5.1
5.2
6.0
PERFORMANCE
6.1
6.2
6.3
6.4
6.5
6.6
6.7
6.8
6.9
6.10
6.11
6.12
4.2.6
4.2.7
4.2.8
6.2.1
6.2.2
6.2.3
6.2.4
6.3.1
6.3.2
6.4.1
6.4.2
6.5.1
6.5.2
6.6.1
6.6.2
6.7.1
6.7.2
6.8.1
6.8.2
6.8.3
6.8.4
6.9.1
6.9.2
6.9.3
6.10.1
6.10.2
6.11.1
6.11.2
6.12.1
Review Setup Hand-Held
&
PREVENTATIVE
Cleaning
Preventative
Test
Equipment,
Mechanical
Equipment
Exterior....................
Bezel………
Mechanical
Simulated
Equipment
Test
Volume
Equipment
Test
Air-Inline
Equipment
Test
Occlusion
Equipment
Test
Static
Electrical
Power
Free
Eguipment
Test
Equipment Test Test Test
Switch
Equipment
TestProcedure.......................................
Test
Main
Battery
Equipment
TestProcedure
Pressure
Equipment
Test
Restriction/Outlet
EguipmentReguvired.......................................
Time & Date
Drugs
Maintenance
CHECK
Inspection
ee
Use
Test
Procedure
Delivery
Procedure
Test
Safety
.................
Procedure Alarm
Procedure
Flow
Procedure
Procedure: Procedure: Procedure:
Over
Procedure:
Capacity
Transducer
Procedure
serrer
................
Scanner
Materials,
Required
User
Required
Accuracy
Required
Required
Test
Required
Test Reguired
Test Reguifed
Test
Required
Required
Required
Valve
Setup
MAINTENANCE
Schedule
and
Consumables
in
Interface
...................
..................
Test
..........
........
...
........
.......
........
.......
........
....
.ee Electrical Power Hi-Pot
Daisy
i...
A/D
iii
Cord
Testing
...................
...
Chain
Test
(optional)
..................
Offset
Yoke
Adjustment
22
for
Standard
SCHEDULE
......................................
.....
.....
Chassis / Case
Resistance
...
Power
..........................-
iii
Test
(Occlusion
Bar
Coding
...
ss
..
Leakage
..................
Verification
iene
Verification)
Test
..
..
..
0.1000
ee
.
nenene
29
30
33 ...34 ...35
35 ...35
...36
...36
...38
...38 ...38 42
42
42
„44
44 ...44 45 ...45
45
„47 AT AT 48
48
…48 49
...49 ...50
50
.…
50 ...50 …51 ...51
nos
51
......53
53
...53 ...55
55
ü
950953
Rev B (9/04)
TABLE
OF
CONTENTS
7.0
6.13
6.14
THEORY
7.1
7.2
7.3
7.4
7.5
7.6
7.7
6.12.2
6.13.1
6.13.2
6.14.1
6.14.2
General
7.1.1
7.1.2
7.13 Functional
7.2.1
7.2.2
7.2.3
7.2.4
7.2.5 Power
7.3.1 Display
7.4.1
7.4.2
7.4.3
7.4.4
7.4.5
Pressure
7.5.1
7.5.2 Mechanism
MainBoardAssy.........................
7.71
7.7.2
7.7.3
7.7.4
7.7.5
7.7.6
TIT
7.7.8
7.7.9
7.7.10
7.7.11
7.7.12
7./.13
7.7.14
Test
Inlet
Procedure
Solenoid
Adjustment
siens
TeSt
.......................................
EguipmentReguired.......................
Test
Procedure
Annual
OF
Preventive
Equipment
Procedure
OPERATION
Design
Patient Ease
Upgradabiliy.........................
Case
Power
Main Half
Мат
Supply
16
Door Key
LCD
LCD LCD
Pressure
Offset,
Management
Door
Display
Communication
Pump
Interprocessor ASIC Transfer
Temperature
External
Power............................
Power Voltage Pressure
Safety
of
Operation
Description
Bottom
Supply
Battery
Door/Bezel
Воага
Volt
Supply
Board
Processor
Panel
Contrast
Backlight Module
Transducer
Gain,
Interface
Interface
Board
Processor
e
Control
Maintenance
Required
ie
.....
Features
.......................
and
..
........
....
.......
Assy.
................
Assy.
Assy.
Transducer
..
......
.....
.................
Interface
Voltage
Supply
Assy.
........................
and
Board
Processor
Bus
Interface
Data
Communications
Valve
Solenoid
Controller
Sensor
Circuits
References
Signal
.
Assy.
......
Circuits
Output
............
Port
...............................
Interface
ire
Circuits
(Every
Repair
.
.
.......
....
........
Circuits
Assy.
....
....
......
....
.
Bus
Driver
.....
...
cemre
nice
12
Months)
.
.
ss
59
59
59
60
60
...
61
.…
61
„61
„61 ...
61 ...65
66 ...
66
66
67
67
68
68
68
69
69
69
70
70
70
71
...71
11
...
72
„72
13
73
73
75
suisse
ee
71
76 79 79
950953
Rev B (9/04)
iii
OUTLOOK™
200
SERVICE
MANUAL
7./.15
7.7.16
7.7.17
7.7.18
7.7.19
7.7.20
7.8
Air-In-Line
7.9
System
7.9.1
7.9.2
7.9.3
7.9.4
7.9.5
8.0
Pump
8.1
HISTORY,
Alarm
8.1.1
8.2
Check
8.3
System
9.0
REMOVAL & INSTALLATION
9.1
Repair
9.1.1
9.1.2
9.2
Batteries..................................
9.2.1
VariablelevelShifter Variable
Time Audible
Overvoltage
LeadAcidBatteyChargerSystem......................................
Operation
Power-Up
Hold Run Keep
Alarm
ALARMS, & TROUBLESHOOTING
and
Accessing
Configuration
AlarmS
Tools & Parts
Hand
Spare
12
Volt
9.2.2 3 Volt
9.2.3
93
9.3.1
9.3.2
9.3.3
9.4
9.4.1
9.4.2
9.4.3
9.5
9.5.1
9.5.2
9.5.3
9.6
9.6.1
9.6.2
4.8
Volt
Caselop.............................
Removal-Case
Installation-Case 1 00
Testing
Internal
Scanner
Removal-Internal
Installation-Internal Testing
Main
Circuit
Removal-Main
Installation-MainCircultBoard
Testind
PoleClampAssembiy..........................
Removal-Pole
Installation-PoleClampAssembly..........................................
AR
9.7
PowerSupply..........................
9.7.1
9.7.2
Removal-Power Installation-Power
Gain
Of
Day
AlarmS
Board
ASSY.
..................
State
State
Vein
State
Operation
Operation
Open
Operation
Operation
Operation
ToolS
Parts
.ee
........
Lead-Acid
Lithium
NICAD
.ee
.........................
.ee
Board
.................
Amplifier
inner
ee
Detection
Circuit
........
ee
usines
.pe
State
Operation
.
ss
Histories
of
Software
PROCEDURES
....................:
and
Alarm
Histories
........................
...........
ee
eine
Battery
Battery
.pe
Battery
Top
Scanner
Scanner
.........................
Circuit
Board
....
.........................................
ee
Clamp
Assembly
Supply
Supply
......
.....
..
eeeeonnnne
cc
.....
con
rn
rnrinnern
nens
e
eee
нение
иененнииинеинитиннин
79 80
80 ...80 ...81
81
res
82
.
...
...
89
...89 90
...91
91
119
...
.
nonné
124 125
..126
126
126
126 „127 127
127
A
127
..128
...128
130
iv
950953
Rev B (9/04)
TABLE
OF
CONTENTS
9.8
9.9
9.10
9.11
9.12
]
9.13
9.14
9.7.3
9.8.1
9.82
9.8.3
9.9.1
9.9.2
9.9.3
9.10.1
9.10.2
9.10.3
9.11.1
9.11.2
9.11.3
9.12.1
9.122
A
9.13.1
9.13.2
9.13.3
9.14.1
9.14.2
9.143
Testing
Doo/BezelAssembly................................................eeeeeee
Removal-Door/Cosmetic
installation-Door/Bezelâssemby.....................................
Testing
Door
Assembly Removai-Door installation-DoorAssembl
Testing
Cosmetic
Removal-Cosmetic Installation-Cosmetic
Testing
DisplayBoardAssembly
Removal-Display Installation-Display
Testing
Main
Bezel
Removai-Main
installation-MainBezel................................................
Air-In-Line
Removai- Installation-Air-In-Line
Testing
PetalCoverAssembiy
Removal-Pefal
Installation-Petal
Testing.................................
RR
Assembly
eee
Bezel & Keypanel
Bezel & Keypanel
Bezel & Keypanel
,es
iii
...........
Assembly
Air-In-Line
renser,
..................................... 2 134
Board
Board
iii
Bezel
Cover
.......................rrrrrcerrrieii
(sensor
Assembly
...........................................
CoverAssemby
130
131
Bezel
Assembly
ASSEmMDly
Assembly
Assembly
Assembly
"ee
.................................................
........................................
.........
Assembly
and
circuit
board)
iene
irene
...................................
eee
.ps
ea
nana n nan
eee
ee
eee n anna a nenene
Assembly
Assembly
....................................
.................
..............
iii
iii
iene
een
eee
eee
...........................
ire
131
131 131
132
132 132 132 133 133
133
133
134 134 134
135
135
135
135
136
136
137
138
139
139
139
139
APPENDİX
950953
Rev B (9/04)
A...
eee
eee
erene
ema
eene
sper
annene
V
sesasesenarsenensenamanaan
rocas
do
141
This
page
intentionally
left
blank.
INTRODUCTION
CHAPTER
1.0
INTRODUCTION
IN
THIS
CHAPTER
*
The
intended
*
1.1
INTENDED
This
manual
Medical
Outlook service
Braun
available Customer fees. service
Inc. Infusion
professional.
Medical
to
Service
B.
Braun
professional
*
*
*
USE
is
intended
trained
Pumps.
Inc.
prior
your
institution
for
Medical
About What
Ofwarnings,
How
service
Although
additional
has
the
is
covered
to
contact
to
be
used
personnel
Repair
to
servicing
upon
Inc.
maintains
been
certified
it
YOU
use
of
this
important
cautions,
by
of
the
is
not
the
information
features
under
the
and
B.
Braun
qualified
for
the
pumps
mandatory
the
Outlook
expiration
on
the
right
by
B.
Braun
WILL
service
Customer
biomedical
troubleshooting,
should
to
LEARN:
manual.
of
the
B.
Braun
disclaimers
be
performed
each
service
pumps,
of
the
offered
it
pump
training
withhold
Medical
Outlook
warranty.
related
Service
service
repair,
professional
is
recommended
warranty.
classes
the
purchase
Inc.
pumps.
to
the
use
of
this
and
Technical
professionals
and
maintenance
Support.
and
by a qualified
be
certified
and
will
Please
and
of
contact
their
associated
parts
until
manual.
B.
Braun
of
the
biomedical
by
B.
be
made
a
K
1.1.1 Certain
text
1.1.1.1
950953
NOTE:
Braun
This
Medical personnel the
information manner or
their
component
contained
infringement
Keys,
Text
conventions
display,
indicators,
Keys
Keys
found
Rev B (9/04)
manual
Inc.
in
maintaining
other
than
herein
of
Display,
are
on
contains
This
information
and
contained
for
the
parts.
may
subject
trade
secrets,
Indicators,
used
within
and
softkeys.
the
pump’s
information
is
servicing
herein
may
maintenance
Any
unauthorized
the
user
copyright,
and
this
manual
key
panel
will
and
data
provided
the
Outlook
be
duplicated
and
servicing use
to
substantial
and
patents.
Softkeys
when
referring
be
graphically
proprietary
solely
for
technical
pumps.
or
used
of
the
of
the
information
liability
to
the
pump’s
depicted
to
B.
None
in
any
pumps
for
within
Page 1 of
of
keys,
the
LCD
144
OUTLOOK™
200
SERVICE
MANUAL
instruction
1.1.1.2 Text When
Information
10.7
mf”
When “.
highlight
1.1.1.3
1.1.1.4
Indicators
All
Indicators
indicate: and
Alarm
Softkeys
In
some
functions appear
text.
Display
referring
Screen,
“DOSE
referring
the
located
AC
Power
Indicator.
screens
and/or
as
follows:
For
example,
to
informational
the
text
will
Mode”
to a specific
CONC
on
Indicator,
there
are
field,
field.”
the top
softkeys
information.
“Clear
Drug”
the
pump’s
text
or
appear
the text
left
of
the
Battery
Softkeys
Indicator,
which
available
power
menu
in
the
will
door
allow
key
will
items
displayed
following
appear
the
in
are
referred
Hold
.
user
on
the
Indicator,
appear
format:
the
to
Information
as
GOWER).
on
the
“Tot
following
to
by
what
Run
access
additional
Screen
pump’s
Infused=
format:
they
Indicator,
will
A
Zin,
se
1.1.1.5
Additionally,
information.
ATTENTION: topic
in
this
NOTE:
this
This
manual.
CAUTIONS
throughout
This
manual.
icon
will
or
WARNINGS:
recommendations the
patient.
Companion
The
following
will
be
Documents
documents
required
icon
will
provide
This
in
order
for a full
this
manual
provide
additional
icon
to
prevent
are
companion
are
additional
information
will
provide
accident
understanding
icons
which
information
safety
or
documents
of
the
pump’s
identify
supplemental
for a given
for a given
injury
to
to
this
Service
the
functionality.
topic
user
or
Manual
in
and
Page 2 of
144
*
Outlook
*
Outlook
200
Operator’s
Windock
Manual - Part
Operation
Manual - Part
No.
950980
No.
951031
950953
Rev B (9/04)
1.2
IMPORTANT
The
Outlook intended anesthetic.
for
continuous
Additional
Flash
200
is a
EEPROMs
PUMP
Class
FEATURES
1-Internally
operation.
features
include:
powered,
Operation
is
Type
not
suitable
CF,
IPX1
drip-proof
in
the
INTRODUCTION
infusion
presence
of a flammable
pump
1.3
1.3.1
Integrated
bar
Hand-held Daisy Easily Time
Subassembly
PUMP
Chaining
accessed
and
date
SAFETY
Warnings
WARNING: component undergo
Performance Check
care.
All pump with
bodily
must
an
harm
code
Scanner
bar
code
Scanner
Power
Main
stamped
Repair
If
the
product
or
assembly
and
pass
final
performance
be
performed
exception
and
will
(Outlook
(Outlook
Cords
Battery
Operation
is
has
all
performance
chapter
with a set that
for
set
sterility
void
ail
200)
200)
and
Alarm
partially
been
or
replaced
checks
of
this
manual
testing
and
only.
warranties
Logs
completely
or
repaired,
outlined
prior
to
calibration
is
acceptable
Failure
and
to
do
liabilities.
disassembled,
the
unit
must
in
the
use
for
patient
of
the
Outlook
for
patient
so
may
cause
or
a
use
950953
WARNING: unless
personnel
soldering
consent Center
Representative.
WARNING:
may
occur
labeling.
Rev B (9/04)
Rework
of
have
techniques
has
been obtained
B.
Braun
if
the
product
multi-layer
attended
for
this
type
from
Failure
will
assume
is
not
boards
and
of
an
to
no
used
should
passed
board
authorized
do
so
will
responsibility
in
accordance
not
an
accredited
and
express
B.
void
be
attempted
written
Braun
all
for
with
Service
warranties.
incidents
product
course
which
Page 3 of
in
144
OUTLOOK™
A7
Zik,
X:
jik,
#/
200
SERVICE
WARNING: the
source
Carefully
the
set,
WARNING:
autoclave
Always
Outlook
WARNING:
Always
container
follow
as
well
To
avoid
or
immerse
the
as
disconnect
pumps
before
Trained
representatives
a
clinical
setting
device.
MANUAL
read
and
and
IV
instructions
the
recommended
mechanical
the
pump
the
power
cleaning
Biomedical
must
perform a full
to
ensure
proper
follow
the
instructions
administration
for
loading,
set
change
or
electronic
in
any
fluids
cord
from
the
to
prevent
professionals
set-up
programming
which
sets
you
removing
intervals.
damage,
or
cleaning
outlet
electrical
of
or
or
B.-Braun
the
pump
and
accompany
are
using.
and
reloading
do
not
solutions.
from
other
shock.
before
function
steam
use
of
in
©
1.3.2
/
Ry
N27
WARNING:
immediately professional
WARNING: the
drip
are
falling
loaded
Cautions
CAUTION:
\EC
60825-1
If
the
remove should
Prior
chamber
and
and
further
The
Beam
CAUTION:
handling. initial
The
Do
self test
pump
has
the
inspect
to
starting
and
the
the
infusion
use
may
Outlook
200
standard.
pump
not
as
may
use
the
patient
been
dropped
pump
from
the
device.
an
infusion,
service. A qualified
programmed
has
not
begun,
result
in
inaccurate
is a class 2 laser
LASER
have
pump
injury
RADIATION - Do
been
damaged
if
it
appears
or
device
or
appears
verify
information
no
the
drops
set
is
may
infusion.
product
Not
during
damaged
damage
may
to
be
damaged,
Biomedical
are
falling
correct.
be
If
improperly
according
Stare
Into
shipping
or
fails
the
occur.
in
drops
to
and
Page 4 of
144
950953
Rev B (9/04)
INTRODUCTION
in
/
A
CAUTION: 60
Hz
CAUTION:
sembling
tools
CAUTION: Outlook
CAUTION:
possible
The
power
grounded
3-prong
Wear a grounding
the
pump.
which
could
It
is
pumps
Do
not
scratch
not
recommended
together.
use
explosion
cord
receptacle
Do
not
in
the
hazard
must
be
wrist
lay
pump
or
damage
presence
exists.
connected
designated
strap
when
face
the
to
Daisy
of
flammable
to a 110-120
as
assembling
down
LCD
Chain
on
or
more
“Hospital
or
components
key
panel.
than
anesthetics,
VAC,
50/
Grade.”
disas-
or
five
as
a
M
7
is, 2
dip,
/
CauTIon: autoclaving device be
disinfected.
CAUTION: clamped
CAUTION:
taken
interference Resonance
CAUTION: inthis unit
Do
and
When
securely
As
to
avoid
Perform
Service
cannot
not
or
void
with
and strong
Imaging
be
clean,
with
ethylene
the
warranty.
attaching
to
avoid
all
medical
exposure
(MRI).
testing
Manual
or
assured.
disinfect,
or
oxide
Only
this
pump
damage
electronic
to
powerful
magnetic
fields
Permanent
and/or
the
preventive
reliability
sterilize
gas.
Doing
external
to a pole,
to
the
device.
equipment,
sources
such
damage
and
proper
any
part
of
the
so
may
damage
parts
of
the
pump
make
of
as
sure
care
electromagnetic
Magnetic
may
result.
maintenance
functionality
device
it
is
must
as
defined
by
the
should
be
of
the
950953
Rev B (9/04)
©
Page 5 of
144
OUTLOOKTM
fii, 82
CAUTION: and has Representative. not
specifications.
200
SERVICE
will
void
been
be
possible
MANUAL
Individual
component
any and
obtained
B.
to
from Braun
test
all
warranties
an
authorized
recommends
complete
replacement
unless
B.
Braun
express
board
circuit
function
is
not
recommended
written
Service
replacement
to
factory
consent
Center
as
it
may
1.3.3
A
ZİN
A
/!\
A
Attentions
ATTENTION:
and
Mechanical
CSA
C22.2
ATTENTION: for
use
by
ATTENTION:
/ N battery!
1.3.4
The
Laser
following
Caution
laser
device.
AVOID
Medical
No.
The
authorized
Only
caution
Equipment,
Hazards
601.1.
DB9
connecior
personnel
use a B.
Labels
label
only
Braun
(enlarged
EXPOSURE - LASER
with
respect
in
accordance
on
the
back
only.
specified
to
show
RADIATION
EMITTED
to
Electrical
with
of
the
lithium
detail)
is
FROM
Shock,
UL2601-1,
Outlook
replacement
affixed
THIS
pump
to
the
DEVICE
Fire
CAN/
is
handle
of
the
Page 6 of
144
P/N
950851
Rev
COMPLIES
STANDARDS EXCEPT TO
JULY
FOR
LASER
26,
WITH
FDA
FOR
LASER
DEVIATIONS
NOTICE
#50,
2001
PERFORMANCE
PRODUCTS
PURSUANT
DATED
950953
Rev B (9/04)
INTRODUCTION
The
following
laser
caution
label
is
affixed
to
the
bottom
of
the
device.
1.3.5
Due vice a
course. not microprocessor
devices,
which
By acknowledge
harmless including
Disclaimers
to
the
criticality
on
the
Outlook
qualified
biomedical
Repairs
be
attempted
testing,
have
not
performing
from
attorney’s
should
controlled
static
specifically
any
and
and
repairs.
CAUTION - CLASS
WHEN
EYE
of
pump
repair
not
by
other
control
repairs
agree
against
fees,
product
that
performance,
be
performed professional
be
attempted
than
qualified
electronics,
and
been
identified
with
or
B.
Braun
any
and
directly
3R
LASER
OPEN,
EXPOSURE.
B.
only
by
who
without a current
personnel
digital
soldering
without
and
all
damages,
or
indirectly
and
techniques.
in
this
use
its
distributors
RADIATION
AVOID
Braun
an
has
of
this
liabilities,
arising
DIRECT
950966A
recommends
authorized
attended
that
analog
Service
the
Service
are
knowledgeable
circuitry,
Repairs
Manual.
manual
shall
actions
out
of
that
all
B.
Braun
B.
Service
Braun
service
Manual.
Repairs
in
electromechanical
should
as
reference,
be
or
resulting
not
be
you
indemnified
or
causes
from
required
ser-
Center
training
should
the
areas
attempted
hereby
and
held
of
action,
such
or
of
You
also
understand
dental
rized and
beyond
B.
additional
accepts during
date
1.4
1.4.1
If
manual,
950953
or
consequential
or
unauthorized
experience
B.
Braun
of
personnel
Braun’s
neither
assumes,
liability
no
financial
the
warranty
of
issue.
SERVICE & TECHNICAL
Contact
the
pump
and
Rev B (9/04)
Information
fails
to
the
cause
that
neither
loss,
repair
control,
nor
or
responsibility
obligations
period.
B.
damage,
of
this
performing
directly
authorizes
for
The information
SUPPORT
respond
to
cannot
the
be
determined,
Braun
device
affect
nor
or
expense
since
repairs,
the
any
in
connection
any
institution’s
operating
its
distributors
directly
the
as
well
device
other
with
contained
or
troubleshooting
discontinue
or
quality
as
other
and
the
person
repair
authorized
in
this
using
shall
be
indirectly
of
repair,
results
to
assume
of
manual
and
factors
obtained
for
this
device.
or
unauthorized
procedures
the
pump.
liable
for
arising
the
knowledge
relating
from
it,
any
B.
is
current
listed
Page 7 of
any
from
to
matters
other
Braun
repair
as
of
in
inci-
autho-
its
use.
or
the
this
144
OUTLOOK™
200
SERVICE
MANUAL
Service
contacting:
Product complaints
and
product
performance
may
address.
1.4.2 Should
preferably
turer
they
Authorization
contact
tion
Pump
it
be
in
cannot
are
in
transit.
Customer
Number.
Returns
necessary
the
original
assume
to
return
Service
B.
Braun
1601
Wallace
Carrollton,
Attn.:
(800)
be
sent
to
return packing,
any
responsibility
products
at
the
information
Medical
Drive,
Texas
Manager
of
627-7867
to
the
Customer
the
pump
and
must
above
to
ship
for
loss
be
received
phone
may
be
obtained
Inc.
Suite
150
75006
Service
Complaints
the
manufacturer,
it
prepaid
or
number
to
damage
from
for a Returned
Supervisor
the
above
to
returned
B.
Braun
from
the
manufacturer
at
carefully
pack
address.
instruments
prior
to
shipment.
Materials
the
above
the
pump,
The
manufac-
Authoriza-
by
while
Please
With
each
question
*
the
pump's
a description *
the
pressure
+
the
rate
*
the
initial
*
the
type
*
the
amount
noticed,
+ © the +
* + © any
message
the
diagnostic
the
catalog
other
serial
of
limit
setting,
volume(s)
of
fluid(s),
of
displayed
and
information
or
complaint,
number
the
difficulty
setting,
to
time
between
code
number
lot
number
(located
be
infused,
at
the
(if
of
which
please
experienced,
the
time
include:
on
the
start
of
the
difficulty
the
applicable),
the
set(s)
in
might
aid
in
rear
of the
pump),
infusion
and
occurred,
use,
and
the
investigation
the
of
time
the
the
complaint.
difficulty
was
Page 8 of
144
950953
Rev B (9/04)
PUMP
DESCRIPTION
CHAPTER
PUMP
2.1
VISUAL
2.1.1
1.
AC
3.
Hold
5.
Alarm
Front
power
Indicator
Indicator
DESCRIPTION
INDICATORS
View
indicator
IN
*
*
»
THIS
The
Of
the
The
2.0
CHAPTER
visual
pump’s
indicators
location
technical
&
YOU
and
and
function
specifications.
CONTROLS
2.
4.
6.
WILL
pump
Battery Run Door
LEARN:
controls.
of
the
Indicator
Indicators
Lever
key panel
keys.
7.
Pole
Clamp
9.
Primary
11.Piggyback
Rate
Rate
13.DoseScan™
Display
Display
Label
Scanner
]
Aperture
8.
Information
10.Primary
'2-Figgyback
Volume
Volume
play
14.
Receptacle
Display
To
for
optional
Be
Delivered
To
Be
Delivered
hand-held
Display
Dis-
Scanner
950953
Rev B (9/04)
Page 9 of
144
OUTLOOK™
200
SERVICE
MANUAL
2.1.2
1.
Daisy
3.
Panel
fae)
\&
Back
Chain
Lock
CAUTION:
a
device
View
Outlet
Out
PC
communications
Access
Door
Switch
Although a DB9
other
than
directed
port,
port
2.
Communication
|
1
the
Communication
and
should
in
applicable
not
be
B.Braun
Data
Port
“23
Data
connected
documentation.
Port
to
any
is
NOT
other
Page
10
of
144
This
space
intentionally
left
blank.
950953
Rev B (9/04)
Pump
DESCRIPTION
2.1.3
1.
Cassette
3.
Refill
5.
Petal
7.
Air-In-Line
9.
Pressure
11.Tube
Inside
Piston
Module
Pusher
Door
Alignment
Pins
Cover
Detector
Transducer
View
2.
Inlet
Valve
4.
Transfer
6.
Restriction / Outlet
8.
Free
10.Free
Valve
Flow
Protection
Flow
Prevention
Valve
Clip
Spring
Receptacle
950953
Rev B (9/04)
Page
11
of
144
OUTLOOK™
2.2
KEY
©
200
PANEL
1.
POWER
on
and
3.
Channel
select or
Piggyback
for
data
5.
HOLD stops
quick
menu, Stop minutes
pushed.
7.
CURSOR
cursor tion
SERVICE
turns
off.
Indicator
either
entry.
silences
an
infusion,
escape
and
extends
state
an
every
keys
on
the
Screen.
MANUAL
the
pump
keys
the
Primary
channels
an
alarm,
is
a
from
the
the
additional
time
it
is
move
the
Informa-
3
GED
.
RUN
sion
mary channels.
.
DATA
of
numbers.
.
MENU
other
ENTER
8.
the
selection
keys
begin
of
either
or
Piggyback
keys
allow
lets
the
modes
accepts
the
the
Pri-
input
user
and
options.
data
made.
infu-
see
or
Page
12
of
144
950953
Rev B (9/04)
2.3
TECHNICAL
Catalog
Type
Classification
Dimensions
Weight
Power Electrical
EMVEMC
Ground
Grounding
Leakage
Plug Pole
Batteries
Number
Outlook
200
of
Unit
Requirements
Shock
Impedance
Current
Size
Range
Compliance
Resistance
SPECIFICATIONS
620-200
Volumetric
placement
Class suitable
intended
Width
approx.
120
Protection
Meets
emissions
Less
Meets
Meets
Hospital
0.85"-1.3"
(Non
Reservoir
1-Internally
in
the
presence
for
continuous
=12",
Height=4.5",
11
Ibs
VAC,
2A,
60
Type CF
1.
or
than
UL
or
12
time rate
shock
exceeds
and
immunities
0.2
ohms
Standard
exceeds
Grade
Volt
sealed
is
dependent
of
fluid
(3
PUMP
Drop-Counting),
Powered,
of
flammable
use
in
Depth=8.75”
Hz excluding
protection
.
EN60601-1-2:2001
standard.
(tested
2601
UL
pin)
lead-acid
delivery
per
Standard
on
the
Positive
CF
type,
accordance
accessories
in
accordance
UL2601-1)
2601
battery
state
of
DESCRIPTION
Pressure
IPX1
drip-proof,
anesthetic
for
Class
for
portable
battery
and
with
UL2601-1.
(domestic)
with
.
B,
|
use.
charge
Dis-
not
UL2601-
Run
and
the
Delivery
Volume
Delivered
Occlusion
Keep Volume
950953
Rate
Range
to
be
Range
Limit
Vein
Open
Delivered
Rev B (9/04)
Settings
Rate
Accuracy
3
Volt
lithium
*
4.8
Volt
0.1
to
999.9
Micro
0.1
75, default
3.0 +5%
Mode:
to
9999.9
100,
200,
is
mi/hr
300
or
battery
NiCAD
ml/hr,
selected
in
0.1
to
ml,
in
300,
400,
mmHg.
0.1
for
battery
0.1
99.9
rate,
for
ml/hr
ml/hr
ml
increments
500
and
whichever
RAM
data
backup
increments
750
alarm
mmHg.
is
lower
retention
Standard
Page
13
of
144
OUTLOOK™
Fluid
Types
Alarms
200
SERVICE
MANUAL
All
standard
Air-In-Line,
tainer
Time Exceeded,
-
TV
Battery
Empty,
fluids
Very
Downstream
KVO,
Low
Low,
Close
Occlusion,
Battery
Roller
Warning,
Door
Clamp,
Open,
Con-
Hold
System
Error.
This
space
intentionally
left
blank.
Page
14
of
144
.
950953
Rev B (9/04)
RECEIVING
PROCEDURE
CHAPTER
RECEIVING
3.1
INTRODUCTION
3.1.1
Any be
control
not
Notices
new
device
replaced
only
numbers,
be
considered
3.0
PROCEDURE
IN
THIS
CHAPTER
*
How
to
properly
*
How
to
perform a receiving
*
What
procedures
that
does
not
if
the
device
or
preventative
an
out-of-box
YOU
receive
to
follow
meet
factory
is
not
altered,
maintenance
failure.
WILL
in
specifications
LEARN:
an
Outlook
inspection.
when
checking
i.e.
hospital
tags.
If a device
pump.
during
identification
in a pump.
incoming
markings,
is
used
on a patient,
inspection
will
asset
it
will
3.1.2
Upon
chandise
extent agent notify an age,
indicating
Service
Damaged
receipt,
should
of
the
damage
of the
transportation
the
transportation
inspection
and
file a claim
the
Center
accompanied company
3.1.3
Items
Service
ordered
Credit
has
Receiving
ordered
is
notified
in
error
will
be
Proper
authorization
Goods
count
and
inspect
not
be
accepted.
or
loss
company.
company
report
by a delivery
properly
item
and
(refer
rendered.
with
the
quantity
to
1.4.1)
receipt
noted
Errors
or
shipped
(refer
must
issued
in
to
be
returned
for
all
has
your
Identify
noted
on
immediately,
B.
Braun
carrier,
damaged
and
received
or
such
damage
error
may
1.4),
and
freight
returns
been
for
obtained
freight
the
If
providing
an
the
before
any
items
customer’s
damage
will
be
happy
your
or
not
within
inspection
or
loss.
be
returned
products
prepaid
goods
received
(refer
the
carrier
that
are
copy
is
discovered
and
request
to
issue a credit
request
received
ten
are
days.
report
to
the
warehouse
are
returned
by
the
customer.
in
error
to
1.4.2).
departs.
damaged
of
the
after
that
and
promptly
The
upon
which
within
provided:
Damaged
or
lost,
and
delivery
receipt
inspection
the
claim
document
of
be
for
the
carrier’s
forwarded request
the
transportation
provided
14
days.
mer-
have
the
by
an
shipment,
made
and
loss
or
dam-
freight
to
must
bill
the
be
Customer
Items
|
+ *
950953
Products Products
Rev B (9/04)
are are
in
the
original
current
undamaged
inventory
items.
cases,
suitable
for
immediate
resale.
Page
15
of
144
OUTLOOK™
e
Merchandise
200
SERVICE
has
been
MANUAL
shipped
and
billed
to
the
customer
by
B.
Braun.
Certain *
*
Except
tomer.
3.1.4 B.
cause
acts
rial
rials,
3.1.5
When
specified added
products
Products Products
in
the
Late
Braun
shall
beyond
of
God,
shortages,
equipment,
Expedited
expedited
by
to
the
are
not
eligible
which
whose
case
have
labels
of a B.
been
or
Braun
Shipments
be
its
or
war,
fire,
excused
its
acts
flood,
or
transportation.
from
suppliers’
of
government,
strike,
Shipments
transportation,
the
customer,
an
invoice.
for
return
used.
seals
have
been
error,
any
items
delay
or
subcontractors’
acts
labor
trouble,
specialized
additional
charge
for
credit.
tampered
must
be
in,
or
impossibility
control,
of
purchase
sabotage
services,
to
cover
These
returned
are:
with
or
freight
of
including,
priorities
or
delay
or
alternate
the
premium
removed.
prepaid
performance
but
or
allocations,
in
obtaining
transportation
expense
by
due
not
labor,
will
limited
raw
modes
the
to
mate-
mate-
be
cus-
any
to:
are
3.2
RECEIVING
Each
Outlook There therefore,
1.
Before
is,
however,
the
opening obvious open
the
to
place
return
damaged
Customer
2.
Carefully case
the
3.
Remove are
no
visible
4.
Complete
INSPECTION
pump
following
damage.
container
responsibility
Service,
remove
pump
the
Section
is
factory
the
possibility
inspection
the
shipping
If
damage
until a representative
equipment
whether
the
ever
has
plastic
cover(s)
dents,
tears
6.2 of
for
pump
to
the
inspected
of
in-transit
is
recommended.
container,
is
present,
shipping
to
or
from
be
on
or
damage
B.
Braun
not
B.
the
shipped
the
Rate
scratches
Performance
and
tested
damage
inspect
notify
of
the
the
after
unless
Braun
is
responsible
shipping
for
repair.
Displays
on
the
pump
Check
prior
to
its
final
which
it
carefully
may
to
freight carrier
shipping
the
specifically
company
shipping
authorized
for
carton.
Retain
and
Information
displays
outlined
in
the
or
Chapter
packaging
or
may
not
assure
the
absence
immediately,
is
present.
container
the
is
opened).
to
do
repair.
packaging
Screen;
key
panel.
6.0.
and
shipment.
be
obvious;
of
and
wait
(It
is
difficult
Do
so
by
materials
make
sure
to
not
in
there
Page
16
of
144
950953
Rev B (9/04)
3.3
CHECK-IN
Prior
to
use, indicated Materials
Each
repair manual
of
the
in
Authorization
pump
work
are
Outlook
PROCEDURE
it
is
recommended
this
Manual.
has
been
inspected
may
have
intended
to
pumps.
If
the
Number
caused
be
used
that
the
pump
fails
(refer
and
tested
some
misalignment.
by
qualified
pump
to
to
1.4.2.)
prior
service
RECEIVING
be
checked-in according
operate,
to
contact
shipment;
Therefore,
personnel
B.
however,
the
in
PROCEDURE
to
the
Braun
procedures the
for a Returned
in-transit
referenced
initial
and
procedures
handling
routine
or
in
this
checks
3.3.1 All
tery use
of
Tn
into a hospital in a storage Chained
3.3.2
Prior
Chapter
Main
Battery
rechargeable
used
in
this
the
pump
the
pump
order
for
to a hospital
to
maintain
grade
facility,
together
Performance
to
patient
6.0.
Check
batteries
pump
the
the
is
no
first
battery
electrical
not
in
for
outlet
Check
use,
every
and
Charging
automatically
exception.
time,
the
grade
electrical
discharge
Therefore,
battery
in a fully
outlet
the
whenever
vicinity
of
conservation
new
pump
must
when
it
is
be
fully
charged
outlet
for
at
charged
condition,
possible.
patients,
up
purposes.
pass
the
entire
not
in
use,
suggested
least
When
that
by
connecting
24
hours.
the
pump
charging
to 5 Outlook
Performance
The
rechargeable
before
attempting
the
should
multiple
pumps
may
Check
bat-
power
be
cord
plugged
devices
be
Daisy
outlined
to
in
950953
Rev B (9/04)
This
space
intentionally
left
blank.
Page
17
of
144
PUMP
CONFIGURATION
CHAPTER
PUMP
4.1
ENTERING
1.
Press
2.
Press
3.
Use
4.
Use
CONFIGURATION
for
(MENU).
the
È
the
Y
4.0
IN
THIS
+
How
to
*
How
to
*
How
to
THE
to
to
BIOMED
two
seconds.
highlight
highlight
CHAPTER
enter
the
Biomed
customize
set
up
the
Drug
OPTIONS
“Alternate
“Biomed
Please
security
and
YOU
WILL
Options
the
pump
software.
Lists
in
MENU
Menu,”
Options,”
Press
then
then
enter 4-digit
code,
such
1342
ENTER
LEARN:
Menu
DOSE
ON
press
(ENTER).
press
(ENTER).
as
of
the
Mode.
THE
pump.
PUMP
5.
Use
the ) (data)
The
Security
Security
the
hundreds
The
Security
950953
Rev B (9/04)
Code
Code
is
“1342,”
key
once,
Code
number
keys
to
enter
not
really a number,
press
the
“+”
then
the
tens
is
created
the
Security
on
the
key
once,
randomly
Code.
but a key
thousands
then
each
press
sequence.
key
press
(ENTER).
time
once,
the
screen
For
then
example,
the
ones key
is
shown.
to
enter
once,
Page
19
the
then
of
144
OUTLOOK™
4.2
CONFIGURATION
The
Outlook Biomed options Your pump the
completed
Options
may
B.
Braun
which
other
settings
by
200
SERVICE
pump
can
be
customized
Menu
be
placed
Sales
details
representatives
contains
in
either
Representative
where
which
the
may
be
from
MANUAL
to
fit
the
configuration
the
Main
or
can
provide
menu
items
customized.
the
Medical,
the
specific
options
Alternate
you
may
be
placed
This
Programming
Nursing,
needs
of
each
facility.
for
the
user
interface. A variety
Menus,
or
the
option
with a Programming
within
the
menu
Worksheet
and
Pharmacy
staff.
The
Outlook’s
may
be
disabled. Worksheet system,
should
along
be
for
with
of
the
According Professional
appropriate audible from
of
configuring
To
access
4.2.1
A
and
this
to
the
first
menus,
pressure
“Golden
the
following
Forcing
NOTE: pump
1.
Use
the
2.
Use
the
facility
wishes
creates a “Golden
as
well
as
data
defaults.
Pump”
multiple
which
pumps.
features,
Factory
lt
is
important
in
order
to
avoid
to
highlight
Y
A
to
highlight
<
expressed
Pump”
retention
Outlook
can
be
enter
WinDock
the
Defaults
to
“Force
corrupted
“Load
“VES,”
on
the
Programming
with
all
modes
defaults,
then
downloaded
Biomed
All
Options
Defauits"
data.
Factory
then
press
Defaults,”
and
drug
selections,
Worksheet,
options
the
placed
concentrations,
allows a configuration
to
the
next
pump,
expediting
°
Menu
before
then
(refer
to
configuring
press
(ENTER).
4.1).
.
ENTER).
Biomedical
in
the
and
file
to
be
made
the
process
a
3.
4.
5.
4.2.2
1.
2.
3.
Page
20
Press
POWER).
Press
and
Enter
the
Configuring
Use
the
Press
Use
the
up
of
144
hold
Biomed
Options
Menus
x
to
highlight
Y
again. A list
and
down
to
turn
on
the
pump.
Menu.
des
“Configure
of
modes
to
move
Menus,”
and
pump
through
then
press
features
the
appears.
list.
(ENTER).
i . | 950953
Rev B (9/04)
PUMP
4.
Press
the
right
and
left
to
select
the
appropriate
CONFIGURATION
choice
of
Main
Menu
(MAIN),
Alternate
5.
4.2.3
A
1.
2.
.
3.
Menu
When
finished,
Default
NOTE:
The
Occlusion
mmHg.
R
Use
the
<
L
To
select a specific
light
“Fixed
100,
When
Value=300
200,
300, 400,
finished,
(ALT),
or
press
Factory
to
highlight
Fixed
or
press
Disabled
Limit
Default
Value
mmHg,”
500
mmHg.
(DIS).
to
return
To
select
as
and
to
return
to
value
“Retain
the
Default
use
the
to
the
for
+]
| |
the
the
keys
Biomed
Last
Menu.
Occlusion
Setting,”,
=.
Occlusion
to
change
Biomed
Menu.
Limit
then
Limit,
use
the
Occlusion
is
300
press
the
ENTER).
Y
Limit
to
hight-
to
75,
4.2.4
1.
Use
2.
Use
3.
When
4.2.5
1.
Use
ble
2.
Use
3.
Press
Data
Retention
the
the
finished,
Audible
A
the
Y,
pump
the
up
the
options
and
right
to
highlight
to
highlight
press
Signal
to
highlight
appears.
down
and
left
Defaults
“Data
either
to
Options
“Audible
A
to
Y
Retention
“Clear
return
Signal
move
to
select
Defaults,” then press
All”
or
“Retain
to
the
Biomed
Options,”
through
“Y”
the
(Yes)
list.
or
“N”
All.”
Menu.
then
press
(No)
(ENTER).
(ENTER). A list
to
enable
or
disable
of
audi-
an
950953
option.
Rev B (9/04)
Page
21
of
144
Loading...
+ 126 hidden pages