B Braun Infusomat fmS User manual

Infusomat®fmS
Instructions for Use
Software IFME/IFMe
Attention: Consult accompanying documents!
Operation
secured.
Prior to use check audible and visual alarms
during self test. Also check the device for possible damage.
If staff call is used we recommend to check the
equipment once after connecting the pump.
Connect to patient only after switching on the
device. Interrupt the connection during changing administration set(s) to prevent incorrect dose delivery.
Select cannula/catheter suitable for use with the
intended medical application.
Position the infusion line free of kinks.Recommended change of disposables after 24 h
(consider national hygiene regulations).
Compare displayed value with entered value.
Start infusion only if values are corresponding.
Installation in medically used rooms must comply
with the appropriate regulations (e.g. VDE 0100, VDE 0107 or IEC-publications.
Possible explosion hazard if used in presence of
flammable anaesthetics!
Air in line cannot be detected by the air detector
at stop-cocks, infusion ports and additional administration set components.
Other components
Variations in pressure (e.g. as caused by change
of level) can affect the accuracy of the device.
Where several infusion lines are connected on
one single vascular access the possibility of their exerting a mutual influence vice-versa cannot be excluded.
Refer to respective manufacturer’s information
for possible incompatibilities of equipment resp. drugs.
Use only compatible combinations of equipment,
accessories, working parts and disposables.
It is recommended to use original Infusomat®
infusion lines only.
The use of not recommended resp. incompatible
disposables may influence the technical specification.
Connected electrical equipment must comply
with the relevant IEC/EN-publications (e.g. IEC/EN 60950 for data-processing equipment). The user/operator is responsible for the system configuration if additional equipment is connected. The international standard IEC/EN 60601-1-1 has to be taken into account.
Safety Standards
The Infusomat® fmS satisfies all safety standards for medical electrical devices in compliance with IEC/EN 60601-1 and IEC/EN 60601-2-24.
The EMC-limits (electro-mangnetic compatibility)
according to IEC/EN 60601-1-2 and IEC/EN 60601-2-24 are maintained. If the equipment is operated in the vicinity of other equipment which may cause high levels of interference (e.g. HF surgical equipment, nuclear spin tomography units, mobile telephones etc.) maintain the recommended protec­tive distances for these devices. Under certain conditions malfunctions may occur which lead to a device alarm with permanent alarm tone (see also alarm conditions, page 13). Interferences may occur e.g. at electro-magnetic fields > 10 V/m resp. electro-magnetic discharges > 8 kV.
Special Function "without drip control”,
see page 10.
Read Instructions for Use prior to use. Application only under regularly supervision by specially trained staff.
Patient Safety
Infusomat®fmS
Contents
Infusomat® fmS / Overview Page 4
Operation Page 6
Additional Settings Page 7
Special Functions Page 8
Alarm Conditions Page 13
Start-up Graphs and Trumpet Curves Page 15
Technical Data Page 16
Warranty / TSC*)/ Service / Cleaning Page 18
Ordering Page 19
*
)
Technical Safety Check
The Infusomat® fmS is according to IEC/EN 60601-1 resp. IEC/EN 60601-2-24 a volumetric infusion pump for infusion of small and large volumes at highest accuracy and is suitable for intravenous and intra-arterial applications, for blood transfusion and for enteral nutrition.
The medical specialist has to decide on the suitability of the application. The decision has to be made on the basis of the specified properties and technical data.
For further details please refer to the Instructions for Use.
3
4
Handle
For easy transport.
Holder for Drop Chamber
Prevents unintended movement (swaying).
Adapter for Drop Sensor
Replaceable, depending on drop chamber size. Press lateral and pull off.
Universal Pole Clamp
Attach the Infusomat® fmS from above, clicking into its place. To release, press the black button.
Overview
Display
All important information in plain text. Green background illumination only if connected to mains or a key is pushed at battery operation.
Correct Input/CLEAR
Display reset to 000.0
Red LED indicates alarm condition
Additionally "AAA.A” is flashing in the display. Possible alarms: "drop alarm, air alarm, pressure alarm, pump-door open, battery alarm, standby-alarm”
Keypad for Input
Operating Indicator
Additional operating control indicator in display.
5
Guide for Short Infusion Pole
VOL Infusion Volume
Press key below VOL. Enter volume (0.1 ... 9999.9 ml). Confirm. Press again key below VOL.
TIME Infusion Time
Press key below TIME. Enter time e.g. 50 min as 5 0 or 2 h 30 min as 2 3 0. Confirm. Press again key below TIME.
RATE Delivery Rate
Only active when rate is calculated automatical­ly. The key below RATE confirms the calculated rate.
SF Special Functions
If activated: dose calculation / bolus function / standby / drug selection / occlusion pressure / drop control / piggyback / battery capacity / data lock / contrast / date, time.
Function Keys
Mains/Power Connection
(protect against ingress of moisture). In case of mains/power failure, the pump switches to battery operation. Battery operation time: > 3.5 h at highest delivery rate. Automatic over­load protection. Mains fuse: directly above the mains/power connector.
Potential Equalisation
To be connected for CF-applications.
Aluminium Housing
Easy-care, drip water protected, resistant to disinfectants.
Mains/Power Supply
For operation with fluid manager system.
Multi-Function­Connector (MFC)
Connection for staff call, ambulances (12 V DC) and inter­face for fm anaes­thesia/fm intensive.
Optical Interface
Infrared Interface for operation on "fluid manager system”.
Flow Inhibitor
Opening door clamps off infusion line automatically.
Peristaltic Pump
For precise and reliable dosage.
Door Opener
Special Functions "SF”
Function Keys (Soft keys)
Mains/Power Switch ON/OFF
Suppress alarm tone for 2 min
Infusion START/STOP
Infusion
1. Ensure reliable installation
Never position infusion bottle below pump level.
Connect staff call.Insert spike vertically into infusion bottle.
Fill lower part of drop chamber to max. 2/3.
Open roller clamp.
2. Filling and Venting
Fill infusion line from bottom to top.
Close roller clamp.
3. Insert Infusion Line
Press door opener.
Insert infusion line: Locate clips first on top,
then on bottom.
Keep infusion line in place at air sensor.Close door. In the area of the peristaltic
pump segments and free-flow clamp the infusion line will be positioned self-acting.
Open roller clamp completely. There may not
be a continuous drip.
Place drop sensor on drop chamber
(if necessary, use an appropriate adapter).
4. Switch On with
Green mains/power control or yellow battery control, alarm tone and display LED’s are briefly activated.
5. Puncture
6. Setting the Delivery Rate
In the range: 0.1 - 999.9 ml/h and check (selectable in 0.1 ml/h-increments). Correction: Press C, then set new rate. Select
additional settings, if desired (see page 7).
7. Press START
Operating symbol appears on the display and the green operating indicator lights.
8. Stop the infusion
Press STOP. Green operating indicator goes
out
Close roller clamp.Press door opener. Infusion line will be
clamped off when opening the door.
Remove infusion line. First bottom, then
top.- Replacement: Insert new infusion line as described. Then press START again.
To end switch off
Press for 2 sec.
Operation
6
Original Infusomat® line with silicone-pump element
upper clip
free flow clamp
lower clip
air sensor
Loading...
+ 14 hidden pages