Lektrafuse HF Generator GN200
Legend
Aesculap
Lektrafuse HF Generator GN20 0
®
Symbols on product and packages
1 Signal lamp HF-ON (surrounding ring)
2 HF activation button (RF on/off) (center)
3 Connection socket (foot switch)
4 Signal lamp regrasp
5 Display
6 Signal lamp error
7 Signal ring instrument
8 Connection socket (instrument)
9 Indicator lamp power ON
10 Equipotential connector
11 Fuse holder (with 2 fuses)
12 OFF/ON switch
13 Mains power input socket
14 Type plate
15 Ventilation slots
Caution, general warning symbol
Caution, see documentation supplied with the product
Error message
Follow the instructions for use
Marking of electric and electronic devices according
to directive 2002/96/EC (WEEE). see Disposal
Foot switch
Regrasp
Type CF applied part, defibrillation-protected
Equipotentialization
LOT
Fuse
Non-ionizing radiation
Alternating current
Batch designation
Serial number
Order number
Date of manufacture
2
Contents
1.Safe handling
1.Safe handling3
1.1Safe for use in conformance with IEC regulations4
2.Product description5
2.1Scope of supply5
2.2Components required for operation5
2.3Intended use5
2.4Operating principle5
2.5Acoustic warning signal6
2.6Monitoring functions6
2.7Output power diagrams7
2.8Maximum peak output voltage (Up)7
3.Preparation and setup7
3.1First use7
4.Working with the Lektrafuse HF Generator GN2008
4.1System set-up8
4.2Function checks10
4.3Safe operation10
5.Validated reprocessing procedure10
5.1General safety instructions10
5.2General information11
5.3Preparations at the place of use11
5.4Preparation before cleaning11
5.5Cleaning/disinfection11
5.6Wipe disinfection for electrical devices without sterilization12
5.7Inspection, maintenance and checks12
5.8Storage and transport12
6.Maintenance12
6.1Safety inspection12
7.Troubleshooting list14
7.1Regrasp error14
7.2Warnings15
7.3Error reports15
7.4Failure messages16
7.5Problem solving16
7.6User maintenance17
7.7Fuse replacement17
8.Technical Service17
9.Accessories/Spare parts18
10.Technical data18
10.1 Ambient conditions19
11.Disposal19
12.Distributor in the US/Contact in Canada for product
information and complaints19
CAUTION
Federal law restricts this device to sale by or on order of a physician!
Note
The present instructions for use only describe the layout, function and
operation of the Lektrafuse HF generator GN200 and are not suitable for
instructing a beginner in high-frequency electrosurgery. A general description of the application of high-frequency electrosurgery can be found in
the respective O.R. manuals and other relevant literature.
These instructions and the device described are intended exclusively for use
by qualified medical specialists who have been trained in the specific technique and the surgical procedures to be carried out.
Risk of injury to patients due to inappropriate
application!
DANGER
► Transport the product only in its original box.
► Prior to use, check that the product is in good working order.
► Observe “Notes on Electromagnetic Compatibility (EMC)“, see
TA022130.
► The Lektrafuse HF generator GN200 fulfills the requirements of
CISPR 11, Class A.
► To prevent damage caused by improper setup or operation, and in order
not to compromise warranty and manufacturer liability:
– Use the product only according to these instructions for use.
– Follow the safety and maintenance instructions.
– Only combine Aesculap products with each other.
– Adhere to application instructions according to relevant norms, see
Safe for use in conformance with IEC regulations.
► Inspect the accessories regularly: Electrode cables and endoscopic
accessories, in particular, must be checked for possible damage to the
insulation.
► Keep the instructions for use accessible for the user.
► Always adhere to applicable standards.
► The product and accessories should only be
operated by qualified or trained and experienced personnel.
3
Aesculap®
Lektrafuse HF Generator GN200
1.1Safe for use in conformance with IEC regulations
The operating environment
Risk of death by electric shock!
► Do not open the product.
DANGER
WARNING
CAUTION
► Only ever connect the product to power mains
with equipment grounding conductor.
Danger of injury from burns and/or explosion from
flammable gases! Sparks can occur even if the HF
surgical device is used properly and according to its
intended use.
► Do not use the device in explosion-hazard zones.
► When operating in the head or thoracic region,
avoid using combustible anesthetics and accelerating gases (e.g. nitrous oxide or oxygen) or,
when using such substances, ensure they are
extracted from the region of operation.
► If possible, use incombustible cleaning and dis-
infecting agents.
► If combustible cleaning and disinfecting agents
or solvents have to be used: Verify that such
agents have evaporated prior to commencing HF
surgery.
► Ensure that no combustible fluids accumulate
under the patient or in body cavities (e.g. the
vagina). Wipe off all fluids before using the HF
surgical device.
► Ensure the absence of any endogenous, combus-
tible gases.
► Check that oxygen-soaked materials (e.g.
absorbent cotton or mull) are kept at a safe distance from the HF field, so that they cannot
ignite.
Risk of interference with other devices!
Even during normal use, the HF surgical device cre-
ates electromagnetic fields that can interfere with
other devices.
► Check that no electronic devices that could be
subject to interference by electromagnetic
fields are set up in the vicinity of the HF surgical
device.
Patient safety
DANGER
DANGER
DANGER
WARNING
WARNING
Danger due to inadequate preparation or faults in
the HF surgical device!
► Check that the HF surgical device is in perfect
working order.
► Ensure that neither the foot switch nor the hand
switch has been penetrated by conductive fluids
(e.g. blood, amniotic fluid).
► Ensure there is no short circuit in the foot or
hand switch cables.
Risk of burns suffered by the patient due to inadvertent activation of the HF surgical device!
► In case of any inadvertent activation of the HF
surgical device, switch off the device immediately at the OFF/ON switch.
► Always exercise particular care when operating
the foot switch.
Risk of injury to the patient due to uncontrolled
rise of the HF output voltage due to some fault in
the HF surgical device!
► Stop using the HF surgical device as soon as it
shows even the slightest anomaly.
Risk of injury to patients/users due to defective
power cord or missing protective ground connections!
► Check the mains power cord/protective ground
connections.
Danger of injuries due to muscle contraction,
caused by stimulation of the nerves and muscles!
► Work with particular care on sensitive struc-
tures.
Restriction to view and/or side-effects due to the
development of steam/smoke during HF surgery!
CAUTION
► Ensure that the device does not come in direct contact with the patient
or in the sterile area respectively.
► Ensure that the user does not come into direct contact with the patient
and the HF device at the same time.
4
► If necessary, use smoke suction.
Always be sure to do the following:
► Position the patient in such a way that s/he is not in contact with any
metal parts that are grounded or have a significant electric capacity
against ground (e.g. operating table, fixtures). If necessary, interpose
antistatic drape.
► Ensure that the patient will not be in contact with any damp cloths,
drapes or bedding.
► Safeguard areas prone to strong perspiration against skin contact with
the trunk of the patient’s body by inserting antistatic tissue between
such areas and the trunk.
► Siphon off urine with a catheter.
► For open heart surgery, ground the HF surgical device via potential
equalization.
► For patients with cardiac pacemaker or other active implants, consult
with the relevant medical specialist prior to applying HF surgery, so
that irreparable damage to the pacemaker or implant can be avoided.
► If possible, remove from the vicinity of HF electrodes any electrodes
from physiological monitoring devices that are not equipped with protective resistors or HF dampers.
► Do not use needle electrodes for intraoperative monitoring.
► Arrange the wires and cables of monitoring devices in such a way that
they do not come in contact with the patient’s skin.
► Keep the leads to the HF electrodes as short as possible, and arrange
them in such a way that they do not touch the patient or any other
wires or cables.
► Should the output power appear insufficient with the usual settings,
check that:
– the working electrodes are clean,
– all plug connections are properly in place.
► Never lay active electrodes on or next to the patient.
► Put down active electrodes that are not needed at any particular
moment, in such a way that they will not touch the patient.
► For operations involving unavoidable, continuous contact between the
electrodes and the patient (e.g. endoscopic procedures), press the OFF/
ON switch 12 at the HF surgical device immediately after any inadver-
tent activation of the electrode.
► Do not remove hot electrodes from the patient’s body immediately
after performing cutting or coagulation procedures.
► During operation of the generator, do not touch the electrodes on the
instrument (on the upper or lower side) as the electrodes can heat up
and this may lead to burns or injuries.
2.Product description
2.1Scope of supply
DesignationArt. no.
High frequency surgical deviceGN200
Instructions for useTA022377
Notes on Electromagnetic Compatibility
2.2Components required for operation
■ Power cord
■ Foot switch (optional)
■ Caiman Instrument
2.3Intended use
The Lektrafuse HF generator GN200 is used for vessel sealing and vessel
division in open and minimally invasive surgery. The instruments can seal
vessels of up to and including 7 mm.
The Lektrafuse HF generator is not suitable for use in tube sterilization/
tube coagulation for sterilization.
With respect to the electric shock hazard, the Lektrafuse HF generator
meets the classification and safety requirements of a type CF device.
The Lektrafuse HF generator is intended for operation and storage in
closed spaces.
2.4Operating principle
The Lektrafuse HF generator GN200 works on a micro-processor controlled basis and converts the voltage of the mains supply into a high-frequency alternating current for bipolar vessel sealing.
The sealing process is carried out via a closed control circuit. The sealing
can be started and stopped via a button on the instrument, the button on
the front of the device or a foot switch.
TA022130
5
Aesculap®
Lektrafuse HF Generator GN200
2.5Acoustic warning signal
StatusAcoustic warning signalNotes
System failure3 repetitions of: 1 long beep 3 KHz, 1 beep 2 KHz Following each system failure, if identified (F001,...)
REGRASP error3 repetitions of: 1 beep 2 KHz,
1 beep 3 KHz
System error 3 beeps, 3 KHzFollowing each failure report, if identified (E001,...)
System warning3 beeps, 3 KHzFollowing each warning, if identified (W001,...)
Start HF discharge2 beeps, 480 HzWhen starting the HF discharge
HF discharge activeRepeating beep 480 HzContinually during HF discharge
HF discharge ended3 beeps, 530 HzWhen a sealing process has been successfully completed (no
Power up self-test1 beep 2 KHzDuring the self-test, in order to check that the signal tone is
Following each regrasp error, if identified
failure reports or errors)
functioning
2.6Monitoring functions
Self-test
As soon as it is switched on, the unit performs a self-test of the control
elements, the acoustic warning signal, the microprocessor and the hardware function. During this phase, the message "Performing Self Test"
appears in the display 5.
Fig. 1
Continuous test cycle during operation
During operation, safety-relevant functions and signals are monitored
through a continuous test cycle. As soon as a critical error is detected, the
HF device terminates the HF activation. In display 5, the error number of
the corresponding error is displayed and an acoustic warning signal is
emitted, see Troubleshooting list.
6
2.7Output power diagrams
Representation of the output power as a function of the load resistance.
3.Preparation and setup
Non-compliance with the following instructions will preclude all responsibility and liability in this respect on the part of Aesculap.
► When setting up and operating the product, adhere to
– national regulations for installation and operation,
– national regulations on fire and explosion protection.
Note
For the safety of patients and users it is essential that the mains power cord
and, especially, the protective earth connection are intact. In many cases
defective or missing protective earth connections are not registered immediately.
► Connect the device via the potential equalization terminal at the rear
panel of the device to the potential equalization system of the room
used for medical purposes.
Fig. 2Output power diagram Caiman 12 mm Seal & Cut instrument
family
Fig. 3Output power diagram Caiman 5 mm Seal & Cut instrument
family
2.8Maximum peak output voltage (Up)
Note
The following data allow the user to determine whether the HF device
GN200 is suitable for a given accessory (insulation rating).
Note
The potential equalization lead can be ordered from the manufacturer as
article no. GK535 (4 m length) or TA008205 (0.8 m length).
► In order to ensure sufficient ventilation and cooling of the HF device,
leave at least 10 cm to 15 cm of space around the HF device.
3.1First use
Risk of injury and/or product malfunction due to
incorrect operation of the electromedical system!
WARNING
► Adhere to the instructions for use of any medi-
cal device.
Risk of injury to patient or user caused by inadequate accessories (insulation rating)!
WARNING
Maximum peak output voltage (Up) 200 V
► Make certain that the accessory voltage rating
specified in the product documentation is
higher than maximum peak output voltage.
7
Aesculap®
Lektrafuse HF Generator GN200
4.Working with the Lektrafuse HF Gener-
ator GN200
4.1System set-up
Connecting the accessories
Risk of injury due to unapproved configuration
using additional components!
DANGER
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the performance and safety characteristics of the products.
Only CAIMAN instruments can be used in conjunction with the Lektrafuse
HF generator GN200, see Accessories/Spare parts.
Connecting the power supply
DANGER
► For all applied components, ensure that their
classification matches that of the application
component (e.g. Type BF or Type CF) of the
respective device.
Risk of death by electric shock!
► Only ever connect the product to power mains
with equipment grounding conductor.
► Set up the device in such a way that switching
it off using the power OFF/ON switch is possible
and straightforward.
► Set up the device in such a way that disconnect-
ing the power cord is straightforwardly possible.
Switching on the device
► Switch on the device with the power OFF/ON switch 12.
Indicator lamp power ON 9 is illuminated. The device performs a selftest, see Continuous test cycle during operation:
If no instruments are connected, the message "Attach Instrument"
appears in the display 5 after the self-test. If the self-test fails, an error
message is shown in the display 5.
Fig. 4
Connecting the foot control
Connecting the foot switch GN201 is optional.
Both the foot switch plug and connection socket (foot switch) 3 have a
dot marking. To connect the plug to the connection socket correctly, these
markings must be aligned.
The voltage of the mains power supply must correspond to the voltage
indicated on the type plate of the device.
The device is fitted with a universal power supply unit, which means that
it automatically adapts to mains voltages ranging from 100 V to 240 V
without having to switch between voltage ranges.
► Ensure that the device is switched off. Also switch off with the power
OFF/ON switch 12.
► Plug in the power cord in mains power input socket 13 in the rear
panel of the HF surgical device.
► Plug in the mains plug at the building mains power socket.
8
Fig. 5
► Align the foot switch plug and insert it into the foot switch connection
socket 3 until it engages.
When the device and the foot switch have been successfully connected, the message "Footswitch attached" appears for 3 seconds in
the display 5.
Fig. 6
Note
Do not activate the foot switch continually during HF discharge. HF discharge is started and stopped by pressing once on the foot switch and then
releasing.
Remove the foot switch
► Turn the locking ring counter-clockwise and withdraw the connecting
plug at the same time
The message "Footswitch removed" appears in the display 5 for 3 sec-
onds.
Connecting the instrument
Only instruments of the CAIMAN Seal & Cut product family can be used
in conjunction with the Lektrafuse HF generator GN200.
The instrument connector has an arrow marking and the connection
socket 8 has a dot marking. To connect the plug to the connection socket
correctly, these markings must be aligned.
Fig. 7
Fig. 8
► Connect the instrument with the connection socket 8.
The green signal ring 7 around the instrument connector lights up. The
message "Ready to Seal" appears in the display 5.
Fig. 9
9
Aesculap®
Lektrafuse HF Generator GN200
Decommissioning
► Switch off the device with the power OFF/ON switch 12.
All the poles of the device are disconnected from the power supply.
► Disconnect power cord
The device is completely disconnected from the power supply.
4.2Function checks
Prior to each use, carry out a functional test of the device:
► Check that the accessories have no visible damage.
► Prepare and set up the unit, see Preparation and setup.
► Check the functionality of the following elements, one after the other,
in the following sequence:
– Switch on the power OFF/ON switch 12; the power on signal lamp 9
lights up
– Automatic self-test after each time the device is switched on: brief
acoustic signal, all display elements light up briefly
– Foot control, see Connecting the foot control
– Instrument, see Connecting the instrument
– Activation with RF on/off 2 on the front of the device, button on
the instrument and foot switch
– Switch off the the power OFF/ON switch 12
To activate/deactivate device
► Switch on/switch off the device with the power OFF/ON switch 12.
Operating mode
The Lektrafuse HF generator GN200 automatically recognizes the connected instrument. The internal device settings are loaded. It is not necessary/possible for the user to set the operating mode or the parameters.
Activating the HF current
► Be sure that the patient is prepared in such a way that the HF unit can
be applied without any risk to the patient.
► Ensure that the instrument and foot switch, if used, are correctly con-
nected.
► Activate the HF device via the hand switch, foot switch or HF activa-
tion button (RF on/off) 2 on the front of the device.
Note
Do not activate the foot switch or button continually during HF discharge.
The vessel sealing process is started and ended by pressing the foot switch
or the HF activation button (RF on/off) 2 on the front of the device or button on the instrument once. After completing the sealing process, the
device stops the HF discharge automatically. The sealing process can be
interrupted by repeatedly pressing the button.
4.3Safe operation
Risk of injury and/or malfunction!
► Always carry out a function check prior to using
WARNING
WARNING
► Start the sealing process (HF discharge) by pressing the button on the
instrument, the foot switch or the HF activation button (RF on/off) 2
on the front of the device once.
After completing the sealing process, the HF discharge is automatically
stopped.
During HF sealing, the message "Sealing in Progress" is shown on the
display 5.
Fig. 10
the product.
Risk of injury when using the product beyond the
field of view!
► Apply the product only under visual control.
5.Validated reprocessing procedure
5.1General safety instructions
Note
Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
Note
For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning
the reprocessing of products.
Note
Mechanical reprocessing should be favored over manual cleaning as it
gives better and more reliable results.
Note
Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this.
The recommended chemistry was used for validation.
10
Note
If there is no final sterilization, then a virucidal disinfectant must be used.
Note
For the latest information on reprocessing and material compatibility see
also the Aesculap extranet at www.extranet.bbraun.com
The validated steam sterilization procedure was carried out in the Aesculap
sterile container system.
5.4Preparation before cleaning
► Disconnect the unit from mains power.
► Remove accessories.
5.5Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
5.2General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating precleaning temperatures >45 °C nor fixating disinfecting agents (active
ingredient: aldehydes/alcohols) should be used.
Excessive measures of neutralizing agents or basic cleaners may result in
a chemical attack and/or to fading and the laser marking becoming
unreadable visually or by machine for stainless steel.
Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must
be removed by rinsing thoroughly with demineralized water and then drying.
Additional drying, if necessary.
Only process chemicals that have been tested and approved (e.g. VAH or
FDA approval or CE mark) and which are compatible with the product’s
materials according to the chemical manufacturers’ recommendations
may be used for processing the product. All the chemical manufacturer's
application specifications must be strictly observed. Failure to do so can
result in the following problems:
■ Optical changes of materials, e.g. fading or discoloration of titanium or
aluminum. For aluminum, the application/process solution only needs
to be of pH >8 to cause visible surface changes.
■ Material damage such as corrosion, cracks, fracturing, premature aging
or swelling.
► Do not use metal cleaning brushes or other abrasives that would dam-
age the product surfaces and could cause corrosion.
► Further detailed advice on hygienically safe and material-/value-pre-
serving reprocessing can be found at www.a-k-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
DANGER
CAUTION
CAUTION
CAUTION
Risk of electric shock and fire hazard!
► Unplug the device before cleaning.
► Do not use flammable or explosive cleaning or
disinfecting solutions.
► Ensure that no fluids will penetrate the product.
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
► Only clean and disinfect the product manually.
► Do not sterilize the product under any circum-
stances.
Damage to the product due to inappropriate cleaning/disinfecting agents!
► Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer’s
instructions for the respective cleaning/disinfecting agent.
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures!
► Use cleaning agents and disinfectants according
to the manufacturer's instructions.
► Observe specifications regarding concentration,
temperature and exposure time.
5.3Preparations at the place of use
► Remove any visible surgical residues to the extent possible with a
damp, lint-free cloth.
► Transport the dry product in a sealed waste container for cleaning and
disinfection within 6 hours.
11
Aesculap®
Lektrafuse HF Generator GN200
Validated cleaning and disinfection procedure
Validated procedureSpecial featuresReference
Wipe disinfection for electrical
devices without sterilization
–Chapter Wipe disinfection for electrical devices
without sterilization
5.6Wipe disinfection for electrical devices without sterilization
► Remove any visible residues with a disposable disinfectant wipe.
► Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
► Observe the specified application time (1 min minimum).
Conc.
[%]
Water qualityChemical
6.Maintenance
No servicing is required for the Lektrafuse HF generator GN200.
Note
Maintenance must only be carried out by authorized personnel. If necessary, the circuit diagrams and the service manual, which contains all the
necessary documents, are provided for this.
5.7Inspection, maintenance and checks
► Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
► Set aside the product if it is damaged.
► Perform a safety inspection annually, see Maintenance.
5.8Storage and transport
Damage to the product caused by switch-on too
early after storage/transport at temperature below
CAUTION
+10 °C (50 °F)!
► Allow the HF device to acclimate at room tem-
perature for about 1 hour.
6.1Safety inspection
A safety inspection must be carried out annually.
The person in charge of the inspection will document the inspection find-
ings and measurement values according to the printed inspection protocol.
► Only have the product and its accessories inspected by persons who
possess the requisite training, knowledge and experience and who are
authorized to do so.
► If there are significant deviations from the inspection protocol values,
or if the prescribed maximum values are exceeded: Send in the device.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
► Transport the product only in its original box.
► For storage and transport conditions, see Ambient conditions.
12
Inspection protocol – safety inspection
TEST INTERVAL:1 Year
TYPE OF DEVICE:
MANUFACTURED BY:
SN: ...............................
SCOPE OF INSPECTION Verdict
1.) VISUAL INSPECTION
1.1Inspection of power cable for visually detectable damage
1.2Inspection of foot switch for visually detectable damage
1.3Type plate still present and readable
1.4Check fuse links for prescribed values
1.5Overall condition of device: dirt, damage
2.) ELECTRICAL INSPECTION
Protective earth resistance incl. mains cable
2.1
according IEC62353:2007
Earth leakage current
2.2
according IEC60601-1*
Enclosure leakage current/touch current
2.3
according IEC60601-1*
Patient leakage current
2.4
according IEC60601-1*
Equipment leakage current - alternative method
2.5
according IEC62353:2007
Applied part leakage current - alternative method
2.6
according IEC62353:2007
2.7HF-leakage current bipolar
according IEC60601-2-2:2009
2.8Insulation resistance
according IEC62353:2007 (Preparation according Service Manual)
2.8.1Mains against HF output
2.8.2Mains against housing
2.8.3HF-output against housing
2.9HF-output power measurement with induction-free resistor
InstrumentChannelR
C5 1 50
C24/44150
= 0,50 mA
= 1,00 mA
= 0,10 mA
= 0,50 mA
= 0,01 mA
= 0,05 mA
= 1,00 mA
= 0,05 mA
=122mA
= 7 M:
= 2 M:
= 2 M:
= 2 M:
Performed:
Verdict
Pass / Fail
* The listed allowable values comply with the requirements of the IEC60601-1:1988+A1:1991+A2:1995 and IEC60601-1:2005
Fig. 11 Inspection protocol form Safety inspection
13
Aesculap®
Lektrafuse HF Generator GN200
7.Troubleshooting list
7.1Regrasp error
A regrasp error, i.e. potentially insufficient sealing of the vessels, is displayed as follows:
■ Clear sound sequence and interruption of the sound sequence of HF
operation, also halts the HF discharge
■ The yellow regrasp error light 4 on the front side of the HF device
flashes
■ Warning message in the display 5 on the front side of the HF device
Regrasp error messageSignificanceRemedy
REGRASP INDICATOR – OPEN
Visually check seal – Manually cut
- Clean jaws
OR
Regrasp thicker tissue
REGRASP INDICATOR – SHORT
Visually check seal – Manually cut
Check for metal or occlusion in jaw
Remove excess fluids
REGRASP INDICATOR – TIME
Visually check seal – Reseal as needed
Seal cycle interrupted
Seal endpoint not reached
Procedure if a regrasp error occurs:
► Do not cut the vessel.
► Open the instrument.
► Check the sealed vessel.
When correct hemostasis has occurred:
► Cut the vessel manually with surgical scissors.
When hemostasis has not occurred in the correct way:
► Place the instrument on the vessel again and lock it in place.
Note
The regrasp error must not be confirmed. The activation of the hand switch,
foot switch or HF activation button (RF on/off) button 2 directly starts the
next sealing process.
The vessel seal must be checked. Contact areas of the jaw
part must be covered with dried vessel residue or bodily
fluids.
- or The gripped bundle of tissue is too thin
There is metal (e.g. staples) or other foreign material
between the electrodes in the jaw part
HF device has reached its maximum sealing time and has
not ascertained any complete seal
- or The user has ended HF discharge during sealing
Clean contact areas with sterile water and
a sponge
Grip a thicker tissue bundle
Open and check jaw part
Remove any foreign material
Remove excess suction fluids in the operating area
Check the quality of the seal
► Start the additional HF cycle.
► If the additional HF cycle leads to a normal seal, the vessel can be cut,
and the instrument can be opened and removed.
► If a regrasp error occurs again, do not cut the vessel, open the instru-
ment and manually cut the sealed vessel with surgical scissors. When
so doing, ensure that the hemostasis is correct.
14
7.2Warnings
A warning is indicated as follows:
■ Individual tone (different from a failure, an error or normal HF opera-
tion)
■ Warning message in the display 5 on the front side of the HF device
Warning messageSignificanceRemedy
Generator Cooling
Please Wait
Generator Too Cold
Please Wait
Excess generator temperatureCheck the airflow around the generator
Generator must first warm upDo not switch off the generator whilst it is warming up
7.3Error reports
An error is indicated as follows:
■ Clear tone sequence (different from a failure or normal HF operation)
■ Error message in the display 5 on the front side of the HF device
Footswitch Error
Release footswitch
Remove and reattach footswitch
"HF activation (RF-on/off) 2" instrument
switch pressed while the instrument is
connected to the HF device
HF device cannot find instrument
- or Instrument is defective and must be
replaced
Foot switch is pressed while the foot
switch is connected to the HF device
Release the "HF activation (RF on/off) instrument switch 2".
Remove instrument and reconnect it
Remove instrument and reconnect it
- or Replace instrument
Release foot switch
Remove foot switch and reconnect it
15
Aesculap®
Lektrafuse HF Generator GN200
7.4Failure messages
A failure is indicated as follows:
■ Clear sound sequence (as necessary, may interrupt the sound of the HF
discharge)
■ The red error light 6 on the front side of the HF device flashes
■ Failure report in the display 5 on the front side of the HF device
Failure codeCauseRemedy
F001Software error on the HF deviceSwitch off HF device, wait 5 s, switch HF device on
F002Dosage error of HF deviceSwitch off HF device, wait 5 s, switch HF device on
F003Hardware errorSwitch off HF device, wait 5 s, switch HF device on
F004Internal communication error of the HF deviceSwitch off HF device, wait 5 s, switch HF device on
Note
If the failure occurs during HF discharge, the vessel sealing may be incomplete or insufficient.
In order to ensure that sufficient hemostasis has been achieved, follow the
procedure for regrasp errors, see Regrasp error.
Note
If the failure continues after turning the device off and on again twice, contact your national B. Braun/Aesculap agency, see Technical Service.
F005Overheating of HF deviceSwitch off HF device and leave to cool
Check the air stream around the generator
F006Error on the RF on/off button 2 (activated when switching on) Switch off HF device, wait 5 s, release RF on/off button 2,
switch on HF device
F007Error with HF energy outputSwitch off HF device, wait 5 s, switch HF device on
F008Fault: HF outside the calibrationSwitch off HF device, wait 5 s, switch HF device on
F009Fault: Internal power supplySwitch off HF device, wait 5 s, switch HF device on
7.5Problem solving
FindingRemedy
HF device does not switch onEnsure that the power cord has been correctly plugged into the power supply 13 of the HF
device.
Ensure that the power cord is connected to a functioning power source.
Ensure that the power OFF/ONN switch 12 of the HF device is in the power ON position.
Check the fuse in the fuse box and replace if necessary, see Fuse replacement.
If the error continues, replace power cord.
The device is switched on, but no displays are illuminated and the self-test is not being carried out
Self-test failsSwitch off HF device, wait 5 s, and switch HF device on again.
Switch off HF device, wait 5 s, and switch HF device on again.
If the error continues, contact your national B. Braun/Aesculap agency, see Technical Service.
If the error continues, contact your national B. Braun/Aesculap agency, see Technical Service.
HF device is active ("power ON signal lamp 9" light
up), and accessories are connected, but there is not HF
discharge
16
Verify that the required foot switch (if used ) is properly connected and plugged in.
Ensure that the Caiman Seal & Cut Instrument is correctly connected and plugged in (green
ring 1 is lit up).
In order to start the HF discharge, activate the "RF on/off" button 2.
If the error continues, replace Lektrafuse instrument.
7.6User maintenance
► In case of error, follow the instructions in the display and note the error
code.
► If the error continues after following the instructions in the display,
contact your national B. Braun/Aesculap agency, see Technical Service.
7.7Fuse replacement
Risk of fatal injury from electric shock!
► Unplug the device before changing the fuses!
DANGER
Specified fuses: 2 units Aesculap art. no. TA021404:
G fuse link, time lag (T) 8 A, breaking capacity H (1 500 A)
► Use a small screwdriver to release the clip on the fuse holder 11 and
tilt up cover.
► Use a small screwdriver to release the fuse holder 11 and pull it out.
► Replace both fuses.
► Reinsert the fuse holder 11.
► Close the cover.
Fig. 12
Note
If the fuses burn out frequently, the device is faulty and should be
repaired, see Technical Service.
8.Technical Service
Risk of injury and/or malfunction!
► Do not modify the product.
WARNING
► For service and repairs, please contact your national B. Braun/Aesculap
agency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:+49 (7461) 95 -1601
Fax:+49 (7461) 14 -939
E-Mail:ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:+1 (800) 214 -3392
Fax:+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
17
Aesculap®
Lektrafuse HF Generator GN200
9.Accessories/Spare parts
Art. no.Designation
TA021404Fuse
PL720SUInstrument Caiman 5
PL730SUInstrument Caiman 12
PL731SUInstrument Caiman 12 plus
GN201Foot switch
Note
Other accessories, applied parts and replacement parts are described in
Aesculap brochure C-304-81 and C-902-02
10.Technical data
Classification acc. to Directive 93/42/EEC
Art.
no.
GN200High frequency surgical deviceIIb
Mains voltage ranges
(Current intake)
Frequency50–60 Hz
Protection class (acc. to
IEC / DIN EN 60601-1)
Foot switch circuitignition-safe acc. to IEC/DIN EN 60601,
Output power2 x 150 W on 50 Ohm
Device protection
Time-Current characteristic
Breaking capacity
Construction
DesignationClass
100V~–240 V~ (6.7 A–2.8 A)
I
approved for use in “medical environment”
T8 AH/250 V
T (Time lag)
H (1 500 A)
5 mm x 20 mm
HF operating frequency460 kHz ± 1 %, Quasi-Sinus
Operating modeInt 10 s/30 s
Weight8.1 kg
Dimensions (L x W x H)400 mm x 355 mm x 100 mm
Application part type acc.
to IEC/DIN 60601-1
Recovery time of the defibrillation-protected output
CISPR11Class A
EMCIEC/DIN EN 60601-1-2
Conforming to standardIEC/DIN EN 60601-1
CF
0 seconds
IEC/DIN EN 60601-2-2
18
10.1Ambient conditions
OperationStorage and transport
Temperature
-10-10
Relative
humidity
10
non-condensingnon-condensing
Atmospheric
pressure
Note
The atmospheric pressure of 700 hPa corresponds to a maximum operating
altitude of 3 000 m.
Note
After transporting or storing at temperature outside the operating temperature range, leave the HF device for one hour before use so that it can reach
room temperature.
50
90
12.Distributor in the US/Contact in Canada for product information and complaints
3773 Corporate Parkway
Center Valley, PA, 18034,
USA
11.Disposal
Note
The user institution is obliged to process the product before its disposal, see
Validated reprocessing procedure.
Adhere to national regulations when disposing of or recycling the product, its components and its packaging!
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions for
the product, as well as information for proper disposal of
components harmful to the environment.)
Products carrying this symbol are subject to separate collection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manufacturer as a free-of-charge service.
► Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical
Service.
4.Arbeiten mit dem Lektrafuse HF-Generator GN20026
4.1Bereitstellen26
4.2Funktionsprüfung28
4.3Bedienung28
5.Validiertes Aufbereitungsverfahren28
5.1Allgemeine Sicherheitshinweise28
5.2Allgemeine Hinweise29
5.3Vorbereitung am Gebrauchsort29
5.4Vorbereitung vor der Reinigung29
5.5Reinigung/Desinfektion29
5.6Wischdesinfektion bei elektrischen Geräten ohne
Sterilisation30
5.7Kontrolle, Wartung und Prüfung30
5.8Lagerung und Transport30
6.Instandhaltung30
6.1Sicherheitstechnische Kontrolle30
7.Fehler erkennen und beheben32
7.1Regrasp-Fehler32
7.2Warnungen33
7.3Störungsmeldungen33
7.4Fehlermeldungen34
7.5Problembehebung34
7.6Fehlerbehebung durch den Anwender35
7.7Sicherungswechsel35
8.Technischer Service35
9.Zubehör/Ersatzteile36
10.Technische Daten36
10.1 Umgebungsbedingungen37
11.Entsorgung37
Hinweis
Diese Gebrauchsanweisung beschreibt nur Aufbau, Funktion und Bedienung des Lektrafuse HF-Generators GN200 und ist nicht dazu geeignet,
einen Anfänger in die Hochfrequenz-Chirurgie einzuführen. Eine allgemeine Beschreibung zur Anwendung der Hochfrequenz-Chirurgie ist in den
OP-Manuals und anderer einschlägiger Literatur beschrieben.
Diese Anleitung und das beschriebene Gerät sind ausschließlich zur Verwendung durch qualifiziertes medizinisches Fachpersonal gedacht, das in
der speziellen Technik und dem durchzuführenden chirurgischen Verfahren
geschult wurde.
Verletzungsgefahr für den Patienten durch unsachgemäße Anwendung!
GEFAHR
► Produkt nur im Originalkarton transportieren.
► Vor der Anwendung des Produkts Funktionsfähigkeit und ordnungsge-
mäßen Zustand prüfen.
► „Hinweise zur elektromagnetischen Verträglichkeit (EMV)“ beachten,
siehe TA022130.
► Der Lektrafuse HF-Generator GN200 erfüllt die Anforderungen nach
CISPR 11 Klasse A.
► Um Schäden durch unsachgemäßen Aufbau oder Betrieb zu vermeiden
und die Gewährleistung und Haftung nicht zu gefährden:
– Produkt nur gemäß dieser Gebrauchsanweisung verwenden.
– Sicherheitsinformationen und Instandhaltungshinweise einhalten.
– Nur Aesculap-Produkte miteinander kombinieren.
– Anwendungshinweise gemäß Norm einhalten, siehe Sichere Hand-
habung gemäß IEC-/VDE-Bestimmungen.
► Zubehör regelmäßig prüfen: Insbesondere sollten Elektrodenkabel und
endoskopisches Zubehör auf mögliche Beschädigungen an der Isolation geprüft werden.
► Gebrauchsanweisung für den Anwender zugänglich aufbewahren.
► Gültige Normen einhalten.
► Produkt und Zubehör nur von Personen betrei-
ben lassen, die die erforderliche Ausbildung,
Fachkenntnis oder Erfahrung haben.
21
Aesculap®
Lektrafuse HF-Generator GN200
1.1Sichere Handhabung gemäß IEC-/VDE-Bestim-
mungen
Operationsumfeld
Lebensgefahr durch elektrischen Stromschlag!
► Produkt nicht öffnen.
GEFAHR
WARNUNG
VORSICHT
► Produkt nur an ein Versorgungsnetz mit Schutz-
leiter anschließen.
Verletzungsgefahr durch Zündung oder Explosion
von brennbaren Gasen! Beim bestimmungsgemäßen
Gebrauch des HF-Geräts können Funken entstehen.
► Nicht im explosionsgefährdeten Bereich einset-
zen.
► Bei Operationen im Bereich von Kopf und Tho-
rax den Einsatz von zündfähigen Anästhetika
und verbrennungsfördernden Gasen (z. B. Lachgas, Sauerstoff) vermeiden oder diese Stoffe
absaugen.
► Wenn möglich, nicht brennbare Stoffe zur Rei-
nigung und Desinfektion verwenden.
► Wenn brennbare Reinigungs-, Desinfektions-
und Lösungsmittel verwendet werden: Sicherstellen, dass diese Stoffe vor dem Einsatz der
HF-Chirurgie verdunstet sind.
► Sicherstellen, dass sich keine brennbaren Flüs-
sigkeiten unter dem Patienten oder in Körperhöhlen (z. B. Vagina) sammeln. Bevor das HFGerät eingesetzt wird, alle Flüssigkeiten abwischen.
► Sicherstellen, dass keine endogenen Gase vor-
handen sind, die sich entzünden können.
► Sicherstellen, dass mit Sauerstoff getränkte
Materialien (z. B. Watte, Mull) so weit vom HFUmfeld entfernt sind, dass sie sich nicht entzünden können.
Gefahr von Beeinträchtigungen anderer Geräte!
Beim bestimmungsgemäßen Gebrauch des HF-
Geräts entstehen elektromagnetische Störfelder.
► Sicherstellen, dass keine elektronischen Geräte
im Umfeld des HF-Geräts aufgestellt sind, die
durch elektromagnetische Störfelder beeinträchtigt werden können.
► Sicherstellen, dass das Gerät nicht in direkten Kontakt mit dem Patien-
ten bzw. in den Sterilbereich gelangt.
► Sicherstellen, dass der Anwender nicht zeitgleich in direkten Kontakt
mit dem Patienten und dem HF-Gerät gelangt.
Sicherheit für den Patienten
Lebensgefahr durch mangelnde Vorbereitung oder
Fehler am HF-Gerät!
GEFAHR
GEFAHR
GEFAHR
WARNUNG
WARNUNG
► Sicherstellen, dass das HF-Gerät fehlerfrei
arbeitet.
► Sicherstellen, dass keine leitfähigen Flüssigkei-
ten (z. B. Blut, Fruchtwasser) in Fußpedal oder
Handschalter eingedrungen sind.
► Sicherstellen, dass im Fußpedal- oder Hand-
schalter-Kabel kein Kurzschluss vorliegt.
Verbrennungsgefahr für den Patienten durch unbeabsichtigtes Aktivieren des HF-Geräts!
► HF-Gerät nach unbeabsichtigtem Aktivieren
sofort am Schalter Netz-AUS/EIN ausschalten.
► Fußpedal oder Handschalter immer mit beson-
ders hoher Aufmerksamkeit bedienen.
Verletzungsgefahr für den Patienten durch unerwünschten Anstieg der HF-Ausgangsleistung bei
Versagen des HF-Geräts!
► Gerät bei kleinsten Unregelmäßigkeiten nicht
mehr verwenden.
Verletzungsgefahr für den Patienten/Anwender
durch defekte Netzzuleitung/nicht vorhandene
Schutzleiterverbindung!
► Netzzuleitung/Schutzleiterverbindungen prüfen.
Gefahr von Verletzungen durch Muskelkontraktion
verursacht durch Stimulation der Nerven und Muskeln!
► An empfindlichen Strukturen mit besonderer
Aufmerksamkeit arbeiten.
VORSICHT
22
Sichtbehinderung und/oder Nebenwirkungen durch
Dampf-/Rauchentwicklung bei HF-Chirurgie!
► Ggf. Rauchabsaugung verwenden.
Folgendes immer sicherstellen:
► Patienten so lagern, dass er keine Metallteile berührt, die geerdet sind
oder beträchtliche Kapazität gegen Erde haben (z. B. Operationstisch,
Halterungen). Bei Bedarf antistatische Tücher dazwischen legen.
► Sicherstellen, dass der Patient keine feuchten Tücher oder Unterlagen
berührt.
► Bereiche mit starker Schweißabsonderung und Haut-an-Haut-Berüh-
rungen am Körperstamm durch Zwischenlegen von antistatischen
Tüchern schützen.
► Urin über Katheter ableiten.
► Bei Operationen am Herzen HF-Gerät über Potenzialausgleich erden.
► Bei Patienten mit Herzschrittmachern oder anderen aktiven Implanta-
ten vor der Anwendung der HF-Chirurgie den entsprechenden Facharzt
konsultieren, um irreparable Schäden oder Fehlfunktionen am Schrittmacher oder dem Implantat zu vermeiden.
► Elektroden von physiologischen Überwachungsgeräten ohne Schutz-
widerstände oder HF-Drosseln möglichst weit entfernt von den HFElektroden anbringen.
► Keine Nadelelektroden für die Überwachung verwenden.
► Leitungen von Überwachungsgeräten so anbringen, dass sie nicht auf
der Haut liegen.
► Leitungen zu den HF-Elektroden möglichst kurz halten und so führen,
dass sie weder den Patienten noch andere Leitungen berühren.
► Bei unzureichender Leistung mit den gewohnten Einstellungen sicher-
stellen, dass:
– die Arbeitselektroden sauber sind,
– die Steckverbindungen korrekt sind.
► Aktive Elektrode nie auf oder neben dem Patienten ablegen.
► Zeitweise nicht benötigte aktive Elektroden so ablegen, dass sie den
Patienten nicht berühren.
► Bei Operationen, bei denen sich ein ständiger Kontakt der Elektroden
zum Patienten nicht vermeiden lässt (z. B. bei endoskopischen Operationen), HF-Gerät bei unbeabsichtigtem Aktivieren der Elektrode sofort
am Schalter Netz-AUS/EIN 12 ausschalten.
► Heiße Elektrode nicht unmittelbar nach dem Schneiden oder Koagulie-
ren aus dem Körper entfernen.
► Während des Betriebs des Generators die Elektroden am Instrument
(auf der oberen oder unteren Seite) nicht berühren, da die Elektroden
sich erwärmen können und dies zu Verbrennungen oder Verletzungen
führen kann.
2.Gerätebeschreibung
2.1Lieferumfang
BezeichnungArt.-Nr.
Hochfrequenz-ChirurgiegerätGN200
GebrauchsanweisungTA022377
Hinweise zur elektromagnetischen
Verträglichkeit
2.2Zum Betrieb erforderliche Komponenten
■ Netzkabel
■ Fußpedal (optional)
■ Caiman Instrument
2.3Verwendungszweck
Der Lektrafuse HF-Generator GN200 wird zur Gefäßversiegelung und
Gefäßtrennung in der offenen und minimal-invasiven Chirurgie eingesetzt. Die Instrumente können Gefäße bis einschließlich 7 mm versiegeln.
Der Lektrafuse HF-Generator ist nicht für den Einsatz bei Tubensterilisation/Tubenkoagulation zur Sterilisation geeignet.
Der Lektrafuse HF-Generator ist zur Anwendung am Herzen zugelassen
(Typ CF).
Der Lektrafuse HF-Generator ist für den Betrieb und die Lagerung in
geschlossenen Räumen bestimmt.
2.4Funktionsweise
Der Lektrafuse HF-Generator GN200 arbeitet mikroprozessorgesteuert
und wandelt die Netzspannung in einen hochfrequenten Wechselstrom
für die bipolare Gefäßversiegelung um.
Der Versiegelungsprozess erfolgt über einen geschlossenen Regelkreis. Die
Versiegelung kann über die Taste am Instrument, den Taster an der
Gerätefront oder ein Fußpedal gestartet und gestoppt werden.
TA022130
23
Aesculap®
Lektrafuse HF-Generator GN200
2.5Signaltöne
Zustand/StatusSignaltonBemerkung
System Fehlermeldung3 Wiederholungen von: 1 langer Piepton 3 KHz,
System Störungsmeldung3 Pieptöne, 3 KHzNach jeder Störungsmeldung, wenn erkannt (E001,..)
System Warnung3 Pieptöne, 3 KHzNach jeder Warnung, wenn erkannt (W001,...)
HF-Abgabe Start2 Pieptöne 480 HzBeim Start der HF-Abgabe
HF-Abgabe AktiveWiederholender Piepton 480 HzKontinuierlich während der HF-Abgabe
HF-Abgabe beendet3 Pieptöne 530 HzBei erfolgreicher Beendigung eines Versiegelungsprozesses
Power Up Selbsttest1 Piepton 2 KHzBeim Selbsttest, um die Funktionalität des Signaltons zu über-
Nach jedem System-Fehler, wenn erkannt (F001,...)
Nach jedem Regrasp-Fehler, wenn erkannt
(keine Störungsmeldung oder Fehler)
prüfen
2.6Überwachungsfunktionen
Selbsttest
Beim Einschalten führt das Gerät einen Selbsttest durch, der Bedienelemente, Signalton, Mikroprozessor und Hardwarefunktion prüft. Während
dieser Phase erscheint in Anzeige 5 die Meldung "Performing Self Test".
Abb. 1
Permanenter Test während des Betriebs
Während des Betriebs werden sicherheitsrelevante Funktionen und Signale zyklisch getestet. Wenn kritische Fehler erkannt werden, beendet das
HF-Gerät die HF-Aktivierung. In Anzeige 5 wird die Fehlernummer des
entsprechenden Fehlers angezeigt und ein Signalton ausgegeben, siehe
Fehler erkennen und beheben.
24
2.7Diagramme Ausgangsleistung
Darstellung der Ausgangsleistung als Funktion des Lastwiderstands.
3.Vorbereiten und Aufstellen
Wenn die folgenden Vorschriften nicht beachtet werden, übernimmt
Aesculap insoweit keinerlei Verantwortung.
► Beim Aufstellen und Betrieb des Produkts einhalten:
– die nationalen Installations- und Betreiber-Vorschriften,
– die nationalen Vorschriften über Brand- und Explosionsschutz.
Hinweis
Die Sicherheit des Anwenders und des Patienten hängt u. a. von einer intakten Netzzuleitung, insbesondere von einer intakten Schutzleiterverbindung
ab. Defekte oder nicht vorhandene Schutzleiterverbindungen werden häufig nicht sofort erkannt.
► Gerät über den an der Geräterückwand montierten Anschluss für
Potenzialausgleichsleiter mit dem Potenzialausgleich des medizinisch
genutzten Raums verbinden.
Abb. 2 Diagramm Ausgangsleistung Caiman 12 mm Seal & Cut Instru-
mentenfamilie
Abb. 3 Diagramm Ausgangsleistung Caiman 5 mm Seal & Cut Instru-
mentenfamilie
2.8Maximale Ausgangsspitzenspannung (Up)
Hinweis
Die folgenden Angaben ermöglichen dem Anwender zu beurteilen, ob das
HF-Gerät GN200 für bestimmtes Zubehör (Qualität der Isolierung) geeignet ist.
Hinweis
Die Potenzialausgleichsleitung ist unter der Artikel-Nr. GK535 (4 m lang)
bzw. TA008205 (0,8 m lang) beim Hersteller erhältlich.
► Um eine ausreichende Belüftung und Kühlung des HF-Geräts sicherzu-
stellen, rund um das HF-Gerät mindestens 10 cm bis 15 cm Raum lassen.
3.1Erstinbetriebnahme
Verletzungsgefahr und/oder Fehlfunktion des Produkts durch Fehlbedienung des medizinisch-elekt-
WARNUNG
rischen Systems!
► Gebrauchsanweisungen aller medizinischen
Geräte einhalten.
Verletzungsgefahr für den Patienten/Anwender
durch Verwendung von unzureichendem Zubehör
WARNUNG
Maximale Ausgangsspitzenspannung (U
(Qualität der Isolierung)!
► Sicherstellen, dass die in den Begleitpapieren
des Zubehörs angegebene Bemessungszubehörspannung größer ist als die maximale Ausgangsspitzenspannung.
): 200 V
p
25
Aesculap®
Lektrafuse HF-Generator GN200
4.Arbeiten mit dem Lektrafuse HF-Gene-
rator GN200
4.1Bereitstellen
Zubehör anschließen
Verletzungsgefahr durch unzulässige Konfiguration
bei Verwendung weiterer Komponenten!
GEFAHR
Zubehörkombinationen, die nicht in der Gebrauchsanweisung erwähnt
sind, dürfen nur verwendet werden, wenn sie ausdrücklich für die vorgesehene Anwendung bestimmt sind. Leistungsmerkmale sowie Sicherheitsanforderungen dürfen nicht nachteilig beeinflusst werden.
Nur CAIMAN Instrumente können in Verbindung mit dem Lektrafuse HFGenerator GN200 eingesetzt werden, siehe Zubehör/Ersatzteile.
Spannungsversorgung anschließen
GEFAHR
► Sicherstellen, dass bei allen verwendeten Kom-
ponenten die Klassifikation mit der Klassifikation des Anwendungsteils (z. B. Typ BF oder
Typ CF) des eingesetzten Geräts übereinstimmt.
Lebensgefahr durch elektrischen Stromschlag!
► Produkt nur an ein Versorgungsnetz mit Schutz-
leiter anschließen.
► Gerät so aufstellen, dass das Ausschalten über
Schalter Netz-AUS/EIN problemlos möglich ist.
► Gerät so aufstellen, dass eine Trennung vom
Netzkabel problemlos möglich ist.
Gerät einschalten
► Gerät mit Schalter Netz-AUS/EIN 12 einschalten.
Die Signallampe Netz-EIN 9 leuchtet. Das Gerät führt einen Selbsttest
durch, siehe Permanenter Test während des Betriebs:
Ist kein Instrument angeschlossen, erscheint nach dem Selbsttest in
Anzeige 5 die Meldung "Attach Instrument". Scheitert der Selbsttest,
wird eine Fehlermeldung auf Anzeige 5 ausgegeben.
Abb. 4
Fußpedal anschließen
Der Anschluss des Fußpedals GN201 ist optional.
Am Fußpedalstecker und an der Anschlussbuchse Fußpedal 3 ist jeweils
ein Punkt als Markierung angezeichnet. Um den Stecker korrekt in die
Anschlussbuchse einstecken zu können, müssen diese Markierungen übereinander stehen.
Die Netzspannung muss mit der Spannung übereinstimmen, die auf dem
Typschild des Geräts angegeben ist.
Das Gerät ist mit einem Weitbereichsnetzteil ausgerüstet, so dass die
Netzspannung zwischen 100 V und 240 V liegen kann, ohne dass der
Netzbereich umgeschaltet werden muss.
► Sicherstellen, dass das Gerät ausgeschaltet ist. Gegebenenfalls mit
Schalter Netz-AUS/EIN 12 ausschalten.
► Netzkabel an der Rückseite des Geräts in Netzeingangsbuchse 13 ein-
stecken.
► Netzstecker in die Steckdose der Hausinstallation stecken.
26
Abb. 5
► Fußpedalstecker ausrichten und bis zum Einrasten in Anschlussbuchse
Fußpedal 3 einstecken.
Bei erfolgreicher Verbindung zwischen Gerät und Fußpedal erscheint in
Anzeige 5 für 3 Sekunden die Meldung "Footswitch attached".
Abb. 6
Hinweis
Fußpedal während der HF-Abgabe nicht kontinuierlich betätigen. Die HFAbgabe wird gestartet und beendet durch einmaliges Drücken und Loslassen des Fußpedals.
Fußpedal entfernen
► Rast-Ring gegen Uhrzeigersinn drehen und gleichzeitig Anschlussste-
cker abziehen
In Anzeige 5 erscheint für 3 Sekunden die Meldung "Footswitch remo-
ved".
Instrument anschließen
Ausschließlich Instrumente der CAIMAN Seal & Cut Produktfamilie können in Verbindung mit dem Lektrafuse HF-Generator GN200 eingesetzt
werden.
Am Instrumentenstecker ist ein Pfeil und an der Anschlussbuchse 8 ein
Punkt als Markierung angezeichnet. Um den Stecker korrekt in die
Anschlussbuchse stecken zu können, müssen diese Markierungen übereinander stehen.
Abb. 7
Abb. 8
► Instrument in Anschlussbuchse 8 stecken.
Der grüne Signalring 7 um den Instrumentenanschluss leuchtet auf. In
Anzeige 5 erscheint die Meldung "Ready to Seal".
Abb. 9
27
Aesculap®
Lektrafuse HF-Generator GN200
Außerbetriebnahme
► Gerät mit Schalter Netz-AUS/EIN 12 ausschalten.
Alle Pole des Geräts werden vom Versorgungsnetz getrennt.
► Netzkabel abziehen
Das Gerät ist vollständig vom Versorgungsnetz getrennt.
4.2Funktionsprüfung
Korrekte Funktion des Geräts vor jedem Einsatz prüfen:
► Sicherstellen, dass das Zubehör keine sichtbaren Schäden hat.
► Gerät vorbereiten und aufstellen, siehe Vorbereiten und Aufstellen.
► Funktion folgender Elemente in der vorgegebenen Reihenfolge prüfen:
– Automatischer Selbsttest nach jedem Einschalten: kurzes akusti-
sches Signal, alle Anzeige-Elemente leuchten kurz auf
– Fußpedal, siehe Fußpedal anschließen
– Instrument, siehe Instrument anschließen
– Aktivierung mit RF-on/off 2 an Gerätefront, Taster an Instrument
und Fußpedal
– Schalter Netz-AUS/EIN 12 ausschalten
4.3Bedienung
Gerät ein-/ausschalten
► Gerät mit Schalter Netz-AUS/EIN 12 einschalten/ausschalten.
Betriebsart
Der Lektrafuse HF-Generator GN200 erkennt automatisch das angeschlossene Instrument. Interne Geräteeinstellungen werden geladen. Das Einstellen der Betriebsart oder der Parameter durch den Benutzer ist nicht
erforderlich/möglich.
Hochfrequenz aktivieren
► Sicherstellen, dass der Patient so vorbereitet ist, dass gefahrlos mit
dem HF-Gerät gearbeitet werden kann.
► Sicherstellen, dass Instrument und Fußpedal, falls verwendet, korrekt
angeschlossen sind.
► HF-Gerät über Handschalter, Fußpedal oder Taster HF-Aktivierung (RF-
on/off) 2 an der Gerätefront aktivieren.
Hinweis
Fußpedal oder Taster während der HF-Abgabe nicht kontinuierlich betätigen. Der Gefäßversiegelungsprozess wird gestartet und beendet durch einmaliges Drücken des Fußpedals, des Tasters HF-Aktivierung (RF-on/off) 2
an der Gerätefront oder des Tasters am Instrument. Nach Beendigung des
Versiegelungsprozesses stoppt das Gerät die HF-Abgabe automatisch.
Durch wiederholten Tastendruck kann der Versieglungsprozess abgebrochen werden.
Verletzungsgefahr und/oder Fehlfunktion!
► Vor jedem Gebrauch Funktionsprüfung durch-
WARNUNG
WARNUNG
► Start des Versiegelungsprozesses (HF-Abgabe) durch einmaliges Drü-
cken des Tasters am Instrument, des Fußpedals oder des Tasters HFAktivierung (RF-on/off) 2 an der Gerätefront.
Nach Beendigung des Versiegelungsprozesses wird die HF-Abgabe
automatisch gestoppt.
Während der HF-Versiegelung erscheint in Anzeige 5 die Meldung
"Sealing in Progress".
Abb. 10
führen.
Verletzungsgefahr durch Verwendung des Produkts
außerhalb des Sichtbereichs!
► Produkt nur unter visueller Kontrolle anwenden.
5.Validiertes Aufbereitungsverfahren
5.1Allgemeine Sicherheitshinweise
Hinweis
Nationale gesetzliche Vorschriften, nationale und internationale Normen
und Richtlinien und die eigenen Hygienevorschriften zur Aufbereitung einhalten.
Hinweis
Bei Patienten mit Creutzfeldt-Jakob-Krankheit (CJK), CJK-Verdacht oder
möglichen Varianten bezüglich der Aufbereitung der Produkte die jeweils
gültigen nationalen Verordnungen einhalten.
Hinweis
Die maschinelle Aufbereitung ist aufgrund eines besseren und sichereren
Reinigungsergebnisses gegenüber der manuellen Reinigung vorzuziehen.
Hinweis
Es ist zu beachten, dass die erfolgreiche Aufbereitung dieses Medizinprodukts nur nach vorheriger Validierung des Aufbereitungsprozesses sichergestellt werden kann. Die Verantwortung hierfür trägt der Betreiber/Aufbereiter.
Zur Validierung wurde die empfohlene Chemie verwendet.