B. Braun Aesculap DSM Instructions For Use/technical Description

Aesculap®
Aesculap Power Systems
USA Instructions for use/Technical description
DSM (Digital Surgical Microscope)
1
2
3
4
5
6
Aesculap®
DSM (Digital Surgical Microscope)
Note
CAUTION: RX only!
Legend
1 3D flat panel display 2 Touchscreen 3 Robotic arm 4 DSM head 5 Medical grade cart 6 Footswitch
The illustrations in these instructions for use (especially illustrations of the software) may deviate from the actual device.
Contents
1. General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1 Symbols on product and packaging . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Applicable to . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4 Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.5 Indications for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.6 Contraindications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. Safe handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1 Light emission risks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Fluorescent surgery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Place of use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.4 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.5 DSM sterile drape PV012SU . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.6 Footswitch PV014 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3. Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.1 Robotic arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.2 DSM head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2.1 LED indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2.2 Handle buttons (default setting) . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3 DSM cart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3.1 Emergency stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3.2 Cart LED indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3.3 Cart connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.4 3D display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.5 3D display stand (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.6 Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.7 Footswitch PV014 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.7.1 Default footswitch button functionality . . . . . . . . . . . . . . . . . . 8
3.7.2 Footswitch status LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.7.3 Backup USB cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.8 Isolation transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.9 Wireless keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.10 Polarized 3D glasses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.11 Additional components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4. Preparation and setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5. Working with the product. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.1 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.2 System set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.2.1 3D display and robotic arm workspace . . . . . . . . . . . . . . . . . . . 9
5.2.2 Connecting the accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.2.3 Connecting the power supply . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.3 Function checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.4 Positioning the 3D display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.5 Turning the system on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.6 Setting up the robotic arm and the DSM head . . . . . . . . . . . . 11
5.6.1 Rotating the DSM using the yaw coupler. . . . . . . . . . . . . . . . . 11
5.6.2 Moving the robotic arm out of storage position and draping the
robotic arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.7 Starting a case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.8 Ending a case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.9 Saving case media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.9.1 Save to default location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.9.2 Copy media to drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.10 Removing the drape . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.11 Shutdown procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.11.1 Software shutdown and robotic arm storage. . . . . . . . . . . . . . 13
5.12 Changing footswitch batteries . . . . . . . . . . . . . . . . . . . . . . . . . 13
6. Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.1 Home screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.2 DSM setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.2.1 Preset positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.2.2 Rebalance scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.2.3 Robot recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.3 Live screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.3.1 Zoom, focus, and light controls . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.4 Fluorescence controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.4.1 Function checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.4.2 DUV 400 imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.5 Touchscreen robot control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.5.1 Lock On Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.5.2 Robot Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.6 Quick Access Bar (QAB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.7 Robotic arm state indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.8 Software settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.8.1 Camera settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.8.2 Screen Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.8.3 Digital Aperture. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.8.4 DUV 400 (Fluorescence) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.8.5 Color. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.8.6 Recording. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6.8.7 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6.8.8 Input Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.8.9 DICOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.8.10 Handles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.8.11 Footswitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
6.8.12 About . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7. Reprocessing procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.1 Single-use products . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2
7.2 Preparation before cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.3 Limitations on reuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.4 Cleaning/disinfection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.4.1 Product-specific safety instructions for the reprocessing
procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.5 Wipe disinfection for electrical devices . . . . . . . . . . . . . . . . . . . 21
7.6 Inspection, maintenance and checks . . . . . . . . . . . . . . . . . . . . . 21
7.7 Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
7.8 Storage and transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
8. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
9. Troubleshooting list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.1 Malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.2 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.2.1 Startup errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.2.2 License error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.2.3 Recording errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.2.4 DSM head errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
10. Technical Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
11. Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
11.1 Performance data, information about standards . . . . . . . . . . . 23
11.2 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
12. Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
13. Distributor in the US/Contact in Canada for product
information and complaints . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Symbol Location Description
Product label Date of manufacture
Product label CSA NRTL Mark
Product label CE marking acc. to DIR 93/42/EEC
Touchscreen mount label
Coupler label Yaw adjustment slider
Cart potential equalization conductor
Product label Marking of electric and electronic
Product label Catalog number
Product label Serial number
USB 3.0 port
Connection point for potential equal­ization conductor
devices according to directive 2002/96/EC (WEEE), see Chapter 12.
1. General information
1.1 Symbols on product and packaging
Symbol Location Description
Base of cart Main power switch.
Press to turn the system ON. Press again to turn the system OFF
Product label AC input
DSM head DC input
Product label Warning
Product label Risk of electric shock! Do not open.
Robotic arm Take care to avoid injury when in the
vicinity of the display arm or robotic arm.
Support arms Do not exceed load ratings for support
arms.
Product label Follow instructions for use.
Product label Revision
1.2 Applicable to
These instructions for use are applicable to the following products:
Art. no. Designation
PV010 Aesculap DSM (Digital Surgical Microscope)
PV022 Software module DUV 400
PV024 Software module DICOM
PV030 White balance card
1.3 Warnings
Warnings make clear the dangers to patient, user and/or product that could arise during the use of the product. Warnings are labeled as follows:
Danger of death or serious injury.
DANGER
Danger of minor injury or product damage.
WARNING
Cart emer­gency stop
Product label System mass
Product label Manufacturer
Emergency stop control device
Indicates a possible threat of material damage. If not avoided, the product may be damaged.
CAUTION
1.4 Intended use
This device is for use with patients undergoing microsurgery within its indications for use.
There is no patient contact intended with this device.
3
1.5 Indications for Use
Aesculap DSM (Digital Surgical Microscope) PV010
The Aesculap DSM (Digital Surgical Microscope) generates a magnified 3D view of the surgical field.
Software module DUV 400 PV022
The DUV 400 is an accessory for the AESCULAP® DSM and is used for view­ing fluorescence of fluorophores, comprising:
An excitation filter for blue spectral range between 390 nm and
420 nm
An observation filter for visible light with spectral range greater than
510 nm
1.6 Contraindications
The DSM system must not be used for ophthalmology.
2. Safe handling
CAUTION Federal law restricts this device to purchase by, or on instruction by a
physician!
Risk of fatal injury from electric shock!
Do not open the product.
DANGER
Manually clean the new product after removing its transport packaging
and prior to first use.
Prior to use, check that the product is in good working order.Observe “Notes on Electromagnetic Compatibility (EMC)“, see
TA015615-EMV.
To prevent damage caused by improper setup or operation, and in order
not to compromise warranty and manufacturer liability: – Use the product only according to these instructions for use. – Follow the safety and maintenance instructions. – Only combine Aesculap products with each other.
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
Keep the instructions for use accessible for the user.Always adhere to applicable standards.Set up and connect any external video input or output before the sys-
tem is turned on.
Ensure that exposed cables for the footswitch, head, any external dis-
plays, and power supply are laid flat on the ground and out of high­traffic areas to minimize potential tripping hazards.
Do not use the robotic arm as an armrest or to support any unapproved
accessories. Do not place other equipment on the system.
Do not place objects around or on top of the E-stop or on the DSM head
and arm.
Ensure the E-stop is not locked prior to surgery.Use the protective stop on the touchscreen if there is unintended robot
arm movement.
Use the E-Stop only if the protective stop malfunctions.Check fluorescence functionality prior to surgery using the provided
reference card.
Ensure that the DSM head working distance is between 20 cm and
45 cm.
The fluorescence reference card and the touchscreen are not sterile.
Perform fluorescence functional tests before surgery.
Take precautions to avoid touching any system components and the
patient at the same time.
Connect the product only to a grounded power
supply.
Do not push or pull the product by the display, head, or power cable.To ensure optimal image quality, only use product with 3D glasses sup-
plied by Aesculap. Contact Aesculap if additional pairs are required.
Do not use polarized prescription glasses as they will distort the 3D
effect.
Always return the robot to storage position before shutdown.Store the head with the lens cap covering the main objective when it
is not in use.
Do not cover the air vents of the system, touchscreen, or 3D display.
Note
When the emergency stop is pressed, the DSM head may move up to 3 cm (depending on position) due to mechanical brakes locking into place.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interfer­ence when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his or her own expense. You are cautioned that any changes or modifications not expressly approved in this manual could void your authority to operate this equipment.
All interface cables used to connect peripherals must be shielded in order to comply with the limits for a digital device pursuant to Subpart B of Part 15 of FCC Rules. This device complies with Part 15 of the FCC Rules. Oper­ation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interfer­ence received, including interference that may cause undesired operation.
Mode of operation: Continuous
Protection against electric shock: Class I
2.1 Light emission risks
Retinal blue light and Near-UV emission risk.
Do not stare at the lamp emission area during operation.Minimize exposure to eyes and skin.Use appropriate shielding.Make sure that no light from the DSM enters the patient’s eyes.
2.2 Fluorescent surgery
Only use fluorescent agents that are approved for the planned applica-
tion.
Danger of injury to the eyes due to possibly hazardous UV light. Do not
look at the DSM illumination, minimize exposure to eyes or skin, and use appropriate shielding.
Use the lowest comfortable light intensity.Ensure that no tissue damage is caused by excessive illumination inten-
sity.
The room lighting impairs the visualization of fluorescence. For surger-
ies using the DSM fluorescence modules, operate in a darkened room, if possible.
Always perform a functional test before using the fluorescence mod-
ules, see Chapter 6.4.1. Use the reference cards to check whether the fluorescence medium can be excited for the DSM wavelength range and whether it emits fluorescent light of sufficient intensity.
Note
As in almost all diagnostic procedures, false-positive and false-negative results can occur in the fluorescence-based digital overlay. Evaluation by the user based on other methods may be necessary.
4
2.3 Place of use
This system is not suitable for use in the presence of flammable anesthet­ics mixture with air or with oxygen.
Do not use system in explosive atmospheres.To prevent fire or shock hazard, do not expose system to rain or mois-
ture.
2.4 Ambient conditions
The following ambient conditions apply to the use of the product:
Temperature
Relative humidity
Atmospheric pressure
10 °C to 35 °C
30 % to 90 %; non-condensing
700hPa to 1060hPa
2.5 DSM sterile drape PV012SU
Infection hazard for patients and/or users and impairment of product functionality due to reuse.
WARNING
The product has been EO-sterilized and is supplied in sterile packaging. The product must not be reused. The product is the only sterile component of the system.
Place a drape prior to surgery. Do not use products from open or damaged sterile packaging.Do not use the product if it is damaged or defective.In case of intraoperative damage of the drape: remove the system from
the operating table and replace the drape.
Do not use the product after its use-by date.
Risk of injury, illness or death due to contamination and/or impaired functionality of the product!
Do not reprocess the product.
Note
Wireless mode is not possible when the foot switch is connected via USB cable.
3. Product description
The Aesculap DSM is a freestanding, digital surgical microscope in a look­over configuration. The DSM does not utilize traditional microscope bin­oculars within the field of use and is intended to be an alternative to tra­ditional optical microscopes. The DSM software is a component of the medical device.
The main components of the DSM system are:
3D flat panel display 1 – live surgical view
Touchscreen 2 – control interface
Robotic arm 3
DSM head 4 – 3D camera, illumination, and handle controls
Medical-grade cart 5 with embedded computers
Footswitch 6
Visualization software
Sterile drape
3.1 Robotic arm
The robotic arm is a six-jointed arm that supports the DSM head. Its touch-guided movement allows the user to manipulate it through the workspace. The robotic arm may be controlled via the handles, footswitch, or touchscreen. On the 3D display, the robot state indicator shows the cur­rent state of the robotic arm. The robotic arm workspace is illustrated below.
2.6 Footswitch PV014
Keep footswitch within a range of max. 10 m to the receiver.Do not place the footswitch in or near strong magnetic fields.Do not use more than four wireless systems in the same area.
The footswitch is regularly powered by three alkaline batteries of type C “Baby” (IEC-LR 14), which are placed behind the backside battery cover.
If they need changing, the terminal must be switched off first in order
to rule out any unintended triggering of functionality.
Only use footswitch with the specified batteries or backup cable.To prevent the footswitch from battery leakage, remove battery from
the case during long periods of non-use.
Never touch the battery contacts and the patient at the same time.
The maximum mechanical load which the operating elements may be sub­jected to is max. 1350 N.
The integrated accelerometer prevents an accidental actuation of switch­ing element if the inclination passes 35°.
Only use footswitch on a firm, even (horizontal) surface.
When using the backup cable:
When connecting the backup cable, always ensure that connecting ele-
ments (plugs) are clean.
Ensure that the plug connections are not subjected to mechanical
loads.
Do not subject backup cable to excessive pulling, pressure or bending.
Do not drive over cable.
Always perform function test in combination with the DSM when the
device is switched back on.
390 mm
425 mm
a = 1,24 m
Fig.1
Legend a Max. radius
5
3.2 DSM head
The DSM head contains two sets of optics and two high-resolution sensors to convert the analog signal to a digital video stream.
b = 200 mm
c = 450 mm
Fig.2
Legend a Bottom of DSM head b Min. working distance c Max. working distance
3.2.1 LED indicator
The DSM head has an LED indicator on the back panel. The indicator colors are explained below.
Indicator Description
3.2.2 Handle buttons (default setting)
The DSM is designed to be controlled primarily through the buttons on the handles of the sides of the head. The default layout of the handles is shown below. In the software Handles settings, the handle button commands can be customized.
a
b c
d
a
e
f
g
h
i
lml
j
n
k
L
Fig.4
Legend
a
Assisted drive
b
Zoom +
c
White light +
d
Focus -
e
Auto focus
f
White light -
g
Zoom -
h
Assisted drive
i
Record
j
Focus +
k
Snapshot
l
Move up
m
Move left
n
Move down
o
Assisted drive
p
Move forward
q
Way point save/delete
r
Move right
s
Lock-on mode
t
Waypoint select/move
u
Move backward
v
Assisted drive
o
p q
r s
t u
v
R
Blinking blue The DSM head is starting up
Solid green The DSM head is ready for use
Solid yellow A minor error has occurred. See software mes-
sage for details.
Solid red A major error has occurred. User should switch
to a backup system immediately
Fig.3
3.3 DSM cart
The images acquired by the DSM head are delivered to the system com­puter (EPU) in the cart. This is a special-purpose computer, optimized to process the images for storage, manipulation, and stereoscopic display.
The signals generated for display employ the Digital Video Interface (DVI), delivering high resolution and high refresh rate stereoscopic display images to the display devices. The DVI signals can be delivered to display devices with DVI or HDMI cable connections.
The system computer is also fitted with a 2 TB storage drive or larger that stores at least 100 hours of 3D HD videos. The videos can be immediately accessed for playback due to real-time compression of the video stream as it is recorded. The compressed video files can easily be transferred to exter­nal storage for backup and archival and to free space on the internal stor­age drive.
The cart also holds a second computer controlling the robotic arm.
3.3.1 Emergency stop
The robotic arm emergency stop is located at the top of the cart. The robotic arm movement is always under the control of the operator.
If there is a malfunction and the robotic arm moves in an unintended
way, first attempt to press the Protective Stop button on the touch­screen.
If the touchscreen does not respond, use the emergency stop to remove
power from the arm and cease all movement.
Note
When the emergency stop is pressed, the DSM head may move up to 3 cm (depending on position) due to mechanical brakes locking into place.
6
3.3.2 Cart LED indicators
There are three LED indicators near the base of the cart that display the status of the DSM, system computer (EPU), and robotic arm components. The indicator colors are explained below.
Indicator Description
Solid green The component is ready for use.
Solid yellow A minor error has occurred.
Solid red Warning – the component is not ready for use.
3.3.3 Cart connections
a
Connect the display to the HDMI output port c at the back of the cart,
see Fig.5.
Configure the additional displays in the DSM software “Display” set-
tings panel, see Chapter 6.8.7.
The individual display settings are specially configured at the factory. When the PV644 or PV015 display is used in a high-temperature environ­ment, the brightness of the display backlight is reduced to lower the tem­perature inside the unit. When this function is activated, the power indi­cator flashes in amber. It is not recommended to drape the displays.
Connecting a video stream from another device (input)
Video streams from other devices may be connected to facilitate display in conjunction with the 3D image from the DSM head.
The following devices are approved for providing video input:
Art. no. Designation
PV460 Full HD 3 chip camera control unit
PV470 Full HD CMOS camera control unit
PV480 2D camera platform
Connect the device to the HDMI input port g or SDI input port h at the
front of the cart, see Fig.5.
Configure the video input in the DSM software “Input Source” settings
panel, see Chapter 6.8.8.
b c d e f g h i j
Fig.5
On the touchscreen arm, there are the following connections:
2 × USB 3.0 ports a
At the back of the cart, there are the following connections:
2 × Ethernet ports b
2 × HDMI output ports c
2 × USB 2.0 ports d
At the front of the cart, there are the following connections:
2 × Ethernet ports e
1 × HDMI output port f
1 × HDMI input port g
1 × SDI input port h
Main power cord connection i
6 mm potential equalization port j
Connecting a secondary display (output)
The system can support simultaneously displaying the live view on multi­ple displays.
The following 3D displays are approved for use with the DSM:
Art. no. Designation
PV008 26” display
PV646 32” display
Connecting to a potential equalization system
Directly above the main power cord connection i, the DSM system is equipped with a 6 mm potential equalization port j that can be connected to the healthcare facility potential equalization system.
Connect the display to the potential equalization port j at the front of
the cart, see Fig.5.
Connecting to a PACS/RIS server
There are two Ethernet ports on the both the front and back panels of the cart.
Ensure that the Ethernet cable and connector comply with at least Cat-
5e EIA/TIA-568A-5, preferably Class D values from ISO/IEC 11801:2002 or EN 50173-I:2002.
Connect an Ethernet cable to one of the Ethernet ports b or e to link
the system with the facility PACS server, see Fig.5.
Configure the connection in the software DICOM settings. See
TA015615-DCS for the complete DICOM Conformance Statement.
3.4 3D display
The default 26” 3D display has a circularly-polarized, micro-polarizing fil­ter that displays the right and left image alternately. Circularly-polarized glasses are worn to ensure that the left eye only sees the left image and the right eye only sees the right image. The 3D display has a refresh rate of 60 frames per second (FPS) simultaneously for each eye. The display comes optimally pre-configured from the factory.
If the display settings are causing inadequate visualization, please con-
tact customer support for assistance.
3.5 3D display stand (optional)
The system can support remote 3D displays on a display stand approved by Aesculap. If the default 26” 3D display does not come with the system, an optional display can be the primary display.
The following display stands are approved for use with the DSM:
Art. no. Designation
PV644 31” 4K display
PV015 55” 4K display
PV016 Monitor stand for 55" monitor
PV818 Mobile monitor stand
7
3.6 Touchscreen
The system can be controlled by using the touchscreen. The touchscreen is 2D only and not sterile.
Position the touchscreen for quick and convenient access.
3.7 Footswitch PV014
The system can be controlled by using the wireless footswitch. The foot­switch is uniquely paired with the DSM system in the DSM software “Foot­switch” settings panel, see Chapter 6.8.11.
3.7.1 Default footswitch button functionality
a
b
c
d
e
Fig.6
Legend
a
Waypoint save/delete
b
Auto focus
c
Focus -
d
Zoom -
e
White light -
The footswitch button commands can be customized in the DSM software “Footswitch” settings panel, see Chapter 6.8.11.
f
Waypoint select/move
g
Lock-on mode toggle
h
Focus +
i
Zoom +
j
White light +
f
g
h
i
j
Change the batteries of the footswitch when the battery indicator
flashes orange, see Chapter 5.12.
3.7.3 Backup USB cable
Fig.7
Alternatively, the footswitch can be powered by the 4.5 m optional backup cable.
3.8 Isolation transformer
The system contains a medical-grade isolation transformer to ensure patient and product safety. The transformer is configured at the factory to accept 100 V AC, 120 V AC, or 220 V AC to 240 V AC from the supply mains. The DSM computers, 3D display, and touchscreen are connected directly to the isolation transformer. The isolation transformer is con­nected to the outlet power source – the isolation transformer outlet plug is the mains disconnect device. The isolation transformer prevents out­ages, sags, surges, and bad harmonics.
3.9 Wireless keyboard
The system includes a wireless keyboard with an integrated touchpad. The wireless keyboard is rechargeable. There are no batteries to be replaced. A three-hour charge provides up to ten days of use when typing about two hours a day with the backlight on, or about one year without backlight. The backlight is off by default. When the keyboard is turned on, the battery indicator shows the charge level.
Indicator Description
Blinking green The battery is charging.
Solid green The battery has adequate charge, or charging is
complete.
Blinking red Battery power is low. Recharge the battery.
To charge the battery, connect the keyboard to a USB port a on the side
of the touchscreen mount using the Micro-USB cable supplied, see Fig.5.
The battery indicator blinks green while charging.
Charge until the indicator light stops blinking.
3.7.2 Footswitch status LEDs
Symbol State Designation
Connection
Battery
8
Solid green
Blinking green
Blinking red
Off Footswitch in power save mode
Solid orange
Blinking orange
Fast blink­ing orange
Off Depending on Connect LED:
Radio connection established, footswitch engaged
No radio connection, e.g. radio range error or receiver not online
Pairing mode active Follow pairing instructions, see Chapter 6.8.11.
Warning 1: “Battery low” Battery voltage = 3.3 V to 3.2 V (± 10 %)
Warning 2: “Battery low” Battery voltage ≤ 3.2 V (± 10 %)
Battery is empty: No further operation is possible. Battery voltage < 3.0 V (± 10 %)
Connect LED ON: Battery voltage > 3.55 V Connect LED OFF: Footswitch in sleep mode
Note
The keyboard can be used while it charges.
Note
If the battery power is low (indicator blinking red), the keyboard backlight is disabled.
3.10 Polarized 3D glasses
Passively, circularly-polarized glasses are needed to see the display in 3D. The glasses may be worn over prescription glasses. Keep the glasses clean and free of scratches. To ensure optimal image quality, the system should only be used with Aesculap-supplied 3D glasses. Polarized prescription glasses will distort the 3D effect and should not be used.
Contact Aesculap if additional pairs are required.
The following 3D glasses are approved for use with the DSM:
Art. no. Designation
PV621 3D polarization glasses
PV622 3D anti-fog glasses
PV623 3D polarization glasses with clip
PV624 3D eye shield glasses kit
3.11 Additional components
The DSM system comes with additional components to help configure the system to work best in the operating room.
Art. no. Designation Function
FS095 Mains cord
FS096 Europe
FS097 Great Britain and Ireland
FS098 Japan, USA, Canada
PV012SU DSM sterile drape Sterile drape for the head and robotic
PV030 White balance
PV032SU DUV 400 refer-
TA015635 Lens cap Protecting the lens of the DSM head
Aesculap doesn’t provide a sterile drape for the DSM cart. If required, the sterile drape should meet the following recommendations:
orange angled
card
ence card
Switzerland
arm, sold separately. Single use only
White-balancing the DSM head image sensors. The reverse side of the card is used for checking the focus and align­ment of the DSM head optics.
For checking the functionality of the DUV 400 imaging module (optional). Single use only for accuracy check
when it is being stored
5. Working with the product
Risk of injury and/or product malfunction due to incorrect operation of the electromedical system!
WARNING
5.1 Transport
Do not push the product while transporting. Only pull the product,
especially over elevator and door thresholds.
5.2 System set-up
5.2.1 3D display and robotic arm workspace
To help with positioning the DSM in the operating room, the cart dimen­sions, the minimum and maximum extensions of the 3D display arm, and the height and maximum extension of the robotic arm are shown below.
Adhere to the instructions for use of any medi-
cal device.
a = 2,34 m
b = 2,03 m
c = 1,47 m
Minimum height
Maximum height 75 cm
Minimum width
Maximum width
Additional recommenda­tions
30 cm
70 cm
115 cm
Must not hang over the castors
Must not cover the cart vents
4. Preparation and setup
Non-compliance with the following instructions will preclude all respon­sibility and liability in this respect on the part of Aesculap.
When setting up and operating the product, adhere to
– national regulations for installation and operation, – national regulations on fire and explosion protection.
Note
For the safety of patients and users it is essential that the mains power cord and, especially, the protective earth connection are intact. In many cases defective or missing protective earth connections are not registered imme­diately.
Connect the device via the potential equalization terminal j at the front
of the cart to the potential equalization system of the room used for medical purposes, see Fig.5.
d = 1,86 m
e = 1,24 m
Fig.8
Legend a Max. length (arm fully extended) b Max. monitor height c Min. monitor height d Highest point (roll through doors) e Radius
Note
The potential equalization lead can be ordered from the manufacturer as art. no. GK535 (4 m length) or TA008205 (0.8 m length).
9
a
b
c
There are two recommended layouts for the operating room.
Neurosurgery operating room layout (example)
b
a
c
Fig.9
Legend a Surgeon b Assistant c Scrub nurse
Spinal surgery operating room layout (example)
c
a
Also note that any equipment connected at the interfaces must demon­strably meet the respective IEC standards (e.g. IEC 60950 for data process­ing equipment, IEC/DIN EN 60601-1 for electromedical devices).
All configurations must comply with basic standard IEC/DIN EN 60601-1. Any individual connecting devices with one another is responsible for such configuration and must ensure compliance with basic standard IEC/DIN EN 60601-1 or applicable national standards.
Please address the B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see Chapter 10.
Every time a new piece of equipment is added and cables are connected
to the DSM system, power down the system prior to connecting cables, then power it back up once the cables are connected.
5.2.3 Connecting the power supply
Risk of fatal injury from electric shock!
Connect the product only to a grounded power
DANGER
Connect the main power cable to a hospital-grade power outlet. Do not
supply.
use un-approved extension cords or power strips. Use uninterruptible power source when loss of power would result in an unacceptable risk.
Lay the power cable in areas with minimal foot traffic.
5.3 Function checks
Before turning the DSM system on
Check that the castors are locked once the cart is in the desired loca-
tion. Make sure the cart is stable and cannot roll in any direction.
Connect the main power cable, see Chapter 5.2.3.Connect any external input or output video or Ethernet cables, see
Chapter 3.3.3.
Check that there are no unnecessary objects or people within the range
of the robotic arm workspace.
Check that the DSM head optical path is clean and free from dust.Ensure that exposed cables for the footswitch, any external displays,
and power supply are laid flat on the ground and out of high-traffic areas to minimize potential tripping hazards.
b
Fig.10
Legend a Surgeon b Assistant c Scrub nurse
5.2.2 Connecting the accessories
Risk of injury due to unapproved configuration using additional components!
DANGER
For all applied components, ensure that their
classification matches that of the application component (e.g. Type BF or Type CF) of the respective device.
Combinations of accessories that are not mentioned in the present instructions for use may only be employed if they are specifically intended for the respective application, and if they do not compromise the perfor­mance and safety characteristics of the products.
After turning the system on and before starting the surgery
Position the yaw coupler as desired, see Chapter 6.2.1.Move the robotic arm from “Storage” to “Drape” position, see
Chapter 6.2.1.
Follow the instructions printed on the drape packaging to correctly
apply the drape to the DSM head and robotic arm.
Position DSM at the OR table.Position the 3D display for the surgeon’s viewing preference.Check that the “Zoom”, “Focus”, and “White Light Level” handle buttons
are functioning properly.
Check that the 3D image colors, focus, and alignment look correct.Check that the robotic arm assisted drive is moving smoothly and pre-
dictably.
Check that the footswitch is paired with the system and the buttons
are working properly.
If using fluorescence visualization (DUV 400), check the fluorescence
module using a reference card to ensure proper working order.
5.4 Positioning the 3D display
For ideal viewing, position the 3D display next to the patient table,
between 4 and 6 feet (1 to 2 meters) away from the surgeon.
Position the screen perpendicular to and at the same height as the sur-
geon’s eyes when he or she is seated at the operating table.
The 3D display arm’s maximum extension and height are shown below:
10
b = 2,03 m
a
c = 1,47 m
Fig.11
Legend a Top of the screen b Max. monitor height (top of the screen) c Min. monitor height (top of the screen)
5.5 Turning the system on
Perform the relevant items of the function checks, see Chapter 5.3.
a
Fig.13
Legend a Sliders of the yaw coupler
Unlock the sliders a and rotate the DSM manually into the desired posi-
tion. The available coupler positions are illustrated below. The geometry lim-
its of the robotic arm will change based on the coupler position.
-90° 90°
Fig.12
a
Legend a Main power button
Press the main power button a located on the back panel of the DSM
cart, next to the power cable connection and LED indicators. The displays, head, computers, robotic arm, and software are set to
automatically start when the system is turned on.
Wait for the system to initialize (about 90 seconds).
The system will be ready for use once the DSM software has opened and the notification “DSM initializing, please wait…” is no longer dis­played.
5.6 Setting up the robotic arm and the DSM head
5.6.1 Rotating the DSM using the yaw coupler
Depending on the surgical approach and operating room layout, it may be advantageous to rotate the DSM head 90° to the left or right.
Note
Only adjust the yaw coupler before draping, using the proper “Reconfigure Coupler” workflow. Do not adjust the yaw coupler during surgery.
Once the system has fully started up, go to the “Robot” tab in the DSM
home screen, see Fig.22.
Under “Coupler”, press the “Reconfigure Coupler” button.Press and hold the “Hold or Reconfigure position” button until the arm
is fully extended.
Fig.14
Press the “Hold for Drape position” button in the DSM software until
the movement is finished.
5.6.2 Moving the robotic arm out of storage position and draping the robotic arm
The DSM head and distal portion of the robotic arm must be draped using the provided Aesculap sterile drape PV012SU prior to surgery.
Note
Additionally, the DSM cart may be draped at the user’s discretion (cart drape not provided by Aesculap). For drape dimensions, see Chapter 3.11.
Once the system has fully started up, move the robotic arm into the
“Drape” position, see Chapter 6.2.1.
Once the DSM is in the “Drape” position, follow the instructions printed
on the drape packaging to correctly apply the drape to the DSM head and robotic arm.
If the drape affects the balance of the robotic arm, rebalance the DSM,
see Chapter 6.2.2.
Perform all the remaining items of the function checks before starting
the surgery, see Chapter 5.3.
11
5.7 Starting a case
Once the DSM arm has been draped and all pre-operative checks are com­plete, a new case can be started.
Fig.15
Select a surgeon profile and a patient in the “Case” tab.
– or –
Add a new Surgeon profile or Patient by using the Add button.
DSM settings and preferences are saved for each Surgeon profile. The Delete button is disabled when the “Default Surgeon” or
“Default Patient” is selected, but it can be used to delete custom pro­files.
When the DICOM settings are configured to a networked database,
select the “DICOM” toggle and search for a patient. To search all listed patients in the database, enter an asterisk (*) in the first and last name search fields.
Once the desired surgeon and patient have been chosen, press “Start
Case” to advance to the live screen, see Chapter 6.3.
Press “Yes” to continue saving the media information.
– or –
Press “No” to permanently delete all case media (this will require con-
firmation). – or –
Press “Cancel” to return to the case.
Selecting “Yes” will allow the patient’s information to be added or updated. Patient information saved prior to starting the case will be retained at the end of the case.
Fig.17
Enter a first and last name, then press Next button to continue, or
Cancel button to return to the “Save Recorded Media?” message.
Press the pencil button to add or update “Optional Patient Data”.
This option is enabled once a first and last name have been entered.
Press OK button when done editing to proceed.
5.8 Ending a case
Select the End Case button to end the current case.
If recordings or snapshots were taken during the case, there will be options to save and/or upload the case media for both DICOM and non­DICOM patients to the DICOM server (if applicable), local drives, or connected external drives.
5.9 Saving case media
If snapshots and/or images were recorded, ending the current case will prompt users to save the media. The patient information can also be updated at this time, even if the case was started with previously saved information.
5.9.1 Save to default location
After selecting to end the case:
Fig.16
Fig.18
From the “Media Information” screen, optionally keyword(s) can be entered or the folder name changed.
Press Back button to return to the previous screen or press OK
button to proceed.
12
5.9.2 Copy media to drive
Fig.19
To copy case media to an available local or externally connected drive,
select the drive letter from the drop-down list, then press “Yes” – or –
Press “No” to save to the default location only.
The save media screens will close and the application returns to the home screen.
There will be progress messages for the copy in progress, as well as con­firmation once it completes.
Note
The location where case media will be saved is F:\TVS\TrueVision\Sur­geons\<SurgeonName>. Selecting an additional drive to copy to will be in addition to the default location.
5.10 Removing the drape
At the end of the case or when the robotic arm drape needs to be removed:
Ensure the area is clear of objects or people and move the robotic arm
into the “Drape” position, see Chapter 6.2.1.
Release the stripes on the drape bag.Detach the adapter from the camera head.Invert and remove the drape
along the robotic arm.
Dispose the sterile drape in compliance with medical waste regulations.
5.11 Shutdown procedure
For safety, it is important to follow proper shutdown procedure when the DSM system is not in use. The software will need to be shut down, the robotic arm will need to be moved into storage position, and the system power should be turned off with the main power button.
5.11.1 Software shutdown and robotic arm storage
After the drape has been removed from the robotic arm, press the Exit
button.
Once the 3D display is cleared from the robotic arm movement path,
move the robotic arm to “Storage” position, see Chapter 6.2.1 or press “Cancel” button (return to the application).
Once the robotic arm is fully in “Storage” position, the software will automatically provide further options to “Shut Down”, or press “Can­cel” button (return to the application).
5.12 Changing footswitch batteries
Battery type: 3 × IEC- LR14 (Type C – “Baby”)
Shut down DSM, see Chapter 5.11.Open battery housing: Rotate both lockers near battery housing by 90°.
While doing so, lightly press down the battery cover.
Lift the empty batteries by the ejector belt and remove them.Insert new batteries. Keep attention to the right polarity (see marking
on battery compartment bottom).
Close battery housing: rotate both lockers back to original position.
6. Software
6.1 Home screen
The DSM software starts on the home screen on the touchscreen.
Fig.21
Symbols
Protective stop
As a risk control, the software has a protective stop button in the unlikely scenario that the robotic arm moves in an unpre­dictable way. The protective stop button halts all robotic arm movement when pressed. The button is available in both the home and live screens. The button will change color from red to gray when engaged (On, movement disabled). Press the button again to remove the protective stop and allow the robotic arm to move again. The button will change back to red.
Eject drive
When an external USB drive is connected to the system, select the Eject button to properly eject the drive. The software will confirm the drive has been ejected and can be physically removed.
TrueMedia application
Select the TrueMedia application playback button to view or edit surgical content. The DSM software will need to close in order to open TrueMedia.
Fig.20
Fullscreen view
Select the fullscreen button to enter fullscreen mode, which displays the surgery view and contains active status messages and the buttons for emergency stop, settings, and exit fullscreen mode. The fullscreen view button is available in both the home and live screens. Select the close fullscreen button to exit fullscreen mode.
13
6.2 DSM setup
The “Setup” panel in the home screen provides the following functions:
Move the robotic arm to a preset position
Rebalance the DSM
Enter recovery state
Reconfigure the coupler.
Note
Setup should always take place before surgery.
Fig.22
6.2.1 Preset positions
The available preset positions are:
Drape
Storage
6.2.2 Rebalance scope
Moving the robotic arm to its limits can affect the DSM’s balance. To cor­rect the presence of any drift, the DSM can be rebalanced.
The DSM robotic arm is factory-calibrated. It is recommended to rebalance the robotic arm once annually or as needed, especially if the robotic arm feels unbalanced during movement with the handle brake-release buttons.
Ensure the area of the robotic arm is clear.Press “Rebalance Scope” in the “Setup” tab.
In rare instances, some configurations can cause the robotic arm to lock up when reaching the joint limits.
If this occurs, move the robotic arm to a preset position.
Note
If proper rebalancing of the DSM still results in unsatisfactory balance, con­tact Aesculap Technical Service for assistance, see Chapter 10.
6.2.3 Robot recovery
If communication is lost with the force-torque sensor or the coupler and the robotic arm becomes difficult or impossible to move, the arm joints need to be manually repositioned.
Press and hold “Robot Recovery” in the “Setup” tab.Manually reposition of the arm joints.
6.3 Live screen
Selecting “Start Case” in the “Case” tab advances to the live screen, intended for live surgery. The Live Screen shows controls for the DSM head, robotic arm, recording, fluorescence, and access to settings.
a = 450 mm
b = 1 240 mm
Fig.23
Legend a Keep this area clear vertically during position change b Keep this area clear horizontally during position change
Risk of injury due to unintended movement of the robotic arm!
WARNING
Select a pre-configured position for the robotic arm in the “Position”
drop-down list.
Press the “Press and Hold to auto-position” button until the robotic
arm movement fully stops in position.
Make sure the area surrounding the robotic arm
is clear of objects or people.
Watch the robotic arm while it is moving.
Fig.24
6.3.1 Zoom, focus, and light controls
The “DSM” tab within the “DSM Control” panel shows controls for zoom, focus, white light levels, and DUV light levels (if licensed). The DSM control values can be adjusted by using the handles, footswitch, or the touch­screen.
Press anywhere on the value bar or press and drag the slider to change
the values.
Select the or buttons for fine tuning.
Setting Range
Zoom 1.00 × to 10.00 ×
Focus 200 mm to 450 mm
White Light 0 % to 100 %
DUV Light 0 % to 100 %
14
6.4 Fluorescence controls
Note
Fluorescence module (DUV 400) requires a license for use with the DSM sys­tem.
DUV 400: The integrated DUV 400 fluorescence module of the DSM is used to visualize near-UV fluorescent areas in the surgical field. The mod­ule has been designed for excitation in the wavelength range from 390 nm to 420 nm and for fluorescence visualization in the wavelength range greater than 510 nm. The module enables the surgeon to switch between white light and blue-violet excitation light for fluorescence. In addition to visualization, users can record fluorescent and auto-fluorescent light emitted by tissue by enabling the “Overlay” feature.
Fluorescence controls can be adjusted by using the handles, footswitch, or touchscreen.
Note
The “Autofocus” feature is disabled while fluorescence function DUV 400 is enabled. Use the touchscreen, handles, or footswitch to manually refocus the target image.
6.4.1 Function checks
The fluorescence reference card and the touch­screen are not sterile!
WARNING
Before starting fluorescence visualization, always check the function-
ality of the integrated fluorescence module.
Ensure that the DSM system is in a dark room, with no other illumina-
tion other than the DSM integrated lighting and the system displays.
Place the reference card on a level surface.Set the focus to 325 mm and the zoom to 5 × on the touchscreen. Refocus the image by moving the DSM head upwards or downwards
using assisted drive.
Activate the fluorescence function by selecting the “DUV 400 On” but-
ton on the touchscreen or the corresponding handle or footswitch but­ton.
In the live image, check the following:
DUV 400: There is distinct contrast (orange/pink) between the DUV
400 reactive compound and the reference card (green/white).
– The reactive compound must be clearly distinguishable and the live
image should not be washed out. Slight deviations in color and brightness are acceptable.
6.4.2 DUV 400 imaging
The following factors influence the visualization of the fluorescence sig­nal:
Perform fluorescence functional tests before
surgery.
Fluorescence medium and its concentration in the tissue
Illumination intensity of the light source in the defined wavelength
range (both blue and white light)
Transmission of the optical system
Working distance and the illuminated field diameter
DSM white light level
Activate the fluorescence function by selecting the “DUV 400 On” but-
ton on the touchscreen or the corresponding handle or footswitch but­ton.
The “DSM Control” panel will still be available and – additionally – the “DUV Light Level” value bar can be adjusted while DUV 400 is on.
Select “Overlay” to show an overlay of captured fluorescence in the live
view.
Select “FL Off” to turn off near-ultraviolet light and return to white
light only.
6.5 Touchscreen robot control
Fig.25
The robotic arm pan and rotation movements can be controlled from the “Robot” tab in the “DSM Control” panel in precise, single-axis movements. Each arrow (Forward, Back, Left, Right, Up, and Down) corresponds to a directional movement in relation to the DSM head.
Holding down one of the directional arrow buttons moves the robotic arm until a joint limit is reached. Directional movements can also be employed by configuring the corresponding commands to the handle buttons (see Chapter 6.8.10) or footswitch buttons (see Chapter 6.8.11).
6.5.1 Lock On Target
The “Lock On Target” function locks the DSM onto a selected focal target in the live image. When enabled, directional movement will rotate around the focal target instead of moving linearly. Up and Down directional movement remains unchanged while “Lock On Target” is enabled.
“Lock On Target” does not apply when moving the DSM by using the han­dle brake-release buttons, even when enabled. After such a movement, Lock On Target movement will resume as long as it is enabled.
Press the “Lock On Target” button to activate/deactivate the function.
6.5.2 Robot Speed
If the slider position is closer to “Min” (minimum), the robotic arm move­ment will be slower, and if the slider position is closer to “Max” (maxi­mum), the robotic arm movement will be faster.
Change the speed of the robotic arm movement by clicking anywhere
on the “Robot Speed” bar or by clicking and dragging the slider.
6.6 Quick Access Bar (QAB)
The Quick Access Bar (QAB) can be configured to show the top four most­used available functions in the home and live screens, as well as while in fullscreen.
Fig.26
Edit the QAB by selecting the pencil button and toggling on or off
the available functions. Selected QAB functions are highlighted in green and are added to the
bar by order of selection.
15
The following QAB functions are available:
Function Description
Viewport Configure the format for multiple inputs on screen.
Select between primary input, secondary input, split screen, or picture-in-picture. Configure a second video source in the input settings.
Swap Sources Switch between the primary and secondary input in the
main view.
Orientation Toggle the orientation of the primary input between
standard and inverted 180 degrees.
Digital Aper­ture
Autofocus Select to automatically focus the object in the center of
Toggle a digital aperture that blocks out the periphery of the image. Adjust the size and transparency in the “Digital Aperture” settings, see Chapter 6.8.3.
the DSM image.Start the autofocus by:
Pressing an appropriately configured handle button,
see Chapter 6.8.10
Pressing an appropriately configured footswitch
button, see Chapter 6.8.11
Note
If “Auto-focus After Move” in the “Camera” settings panel is disabled, the focal target will not be reassessed after manual movement while “Lock On Target” is enabled. To refocus image manually, use the touchscreen, han­dles, or footswitch.
6.7 Robotic arm state indicators
The robotic arm has defined force, speed, and geometric limits for safety. As the arm approaches a limit, the indicator shown in the lower right of the live screen will change its icon and color.
State Icon Description
Footswitch Directional Move with Restrictions
Handle Directional Move with No Restrictions
Handle Directional Move Approaching Restric­tions
Free Drive State Displayed when robot arm is in “Free
Pre-Programmed Posi­tion Move
Assisted Drive Move with No Restrictions
Assisted Drive Move Approaching Restric­tions
Assisted Drive Move with Restrictions
Uncalibrated State Displayed when arm hasn’t been cal-
Handle Directional Move with Restrictions
Displayed when moving the arm with the footswitch and arm reaches a directional limit
Displayed when moving the arm with the handle buttons and there are no movement restrictions
Displayed when moving the arm with the handle buttons and arm approaches directional limit
Drive State”
Displayed when arm is moving to a pre-programmed position
Displayed when moving the arm with the handles and there are no movement restrictions
Displayed when moving the arm with the handles and arm approaches directional limit
Displayed when moving the arm with the handles and arm reaches a directional limit
ibrated
Displayed when moving the arm with the handles buttons and arm reaches a directional limit
Specification Limit
Force 250 N
Speed 1 m/s
The Indicator states are pictured and described in the table below:
State Icon Description
Initializing State Displayed during initial power ON, or
when arm is initializing
Ready State Displayed when arm is ready for
movement
Emergency Stop State Displayed when Emergency Stop
button is engaged
Protective Stop State Displayed when the robot is in a pro-
tective stop state after the arm movement is interrupted
Soft Stop State Displayed when soft stop state is
selected
Footswitch Directional Move with No Restric­tions
Footswitch Directional Move Approaching Restrictions
Displayed when moving the arm with the footswitch and there are no movement restrictions
Displayed when moving the arm with the footswitch and arm approaches a directional limit
GUI Directional Move with No Restrictions
GUI Directional Move Approaching Restric­tions
GUI Directional Move with Restrictions
Failure State Displayed during a hardware mal-
Coupler in Unlocked Position
Note
In the unlikely event that the robotic arm becomes difficult to move near a geometric limit, it is recommended to move the robotic arm back to a preset position (e.g. Drape).
Displayed when moving the arm from the user interface and there are no movement restrictions
Displayed when moving the arm from the user interface and arm approaches a directional limit
Displayed when moving the arm from the user interface and arm reaches a directional limit
function
Displayed when coupler in Unlocked position and while changing coupler orientation
6.8 Software settings
Select the Settings button to open system settings panels.
The “Camera” settings panel will open by default the first time the set­tings are accessed.
After that, the last used panel will be opened in subsequent accesses. Selecting a different Settings button will automatically save any changes made and switch panel.
Selecting the More Settings button reveals additional settings panels.
16
Commonly found buttons in the settings panels include:
Icon Description
Exit panel without saving changes made to current panel.
Reset the current panel to its default settings. (Not present for DICOM or Input panels.)
Save changes and close panel.
6.8.1 Camera settings
In the Camera settings panel the following settings of the DSM camera can be modified:
Orientation
Auto-exposure
Auto-focus After Move
Orientation
The Orientation drop-down menu has the following options:
Standard (default, used for most configurations of the DSM head to
match the surgeon’s view)
Rotated (180° view from Standard orientation)
The orientation applies to all connected monitors, unless “Assistant” view is being used; see Chapter 6.8.7 for more information on “Assistant” view settings.
Note
The DSM handle button layout will rotate if the view is set to "Rotated" in software settings.
Auto-exposure
When “Auto-exposure” is enabled, the live view automatically adjusts for changing lighting conditions and is especially effective in reducing red color saturation.
When “Auto-exposure” is disabled, no adjustments to the live image will be made.
Press the “Auto-exposure” toggle to disable (default) or enable the fea-
ture.
Auto-focus After Move
If “Auto-focus After Move” is enabled and the DSM is moved via any method, once the DSM stops movement the software will automatically refocus on the current focal point.
When this feature is disabled, manual adjustment to achieve ideal focus will be required (using the handles, footswitch, or touchscreen).
Select the “Auto-focus After Move” toggle to enable (default) or dis-
able the feature.
6.8.2 Screen Layout
In the Screen Layout settings panel the following settings of the DSM camera can be modified:
Picture in Picture (PiP) Location
DSM Settings
Picture in Picture (PiP) Location
The “PiP Location” drop-down menu has the following options:
Bottom Right (default)
Top Left
Top Right
Bottom Left
DSM Settings
The “DSM Settings” drop-down list has the following options:
Option Description
Auto-Hide (default)
Show The DSM control values will always been shown on the
Hide The DSM control values will never be shown on the sur-
6.8.3 Digital Aperture
The Digital Aperture feature enables a digital overlay where the live image can be restricted to a circle in the center of the screen, with a dark mask covering the rest of the image.
Press the toggle button to disable (default) or enable the feature.
Diameter
Use the “Diameter” bar to select or slide the value anywhere between
0.05 (smallest diameter) and 1.00 (largest diameter). The default Diameter is 0.75.
Opacity
Use the “Opacity” bar to select or slide the value anywhere between
0.05 (most transparent mask) and 1.00 (most opaque mask). The default Opacity is 0.50.
6.8.4 DUV 400 (Fluorescence)
In the DUV 400 panel the following settings for the DUV 400’s overlay can be modified:
The DSM control values (Zoom, Focus, and White Light) will be displayed over the live surgery screen for ~5 sec­onds and then hidden after adjusting the values.
surgery screen(s).
gery screen(s).
Hue: can be adjusted on a range of yellow to blue (default: green)
Opacity: can be adjusted from 0 to 100 (default: 100)
“Auto-Overlay”
“Auto-Overlay”, when enabled, transitions from DUV 400 to Overlay after a set amount of time. “Overlay Mode Timer” can be adjusted from 1 to 60 seconds. By default, “Auto-Overlay” is disabled and “Overlay Mode Timer” is set to 30 seconds.
6.8.5 Color
In the Color panel the following settings can be modified:
Gain: can be adjusted from 0 to 80 (default: 0)
Picture: adjust Brightness, Contrast, Gamma, Hue and Saturation
White Balance
Picture
Note
These are advanced settings. It is not recommended to change these set­tings.
Select the desired quality setting by pressing or pressing and dragging
the slider on any of these settings: – Brightness: determines the intensity of the color in the live image
(range: 20 to 80, default: 48)
– Contrast: difference between the lightest and darkest regions of the
live image (range: 20 to 80, default: 55)
– Gamma: non-linear operation used to encode and decode lumi-
nance of the live image (range: 0.50 to 5.00, default: 1.20)
– Hue the degree to which a stimulus can be described as similar to
or different from stimuli that are described as red, green, blue, and yellow (range: -180 to 180, default: 2)
– Saturation: the amount of white mixed into the hue (range: 0 to
300, default: 90)
To return to the original settings, press the “Default” button.
17
White Balance
Every DSM is shipped with pre-configured white balance settings.
Note
These settings may not be optimal for the specific lighting conditions.
To reset the white balance to the factory settings, press “Reset to Fac-
tory WB” button.
For the best image, the DSM should be white-balanced.
Set the DSM light to the surgeon’s normal working conditions.Place the white balance card PV030 under the light. Ensure that the
entirety of the live view is filled by the card. Never use gauze, tape, sheets, or other fabrics which may have been bleached.
Tilt the card and move it either closer or farther from the light as the
software instructs.
Select White Balance button to perform white balance and follow
the on-screen instructions to complete the white balance process.
Test the colors by viewing your hand under the DSM. Repeat the pro-
cess if the colors look incorrect.
6.8.6 Recording
In the Recording panel the following settings can be modified:
Movie or snapshot file formats
Auto-record
Fullscreen recording
Frame rate
Movie quality
DICOM 3D recording
Note
Recording settings cannot be changed while recording a case.
– Squeezed SxS (side by side): checking the checkbox reduces resolu-
tion of saved media to 1920 × 1080 (default: unchecked). Select this option when smaller file sizes and/or the ability to edit with 3rd party editing software are required.
– DICOM Record 3D: when a DICOM patient is selected at the start of
a case, all media recordings will be in 2D by default. Check the checkbox to record all DICOM case media in 3D. Uncheck the check­box to record in 2D again.
Note
Non-DICOM patient case media are in 3D by default, therefore the “DICOM Record 3D” feature is disabled when a non-DICOM patient is selected.
6.8.7 Display
In the Display panel certain display settings can be modified, including the view type, 3D format, and polarity of connected monitors (see Chapter 3.3.3 for a list of compatible displays).
Identification information (numeric identifier, model number, 3D format, polarity, resolution, and view type) will automatically appear on the live image of each connected monitor while the Display settings panel is open.
Note
For displays intended for use (see Chapter 3.3.3), the 3D format and polarity (“Swap 3D”) are automatically detected and set to their ideal pre-configu­rations, as long as “Detect” remains enabled with the default panel settings.
To return to the ideal pre-configuration after making changes, select
“return to defaults” button, then ensure “Detect” is enabled.
Detect
Press the “Detect” button to turn this feature on or off.
Detect is set to “Off” by default to allow the software to automatically configure the connected displays to their saved settings.
Basic settings
Note
Fullscreen recording is not recommended due to possible delays in the image.
Select the desired recording settings:
– Movie format: Press the radial button of either “AVI” or “MOV”
(default: MOV).
– Auto-Record: check the checkbox to have the software start record-
ing as soon as “Start Case” is selected (default: unchecked).
– Fullscreen: Check the checkbox to record the surgery screen with all
annotations and overlays. If unchecked, the software will only record the raw DSM input (default: unchecked).
– Snapshot format: Press the radial button of either “JPEG” or “PNG”
(default: JPEG).
Advanced settings
Note
Increases or decreases to frame rate and quality can be expected to have relative increases or decreases to file size.
Select the desired advanced recording settings:
– Frame rate: frames per second that will be captured during a
recording. Select the desired frame rate by pressing or pressing and dragging the slider from any value between 30 fps and 58 fps (default: 45 fps).
– Quality: can be adjusted from 0 to 10, resulting in respective
changes to recording quality. Notably, minimizing the slider can introduce compression elements such as haloing or blurry frames to the recording with the benefit of reducing file size. Select the desired quality setting by pressing or pressing and dragging the slider from any value between 1 and 10 to change the quality for recordings (default: 5).
Display List
To view or manually modify settings for a specific display, select the
corresponding numeric identifier for that display from the “Display” drop-down list.
View
Press the “View” drop-down list to assign a view type to the selected
display.
The available view types are:
Control
Surgery
Assistant 180
Assistant 90 CW
Assistant 90 CCW
“Surgery” view is a non-interactive secondary view with the primary func­tion of displaying the live image. “Surgery” view will display relevant noti­fications as they appear on the “Control” view.
“Assistant 180", “Assistant 90 CW” (clockwise), and “Assistant 90 CCW” (counter-clockwise), are “Surgery” views that are rotated as indicated.
Note
At least one display must be assigned a “Control” view type. The touch­screen display is assigned as the “Control” view by default.
3D Format
When “Detect” is set to “Off”, press the “3D Format” drop-down list to
change the 3D format for the selected display from the following avail­able options:
– Monoscopic – Row-Interleaved – Side-by-Side – Top-Bottom –Dimenco
18
Swap 3D
“Swap 3D” switches the left and right eye polarity and can be used to help determine if the monitor is using the correct 3D polarity setting.
To check the polarity, close the right and left eye alternatively while
wearing 3D glasses and ensure the on-screen “Left” and “Right” text matches the eye opened.
6.8.8 Input Source
In the Input Source panel external display devices, such as an endo­scope, can be configured to the correct input type.
The available input source types are:
SDI
HDMI
“Viewport 1" is set to the 3D live image from the DSM head.“Viewport 2" can be configured to display either SDI or HDMI input from connected external devices. Both SDI and HDMI devices can be connected simultane­ously, however only one type can be displayed at a time.
Press the corresponding radial button of the desired input type.
The following Aesculap camera systems are supported by the DSM:
Art. no. Supported settings
PV460 60 Hz signal, DVI output on PV460, SDI input on DSM
PV470 50 Hz signal, DVI output on PV470, HDMI input on DSM
PV480 50 Hz or 60 Hz signal, DVI output on PV480, HDMI or SDI
input on DSM
6.8.9 DICOM
Fig.27
In the DICOM panel the networking for uploading and downloading media to and from a configured server database can be set up.
Note
The DICOM module requires a license for use with the DSM system. See DICOM Conformance Statement TA015615-DCS for information
related to the use of DICOM data exchange.
Network and DICOM addresses must be provided by the IT administra-
tor responsible
AE titles are case-sensitive
A connection test is automatically run prior to each DICOM upload to
verify both servers are running and accepting the data type being sent
The information on the “DSM” tab is static. The settings in “HIS/RIS”
and “PACS” tabs are editable and can be updated at any time.
Setting up network connection
To set up the initial connection with the hospital network, consult with
local network administrator.
Connect an Ethernet cable to the hospital network installation and to
one of the Ethernet ports b/e on the cart, see Fig.5 the system.
Turn the DSM on, see Chapter 5.5.
Exit out of the DSM software onto the Windows Desktop, see
Chapter 5.11.
Configure Ethernet connection using the Windows Control Panel.
6.8.10 Handles
In the Handles panel the layout and functions of the DSM head handles using can be customized.
Press the “Left–Right” toggle to switch between handles. See below for a list and description of each button’s function.
Dual Mode
Press the “Dual Mode” checkbox to enable/disable this function.
When Dual mode is enabled, both handles share the same configura­tion. Two Dual mode setups can be configured, A and B, and toggled between. The A–B toggle is only accessible when “Dual Mode” is enabled.
Available handle button functions
Button description Functionality
None No Function
Autofocus Microscope will automatically focus on the
center of the field of view.
Button Layout Map Displays the current button layout.
Cycle DUV 400 Toggle to a blue-light filter to intraoperatively
differentiate between diseased and healthy tis­sue after a 5-ALA injection.
DUV Light Down Reduce the amount of light while DUV 400
mode is on.
DUV Light Up Increase the amount of light while DUV 400
mode is on.
Digital Aperture Tog­gle
Focus (+) Move the objective lens closer.
Focus (-) Move the objective lens farther away.
Move (Down) Move the DSM head down.
Move (Up) Move the DSM head up.
Record Start/stop video recording.
Waypoint Save/Delete Press button once to save current location and
Waypoint Select/Move
Snapshot Take a snapshot.
Swap Sources Toggle between input sources. Can be config-
Lock-on Mode Toggle Toggle lock to target mode.
Viewport Layout Cycle
White Light (-) Decrease white light.
White Light (+) Increase white light.
Zoom (+) Increase zoom.
Zoom (-) Decrease zoom.
Toggle digital aperture overlay. Diameter and Opacity can be changed in settings.
configuration as a waypoint. Hold down button to delete selected waypoint.
Press button to toggle through saved way­points. Hold down button to move to selected waypoint.
ured to either SDI or HDMI.
Cycle through available viewport settings. Can be configured to “Source 2”, “Picture in Pic­ture”, or “Side by Side” view.
19
6.8.11 Footswitch
In the Footswitch panel the footswitch can be paired with the DSM and the layout and function of the footswitch can be configured.
See below for a list and description of each button’s function.
Note
The footswitch has a red colored status indicator when unpaired.
Pairing the footswitch
Ensure that the footswitch is in sleep mode and there is no active radio
connection: – Do not press switches for at least 2.5 seconds. – Ensure that all LEDs on the footswitch are off.
Simultaneously press and hold buttons 1 and 2 on footswitch until the
LED slowly blinks red, then select the Pair button in the DSM soft­ware.
The fooswitch is ready for pairing. The footswitch LED will blink red until a new dongle is found or until pairing mode times out after ~30 seconds.
To stop the pairing process, press “Cancel”.
Once successfully paired, the status indicator in the software will be green and the footswitch will have a green LED when commands are sent to the software.
A yellow status indicator means the footswitch is paired but inactive (no current command is being sent to the software).
A green status indicator means the footswitch is paired and in use (a current command is being sent to the software).
By pressing any of the control components the footswitch changes into active mode and tries to establish a radio connection with the receiver. If there is no pressed button or switch for more than 120 seconds, the footswitch falls back to sleep mode. In this mode all indication LEDs are turned off.
Available footswitch button functions
The same button functions as for the handle are also available for the foot­switch, see Chapter 6.8.10
6.8.12 About
In the About panel the following system information is shown:
Software version
Technical support contact information
Language selection
Link to the user manual
Other software and regulatory information
Language selection
From the drop-down list, select the desired language, then select Save
button.
The DSM application will need to restart after changing the language.
Select Cancel button to keep the current language.
Note
Patient data saved in a language other than the current language will not be displayed.
User manual
Press “User Manual” to open the user manual folder found on the desk-
top.
Double-click on the DSM user manual to open the PDF version.
Note
The DSM application does not need to be closed in order to view the user manual.
More
Press “More” to open a separate panel with two pages.
Page 1 contains system information, including Computer Name, Mac Address, Camera, Licenses, and Software Part Number.
Page 2 contains regulatory information.
Press Back button to return to the main “About” panel.
7. Reprocessing procedure
7.1 Single-use products
Art. no. Designation
PV012SU DSM sterile drape
Risk of infection for patients and/or users and impairment of product functionality due to reuse.
WARNING
7.2 Preparation before cleaning
Remove any visible surgical residues as much as possible with a damp,
lint-free cloth.
7.3 Limitations on reuse
The camera housing can be disinfected up to 140 times with the proper care. Any additional reprocessing falls under the responsibility of the user.
For the other parts of the DSM there is no set maximum number of uses and reprocessing cycles.
The life of the product is limited by damage, normal wear and tear, type and duration of use, as well as handling, storage and transport of the prod­uct.
A careful visual and functional inspection before the next use is the best opportunity to recognize a product that is no longer functional.
7.4 Cleaning/disinfection
7.4.1 Product-specific safety instructions for the reprocessing pro­cedure
DANGER
CAUTION
CAUTION
Risk of injury, illness or death due to contamination and/or impaired functionality of the product!
Do not reprocess the product.
Risk of electric shock and fire hazard!
Unplug the device before cleaning.Do not use flammable or explosive cleaning or
disinfecting solutions.
Ensure that no fluids will penetrate the product.
Damage to, or destruction of the product caused by mechanical cleaning/disinfection!
Only clean and disinfect the product manually.Do not sterilize the product under any circum-
stances.
Damage to the product due to inappropriate clean­ing/disinfecting agents!
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer’s instructions for the respective cleaning/disin­fecting agent.
20
Damage to the product due to inappropriate clean­ing/disinfecting agents and/or excessive tempera-
CAUTION
tures!
Use cleaning and disinfecting agents according
to the manufacturer’s instructions which – are approved for plastic material and high-
grade steel,
– do not attack softeners (e.g. in silicone).
Observe specifications regarding concentration,
temperature and exposure time.
7.5 Wipe disinfection for electrical devices
Phase Step T
[°C/°F]t[min]
I Cleaning RT - - - 55% isopropyl alcohol,
II Wipe disinfection RT 2 - - 55% isopropyl alcohol,
RT: Room temperature * The manufacturer recommends Super Sani-Cloth® Germicidal Disposable Wipe (PDI, EPA register number 9480-4)
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Phase II
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (2 min minimum).
Conc. [%]
Water quality Chemical
0,25% n-alkyl dimethyl ethyl benzyl ammonium chloride, 0,25% n-alkyl dimethyl benzyl ammonium chloride*
0,25% n-alkyl dimethyl ethyl benzyl ammonium chloride, 0,25% n-alkyl dimethyl benzyl ammonium chloride*
7.8 Storage and transport
7.6 Inspection, maintenance and checks
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
Set aside the product if it is damaged.Periodically download or delete video files on the system computer to
avoid filling up the storage drive. The DSM system will not allow the user to make new recordings if the
storage drive is filled up.
7.7 Packaging
Ensure that the packaging will prevent a recontamination of the prod-
uct.
Fig.28
Ensure that the robotic arm is in “Storage” position before moving the
cart, see Chapter 6.2.1.
Wrap the power cable around its wrapping post and hang the wireless
footswitch from its on the front of the cart.Release the brakes on each of the four casters and move the 3D display into the position illustrated above.
Twist the arm joint handles to lock the arm in place. Lock the cart castors after moving the system.
8. Maintenance
Preventive maintenance The Digital Surgical Microscope does not consist of any device that
requires a regularly maintenance within a defined time frame. However, in an interval of 12 month a regular repeat inspection is manda-
tory. The repeat inspection has also to be carried out after each kind of repair, if the device or one component was dropped or damaged or if the medical device was not used according to its intended use.
21
9. Troubleshooting list
9.1 Malfunctions 9.2 Error messages
Malfunction Remedy
Colors look wrong Perform a white balance of the DSM head in the
software settings.
3D looks inverted Switch the 3D polarity in the software settings.
Image is upside­down
Hard to see 3D Position the display so that it is directly in front
DSM head not detected
Light, zoom, or focus not working
Display not recog­nized
Spots on the display If the displayed image develops spots that do not
Robotic arm not working
Touchscreen not working
Rotate the DSM head orientation in the software settings.
of the surgeon. Some displays have a small ideal viewing angle. Adjust the height and tilt of the display for the best view.
If the software reports that the DSM head is not connected, check the DSM head cable connection at the EPU and at the DSM head. Unplug the cable connection at the DSM head and plug it back in to check if the software is operating correctly now. Ensure that the DSM head indicator is green.
Check that the DSM head power cable is con­nected. Ensure that the DSM head indicator is green. Restart the system. If issues remain, contact Aesculap Technical Ser­vice, see Chapter 10.
If a display is turned on after the EPU has started, it may not recognize the display. Check that the display cable is securely plugged into the back of the display. Restart the EPU.
move as the viewing target is moved, the DSM head optics may have dust on the viewing sur­faces. Clean the optics and screen. If the spots remain, the dust may be inside the DSM head. Contact Aesculap Technical Service, see Chapter 10.
Reboot the system or follow the error messages shown on the display.
Re-connect the touchscreen.
9.2.1 Startup errors
Error message Remedy
Already initialized Restart the system.
DSM Head Manager failed
Cursor Manager failed
DSM failed
Event Manager failed
Message System failed
Core functions failed
Patient Manager failed
Playback Manager failed
Procedure Manager failed
Recording Manager failed
Storage Manager failed
Surgeon Manager failed
QAB settings failed
Video Capture Manager failed
Widget Manager failed
Invalid init parameters
Failed to run
9.2.2 License error
Error message Remedy
DSM license not found. Please contact Aesculap for support.
9.2.3 Recording errors
Error message Remedy
Recording failed. Not enough disk space avail­able.
Erasing recordings less than [3] seconds.
Insufficient space on external drive. Internal drive has approx. [time] remaining.
If that does not help, contact the local Aesculap representative or Aesculap Technical Service, see Chapter 10.
Contact the local Aesculap representative or Aesculap Technical Service, see Chapter 10.
Download any videos from the system that should be kept and delete any extra files to free up space on the computer.
Any recordings shorter than three seconds are automatically deleted. Create a longer recording.
Download any videos from the system that should be kept and delete any extra files to free up space on the connected drive. Recordings will only save locally.
22
[Drive volume name] drive connected. Insuffi­cient recording space.
Download any videos from the system that should be kept and delete any extra files to free up space on the connected drive.
9.2.4 DSM head errors
Error message Remedy
DSM head is too hot. Maximum white light reduced to 50%.
Connect DSM head. The system is not connected to the DSM
Possible image degrada­tion detected.
The DSM head is overheating. The maximum white light setting has been capped at 50%. Turn the system off and allow it to cool when safe.
head. Connect the DSM head while the sys­tem is off. Wait 30 seconds and restart. If the connection is failing, contact Aesculap Technical Service, see Chapter 10.
Frame rate from the DSM head is not accept­able. Check DSM head cable connections. Contact Aesculap Technical Service, see Chapter 10.
10. Technical Service
Risk of death for patient and operator due to mal­function and/or failure of safety measures!
WARNING
CAUTION
Service addresses
Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1601 Fax: +49 (7461) 16-2887 E-Mail: ats@aesculap.de Or in the US: Aesculap Inc. Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 USA Aesculap Repair Hotline Phone: +1 (800) 214 -3392 Fax: +1 (314) 895 -4420 Other service addresses can be obtained from the address indicated above.
Do not perform any service or maintenance
while using the product on the patient.
Do not modify the product.
Modifications carried out on medical technical equipment may result in loss of guarantee/war­ranty rights and forfeiture of applicable licenses!
For service and repair, contact the local
B. Braun/Aesculap representative.
11. Technical data
11.1 Performance data, information about standards
Weight (unpacked) 454 lbs. / 206 kg
Weight (in crate)
Dimensions (unpacked, L × W × H)
Dimensions (in crate, L × W × H)
Power input
EMC
Conforming to standard IEC/DIN EN 60601-1
600 lbs. / 272 kg
32.5” × 48.3” × 76.3” / 826 mm × 1 227 mm × 1 938 mm
39.7” × 48” × 71.6” / 1 008 mm × 1 219 mm × 1 819 mm
100/120/240 V AC, 850 W, 50 Hz / 60 Hz
IEC/DIN EN 60601-1-2
11.2 Ambient conditions
Operation Storage and transport
Temperature 10°C to 35°C -10°C to 50°C
Relative humidity
Atmospheric pressure
30 % to 90 % non-condensing
700 hPa to 1 060 hPa 500 hPa to 1 060 hPa
10 % to 90 % non-condensing
12. Disposal
Note
The user institution is obliged to process the product before its disposal, see Chapter 7.
Adhere to national regulations when disposing of or recy­cling the product, its components and its packaging! The recycling pass can be downloaded from the Extranet as a PDF document under the respective article number. (The recycling pass includes disassembling instructions for the product, as well as information for proper disposal of components harmful to the environment.) Products carrying this symbol are subject to separate col­lection of electrical and electronic devices. Within the European Union, disposal is taken care of by the manufac­turer as a free-of-charge service.
Detailed information concerning the disposal of the product is avail-
able through the national B. Braun/Aesculap agency, see Chapter 10.
13. Distributor in the US/Contact in Canada for product information and complaints
Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA
23
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com
Aesculap
®
– a B. Braun brand
SOP-AIC-5001887 (TA015615) 2019-02
Loading...