The ANRITSU product(s) listed on the title page is (are) warranted
against defects in materials and workmanship for one year from the
date of shipment.
ANRITSU’s obligation covers repairing or replacing products which
prove to be defective during the warranty period. Buyers shall prepay
transportation charges for equipment returned to ANRITSU for warranty repairs. Obligation is limited to the original purchaser. ANRITSU
is not liable for consequential damages.
LIMITATION OF WARRANTY
The foregoing warranty does not apply to ANRITSU connectors that
have failed due to normal wear. Also, the warranty does not apply to
defects resulting from improper or inadequate maintenance by the
Buyer, unauthorized modification or misuse, or operation outside the
environmental specifications of the product. No other warranty is expressed or implied, and the remedies provided herein are the Buyer’s
sole and exclusive remedies.
TRADEMARK ACKNOWLEDGEMENTS
MS-DOS, Windows, and Windows for Workgroups are registered
trademarks of the Microsoft Corporation.
Site Master, Site Master 110, Site Master S110, Site Master S111, Site
Master S112, Site Master S113, Site Master 330, Site Master S330, and
Site Master S331 are trademarks of ANRITSU Company.
NOTICE
ANRITSU Company has prepared this manual for use by ANRITSU
Company personnel and customers as a guide for the proper installation, operation and maintenance of ANRITSU Company equipment
and computer programs. The drawings, specifications, and information contained herein are the property of ANRITSU Company, and
any unauthorized use or disclosure of these drawings, specifications,
and information is prohibited; they shall not be reproduced, copied, or
used in whole or in part as the basis for manufacture or sale of the
equipment or software programs without the prior written consent of
ANRITSU Company.
Table of Contents
General Information ..........1-1
Introduction..................1-1
Description ..................1-1
Standard Accessories.............1-2
Options ....................1-3
Optional Accessories .............1-3
Spares .....................1-4
Performance Specifications .........1-5
Operation ................2-1
Introduction..................2-1
Control Descriptions .............2-1
Test Panel ........................2-1
Keypad .........................2-2
Soft Keys ........................2-6
Remote Operation...................2-20
Operating Procedures............2-20
Making Common Function Selections . . . 2-21
Applying Power ....................2-21
Setting Options ....................2-22
Setting Scale/Limits..................2-22
Determining Remaining Battery Life ........2-23
Making Frequency-Domain Measurements2-24
Return Loss or SWR Measurement ....2-24
Selecting a Frequency Range.............2-24
Performing a Calibration...............2-24
Making a Measurement ...............2-26
Scaling the Display ..................2-27
October 199710580-00009
Copyright 1996, ANRITSU Co.Revision: B
The operation of the Site Master ™ is straightforward and intuitive.
However, you may find it helpful to review the operation of the keys
and menus prior to first-time use.
Descriptions of the keys and menus, along with measurement procedures, are provided in Chapter 2.
First-time users and maintenance supervisors will benefit from perusing
the material in Chapter 1. This chapter describes the instrument and
provides listings of options and performance specifications.
Please Recycle
This product contains a rechargeable battery. This battery may be recyclable. For environmental reasons, you should dispose of
the battery in a proper manner at the end of
equipment life. If you do not have access to
proper disposal methods, return the battery
to your ANRITSU service center.
ANRITSU will dispose of the unit at no
charge. ANRITSU service centers are listed
in Table 2-4 (page 2-50).
v
Interface
Serial
12V
External
Charging
Battery
(45C mA)
12-15VDC
Site Master S112
START
CAL
SAVE
SETUP
LIMIT
SAVE
DISPLAY
ON
OFF
User's Guide
Port
Test
Det
RF
AUTO
SCALE
1
2
ESCAPE
CLEAR
CAL
3
4
RECALL
SETUP
5
6
MARKER
7
8
ENTER
RECALL
DISPLAY
9
0
RUN
HOLD
Figure 1-1.
1-0
Site Master System
Chapter 1
General Information
Introduction
This chapter provides description, specification, and optional accessories for the Site Master Series instruments. This series has two members, as shown below. Throughout this manual, the term Site Master
will refer to the series; whereas, the terms Site Master S112, S113 will
refer to the applicable individual models.
Model
S1125 to 1000 MHz
S1135 to 1200 MHz
Frequency Range
Description
The Site Master (Figure 1-1) is a hand held SWR/RL(standing wave
ratio/return loss) and Distance-To-Fault measurement instrument that
includes a built-in synthesized signal source and an optional power
monitor. It uses a keypad to enter data and a liquid crystal display
(LCD) to provide a graphical indication of SWR or RL over the selected frequency range. The Site Master S113 has a built-in distanceto-fault capability. The Site Master S112, as well as model S113, allows measurement data to be converted to Fault Location via the companion Software Tools program. The Site Master is capable of up to
two hours of continuous operation from a fully charged internal battery.
It can also be operated from a 12.5 dc source (which will also simulta-
1
1-1
Chapter 1 General Information
neously charge the battery). Built-in energy conservation features can
be used to extend battery life over an eight-hour work day.
The Site Master is designed for measuring SWR, return loss, or cable
insertion loss and locating faulty RF components in antenna systems.
Power monitoring capability is available as an option. The displayed
trace can be scaled and/or enhanced with settable frequency markers
and/or a limit line. A menu option provides for an audible “beep” when
the limit value is exceeded. To permit use in low-light environments,
the LCD can be back lit using a front panel key.
Standard Accessories
A PC based software program (called Software Tools) provides an online database record for storing measurement data. Site Master Software Tools can also convert the Site Master display to a Microsoft
Windows 3.x graphic. Measurements stored in the Site Master internal
memory are down-loaded to the PC using the included serial cable.
This null-modem serial cable connects between the Serial Interface
connector on the Site Master and a Com Port on a DOS/Windowsbased PC. Once stored, the graphic trace can then be displayed, scaled,
and/or enhanced with markers and limit lines. Historical graphs can be
overlaid with current data by using the PC’s mouse in “drag-n-drop”
fashion. The underlying data can be extracted and used in spreadsheets
or for other analytical tasks.
The Software Tools program also performs DTF (Distance To Fault) or
Fault Location by clicking on the appropriate icon.
1-2
Chapter 1 General Information
The following items are supplied with the basic hardware.
Soft carrying case
•
Standard Calibration Components
•
AC-DC adapter
•
Automotive Cigarette Lighter 12 Volt DC Adapter
•
3 1/2-inch floppy disk containing the Software Tools program.
•
This program contains Fault Location (DTF) and Smith Chart
functions.
Serial Interface Cable, Part Number 800-441
•
One year Warranty (includes battery, firmware, and software)
•
•User’s Guide
•Programming Manual
Options
•
Option 1 — Switch Standard Calibration Components to
Precision N type components
•
Option 5 — Add RF Wattmeter Power Monitor
Optional Accessories
•
ANRITSU precision N type Short/Open, Part No. 22N50
•
Site Master precision N Load, 42 dB Part No. SM/PL
•
Test Port Extension Cable, 1.5 meter, Part No. TP/ECN 1.5
•
Test Port Extension Cable, 3.0 meter, Part No. TP/ECN 3.0
1-3
Chapter 1 General Information
Test Port Extension Cable, 5.0 meter, Part No. TP/ECN 5.0
•
Test Port Extension Cable, 1.5 meter, Part No. TP/EC 1.5
•
Test Port Extension Cable, 3.0 meter, Part No. TP/EC 3.0
•
Test Port Extension Cable, 5.0 meter, Part No. TP/EC 5.0
Part No. 5400-71N50
Transit Case for Site Master, Part No. 760-194A
•
HP Deskjet 340 Printer, Part No. 2000-766
•
Serial-to-Parallel Converter Cable (use with the
•
HP 340 Printer), Part No. 2000-753
•Seiko DPU-411 Thermal Printer, Part No. 2000-754 or
2000-761
•Serial Interface Cable (use with the DPU-411 Printer),
Part No. 2000-756
•
Thermal Paper (use with the DPU-411 Printer),
Part No. 2000-755
Spares
•
N type Standard Short, Part No. SM/STS
•
N type Standard Load, 35 dB, Part No. SM/STL
•
Soft Carrying Case, Part No. D41955
•
AC/DC Adapter, Part No. 40-97
•
Serial Interface Cable, Part Number 800-441
1-4
Chapter 1 General Information
Performance Specifications
Performance specifications are provided in Table 1-1.
Table 1-1.Performance Specifications (1 of 2)
Specifications are valid when the unit is calibrated at ambient temperature after a 5 minute warmup.
DescriptionValue
Frequency Range:
Site Master S112
Site Master S113
Frequency Accuracy (CW Mode)75 parts per million @25°C*
Frequency Resolution10 kHz
SWR:
Range
Resolution
Return Loss:
Range
Resolution
Cable Insertion Loss:
Range
Resolution
**Distance-To-Fault (DTF):
Range
Resolution (in meters)
(Rectangular Windowing)
5 to 1000 MHz
5 to 1200 MHz
1.00 to 65.00
0.01
0.0 to 54.00 dB
0.01 dB
0.0 to 20.00 dB
0.01 dB
0 to (Resoution x 129)
(.)( )15 108´V
Frequency
D
Where
propagation velocity.
p
V
is the cable’s relative
p
1-5
Chapter 1 General Information
Table 1-1.Performance Specifications (2 of 2)
Wattmeter Power Monitor:
Range
–50.0 to +20 dBm
10.0 nW to 100.0 mW
Offset Range
Resolution
0 to +60.0 dB
0.1 dB
or
0.1 xW
Test Port, Type N50 Ohms
***Immunity to Interfering signals
up to the level of+10 dBm
Maximum Input (Damage Level):
Test Port, Type N
RF Detector
+22 dBm
+20 dBm
Measurement Accuracy:
Measurement accuracy depends on calibration components. Standard calibration components have a directivity of 35 dB. Precision
calibration components have a directivity of 42 dB.
Temperature:
Storage
Operation
–20° C to 75° C
0° C to 50° C
Weight:2.2 pounds
1
Size:8x7x2
inches
4
or
* ±2 ppm/D°C from 25°C
** Fault location is accomplished by inverse Fourier Transformation of data taken with
the Site Master. Resolution and maximum range depend on the number of frequency
data points, frequency sweep range and relative propagation velocity of the cable being
tested.
*** Immunity measurement is made in CW mode with incoming intefering signal exactly
at the same frequency (worst case situation). Typical immunity is better when swept frequency is used.
1-6
Chapter 1 General Information
This page is intentionally blank
1-7
w
Battery
Charging
Back Lighting
START
CAL
SAVE
SETUP
Soft Keys
Serial
Interface
LIMIT
SAVEDISPLAY
Interface
Serial
External
Power
Battery
ON/OFF
12.5-15V DC
12V
External
Charging
(45C mA)
12-15V DC
Site Master S112
Test Port
RF
Detector
Input
Det
AUTOSCALE
ESCAPE
CLEAR
CAL
Up/Down Arro
RECALL
SETUP
MAKER
ENTER
RUN
HOLD
Port
Test
RF
AUTO
SCALE
1
START
CAL
2
ESCAPE
CLEAR
CAL
3
SAVE
SETUP
4
RECALL
SETUP
5
LIMIT
6
MARKER
7
SAVE
DISPLAY
8
ENTER
RECALL
DISPLAY
9
0
RUN
ON
HOLD
OFF
PRINT
Figure 2-1.
2-0
PRINT
Site Master Controls and Connectors
RECALL DISPLAY
Chapter 2
Operation
Introduction
This chapter provides a description of each control and describes how
to calibrate the Site Master and make a measurement.
Control Descriptions
Control descriptions are given below; the test panel controls and connectors are listed first. The keypad controls follow and are listed alphabetically. Then, the soft keys and menu structure are described using
Figure 2-3 (page 2-7).
Test Panel
2
12.5-15VDC
(600 mA)
Battery
Charging
External
Power
Serial
Interface
Provides input for battery charging the unit. Input is
12.5 to 15 Vdc @ 600 mA.
Indicator light to show that the battery is being
charged.
Indicator light to show that the Site Master is being
powered by the external charging unit.
Provides an RS232 DB9 interface with a Com Port
on a personal computer (for use with the ANRITSU
Software Tools program). Also provides an interface
2-1
Chapter 2 Operation
to a HP Deskjet 340 printer or a Seiko DPU-411
Thermal printer.
Test
Port
RF DetProvides RF detector input for the Power Monitor.
Provides RF output, 50W impedance.
Keypad
Turns the liquid crystal display (LCD) back-lighting
ON or OFF. (Leaving back lighting off conserves
battery power.)
AUTO
SCALE
CALOpens a calibration menu. Use Up/Down Arrow key
ENTERImplements certain menu and key selections.
ESCAPE
CLEAR
LIMITCalls up the Scale Menu.
Automatically scales the display for optimum resolution.
and ENTER key to select a stored calibration (A or
B) or turn CAL off.
Exits the present operation and/or clears the display.
If a parameter is being edited, pressing this key will
clear the value currently being entered and restore the
last valid entry. Pressing this key again will close the
parameter. During normal sweeping, pressing this
key will move up one menu level.
MARKERCalls up the Marker Menu.
ON
OFF
Turns the Site Master on or off. When turned on, the
system state at the last turn-off is restored. If the
2-2
Chapter 2 Operation
ESCAPE key is held down, the factory preset state is
restored.
PRINTPrints the current display to the selected printer.
RECALL
DISPLAY
RECALL
SETUP
RUN
HOLD
SAVE
DISPLAY
Recalls a previously saved trace from memory location 1 through 40. When the key is pressed, “Recall
display:” appears on the display. Select an appropriate number from the keypad and press the ENTER
key to implement.
Recalls a previously saved setup from memory location 0 through 9. When the key is pressed, “Recall
Setup:” appears on the display. Select an appropriate
number from the keypad and press the ENTER key
to implement. Setup 0 recalls the factory preset state.
When in the Hold mode, this key starts the Site Mas-ter sweeping and provides a Single Sweep Mode
trigger; when in the Run mode, it pauses the sweep.
When in the Hold mode, the hold symbol (Table 2-1,
page 2-45) appears on the left side of the LCD.
(HOLD conserves considerable battery power.)
Saves the displayed trace to 1 of 40 internal nonvolatile memory locations. When the key is pressed,
“Save display:” appears on the display. Select an appropriate number from the keypad and press the
ENTER key to implement.
CAUTION:The selected memory location will be overwritten by the SAVE DISPLAY operation. No warning
is given.
2-3
Chapter 2 Operation
SAVE
SETUP
Saves the current system setup to 1 of 9 internal
non-volatile memory locations. When the key is
pressed, “Save Setup:” appears on the display. Select
an appropriate number from the keypad and press the
ENTER key to implement.
CAUTION:The selected memory location will be overwritten by the SAVE SETUP operation. No warning is
given.
START
CAL
Opens a calibration menu. Use the Up/Down Arrow
key and ENTER key to select the location to store
the calibration (A or B) or cancel the command. The
existing calibration in the location chosen will be
overwritten once the calibration is performed.
Follow the text in the message area that instructs you
to do the following:
q
Connect OPEN, Press ENTER
The Site Master then measures the “open” test
port (standard calibration), or the calibration
“open” (precision calibration kit) that you must attach to the end of the test port or transmission line.
2-4
q
Connect SHORT, Press ENTER
The Site Master then measures the calibration
“short” that you must attach to the end of the test
port or transmission line.
Chapter 2 Operation
Connect LOAD, Press ENTER
q
The Site Master then measures the 50W termination (load) that you must attach to the end of the
test port or transmission line.
NOTES:
The order of the “Open,” “Short,” and “Load” measurements can be changed or the selection repeated
using the Up/Down Arrow Key.
The combined measurements of an open, a short,
and a known-impedance load normalizes the measurement system, to account for uncertainties introduced by measurement-system components (e.g.,
cables, connectors, etc.).
Up/Down
Arrow Key
Increments or decrements a parameter value.
2-5
Chapter 2 Operation
Display Area
Message Area
FREQ
DIST
SCALE
OPT
POWER
START
CAL
SAVE
SETUP
LIMIT
SAVE
DISPLAY
ON
OFF
AUTO
SCALE
ESCAPE
CAL
RECALL
SETUP
MARKER
RECALL
DISPLAY
PRINT
CLEAR
2
4
6
ENTER
8
RUN
HOLD
0
1
3
5
7
9
.
Soft Key
Labels
Soft Keys
Figure 2-2.Site Master Front Panel Display Layout
Soft Keys
Each of the soft keys has a corresponding soft key label area on the display. The label identifies the function of the soft key for the current
menu selection. Figure 2-3, on the following page, shows the menu
structure, the soft key labels for each menu selection, and the page
where a description of the menu can be found.
2-6
MAIN MENU
Chapter 2 Operation
FREQUENCY MENUDISTANCE MENU
MARKER MENU
DISTANCE SUB-MENU
MENU TITLE
MAIN MENU
FREQUENCY MENU
DISTANCE MENU
DISTANCE SUB-MENU
MARKER MENU
SCALE MENU
OPTION MENU
OPTION SUB-MENU
POWER MENU
FREQ
F1
D1
LOSS
M1M2
TOP
B1
B5
UNITSREL
SCALE MENU
SOFT KEY LABELS
DIST
F2
D2
PROP V
M1PEAK
BOTTOM
B2
B6
OFFSET
Figure 2-3.Site Master Menu Structure
SCALE
MKRS
MKRS
CABLE
LIMIT
B3
B7
OPTION MENU
OPTION SUB-MENU
POWER
OPT
MAIN
MORE
B4
B8
ZERO
MAIN
MAIN
MAIN
MORE
MAIN
MAIN
WINDW
M2PEAK
POWER MENU
PAGE
2-8
2-9
2-11
2-12
2-10
2-16
2-17
2-18
2-19
2-7
Chapter 2 Operation
Main Menu At turn on, the Main Menu soft keys, below, are dis-
played, and the Marker status is shown in the message area.
FREQ
DIST
SCALE
OPT
POWER
These soft keys provide the following menu selections:
FREQ — Selects the Frequency Menu, described
q
on page 2-9.
DIST — Selects the Distance Menu, described on
q
page 2-11.
q SCALE —Selects the Scale Menu, described on
page 2-16.
q OPT — Selects the Option Menu, descibed on
page 2-17.
q
POWER — Selects the Power Menu, described
on page 2-19.
2-8
Chapter 2 Operation
Frequency
Menu
(FREQ)
Provides for setting sweep frequency end-points.
Also provides access to the Marker Menu. Selected
frequency values may be changed using the keypad
or Up/Down Arrow key.
Choosing FREQ causes the soft keys, below, to be
displayed and the corresponding values to be shown
in the message area.
F1
q F1 — Opens the F1 parameter for data entry. This
is the start value for the frequency sweep. Press
ENTER when data entry is complete.
q F2 — Opens the F2 parameter for data entry. This
is the stop value for the frequency sweep. Press
ENTER when data entry is complete.
q MKRS — Selects the Markers Menu, described
on page 2-10.
q
MAIN — Returns to the Main Menu.
F2
MKRS
MAIN
2-9
Chapter 2 Operation
Marker
Menu
(MKRS)
Provides for setting marker values. Selected frequency or distance values may be changed using the
keypad or Up/Down Arrow key.
Choosing MKRS causes the soft keys, below, to be
displayed and the corresponding values to be shown
in the message area.
Pressing the MARKER key will also call up this
menu.
M1
M1 — Turns M1 marker OFF, if it is currently
q
M2
M1PEAK
M2PEAK
MAIN
ON. If M1 is currently OFF, turns it ON and opens
the M1 marker parameter for data entry, displaying either frequency or distance and the corresponding SWR, RL, or CL. Press ENTER when
data entry is complete.
q M2 — Turns M2 marker OFF, if it is currently
ON. If M2 is currently OFF, turns it ON and opens
the M2 marker parameter for data entry, displaying either frequency or distance and the corresponding SWR, RL, or CL. Press ENTER when
data entry is complete.
2-10
q
M1PEAK — Places the M1 marker at the frequency or distance with the peak SWR, RL, or CL
value.
q
M2PEAK — Places the M2 marker at the frequency or distance with the peak SWR, RL, or CL
value.
MAIN — Returns to the Main Menu.
q
Chapter 2 Operation
Distance
Menu
(DIST)
Provides for setting Distance to Fault parameters.
Selected distance values may be changed using the
keypad or Up/Down Arrow key. Entry can be in feet
or meters, depending on the setting of the B5 soft key
in the Option Sub-Menu (page 2-18).
Choosing DIST causes the soft keys, below, to be
D1
D2
MKRS
MORE
displayed and the corresponding values to be shown
in the message area.
q D1 — Opens the start distance (D1) parameter for
data entry. This is the start value for the distance
range. Press ENTER when data entry is complete.
q D2 — Opens the end distance (D2) parameter for
data entry. This is the end value for the distance
range. Press ENTER when data entry is complete.
q
MKRS — Selects the Marker Menu, described on
page 2-10.
q
MORE — Selects the Distance Sub-Menu, described on page 2-12.
2-11
Chapter 2 Operation
NOTE: Refer to the table on the inside of the manual
back cover for a listing of common coaxial cables
showing values for “Dielectric,”“Relative Velocity,”
“Dielectric Constant,” and “Nominal Attenuation in
dB/m@ 1000 MHz”.
Distance
Sub-Menu
Provides for setting the cable loss and relative propagation velocity of the coaxial cable. Selected values
may be changed using the keypad or Up/Down Arrow key.
LOSS
Choosing MORE causes the soft keys, below, to be
displayed and the corresponding values to be shown
in the message area.
q LOSS — Opens the Cable Loss parameter for data
entry. Enter the loss per foot (or meter) for the
type of transmission line being tested. Press EN-TER when data entry is complete. (Range is 0.000
to 5.000 dB/m)
q
PROP V (relative propagation velocity) — Opens
the Propagation Velocity parameter for data entry.
Enter the propagation velocity for the type of
transmission line being tested. Press ENTER
when data entry is complete. (Range is 0.010 to
1.000)
q
CABLE — Opens a menu of common coaxial cables. Use the up/down arrow key and ENTER key
PROP V
CABLE
WINDW
MAIN
2-12
Chapter 2 Operation
to make a selection. This feature is provided as a
rapid means of setting both cable loss and propagation velocity.
WINDW — Opens a menu of FFT windowing
q
types for the DTF calculation. Use the Up/Down
Arrow key and ENTER key to make a selection.
NOTE: Using Windowing
The theoretical requirement for inverse FFT is for the
data to extend from zero frequency to infinity. Side
lobes appear around a discontinuity due to the fact
that the spectrum is cut off at a finite frequency. Windowing reduces the side lobes by smoothing out the
sharp transitions at the beginning and end of the frequency sweep. As the side lobes are reduced the
main lobe widens thereby reducing the resolution.
In situations where there may be a small discontinuity
close to a large one, side lobe reduction Windowing
should be used. When distance resolution is critical
Windowing can be reduced. The types of Windowing
in order of increasing side lobe reduction are: rectangular, nominal side lobe, low side lobe, minimum side
lobe. Figures 2-4 thru 2-7, on pages 2-14 and 2-15,
are examples of the types of Windowing.
q
MAIN — Returns to the Main Menu.
2-13
Chapter 2 Operation
-10
-15
-20
-25
Return Loss (dB)
-30
-35
-40
-45
-50
Distance To Fault
-5
10
5
25
20
15
30 35 40
Feet
Figure 2-4.Rectangular Windowing Example
Distance To Fault
-5
-10
-15
-20
-25
Return Loss (dB)
-30
-35
-40
-45
-50
51015202530
Feet
35
Figure 2-5.Nominal Side Lobe Windowing Example
2-14
45 50
55 60
55
45 50
40
60
Return Loss (dB)
Chapter 2 Operation
Distance To Fault
-5
-10
-15
-20
-25
-30
-35
-40
-45
-50
510
15
354045
Feet
50 55
20 25 30
Figure 2-6.Low Side Lobe Windowing Example
Distance To Fault
-5
-10
-15
-20
-25
Return Loss (dB)
-30
-35
-40
-45
-50
5
20 25
15
10
30 35 40
Feet
45
50
Figure 2-7.Minimum Side Lobe Windowing Example
60
55 60
2-15
Chapter 2 Operation
Scale Menu
(SCALE)
Provides for changing the display scale. Selected
values may be changed using the keypad or
Up/Down Arrow key.
Choosing SCALE causes the soft keys, below, to be
displayed and the corresponding values to be shown
in the message area.
Pressing the LIMIT key on the keypad will also call
up this menu.
TOP
TOP — Opens the TOP parameter for data entry
q
BOTTOM
LIMIT
MAIN
and provides for setting the top scale value. Press
ENTER when data entry is complete.
q BOTTOM — Opens the BOTTOM parameter for
data entry and provides for setting the bottom
scale value. Press ENTER when data entry is
complete.
q
LIMIT — Turns Limit OFF, if currently ON. If
Limit is currently OFF, turns it ON and opens the
Limit parameter for data entry. Press ENTER
when data entry is complete.
2-16
q
MAIN — Returns to the Main Menu.
Chapter 2 Operation
Option
Menu
(OPT)
Provides for selecting Site Master options.
Choosing OPT causes the soft keys, below, to be dis-
played and the corresponding values to be shown in
the message area.
B1
B1 MODE — Opens a menu of measurement
q
B2
B3
B4
MORE
modes; Return Loss (RL), Standing Wave Ratio
(SWR), and Cable Loss(CL). Use the Up/Down
Arrow key and ENTER key to make a selection.
B2 LIMIT BEEP ON/OFF — Toggles the limit
q
beeping sound on or off. When on, the Site Master sounds a beep when the measured value is
above the limit line.
q B3 KEYBD LOCK ON/OFF — Toggles the key-
board locking feature on or off. When on, the keyboard (except for the keys listed below) is locked
to prevent inadvertant data entry.
When locked, pressing any key (except ON/OFF,RUN/HOLD, MORE, MAIN, OPT, and B3) will
cause the lock-out symbol (Table 2-1, page 2-45)
to display along the left edge of the LCD and an
error beep to sound.
q
B4 SINGLE SWP ON/ OFF — Toggles the single sweep function on or off. When on, the SiteMaster will have to be manually triggered using
the RUN/HOLD key on the keypad for each
sweep.
2-17
Chapter 2 Operation
q
MORE — Selects the Option Sub-Menu, described below.
Option
Sub-Menu
Provides for selecting additional Site Master options.
Choosing MORE causes the soft keys, below, to be
displayed and the corresponding values to be shown
in the message area.
B5
B5 UNITS — Toggles between ENGLISH and
q
B6
B7
B8
MAIN
METRIC units.
q B6 FIXED CW — Toggles the fixed CW function
on or off. When on, if F1 is set equal to F2, a CW
frequency is fixed at the selected value. When off,
a narrow band of frequencies, centered on the selected frequency, is generated.
q B7 PRINTER — Displays a menu of supported
printers—None, Seiko DPU-411, and HP Deskjet
340. Use the up/down arrow key and ENTER key
to make the selection.
q
B8 CONTRAST — Enables adjustment of the
LCD contrast. Use the up/down arrow key and
ENTER key to set the contrast.
2-18
q
MAIN — Returns to the Main Menu.
Chapter 2 Operation
Power
Menu
Provides for setting power monitor modes and parameters.
Choosing POWER causes the soft keys, below, to be
displayed and the corresponding values to be shown
in the message area.
UNITS
UNITS — Toggles between dBm and Watts.
q
REL — Turns relative mode OFF , if currently
q
REL
OFFSET
OFFSET
ZERO
MAIN
ON. If relative mode is currently OFF, turns it ON
and causes the power level to be measured and
saved as the base level. Subsequent measurements
are then displayed relative to this saved value.
With units of dBm, relative mode displays dBr;
with units of Watts, relative mode displays %
(percent).
q OFFSET — Turns Offset OFF, if currently ON.
If Offset is currently OFF, turns it ON and opens
the Offset parameter for data entry. Press ENTER
when data entry is complete.
Offset is the attenuation (in dB) inserted in the line
between the DUT and the RF detector. The attenuation is added to the measured input level
prior to display.
q
ZERO — Turns Zero OFF, if currently ON. If
Zero is currently OFF, turns it ON and initiates
collection of a series of power level samples,
which are averaged and saved. This saved value is
2-19
Chapter 2 Operation
then subtracted from subsequent measurements
prior to display.
MAIN — Returns to the Main Menu.
q
Remote Operation
All Site Master functions, settings, and operating modes (except for
power on/off) are controllable using commands sent from an external
computer via the serial port. A detailed description of the programming
commands is available in the Site Master Programming Manual, P/N
10580-00012.
Operating Procedures
Site Master operation can be divided into four functional areas: common functions, frequency domain, distance domain, and power monitor.
•Common functions are options and scale functions that are
common to both frequency- and distance-domain measurements.
•
Frequency domain measurements consist of Return Loss (RL),
Standing Wave Ratio (SWR), and Cable Loss (CL) made over
a selectable frequency range. RL and SWR characterize the
magnitude of reflections present in a device or transmission
line. CL measurements display the insertion loss of a transmission line over frequency.
•
Distance domain measurements—commonly known as
distance-to-fault (DTF)—are made over a selectable distance
range. They include RL or SWR, but they also return informa-
2-20
Chapter 2 Operation
tion that can help locate discontinuities in a transmission line.
CL cannot be measured in the distance domain.
Power monitor measurements can be either absolute or rela-
•
tive to some base power level, and can be displayed in either
dBm or Watts. To allow measurement and display of power
levels above the unit’s specified input, the user may attenuate
the signal and enter the corresponding offset. The offset is
added to the measured input power prior to display. Finally, a
zero adjust function is provided to allow substraction of an inherent “noise” level prior to display.
CAUTION: The measurement system MUST be calibrated at the ambient temperature prior to making a
measurement. You must calibrate when the setup frequency is changed. The instrument can store up to
two calibrations. These stored calibrations are independent of any stored setups. When you recall a
stored setup, perform a new calibration or recall a
stored calibration.
Making Common Function Selections
Common functions are options and scale functions that are common to
both frequency- and distance-domain measurements.
Applying Power
Step 1.
Turn the Site Master on using the ON/OFF key.
2-21
Chapter 2 Operation
Step 2.
Observe that the Site Master opening screen appears
and that it performs a self test. The firmware version
and model number are shown in the opening screen.
Setting Options
Step 3.
Step 4.
Press the OPT soft key, from the Main Menu
(page 2-8).
Accept or change the settings of the B1 through B8
options. Refer to pages 2-17 and 2-18 for a description of each option.
Setting Scale/Limits
Step 5.
Step 6.Accept or change the settings. To change, press the
Press the SCALE soft key, from the Main Menu.
applicable TOP, BOTTOM, or LIMIT soft key and
either:
q Enter a numeric value from the keypad and press
the ENTER key.
q
Or press the Up/Down Arrow key and press the
ENTER key.
2-22
Chapter 2 Operation
Determining Remaining Battery Life
A symbol that denotes the charge state of the battery is continuously
displayed in the top-right corner of the display (Figure 2-8). A totally
black bar indicates a fully charged battery.
Remaining Battery Life
Sliding Gauge Symbol
0
CAL A
LIM ON
30
900.01100.0FREQ (MHz)
M1 = 30.75 dB AT 1000 MHz
M2 = 21.10 dB AT 1050 MHz
FREQ
DIST
Figure 2-8.Battery Monitor
RETURN LOSS
SCALE
OPT
POWER
2-23
Chapter 2 Operation
Making Frequency-Domain Measurements
Frequency domain measurements consist of Return Loss (RL), Standing Wave Ratio (SWR), or Cable Loss (CL) measurements made over a
selectable frequency range.
Return Loss or SWR Measurement
Selecting a Frequency Range
Step 1.
Step 2.
Step 3.Enter the desired numerical value using the keypad or
Step 4.
Step 5.Enter the desired numerical value using the keypad or
Step 6.Check that the FREQ (MHz) scale in the display area
Press the FREQ soft key, from the Main Menu
(page 2-8).
Press the F1 soft key from the Frequency Menu (page
2-9).
the Up/Down Arrow key (Figure 2-1, page 2-0).
Press ENTER when data entry is complete.
Press the F2 soft key.
the Up/Down Arrow key. Press ENTER when data
entry is complete.
indicates the new frequency start and stop values.
Performing a Calibration
Step 7.Perform a measurement calibration, using the
START CAL key. Select CAL A or CAL B.
Follow the ensuing “Connect OPEN, Press EN-
TER,” “Connect SHORT, Press ENTER,” and
2-24
Chapter 2 Operation
“Connect LOAD, Press ENTER” instructions that
appear in the message area. Connect the respective
Open, Short, and Load component to the end of the
Test Port Extension Cable (Figure 2-9) prior to pressing ENTER.
After each selection, one of the messages “Measur-
ing OPEN,” “Measuring SHORT,” or “Measuring
LOAD” appears while the measurement is in prog-
ress.
DUT
Antenna
Interface
Serial
or Antenna Feed
12V
External
Charging
Battery
(45C mA)
12-15VDC
Site Master S112
TestPort Extension Cable
(Optional)
Standard TestPort
Port
Test
AUTO
SCALE
1
START
CAL
2
ESCAPE
CLEAR
CAL
3
SAVE
SETUP
4
RECALL
SETUP
5
LIMIT
6
MARKER
7
SAVE
DISPLAY
8
ENTER
RECALL
DISPLAY
9
0
RUN
ON
HOLD
OFF
Det
RF
Open
Short
Load
Figure 2-9.Measurement/Calibration Test Setup
2-25
Chapter 2 Operation
If, during a calibration, you wish to repeat a measurement or change the order in which the Open,
Short, and Load are connected, press Up/Down Arrow key.
NOTES:
For best calibration results — compensation for all
measurement system uncertainties — ensure that the
Open/Short/Load is at the end of the test port or optional extension cable; that is, at the same point that
you will connect the antenna or device to be tested.
For best results, use a phase stable Test Port Extension Cable (see Optional Accessories). If you use a
typical laboratory cable to extend the
port to the device-under-test, cable bending subsequent to the Open/ Short/Load calibration will cause
uncompensated phase reflections inside the cable.
Thus, cables which are NOT phase stable may cause
measurement errors that are more pronounced as the
test frequency increases.
Site Master
test
For optimum calibration, ANRITSU recommends using
precision calibration components.
Making a Measurement
Step 8.
Step 9.Observe the displayed waveform.
Connect the device-under-test to the Site Master
Test Port or the Test Port Extension Cable.
2-26
Chapter 2 Operation
Scaling the Display
Step 10.The display can be scaled using either of the follow-
ing methods:
Automatically scale the display using the AUTO
q
SCALE key (Figure 2-1, page 2-0).
Manually scale the display using the SCALE soft
q
key—from the Main Menu—and the TOP and
BOTTOM soft keys from the ensuing soft key
menu (page 2-16).
Adjusting Markers
Step 11.
Step 12.
Step 13.
Step 14.
Press the MARKER key (Figure 2-1, page 2-0) to
call up the Marker Menu.
If M1 needs to be turned OFF, press the M1 soft key
on the Markers Menu. If M1 needs to be turned ON,
press the M1 soft key again. This will open the M1
parameter for entry. If needed, enter the desired numerical value using the keypad or the Up/Down Arrow key (Figure 2-1, page 2-0). Press ENTER when
the data entry is complete.
If M2 needs to be turned OFF, press the M2 soft key
on the Markers Menu. If M2 needs to be turned ON,
press the M2 soft key again. This will open the M2
parameter for entry. If needed, enter the desired numerical value using the keypad or the Up/Down Arrow key (Figure 2-1, page 2-0). Press ENTER when
the data entry is complete.
Press the MAIN soft key to return to the Main Menu.
2-27
Chapter 2 Operation
Adjusting a Limit
Step 15.
Step 16.
Step 17.If you wish to have an audible beep when the trace is
Step 18.
Press the LIMIT key (Figure 2-1, page 2-0) to call up
the Scale Menu.
If the Limit needs to be turned OFF, press the LIMIT
soft key on the Scale Menu. If the Limit needs to be
turned ON, press the LIMIT soft key again. This will
open the Limit parameter for entry. If needed, enter
the desired numerical value using the keypad or the
Up/Down Arrow key (Figure 2-1, page 2-0). Press
ENTER when the data entry is complete.
above the limit line, press the OPT soft key from the
Main Menu (page 2-8).
Press the B2 soft key from the Option Menu (page
2-17) to toggle the limit beep indicator ON.
Saving a Setup
Step 19.
Step 20.Enter the desired numerical value using the keypad (1
Press the SAVE SETUP key (Figure 2-1, page 2-0).
to 9) or the Up/Down Arrow key. Press ENTER
when data entry is complete.
Recalling a Setup
Step 21.
Step 22.Enter the desired numerical value using the keypad (1
Press the RECALL SETUP key.
to 9) or the Up/Down Arrow key. Press ENTER
2-28
when data entry is complete. To restore the factory
defaults, enter 0, then press ENTER.
Storing a Display
Chapter 2 Operation
Step 23.
Step 24.Enter the desired numerical value using the keypad (1
Press the SAVE DISPLAY key.
to 40) or the Up/Down Arrow key. Press ENTER
when data entry is complete.
Recalling a Display
Step 25.
Step 26.Enter the desired numerical value using the keypad (1
Press the RECALL DISPLAY key.
to 40) or the Up/Down Arrow key. Press ENTER
when data entry is complete.
2-29
Chapter 2 Operation
Cable Loss Measurement
Selecting a Frequency Range
Step 1.
Step 2.
Step 3.Enter the desired numerical value using the keypad or
Step 4.
Step 5.Enter the desired numerical value using the keypad or
Step 6.Check that the FREQ (MHz) scale in the display area
Press the FREQ soft key, from the Main Menu
(page 2-8).
Press the F1 soft key from the Frequency Menu (page
2-9).
the Up/Down Arrow key (Figure 2-1, page 2-0).
Press ENTER when data entry is complete.
Press the F2 soft key.
the Up/Down Arrow key. Press ENTER when data
entry is complete.
indicates the new frequency start and stop values.
Performing a Calibration
Step 7.Perform a measurement calibration, using the
START CAL key. Select CAL A or CAL B. Follow
the ensuing “Connect OPEN, Press ENTER,”
“Connect SHORT, Press ENTER,” and “Connect
LOAD, Press ENTER” instructions that appear in
the message area. Connect the respective Open,
Short, and Load component to the end of the Test
Port Extension Cable (Figure 2-9, page 2-25) prior to
pressing ENTER.
2-30
Chapter 2 Operation
After each selection, one of the messages “Measuring OPEN,” “Measuring SHORT,” or “Measuring
LOAD” appears while the measurement is in prog-
ress.
If, during a calibration, you wish to repeat a measurement or change the order in which the Open,
Short, and Load are connected, press Up/Down Arrow key.
NOTES:
For best calibration results — compensation for all
measurement system uncertainties — ensure that the
Open/Short/Load is at the end of the test port or optional extension cable; that is, at the same point that
you will connect the antenna or device to be tested.
For best results, use a phase stable Test Port Extension Cable (see Optional Accessories). If you use a
typical laboratory cable to extend the
port to the device-under-test, cable bending subsequent to the Open/ Short/Load calibration will cause
uncompensated phase reflections inside the cable.
Thus, cables which are NOT phase stable may cause
measurement errors that are more pronounced as the
test frequency increases.
Site Master
test
For optimum calibration, ANRITSU recommends using
precision calibration components.
2-31
Chapter 2 Operation
Making a Measurement
Step 8.
Step 9.Place an Open or Short at the other end of the cable.
Step 10.Observe the displayed cable loss vs frequency.
Connect the cable to be tested to the Site Master
Test Port or the end of the Test Port Extension Cable
(if used).
Scaling the Display
Step 11.The display can be scaled using either of the follow-
ing methods:
Automatically scale the display using the AUTO
q
SCALE key (Figure 2-1, page 2-0).
q Manually scale the display using the SCALE soft
key—from the Main Menu—and the TOP and
BOTTOM soft keys from the ensuing soft key
menu (page 2-16).
Adjusting Markers
Step 12.
Step 13.
Press the MARKER key (Figure 2-1, page 2-0) to
call up the Marker Menu.
If M1 needs to be turned OFF, press the M1 soft key
on the Markers Menu. If M1 needs to be turned ON,
press the M1 soft key again. This will open the M1
parameter for entry. If needed, enter the desired numerical value using the keypad or the Up/Down Arrow key (Figure 2-1, page 2-0). Press ENTER when
the data entry is complete.
2-32
Chapter 2 Operation
Step 14.
Step 15.
If M2 needs to be turned OFF, press the M2 soft key
on the Markers Menu. If M2 needs to be turned ON,
press the M2 soft key again. This will open the M2
parameter for entry. If needed, enter the desired numerical value using the keypad or the Up/Down Arrow key (Figure 2-1, page 2-0). Press ENTER when
the data entry is complete.
Press the MAIN soft key to return to the Main Menu.
Adjusting a Limit
Step 16.
Step 17.
Step 18.If you wish to have an audible beep when the trace is
Press the LIMIT key (Figure 2-1, page 2-0) to call up
the Scale Menu.
If the Limit needs to be turned OFF, press the LIMIT
soft key on the Scale Menu. If the Limit needs to be
turned ON, press the LIMIT soft key again. This will
open the Limit parameter for entry. If needed, enter
the desired numerical value using the keypad or the
Up/Down Arrow key (Figure 2-1, page 2-0). Press
ENTER when the data entry is complete.
above the limit line, press the OPT soft key from the
Main Menu (page 2-8).
Step 19.
Press the B2 soft key from the Option Menu (page
2-17) to toggle the limit beep indicator ON.
Saving a Setup
Step 20.
Press the SAVE SETUP key (Figure 2-1, page 2-0).
2-33
Chapter 2 Operation
Step 21.Enter the desired numerical value using the keypad (1
to 9) or the Up/Down Arrow key. Press ENTER
when data entry is complete.
Recalling a Setup
Step 22.
Step 23.Enter the desired numerical value using the keypad (1
Press the RECALL SETUP key.
to 9) or the Up/Down Arrow key. Press ENTER
when data entry is complete. To restore the factory
defaults, enter 0, then press ENTER.
Storing a Display
Step 24.
Step 25.Enter the desired numerical value using the keypad (1
Press the SAVE DISPLAY key.
to 40) or the Up/Down Arrow key. Press ENTER
when data entry is complete.
Recalling a Display
Step 26.
Step 27.Enter the desired numerical value using the keypad (1
Press the RECALL DISPLAY key.
to 40) or the Up/Down Arrow key. Press ENTER
when data entry is complete.
2-34
Chapter 2 Operation
Making Distance-Domain Measurements
Distance domain measurements—commonly known as distance-tofault (DTF)—are made over a selectable distance range. They return information that can help locate discontinuities in a transmission line.
Selecting a Frequency Range
NOTE:
The maximum distance range is determined by the
frequency span, number of data points, and relative
propagation velocity:
Maximum Unaliased Range
Where: F1 is start frequency
F2 is stop frequency
Vf is relative propagation velocity
maximum unaliased range is in meters.
Refer to Figure 2-10, page 2-36, for a plot of maximum
distance range versus frequency span.
=
8
´-(.)()( )15 10129
FF
21
V
f
Step 1.
Press the FREQ soft key, from the Main Menu (page
2-8).
Step 2.
Press the F1 soft key from the ensuing soft key menu.
Step 3.Enter the desired numerical value using the keypad or
the Up/Down Arrow key (Figure 2-1, page 2-0).
Press ENTER when data entry is complete.
Step 4.
Press the F2 soft key.
2-35
Chapter 2 Operation
Figure 2-10.Maximum Distance and Resolution vs Frequency Span
2-36
Chapter 2 Operation
Step 5.Enter the desired numerical value using the keypad or
the Up/Down Arrow key. Press ENTER when data
entry is complete.
Step 6.Check that the FREQ (MHz) scale in the display area
indicates the new frequency start and stop values.
Performing a Calibration
Step 7.Perform a measurement calibration, using the
START CAL key. Select CAL A or CAL B. Follow
the ensuing “Connect OPEN, Press ENTER,”
“Connect SHORT, Press ENTER,” and “Connect
LOAD, Press ENTER” instructions that appear in
the message area. Connect the respective Open,
Short, and Load component to the end of the Test
Port Extension Cable (Figure 2-9, page 2-25) prior to
pressing ENTER.
After each selection, one of the messages “Measur-
ing OPEN,” “Measuring SHORT,” and “Measur-
ing LOAD” appears while the measurement is in
progress.
If, during a calibration, you wish to repeat a meas-
urement or change the order in which the Open,
Short, and Load are connected, press Up/Down ar-
row key.
2-37
Chapter 2 Operation
NOTES:
For best calibration results — compensation for all
measurement system uncertainties — ensure that the
Open/Short/Load is at the end of the test port or optional extension cable; that is, at the same point that
you will connect the antenna or device to be tested.
For best results, use a phase stable Test Port Extension Cable (see Optional Accessories). If you use a
typical laboratory cable to extend the
port to the device-under- test, cable bending subsequent to the Open/ Short/Load calibration will cause
uncompensated phase reflections inside the cable.
Thus, cables which are NOT phase stable may cause
measurement errors that are more pronounced as the
test frequency increases.
For optimum calibration, ANRITSU recommends using
precision calibration components.
Site Master
Performing a DTF Measurement
test
Step 8.
Step 9.
Step 10.Enter the desired numerical value using the keypad or
Step 11.
Press the DIST soft key, from the Main Menu (page
2-8).
Press the D1 soft key from the ensuing soft key menu
(page 2-11).
the Up/Down Arrow key (Figure 2-1, page 2-0).
Press ENTER when data entry is complete.
Repeat the above two steps for D2.
2-38
Chapter 2 Operation
Step 12.
Step 13.
Step 14.Enter the desired numerical value using the keypad or
Step 15.
Step 16.
Step 17.
Step 18.
Step 19.
Step 20.Enter the desired numerical values using the keypad
Press the MKRS soft key to go to the Markers Menu
(page 2-10).
Press the M1 soft key.
the Up/Down Arrow key. Press ENTER when data
entry is complete.
Repeat the above two steps for M2.
Press the MAIN soft key to return to the Main Menu.
Press the DIST soft key to go to the Distance Menu.
Press the MORE soft key to go to the Distance Sub-
Menu.
Press the CABLE soft key to select a coaxial cable
from the menu or press the LOSS and PROP V soft
keys to enter cable loss and relative propagation ve-
locity parameters manually (page 2-12). The loss of
the cable shown in the cable menu is @1000 MHz.
or the Up/Down Arrow key. Press ENTER when
data entry is complete.
Step 21.
Step 22.After calculations are complete (approximately 8 sec-
Press the MAIN soft key to return to the Main Menu.
onds), observe that the message area displays the re-
turn loss data at the two marker points.
2-39
Chapter 2 Operation
NOTES:
Press theB5soft key from the Option Sub-Menu
(page 2-18) to toggle between feet and meters. Values entered in either will freely convert to the other.
Loss and relative propagation velocity values for
many common cable types are listed in the table on
the inside of the manual back cover.
Making Power Measurements
Power measurement is accomplished using a broadband (1 MHz to
3 GHz) RF detector, P/N 5400-71N50. The power monitor displays the
measured power in dBm or Watts.
Entering Power Monitor Mode
Step 1.
Press the POWER soft key, from the Main Menu
(page 2-8)
Zeroing the Power Monitor
Step 2.
With no power applied to the DUT, press the ZERO
soft key from the Power menu (page 2-19). Wait a
few seconds while the Site Master accumulates sam-
ples of the quiescent power level. When complete,
ZERO ADJ: ON is displayed in the message area.
2-40
Chapter 2 Operation
Measuring High Input Power Levels
Step 3.Insert an attenuator between the DUT and the RF de-
tector, sufficient to insure that the input power to the
Site Master is no greater than 20 dBm.
Step 4.
Step 5.Enter the attenuation in dB using the keypad. Press
Press the OFFSET soft key.
ENTER to complete the entry. The message area will
show OFFSET is ON along with the entered value in
dB.
Displaying Power in dBm and Watts
Step 6.
Press the UNITS soft key to display power in Watts.
Displaying Relative Power
Step 7.
Step 8.
With the desired base power level input to the Site
Master, press the REL soft key. The message area
will show REL: ON and the power reading will indi-
cate 100%.
Press the UNITS soft key to display power in dBm.
Since REL is ON, the power reading will be in dBr,
relative to the base power level.
2-41
Chapter 2 Operation
Printing
Printing is accomplished with either of two printers: the Seiko DPU411 thermal printer or the Hewlett Packard DeskJet 340 ink jet printer.
Figure 2-11 shows a setup diagram for these two printers.
Printer Switch Settings
Set the switches, SW1 and SW2, on the Seiko DPU-411 thermal printer
as follows:
Switch12345678
SW1OFFONONONONOFFONON
SW2ONONONOFFOFFOFF
Set the switches on the serial-to-parallel interface cable to the HP Deskjet 340 ink jet printer as follows:
SW1
OFFONOFFOFFOFFOFFOFFOFF
SW2SW3SW4SW5SW6SW7SW8
2-42
Chapter 2 Operation
Printing a Screen
Step 1.Connect the printer as shown in Figure 2-11.
Step 2.Obtain a SWR, RL, CL, or Distance-to-Fault
measurement display.
Step 3.
Step 4.
Select the printer using the B7 soft key from the Op-
tion Sub-Menu (page 2-18).
Press the PRINT key (Figure 2-1. page 2-0).
SEIKO
2000-754, or
2000-761
SERIAL CABLE
2000-756
HP 340
DESKJET
SERIAL-TO-PARALLEL
INTERFACE CABLE
2000-753
Figure 2-11.Printer Setup
2-43
Chapter 2 Operation
Symbols
Table 2-1 provides a listing of the symbols used as condition indicators
on the LCD display.
Self Test
At turn-on, the Site Master runs through a series of quick checks to ensure the system is functioning properly. Note that the battery voltage
and temperature are displayed in the lower left corner below the self
test message. If the battery is low, or if the ambient temperature is not
within the specified operational range, Self Test will fail. If Self Test
fails AND the battery is fully charged AND the Site Master is within
the specified operating range, call your ANRITSU Service Center.
Error Codes
Self Test Errors
A listing of Self Test Error messages is given in Table 2-2.
Range Errors
A listing of Range Error messages is given in Table 2-3.
Replacing the Battery
Replacing the battery is the only recommended field-level maintenance
action. If your battery fails, contact your ANRITSU Sales Office or
Service Center. Table 2-4 provides a listing of current service centers.
2-44
Table 2-2.LCD Symbols
IconSymbol
Chapter 2 Operation
Site Master is in Hold or Power Conservation
mode. To resume sweeping, press the
RUN/HOLD key. After 10 minutes without a
key press, the Site Master will automatically
enter into its power conservation mode.
Site Master is in keypad lockout mode. To turn
off keypad lockout, use the B3 soft key (page
2-17).
Lock fail indication. Check battery. (If SiteMaster fails to lock with a fully charged battery, call your ANRITSU Service Center.)
Processor timeout failure. Symbol appears at
the frequency that causes an input RF overload;
it then disappears as the sweep continues past
that point.
When calibration is performed, the Site Master
stores the ambient temperature. If the temperature drifts outside the specified range, this indicator will flash. A recalibration at the current
temperature is recommended.
Indicates the remaining charge on the battery.
The inner white rectangle grows longer as the
battery charge depletes.
2-45
Chapter 2 Operation
Table 2-3.Self Test Error Messages
Error Message
BATTERY LOW
Battery voltage is less than 12.5 volts.
Description
Charge battery. If condition persists, call
your ANRITSU Service Center.
EXTERNAL POWER
LOW
PLL FAILED
External supply voltage is less than 12.5
volts. Call your ANRITSU Service Center
Phase-locked loops failed to lock. Charge
battery. If condition persists with a fully
charged battery, call your ANRITSU Service Center
INTEGRATOR
FAILED
Integration circuit could not charge to a
valid level. Charge battery. If condition per-
sists with a fully charged battery, call your
ANRITSU Service Center.
EEPROM R/W
FAILED
OUT OF TEMP.
RANGE
Non-volatile memory system has failed.
Call your ANRITSU Service Center.
Ambient temperature is not within the
specified operating range. Return temperature to specified operating range. If condi-
tion persists, call your ANRITSU Service
Center.
Note: A listing of current ANRITSU service centers is given in Table 2-4.
2-46
Table 2-4.Range Error Messages (1 of 2)
Chapter 2 Operation
Error Message
RANGE ERROR:F1 >
F2
RANGE ERROR:D1 >
D2
RANGE ERROR:D2 >
DMax=xx.x ft (m)
RANGE ERROR:
TOP<=BOTTOM
RANGE ERROR:
TOP>=BOTTOM
CAL
INCOMPLETE
CAL VALID FROM:
xxxx.x to
xxxx.x MHz
Description
The start (F1) frequency is greater than the
stop (F2) frequency.
The start (D1) distance is greater than the
stop (D2) distance.
The stop distance (D2) exceeds the maximum unaliased range. This range is determined by the frequency span, number of
points, and relative propagation velocity:
Maximum Unaliased Range
Where: F1 is start frequency
F2 is stop frequency
Vfis relative propagation velocity
maximum unaliased range is in meters
Refer to Figure 2-10, page 2-35, for a plot of maximum
range versus frequency span.
=
8
´-(.)()( )1 5 10129
FF
21
V
f
The SWR scale parameter top value is less
than or equal to its bottom value.
The RL scale parameter top value is greater
than or equal to its bottom value.
A complete open, short, and load calibration
must be performed before calibration can be
turned on.
The calibration was performed at a frequency range that is different from the current range.
2-47
Chapter 2 Operation
Table 2-3.Range Error Messages (2 of 2)
Error Message
DIST REQUIRES
F1 < F2
DIST REQUIRES
CAL
NO STORED SWEEP
AT THIS LOCATION
Valid distance to fault plots require a nonzero frequency span.
Distance-to-fault measurements do not provide usable data with CAL OFF.
Attempting to recall a display from a location that has not been previously written to.
That is, the location does not contain stored
sweep.
USE OPTIONS
MENU TO SELECT
A PRINTER
DISTANCE AND
CABLE
INSERTION LOSS
MODE ARE
INCOMPATIBLE
CANNOT ZERO
NO DETECTOR
INSTALLED
CANNOT ZERO
INPUT SIGNAL
TOO HIGH
POWER MONITOR
OPTION NOT
INSTALLED
Attempting to print a display with no printer
selected. Select a printer, then retry.
DTF measurements only display RL or
SWR versus distance.
Attempting to perform a Power Monitor
zero adjust function with no RF detector
connected to the Site Master.
Attempting to perform a Power Monitor
zero adjust function with an input of greater
than –20 dBm.
Attempting to enter Power Monitor mode
with no option 5 installed.
Description
2-48
Chapter 2 Operation
Using the Soft Carrying Case
The soft carrying case has been designed such that the strap can be unsnapped to allow the case to be easily oriented horizontally; thus allowing the Site Master controls to be more easily accessed (Figure 2-11).
Figure 2-11.Using Soft Carrying Case
2-49
Chapter 2 Operation
Table 2-4.ANRITSU Service Centers (1 of 2)
UNITED STATES
ANRITSU COMPANY
685 Jarvis Drive
Morgan Hill, CA 95037-2809
Telephone: (408) 776-8300
FAX: 408-776-1744
ANRITSU BEIJING SERVICE CENTER
Beijing Fortune Building
416W, 5 Dong San Huan Bei Lu
Chaoyang qu
Beijing 100004, China
Telephone: 010-501-7559
FAX: 010-501-7558
FRANCE
ANRITSU S.A
9 Avenue du Quebec
Zone de Courtaboeuf
91951 Les Ulis Cedex
Telephone: 016-44-66-546
FAX: 016-44-61-065
ANRITSU LTD.
200 Capability Green
Luton, Bedfordshire
LU1 3LU, England
Telephone: 015-82-41-88-53
FAX: 015-82-31-303
2 - 51/2-52
Chapter 3
Software Tools
Program
Description
The Site Master Software Tools program provides the means for transferring the measured trace, along with any applied markers and/or a
limit, to the screen of an MS-DOS based personal computer (PC) running Windows, Windows for Workgroups 3.xx, Windows NT, or Windows 95.
Requirements
The Site Master Software Tools program is a standard Windows 3.x
program and will run on any computer that will run Windows, Windows for Workgroups 3.xx, Windows NT, or Windows 95. Typically,
this means having a PC with the following characteristics:
•
386 or better microprocessor running enhanced mode (386DX,
33 MHz or better, recommended).
3
•
8 MBytes of memory, minimum.
•
Hard Disk Drive, with approximately 3 MBytes of available
space.
3-1
Chapter 3 Screen Capture Program
Communication Port Setting
The Site Master Software Tools communicates with the Site Master
through a standard COM port on the PC. It is important that your Windows COM port settings conform to the actual hardware settings. Table
3-1 provides a listing of standard COM port settings for most IBM
AT-Compatible computers.
Table 3-1.Standard COM Port Settings
PortIRQAddress
COM143F8
COM232F8
COM343E8
COM432E8
Since various add-in devices such as sound cards, modems, and network cards use IRQ (Interrupts), it is possible that your computer has
non-standard COM port settings. Please consult your computer vendor
for COM port address and IRQ information.
Changing COM Port Settings–Windows 3.1
Refer to Figure 3-1 while performing the following procedure.
Step 1.
Step 2.
Step 3.Click on the icon for the COM Port whose communi-
Open the Windows Control Panel.
Double click on the Ports icon. The Ports dialog box
appears.
cations settings you wish to specify, then click the
Settings... button. The Settings dialog box appears.
3-2
Chapter 3 Screen Capture Program
Figure 3-1.Windows 3.1 COM Port Setting Dialog Boxes
Step 4.Change to the following settings if necessary:
Baud Rate:9600
Data Bits:8
Parity:None
Stop Bits:1
Flow Control: None
Step 5.
Step 6.
Click on the Advanced... button. At the Advanced
Settings dialog box verify the COM Port Base Address and IRQ. Click OK when done.
Click OK again and then Close to close the COM
Port Setting Windows.
3-3
Chapter 3 Screen Capture Program
Note:
If you changed the COM Port Base Address and/or
IRQ, you will need to restart Windows.
Figure 3-2.Windows 95 COM Port Setting Dialog Boxes
Changing COM Port Settings–Windows 95
Refer to Figure 3-2 while performing the following procedure.
Step 1.
Step 2.
Step 3.
Open the Windows Control Panel.
Double click on the SYSTEM icon. The System
Properties window appears.
Select Device Manager. The Device List appears.
3-4
Chapter 3 Screen Capture Program
Step 4.
Step 5.
NOTE:
If Windows doesn’t show any available COM Ports,
consult your computer manufacturer.
Step 6.
Step 7.
Step 8.
Double click on the item Ports (COM & LPT) in the
device list.
Double click on the Communications Port you want
to set. The Communications Port Properties window
appears.
Choose Port Settings, then change to the following
settings if necessary.
Baud Rate:9600
Data Bits:8
Parity:None
Stop Bits:1
Flow Control: None
Choose Resource and verify the COM Port Base Address and IRQ. Click OK when done.
Click OK again to close the System Properties window.
Note:
If you changed the COM Port Base Address and/or
IRQ, you will need to restart Windows.
3-5
Chapter 3 Screen Capture Program
Software Installation
The Site Master Software Tools program is a conventional Microsoft
Windows 3.x program. Installation is similar to all other such programs. For users new to Windows 3.x, a detailed procedure is given
below.
Step 1.
Step 2.In the Windows Program Manager, pull down the
Step 3.Type A: (or B:)\SETUP, when the dialog box ap-
Step 4.Press the Enter key to select the default directory,
Step 5.Select “OK” or press the Enter key to exit the Setup
NOTE: The Setup routine will create a new Program
Manager Group named “
This group will contain three file icons, “Read Me,”
“Site Master Help,” and “Site Master Software Tools.”
Step 6.Double-click on the “Read Me” icon to read about (1)
Insert the Wiltron Site Master Software Tools For
Windows disk in floppy drive A:\ or B:\.
File menu and select Run.
pears.
C:\SITEMSTR, and begin the Setup routine.
routine, when its completes.
Site Master Software Tools
recent changes that did not get into this manual
and/or (2) important features or problems that you
should know about.
.”
3-6
Chapter 3 Screen Capture Program
Step 7.Double-click on the “Site Master Help” icon to ac-
quaint yourself with the comprehensive on-line manual. This manual provides descriptive narrative for
the various program features and controls.
Step 8.Double-click on the “Site Master Software Tools ”
icon to open the Software Tools program.
Plot Capture
Plots (traces) can be captured either singly from the Site Master display or in multiples from one or more stored-display locations. Both
methods are described below.
The recommended method is the Capture Multiple Traces option in
the Capture pull-down menu. Using this method, you can download to
the Software Tools program in one operation all of the data residing in
the up-to-40 Stored Display memory locations. The downloaded traces
will appear in cascade on the PC screen.
Multiple trace capture (preferred method)
Step 1.Connect the supplied cable as shown in Figure 3-3.
Step 2.
Step 3.
Open the “Site Master Software Tools” group, in
the Windows Program Manager.
(Select the Windows pull-down menu from the Pro-
gram Manager menu bar, and select Site Master
Software Tools.)
Double-click on the “Site Master Software Tools ”
icon to open the program.
3-7
Chapter 3 Screen Capture Program
Serial Interface
Com port
(Program defaults to Com2)
Figure 3-3.Equipment Setup for Site Master Tools Operation
3-8
Step 4.
Click on Capture, in the top menu bar, and select
Capture Multiple Traces from the drop-down
menu.
Step 5.Enter the number (or numbers) of the stored-display
memory location(s) (1 to 40) from which you wish to
display traces in Site Master Software Tools.
Step 6.Select “OK.”
Step 7.Observe that the “Acquiring Control” box appears on
the screen, then disappears as the traces are automatically acquired.
Chapter 3 Screen Capture Program
Single trace capture
Step 1.Perform steps 1, 2, and 3 of the multiple-trace-
capture procedure.
Step 2.
Step 3.Observe that a “Waiting for Data” box appears on the
Step 4.
Click on Capture, in the top menu bar, and select
Start Capture from the drop-down menu.
screen.
On the Site Master,
Capture a new trace and store it to a memory loca-
q
tion.
q Recall the stored plot to the screen.
q Press the ENTER key (on the Site Master) to start
the screen capture.
NOTE: The ENTER key on the Site Master must be
pressed while the PC screen shows “Waiting for
Data” for the transfer of information to be complete.
Step 5.On the PC, observe that the “Waiting for data” box
disappears and the trace capture process begins.
Within a short time, the Site Master trace will appear.
Step 6.If the trace differs from the Site Master display, it is
being autoscaled by the capture program. To turn off
autoscaling, proceed as follows:
q
Click on the Settings menu, in the top menu bar.
3-9
Chapter 3 Screen Capture Program
Click on Default Plot Settings, pull down menu.
q
Click on Per Site Master, in the dialog box, and
q
select OK to save changes.
On the Site Master, press the ENTER key to re-
q
send the plot.
Program Operation
The captured trace on the PC can be scaled and have its limit line,
markers, and properties changed. (Select Plot Properties under the
View menu to make these changes.) The operation of the various
menus that allow these operations to be accomplished is straightforward. To read about the operation of the menus, refer to the on-line
help screens. They can be accessed from the Help menu, in the top
menu bar.
3-10
Chapter 3 Screen Capture Program
Fault Location Software
A captured RL or SWR trace can be transformed to a Distance to Fault
display. This is useful for determining the location of faults, connections, and other discontinuities within the cable.
To transform a plot, select Fault Find from the Tools menu (or click
the toolbar button). A drop down menu will appear that asks you to
supply start and stop distances, along with dielectric constant or propagatiion velocity and insertion loss values. The table on the inside of the
manual back cover provides these values for some of the more popular
cables. The values are given at a nominal frequency of 1000 MHz. If
values are needed at a different frequency, or if you need the exact values for more accurate measurement, please contact the cable manufac-
turer. Cables may be added to the list be editing the
file and following the existing format. The distance may be entered in
feet or meters. Propagation velocity or dielectric constant may be used.
The units may be changed by selecting Default Plot Settings under the
Settings menu and changing “Distance Units” or “Cable Units.”
After supplying the needed information and clicking OK, a new plot
will open showing cable match vs. distance.
cables.lst
Smith Chart Software
After obtaining a frequency SWR or RL plot on the computer display,
click the Smith Chart icon. Read the Help file to see how this feature
functions (under Smith Chart and Data Readout).
3-11
Chapter 3 Screen Capture Program
Saving a Plot as a Windows Metafile
Plots can be saved as Windows Metafiles (.WMF). The metafile may
be imported into other graphic programs, but cannot be reloaded into
the Site Master Software Tools program.
To save a plot as a Windows Metafile, click on File, in the top menu
bar, and select Save as Metafile from the drop down menu.
Saving Data to a Spreadsheet
The data points from a plot can be exported to a spreadsheet via the
clipboard. To transfer data to a Windows spreadsheet program:
3-12
Step 1.
Step 2.Capture or load the desired plot.
Step 3.
Step 4.Open the spreadsheet program and place the cursor
Step 5.
Select Clipboard Format from the Settings menu
and choose Formatted Text.
Copy the data to the clipboard by selecting Copy
from the Edit menu.
where the first data point should appear.
Select Paste from the spreadsheet program’s Edit
menu.
Chapter 3 Screen Capture Program
“Drag-n-Drop”
Site Master Software Tools is Windows based. Graphs can “Drag-nDrop” onto each other. Site Master Software Tools allows quick comparison of “before” and “after” Distance-To-Fault measurements. Recent data is compared to a historical PC database record, which is
usually recorded during site installation/commissioning.
Each cable/antenna tends to have a unique Distance-To-Fault (DTF)
“Signature” because differing cable electrical lengths, cable types, dielectric thickness variations, and the positions of components
(connectors, adapters, and lightning arrestors) will cause different reflections at differing positions in the transmission line. Variations in the
“signature” between maintenance intervals offer a good indication of
damage or damage causing conditions.
Printing
Captured traces may be printed from a PC using Site Master Software
Tools. Once a captured trace has been downloaded choose Print under
the File menu for printing options. The printer setup can be altered,
plots can be scaled, and multiple plots can be printed from the Print
dialog box.