Anritsu S251C User Manual

MAINTENANCE MANUAL
Site Master™ Model S251C
Antenna and Cable Analyzer
Figure 1. Site Master Model S251C
1. INTRODUCTION
This manual provides maintenance instructions for the Site Master S251C Antenna and Cable Analyzer. It describes the product and provides performance verification procedures, parts replacement proce
dures, and a replaceable parts list.
2. DESCRIPTION
The Site Master (Figure 1) is a hand held S mission gain or loss), SWR/RL (standing wave ra
21
(trans
3. PERFORMANCE VERIFICATION
Paragraphs 4 through 10 contain tests that can be used to verify the performance of the Site Master model S251C having any version of firmware.
3.1. Initial Setup for Testing
1. Press and hold the press the
ter. (This sets the instrument to the factory
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
preset state.)
2. If necessary, use the Up/Down arrow key to adjust the contrast to give a readable display. Press
ENTER to accept.
490 JARVIS DRIVE ¨ MORGAN HILL, CA 95037-2809
P/N: 10580-00067
REVISION B
PRINTED: FEBRUARY 2003
COPYRIGHT 2001-2003 ANRITSU CO.
R F
S o u r c e
C o u p l e r
R F O u t
P o r t
R F I n
P o r t
I n c i d e n t
L o c a l
O s c i l l a t o r
D i s p l a y
Figure 2. Site Master Block Diagram
4. FREQUENCY ACCURACY
The following test can be used to verify the CW fre­quency accuracy of the Site Master. Measurement calibration of the Site Master is not required for this test.
a. Equipment Required:
·
Spectrum Analyzer
Anritsu Model MS2668C or equivalent
b. Procedure:
1. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
NOTE
Before continuing, allow a five minute warm up for the internal circuitry to stabilize.
2. Press the
3. Use the Up/Down Arrow key to highlight
SOURCE
4. Press the
MODE key.
RF
, then press ENTER.
FREQ soft key.
R e f l e c t e d
S y n c h r o n o u s
D e t e c t o r
A / D C o n v e r t e r
C P U
T r a n s m i t t e d
S e r i a l
P o r t
5. Using the keypad or Up/Down Arrow key en­ter 1000.0 MHz then press the
ENTER key.
6. Connect the RF cable from the Site Master RF Out Test Port to the RF Input on the Spec­trum Analyzer.
7. Set up the Spectrum Analyzer as follows: (a) Press
Preset
(b) Press the PRESET ALL soft key (c) Press (d) Press
(e) Press
Frequency and enter 1 GHz Span and enter 200 kHz Amplitude
(f) Press the REFERENCE LEVEL soft key and
enter 10 dB
8. If the Site Master has gone into the hold mode, press the
RUN/HOLD key to make the
measurement.
9. On the Spectrum Analyzer: (a) Press
Marker
(b) Press PEAK SEARCH. The frequency
should be 1000 MHz ±75 kHz.
2 Site Master S251C MM
5. TRANSMISSION/ISOLATION VERIFICATION
The following test can be used to verify transmission test port isolation and the accuracy of transmission measurements. Measurement calibration of the Site Master is required for this test.
TransTest Port
TestPort Extension Cable
Refl Test Port
a. Equipment Required:
10 dB Attenuator, Weinshel 1-10
·
30 dB Attenuator, Weinshel 1-30
·
Open/Short, Anritsu 22NF50
·
50 Ohm Terminations,
·
Anritsu SM/PL or 28N50-2 Anritsu SM/PLNF or 28NF50-2
Armored Test Port Extension Cable,
·
1.5 Meter, N(m) to N(m), Anritsu 15NN50-1.5A
b. Procedure:
1. Press and hold the press the
ON/OFF key to turn on the Site Mas-
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
NOTE
Before continuing, allow a five minute warm up for the internal circuitry to stabilize.
2. Press the
MODE key.
3. Use the Up/Down Arrow key to highlight
SERTION LOSS
4. Press the
5. Press the
, then press ENTER.
AMPLITUDE key. TOP soft key.
6. On the keypad, press “0”, then press (Verify the bottom limit is set to –120 dB.)
7. Press the
LIMIT key.
IN
ENTER.
Site Master S251C
MODE
FREQ/DIST
SWEEP
AMPLITUDE
Figure 3. OSL-THRU-ISOL Calibration Setup
12. Connect a Load (28N50-2 or SM/PL) to the RF In Test Port and verify that the noise floor (isolation) is below –80 dB.
13. Press the
14. Press the
MARKER key. M1 soft key, then the EDIT soft key.
15. Using the keypad or the Up/Down Arrow key, enter 1000.0 MHz, and press
NOTE
For the following steps in the procedure, use only attenuators that have Type N connectors. The use of attenuators with other type connectors and adapters will cause measurement errors.
16. Connect the 10 dB attenuator to the RF In
-
Test Port (refer to Figure 4) and verify that the reading is –10 dB ±2.25 dB @ 1000 MHz.
17. Connect the 10 dB and 30 dB attenuators in series to the RF In Test Port and verify that the reading is –40 dB ±3.25 dB @ 1000 MHz.
Attenuator
ESCAPE CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
RUN
SAVE
DISPLAY
DISPLAY
HOLD
+
9
0
-
/
PRINT
ON
SYS
OFF
.
ENTER.
TestPort Extension Cable
8. Press the
9. Press the the
10. Press the
11. Follow the instructions on the display to per form a OSL-THRU-ISOL calibration using a 22NF50 Open/Short, 28NF50-2 or SM/PLNF Terminations, and 15NN50-1.5ATest Port Ex tension Cable (refer to Figure 3).
SINGLE LIMIT soft key. EDIT soft key, enter –80, and press
ENTER key.
START CAL key.
Figure 4. Test Setup
TransTest Port
Refl Test Port
Site Master S251C
MODE
FREQ/DIST
AMPLITUDE
SWEEP
ESCAPE CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
RUN
SAVE
DISPLAY
DISPLAY
HOLD
+
9
0
-
/
PRINT
ON
SYS
OFF
.
Site Master S251C MM 3
6. RETURN LOSS VERIFICATION
The following test can be used to verify the accuracy of return loss measurements. Measurement calibra tion of the Site Master is required for this test.
a. Equipment Required:
5. Follow the instructions on the screen to per
form a calibration using a 22N50 Open/Short
and 28N50-2 or SM/PL Termination.
6. Connect the 20 dB offset to the RF Out Test Port and verify that the reading is 20 dB ±1.7 dB
20 dB offset, Anritsu SC5270
·
6 dB offset, Anritsu SC5237
·
Open/Short, Anritsu 22N50
·
50 Ohm Termination,
·
Anritsu 28N50-2 or SM/PL
b. Procedure:
1. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
NOTE
Before continuing, allow a five minute warm up for the internal circuitry to stabilize
2. Press the
MODE key.
3. Use the Up/Down Arrow key to highlight
TURN LOSS
4. Press the
, then press ENTER.
START CAL key.
RE-
7. Connect the 6 dB offset to the RF Out Test Port and verify that the reading is 6 dB ±1.2 dB
7. POWER MONITOR VERIFICATION
If the Power Monitor (Option 5) is installed in the Site Master, the following test can be used to verify the accuracy of the power measurements. Measure ment calibration of the Site Master is not required for this test.
a. Equipment Required:
RF Detector, 10 MHz to 20 GHz,
·
Anritsu 560-7N50B
· 10 dB Attenuator, Weinschel 1-10
· 30 dB Attenuator, Weinschel 1-30
· RF Reference Source, 0.050 GHz,
Anritsu MA2418A
· DC Power Supply, Anritsu 2000-933
Attenuators
560-7N50B
RF Detector
MA2418AReference Source
W
50MHz, 50 0dBm
Site Master S251C
MODE
FREQ/DIST
AMPLITUDE
SWEEP
ESCAPE CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
RUN
SAVE
DISPLAY
DISPLAY
HOLD
+
9
0
-
/
PRINT
ON
SYS
OFF
.
DC Power Supply 2000-933
20mA
15-24VDC
MA2418A Reference Source
Site Master w/ Option 5 Installed
Figure 5. Power Monitor Verification
4 Site Master S251C MM
b. Procedure
1. Connect the DC power supply to the MA2418A Reference Source. (Refer to Figure
5.)
2. Connect the MA2418A Reference Source to the input of the 560-7N50B RF detector.
3. Connect the RF Detector output to the RF De tector input of the Site Master.
(Option 10A only works with the AC-DC Adapter).
2. Apply AC power to the power supply.
3. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory
preset state.)
4. Press the
SYS key.
4. Connect the DC power supply to the appropri ate line voltage to supply power to the MA2418A Reference Source.
5. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
6. Press the
MODE key.
7. Use the Up/Down Arrow key to highlight
POWER MONITOR, then press ENTER.
8. Press the
ZERO soft key to zero the power
monitor. When complete,
ZERO ADJ:ON is displayed in
the message area.
9. Verify that the power monitor reading is
0.0 dBm ±1 dB.
10. Connect the output of the MA2418A Refer­ence Source to the two attenuators so as to add 40 dB of attenuation (Figure 5).
11. Connect the MA2418A Reference Source and the attenuators to the input of the 560-7N50B RF detector.
5. Press the
6. Press the
OPTIONS soft key. BIAS-T soft key to activate the Bias
Tee option.
7. Verify that the display shows
15.0V, 0 mA.
8. Connect the 22NF50 short to the RF In test port.
9. Verify the test port voltage relay clicks on and off, and
FAULT is shown in the upper left cor
ner of the display.
10. Remove the short and install the 100 Ohm load.
CAUTION
The 100 Ohm resistor will get hot. Ver­ify the readings and remove the 100 Ohm load immediately.
11. Verify that the display shows
BIAS, with a
reading of 14.8 to 15.3 Volts, and 120 to 160 milliamps.
9. BIAS T (OPTION 10B) VERIFICATION a. Equipment Required:
·
12. Verify that the power monitor reading is now –40.0 dBm ±2 dB.
8. BIAS T (OPTION 10A) VERIFICATION a. Equipment Required:
·
AC-DC Adapter Power Supply 40-115
·
Open/Short, Anritsu 22N50
·
100 Ohm, 1 Watt load (100 Ohm, 1 Watt re sistor soldered from the center pin to ground of an N-type connector.)
b. Procedure:
1. Connect the AC-DC Adapter Power Supply to the battery charging port on the Site Master
Open/Short, Anritsu 22N50
·
100 Ohm, 1 Watt load (100 Ohm, 1 watt resis tor soldered from the center pin to ground of a N-type connector)
·
Anritsu T2904
b. Procedure:
1. Ensure the battery is fully charged. (Option 10B only works with the battery).
2. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
3. Press the
4. Press the
SYS key. OPTIONS soft key.
Site Master S251C MM 5
5. Press the BIAS-T soft key to activate the Bias Tee option.
NOTE
When the Bias Tee option is selected, the unit is automatically forced into sin
gle sweep mode.
6. Press the
ON/OFF soft key.
7. Verify that 15.1V, 0 mA is shown in the upper left corner of the display.
10. TERMINATION VERIFICATION
This test verifies the accuracy of the Site Master SM/PL termination using the precision return loss mode of the 541XXA Scalar Measurement System. Measurements of terminations using this mode pro
a. Equipment Required:
8. Connect the 22N50 SHORT to the RF In test port.
9. Press the
RUN/HOLD key.
NOTE
The message “ displayed whenever the
Powering up TMA…” will be
RUN/HOLD key
is pressed.
10. Verify the test port voltage relay clicks on and off, and
FAULT is shown in the upper left cor
ner of the display.
11. Remove the 22N50.
12. Select LOW VOLTAGE (12V) and LOW CUR­RENT (460 mA(), 244 mA(–)) by pressing
LOW/HIGH VOLTAGE soft key.
the
13. Connect the 100 Ohm load to the
RF In test
port.
14. Press the
RUN/HOLD key and verify the read-
ing displayed in upper left corner is approximately 11.9V to 12.9V, 120 mA to 160 mA.
15. Remove the 100 Ohm load.
16. Select HIGH VOLTAGE (15V) and HIGH CURRENT (1A(), 650 mA(–)) by pressing the
LOW/HIGH VOLTAGE and LOW/HIGH CURRENT
soft keys.
17. Connect the T2904 to the
18. Press the
RUN/HOLD key and verify the read
ing displayed in upper left corner is approxi
RF In test port.
mately 14.8V to 15.3V, 343 mA to 379 mA.
Scalar Measurement System,
·
Anritsu 541XXA
Offset SWR Autotester,
·
Anritsu 560-97A50-20
Precision Airline, Anritsu 18N50
·
Open/Short, Anritsu 22N50
·
50 Ohm Termination, Anritsu 26N50
·
Source Adapter, Anritsu 34NN50A
·
b. Procedure
1. Connect the test equipment as shown in Fig­ure 6.
2. Press the
3. Press the
Power key on the 541XXA to On. System Menu key.
4. Using the Menu up-down keys: Highlight
SET
, then press the Select key.
5. At the up-down keys to highlight
DEFAULTS
RESET MENU display, use the Menu
RESET TO FACTORY
, then press the Select key.
6. Set the signal source for the frequency range as follows:
(a) Press the
Frequency key.
(b) Using the Data Entry Keypad or Data
Entry Knob, set the Start frequency to
0.01 GHz. Press the
Enter key.
(c) Using the Data Entry Keypad or Data
Entry Knob, set the Stop frequency to
4.0 GHz. Press the
Enter key.
RE-
7. Press the Channel 2 Display On/Off key to Off.
8. Press the Channel 1 Menu key.
9. Using the Menu up-down keys: Highlight
PRECISION RL, then press the Select key.
6 Site Master S251C MM
10. At the PRECISION RETURN LOSS menu display, use the Menu up-down keys to high light
FINAL, then press the Select key.
11. Press the
12. At the CALIBRATION menu display, use the Menu up-down keys to highlight then press the
13. At the PRECISION RETURN LOSS CALI BRATION menu display prompt, connect the Offset SWR Autotester to Input A, if you have not done so yet.
14. Connect the precision air line to the Offset SWR Autotester test port. Position the air line pointing vertically upward. Downward or horizontal positions make connector pin alignment difficult.
Calibration key.
START CAL,
Select key.
16. At the PRECISION RETURN LOSS CALI
BRATION menu prompt, connect the Open to the beadless end of the airline. Press the
lect
key to start the calibration.
17. Verify that the display resembles that shown in Figure 8.
CAUTION
During both calibration and measure ment, be sure to properly align the beadless connector of the airline. When the connectors are mis-aligned, a spike will usually be visible on the display.
18. At the next menu prompt, remove the Open and connect the Short to the beadless end of the airline. Press the calibration process.
Select key to start the
Se
-
NOTE
Ensure that the beadless end of the pre cision airline is at the measurement con nection point.
15. Press the
Select key when ready.
50 Ohm
Termination
541XXA Scalar Measurement System
Open/Short
Precision
Air Line
19. At the next menu prompt, remove the Short and connect the 50 Ohm Termination to the beadless end of the air line. Press the key to start the calibration process.
20. When the calibration is complete, remove the 50 Ohm Termination.
Select
Termination
under Test
Beadless End
Connect Dashed Line Connections When Directed By Procedure
B
A
Source Adapter
Figure 6. 541XXA Precision Return Loss Setup
Offset SWR
Autotester
Site Master S251C MM 7
2: OFF
10.0 dB/DIV OFFSET1: PRECSN RL (A)
2: OFF
+0.0dB
2 GHz/DIV
START: 0.0100 GHz 401 pts
STOP: 20.0000 GHz
LEVEL +7.0 dBm
54147A
CURSOR
1: -46.02 dB
PRESS SELECT
FOR
CURSOR MENU
10.0050 GHz
1
10.0 dB/DIV OFFSET1: PRECSN RL (A)
+0.0dB
54147A
PRECISION
RETURN LOSS
CALIBRATION
bles that shown in Figure 7.
22. Observe that the waveform displayed resem
1
START: 0.0100 GHz 401 pts
2 GHz/DIV
Figure 7. Example of a Good Connection
21. Connect the SM/PL termination to the beadless end of the air line and press the Se­lect key to begin the measurement.
STOP: 20.0000 GHz
LEVEL +7.0 dBm
CONNECT OPEN
TO AIRLINE
PRESS SELECT
WHEN READY
Figure 8. Direct Readout of the Precision Return Loss
23. Press the
Cursor On/Off key to On.
24. Observe the Cursor menu readout. The mini­mum return loss reading for the SM/PL ter­mination should be 42 dB.
8 Site Master S251C MM
11. BATTERY PACK REMOVAL AND REPLACEMENT
This procedure provides instructions for removing and replacing the Site Master battery pack.
1. With the Site Master standing upright on a stable surface, locate the battery access door (Figure 9).
BATTERYACCESS DOOR
Figure 9. Battery Access Door Location
Figure 11. Removing the Battery Access Door
2. Lift up the access door handle and rotate it 90 degrees counterclockwise, as illustrated in Figure 10.
Figure 10. Rotate the Battery Access Door Handle
3. Lift the door and remove, as illustrated in Figure 11.
4. Grasp the battery lanyard and pull the bat tery straight up and out of the unit, as illus trated in Figure 12.
Figure 12. Removing the Battery
5. Replacement is the opposite of removal. Note the orientation of the battery contacts, and be sure to insert the new battery with the contacts facing the rear of the unit (Figure
13).
BATTERYCONTACTS
Figure 13. Battery Orientation
Site Master S251C MM 9
12. BATTERY INFORMATION
The following information relates to the care and handling of the Site Master battery, and NiMH bat teries in general.
Figure 14. Site Master Battery
The Nickel Metal Hydride (NiMH) battery sup
·
plied with the Site Master is shipped in a dis charged state. Before using the Site Master, the internal battery must first be charged for three hours, either in the Site Master or in the op­tional battery charger (Anritsu part number: 2000-1029).
The battery may need to be replaced when the
·
operating time between charging becomes no
ticeably shorter than normal. Never use a damaged or worn out charger or
·
battery. Storing the battery in extreme hot or cold
·
places will reduce the capacity and lifetime of the battery.
Never short-circuit the battery terminals.
·
Do not drop, mutilate or attempt to disassemble
·
the battery. Do not dispose of batteries in a fire!
·
Batteries must be recycled or disposed of prop
·
erly. Do not place batteries in household gar bage.
Always use the battery for its intended purpose
·
only.
· Use only Anritsu approved battery packs.
· Recharge the battery only in the Site Master or
in an Anritsu approved charger.
· With a new NiMH battery, full performance is achieved after three to five complete charge and discharge cycles.
·
When the Site Master or the charger is not in use, disconnect it from the power source.
·
Do not charge batteries for longer than 24 hours; overcharging may shorten battery life.
·
If left unused a fully charged battery will dis charge itself over time.
·
Temperature extremes will affect the ability of the battery to charge: allow the battery to cool down or warm up as necessary before use or charging.
·
Discharge an NiMH battery from time to time to improve battery performance and battery life.
·
The battery can be charged and discharged hundreds of times, but it will eventually wear out.
10 Site Master S251C MM
13. FRONT PANEL ASSEMBLY REMOVAL AND REPLACEMENT
This procedure provides instructions for removing and replacing the Site Master front panel assembly. With the front panel assembly removed, the LCD display, keypad PCB, keypad membrane, and main PCB assemblies can be removed and replaced.
1. Place the Site Master face up on a work sur
face.
2. Remove the four rubber corner bumpers by carefully sliding the bumpers off of the case corners (Figure 15).
7. Carefully disconnect the LCD display backlight cable from J15 on the main PCB.
J15
J1
J12
Figure 16. Site Master Front Panel Cable Connections
8. Remove the front panel assembly.
9. Reverse the above steps to replace the front
panel assembly.
Figure 15. Removing the Corner Bumpers
3. With the bumpers removed, the access holes for the case screws are revealed. Use a Phil­lips screwdriver to remove the four screws securing the two halves of the Site Master case together.
4. Carefully lift up on the right side (as viewed from the front) of the front half of the case and begin to separate the two halves.
CAUTION
Do not force or pull the two halves of the case apart as there are delicate cables attached between the two halves that must be disconnected first.
5. Carefully depress the latch tab and discon nect the LCD display cable from J12 on the main PCB.
The corner bumpers only mount one way. That is, the raised area inside one end of the bumper (Figure 17) is made to conform to the contour of the front cover only.
Figure 17. Corner Bumper Detail
NOTE
RAISED AREA
6. Carefully disconnect the keypad interface ca
ble from J1 on the main PCB.
Site Master S251C MM 11
14. LCD ASSEMBLY REPLACEMENT
15. KEY PAD PCB REPLACEMENT
This procedure provides instructions for removing and replacing the Liquid Crystal Display (LCD) once the front panel assembly has been separated from the Site Master.
1. Remove the front panel assembly as directed in section 13.
2. Place the front panel assembly face down on a protected work surface.
3. Remove the 14 Phillips screws that attach the LCD retainer plate to the front panel as sembly.
4. Release the LCD display cable from the cable clamp on the LCD retainer plate.
LCD RETAINER PLATE LCD CABLE CLAMP
This procedure provides instructions for removing and replacing the key pad PCB.
1. Remove the front panel assembly as directed in section 13.
2. Place the front panel assembly face down on a protected work surface.
3. Remove the 14 Phillips screws that attach the backing plate to the front panel assem bly.
4. Release the LCD display cable from the re taining clip on the front panel backing plate (Figure 19).
5. Remove the front panel backing plate, care fully feeding the LCD cable through the ac cess hole to avoid damage to the cable or con nector.
6. Remove the rubber cushion pad from the key pad PCB and remove the PCB.
Figure 18. Front Panel Backing Plate
5. Remove the LCD retainer plate, carefully feeding the LCD cable through the access hole to avoid damage to the cable or connector.
6. Remove the rubber cushion pad from the LCD assembly and remove the assembly.
7. Reverse the above steps to install the re placement assembly.
KEYPAD PCB
Figure 19. Front Panel Keypad PCB Location
7. Reverse the above steps to install the re
placement assembly.
12 Site Master S251C MM
16. KEY PAD MEMBRANE REPLACEMENT
C
17. MAIN PCB ASSEMBLY REPLACEMENT
This procedure provides instructions for replacing the key pad membrane.
1. Remove the front panel assembly as directed in section 13.
2. Remove the key pad PCB as directed in sec tion 15.
3. Remove the keypad membrane by gently pulling the membrane up and out of the holes in the front panel.
KEYPAD MEMBRANE
This procedure provides instructions for replacing the main PCB assembly with the connector panel at tached.
1. Remove the front panel assembly as directed in section 13.
2. Disconnect the battery connector from J13 on the main PCB.
3. Remove the three PCB mounting screws and remove the PCB assembly with the connector panel attached.
BATTERY
ONNECTOR
Figure 20. Front Panel Keypad Membrane
4. Reverse the above steps to install the re­placement membrane.
PCB MOUNTING
SCREWS (3)
Figure 21. Main PCB
OPTION 10A
4. If Option 10A is installed (Figure 21) remove the option PCB by gently squeezing the standoff mounts to release the option PCB and disconnect it from the main board. Re move the standoffs and install them on the new main PCB, then install the Option 10A PCB on the new main PCB.
5. Reverse the above steps to install the new main PCB.
NOTE
The main PCB connector panel fits into grooves in the two halves of the Site Master case. Make sure the panel is cor rectly aligned with the grooves before reassembling the two halves together.
Site Master S251C MM 13
18. REPLACEABLE PARTS
Replaceable parts for the Site Master Model S251C are listed below.
Table 1. Replaceable Parts List
Part Number Description Qty
Accessories
10580-00065
10580-00066
2300-347 Software Tools, Site Master 1 40-115 Power Supply 1 2000-1029 Battery Charger 22N50 Precision Short/Open, N Male 1 SM/PL Connector, RF Termination 1 806-62 Cable Assy, Cig Plug, Female 1 800-441 Serial Interface Cable Assembly 1 48258 Soft Carrying Case 1
510-87 N-Connector 2 ND57371 Opt. 05 Input Connector w/cable 1 15-102 Liquid Crystal Display Assy 1 633-27 Rechargeable Battery, NiMH 1 ND57372 Main PCB Assembly,S251C 1
ND57373
ND57374
ND57375
ND60688
ND60690
52737-3 Keypad PCB Assembly 1 46649-1 Membrane Keypad, Main 1
User's Guide, Site Master S251C
Programming Manual, Site Mas ter S251C (available on disk only)
Replaceable Parts
Main PCB Assembly,S251C with Option 05
Main PCB Assembly,S251C with Option 10A
Main PCB Assembly,S251C with Option 05 and Option 10A
Main PCB Assembly,S251C with Option 10B
Main PCB Assembly,S251C with Option 05 and Option 10B
-
Part Number Description Qty
Hardware
1
1
1
1
900-160 Pan Head Screw, 4-40, 0.875 4 900-861 Pan Head Screw, 4-20, 0.365 15 900-869 Screw, 4-40, 0.875 4 900-720 Screw, 4-40, 0.187 3 900-754 Pan Head Screw, 4-40, 0.750 4 900-697 Screw, 4-40, 0.312 3 785-929 M-F Stand off, 4-40, 11/16 3 900-326 Kep Nut, 4-40, 0.187 8 790-516 Hole Plug, 0.6875L 1 790-42 Hole Plug, 0.625 1 761-79 Cap Vinyl, Black, round 1
Case Parts
46652-1 Top Case only 1 46653-1 Bottom Case only 1 48231-1 Battery Door 1 790-509
790-510 790-511
46655 Case Corner Bumpers 4 46662 LCD Retainer Plate 1 48241 Foam, LCD Corners 8 48278 Foam, LCD Window 1 46659 Foam, LCD Backing 1 46661 Foam, Keypad Backing 1 48246 Foam, Battery Door 1 48271 Foam, Battery Compartment 1 720-19 Cable Clamp 1 790-515 Spring, Battery Compartment 1 55217 ID Label, Model S251C 1
Battery Door Latch (3 pieces)
1
14 Site Master S251C MM
Table 2. Anritsu Service Centers
UNITED STATES
ANRITSU COMPANY 490 Jarvis Drive Morgan Hill, CA 95037-2809 Telephone: (408) 776-8300 FAX: 408-776-1744
ANRITSU COMPANY 10 NewMaple Ave., Suite 305 Pine Brook, NJ 07058 Telephone: 973-227-8999 FAX: 973-575-0092
ANRITSU COMPANY 1155E. Collins Blvd Richardson, TX 75081 Telephone: 1-800-ANRITSU FAX: 972-671-1877
AUSTRALIA
ANRITSU PTY. LTD. Unit 3, 170 Foster Road Mt Waverley, VIC 3149 Australia Telephone: 03-9558-8177 FAX: 03-9558-8255
BRAZIL
ANRITSU ELECTRONICA LTDA. Praia de Botafogo 440. Sala 2401 CEP22250-040,Rio de Janeiro,RJ, Brasil Telephone: 021-527-6922 FAX: 021-53-71-456
CANADA
ANRITSU INSTRUMENTS LTD. 700 Silver Seven Road, Suite 120 Kanata, Ontario K2V 1C3 Telephone: (613) 591-2003 FAX: (613) 591-1006
CHINA (SHANGHAI)
ANRITSU ELECTRONICS CO LTD 2F,Rm.B, 52 Section Factory Bldg. NO 516 Fu Te Road (N) Waigaoqiao Free Trade Zone Pudong, Shanghai 200131 PR CHINA Telephone: 86-21-58680226 FAX: 86-21-58680588
FRANCE
ANRITSU S.A 9 Avenue du Quebec Zone de Courtaboeuf 91951 Les Ulis Cedex Telephone: 016-09-21-550 FAX: 016-44-61-065
GERMANY
ANRITSU GmbH Grafenberger Allee 54-56 D-40237 Dusseldorf Germany Telephone: 0211-968550 FAX: 0211-9685555
INDIA
MEERA AGENCIES (P) LTD 23 Community Centre Zamroodpur, Kailash Colony Extension New Delhi, India 110 048 Telephone: 011-644-2700 FAX: 011-644-2500
ISRAEL
TECH-CENT, LTD 4 Raul ValenbergSt. Tel-Aviv, Israel 69719 Telephone: 972-36-478563 FAX: 972-36-478334
ITALY
ANRITSU Sp.A Rome Office Via E. Vittorini, 129 00144 Roma EUR Telephone: (06) 50-2299-711 FAX: 06-50-22-4252
JAPAN
ANRITSU CUSTOMER SERVICES LTD. 1800 Onna Atsugi—shi Kanagawa-Prf. 243 Japan Telephone: 0462-96-6688 FAX: 0462-25-8379
KOREA
ANRITSU SERVICE CENTER 8F Hyunjuk-Bldg, 832-41 Yeoksam-Dong Kangnam-Gu Seoul, 135-080, Korea Telephone: 82-2-553-6603 FAX: 82-2-553-6605
SINGAPORE
ANRITSU (SINGAPORE) PTE LTD 10, Hoe Chiang Road #07-01/02 Keppel Towers Singapore 089315 Telephone:65-282-2400 FAX:65-282-2533
SOUTH AFRICA
ETESCSA 12 Surrey Square Office Park 330 Surrey Avenue Ferndale, Randburt, 2194 South Africa Telephone: 011-27-11-787-7200 Fax: 011-27-11-787-0446
SWEDEN
ANRITSU AB Botvid Center Fittja Backe 13A 145 84 Stockholm, Sweden Telephone: (08) 534-707-00 FAX: (08)534-707-30
TAIWAN
ANRITSU COMPANY 7F, NO.316, Sec.1, Nei Hu Road Taipei, Taiwan, R.O.C. Telephone: 886-2-8751-2126 FAX: 886-2-8751-1817
UNITED KINGDOM
ANRITSU LTD. 200 Capability Green Luton, Bedfordshire LU1 3LU, England Telephone: 015-82-43-3200 FAX: 015-82-73-1303
Loading...