The Anritsu product(s) listed on the title page is (are) warranted against defects in
materials and workmanship for one year from the date of shipment.
Anritsu's obligation covers repairing or replacing products which prove to be defec
tive during the warranty period. Buyers shall prepay transportation charges for
equipment returned to Anritsu for warranty repairs. Obligation is limited to the origi
nal purchaser. Anritsu is not liable for consequential damages.
-
-
LIMITATION OF WARRANTY
The foregoing warranty does not apply to Anritsu connectors that have failed due to
normal wear. Also, the warranty does not apply to defects resulting from improper or
inadequate maintenance by the Buyer, unauthorized modification or misuse, or op
eration outside the environmental specifications of the product. No other warranty is
expressed or implied, and the remedies provided herein are the Buyer's sole and
exclusive remedies.
-
TRADEMARK ACKNOWLEDGMENTS
MS-DOS, Windows, Windows 95, Windows NT, Windows 98, Windows 2000 and
Windows ME are registered trademarks of the Microsoft Corporation.
Anritsu and Site Master are trademarks of Anritsu Company.
NOTICE
Anritsu Company has prepared this manual for use by Anritsu Company personnel
and customers as a guide for the proper installation, operation and maintenance of
Anritsu Company equipment and computer programs. The drawings, specifications,
and information contained herein are the property of Anritsu Company, and any unauthorized use or disclosure of these drawings, specifications, and information is
prohibited; they shall not be reproduced, copied, or used in whole or in part as the
basis for manufacture or sale of the equipment or software programs without the
prior written consent of Anritsu Company.
October 200110580-00060
Copyright ã 2001 Anritsu Co.Revision: A
Table of Contents
Chapter 1 - General Information
Introduction..................................1-1
Description ..................................1-1
Standard Accessories .............................1-1
This chapter provides a description, performance specifications, optional accessories, pre
ventive maintenance, and calibration requirements for the Site Master models S113C,
S114C, S331C, and S332C. Throughout this manual, the term Site Master will refer to the
models S113C, S114C, S331C, and S332C.
Model
S113C
S114C
S331C
S332C
Frequency Range
2 to 1600 MHz
2 to 1600 MHz, 100 kHz to 1600 MHz Spectrum Analysis
25 to 4000 MHz
25 to 4000 MHz, 100 kHz to 3000 MHz Spectrum Analysis
Description
The Site Master is a hand held SWR/RL (standing wave ratio/return loss), and Distance-ToFault (DTF) measurement instrument that includes a built-in synthesized signal source. All
models include a keypad to enter data and a liquid crystal display (LCD) to provide graphic
indications of SWR or RL over the selected frequency range and selected distance. The Site
Master is capable of up to 2.5 hours of continuous operation from a fully charged
field-replaceable battery and can be operated from a 12.5 dc source. Built-in energy conservation features can be used to extend battery life over an eight-hour work day.
The Site Master is designed for measuring SWR, return loss, or cable insertion loss and locating faulty RF components in antenna systems. Power monitoring is available as an option. Site Master models S114C and S332C include spectrum analysis capability. The
displayed trace can be scaled or enhanced with frequency markers or limit lines. A menu
option provides for an audible “beep” when the limit value is exceeded. To permit use in
low-light environments, the LCD can be back lit using a front panel key.
-
1
Standard Accessories
The Software Tools PC-based software program provides a database record for storing
measurement data. Site Master Software Tools can also convert the Site Master display to a
Microsoft Windowsä 95/98/NT4/2000/ME workstation graphic. Measurements stored in
the Site Master internal memory can be downloaded to the PC using the included null-mo
dem serial cable. Once stored, the graphic trace can be displayed, scaled, or enhanced with
markers and limit lines. Historical graphs can be overlaid with current data, and underlying
data can be extracted and used in spreadsheets or for other analytical tasks.
The Site Master Software Tools program can display all of the measurements made with the
Site Master (SWR, return loss, cable loss, distance-to-fault) as well as providing other func
tions, such as converting display modes and Smith charts. Refer to Chapter 7, Software
Tools, for more information.
-
-
1-1
Chapter 1 General Information
The following items are supplied with the basic hardware.
Soft Carrying Case
·
AC-DC Adapter
·
Automotive Cigarette Lighter 12 Volt DC Adapter,
·
CDROM disk containing the Software Tools program. This program contains Fault Lo
·
cation (DTF) and Smith Chart functions
Serial Interface Cable (null modem type)
·
InstaCal Module
·
One year Warranty (includes battery, firmware, and software)
·
User's Guide
·
Options
Option 5 — Add Power Monitor
·
Optional Accessories
Anritsu Precision N (m) Open/Short/Load, 42 dB, Part No. OSLN50LF
·
Anritsu Precision N (f) Open/Short/Load, 42 dB, Part No. OSLNF50LF
·
Anritsu Precision N (m) Short/Open, Part No. 22N50
·
Anritsu Precision N (f) Short/Open, Part No. 22NF50
·
·
Site Master Precision N (m) Load, 42 dB, Part No. SM/PL
·
Site Master Precision N (f) Load, 42 dB, Part No. SM/PLNF
·
7/16 (m) Precision Open/Short/Load, Part No. 2000-767
·
7/16 (f) Precision Open/Short/Load, Part No. 2000-768
·
Adapter, Precision N (m) to N (m), Part No. 34NN50A
·
Adapter, Precision N (f) to N (f), Part No. 34NFNF50
·
Adapter, 7/16 (f) to N (m), Part No. 510-90
·
Adapter, 7/16 (f) to N (f), Part No. 510-91
·
Adapter, 7/16 (m) to N (m), Part No. 510-92
·
Adapter, 7/16 (m) to N (f), Part No. 510-93
·
Adapter, 7/16 DIN (m) to 7/16 DIN (m), Part No. 510-96
·
Adapter, 7/16 DIN (f) to 7/16 DIN (f), Part No. 510-97
·
Armored Test Port Extension Cable, 1.5 meter, N (m) to
N (f), Part No. 15NNF50-1.5C
·
Armored Test Port Extension Cable, 3.0 meter, N (m) to
N (f), Part No. 15NNF50-3.0C
·
Armored Test Port Extension Cable, 5.0 meter, N (m) to
N (f), Part No. 15NNF50-5.0C
·
Armored Test Port Extension Cable, 1.5 meter, N (m) to
N (m), Part No. 15NN50-1.5C
-
1-2
Chapter 1General Information
Armored Test Port Extension Cable, 3.0 meter, N (m) to
·
N (m), Part No. 15NN50-3.0C
Armored Test Port Extension Cable, 5.0 meter, N (m) to
·
N (m), Part No. 15NN50-5.0C
Armored Test Port Extension Cable, 1.5 meter, N (m) to
·
7/16 DIN (f), Part No. 15NDF50-1.5C
Armored Test Port Extension Cable, 1.5 meter, N (m) to
·
7/16 DIN (m), Part No. 15ND50-1.5C
RF Detector, 1 to 3000 MHz, N(m) input connector, 50 Ohms, Part No. 5400-71N50
·
RF Detector, 0.01 to 20 GHz, N(m) input connector, 50 Ohms, Part No. 560-7N50B
·
Transit Case for Site Master, Part No. 760-215A
·
Antenna SMA (f), 50 W, 1.71 to 1.88 GHz, Part No. 2000-1030
·
Antenna SMA (f), 50 W, 1.85 to 1.99 GHz, Part No. 2000-1031
·
Antenna SMA (f), 50 W, 2.4 to 2.5 GHz, Part No. 2000-1032
·
Antenna SMA (f), 50 W, 806 to 869 MHz, Part No. 2000-1034
·
Antenna SMA (f), 50 W, 896 to 941 MHz, Part No. 2000-1035
·
HP Deskjet 350 Printer, Part No. 2000-766
·
Serial-to-Parallel Converter Cable (use with the HP 350 Printer), Part No. 2000-753
·
Seiko DPU-414 Thermal Printer, Part No. 2000-754 (U.S.) or
·
2000-761 (Europe)
·US Adapter (use with the DPU-414 Printer),
Part No. 2000-1002
·
Europe Adapter (use with the DPU-414 Printer)
Part No. 2000-1003
·
Japan Adapter (use with the DPU-414 Printer)
Part No. 2000-1194
·
Battery Pack (use with the DPU-414 Printer)
Part No. 2000-1004
·
Serial Interface Cable (use with the DPU-414 Printer),
Part No. 2000-1012
·
Thermal Paper (use with the DPU-411/DPU-414 Printer),
Part No. 2000-755
·
Rechargeable Battery, NiMH
Part No. 633-27
·
Battery Charger with universal power supply, NiMH only
Part No. 2000-1029
Performance Specifications
Performance specifications are provided in Table 1-1, on the following page.
1-3
Chapter 1 General Information
Table 1-1.Performance Specifications (1 of 2)
Specifications are valid when the unit is calibrated at ambient temperature after a 5 minute
warmup. Typical values are provided for reference only and are not guaranteed.
Description
Site Master:Frequency Range:
S113C, S114C
S331C, S332C
Frequency Accuracy (RF Source Mode)£75 parts per million @ 25°C*
Frequency Resolution:S113C, S114C
S331C, S332C
SWR:Range
Resolution
Return Loss:Range
Resolution
**Distance-To-Fault (DTF):
Vertical RangeReturn Loss:
SWR:
Horizontal Range
Horizontal Resolution for Coax
(rectangular windowing)
2 to 1600 MHz
25 to 4000 MHz
10 kHz
100 kHz
1.00 to 65.00
0.01
0.00 to 54.00 dB
0.01 dB
0.00 to 54.00 dB
1.00 to 65.00
0 to ((# of data points –1) ´ resolution) a maximum
of 1000m (3281 ft.) with a maximum of 517 points
resolution, # of data pts. = 130, 259, 517
(.)( )15 108´Vp
FD
Where V
the cable; dp is the number of data points
(130, 259, 517); DF is the stop frequency minus the
start frequency (Hz)
is the relative propagation velocity of
p
Value
Horizontal Resolution for Waveguide
RF Power Monitor:
Display Range
Detector Range
Offset Range
Resolution
Test Port ConnectorType N, 50W, female
***Immunity to Interfering signals
up to the level of: S113C, S114C
S331C, S332C
Maximum Input without Damage:
Test Port, Type N (f)
RF Power Detector
Measurement Accuracy:
Measurement accuracy depends on calibration components. Precision calibration components
have a directivity of 42 dB.
Cable Insertion Loss:Range
Resolution
Transmission Line Loss (one-port)
Range
Resolution
8
15 101
.((/))´-FF
F
D
Where F
is the start frequency (Hz); DF is the stop fre-
F
1
quency minus the start frequency (Hz)
–80.0 to +80 dBm or
10.0 pW to 100.0 kW
–50 dBm to +20 dBm, or 10 mW to 100 mW
0 to +60.0 dB
Measurement Range–97 dBm to +20 dBm typical
Dynamic Range³ 65 dB typical
Maximum Safe Input Level+20 dBm, maximum measurable safe input
Displayed Average Noise Level:£ –80 dBm (100 kHz to 300 kHz) typical
Display Range, Log Scale2 to 15 dB/div. in 1 dB steps; 10 divisions displayed.
Frequency Response
RF Input VSWR2.0:1
Resolution (Ref. Level)1.0 dB
Total Level Accuracy±2 dB ³ 300 kHz typical
General
Internal Memory:
Trace Memory
Instrument configuration
RS-232
Electromagnetic CompatibilityComplies with European community requirements
Power Supply:
External DC Input+12.5 to +15 Vdc, 1100 mA max.
Temperature:Storage
Operating
Weight:2.15 kg (4.76 pounds)
Dimensions:25.4 x 17.8 x 6.1 cm
100 kHz to 1.6 GHz
100 kHz to 3.0 GHz
±1 ppm/yr
±2 ppm
0 Hz (zero span) 100 kHz to 1.6 GHz
0 Hz (zero span) 100 kHz to 3.0 GHz
500 ms (zero span)
10 kHz, 30 kHz, 100 kHz, 1 MHz
± 20% typical
+27 dBm, maximum input (damage)
+27 dBm, peak pulse power
+50 Vdc
£ –97 dBm (>300 kHz) typical
±3 dB <300 kHz typical
200 maximum
10 setup locations
9 pin D-sub, three wire serial
for CE marking
–20° C to 75° C
0° C to 50° C
(10x7x2.4inches)
* ±2 ppm/D°C from 25°C
** Fault location is accomplished by inverse Fourier Transformation of data taken with the Site Master. Resolution and
maximum range depend on the number of frequency data points, frequency sweep range and relative propagation velocity
of the cable being tested.
*** Immunity measurement is made in CW mode with incoming interfering signal exactly at the same frequency (worst
case situation). Typical immunity is better when swept frequency is used.
1-5
Chapter 1 General Information
Preventive Maintenance
Site Master preventive maintenance consists of cleaning the unit and inspecting and clean
ing the RF connectors on the instrument and all accessories.
Clean the Site Master with a soft, lint-free cloth dampened with water or water and a mild
cleaning solution.
CAUTION: To avoid damaging the display or case, do not use solvents or abra
sive cleaners.
Clean the RF connectors and center pins with a cotton swab dampened with denatured alco
hol. Visually inspect the connectors. The fingers of the N (f) connectors and the pins of the
N (m) connectors should be unbroken and uniform in appearance. If you are unsure whether
the connectors are good, gauge the connectors to confirm that the dimensions are correct.
Visually inspect the test port cable(s). The test port cable should be uniform in appearance,
not stretched, kinked, dented, or broken.
Calibration
The Site Master is a field portable unit operating in the rigors of the test environment. An
Open-Short-Load (OSL) calibration should be performed prior to making a measurement in
the field (see Calibration, page 3-2). A built-in temperature sensor in the Site Master advises the user, via an icon located on the right side of the LCD screen, that the internal temperature has exceeded a safety window, and the user is advised to perform another OSL
calibration in order to maintain the integrity of the measurement.
-
-
-
NOTES:
For best calibration results—compensation for all measurement uncertain
ties—ensure that the Open/ Short/Load is at the end of the test port or optional
extension cable; that is, at the same point that you will connect the antenna or
device to be tested.
For best results, use a phase stable Test Port Extension Cable (see Optional
Accessories). If you use a typical laboratory cable to extend the Site Master test
port to the device under test, cable bending subsequent to the OSL calibration
will cause uncompensated phase reflections inside the cable. Thus, cables
which are NOT phase stable may cause measurement errors that are more pro
nounced as the test frequency increases.
For optimum calibration, Anritsu recommends using precision calibration com
ponents.
1-6
-
-
-
Chapter 1General Information
InstaCal Module
The Anritsu InstaCal module can be used in place of discrete components to calibrate the
Site Master. The InstaCal module can be used to perform an Open, Short and Load (OSL)
calibration procedure. Calibration of the Site Master with the InstaCal takes approximately
45 seconds (see Calibration, page 3-2). Unlike a discrete calibration component, the
InstaCal module can not be used at the top of the tower to conduct load or insertion loss
measurements. The module operates from 2 MHz to 4 GHz and weighs eight ounces.
Anritsu recommends annual re-characterization of the InstaCal module to verify perfor
mance with precision instrument data. The re-characterization can be performed at a local
service center or at the factory.
Annual Verification
Anritsu recommends an annual calibration and performance verification of the Site Master
and the OSL calibration components and InstaCal module by local Anritsu service centers.
Anritsu service centers are listed in Table 1-2 on the following page.
The Site Master itself is self-calibrating, meaning that there are no field-adjustable compo
nents. However, the OSL calibration components are crucial to the integrity of the calibra
tion and therefore, must be verified periodically to ensure performance conformity. This is
especially important if the OSL calibration components have been accidentally dropped or
over-torqued.
-
-
-
1-7
Chapter 1 General Information
Table 1-2.Anritsu Service Centers
UNITED STATES
ANRITSU COMPANY
685 Jarvis Drive
Morgan Hill, CA 95037-2809
Telephone: (408) 776-8300,
1-800-ANRITSU
FAX: 408-776-1744
ANRITSU COMPANY
10 New Maple Ave., Unit 305
Pine Brook, NJ 07058
Telephone: (201) 227-8999,
ETESCSA
12 Surrey Square Office Park
330 Surrey Avenue
Ferndale, Randburt, 2194
South Africa
Telephone:01-11-787-7200
Fax: 01-11-787-0446
SWEDEN
ANRITSU AB
Botvid Center
S-1585
Stockholm, Sweden
Telephone: (08) 534-717-00
FAX: (08) 534-717-30
TAIWAN
ANRITSU CO., LTD.
8F, No. 96, Section 3
Chien Kuo N. Road
Taipei, Taiwan,R.O.C.
Telephone: (02) 515-6050
FAX: (02) 509-5519
UNITED KINGDOM
ANRITSU LTD.
200 Capability Green
Luton, Bedfordshire
LU1 3LU, England
Telephone: 015-82-41-88-53
FAX: 015-82-31-303
1-8
Chapter 2
Functions and Operations
Introduction
This chapter provides a brief overview of the Site Master functions and operations, provid
ing the user with a starting point for making basic measurements. For more detailed infor
mation, refer to Chapter 4, Measurements and Chapter 7, Software Tools.
The Site Master is designed specifically for field environments and applications requiring
mobility. As such, it is a lightweight, handheld, battery operated unit which can be easily
carried to any location, and is capable of up to 2.5 hours of continuous operation from a
fully charged battery. Built-in energy conservation features allow battery life to be extended
over an eight-hour workday. The Site Master can also be powered by a 12.5 Vdc external
source. The external source can be either the Anritsu AC-DC Adapter (P/N 40-115) or 12.5
Vdc Automotive Cigarette Lighter Adapter (P/N 806-62). Both items are standard accesso
ries.
Test Connector Panel
The connectors and indicators located on the test panel are listed and described below.
12.5-15VDC
(1100 mA)
When using the AC-DC Adapter, always use a three-wire power cable connected
to a three-wire power line outlet. If power is supplied without grounding the equip
ment in this manner, there is a risk of receiving a severe or fatal electric shock.
12.5 to 15 Vdc @ 1100 mA input to power the unit or for battery charging.
WARNING
-
-
2
-
-
Battery
Charging
External
Power
Serial
Interface
RF OutRF output, 50 W impedance, for reflection measurements.
RF InRF input, 50 W impedance, for spectrum analysis measurements.
RF DetectorRF detector input for the Power Monitor.
Illuminates when the battery is being charged. The indicator automatically shuts
off when the battery is fully charged.
Illuminates when the Site Master is being powered by the external charging unit.
RS232 DB9 interface to a COM port on a personal computer (for use with the
Anritsu Software Tools program) or to a supported printer.
2-1
Chapter 2 Functions and Operations
Front Panel Overview
The Site Master menu-driven user interface is easy to use and requires little training. Hard
keys on the front panel are used to initiate function-specific menus. There are four function
hard keys located below the display, Mode, Frequency/Distance, Amplitude and Sweep.
There are seventeen keypad hard keys located to the right of the display. Twelve of the key
pad hard keys perform more than one function, depending on the current mode of operation.
The dual purpose keys are labeled with one function in black, the other in blue.
There are also six soft keys that change function depending upon the current mode selec
tion. The current soft key function is indicated in the active function block to the right of
the display. The locations of the different keys are illustrated in Figure 2-1.
Active
Function
MODE
Block
FREQ/DIST
Site Master S332C
AMPLITUDE
SWEEP
START
CAL
SAVE
SETUP
LIMIT
SAVE
DISPLAY
ON
OFF
1
3
5
7
9
Soft Keys
2
AUTO
SCALE
4
RECALL
SETUP
6
MARKER
8
RECALL
DISPLAY
0
PRINT
.
ESCAPE
CLEAR
ENTER
RUN
HOLD
SYS
Keypad
Hard
Keys
+
-
/
-
Function Hard Keys
Figure 2-1.Site Master Soft Keys
The following sections describe the various key functions.
2-2
Chapter 2 Functions and Operations
Function Hard Keys
MODEOpens the mode selection box (below). Use the Up/Down arrow key to select a
mode. Press the
MEASUREMENT MODE
FREQ - SWR
DTF -SWR
POWER MONITOR
SPECTRUM ANALYZER
Figure 2-2.Mode Selection Box
FREQ/DISTDisplays the Frequency or Distance to Fault softkey menus depending on the
measurement mode.
ENTER key to implement.
RETURN LOSS
CABLE LOSS - ONE PORT
RETURN LOSS
AMPLITUDEDisplays the amplitude softkey menu for the current operating mode.
SWEEPDisplays the Sweep function softkey menu for the current operating mode.
2-3
Chapter 2 Functions and Operations
Keypad Hard Keys
This section contains an alphabetical listing of the Site Master front panel keypad controls
along with a brief description of each. More detailed descriptions of the major function
keys follow.
The following keypad hard key functions are printed in black on the keypad keys.
0-9These keys are used to enter numerical data as required to setup or per
form measurements.
+/–
·
ESCAPE
CLEAR
UP/DOWN
ARROWS
NOTE: At turn on, before any other keys are pressed, the Up/Down arrow key
may be used to adjust the display contrast. Press
eration.
ENTERImplements the current action or parameter selection.
The plus/minus key is used to enter positive or negative values as required
to setup or perform measurements.
The decimal point is used to enter decimal values as required to setup or
perform measurements.
Exits the present operation or clears the display. If a parameter is being
edited, pressing this key will clear the value currently being entered and
restore the last valid entry. Pressing this key again will close the parame
ter. During normal sweeping, pressing this key will move up one menu
level.
Increments or decrements a parameter value. The specific parameter value
affected typically appears in the message area of the LCD.
ENTER to return to normal op-
-
-
ON
OFF
SYSAllows selection of display language and system setup parameters.
Turns the Anritsu Site Master on or off. When turned on, the system state
at the last turn-off is restored. If the ESCAPE/CLEAR key is held down
while the ON/OFF key is pressed, the factory preset state will be restored.
Choices are OPTIONS, CLOCK, SELF TEST, STATUS andLANGUAGE.
2-4
Chapter 2 Functions and Operations
The following keypad hard key functions are printed in blue on the keypad keys.
Turns the liquid crystal display (LCD) back-lighting ON or OFF.
(Leaving back lighting off conserves battery power.)
LCD Contrast adjust. Use the Up/Down arrow key and
ENTER to adjust
the display contrast.
AUTO
Automatically scales the display for optimum resolution.
SCALE
LIMITDisplays the limit line menu for the current operating mode.
MARKERDisplays the marker menu of the current operating mode.
PRINTPrints the current display to the selected printer via the RS232 serial port.
RECALL
DISPLAY
Recalls a previously saved trace from memory. When the key is pressed, a
Recall Trace selection box appears on the display. Select a trace using the
Up/Down arrow key and press the
To erase a saved trace, highlight the trace and select the
ENTER key to implement.
DELETE TRACE
softkey. To erase all saved traces, select the DELETE ALL TRACES
softkey.
RECALL
SETUP
Recalls a previously saved setup from memory location 1 through 10.
When the key is pressed, a
RECALL SETUP selection box appears on the
display. Select a setup using the Up/Down arrow key and press the
ENTER key to implement. Setup 0 recalls the factory preset state.
RUN
HOLD
SAVE
DISPLAY
SAVE
SETUP
START
CAL
When in the Hold mode, this key starts the Site Master sweeping and provides a Single Sweep Mode trigger; when in the Run mode, it pauses the
sweep. When in the Hold mode, the hold symbol (page 2-19) appears on
the display. Hold mode conserves battery power.
Saves up to 200 displayed traces to non-volatile memory. When the key is
pressed,
TRACE NAME: appears in the lower left of the display. Save the
display with up to 16 alphanumeric characters for that trace name and
press the
ENTER key to implement.
Saves the current system setup to 1 of 10 internal non-volatile memory lo
cations. When the key is pressed, a
SAVE SETUP selection box appears
on the display. Use the Up/Down arrow key to select a setup and press the
ENTER key to implement.
Starts the calibration in SWR, Return Loss, Cable Loss, or DTF measure
ment modes (not available in Spectrum Analyzer mode).
-
-
2-5
Chapter 2 Functions and Operations
Soft Keys
Each keypad key opens a set of soft key selections. Each of the soft keys has a correspond
ing soft key label area on the display. The label identifies the function of the soft key for the
current Mode selection.
Figures 2-3 through 2-7 show the soft key labels for each Mode selection.
FREQ/DISTDisplays the frequency and distance menu depending on the measurement mode.
Frequency
Menu
Distance
Menu
Provides for setting sweep frequency end points when
FREQ mode is selected.
Selected frequency values may be changed using the keypad or Up/Down arrow
key.
F1 — Opens the F1 parameter for data entry. This is the start value for the
q
frequency sweep. Press
F2 — Opens the F2 parameter for data entry. This is the stop value for the
q
frequency sweep. Press
ENTER when data entry is complete.
ENTER when data entry is complete.
Provides for setting Distance to Fault parameters when a DTF mode is selected.
Choosing
DIST causes the soft keys, below, to be displayed and the correspond
ing values to be shown in the message area. Selected distance values may be
changed using the keypad or Up/Down arrow key.
D1 — Opens the start distance (D1) parameter for data entry. This is the start
q
value for the distance range (D1 default = 0). Press
ENTER when data entry
is complete.
D2 — Opens the end distance (D2) parameter for data entry. This is the end
q
value for the distance range. Press
DTF AID — Provides interactive help to optimize DTF set up parameters.
q
Use the Up/Down arrow key to select a parameter to edit. Press
ENTER when data entry is complete.
ENTER when
data entry is complete.
MORE — Selects the Distance Sub-Menu, detailed below.
q
-
Distance
Sub-Menu
Provides for setting the cable loss and relative propagation velocity of the coaxial cable. Selected values may be changed using the Up/Down arrow key or keypad.
q
LOSS — Opens the Cable Loss parameter for data entry. Enter the loss per
foot (or meter) for the type of transmission line being tested. Press
ENTER
when data entry is complete. (Range is 0.5 to 5.000 dB/m, 1.524 dB/ft)
q
PROP VEL (relative propagation velocity) — Opens the Propagation Veloc
ity parameter for data entry. Enter the propagation velocity for the type of
transmission line being tested. Press
ENTER when data entry is complete.
(Range is 0.010 to 1.000)
q
CABLE — Opens a list of cable folders for selection of a common coaxial ca
ble folder or custom coaxial cable folder. Select either folder and use the
Up/Down arrow key and
ENTER to make a selection. This feature provides a
rapid means of setting both cable loss and propagation velocity. (Refer to Ap
pendix A for a listing of common coaxial cables showing values for Relative
Propagation Velocity and Nominal Attenuation in dB/m or dB/ft @
1000 MHz, 2000 MHz and 2500 MHz.) The custom cable folder can consist
of up to 24 user-defined cable parameters downloaded via the Site Master
Software Tools program.
q
WINDOW — Opens a menu of FFT windowing types for the DTF calcula
tion. Scroll the menu using the Up/Down arrow key and make a selection
with the
ENTER key.
-
-
2-10
q
BACK — Returns to the Distance Menu.
Chapter 2 Functions and Operations
Choosing FREQ/DIST in Spectrum Analyzer mode causes the soft keys, below, to be dis
played and the corresponding values to be shown in the message area.
CENTER ¾ Sets the center frequency of the Spectrum Analyzer display . En
q
ter a value using the Up/Down arrow key or keypad, press
ESCAPE to restore previous value.
SPAN ¾ Sets the user-defined frequency span. Use the Up/Down arrow key
q
or keypad to enter a value in MHz. Also brings up
EDIT allows editing of the frequency span. Enter a value using the num
q
FULL and ZERO softkeys.
ENTER to accept,
ber keys.
FULL span sets the Spectrum Analyzer to its maximum frequency span.
q
ZERO span sets the span to 0 Hz. This displays the input signal in an am
q
plitude versus time mode, which is useful for viewing modulation.
SPAN UP 1-2-5 activates the span function so that the span may be in
q
creased quickly in a 1-2-5 sequence.
SPAN DOWN 1-2-5 activates the span function so that the span may be
q
reduced quickly in a 1-2-5 sequence.
BACK returns to the previous menu level.
q
START ¾ Sets the Spectrum Analyzer in the START-STOP mode. Enter a
q
start frequency value (in kHz, MHz, or GHz) using the Up/Down arrow key
or keypad, press
ENTER to accept, ESCAPE to restore.
-
-
-
-
-
q STOP ¾ Sets the Spectrum Analyzer in the START-STOP mode. Enter a
stop frequency value (in kHz, MHz, or GHz) using the Up/Down arrow key
or keypad, press
ENTER to accept, ESCAPE to restore.
2-11
Chapter 2 Functions and Operations
AMPLITUDEDisplays the amplitude or scale menu depending on the measurement mode.
Amplitude
Menu
Provides for changing the display scale. Selected values may be changed using
the Up/Down arrow key or keypad.
Choosing
AMPLITUDE in FREQ or DTF measurement modes causes the soft
keys, below, to be displayed and the corresponding values to be shown in the
message area.
TOP — Opens the top parameter for data entry and provides for setting the
q
top scale value. Press
BOTTOM — Opens the bottom parameter for data entry and provides for set
q
ting the bottom scale value. Press
Choosing
AMPLITUDE in SPECTRUM ANALYZER mode causes the soft keys,
ENTER when data entry is complete.
ENTER when data entry is complete.
below, to be displayed and the corresponding values to be shown in the message
area.
REF LEVEL — Activates the amplitude reference level function.
q
SCALE — Activates the scale function in a 2 through 15 dB logarithmic am
q
plitude scale.
ATTEN — Sets the Anritsu input attenuator so that it is either coupled auto
q
matically to the reference level (
UNITS — Choose from the menu of amplitude related units. Selection of
q
AUTO) or manually adjustable (MANUAL).
-
dBm sets absolute decibels relative to 1 mW as the amplitude unit. Selection
of dBV, dBmV or dBmV sets absolute decibels relative to 1 volt, 1 millivolt,
or 1 microvolt respectively as the amplitude unit.
-
-
2-12
Chapter 2 Functions and Operations
SWEEPDisplays the Sweep function soft key menu for the current operating mode.
Sweep MenuProvides for changing the display resolution, single or continuous sweep, and
access to the Trace Math functions.
Choosing
SWEEP in FREQ or DTF measurement modes causes the soft keys
below to be displayed.
RESOLUTION — Opens the display to change the resolution. Choose 130,
q
259, or 517 data points. (In DTF mode, resolution can be adjusted through
the DTF-AID table.)
SINGLE SWEEP — Toggles the sweep between single sweep and continu
q
ous sweep. In single sweep mode, each sweep must be activated by the
RUN/HOLD button.
TRACE MATH — Opens up the Trace Math functions (trace-memory or
q
trace+memory) for comparison of the real time trace in the display with any
of the traces from memory. (Not available in DTF mode.)
TRACE OVERLAY — Opens up the Trace Overlay functions menu to allow
q
the current trace to be displayed with a trace in memory overlaid on it.
Choose ON or OFF and SELECT TRACE to select the trace from memory to
be overlaid.
Choosing
SWEEP in SPECTRUM ANALYZER mode causes the soft keys below
to be displayed.
RBW — Sets the resolution bandwidth so that it is either coupled automati-
q
cally to the span (
q
VBW — Sets the video bandwidth so that it is either coupled automatically to
the span (
AUTO) or manually adjustable (MANUAL).
AUTO) or manually adjustable (MANUAL).
-
q
MAX HOLD — Displays and holds the maximum responses of the input sig-
nal.
q
CONT/SINGLE — Toggles between continuous and single sweep.
q
MEASURE — Activates a menu of measurement related functions. Use the
corresponding softkey to select the measurement function.
q
FIELD STRENGTH — Accesses a menu of field strength measurement
options.
ON/OFF - Turns field strength measurements on or off.
SELECT ANTENNA - Selects an antenna profile to be used for field
strength measurements.
BACK - Returns to the previous menu.
q
OBW — Activates the occupied bandwidth menu. Select either
METHOD, % or dBc method of occupied bandwidth measurement.
METHOD allows selection of either % of power or dB Down.
Selecting % allows entry of the desired % of occupied bandwidth to be
measured.
Selecting dBc allows entry of the desired power level (dBc) to be mea
sured.
q
DETECTION — Accesses a menu of detector modes including POSI
TIVE PEAK detection, AVERAGE detection and NEGATIVE PEAK
detection.
-
-
2-13
Chapter 2 Functions and Operations
AVERAGE (1-25) — The display will be an average of the number of
q
sweeps specified here. For example, if the number four is entered here,
the data displayed will be an average of the four most recent sweeps.
CHANNEL POWER — Activates Channel Power measurement. Chan
q
nel power is measured in dBm. Channel Power density is measured in
dBm/Hz.
ACP — Accesses a menu of Adjacent Channel Power ratio measurement
q
options:
CENTER FREQ - Activates the center frequency function and sets the
Anritsu Site Master to the center frequency. A specific center frequency
can be entered using the keypad or Up/Down arrow key. Select the GHz,
MHz, kHz,orHz softkey to accept the center frequency input.
MAIN CHANNEL BW - Sets the bandwidth of the main channel.
ADJACENT CHANNEL BW - Sets the bandwidth of the adjacent chan
nel.
CHANNEL SPACING - Sets the channel spacing.
MEASURE - Begins the ACP measurement.
BACK - Returns to the previous menu.
BACK - Returns to the previous menu.
q
TRACE — Activates a menu of trace related functions. Use the correspond-
q
ing softkey to select the desired trace function.
NOTE: For this function, Trace A is always the runtime trace, and Trace B is always a saved trace. Refer to page for information on saving and recalling
traces.
-
-
q
RESET A — Clears the current runtime trace.
q
A ® B — Stores the current runtime trace into the Trace B buffer.
q
A–B® A — Performs a subtraction trace math operation.
q
A+B®A — Performs an addition trace math operation.
q
TRACE B — Accesses a menu of saved trace options.
VIEW B - Provides a trace overlay of the saved trace with the current
runtime trace.
CLEAR B - Turns off trace overlay.
RECALL TRACE ® B - Recalls the saved trace B buffer.
q
BACK — Returns to the previous menu level.
2-14
Chapter 2 Functions and Operations
Power Monitor Menu
Selecting POWER MONITOR from the Mode menu causes the soft keys, described below,
to be displayed and the corresponding values shown in the message area.
UNITS — Toggles between dBm and Watts.
q
REL — Turns relative mode OFF, if currently ON. If relative mode is cur
q
rently OFF, turns it ON and causes the power level to be measured and saved
as the base level. Subsequent measurements are then displayed relative to this
saved value. With units of dBm, relative mode displays dBr; with units of
Watts, relative mode displays % (percent).
OFFSET — Turns Offset OFF, if currently ON. If Offset is currently OFF,
q
turns it ON and opens the Offset parameter for data entry. Press
when data entry is complete.
Offset is the attenuation (in dB) inserted in the line between the DUT and the
RF detector. The attenuation is added to the measured input level prior to dis
play.
ZERO — Turns Zero OFF, if currently ON. If Zero is currently OFF, this
q
softkey turns it ON and initiates collection of a series of power level samples,
which are averaged and saved. This saved value is then subtracted from subsequent measurements prior to display.
-
ENTER
-
2-15
Chapter 2 Functions and Operations
MARKERChoosing MARKER causes the soft keys, below, to be displayed and the corre
sponding values to be shown in the message area. Selected frequency marker or
distance marker values may be changed using the keypad or Up/Down arrow
key.
M1 — Selects the M1 marker parameter and opens the M1 marker second
q
level menu.
ON/OFF — Turns the selected marker on or off.
q
EDIT — Opens the selected marker parameter for data entry. Press
q
ENTER when data entry is complete or ESCAPE to restore the previous
value.
MARKER TO PEAK — Places the selected marker at the frequency or
q
distance with the maximum amplitude value.
MARKER TO VALLEY — Places the selected marker at the frequency or
q
distance with the minimum amplitude value.
BACK — Returns to the Main Markers Menu.
q
M2 through M4 — Selects the marker parameter and opens the marker second
q
level menu.
ON/OFF — Turns the selected marker on or off.
q
EDIT — Opens the selected marker parameter for data entry. Press
q
ENTER when data entry is complete or ESCAPE to restore the previous
value.
-
q DELTA (Mx-M1) — Displays delta amplitude value as well as delta fre-
quency or distance for the selected marker with respect to the M1 marker.
q
MARKER TO PEAK — Places the selected marker at the frequency or
distance with the maximum amplitude value.
q
MARKER TO VALLEY — Places the selected marker at the frequency or
distance with the minimum amplitude value.
q
BACK — Returns to the Main Markers Menu.
q
M5 — Selects the M5 marker parameter and opens the M5 second level
menu.
q
ON/OFF — Turns the selected marker on or off.
q
EDIT — Opens the selected marker parameter for data entry. Press
ENTER when data entry is complete or ESCAPE to restore the previous
value.
q
PEAK BETWEEN M1 & M2 — Places the selected marker at the fre
quency or distance with the maximum amplitude value between marker
M1 and marker M2.
q
VALLEY BETWEEN M1 & M2 — Places the selected marker at the fre
quency or distance with the minimum amplitude value between marker
M1 and marker M2.
q
BACK — Returns to the Main Markers Menu.
-
-
2-16
q
M6 — Selects the M6 marker parameter and opens the M6 second level
menu.
Chapter 2 Functions and Operations
ON/OFF — Turns the selected marker on or off.
q
EDIT — Opens the selected marker parameter for data entry. Press
q
ENTER when data entry is complete or ESCAPE to restore the previous
value.
LIMIT
PEAK BETWEEN M3 & M4 — Places the selected marker at the peak be
q
tween marker M3 and marker M4.
VALLEY BETWEEN M3 & M4 — Places the selected marker at the valley
q
between marker M3 and marker M4.
BACK — Returns to the Main Markers Menu.
q
Pressing LIMIT on the data keypad activates a menu of limit related functions.
Use the corresponding softkey to select the desired limit function. Then use the
Up/Down arrow key to change its value, which is displayed in the message area
at the bottom of the display.
Choosing
LIMIT in FREQ or DTF measurement modes causes the soft keys be
low to be displayed.
SINGLE LIMIT — Sets a single limit value in dBm. Menu choices are:
q
ON/OFF
EDIT
BACK
MULTIPLE LIMITS — Sets multiple user defined limits, and can be used to
q
create a limit mask for quick pass/fail measurements. Menu choices are:
LIMIT BEEP — Turns the audible limit beep indicator on or off.
q
SYSDisplays the System menu softkey selections.
OPTIONS — Displays a second level of functions:
q
UNITS — Select the unit of measurement (English or metric).
q
PRINTER — Displays a menu of supported printers. Use the Up/Down
q
arrow key and
FIXED CW — Toggles the fixed CW function ON or OFF. When OFF, a
q
ENTER key to make the selection.
narrow band of frequencies centered on the selected frequency is gener
ated. When CW is ON, only the center frequency is generated. Output
power is pulsed in all modes.
CHANGE DATE FORMAT — Toggles the date format between
q
MM/DD/YY, DD/MM/YY, and YY/MM/DD.
-
BACK — Returns to the top-level SYS Menu.
q
CLOCK — Displays a second level of functions:
q
q HOUR — Enter the hour (0-23) using the Up/Down arrow key or the key-
pad. Press
ENTER when data entry is complete or ESCAPE to restore the
previous value.
q
MINUTE — Enter the minute (0-59) using the Up/Down arrow key or the
keypad. Press
ENTER when data entry is complete or ESCAPE to restore
the previous value.
q
MONTH — Enter the month (1-12) using the Up/Down arrow key or the
keypad. Press
ENTER when data entry is complete or ESCAPE to restore
the previous value.
q
DAY — Enter the day using the Up/Down arrow key or the keypad. Press
ENTER when data entry is complete or ESCAPE to restore the previous
value.
q
YEAR — Enter the year (1997-2036) using the Up/Down arrow key or
the keypad. Press
ENTER when data entry is complete or ESCAPE to re
store the previous value.
q
BACK — Returns to the top-level SYS menu.
q
SELF TEST — Start an instrument self test.
q
STATUS — Displays the current instrument status, including calibration sta
tus, temperature, and battery charge state. Press ESCAPE to return to opera
tion.
-
-
-
2-18
q
Language — Pressing this soft key immediately changes the language used
to display messages on the Site Master display. Choices are English, French,
German, Spanish, Chinese, and Japanese. The default language is English.
Chapter 2 Functions and Operations
Symbols
Table 2-1 provides a listing of the symbols used as condition indicators on the LCD display.
Table 2-1.LCD Icon Symbols
IconSymbol
Site Master is in Hold for power conservation. To resume sweeping, press
HOLD
the
RUN/HOLD key. After 10 minutes without a key press, the Site Master
will automatically activate the power conservation mode.
ò
dx
T
6
CAL ON
CAL ON!
CALL OFF
Integrator Failure. Intermittent integrator failure may be caused by inter
ference from another antenna. Persistent integrator failure indicates a need
to return the Site Master to the nearest Anritsu service center for repair.
Lock fail indication. Check battery. (If the Site Master fails to lock with a
fully charged battery, call your Anritsu Service Center.)
When calibration is performed, the Site Master stores the ambient temper
ature. If the temperature drifts outside the specified range, this indicator
will flash. A recalibration at the current temperature is recommended.
Indicates the remaining charge on the battery. The inner white rectangle
grows longer as the battery charge depletes.
Indicates internal data processing.
The Site Master has been calibrated.
The Site Master has been calibrated with the InstaCal Module.
The Site Master has not been calibrated.
-
-
Self Test
At turn-on, the Site Master runs through a series of quick checks to ensure the system is
functioning properly. Note that the battery voltage and temperature are displayed in the
lower left corner below the self test message. If the battery is low, or if the ambient temper
ature is not within the specified operational range, Self Test will fail. If Self Test fails and
the battery is fully charged and the Site Master is within the specified operating range, call
your Anritsu Service Center.
Error Codes
Self Test Errors
A listing of Self Test Error messages is given in Table 2-2.
-
2-19
Chapter 2 Functions and Operations
Table 2-2.Self Test Error Messages
Error Message
BATTERY LOW
EXTERNAL
POWER LOW
PLL FAILED
INTEGRATOR
FAILED
EEPROM R/W
FAILED
OUT OF TEMP.
RANGE
RTC BATTERY
LOW
LO LOCK FAILThe local oscillator in the spectrum analyzer has phase lock loop er-
Battery voltage is less than 9.5 volts. Charge battery. If condition per
sists, call your Anritsu Service Center.
External supply voltage is less than 12.5 volts. Call your Anritsu Ser
vice Center
Phase-locked loops failed to lock. Charge battery. If condition persists
with a fully charged battery, call your Anritsu Service Center
Integration circuit could not charge to a valid level. Charge battery. If
condition persists with a fully charged battery, call your Anritsu Ser
vice Center.
Non-volatile memory system has failed. Call your Anritsu Service
Center.
Ambient temperature is not within the specified operating range. If the
temperature is within the specified operating range and the condition
persists, call your Anritsu Service Center.
The internal real-time clock battery is low. A low or drained clock battery will affect the date stamp on saved traces. Contact your nearest
Anritsu Service Center.
rors. If condition persists with a fully charged battery, call your Anritsu
Service Center.
Description
-
-
-
POWER MONITOR
FAIL
BATTERY CAL
LOST
MEMORY FAIL
The time and date
Have not been set
on this Site Master.
To set it, after exit
ing here press
the <SYS>
[CLOCK] keys.
Press ENTER or
ESC to continue
Note: A listing of Anritsu Service Centers is provided in Table 1-2 , page 1-8.
Failure of the power monitor system board voltages. If condition persists, call your Anritsu Service Center.
Battery communication failed. The indicated battery charge status may
be invalid. If condition persists, call your Anritsu Service Center.
The EEPROM test on the Site Master main board has failed. If condi
tion persists, call your Anritsu Service Center.
The time and date are not properly set in the Site Master. If condition
persists, call your Anritsu Service Center.
-
-
2-20
Range Errors
A listing of Range Error messages is given in Table 2-3.
Table 2-3.Range Error Messages (1 of 2)
Error Message
RANGE
ERROR:F1 > F2
RANGE
ERROR:D1 > D2
RANGE
ERROR:D2 >
DMax=xx.x ft (m)
The start (F1) frequency is greater than the stop (F2) frequency.
The start (D1) distance is greater than the stop (D2) distance.
The stop distance (D2) exceeds the maximum unaliased range. This
range is determined by the frequency span, number of points,
and relative propagation velocity:
MaximumUnaliased Range
Where:dp is the number of data points (130, 259, or 517)
V
is the relative propagation velocity
f
F2 is the stop frequency in Hz
F1 is the start frequency in Hz
Maximum Unaliased Range is in meters
Chapter 2 Functions and Operations
Description
8
´-
( .)()()15 10
=
dpV
1
FF
-
21
f
RANGE ERROR:
TOP<=BOTTOM
RANGE ERROR:
TOP>=BOTTOM
CAL
INCOMPLETE
DIST REQUIRES
F1<F2
DIST REQUIRES
CAL
NO STORED
SWEEP AT THIS
LOCATION
USE OPTIONS
MENU TO
SELECT A
PRINTER
DISTANCE AND
CABLE
INSERTION LOSS
MODE ARE
INCOMPATIBLE
The SWR scale parameter top value is less than or equal to its bottom
value.
The RL scale parameter top value is greater than or equal to its bottom
value.
A complete open, short, and load calibration must be performed before calibration can be turned on.
Valid distance to fault plots require a non-zero frequency span.
Distance-to-fault measurements cannot be performed with CAL OFF.
Attempting to recall a display from a location that has not been previ
ously written to. That is, the location does not contain stored sweep.
Attempting to print a display with no printer selected. Select a printer,
then retry.
DTF measurements only display RL or SWR versus distance.
-
2-21
Chapter 2 Functions and Operations
Table 2-3.Range Error Messages (2 of 2)
Error Message
CANNOT ZERO
NO DETECTOR
INSTALLED
CANNOT ZERO
INPUT SIGNAL
TOO HIGH
POWER MONITOR
OPTION NOT
INSTALLED
Attempting to perform a Power Monitor zero adjust function with no
RF detector connected to the Site Master.
Attempting to perform a Power Monitor zero adjust function with an
input of greater than –20 dBm.
Attempting to enter Power Monitor mode with no Option 5 installed.
Description
InstaCal Error Messages
If an InstaCal module that has not been characterized in more than one year is connected to
the Site Master, the follow message will be displayed:
The InstaCal Module Connected, S/N XXXXX
Was Characterized more than 1 year ago.
Characterization date
Current SiteMaster Date
You are advised to return this InstaCal module
to the factory for re-characterization.
Press ENTER to continue, see manual for details.
Return the InstaCal module to the factory for re-characterization in order to ensure contin
ued accurate measurements.
If the serial number of the connected InstaCal module does not match the serial number
stored in the Site Master, the following message is displayed:
The InstaCal characterization data stored in the SiteMaster
is for a module different than the one currently connected.
SiteMaster contains data for InstaCal module S/N:
Currently connected InstaCal Module S/N:
Would you like to overwrite the previously
loaded InstaCal characterization?
Press the YES soft key to update the stored InstaCal characterization to use the currently
connected module.
Press the
NO soft key to keep the stored InstaCal characterization.
2-22
-
A listing of InstaCal error messages is given in Table 2-4.
Table 2-4.InstaCal Error Messages
Chapter 2 Functions and Operations
Error Message
FAILED TO READ
SERIAL NUMBER
OF INSTACAL
MODULE
FAILED TO SUC
CESSFULLY
TRANSFER
INSTACAL MOD
ULE DATA TO
SITE MASTER
FAILED TO SET
BAUD WITH
INSTACAL
MODULE
Description
The Site Master was unable to read the InstaCal module serial number.
The Site Master was unable to transfer all necessary data between the
-
InstaCal module and the Site Master.
-
The Site Master was unable to establish communication with the
InstaCal module.
2-23
Chapter 2 Functions and Operations
Battery Information
Charging a New Battery
The NiMH battery supplied with the Site Master has already completed three charge and
discharge cycles at the factory and full battery performance should be realized after your
first charge.
NOTE: The battery will not charge if the battery temperature is above 45° Cor
below 0° C.
Charging the Battery in the Site Master
The battery can be charged while installed in the Site Master.
Step 1. Turn the Site Master off.
Step 2. Connect the AC-DC adapter (Anritsu part number: 40-115) to the Site Master
charging port.
Step 3. Connect the AC adapter to a 120 VAC or 240 VAC power source as appropriate
for your application.
The green external power indicator on the Site Master will illuminate, indicating
the presence of external DC power and the battery will begin charging. The indicator will remain lit as long as the battery is charging. Once the battery is fully
charged, the battery indicator will turn off. If the battery fails to charge, contact
your nearest Anritsu service center.
Charging the Battery in the Optional Charger
Up to two batteries can be charged simultaneously in the optional battery charger.
Step 1. Remove the NiMH battery from your Site Master and place it in the optional
charger (Anritsu part number 2000-1029).
Step 2. Connect the lead from the AC-DC adapter to the charger.
Step 3. Connect the AC-DC adapter to a 120 VAC or 240 VAC power source as appro
priate for your application.
Each battery holder in the optional charger has an LED charging status indicator. The LED
color changes as the battery is charged:
Red indicates the battery is charging
Green indicates the battery is fully charged
Yellow indicates the battery is in a waiting state (see below).
A yellow light may occur because the battery became too warm during the charge cycle.
The charger will allow the battery to cool off before continuing the charge. A yellow light
may also indicate that the charger is alternating charge to each of the two batteries.
-
A blinking red light indicates less than 13 VDC is being supplied to the charger stand.
Check that the correct AC charger adapter is connected to the charger stand. If the battery
fails to charge, contact your nearest Anritsu Service Center.
2-24
Chapter 2 Functions and Operations
The NiMH battery will last longer and perform better if allowed to completely discharge
before recharging. For maximum battery life, it is recommended that the NiMH battery be
completely discharged and recharged once every three months.
Battery Charge Indicator
When the AC-DC adapter is unplugged from the Site Master, the battery indicator symbol
will be continuously displayed at the top left corner of the Site Master display (Figure 2-8).
BATTERY INDICATOR
Figure 2-8.Site Master Battery Indicator
A totally black bar indicates a fully charged battery. When LOW BATT replaces the battery
indicator bar at the top left corner, a couple of minutes of measurement time remains. If a
flashing LOW BATT is accompanied by an audio beep at the end of each trace, the battery
has approximately one minute of useable time remaining.
Once all the power has drained from the battery, the Site Master display will fade. At this
point, your Site Master will switch itself off and the battery will need to be recharged.
Battery Life
It is normal for NiMH batteries to self-discharge during storage, and to degrade to 80% of
original capacity after 12 months of continuous use.
Figure 2-9.NiMH Battery Storage Characteristics
2-25
Chapter 2 Functions and Operations
The battery can be charged and discharged 300 to 500 times, but it will eventually wear out.
The battery may need to be replaced when the operating time between charging is notice
ably shorter than normal.
Important Battery Information
With a new NiMH battery, full performance is achieved after three to five complete
·
charge and discharge cycles. The NiMH battery supplied with the Site Master has already
completed three charge and discharge cycles at the factory.
Recharge the battery only in the Site Master or in an Anritsu approved charger.
·
When the Site Master or the charger is not in use, disconnect it from the power source.
·
Do not charge batteries for longer than 24 hours; overcharging may shorten battery life.
·
If left unused a fully charged battery will discharge itself over time. Storing the battery in
·
extreme hot or cold places will reduce the capacity and lifetime of the battery. The bat
tery will discharge faster at higher ambient temperatures.
Discharge an NiMH battery from time to time to improve battery performance and bat
·
tery life.
The battery can be charged and discharged hundreds of times, but it will eventually wear
·
out.
The battery may need to be replaced when the operating time between charging is notice-
·
ably shorter than normal.
-
-
-
Do not short-circuit the battery terminals.
·
Do not drop, mutilate or attempt to disassemble the battery.
·
·
Never use a damaged or worn out charger or battery.
·
Always use the battery for its intended purpose only.
·
Temperature extremes will affect the ability of the battery to charge: allow the battery to
cool down or warm up as necessary before use or charging.
·
Batteries must be recycled or disposed of properly. Do not place batteries in garbage.
·
Do not dispose of batteries in a fire!
2-26
Chapter 3
Getting Started
Introduction
This chapter provides a brief overview of the Anritsu Site Master. The intent of this chapter
is to provide the user with a starting point for making basic measurements.
Power On Procedure
The Anritsu Site Master is capable of up to 2.5 hours of continuous operation from a fully
charged, field-replaceable battery. Built-in energy conservation features allow battery life to
be extended over an eight-hour workday.
The Site Master can also be operated from a 12.5 Vdc source (which will also simulta
neously charge the battery). This can be achieved with either the Anritsu AC-DC Adapter
(P/N 40-115) or 12.5 Vdc Automotive Cigarette Lighter Adapter (P/N 806-62). Both items
are included as standard accessories (see Chapter 1).
To power on the Site Master:
Step 1.
Figure 3-1.Site Master On/Off Key
Press the ON/OFF front panel key (Figure 3-1).
Site Master S332C
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
8
7
RECALL
SAVE
DISPLAY
DISPLAY
9
0
PRINT
ON
OFF
MODE
FREQ/DIST
AMPLITUDE
SWEEP
.
ESCAPE
CLEAR
ENTER
RUN
HOLD
SYS
ON/OFF
KEY
+
-
/
-
3
Step 2.
The Site Master takes about five seconds to perform a series of self-diagnostic
and adjustment routines. At completion, the screen displays the Anritsu logo, the
model number, and the firmware version. The Site Master will continue after a
10 second timeout, or:
Press ENTER to continue.
The Site Master is now ready for operation.
3-1
Chapter 3 Getting Started
Select the Frequency or Distance
Regardless of the calibration method used, the frequency range for the desired measure
ments must be set before calibrating the Site Master. The following procedure selects the
frequency range for the calibration.
Step 1. Press the
Step 2. Press the
Step 3. Enter the desired start frequency using the key pad or the Up/Down arrow key.
Step 4. Press
Step 5. Press the
Step 6. Enter the desired stop frequency using the keypad or the Up/Down arrow key.
Step 7. Press
Check that the start and stop frequencies displayed match the desired measure
ment range.
FREQ/DIST key.
F1 soft key.
ENTER to set F1 to the desired frequency.
F2 soft key.
ENTER to set F2 to the desired frequency.
Calibration
For accurate results, the Site Master must be calibrated before making any measurements.
The Site Master must be re-calibrated whenever the setup frequency changes, the temperature exceeds the calibration temperature range or when the test port extension cable is removed or replaced.
The Site Master may be calibrated manually with Open, Short, Load (OSL) calibration
components, or by using the InstaCal module provided with the system.
-
-
If a Test Port Extension Cable is to be used, the Site Master must be calibrated with the
Test Port Extension Cable in place. The Test Port Extension Cable is a phase stable cable
and is used as an extension cable on the test port to ensure accurate and repeatable measurements. This phase stable cable can be moved and bent while making a measurement with
out causing errors in the measurement.
NOTE: The test port extension cable should have the appropriate connectors
for the measurement. Use of additional connector adapters after the test port
extension cable can contribute to measurement errors not compensated for
during calibration.
-
3-2
Calibration Verification
Chapter 3 Getting Started
During the calibration process in Return Loss mode, either with discrete calibration compo
nents or with the InstaCal module, there are typical measurement levels expected. Verifying
the measurement levels displayed on the screen during calibration can save valuable time in
the field.
Trace Characteristics in Return Loss Mode
As the discrete calibration components are connected to the Site Master RF out port, the
following measurement levels will be displayed on the screen:
When an OPEN is connected, a trace will be displayed between 0-10 dB.
q
When a SHORT is connected, a trace will be displayed between 0-10 dB.
q
When a LOAD is connected, a trace will be displayed between 0-50 dB.
q
When an InstaCal module is connected to the Site Master RF out port, the following mea
-
surement levels will be displayed on the screen:
When the Site Master is measuring an equivalent OPEN, a trace will be displayed
q
between 0-20 dB.
When the Site Master is measuring an equivalent SHORT, a trace will be displayed
q
between 0-20 dB.
When the Site Master is measuring an equivalent LOAD, a trace will be displayed
q
between 0-50 dB.
-
The following procedures explain Manual and InstaCal calibration methods. Refer to Figure
3-2 for a calibration setup diagram.
3-3
Chapter 3 Getting Started
Manual Calibration Procedure
If the “CAL OFF” message is displayed, or the test port cable has been changed, a new cali
bration is required. The following procedure details how to perform the calibration.
OPEN
RFOUT/REFLECTION
TEST PORT
Site Master S332C
MODE
FREQ/DIST
AMPLITUDE
SWEEP
SHORT
LOAD
TEST PORT CABLE (OPTIONAL)
ESCAPE
CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
RUN
SAVE
DISPLAY
DISPLAY
HOLD
+
9
0
-
/
PRINT
ON
SYS
OFF
.
CALIBRATION
-
Figure 3-2.Calibration Setup
Step 1. Select the appropriate frequency range, as described in the procedure above.
Step 2. Press the
Out PORT” will appear in the display.
Step 3. Connect the calibrated Open and press the
suring OPEN” and “CONNECT SHORT TO RF Out” will appear.
Step 4. Remove the Open, connect the calibrated Short and press the ENTER key. The
messages “Measuring SHORT” and “CONNECT LOAD TO RF Out” will ap
pear.
Step 5. Remove the Short, connect the calibrated Termination and press the
key. The messages “Measuring LOAD” will appear.
Step 6. Verify that the calibration has been properly performed by checking that the
“CAL ON” message is now displayed in the upper left corner of the display.
START CAL key. The message “CONNECT OPEN or InstaCal TO RF
ENTER key. The messages “Mea
-
-
ENTER
3-4
Chapter 3 Getting Started
InstaCal Module Verification
Verifying the InstaCal module before any line sweeping measurements is critical to the
measured data. InstaCal module verification identifies any failures in the module due to cir
cuitry damage or failure of the control circuitry. This test does not attempt to characterize
the InstaCal module, which is performed at the factory or the service center.
The performance of the InstaCal module can be verified by the Termination method or the
Offset method. The termination method is the preferred method in the field, and is similar
to testing a bad load against a known good load.
Termination Method
-
The Termination method compares a precision load against the InstaCal Module and pro
vides a baseline for other field measurements. A precision load provides better than 42 dB
directivity.
Step 1. Set the frequency according to the device under test (cellular, PCS, GSM).
Step 2.
Step 3. Connect the InstaCal module to the Site Master RF OUT port and calibrate the
Step 4. Remove the InstaCal module from the RF OUT port and connect the precision
Step 5. Measure the return loss of the precision load. The level should be less than 35
Step 6.
Step 7. Press
Offset Method
An alternative to the termination method is to measure the return loss of a 20 dB offset.
This is similar to measuring an antenna that has been specified to have a 20 dB return loss
across the frequency of operation. A 20 dB offset provides a 20 dB return loss across a very
wide frequency range. Measuring the return loss with the 20 dB offset will provide a rela
tively flat response across the operating frequency range of the Site Master models (S113C,
S114C, S331C, and S332C).
Press the
Site Master using the InstaCal module (page 3-6).
load to the RF OUT port.
dB across the calibrated frequency range.
Press the
M1 value should be less than 35 dB return loss.
MODE key and select FREQ-RETURN LOSS mode.
MARKER key and set the M1 marker to MARKER TO PEAK. The
SAVE DISPLAY (page 3-8) name the trace, and press ENTER.
-
-
Step 1. Set the frequency according to the device under test (cellular, PCS, GSM).
Step 2.
Step 3. Connect the InstaCal module to the Site Master RF OUT port and calibrate the
Step 4. Remove the InstaCal module from the RF OUT port and connect the 20 dB Off
Step 5. Measure the return loss of the 20 dB Offset. The level should be 20 dB, ±2dB
Step 6.
Step 7. Press
Press the
Site Master using the InstaCal module (page 3-6).
set to the RF OUT port.
across the calibrated frequency range.
Press the
M1 value should be approximately 20 dB return loss.
MODE key and select FREQ-RETURN LOSS mode.
MARKER key and set the M1 marker to MARKER TO PEAK. The
SAVE DISPLAY (page 3-8) name the trace, and press ENTER.
3-5
-
Chapter 3 Getting Started
InstaCal Module Calibration Procedure
The InstaCal module automatically calibrates the Site Master using the OSL method.
NOTE: The InstaCal module is not a discrete calibration component and it can
not be used at the top of the tower to perform line sweep measurements.
Check that the “CAL OFF” message is displayed in the upper left corner of the display.
This indicates that the Site Master has not been calibrated. The following procedure details
how to perform the calibration using the InstaCal module.
Step 1. Select the appropriate frequency range, as described in the procedure above.
Step 2. Press the
START CAL key. The message “CONNECT OPEN or InstaCal TO RF
Out PORT” will appear in the display.
Step 3. Connect the InstaCal module to the RF Out port.
Step 4. Press the
ENTER key. The Site Master senses the InstaCal module and automati
cally calibrates the unit using the OSL procedure. The calibration should take
about 45 seconds.
Step 5. Verify that the calibration has been properly performed by checking that the
CAL ON! message is displayed in the upper left corner of the display.
InstaCal
1
START
CAL
3
SAVE
SETUP
5
LIMIT
7
SAVE
DISPLAY
9
ON
OFF
MODULE
ESCAPE
CLEAR
2
AUTO
SCALE
4
RECALL
SETUP
6
MARKER
ENTER
8
RECALL
RUN
DISPLAY
HOLD
+
0
PRINT
SYS
.
-
/
ModelICN50
10M
z-4.0G
H
InstaCal
z
H
Site Master S332C
MODE
FREQ/DIST
AMPLITUDE
SWEEP
-
Figure 3-3.InstaCal Module Calibration
Calibration with the Test Port Extension Cable
If a Test Port Extension Cable is to be used, the Site Master must be calibrated with the
Test Port Extension Cable in place. Follow the same calibration procedures as above with
the OSL components or the InstaCal module in place at the end of the test port extension
cable.
3-6
Chapter 3 Getting Started
Setting the Scale
Auto Scale
The Site Master can automatically set the scales to the minimum and maximum values of
the measurement on the y-axis of the display. This function is particularly useful for mea
surements in SWR mode. To automatically set the scales:
-
Step 1. Press the
The Site Master will automatically set the top and bottom scales to the minimum
and maximum values of the measurement on the y-axis of the display.
AUTO SCALE key.
Amplitude Scale
The following procedure sets the top and bottom scale display.
Step 2. Press the
Step 3. Press the
scale value.
Step 4. Press the
the bottom scale value.
NOTE: Typically the y-axis scale of the display is 0-54 dB (return loss) but for
some measurements (for example, insertion loss) the scale should be changed
to 0-10 dB. If the scale is not changed, some measurement results may not be
easily displayed on the screen.
AMPLITUDE key to call up the Scale Menu.
TOP soft key and use the keypad or Up/Down arrow key to edit the top
BOTTOM soft key and use the keypad or Up/Down arrow key to edit
Save and Recall a Setup
Saving a Setup
Saving a setup configuration in memory will preserve the calibration information.
Step 1. To save the configuration in one of the 10 available user setup locations, press
SAVE SETUP .
Step 2. Use the key pad or the Up/Down arrow key to select a location (1 - 10).
Step 3. Press
NOTE: A manual calibration setup will be saved with an OSL designator. An
InstaCal setup will be saved with an ICAL designation.
ENTER to save the setup.
3-7
Chapter 3 Getting Started
Recalling a Setup
The following procedure recalls a setup from memory.
Step 1. Press the
Step 2. Select the desired setup using the Up/Down arrow key.
Step 3. Press
RECALL SETUP key.
ENTER to recall the setup.
Save and Recall a Display
Saving a Display
The following procedure saves a display to memory.
Step 1. Press the
age.
Step 2. Use the soft keys to enter a label for the saved trace.
For example, to save a display with the name “TX1 RETURN LOSS” press the
soft key group that contains the letter “T” then press the “T” soft key. Press the
soft key group that contains the letter “X” then press the “X” soft key. Press the
number “1” key on the numeric keypad. Use the softkeys and keypad as necessary to enter the entire name, then press
NOTE: More than one trace can be saved using the same alphanumeric name,
as traces are stored chronologically, using the time/date stamp.
SAVE DISPLAY key to activate the alphanumeric menu for trace stor
ENTER to complete the process.
-
Recalling a Display
The following procedure recalls a previously saved display from memory.
Step 1. Press the
Step 2. Select the desired display using the Up/Down arrow key.
Step 3. Press
RECALL DISPLAY key.
ENTER to recall the display.
3-8
Chapter 3 Getting Started
Set the Distance and Cable Type
In Distance-To-Fault (DTF) mode, the length of the transmission line (distance) and cable
type are selected. The cable type determines the velocity propagation and cable attenuation
factor. The following procedure can be used to set the distance and select the appropriate
cable type.
NOTE: Selecting the correct cable is very important for accurate measurements
and for identifying faults in the transmission line. Selecting the incorrect cable
type will shift the DTF trace vertically and horizontally making it difficult to accu
rately locate faults.
-
Step 1. Press the
Step 2. Select
sets D1 to zero.
Step 3. Press the
Step 4. Enter the appropriate D2 value for the maximum length of the transmission line
and press the
Step 5. Press the
Step 6. Using the
stored in the Site Master cable list. This standard list may not be edited. A custom cable list can also be created.
Step 7. Using the
cable. The selected cable type, PROP VEL and CABLE LOSS in dB/ft (or
dB/m) will be displayed.
Step 8. Press
MODE key.
DTF RETURN LOSS or DTF SWR mode. The Site Master automatically
D2 soft key.
ENTER key to set the D2 value.
DTF AID soft key.
Up/Down arrow key, select CABLE TYPE. Standard cable types are
Up/Down arrow key, select the appropriate list and select the type of
ENTER.
Changing the Units
By default, the Site Master displays information in metric units. Use the following proce
dure to change the display to English units.
Step 1.
Press the SYS key.
-
Step 2. Select the
Step 3. Press
The current selection is displayed at the bottom left corner of the screen.
OPTIONS soft key.
UNITS to change from metric to English measurement units, or vice versa.
Changing the Display Language
By default, the Site Master displays messages in English. To change the display language:
Step 1.
Step 2. Select the
Step 3. Select the desired language. Choices are English, French, German, Spanish, Chi
Press the SYS key.
Language soft key.
nese, and Japanese. The default language is English.
-
3-9
Chapter 3 Getting Started
O
S
Printing
Printing is accomplished by selecting an available printer and pressing the print key as de
scribed below. Refer to the particular printer operating manual for specific printer settings.
Printing a Screen
Step 1. Connect the printer as shown in Figure 3-4.
SEIKO
PRINTER
Site Master S332C
MODE
FREQ/DIST
SWEEP
AMPLITUDE
Figure 3-4-.Site Master Printer Setup
ESCAPE
CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
RUN
SAVE
DISPLAY
DISPLAY
HOLD
+
9
0
-
/
PRINT
ON
SYS
OFF
.
SERIAL CABLE
2000-1012
SERIAL-TO-PARALLEL
INTERFACECABLE
2000-753
HP 350
DESKJET
-
Step 2. Obtain the desired measurement display
Step 3. Press the
Figure 3-5-.SYS Key and OPTIONS Soft Key
SYS key and the OPTIONS soft key (Figure 3-5) .
Site Master S332C
FREQ/DIST
MODE
AMPLITUDE
SELF TEST
SWEEP
OPTIONS
CLOCK
STATUS
PTION
ESCAPE
CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
RECALL
SAVE
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
SAVE
DISPLAY
ON
OFF
RUN
DISPLAY
HOLD
+
9
0
-
/
PRINT
SYS
.
SYS
3-10
Chapter 3 Getting Started
PRINTER
Step 4. Press the PRINTER soft key and select from the displayed menu of supported
printers.
Step 5. Press the
PRINT key. (Figure 3-6).
MODE
FREQ/DIST
Site Master S332C
PRINTER
CHANGE
DATE
AMPLITUDE
SWEEP
START
CAL
SAVE
SETUP
LIMIT
SAVE
DISPLAY
OFF
ON
ESCAPE
CLEAR
2
1
AUTO
SCALE
3
4
RECALL
SETUP
6
5
MARKER
ENTER
8
7
RECALL
RUN
DISPLAY
HOLD
PRINT
+
0
-
/
SYS
.
9
PRINT
Figure 3-6-.PRINTER Soft Key and PRINT Key
Printer Switch Settings
Set the switches, SW1, SW2, and SW3, on the Seiko DPU-414 thermal printer as follows:
Set the switches on the serial-to-parallel interface cable to the HP Deskjet 350 ink jet
printer as follows:
SW1
OFFONOFFOFFOFFOFFONOFF
SW2SW3SW4SW5SW6SW7SW8
3-11
Chapter 3 Getting Started
Using the Soft Carrying Case
The soft carrying case has been designed such that the strap can be unsnapped to allow the
case to be easily oriented horizontally; thus allowing the Anritsu controls to be more easily
accessed (Figure 3-7).
Figure 3-7.Using the Site Master Soft Carrying Case
3-12
Chapter 4
Cable & Antenna
Measurements
Introduction
This chapter provides a description of cable and antenna measurements, including line
sweeping fundamentals and line sweeping measurement procedures, available when the
Site Master is in frequency or DTF mode.
Line Sweep Fundamentals
In wireless communication, the transmit and receive antennas are connected to the radio
through a transmission line. This transmission line is usually a coaxial cable or waveguide.
This connection system is referred to as a transmission feed line system. Figure 4-1 shows
an example of a typical transmission feed line system.
Figure 4-1.A Typical Transmission Feed Line System
4
4-1
Chapter 4 Cable & Antenna Measurements
The performance of a transmission feed line system may be effected by excessive signal re
flection and cable loss. Signal reflection occurs when the RF signal reflects back due to an
impedance mismatch or change in impedance caused by excessive kinking or bending of
the transmission line. Cable loss is caused by attenuation of the signal as it passes through
the transmission line and connectors.
To verify the performance of the transmission feed line system and analyze these problems,
three types of line sweeps are required:
Return Loss Measurement¾Measures the reflected power of the system in decibels (dB).
This measurement can also be taken in the Voltage Standing Wave Ratio (VSWR) mode,
which is the ratio of the transmitted power to the reflected power. However, the return loss
measurement is typically used for most field applications.
Insertion Loss Measurement¾Measures the energy absorbed, or lost, by the transmission
line in dB/ft or dB/meter. Different transmission lines have different losses, and the loss is
frequency and distance specific. The higher the frequency or longer the distance, the greater
the loss.
Distance-To-Fault (DTF) Measurement¾Reveals the precise fault location of compo
nents in the transmission line system. This test helps to identify specific problems in the
system, such as connector transitions, jumpers, kinks in the cable or moisture intrusion.
The different measurements are defined as:
-
-
Return Loss - System Sweep¾A measurement made when the antenna is connected at the
end of the transmission line. This measurement provides an analysis of how the various
components of the system are interacting and provides an aggregate return loss of the entire
system.
Distance To Fault - Load Sweep¾A measurement made with the antenna disconnected
and replaced with a 50W precision load at the end of the transmission line. This measure
ment allows analysis of the various components of the transmission feed line system in the
DTF mode.
Cable Loss - Insertion Loss Sweep¾A measurement made when a short is connected at
the end of the transmission line. This condition allows analysis of the signal loss through
the transmission line and identifies the problems in the system. High insertion loss in the
feed line or jumpers can contribute to poor system performance and loss of coverage.
This whole process of measurements and testing the transmission line system is called Line
Sweeping.
Information Required for a Line Sweep
The following information must be determined before attempting a line sweep measure
ment:
q
System Frequency Range, to set the sweep frequency
q
Cable Type, to set the cable characteristics for DTF measurements
-
-
4-2
q
Distance of the Cable Run, to set the distance for DTF measurements
Chapter 4 Cable & Antenna Measurements
Typical Line Sweep Test Procedures
This section provides typical line sweep measurements used to analyze the performance of
a transmission feed line system.
System Return Loss Measurement
System return loss measurement verifies the performance of the transmission feed line sys
tem with the antenna connected at the end of the transmission line. To measure the system
return loss:
Optional 510-90 Adapter, DC to 7.5 GHz, 50 ohm, 7/16(F)-N(M)
q
Anritsu InstaCal Module, ICN50
q
Device Under Test
Transmission Feed Line with Antenna
q
Procedure
Step 1. Press the
Step 2.
Select FREQ-RETURN LOSS using the Up/Down arrow key and press
ENTER.
MODE key.
-
Step 3. Set the start and stop frequencies, F1 and F2, as described on page 3-2.
Step 4. Calibrate the Site Master as described on page 3-2.
Step 5. Connect the Device Under Test to the Site Master. A trace will be displayed on
the screen when the Site Master is in the sweep mode.
Step 6. Press
SAVE DISPLAY (page 3-8) name the trace, and press ENTER.
NOTE: The antenna must be connected at the end of the transmission feed line
when conducting a System Return Loss measurement.
4-3
Chapter 4 Cable & Antenna Measurements
Figure 4-2 is an example of a typical system return loss trace:
Figure 4-2.Typical System Return Loss Trace
NOTE: The system sweep trace should appear at an approximate return loss of
15 dB (±3 dB) on the display. In general, a 15 dB return loss is measured in the
passband of the antenna system.
Insertion Loss Measurement
The transmission feed line insertion loss test verifies the signal attenuation level of the cable system in reference to the specification. This test can be conducted with the Site Master
in either FREQ–CABLE LOSS or FREQ–RETURN LOSS mode. In Cable Loss mode, the
Site Master automatically considers the signal traveling in both directions when calculating
the insertion loss, making the measurement easier for the user in the field. Both methods are
explained below.
Required Equipment
q
Site Master Model S113C, S114C, S331C, or S332C
q
Precision Open/Short, Anritsu 22N50 or
Precision Open/Short/Load, Anritsu OSLN50LF
q
Precision Load, Anritsu SM/PL
q
Test Port Extension Cable, Anritsu 15NNF50-1.5C
q
Optional 510-90 Adapter, DC to 7.5 GHz, 50 ohm, 7/16(F)-N(M)
q
Anritsu InstaCal Module, ICN50
Device Under Test
q
Transmission Feed Line with Short
Procedure - Cable Loss Mode
4-4
Step 1. Press the
Step 2.
Select FREQ-CABLE LOSS using the Up/Down arrow key and press
MODE key.
ENTER.
Chapter 4 Cable & Antenna Measurements
Step 3. Set the start and stop frequencies, F1 and F2, as described on page 3-2.
Step 4. Connect the Test Port Extension cable to the RF port and calibrate the Site Mas
ter as described on page 3-2.
Step 5. Save the calibration set up (page 3-7).
Step 6. Connect the Device Under Test to the Site Master phase stable Test Port Exten
sion cable. A trace will be displayed on the screen as long as the Site Master is
in sweep mode.
Step 7.
Press the
AMPLITUDE key and set the TOP and BOTTOM values of the dis
-
play. In the example below, the TOP is set to 2, and the BOTTOM is set to 5.
Step 8. Press the
Step 9.
Step 10.
Set M1 to MARKER TO PEAK.
Set M2 to MARKER TO VALLEY.
MARKER key.
Step 11. Calculate the measured insertion loss by averaging M1 (marker to peak) and M2
(marker to valley) as follows:
MM
InsertionLoss
Step 12. Press
SAVE DISPLAY (page 3-8) name the trace, and press ENTER.
=
+12
2
Step 13. Verify the measured insertion loss against the calculated insertion loss. For
Number of connector pairs (3) times the loss per pair (in dB) equals the connec
tor loss: 3 ´ 0.14 = 0.42.
The insertion loss of the transmission system is equal to:
First Jumper loss + Main Feeder Loss + Top Jumper Loss + Connector Loss:
0.65+2.79+0.325+0.42 = 4.19 dB
4-5
-
Chapter 4 Cable & Antenna Measurements
Figure 4-3 is an example of a typical transmission line cable loss trace.
Figure 4-3.Typical Transmission Line Cable Loss Trace
Procedure - Return Loss Mode
Step 1. Press the
Step 2.
Select FREQ-RETURN LOSS using the Up/Down arrow key and press
ENTER.
MODE key.
Step 3. Set the start and stop frequencies, F1 and F2, as described on page 3-2.
Step 4. Connect the Test Port Extension cable to the RF port and calibrate the Site Mas-
ter as described on page 3-2.
Step 5. Save the calibration set up (page 3-7).
Step 6. Connect the Device Under Test to the Site Master phase stable Test Port Exten-
sion cable. A trace will be displayed on the screen as long as the Site Master is
in sweep mode.
Step 7.
Press the
AMPLITUDE key and set TOP and BOTTOM values of the display.
In the example below, the TOP is set to 4, and the BOTTOM is set to 10.
Step 8. Press the
Step 9.
Step 10.
Set M1 to MARKER TO PEAK.
Set M2 to MARKER TO VALLEY.
MARKER key.
Step 11. Calculate the insertion loss by averaging M1 (marker to peak) and M2 (marker
to valley) and dividing by two as follows:
MM
+12
InsertionLoss
2
=
2
4-6
Step 12. Press
SAVE DISPLAY (page 3-8) name the trace, and press ENTER.
Chapter 4 Cable & Antenna Measurements
Figure 4-4 is an example of a typical insertion loss measurement in return loss mode.
Figure 4-4.Typical Return Loss Trace
4-7
Chapter 4 Cable & Antenna Measurements
Distance-To-Fault (DTF) Transmission Line Test
The Distance-To-Fault transmission line test verifies the performance of the transmission
line assembly and its components and identifies the fault locations in the transmission line
system. This test determines the return loss value of each connector pair, cable component
and cable to identify the problem location. This test can be performed in the DTF–RE
TURN LOSS or DTF–SWR mode. Typically, for field applications, the DTF–RETURN
LOSS mode is used. To perform this test, disconnect the antenna and connect the load at the
end of the transmission line.
Optional 510-90 Adapter, DC to 7.5 GHz, 50 ohm, 7/16(F)-N(M)
q
Anritsu InstaCal Module, ICN50
q
Device Under Test
Transmission Feed Line with Load
q
-
Procedure - Return Loss Mode
The following steps explain how to make a DTF measurement in return loss mode.
Step 1. Press the
Step 2.
Select DTF-RETURN LOSS using the Up/Down arrow key and press
MODE key.
ENTER.
Step 3. Connect the Test Port Extension cable to the RF port and calibrate the Site Mas-
ter as described on page 3-2.
Step 4. Save the calibration set up (page 3-7).
Step 5. Connect the Device Under Test to the Site Master phase stable Test Port Exten
sion cable. A trace will be displayed on the screen as long as the Site Master is
in sweep mode.
Step 6. Press the
Step 7. Set the
Step 8.
Press the
FREQ/DIST key.
D1 and D2 values. The Site Master default for D1 is zero.
DTF AID soft key and select the appropriate CABLE TYPE to set the
correct propagation velocity and attenuation factor.
NOTE: Selecting the right propagation velocity, attenuation factor and distance
is very important for accurate measurements, otherwise the faults can not be
identified accurately and insertion loss will be incorrect.
-
4-8
Step 9. Press
SAVE DISPLAY (page 3-8) name the trace, and press ENTER.
Step 10. Record the connector transitions.
Chapter 4 Cable & Antenna Measurements
Figure 4-5 shows an example of a typical DTF return loss trace.
Figure 4-5.Typical DTF Return Loss Trace
In the above example:
Marker M1 marks the first connector, the end of the Site Master phase stable Test
q
Port Extension cable.
q
Marker M2 marks the first jumper cable.
q
Marker M3 marks the end of the main feeder cable.
q
Marker M4 is the load at the end of the entire transmission line.
Procedure - DTF-SWR Mode
The following steps explain how to measure DTF in SWR mode.
Step 1. Press the
Step 2. Select the
MODE key.
DTF-SWR using the Up/Down arrow key and press ENTER.
Step 3. Follow the same procedure as DTF-Return Loss Mode, above.
4-9
Chapter 4 Cable & Antenna Measurements
Antenna Subsystem Return Loss Test
Antenna Subsystem return loss measurement verifies the performance of the transmit and
receive antennas. This measurement can be used to analyze the performance of the antenna
before installation. The antenna can be tested for the entire frequency band, or tested to a
specific frequency range. Transmit and receive frequency measurements are conducted sep
arately. The following steps explain how to measure the antenna loss in return loss mode.
Optional 510-90 Adapter, DC to 7.5 GHz, 50 ohm, 7/16(F)-N(M)
q
Anritsu InstaCal Module, ICN50
q
Device Under Test
Antenna Sub Assembly
q
-
Procedure
Step 1. Press the
Step 2.
Select FREQ-RETURN LOSS using the Up/Down arrow key and press
ENTER.
MODE key.
Step 3. Connect the Test Port Extension cable to the RF port and calibrate the Site Mas-
ter as described on page 3-2.
Step 4. Press
SAVE SETUP and save the calibration set up (page 3-7).
Step 5. Connect the Device Under Test to the Site Master phase stable Test Port Exten
sion cable.
Step 6. Press the
MARKER key.
Step 7. Set markers M1 and M2 to the desired frequency.
Step 8. Record the lowest return loss over the specified frequency.
Step 9. Press
SAVE DISPLAY (page 3-8) name the trace, and press ENTER.
-
4-10
Chapter 4 Cable & Antenna Measurements
The following trace is an example of an antenna return loss trace.
Figure 4-6.Antenna Return Loss Trace
Calculate the threshold value and compare the recorded Lowest Return Loss to the calculated threshold value.
--
201
log VSWR
Maximum Return Loss =
NOTE: VSWR is the antenna manufacturer’s specified VSWR.
If the measured return loss is less than the calculated threshold value, the test fails and the
antenna must be replaced.
()
+
VSWR
()
1
4-11/4-12
Chapter 5
Spectrum Analyzer
Measurements
Introduction
This chapter provides a description of field measurements available when the Model S114C
or S332C Site Master is in Spectrum Analyzer mode.
Occupied Bandwidth
A common measurement performed on radio transmitters is that of occupied bandwidth
(OBW). This measurement calculates the bandwidth containing the total integrated power
occupied in a given signal bandwidth. There are two different methods of calculation de
pending on the technique to modulate the carrier.
99% Method — The occupied frequency bandwidth is calculated as the bandwidth
q
containing 99% of the transmitted power.
X dB Down Method — The occupied frequency bandwidth is defined as the band-
q
width between the upper and lower frequency points at which the signal level is X
dB below the peak carrier level.
Frequency Span900 MHz900 MHz
Reference Level–15 dBm–10 dBm
Input Attenuation LevelAutoAuto
Resolution BandwidthAuto10 kHz
Video BandwidthAuto30 kHz
Channel Power Measurement
One of the most common measurements for a transmitter is that of channel power. This test
measures the output power, or channel power, of a transmitter over a frequency range and a
specific time interval. Channel Power measurements are used to validate transmitter performance, verify compliance with FCC regulations, and keep overall system interference at a
minimum.
Out-of-specification power measurements indicate system faults in the power amplifier or
in-filter circuits.
Channel Power Measurement with the Site Master
The Site Master can be used to measure CDMA signals in terms of channel power. It is nec
essary to use precise and specific settings to accurately measure the CDMA signal because
the CDMA is different from CW signals.
For accurate IS-95 CDMA measurements, use the following settings:
Using the test port extension cable and 30 dB attenuator, connect the Site Master to the
RF_IN test port. The measurement can be accomplished with or without the 30 dB attenu
ator. The attenuator is recommended for additional protection.
-
Step 1. Press the
Step 2. Press the
ON/OFF key on the Site Master.
MODE key.
Step 3. Use the Up/Down arrow key to select the Spectrum Analyzer mode and press
Step 5. Press the
Step 6. Once in the MEAS menu, press the
Step 7. Select the center frequency by pressing the
MEAS key.
CHANNEL POWER soft key.
CENTER FREQ soft key and enter
900 MHz as the center frequency of the CDMA signal.
Step 8. Select the integration bandwidth by pressing the
INT BW soft key and enter 1.23
MHz for the integration bandwidth.
Step 9. Select channel span by pressing the
CHANNEL SPAN soft key and enter 1.23
MHz as the desired integration bandwidth.
Step 10. Make the measurement by pressing the
MEASURE soft key. The Site Master
will display the measurement results in the display area.
Integration bandwidth is defined as the frequency span in which the spectrum analyzer inte
grates measured power readings.
Channel span is the frequency range the analyzer sweeps. Channel span must be set to equal
or greater than the integration bandwidth, or the Site Master will by default set the channel
span equal to the integration span. The Site Master also keeps the ratio of the integration
bandwidth to channel span constant. For example, if the channel span is doubled, the Site
Master will double the integration span.
When the integration bandwidth and channel span are set to the same value, the Site Master
provides the most accurate measurement, as it uses all the sampling points for integration.
5-3
-
Chapter 5 Spectrum Analyzer Measurements
Adjacent Channel Power Measurement
Another common transmitter measurement is that of adjacent channel power ratio. This is
defined as the ratio of the amount of leakage power in an adjacent channel to the total trans
mitted power in the main channel. The current trend is to use this measurement to replace
the traditional two-tone intermodulation distortion (IMD) test for system non-linear behav
ior.
The specification for ACPR requires comparing the power in the RF channel to the power
at several offset frequencies. The result can be expressed either in power ratio or power
density. In order to calculate the upper and lower adjacent channel values, the Site Master
requires four parameters:
Step 1. Using the test port extension cable and 30 dB attenuator connect the Site Master
to the appropriate transmitter test port. The measurement can be accomplished
with our without the 30 dB attenuator. The attenuator is recommended for addi
tional protection.
30 kHz
-
5-4
Step 2. Press the
ON/OFF key on the Site Master.
Chapter 5 Spectrum Analyzer Measurements
Step 3. Press the MODE key.
Step 4. Use the Up/Down arrow key to select the Spectrum Analyzer mode and press
ENTER.
Step 5. Select the appropriate reference level, input attenuation level, resolution band
width, and video bandwidth.
Step 6. Press the
Step 7. Once in the MEAS menu, press the
Step 8. Select the center frequency by pressing the
MEAS key.
ACP soft key.
CENTER FREQ soft key and enter
the desired center frequency.
Step 9. Select the main channel bandwidth by pressing the
MAIN CHANNEL BW soft
key and enter the desired main channel bandwidth. For an IS-95 signal enter
1.23 MHz.
Step 10. Select the adjacent channel bandwidth by pressing the
ADJ CHANNEL BW soft
key and enter the desired adjacent channel bandwidth. For an IS-95 signal enter
30 kHz.
Step 11. Select the channel spacing by pressing the
CHANNEL SPACING soft key and
enter the desired channel spacing. For an IS-95 signal, enter 885 or 1.98 MHz.
Step 12. Make the measurement by pressing the
MEASURE soft key. The Site Master
will display the measurement results in the display area.
The table below provides typical settings for measuring a CW and a IS-95 CDMA signal.
Out-of-band spurious measurements are made on signals outside the system main band.
These signals, which can interfere with other communication systems, can be categorized
into harmonics and random spurious emissions. Real time monitoring of spurious emissions
from a transmitter can uncover unwanted signals before they interfere with other users on
the other channels.
Step 1. Using the test port extension cable and 30 dB attenuator connect the Site Master
to appropriate transmit test port.
Step 2. Press the
Step 3. Press the
Step 4. Use the Up/Down arrow key to select the Spectrum Analyzer mode and press
ENTER.
Step 5. Select the center frequency for the channel of interest.
Step 6. Set the frequency span wide enough to include the primary channel and upper
and lower channel.
Step 7. Set the following parameters:
Step 8. Use the limit line and markers to read and compare the observed signals to the
specified allowable level of out-of-band spurious emissions for the correspond
ing channel transmitted frequency.
ON/OFF key on the Site Master.
MODE key.
reference level=0dBm
input attenuation level = auto
resolution bandwidth =10 kHz
video bandwidth =300 Hz.
-
5-6
Chapter 5 Spectrum Analyzer Measurements
In-Band/Out-of-Channel Measurements
The in-band/out-of-channel measurements are those measurements that measure distortion
and interference within the system band, but outside of the transmitting channel. These
measurements include in-band spurious emissions and adjacent channel power ratio (also
known as spectral regrowth). There are stringent regulatory controls on the amount of inter
ference that a transmitter can spill to neighboring channels. In order to determine compli
ance with the allowable level of spurious emissions, the spectrum analyzer needs two
parameters to be specified:
Step 1. Using the test port extension cable and 30 dB attenuator connect the Site Master
to appropriate transmit test port.
-
-
Step 2. Press the
Step 3. Press the
ON/OFF key on the Site Master.
MODE key.
Step 4. Use the Up/Down arrow key to select the Spectrum Analyzer mode and press
ENTER.
Step 5. Select the center frequency for the channel of interest.
Step 6. Set the frequency span wide enough to include the primary channel and upper
and lower channel.
Step 7. Set the following parameters as specified:
reference level=0dBm
input attenuation level = auto
resolution bandwidth =10 kHz
video bandwidth =300 Hz
Step 8. Use the limit line and markers to read and compare observed signals to the spec
ified allowable level of in-band spurious emissions for the corresponding chan
nel transmitted frequency.
NOTE: The resolution bandwidth of a spectrum analyzer is determined by the
intermediate frequency (IF) filter bandwidth. Site Master, as do conventional
spectrum analyzers, traces the shape of the IF filter as it sweeps past a signal.
Therefore, if two equal-amplitude signals are very close to each other, the mea
surement result can appear to be one single response because the IF or resolu
tion bandwidth is not small enough to resolve the two signals. Similarly, if two
signals are not equal in amplitude but are very close together, the smaller sig
-
nal may not be seen because it is hidden under the large response.
-
-
-
-
5-7
Chapter 5 Spectrum Analyzer Measurements
Field Strength Measurement
The procedure below details steps required making a field strength measurement with the
Site Master.
Required Equipment
Site Master Model S114C or S332C with tracking generator option
q
Antenna of your choice
q
Procedure
Step 1. Press the
ON/OFF key on the Site Master.
Step 2. Use the antenna editor feature of the Site Master Software Tools to define an an
tenna.
Step 3. Upload the antenna information to the Site Master.
Step 4. Press the
Step 5. Select the
Step 6. Press the
MEAS key.
FIELD STRENGTH soft key.
SELECT ANTENNA soft key.
Step 7. Use the Up/Down arrow key to choose the desired antenna and then press the
ENTER key
Step 8. To change the unit of measurement, press the
press the
UNITS soft key to select dB, dBV, dBmV or dBmV.
AMPLITUDE hard key, and then
The Site Master will automatically scale the display by the antenna factors selected.
-
5-8
Antenna Calculations
Chapter 5 Spectrum Analyzer Measurements
The following is a list of various antenna calculations should you find it necessary to con
vert from one to another:
Conversion of signal levels from mW to mV in a 50-ohm system:
2
V
P
=
R
where:P = power in Watts
V = voltage level in Volts
R = resistance in Ohms
-3
For power in milliwatts (10
V
dB V
()( )
Power density to field strength. An alternate measure of field strength to electric field is
power density:
Pd
=
where:E = field strength in V/m
P
=+107
dBmm
2
E
120p
P = Power density in W/m
W), and voltage in microvolts (10-6V):
2
-
Power density at a point:
PtG
P
d
In the far field, where electric and magnetic fields are related by the impedance of free
space:
P
d
Pt= power transmitted in Watts
G
r = distance from the antenna in meters
t
2
r=4
p
= power density in W/m
= gain of transmitting antenna
t
2
5-9/5-10
Chapter 6
Power Measurement
Introduction
The Site Master with Option 5 installed can be used for making power measurements with a
broadband (1 MHz to 3000 MHz) RF detector, Anritsu P/N 5400-71N50. The power moni
tor displays the measured power results in dBm or Watts.
Power Measurement
Required Equipment
Site Master Model S114C or S332C with Power Monitor option
Step 2. Press the
Step 3. Use the Up/Down arrow key to select the Power Monitor mode and press
ENTER.
Zeroing the Power Monitor
Step 4. With no power applied to the DUT, press the
menu. Wait for a few seconds while the Site Master accumulates data samples of
the quiescent power. When complete,
sage area.
Measuring High Input Power Levels
Step 5. Insert an attenuator between the DUT and the RF detector to protect the Site
Master so that the input power level is less than or equal to 20 dBm.
Step 6. Press the
Step 7. Enter the attenuation in dB using the keypad. Press the
the entry. The message area will show
dB.
Displaying Power in dBm and Watts
Step 8. Press the
MODE key.
ZERO soft key from the Power
ZERO ADJ: On is displayed in the mes
OFFSET soft key.
ENTER key to complete
REL: ON along with the entered value in
UNITS soft key to display power in Watts.
-
6
6-1
Chapter 6 Power Measurement
Displaying Relative Power
Step 9. With the desired base power level input to the Site Master, press the
key. The message area will show
100%.
REL soft
REL: ON and the power reading will indicate
Step 10. Press the
reading will be in dBr, relative to the base power level.
UNITS soft key to display power in dBm. Since REL is ON, the power
6-2
Chapter 7
Site Master Software Tools
Introduction
This chapter provides a description of the Site Master Software Tools provided with the
Site Master. Site Master Software Tools (SMST) is a Windows 95/98/2000/ME/NT4 pro
gram for cable and antenna trace analysis. The program help function provides on screen
instructions on display modification, trace overlay, and multiple plot printing.
Features
Site Master Software Tools provides the following features:
Download traces saved in the Site Master memory to the PC for storage and analysis
q
DTF and Smith Chart analysis capability
q
Trace Overlay allows the viewing of two plots simultaneously
q
Ability to save captured plots as data files (.dat file format) or as records representing
q
site information in a database file
Ability to convert Return Loss measurements to Distance-To-Fault measurements
q
-
Ability to zoom in or out to analyze a particular region of the plot
q
q
Ability to modify Plot Properties
q
Ability to output plots to a printer for hard copy records
q
Capture of the current trace in the unit
System Requirements
The Site Master Software Tools program will run on most any computer running Windows
95/98/2000/ME/NT4. Minimum requirements and recommendations are:
q
Intel Pentium 100 MHz microprocessor (Pentium II 350 MHz or better recom
mended)
q
16 MB of RAM (64 MB or above recommended)
q
Hard disk drive with approximately 15 MB of available space (An additional 20 MB
free space for storage of captured plots is recommended.)
q
A serial (COM) port for communication with the Site Master
NOTE: It is recommended that Windows NT 4.0 users install NT 4.0 Service
Pack 3 (SP3) or above. In addition, Windows 2000 and Windows ME may re
quire installation of the latest Service Pack. Please contact Microsoft Corpora
tion for further information on obtaining and installing service packs.
-
-
-
7
7-1
Chapter 7 Software Tools Program
Installation
To install the Site Master Software Tools program:
Step 1. Insert the Anritsu Site Master Software Tools disk in the CDROM drive.
Step 2. From the Windows Start menu, select
Step 3. Type:
Step 4. When prompted, press the
Step 5. When the installation is complete, press
The readme.doc file on the disk provides updated information about the program, and the
Help function provides detailed operating information.
X:\Setup.exe where X is the drive letter of your CDROM drive.
Enter key to accept the default directory C:\Program
Files\Site Master Software Tools and the installation will begin.
Run .
Enter to restart Windows.
Communication Port Setting
The Site Master Software Tools communicates with the Site Master through a standard se
rial COM port on the PC. Set the baud rate of the COM port to 115200.
Step 1. Select
Step 2. When the program has loaded, select
Step 3. Select the appropriate PC COM port and Transfer Baud Rate for your system,
Start, Programs and select Site Master Software Tools.
Settings, Communication.
and click
OK.
-
Figure 7-1.Communication Setting Dialog Box
7-2
Interface Cable Installation
Chapter 7 Software Tools Program
Communication between the Site Master and the PC is accomplished over a null modem se
rial cable provided with the Site Master (Anritsu part number 800-441).
Step 1. Install the null modem serial interface cable to the Serial Interface connector on
the Site Master Test Connector Panel.
Step 2. Connect the other end of the serial interface cable to the appropriate COM port
connector on the PC.
Step 3. Turn on both the Site Master and the PC.
ERIAL INTERFACE
Site Master S332C
MODE
FREQ/DIST
AMPLITUDE
SWEEP
ESCAPE
CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
RUN
SAVE
DISPLAY
DISPLAY
HOLD
+
9
0
-
/
PRINT
ON
SYS
OFF
.
COM PORT
-
Figure 7-2.Serial Cable Connection
Using Software Tools
Select Start, Programs and select Site Master Software Tools.
If the proper command was sent to the Site Master from the PC a “remote” status will be
displayed on the Site Master display.
Downloading Traces from the Site Master
Downloaded traces from the Site Master can be selected as a entire block or as individual
files. Files are categorized by measurement type:
q
Return Loss
q
VSWR
q
DTF (Distance-to-Fault)
q
Time/Date stamp
q
SPA Traces with ACPR and CP
7-3
Chapter 7 Software Tools Program
Plot Capture
To open the plot capture menus, select the capture icon on the button bar, or select the Cap
ture
drop down menu from the menu bar.
Select
Capture to Screen to download traces to the PC.
Highlight a folder, or highlight individual traces within a folder, to be downloaded to the
PC. The traces will appear on the PC display as they are downloaded.
Plot Properties
After downloading, certain plot properties and information can be modified. Select the Plot
Properties or Plot Information icon.
After downloading the plot, the Main Title can be changed to reflect the site name or other
descriptive information. The Sub Title field can be used to describe the specifics of the
measurement and configuration.
-
Display Mode
Display Mode allows changing the display type without having to retest. Plots can be
changed from
Scale/Limit
The scale of the displayed plot can be modified to help analyze whether the plot meets
pass/fail criteria. The Scale/Limit sub menu activates options to manually enter the scale
limits, or to use the Auto Scale mode.
Return Loss to VSWR with single click of mouse button.
Manual adjustment sets the upper and lower limits of the display under the
Scale/Limit Submenu.
Auto Scale automatically adjusts the scale for maximum and minimum mea
surement readings.
Limit Line Off turns off all limit lines.
Single Limit Line can be activated to help identify faults. Typical antenna speci
fications require better than –15 dB return loss.
Multi-Segment Limit Lines can be activated to set different limits within the
same measurement display for specification requirements.
-
-
7-4
Chapter 7 Software Tools Program
Markers
Markers M1 through M6 can be activated from the Plot Properties menu.
The six active markers can be displayed on the plot to help identify faults from the line
sweep data.
Misc.
The Miscellaneous tab allows adjustment of the Plot/Limit Line Width and the setting of
the Plot Footer.
Plot/Limit Line Width can be used when preparing reports for the carrier, ser
vice providers, and network operators where the trace data must be visible and
legible for documentation. The line density of the trace can be set to
(Thin Line)
Plot Footer allows selection of information to be displayed in the trace display.
or Thick Line to make the data appear easily when copies are made.
Selections are:
Bias Tee
q
Time/Date
q
Model and Serial Number
q
Date Format
q
Impedance
q
Measurements
q
DTF Parameters
q
Overlay Indicator
q
Trace Overlay or Plot Overlay
Trace Overlay is activated by the Mouse Function icon. Single-click on the Mouse Function
icon to toggle. Toggling this icon switches the mouse function between Marker/Limit/Zoom
and Plot Overlaying.
To overlay two plots, click-and-drag from one plot to the other plot. The final display will
be the two plots superimposed on one another.
Refer to the SMST
trace overlay.
HELP function for more information on the steps necessary to perform a
-
Normal
Saving Traces
Once the plots are captured to the PC they can be saved as individual files or as a block of
files to a database.
Saving a Plot as a Windows Metafile or to a Spreadsheet
Plots can be saved as a Windows metafile (.wmf) or as a text file (.txt). The metafile may
be imported into graphic programs and the text file can be imported into spreadsheet.
To save a plot as a Windows metafile, click on the
from the pull down menu. Once the trace is saved as a metafile, it can be copied and trans
ferred into other applications as a picture or graphic file.
To copy a metafile:
Step 1. Select the trace to be copied with the mouse cursor.
Step 2. Select
Edit and then Copy. The file will copied to clipboard.
File menu and select Save as Metafile
-
7-5
Chapter 7 Software Tools Program
Step 3. Exit the Site Master Software Tools program and open the target application
(Microsoft Word, etc.).
Step 4. Select
Saving a Plot to a Spreadsheet
Plots can be saved as a text file (.txt) which can then be imported into a spreadsheet
program.
To save a plot as a text file:
Step 1. Select the trace to be copied with the mouse cursor.
Step 2. Click on the
Step 3. Save the file to a local directory.
Step 4. Exit the Site Master Software Tools program and open the spreadsheet applica
Step 5. Import the .txt file into the spreadsheet program.
Edit and then Paste. The file will be inserted as a low-resolution graphic
file or a bitmap.
File menu and select Export to Text File for a Spreadsheet from
the pull down menu.
tion.
Custom Cable List
A custom cable list can be created in Site Master Software Tools and uploaded to the Site
Master. Some standard 1000 MHz, 2000 MHz, and 2500 MHz cables are stored in the Site
Master and are listed in Appendix A of this User’s Guide.
-
The cable list contains vital information such as propagation velocity, Vp and insertion loss
(dB/m or dB/ft). These parameters are important when identifying faults and discontinuities
of transmission lines.
NOTE: Values for the cable list are taken from manufacturers specifications and
are normalized for a sweep frequency setting of 1 GHz. Complex digital com
munications systems may require adjustment to attenuation values at the
higher frequencies. Contact the manufacturer for the appropriate cable losses.
The Cable Editor is specifically designed to provide contractors, service providers, and net
work operators with a list of cable types that are tailored to their use.
To open the Cable Editor, select the
To open an existing cable list file, select
Enter.
press
Creating a Custom Cable List
To create a custom cable list:
Step 1. Select the File menu and select New.
Step 2. Type in or cut and paste the cable types and specifications for the desired cables.
Tools menu and select Cable Editor.
File and Open. Select the desired cable list and
-
-
7-6
Chapter 7 Software Tools Program
Uploading a Cable List
After custom cable list has been created in Site Master Software Tools it can upload it to
the Site Master. The serial communication null modem cable must be connected from the
Site Master to the PC.
In the Cable Editor, select
load with a message " Upload Complete" when the operation is finished.
Querying the Site Master Cable List
Query Cable List allows the user to check and review the contents of the custom cable list
in the Site Master.
In the Cable Editor, select
ble list from the Site Master to the PC.
Tools and Upload Cable List. The software will confirm the up
Tools and Query Cable List. The software will download the ca
Entering Antenna Factors
The antenna factor is a parameter of an antenna that is used in the calculation of field
strength during radiated emissions measurement. It relates the voltage output of the mea
surement antenna to the value of the incident field producing that voltage. The units are
volts output per volt/meter incident field or reciprocal meters. Antennas used for radiated
emissions testing are individually calibrated (the antenna factor is directly measured) at all
appropriate distances. The calibrations produce values that are defined as the “equivalent
free space antenna factor.” The calibration procedure corrects for the presence of the reflection of the antenna in the ground plane, giving the value that would be measured if the
antenna were in “free space.”
Step 1. Once the Site Master Software Tools has been opened, select Antenna Editor
from the Tools menu on the tool bar. A pop-up box will appear on the screen of
the PC.
-
-
-
Step 2. Click on “Edit Antenna” to enter an antenna name, description, frequencies, and
antenna factors. Enter the frequencies in ascending order, starting with lowest
frequency first. A maximum of 60 antenna factors may be entered. Use the ar
row keys on the keyboard to move between entries.
NOTES: Only one input of frequency and antenna factor is allowed per row.
Multiple antenna factors for a single antenna must be entered individually. For
example, an antenna having an antenna factor of 5 from 2.0 to 2.25 GHz and
an antenna factor of 4 from 2.25 to 2.5 GHz should be entered as follows:
Frequency (MHz)Antenna Factor
20005
22514
25004
If necessary, an antenna factor of zero (0) may be entered.
Step 3. Select Save from the File menu to save the antenna factors to the hard disk.
-
7-7
Chapter 7 Software Tools Program
Uploading Antenna Factors
To upload antenna information from Software Tools to the Site Master:
Step 1. Connect the RS232 cable between the PC and the Site Master.
Step 2. Click on the
menu). It is important to note that the Site Master must sweep very quickly dur
ing the data transfer, at least every 5 seconds, or the Software Tools program
may “time-out.” To improve the chances of a successful upload, increase the
RBW and VBW settings to the maximum.
Step 3. To determine if the antenna information has been successfully uploaded to the
Site Master, press the measure function key (#4 key) and activate the field
strength measurement by pressing the ON/OFF soft key.
Upload button on the tool bar (or select Upload from the Tools
Creating a Database
A single trace or a block of traces can be transferred from the Site Master to the PC. A sep
arate database can be created for each cellular site. The site name can be used as the data
base name.
Step 1. Select the
Step 2. Open an existing database or create a new database with a descriptive filename
that represents the site name.
Once the database has been created, files can be saved and site information can be added,
such as Plot Description, Date/Time, Operator, record/trace number and session description
(Transmitter type etc.)
Individual plots will be labeled with the Database Site Name. Each record has its own plot
description and measurement type (for example, Return Loss-antenna, Return Loss-cable,
and DTF-Return Loss).
File menu and select the Database.
-
-
-
Printing Formats
One to multiple plots per page can be set up under the print format in Site Master Software
Tools. Orientation of how the plots are printed on the page can be changed from vertical to
horizontal.
7-8
Appendix A
Reference Data
Coaxial Cable Technical Data
Table A-1 provides a standard listing of common coaxial cables along with their Relative
Propagation Velocity and Nominal Attenuation values in dB/m @1000, 2000, and
2500 MHz. (N/A indicates that the specification is not applicable to the listed cable.)
The Distance sub-menu (page 2-10) provides for setting the cable loss and relative propaga
tion velocity of the coaxial cable. The WINDOW key opens a menu of FFT windowing
types for the DTF calculation.
The theoretical requirement for inverse FFT is for the data to extend from zero frequency to
infinity. Side lobes appear around a discontinuity due to the fact that the spectrum is cut off
at a finite frequency. Windowing reduces the side lobes by smoothing out the sharp transi
tions at the beginning and at the end of the frequency sweep. As the side lobes are reduced
the main lobe widens thereby reducing the resolution.
In situations where there may be a small discontinuity close to a large one, side lobe reduc
tion windowing should be used. When distance resolution is critical windowing can be re
duced.
Examples
The types of windowing in order of increasing side lobe reduction are: rectangular, nominal
side lobe, low side lobe, and minimum side lobe. Figures B-1 thru B-4 show examples of
the types of windowing.
D i s t a n c e T o F a u l t
-
-
-
-
- 5
- 1 0
- 1 5
- 2 0
- 2 5
R e t u r n L o s s ( d B )
Figure B-1.Rectangular Windowing Example
- 3 0
- 3 5
- 4 0
- 4 5
- 5 0
1 0
5
1 5
2 5
2 0
3 03 54 0
F e e t
4 55 0
5 56 0
B-1
Appendix B Windowing
R e t u r n L o s s ( d B )
D i s t a n c e T o F a u l t
- 5
- 1 0
- 1 5
- 2 0
- 2 5
- 3 0
- 3 5
- 4 0
- 4 5
- 5 0
1 0
5
2 02 5
1 5
Figure B-2.Nominal Side Lobe Windowing Example
- 5
- 1 0
- 1 5
- 2 0
- 2 5
R e t u r n L o s s ( d B )
- 3 0
- 3 5
- 4 0
- 4 5
3 0
F e e t
D i s t a n c e T o F a u l t
3 5
4 5
4 0
5 5
5 0
6 0
- 5 0
51 0
1 5
Figure B-3.Low Side Lobe Windowing Example
B-2
2 02 53 0
3 54 04 5
F e e t
5 05 5
6 0
Appendix B Windowing
D i s t a n c e T o F a u l t
- 5
- 1 0
- 1 5
- 2 0
- 2 5
R e t u r n L o s s ( d B )
- 3 0
- 3 5
- 4 0
- 4 5
- 5 0
5
1 0
2 02 5
1 5
Figure B-4.Minimum Side Lobe Windowing Example
3 03 54 0
F e e t
4 5
5 56 0
5 0
B-3/B-4
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.