permission is prohibited, except as allowed under the copyright laws.
Printed in the UK.
iiAgilent E4418B/E4419B Service Guide
Legal Informat io n
Legal Information
Notice
Information contained in this document is subject to change without notice. Agile nt
Technologies makes no warranty of any kind with regard to this material, including,
but not limite d to, the implied warranties of mercha ntability and fitness for a
particula r purpose. Agilent T echnologies shall not be liable for errors contained
herein or for incidental or consequential damages in conn ection with the furnishings,
performance, or use of this material. No part of this document may be photocopi ed,
reproduced, or tr anslated to another language without the prior written consent of
Agilent Technologie s.
Certification
Agilent Technologies certifies that th is product met its published specifications a t the
time of shipment from the factory. Agilent Technologies further certifies that its
calibration measurements are traceable to the United States National Institute of
Standards and Technology, to the extent allowed by the Institute’s calibration facility,
and to the calib ration facilities of other International Standards Or ganization
members.
Warranty
This Agilent Technologies instrument pr oduct is war ranted a gainst de fe cts in ma teria l
and workmanship for a per iod of one ye ar f rom date of shipm ent. During the war ranty
period, Agile nt Technologies will at its opt ion , eith er repai r or repla ce produc ts whi ch
prove to be defective. For warranty service or repai r, this product must be returned to
a service facility designated by Agilent Technologies. Buyer shall prepay shipping
charges to Agilent T echn ologies and Agile nt T echnologi es shall pay shippi ng char ges,
duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies warrant s that its software and firmware designated by Agilent
Technologies for use with an instrument will execute its programming instructions
when properly ins talled on that instrument. Agil ent Technologies does not warrant
that the operation of the instrument, or firmware will be uninterrupted or error free.
Agilent E4418B/E4 419B S ervice Guideiii
Legal Information
Limitation of Warranty
The foregoing warrant y sha ll not apply to defects resulting from improper or
inadequate maintenance by Buyer, Buyer-supplied software or interfacing,
unauthorize d modi fication or misuse, operation outside of the environmental
specifications for the product, or improper site preparation or maintenance. NO
OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT
TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Exclusive Remedies
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND E XCLUSIVE
REMEDIES. HP SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT,
SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER
BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ivAgilent E4418B/E4419B Service Guide
Equipment Operation
Equipment Operation
Warnings and Cautions
This guide uses warnings and cautions to denote hazards.
WARNING A warning calls attention to a procedure, practice or the like, which, if
not correctly performed or adhered to, could result in injury or the loss
of life. Do not proceed beyond a warning until the indicated conditions
are fully understood and met.
Caution A caution calls attention to a procedure, practice or the like which, if not
correctly perf orm ed or adhered to, could result in damage to or the
destruction of part or all of the equipment. Do not pr oce ed beyond a caution
until the indicated conditions are fully understood and met.
Personal Safety Considerations
WARNING This is a Safety Class I product (provided with a protective earthing
ground inco r p o ra t ed in the pow er co rd ). The main s p lu g sh a l l onl y be
inserted in a socket outlet provided with a protective earth co nt act. Any
interruptio n of the protective conductor, inside or outside the
instrument, is likely to make the instrument dangerous. Intentional
interrup t ion is prohibited .
If this instrument is not used as specified, the protection provided by the
equipment could be impaired. This instrument must be used in a normal
condition (in which all means of protection are intact) only.
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
For continued protection against fire hazard, replace the line fuse(s)
only with fuses of the same type and rating (for example, normal blow ,
time delay, etc.). The use of other fuses or material is prohibited.
Agilent E4418B/E4 419B S ervice Guidev
Equipment Operation
ISM
General Safety Considerations
WARNING Before this instr u m en t is s w i t ched on, ma k e su re it has been prop erly
grounded through the protective conductor of the ac power cable to a
socket outlet provided with protective earth contact.
Any interruption of the protective (grounding) conductor, inside or
outside the instrument, or disconnection of the protectiv e earth terminal
can result in personal inju ry.
Caution Any adjustments or service procedures that require operation of the
instrument with protective covers removed should be per formed only by
trained service personnel.
Markings
The CE mark shows that the product complies with all the
relevant European legal Directives (if accompanied by a
year, it signifies when the design was proven.
Group 1
Class A
viAgilent E4418B/E4419B Service Guide
This is the symbol of an Indust rial Scientific and Medical
Group 1 Class A product.
The CSA mark is a registered trademark of the Canadian
Standards Association.
External Protective Earth Terminal.
While this is a Class I product, provided with a protectiv e
earthing conductor in a power cord, an external protective
earthing terminal has also been provided. This terminal is for
use where the ea rthing c annot be assured . At lea st an 18AWG
earthing conductor should be used in such an instance, to
ground the instrument to an assured earth terminal.
Equipment Operation
IEC 1010-1 Compliance
This instrument has been designed and tested in accordance with IEC Publication
1010-1 +A1:1992 Safet y Requi rements for Electrical Equipment for Measur ement,
Control and L aborat ory Use and has be en suppl ied in a sa fe condi tion. The i nstruc tion
documentation contains information and warnings which must be followed by the
user to ensure safe operation and to maintain the instrument in a safe condition.
Statement of Compliance
This product has been designed and tested for compliance with IEC 60529 (1989)
Degrees of Protection Provided by Enclosures (IP Code). Leve l IPx 4 is attained if,
and only if, the carry ca s e (Agilent part number 34141A) is fitted.
User Environment
This product is desi gned for use in a sheltered environment (avoiding extreme
weather conditions) in accordance with Pollution De gree 3 defined in IEC 60664-1,
with the carry case ( Agilent part number 34141A) fitted over the inst rument.
The product is suitable for indoor use only, when this carry case is not fitted.
Installation Instructions
To avoid unnecessary over-temperature conditions, while this carry case is fitted do
not apply an ac mains suppl y volt age, only operat e your power mete r from the batt ery
pack.
Agilent E4418B/E4 419B S ervice Guidevii
List of Related Publications
List of Rel ate d P ub li c ati on s
The Agilent E4418B and Agile ntE4419B User’s Guides are also available in the
following languages:
•English Langua ge Us er’s Guide - Standard
•German Language User’ s Guide - Option ABD
•Spanish Language Use r’s Guide - Option ABE
•French Language User’s Guide - Option ABF
•Italian Language User’s Guide - Option ABZ
•Japanese Language User’s Guide - Option ABJ
Agilent E4418B/E4419B Service Guide is available by ordering Option 915.
Agilent E4418B/E4419B CLIPs (Component Location and Information Pack) is
available by ordering E4418-90031.
viiiAgilent E4418B/E4419B Service Guide
Sales and Service Offi ces
Sales and Service Offices
For more information about Agilent Technologies test and measurement
products, applications, services, and for a current sales office listing, visit our
web site: http://www.agilent.com
You can also contact one of the following center s and ask for a test and
measurement sales representative.
Asia Pacific:
Agilent Te chnologies
19/F, Cityplaza One, 1111 King’s Road,
Taikoo Shing, Hong Kong, SAR
(tel) (852) 2599 7889
(fax) (852) 2506 9233
Agilent Te chnologies
Test & Measurement
European Marketing Organization
P. O. Box 999
1180 AZ Amstelvee n
The Netherlands
(tel) (31 20) 547 9999
Latin America:
Agilent Te chnologies
Latin American Region Headquarters
5200 Blue Lagoon Drive, Suite #950
Miami, Florida 33126
U.S.A.
(tel) (305) 267 4245
(fax) (305) 267 4286
United States:
Agilent Te chnologies
Test and Measurement Call Center
P. O. Box 4026
Englewood, CO 80155-4026
(tel) 1 800 452 488
In any correspondence or telephone conversations, refer to the power sensor
by its model nu mber and f ull s erial n umber. With t his i nformation, the Agi le nt
T echnologies representative can quickl y determine whether your unit is still
within its warranty peri od.
xAgilent E4418B/E4419B Service Guide
Sales and Service Offi ces
Returning Your Power Meter for Service
Use the information in this section if you need to return your power meter to Agilent
Technologies.
Package the Power Meter for Sh ipment
Use the foll owing steps to package the power meter for shipment to Agilent
Technologies for service:
1.Fill in a blue ser vice tag (available at the end of this guide) and attach it to
the power meter. Please be as specific as possible about the nature of the
problem. Send a copy of any or all of the following information:
nA n y erro r mes s ag es that appeared on th e p ow er m et er dis p l ay.
nAny information on the performance of the power meter.
Caution Power meter damage can result from using packaging materials other than
those specified. Never use styrene pellets in any shape as packaging
materials. They do not adequately cushion the power meter or prevent it
from shifting in the carton. Styrene pellets cause power meter damage by
generating static electricity and by lodging in the rear panel.
2.Use the original packaging materials or a strong shipping con tainer that is
made of double-wall ed, corrugated cardboard with 159 kg (350 lb) burstin g
strength. The carton must be both large enough and strong en ough to
accommodate the power meter and allow at least 3 to 4 inches on all sides of
the power meter for packing material.
3.Surround the power meter with at least 3 to 4 inches of packing material, or
enough to prevent the powe r meter from moving in the carton. If packing
foam is not available, t he best alternative is SD-240 Air Cap
TM
from Seal ed
Air Corporation (Commerce, CA 90001). Air Cap looks like a plastic sheet
covered with 1-1/4 inch air filled bubbles. Use the pink Air Cap to reduce
static electricity. Wr ap the power meter several times in the material to both
protect the power meter and prevent it from moving in the carton.
4.Seal the shipping container securely with strong nylon adhesive tape.
5.Mark the shipping container “FRAGILE, HANDLE WITH CARE” to
ensure careful handling.
6.Retain copies of all shipping papers.
Agilent E4418B/E4 419B S ervice Guidexi
About this Guide
About this Guide
Chapter 1: Specifications
This chapter li st s the power meter’s specifications and describes how to interpret
these sp ecifica ti o n s .
Chapter 2: Performance Test s
This chapter contains procedures which allow you to tes t the power meter’s electrical
performance to it’s specifications.
Chapter 3: Adjustments
This chapter conta ins checks and adjustments that ensure proper performance of the
power meter.
Chapter 4: Theory of Operation
This chapter describes how each of the power meter’s individual assemblies operate.
Chapter 5: Replaceable Parts
This chapter detai ls the power meter’s replaceable parts. It also explains how to
assemble and dis assemble the power meter.
Chapter 6: Trou blesho oting
This chapter con tains troubleshooting flow charts designed to isolate faults in the
Rmt I/O, GP-IB and RS232/422 interface ports.
This chapter details the power meter’s specificatio n s a nd supp le m ental
characteristics.
Specifications descri be the warranted perfo rma nc e and apply afte r a 30
minute warm-up . The se specifications are valid over the power meter’ s
operating and environmental range unless otherwise stated and after
performing a zero and calibration.
Supplement al char act eristics, which are shown in italics, are inten d ed to
provide infor m ation useful in applyin g the power meter by givi ng ty pi cal,
but non-warranted performance parameters. These characteristics are
shown in italics or denoted as “typical”, “nominal” or “approximate”.
For information on measurement uncertainty calculations, refer to Agilent
Application Note 64-1A, “Fundam e nt als of RF and Micro wave Power
Measurements”, Literature Number 5965-6630.
1-2Agilent E4418B/E4419B Service Guide
Specifications
Power Meter Specifications
Power Meter Specifications
Meter
Frequency R ange
100 kHz to 110 GH z, power sensor dependent
Power Range
-70 dBm to +44 dBm (100 pW to 25 W), power sensor dependent
Power Sensors
Compatible with all Agile nt 8480 serie s pow e r sensors and
Agilen t E-series powe r s e nsors.
Singl e Sensor Dynamic Range
90 dB maximum (Agilent E-series power sensors)
50 dB maximum (Agilent 8480 series power sensors)
Display Units
Absolute: Watts or dBm
Relative: Percent or dB
Display Resolution
Selectable resolution of:
1.0, 0.1, 0.01 and 0.001 dB in logar ith mic mo de, or
1, 2, 3 and 4 significant digits in linear mode
Default Resolution
0.01 dB in logarithmic mode
3 digits in linear mo de
Agilent E4418B/E4419B Service Guide1-3
Specifications
Powe r Me te r S p ecifications
Accuracy
Instrumentation
Absolute:±0.02 dB (Logarithmic) or ±0.5% (Linear). (Refer to the power
sensor linearity specification in your power sensor manual to assess
overall system accu racy.)
Relative:±0.04 dB (Logarithmic) or ±1 .0% (Linear). (Refer to the power
sensor linearity specification in your power sensor manual to assess
overall system accu racy.)
Zero Set (digital settabi li ty of zero) : Power sensor dependent (refer to
Table 1-1). For Agilent E-series power sensors, this specification applies
when zeroing is performed with the sensor input disconnected from the
POWER REF.
1. National met rology institutes of member states of the Metre Convention,
such as the National Institute of Standards and Technology in the USA, are
signator ies to the Comi t é Inte r na tio nal de s Poids et Me su res Mutual
Recognition Arrangem ent. Further information is available from the Bureau
Interna tio na l de s Poids e t Me su re s, at ht tp :/ /w ww.bipm.fr/
2. This SWR specification is only warranted for the E4418B and the E4419B
Power Meters with serial prefix GB4331xxxx and above. Prior to this prefix
the values shown are sup plemental characteristics.
Agilent E4418B/E4419B Service Guide1-5
Specifications
Power Meter Supplemental Characteristics
Power Meter Supplemental Characteristics
Measurement Speed
Over the GP-IB, three mea s urement sp ee d modes are a v ai l a bl e as shown ,
along wit h th e t yp ical maxi m um measurem ent speed fo r ea ch mo de:
•Normal: 20 readings/sec ond
•x2: 40 readi ngs/second
•Fast: 20 0 rea d i ngs/secon d, for Agilen t E-s eries powe r se nsors onl y
Maximum measurement speed is obtained using binary output in free run
trigger m o de.
Zero Drift of Sensors
Power sensor dependent (refer to Table 1-3).
1-6Agilent E4418B/E4419B Service Guide
Specifications
Power Meter Supplemental Character istics
Measurement Noise
Power sensor dependent (refer to Table 1-2 and Table 1-3).
Averaging effects on measurement nois e. Averaging over 1 to 1024
readings is available for redu cin g noise. Table 1-3 provides the
measurement noise for a particular power sensor with the number of
averages set to 16 for normal mode and 32 for x2 mode. Use the “Noise
Multiplier” for the appropriate mode (normal or x2) and number of
averages to determi ne the tot al measurement noise value.
For example, for an Agilent 8481D power sensor in normal mode with the
number of averages set to 4, the measurement noise is equal to:
1. Within 1 hour after zero set, at a constant te mperature, after a 24 hour
warm-up of the power meter.
2. The number of av erages at 16 (for normal m ode) and 32 (for x2 mode), at
a constant temperature, measu red over a 1 minute interval and 2 standard
deviations. For Agilent E-series power sensors the measurement noise is
measured within t h e low range. Refer to t he relevant power sensor manual
for furthe r in formation.
1-8Agilent E4418B/E4419B Service Guide
Power Meter Supplemental Character istics
Settling Time
0 to 99% settled readings over the GP-IB.
For Agilent 8480 series power sensors
Manual filter, 10 dB decreasing power step:
Table 1-4: Settling Time
Specifications
Number of
Averages
Settling Time (s)
(Normal Mode)
Response Time (s)
(x2 mode)
Normal Modex2 Mode
Maximum dBm
Maximum dBm
Maximum dBm
150 ms
Typical
Settling
Settling
Settling
Times
Times
Times
200 ms
500 ms
12481632641282565121024
0.150.20.30.51.11.93.46.6132757
0.150.180.220.350.551.11.93.56.914.533
Auto filter, def a ul t res o lu t ion, 10 dB decreasing powe r st ep,
normal and x2 speed modes:
Maximum dBm
20 dB
20 dB
20 dB
10 dB
10 dB
10 dB
10 dB
10 dB
10 dB
Power
Power
Power
Sensor
Sensor
Sensor
Dynamic
Dynamic
Dynamic
Range
Range
Range
Typical
Settling
Settling
Settling
Times
Times
Times
150 ms
180 m s
350 m s
20 dB
20 dB
20 dB
10 dB
10 dB
10 dB
10 dB
10 dB
10 dB
Power
Power
Power
Sensor
Sensor
Sensor
Dynamic
Dynamic
Dynamic
Range
Range
Range
6.6 s
Minimum dBm
Minimum dBm
Minimum dBm
10 dB
10 dB
10 dB
3.5 s
Minimum dBm
10 dB
10 dB
10 dB
Agilent E4418B/E4419B Service Guide1-9
Specifications
Power Meter Supplemental Characteristics
For Agilent E-series power sensors
Number of
Averages
Settling Time (s)
(Normal Mode)
Response Time (s)
(x2 mode)
In FA ST m o de, within the range -50 dBm to +17 dBm, for a 10 dB
decreasing power step, the settling time is 10 ms
and 20 ms
1
When a decreasing power step crosses the power sensor’s auto-range
1
for the Agilent E4419B
1
for the Agilent E4418B
switch po in t , a dd 25 ms. Refer to th e re l ev a nt power sensor manual for
further information.
F o r Agilent E-series p ower sensors i n normal and x2 speed mode s, manua l
filter, 10 dB decreasing p o wer step:
Table 1-5: Settling Time
12481632641282565121024
0.070.120.210.411.83.36.5132757
0.040.070.120.210.411.83.46.814.232
1-10Agilent E4418B/E4419B Service Guide
Power Meter Supplemental Character istics
c
Auto filter, def a ul t res o lu t ion, 10 dB decreasing powe r st ep,
normal and x2 speed modes:
Normal Modex2 Mode
Specifications
Typical
Settling
Times
Maximum dBm
70 ms
120 ms
400 m s
6.5 s
Minimum dBm
60 dB
10 dB
10 dB
10 dB
Power
Sensor
Dynamic
Range
Typical
Settling
Times
Maximum dBm
40 ms
70 ms
210 ms
3.4 s
60 dB
10 dB
10 dB
10 dB
Minimum dBm
Power
Sensor
Dynami
Range
Agilent E4418B/E4419B Service Guide1-11
Specifications
Power Meter Supplemental Characteristics
Power Sensor Specifications
Definitions
Zero Set
In any power meas ur ement, the power meter must initially be set to zero
with no power applied to the power senso r. Zero setting is accomplished
within the power meter by digitally correcting for residual offsets.
Zero Drift
This parame ter is also called lo ng term stability and is the chan ge in the
power meter indication over a long time (usually one hour) for a constant
input power at a constant temperature , after a defined war m-u p interv al.
Measurement Noise
This paramete r is also kn o wn as sh or t term stability and is specified as
the change in the po wer m ete r in di cation over a short time interval
(usually one minute) for a co nstant inp ut power at a co nstant
temperature.
1-12Agilent E4418B/E4419B Service Guide
Specifications
Battery Option 001 Operational Characteristics
Battery Option 001 Operational Characteristics
The following information describes characteristic performance based at a
temperature of 25
product performance that is useful in the applicati on of the product, but is
not covered by the product warranty.
Typical Operating Time
Up to 2 hours with LED backlight on; up to 3 hou rs with LED backligh t
off.
Charge Time
< 2 hours to charge fully from an empty state: 50 minutes charging
enables 1 hour of operation with L ED backlight ON; 35 minutes charging
enables 1 hour of operation with th e LED backlight OFF. Power meter is
operational whilst charging.
Service Life
°C unless otherwise note d. Ch aracte ri stics describe
To 70% of initial capacity at 25
cycles.
Chemistry
Nickel Metal Hydride.
Weight
1 kg.
°C: approximately 450 charge/discharge
Agilent E4418B/E4419B Service Guide1-13
Specifications
General Characteristics
General Characteristics
Rear Panel Connectors
Recorder Output(s)
Analog 0-1 Volt, 1 kΩ output impedance, BNC connector
GP-IB
Allows communication with an external GP-IB controller.
RS-232/422
Allows communication with an external RS-232 or RS422 controller. Male
Plug 9 position D-subminiature connector.
Remote Input/Output
A TTL log ic le ve l i s out put whe n the m ea sure men t e xce eds a p red et er mine d
limit. TTL inputs are provided to initiate zero and calibration cycles.
RJ-45 se ri es shielde d m o dular jack ass em b l y.
TTL output: high = 4.8 V max; low = 0.2 V max
TTL input: high = 3.5 V min, 5 V max; low = 1 V max, -0.3 V min
Ground
Binding post, accepts 4 mm plug or bare-wire connection
Line Power
•Input Voltage Range: 85 to 264 Vac, automatic selection
•Input Frequency Ra nge: 50 to 440 Hz
•Power Requirement: approximately 50 VA (14 Watts)
1-14Agilent E4418B/E4419B Service Guide
Specifications
Environmental Characteristics
Environmental Characteristics
General Conditions
Comp lies with th e requirem ents of th e E M C D i rective 8 9 /3 3 6/EEC. This
includes Generic Immunity Standard EN 50082-1: 1992 and Radiated
Interference Standard EN 55011:1991/CISPR11:1990, Group 1 - Class A.
Operating Environment
Temperature
0ºC to 55ºC
Maximum Humidity
95% at 40ºC (non-condensing)
Minimum Humidity
15% at 40ºC (non-condensing)
Maximum Altitude
3,000 meters (9,840 feet)
Storage Conditions
Storage Temperature
-20ºC to +70ºC
Non-Operating Maximum Humidity
90% at 65ºC (non-condensing)
Non-Operating Maximum Altitude
15,240 meters (50,000 feet)
Agilent E4418B/E4419B Service Guide1-15
Specifications
General
General
Dimensions
The following dimensions exclude front and rear panel protrusions:
212.6 mm W x 88.5 mm H x 348.3 mm D (8.5 in x 3.5 in x 13.7 in)
Weight
Net
Agilent E4418B, 4.0 kg (8.8 lb) - 5.0 kg (11.0 lb) with option 001
Agilent E4419B, 4.1 kg (9.0 lb) - 5.1 kg (11.2 lb) with option 001
Shipping
Agilent E4418B, 7.9 kg (17.4 lb) - 8.9 kg (19.6 lb) with option 001
Agilent E4419B, 8.0 kg (17.6 ib) - 9.0 kg (19.8 lb) with option 001
Lithium Polycarbon Monoflouride, approximate lifetime 5 years at 25ºC.
Agilent E4418B/E4419B Service Guide1-17
Specifications
General
1-18Agilent E4418B/E4419B Service Guide
2
Performance Tests
Per formance Tests
Introduction
Introduction
The procedures in this chapt er test the power met er’s electrical
performance using the specifications in Chapter 1 as the performance
standards. All tests can be performed without access to the interior of the
power meter. A simpler operational test is inc lude d in “ Se lf Test” of the
User’s Guide.
Note For valid performance tests, the follo win g cond itions must be met:
a. The power meter and test e quipm ent must have a 30 minute
warm-up for all specific ations.
b. The line voltage for the power meter must be 85 to 264 Vac and
the line frequency must be 50 to 440 Hz.
c. The ambient temperature must be 00 to 550C.
2-2Agilent E4418B/E4419B Service Guide
Performance Tests
Introduction
Equipment Required
Table 2-1 lists all the equipment required for the adjustments and
performance tests. If substitutions must be made, the equipment used
must meet the critic al sp e cific atio ns.
Table 2-1: Required Equipment
Equipment
Required
Model NumberCritical SpecificationUsage
Digital MultimeterAgilent 3458ARange 0 to 20 V
Resolution 0.01 V
DC CalibratorDatron 4000A
or
Fluke 5720A
Range Calibrator
Agilent 11683A
or
Agilent 11683A
Opt H01
Option H01 is essential for
accurate verification. (See
analysis in
Standard 11683A can be used
for indicati on on l y.
Appendix A). The
Frequency CounterAgilent 53132A Range 10 Hz to 50 MHz
Resolution 1 Hz
Power MeterAgilent 432ARange 1 mW
Transfer Accuracy 0.2%
Thermistor MountAgilent 478A Opt H75/H76
or
Agilent 8478B Opt H01
SWR<1.05 at 50 MHz
Accuracy ±0.2% at 50 M Hz.
Power Sensor Agilent E-series power
sensor
1
P and A
P
P
P and A
P and A
P and A
P
30 dB AttenuatorAgilent 11708A
or
Agilent 8491A Option 030
Power Sensor
Cable
1. P = Performance Tests, A = Adjustments
Agilent 11730A
or
Agilent
E9288A
Agilent E4418B/E4419B Service Guide2-3
P
P and A
Per formance Tests
Introduction
Performance Test Record
Results of the performance tests may be tabulated in Table 2-7,
“Performance Test Record,” on page 2-23. The Performance Test Record
lists all of the performance test specifications and the acceptable limits for
each specification. If performance test results are recorded during an
incoming inspection of the power meter, they can be used for comparison
during periodic maintenance. The test results may also prove useful in
verifying proper adjustments after repairs are made.
Performance Tests
The performance tests given in this chapter are suitable for incoming
inspection or preventive maintenance. During any performance test, all
shields and connecti ng hardware must be in place.
The tests are designed to verify published power meter specifications.
Perform t he tes ts in the order give n and record the dat a in Ta ble 2-7 on
page 2-23 and/or in the data spaces provided at the end of each procedure.
Calibration Cycle
This power meter requires periodic verification of performance to ensure
that it is operating within specified tolerances. The performance tests
described in t his ch apter s hould be perf ormed once every t wo yea rs. Under
conditions of heavy usa ge or seve re oper ating environments, the tests
should be more frequent. Adjustments that may be required are described
in Chapter 3.
Test Procedures
It is assumed that the per so n p erfo rm in g the following tests un d e rst and s
how to operate the specified test eq uipmen t . Equi pment se tti ngs, other
than those for the power meter, are stated in general terms. It is assumed
that the person will select the proper cables, adapters, and probes
required for test setups illustrated in this chapter.
2-4Agilent E4418B/E4419B Service Guide
Zero Tes t
Specification
Electric al Charac te risticsPerform ance Lim its
Performance Tests
Zero Test
Accuracy: Zero set (Digital settability of zero)±0.0764 µW
1. This performance limit is determined by the zero set specification of the powe r
sensor used in the measur ement plus the measurement noise. The rang e
calibrator ha s a zero s et specification of
noise specification is 110 nW at 16 averages. At 512 av erages a noise multiplier of
0.24 is required, gi vi ng a measurement noise specification of 0.0264 µW
(0.24x512).
±0.05 µW . The calibrator measurement
1
Description
After the power meter is initially calibrated and zeroed, the change in the
digital readout is m onitored. This test als o takes d rift and noise into
accoun t , si nce d rif t , no is e a nd z e ro re a d ings canno t be s eparated.
Equipment
Range Calibrator .................................. Agilent 11683A
Power Sensor Cable .............................Agilent 11730A
Test Setup
Power M e t e r
Range
Calibrator
CHANNEL A
Agilent 11730A
POWER
METER
Figure 2-1: Zero Test Setup
Agilent E4418B/E4419B Service Guide2-5
Per formance Tests
Preset
dBm/W
System
Zero
Zero Test
Procedure
The following procedure should be performed for the Zero Test.
Note The procedure deta i l s the key pres se s required on the
Agilent E4418B. For the Agilent E4419B the equivalent key
presses should be performed on both chan nels.
Note When switching the rang e ca libra to r to S TANDB Y, allow enough
time for the range calibrator to settle to its zero value before
attempting to zero the power meter . This settling would appear on
the power meter display as downward drift. When the drift has
reached minimum , (typ ica lly le ss th an 60 se cond s), the range
calibrator is settled.
6. On the power meter press ,
Filter Filter On Mode Man Length
, , , , and set the filter
Inputs
length to 512 in the pop up window then press .
7. On the power meter press , . Wait approximately 10
Cal
Input Settings
Enter
Zero
More
, ,
seconds for the wait symbol to disappear.
8. Wait 30 seconds then verify that the power mete r’s reading is
within ±0.05 µW. Record the re a d i ng.
1. This performance limit does not include the corresponding sensor power
linearity speci fication.
1
Description
The power meter accuracy is verified for variou s power inpu t s. There are
two different methods described here that can be used to perform this test,
although one of these methods has a relatively high measurement
uncertainty values associated with it.
Test Procedure 1
Note The measurement uncertainty of this test procedure is relatively
high at power levels of 30uW and below, as it uses a stan dard
11683A Range Calibrator. As such, this test procedure should be
used for indication only.
Equipment
Range Calibrator...................................Agilent 11683A
Power S ens or Cable...............................Agilent 11730A
Agilent E4418B/E4419B Service Guide2-7
Per formance Tests
Preset
dBm/W
Instrument Accuracy Test
Test Setup
Pow er Meter
Range
Calibrator
CHANNEL A
Agilent 11730A
Figure 2-2: Instrument Accu ra cy Test Setup
Procedure
The following procedu re should be perf ormed for the Instr ument Accurac y
Test.
Note The procedure deta i l s the key pres se s required on the
Agilent E4418B. For the Agilent E4419B the equivalent key
presses should be performed on both chan nels.
LINE ............................................... ON
Confirm
W
POWER
METER
2-8Agilent E4418B/E4419B Service Guide
Performance Tests
System
Zero
Instrument Accuracy Test
Note When switching the rang e ca libr ato r to S TANDB Y, allow enough
time for the range calibrator to settle to its zero value before
attempting to zero the power meter . This settling would appear on
the power meter display as downward drift. When the drift has
reached minimum , (typ ica lly le ss th an 60 se cond s), the range
calibrator is settled.
6. On the power meter press ,
Filter Filter On Mode Man Length
, , , , and set the filter
Inputs
length to 512 in the pop up win d ow then p ress .
7. On the power meter press , . Wait approximately 10
Cal
Input Settings
Enter
Zero
More
, ,
seconds for the wait symbol to disappear. V erify that the display
reads 0 ±0.05 µW.
8. Set the range calibrator’s FUNCTION switch to CALIBRATE.
9. Set th e ra nge calibrator ’s RANGE sw it ch to 1 mW.
10. Press to calibrate the power meter.
Cal
11. Set the range calibrator’s RANGE switch to the positions shown in
Table 2-3 on page 2-10. For each setting, verify that the power
meter’s reading is within the limits shown.
Agilent E4418B/E4419B Service Guide2-9
Per formance Tests
Instrument Accuracy Test
Table 2-3: Instru m e nt Accuracy Results (For Indication Only)
1. These performance limits are determined by the zero set specification of the power sensor
used in the measurement plus the measurement noise.
Note The nominal outputs for the 3µW, 30 µW, 300 µW, 3 mW, 10 mW,
30 m W and, 100 mW settings are 3.16 µW, 31.6 µW, 316 µW,
3.156 mW, 10.03 mW, 31.78 mW, and 101.3 mW respectively.
Note It is not neces sary to ch ec k instrument accur acy in dBm . T he
power meter uses the same internal circuitry to measure power
and mathematically converts watts to dBm .
2-10Agilent E4418B/E4419B Service Guide
Performance Tests
Instrument Accuracy Test
Test Procedure 2
Note This test procedure makes use of a 11683A Range Calibrator fitted
with option H01. The measurement uncer t a i nty of this test
procedure is significantly lower than the previously described
method (refer to Appendix A f or a des cr i pt i on of the mea s ur ement
uncertainty analysis). As such, this test procedure should be used
to ensure the power meter meets it's published specification.
Equipment
•Range Calibrator: Agilent 11683A Option H01
•Power Sensor Cable: Agilent 11730A
•DC Range Calibrator: Datron 4000A
Test Setup
DC Calibrator
Power Meter
CHANNEL A
Power sensor cable
Range
Calibrator
Figure 2-3: Instrum ent Ac curacy (Average Path) Test Setup
Agilent E4418B/E4419B Service Guide2-11
Per formance Tests
Preset
dBm/W
Meas
n
Instrument Accuracy Test
Procedure
The following procedu re should be perf ormed for the Instr ument Accurac y
Test.
Note The procedure deta i l s the key pres se s required on the
Agilent E4418B. For the Agilent E4419B the equivalent key
presses should be performed on both chan nels.
1. Connect the equipment as shown in Figur e 2-3.
2. Unplug the range calibrator's power cord. Due to a ground loop
design problem with th e 1168 3A option H01 ra nge cali brat or, it
will not operate properly unl ess gr ound lo ops are elimi nat ed. The
instrument will continue to operat e with no pow er appl ie d, and
disconnecting the power cord eliminates the problem.
WARNI NG There is a small risk of mild electrical shock from the
chassis of the 11683A option H01 range calibrator unless
the power meter is connected.
3. Figure 2-3 shows the HI/LO outputs of the DC calibrator
connected to the DC reference input of the range calibrator.
Ensure that the Guard and LO output of the DC calibrator are
floating, and are not connected to the earth ground.
4. Swi t ch the power m e t e r o n.
5. Press then .
Local
6. Press , for a reading in watts.
7. Press on the power meter, and set the value
Setup
Confirm
W
Resolutio
to 4.
8. Set th e range calibrator as follows:
•RANGE: 100 mW
•POLARITY: NORMAL
•FUNCTION: STANDBY
•REF SWITCH: EXT
•LINE: OFF
2-12Agilent E4418B/E4419B Service Guide
Performance Tests
System
Zero
Instrument Accuracy Test
Note When switching the rang e ca libr ato r to S TANDB Y, allow enough
time for the range calibrator to settle to its zero value before
attempting to zero the power meter . This settling would appear on
the power meter display as downward drift. When the drift has
reached minimum , (typ ica lly le ss th an 60 se cond s), the range
calibrator is settled.
9. Set the DC calibrator as follow s:
• SET VALUE: 0 (ZERO)
• FUNCTION: DC
• OUTPUT: OFF
10. On the power meter press ,
Filter Filter On Mode Man Length
, , , , and set the filter
Inputs
length to 512 in the pop up win d ow then p ress .
11. On the power meter press , . Wait approximately 10
Cal
Input Settings
Enter
Zero
More
, ,
seconds for the wait symbol to disappear. V erify that the display
reads 0 ±0.05 µW.
12. Set the DC cali brator output to 144.93mV, and set the output
switch to '+'.
13. Press to calibrate the power meter.
Cal
Agilent E4418B/E4419B Service Guide2-13
Per formance Tests
Instrument Accuracy Test
14. Set the DC calibrator output to the values shown in Table 2-4. For
each setting, set the filte r length of the power mete r to the
corresponding value, and veri f y that the power me ter reading is
within the limits sho wn. Refer to step 10 for details o n how to
setup the filter len gth
.
Table 2-4: Instrument Accuracy Results
DC
Calibrator
Setting
458.31uV3.1623 uW2560.028 uW3.0977uW3.2269uW
1.4493mV10uW2560.027 uW9.904 uW10.096uW
4.5831mV31.6228uW2560.027 uW31.428uW31.819uW
14.493mV100uW640.073 uW99.49uW100.51uW
45.831mV316.2278uW640.104 uW314.72uW317.74uW
144.93mV1mW160.2 uW0.9954mW1.0047mW
459.12mV3.1623mW160.9 uW3.1477mW3.1769mW
1.4588V10mW161.6 uW9.954mW10.046mW
4.6872V31.6228mW164.8 uW31.477mW31.769mW
15.588V100mW1618.5 uW99.54mW100.46mW
Effective
Power
Applied
Power
Meter
Filter
Setting
Meas
Unc
Lower
Limit
Ch A
Result
Ch B
Result
Upper
Limit
Note The test limits shown in th is table d o not tak e into acc o unt the
correspo nding measure m e nt uncertaint y va lu e s.
2-14Agilent E4418B/E4419B Service Guide
Po wer Reference Level Test
Performance Tests
Power Reference Level Test
Electrical
Characteristics
Power reference1 mWInternal 50 MHz oscillator
Power refere nce
accura cy setting
Power refere nce
accura cy after
2 years
Performance
Limits
facto ry set to ±0 .4% tr acea ble to
National Phy sica l Labora tory
(NPL), UK.
±0.1%Guardban ded Test Line Limits.
±0.5%
(23 ±3
o
C)
Worst case power me ter
accuracy, taking into account
the possibilit y of oscil lato r drift
and operation at 55 degree s.
Conditions
Description
This test verifies the power level accuracy of the internal 50 MHz
osci ll ator—also cal led the power reference oscillator. The power reference
oscillator output is factory adjusted to 1 mW ±0.4%. This accuracy
includes a performance limit of 0.1% and a system measurement
uncertainty figure of 0.3% (traceable to the National Physical Laboratory
(NPL), UK). Refer to Appendix B for the measurement uncertainty of the
system.
Note The setting of the oscillator output accuracy is determined from the
system measurement uncertainty plus any chosen test line limit.
For example, if a test line limit of 1 mW ±0.2% was chosen while
maintaining a system measur ement uncer t a int y of ±0.3%, the overall
accuracy of the test setup for the 1 mW power out put be com es
0.3% + 0.2%, giving a setting accuracy of 0.5%.
Agilent E4418B/E4419B Service Guide2-15
Per formance Tests
Power Reference Level Test
If an equivalent measurement system is employ ed f or post-factory test,
the power reference oscillator output can again be set to 1 mW ±0.4%. This
setting is sufficient to ensure that the power reference oscillator meets the
accuracy specification of ±0.5% after 2 years. This specification includes
the ±0.4% settin g and also tak es into acc o un t othe r facto r s that will
contribute to determining the accuracy after 2 years. The complete
breakdown of the ±0.5% specific atio n is as fo llo ws:
• Test line limit: 0.1%.
• System uncertainty: 0.3%.
• 2 year oscillator drift: 0.04% .
• Delta environment: <0.06%.
The same equipment used to measure the accuracy of t he power reference
oscillator can also be used to set it to the post-factory accuracy of ±0.4%
(for further information re f e r t o chap t e r 3, Adjust m e nts).
The following procedure ensures maximum accuracy in measuring power
reference oscillator output using Agilent test instruments of known
capability. Signal acqu isition criteria may vary if alternative test
instr umen t s ar e us ed. In s u ch ca se s re f er to the man uf act ur er’s guide li ne s
for furthe r i nf o r m a ti o n.
The test is performed using Equatio n 2-4:
Equation 2-4: Power Level Provided By The Reference Oscillator
The components of this equation are defined as foll ows:
is the voltage measured between V
•V
0
comp
and V
with no powe r
RF
applied and after the Agilent 432A has been zeroed.
•V
is the voltage measured between V
1
comp
and V
with power applied.
RF
This power is approx im ately 1 m W.
•V
is the voltage between th e Agilent 432A V
comp
connector and
comp
chassis ground with power applied.
• R is the resistance of the mount resistor in the Agilent 432A power
meter. For fu rther information re f e r to A p pe ndix A.
• Calibration Factor is the value of the thermistor mount at 50 MHz
2-16Agilent E4418B/E4419B Service Guide
Power Reference Level Test
Equipment
• Test power mete r: Agilent 432A .
• Thermistor moun t: Agil ent 478A opt ion H75 or H76.
• Digital voltmeter (DVM): Agilent 3458A.
Test Setup
Performance Tests
Pow er Meter
Te st Powe r Mete r
Power
ref
Digital Voltmeter
V
RF
Vcomp
Thermistor
mount
Figure 2-4: Power Reference Level Test Setup
Procedure
Use the following procedure to perfo rm the power reference level test:
Note The procedure detail s the key presses required on the
Agilent E4418B. For the Agilent E4419B the equivale nt key presse s
should be performed on both channels.
1. Set the DVM to me asu re resistance:
Connect the DVM betw e en the V
Agilent 432A and pin 1 on the thermistor mou nt end of the sen sor cab le .
V erify that no power is applied to the Agilent 432A.
2. Round off the DVM reading to two decimal places and note this valu e
as the internal bridge resistance (R) of the Agilent 432A
(approximately 200 Ω).
•R ____________Ω
3. Connect the equipment as shown in Figure 2-4, ensuring that the DVM
input leads are isolated from chassis ground.
connector on th e rear pa ne l of the
RF
+ input
-input
Note Ensure that the Agile nt 432A and the power meter under test ha ve
been powered on for at least 30 minutes be fo re mov ing to step 4.
Agilent E4418B/E4419B Service Guide2-17
Per formance Tests
Preset
n
Power Reference Level Test
4. Preset the power meter:
Press , then on the power m e ter.
Local
Confirm
Set the Agilent4 32A RANGE switch to Coarse Zero. Adjust the front
panel Coarse Zero control to obtain a zero meter indication.
5. Zero the Agilent 432A test power me te r:
•Fine zero the Agilen t 432A on the most sensit ive ran ge.
•Set the RANGE switch to 1 mW.
6. Set the DVM to measure m icr ovo lts.
7. Connect the p ositive and negative input leads to the V
comp
and VRF
connectors respectively, on the rear panel of the test power meter.
8. Observe the reading on the DVM. If less than 400 µV, proceed to the
next step. If 400 µV or greater, press and hold the test power meter
Fine Zero switch and adjust the Coarse Zero control so that the DVM
indicates 200 µV or less. Release the Fine Zero switch and proceed to
the next step.
9. Round the DVM reading to two decimal places.
Record t h i s re a d i n g a s V
___________
•V
0
.
0
10.Turn the po wer refe re nce on:
Press ,.
Zero
Cal
Power Ref O
11.Round the DVM reading to two decimal places.
Record t h i s re a d i n g a s V
___________
•V
1
12.Disconnect the DVM negative input lead from the V
.
1
connector on the
RF
Agilent 432A. Reconnect it to the Agilent 432A chassis ground.
13.Obs erve the DVM read ing. Record the readin g as V
•V
__________
comp
comp
.
14.C alc ulat e th e Power Referen ce Osc illato r power usin g eq uatio n 2-4.
15.Verify that P
is within the limits shown in Table 2-5 and record the
meas
reading in the table.
Table 2-5: Power Reference Level Test Result
MinMaxActual Result
0.999 mW1.001 mW________________
2-18Agilent E4418B/E4419B Service Guide
Performance Tests
Agilent E-Series Power Sensor Interface Test
Agilent E-Series Power Sensor Interface Test
The Agilent E-series power sensors have their sensor calib ration tables
stored in EEPROM which enables the frequency and calibrati on factor
data to be downloa ded automatically by the p o wer m e ter. The frequency
and calibration factor data have checksums which are compared with the
data downloaded by the powe r mete r. For the Agilent E4419B dual
channel p o w e r m e ter, this test shoul d b e perf o rmed on both chann els.
Procedure
1. When there is no power sensor connected to the Channel A input
the message “No Sensor” (“No Sensor ChA” on the
Agilent E4419B) appears on the measurem ent window.
2. Connect the A gilent E-series power sensor to the power meter’s
Channel A input. The message “Reading Sensor ChA” appears.
When all the data is downloaded this message disappears and a
reading is displayed on the measurement window. If the download
fails, a war ning message and error flag are displayed on the front
panel. Also, SCPI error message “-310, System Error” is
generated.
3. For the Agilent E4419B repeat these steps for Channel B.
Agilent E4418B/E4419B Service Guide2-19
Per formance Tests
Agilent E-Series Power Sensor Functional Test
Agilent E-Series Power Sensor Functional Test
Description
This test verifie s that the mete r /sen so r com bin atio n can make RF
measurements.
The Agilent E-series power sensors operate over a 90 dB dynamic range
(-70 to +20 dBm). However, since there is an amplifier in these power
sensors, the voltages presented to the power sensor connector are always
within the range of voltages available from the Agilent 11683A range
calibrator.
Equipment
Power Sensor Cable...............................Agilent 11730A
T o complet e th e Functi onal Test the following procedu re should be
performed.
Note The procedure deta i l s the key pres se s required on the
Agilent E4418B. For the Agilent E4419B the equivalent key
presses should be performed on both chan nels.
1. Connect the equipment as shown in Figur e 2-5.
Powe r Meter
Figure 2-5: Sensor Functional Test Setup
2-20Agilent E4418B/E4419B Service Guide
Agilent E-Series
Power Sensor
POWER
REF
CHANNEL A
Agilent 11730A
Performance Tests
Preset
Zero
Zero
Zero
System
Agilent E-Series Power Sensor Functional Test
2. Press then .
Local
3. On the power meter press , . Wait approxima t el y
Confirm
Cal
Zero
10 seconds for the wait symbol to disappear. Verify that the
display reads 0 ±50 pW.
4. Press to calibr ate the p ow er m ete r.
Cal
5. Turn the pow e r referenc e on by pressi ng
■, on the Agilent E4418B.
■, , , on the
Power Ref Off On
Cal
More
Cal
Cal
Power Ref Off On
Agilent E4419B.
6. Connect the 30 dB attenuator as shown in Figure 2-6.
Power M e t e r
Agilent E-Series
Pow er Sens or
Agilent 11708A
30 dB ATTENUATOR
CHANNEL A
HP 11730A
Figure 2-6: Sensor Function al Test Setup with Attenuator
7. Press , , and select to “LOWER”.
Inputs
Input SettingsRange
This s witch es to th e power sensor’s l ow rang e.
8. Verify that the display reads -30 dBm ±1 dB. Rec o rd the read i ng.
9. Press and set to “UPPER”. This switches to the power
Range
sensor’s upper range.
10. Verify that the display reads -30 dBm ±1 dB. Record t he re a d i ng.
11. Press and set to “LOWER”.
Range
12. Disconn e ct th e 30 dB attenuato r and rec onnect as shown in
Figure 2-5.
13. Verify that an overload error is displayed on the status line at the
top of the power meter’s display.
14. Press and set to “UPPER”.
Range
15. Verify that the display reads 0 dBm ±1 dB. Record the reading.
Agilent E4418B/E4419B Service Guide2-21
Per formance Tests
Agilent E-Series Power Sensor Functional Test
Table 2-6: Functional Test Result
Test
ChannelA
Overload Error
Low Range, 30 dB pad
Upper Range, 30 dB pad
Low Rang e, no pad
High Range, no pad
Channel B
Overload Error
(Agilent E4419B
Only)
2-22Agilent E4418B/E4419B Service Guide
Performance Test Record
Performance Test Record
Model Agilent E4418B/E4419B Power Meter
Tested by ___________________________________________
Serial Number ________ _______________________________
Date _______________
Table 2-7: Performance Test Record
Performance Tests
Test
Zero Test−76.40 nW+76.40 nW
Instrument Accuracy
3.1623 µW
10 µW
31.6228 µW
100 µW
316.227 µW
1 mW
3.1623 mW
10 mW
31.6228 mW
100 m W
Power Reference
P
meas
Interface TestPass/Fail______________________Pass/Fail
Functional Test
Low Range, 30 dBm pad
Low Range, 30 dBm pad
Low Range, no pad
Upper Range, no pad
2. For instruments with the 50MHz 1mW output (power ref) connector mounted on the
front panel .
For instruments with the 50MHz 1mW output (power ref) connector mounted on the rear
3.
panel. This applies to models E4416A and E4417A with option 003 fitted.
1
Min
Result
N/A
Channel A
Actual
Result
______________________
Channel B
Actual
Result
Result
1.06
1.08
Max
1, 2
1, 3
2-24Agilent E4418B/E4419B Service Guide
3
Adjustments
Adjustments
Introduction
Introduction
This chapter descr ibe s adjustments and che cks which ensure prope r
performance of the power meter. Adjustments are not normally required
on any fixed periodic basis , a nd normal l y are perf ormed only af ter a
performance test has indicated that some parameters are out of
specification . Performanc e tests should be complete d after any rep air s
that may have altered the characte r istic s of th e po wer m e ter. The test
results will make it pos sible to determine if adjustments are required.
Allow 30 minutes for the power meter to warm up, and then remove the
cover, for access to the test and adjustment points.
T o determine which performan ce tests and adjustments to perform after a
repair, see “Post -Repa i r Adjustments”, on page 3-3.
Safety Considerations
This warni ng must b e fol l owe d for your protection and to avoid damage to
the equipment being used.
WARNI NG Adjustments described in this chapter are performed with
power applied to the instrument and with protective
cover s rem oved . Mai nt ena nce sh ou ld be perf orme d onl y by
trained personnel who are aware of the hazards involved.
When the maintenance procedure can be performed
without power, the power should be removed.
3-2Agilent E4418B/E4419B Service Guide
Adjustments
Introduction
Equipment Required
The adjustment procedu res incl ude a list of recommen ded test equ ipment.
The test equipment is also identified on the test setup diagrams.
Post-Repair Adjustments
Table 3-1 lists the adjustments related to repairs or replacement of any of
the assemblie s.
Table 3-1: Post Repair Adjustments, Tests, and Checks
Assembly Replaced
A1 Power SupplySelf Test .
A2 Processor AssemblyFrequency and Level Adjustments,
A3 Front Panel AssemblyInstrument Accuracy Test, Display
A4 Comms AssemblyAll automated T est s and Serial Interface
A5 Daughter AssemblyInstrument Accuracy Test.
A6 Mea sur e m e nt Assem b lyZero Test, Instru m e nt A cc uracy Test.
W1 or W2 Power Sensor
Power Reference Level Test, Display
Brightness and C ontrast Adjustme nt.
Brightness and C ontrast Adjustme nt.
Self Test.
Instr ument Accuracy Test, Ze ro Test.
Agilent E4418B/E4419B Service Guide3-3
Adjustments
Power Reference Oscillator Frequency Adjustment
Power Reference Oscillator Frequency Adjustment
Note Adjustment of the power reference oscillator frequency may also
affect the output lev el of the os cilla to r. Therefore, after the
frequency is adjusted to 50.0 ±0.5 MHz, the out put level should be
checked as describe d in “Power Refer e nc e Oscillator Level
Adjustment”, on page 3-6.
The power reference oscillator frequency is a nominal
specification .
Description
Variable inductor A2L9 is adjusted to set the power reference oscillator
output frequency to 50.0 ±0.5 MHz. This frequ en cy is a sup pleme n tal
characteristic.
Powe r Meter
POWER
REF
Frequency Counter
Figure 3-1: Adjustment Setup
Equipment
Frequency Counter. ............................... Agilent 53132A
Procedure
1. Ensure that the power m eter has bee n powe red on for at least
30 m inu te s befor e making any measurement s.
2. Connect the equipment as shown in Figure 3-1. Set up the counter
to measure frequency.
3-4Agilent E4418B/E4419B Service Guide
Power Reference Oscillat or Frequency Adjustment
Zero
Zero
3. Turn the pow e r referenc e on by pressi ng
■, on the Agilent E4418B.
■, , , on the
Power Ref Off On
Cal
More
Cal
Cal
Power Ref Off On
Agilent E4419B.
4. Observe the reading on the frequency counter. If it is
50.0 ±0.5 MHz, no adjustment of the power reference oscillator
frequency is ne cessary. If it is not within these limits, adjust th e
power reference oscillator frequency as desc ribed in step 5 and
step 6.
5. Remove the power meter cov e r.
6. Adjust A2L9 to obtain a 50.0 ±0.5 MHz indication on the
frequency counter. Refer to Figure 3-2 for the position of A2L9.
A2R90
BT1
Adjustments
A2 Assembly
A2J13
J12
A2L9
Front Panel
Figure 3-2: A2L9 Adjustment Location
Agilent E4418B/E4419B Service Guide3-5
Adjustments
Power Reference Oscillator Level Adjustment
Power Reference Oscillator Level Adjustment
Description
This test adjusts the power level accuracy of the internal 50 MHz
oscillator —also called t he power ref eren c e osc illa t or.
The power refere nc e osc illator ou tp ut is fact or y adju ste d to 1 mW ±0.4% .
This accuracy includes a perfor ma nce limit of 0.1% and a sys te m
measurement un cer taint y figure of 0.3% (traceable to the Nat ional
Physical Laboratory (NPL), UK).
Note The setting of the oscillator output accuracy is determined from the
system measurement uncertainty plus any chosen test line limit. For
example, if a test line limit of 1 mW ±0.2 % was chosen while
maintaining a system measur ement uncer t a i nty of ±0.3 %, the overal l
accuracy of the test setup for the 1 mW power output be com e s 0.3 % +
0.2%, giving a setting accuracy of 0.5%.
If an equivalent measurement system is employ ed f or post-factory test,
the power re ference os cillat or ou tput can a gain be set to 1 mW ±0.4%. T his
setting is sufficient to ensure that the power reference oscillator meets the
accuracy specific ation o f ±0 .5% afte r 2 years. This specific atio n includes
the ±0.4% setting and also takes into account other factors that will
contribute to determining the accuracy after 2 years. The complete
breakdown of the ±0.5% specific a t io n is as follows:
• Test line limit: 0.1%.
• System uncertainty: 0.3%.
• 2 year oscillator drift: 0.04% .
• Delta environment: <0.06%.
The following adjustment procedure ensure the power reference oscillator
meets th e p ost-factory ac curacy setting of ±0.4%.
Note To ensure maximum accuracy in re-adjusting the power reference
oscillator output, this procedure provides s t ep-by-step instructions for
using Agilent test instruments of known capability. Signal acquisition
criteria may vary if alte rn ativ e test instruments are used . In su ch
cases re f er to the manuf a cturer’s guidelines f o r f urther information.
3-6Agilent E4418B/E4419B Service Guide
Adjustments
T
V
–
2V0V1V0–
2V
–
«
V0V
p
«
V
4Volts~
V0400µV<
Power Reference Oscillator Level Adjustment
Note The power meter may be returned to the nearest Agilent Technologies
office to have the power reference oscillator checked and/or adjusted.
Agilent 432A
Power Me t e r
Power
ref
Thermistor
Mount
Powe r Meter
RF
V
Vcomp
Digital
Voltmeter
+
INPUT
INPU
Figure 3-3: Adjustment Setup
For this adjustment the following mathematical assumptions are made:
By substitutin g equation 3 into equation 1 and m anipulating the result
you get:
V1V0–()
2
2– V
compV1V0
–()=410
()R CalibrationFactor()+
This qu ad ra t ic can be sol ve d t o give equati on 2 .
The definitions of the terms in equation 2 are:
is the voltage measured between V
•V
0
comp
and V
RF
applied and after the Agilent 432A has been zeroed.
•V
is the voltage measured between V
1
comp
and V
RF
This power is approx im ately 1 m W.
•V
is the voltage between th e Agilent 432A V
comp
comp
chassis ground with power applied.
• R is the resistance of the mount resistor in the Agilent 432A power
meter.
• Calibration Factor is the value of the thermistor mount at 50 MHz.
Equipment
• Test power mete r: Agilent 432A .
• Thermistor Mount: Agilent 478A Opt ion H75 or H 76.
• Digital Voltmeter (DVM): Agilent 3458A.
Procedure
Note The procedure detail s the key presses required on the
Agilent E4418B. For the Agilent E4419B the equivale nt key presse s
should be performed on both channels.
with no powe r
with power applied.
connector and
1. Set up the DVM to measure resistance. Connect the DVM between the
V
connector on the rear panel of the Agilent 432A and pin 1 on the
RF
thermistor mount end of the sensor cable. Verify that no power is
applied to the Agilent 478A
2. Round off the DVM reading to two decimal places and record this value
as the internal brid ge re sis tanc e (R) o f the test powe r mete r
(approximately 200 Ω).
•R (Internal Bridge Resistance)_______________Ω
3-8Agilent E4418B/E4419B Service Guide
Adjustments
Zero
n
Power Reference Oscillator Level Adjustment
3. Connect the equipmen t as shown in Figure 3-3. The leads should be
isolated from ground. Ensure that the power reference oscillator is off.
Ensure that both the pow e r me te r un der tes t and the A gilent 43 2A
have been powered on for at least 30 minutes before proceeding to the
next step.
4. Set the Agilent 432A range switc h to coa rse ze ro and adjust the front
panel coarse zero control to obtain a zero meter indication.
5. Fine zero the Agilent 432A on the most sensitive ran ge, then set the
range sw itch to 1 mW.
Note Ensure that the DVM input leads are isolated from chassis ground
when performing the next step.
6. Set up the DVM to measu re mi cr ovo lts.
7. Connect the p ositive and negative input leads, respectively, to the
V
and VRF connectors on the rear panel of the Agilent 432A.
comp
8. Observe the readi ng on th e D VM. If l es s th an 400 µV, proceed with the
next step. If 400 µV or greater, press and hold the Agilent 432A fine
zero switch and adju st the coar se zer o contr ol so tha t the DVM
indicates 200 µV or less. Then release the fine zero switch and proceed
to the next step.
9. Round off the DVM reading to the nearest microvolt and record
this value as V
________________µV
•V
0
•Disconnect the DVM negative input lead from the V
.
0
connector on
RF
the test p o we r meter and reconnect it to chassis ground .
10. Turn the power refe rence on by pressin g
■, .
Power Ref O
Cal
11.Record t he readin g on the DVM to two decim a l places. Th is is V
•V
12.R eco nn e ct th e neg ative le ad to th e V
the Agilent 432A. The DVM is now set up to measure V
_____________V
comp
connector on the rear panel of
RF
which
1
represents th e p o we r reference oscillator ou tp ut level .
13.Calc ulat e the required valu e of V
using equa tio n 2.
1
comp
.
V
1
V
compV0
V
comp
2
–4R()CalibrationFactor()–+=
10()
3–
Agilent E4418B/E4419B Service Guide3-9
Adjustments
Power Reference Oscillator Level Adjustment
14.Re mo ve the power meter ’s cover and adjust A2R90 un til the DVM
indicates th e calculated value of V
. Refer to Figure 3-4 for the position
1
of A2R90:
A2R90
BT1
A2 Assembly
A2J13
J12
A2L9
Front Panel
Figure 3-4: A2R90 Adjustment Location
Measurement Uncertainty
Refer to Appendix B “Measurement Uncertainty Analysis – Powe r Reference Level
Test”
3-10Agilent E4418B/E4419B Service Guide
Adjustments
Display Brightness and Contrast Adjustment
Display Brightness and Contrast Adjustment
Introduction
The following procedure should be performed whenever a front panel
assembly or processor assemb l y are rep lac ed.
The brightness is controlled automatically after executing the
Set BrightnessService
Note The contrast adjustment is subjective and varies according to
individual user requirements.
Procedure
softkey, located under the softkey menu.
1. Access the softkey menu and use the softkey to
increase the contrast, or the softkey to decrease the contrast
of the display.
2. When the de sir ed di splay co n tra st le ve l is obt aine d , press the
Set ContrastService
menu.
Display
softkey, located under the softkey
Agilent E4418B/E4419B Service Guide3-11
Adjustments
Display Brightness and Contrast Adjustment
3-12Agilent E4418B/E4419B Service Guide
4
Theory of Operation
Theory of Operation
Introduction
Introduction
This chapter describ es how each of the power mete r’s assemblies operate.
A block diagram is inc luded at the en d of the chapt er giving y ou an overall
view of the power meter’s operat io n.
4-2Agilent E4418B/E4419B Service Guide
Theory of Operation
A1 Power Supply/Battery Charger
A1 Power Supply/Battery Charger
The A1 power supply/battery charger is a 20 W, 47 to 440 Hz switching
power supply produci ng three dc voltages, (+5 V, +12 V, -12 V) used to
power the subasse mb l ie s, and a constant current supply to re cha rge the
optional +12 V rechargeable Battery M odule (Agilent Part Number
E9287A) when operatin g from an ac powe r source. The power supp ly can
be driven either by AC power voltages in the range 85 to 264 V or by the
A8 Rechargeable Battery Module.
The signal line Power_on, from the A2 proces sor assembl y to the A1 pow er
supply assem bly, indicates the status of the front pa ne l PO W E R switc h.
This signal line enables and disables the dc voltage outputs of the power
supply assembly to the A2 processor assembly when driven by the A8
battery assembly. The chargin g cu rrent for the re chargeable battery
module is independent of the state of the Pow er_on line.
A logic low on the Power_on line indicates that the front panel POWER
switch is in s tandby mo de and the thr ee dc vol tage out puts from t he power
supply are disable d. A lo gic high on this line indicates that the POWER
switch is in the On mode and the three dc voltages from the power suppl y
are enabled. Power is distributed to the m eter subassemblies via the
processor assembly (See “A2 Processor Assembly”, on page 4-4).
If fitted, the A8 rechargeable battery module controls the charging current
while ac power is connected to the meter.The fan assembly is active under
the following cond itions:
1. ac power connec te d an d the POWE R switch set to On.
or
2. ac power connecte d, POW ER switc h set to Standby and optional
+12 V rechargeable battery module fitted .
The ac line fuse is located in th e lin e in put m o dule on the power meter’s
rear panel (Agilent Part Number 2110-0957). The fuse holder contains a
spare fuse as standard on ship m ent.
Agilent E4418B/E4419B Service Guide4-3
Theory of Operation
A2 Processor Assembly
A2 Processor Assembly
The processor assembl y cont ai ns the microc ontroller and associated
circuits, the po w e r-on/ standby contro l a nd switching, the 1 mW reference
calibrato r, the recorder outputs, TTL inpu t /output, and the front panel
drivers. It provides tha t platform on which the power meter can run,
facilitating the system inputs and outputs.
Regulated DC voltages at +12 V, -12 V and +5 V are converted by the A1
power supply asse mb ly whe n AC pow e r is connected to the rear pane l o r
the optional rechargeable battery is fitted. The DC voltages are connected
and distributed to the rest of the system by circuits on the processor
assembly. When the key on the front panel is pressed a bis table latch
changes state. This, in turn, app lie s the corr e ct ga te voltage s to tur n on
three MOSFET switches, which connects the power supply unit voltages
to the distributed p ow er buses. The bistable latch is connecte d to a
permanent power-supply which has the b acku p of battery power when AC
mains is removed from the power meter. The +12 V supply is used to
power the fan. This is filtered to decouple it from the rest of the system.
When the optional recharge able battery is fitted and the meter is in
standby and connect ed to the ac powe r su ppl y, the recharg eable batte ry
will recharge and the fa n will be pow ere d on.
The reference osc illato r has a 50 MHz oscillato r circuit with automatic
level control (ALC). The oscillator output is level detected and that level is
compared to a temperatur e stable precis io n referenc e vol tage. This
comparison produces an error signal that gives negative feedback control
of the oscillator output power. The frequency and power level of the
calibrator are factory set to provide a 50 MHz 1 mW transfer standard.
The precision reference v ol tage and the ALC control sign a l are both
measured in the calibrator self test. The calibr ato r is sw itch ed o n or off
using a signal from the microcontroller. The front panel LED indicator is
switched with a separ a te mic roc on tr ol l er signal. The recorder outputs are
driven from a dual 12 bit DAC which is driven by the microcontroller. The
dual DAC outputs are buffered, filte red and scal ed to give a 1 V full scale
outp ut nominal .
The recorder is a 12 bit DAC driven by the microcontroller. The DAC
output is b uffered, filtered and scaled t o g i ve a 1 V full sca l e ou t put with a
nominal 1 kΩ output impedance.
Circui try for the k ey b o ard driver includes so m e damage pr otection, bu t i t
is basically a direct connection from the keypad row and column matrix to
the microcon trolle r ’s control lin es.
4-4Agilent E4418B/E4419B Service Guide
Theory of Operation
A2 Processor Assembly
The LCD contr ol le r on the A 3 fron t p ane l asse m bly is configured as a
memory mapped peripheral, and as such requires only to be fed with the
appropriate address, data and control lines from the microcontroller
circuits . Th e bias volta ge for the LCD is produced by a DC to DC converter
that takes the +5 V (DIST) voltage and converts that to a nominal +21 V.
The DC to DC converter is adjusted by a combinat io n of the contr ast
control signa l fro m t he microcontroller and the tempe rat ur e s en se voltage
that is generated on the A3 front panel assembly. The temperature sense
voltage helps to compensate for the normal variation of LCD contras t with
temperature. Current to the LCD’s LED backlight array is sourced from a
constant current circuit that compe ns ate s f or vari atio n in L E D k nee
voltage. A control input to this circuit from the front p ane l te m pe ratu re
sense voltage allo ws f o r de-r a tin g of the maximum LED current at high
operating temperatures. A control input from the microcontroller allows
the optimum backlight brightness to be factory set. The temperature
sense voltage, LED current and LCD bias voltage are all measured at self
test.
An analog multiplexer provides means by which several diagno stic poin ts
can be switched to the A 6 meas urement assembly ( s) for an alogue to
digital conversion. The microcontroller uses this function to perform self
test.
A6 measurement assembly(s) to microcontroller communication is
conducted on a serial interface bus, the Internal Serial Bus (ISB). The ISB
is connected to the bi-direc tion al ser ial port on the micr ocon t roller via
buffers. The system clock and some control signals are connected to the
measurement modules. The measurement module interface is completed
with the distribution of switched and direct power (+12 V, -12 V and +5 V).
The TTL outputs are conn ected to the A6 processor assembly via in line
resistors locate d on the A4 int e rfac e ass emb ly. A TTL lo gic level is outp u t
from the processor assembly when the user defined windows based limits
are exceeded.
The TTL inputs are connected to the processor card via in line resistors
located on the interf ace car d and can be used to initiat e zero and
calibration cycl es on the power meter.
Agilent E4418B/E4419B Service Guide4-5
Theory of Operation
A2 Processor Assembly
The microcontroller circuits that control all the above functions, and
provide platform for the system software to run on, comprise the
microcontroller itself, memory, and clock and logic circuits. The logic
circuits ha v e t he function of ensur i ng t h e correct sequencing and decoding
of the control signals for the various peripherals. The crystal oscillator
clock circuit is buffered and distributed to the A4 interface assembly and
the A5 daughter asse mbly. The program memory for the microcontro ller is
FLASH EEPR OM to allow for in-circuit prog ram ming. The static RAM is
split into volatile and no n- volatile blocks, with the pow e r for the
non-volatile RA M be ing conne c te d to the p erm an en t (batte r y backe d)
power supply.
Note It is advisable to replace the RAM battery A2BT1 every three
years. If the battery fails or drops below 3 V the RAM will lose the
following information when power is disconnected from the meter:
Unit Serial Number
Revision letter of Processor Card
Unit Option
User defined setups
User defined sensor calib rat i on tables
4-6Agilent E4418B/E4419B Service Guide
Theory of Operation
System
A3 Front Panel Assembly
A3 Front Panel Assembly
The front panel ass emb ly is m ad e up of a liquid crystal display (LCD), a
keypad and, depending on the power meter option, a power reference cable
assembly and a sensor cable assembly(s).
There are two inputs to the front panel assembly:
•the flex circuit from connector A2J4 of the pro cessor assembly
which controls the keypad.
•the ribbon cable from A2J3 of the processor assem bly wh ich
contro ls t he LCD dis p la y.
The power meter self tests cont ain display tests. These are accessed by
More
pressing , , , , ,
Display
Inputs
.
Service Self Test Individual
Agilent E4418B/E4419B Service Guide4-7
Theory of Operation
A4 Comms Assembly
A4 Comms Assembly
The comms assembly contains the circuitry required for remote control of
the power meter. This assembly supports parallel and serial interfaces.
The GP-IB interface is supported by a protocol controller integrated
circuit and two physical interface buffers. The system clock is divided by
four to provide the GP-IB controller integrated circuit clock signal.
The RS232 and RS422 interface is supported by a single Universal
Asynch ro no us Recei ve r/Transceiver ( UAR T) in tegrated circuit and a
programmable transceiver. The programmable transceiver can be
configured for either RS232 or RS422. The remote interface is d esign ate d
a DTE (Data Terminal Equipment).
The programmable baud rate gene rat or in the UART is driven from the
system clock (16.67 MHz) which allows baud ra tes in the ran ge 50 to
115.2 K.
The whole assembly is connected to data, address, and control signals
from the A2 processor assembly.
The comms assembly also provides a filtered path for the single/dual
recorder out put signal and the TTL input and output signals . It als o
provides ESD protection for the RS232.422, TTL I/O, and Recorder
outputs.
4-8Agilent E4418B/E4419B Service Guide
Theory of Operation
A5 Daughter Assembly
A5 Daughter Assembly
The A5 daughter assembly is loaded vertically into the A2 processor
assembly. The Agilent E4418B has five connectors on the A5 daughter
assembly:
•two 6-way co nnect ors route the signal lines between the
rechargeable battery assembly and the A1 power supply/charger
assembly, and provide an inte rface for the signal lin e s that are
monitored by the A2 proc es sor ass embly.
•a 48-way connector provides the interface for the A2 processor
assembly signals.
•a 36-way slotted post connector provides the interface for the A6
measurement assembly.
•a 2-way connector for production test only.
The Agilent E4419B has an additional 36-way slotted post connector.
Agilent E4418B/E4419B Service Guide4-9
Theory of Operation
A6 Measurement Assembly
A6 Measurement Assembly
There is one measu rement assemb ly i n the Agilent E4418B and two in the
Agilent E4419B.
The measurement ass embly amplif ies and c onvert s the chop ped AC signal
produced by the power sensor (either Agilent 8480 series power sensors or
Agilent E-series powe r sens or s) into a 32 bit di gital wor d. T his digital
word is proportional to the input RF power level applied to the power
sensor. The measurement asse mbl y also mea sures the sensor resistor
voltage to detect whether a power sensor is present. If a power sensor is
connected it determines the model, a nd w heth er it’s connected t o the front
or rear panel. This assembly produces the differential drive signal for the
power sensor’ s ch opper. When an Agilent E-series powe r sensor is present
this assembly converts and corrects the power sensor’s temperature
dependent voltage.
With refe rence to t he block diagr am at th e end of this c hapter, the chopped
AC input signal is amplified by the input amplifier . This operational
amplifier in combination with a single bipolar transistor in the power
sensor, produce a feedback amplifier which has a gain of approximately
500. The “Feedback” line shown sets the gain and also biases the
transistor. The differential ampl ifie r rem o ves any common mode no ise o r
interference. It uses a special ground wire (“Sensor Ground”), connected
close to the RF bulkhead in the power sensor as it’s second input.
The band pass filter represent s the filte ri n g acti on of the input ampli f i er.
When the power m ete r is working with an Agilen t E-serie s po we r sensor
the equalizer is switched on to reduce the high pass cut off frequency of
the input amplif ie r. This improves the settling time in f ast m od e
operation.
The output of the equalizer is split into two paths. One path is amplified
by a gain of 100. Both signals are each converted to dig ital wor d s by the
dual analog to digital converter (ADC). The ADC’s sampling rate is set to a
frequency 200 times above the chop rate for the Agilent 8480 series power
sensors and 100 times above the ch op rat e for the Agilen t E-series powe r
sensors.
The Digital Signal Proces sor (DSP) converts the chopped signals into
digital words proportional to the input level. It compares both levels,
calibrates the gain and d ecid es which level is output to the ho st p roc es so r
(after additional filte rin g) via the “Intern al Serial Bus” .
4-10Agilent E4418B/E4419B Service Guide
Theory of Operation
A6 Measurement Assembly
The DSP chip controls the logic which sets the chopper driver voltage and
frequency control. For the:
•Agilent 8480 series power sensors this is 0 V and -10 V at 217 Hz.
•Agilent E-series power sensors this is +7 V and -3 V at 434 Hz.
The DSP also control s logic to allow t he A UX ADC to measure a number of
voltages, and when requ este d , send the relevant data to the host
processor. The AUX ADC digitizes a voltage which is produced by the
Agilent E-series power sensors and is proportional to temperature. This
voltage is required for temperat ur e compen sat i on . This AUX ADC also
tests the sensor resistor wires to check sensor continuity. Lastly, use is
made of this AUX ADC for the pow e r me te r’s self test of both the
measurement assembly and the A2 processor assembly. The results are
reported back to the A2 processor ass e mbly via th e “Inte rn al Serial Bu s”.
The Agilent E-series power sensors have built in serial EEPROM and
range switching which is controlled or acce sse d via the m easurement
assembly. The “N-Chop” sensor wire used by the Agilent 8480 series power
sensors double s as a serial clock and the sensor resis to r line d ouble s as a
bi-directional serial da ta line. The “Auto Zero” line is also used as a serial
bus enable control.
The wire labelled “Auto Zero” is grounded to the “Sensor Ground” line
when an Agilent 8480 series power senso r is used. Auto zeroing is
accomplished without the need for this sign al i n the power meter.
All clocks are derived from t he 16.67 MHz system cl ock w hich i s g enerated
on the A2 processor assembly and distributed on the A5 daughter
assembly.
Agilent E4418B/E4419B Service Guide4-11
Theory of Operation
A8 Rechargeable Battery As sem bly
A8 Rechargeable Battery Assembly
The A8 Rechargeable Battery Assembly all ows the EPM power meter with
option 001 fitted to operate when no AC power in p ut is availab le. The
battery pack contain s a he alth mon ito rin g cir cu it wh ich the A2 p roce s so r
assembly can interr ogate. Th e A8 batte ry assembly connects to the A2
proces s o r a ssembly an d the PSU/battery cha rger circuit via the A 5
Daughter Assembly.
There are th ree c onnec tions fr om the rech argeab le bat tery as sembly to t he
A1 psu/charger assembly, +12V, 0V , and Signal 1. The Signal 1 line is an
indicator line which signals when the battery is below a threshold limit.
When this line go es low, indicating there is insuffic ie nt cha rg e lef t in the
Battery Assembly to continue operating the power meter, it will cause the
A1 psu/charger assembly to shut down.
There is one connection, the communications line, from the A8 Battery
Module to the A2 processor assembly. This line provides serial
communic atio ns with the processor asse m bly o n batter y pa ck status
information, such as amount of charge remaining. When there is less than
10 minutes run time char ge rem ainin g in th e bat tery an error me ssage
"Battery Power Low’" is displayed on the power meter front panel display .
The battery assembly ca n only be recharged when fitted to a power meter
with the battery option fitted. The charge current from the charging
circuit is controlled by a mod u le con tain e d within the Battery assem bly.
When the ac power is connected the battery assembly regulates the
charging current based on the discharge state of the battery.
The battery should fully charge from empty within 2 hours. At 25 degrees
C this will give the me te r 2 hour s ope ra tio nal tim e. The A8 Battery
module has a lifetime of approximately 450 charge/discharge cycles after
which the battery can still be use d, but will o nly have appr oxim ate ly 70 %
of its original capacity. The battery capacity may be reduc e d wh ilst
charging at temperatures abov e 35 degrees C.
Note:- partially d ischa rg ing and recharging the battery module will
reduce its lifetime. It is advis ed that periodically the b attery assembly
should have its charge fully cycled i.e. fully discharged and then fully
recharged.
4-12Agilent E4418B/E4419B Service Guide
A8 Rechargeable Battery Assembly
Remove this page and replace with pullout.fm.
Theory of Operation
Agilent E4418B/E4419B Service Guide4-13
Theory of Operation
A8 Rechargeable Battery As sem bly
4-14Agilent E4418B/E4419B Service Guide
5
Replaceable Parts
Replaceable Parts
Introduction
Introduction
This chapter con tains details of some of the higher level components and assemblies
which can be ordered from Agilent Te chnologies. It also details how to assemble and
disassemble the power meter for repair.
To order parts co ntact your local Agilent Technolo g ies Sales and S er vice Office.
To return your power meter for servicing at a qualified service center refer to
“Contacti ng Agilent Technologies” in chapter 2 of the User’s Guide.
5-2Agilent E4418B/E4419B Service Guide
Assembly Part Numbers
Replaceable Parts
Assembly Part Numb ers
Reference
Designator
A1Power Supply Assembly
Refer to “Po wer Supp ly As sembly Types”,
on page 5-5 to help with the selection.
push-fit line module)
MP6J1Push-F it Line ModuleE4418-61002
Battery Module CoverE4418-61019
W3Recorder Output Cable AssyE4418-61015
Sensor Cable Assembly KitE4418-61036
5-4Agilent E4418B/E4419B Service Guide
Replaceable Parts
Assembly Part Numb ers
Reference
Designator
Front Power Reference Cable KitE4418-61811
Rear Power Reference Cable KitE4418-61813
Description
Agilent Part
Number
Power Supply Assembly Types
Before ordering an A1 Power Supp ly Assembly, check connector J2 and compare it
to Figure 5-1 and Figure 5-2 and select the appropriate mod el. The J2 connector is
found nearest the fan assembly.
Figure 5-1: 0950-26 49 Powe r Supply Assembly
Figure 5-2: 0950-3 681 Power Su pply and Charger Assembly
Agilent E4418B/E4419B Service Guide5-5
Replaceable Parts
Front Panel Connector Repair
Front Panel Connector Repair
The front panel assembl y is an exchange assembly. However, if front panel sensor
cables or power referenc e cables are faulty they should be replaced by ordering the
appropriate kit and following the assembly/di sa ssembly procedure. The instrument
accuracy test should then be carried out to verify the functionality of the new part.
5-6Agilent E4418B/E4419B Service Guide
Replaceable Parts
System
Firmware Upgrades
Firmware Upgrades
The power meter’s firmware can be upgraded using a PC. The current firmware is
available on the World Wide Web. A firmware upgrade package (Part Number
E4418-61035)can be purchased for users who do not have access to the World Wide
Web. The upgrade package contains instructions and a disc pack. The firmware can
be upgraded via the GP-I B connector or the serial RS232/422 connec tor.
To determine the firmware version in your power meter, press:
•, , , .
Inputs
More
Contact your local Service Center if you require a firmware upgrade. Refer to “Sales
and Service Offices”, on page -ix for details of your local Service Center.
Downloading Firmware
Instrument Firmware should be downloaded after the processor board is repl ac ed.
Firmware can be access ed by Agilent Technologies Service Centers using the World
Wide Web.
Service Version
Agilent E4418B/E4419B Service Guide5-7
Replaceable Parts
Assembly and Disassembly Guidelines
Assembly and Disassembly Guidelines
The guidelin es in this section describe the removal and rep lac ement of the major
assemblies in the Agilent E4418B and AgilentE4419B power meters.
Once an assembly has b een re place d, refer t o “ Post-Re pair Adj ustment s”, on page 3-3
to ensure that the correct performance tests and adjustments are carried out.
WARNING Refer to the safety symbols on the power meter and “Equipment
Operation”, on page -v before ope rating t his powe r met er wit h the cover
removed. Failure to comply with the safety precautions can result in
severe or fatal injury.
WARNING Before carry ing ou t any ass em b ly or disassembly of the power meter
ensure that you disconnect the power cord. Even with the power meter
switch ed off there are potentially dangerous voltages present on the
power supply assembly.
Removing the Power Meter Cover
Remove the optional battery if fitted before removing the cover.
When replacing the power meter cover retighten the captive scre ws in th e rear bez el
using a 9 lb/in T15 screw driver.
5-8Agilent E4418B/E4419B Service Guide
Replaceable Parts
Assembly and Disassembly Guidelines
Removing the A1 Power Supply Assembly
1.Remove the power supp ly cover by lifting it out.
2.Disconnect the line input module from the chassis and power supply
assembly.
3.Disconnect the cable assembly from the power supply which connects to the
A2 processor assembly.
4.Unscrew the power supply assembly and lift out the power supply. (When
replacing these screws use a 9 lb/in T15 screw driver.)
WARNING When replacing the power supply unit in the power me ter ensure that
all earth wi r ing is reco nnected . T here are two terminals to check:
The first is the force fit connector to the power supply unit itself. It is
essential that the gap between the terminal and the adjacent large
capacitor is maximized. Fit the connector so that its flat side faces
towards the large capacitor.
The second, a closed loop terminal bonded to the chassis with an M3.5
machine screw (use a 9 lb/in T15 screw driver).
All the p rotective earth wiring can be identified by the insulation co lor
green with a yellow stripe.
Agilent E4418B/E4419B Service Guide5-9
Replaceable Parts
Assembly and Disassembly Guidelines
Removing the A2 Processor Assembly
1.Remove the A5 daughte r and A6 measurement assemblies as described on
page 5-12.
2.Move the A2 plastic support bracket to its forward posit ion using the two
side levers, unc li p the fle xi-cabl e reta ining ba r on the front pane l keypa d and
front panel LCD cable connectors and disconnect the cable.
3.Disconnect the following cables from the A2 processor assembly:
4.Turn the power meter upside down and remove the three screws as shown.
(When replacing these screws use a 6 lb/in T10 screw driver.)
5.Pu sh th e A 2 pro c es s o r assembly to w ar d s the f r on t pan e l to re l ease it from
the 50-pin connector. Lift it upwards to remove.
screws
Note When re-assembling the processor board, ensure the A2 plastic support
bracket is returned to its locked position.
Note After repla cing a processor board, the Power Reference Frequency and
Level must be checked. For further informati on refe r to “Pe rform anc e
Tests”, on page 2-1.
Note After repla cing a processor board, the display bright nes s and co ntrast must
be adjusted. See “Adjustments”, on page 3-1.
Note Firmware should be downloaded to the instrument after the processor board
is replaced. Refe r to “Downloading Firmware”, on page 5-7.
5-10Agilent E4418B/E4419B Service Guide
Replaceable Parts
Assembly and Disassembly Guidelines
Removing the A4 Comms Assembly
1.Remove the A2 process or, A5 daughter and A6 measurement ass emblies as
shown on page 5-10 and page 5-12.
2.Disconnect the line power module from the A1 power supply and the
chassis.
3.Disconnect the earth wire screw. (When replacing this screw use a 9 lb/in
T15 screw driver.)
4.Remove the two screws on the underside of the deck assembly. Remove the
assembly by sliding forward and tilting up from the rear (when replacing
these screws use a 21 lb/in T15 screw driver).
screws
5.Unscrew the GP-IB and RS232/422 connectors from the rear panel. (When
replacing these screws use a 6 lb/in 9/32 in socket.)
6.Disconnect the Reco rder cable for channel A from J23 pins 1,2,3.
Disconnect the Recorder cable for channel B from J23 pins 4,5,6.
7.Lift the A4 Comms assembly from the two standoffs and slide it out.
Agilent E4418B/E4419B Service Guide5-11
Replaceable Parts
Assembly and Disassembly Guidelines
Removing the A5 Daughter or A6 Measurement Assemblies
1.Disconnect the flex circuit from measurement assembly.
Note Care shou ld be taken when disconnecting the fle x circuit from the
measurement assembly.
The flex circuit assembly is released by pushing the conne ctor tab forward
and lifting.
To replace the flex circuit, loop it as shown on page 5-17, and connect the
flex circuit as shown in the figures below.
2.Slide out the measurement assembly from the side of the power meter.
3.The daughter asse mbl y is removed vertically.
A5
A6
1. Connector closed
2. Slide connector clamp
in direction of arrow
4. Position sensor flex
circuit into connector
in direction of arrow
5-12Agilent E4418B/E4419B Service Guide
5.
6. Push clamp closed in
direction of arrow
3. Rotate connector
clamp in direction of
arrow
Replaceable Parts
Assembly and Disassembly Guidelines
Removing the A7 Fan Assembly
1.Remove the A1 power supply ass em bly as shown on page 5-9.
2.Remove the pins whic h attach the fan to the chassis.
3.Remove the fan cable co nnec tor from the A2 processor assembly to release
the fan.
Agilent E4418B/E4419B Service Guide5-13
Replaceable Parts
Assembly and Disassembly Guidelines
Removing the Power Meter RAM Battery (BT1)
1.Remove the A2 process or as sembly as described on page 5-10.
2.Remove the A2J 1 link to disconnect the battery from the rest of the circuitry.
3.Verify the battery protection circuitry by:
nensuring that there are no electrical short circuits across the battery
terminals.
nensure that there are no voltages present which could apply a charging
voltage.
4.Once the prot ection circuit has bee n verified remove the battery. The battery
is siliconed to the assembly. It may be necessary to remove it using a scalpel.
5.Store the b attery individually in an anti-static (dissipative) ba g or sui table
non-conductive packaging.
6.After replacing the battery secure it using a tie wrap. This should be se cured
from the top of the A2 proces sor assembly. The recomme nded tie wrap part
number is 1400-1154.
7.Replace the A2J1 link.
WARNING This power meter uses a lithium battery which may explode if
mishandled. The battery should not be subjected to short circuit of the
battery terminals or to excessive heat.
Do not recharge this battery or dispose of it by burning. Check local
country regulatory requirements on the disposal of lithium batteries.
5-14Agilent E4418B/E4419B Service Guide
Replaceable Parts
Assembly and Disassembly Guidelines
Removing the A3 Front Panel Assembly
1.Disconnect the following cables from the A2 processor assembly:
npower reference semi-rigid (When replacing use the torque s d etailed on
page 5-18.)
nfront panel keypad
nfront panel LCD
Note Care s hould be t aken when dis conne cting the front pan el ke yboard a nd LCD.
Move the A2 plastic support bracket to its forward posit ion using the two
side levers, unc li p the fle xi-cabl e reta ining ba r on the front pane l keypa d and
front panel LCD cable connectors and disconnect the cable.
2.Disconnect the flex circuit from the measurement assembly.
Note Care shou ld be taken when disconnecting the fle x circuit from the
measurement assembly.
The flex circuit assembly is released by pushing the conne ctor tab forward
and lifting.
To replace the flex circuit, loop it as shown on page 5-17, and connect the
flex circuit as shown in the figures on page 5-12.
3.Remove the center screw from the right hand side of the front panel.
4.Remove the front pane l by pres sing in the metal tab on the front panel as
shown and push down on the side of the chassis until the standoffs are
cleared from the holes.
screw
tab
Note After repla cing a front panel assembly, the display brightness an d contrast
must be adjusted. Refer to Chapter 3 “Adjustments”.
Agilent E4418B/E4419B Service Guide5-15
Replaceable Parts
Assembly and Disassembly Guidelines
Replacing the Front Panel Power Sensor Cable Assemblies
1.Remove the front panel from the power meter. (Refe r to “Removing the A3
2.Replace the power sensor cable assembly(s) using the replac ement kit.
Front panel sensor
retaining clip recess
Front Panel Assembl y”, on page 5-15 for details on removing the front
panel.)
5-16Agilent E4418B/E4419B Service Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.