The issue date for the Propaq MD Operator's Guide (REF 9650-0805-01 Rev. A) is March, 2012.
0197
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA USA
01824-4105
ZOLL International Holding B.V.
Newtonweg 18
6662 PV ELST
The Netherlands
If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to det ermine if
additional product information updates are available.
Propaq, Smartcuf, and SureBP are trademarks or registered trademarks of Welch Allyn or its subsidiaries in the
United States and/or other countries.
Nellcor and OxiMax are trademarks or registered trademarks of Covidien in the United States and/or other
countries.
Oridion, Microstream, FilterLine, and CapnoLine are trademarks or registered trademarks of Oridion Systems,
Ltd. in the United States and/or other countries.
Synchronized Cardioversion of Atrial Fibrillation ......................................................A-26
Electromagnetic Compatibility Guidance and Manufacturer’s Declaration.......................A-27
Appendix BAccessories
viiiwww.zoll.com9650-0805-01 Rev. A
Product Description
The ZOLL® Propaq® MD unit is an easy-to-use portable defibrillator that combines
defibrillation and external pacing with the following monitoring capabilities: ECG, Pulse
Oximeter, Non-invasive Blood Pressure, IBP, CO2, Temperature, and Respiration. It has been
designed for all resuscitation situations and its rugged, compact, lightweight design makes it
ideal for transport situations. It is powered by auxiliary power and an easily replaced battery
pack that is quickly recharge d in the device when it is connected to aux iliary power . In addit ion,
the unit’s battery may be recharged and tested using a ZOLL SurePower™ Battery Charger Station.
Note:The Propaq MD has defibrillation and pacing functionality , but some of the monitoring
functions are optional features. See the complete list of options in Fig. 1-1. Optional
features are specified as “optional” within this guid e.
The unit has a large colorful LCD display of numerics and waveform data that provides easy
visibility from across the room and at any angle. ECG, plethysmograph, and respiration
waveform traces can be displayed simultaneously, giving easy access to all patient monitoring
data at once. The display screen is configurable, so you can choose the best visual layout to fit
your monitoring needs. The Propaq MD includes a transcutaneous pacemaker consisting of a
pulse generator and ECG sensing circuitry . Pacing supports both deman d and fixed noninvasive
pacing for adult patients and adolescent, child, and infant pediatric patients.
Chapter 1
General Information
The Propaq MD has a patient data review and collection system that allows you to view, store,
and transfer patient data. The Propaq MD unit contains a printer and USB po rt , whi ch you can
use to print the data and transfer it to a PC.
9650-0805-01 Rev. APropaq MD Operator’s Guide1-1
CHAPTER 1GENERAL INFORMATION
Propaq MD Optional Features
The following features are optional in the Propaq MD unit.
Figure 1-1 Propaq MD Optional Features
Optional Feature
12 Lead ECG
(NellcorTM)
SpO
2
NIBP (with Smartcuf
EtCO2 (Oridion
Temperature
Invasive Pressures (3 Channels)
Printer
®
and SureBPTM)
®
Microstream®)
How to Use This Manual
The Propaq MD Operator's Guide provides information operators need for the safe and
effective use and care of the Propaq MD product. It is important that all persons using this
device read and understand all the information contained within.
Please thoroughly read the safety considerations and warnings section.
Procedures for daily checkout and unit care are located in the Chapter 18, "Maintenance".
Operator ’s Guide Updates
An issue or revision date for this manual is shown on the front cover. If more than three years
have elapsed since this date, contact ZOLL Medical Corporation to determine if additional
product information updates are available.
All users should carefully review each manual update to understand its significance and then
file it in its appropriate section within this manual for subsequent reference.
Product documentation is available through the ZOLL website at www.zoll.com. From the
Products menu, choose Product Manuals.
Unpacking
Carefully inspect each container for damage. If the shipping container or cushion material is
damaged, keep it until the contents have been checked for completeness and the instrument has
been checked for mechanical and electrical integrity. If the contents are incomplete, if there is
mechanical damage, or if the defibrillator does n ot pass its electrical self-t est, U.S.A. customers
should call ZOLL Medical Corporation (1-800-348-9011). Customers outside of the U.S.A.
should contact the nearest ZOLL authorized representative. If the shipping container is
damaged, also notify the carrier.
1-2www.zoll.com 9650-0805-01 Rev. A
Symbols Used on the Equipment
Any or all of the following symbols may be used in this manual or on this equipment:
SymbolDescription
Dangerous voltage.
Attention, consult accompanying documents.
Fragile, handle with care.
Keep dry.
This end up.
Symbols Used on the Equipment
Temperature limitation.
Conformité Européenne Complies with medical device directive 93/42/EEC.
Type B patient connection.
Type BF patient connection.
Type CF patient connection.
Defibrillator-proof type BF patient connection.
Defibrillator-proof type CF patient connection.
Fusible link.
9650-0805-01 Rev. APropaq MD Operator’s Guide1-3
CHAPTER 1GENERAL INFORMATION
2%452.
,I)/.
RECYCLE
,I)/.
SymbolDescription
Equipotentiality.
Alternating current (ac).
Direct current (dc).
Auxiliary power adapter operation.
Caution, high voltage.
Earth (ground).
Negative input terminal.
Positive input terminal.
Power On/Off.
Protective earth (ground).
Contains lithium. Recycle or dispose of properly.
Keep away from open flame and high heat.
Do not open, disassemble, or intentionally damage.
Do not crush.
1-4www.zoll.com 9650-0805-01 Rev. A
SymbolDescription
Do not discard in trash. Recycle or dispose of properly.
Return to a collection site intended for waste electrical and electronic
equipment (WEEE). Do not dispose of in unsorted trash.
Date of manufacture.
Use by.
Latex-free.
Do not reuse.
Symbols Used on the Equipment
Do not fold.
Not sterile.
Manufacturer.
Authorized representative in the European Community.
Serial Number.
Catalogue number.
Consult instructions for use.
Prescription only.
9650-0805-01 Rev. APropaq MD Operator’s Guide1-5
CHAPTER 1GENERAL INFORMATION
MR
SymbolDescription
Battery charging status.
Do not use device, cables, or probes in an MRI environment.
Conventions
This guide uses the following conventions:
Within text, the names and labels for physical buttons and softkeys appear in boldface type (for
example, “Press the Charge button or press the Pacer button”).
This guide uses uppercase italics for audible prompts and for text messages displayed on the
screen (for example, LEAD FAULT).
War ning !Warning statements alert yo u to cond itions o r actio ns that can result in personal injury
or death.
CautionCaution statements alert you to conditions or actions that can result in damage to the unit.
Propaq MD Indications for Use
The Propaq MD is intended for use by trained medical personnel who are familiar with basic
monitoring, vital sign assessment, emergency cardiac care and the use of the Propaq MD. The
Propaq MD is also intended for use by (or on the order of) physicians at the scene of an
emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other
similar areas of a hospital. The usage may be in an ambulance or at the scene of an emergency.
It is also intended to be used during the transport of patients. The Propaq MD will be used
primarily on patients experiencing symptoms of cardiac arrest or in post trauma situation. It
may also be used whenever it is required to monitor any of th ose functions that are included (as
options) in the device. The Propaq MD can be used on pediatric patients (as described in the
following table) and on adult patients (21 years of age or older) with and without heart
dysfunction.
Pediatric Patient SubpopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age.
Infant1 month to 2 years of age.
Child2 to 12 years of age.
Adolescent12 to 21 years of age.
1-6www.zoll.com 9650-0805-01 Rev. A
When the pediatric patient is less than 8 years of age or weighs less than 55 lbs. (25 kg.), use
ZOLL pedi-padz
patient’s exact age or weight.
Manual Defibrillation
Use of the Propaq MD in the manual mode for external and internal defibrillation is indicated
on victims of cardiac arrest wher e there is apparent lack of circulation as indicated by:
• Unconsciousness.
• Absence of breathing.
• Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular
fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable
of producing hemodynamically significant heart beats.
The unit can also be used for synchronized cardioversion of certain atrial or ventricular
arrhythmias. Qualified medical personnel must decide when synchronized cardioversion is
appropriate.
The patient population will range from newborn (neonate) to adul t.
Propaq MD Indications for Use
®
pediatric defibrillation electrodes. Do not delay therapy to deter m ine the
ECG Monitoring
The Propaq MD is intended for use to monitor and/or record 3-, 5-, or 12-lead ECG waveform
and heart rate, and to alarm when heart rate is above or below limits set by the operator. The
patient population will range from newborn (neonate) to adult, with and without heart
dysfunction.
External Transcutaneous Pacing
This product can be used for temporary external cardiac pacing in conscious or unconscious
patients as an alternative to endocardial stimulation.
The purposes of pacing include:
• Resuscitation from standstill or bradycardia of any etiology:
• As a standby when standstill or bradycardia might be expected:
• Suppression of tachycardia.
• Pediatric pacing.
Non-Invasive Blood Pressure Monitoring
The Propaq MD is intended for use to make no n-invasive measurement s of arterial pressure and
heart rate, and to alarm if either parameter is outside of the limits set by the user.
Measurements are made using an inflatable cuff on the patient's arm or leg. The patient
population will range from newborn (neonate) to adult.
9650-0805-01 Rev. APropaq MD Operator’s Guide1-7
CHAPTER 1GENERAL INFORMATION
Temperature Monitoring
The Propaq MD is intended for use to make continuous temperature measurements of rectal,
esophageal, or surface temperatures, and to alarm if the temperature is outside of the limits set
by the user. The patient population will range from newborn (neonate) to adult.
SpO2 Monitoring
The Propaq MD is intended for use to monitor pulse rate and oxygen saturation of arteriolar
hemoglobin, and to alarm if either parameter is outside of the limits set by the user.
Measurements are made non-invasively at remote sites such as a finger, toe, ear lobe, bridge of
nose, etc. The patient population will range from newborn (neonate) to adult.
Respiration Monitoring
The Propaq MD is intended for use to continuously monitor respiration rate an d to ala r m if the
rate falls outside of the range set by the operator. Because the measurement method actually
measures respiratory effort, apnea episodes with continued respiratory effort (such as
obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The
patient population will range from newborn (neonate) to adult.
CO2 Monitoring
The Propaq MD is intended for use to make continuous noninvasive measurement and
monitoring of carbon dioxide concentration of the expired and inspired breath and breat h rate.
The patient population will range from newborn (neonate) to adult.
Invasive Pressure Monitoring
The Propaq MD is intended for use to display and make continuous invasive pressure
measurements from any compatible pressure transducer. The primary intended uses are arterial
blood pressure, central venous pressure and intracranial pressure monitoring. Any contraindications of the particular transducer selected by the user shall apply. The patient population
will range from newborn (neonate) to adult.
1-8www.zoll.com 9650-0805-01 Rev. A
Propaq MD Product Functions
Defibrillator Function
The Propaq MD contains a direct current (dc) defibrillator capable of delivering up to 200
joules. It may be used in synchronized mode to perform synchronized cardioversion using the
patient’s R-wave as a timing reference. The unit uses paddles or disposable, pregelled
electrodes for defibrillation.
Defibrillator Output Energy
Propaq MD defibrillators can deliver biphasic energy from 1 joule to 200 joules. The energy
delivered through the chest wall, however, is determined by the patient’s transthoracic
impedance. An adequate amount of e lectrolyte gel mu st be applied t o the paddles and a force of
10 to 12 kilograms (22 to 26.4 pounds) must be applied to each paddle in order to minimize this
impedance. If hands-free therapy electrodes are used, make sure that they are properly applied.
(Refer to the instructions on the electrode package).
Propaq MD Product Functions
External Pacemaker
Propaq MD defibrillators include a transcutaneous pacemaker consisting of a pulse generator
and ECG-sensing circuitry. Noninvasive transcutaneous pacing (NTP) is an established and
proven technique. This therapy is easily and rapidly applied in both emergency and
nonemergency situations when temporary cardiac stimulation is indicated.
The output current of the pacemaker is continuously variable from 10 to 140 mA (the output
current is 0 mA when paused). The rate is continuously variable from 30 to 180 pulses per
minute (ppm), by increments of 5 ppm (10 bpm when greater than 100 ppm).
The pacing output pulse is delivered to the heart via ZOLL hands-free defibrillation/pacing
electrodes placed on the patient’s back and the precordium.
Proper operation of the device, together with correct electrode placeme nt, is critical to
obtaining optimal results. Every operator must be thoroughly familiar with these operating
instructions.
ECG Monitoring
The patient’s ECG is monitored by connecting the patient to the unit via a 3-, 5-, or 12-lead
patient cable or hands-free therapy electrodes. The ECG waveform is presented on the display
along with the following information:
• averaged heart rate, derived by measuring R to R intervals
• lead selection - I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (with ECG cable),
The Propaq MD will defibrillate, cardiovert, and monitor ECG using hands-free therapy
electrodes. The Propaq MD will pace using ZOLL hands-free therapy electrodes.
Energy Select, Charge and Shock controls are located on the paddles and front panel. When
using hands-free therapy electrodes, you must use the controls on the front panel of the unit. To
switch between paddles and hands-free therapy electrodes, remove the multifunction cable
(MFC) from the apex paddle and connect the hands-free therapy electrodes to the cable.
You should always check the expiration date on the electrode packaging. Do not use expired
electrodes, which might result in false patient impedance readings and affect the level of
delivered energy, or cause burns.
This symbol on the electrode package is accompanied by the expiration date.
For stat-padz® II, this symbol does not appear; the expiration date appears on the
lower right corner of the label, below the lot number.
Note:ZOLL electrodes contain no hazardous materials and may be disposed of in general
trash unless contaminated with pathogens. Use appropriate precautions when
disposing of contaminated electrodes.
Batteries
Propaq MD models use an easily replaced rechargeable lithium-ion battery pack (the
SurePower II Battery Pack). A new, fully charged battery pack typically delivers more than 6
hours of ECG monitoring. Use of other functions (such as the defibrillator, printer, or
pacemaker) reduces this time.
When a LOW BATTERY icon appears on the display and the unit emits three beeps in
conjunction with the displayed battery icon, the battery must be replaced and recharged.
You can charge the battery by either of the following methods:
• Internal charging — plug the Propaq MD into an auxiliary power adapter to automatically
begin charging the installed battery pack. The front panel battery indicator operates as
follows:
When the indicator is:It means:
Steady yellowBattery is charging.
Steady greenBattery is charged.
Alternating yellow and
green
Not litNo battery in device.
The charge state cannot be
determined or a battery charging
fault has been detected.
Note:Upon power up, it takes approximately 45 seconds for the LEDs on the battery to
accurately display run time.
1-10www.zoll.com 9650-0805-01 Rev. A
Propaq MD Product Functions
• External charging — use the ZOLL SurePower™ Battery Charger with the Propaq MD
battery adapter to charge the battery pack and test the battery’s capacity. For details, refer to
the SurePower II Battery Pack Guide.
The Recalibration LED icon () lights for approximately 10 seconds (after you press and
release the Display button) if the battery needs to be calibra ted. If the Recalibrat ion LED lights,
the runtime indicator will not display run time for that battery. For best performance of the
battery, you should recalibrate the battery as soon as possible.
To manually recalibrate the SurePower Battery Pack, you can insert the battery into the
SurePower Charger Station and perform a Manual Test (for more information, see the ZOLL
SurePowerCharger Station Operator’s Guide).
After you recalibrate the battery, the Recalibration LED will only flash when you press the
Display button.
9650-0805-01 Rev. APropaq MD Operator’s Guide1-11
CHAPTER 1GENERAL INFORMATION
Ready For Use (RFU) Indicator
The Propaq MD has an RFU indicator on the front panel that indicate s if the device is ready for
use. The RFU indicator has three states which are described in the following table.
StateDescriptionAction
Ready for UseThe device is ready for use. Patient
monitoring, defibrillation, and
pacing parameters are functional
and the battery is above the low
battery capacity.
Note: If the device is plugged into
the auxiliary power adapter, the
Ready for Use indicator may display
even if the battery is depleted.
Check the status of the battery
before removing the device from the
auxiliary power adapter.
FlashingOne or more of the following has
occurred:
• The battery is not properly
installed.
• A low battery is installed.
• A battery fault has occurred.
• There is no battery installed
while connected to auxiliary
power.
• One or more patient monitoring
parameters have failed self-test
(NIBP, SpO2, CO2, IBP, or
Temp).
• The front panel button self-test
failed.
• The speech database self-test
failed.
None required.
Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to flash, remove the unit
from service and contact the
appropriate technical personnel or
the ZOLL Technical Service
Department.
Do Not UseOne or more of the following has
occurred:
• The battery is not properly
installed.
• No battery is installed and
auxiliary power is not present.
• A very low battery (below
software shutdown limit) was
installed.
• ECG, defibrillator, or pacer selftests have failed, or other critical
self-tests have failed.
Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to display the Do Not
Use symbol, remove the unit from
service and contact the appropriate
technical personnel or the ZOLL
Technical Service Department.
1-12www.zoll.com 9650-0805-01 Rev. A
Safety Considerations
All operators should review these safety considerations before using the Propaq MD unit.
Propaq MD units are high-energy defibrillators capable of delivering 200 joules. To completely
deactivate the unit, press the power switch to turn the unit off.
To manually disarm a charged (or charging) defibrillator, do one of the following:
• Press the Disarm quick access key .
• Change the selected energy.
• Press the power switch to turn the unit off.
For safety , the Propaq MD a utomatically disarms i f left char ged for more than 60 seconds if the
shock button () is not pressed.
Warnings
General
Federal (U.S.A.) law restricts this defibrillator to sale by or on the order of a physician.
Safety Considerations
Only appropriately trained, skilled personnel who are familiar with equipment operation should
perform emergency defibrillation. The prescribing physician should determine what training,
such as Advanced Cardiac Life Support (ACLS) or Basic Life Support (BLS) certification, is
appropriate.
Only skilled personnel trained in Advanced C ardiac Life Support (ACLS) and who are familiar
with equipment operation should perform synchronized cardioversion. The precise cardiac
arrhythmia must be determined before attempting defi bril lat io n.
These operating instructions describe the functions and proper operation of the Propaq MD
products. They are not a substitute for a formal patient care training course. Operators should
obtain formal training from an appropriate authority before using this defibrillator for patient
care.
Proper operation of the unit and correct electrode placement is critical to obtaining optimal
results. Operators must be thoroughly familiar with proper device operation.
The use of external pacing/defibrillation electrodes, accessories, or adapter devices from
sources other than ZOLL is not recommended. ZOLL makes no representations or warranties
regarding the performance or effectiveness of its products when used with pacing/defibrillation
electrodes or adapter devices from other sources. Defibrillator failures attributable to the use of
pacing/defibrillation electrodes or adapters not manufactured by ZOLL might void ZOLL’s
warranty.
At receipt of shipment, check pacing/defibrillation electrodes to ensure compatibility.
Allow ample slack in cables to make sure that cables do not tug at electrodes.
Do not disassemble the unit. A shock hazard exists. Refer all problems to authorized service
personnel.
Follow all recommended maintenance instructions. If a problem occurs, obtain service
immediately. Do not use the defibrillator until it has been inspected by appropriate personnel.
9650-0805-01 Rev. APropaq MD Operator’s Guide1-13
CHAPTER 1GENERAL INFORMATION
The Propaq MD unit might not perform to specifications when stored at the upper or lower
extreme limits of storage temperature and then immediately put into use. The Propaq unit
should not be stored or used outside of the environmental limits p rovided in Appendix A of this
manual.
Avoid using the Propaq MD adjacent to, or stacked on, other equipmen t. If unav oidab le, verify
that the unit operates normally in this configuration before clinical use.
The Propaq MD should be installed and put into service according to the EMC information in
Appendix A of this manual.
Do not use internal paddles while the Propaq MD auxiliary power source is connected to an
aircraft AC power operating at a frequency of 400 Hz.
The use of accessories, transducers, and cables other than those specified in this manual and
related Propaq MD option manual inserts may result in increased emissions or decreased
immunity of the Propaq MD.
Perform functional test of internal paddles prior to use.
Do not use or place the unit in service if the Ready For Use indicator (at the upper right of the
front panel) displays a red circle with a line through it.
Carefully route patient cabl es to avoid tri pping over the m, or inadverte ntly pulli ng the unit onto
the patient.
Always inspect the unit for damage if it has been dropped.
Only authorized personnel should use the Supervisor menus.
If uncertain about the accuracy of any measurement, first check the patient’s vital signs by
alternate means, and then make sure the monitor is functioning correctly.
ECG Monitoring
Implanted pacemakers might cause the heart rate meter to count the pac emak er rate during
incidents of cardiac arrest or other arrhythmias. Dedicated pacemaker detection circuitry may
not detect all implanted pacemaker spikes. Check the patient's pulse; do not rely solely on heart
rate meters. Patient history and physical examination are important factors in determining the
presence of an implanted pacemaker. Pacemaker patients should be carefully observed. See
“Pacemaker Pulse Rejection:” on page A-15 of this manual for disclosure of the pacemaker
pulse rejection capability of this instrument.
Use only ECG electrodes that meet the AAMI standard for electrode performance
(AAMI EC-12). Use of electrodes not meeting this AAMI standard could cause the ECG trace
recovery after defibrillation to be significantly delayed.
Prior to attempting synchronized cardioversion, ensure the ECG signal quality is good and that
sync markers are displayed above each QRS complex.
Do not place electrodes directly over an implanted pacemaker.
The Propaq MD unit detects ECG electrical signals only. It does not detect a pulse (effective
circulatory perfusion). Always verify pulse and heart rate by physical assessment of the patient.
Never assume that the display of a nonzero heart rate means that the pat ient has a pulse.
Excessive artifact can result due to improper skin preparation of the electrode site s. Follow skin
preparation instructions in Chapter 6: “Monitoring ECG.”
1-14www.zoll.com 9650-0805-01 Rev. A
Equipment such as electrocautery or diathermy equipment, RFID readers, electronic ar ticle
surveillance (EAS) systems, or metal detectors that emit strong radio frequency signals can
cause electrical interference and distort the ECG signal displayed by the monitor, thereby
preventing accurate rhythm analysis. Ensure adequate separation between such emitters, the
device, and the patient when performing rhythm analysis.
Shock Hazard: Use of accessories, other than those specified in the operating instructions, may
adversely affect patient leakage currents.
Certain line-isolation monitors may cause interference on the ECG display and may inhibit
heart rate alarms.
Monitoring ECG through the paddles may result in inaccurate heart rate display due to artifact.
Defibrillation
The ZOLL Propaq MD can deliver 200 joules of electrical energy . If this electrical ener gy is not
discharged properly, as described in the this manual, the electrical energy could cause persona l
injury or death to the operator or bystander.
To avoid possible damage to the Propaq MD unit, turn off pacing before defibrillating the
patient with a second defibrillator.
Warnings
After a synchronized cardioversion, the SYNC mode may be cleared after each shock or
disarm. The user may have to reselect (press) the SYNC button after each synchronized
cardioversion shock performed on a patient. In Defib/Pacer Default settings in the Supervisor
Setup menu, the Propaq MD can be configured to remain in the SYNC mode after each
synchronized cardioversion.
Synchronized cardioversion can be performed in the paddle monitoring mode. However, it is
possible that artifact can be produced by the moving paddles, which could cause the
defibrillator to trigger on the artifact. It is recommended that monitoring in leads I, II or III be
used during synchronized cardioversion. Paddle monitoring should not be used for elective
cardioversions procedures.
To avoid stress to the defibrillator or the tester, never attempt to repeatedly charge and
discharge the defibrillator in rapid succession. If a need for repetitive testing arises, allow a
waiting period of at least 2 minutes after every third discharge.
In the SYNC mode, the defibrillator does not discharge without a command signal (R-w ave
detection) from the ECG monitor indicate d by a SYNC marker on the trace and a flashing
SYNC indicator.
If conductive gel forms a continuous path between the defibrillator electrodes, delivered energy
may be dramatically reduced to zero. In this case, reposition the electrodes to eliminate the
shunting path before attempting additional shocks.
Improper defibrillation technique can cause skin burns. To limit possible skin burns, use only
ZOLL defibrillation gel on paddles, ensure the gel covers the entire paddle surface and press
firmly against patient’s chest.
If a new energy level is selected after the CHARGE button is pushed and while the defibrillator
is charging or charged, the defibrillator will disarm. The CHARGE button will need to be
pressed again to charge to the new energy level.
Prior to defibrillation, disconnect from the patient any medical electronic device that is not
labeled “defibrillation protected.”
9650-0805-01 Rev. APropaq MD Operator’s Guide1-15
CHAPTER 1GENERAL INFORMATION
Before charging the defibrillator, verify that the energy selected on the display is the desired
output.
Defibrillation takes priority over external pacing. Should the defibrillator be charged during the
administration of external pacing, the pacer turns off and the defibrillator charges to the
selected energy.
Pacing
Ventricular fibrillation does not respond to pacing and requires immediate defibrillation.
Therefore, the patient’s dysrhythmia must be determined immediately, so that you can employ
appropriate therapy . If t he patient is in ventricula r fibrillation and defibrilla tion is successful but
cardiac standstill (asystole) ensues, you should use the pacemaker.
Ventricular or supraventricular tachycardias can be interrupted with pacing, but in an
emergency or during circulatory collapse, synchronized cardioversion is faster and more
certain.
Pulseless electrical activity (PEA) can occur following prolonged cardiac arrest or in other
disease states with myocardial depression. Pacing might then produce ECG responses wi th out
effective mechanical contractions, making other effective treatment necessary.
Pacing can evoke undesirable repetitive resp onses, tachycardia, or fibrillation in the presence of
generalized hypoxia, myocardial ischemia, cardiac drug toxicity, electrolyte imbalance, or
other cardiac diseases.
Pacing by any method tends to inhibit intrinsic rhythmicity. Abrupt cessation of pacing,
particularly at rapid rates, can cause ventricular standstill and should be avoided.
Noninvasive temporary pacing can cause discomfort of varying intensity, which occasionally
can be severe and preclude its continued use in conscious patients.
Similarly, unavoidable skeletal muscle contraction might be troublesome in very sick patients
and might limit continuous use to a few hours. Erythema or hyperemia of the skin under the
hands-free therapy electrodes often occurs; this effect is usually enhanced along the perimeter
of the electrode. This reddening should lessen substantially withi n 72 hours.
There have been reports of burns under the anterior electrode when pacing adult patients with
severely restricted blood flow to the skin. Prolonged pacing should be avoided in these cases
and periodic inspection of the underlying skin is advised.
There are reports of transient inhibition of spontaneous respiration in unconscious patients with
previously available units when the anterior electrode was placed too low on the abdomen.
The pacing rate determination can be adversely affecte d by arti fact. If the patient’s pulse and
the heart rate display are sig nificantly different, external pacing pulses may not be delivered
when required.
Artifact and ECG noise can make R-wave detection unreliable, affecting the HR meter and the
demand mode pacing rate. Always observe the patient closely during pacing operations.
Consider using asynchronous pacing mode if a reliable ECG trace is unob tai nab le .
Transcutaneous pacing should not be used to treat V FIB (ventricular fib rillation). In cases of V
FIB, immediate defibrillation is advised.
1-16www.zoll.com 9650-0805-01 Rev. A
Warnings
Transcutaneous pacing may cause discomfort ranging from mild to severe, depending on the
patient’s tolerance level, muscle contractions and electrode placement. In certain cases,
discomfort may be decreased by slightly relocating the pacing pads.
It is important to monitor the patient closely to verify that both mechanical and electrical
capture are occurring. Electrical capture can be verified by observing the presence of a large
ectopic beat after the pacing pulse is delivered. The size and morphology of the beat are
dependent on the patient. In some instances the be at may appe ar as a rel ati vel y normal looking
QRS pulse. Mechanical capture can be verified by checking for signs of increased blood flow
i.e., reddening of the skin, palpable pulses, increased blood pressure, etc. Continuously observe
the patient during pacing administration , to insure capture retention. Do not leave the patient
unattended when administering external pacing therapy.
Warning!This device can only be used for external pacing of patients and cannot be used for
internal pacing. Do not conn ect internal pacing lead wires to the Propaq MD defibrillator.
Pulse Oximeter
Keep the ZOLL finger probe clean and dry.
SpO2 measurements may be affected by certain patient conditions: severe right heart failure,
tricuspid regurgitation or obstructed venous return.
SpO
measurements may be affected when using intravascular dyes, in extreme
2
vasoconstriction or hypovolemia or under conditions where there is no pulsating arterial
vascular bed.
SpO
measurements may be affected in the presence of strong EMI fields, electrosurgical
2
devices, IR lamps, bright lights, improperly applied sensors; the use of non-ZOLL sensors, or
damaged sensors; in patients with smoke inhalation, or carbon monoxide poisoning, or with
patient movement.
Tissue damage can result if sensors are applied incorrectly, or left in the same location for an
extended period of time. Move sensor every 4 hours to reduce possibility of tissue damage.
Do not use any oximetry sensors during MRI scanning. MRI procedures can cause conducted
current to flow through the sensors, causing patient burns.
Do not apply SpO
when the arterial circulation is cut off during NIBP measurements, and may affect SpO
sensor to the same limb that has an NIBP cuff. The SpO2 alarm may sound
2
2
measurements.
In some instances, such as obstructed airway, the patient's breathing attempts may not produce
any air exchange. These breathing attempts can still produce chest size changes, creating
impedance changes, which can be detected by the respiration detector. It is best to use the pulse
oximeter whenever monitoring respirations, to accurately depict the patient's respiratory
condition.
9650-0805-01 Rev. APropaq MD Operator’s Guide1-17
CHAPTER 1GENERAL INFORMATION
Noninvasive Blood Pressure
Only a physician can interpret pressure measurements.
Blood pressure measurement results may be affected by the position of the patient, his or her
physiological condition and other factors.
Substitution of a component different from that supplied by ZOLL (e.g., cuff, hoses, etc.) may
result in measurement error. Use only ZOLL-approved cuffs and hoses. To avoid the risk of
intravenous line misconnection and possible introduction of air into a patient’s blood, do not
modify the NIBP system or hoses with Luer Lock adapters.
IBP
Do not use a blood pressure cuff on the limb being used for IV infusion or for SpO
Accurate pressure readings may not be achieved on a person experiencing arrhythmias,
shaking, convulsions or seizures. Medication may also affect pressure readings. The correct
size cuff is essential for accurate blood pressure readings.
Blood pressure hoses must be free of obstructions and crimps.
If the patient’s cuff is not at heart level, an error in measurement may result.
When monitoring blood pressure at frequent intervals, observe the cuffed extremity of the
patient for signs of impeded blood flow.
Do not monitor one patient’s NIBP while monitoring another patient’s ECG.
Blood pressure measurement may be inaccurate if tak en while acc elera ting or decelera ting in a
moving vehicle.
If an NIBP measurement result is questionable or “motion” indication is displayed, repeat the
measurement. If the repeated measurement result is still questionable, use another blood
pressure measurement method.
Do not use the NIBP on cardiopulmonary bypass patients.
To ensure compatibility and electrical safety, accessory pressure sensors should comply with
ANSI/AAMI BP-22 and IEC 60601-2-34 for IBP or ANSI/AAMI NS28 for ICP.
monitoring.
2
Follow instructions supplied with any accessory pressure sensor regarding calibration and
removal of trapped air.
Avoid touching metal parts of any transducer while it is in contact with the patient.
Do not reuse any components that are labeled for single use only.
Transducers should be rated to withstand an accidental drop of at least a meter onto a hard
surface.
Transducers that are subject to immersion in liquids should be rated as watertight.
1-18www.zoll.com 9650-0805-01 Rev. A
CO
Warnings
2
During MRI scanning, the monitor must be placed outside the MRI suite. When the monitor is
used outside the MRI suite, EtCO
which permits placement of the monitor outside the MRI suite.
When using the monitor with anesthetics, nitrous ox ide or high concentrations of oxygen,
connect the gas outlet to a scavenger system.
monitoring can be implemented using a long FilterLine®
2
Use only Oridion Microstream CO
Microstream CO
lines.
If using the CO
or when it becomes occluded.
CO
readings and respiratory rate can be affected by sensor application errors, certain ambient
2
environmental conditions, and certain patient co nd itions.
sampling lines are labeled for single patient use onl y. Do not reuse sampling
2
Monitor for extended critical care, replace the airway adapter every 24 hours
2
Respiration
Do not operate the Propaq MD with any other monitor with respiration measurements on the
same patient. The two devices could affect the respiration accuracy.
The device should not be used as an apnea monitor.
Ferromagnetic Equipment
Biomedical equipment and accessories, such as ECG electrodes, cables, and oximeter probes
contain ferromagnetic materials. Ferromagnetic equipment mu st no t be used in the presence of
high magnetic fields created by magnetic resonance imaging (MRI) equipment.
sampling lines.
2
The large magnetic fields generated by an MRI device can attract ferromagnetic equipment
with an extremely violent force, which could cause serious personal injury or death to persons
between the equipment and the MRI device.
Battery
Although the device can operate with auxiliary power alone, ZOLL strongly recommends that
you operate the unit with a battery installed at all times. Operating the unit with a battery
provides a backup in case of ac power shortage, and results in faster charge time. The battery
can be automatically recharged while it is installed in the unit. Keep a fully charged spare
battery pack with the defibrillator at all times.
Test battery packs regularly. A battery that does not pass the ZOLL charger’s capacity test
might cause the Propaq MD unit to shut down unexpectedly.
If the Low Battery indication occurs at any time during operation, immediately replace the
battery pack.
9650-0805-01 Rev. APropaq MD Operator’s Guide1-19
CHAPTER 1GENERAL INFORMATION
If the LOW BATTERY icon appears, plug the Propaq MD unit into a power source or install a
fully charged battery pack. When the warning low battery shutdown prompt appears,
immediately replace the battery pack with a fully char ged pack or plug the Propaq MD unit into
a power source, as unit shut down due to a low battery condition is imminent.
If mistreated, a battery pack might explode. Do not disassemble a battery pack or dispose of it
in fire.
Operator Safety
The Propaq MD can deliver 200 joules of electrical energy. If this electrical energy is not
discharged properly, as described in this manual, the electrical energy could cause personal
injury or death to the operator or bystanders.
Do not use the Propaq MD in the presence of oxygen-rich atmospheres, flammable anesthetics,
or other flammable agents (such as gasoline). Using the unit in such environments might cause
an explosion.
Do not use the unit near or within standing water. Electrical safety might be compromised wh en
the defibrillator is wet.
Never discharge the unit with the defibrillation electrodes or paddles shorted together or in
open air.
Do not discharge the defibrillator except as indicated in the instructions. Discharge the
defibrillator only when defibrillation electrodes or paddles are properly applied to the patient.
To avoid risk of electrical shock, do not touch the gelled area of the hands-free therapy
electrodes during pacing or defibrillation.
T o avoid risk of electrical shock, do not allow electrolyte gel to accumulate on hands or paddle
handles.
To avoid risk of electrical shock, do not allow pa tie nt connectors to contact other conductive
parts, including earth.
For defibrillation using paddles, use only high-conductivi ty electrolyte gel specified for such
use by the manufacturer.
When using paddles for defibrillation, use your th umb s to operat e th e SHOCK buttons. Doing
so avoids inadvertent shock to the operator.
The use of accessory equipment that does not comply with the equivalent safety requirements
of the Propaq MD defibrillator could reduce the level of safety of the combined system. When
choosing accessory equipment, consider the following:
• Use of the accessory in the patient vicinity.
• Evidence that the safety certification of the accessory has been performed in accordance
with the appropriate IEC (EN) 60601-1 and/or IEC (EN) 60601-1-1 harmonized national
standards.
Always check that the equipment functions properly and is in proper condition before use.
Disconnect all electro-medical equipment that is not defibrillation-protected from the patient
prior to defibrillation.
Before discharging the defibrillator, warn everyone to STAND CLEAR of the patient.
1-20www.zoll.com 9650-0805-01 Rev. A
Loading...
+ 208 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.