(a) ZOLL Medical Corporation warrants to the original equipment purchaser that beginning on the date of installation, or thirty (30) days after the date of shipment from
ZOLL Medical Corporation's facility, whichever first occurs, the equipment (other than accessories and electrodes) will be free from defects in material and workmanship
under normal use and service for the period of one (1) year. Durin g such peri od ZOLL Medical Corpor ation will, at no charge to the customer, either repair or replace (at
ZOLL Medical Corporation's sole option) any part of the equipment found by ZOLL Medical Corporation to be defective in material or workmanship. If ZOLL Medical
Corporation's inspection detects no defects in material or workmanship, ZOL L Medical Corporation's regular service char ges shall apply. (b) ZOLL Medical Corporation
shall not be responsible for any equipment defect, the failure of the equipment to perform any function, or any other nonconformance of the equipment, caused by or
attributable to: (i) any modification of the equipment by the customer, unless such modification is made with the prior written approval of ZOLL Medical Corporatio n; (ii)
the use of the equipment with any associated or complementary equipment, (iii) installation or wiring of the equipment other than in accordance with ZOLL Medical
Corporation's instructions. (c) This warranty does not cover items subject to normal wear and burnout during use, including but not limited to lamps, fuses, batteries, patient
cables and accessories. (d) The foregoing warranty constitutes the exclusive remedy of the customer and the exclusive liability of ZOLL Medical Corporation for any
breach of any warranty related to the equipment supplied hereunder. (e) Limitation of Liability: ZOLL shall not in any event be liable to Purchaser, nor shall Purchaser
recover, for special, incidental or consequential damages resulting from any breach of warranty, failure of essential purpose, or under any other legal theory including but
not limited to lost profits, lost savings, downtime, goodwill, damage to or replacement of equipment and property, even if ZOLL has been advised of the possibility of such
damages.
THE WARRANTY SET FORTH HEREIN IS EXCLUSIVE AND ZOLL MEDICAL CORPORATION EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES
WHETHER WRITTEN, ORAL, IMPLIED, OR STATUTORY, INCLUDING BUT NOT LIMITED TO ANY WARRANTIES OF MERCHANTABILITY OR FITNESS
FOR A P ARTICULAR PURPOSE.
For additional information, please call ZOLL Medical Corporation at 1-800-348-9011 (in Massachusetts 1-978-421-9655). International customers should call the nearest
authorized ZOLL Medical Corporation service center.
Software License
Read this License agreement carefully before operating any of the M Series products.
Software incorporated into the system is protected by copyright laws and international copyright treaties as well as other intellectual property laws and treaties. This
software is licensed, not sold. By taking delivery of and using this system, the Purchaser signifies agreement to and acceptance of the following terms and conditions:
Grant of License: In consideration of payment of the software license fee which is part of the price paid for this product ZOLL Medical Corporation grants the Purchaser
a non-exclusive license, without right to sublicense, to use the system software in object-code form only.
Ownership of Software/Firmware: Title to, ownership of and all rights and interests in the system software and all copies thereof remain at all times vested in the
manufacturer, and Licensors to ZOLL Medical Corporation and they do not pass to Purchaser.
Assignment: Purchaser agrees not to assign, sub-license or otherwise transfer or share its rights under the license without the express written permission of ZOLL Medical
Corporation.
Use Restrictions: As the Purchaser, you may physically transfer the products from one location to another provided that the software/firmware is not copied. You may not
disclose, publish, translate, release or distribute copies of the software/firmware to others. You may not modify , adapt, translate, reverse engineer, decompile, crosscompile,
disassemble or create derivative works based on the software/firmware.
No Implied License
Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would, alone, or in combination with
this device, fall within the scope of one or more of the patents relating to this device.
TABLEOF CONTENTS
PREFACE ....................................................................................................................................................................................... V
Additional Reference Material ........................................................................................................................................................... vi
Conventions ....................................................................................................................................................................................... vii
Service Policy Warranty .................................................................................................................................................................... vii
Technical Service .............................................................................................................................................................................. vii
Technical Service for International Customers ................................................................................................................................ viii
Before You Begin the Maintenance Tests ............................................................................................................................................2
Equipment You Need to Perform the Maintenance Tests ....................................................................................................................2
Equipment You Need for the M Series Options Maintenance Tests ....................................................................................................3
Zoll M Series Error Messages ............................................................................................................................................................42
Parts That May Need Replacing After Disassembly ..........................................................................................................................60
Overview of Modules .........................................................................................................................................................................62
1. Removing the ZIF Keeper ............................................................................................................................................65
2. Removing the Front Panel ............................................................................................................................................66
2A.Removing the Display .................................................................................................................................................67
2B.Removing the Control Board ......................................................................................................................................68
3. Removing the Upper Housing Assembly .....................................................................................................................70
i
M Series Service Manual
4. Removing the System Board Assembly .......................................................................................................................71
5. Removing the Battery Interconnect Board Assembly ..................................................................................................73
6. Removing the High Voltage/Charger Assembly .........................................................................................................74
7. Removing the High Voltage Module Assembly ..........................................................................................................76
8. Removing the High Voltage Capacitor Assembly .......................................................................................................77
9. Removing the System Interconnect Board ...................................................................................................................78
10.Removing the Printer/Recorder Motor ........................................................................................................................79
11.Removing the Lower Housing Assembly ....................................................................................................................80
12.Removing the Print Head Assembly ............................................................................................................................81
13.Removing the PCMCIA Card Slot Assembly .............................................................................................................82
14.Removing the Paddle Release Latch ...........................................................................................................................83
CHAPTER 4 REPLACEMENT PARTS ...........................................................................................................................................85
Replacement Parts ..............................................................................................................................................................................86
Field Replacement Parts .....................................................................................................................................................................90
Main System Board ............................................................................................................................................................................94
Main System Board Functions ...........................................................................................................................................................96
Power Supply ................................................................................... ...................................................................................................98
ECG Front End ...................................................................................................................................................................................99
Multifunction Electrode (MFE)/PADS (System Board and High Voltage Module) .........................................................................99
CPU and EPU .....................................................................................................................................................................................99
High Voltage Module .......................................................................................................................................................................100
Defibrillator Charging and Discharging ...........................................................................................................................................101
High Voltage Capacitor Monitor ......................................................................................................................................................102
Pacer/Defibrillator Control Signals ..................................................................................................................................................103
System Interconnect Module ............................................................................................................................................................105
Front Panel and Controls PWBA ......................................................................................................................................................106
M Series Options ............................................................... ...............................................................................................................106
ii
M Series Service Manual
Isolated Power Supply Module .........................................................................................................................................................107
12 Lead Option .................................................................................................................................................................................107
INDEX ..........................................................................................................................................................................................113
Interconnect Diagram for the M Series Monophasic Unit ...............................................................................................................118
Interconnect Diagram for the M Series Biphasic Unit .....................................................................................................................119
Interconnect Diagram for the M Series CCT Biphasic Unit ............................................................................................................120
iii
M Series Service Manual
iv
M Series Service Manual
Overview
ZOLL Medical Corporation’s M Series Service Manual is intended for the service technician whose responsibility is
to identify malfunctions and/or make repairs at the subassembly level. The Zoll M Series Service Manual has five
main sections and one appendix.
Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to review this section
thoroughly before attempting to use or service the M Series unit.
Chapter 1—Maintenance T ests explains how to check the defibrillator’s performance using a series of recommended
checkout procedures to be conducted every six months.
Chapter 2—Troubleshooting provides a listing of the procedures and error messages to help the service technician
detect faults and repair them.
Chapter 3—Disassembly Procedures describes step-by-step procedures for removing subassemblies from the
M Series unit.
Chapter 4—Replacement Parts List displays a complete list of ZOLL part numbers for field replaceable parts
available for the M Series unit, allowing the service person to identify and order replacement parts from ZOLL.
Chapter 5—Functional Description provides technical descriptions for the M Series major subassembly modules.
Appendix A—M Series Operator’s Manual.
Preface
Safety Considerations
The following section describes general warnings and safety considerations for operators and patients. Service
technicians should review the safety considerations prior to servicing any equipment and read the manual carefully
before attempting to disassemble the unit. Only qualified personnel should service the M Series unit.
Federal (U.S.A.) law restricts this unit for use by or on the order of a physician.
Safety and effectiveness data submitted by ZOLL Medical Corporation to the Food and Drug Administration (FDA)
under section 510(K) of the Medical Device Act to obtain approval to market is based upon the use of ZOLL
accessories such as disposable electrodes, patient cables and batteries. The use of external pacing/defibrillation
v
M Series Service Manual
electrodes and adapter units from sources other than ZOLL is not recommended. ZOLL makes no representations or
warranties regarding the performance or effectiveness of its products when used in conjunction with pacing/
defibrillation electrodes and adapter units from other sources. If unit failure is attributable to pacing/defibrillation
electrodes or adapter units not manufactured by ZOLL, this may void ZOLL's warranty.
Only qualified personnel should disassemble the M Series unit.
WARNING! This unit can generate up to 4500 volts with sufficient current to cause lethal shocks.
All persons near the equipment must be warned to “STAND CLEAR” prior to discharging the defibrillator.
Do not discharge the unit’s internal energy more than three times in one minute or damage to the unit may result.
Do not discharge a battery pack except in a Base PowerCharger
Do not use the M Series in the presence of flammable agents (such as gasoline), oxygen-rich atmospheres, or
flammable anesthetics. Using the unit near the site of a gasoline spill may cause an explosion.
Do not use the unit near or within puddles of water.
NOTEThe M Series is protected against interference from radio frequency emissions typical of two-way radios and cellular
phones (digital and analog) used in emergency service/public safety activities. Users of the M Series should assess the
unit’s performance in their typical environment of use for the possibility of radio frequency interference from
high-power sources. Radio Frequency Interference (RFI) may be observed as shifts in monitor baseline, trace
compression, or transient spikes on the display.
4x4
or compatible ZOLL Battery Charging/Testing unit.
Additional Reference Material
In addition to this guide, there are several other components to the Zoll M Series documentation. They include:
• Operator’s Guide - A comprehensive reference work that describes all the user tasks needed to operate the M Series.
• Configuration Guide - Describes the M Series features and functions whose operation can be customized by
authorized users.
vi
M Series Service Manual
Conventions
WARNING!Warning statements describe conditions or actions that can result in personal injury or death.
CAUTIONCaution statements describe conditions or actions that can result in damage to the unit.
NOTENotes contain additional informationon using the defibrillator.
Service Policy Warranty
In North America: Consult your purchasing agreement for terms and conditions associated with your warranty. Outside
of North America, consult ZOLL authorized representative.
In order to maintain this warranty , the instructions and procedures contained in this manual must be strictly followed.
For additional information, please call the ZOLL Technical Service Department 1-800-348-9011 in North America.
Technical Service
If the ZOLL M Series unit requires service, contact the ZOLL Technical Service Department:
Telephone: 1-978-421-9655; 1-800-348-9011
Fax 1-978-421-0010
Have the following information available for the Technical Service representative:
• Unit serial number.
• Description of the problem.
• Department where equipment is used.
• Sample chart recorder strips documenting the problem, if applicable.
• Purchase Order to allow tracking of loan equipment.
• Purchase Order for a unit with an expired warranty.
vii
M Series Service Manual
If the unit needs to be sent to ZOLL Medical Corporation, obtain a service order request number from the Technical
Service representative. Return the unit in its original container to:
Technical Service for International Customers
International customers should return the unit in its original container to the nearest authorized ZOLL Medical
Corporation Service Center. To locate an authorized service center, contact the International Sales Department at
ZOLL Medical at the above address.
Units are available on loan while your unit is being repaired.
ZOLL Medical Corporation
269 Mill Road
Chelmsford, Massachusetts 01824-4105
Attn: Technical Service Department
Telephone: 1-800-348-9011; 1-978-421-9655 FAX: 978-421-0010
viii
M Series Service Manual
Overview
The M Series has two checkout procedures: the operator’s shift checklist and the extensive six-month maintenance
tests checkout procedures.
Because the M Series units must be maintained ready for immediate use, it is important for users to conduct the
Operator’s Shift Checklist procedure at the beginning of every shift. This procedure can be completed in a few minutes
and requires no additional test equipment. (See the ZOLL M Series Operator’s Guide for the Operator’s Shift
Checklist.)
A qualified biomedical technician must perform a more thorough maintenance test checkout every six months to
ensure that the functions of the M Series unit work properly. This chapter describes the step by step procedures for
performing the six month maintenance test checkout. Use the checklist at the back of this document (ZOLL M Series
Maintenance Tests Checklist) to record your results of the maintenanc e tests.
Chapter 1
Maintenance Tests
This chapter describes the following maintenance tests:
• 1. Physical Inspection of the Unit
• 2. Front Panel Button Test
• 3. 3, 5, and 12 Leads Test
• 4. Power Supply Test
• 5. Leakage Current Test
• 6. Paddles Test
• 7. Heart Rate Display Test
• 8. Calibrating Pulses on Strip Chart Test
• 9. Notch Filter Test
• 10. Heart Rate Alarm Test
1
M Series Service Manual
• 11. Defibrillator Self Test
• 12. Synchronized Cardioversion Test
• 13. Shock Test
• 14. Summary Report Test
• 15. Advisory Message Test
• 16. Pacer Test
•17. SpO
•18. EtCO
Before You Begin the Maintenance Tests
• Assemble the tools or specialized testing equipment listed in the “Equipment You Need to Perform the Maintenance
• Keep an extra fully charged ZOLL M Series battery available.
• Schedule an hour to conduct the entire maintenance test.
• Photocopy the checklist at the back of this document and use the copy to record your results. As you conduct each
Monitor Test
2
Monitor Test
2
Tests” section shown below.
step of a procedure, mark the Pass/Fail/NA check boxes on your checklist and then save it for your maintenance file.
• Perform the tests in the order presented.
• Perform all the steps of each test procedure.
• Complete all the steps of the procedure before evaluating the test results.
Equipment You Need to Perform the Maintenance Tests
For testing purposes, you can substitute an equivalent device.
• Zoll Medical Electrode Adapter from Dynatech Nevada Inc. (DNI part number 3010-0378).
• Dynatech Impulse 4000 Defibrillator Analyzer with 1.06 software or higher.
•Bio-Tek
•Bio-Tek
• Novametrix Medical Systems, Inc. Capnostat Simulator TB1265/7100 or equivalent. (For EtCO
®
601 Pro Series International Safety Analyzer.
®
Index 2PFE SpO2 Simulator or equivalent. (For SpO2 units only.)
2
2
units only.)
M Series Service Manual
• ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201).
• Stop watch.
• Standard series II PC flash memory cards.
• PCMCIA card reader and PC.
• Zoll Data Control (ZDC) for Windows
• Phillips #1 screwdriver.
• Phillips #2 screwdriver.
• Flatblade screwdriver.
• Needle nose pliers without teeth.
• Orange (wooden) sticks.
Equipment You Need for the M Series Options Maintenance
®
software from Pinpoint T echnologies, Version 1.5 or higher (no equivalent)
or Zoll Data Control (ZDC) for DOS software, Version 5.5 or higher (no equivalent).
Tests
•SpO2 cable and sensor (if option is installed).
•EtCO
• Paddles.
•Printer Paper.
• Battery.
• AC line cord.
• 3 lead, 5 lead and 12 lead ECG cables. (12 lead cable needed if 12 lead option is installed.)
cable and sensor (if option is installed).
2
3
M Series Service Manual
1.0 Physical Inspection of the Unit
Tools NeededNone.
Test SetupNone.
Observe this...Pass/Fail
1.1Housing
Is the unit clean and undamaged?
1.2Does the unit show signs of excessive wear?
1.3Does the handle work properly?
1.4Does the recorder drawer open and close properly?
1.5Are input connectors clean and undamaged?
1.6Are there any cracks in the housing?
1.7Do the front panel or selector switches have any damage or cracks?
1.8Are there any loose housing parts?
1.9Do the paddle latches work properly?
1.10Paddles
Do the adult and pedi plates have major scratches or show signs of damage?
1.11Do the adult shoes slide on and off easily to expose the covered pedi plates?
oo
oo
oo
oo
oo
oo
oo
oo
oo
oo
oo
1.12Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
1.13Cables
Are all cables free of cracks, cuts, exposed or broken wires?
1.14Are all bend/strain reliefs undamaged and free of excessive cable wear?
4
oo
oo
oo
M Series Service Manual
Observe this...Pass/Fail
1.15Battery
Is the ZOLL battery fully charged?
1.16Is the battery seated in the battery well correctly?
Record your results on the Maintenance Test Checklist.
oo
oo
5
M Series Service Manual
2.0 Front Panel Button Test
Tools NeededNone.
Test SetupDo the following:
• Install strip chart paper into the recorder tray.
• Install the battery in the unit or connect the A/C power cord to the unit and then plug the cord into an electrical outlet.
• Connect the universal cable and ECG cable (3 lead, 5 lead, or 12 lead) to the ZOLL simulator, or Dynatech Impulse
4000 Analyzer (or equivalent).
Do this...Observe this...Pass/Fail/NA
2.1Turn the selector switch to
MONITOR.
(For AED units, turn the
selector switch to ON and
select Manual mode.)
2.2Press the LEAD button;
three times for the 3 lead
cable and seven times for
the 5 lead cable.
2.3Connect the 12 lead cable to
unit and simulator. Press the
LEAD button and select the
lead for each of the 12 lead
settings.
2.4Set the simulator to NSR of
120 BPM. T o check the size
of the ECG waveform,
press the SIZE button.
Listen for 4 beep tones. PADS and MONITOR display on the
monitor.
NOTE: PADS is a factory default setting.
Each time you press the LEAD button, a different lead number
appears under the LEAD heading on the display.
PADS, I, II, III will display a 3 lead ECG cable if connected or no
ECG cable is connected.
PADS, I, II, III, AVR, AVL, AVF, V1 will display a 5 lead ECG
cable.
A 12 Lead cable will display PADS, I, II, III, AVR, AVL, AVF,
VI, V2, V3, V4, V4, V5, V6.
As you press the SIZE button five times (0.5, 1.0, 1.5, 2.0, 3.0),
note that the size of the ECG waveform appropriately changes on
the display.
oo
oo
ooo
ooo
6
M Series Service Manual
Do this...Observe this...Pass/Fail/NA
2.5Press the ALARM
SUSPEND button.
2.6Press the RECORDER
button (if equipped).
2.7Open the paper tray.
Press RECORDER button.
2.8Close the paper tray.
Press RECORDER button.
2.9Press RECORDER button.Strip chart paper stops flowing out of paper tray.
2.10Press the VOLUME
softkey.
To increase the volume of
the beep, press the Inc.
softkey.
Bell changes from disabled to enabled. If the alarm sounds, press
the ALARM SUSPEND button to turn it off. The alarm will only
be suspended for 90 seconds at this point. Press and hold the
ALARM SUSPEND button for 3 seconds to disable alarms.
The strip chart paper moves out of the unit from the paper tray.
Check that the correct time, date, ECG lead annotation and
waveform are recorded on the paper. (Set Time and Date, if
necessary.)
CHECK RECORDER message appears on the monitor.
Strip chart paper flows out of paper tray. Verify that the CHECK
RECORDER message no longer displays.
The volume bar graph displays.
Audible beep when the QRS wave displays. The bar graph
increases on the display indicating an increase in volume. This
action does not increase the volume which is normal.
Note: The QRS tone is on or off. There is no gradual change in
volume. If equipped, voice prompts are gradual. Note: The voice
volume has 5 settings. Setting 3 is in the mid-range.
ooo
ooo
ooo
ooo
ooo
oo
2.11To decrease the volume of
the beep, press the Dec.
softkey.
2.12Press the CONTRAST
button.
The bar graph decreases on the display indicating a decrease in
volume. The volume shuts off at the last bar; otherwise, the
volume is the same as originally set.
Contrast menu displays.
7
oo
oo
M Series Service Manual
Do this...Observe this...Pass/Fail/NA
2.13(For LCD monitors only.)
Press the CONTRAST
button.
To increase the contrast of
the display, press the Inc.
softkey.
2.14To decrease the contrast of
the display, press the Dec.
softkey.
2.15Press the SUMMARY
button (if available).
2.16Press the CODEMARKER
button (if available).
2.17Connect A/C current and
install the battery. Turn the
unit off.
Background light and characters display.
The contrast increases on the monitor display (LCD).
The brightness increases on the monitor display (EL).
Thebar graph increases on the display indicating an increase in
contrast.
NOTE Electro luminescence (EL) displays have only two settings
independent of the bar graph without any gradual changes.
If EL brightness is already set to its highest level,
brightness will not change.
The bar graph decreases on the display indicating a decrease in
contrast (LCD) and brightness (EL). The display contrast and
brightness changes.
Summary menu displays on the monitor showing the summary
report options.
Code marker menu displays.
CHARGER ON indicator lights.
The amber or green lights illuminate.
Note: If both lights flash ON/OFF, the unit is defective or no
battery is installed.
For LCDmonitors
ooo
ooo
oo
ooo
ooo
ooo
2.18If applicable, connect D/C
current and install the
battery. Turn the unit off.
2.19Remove the battery.Note that both charge lights (green and amber) flash alternately.
2.20Replace the battery and the
turn unit on.
CHARGER ON indicator lights.
The amber or green lights illuminate.
The yellow light indicates the battery is being charged. The green
light indicates the battery is fully charged to present capacity.
NOTE If both lights flash ON/OFF, the unit is defective or no
battery is installed.
Note that the yellow charge light illuminates.
8
ooo
oo
oo
M Series Service Manual
Do this...Observe this...Pass/Fail/NA
2.21Press the ANALYZE
button (if available).
2.22Move the selector switch to
DEFIB. Select 2J. Press the
CHARGE button.
2.23Press and hold the
ENERGY SELECT down
arrow.
2.24Press and release the
ENERGY SELECT up
arrow 19 times.
2.25Press the CHARGE button.Note the disp lay shows the unit charged up to 360J (200J -
2.26Press the SHOCK button.The unit discharges and the SHOCK button is no longer lit. A 15
Record your results on the Maintenance Test Checklist.
SELECT DEFIB MODE message appears on the monitor. (For
manual devices.)
The display shows that the unit is charging. The SHOCK button
lights when the unit is charged. Ready tone for DEFIB sounds.
Unit discharges internally and selected energy decrements to 1J.
second strip chart automatically prints, displaying the number of
joules delivered (if equipped with recorder).
ooo
ooo
ooo
ooo
ooo
ooo
9
M Series Service Manual
3.0 3, 5, and 12 Leads Test
Tools Needed3 lead, 5 lead, and 12 lead cables.
Test each cable separately.
Test SetupNote: The M Series unit must be configured to display ECG LEAD OFF message.
Connect the lead wires appropriate to each test to the Dynatech Impulse 4000 or equivalent (Symbio CS1201).
Do this...Observe this...Pass/Fail/NA
3.1Turn the selector switch to
MONITOR. Select leads.
3.2Disconnect one lead from
the simulator .
3.3Reconnect the lead. Repeat
step 3.2 with the remaining
leads.
3.4Repeat 3.2 and 3.3 for 5
lead and 12 lead cables.
Record your results on the Maintenance Tests Checklist.
NO ECG LEADS OFF message displayed.
The ECG LEAD OFF message displays within 3 seconds (if
configured).
Wait for ECG LEAD OFF message to clear from the display (if
configured).
NOTE: If heart rate alarm sounds, press and hold the ALARM
SUSPEND button for 4 seconds to disable the alarms.
NOTE: When testing the 12 lead cable, the ECG LEAD OFF
message displays when you pull off a limb lead. When you pull
off a V lead, the ECG VX LEAD OFF message displays where
“X” is the number between 1 and 6.
ooo
ooo
ooo
ooo
10
M Series Service Manual
4.0 Power Supply Test (Optional)
Tools Needed2 red miniature alligator to miniature alligator leads.
1 black miniature alligator to miniature alligator test lead.
DC power supply (15 Amp minimum).
0.1
Ω 1% resistor (¼W or greater).
1000
Ω 1% ¼W resistor.
Fluke 75 multimeter or equivalent.
Test SetupMake sure the unit and power supply are turned off.
Connect one end of the black lead to the “-” terminal in the battery well.
Connect the other end of the black lead to the “-” terminal of the power supply.
Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it.
Connect the other end of the red lead to the “+” terminal of the power supply.
Set the power supply voltage to 7V.
CAUTIONBe sure to connect the power supply properly to the M Series battery well terminals or damage to the unit may result.
Do NOT raise the power supply voltage above 12V.
Red
+
Battery Well
+
_
15 Amp
Supply
_
Black
11
M Series Service Manual
Do this...Observe this...Pass/Fail
4.1Turn the selector switch to
MONITOR.
(For AED units, turn the
selector switch to ON and
select Manual mode.)
4.2Turn the unit off.
4.3Adjust the power supply
voltage to 10.3V and turn
the selector switch to
MONITOR (for AED
units, turn the selector
switch to ON).
4.4Low Battery Test
Set voltage to 9.8V.
4.5Set voltage to 9.3V.LOW BATTERY message displays within 30 seconds.
4.6Shut Down Voltage Test
Set voltage to 8.5V.
Record your results on the Maintenance Tests Checklist.
The unit should not turn on.
The unit should turn on.
No LOW BATTERY message displays.
Unit should shut off within 30 seconds.
oo
oo
oo
oo
oo
oo
12
M Series Service Manual
Test SetupRemove red lead from power supply and connect to 0.1Ω resistor.
Connect other end of resistor to “+” terminal of power supply using a second red lead.
Connect multimeter across the resistor.
Set voltage scale (if DVM is not autoranging) to 220 mV.
Red
+
Battery Well
_
Black
Do this...Observe this...Pass/Fail/NA
_
+
+
_
DMM
15 Amp
Supply
4.7System Current Test
Set power supply to 10.3V.
4.8Turn the selector switch to
MONITOR.
(For AED units, turn the
selector switch to ON and
select Manual mode.)
4.9Turn unit off.
Record your results on the Maintenance Tests Checklist.
Voltage across resistor should be 80 mV or less (<800 mA of ON
current). NOTE: Without optional parameters.
a)With green screen or LCD and no options <80mV
b)With yellow screen and no options <81mV
c)With yellow screen and SpO
d)With yellow screen and voice recording <91mV
e)With yellow screen, voice recording and SpO
f)All devices with EtCO
13
<121mV
2
<104mV
2
<114mV
2
ooo
ooo
ooo
ooo
ooo
ooo
ooo
oo
M Series Service Manual
Test Setup for Off Current
Test
Remove 0.1Ω resistor and replace with 1KΩ.
Connect DMM across resistor.
Set voltage scale to DCV.
Measure voltage across resistor.
Do this...Observe this...Pass/Fail
4.10Off Current Test
Measure across resistor
with unit turned off.
Record your results on the Maintenance Tests Checklist.
Voltage should be less than 450 mV (<450
µA of current).
oo
14
M Series Service Manual
5.0 Leakage Current Test
Tools NeededSee the manufacturer’s instructions or supplied specifications for the leakage tester you use.
SetupSee the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with
accessories: MFC, external paddles, internal paddles, and anterior/posterior paddles.
Maximum Leakage Acceptance Limits
Normal ConditionSingle Fault Condition*
ECG10µΑ50µΑ
MFC100µΑ100µΑ
Earth500µΑ1000µΑ
*Single fault considered AC mains on applied part.
15
M Series Service Manual
6.0 Paddles Test
Tools NeededNone.
Test SetupIf applicable, connect the universal cable to the paddles. Place paddles in paddle wells.
Do this...Observe this...Pass/Fail/NA
6.1
6.2Press and release the
6.3Press and release the
6.4Select 30J using the paddle
6.5
6.6
6.7
Turn the selector switch to
DEFIB. Press and hold the
ENERGY DOWN button on the sternumpaddle.
The recorder turns on. Press and release again to turn off.
The unit charges to 30J, then the red LED charge indicator
illuminates and the charge tone sounds. (Note that the front panel
shock button does not illuminate).
No discharge.
No discharge.
The unit discharges. The TEST OK message displays and the red
LED turns off. The recorder runs.
oo
oo
ooo
oo
oo
oo
oo
Record your results on the Maintenance Tests Checklist.
16
M Series Service Manual
7.0 Heart Rate Display Test
Tools NeededCalibrated ECG simulator with 60Hz sine wave output capability.
Mini-phone plug for measuring output signal from 1 Volt ECG OUT jack (optional).
ECG Cable (3 or 5 leads).
Test SetupTurn the selector switch to MONITOR. Press LEAD button until “I” displays.
Connect the ECG leads to the DYNATECH Impulse 4000 or equivalen t .
Connect the ECG cable to the unit.
Do this...Observe this...Pass/Fail/NA
7.1Set the ECG Simulator to
120BPM.
Record your results on the Maintenance Tests Checklist.
The Heart Rate displays as 120 +/- 2 bpm
ooo
17
M Series Service Manual
8.0 Calibrating Pulses on Strip Chart Test
Tools NeededNone.
Test SetupNone.
Do this...Observe this...Pass/Fail/NA
8.1
8.2Press and hold SIZE button
Record your results on the Maintenance Tests Checklist.
Press the RECORDER
button.
to activate the calibration
signal.
The strip chart displays a signal of 300 ppm with an amplitude of
10 mm +/- 1 mm. The signal also appears on the video display.
Test SetupConnect the ECG cable to the DYNATECH Impulse 4000 or equivalent.
Connect the ECG cable to the unit
Do this...Observe this...Pass/Fail/NA
9.1
9.2
9.3Press RECORDER button.Verify that the waveform amplitude on the strip chart is less than
9.4Turn the ECG simulator off.
Record your results on the Maintenance Tests Checklist.
Turn the selector switch to MONITOR mode.
(For AED units, turn the
selector switch to ON and
select Manual mode.)
Select lead I, size 3x.
Select 60Hz (or 50 Hz for a
50Hz unit) on the Dynatech
Impulse 4000.
.
1.5 mm.
ooo
19
M Series Service Manual
10.0 Heart Rate Alarm Test
Tools NeededDynatech Impulse 4000.
Do this...Observe this...Pass Fail/NA
10.1
10.2Press ALARMS.
10.3Press SELECT PARAM softkey
10.4Press INC> for state.
10.5Press DEC>for state.
10.6
10.7
10.8
Turn the selector switch to MONITOR mode.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Connect the ECG leads to the
Dynatech Impulse 4000. Set the
simulator to 120 BPM and the
defibrillator to lead II.
until ECG HR displays.
Press INC> until ENABLE displays.ENABLE displays.
Set LOW limit to 30, HIGH limit to
150 then, press the RETURN
softkey.
Press ALARM SUSPEND button.No alarm sounds.
Lead II message displays.
NSR ECG at 120 BPM +/- 2 displayed.
The alarm menu displays.
Cursor scrolls through parameters.
Cursor scrolls through ENABLE, AUTO and
DISABLE.
Cursor scrolls through ENABLE, DISABLE, AND
AUTO.
MONITOR displays.
ooo
ooo
ooo
ooo
ooo
ooo
ooo
ooo
10.9Remove a lead wire from the
Dynatech Impulse 4000.
The bell symbol flashes and the heart symbol stops
flashing. The ECG LEAD OFF alarm tone sounds.
Recorder prints a stripchart showing a low heart rate, if
enabled.
20
ooo
Loading...
+ 108 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.