ZOLL shall not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this material.
User Guide
applies to the ZOLL AutoPulse Resuscitation System Model 100.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form
or by any means, mechanical, electronic, photocopying, recording, or otherwise, without prior written
permission of ZOLL.
AutoPulse and LifeBand® are trademarks of ZOLL. All other trademarks mentioned herein belong to
their respective owners.
0344
USA
ZOLL Circulation
650 Almanor Avenue
Sunnyvale, CA 94085 USA
t: +1.800.348.9011
(or: +1.978.421.9655)
Page iiP/N 11440-001 Rev. 3
EU Authorized Representative
ZOLL International Holding B.V.
Newtonweg 18
6662 PV ELST
The Netherlands
Table B-1 AutoPulse Parts and Accessories .......................................................................B-1
Page viiP/N 11440-001 Rev. 3
User Guide
[This page left intentionally blank.]
Page viiiP/N 11440-001 Rev. 3
User Guide
Preface
This document describes the operating steps and maintenance requirements for the AutoPulse®
Resuscitation System Model 100 (also known as the
Proper use of the AutoPulse requires a thorough understanding of the product, appropriate training and
practice.
AutoPulse
).
Please read the entire
User Guide
before operating the AutoPulse.
Who Should Read this Guide0
This document should be read by personnel who will use this product and who are trained in Basic Life
Support (BLS) and/or Advanced Life Support (ALS) techniques. This includes emergency medical
technicians, paramedics, nurses, physicians, police, and fire rescue personnel, and people certified to
administer cardiopulmonary resuscitation (CPR).
General Warnings and Precautions0
Warning:
•
The AutoPulse is intended for use on adults, 18 years of age or older.
The AutoPulse is
•
sudden physical injury or violence).
When CPR is indicated, it should start immediately and should not be postponed.
•
•
The AutoPulse must be used
Personnel certified in manual CPR must always be present during the AutoPulse System
operation.
not
intended for patients with traumatic injury (wounds resulting from
only
in cases that manual CPR would normally be initiated.
The AutoPulse Platform is
•
AutoPulse Platform should be secured to the top of a backboard or other equipment used to
carry or transport the patient, if necessary. During transport, regular checks of the patient’s
alignment should be performed.
Failure to properly position the LifeBand® at the patient’s armpit line may cause injury to
•
the patient.
•
Do not place or position the patient on the AutoPulse Platform in either a facedown
orientation or on the patient’s side.
•
If a system error occurs, immediately revert to manual CPR.
P/N 11440-001 Rev. 3Page ix
not
intended to be the sole means of carrying a patient. The
Warning:
•
•
•
•
•
•
User Guide
If a user advisory or fault cannot be cleared or a system error occurs during active operation,
immediately revert to manual CPR.
Do not strap across, or otherwise constrain, the LifeBand. Constraining the movement of the
LifeBand can damage or break the LifeBand.
Do not touch the patient while the AutoPulse is analyzing the patient’s size.
Failure to properly position a patient, both vertically and laterally with respect to the
AutoPulse Platform, may cause injury to the patient.
If you must move or realign the patient, you must press the Stop/Cancel button before
adjustment.
Do not place your hands or any other objects on or under the LifeBand while the AutoPulse
is analyzing the patient or during active operation.
•
•
Caution:
Caution:
Caution:
Caution:
Caution:
Caution:
Caution:
Caution:
Caution:
Operating the AutoPulse on a patient for extended periods of time may result in minor skin
irritation to the patient.
Check the patient’s chest rise during ventilation during active operation.
United States federal law restricts this device to sale by or on the order of a licensed physician.
The AutoPulse is designed to be used only with ZOLL-approved accessories. The AutoPulse
will perform improperly if non-approved accessories are used.
Only use ZOLL Batteries specifically designed for use with the AutoPulse. The use of other
batteries may cause permanent damage to the AutoPulse and will void the warranty.
Make sure that the LifeBand is
Do
not
submerge the AutoPulse Platform in liquid.
not
twisted before automatic compressions begin.
Use care while using sharp instruments around the LifeBand.
Do not block the vents of the AutoPulse.
Do not use the AutoPulse alone as a patient transportation aid.
Straps or restraints used for transportation purposes
must not interfere
with the operation of
the AutoPulse. Specifically, straps across the patient’s chest may restrict the compression/
decompression of the chest. In general, strapping schemes must not alter the alignment of the
patient to the AutoPulse.
Page xP/N 11440-001 Rev. 3
User Guide
Caution:
Caution:
Motion can cause the patient to shift and restraints to loosen, so care should be given to the
initial strapping for alignment of the patient to the AutoPulse. Regular checks of patient
alignment to the AutoPulse and alignment of the LifeBand to the patient's mid-axillary line
should be made if the AutoPulse is performing active compressions, or before active
compressions are restarted.
Remove the protective plastic cap from the Battery before attempting to charge the Battery.
Symbols0
The symbols below may be found in this
Attention: Consult Accompanying Documents
User Guide
, on the AutoPulse Platform, or on the LifeBand.
!
Date of Manufacture
Manufacturer
Authorized Representative
SN
Serial Number
Defibrillation Protected, Type BF Patient Connection
Degree of Protection Provided by Enclosure Per IEC 60529
Do Not Reuse—Single Use Only
Charging
Ready
Fail
P/N 11440-001 Rev. 3Page xi
Start Test
Power
User Guide
Page xiiP/N 11440-001 Rev. 3
User Guide
1Introduction of the AutoPulse
For years, a variety of attempts have been made to overcome the limitations of manual cardiopulmonary
resuscitation (CPR). ZOLL has developed a practical solution as an adjunct to the method used by
medical professionals to perform CPR chest compressions. The AutoPulse is the resulting product.
®
1.1Indication for Use
The AutoPulse is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of
clinical death as defined by a lack of spontaneous breathing and pulse.
1.2Description of the System
The AutoPulse is an automated, portable, battery-powered chest compressor, which provides chest
compressions as an adjunct to performing manual CPR (see Figure 1-1). Use of the AutoPulse is intended
to reduce the impact of rescuer fatigue and will enable the rescuer to address additional patient needs.
The AutoPulse has the following operating parameters:
Chest displacement: equal to 20% reduction in anterior-posterior chest depth.
•
•
Consistent compression rate and depth.
•
Physiological duty cycle: fixed at 50 ± 5%.
Standardized 30:2, or 15:2 compressions (30 or 15 compressions followed by two consecutive 1.5
•
second ventilation pauses) or continuous compressions (user selectable).
Table 1-1 provides the patient/AutoPulse operating parameters.
Table 1-1Patient/AutoPulse Operating Parameters
Patient ParameterAutoPulse Specification
Patient chest circumference permitted29.9 to 51.2 in. (76 to 130 cm)
Patient chest width permitted9.8 to 15 in. (25 to 38 cm)
Maximum patient weight permitted300 lbs.(136 kg)
P/N 11440-001 Rev. 3Page 1-1
AutoPulse Platform
LifeBand
Figure 1-1AutoPulse
User Guide
Battery Charger User Control Panel
Batteries
(Primary and Spare)
1.3System Components
Figure 1-1 shows the main components of the AutoPulse.
The AutoPulse consists of the following:
•
AutoPulse Platform
LifeBand
•
•
AutoPulse Battery
•
AutoPulse Battery Charger
1.3.1AutoPulse Platform
The AutoPulse Platform contains the mechanical drive mechanism, control system, and electronics
necessary to generate and control the force required to perform mechanical chest compressions. User
controls and indicators are contained in the User Control Panel.
Figure 1-2 shows the patient surface (front) and back surface details of the AutoPulse Platform. The
AutoPulse Platform features carry handles to facilitate transporting it to the scene of the arrest.
Page 1-2P/N 11440-001 Rev. 3
User Guide
ON/OFF Button
Battery
Patient Alignment
Reference
User Control Panel
Start/Continue Button
Stop/Cancel Button
Carry Handle
Cooling Vent
AutoPulse
Slotted Drive Shaft
Back Surface
Cooling Vent
Patient Surface
Figure 1-2AutoPulse Platform (Patient and Back Surfaces)
1.3.2LifeBand Load-distributing Band (LDB)
LifeBand
Roller Guide
®
The LifeBand is a load-distributing band (LDB) that consists of a cover plate and two bands integrated
®
with a compression pad with a Velcro
fastener. Attached to the AutoPulse Platform, the LifeBand is
automatically adjusted to the patient and provides compressions to the patient's chest in the region of the
heart. The latex-free LifeBand is a single-use component that is attached to the AutoPulse Platform
before each use.
1.3.3AutoPulse Power System Battery
The AutoPulse Battery is a removable component that supplies power for the AutoPulse operation. The
Battery is a proprietary, rechargeable, nickel-metal hydride (NiMH) battery that is the exclusive power
source for the AutoPulse.
The Battery is mechanically keyed to the AutoPulse Platform and Battery Charger to facilitate correct
installation. The Battery’s back end contains connections for power and communications to the Battery
Charger and to the AutoPulse Platform. A Battery Status Check button illuminates the Battery’s status
light-emitting diodes (LEDs).
For more information on the Battery, refer to the
AutoPulse Power System User Guide
.
P/N 11440-001 Rev. 3Page 1-3
User Guide
1.3.4AutoPulse Power System Battery Charger
The AutoPulse Battery Charger is a stand-alone unit intended to charge and test-cycle the Batteries. The
Battery Charger has two charging bays; each with its own indicators. The Battery Charger automatically
maintains both the state of charge and tests and maintains the Battery to its highest possible capacity.
Batteries should always be fully charged and ready for use before deploying the AutoPulse.
For more information on the Battery Charger, refer to the AutoPulse Power System User Guide.
1.4User Controls and Indicators
1.4.1ON/OFF Button
The ON/OFF button is located adjacent to the Battery on the AutoPulse Platform (see Figure 1-3).
Pressing the ON/OFF button once powers up the AutoPulse Platform and initiates a self-test. The User
Control Panel’s green Power LED lights up. Pressing the ON/OFF button again powers down the
AutoPulse Platform.
User Control Panel
ON/OFF button
Figure 1-3ON/OFF Button Location
1.4.2User Controls
All user controls and indicators (except the ON/OFF button) are contained in User Control Panel (see
Figure 1-4).
Page 1-4P/N 11440-001 Rev. 3
Power
User Guide
Alert LED (red)
Display Panel
LED (green)
Increase Contrast
Decrease Contrast
Tone Mute
Stop/Cancel LED
Stop/Cancel
Figure 1-4User Control Panel
1.4.2.1Start/Continue Button
The green Start/Continue button is active when “Start” or “Continue” appears on
the display panel above the button and its green LED is illuminated.
Use the Start/Continue button to start or continue:
•Analyzing patient size
Move Up
Select Choice
Move Down
Start/Continue
Menu/Mode Switch
Start/Continue LED
•Chest compressions
1.4.2.2Stop/Cancel Button
The orange Stop/Cancel button is active when “Stop”, “Quit”, “No Realign” or
“Cancel” appears on the display panel above the button and its orange LED is
illuminated.
Use the Stop/Cancel button to stop or cancel:
•Analyzing patient size (see Figure 3-9)
•Chest compressions (the AutoPulse releases the tension on the LifeBand)
(see Figure 3-11)
•Verify patient alignment pause (see Figure 3-10)
P/N 11440-001 Rev. 3Page 1-5
1.4.2.3Menu/Mode Switch Button
On initial power-up the gray Menu/Mode Switch button functions as the Menu
button. Pressing this button allows you to:
1.Enter the communication mode
2.View last patient session information
3.View AutoPulse Platform information
4.View AutoPulse Battery information
For a complete description of the available information and how to access it,
refer to section Section 3.7, “Viewing AutoPulse Platform Information,” on page
3-19. For more information on the Communication Mode, see Section 3.8,
“Uploading AutoPulse Information to your PC,” on page 3-20.
While the AutoPulse is actively doing compressions this button may function as
the Mode Switch button. The Mode Switching feature is only active when the
“30:2 or Continuous” or “15:2 or Continuous” option has been set in Mode
section of the Administration Set-up (refer to section Section 2.3,
“Administrative Menu: User Pre-set Options,” on page 2-10). If the “30:2 or
Continuous” or “15:2 or Continuous” option has been set, the Mode Switch
button will allow you to switch on-the-fly between 30:2 and Continuous or 15:2
and Continuous compression modes respectively.
User Guide
The current mode (either “30:2”, “15:2” or “CONTINUOUS”) is displayed on
the upper left portion of the User Display.
1.4.2.4Move Up/Move Down Button
These buttons allow you to highlight for selection different menu or list items.
Pressing the Move Up button (upward pointing triangle) moves up to the next
menu item.
Pressing the Move Down button (downward pointing triangle) moves down to
the next menu item.
Page 1-6P/N 11440-001 Rev. 3
User Guide
1.4.2.5Select Choice Button
Pressing the Select Choice button selects the currently highlighted menu or list
item.
1.4.2.6Tone Mute Button
c
The ventilation and pause tones are always generated at the appropriate times by
the system and cannot be turned off, but they can be muted for a short period of
time as set in the Administration Menu (refer to section Section 2.3,
“Administrative Menu: User Pre-set Options,” on page 2-10). Selecting the Tone
Mute button will mute or enable the audible tone feedback generated by the
AutoPulse. When the audio feedback is audible, pressing the Tone Mute button
will mute it. When the audio feedback is muted, pressing the Tone Mute button
will make it audible once again.
Once audio feedback has been muted with the Tone Mute button, it will
automatically be reactivated when one of the following occurs:
1.The Mute Duration time set in the Administrative Menu has expired.
2.You press the Tone Mute button again.
3.You press the Start/Continue button to begin compressions.
4.You press the Stop button during compressions.
5.A low battery condition is reached (refer to section Section 1.4.3, “Low
Battery Warning,” on page 1-9).
6.You switch between compression modes (refer to section Section 3.2,
“Mode Change Confirm Display Panel Screen,” on page 3-13).
The icon displayed on the User Control Panel display when tones are audible.
The icon displayed on the User Control Panel display when tones are muted.
P/N 11440-001 Rev. 3Page 1-7
1.4.2.7Increase/Decrease Contrast Button
Pressing the Increase Contrast button (plus sign) increases the contrast of the
display panel screen. Each key press increases the contrast of the display panel
screen by one level.
There are a total of eight contrast levels.
Pressing the Decrease Contrast button (minus sign) decreases the contrast of the
display panel screen. Each key press decreases the contrast of the display panel
screen by one level.
You can adjust the contrast of the display panel screen at any time the AutoPulse
Platform is powered up.
User Guide
Page 1-8P/N 11440-001 Rev. 3
User Guide
1.4.2.8Power (Green LED)
The green Power LED lights whenever the AutoPulse is powered on and able to respond to user input.
1.4.2.9Alert (Red LED)
The red Alert LED lights whenever a user advisory, fault or system error condition exists for the
AutoPulse. For a list of advisory, fault and error conditions, refer to Chapter 5, “Troubleshooting
Procedures”.
1.4.3Battery Charge Status
The User Control Panel displays the battery charge status. The Battery charge status icon only appears
when the AutoPulse is powered up.
Indicates the level of charge of the Battery. A graphic battery icon indicating four
proportional levels of battery charge is displayed.
30:2
LOW BATTERY
Figure 1-5Low Battery Warning
When at least five minutes of active operation remain on a Battery, the User Control Panel will give a
“Low Battery” indication (see Figure 1-5). The “Low Battery” indication will remain on until the Battery
is replaced or depleted. The Low Battery warning display will be accompanied with an audio warning of
four rapid beeps which will be followed by two beeps every 30 seconds until the battery is replaced or
depleted. It is recommended that a fully-charged Battery be exchanged for the Battery with the low
charge (if available).
P/N 11440-001 Rev. 3Page 1-9
User Guide
To exchange Batteries:
1.Press the Stop/Cancel button.
2.Press the ON/OFF button.
3.Remove the Battery (refer to Section 2.2, “Battery Installation and Removal” for more information).
4.Install the fully-charged Battery (refer to Section 2.2, “Battery Installation and Removal” for more
information).
5.Resume chest compressions (refer to Section 3.2, “Starting Chest Compressions” for more
information).
Table 1-2 describes the battery charge status indicator specifics. The Battery, when fully charged, has the
capacity for a minimum run time of 30 minutes on a patient of nominal size and nominal resistance to
compressions (a nominal patient). This is referred to in Table 1-2 as the nominal run time.
Table 1-2Battery Charge Status Indicator Specifics
Battery Charge Icon Bars ShowingCharge Level
No bars showing.The Battery has been depleted. Replace the Battery
immediately.
One bar showing.The Battery’s capacity is less than one-third of its nominal
run time. Be prepared to exchange this Battery with a fullycharged Battery.
Two bars showing.The Battery’s capacity is between 33% and 66% of its
nominal run time.
Three bars showing. The Battery’s capacity is between 66% and 100% of its
nominal run time.
All bars showing.The Battery’s capacity is at least 100% of its nominal run
time.
Page 1-10P/N 11440-001 Rev. 3
User Guide
1.4.4Performance Characteristics
The basic operating characteristics of the AutoPulse are shown in Table 1-3.
Table 1-3Operating Characteristics
Operating PerformanceSpecification
Compression rate80 (± 5 compressions per minute)
Compression modes (user selectable)
•30:2 (30 compressions with two 1.5 second ventilation pauses)
•Continuous compressions
Duty cycle50 (± 5%)
Compression depth20% of chest depth, +0.25/-0.5 inch
P/N 11440-001 Rev. 3Page 1-11
User Guide
[This page left intentionally blank.]
Page 1-12P/N 11440-001 Rev. 3
User Guide
2Preparing the AutoPulse for Use
The AutoPulse System is delivered fully assembled, except for the LifeBand and the AutoPulse Battery.
2.1LifeBand Load-distributing Band (LDB)
Note:Do not cut the LifeBand before removing it from the AutoPulse. Cutting the LifeBand may
cause the AutoPulse to report a Fault and will require specific steps to clear that Fault.
2.1.1Installing the LifeBand
1.Power off the device.
2.Place the AutoPulse Platform with the patient surface facing down on a smooth, flat surface.
Note:The driveshaft should be oriented so that the slot faces directly upward.
3.Insert the head end of the LifeBand band clip into the driveshaft slot. The correct direction is
1
towards the LifeBand plate alignment arrow seen on the Platform (see Figure 2-1).
LifeBand plate alignment arrow
Head End
Cover Plate
Slotted Drive Shaft
Guide Plate
1
2
Band Clip
Tail End
Figure 2-1Sliding the LifeBand Band Clip into the Driveshaft Slot
4.Once the head end of the LifeBand band clip is positioned into the slot, press the tail end of
2
the band clip into the slot of the guide plate until the band clip is fully seated in the driveshaft. You
should feel it lock into place.
P/N 11440-001 Rev. 3Page 2-1
User Guide
5.Make sure that the band clip is seated properly and fully into the slot on the driveshaft (see Figure
2-2).
Note:If the band clip is properly seated, you should be able to turn the driveshaft in each
direction by hand.
Figure 2-2Seating the LifeBand Band Clip Properly into the Driveshaft Slot
6.Ensure that both free ends of the LifeBand are oriented flat (not twisted) and away from the
AutoPulse Platform. Inspect the LifeBand for any cuts or tears. Do not use the LifeBand if cuts or
tears are present.
7.Line up the arrow on the LifeBand cover plate with the matching arrow on the Platform.
Page 2-2P/N 11440-001 Rev. 3
User Guide
8.Snap the LifeBand cover plate in place by fully inserting the locking tabs into the slots on the
AutoPulse (see Figure 2-3).
LifeBand Cover Plate
Alignment Arrows
Figure 2-3Snapping the LifeBand Cover Plate into Place
9.Flip down and snap into place the hinged belt guards of the LifeBand cover plate to engage the
LifeBand chest bands to the rollers (see Figure 2-4).
Belt Guard
Figure 2-4Flip Down the Hinged Belt Guards of the LifeBand
P/N 11440-001 Rev. 3Page 2-3
User Guide
10. Turn the AutoPulse over and press the ON/OFF button to power it up. If the User Control Panel
reports a user advisory, check the installation of the LifeBand band clip into the slot in the
driveshaft.
2.1.2Removing the LifeBand
Note:Please do NOT cut the LifeBand before removing it from the AutoPulse. Cutting the LifeBand
may cause the AutoPulse to report a Fault and will require specific steps to clear that Fault.
1.Place the AutoPulse Platform with the patient surface facing down on a smooth, flat surface.
2.Flip up the hinged belt guards of the LifeBand cover plate to disengage the LifeBand from the rollers
(see Figure 2-5). You will hear the guards “snap.” This is normal.
LifeBand Cover Plate
Belt Guard
Figure 2-5Flip Up the Hinged Belt Guards of the LifeBand
Page 2-4P/N 11440-001 Rev. 3
User Guide
3.Using both hands, pinch together the locking tabs of the LifeBand cover plate and firmly pull the
plate straight up and away from the AutoPulse (see Figure 2-6).
Figure 2-6Pinching the Locking Tabs of the LifeBand
P/N 11440-001 Rev. 3Page 2-5
2
2
User Guide
Guide Plate
1
Figure 2-7Removing the LifeBand from the AutoPulse Platform.
Warning: Removing the band clip when the driveshaft is not at its home position (see Figure 2-10)
will result in a permanent User Advisory (45) that the user will not be able to clear. This
may be the case if the LifeBand has been cut. Please refer to Section 2.1.2.1, “Removing a
LifeBand that is Cut or Not in the Home Position” for additional specific instructions before
removing it. Do not continue to remove the LifeBand using the following procedure
without first consulting Section 2.1.2.1.
4.Remove the LifeBand from the AutoPulse Platform by grasping the LifeBand chest band with the
thumb and index finger of both hands, on either side of the LifeBand band clip. Push in the
guide plate, using both middle fingers. Keeping the guide plate pushed in, pull up the band to
remove the clip from the shaft (see Figure 2-7).
1
5.Discard the LifeBand as it is a single-use component after patient use. Treat the LifeBand as
contaminated medical waste and dispose of it accordingly. There are no user-serviceable
parts.
Page 2-6P/N 11440-001 Rev. 3
User Guide
6.Inspect the new, replacement LifeBand for cuts or tears.
7.Install the new LifeBand following the procedures in Section 2.1.1, “Installing the LifeBand”.
2.1.2.1Removing a LifeBand that is Cut or Not in the Home Position
The chest band must be completely unwound from the driveshaft before it is removed. When the chest
band is completely unwound the seam is visible and the driveshaft rests in the Home Position (see Figure
2-10). Until a LifeBand is installed a shaft lock mechanism secures the driveshaft in the Home Position:
do not attempt to defeat the shaft lock mechanism.
Figure 2-8LifeBand NOT in the Home Position: Do Not Remove!
Removing the band clip when the driveshaft is not at its home position will result in a permanent User
Advisory (45) that the user will not be able to clear. The LifeBand should be removed from the driveshaft
ONLY from its home position.
If the chest bands have been cut it is quite possible that the chest band is still wound onto the driveshaft.
Care should be taken to ensure that the bands are fully extended before the cover plate is opened and the
band clip is removed.
P/N 11440-001 Rev. 3Page 2-7
User Guide
If the chest band is not completely unwound from the driveshaft follow this procedure
Figure 2-9Positioning the LifeBand for Removal
1.Once the cover plate is removed, take hold of the chest band on either side of the driveshaft and pull
both bands outward.
LifeBand Seam
Figure 2-10 LifeBand in the Home Position: Ready for Removal
2.The chest bands should now be fully extended. The seam securing the band clip should easily be
seen and the driveshaft is centered. The LifeBand may now be removed continuing with the
procedure in Section 2.1.2 on page 2-6. Do not remove the LifeBand if it is not in this position.
Page 2-8P/N 11440-001 Rev. 3
User Guide
2.2Battery Installation and Removal
For more information on the AutoPulse Battery and Battery Charger refer to the AutoPulse Power System
User Guide.
Remove a charged Battery from the Battery Charger. Ensure the Battery is fully charged before insertion
into the AutoPulse Platform (refer to the “Performing a Battery Status Check” section in the AutoPulse Power System User Guide for more information).
Caution: Remove the protective plastic cap from the Battery connector before attempting to charge the
Battery.
The Battery remaining in the AutoPulse Platform should be replaced with a charged Battery from the
Battery Charger.
Finger Latch
Battery Latch Opened
Figure 2-11 Battery Installation and Removal
Locking Bar
Battery Latch
Battery Front
Battery Latch Closed
To install the Battery, first make sure the Battery Latch is rotated out of the way, so that the Battery can
slide into the Battery compartment (see Figure 2-11). Then slide the Battery into the Battery
compartment in the AutoPulse Platform. Push the Battery most of the way into place, then advance the
Battery the rest of the way with your thumbs. The Battery should snap into place and mount flush with
the AutoPulse Platform. Then rotate the Battery Latch into position to ensure the Battery is retained.
P/N 11440-001 Rev. 3Page 2-9
User Guide
The Battery is mechanically keyed so that it can only be inserted in one orientation. Do not force or slam
the Battery into the Battery compartment. If resistance is met, check for appropriate orientation, and
check to insure there are no obstructions to Battery insertion and that the Battery Latch is open.
Ensure that the Battery is securely latched (snaps into place) before moving the AutoPulse Platform or
initiating chest compression.
To remove the Battery, first rotate the Battery Latch into the open position as shown in Figure 2-11. Then
hold the AutoPulse Platform firmly and grip the Battery while pressing the finger latch upwards to
disengage the locking bar (see Figure 2-11), then pull the Battery straight out until it fully clears the
Battery compartment.
Caution: Only use ZOLL Batteries specifically designed for use with the AutoPulse. The use of other
batteries may cause permanent damage to the AutoPulse and will void the warranty.
2.3Administrative Menu: User Pre-set Options
There are several options that may be pre-set by the user prior to deployment of the AutoPulse. These
options are:
•Compression mode
•Mute duration
•Tone volume
Additionally the Administrative Menu allows you to access the following information (refer to section
Section 3.7, “Viewing AutoPulse Platform Information,” on page 3-19):
•The last patient session
•The AutoPulse Platform
•The AutoPulse Battery
Page 2-10P/N 11440-001 Rev. 3
User Guide
S e t M o d e ( 3 0 : 2 o r C o n t )
30:2
ADMIN MENU
Set Mode (30:2 or Cont)
Set Mute Duration (30s)
Set Tone Volume (High)
View Last Patient Info
RESTART
Figure 2-12 Administrative Menu
To access the Administrative Menu the AutoPulse must be powered off. The Administrative Menu is
activated by pressing the ON/OFF switch while both the Stop (orange) and Start (green) buttons are
being depressed. Once the Administrative Menu is active use the Move Up and Down arrow buttons to
highlight the desired menu item and the Select Choice Button to select it.
To exit from the Administrative Menu, press the “START” (green) button under the word “RESTART.”
The AutoPulse will restart and place you into the idle state, ready for patient alignment or for system
power-down.
Note:The current setting is displayed in the parenthesis after the main menu item.
P/N 11440-001 Rev. 3Page 2-11
User Guide
3 0 : 2 o r C o n t i n u o u s
3 0:2
MODE
15: 2 or Continuous
15: 2 Onl
3 0 : 2 or Continuous
3 0 : 2 Onl
Continuous Onl
y
y
y
b a ck
Figure 2-13 Compression Mode Menu
The “Set Mode” menu item allows you to restrict the AutoPulse Platform operation to a single
compression mode or to allow on-the-fly mode switching. Selecting “30:2 or Continuous” or “15:2 or
Continuous” will allow on-the-fly mode switching between 30:2 and continuous compressions or 15:2
and continuous compressions respectively, while the system is actively doing compressions. Selecting
“30:2 Only” or “15:2 Only” will restrict the system operation to the 30:2 mode or 15:2 mode,
respectively. Selecting “Continuous Only” will restrict the system operation to continuous compressions.
Highlight the desired setting using the Move Up and Down arrow buttons, press the Select Choice Button
to select it (a check will appear beside the selected item) and then press the gray Menu/Mode switch
button under the word “back” to return to the main Administrative Menu.
Page 2-12P/N 11440-001 Rev. 3
User Guide
3 0 s e c o n d s
3 0:2
MUTE DURATION
3 0 s econds
60 s econds
120 s econds
Disabled
b a ck
Figure 2-14 Mute Duration Menu
The “Set Mute Duration” menu item allows you to set the length of time that the audio tone mute will be
sustained when it is activated by pressing the Tone Mute Button, or to disable the button’s function
altogether. The time options are to set the mute to last for “30 seconds”, “60 seconds” (one minute) or
“120 seconds” (two minutes). Highlight the desired setting using the Move Up and Down arrow buttons,
press the Select Choice Button to select it (a check will appear beside the selected item) and then press
the gray Menu/Mode switch button under the word “back” to return to the main Administrative Menu. If
the pause alert tone is disabled here it cannot be re-enabled, during operation, by using the Tone Mute
Button (refer to Section 1.4.2.6, “Tone Mute Button,” on page 1-7).
P/N 11440-001 Rev. 3Page 2-13
User Guide
H i
g
h
3 0:2
VENT/PAUSE TONE VOLUME
Hi
Low
b a ck
Figure 2-15 Ventilation/Pause Tone Volume Menu
The “Set Tone Volume” menu item allows you to select the volume of the audible tone sequence that is
used to cue for ventilation during active compressions and to alert the operator that the system has been
deliberately stopped (paused) while actively doing compressions. Choices are “High” and “Low.”
Highlight the desired setting using the Move Up and Down arrow buttons, press the Select Choice Button
to select it (a check will appear beside the selected item) and then press the gray Menu/Mode switch
button under the word “back” to return to the main Administrative Menu.
Page 2-14P/N 11440-001 Rev. 3
User Guide
3Using the AutoPulse
This chapter describes how to use the AutoPulse in an emergency situation. The AutoPulse User Control
Panel automatically provides display prompts to guide you.
Before deploying the AutoPulse, note the following warnings and precautions:
Warning:
•The AutoPulse is intended for use on adults, 18 years of age or older.
•The AutoPulse is not intended for patients with traumatic injury (wounds resulting from
sudden physical injury or violence).
•When cardiopulmonary resuscitation (CPR) is indicated, it should start immediately and
should not be postponed.
•The AutoPulse must be used only in cases that manual CPR would normally be initiated.
Personnel certified in manual CPR must always be present during the AutoPulse operation.
Caution: Use care while using sharp instruments around the LifeBand.
Caution: Do not block the vents of the AutoPulse Platform.
3.1Deploying the AutoPulse System
In order to deploy the AutoPulse quickly and with the least interruption in cardiac compressions, a pit
crew model - similar to that which is used in auto racing - is suggested for roles and positions of the staff
involved in performing defibrillation and using the AutoPulse. Your local ZOLL representative can
provide you with appropriate detailed instructions based upon the setting in which you work (EMS or
hospital) and the number of clinicians that are typically involved in dealing with sudden cardiac arrest.
Each organization should determine how this type of model can be integrated into the typical roles
performed by members of their resuscitation team. Practice as a team using this model will help to
streamline actions and ensure rapid, efficient deployment.
P/N 11440-001 Rev. 3Page 3-1
User Guide
1.Power up the AutoPulse. The ON/OFF button is located on the top (“head”) edge of the AutoPulse
Platform (see Figure 3-1).
User Control Panel
ON/OFF button
Figure 3-1ON/OFF Button Location
2.The AutoPulse illuminates the green Power light-emitting diode (LED) on the User Control Panel
and performs its self-tests (see Figure 3-2). Refer to the User Control Panel and its display panel
during the operation of the AutoPulse. All operating information is available on the User Control
Panel.
Note:Make sure that no User Advisory, Fault or System Error messages display.
3 0:2
Initia lizing...
Figure 3-2Self-Test Display Panel Screen
Page 3-2P/N 11440-001 Rev. 3
User Guide
3.The AutoPulse indicates that it is ready for use (see Figure 3-3).
3 0:2
Align pa tient on pl a tform,
clos e LifeBa nd, a nd then
p
ress CONTINUE
menuCONTINUE
Figure 3-3Patient-Readiness Display Panel Screen
P/N 11440-001 Rev. 3Page 3-3
User Guide
4.After assessing the patient’s condition, sit the patient up and make a single cut down the back of the
patient’s clothing (see Figure 3-4), or undo any ties on the back of the gown in a hospital. The
posterior defibrillation/pacing pad may be placed on the patient's back at this time if the local
protocol calls for anterior-posterior placement. Use of standard pads in both the a-p or anterioranterior/apex-sternum placement is acceptable and will not affect the operation of the AutoPulse or
defibrillator.
5.Slide the AutoPulse Platform into position behind the sitting patient and lay the patient down onto
the Platform. Placing the AutoPulse to the patient's side and "log rolling" him or her onto the
Platform is an acceptable alternative.
Figure 3-4Cutting Patient Clothing and Positioning of AutoPulse
Page 3-4P/N 11440-001 Rev. 3
User Guide
6.Grasp the clothing by the sleeves and pull down toward the ankles to remove all of the clothing from
both the front and back of the torso. (see Figure 3-5). The anterior pad(s) may be placed at this time.
Figure 3-5Removal of all Clothing from the Torso
P/N 11440-001 Rev. 3Page 3-5
User Guide
7.Position the patient so that he/she is centered laterally (from left to right) and that the armpits are
aligned with the AutoPulse using the yellow line positioning guides on the platform as shown in
Figure 3-6.
Warning: Do not place or position the patient on the AutoPulse Platform in either a facedown
orientation or on the patient's side. Also ensure the head orientation is correct. The head
label arrow points towards the patient's head.
Center the patient on the
AutoPulse Platform.
ON/
OFF
Line up armpits to
the yellow Patient
Alignment Reference.
Figure 3-6Patient Alignment
Page 3-6P/N 11440-001 Rev. 3
User Guide
Band 1
LifeBand
Alignment Tab
ON
/
OFF
Band 2
LifeBand®
Alignment Slot
Figure 3-7Aligning the LifeBand
8.Close the LifeBand around the patient's chest.
Note:AED or defibrillation pads are not components of the LifeBand.
To properly align the two sides of the LifeBand:
a)Place band
b)Locate mating slot of band
➊ on top of patient’s chest (see Figure 3-7).
➋ over the alignment tab ➊ (see Figure 3-7).
c)Press the bands together to engage and secure the Velcro
d)Lift up the LifeBand to its fullest, ensuring that the side bands are at a 90 degree angle to
the platform, that they are not twisted and that there are no obstructions.
e)Center the LifeBand on the patient's chest, placing it such that its center is over the area
upon which manual compressions are conducted.
Note: If the bands cannot be closed, use manual CPR.
®
fastener (see Figure 3-8).
P/N 11440-001 Rev. 3Page 3-7
User Guide
ON
/
OFF
Figure 3-8Fastening the LifeBand
3.2Starting Chest Compressions
1.Make sure that the yellow upper edge of the LifeBand is aligned with the patient's armpits, and is
directly over the yellow line on the AutoPulse Platform. Also make sure that there are no
obstructions, such as clothing, straps or equipment, with the bands.
Warning:
•Failure to properly position the LifeBand at the patient’s armpit line may cause injury to the
patient.
•Failure to properly position a patient, both vertically and laterally with respect to the
AutoPulse Platform, may cause injury to the patient.
•Do not strap across, or otherwise constrain, the LifeBand. Constraining the movement of the
bands can damage or break the LifeBand.
•Operating the AutoPulse on a patient for extended periods of time may result in minor skin
irritation to the patient.
Page 3-8P/N 11440-001 Rev. 3
User Guide
2.Press and release the Start/Continue button once. The AutoPulse automatically adjusts the bands to
the patient's chest (see Figure 3-9).
Warning: Do not touch the patient or the LifeBand while the AutoPulse Platform is analyzing the
3.The AutoPulse will pause for 3 seconds to allow you to verify that the patient is properly aligned
and that the LifeBand has taken up any slack in the bands (see Figure 3-10).
If the patient is not properly aligned, press the Stop/Cancel button, realign the patient, and begin
compressions again with step 1 on page 3-8.
Warning: If the Stop/Cancel button is not pressed within 3 seconds, compressions will automatically
begin. Press the Stop/Cancel button to immediately stop the compressions.
4.After the 3 second verify patient alignment pause is complete, compressions will automatically
begin. You may press the Start/Continue button to immediately initiate compressions ahead of that
time.
Warning:
•Do not lean on the patient after pressing the Start/Continue button.
•If you must move or realign the patient, you must press the Stop/Cancel button before
adjustment.
•Do not place your hands or any other objects on or under the LifeBand while the AutoPulse
is analyzing the patient or during active operation.
5.Depending on the Mode setting in Administrative Menu (refer to section Section 2.3,
“Administrative Menu: User Pre-set Options,” on page 2-10), the AutoPulse will perform 30:2, 15:2
or Continuous compressions. In 30:2 mode it performs 30 compressions and then pauses for three
seconds to permit the user to ventilate the patient before automatically resuming compressions (see
Figure 3-12). In 15:2 mode it performs 15 compressions and then pauses for three seconds to permit
the user to ventilate the patient before automatically resuming compressions (see Figure 3-12). In
Continuous mode it performs continuous compressions. If 30:2 on-the-fly mode switching has been
enabled (in the Mode setting within the Administrative Menu) then the AutoPulse will perform in
the mode (either 30:2 or Continuous) that was used last until powered down; on power up 30:2 will
be the initial selection. If 15:2 on-the-fly mode switching has been enabled (in the Mode setting
within the Administrative Menu) then the AutoPulse will perform in the mode (either 15:2 or
Continuous) that was used last until powered down; on power up 15:2 will be the initial selection.
In 30:2 mode three audio cue tones will sound prior to the ventilation pause: one during each of the
28th, 29th, and 30th compressions. In 15:2 mode the three audio cue tones will sound prior to the
ventilation pause during each of the 13th, 14th, and 15th compressions. In Continuous mode, an
audio cue tone for ventilation will sound 8 times per minute. The tones can be temporarily disabled
(and re-enabled) by pressing the Tone Mute Button (refer to Section 1.4.2.6, “Tone Mute Button,”
on page 1-7).
At the initiation of compressions, the counter at the right center of the display panel screen will be
set to 00:00 and will automatically begin recording the elapsed time until the Stop/Cancel button is
pressed. The format of the counter is minutes:seconds. When the Stop/Cancel button is pressed, the
counter will immediately reset to zero and begin recording the elapsed "no-flow" time. The counter
will reset to zero when chest compressions are started again.
6.If 30:2 on-the-fly mode switching has been enabled you may press the gray Menu/Mode switch
button to switch between 30:2 and continuous compressions. If 15:2 on-the-fly mode switching has
been enabled you may press the gray Menu/Mode switch button to switch between 15:2 and
continuous compressions. The current mode is displayed in the upper left corner of the screen. The
words above the gray Menu/Mode switch button indicate the alternate mode that the AutoPulse will
switch to. If there are no words above the gray button then on-the-fly mode switching is not enabled,
and the device will only operate in the current mode and pressing the gray button will have no effect.
Once the gray Menu/Mode switch button has been pressed you will be asked to confirm the mode
switch (see Figure 3-13) by pressing the gray Menu/Mode switch button twice in rapid secession.
A single tone will sound to confirm that the mode change has been accepted.
7.To access the patient or to pause the AutoPulse for any reason, press the Stop/Cancel button. The
AutoPulse Platform releases the tension on the LifeBand, allowing the user to pull the bands to the
maximum extended position. 10 seconds after the Stop/Cancel button has been pressed a single
audio alert tone will sound. Three audio alert tones will sound 20 seconds after the pause was
initiated. Audio alert tones will sound continuously after 30 seconds into the pause. The tones can
be temporarily disabled (and re-enabled) by pressing the Tone Mute button (refer to section Section
1.4.2.6, “Tone Mute Button,” on page 1-7) if this function is allowed in the Administrative Menu.
Pressing the Stop/Cancel button while paused will exit the paused state and the alert tone will be
stopped.
Note:Opening the bands during active operation will cause the AutoPulse to stop operation
®
immediately. To restart compressions, re-fasten the Velcro
fastener, clear the Fault by
pulling up on the LifeBand and pressing Start/Continue and then follow the normal
operating steps beginning with step 1 on page 3-8.
8.To restart compressions, press the CONTINUE button as described in the procedure starting at step
1 on page 3-8 (see Figure 3-15).
Warning: If a User Advisory or Fault cannot be cleared, or a System Error occurs during active
operation, immediately revert to manual CPR.
Page 3-14P/N 11440-001 Rev. 3
User Guide
3.3Ending Active Device Use
1.After either successful resuscitation or termination of activities, press the Stop/Cancel button
followed by the ON/OFF button. The Stop/Cancel button action will cease the compression cycles
and relax the LifeBand (see Figure 3-14). The ON/OFF button action will power down the
AutoPulse.
®
2.Open the Velcro
necessary.
fastener and lift or log roll off the patient from the AutoPulse Platform, as
3.4Preparing the AutoPulse for Its Next Use
1.Remove the LifeBand from the AutoPulse Platform. Refer to Section 2.1.2, “Removing the
LifeBand” for more information.
2.Discard the LifeBand as it is a single-use component. Treat the LifeBand as contaminated
medical waste and dispose of it accordingly.
3.Clean the AutoPulse Platform before its next use. Refer to Section 4.2, “Cleaning the AutoPulse
Platform” for more information.
4.Replace the LifeBand before returning the AutoPulse to service. Refer to Section 2.1.1, “Installing
the LifeBand” for more information.
5.Remove the AutoPulse Battery.
Note:Ensure that the AutoPulse is powered down before removing and replacing the Battery.
6.Replace the Battery with a fully charged Battery before returning the AutoPulse to service.
7.Recharge the used Battery as necessary for future use.
When the AutoPulse is used in conjunction with defibrillators or with other therapeutic devices that must
monitor an ECG signal, interruption of the compression cycles may be required to avoid ECG motion
artifact associated with mechanical chest compressions.
To temporarily interrupt the AutoPulse’s active operation, press the Stop/Cancel button.
To restart the AutoPulse, follow the procedures in Section 3.2, “Starting Chest Compressions”.
P/N 11440-001 Rev. 3Page 3-15
User Guide
3.6Patient Alignment and Securing for Transport
Warning: The AutoPulse is not intended for carrying or transporting a patient. The AutoPulse should
be secured to the top of a backboard or other equipment used to carry or transport the
patient, if necessary. During transport, regular checks of the patient’s alignment should be
performed.
The AutoPulse does not require any patient restraints to perform compressions while the patient is lying
on a flat surface. However, patient restraints should be used to maintain alignment of the patient to the
AutoPulse:
•If the AutoPulse cannot be set on a flat level surface
•If the AutoPulse is used during extrication or during transport
The AutoPulse is designed to accept standard restraints to maintain patient alignment. The rescuer can
secure a patient of up to 300 pounds to a backboard and maneuver the patient as necessary while the
AutoPulse is performing active compressions.
Caution: Motion can cause the patient to shift and restraints to loosen, so care should be given to the
initial strapping for alignment of the patient to the AutoPulse. Regular checks of patient
alignment to the AutoPulse and alignment of the LifeBand to the patient's mid-axillary line
should be made if the AutoPulse is performing active compressions, or before active
compressions are restarted.
When transporting the patient, lift by supporting the patient and the AutoPulse onto the transportation
device (for example, a gurney or backboard) and place the AutoPulse and patient within the vehicle
during AutoPulse operation. Secure the AutoPulse and patient to the transportation device.
The AutoPulse can be used in conjunction with a transportation device during transport to hospital (a
gurney or backboard). However, care must be taken to ensure that the patient is properly strapped to the
transportation device using locally-approved procedures for safe transport.
Caution: Straps or restraints used for transportation purposes must not interfere with the operation of
the AutoPulse. Specifically, straps across the patient’s chest may restrict the compression/
decompression of the chest. In general, strapping schemes must not alter the alignment of the
patient to the AutoPulse.
Page 3-16P/N 11440-001 Rev. 3
User Guide
3.6.1Recommended Method of Patient Extrication
This extrication method includes the AutoPulse Shoulder Restraint, Head Immobilizer, and Soft Carry
Stretcher.
ON
/
O
FF
Head Immobilizer
Shoulder Restraint
Figure 3-16 Securing the Patient for Transport
P/N 11440-001 Rev. 3Page 3-17
User Guide
1.Attach the Shoulder Restraint to keep the patient properly aligned on the AutoPulse Platform.
2.The Head Immobilizer assists in keeping the patient's head from moving, especially when
combined with a cervical collar. A cloth may also be placed under the patient's head.
Soft Stretcher
Figure 3-17 Transporting the Patient
3.When lifted, the Soft Stretcher has a cradling effect that helps maintain alignment of the patient
on the AutoPulse. Users can also allow the patient's lower legs to bend freely at the knees,
facilitating moving around tight corners and stairwells. The AutoPulse platform may also be
secured to a backboard using standard straps or cable ties.
Always ensure the following:
1.Make sure that the patient's armpits and the upper edge of the LifeBand are aligned with the
yellow line on the AutoPulse.
®
2.Make sure that the LifeBand is not twisted and properly mated with the Velcro
.
3.Maintain the LifeBand at 90 degrees with the AutoPulse Platform. Ensure that the LifeBand is
not impeded by anything such as the patient's arms, clothing, straps, and buckles that may
interfere with the movement of the LifeBand.
For more information about strapping and patient restraint options, contact ZOLL at +1.800.348.9016
(or +1.978.421.9655).
Page 3-18P/N 11440-001 Rev. 3
User Guide
3.7Viewing AutoPulse Platform Information
On initial power-up, pressing the Menu/Mode Switch button allows you to:
1.Enter Communication mode (refer to Section 3.8.1, step 3 on page 3-21)
2.View last patient session information
3.View AutoPulse Platform information
4.View AutoPulse Battery information
Items 2-4 above may also be accessed from the Administrative Menu (refer to section Section 2.3,
“Administrative Menu: User Pre-set Options,” on page 2-10).
Once the Menu or Administrative Menu is active use the Move Up and Down arrow buttons to highlight
the desired menu item and the Select Choice Button to select it.
Information presented about the Last Patient session is:
1.Total compressions
2.Total active time (min:sec)
3.Total pause time (min:sec)
The last patient session data is updated after the AutoPulse is power cycled and one complete
compression occurs.
Information presented about the AutoPulse is:
1.Model number
2.Serial number
3.Software version
4.Name of manufacturer
5.Manufacturer location (city, state, country)
Information presented about the AutoPulse Battery is:
1.Battery serial number
2.Number of charge cycles performed (refer to the AutoPulse Power System User Guide for more
information).
P/N 11440-001 Rev. 3Page 3-19
User Guide
From any of the information displays, press the gray Menu/Mode Switch button under the word “back”
to return to the main Menu or Administrative Menu.
To exit from the menu, press the gray Menu/Mode Switch button under the word “back.” The AutoPulse
returns to the idle state, ready for patient alignment.
To exit from the Administrative Menu, press the “START” (green) button under the word “RESTART.”
The AutoPulse will restart and place you into the idle state, ready for patient alignment.
3.8Uploading AutoPulse Information to your PC
The AutoPulse has an infrared communication port located on the top (“head”) edge of the unit just
below the ON/OFF button (see Figure 3-1). This port is provided to allow upload of the information from
the platform. Currently the platform will upload the patient information from at least the last three patient
sessions (and potentially many more depending on the length of the resuscitation efforts). The uploaded
patient information consists of the key timing information on the AutoPulse use.
Prior to using the AutoPulse infrared communication port, the Code Summary software must be installed
on your host computer. This software may be acquired from ZOLL Data Systems (www.zolldata.com)
or from your ZOLL Data Systems representative (+1.978.421.9655 or +1.301.801.0000). Installation
instructions, the software's User Guide and system requirements are provided with the software. The
computer with Code Summary installed must be equipped with an infrared communication adapter listed
below.
•iFoundry InfraRed Communication Module (serial port 8001A or USB port 8003A)
•Actisys InfraRed Communication Module
To purchase an infrared adaptor, please contact ZOLL.
Page 3-20P/N 11440-001 Rev. 3
User Guide
3.8.1Information Upload Procedure
1.Turn on your PC (the host computer).
2.Start the Code Review software.
30:2
MENU
Enter Comm Mode
View Last Patient Info
View Platform Info
View Battery Info
exit
Figure 3-18 Main Menu
3.Power up the AutoPulse and when the initialization is complete press the Menu/Mode Switch button
and then select the “Enter Comm. Mode” option from the Main Menu. Alternatively you may enter
directly into the Communication Mode by depressing the Menu/Mode Switch button as you power
up the platform.
AutoPulse Platform
Infrared Communication Port
Computer Infrared
Communication Adapter
Figure 3-19 Infrared Communication Set-up
P/N 11440-001 Rev. 3Page 3-21
User Guide
4.Aim the host computer’s infrared communication unit at the AutoPulse infrared communication
port.
W a iting for Connection...
RESTART
Figure 3-20 Waiting for Connection... Display Panel Screen
5.Upon entering the Communication Mode the platform will immediately attempt to establish a
communication connection with the host computer. While this is occurring the “Waiting for
Connection...” panel screen is displayed. If a connection can not be established within ten minutes
the AutoPulse will automatically power down.
Connected.
RESTART
Figure 3-21 Connected Display Panel Screen
Page 3-22P/N 11440-001 Rev. 3
User Guide
6.When a connection is established between the platform and your computer the “Connected” panel
screen will be displayed. If a connection cannot be made within ten minutes the AutoPulse will
automatically power down.
Uploa ding Da t a ...
Figure 3-22 Transmitting... Display Panel Screen
7.From your computer send the start transmission of data command. See the Code Review software
User Guide for instructions. The “Transmitting...” panel screen will be displayed while the patient
information is being transferred between the platform and your computer. When the data transfer is
complete the panel screen will return to the Connected display (see Figure 3-21). If while
transmitting data the AutoPulse loses communication with the host computer it will automatically
try to reestablish the connection: repeat procedure starting at step 5 on page 3-22 (see Figure 3-20).
8.To exit from the Communication mode, either power down the platform by pressing the ON/OFF
button or press the “START” (green) button under the word “RESTART.” The AutoPulse will restart
and place you into the idle state, ready for patient alignment.
P/N 11440-001 Rev. 3Page 3-23
User Guide
[This page left intentionally blank.]
Page 3-24P/N 11440-001 Rev. 3
User Guide
4Maintaining the AutoPulse System
4.1Charging the Battery
For more information on the AutoPulse Power System and its components, refer to the AutoPulse Power
System User Guide.
To charge a Battery, follow these steps:
1.Slide the Battery into an available charging bay (see Figure 4-1). Make sure that the Battery locks
into place (locking bar engaged).
Caution: Remove the protective plastic cap from the Battery connector before attempting to charge
the Battery.
Note:For optimal charging, make sure that the Battery is at room temperature before insertion
into the Battery Charger.
Note:Do not slam the Battery into the Battery Charger because doing so may cause damage to
the Battery’s connector.
Note:The Battery is mechanically keyed so that it can only be inserted in one orientation. Do
not force the Battery into the charging bay. If resistance is met, check for appropriate
orientation, and check to ensure that there are no obstructions to Battery insertion.
Figure 4-1Battery Charger with Charging Bay
2.The Battery Charger automatically detects the presence of an AutoPulse Battery.
3.The Battery Charger’s status will be indicated on the control panel (see Figure 4-2).
P/N 11440-001 Rev. 3Page 4-1
User Guide
Figure 4-2Battery Charger Control Panel
4.The Battery Charger will automatically perform a battery test-cycle when needed. To manually run
a test-cycle on a Battery, simply press the Battery Charger’s Start Test button at any time that the
Battery Charger is in Charge mode.
Note:Do not remove a Battery from the Battery Charger until its charging completes or the
Battery’s runtime will be reduced.
Note:Do not remove a Battery during a test-cycle or the Battery’s runtime will be unknown.
Removing a Battery during a test-cycle may cause the Battery Charger to automatically
enter a test-cycle mode the next time a Battery is inserted into it.
5.Remove the Battery from the Battery Charger by gripping the Battery while pressing the finger latch
to disengage the locking bar. Pull the Battery straight out until it fully clears the charging bay.
Note:Recently-charged Batteries can be warm to the touch. This is a consequence of normal
operation.
4.2Cleaning the AutoPulse Platform
1.Remove and dispose of the LifeBand.
2.Wipe all the surfaces of the AutoPulse Platform free of foreign matter and spills with a disinfectant
or bactericidal wipe. Check the vents to ensure that they are free and clear of any obstructive matter.
Caution: Do not submerge the AutoPulse in liquid.
3.Ensure that the AutoPulse is dry before storing.
4.3Storing the AutoPulse Platform
1.The AutoPulse should be powered down before it is stored.
Note:The AutoPulse will automatically power down after 10 minutes of inactivity.
2.The AutoPulse should be returned to a Carry Case, Hygiene Barrier or Transporter before being
stored. To purchase a Carry Case, Hygiene Barrier or Transporter, please contact your ZOLL
representative by calling +1.800.348.9011 or +1.978.421.9655.
3.The AutoPulse should be stored in a cool, dry place.
4.The Battery should be stored in the Battery Charger unless the Battery is deployed in an AutoPulse.
Page 4-2P/N 11440-001 Rev. 3
User Guide
4.4Maintenance
There are no user-serviceable parts in the AutoPulse . There are no AutoPulse components that require
calibration. In case of repair or service, contact ZOLL at +1.800.348.9011 or +1.978.421.9655.
Consideration should be given to the frequency of product use, storage conditions and knowledge of all
product users to determine your local operator maintenance schedule.
•If the AutoPulse is used more than once per week, daily inspections may be appropriate.
•If the AutoPulse is used less than once per week, weekly inspections may be appropriate.
It is the customer's decision to determine the frequency of preventative maintenance.
Inspect the AutoPulse in accordance with the inspection checklist provided. Follow the recommended
actions as appropriate. The AutoPulse conducts a self-test each time the device is powered on. The selftest is designed to perform a variety of internal system self-checks for the platform components, and is
fully automated.
1.Inspect the platform for physical damage, including cracks, tears and missing or broken pieces.
Contact ZOLL as needed.
2.Remove the Battery from the AutoPulse.
3.Inspect the battery compartment for foreign substances, be sure it is free and clear for battery
insertion. Remove any foreign matter or debris.
4.Check the vent openings on the bottom-side of the platform, insure they are free and clear to allow
air flow. Remove any foreign matter or debris that may block the flow of air.
5.Check the shaft opening where the belt clip is inserted for foreign substances. Blow out the opening
using compressed air or micro-duster. Do not insert tools or other objects into the opening.
6.Install a fully charged Battery from the Battery Charger. If a spare or second battery is stored with
the AutoPulse, place the spare in the Battery Charger and place the Batter removed from the
AutoPulse into the spare location. If a spare Battery is not stored, place the Battery into the Battery
Charger.
7.Perform System Self Test - Install a LifeBand and power on the AutoPulse. On Power up, all LEDs
on the display will momentarily light up, then only the Green POWER LED will remain lit. The
battery gauge should also be visible with 4 bars, if not replace Battery with a fully charged Battery
from the Battery Charger. If the AutoPulse Platform red ALERT LED remains on, refer to the
Troubleshooting Guide. If the RED LED can not be resolved, contact ZOLL.
P/N 11440-001 Rev. 3Page 4-3
User Guide
Inspect the Battery in accordance with the inspection checklist provided below. Follow the recommended
actions as appropriate. The Battery is automatically checked each time the device is powered on.
1.Inspect the Battery Pack, including the connector for physical damage that would preclude its
insertion in either the platform or the charger. Contact ZOLL as needed.
2.Inspect the pack enclosure for cracks that would expose the internal volume of the battery. Discard
and replace battery.
3.Place the Battery in the Battery Charger where a self-test will be performed during the battery
charging cycle.
Inspect the Battery Charger in accordance with the inspection checklist provided below. Follow the
recommended actions as appropriate.
1.Inspect the Battery Charger for physical damage. Contact ZOLL as needed.
2.Inspect the battery compartments for foreign substances, be sure they are free and clear for battery
insertion. Remove any foreign matter or debris.
If the product malfunctions or fails any of these inspections, contact ZOLL at +1.800.348.9011 or
+1.978.421.9655.
Page 4-4P/N 11440-001 Rev. 3
User Guide
5Troubleshooting Procedures
In the event of a user advisory, fault or system error, the User Control Panel Alert light-emitting diode
(LED) illuminates (refer to Section 1.4, “User Controls and Indicators”). Consider the information in this
chapter.
Warning: If a persistent fault or system error occurs during active operation, immediately revert to
manual CPR.
5.1Troubleshooting Batteries
If a Battery’s charge falls too low, a Low Battery warning will appear on the Control Panel Display. The
Low Battery warning display will be accompanied with an audio warning of four rapid beeps which will
be followed by two beeps every 30 seconds until the battery is replaced or depleted. If operation
continues without changing the Battery, a Replace Battery screen will appear (see Figure 5-1 and Figure
5-2). If this is the case:
•Press the ON/OFF button to power down the AutoPulse.
•Replace the AutoPulse Battery with a new, fully-charged Battery. Press the ON/OFF button
followed by the Start/Continue button again.
30:2
LOW BATTERY
Figure 5-1Low Battery Warning
P/N 11440-001 Rev. 3Page 5-1
User Guide
CONTINUOUS
Replac e/rec harge battery.
REPLACE BATTERY
menu
RESTART
Figure 5-2Replace Battery Screen
5.2Troubleshooting User Advisories and Faults
The AutoPulse enters a User Advisory state (see Figure 5-3) or the Fault State (see Figure 5-4) when one
of several conditions is detected. A User Advisory generally indicates that a misalignment or
inappropriate movement of the patient or the LifeBand has occurred. A Fault generally indicates that the
AutoPulse has detected an inappropriate internal condition. Both conditions are typically correctable by
the operator. Follow the instructions on the screen and then attempt to RESTART active operation by
pressing the Start/Continue button. If that does not work you should follow these general steps to
troubleshoot Advisories and Faults:
1.Check for correct patient alignment (refer to Section 3.1, “Deploying the AutoPulse System” for
more information), fully extend LifeBand and attempt to RESTART active operation by pressing
the Start/Continue button.
2.If the user advisory or fault persists:
a)Remove the LifeBand from, and fully re-insert into, the AutoPulse (refer to Section 2.1,
“LifeBand Load-distributing Band (LDB)”) and then press the Start/Continue button again.
b)Remove and replace the LifeBand with a new LifeBand and then press the Start/Continue
button again.
c)Check the AutoPulse Platform for blocked vents.
3.If a User Advisory or Fault indicator cannot be cleared, record the User Advisory or Fault number
and contact ZOLL at +1.800.348.9011 or +1.978.421.9655.
Page 5-2P/N 11440-001 Rev. 3
User Guide
In either User Advisory state or the Fault State, pressing the gray Menu/Mode Switch button, under the
word “menu” allows you to access the AutoPulse Platform information menu (refer to section Section
3.7, “Viewing AutoPulse Platform Information,” on page 3-19).
3 0:2
Pull up LifeBa nd,
check pa tient a lignment,
a nd press RESTART.
Figure 5-3A User Advisory Screen
3 0:2
Pull up LifeBand and
p
ress RESTART.
USER ADVISORY(02)
menuRESTART
FAULT(08)
menuRESTART
Figure 5-4A Fault Screen
P/N 11440-001 Rev. 3Page 5-3
5.2.1User Advisory (45)
User Guide
3 0:2
Pull up ches t ba nd,
a nd press RESTART.
USER ADVISORY(45)
menu RESTART
Figure 5-5User Advisory (45)
The AutoPulse driveshaft has a “home” position that is a point of reference for AutoPulse operation. If
the driveshaft is not at its home position when the AutoPulse is powered on, a User Advisory (45) will
occur. This User Advisory will persist until the driveshaft is returned to its home position.
To clear a User Advisory (45) pull up on the LifeBand until the chest bands are fully extended (thereby
moving the driveshaft back to its home position), and then RESTART.
Warning: Removing the band clip when the driveshaft is not at its home position (see Figure 2-8 on
page 2-7) will result in a permanent User Advisory (45) that the user will not be able to
clear. To avoid this situation, the following guidelines should be adhered to:
1.The LifeBand should be removed from the driveshaft ONLY from its home position.
2.The LifeBand must be completely unwound with the seam visible (see Figure 2-10 on
page 2-8).
3.If the LifeBand is cut, care should be taken to ensure that the bands are fully extended
before the cover plate is opened and the band clip is removed.
4.Do not attempt to defeat the shaft lock mechanism, which keeps the driveshaft at its
home position when a LifeBand is not installed.
If a User Advisory (45) can't be cleared; Power off the AutoPulse and contact ZOLL at +1.800.348.9011
or +1.978.421.9655
Page 5-4P/N 11440-001 Rev. 3
User Guide
SYSTEM ERROR
5.3Troubleshooting Errors
In the event of a System Error (see Figure 5-6), you can take no steps to return the AutoPulse to normal
operation. The AutoPulse has detected an unrecoverable problem and cannot be restarted. Therefore, you
must:
1.Begin manual CPR immediately.
2.Contact ZOLL at +1.800.348.9011 or +1.978.421.9655.
3 0:2
Figure 5-6System Error Screen
SYSTEM ERROR
OUT OF SERVICE
REVERT TO MANUAL CPR
menu
P/N 11440-001 Rev. 3Page 5-5
User Guide
[This page left intentionally blank.]
Page 5-6P/N 11440-001 Rev. 3
User Guide
Appendix A Technical Specifications
The specifications provided in this chapter apply to the AutoPulse Resuscitation System Model 100.
A.1Patient Parameters
The AutoPulse is designed for adults with weight of no more than 300 lbs. (136 kg) with chest
circumference of 29.9 to 51.2 in. (76 to 130 cm) and chest width of 9.8 to 15 in. (25 to 38 cm).
Warning: The AutoPulse System is intended for use on adults, 18 years of age or older.
A.2LifeBand
The latex-free LifeBand is for single-patient use only. The LifeBand consists of a cover plate and two
®
bands integrated with a patient liner and compression pads with a Velcro
fastener.
A.3Operating Parameters
Table A-1 Operating Parameters
CategorySpecifications
Chest displacementEqual to 20% reduction in anterior-posterior chest depth.
Physiological duty cycle50 ± 5%.
Compression rate80 ± 5 compressions per minute.
Compression modes
(user selectable)
•30:2 (30 compressions with two 1.5 second ventilation pauses)
•15:2 (15 compressions with two 1.5 second ventilation pauses)
•Continuous compressions
Ventilation pause
(30:2 and 15:2 mode)
Two pauses of 1.5 seconds.
A.4Platform Physical
Table A-2Physical Specifications
CategorySpecifications
Size (L×W×H)32.5 in. by 17.6 in. by 3.0 in. (82.6 cm by 44.7 cm by 7.6 cm).
Free fall (non-operational)Meets IEC 60068-2-32 Basic Environmental Testing Procedures – Free Fall –
Safety requirementsSafety certified to UL2601, CSA 22.2 No. 601.1-M90, EN60601-1
2
m/s
Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)
Procedure 1.
P/N 11440-001 Rev. 3Page A-5
Table A-9Battery Charger EMI/EMC Specifications
TitleStandard
Electrostatic dischargeIEC 61000-4-2, Level 3
User Guide
RF electromagnetic fields
immunity
EFT/burstIEC 61000-4-4, Level 2
Surge immunityIEC 61000-4-5, Level 2
Conducted RF disturbances
immunity
Dips, interruptions, and
variations
Harmonics current emissions IEC 61000-3-2, Class A
Radiated emissionsCISPR 11/EN55011, Group 1, Class A
IEC 61000-4-3, Level 2
IEC 61000-4-6, Class A
IEC 61000-4-11
FCC part 15, Class A
A.10Limited Warranty
ZOLL warrants to the Customer that from the date of shipment from ZOLL’s facility, the Equipment,
(constituting the AutoPulse Platform, Battery Charger, and Batteries) will be free from defects in
material and workmanship under normal use and service for a period of one (1) year from date of
shipment. Accessories and disposables shall be warranted for ninety (90) days from date of shipment.
During such period ZOLL will at no charge to the Customer, either repair or replace (at ZOLL’s sole
option) any part of the Equipment found to be defective in material or workmanship. ZOLL will pay for
the shipping, insurance cost and a service loaner at no charge for use during the repair.
ZOLL shall not be responsible for any Equipment defect, the failure of the Equipment to perform any
specified function, or any other nonconformance of the Equipment caused by or attributable to: (i) any
modification of the Equipment by the Customer, (ii) the use of the Equipment with any associated or
complementary Equipment, accessory or software not supplied by ZOLL; (iii) any misuse or abuse of
the Equipment; (iv) exposure of the Equipment to conditions beyond the environmental, power or
operating constraints specified by ZOLL; or (v) installation or use of the Equipment other than in
accordance with ZOLL’s instructions.
This warranty does not cover items subject to normal wear and burnout during use, including but not
limited to lamps, fuses, internal batteries, LifeBands and accessories. The foregoing warranty does not
apply to software included as part of the Equipment (including software embodied in read-only memory,
known as “firmware”).The foregoing warranty constitutes the exclusive remedy of the Customer and the
exclusive liability of ZOLL for any breach of any warranty related to the Equipment supplied hereunder.
Page A-6P/N 11440-001 Rev. 3
User Guide
THE WARRANTY SET FORTH HEREIN IS EXCLUSIVE AND ZOLL EXPRESSLY DISCLAIMS
ALL OTHER WARRANTIES WHETHER WRITTEN, ORAL, IMPLIED, OR STATUTORY,
INCLUDING BUT NOT LIMITED TO ANY WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE.
ZOLL’S MAXIMUM LIABILITY ARISING OUT OF THE SALE OF THE PRODUCTS (i.e.
EQUIPMENT AND RELATED ACCESSORIES AND DISPOSABLES) OR THEIR USE, WHETHER
BASED UPON WARRANTY, CONTRACT, TORT OR OTHERWISE, SHALL NOT EXCEED THE
ACTUAL PAYMENTS RECEIVED BY ZOLL IN CONNECTION THEREWITH. ZOLL SHALL
NOT BE LIABLE FOR ANY INCIDENTAL, SPECIAL OR CONSEQUENTIAL LOSS, DAMAGE
OR EXPENSE (INCLUDING WITHOUT LIMITATION LOST PROFITS) DIRECTLY OR
INDIRECTLY ARISING FROM THE SALE, INABILITY TO SELL, USE OR LOSS OF USE OF ANY
PRODUCT (HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY), EVEN IF ZOLL HAS
BEEN ADVISED OF THE POSSIBILITY OF SUCH LOSS. THE FOREGOING LIMITATIONS
SHALL NOT APPLY TO ANY CLAIMS FOR BODILY INJURY OR DEATH TO THE EXTENT
THAT LIMITATION OF DAMAGES FOR SUCH CLAIMS ARE UNENFORCEABLE OR AGAINST
PUBLIC POLICY UNDER ANY APPLICABLE STATUTE OR RULE OF LAW.
P/N 11440-001 Rev. 3Page A-7
User Guide
[This page left intentionally blank.]
Page A-8P/N 11440-001 Rev. 3
User Guide
Appendix BAutoPulse Parts and Accessories
Table B-1AutoPulse Parts and Accessories
The AutoPulse components and accessories listed in Table B-1 can be ordered from ZOLL at
+1.800.348.9011 or +1.978.421.9655.
DescriptionZOLL Part No.
AutoPulse
AutoPulse Battery Charger, U.S.8700-0703-01
AutoPulse Battery8700-0702-01
LifeBand, 1 pack8700-0701-01
LifeBand, 3 pack8700-0706-01
LifeBand Trainer (Reusable, Not for Patient Use)8700-0707-01
AutoPulse Battery Charger power cord8700-0704-01
Platform8700-0700-01
User Guide, AutoPulse (English)9650-0714-01
User Guide, AutoPulse Power System (English)9650-0715-01
AutoPulse Soft Carry Case8700-0705-01
AutoPulse Grip Strips8700-0708-01
AutoPulse Shoulder Restraints8700-0709-01
AutoPulse Head Immobilizer, 5 pack8700-0710-01
Back board Cable Ties, 25 pack8700-0711-01
AutoPulse In-Service Video, DVD Format (English)9658-0716-01
Soft Stretcher8700-0712-01
AutoPulse Training System8700-0713-01
AutoPulse Transporter8700-0716-01
AutoPulse Hygiene Barrier8700-0717-01
AutoPulse, 1 Year Extended Warranty (Includes 1 Preventive Maintenance)8700-001
AutoPulse, 2 Year Extended Warranty (Includes 1 Preventive Maintenance Per Year) 8700-002
P/N 11440-001 Rev. 3Page B-1
User Guide
[This page left intentionally blank.]
Page B-2P/N 11440-001 Rev. 3
User Guide
IndexX
A
accessories B-1
active device use
ending 3-15
Administrative Menu 2-11
Alert LED 1-9
AutoPulse
accessories B-1
battery 1-3
cleaning 4-2
components 1-2
deploying 3-1
description 1-1, B-1
ending active device use 3-15
interrupting 3-15
introduction 1-1
maintenance 4-3
operating characteristics 1-11
operating parameters A-1
parts B-1
platform 1-2
starting chest compressions 3-8
storing 4-2
technical specifications A-1
troubleshooting 5-1
viewing system information 3-19
warranty A-6
AutoPulse Platform
viewing system information 3-19
AutoPulse Power System 2-9
B
Batteries
troubleshooting 5-1
Battery 1-3
capacity A-3
charge time A-3
charging 4-1
description B-1
specifications A-3, A-4
test-cycle time A-3
Battery charge icon 1-10
Battery charge status 1-9