VOLUME CONTROL
CPR FEEDBACK OPERATION
To increase the prompt volume, insert a small flat blade screwdriver into the slotted
opening on the right side of the unit and turn clockwise.
POWER
Insert six new "C" batteries into the battery well in the Trainer. Replace
batteries before expiration date, if the unit fails to power on, or if the unit returns
to "Select Mode" during voice prompts.
Alternatively, you can power the unit using the supplied AC adapter.
TROUBLESHOOTING
AED Trainer problem / display: Action:
The Trainer does not respond
when you press .
CHECKSUM ERROR
PROCESSOR ERR
Check and replace batteries. If the unit
still does not respond, call Zoll Technical
Service.
Call Zoll Technical Service.
Call Zoll Technical Service.
The AED PLUS Trainer includes controls that allow instructors to simulate the voice prompts
and text messages issued by the AED PLUS when rescuers perform CPR. These controls
include:
Handheld control: Description:
Metronome On/Off
Trainer
This control activates the training unit's metronome which begins
to beep at 60 compressions per minute (CPM) then automatically
accelerates to 100 CPM after a few seconds. Metronome beeps
are independent of the student's actual compression rate and
continue until the CPR period times out or until the Metronome
On/Off key is pressed again. The control and the metronome
function are only active during periods when the AED Trainer has
prompted the rescuer to perform CPR.
Push Harder Control
Pressing this control button causes the training unit to emit a
single "Push Harder" voice and text prompt. Instructors should
depress this button when the student's chest compressions are
too shallow.
Note: The AED PLUS will issue only two "Push Harder" prompts
during each CPR session when chest compressions are less than
1 ½ inches (3.75 cm) deep.
MAINTENANCE
1. Keep the unit clean, undamaged, and free of excessive wear.
2. Verify that there are no cracks or loose parts in the housing.
3. Verify that all cables are free of cracks, cuts, and exposed or broken
wires.
4. Replace Training electrodes as necessary.
CLEANING
Clean unit with soft, damp cloth using 90% isopropyl alcohol, or soap
and water, or chlorine bleach (30 ml/liter water).
Do not immerse any part of the unit in water.
Do not use ketones (e.g., MEK, acetone).
Avoid using abrasives (e.g., paper towels) on the displays.
Do not sterilize the unit.
OK
Good Compressions Control
Pressing this control button causes the training unit to emit a
"Good Compressions" voice prompt and text message. To
simulate operation of the AED PLUS defibrillator, instructors
should use this control to provide students with positive feedback
about improvements in chest compression depth when they
respond to "Push Harder" prompts.
Note: The AED PLUS emits a maximum of one "Good
Compressions" prompt during each CPR session.
Automated External Defibrillator Trainer
Operator's Guide
For technical support or repair:
Technical Service Department
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA USA 01824-4105
978-421-9655 800-348-9011
Fax: 978-421-0010
WARNING
WARNING
WARNING
Product intended for indoor use only.
Use only Safety Agency-approved AC/DC or DC/DC power
adapter rated 12 VDC at 800 mA output.
This device is for training use only. It will NOT deliver
defibrillation energy.
NOTE
For replacement parts, see the AED PLUS Administrator's Guide.
International Service: Contact your local distributor.
© ZOLL Medical Corporation 2006
TRAINING USE ONLY - NOT FOR CLINICAL USE
REF 9650-0304-01 Rev. E
CONTROLS / INDICATORS
2
SELECT MODE
SCENARIOS
Handheld:
OK
Handheld display: Action:
1
Change mode on the text display.
Select displayed mode.
+
End current mode.
Pause operation / Resume operation.
CPR metronome on / CPR metronome off.
CPR "Push Harder" prompt.
CPR "Good Compressions" prompt.
“Attach defib pads to patient’s bare chest” prompt.
Noisy ECG / Selected ECG.
Low battery / Normal battery.
SELECT MODE
MANUAL
SCENARIO 1
SCENARIO 4
2a
Handheld display: Action:
ANALYSIS - VF
SELECT VF
SELECT NSR
SELECT ASYS
MANUAL
Press to change the mode.
Press to select the displayed mode.
VF is the default simulated heart rhythm for manual mode.
Press to change the heart rhythm.
Press to select the displayed rhythm.
2
3
Simulated heart rhythm:
VF 1
VF 2
VF 3
VF 4
Done
Simulated heart rhythm: Student Response:
ASYSTOLE 1
ASYSTOLE 2
VF 1
VF 2
Done
Simulated heart rhythm: Student Response:
VF 1
VF 2
NSR 1
NSR 2
Done
Simulated heart rhythm:
Student Response:
Shock
CPR
Shock
CPR
No Shock
CPR
Shock
CPR
Shock
CPR
No Shock
CPR
Student Response:
Unit Failed / Unit OK.
Zoll AED Plus Trainer:
Power on / Power off.
Status indicator: is ready for use; is fault
conditon.
1
AED Trainer control: Action:
SETUP
Install six "C" batteries or attach the AC adapter to the
AED PLUS Trainer.
Turn the unit on. If the handheld is connected, the status
indicator will switch to then back to , signifying
that the self-test phase is complete and the unit is OK.
ANALYSIS - NSR
2b
Handheld display: Action:
ANALYSIS - VF
SCENARIO DONE
SCENARIO 1 TO 4
The Handheld will display the selected heart rhythm and the
Trainer will behave accordingly. You can change the heart
rhythm at any time by following the instructions above.
The Handheld will display the preprogrammed heart rhythm.
The AED PLUS Trainer will operate according to the
the chosen scenario. Instruct the student to follow the AED
PLUS Trainer voice instructions and text prompts.
Indicates that the scenario is done. The Trainer will
continue to cycle through analysis or CPR, depending on the
4
ASYSTOLE 1
ASYSTOLE 2
ASYSTOLE 3
ASYSTOLE 4
Done
No Shock
CPR
No Shock
CPR
PLUG IN HANDHELD
NOTE
The Handheld display indicates the current training settings while the
Trainer simulates the actual Zoll AED PLUS behavior.
If the handheld is not connected, the status indicator will
remain . Connect the Handheld to the Trainer. The
status indicator will turn and the software version
number will display.
NOTE
scenario. Press to return to the
SELECT MODE prompt.
Refer to the Zoll AED PLUS Operator and Administration Guides for additonal
information.
+