ZOLL AED PLUS Simulator Operator's Manual

Specifications
Size: 6.2"L x 3.7"W x 1.8"H (15.7 cm x 9.4 cm x 4.6 cm) Weight: 1.1 lb (0.5 Kg) Power Requirements: 9 Volt alkaline battery Temperature: 40 F to 95 F (5 C to 35 C), operating 14 F to 122 F (-10 C to 50 C), storage
Cleaning and Inspecting
Clean the ZOLL AED PLUS Simulator with a soft cloth, mild soap, and water. Do not immerse any part of the ZOLL AED PLUS Simulator in water. Do not use alcohol or ketones (MEK, acetone, etc.). Do not autoclave the ZOLL AED PLUS Simulator.
Check that fluid has not been spilled on the unit and that nothing is stored on the unit. Inspect the ZOLL AED PLUS Simulator for physical damage. Check all cables, cords, and connectors for cuts in the insulation, or bent and broken connector pins.
Warranty
(a) The ZOLL AED PLUS Simulator is warranted for 90 days from date of shipment. During such period ZOLL Medical Corporation will, at no charge to the customer, either repair or replace (at ZOLL Medical Corporation's sole option) any part of the equipment found by ZOLL Medical Corporation to be defective in material or workmanship. If ZOLL Medical Corporation's inspection detects no defects in material or workmanship, ZOLL Medical Corporation's regular service charges shall apply. (b) ZOLL Medical Corporation shall not be responsible for any equipment defect, the failure of the equipment to perform any function, or any other nonconformance of the equipment, caused or attri­buted to: (i) any modification of the equipment by the customer, unless such modification is made with the prior written approval of ZOLL Medical Corporation; (ii) the use of the equipment with any associated or complimentary equipment, accessory or software not supplied by ZOLL Medial Cor­poration (iii) any misuse or abuse of the equipment; (iv) exposure of the equipment to conditions beyond the environmental, power or operating constraints specified by ZOLL Medical Corporation; or (v) installation or wiring of the equipment other than in accordance with ZOLL Medical Corporation's instructions. (c) This warranty does not cover items subject to normal wear and burnout during use, including but not limited to lamps, fuses, batteries, patient cables and accessories. (d) The foregoing warranty does not apply to software included as part of the equipment (including software embodied in read-only memory, known as "firmware"). (e) The foregoing warranty constitutes the exclusive remedy of the customer and the exclusive liability of ZOLL Medical Corporation for any breach of any warranty related to the equipment supplied hereunder. THE WARRANTY SET FORTH HEREIN IS EXCLUSIVE AND ZOLL MEDICAL CORPORATION EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES WHETHER WRITTEN, ORAL, IMPLIED, OR STATUTORY, INCLUDING BUT NOT LIMITED TO ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
Service
U.S.A. Customers
Should the ZOLL AED PLUS Simulator require service, contact ZOLL Medical at 1-800-348-9011 (in Massachusetts: 1-978-421-9655). You will be given a Service Request number under which to return the ZOLL AED PLUS Simulator. Send the unit in its original packaging to:
ZOLL Medical Corporation
269 Mill Road Chelmsford, MA 01824-4105 Attn: Technical Service Department
International Customers
Should the ZOLL AED PLUS Simulator require service, it should be returned, in its original container, or equivalent, to the nearest authorized ZOLL Medical Corporation service center.
REF 9651-0800-01 Rev. E Copyright 2008 by ZOLL Medical Corporation. All rights reserved. Printed in U.S.A.
OPERATO R ' S G U I D E
Warnings . . . . . . . . . . . . . . . . . . . . . . 1
Controls and Indicators . . . . . . . . . . . . . 1
Getting Started . . . . . . . . . . . . . . . . . . 2
Simulator Use . . . . . . . . . . . . . . . . . . . 2
Battery Saver Feature . . . . . . . . . . . . . . 2
S pec if ic ati on s. . . . . . . . . . . . . . . . . . 3
C lea ni ng an d Ins pe cti ng . . . . . . . . . . 3
W arr an ty . . . . . . . . . . . . . . . . . . . . . 3
Service . . . . . . . . . . . . . . . . . . . . . . . 3
b/ w p hot o o f s imu lat or
3
VF
NSR
ASYS
CPR AUTO convert
2/30
VF NSR
1 5
SIMULATOR
Getting Started
To use the ZOLL AED PLUS Simulator:
1. Install the 9 Volt alkaline battery.
2. Connect to the ZOLL AED PLUS device.
3. Press ON OFF key to power-on simulator. (If simulator powered-on after inserting the battery, press ON OFF key to power-off simulator, then press again to power-on.)
Simulator Use
ECG rhythms (any one at a time)
VF Ventricular Fibrillation NSR Normal Sinus Rhythm at 72 bpm ASYS Asystole
convert
The convert feature allows you to convert automatically from one rhythm (running rhythm) to another rhythm (waiting rhythm) when a discharge is delivered from the AED PLUS device.
To perform convert operation:
1. Press convert key. Indicator of running rhythm pulses brighter.
2. Press key of rhythm to be simulated immediately after defib discharge.
Indicator of this (waiting) rhythm blinks on and off.
3. Discharge defibrillator. The waiting rhythm becomes the running rhythm.
To cancel convert operation before it's completed, either press convert key again or press key of running rhythm. If convert operation is started, but a discharge is not sensed within two minutes, the convert operation cancels automatically.
CPR
Toggles CPR signal on/off.
AUTO
Provides an automated test sequence of one shockable rhythm, CPR, and one non-shockable rhythm when attached to the ZOLL AED PLUS and user follows voice prompts from defibrillator.
Battery Saver Feature
The simulator powers-off automatically about 30 minutes after the last key press, or 30 minutes after defibrillator energy was last detected. This is normal operation.
To allow simulator to remain powered-on after 30 minutes, follow these instructions:
1. Power-off simulator.
2. Press and hold down convert key, then press and release ON OFF key. The simulator powers-on.
3. Observe NSR indicator. When indicator blinks on and off, release convert key. NSR indicator continues to blink for about one second.
The battery saver feature is now disabled. When you power-off the simulator, it will return to normal battery saver operation.
Low Battery
Red indicator illuminates when battery needs to be replaced.
Defib Discharge
Green indicator illuminates for two seconds when defib discharge is detected.
convert
Simulate cardioversion by activating convert feature. AED PLUS Simulator responds by changing the ECG rhythm following defib discharge.
ON OFF
Press to power-on and to power-off.
VF
NSR
ASYS
CPR AUTO convert
2/30
VF NSR
1 5
Warnings
Keep hands and all other objects clear of simulator while discharging the defibrillator.
Shock hazard exists. Be sure simulator cable is attached securely to the ZOLL® AED PLUS®, that there are no nicks, cracks, or cuts in the cable and that the simulator housing is intact.
Do not discharge into the ZOLL AED PLUS simulator more than three times in one minute. Do not discharge more than 360J into simulator at any time. ZOLL AED PLUS automatically limits discharges to 5J when discharged into the simulator.
Do not immerse or set the ZOLL AED PLUS simulator in liquid. Using the device near or within puddles of liquid may present a shock hazard to the operator, patient, and nearby personnel.
Note: Refer to the appropriate Operator's Guide for proper defibrillator operation.
Product Description
The ZOLL AED PLUS Simulator is a 9V battery operated cardiac rhythm and CPR simulator for use with ZOLL AED PLUS devices.
The ZOLL AED PLUS Simulator attaches to the ZOLL AED PLUS device through a cable on the simulator. The ZOLL AED PLUS Simulator provides ECG and CPR signals to the defibrillator unit via the attached cable.
Controls and Indicators
1
2
CPR
Simulate CPR signal.
AUTO
Generate automated test sequence of ECG rhythms and CPR signal.
REF 9651-0800-01 Rev. E