Xylem WL430 User Manual

Quick Reference
x
Standard Features:
¾ UL listed ¾ Flush clog-free sensor; 316L SST ¾ Enhanced internal surge and lightning protection ¾ ± 0.5% FSO Static Accuracy ¾ 40 ft. polyurethane cable, vented and shielded ¾ Double-sealed submersible cable exit; ¾ 200 lb. tensile strength
TECHNICAL SPECIFICATIONS
WL430-xx
Waste Water Sensor
Rev. E
PERFORMANCE
Static Accuracy: ± 0.5% FSO ( BFSL, RSS)
(combined effects of non-linearity, hysteresis & repeatability)
Repeatability: ± 0.1% FSO Temperature Effects: ± 1.5% FSO over comp
range
(combined effects of Zero & FSO with reference at 70ºF )
Long Term Stability: ± 0.25% FSO per year
MECHANICAL
Pressure Range: 0-10 psi, 0-15 psi Proof Pressure: 2x Full Scale Burst pressure: 5x Full Scale Materials: 316L SST plus cable Pressure Port: Flush mount per outline Electrical Connector: 1/2”-NPT Male submersible
conduit fitting with 40ft of polyurethane cable
Dimensions: Per outline Weight: Nominal 20 oz Compensated Temp Range: 20 to 170°F Operating Temp Range: 0 to 200°F
Pressure
Psig
0-10 psi 276.8 23 20 psi 0-15 psi 415.2 34.6 30 psi
Level In. of Water
Level Ft.
of water
Max.
Pressure
psig
ELECTRICAL
Excitation: 8 - 38 Vdc Output: 1 - 5 Vdc or 4–20 mAdc Zero Balance: ± 1% FSO FSO Setting: ± 1% FSO Resolution: Infinite ( ±.001% FSO) Response Time: < 5mS Insulation Resistance: 1000 MO @ 50 Vdc Reverse Polarity: Protected Warm-Up: < 10 mS Power Supply Effect: ± .002% FSO per V input EMI/RFI: Internal Filtering Lightning Protected: MOV and dual gas discharge
tube
Short Circuit Protected: Up to 40 Vdc
Quick Reference
___________________________________________________________________________________
WL430-xxx Waste Water Sensor
Rev. E
Installation Procedures:
The STEELCAGE assembly can be lowered and placed directly on the floor of the wet well/lift station. No stilling well is necessary. The cable should be routed into the logger or control panel for connection. The assembly does not need to be suspended; it can be placed directly on the wet well floor. The cable should be snug with minimal slack to prevent the assembly from “traveling” about in any turbulence in the wet well. Any excess cable can be coiled with a tie wrap and hung on the wall. This will work fine as long as there is no sharp bend in the cable that could cause it to kink.
DO NOT kink the cable, as this will cause blockage to the breather tube inside.
Electrical Installation:
There are two wires, a vent tube and ground inside the cable.
TURN OFF POWER BEFORE CONNECTING THE WIRES!
1. Connect the ground wire first (shield to earth ground to appropriate controller ground).
2. Then connect the red wire to a 13-30 VDC (24 VDC nominal) power supply. Note: An external power supply with a minimum current capability of 40 mA per transmitter, is required.
3. Connect the black lead to the 4-20mA input on the logger or other equipment that will be monitoring the sensor.