Van Air Systems 46-2529 2-02 User Manual

INSTALLATION, OPERATION, & MAINTENANCE MANUAL
FOR
DEW POINT METER
P/N 46-2529
WARNING
BEFORE STARTING INSTALLATION AND/OR MAINTENANCE PROCEDURES, TURN OFF POWER AND DEPRESSURIZE UNIT BEFORE SERVICING TO PREVENT SERIOUS INJURY. SERIOUS PERSONAL INJURY MAY RESULT IF THESE SAFETY RULES ARE NOT FOLLOWED.
WHEN MAKING ANY ELECTRICAL CONNECTIONS TO THE METER, COMPLY WITH ALL LOCAL, STATE, AND FEDERAL CODES I.E. THE NATIONAL ELECTRICAL CODE. ALWAYS USE COMPONENTS AND WIRING RATED FOR THIS INSTALLATION.
INTRODUCTION SECTION 1
1.1 SPECIFICA TIONS
Dew Point Range ........ -40oF to +15oF (-40oC to -9oC)
Accuracy..................... +2oF
1.2 DIMENSIONS
Alarm Set Point........... -10oF (-23oC) (Adjustable)
Remote Alarm Output
Normally open and closed contacts 5A @ 115V rated
Recorder Output ......... 4-20 mA or 0-5V (jumper selectable)
Power Requirement.....115VAC (+10%) 50/60Hz
Power Consumption ... 4 Watts
Enclosure .................... Nema 4, Fiberglass w/window
INSTALLATION SECTION 2
2.1 INST ALLA TION
The meter can be mounted on any flat vertical surface. The enclosure has threaded inserts on the back for mounting.
Once the unit has been mounted, the sample line can be connected to the sample cell isolation needle valve. The sample line should be 1/4" copper tubing. The pressure of the sample line should not exceed 150 PSIG.
NOTE:
See Model 6392N and 6392N2 instructions for operation, maintenance, printed circuit board details, and calibration of equipment.
The unit will require 115 volts for operation. Run the necessary conduit and wiring to the enclosure.
If the alarm relays and the recorder output are to be used, they will require several procedures for installation. Otherwise, no additional installation steps are required to use the meter.
WIRING DIAGRAM SECTION 3
REPLACEMENT PARTS SECTION 4
ITEM
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15 16
QTY
1 2 4 1 1 1 2
.5'
2 2 1 1 1 1 1 1
DESCRIPTION
Enclosure Panel Terminal, Grey Terminal, Blue Terminal, Ground End piece End clamp Channel Screw, 8-32UNF x 1/2"lg Lockwasher, #8 nom Meter, (humistat) Sensor Sensor mount Meter bezel Cable O-Ring
PART NO.
27-1114
27-1082
26-0380 26-1391 26-1392 26-2053 26-0388 26-0389 28-0204 28-0314 26-5977 26-5978 26-0816 26-5979 46-2185 18-0249
ITEM
17 18 19 20 21 22 23 24 33 34 35 36 37 38 39 40
QTY
1 2 1 1 1 1 1 1 4 4 2 6 6 4 2 2
DESCRIPTION
Tee 1" NPT Bushing 1" x 1/4" NPT Choke fitting Short coupling 10-32 Reducing fitting 1/4" x 10-32 Needle valve Fitting nut & sleeve assy Sintered filter Standoff Screw, 10-32UNF x 1/2"lg Nut, 10-32UNF Washer flat, #10 nom Lockwasher, #10UNF nom Screw, 6-32UNF x 1/4"lg Standoff Lockwasher, #6 nom
PART NO.
14-1807 14-1808 26-0490 26-0657 26-0296 14-1306 26-3043 26-0623 26-6805 28-0203 28-1062 28-0323 28-0312 28-1321 26-6006 28-0324
HYGRODYNAMICS
DIGITAL DEW POINT MONITOR
PC BOARD
MODEL 6392N & 6392N2
TABLE OF CONTENTS
INTRODUCTION 2 SPECIFICATIONS 2 INSTALLATION 2
WIRING SUMMARY 3 OPERATION & MAINTENANCE 4 CIRCUIT CALIBRATION 4 ALARM SETPOINT ADJUSTMENT 4
REMOTE ALARM CONNECTION 5 AUXILIARY ALARM INDICATION 5 AUXILIARY NON-ALARM INDICATION 5 RECORDER OUTPUT 5
AUXILIARY PUMP POWER 5 °F TO °C DISPLAY CHANGE 5 115 TO 230V CHANGE 5
WARRANTY 6
DIAGRAMS
MOUNTING & WIRING DIAGRAM DWG. # 6392NWD
NEWPORT SCIENTIFIC, INC.
8246-E SANDY COURT
JESSUP, MARYLAND 20794-9632
PHONE: (301) 498-6700 FAX (301) 490-2313
E-MAIL: NEWPORT888@AOL.COM
WEBSITE: WWW.NEWPORT-SCIENTIFIC.COM
ADMIN/MANUAL/6392N-N2 JANUARY 2002