The technical documentation supplied is also part of the
product. You should always keep this documentation
within reach.
These Operating Instructions describe your treatment
center with maximum possible equipment.
For daily care of your treatment center, please observe
the instructions provided in “Care and Clearing by the Practice Team”.
To safeguard your warranty claims, please complete the
attached “Installation Report / Warranty Passport”
together with the service engineer immediately after the
installation of your treatment center.
Separate Operating Instructions with corresponding
instructions for care are enclosed with the dental
instruments, SIROLUX FANTASTIC, SIROCAM 3, tray
and X-ray image viewer.
Prior to start-up, you must read all Operating Instructions
to familiarize yourself with the treatment center.
To prevent any personal injury or material damage,
pay special attention to any notes printed in bold type or
marked in one of the ways indicated below:
+
dental treatment
i
NOTE
CAUTION
WARNING
Additional information, explanation or supplement
The information provided under this keyword directly
concerns the functioning of the product and/or warns
against possible operating errors.
This warning symbol warns against possible hazards for
the product or the user.
In case you get stuck despite having thoroughly studied
the Operating Instructions, please contact your dental
depot.
Your C2+ Team
2D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH Table of Contents
bеЦдблЬ
Operating Instructions C2
+
Table of Contents
1General information .................................................................................................... 2
2Warning and safety information ................................................................................ 7
17.4Seat cushion C ................................................................................................................................. 93
Sirona Dental Systems GmbH2 Warning and safety information
Operating Instructions C2
+
2.1 General safety information
2 Warning and safety information
2.1General safety information
Intended useThis dental treatment center is intended for diagnosis,
therapy and dental treatment of humans by properly
trained personnel.
This unit is not intended for operation in areas subject to
explosion hazards.
On-site installationThe 'On-site installation' must be performed according to
our requirements. Details are described in the document
"Installation requirements".
Maintenance and repairAs manufacturers of dental medical equipment and in
the interest of the operational safety of your system, we
stress the importance of having maintenance and repair
of your treatment center performed only by ourselves or
by agencies expressly authorized by us. Furthermore
components influencing the safety of the device should
always be replaced with original spare parts upon failure.
When having such work done, we suggest that you
request a certificate stating the type and extent of work
performed,
cations
range, as well as the date, name of organization and signature.
includingstatementsconcerninganymodifi-
of therated parametersorof theoperating
Changes to the unitChanges to this unit which could impair the safety of the
system owner, patients or other persons are prohibited
by legislation!
For reasons of product safety, only original Sirona
accessories approved for this product, or accessories
from third parties approved by Sirona, may be used. The
user is responsible for dangers resulting from the use of
non-approved accessories.
If any devices not approved by Sirona are connected,
they must comply with the applicable standards:
IEC 60950 for information technology equipment (e.g.
PCs), and IEC 60601-1 for medical electrical equipment.
The loudspeaker socket of the monitor may be connected only to a device which complies with IEC 60950
(e.g. PC) or IEC 60601-1, and under no circumstances
e.g. to a stereo system etc.
59 57 928 D 3370
D 3370.201.01.15.02
7
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General safety informationOperating Instructions C2
Combination with other unitsAny person who assembles or modifies a medical elec-
trical system complying with the standard IEC 60601-1-1
(Safety requirements for electromedical systems) by
combining it with other equipment (e.g., by connecting it
with a PC) is responsible for ensuring that the requirements of this regulation are met to their full extent for the
safety of the patients, operators and environment.
In case of doubt, contact the manufacturer of the system
components.
+
Electromagnetic compatibility (EMC)The C2
+
complies with the requirements of
IEC 60601-1-2:2001.
Medical electrical devices are subject to special precautionary measures regarding EMC. They must be
installed and operated as specified in the document
"Installation requirements".
Portable and mobile HF communication devices can
influence medical electrical equipment. The use of
mobile telephones in the practice or hospital area therefore must be prohibited.
For video systems:
Electromagnetic disturbances in the environment of the
+
may result in reduced image quality. In such cases,
C2
it is advisable to repeat the exposure before saving the
image.
In case of line voltage fades, öimage data stored in the
video camera may be lost. Deposit the camera and
repeat the exposure.
Image artifacts may possibly occur in the event of
radio-frequency interference caused by the transmitters
of e.g. radio services or radio amateurs.
Influence on cardiac pacemakersThe workstation includes strong magnets which can
affect cardiac pacemakers of patients, users and
techni-cal personnel.Therefore, please pay attention to
the safety information on pages 32 and 67.
Treatment of highly immunosuppressed patientsHighly immunosuppressed patients should not come in
contact with water from the treatment center. The use of
sterile solutions is recommended.
Quality of water / air supplyAir and water supply must meet the requirements speci-
fied in the Installation Instructions. Use only clean
water.
Water quality complianceTo ensure compliance with the medical and national
legal requirements for water from treatment centers,
Sirona recommends equipping the treatment center with
a disinfection system. As owner of the treatment center,
you are responsible for the water quality and may have
to take alternative measures to ensure its compliance if
you operate the treatment center without a disinfection
system.
59 57 928 D 3370
8D 3370.201.01.15.02
Sirona Dental Systems GmbH2 Warning and safety information
Operating Instructions C2
+
Please contact your specialized dealer or your relevant
dental association for the respective national requirements and measures.
2.1 General safety information
HF surgeryThis dental treatment center is available with a high-fre-
quency surgical device (optional).
Only in the Federal Republic of Germany: The user is
obligated to keep a “medical product log” if a HF surgical
device is installed! See section 18.3 "Safety checks" on
page 98.
Identification of warning and safety informationTo avoid personal injury and material damage, you must
also observe the warning and safety information provided in the present Operating Instructions. They are
highlighted by the captionNOTE, CAUTION or
WARNING.
i
Symbols usedObserve accompanying documents
(on name plate of chair)
Ventilation slotsUnder no circumstances may the ventilation slots on the
unit be covered, since otherwise the air circulation will be
obstructed.
Do not spray with disinfectants or the like into ventilation
slots.
Vacuum systemThe suction removal of aluminum and other metal oxides
from blasting devices via the automatic separator integrated in the treatment center and the amalgam separator is prohibited!
This would cause extreme wear and clogging of the suction and water paths.
A separate vacuum system must be used in connection
with metal oxide blasting devices.
Treatment centers equipped with a central wet suction
system are generally suitable for suction removal of the
above material. However, please be sure to observe the
instructions provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in
connection with Sirona treatment centers. However, in
such cases, make sure that the system is subsequently
rinsed with an adequate amount of water.
Patient chairPlease observe the maximum load capacity of the chair
of 135kg according to EN ISO 6875 (tested with a
four-fold safety factor according to IEC 60601-1)
Electric micromotors in dentist elementIntermittent operation: 1 min. ON – 7 min. OFF with a
motor current of 1.5A.
Drive motors for chair and backrest6% duty time, cycle time 250s.
59 57 928 D 3370
D 3370.201.01.15.02
9
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General safety informationOperating Instructions C2
Maintenance of the treatment centerDespite the outstanding quality of your treatment center
and regular care by the practice team, it is required in the
interest of operational safety that preventive maintenance be performed at predetermined intervals.
In order to guarantee the operational safety and reliability of your treatment center and to avoid damage
due to natural wear, you as the system owner must
have your system checked through regularly by an
authorized service engineer from your dental depot.
Furthermore, safety checks must be performed.
Please contact your dental depot to obtain a maintenance offer.
For details please refer to chapter 18"Maintenance" on
page 97.
Dismantling and reassemblyWhen dismantling and reassembling the system, pro-
ceed according to the installation instructions for new
installation in order to guarantee its functioning and stability.
DisposalIt applies generally that the national regulations have to
be complied with when disposing of this product. Please
observe the regulations applying in your country.
Within the European Economic Community the directive
2002/96/EEC (WEEE) for electrical and electronic
devices requires environmentally compatible recycling /
disposal.
+
Please observe the disposal regulations applicable in
your country.
Within the European Economic Area, this product is subject to Directive 2002/96/EC as well as the corresponding national laws. This directive requires environmentally
sound recycling/disposal of the product.
The product must not be disposed of as domestic refuse!
Please contact your dealer if final disposal of your product is required.
10D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH3 Technical description
Operating Instructions C2
+
3 Technical description
Model designationC2
+
Power supply connection230VAC 50Hz,
115VAC 50/60Hz
100VAC 50/60Hz
Nominal current4.5A at 230V
9.5A at 115V
11.5A at 100V
Main unit fusefor 230VAC: SB 6.3A H , 250VAC, REF 10 77 452
Relative humidity: 30% – 75%
Air pressure: 700hPa – 1060hPa
Transport and storage conditionsTemperature: -40°C – +70°C (-40°F – 158°F)
Relative humidity: 10% – 95%
Air pressure: 500hPa – 1060hPa
Protection classClass I equipment
Degree of protection against electrical shock:Type B, applied parts
except Cerec 3D camera, SIROTOM, SIROCAM 3 /
SIROCAM C:
Type BF applied parts
Degree of protection against ingress of waterOrdinary equipment (not protected)
The foot switch is protected against dripping water IPX
1.
Mode of operation: Continuous operation with intermittent loading corre-
sponding to the dental mode of working.
Permanently connected unit.
Year of manufacture(on name plate of chair)
Tests / approvalsThis dental treatment center complies with the require-
ments of
IEC 60601-1: 1988 + A1: 1991 + A2: 1995 and of
IEC 60601-1-2: 2001 .
DVGW: This unit complies with the technical rules and
requirements on safety and hygiene for connection to
the drinking water supply, provided that a disinfection
unit is installed.
This product bears the CE marking in accordance with
the provisions of the Council Directive 93/42/EEC of
June 14, 1993 concerning medical devices.
0123
59 57 928 D 3370
D 3370.201.01.15.02
11
4 Operating and functional elements Sirona Dental Systems GmbH
S
0
4.1 System overview C2
+
Operating Instructions C2
4 Operating and functional elements
+
4.1System overview C2
20
1
21
22
234
56
S
0
SAN
1
2
L
7
+
8
910
11 12 13 1416 1715
+
C2
4
0
1
0
0
3
0
7
5
SS
2
0
5
0
–
M
ode
11
1
0
+
22
2
2
5
5
A
/
B
0
.2
r
p
m
x
1
1
0
0
0
00
24
18
19
23
1
Assistant element
2
Holder 1: Polylight Mini LED or 3rd suction hose
3
Holder 2: SPRAYVIT in the assistant element
4
Holder 3: Suction handpiece
5
Holder 4: Saliva ejector
6
Control panel on the assistant element
7
Swiveling cuspidor
8
Tu m bl er f il le r
9
Water unit with amalgam rotor, disinfection system,
automatic separator, wet suction
10
Patient chair, option: folding armrest
11
Holder 1: SPRAYVIT in the dentist element
12
Holder 2 and 3: Electric motor / highspeed handpiece burr drives
13
Holder 4: 3. Electric motor / highspeed handpiece
burr drive or SIROSONIC L scaler
14
Holder 5: 4. Burr drive (highspeed handpiece only)
or SIROSONIC L
or Polylight Mini LED
15
Holder 6: SIROSONIC L or HF surgery
or SIROCAM 3 / SIROCAM C
or Polylight Mini LED
16
Control panel on the dentist element
17
Dentist element
18
Unit main switch
19
Locking brake on the dentist and assistant element
20
Headrest adjustable by motor drive
21
Manual switch for chair movements
22
4-way foot control of chair
23
unit foot switch
24
Additional holder (for SIROCAM 3 / SIROCAM C
only)
59 57 928 D 3370
12D 3370.201.01.15.02
Sirona Dental Systems GmbH4 Operating and functional elements
AMALG
DESINF
Operating Instructions C2
+
4.2 Control panel on the dentist element
4.2Control panel on the dentist element
Control panel
40
100
SS
00
A
RPM
30
20
75
50
11
22
A/B
10
25
–
–
Mode
+
0.2
rpm x 1000
AMALG
1
DESINFAB
RPM
System status indication display and Mode
key
AMALG appears if the amalgam rotor needs to be
replaced.
DESINF flashes if DENTOSEPT P must be refilled.
A appears if user A is selected.
B appears if user B is selected.
Mode button with
tings.
Apart from the settings in the Mode dialog, it is also possible to make settings with the
•Intensity of instrument lighting
•Intensity of the operation light
•Tumbler filling time
•Cuspidor flushing time
+ / – keys for programming basic set-
+ / – keys:
59 57 928 D 3370
D 3370.201.01.15.02
13
4 Operating and functional elements Sirona Dental Systems GmbH
4.2 Control panel on the dentist elementOperating Instructions C2
Main functions – dark blue keys
40
40
100
100
Quick setting keys for instrument intensity and
speed
+
30
20
10
0.2
rpm x 1000
75
50
25
1
for electric motors, SIROSON L ultrasound scalers and
SIROTOM HF electrosurgery.
Spray ON / OFF
on burr drives
Chip blower
on burr drives
Counterclockwise rotation
for electric motor
A
INSTR. LIGHT
11
Instrument light ON / OFF
%
SS
Pressing this key for some time displays the light intensity settings dialog.
Mouth rinsing position chair program.
with last position memory function (freely programmable)
Chair program 0
Entry/exit position (freely programmable)
00
Chair programs 1 and 2
(freely programmable)
22
i
NOTICE
The four program keys are also used for reprogramming
the chair programs.
14D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH4 Operating and functional elements
Operating Instructions C2
+
4.2 Control panel on the dentist element
Additional functions – light blue keys
Preselection of user A or B
A
OPERATING LIGHT
A/B
All instruments must be in place.
The user cannot be changed as long as one of the instruments is removed
.
Saving the instrument settings
The instrument to be programmed must be removed.
X-ray image viewer ON/OFF
For SIVISION 3 also WHITE SCREEN activation
Composite key ON / OFF
for SIROLUX operation light, reduced light intensity
8,000 lux for composite fillings
<
SIROLUX operation light ON / OFF
%
for normal light intensity > 24,000 lux
Pressing this key for a while displays the light intensity
settings dialog:
OPERATING LIGHT
A
CUP FILL TIMER
A
BOWL FLUSH TIMER
Tumbler filling
Pressing this key for a while displays the settings dialog:
CUP FILL TIMER
Cuspidor flushing
Pressing this key for a while displays the settings dialog:
BOWL FLUSH TIMER
Freely selectable function
e.g. call key, always functions as button
freely available relay 230 V, 6 A
(connected by the service engineer).
Freely selectable function
The function can be preselected in the Mode dialog as
button or switch.
freely available relay 230V, 6 A
(connected by the service engineer).
59 57 928 D 3370
D 3370.201.01.15.02
15
4 Operating and functional elements Sirona Dental Systems GmbH
4.2 Control panel on the dentist elementOperating Instructions C2
Chair functions – gray keys
for manual chair settings:
Move headrest out / in
(cannot be used with MultiMotion head rest)
Tilt headrest backward or forward.
(cannot be used with MultiMotion head rest)
Tilt backrest backward or forward.
Vertical adjustment down / up
+
Additional functions of the panel keys
for video
i
NOTICE
Apart from the previously described functions, the keys
of the dentist element have additional functions in connection with the video application. These are described
in chapter 15 starting on page 71.
16D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH4 Operating and functional elements
Operating Instructions C2
+
4.3 Control panel on the assistant element
4.3Control panel on the assistant element
Control panel
SAN
S
1
00
SS
0
2
Main functions – dark blue keys
Mouth rinsing position chair program.
with memory function (freely programmable)
Chair program 0
Entry/exit position (freely programmable)
L
59 57 928 D 3370
D 3370.201.01.15.02
11
22
Chair programs 1 and 2
(freely programmable)
i
NOTICE
The four program keys are also used for reprogramming
the chair programs.
17
4 Operating and functional elements Sirona Dental Systems GmbH
4.3 Control panel on the assistant elementOperating Instructions C2
Additional functions – light blue keys
Tumbler filling function
ON / OFF
Cuspidor flushing function
ON / OFF
X-ray image viewer ON / OFF
for SIVISION 2 also WHITE SCREEN activation
Sanitation of the treatment center
+
SAN
L
This key can be used to start the treatment center sanitation program (see instructions “Care and Cleaning by
the Practice Team”).
Freely selectable function
The function can be preselected in the Mode dialog as
button or switch.
Light ON/OFF
With SPRAYVIT removed:
•Instrument lighting
With SPRAYVIT deposited:
•Operation light
Chair functions – gray keys
Manual adjustment of headrest
Move headrest out/in
(cannot be used with MultiMotion head rest)
18D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH5 Putting the system into operation
Operating Instructions C2
+
5 Putting the system into operation
Initial start-up
The disinfection system adds a disinfectant to the water
(1:100) to prevent the formation of microorganisms in
the water system.
Prior to initial start-up of your treatment center, sanitation must be performed.
If, on the basis of an agreement with you, sanitation
was skipped by the service engineer following installation of your treatment center, please perform sanitation
yourself as described in the separate instructions “Care
and Cleaning by the Practice Team”.
Sanitation takes approx. 24 hours.
Main switch
A
READY
10:04:27
ON
OFF
Switch the main switch ON ( l ).
Following power-up, the treatment center automatically
performs a self-test.
CAUTION
For safety reasons, always switch the treatment center
OFF (O) after consulting hours. This cuts off the air and
water supply and the line voltage.
Ready to operate
After the end of the self-test, the message READY the
time and the last selected user, A or B, appear on the
display.
After the treatment center has been switched on, the
cursor is between the headrest in/out keys.
If a program LED lights up, then the chair was in a programmed position before the treatment center was
switched off. If no program LED lights up, then the chair
was in a non-programmed position before the treatment
center was switched off.
59 57 928 D 3370
D 3370.201.01.15.02
The unit is now ready to operate.
19
5 Putting the system into operation Sirona Dental Systems GmbH
A
MAINTENANCE
IN 30 DAYS
Operating Instructions C2
Display of next maintenance date
the next maintenance date is less than 30 days away,
this message appears each time the unit is switched on.
MAINTENANCE
IN 30 DAYS
(see MAINTENANCE Mode dialog).
After this date has expired, the following message is displayed:
MAINTENANCE
REQUIRED
The display message then disappears as soon as an
instrument is removed from its holder.
i
NOTICE
This message can be reset only by the service technician after completion of maintenance work.
CAUTION
wrongright
The dental chair has a maximum load capacity of 135kg
according to EN ISO 6875 (tested with a four-fold safety
factor according to IEC 60601-1).
The patient’s arms and legs must rest on the upholstery
of the chair!
+
A
B
READY
10:04:27
A/B
Adjusting the headrest
Move headrest out
Move headrest in
Hygienic headrest protection and seat cushion for children: see page 93.
Preselecting the user
The treatment center offers the possibility of managing
two different chair and instrument programs for two
users.
Preselect user A or B using the A/B key, with all instruments in place.
The preselected user is shown in the top left corner of
the display. All settings which have been stored for that
user are then activated.
20D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH6 Foot switch
Operating Instructions C2
+
6.1 C+ foot switch
6 Foot switch
6.1C
+
foot switch
1.
B
C
3.
2.
4.
A
D
This foot switch allows you to work with both hands free
using the cursor control or manually by key operation.
The foot switch has a 4-way foot control plate for cursor
navigation, which works independently of the pedal.
Functions
1. Step on foot pedal – all instruments in place
The dentist element moves toward the operator as
long as the foot switch is actuated
(or toward the foot end of the patient if reversed by
the service engineer, see page 28).
Step on foot pedal – instrument removed
Activation of the instrument.
If appropriate, intensity control relative to pedal
movement (if “speed controller foot switch” has been
preset under Mode).
2. Shift 4-way foot control plate
With cursor control enabled
(Mode dialog 11.8):
A forward – cursor up
B backward – cursor down
C to the left – cursor to the left
D to the right – cursor to the right
Cursor navigation along the marked path and re-
lease of the corresponding key function.
With cursor control disabled
(Mode dialog 11.8):
A forward – without function
B backward – the dentist element moves away as
long as the switch is actuated by the user.
C to the left – without function
D to the right – without function
59 57 928 D 3370
D 3370.201.01.15.02
3. Actuate left button – all instruments in place
Programmed movement of the chair into mouth rins-
ing position S or into last treatment position (accord-
ing to starting situation).
Actuate left button – instrument removed
Toggle between spray ON and spray OFF, or
SIVISION function.
4. Actuate right button – all instruments in place
Programmed movement of the chair into the entry/
exit position 0.
Actuate right button – instrument removed
Chip blower active for duration of actuation, or
SIVISION function.
21
6 Foot switch Sirona Dental Systems GmbH
6.1 C+ foot switchOperating Instructions C2
Automatic cursor control with C+ foot switch
1.
A yellow diode on the dentist element indicates the position of the cursor. The cursor path is marked by lines.
1. After the unit is switched on or after the entry/exit position 0 has been reached, the cursor is between the
move headrest out / in keys in the right key block.
To simplify hand-free headrest positioning, the cursor again jumps between the move headrest out / in
keys if the instruments are removed only for wipe
disinfection or for changing straight/contra-angle
handpieces, but are not activated.
2. After the dentist element is moved via the foot switch
pedal, the cursor jumps between chair program keys
1 and 2.
3. On instrument removal or on activation of any instru-
ment, the cursor jumps to the horizontal navigation
path (position between 1st and 2nd row of keys) in
the instrument block.
4. After the instrument is deposited the cursor jumps
back to its position between chair program keys 1
and 2.
0.2
rpm x 1000
3.
40
100
30
75
20
50
10
25
1
A/B
2./4.
SS1100
22
+
CEREC Chairline foot switch
This foot switch includes all of the functions of the
+
foot switch.
C
It also features an additional pedal for operation of the
CEREC Chairline.
For details please refer to the CEREC Chairline Operating Instructions, REF 60 46 028.
22D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH6 Foot switch
Operating Instructions C2
+
6.2 C foot switch
6.2C foot switch
With this foot switch, all functions can be controlled
A*
C
using the entire pedal surface for hands-free working.
However, it is also possible to work manually by key
operation.
For cursor navigation, place your foot fully on the pedal
surface.
Functions (factory settings)
i
NOTICE
The foot switch functions A and B can be reversed by the
service engineer upon request.
B*
D
With cursor control enabled (Mode dialog 11.8)
Actuate foot switch in direction A – cursor upwards.
Actuate foot switch in direction B – cursor downwards.
Swing foot switch to the left C – cursor to the left.
Swing foot switch to the right D – cursor to the right.
Cursor navigation along the marked path and release of
the corresponding key function.
Instrument start
The cursor must be between the spray and the chip
blower key.
With instrument removed, actuate the foot switch in
direction A.
Moving the dentist element
If the cursor is between the headrest
in/out keys:
The dentist element moves toward the operator as long
as the foot switch is actuated (or toward the foot end of
the patient if reversed; this can be set by the service
engineer).
59 57 928 D 3370
D 3370.201.01.15.02
With cursor control disabled (Mode dialog
11.8)
Actuate foot switch in direction A:
With instrument removed – Activation of the instrument.
With instruments in place – The dentist element moves
toward the operator as long as the foot switch is actuated
(or toward the foot end of the patient if reversed; this can
be set by the service engineer).
Actuate foot switch in direction B:
With instrument removed – No function
23
6 Foot switch Sirona Dental Systems GmbH
6.2 C foot switchOperating Instructions C2
With instruments in place – The dentist element moves
away from the operator as long as the foot switch is actuated (or toward the operator if reversed; this can be set
by the service engineer).
Swing foot switch to the left C:
With instrument removed – Spray ON/OFF for burr
instruments or SIVISION function.
With instruments in place – SIVISION function.
Swing foot switch to the right D:
With instrument removed – Chip blower ON for burr
instruments or SIVISION function.
With instruments in place – SIVISION function.
Automatic cursor control with C foot switch
+
0.2
rpm x 1000
2.
3.
40
100
30
75
20
50
10
25
1
A/B
4.
SS1100
22
1.
A yellow diode on the dentist element indicates the position of the cursor. The cursor path is marked by lines.
1. After the unit is switched on or after entry/exit position 0 has been reached, the cursor is between the
move headrest out / in keys in the right key block. To
simplify hands-free headrest positioning, the cursor
jumps back between the move headrest out / in keys
if the instruments are removed only for wipe disinfection or for changing straight/contra-angle handpieces, but are not activated.
2. After the dentist element is moved using the foot
switch pedal, the cursor jumps between chair program keys 1 and 2.
3. On instrument removal or on activation of any instrument, the cursor jumps to a position between the
Spray and Chip blower keys in the first instrument
block.
If the cursor is in the quick setting key column at the
very left, it also jumps to a position between the
spray and chip blower keys in the instrument block
when the foot switch is swung to the right (D).
4. After the instrument is deposited, the cursor jumps
back to its position between chair program keys S
and 0.
24D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH7 Program selection
Operating Instructions C2
+
7.1 Safety
7 Program selection
7.1Safety
CAUTION
The patient’s arms and legs must be resting on the chair
upholstery during the program run!
CAUTION
Make sure that no obstacles (e.g. window wings, drawers, devices, …) extend into the movement range.
Safety stop
A
BACKREST
10:04:27
A built-in safety circuit stops the chair movement in the
following situations:
•The foot support collides with an obstacle.
•The backrest collides with an obstacle.
•The motorized headrest collides with an obstacle
•The swiveling cuspidor is swiveled in during chair
movement.
At the same time, a double-beep warning signal is
issued.
The chair then automatically moves upwards a short distance until the path has been cleared (except for the
swiveling cuspidor)
With instruments in place, the last activated safety
switch is indicated in plain text on the display:
•TILTING PART for motorized headrest
•BACKREST
•FOOT SUPPORT
CUSPIDOR appears on the display after 10 seconds if
the automatic return movement of the cuspidor to its
position is obstructed.
59 57 928 D 3370
D 3370.201.01.15.02
25
7 Program selection Sirona Dental Systems GmbH
7.2 Program selectionOperating Instructions C2
7.2Program selection
Keys for selecting the programmed patient positions.
+
SS
11
00
22
00
SS
With swivelable cuspidor:
Before the chair moves into the selected program position, the swiveled in cuspidor moves back automatically
to its starting position.
Factory settings:
The operation light switches OFF automatically before
the chair moves into the entry/exit position 0 or into the
mouth rinsing position S.
The operation light switches ON automatically after the
chair has reached patient position 1 or 2.
Changing the factory settings:
When programming the patient positions you can preselect whether the operation light should switch OFF or ON
in the different chair programs 0, S, 1 or 2 (refer to page
29).
Program key 0
Intended for the entry/exit position.
Program key S with memory function.
In the S program it is possible to program the cuspidor flushing and / or the tumbler filling function (see Mode
dialog 10.8 and 10.9).
If this key is pressed a second time, the treatment center
returns to its previous position (memory function).
Example: You have programmed this key for the mouth
rinsing position. When you press
ment center moves into the programmed position.
Memory function: When you press this key again, the
treatment center returns to its previous position.
26D 3370.201.01.15.02
thiskeythenthetreat-
59 57 928 D 3370
Sirona Dental Systems GmbH7 Program selection
Operating Instructions C2
+
7.3 MOVEMENT STOP!
7.3MOVEMENT STOP!
Movement of the chair into a programmed position
can be stopped as follows:
•By pressing one of the chair-related keys located on the dentist and assistant elements.
•By actuating the 4-way foot control.
•By actuating the 4-way switch on a motor-adjustable headrest in any direction.
All movements of the treatment center are stopped
immediately.
At the same time, a double-beep warning signal is
issued.
Please note that pressing the program key of the
program which is just running once again does NOT
cause the movement to stop!
59 57 928 D 3370
D 3370.201.01.15.02
27
8 Patient positions Sirona Dental Systems GmbH
8.1 Headrest adjustable by motor driveOperating Instructions C2
8 Patient positions
8.1Headrest adjustable by motor drive
The keys located on the right key block of the dentist element control panel can be used to trigger the different
chair movements manually.
Move headrest out/in
The headrest can also be moved out or in with the two
upper keys in the right block on the assistant element
control panel and on the upper 4-way switch (up/down)
on the headrest.
Headrest tilt
The headrest can also be tilted to the back or front with
the upper 4-way switch on the headrest (left/right).
+
Headrest functions are also possible with the cuspidor
swung in.
Backrest tilt
The backrest can also be tilted with the 4-way foot control on the chair base
switch on the headrest.
Lever to the left:
Backrest tilts backwards.
Lever to the right:
Backrest tilts forward.
Height adjustment
The height can also be adjusted with the 4-way foot control on the chair base as well as with the lower 4-way
switch on the headrest.
Lever up:
The chair moves up.
Lever down:
The chair moves down.
*
as well as with the lower 4-way
Before backrest tilting or up/down movements are
started, the swiveled-in cuspidor automatically returns to
its starting position.
59 57 928 D 3370
28D 3370.201.01.15.02
Sirona Dental Systems GmbH8 Patient positions
Operating Instructions C2
+
8.2 Programming Patient Positions
8.2Programming Patient Positions
Chair and dentist element
SS
11
00
22
The four factory-set programs can be changed individually by users A and B (observe the display) .
•Programs 1 and 2
•Entry/exit program 0
•Mouth rinsing program S
1. Move the chair into the desired
pressing the different setting keys (see page 28).
2. To have the SIROLUX FANTASTIC operation light
switch on or remain off when the chair reaches the
programmed treatment position, you must switch the
lamp ON or OFF now. This setting is then also programmed.
3. Now move the dentist element to the desired treatment position by hand.
CAUTION
In order to avoid damage to the dentist and assistant elements and to the chair upholstery, make sure that these
elements do not protrude into the movement range of the
chair.
treatment position by
4. To save the program settings, press the desired program key approx. 3 seconds until an acoustic signal
sounds and the LED of the corresponding key lights
up.
Programming is completed now.
i
NOTICE
Programming is not possible by activating the program
keys with the foot switch (cursor control). Thus programming errors are avoided.
59 57 928 D 3370
D 3370.201.01.15.02
29
8 Patient positions Sirona Dental Systems GmbH
A
8.3 MultiMotion headrestOperating Instructions C2
8.3MultiMotion headrest
Adjustment to patient's height
The headrest can be adjusted to the patient's height by
pulling out or pushing in the headrest extension. This
adjustment is usually required at the beginning of treatment!
i
NOTICE
Before positioning the patient on the headrest, make
sure that the head support has been optimally adjusted
to the patient's height. This simplifies all subsequent
work with the MultiMotion considerably, since readjustment to the patient's height can thus be omitted when
changing over from mandibular to maxillary treatment.
The MultiMotion enables you to adjust the patient's head
in a way that optimally supports viewing of areas of the
mouth which are difficult to access.
+
Adjustment of hyperextension
Mandibular position
The mandibular position can be set by pulling the operating handle (A). The anatomical movement of the
arched extension keeps the patient's head in the support.
Pull the headrest out of the guide by pulling
handle A.
i
NOTICE
You can reduce the adjustment noise by pressing
release A.
30D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH8 Patient positions
A
B
B
Operating Instructions C2
+
8.3 MultiMotion headrest
Maxillary position
Take the load off the headrest by lifting it slightly.
Press release A on the handle.
Let the headrest slide down slowly.
Let go of the release when the headrest reaches the
desired position.
Rotation and tilt
The MultiMotion enables rotation of the patient's head
about the longitudinal axis of his body as well as lateral
tilting of his head.
The MultiMotion is set to the desired position as follows:
1. Press and hold one or both of the two side control
elements B.
The rotation and tilt adjustment locks of the headrest
are now released.
2. Set the headrest to the desired position.
3. Let go of the control elements B.
The supporting surface is fixed in place.
Make sure that the headrest is indeed fixed in place
after you let go of the control elements!
59 57 928 D 3370
D 3370.201.01.15.02
31
8 Patient positions Sirona Dental Systems GmbH
C
E
D
F
8.3 MultiMotion headrestOperating Instructions C2
Removing the MultiMotion
For certain treatments (e.g. of children) it may be expedient to remove the MultiMotion completely in order to
obtain better access to the patient. The patient's head
will then be supported on the contact surface of the
headrest extension F.
To remove the MultiMotion, proceed as follows:
1. Press release button C.
2. Pull the complete headrest out of the headrest ex-
tension.
3. Cover the opening of the headrest extension with
cover cap E.
4. If a children's head pad D is available, place it on
the headrest extension. The head pad is held magnetically.
i
NOTICE
The children's head pad (D) is not included in the scope
of supply, however, is available from your dental dealer.
+
DANGER
The children's head pad (D) contains a strong magnet on its bottom side. This magnet could affect any
cardiac pacemaker located nearby.
Therefore, do not allow patients, users and technical
personnel with a cardiac pacemaker to be located
near this magnet.
Furthermore, direct contact of the head pad with
magnetic cards can lead to data loss.
i
NOTICE
Deposit the removed MultiMotion in a safe place where it
cannot fall onto the floor.
Inserting the MultiMotion
To insert the MultiMotion, proceed as follows:
Check the guide of the removed headrest for con-
tamination. Remove any contamination.
Remove the cover cap E.
Reinsert the MultiMotion in the guide from above until it
audibly locks in place.
Pull on the headrest again to make sure that it is
locked securely in place.
32D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH9 Dentist element and instrument functions
AMALG
DESINF
S
0
Operating Instructions C2
+
9.1 Dentist element
9 Dentist element and instrument
functions
9.1Dentist element
Maximum load capacity
The maximum load of the dentist element is 2kg (4.4
4.4 lbs
lbs).
Entry/exit position
00
S
0
S
A
N
1
2
L
When the 0 key (factory setting) is actuated, the dentist
element moves to the entry/exit position.
CAUTION
+
F
2
N
I
S
E
C
D
M
P
G
R
L
A
M
A
A
+
e
d
o
M
–
S
0
SA
N
1
2
L
+
C
2
4
0
1
0
0
3
0
7
5
SS
00
2
0
5
0
–
11
M
1
o
d
0
e
22
+
2
2
5
5
A
/B
0
.
2
r
p
m
x
1
1
0
0
0
Tripping hazard! Turn the dentist element outward be-
fore the patient enters or leaves it. This prevents the patient's legs from getting tangled in the instrument hoses.
Travel direction of track:
Changing the factory setting
In the factory setting, the dentist element moves toward
the operator if the foot switch is actuated. The factory
setting can be changed by the service engineer at the
customer's request.
When the foot switch is actuated, the dentist element
moves to the opposite end position.
59 57 928 D 3370
D 3370.201.01.15.02
33
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.2 General instrument functionsOperating Instructions C2
9.2General instrument functions
Spray
The spray cooling can be switched ON/OFF by activat-
3
C
ing the spray key on the control panel of the dentist element. If spray is preselected, the green LED in the key
lights up.
With the C+ foot switch
the spray can be switched ON/OFF with the instrument
removed by pressing the left button (3) on the foot
switch.
With the C foot switch
the cursor jumps directly between the spray and chip
blower keys after removing an instrument. The spray
cooling can be switched ON/OFF by swinging the foot
switch to the left (C).
Spray amount
+
The spray amount is preset at the factory. However, it
can be adjusted using the control valve at the bottom
front part of the dentist element.
To make the adjustment loosen the ring (5) counterclockwise, adjust the spray by turning the screw (6) and
5
screw the ring (5) tight again.
This setting is then valid for all burr drives.
6
Chip blower
C+ foot switch
4
With the C
the instrument removed by pressing the right button (4)
on the foot switch. As long as the button is pressed, an
air jet escapes from the nozzle of the burr instrument.
C foot switch
With the C foot switch, the cursor jumps directly between
D
the spray and chip blower keys after removing an instrument. When swinging the foot switch to the right (D), an
air jet escapes from the nozzle of the burr instrument for
the duration of actuation. If the chip blower is active, the
green LED in the key lights up.
+
foot switch, the chip blower is activated with
34D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH9 Dentist element and instrument functions
A
%
A
RPM
Operating Instructions C2
+
INSTR. LIGHT
9.3 Electric motor
Instrument light ON / OFF
With this key on the control panel of the dentist element,
you preselect the instrument light for the removed instrument.
If light ON is preselected, the green LED in the key is illuminated.
When you press the key for more than 3 seconds, the
INSTR. LIGHT settings dialog is displayed. The light
intensity can be set between 60% and 100% here with
– / + keys.
the
i
NOTICE
If more than one instrument is removed, only the instrument which was removed first is operative.
9.3Electric motor
Speed setting
40
40
30
20
10
0.2
rpm x 1000
100
100
75
50
25
1
After an electric motor is removed, the programmed
speed of the motor in RPM (revolutions per minute)
appears on the display.
The speed is set with the quick setting keys on the left of
the control panel of the dentist element.
If the cursor is in the intensity setting block, then the
speed can also be set by moving the cursor forward/
back with the foot switch.
If you press the corresponding key briefly (< 0.5s), then
the value in RPM indicated next to the actuated key
appears on the display:
Key1–200rpm
Key25–10,000 rpm
Key50–20,000rpm
Key75–30,000rpm
Key100–40,000rpm
When the speed is set with the cursor, the displayed
value increments or decements in steps of 10,000 rpm
when the foot switch is briefly actuated (< 1s) forward or
back.
The green LED in the corresponding key lights up.
When you press and hold down a key for > 0.5s, you
adjust the speed in predefined increments:
Increments of200:from 200– 2000rpm
Increments of400:from 2000– 10,000rpm
59 57 928 D 3370
D 3370.201.01.15.02
35
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.4 Highspeed handpieceOperating Instructions C2
Increments of1000: from 10000 – 40,000rpm
When you press a key (> 0.5s) whose value is > the
value shown on the display, the speed is increased.
When you press a key (> 0.5s) whose value is less than
the value shown on the display, the speed is reduced.
When you actuate the foot switch (> 0.5s) forward or
back, the speed is increased or reduced in increments
proceeding from the displayed value.
If intermediate values are set, the green key LED does
not light up.
Counterclockwise rotation key
Counterclockwise rotation key, only for electric motor.
The green LED in the key lights up when this key is activated.
After the electric motor is started with the foot switch, an
acoustic warning signal sounds 6 times.
+
9.4Highspeed handpiece
A
bar
PSI
Counterclockwise rotation of the electric motor can also
be activated via the foot switch:
•Foot switch plate to the left – counterclockwise
•Foot switch plate to the right – clockwise
In this case the corresponding motor must be removed.
Starting the highspeed handpiece
After the highspeed handpiece is removed, 0 bar or PSI
(according to presetting in Mode dialog 11.15) appears
on the display. When the highspeed handpiece is started
and operated, the current air pressure in bar or PSI
always appears on the display.
36D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH9 Dentist element and instrument functions
A
Operating Instructions C2
+
9.5 SIROSONIC L scaler
9.5SIROSONIC L scaler
Tip protector
The tip protector is used as a tool for screwing in instrument tips.
CAUTION
With the handpiece in its holder, the tip protector must remain fitted to prevent injuries.
Intensity setting
40
40
30
20
10
0.2
rpm x 1000
100
100
75
50
25
1
After the SIROSONIC L handpiece is removed, the programmed intensity value appears on the display.
The intensity is set by briefly (< 0.5s) activating the quick
setting keys in increments of 1 / 25 / 50 / 75 / 100.
If the cursor is in the intensity setting block, then the
speed can also be set by moving the cursor forward/
back with the foot switch.
When the intensity is set with the cursor, the displayed
value is incremented/decremented in steps of 25 if the
foot switch is briefly actuated (< 1s) forward or back.
The green LED in the corresponding key lights up.
When you press and hold down a key for > 0.5s, you
adjust the intensity in predefined increments:
Increments of 1 from 1 – 4
Increments of 5 from 10 – 100
If a key whose value is >
is pressed (> 0.5s), the intensity is increased.
When you press a key (> 0.5s) whose value is less than
the value shown on the display, the intensity is reduced.
the value shown on the display
When you actuate the foot switch (> 0.5s) forward or
back, the speed is increased or reduced in increments
proceeding from the displayed value.
If intermediate values are set, the green key LED does
not light up.
59 57 928 D 3370
D 3370.201.01.15.02
37
9 Dentist element and instrument functions Sirona Dental Systems GmbH
The cooling water flow is always switched on automatically.
Endodontics setting
An intensity range of 1 – 4 is provided for endodontics
mode. It can be adjusted in increments of 1.
ENDO
The warning message ENDO appears on the display.
For safety reasons, we recommend that you program
the set value for endodontics. Otherwise, the previously
programmed value becomes active again if the instrument is deposited in its holder in the meantime.
CAUTION
Temperature damage and needle breakages can occur
on exceeding the value of 4 in endodontics therapy.
9.6SIROTOM HF electrosurgery
After the SIROTOM handpiece is removed, the pro-
A
40
40
100
100
grammed intensity value appears on the display.
The foot switch is automatically set to direct starter operation.
+
30
20
10
0.2
rpm x 1000
75
50
25
1
The intensity is set in increments of 1 / 25 / 50 / 75 / 100
by briefly pressing (<0.5 s) the quick setting keys.
If the cursor is in the intensity setting block, then the
intensity can be set by moving the cursor forward/back
with the foot switch.
For intensity setting with the cursor, the next key value is
set, proceeding from the displayed value, by briefly (<0.5
s) actuating the foot switch forward or back.
The green LED in the corresponding key lights up.
If you press and hold down a key (>0.5 s), you adjust the
intensity in predefined increments:
Steps of 1 from 1 – 10
Steps of 5 from 10 – 100
If a key whose value is >
is pressed (> 0.5 s), the intensity is increased.
If a key whose value is less than the value shown in the
display is pressed, the intensity is reduced.
When the foot switch is actuated (> 0.5s) forward or
back, the intensity is increased or reduced in increments, proceeding from the displayed value.
If intermediate values are set, the green key LED does
not light up.
the value shown on the display
38D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH9 Dentist element and instrument functions
A
A
RPM
Operating Instructions C2
+
9.7 SPRAYVIT
9.7SPRAYVIT
If the SPRAYVIT is the only instrument removed, the
instrument light can be switched on or off. The brightness can also be set.
The instrument light switches on when the SPRAYVIT is
activated (if preselected).
The instrument light is switched off after a time lag of 10s
when the SPRAYVIT is no longer activated.
When the SPRAYVIT is deposited in its holder, the
instrument light switches off immediately.
If the SPRAYVIT is activated together with another
instrument, then the light of the SPRAYVIT is not
switched on.
DANGER
After changing hoses, press the water key of the
SPRAYVIT repeatedly until water flows out of it! Only
then can you begin treatment.
9.8Saving the instrument programs
The factory-set instrument programs can be changed
individually by user A and user B.
Select the corresponding user, A or B, remove an instrument and set it according to your wishes.
Example for electric motor:
•40,000rpm
•CW rotation
40
100
A/B
59 57 928 D 3370
D 3370.201.01.15.02
•with cooling spray
•with instrument light
The settings are stored by pressing key A/B for a longer
time (approx. 3s). An acoustic warning signal is issued.
These settings are then activated whenever the instrument is removed.
The user cannot be changed with the instrument
removed.
i
NOTICE
If counterclockwise rotation was programmed, the program is reset to clockwise rotation after the treatment
center is switched off and back on.
39
9 Dentist element and instrument functions Sirona Dental Systems GmbH
A ball stopper for an unoccupied instrument holder is
enclosed with the dentist element.
Plug the ball stopper (1) into the unoccupied instrument
holder to prevent a treatment instrument from being
deposited inadvertently in this holder.
Additional ball stoppers can be ordered (REF 58 99 575)
to seal any further unused instrument holders.
1
+
40D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH9 Dentist element and instrument functions
Operating Instructions C2
+
9.10 Surgery with saline solution (NaCl)
9.10Surgery with saline solution (NaCl)
Preparation of the unit
•Hook NaCl bottle (1) in place.
•Attach peristaltic pump-hose set (2).
9
•Push short end of hose (3) with cannula as shown
8
7
1
through stopper and into NaCl bottle.
•The regulator in the hose clip (4) must be in the top
position (completely open).
•Run long hose (5) along the corresponding motor
hose up to the angle piece and fasten with clips (6).
•Fit coupling (7) onto hose and connect it with the thin
silicone hose (8).
•Connect spray clip (9) with the thin hose and attach
to angle piece.
2
4
3
10
i
6
5
2
NOTICE
The one-way NaCl pump must be secured with a cap after it is attached to the drive (10). Only then is proper operation of the pump guaranteed. The safety cap is supplied with every one-way pump.
Pump-hose set
The peristaltic pump-hose set (2) is a disposable article
and can be purchased as consumable material under
article number F 58707 directly from the manufacturer in
packs of 10 pcs.4Holder preselection for the instrument
Ordering address:Satelec
Industriestr. 9
D-40822 Mettmann, Germany
with saline solution is described in the Mode dialog
"Basic treatment center settings, Mode" 11.2 on page
53.
CAUTION
The pump flow rate must be at least 70ml/min for reasons of safe cooling and to prevent a pressure rise in the
hose (risk of bursting). This is not always given when using third-party ultrasonic tips.
A
59 57 928 D 3370
D 3370.201.01.15.02
RPM
NACL
Switching the NaCl pump on / off
With the handpiece removed, the NaCl pump can be
switched ON/OFF by pressing the spray key(or by actuating the left button on the C + foot switch or swinging the
pedal plate to the left on the C foot switch) .
When the green LED of the key lights up, the NaCl pump
is switched ON.
NACL appears on the display in the instrument dialog of
the instrument to which the NaCl pump has been
assigned.
41
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.11 Satalec Mini LED curing lightOperating Instructions C2
9.11Satalec Mini LED curing light
Safety information
CAUTION
Use the glare shield!
The light beams emitted by this instrument can be dangerous and must not be aimed directly at anyone's eyes,
even if the person concerned is wearing protective goggles. The light may only be directed at the part of the patient's mouth being treated.
+
right
DO NOT STARE INTO THE BEAM OR VIEW IT
LED RADIATION
DIRECTLY WITH
OPTICAL INSTRUMENTS
CLASS 2M LASER PRODUCT
P < 1 mW, λ = 420nm - 480nm
IEC 60825-1 Ed. 1.2
!
wrong
LED beam
exit aperture
!
CAUTION
Do not stare into the beam path with the glass rod removed (Class 2M laser product).
Viewing the LED light exit aperture with certain optical instruments (e.g. a magnifying lens) within a distance of
100mm may constitute an eye hazard.
42D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH9 Dentist element and instrument functions
Operating Instructions C2
+
9.11 Satalec Mini LED curing light
Additional safety information
Any condensation occurring in the handpiece of the Mini
LED may cause impairments (e.g. fogging of the LED).
If the handpiece is taken out of a cool environment and
placed in a warm room, always wait for it to reach room
temperature before putting it into operation.
Curing lights must not be used on persons who are suffering from or have in the past been afflicted by photobiological reactions (including solar urticaria and erythropoetic porphyria). Nor should they be used on persons
currently being treated with any medicine which
increases one's sensitivity to light (including methoxsalene und chlorotetracycline).
Any persons who have in the past suffered from retinal
or eye lens disorders or have undergone eye surgery,
especially for gray cataracts, must consult their medical
eye specialist before using the light or undergoing treatment with the Mini LED. Caution is advisable even if the
patient gives his or her consent, since the light intensity
can cause accidents. It is especially advisable to always
wear the appropriate protective goggles (UV filter).
Installation
1. Connect the power cable to the treatment center.
3
4
2
1
2. Screw on the handpiece.
3. Plug the sterilized light guide into the handpiece.
Make sure that the light guide is inserted correctly.
The light guide must engage with a "click".
4. Slide the eye protector (glare shield) onto the light
guide. It protects your eyes against reflected light.
59 57 928 D 3370
D 3370.201.01.15.02
43
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.11 Satalec Mini LED curing lightOperating Instructions C2
Versions of installation
+
C2
40
100
30
75
SS
00
20
50
–
Mode
11
10
+
22
25
25
A/B
0.2
rpm
x
1
1000
S
0
SAN
1
2
•iin the dentist element, holder 5 or 6
L
•in the assistant element, holder 1
+
4
7
1
Handpiece and accessories
1Handpiece
6
2Light guide
3Power cable
3
4Status indicator lamps
2
5Eye protector
6ON/OFF key
7Function key
5
Operating modes
The Mini LED features three different modes. Press the
function key to select the desired menu.
4.1 Status indicator lamp
The status indicator lamp displays the following statuses:
4.14.2 4.3 4.4
ColorMode
greenNormal operation
flashing redoverheat protection
4.2 Fast curing mode
The Mini LED runs at full power for 10 seconds in this
mode.
Indicator lamp 4.2 displays your selection.
0510
Sec.
Roughly the following power output is attained in this
mode:
59 57 928 D 3370
44D 3370.201.01.15.02
Sirona Dental Systems GmbH9 Dentist element and instrument functions
Operating Instructions C2
+
9.11 Satalec Mini LED curing light
•1100mW/cm² (± 10%) with the standard light guide,
dia. 7.5 mm .
•2000mW/cm² (± 10%) with the optional booster light
guide, dia. 5.5mm.
4.3 Pulsed curing mode
Selection is made by actuating the function key. The
lamp operates at full power in the "pulsed" mode here,
012.5
Sec.
emitting the radiation in 10 consecutive 1s light intervals.
The pulses are separated by pauses of 250ms. Indicator
4.3 displays your selection.
4.4 "Soft start" mode
Selection is made by actuating the function key. The
"soft start" mode features:
05101520
Sec.
•a "soft start" in 10 seconds from 0 to 1100 mW/cm² or from 0 to 2000 mW/cm² with the "booster light
guide" (optional) dia. 5.5mm.
•full power for a period of 10 seconds.
Indicator 4.4 displays your selection.
Handling
After the treatment center is switched on, the Mini LED
is in its default setting, the fast curing mode (4.2).
If the operating mode of the Mini LED is changed with
the treatment center switched on, the mode also
remains saved after the Mini LED is deposited.
After the mode is selected, the Mini LED is ready for use.
Place the light guide as close as possible to the composite material surface to be photopolymerized. Make sure
that the light guide does not contact the material being
cured, since this could damage the light guide and
reduce its effectiveness.
•Never work without the eye protector (5).
•Never look directly into the light reflected by the
tooth surface.
Briefly actuate the ON/OFF key (6) to start the curing
cycle. This is confirmed by an acoustic signal.
6
An acousing signal is sounded every 5 seconds in each
mode.
The end of the cycle is also signaled by an acoustic signal. However, you can also interrupt the cycle at any
time by gently pressing the ON/OFF key.
59 57 928 D 3370
D 3370.201.01.15.02
45
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.11 Satalec Mini LED curing lightOperating Instructions C2
10.3 Tumbler filling with automatic sensor control (Special equipment, not available in all countries)Operating Instructions C2
Hydrocolloid (continuous operation)
After the filling time of 10 seconds or before the filling
time of 2 seconds,
CUP FILL TIMER
HYDROCOLLOID
CUP FILL TIMER HYDROCOLLOID appears on the display.
An unlimited time is then set, as is required for the hydrocolloid function.
Nevertheless, the last saved tumbler filling time (2 – 10s)
is always started when the chair program S is activated,
i.e. no continuous operation!
Starting continuous operation is possible only using the
tumbler filling key.
10.3Tumbler filling with automatic sensor control
(Special equipment, not available in all countries)
With this tumbler filler, the automatic sensor control recognizes the filling level of the tumbler.
When you place the tumbler beneath the tumbler filler,
the tumbler will be filled automatically.
After the preset filling level has been reached, the water
flow is shut off automatically.
If the tumbler is removed before the preset filling level
has been reached, the water flow is shut off immediately.
+
If required, the tumbler can be refilled manually by
pressing the corresponding “tumbler filling” key.
The filling level of the tumbler can be preset with the
knob on the base of the water unit (foot end).
• The tumbler must always be placed in the center of the
area provided for it.
• If an empty tumbler is standing below the tumbler filler
when the treatment center is switched on with the main
switch, the tumbler will not be filled automatically. To activate automatic tumbler filling, remove the tumbler briefly and then put it back again.
• The Mode dialog “CUP FILL ON S - YES – NO” must be
set in such a way that “NO” flashes.
• Hydrocolloid connection is not possible if tumbler filling
with automatic sensor control is installed.
Briefly pressing this key switches the flushing function
ON and causes the cuspidor to be flushed for a preset
period of time.
Pressing the key again during flushing switches the
flushing function off.
Only for cuspidor flushing key of the dentist element:
Pressing this key for > 3 seconds displays the settings
dialog
59 57 928 D 3370
D 3370.201.01.15.02
BOWL FLUSH TIMER
The time can be set in increments of 20 to 420s using the
10.5 Purge / Auto Purge function (purging the water paths)
Auto Purge (APURGE) function
The Auto Purge function offers possibility of automated
purging of all water carrying instruments in the dentist
element, of the assistant Sprayvit and of the tumbler filling unit.
All of the instruments inserted in the sanitation bowl will
be purged when APURGE is activated. If the instruments remain in the sanitation bowl after the treatment
center is switched off, the purging process will automatically be started again the next time the treatment center
is switched on.
We recommend the APURGE function in the following
cases:
•APURGE function before beginning and at the end
of the work day:
A complete purge cycle is performed following activation. The instruments remain in the sanitation
bowl and the treatment center can be switched off.
On the next day, the APURGE function is automatically performed again on all of the instruments remaining in the sanitation bowl as well as on the tumbler filling unit immediately after the treatment center
is switched on.
Then you can prepare the treatment center for daily
practice operation.
SELECT MODE
PURGE / APURGE / SAN
1
•APURGE function after longer periods of disuse:
A complete purge cycle is performed following activation. The instruments remain in the sanitation
bowl and the treatment center can be switched off.
Now you can switch on the treatment center briefly
every day and then switch it off again after the purging process.
Setting the APURGE function
Activate the selection dialog box by pressing quick setting key 1 (> 3 s) when all of the instruments are deposited in their holders.
You will be prompted to choose between PURGE/APURGE and SANITATION*.
Select APURGE with the
flash.
– or + key. APURGE must
59 57 928 D 3370
D 3370.201.01.15.02
*. The SANITATION function is described in the instructions
10.5 Purge / Auto Purge function (purging the water paths)Operating Instructions C2
If you press quick setting key 1 briefly (< 3 s), the following prompt will appear if all instruments are in their holders:
+
SELECT ITEMS
FOR PURGING
1
wrong
SELECT ITEMS FOR PURGING
Remove the instruments to be purged from the dentist
and assistant elements and insert them in the adapters
of the sanitation bowl as far as they will go.
CAUTION
Be careful not to snap off the instrument hoses when inserting the instruments.
The maximum water flow must be set on the instruments.
Press quick setting key 1 again (< 3 s), the instruments
are then purged one after the other.
The purging time in seconds set previously in the Mode
dialog (11.14) runs down for each instrument individually.
An audible signal sounds after the last instrument has
been flushed.
You now can continue in either of the following ways:
•
Leave the instruments in the sanitation bowl:
right
The treatment center can then be switched off.
The instruments will automatically be purged as
soon as the treatment center is switched on
again.
The APURGE function remains activated.
•
The instruments are returned to their holders:
The treatment center can be prepared for daily
practice operation.
The APURGE function is completed.
i
NOTICE
When removing the sanitation bowl, please notice that it
is filled with water.
52D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH11 Basic treatment center settings in the Mode dialog
A
NACL – PUMP
OFF
A
NACL – PUMP
ON FOR HANGER 2
Operating Instructions C2
+
11.1 Mode key
11 Basic treatment center settings in
the Mode dialog
11.1Mode key
After pressing the Mode key, you can call up and change
various basic treatment center settings in a specific
order.
They usually apply both to user A and to user B.
The corresponding display disappears automatically
after approx. 10 seconds. You can also exit the Mode
dialog more quickly by pressing the Mode key for > 3
seconds.
The flashing text is selected and is saved on exiting the
Mode dialog.
The desired setting can be selected by pressing the
+ key.
Key lock
The key lock can be activated and deactivated again by
pressing the - and + keys simultaneously.
The key lock affects the entire control panel of the dentist
element
i
NOTICE
The key lock can be deactivated only as described above. Switching the treatment center off and on again has
no effect on this state.
11.2Mode: NACL PUMP OFF or ON for instrument holder
Saline solution for the desired burr instrument
or the SIROSONIC L scaler (instrument holder
2 − 6)
NACL – PUMP
OFF
In this mode, you can preselect the NaCl pump for a
specified instrument using the
– / + keys.
– or
NACL – PUMP
ON FOR HANGER 2
59 57 928 D 3370
D 3370.201.01.15.02
For preparation of the NaCl pump, see page 41.
i
NOTICE
NACL appears on the display in the instrument dialog of
the instrument to which the NaCl pump has been assigned.
53
11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
A
MAINTENANCE
360 8
A
SERVICE ONLY
START – STOP
A
FC CONTROL MODE
ON/OFF – VARIABLE
A
TIME – HOURS : MIN
11 : 21
11.3 Mode: MAINTENANCEOperating Instructions C2
11.3Mode: MAINTENANCE
Display of next maintenance date
This dialog displays the number of days until the next
maintenance date. You should have maintenance per-
MAINTENANCE
360 8
formed by your service engineer at least once annually.
The instrument running time is indicated in the lower
right-hand corner of the display (for service engineers
only).
As soon as the time remaining until the next maintenance date is less than 30 days, the message is automatically displayed each time the unit is switched on.
The display message then disappears as soon as an
instrument is removed from its holder.
11.4Mode: SERVICE ONLY START - STOP
For service engineers only!
+
SERVICE ONLY
START – STOP
11.5Mode: FC CONTROL MODE ON/OFF − VARIABLE
Direct starter or speed controller foot switch
Preselection of the foot switch control mode for instrument start.
FC CONTROL MODE
ON/OFF – VARIABLE
With direct starter presetting (ON/OFF), the bar display always jumps immediately to the maximum value
after actuating the foot switch. You work with the displayed maximum value.
With speed controller foot switch presetting (VARI-ABLE), the bar display depends on the pedal position.
You work in the range between 0 and the displayed maximum value.
This setting can be chosen individually for user A or
B!
11.6Mode: TIME − HOURS : MIN
TIME – HOURS : MIN
11 : 21
54D 3370.201.01.15.02
Setting the time
Here you can set the time of day in hours and minutes.
The hours display initially flashes, and can be decremented or incremented with the – / + keys.
When the Mode key is pressed again, the minute display
starts flashing and can be adjusted in the same way.
59 57 928 D 3370
Sirona Dental Systems GmbH11 Basic treatment center settings in the Mode dialog
A
DAY : MON : YEAR
21 : JUN : 00
A
FOOT – CONTROL
NORMAL – EXTENDED
A
BOWL FLUSH ON S
YES – NO
A
CUP FILL ON S
YES – NO
Operating Instructions C2
+
11.7 Mode: DAY : MON : YEAR
11.7Mode: DAY : MON : YEAR
Setting the date
With this display you can set the year, month and day.
DAY : MON : YEAR
21 : JUN : 00
The setting is as for Mode
TIME − HOURS : MIN in the order of year, month, day.
The date is not displayed during operation of the unit.
It is required only for internal treatment center functions,
e.g. for service information.
11.8Mode: FOOT – CONTROL NORMAL − EXTENDED
Switching the cursor on/off
If you want to operate the treatment center only manually, you can switch the cursor control off.
FOOT – CONTROL
NORMAL – EXTENDED
The then changed functions of the relevant foot switch
are described in the chapter entitled “Foot switch”.
This setting can be chosen individually for user A or
B!
11.9Mode: BOWL FLUSH ON S YES − NO
Cuspidor flushing for chair program S
In program S of the treatment chair, cuspidor flushing
switches on.
BOWL FLUSH ON S
YES – NO
11.10Mode: CUP FILL ON S YES − NO
CUP FILL ON S
YES – NO
59 57 928 D 3370
D 3370.201.01.15.02
Tumbler filling function with chair program S
In program S of the treatment chair, the tumbler is filled.
55
11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
A
TAKE VCAM
A
WHITE BALANCE
HOME – USER – EXIT
11.11 Mode: WHITE BALANCE (only with an integrated SIROCAM 3)Operating Instructions C2
11.11Mode: WHITE BALANCE (only with an integrated SIROCAM 3)
White balance for SIROCAM 3
This Mode dialog is only displayed in connection with an
integrated SIROCAM 3.
The camera handpiece contains data making it possible
to restore the factory-set color definitions regardless of
the dental treatment center where the camera is used.
This makes it possible to use the camera at different
dental treatment centers.
For adjustments, e.g. if you wish to set individual colors,
you must perform a while balance.
+
WHITE BALANCE
HOME – USER – EXIT
2
Select the "REMOVE VCAM" display in the Mode dialog.
When the camera has been removed, the dialog used to
perform the white balance is displayed.
There are 3 options available:
USER - individual white balance
HOME - factory settings (can be restored at any
time)
EXIT- to exit the WHITE BALANCE dialog.
The display shows the current camera status, i.e., if an
individual white balance was performed the last time,
USER is displayed in bold type.
If the factory settings are valid, HOME is displayed in
bold type.
The selected item is flashing.
Select USER with the
The slide switch (2) of the camera must be set to its middle position (overview exposure).
The camera then must be held so that its lens is located
2 – 5 cm above a color pattern sheet of your own choice
(A) in the live mode.
– / + keys, USER starts flashing.
56D 3370.201.01.15.02
A
The screen must be completely filled by the focussed
color area.
•Press the Mode key, the white balance is started
(takes several seconds).
If the white balance has been completed successfully,
USER is displayed in bold type and an acoustic confirmation signal is issued.
If the white balance is not successful, USER will not be
displayed in bold type. White balance may be repeated
as often as necessary.
59 57 928 D 3370
Sirona Dental Systems GmbH11 Basic treatment center settings in the Mode dialog
A
TAKE INSTRUMENT
A
MAX LIGHT VOLT
3.9 V WARNING
Operating Instructions C2
+
11.12 Instrument settings via Mode dialog: REMOVE INSTRUMENT.
Different color patterns will produce different results with
respect to the color settings. For example, a color area
with a high amount of blue will reduce the amount of blue
color in the video image.
If you want to reset the camera to its factory settings,
select HOME with the
pressing the Mode key, HOME is displayed in bold type.
To exit this Mode dialog, select EXIT and press the
Mode key or put the camera back in place. Automatic
exit from the dialog is disabled as soon as USER or
HOME is selected. This allows you to repeat the white
balance as many times as necessary until you obtain the
desired result.
– / + keys. When you confirm by
11.12Instrument settings via Mode dialog: REMOVE INSTRUMENT.
This Mode dialog enables the following settings
• Instrument light voltage
and
• Maximum torque of SL motors
MAX LIGHT VOLT
3.9 V WARNING
Setting the instrument light voltage
You can change the voltage setting for instruments in
this Mode dialog.
First a message prompting removal of an instrument
appears.
Then the current voltage setting of the (removed) instrument is displayed.
Using the -/+ keys, you can adjust the instrument light
voltage from 3.0 to 5.6V in steps of 0.1V.
Starting at 3.9V, WARNING also appears.
After depositing the instrument, please remove and set
the next instrument.
Observe the voltage specified by the instrument manufacturer!
i
NOTICE
The voltage (3.6V) is already preselected for Sirona instruments.
59 57 928 D 3370
D 3370.201.01.15.02
57
11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
A
TAKE INSTRUMENT
A
MAX. TORQUE
YES – NO
A
# FUNCTION
TOGGLE – MOMENTARY
A
PURGE TIME
120 SEC
A
PRESSURE
PSI/BAR
11.13 Mode: # FUNCTION TOGGLE - MOMENTARYOperating Instructions C2
Setting the torque of the electric motors
This Mode dialog is used to set the automatic controller
action of the SL or SL ISO electric motors. Individual settings are possible for each motor holder.
First a message prompting removal of an instrument
appears.
After removing the instrument, select the setting dialog
by pressing the Mode key again:
MAX TORQUE
+
MAX. TORQUE
YES – NO
If Yes is flashing:
The motor will attain a higher torque at speeds > 15,000
rpm.
If No is flashing:
The running smoothness of the motor will be higher at
speeds > 15,000 rpm.
i
NOTICE
The new setting becomes effective only after the corresponding motor is removed again.
11.13Mode: # FUNCTION TOGGLE − MOMENTARY
Free function key # as button or switch
In this mode you preselect whether you want to use the
# as a button or as a switch (see page
# FUNCTION
TOGGLE – MOMENTARY
free function key
15 ).
11.14Mode: PURGE TIME... SEC
PURGE TIME
120 SEC
11.15Mode: PRESSURE PSI / BAR
PRESSURE
PSI/BAR
58D 3370.201.01.15.02
Setting the purge time for purging the water
paths
(Only for dentist element)
The purging time per instrument can be preset from 60
to 120 seconds with the
tion (see page 50).
Indication of pressure in PSI or bar
This mode is used to set the driving air pressure display
of the highspeed handpiece to your preferred unit of
measurement, i.e., PSI or bar.
– /+ keys for the purge func-
59 57 928 D 3370
Sirona Dental Systems GmbH11 Basic treatment center settings in the Mode dialog
A
SPRACHE/LANGUAGE
ENG/D/I/F/E
A
SIROLUX U – POS.
ON – OFF
A
POLYLUX WITH CFS
YES – NO
A
SUCTION WITH CFS
YES – NO
Operating Instructions C2
+
11.16 Mode: SPRACHE / LANGUAGE ENG / D / I / F / E
11.16Mode: SPRACHE / LANGUAGE ENG / D / I / F / E
Language setting
In this mode, you can select the desired display language: ENG=English, D=German, I=Italian, F=French
SPRACHE/LANGUAGE
ENG/D/I/F/E
and E=Spanish.
11.17Mode: SIROLUX U - POS. ON – OFF
SIROLUX in the examination position
If ON is flashing, the SIROLUX FANTASTIC operation
light is switched on automatically when you bring up the
SIROLUX U – POS.
ON – OFF
dentist element with the foot switch (with instruments
placed in their holders). (see page 33).
11.18Mode: POLYLUX WITH CFS YES – NO
This dialog has no function in connection with the Mini
LED curing light.
Select NO.
POLYLUX WITH CFS
YES – NO
11.19Mode: SUCTION WITH CFS YES – NO
Interrupting or restarting the suction flow with
the 4-way foot control
(function available on units as of serial no. 52025)
SUCTION WITH CFS
YES – NO
If YES is flashing, the suction flow of the large suction
hose can be interrupted or restarted (not at the saliva
ejector hose) by pushing the 4-way foot control on the
chair in any direction. After depositing the large suction
hose in the interrupted condition, the suction flow
switches back on automatically after renewed removal.
59 57 928 D 3370
D 3370.201.01.15.02
59
11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
Preselecting a white screen for the SIVISION
monitor
If YES flashes –
WHITESCREEN
YES – NO
The “X-ray image viewer” key is used to switch the X-ray
image viewer and the WHITE SCREEN on the SIVISION
3 monitor ON/OFF.
If NO flashes –
The “X-ray image viewer” key only switches the X-ray
image viewer ON and OFF.
With the X-ray image viewer and SIVISION 3 installed,
always select “NO”.
11.21Mode: A/B VIDEO MODE ON – OFF
Locking the video mode switchover
+
A / B VIDEO MODE
ON – OFF
A/B
11.22Mode: BLOW OUT SPRAY
BLOW OUT SPRAY
YES – NO
This Mode is only displayed in connection with an integrated SIVISION 3.
If ON is flashing –
You can use the “A/B” key to switch between the VIDEO
mode and the PC mode when the SIVISION 3 camera is
removed.
If OFF is flashing –
The “A/B” key can no longer be used to switch between
the VIDEO and the PC mode.
This prevents the selected mode from being changed
inadvertently.
Prevents drop formation
You can activate the instrument blow-out function in this
setting option. This removes any residual water (one to
two drops) by blowing air through the instrument heads.
Just press the
OFF.
The selected item flashes.
– or + key to switch this function ON or
60D 3370.201.01.15.02
i
NOTICE
This function is switched OFF in the factory setting.
i
NOTICE
Sirona recommends activating this function especially
for the C3+ treatment center.
59 57 928 D 3370
Sirona Dental Systems GmbH12 Assistant eleme nt
A
SUCTION WITH CFS
YES – NO
Operating Instructions C2
+
12.1 Suction handpieces on the assistant element
12 Assistant element
12.1Suction handpieces on the assistant element
The vacuum system is switched on after a suction hose
is removed from the holder.
CAUTION
Observe the warning and safety information on page 9
(vacuum system).
The suction handpiece can be angled by turning it.
CAUTION
If the vacuum system is switched off unintentionally, remove the suction cannula from the patient’s mouth immediately.
If the unit has no second saliva ejector hose for the surgical cannula, the thick suction hose must be used for
surgical aspiration.
SUCTION WITH CFS
YES – NO
To insert a surgical cannula, please attach the adapter
supplied.
If the unit is equipped with a second saliva ejector hose,
the surgical cannula can be inserted directly into the
saliva ejector handpiece.
The suction flow of the large suction hose can also be
interrupted or restarted by pushing the 4-way foot control
on the chair in any direction.
You can preselect this in Mode 11.19. After depositing
the large suction hose in the interrupted condition, the
suction flow switches back on automatically after
renewed removal.
(function available on units as of serial no. 52025)
59 57 928 D 3370
D 3370.201.01.15.02
61
12 Assistant element Sirona Dental Systems GmbH
12.2 SPRAYVIT on the assistant elementOperating Instructions C2
CAUTION
Make sure to hold the suction cannula in such a way that
the cannula opening cannot attach itself to the oral mucous membranes by accident.
i
NOTICE
The factory-set suction power can be set in the water unit
by a service engineer.
12.2SPRAYVIT on the assistant element
If the SPRAYVIT is the only instrument removed, the
instrument light can be switched on or off.
L
Its brightness is set to an optimum value and cannot be
changed.
+
The instrument light switches on when the SPRAYVIT is
activated (if preselected).
The instrument light is switched off after a time lag of 10s
when the SPRAYVIT is no longer activated.
When the SPRAYVIT is deposited in its holder, the
instrument light switches off immediately.
DANGER
After changing hoses, press the water key of the
SPRAYVIT repeatedly until water flows out of it! Only
then can you begin treatment.
12.3Satalec Mini LED curing light on the assistant element
S
0
S
A
N
1
2
L
For safety instructions and operation, see pages 42 ff.
62D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH12 Assistant eleme nt
Operating Instructions C2
+
12.4 Manual travel track
12.4Manual travel track
You can shift the position of the assistant element.
59 57 928 D 3370
D 3370.201.01.15.02
63
13 Water unit Sirona Dental Systems GmbH
13.1 Swiveling the cuspidorOperating Instructions C2
13 Water unit
13.1Swiveling the cuspidor
The cuspidor on the water unit can be swiveled manually
approx. 110mm toward the patient chair.
Collision protection
Certain chair movements are locked with the cuspidor
swiveled in. If a programmed movement S, 0, 1 or 2 is
started, then the cuspidor first moves back into its starting position before the chair moves into the programmed
position.
When a key is actuated to individually set the chair
height adjustment up/down or the backrest tilt to the
back/front, the cuspidor swivels back first. Then, if the
corresponding key is still pressed, the chair moves.
You can make individual adjustments of the headrest
even with the cuspidor swung in.
If the cuspidor is swiveled in while the chair is moving,
the program is then aborted. This is signaled by a double
beep.
+
13.2Adjusting the water amount for the cuspidor
•Open the flap (1) on the base of the water unit.
•Adjust the amount of water with the valve (2).
1
2
64D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH13 Water unit
Operating Instructions C2
+
13.3 Amalgam rotor
13.3Amalgam rotor
The message AMALG appears on the display if the
amalgam rotor needs replacement, i.e., when having
reached a filling level of 95% (see instructions “Care and
Cleaning by the Practice Team”).
A
AMALG
RPM
A continuous sound indicates that the rotor is full and no
longer functioning properly.
An intermittent sound indicates a system error. The
water supply is interrupted. Please contact your service
engineer.
i
NOTICE
The rotor inside the amalgam separator must be replaced at least once a year, even if the AMALG message
does not light up. Otherwise, unpleasant odors can develop.
59 57 928 D 3370
D 3370.201.01.15.02
65
13 Water unit Sirona Dental Systems GmbH
13.4 Disinfection unit (optional)Operating Instructions C2
13.4Disinfection unit (optional)
If the message DESINF starts flashing on the dentist element display during treatment, then 1 liter of DENTOSEPT P should be refilled not later than the end of the
treatment day.
A
DESINF
RPM
If the message is flashing, it is not possible to perform
sanitation.
Remove the cover behind the cuspidor and fill in 1 liter
of DENTOSEPT P.
The filler trough should always be kept clean. Always
make sure that no tap water enters the filler trough!
Microbiological water test (after an operating pause of
more than 1 week)
+
Sanitation of the unit
Perform sanitation every 4 weeks. This is the only way
to effectively combat the formation of biofilm in the water
paths (see instructions “Care and Cleaning by the Practice Team”).
CAUTION
In Japan the disinfectant OXYDOL available there must
be used instead of DENTOSEPT P.
OXYDOL must be mixed 1:1 with distilled water before it
is filled into the disinfection unit.
66D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH14 SIROTOM HF electrosurgery
Operating Instructions C2
+
14.1 Safety
14 SIROTOM HF electrosurgery
14.1Safety
CAUTION
Incorrect operation and the nonobservance of precautionary measures can cause serious accidents when
working with the electrosurgical unit.
DANGER
There is a risk in the case of patients with cardiac
pacemakers that the pacemaker can be disturbed
duringreatmentwiththeelectrosurgical instrument.
We recommend not using the electrosurgical unit for
these patients.
Further WARNINGS for working with the
surgical unit
•The patient should not come into contact with metallic parts which are grounded or have substantial capacitance to ground.
•The power output should be set to the lowest possible value for the relevant purpose.
•If the surgical unit appears to deliver little power or
does not work properly in its normal setting, this may
be caused by a poor contact in the supply cable.
•Combustible substances, which are used for instance as cleaning or disinfecting agents, should
have evaporated before applying surgery. Cotton
wool may ignite.Endogenous gases may ignite.
•Contact between the HF handpiece and metal implants or supraconstructions must be avoided.
•The HF can impair the function of other electronic
devices.
•Test the electrode cable regularly for possible damage to the insulation.
•If a line voltage fade occurs, the HF electronics
switch off automatically. Deposit the handpiece
briefly. Then you can continue working as usual.
59 57 928 D 3370
D 3370.201.01.15.02
67
14 SIROTOM HF electrosurgery Sirona Dental Systems GmbH
14.2 OperationOperating Instructions C2
14.2Operation
The electrosurgical handpiece is used for electrotomy
(cutting), coagulation and desiccation in biterminal technique.
CAUTION
The patient must always hold the neutral electrode in his
hand during this procedure!
The connection socket is located on the bottom side of
the dentist element.
The handpiece is started by actuating the foot switch. In
this case an interrupted acoustic signal sounds.
CAUTION
Active
electrode
For interference suppression reasons, the handpiece
may be used only for a few seconds.
+
Neutral
electrode
If the foot switch is pressed for more than one minute,
the device switches off automatically. The activation
must be interrupted.
Let go of the foot switch before depositing the HF instrument in its holder. The programmed values of the previously selected user level appear again the next time the
handpiece is removed.
68D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH14 SIROTOM HF electrosurgery
Operating Instructions C2
+
14.3 Technical description
14.3Technical description
Power characteristic
Watt
50
40
30
20
10
Watt
50
40
30
20
5001000
Intensity=100
Intensity=50
1500Ohm
2000
R=500Ω
for cutting and coagulation
Power measured between handpiece and protective
ground wire.
Power curve
for cutting and coagulation
Power measured between handpiece and protective
ground wire.
10
10
20
40
30
50
14.4Technical data
6070
Intensity
8090100
Maximum output voltage peak – peak between handpiece and protective ground wire when cutting: 1500 V
Frequency of operating and alarm tone: 1200Hz
The neutral electrode is connected to the protective
ground wire through a capacitor.
Connection for the neutral electrode.
Neutral electrode at high frequency relative to ground.
The SIROTOM handpiece is a type BF application part
59 57 928 D 3370
D 3370.201.01.15.02
69
14 SIROTOM HF electrosurgery Sirona Dental Systems GmbH
14.5 Safety checksOperating Instructions C2
+
14.5Safety checks
DAB 25% ED 10s SD 40s
1MHz / 50W / 500Ohm
Intermittent operation25%
Duty time10s
Cycle time40s
HF frequency1MHz
Output powermax. 50W across 500
ohms
Symbol for non-ionizing radiation
In Germany, medical devices are subject to the provisions of the Ordinance on the Installation, Operation and
Use of Medical Devices (Medizinprodukte-Betreiberverordnung –
formed and a medical product log must be kept.
For details please refer to Section 18.3 Safety checks.
MPBetreibV). Safety checks must be per-
70D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
+
15.1 System overview of SIVISION 3
15 SIVISION 3
15.1System overview of SIVISION 3
SIVISION on the lamp support tube
• Rotatable
•swivelable
Monitor on tray arm
• Rotatable
•swivelable
59 57 928 D 3370
D 3370.201.01.15.02
Monitor on swivel arm
• Rotatable
•swivelable
• Height adjustable
71
15 SIVISION 3 Sirona Dental Systems GmbH
15.1 System overview of SIVISION 3Operating Instructions C2
CAUTION
Make sure that the patient does not collide with the monitor during movements of the dental treatment chair
and/or operation of the cuspidor. Swivel the monitor out
of the collision area if necessary.
SIVISION 3 monitor versions
• 15“ Monitor
• with glass pane
+
• 19“ Monitor
• with glass pane
• with loudspeaker
CAUTION
The loudspeaker socket of the monitor may be connected only to a device which complies with IEC 60950 (e.g.
PC) or IEC 60601-1, and under no circumstances e.g. to
a stereo system etc.
72D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
+
15.2 Camera versions
15.2Camera versions
SIROCAM 3
• Integrated 4-view image memory
• 3-fold adjustable focusing range
• Titanium case
SIROCAM C
• Integrated 4-view image memory
• 2-fold adjustable focusing range
• Plastic case
59 57 928 D 3370
D 3370.201.01.15.02
73
15 SIVISION 3 Sirona Dental Systems GmbH
15.3 SIROCAM 3 intraoral camera (additional equipment)Operating Instructions C2
15.3SIROCAM 3 intraoral camera (additional equipment)
The SIROCAM 3 intraoral camera in conjunction with
SIVISION 3 is a perfect system for effective communication with the patient.
+
4
3
*
t 1 / t 2
2
1
The camera images cannot be used for diagnosis!
The SIROCAM 3 intraoral camera is a sensitive optical
instrument and must therefore always be handled with
care.
The camera illuminates the parts of the teeth to be
exposed with 7 white-light LEDs (1).
These are arranged around the entrance prism (2).
The LEDs and the lens are protected by a scratch-proof
antiglare glass cover.
CAUTION
* = Intermittent camera operation:
t1= 1 min ON, t2 = 3min OFF.
The camera heats up during operation.
To prevent overheating of the camera, the camera must
be deposited for a time interval t1 = 3 min after an operating time of t2 = 3 min.
The camera will heat up even when it is placed in its
holder, as one of the states "black screen", "white
screen" or "quad image" is always active.
The slide switch (3) is used for presetting the depth of
focus.
The camera button (4) is intended for controlling different
SIVISION 3 functions.
i
NOTICE
For camera operation and further information, please refer to the Operating Instructions for the SIROCAM 3.
i
NOTICE
The camera handpiece contains data making it possible
to restore the factory-set color definitions regardless of
the dental treatment center where the camera is used.
This makes it possible to use the camera at different dental treatment centers. To customize the color settings,
you can perform a white balance as long as the camera
is active, (see Mode dialog, Section 11.11).
74D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
1
+
15.3 SIROCAM 3 intraoral camera (additional equipment)
Hygienic protective covers for SIROCAM 3
Hygienic protective covers are available for the camera.
They can be ordered under REF 59 38 654. Packaging
unit: 500 pcs.
How to use the hygienic protective covers:
1. Hold the hygienic protective cover with its paper side
face-down and insert the camera with the optical
system downward into the cover.
2. Pull off the upper protective foil with the blue stripe.
2
3
4
3. Then pull off the backing paper.
4. Finally, stretch the hygienic protective cover over
the camera from front to rear.
CAUTION
In order to achieve optimum image quality, there must
not be any folds in the transparent area of the cover
above the lens window.
CAUTION
Use only the hygienic protective covers offered by
Sirona.
Prior to each application on a new patient, the camera
must be covered with a new hygienic protective cover.
59 57 928 D 3370
D 3370.201.01.15.02
75
15 SIVISION 3 Sirona Dental Systems GmbH
15.4 SIROCAM C intraoral camera (additional equipment)Operating Instructions C2
15.4SIROCAM C intraoral camera (additional equipment)
The SIROCAM C intraoral camera in conjunction with
SIVISION 3 is a perfect system for effective communication with the patient.
The camera images cannot be used for diagnosis!
3
*
t 1 / t 2
2
1
The SIROCAM C intraoral camera is a sensitive optical
instrument and must therefore always be handled with
care.
The camera illuminates the parts of the teeth to be
exposed with 6 white-light LEDs (1).
These are arranged around the single-lens optical system (2).
The LEDs and the lens are protected by a scratch-proof
antiglare glass cover.
CAUTION
* = Intermittent camera operation:
t1= 1 min ON, t2 = 3min OFF.
The camera heats up during operation.
To prevent overheating of the camera, the camera must
be deposited for a time interval t1 = 1 min after an operating time of t2 = 3 min.
+
When the camera is in LIVE IMAGE mode, it is possible
to toggle between close range and far range by pressing
button (3) on the camera.
Button not pressed – close range 5 – 15mm
Button pressed – far range from 15mm to smile line.
76D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
+
15.5 Operating modes of SIVISION 3
Hygienic protective covers for SIROCAM C
Hygienic protective covers are available for the camera.
They can be ordered under REF 59 14 705. Packaging
unit: 500 pcs.
How to use the hygienic protective covers:
1. Hold the hygienic protective cover with its paper side
face-down and insert the camera with the optical
system downward into the cover.
2. Pull off the upper protective foil with the blue stripe.
3. Then pull off the backing paper.
15.5Operating modes of SIVISION 3
4. Finally, stretch the hygienic protective cover over
the camera from front to rear.
CAUTION
In order to achieve optimum image quality, there must
not be any folds in the transparent area of the cover
above the lens window.
CAUTION
Use only the hygienic protective covers offered by
Sirona.
Prior to each application on a new patient, the camera
must be covered with a new hygienic protective cover.
SIVISION 3 with SIROCAM 3 or SIROCAM C can be
used basically in two different operating modes:
59 57 928 D 3370
D 3370.201.01.15.02
•VIDEO mode
The camera image is directly displayed on the
SIVISION monitor.
The quad image memory of the camera is activated.
•PC mode
The camera image is first processed on the PC and
then the image generated on the PC is displayed on
the SIVISION monitor.
77
15 SIVISION 3 Sirona Dental Systems GmbH
A
A / B VIDEO MODE
ON – OFF
15.5 Operating modes of SIVISION 3Operating Instructions C2
SWITCHING TO PC MODE
+
A
VIDEOCAM
PC MODE
A
VIDEOCAM
VIDEO MODE
A/B key on the panel
Only:
•with the camera removed and active,
•manually or by cursor (using the foot switch),
“VIDEOCAM” and the operating mode “PC MODE”
now appear on the display.
The image from the PC monitor, e.g. SIDEXIS/VI-
DEXIS, appears on the SIVISION monitor. The camera switches to LIVE IMAGE.
This mode remains active even if the treatment center is
switched OFF and back ON.
SWITCHING TO VIDEO MODE
A/B key on the panel
Only:
•with the camera removed and active,
•manually or by cursor (using the foot switch),
78D 3370.201.01.15.02
A/B
“VIDEOCAM” and the operating mode
“VIDEO MODE” now appear on the display.
The SIVISION monitor displays the camera image.
This mode remains active even if the treatment center is
switched OFF and back ON.
This switchover function of the "Mode switchover" key
can be disabled in Mode dialog 11.21Mode: A/B VIDEO
MODE ON – OFF to prevent inadvertent switching.
A / B VIDEO MODE
ON – OFF
59 57 928 D 3370
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
+
15.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)
15.6SIROCAM 3 / SIROCAM C integrated in the dentist element
(video)
Operation in video mode
A
VIDEOCAM
VIDEO MODE
+
C2
4
0
1
0
0
3
0
7
5
SS
00
2
0
5
0
–
M
o
d
e
+
11
1
0
22
2
25
5
A/B
0
.2
r
p
m
x
1
1
0
0
0
In this version the SIROCAM 3 / SIROCAM C intraoral
camera is positioned in holder 6 of the dentist element
or in an additional holder on the dentist element.
The camera can be operated by the unit foot switch if no
other instrument is active (removed and operable by foot
switch) when the camera is removed.
If the camera is the active instrument “VIDEOCAM”
and the operating mode “VIDEO MODE” appear on
the display.
59 57 928 D 3370
D 3370.201.01.15.02
79
15 SIVISION 3 Sirona Dental Systems GmbH
15.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)Operating Instructions C2
SAVE IMAGE
In VIDEO mode, the camera can save 4 still images consecutively (identified with I, II, III and IIII).
+
A 5th view shows a QUAD IMAGE in which these 4 still
images are tiled on the screen.
Remove the camera
If a STILL IMAGE or QUAD IMAGE was displayed with
the camera deposited, this view is retained.
If a BLACK SCREEN or a WHITE SCREEN was displayed with the camera deposited, the view preceding
the BLACK or WHITE SCREEN, i.e. a LIVE, STILL or
QUAD IMAGE, appears.
80D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
C
+
+
15.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)
TOGGLING BETWEEN STILL IMAGE AND LIVE
IMAGE
C+ or C foot switch
With the camera removed, the cursor is in the instruments key block.
With the C+ foot switch:
Step on foot pedal.
With the C foot switch:
C
Actuate foot switch in direction A – cursor upwards.
Toggling between a STILL IMAGE and a LIVE IMAGE.
C
+
C
IMAGE SWITCHING
Spray ON / OFF key on the panel
Manually or by cursor (using the foot switch) with instruments deposited or camera removed.
With C+ foot switch, also left foot switch button (only with
camera removed).
i
NOTICE
If the camera is in the LIVE IMAGE mode, the LIVE
IMAGE is stored in the current memory location during
frame indexing (image switching).
With the camera deposited, a BLACK SCREEN is stored
in the current memory location on frame indexing and
with an empty memory.
In the SINGLE IMAGE mode, switching to the next memory location.
•If a STILL IMAGE is stored there, it will be displayed.
•If no STILL IMAGE is stored there, a LIVE IMAGE
will appear with the camera removed.
In QUAD IMAGE mode, switching to memory location I,
II, III or IIII for image selection.
59 57 928 D 3370
D 3370.201.01.15.02
The number of the currently selected image lights up
in the upper right corner of the image.
81
15 SIVISION 3 Sirona Dental Systems GmbH
15.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)Operating Instructions C2
TOGGLING BETWEEN QUAD IMAGE AND
SINGLE IMAGE
Chip blower key on the control panel
Manually or by cursor (using the foot switch) with instruments deposited or camera removed.
With C+ foot switch also right foot switch button (only
with camera removed).
+
C
+
C
Toggling between QUAD IMAGE and SINGLE IMAGE.
When switching to a SINGLE IMAGE, the image previously selected in the QUAD IMAGE is displayed.
CLEARING THE MEMORY
Electric motor CCW rotation key on the control
panel
Only with the camera removed and active, manually or
by cursor (using the foot switch).
Clears all image memories of the camera.
The LIVE IMAGE for image I then appears.
59 57 928 D 3370
82D 3370.201.01.15.02
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
+
15.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)
Deposit the camera
+
C
2
40
100
30
75
SS
00
20
50
–
M
o
d
e
+
11
10
22
25
2
5
A/B
0.2
r
p
m
x
1
1
0
0
0
•If a LIVE IMAGE was displayed, a BLACK SCREEN
appears.
C
+
C
•STILL IMAGE and QUAD IMAGE, BLACK
SCREEN and WHITE SCREEN remain displayed.
IMAGE SWITCHING WITH THE CURSOR
Special function only with cursor:
The cursor is in the chair programs key block, level 1.
With the C
+
foot switch:
4-way foot control plate forward.
With the C foot switch:
Actuate foot switch in direction A – cursor upwards.
•Switching images:
STILL IMAGE I, STILL IMAGE II, STILL IMAGE III,
STILL IMAGE IIII, QUAD IMAGE, STILL IMAGE I,
STILL IMAGE II ...
59 57 928 D 3370
D 3370.201.01.15.02
83
15 SIVISION 3 Sirona Dental Systems GmbH
A
WHITESCREEN
YES – NO
15.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)Operating Instructions C2
WHITE SCREEN
X-ray image viewer ON / OFF key on the control
panel
•Switch X-ray image viewer or WHITE SCREEN on
the SIVISION monitor ON/OFF.
Presetting possible in the Mode menu (11.20) :
•Only X-ray image viewer ON/OFF if NO flashes
or
•X-ray image viewer and WHITE SCREEN on the
SIVISION monitor ON/OFF if YES flashes.
i
NOTICE
WHITESCREEN - NO should be selected if an X-ray image viewer is installed. If no X-ray image viewer is installed, select WHITESCREEN - YES.
+
WHITESCREEN
YES – NO
BLACK SCREEN
“100” quick setting key on the control panel
Manually or by cursor (using the foot switch) with instruments deposited or camera removed. Cursor forward if
the cursor LED next to the quick setting keys lights up.
•Toggling between any view and a BLACK SCREEN.
•The previous view reappears when this key is
pressed again.
84D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
+
15.7 SIROCAM 3 / SIROCAM C – PC mode
15.7SIROCAM 3 / SIROCAM C – PC mode
A
VIDEOCAM
PC MODE
+
C
2
40
1
0
0
30
7
5
SS
00
20
5
0
–
M
o
d
e
11
1
0
+
22
2
25
5
A
/B
0
.2
r
p
m
x
1
1
0
0
0
Operation in PC mode using a PC connection
and SIDEXIS/VIDEXIS
The monitor at the treatment center is controlled by the
PC in this case. The functional scope of the PC can be
used in this mode.
If the camera is the active instrument “VIDEOCAM” and
the operating mode
“PC MODE” appear on the display.
Remove the camera
The LIVE IMAGE window opens.
C
+
C
TOGGLING BETWEEN STILL IMAGE AND LIVE
IMAGE
C+ or C foot switch
With the camera removed, the cursor is in the instru-ments key block.
With the C+ foot switch:
Step on foot pedal.
With the C foot switch:
Actuate foot switch in direction A – cursor upwards.
Toggling between a STILL IMAGE and a LIVE IMAGE.
59 57 928 D 3370
D 3370.201.01.15.02
85
15 SIVISION 3 Sirona Dental Systems GmbH
15.7 SIROCAM 3 / SIROCAM C – PC modeOperating Instructions C2
USING THE FUNCTION KEYS
Additional functions, e.g. SIDEXIS/VIDEXIS functions:
Tile, Next Image, Full Frame, Zoom In, Zoom Out,
Rotate Image or other PC controls can be assigned as
desired to the following function keys on the panel of the
dentist element:
•Spray
•Chip blower
•Electric motor CCW rotation
•Instrument light
•X-ray image viewer
•Call key (if not already used for relay functions)
•# key (if not already used for relay functions)
•Cursor forward, if the cursor is between Spray and
Chip blower
(With C foot switch: only if the camera is deposited)
•Cursor forward, if the cursor is between the S and 0
chair program keys
(With C+ foot switch – cursor forwards. With C foot
switch – Actuate foot switch in direction A – cursor
upwards)
With instruments deposited or camera removed:
+
•Quick setting keys 100, 75, 50, 25 and 1
Additional possibilities as well as the configuration procedure are described in the “Installation manual for PC software“, REF 59 11 719, Operator’s Manual.
The most important functions based on the factory settings are described in the following.
86D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
+
15.7 SIROCAM 3 / SIROCAM C – PC mode
SAVE IMAGE
Key example: Spray ON / OFF key on the panel
For this purpose, a patient must be registered in
SIDEXIS/VIDEXIS.
Manually or by cursor (using the foot switch) with instruments deposited or camera removed.
+
With C
camera removed).
foot switch also left foot switch button (only with
C
+
C
•Save STILL IMAGE.
A LIVE IMAGE then appears with the camera removed.
Depositing the camera
+
C2
4
0
1
0
0
30
7
5
SS
00
2
0
5
0
–
M
o
d
e
+
11
1
0
22
2
2
5
5
A/B
0
.2
r
p
m
x
1
1
0
0
0
59 57 928 D 3370
D 3370.201.01.15.02
• LIVE IMAGE window is closed.
Displayed STILL IMAGE or saved images remain
open.
87
15 SIVISION 3 Sirona Dental Systems GmbH
A
WHITESCREEN
YES – NO
15.7 SIROCAM 3 / SIROCAM C – PC modeOperating Instructions C2
WHITESCREEN
X-ray image viewer ON / OFF key on the panel
•Switch X-ray image viewer or WHITE SCREEN
on the SIVISION monitor ON/OFF.
•Only X-ray image viewer ON/OFF if NO flashes
or
•X-ray image viewer and WHITE SCREEN on the
SIVISION monitor ON/OFF if YES flashes.
i
NOTICE
WHITESCREEN - NO should be selected if an X-ray image viewer is installed. If no X-ray image viewer is installed, select WHITESCREEN - YES.
+
WHITESCREEN
YES – NO
BLACK SCREEN
“100” quick setting key on the panel
Manually or by cursor (using the foot switch) with instruments deposited or camera removed. Cursor forward if
the cursor LED next to the quick setting keys lights up.
• Toggling between any view and a BLACK SCREEN.
• The previous view reappears when this key is pressed
again.
i
NOTICE
WHITE SCREEN/BLACK SCREEN is only possible
when “AUTO” is set on the monitor (see SIVISION 3 “Operating Instructions for the flat-screen monitor”)
88D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH15 SIVISION 3
Operating Instructions C2
+
15.8 SIVISION 3–second monitor function without SIROCAM3/SIROCAM C
15.8SIVISION 3–second monitor function without SIROCAM3/
SIROCAM C
The monitor integrated in the treatment center serves as
primary or secondary monitor for the externally connected PC.
The following function keys of the control panels as well
as foot switch functions are available for controlling PC
functions:
•Spray
•Chip blower
•Electric motor CCW rotation
•Instrument light
•Call key (if not already used for relay functions)
•# key (if not already used for relay functions)
•All quick setting keys 100, 75, 50, 25 and 1
•Cursor forward, if the cursor is between the
spray and chip blower keys
•Cursor forward, if the cursor is between the S
and 0 chair program keys
+
(With C
switch – Actuate foot switch in direction A – cursor
up.)
No instrument may be removed in this case!
foot switch – cursor forward. For C foot
as well as the function keys:
•X-ray image viewer
With instruments deposited:
They can be used to control e.g. SIDEXIS/VIDEXIS
functions such as Tile, Next Image, Full Frame, Zoom In, Zoom Out, Rotate Image, but also other PC applications.
The factory settings as well as the configuration procedure are described in the “Installation manual for PC software“, REF 59 11 719 Operator’s Manual.
59 57 928 D 3370
D 3370.201.01.15.02
89
16 CEREC Chairline (Option) Sirona Dental Systems GmbH
16.1 CEREC ChairlineOperating Instructions C2
16 CEREC Chairline (Option)
16.1CEREC Chairline
CEREC Chairline is used for 3-dimensional acquisition
of the preparation, the occlusion and the antagonist.
For details on its function and operation, please refer to
the CEREC Chairline Operating Instructions
(REF 60 46 028) and Operator Manual (REF 59 56 458).
+
CEREC Chairline foot switch
A Bluetooth radio module for transmitting the CEREC
foot switch signal to the CEREC PC is integrated in the
CEREC Chairline foot switch.
For details refer to Chapter 6.1.
Model designation of radio moduleBlue RS+I (Stollmann Co.)
Frequency2.4GHz - 2.4835GHz (ISM band)
Bluetooth class2
Rangeapprox. 10m
ApprovalsCE conformity according to R&TTE Directive (1999/5/
EC)
Part 15 of the FCC Rules,
RSS-210 of Industry Canada
90D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH17 Extra equipment and accessories
Operating Instructions C2
+
17.1 Media block on the patient chair
17 Extra equipment and accessories
17.1Media block on the patient chair
Additional equipment
Only approved additional devices (e.g. supplied by
EMS) may be connected to a media block on the base of
the lower part of the chair (the manufacturer's instructions must be complied with).
The media block is equipped with quick couplings for air
and water and an outlet for non-heating appliances with
a separate fuse (6.3A).
For treatment centers with a disinfection unit, the integration of a media block complies with the DVGW
approval (EN1717/ DIN1988). In this case, the additional devices connected are also compliant with the
requirements of EN1717 / DIN1988.
Fuse
Line voltage
230V
Water
Air
Technical data and requirements•All connected devices must be supplied with disin-
fected water if a disinfection unit is installed.
0.1‰ -0.2‰ hydrogen peroxide are required for regular operation and 1.4% for sanitization.
CAUTION
Before connecting any additional devices, first check
whether they are suitable for hydrogen peroxide.
If these devices are not suitable for the higher concentration used for sanitization, Sirona recommends separating them from the treatment center during sanitization
and then sanitizing them separately according to the
manufacturer's instructions.
Sirona shall not be held liable for any damage to connected devices!
Pressure and flow rate specifications
PressureFlow rate
Water2,2 ± 0.2 barmax. 300 ml/min
Air4,4 ± 0.5 barmax. 70 Nl /min
59 57 928 D 3370
D 3370.201.01.15.02
91
17 Extra equipment and accessories Sirona Dental Systems GmbH
Two different hygienic covers are available for the headrest of the patient chair:
•Washable textile headrest cover with Velcro closure
and neck pad
REF:59 05 588 (5 pcs.)
•Disposable headrest cover with a self-adhesive fastening
REF: 58 72 820 (5 pcs.)
Slip headrest cover over headrest and fasten it as illustrated.
17.3Children's headrest
Additional equipment
The children's headrest is held magnetically.
+
To attach it, first insert the two guide pins of the support
plate in the bellows of the headrest as shown.
Then attach the children's headrest.
•Children's head pad set for motor-adjustable headrest
REF: 58 94 907
92D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH17 Extra equipment and accessories
Operating Instructions C2
+
17.4 Seat cushion C
17.4Seat cushion C
Additional equipment
The back part is fastened with a Velcro closure and can
be removed.
Covering material: 100% cotton.
The covers can be removed after opening the Velcro
strap on the back.
The covers are Teflon-coated.
Instructions for care:
The Teflon coating is dirt-repellent and slightly
water-repellent; remove any dirt with lukewarm soap
suds (fine washing powder).
CAUTION
Removing stains with stain remover is not suitable, since
this damages the Teflon coating!
17.5Folding armrest
This comfortable seat is simply placed on top of the
upholstery of the patient chair. Thus it is possible to position the head even of small patients on the motor-adjustable headrest.
Additional equipment
If a folding armrest is attached to the patient chair, it can
be folded up or down by pressing both buttons on the
lower end of the armrest at the same time.
When it is swung up the armrest locks automatically into
its position for use.
59 57 928 D 3370
D 3370.201.01.15.02
93
17 Extra equipment and accessories Sirona Dental Systems GmbH
If you want to work with a hydrocolloid coolant supply,
you can obtain an adapter, Order No: 33 15 814, free of
1
3
2
charge with the enclosed voucher.
Please state your address and the serial number of the
chair (see “Installation Report / Warranty Passport”).
The adapter can be connected to the cuspidor as follows:
Pull out the tumbler outlet (1), turning it slightly at the
same time, and plug the adapter (2) into the self-closing
valve until it snaps into place. Press the locking clip (3)
before pulling out the connecting piece.
+
17.7Tray (additional equipment)
For water cooling please press the tumbler filling key.
The tumbler filling time must be set to unlimited HYDROCOLLOID operation in this case (see chapter 10.2 “Tumbler filling” on page 47).
CAUTION
During the hydrocolloid procedure the disinfection of the
tumbler water and the instruments is switched off.
In order to rinse residual non-disinfected water out of the
lines, please push the button for TUMBLER FILLING one
time before refilling the tumbler.
i
NOTICE
Maximum load: 1 kg
For further details, please refer to the operating instructions for the tray, REF 33 33 718.
94D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH17 Extra equipment and accessories
Operating Instructions C2
+
17.8 X-ray image viewer on the dentist element
17.8X-ray image viewer on the dentist element
Additional equipment
1
This X-ray image viewer is switched ON or OFF using
the shown key on the control panel of the dentist or
assistant element.
To view intraoral dental images the anti-glare film (1)
must first be attached with two clips. Then fasten the
+
C
2
40
1
00
30
75
SS
2
0
–
M
10
o
d
e
+
25
2
5
0.2
rpmx 1000
1
00
50
11
22
A/B
intraoral dental images with the third clip over the
cut-out.
17.9X-ray image viewer on the lamp support tube
Additional equipment
This X-ray image viewer is switched ON or OFF using
the shown key on the control panel of the dentist or
assistant element.
To view intraoral dental images the anti-glare film (1)
must first be attached with two clips. Then fasten the
intraoral dental images with the third clip over the
1
cut-out.
59 57 928 D 3370
D 3370.201.01.15.02
95
17 Extra equipment and accessories Sirona Dental Systems GmbH
17.10 X-ray image view on the trayOperating Instructions C2
17.10X-ray image view on the tray
Additional equipment
This X-ray image viewer is switched ON or OFF using
the shown key on the control panel of the dentist or
assistant element.
To view intraoral dental images the anti-glare film (1)
must first be attached with two clips. Then fasten the
intraoral dental images with the third clip over the
cut-out.
1
+
96D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH18 Maintenance
A
MAINTENANCE
IN 30 DAYS
Operating Instructions C2
+
18 Maintenance
Despite the outstanding quality of your treatment center,
it is necessary to perform inspections and preventive
maintenance as well as safety tests at predetermined
intervals in order to ensure its operational safety.
These events are documented in the Maintenance Manual.
Care and cleaning is done by the practice team.
For details please refer to the following sections.
18.1Care and cleaning by the practice team
The practice team is responsible for regular cleaning
and care of the treatment center, with the aim of minimizing the risk of contamination for patients and users.
The work to be performed is specified in the document
“Care and Cleaning by the Practice Team”.
18.2Inspection and maintenance
MAINTENANCE
IN 30 DAYS
In order to guarantee the operational safety and reliability of your treatment center and to avoid damage due to
natural wear, yearly inspection and maintenance must
be performed on your treatment center. This is done by
an authorized service engineer from your dental depot.
As soon as the next maintenance date is less than 30
days away, the maintenance prompt automatically
appears on the display each time the unit is switched on.
You should then contact your local dental depot immediately for a maintenance appointment.
After this date has expired, the following message is displayed:
MAINTENANCE
REQUIRED
The steps to be performed as well as the parts which
must be replaced are specified in the document “Maintenance Certificate”.
An overview of the inspection and maintenance work
performed is also entered in the “Installation Report /
Warranty Passport” by the service engineer. This document is part of the “Maintenance Manual”.
59 57 928 D 3370
D 3370.201.01.15.02
97
18 Maintenance Sirona Dental Systems GmbH
18.3 Safety checksOperating Instructions C2
18.3Safety checks
Medical products are designed in such a way that the
first occurrence of a fault does not create a hazard to the
safety of patients, users or other persons. Hence it is
important to detect such faults before a second fault
occurs, which might then lead to safety hazards. It is
therefore essential to perform safety tests primarily
aimed at detecting electrical faults (e.g. isolation
defects) every 2 years. This must be done by an authorized service engineer from your dental depot, preferably
in connection with the work specified in section 18.2.
The check includes a visual inspection as well as measurements of the protective ground wire connections
and the equivalent leakage currents.
The inspections and measurements to be performed are
specified in the “Maintenance Manual”. The measured
values must be documented there by the service engineer.
Safety tests must also be performed and documented
during initial start-up, after extensions/upgrades (conversion) of your treatment center and after any repair
work which might affect the electrical safety of the system.
+
i
NOTICE
The treatment center must not be operated if it has failed
to pass the safety tests!
Safety tests for units with HF surgery equipment:In Germany, medical devices are subject to the provi-
sions of the Ordinance on the Installation, Operation and
Use of Medical Devices (Medizinprodukte-Betreiberverordnung – MPBetreibV) of June 29, 1998.
According to Section 6, safety tests are required for
devices with HF surgery equipment.
According to Section 7, a “Medical Product Log” must be
kept, in which the measured values as well as the tests
conducted must be documented.
These tests for devices with HF surgery equipment are
identical to the safety tests described above.
The "Maintenance Manual" is thus at the same time the
"Medical Product Log".
The system owner is obligated to maintain this Medical
Product Log.
In order to comply with the provisions of the Ordinance
on the Installation, Operation and Use of Medical
Devices (MPBetreibV), the following documentation
must be maintained for treatment centers with HF surgery equipment in Germany:
• Performance of safety tests
98D 3370.201.01.15.02
59 57 928 D 3370
Sirona Dental Systems GmbH18 Maintenance
Operating Instructions C2
+
• Personnel who have been trained in the use of
HF surgical equipment (§5 MPBetreibV)
• Repair work on the HF module
• Effects of malfunctions and repeated, similar operator
errors
• Reporting of incidents to authorities and manufacturers
The Medical Product Log must be safekept for a period
of at least 5 years after putting the system out of service.
Upon request, the Medical Product Log must be made
available to the competent authority for inspection purposes at any time.
i
NOTICE
As a system user outside of Germany, you must observe
the legal requirements of your country.
18.4 Maintenance Manual
18.4Maintenance Manual
Keep this document near your treatment center.
Any inspection and maintenance work as well as all
safety tests are documented by the service engineer in
the Maintenance Manual.
For units with HF surgery equipment, the consequences
of malfunctions and reports to the competent authorities
must be documented by the user.
The "Maintenance Manual" is thus at the same time the
"Medical Product Log".
We recommend that the user always keep the documentation in the chapter “Reporting of incidents to authorities
/ manufacturers” up-to-date, regardless of any legal
requirements.
59 57 928 D 3370
D 3370.201.01.15.02
99
18 Maintenance Sirona Dental Systems GmbH
18.4 Maintenance ManualOperating Instructions C2
+
100D 3370.201.01.15.02
59 57 928 D 3370
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.