Parr Instrument A1960E User Manual

INSTRUCTIONS FOR FIELD INSTALLATION OF AN A1960E SERIES PRESSURE DIGITAL PANEL METER (PDM)
1. Unplug the power cord of the modular controller. Remove the (2) screws from the back side (at
2. Remove and discard the uppermost 2" x 4" metal hole plug from the sloped front panel after
3. Use a small screwdriver to pry the red front panel off of the pressure DPM (digital panel meter) by
4. Insert the DPM into the rectangular opening previously covered with the metal hole plug removed
5. Slide the rectangular-shaped ring over the rear side of the DPM from inside the sloped metal
6. Insert (2) small metal tabs (bent edge toward screw head) over #4 x 5/8 flat head machine
7. Install edge connector of internal pressure harness assembly to the back of the pressure DPM.
8. Secure connector to the DPM with the furnished #4 x 3/4 pan head sheet metal screws through
9. Secure the pressure harness to the DPM by inserting cable clamp (186DD) around the pressure
10. Route pressure harness under formed metal wire retainer (near hinged area) by removing one
11. Attach the white wire of the pressure harness assembly to any available position on terminal #9
12. Attach the black wire of the pressure harness assembly to any available position on terminal #10 of the terminal block located on R.H. bottom side of the controller. Attach the violet wire of the
13. Attach the orange wire of the pressure harness assembly to any available position on terminal #6
removing (2) attaching nuts from inside surface of the sloped front panel.
inserting the tip of the screwdriver into the notch at the bottom center of the DPM case.
in step 2. Take care to install the DPM with the printed circuit board, which protrudes from the rear, to the bottom side.
cover.
screws and insert (2) screws with tabs into holes on each side of the DPM. Turn screws with tabs into holes on each side of the DPM. Turn screws to thread into metal rectangular-shaped ring installed on the back side of the cover. Tighten screws securely.
Note, the keying slot in the printed circuit board and the keying plug in the edge connector of the harness must line up to properly install the harness.
the hole in each end of the connector, slide on the 3/16" long nylon spacer (1852HC), and attach to the base of the DPM. Do not over tighten screws as it could bulge or push off the front panel.
harness. Insert the remaining #4 x 3/4 fillister head machine screw through #6 flat washer, cable clamp, and into upper R.H. boss (as viewed from the front) on the back side of the DPM. Tighten to secure.
wire retainer attaching screw. Reposition wire retainer, reinstall screw and tighten securely.
of the terminal block located on R.H. bottom side of the controller.
pressure harness assembly( on EW meters only) to any available position on terminal #8 of the terminal block located on R.H. bottom side of the controller.
of the terminal block located on R.H. bottom side of the controller.
No. 533M
- 1 -
14. Attach the blue wire of the pressure harness assembly to any available position on terminal #7 of the terminal block on R.H. bottom side of the controller.
15. Remove and discard hole plug in the "press. input" opening on rear panel of 4840 chassis.
16. Install the pressure input connector on the pressure harness into the opening previously covered with a hole plug removed in step 15. Short green wire with ring terminal goes on one of the connector mounting studs.
17. If DPM has an analog out connector install the 896E BNC connector into the appropriate hole on the back panel of the controller. Solder the red wire from position 5 of the pressure harness to the center post of the BNC connector . Solder the black wire from position 9 of the pressure harness to the outer tab of the BNC connector.
18. Connect the pressure transducer to the controller with the (A1905E Series) external pressure harness.
19. Reinstall red plastic cover to DPM.
20. Close hinged controller cover and reinstall the two retaining screws.
21. The A1960E Series pressure DPM is calibrated at Parr Instrument Company. No further adjustments are required.
PARR INSTRUMENT COMPANY
211 Fifty-Third Street
Moline, Illinois 61265 USA
309/762-7716 800/872-7720
Fax 309/762-9453
http://www.parrinst.com
E-Mail: parr@parrinst.com
Printed 11/02/06
Loading...