Parr Instrument 6200 User Manual

6200
Isoperibol Calorimeter
Operating Instruction Manual
For models produced after October 2010
585M
Table of Contents6200
Scope 4
Related Instructions 4
Purpose 4
Customer Service 4
Explanation of Symbols 5
Safety Information 5
Intended Usage 5
General Specifications 5
Environmental Conditions 6
Provisions for Lifting and Carrying 6
Cleaning & Maintenance 6
Getting Started 6
Chapter 1 7
Concept of Operation 7
Overview 7
Removable Bomb 7
Removable Bucket 7
Dynamic Operation 7
Full Microprocessor Based Process Control 8
Full Microprocessor Based Data Acquisition and Handling 8
Flexible Programming 8
Chapter 2 9
Installation 9
Environmental Conditions 9
Required Consumables, Utilities and Power Requirements 9
Filling the Jacket Reservoir 9
Power Connection 9
Jacket Cooling Water Connection 9
Tap Water Cooling 10
Cooling with the Water Handling System 10
Oxygen Filling Connection 10
Printer and Balance Connections 10
Standardizing the Calorimeter 10
Swagelok Tube Fittings 11
Chapter 3 15
Instrument Description 15
Types of Controls 15
Menu Keys 15
Control Keys 15
Chapter 4 17
Program Installation & Control 17
Software Installation 17
Default Settings 17
Revising Default Settings 17
Chapter 5 21
Operating Instructions 21
Operating the 1108P Oxygen Combustion Vessel 21
Operating the Filling Connection 21
Operating the Calorimeter 21
Samples and Sample Holders 23
Combustion Aids 23
Oxygen Charging Pressure 23
Combustion Capsules 23
Foodstuffs and Cellulosic Materials 24
Coarse Samples 24
Corrosive Samples 24
Explosives and High Energy Fuels 24
Volatile Sample Holders 24
Poor Combustion 25
Chapter 6 27
Corrections & Final Reports 27
Entering Corrections and Obtaining the Final Report 27
Manual Entry 27
Fixed Corrections 27
Chapter 7 29
Reporting Instructions 29
Report Option Section 29
Report Generation 29
Net Heat of Combustion 29
www.parrinst.com
1
Table of Contents
Chapter 8 31
File Management 31
Clearing Memory 31
Removable SD Memory 31
Chapter 9 33
Maintenance & Troubleshooting 33
Oxygen Bomb 33
Jacket Temperature Troubleshooting 33
Fuses 33
6200 Calorimeter Error List 33
Appendix A 35
Menu Operating Instructions 35
Calorimeter Operation Menu 35
Temperature vs. Time Plot Screen 35
Temperature Plot Setup 36
Operating Controls Menu 37
Program Information and Control Menu 38
Calibration Data and Controls Menu 39
Thermochemical Calculations Menu 41
Data Entry Controls Menu 43
Reporting Controls Menu 45
Communication Controls Menu 45
File Management 47
Diagnostics Menu 48
Appendix B 49
Calculations 49
Calculating the Heat of Combustion 49
General Calculations 49
Thermochemical Corrections 49
ASTM and ISO Methods Differ 50
Thermochemical Calculation Details 51
Acid and Sulfur Corrections 51
ASTM Treatment for Acid and Sulfur 53
ISO Calculations 53
Spiking Samples 54
Conversion to Other Moisture Bases 54
Conversion to Net Heat of Combustion 54
Appendix C 55
Standardization 55
Standardizing the Calorimeter 55
Standard Materials 55
Automatic Statistical Calculations 55
Appendix D 59
Communications Interfaces 59
Printer Port 59
Balance and Port Input Driver Specifications 59
Mettler 011/012 Balance Interface 59
Sartorius Balance Interface 59
Generic Interface 60
Ethernet Interface 61
Samba Server Feature (Optional) 62
Bar Code Port 70
Network Data Services 70
Appendix E 71
Technical Service 71
Return for Repair 71
Appendix F 73
Parts Lists & Drawings 73
Principal Assemblies in Calorimeter 73
Parts List for A1279DD2 Controller Assembly 73
Parts List for Temperature Control Assembly 73
Parts List for A1284DD2 Stirrer Hub Assembly 74
Parts List for Water Tank Assembly 74
Parts List for Cooling Water Supply 74
Parts List for Oxygen Filling System 74
Part List for 6200 Stirrer Motor and Drive 74
Spare and Installation Parts List* 74
2
Parr Instrument Company
Tables
Table 4-1 6200 Factory Default Settings 18
Table B-1 Settings for ISO & BSI Methods 52
Table C-1 Calorimeter Control Limit Values in J/g 56
Table C-2 Calorimeter Control Limit Values in cal/g 57
Table C-3 Calorimeter Control Limit Values in BTU/lb 58
Table D-1 6200 Data File Naming Convention 60
Table D-2 6200 Calorimeter Run Data Template 60
Figures
Figure 2-1 Swagelok Tube Fittings 11
Figure 2-2 6200 Calorimeter Back Panel 12
Figure 2-3 Closed Loop Configuration with 6510 13
Figure 2-4 Open Loop Configuration with Tap Water 13
Figure 2-5 Open Loop Configuration with 1552 14
Figure 5-7 Combustion Capsule with Adhesive Tape Seal 25
Figure F-1 6200 Isoperibol Calorimeter Cutaway Front 75
Figure F-2 6200 Isoperibol Calorimeter Cutaway Rear 76
Figure F-3 A1279DD2 Control Schematic 77
Figure F-4 Oxygen Solenoid Assembly & Fittings 78
Figure F-5 Water Tank Assembly 79
Figure F-6 A1311DD Circulating Pump Assembly 80
Figure F-7 Temperature Control Assembly with Fittings 81
Figure F-8 A1284DD2 Stirrer Hub Assembly 82
Figure F-9 Stirrer Motor Assembly 83
Table of Contents6200
www.parrinst.com
3

Preface

Preface

Scope

This manual contains instructions for installing and operating the Parr 6200 Calorimeter. For ease of use, the manual is divided into nine chapters.
Concept of Operation Installation Instrument Description Program Installation & Control Operating Instructions Corrections & Final Reports Reporting Instructions File Management
Maintenance & Troubleshooting

Related Instructions

Additional instructions concerning the installation and operation of various component parts and pe­ripheral items used with the 6200 Calorimeter have been included and made a part of these instructions.
No. Description
201M Limited Warranty
418M 1108P Oxygen Combustion Vessel
207M Analytical Methods for Oxygen Bombs 230M Safety in the Operation of Laboratory
and Pressure Vessels
483M Introduction to Bomb Calorimetry
Additional instructions for the printer, cooler, and water handling systems are found in the respective package and should be made a part of this book.
Subsections of these chapters are identified in the Table of Contents.
To assure successful installation and operation, the user must study all instructions carefully before starting to use the calorimeter to obtain an under­standing of the capabilities of the equipment and the safety precautions to be observed in the operation.
Note About Nomenclature:
Historically, burning a sample enclosed in a high pressure oxygen environment is known as Oxygen Bomb Calorimetry and the vessel containing the sample is known as an Oxygen Bomb. The terms bomb and vessel are used interchangeably.
Note: The unit of heat used in this manual is the International Table calorie, which is equal to 4.1868 absolute joules.

Purpose

Heats of combustion, as determined in an oxygen bomb calorimeter such as the 6200 Isoperibol Calorimeter, are measured by a substitution proce­dure in which the heat obtained from the sample is compared with the heat obtained from a standard­izing material. In this test, a representative sample is burned in a high-pressure oxygen atmosphere within a metal pressure vessel or “bomb”. The energy released by the combustion is absorbed within the calorimeter and the resulting temperature change is recorded.

Customer Service

Questions concerning the installation or operation of this instrument can be answered by the Parr Customer Service Department:
1-309-762-7716 • 1-800-872-7720 • Fax: 1-309-762-9453
E-mail: parr@parrinst.com • http://www.parrinst.com
4
Parr Instrument Company
6200

Explanation of Symbols

I On Position
O Off Position
~ Alternating Current
Preface
This CAUTION symbol may be present on the Product Instrumenta­tion and literature. If present on the product, the user must consult the appropriate part of the accompanying product literature for more information.
ATTENTION, Electrostatic Discharge (ESD) hazards. Observe precau­tions for handling electrostatic sensitive devices.
Protective Earth (PE) terminal. Provided for connection of the protec-
tive earth (green or green/yellow) supply system conductor.
Chassis Ground. Identifies a connection to the chassis or frame of the equipment shall be bonded to Protective Earth at the source of supply in accordance with national and local electrical code requirements.
Earth Ground. Functional earth connection. This connection shall be bonded to Protective earth at the source of supply in accordance with national and local electrical code requirements.

Safety Information

To avoid electrical shock, always:
1. Use a properly grounded electrical outlet of correct voltage and current handling capability.
2. Ensure that the equipment is connected to electrical service according to local national electrical codes. Failure to properly connect may create a fire or shock hazard.
3. For continued protection against possible hazard, replace fuses with same type and rating of fuse.
4. Disconnect from the power supply before maintenance or servicing.
To avoid personal injury:
1. Do not use in the presence of flammable or combustible materials; fire or explosion may result. This device contains components which may ignite such material.
2. Refer servicing to qualified personnel.

Intended Usage

If the instrument is used in a manner not specified by Parr Instrument Company, the protection pro­vided by the equipment may be impaired.
General Specifications
Electrical Ratings
120VAC, 6.0 Amps. 50/60 Hz 240VAC, 3.0 Amps. 50/60 Hz
Before connecting the calorimeter to an electrical outlet the user must be certain that the electrical outlet has an earth ground connection and that the line, load and other characteristics of the installation do not exceed the following limits:
Voltage: Fluctuations in the line voltage should not exceed 10% of the rated nominal voltage shown on the data plate.
www.parrinst.com
5
Preface
Frequency: Calorimeters can be operated from either a 50 or 60 Hertz power supply without affect­ing their operation or calibration.
Current: The total current drawn should not exceed the rating shown on the data plate on the calorim­eter by more than 10 percent.

Environmental Conditions

Operating: 15 ºC to 30 ºC; maximum relative humid­ity of 80% non-condensing. Installation Category II (over voltage) in accordance with IEC 664. Pollution degree 2 in accordance with IEC 664.
Altitude Limit: 2,000 meters.
Storage: -25 ºC and 65 ºC; 10% to 85% relative
humidity.

Provisions for Lifting and Carrying

Before moving the instrument, disconnect all con­nections from the rear of the apparatus. Lift the instrument by grabbing underneath each corner.

Cleaning & Maintenance

Periodic cleaning may be performed on the exterior surfaces of the instrument with a lightly dampened cloth containing mild soap solution. All power should be disconnected when cleaning the instru­ment. There are no user serviceable parts inside the product other than what is specifically called out and discussed in this manual. Advanced trouble­shooting instructions beyond the scope of this manual can be obtained by calling Parr Instrument Company in order to determine which part(s) may be replaced or serviced.

Getting Started

These steps are offered to help the user become familiar with, install, operate and develop the full capabilities of the Parr 6200 Calorimeter.
1. Review the Concept of Operations, Chapter 1, to get an understanding of the overall capabilities of the calorimeter and microprocessor control.
2. Unpack and install the calorimeter in accordance with Installation, Chapter 2. This simple, step­wise procedure will acquaint the user with the
various parts of the calorimeter and make it easier to understand the operating instructions which follow.
3. Turn the power switch ON (located on the back).
Turn to the Instrument Description, Chapter 3, to review the touch screen controls.
4. Review the Program Installation and Control,
Chapter 4, to match the factory settings to the in­tended mode of operation. Any required changes can be made to the program parameters located in the Main Menu.
5. Review the Reporting Instructions, Chapter 7, to
become familiar with the manner in which calo­rimetry corrections are entered. Also discussed are generating final reports, editing and clearing memory.
6. Turn to the Menu Operating Instructions, Ap-
pendix A, to review the menu functions used to modify the program contained in the 6200 Calorimeter. A review of the menus will provide a good idea of the capabilities and flexibility designed into this instrument.
7. Review the Calculations, Appendix B. This pro-
vides information about calculations performed by the 6200 Calorimeter.
8. Review Standardization, Appendix C. This will
serve two important functions. First, it provides instructions on generating the energy equivalent factor required to calculate the heat of combus­tion of unknown samples. Secondly, it will give the user the opportunity to run tests on a mate­rial with a known heat of combustion to become familiar with the instrument and confirm that the instrument and operating procedures are pro­ducing results with acceptable precision. Most 6200 Calorimeters will have an energy equiva­lent of approximately 2400 calories per ºC. The runs for standardization and determinations are identical, except for the setting of the instrument to the standardization or determination mode.
9. Review the Communication Interfaces, Appendix
D, for the correct installation of any peripherals connected to the 6200 Calorimeter.
10. After successful standardization, the 6200 Calo­rimeter should be ready for testing samples.
6
Parr Instrument Company
6200

chaPter 1

Concept of Operation

Overview

The 6200 Calorimeter has been designed to provide the user with:
Concept of Operation
The 1108P Oxygen Combustion Vessel is made of high-strength, high nickel stainless steel designed to resist the corrosive acids produced in routine fuel testing. An alternative 1108PCL vessel is available, constructed of an alloy containing additional cobalt and molybdenum to resist the corrosive conditions produced when burning samples containing chlori­nated compounds.
1
A traditional design calorimeter with removable oxygen bomb and bucket.
A moderately priced calorimeter which uses real time temperature measurements to determine heat leaks using a controlled calorimeter jacket.
A high precision calorimeter capable of exceeding the repeatability and reproducibility requirements of all international standard test methods.
A compact calorimeter requiring minimum laboratory bench space.
A modern intuitive graphical user interface for ease of operation and training.
A calorimeter with up to date digital hardware, software and communication capabilities.
A calorimeter that is cost effective and which can incorporate a user’s current bombs, buckets, and accessories.

Removable Bomb

The Model 6200 Calorimeter utilizes the Parr 1108P Oxygen Combustion Vessel. More than 20,000 of these reliable 1108 style oxygen combustion ves­sels have been placed in service on a world wide basis. This bomb features an automatic inlet check valve and an adjustable needle valve for controlled release of residual gasses following combustion. They are intended for samples ranging from 0.6 to
1.2 grams with a maximum energy release of 8,000
calories per charge.
The Model 6200 can also be equipped with a variety of special purpose oxygen combustion vessels for
unusual samples and/or applications. The 1104 High
Strength Oxygen Combustion Vessel is designed for testing explosives and other potentially hazardous materials. The 1109A Semimicro Oxygen Combus­tion Vessels can be fitted along with its unique bucket to test samples ranging from 25 to 200 mg.

Removable Bucket

The A391DD removable bucket has been designed to hold the bomb, stirrer and thermistor with a mini­mum volume of water and to provide an effective circulating system which will bring the calorimeter to rapid thermal equilibrium both before and after firing.

Dynamic Operation

In its Dynamic Operating Mode, the calorimeter uses a sophisticated curve matching technique to compare the temperature rise with a known ther­mal curve to extrapolate the final temperature rise without actually waiting for it to develop. Repeated testing, and over 20 years of routine use in fuel laboratories, has demonstrated that this technique can cut the time required for a test by one-half without significantly affecting the precision of the calorimeter.
www.parrinst.com
7
1
Concept of Operation

Full Microprocessor Based Process Control

The microprocessor controller in this calorimeter has been pre-programmed to automatically prompt the user for all required data and control input and to:
Generate all temperature readings in the calorimeter.
Monitor jacket as well as bucket temperature.
Confirm equilibrium conditions.
Fire the bomb.
Confirm that ignition has occurred.
Determine and apply all necessary heat leak
corrections.
Perform all curve matching and extrapolations required for dynamic operation.
Terminate the test when it is complete.
Monitor the conditions within the calorimeter
and report to the user whenever a sensor or operating condition is out of normal ranges.

Full Microprocessor Based Data Acquisition and Handling

In addition to its process control functions, the microprocessor in the calorimeter has been pre­programmed to:
Collect and store all required test data.
Apply all required corrections for combustion
characteristics.

Flexible Programming

The fifth generation software built into this calorim­eter and accessed through the screen menus permit the user to customize the operation of the calorim­eter to meet a wide variety of operating conditions including:
A large selection of printing options.
Choice of accessories and peripheral equipment.
Multiple options in regard to handling
thermochemical corrections.
Choice of ASTM or ISO correction procedures.
A variety of memory management and reporting
procedures.
Complete freedom for reagent concentrations and calculations.
Unlimited choice of reporting units.
Automatic bomb usage monitoring and
reporting.
A choice of Equilibrium or Dynamic test methods.
Automatic statistical treatment of calibration runs.
Enhanced testing and trouble shooting procedure.
The 6200 Calorimeter is equipped with one USB connection for direct communication with its printer and other peripherals. It is also equipped with an Ethernet network connection for connections to laboratory computers.
Compute and report the heat of combustion for the sample.
8
Parr Instrument Company
6200

chaPter 2

Installation

Note: Some of the following manual sec­tions contain information in the form of warnings, cautions and notes that require special attention. Read and follow these instructions carefully to avoid personal injury and damage to the instrument. Only quali­fied personnel should conduct the instal­lation tasks described in this portion of the manual.

Environmental Conditions

The 6200 Calorimeter is completely assembled and given a thorough test before it is shipped from the factory. If the user follows these instructions, instal­lation of the calorimeter should be completed with little or no difficulty. If the factory settings are not disturbed, only minor adjustments will be needed to adapt the calorimeter to operating conditions in the user’s laboratory.
This apparatus is to be used indoors. It requires at least 8 square feet of workspace on a sturdy bench or table in a well-ventilated area with con­venient access to an electric outlet, running water and a drain. The supply voltage must be within ± 10% of marked nominal voltage on the apparatus. The supply voltage receptacle must have an earth ground connection.

Required Consumables, Utilities and Power Requirements

The 6200 Calorimeter System requires availability of Oxygen, 99.5% purity, 2500 psig maximum.
The power requirements for the subassemblies of the 6200 Calorimeter are:
Calorimeter
120VAC, 6.0 Amps. 50/60 Hz 240VAC, 3.0 Amps. 50/60 Hz
Installation

Filling the Jacket Reservoir

The water reservoir of the calorimeter must be filled with approximately 1.4 liters of water (distilled or de-ionized preferred). This must be done prior to turning on the heater and the pump. The reservoir is filled through the tank fill elbow on the back of the calorimeter. The tank is full once water stands in the horizontal run of the filling elbow.

Power Connection

Plug the power line into any grounded outlet provid­ing proper voltage that matches the specification on the nameplate of the calorimeter. The calorim­eter will draw approximately 300 watts of power. Grounding is very important not only as a safety measure, but also to ensure satisfactory controller performance. If there is any question about the reliability of the ground connection through the power cord, run a separate earth ground wire to the controller chassis.
Turn the power switch to the on position. After a short time, the Parr logo will appear on the LCD display followed by a running description of the instrument boot sequence. When the boot sequence is complete, the calorimeter Main Menu is dis­played. Go to the Calorimeter Operation page and turn the heater and pump on. This begins circulating and heating the calorimeter jacket water. Add water to the filling elbow at the rear of the instrument as required in order to keep it full.

Jacket Cooling Water Connection

It becomes necessary to use the jacket cooling water connection only if the calorimeter operating room temperature exceeds 24 °C (75 °F).
When required, an external water source is used to cool the jacket of the 6200 Calorimeter. This is done in either of the following ways:
1. Tap water is used for cooling and then run to a drain.
2. Cooling water is re-circulated to the calorimeter from a Parr 6510 Water Handling System.
2
Printer
100 to 240 VAC, 0.35 Amps 50/60 Hz
Printer Supplies
334C Printer Paper 335C Printer Ribbon
The water that provides the cooling goes through a heat exchanger and does not mix with the water in the jacket and its reservoir. There is a very low cool­ing load and tap water up to a temperature of 27 °C should be adequate.
www.parrinst.com
9
2
Installation

Tap Water Cooling

Connect the tap water supply to the cold water inlet
on the back of the calorimeter using either 1/4”
copper or nylon tubing (HJ0025TB035). A 196VB metering valve is provided with the calorimeter. This valve should be installed in this inlet line near the calorimeter. This valve is used to adjust the flow of water to the heat exchanger to compensate for differences in tap water temperatures and water line pressures. Once the calorimeter is operating at equi­librium, check the jacket temperature that is displayed on the operating page. If this temperature is cycling significantly, close down on the metering valve to reduce the flow of cooling water. If the jacket rises above its 30 °C set point, open this valve to increase
the cooling. A ow rate of 100 ml/ minute is generally
all that is required.
Connect the cooling water outlet on the back of the calorimeter to a drain using either nylon
(HJ0025TB035) or copper 1/4” tubing. A shut off
valve in tap water supply line is also a good idea if the calorimeter will not be used for an extended period.

Cooling with the Water Handling System

If the calorimeter is to be operated with a Parr Water Handling System, connect the pump output to the cooling water inlet and connect the cooling water outlet to the return connection on the water handling system. With this installation it is neither necessary nor desirable to install the 196VB metering valve in the inlet line. It is a good idea to keep all water line runs as short as practical to avoid unwanted tempera­ture changes in the water between the source and the calorimeter.

Oxygen Filling Connection

The 6200 Calorimeter is equipped with an automatic bomb oxygen filling system. This system consists of an oxygen pressure regulator with a relief valve that mounts on an oxygen tank and a controlled solenoid inside the calorimeter. To install the regulator on the oxygen supply tank, unscrew the protecting cap from the oxygen tank and inspect the threads on the tank outlet to be sure they are clean and in good condi­tion. Place the ball end of the regulator in the outlet and draw up the union nut tightly, keeping the gages tilted slightly back from an upright position. Connect the regulator to the oxygen inlet fitting on the back of the calorimeter case. This hose should be routed so that it will not kink or come in contact with any hot surface. Connect the high-pressure nylon hose with the push on connector to the oxygen outlet fitting on the back of the calorimeter.
All connections should be checked for leaks. Any leaks detected must be corrected before proceeding. Instructions for operating the filling connection are in the Operating Instructions chapter.
Adjust the pressure regulator to deliver 450 psi of O2. Assemble the oxygen bomb without a charge and attach the filling hose to the bomb inlet valve. Press the O2 Fill key on the Calorimeter Operation page and observe the delivery pressure on the 0 – 600 psi gage while the oxygen is flowing into the bomb. Adjust the regulator, if needed, to bring the pressure to 450 psi. If there is any doubt about the setting, release the gas from the bomb and run a second check.

Printer and Balance Connections

Connect the printer to the calorimeter at this time. The Parr 1758 Printer is configured and furnished with a cord to connect directly to the USB port on the back of the calorimeter.
If a balance is to be attached to the calorimeter it will be necessary to use a USB hub so that multiple devices can be connected. Any standard USB hub can be used.

Standardizing the Calorimeter

The calorimeter must be accurately standardized prior to actually performing calorimetric tests on sample materials. Review Appendix C - Standardization, in order to become familiar with the general procedure and calculations. The user should configure the calorimeter at this time to accommodate the desired sample weight entry mode. The calorimeter can be placed into standardization mode on the Calorimeter Operation Page, with the OPERATING MODE key. If two bombs and buckets are being used with the calorimeter to maximize sample throughput, the calorimeter can be configured to prompt for a Bomb ID at the start of each test. The Bomb ID can also be selected on the Calorimeter Operations Page, using
the Bomb/EE key. All bomb and bucket combinations
will need to be standardized separately. The end result of a standardization test is an energy equivalent value, or the amount of energy required to raise the temperature of the calorimeter one degree. Repeated standardization with any given bomb and bucket combination should yield an energy equivalent value with a range of 14 calories per degree, centered around the mean value for all tests using that bomb and bucket combination. The calorimeter is ready for testing samples after an energy equivalent value has been obtained.
10
Parr Instrument Company
6200
Installation
2

Swagelok Tube Fittings

When Swagelok Tube Fittings are used, the instruc­tions for installation are:
1. Simply insert the tubing into the Swagelok Tube Fitting. Make sure that the tubing rests firmly on the shoulder of the fitting and that the nut is finger-tight.
2. Before tightening the Swagelok nut, scribe the nut at the 6 o’clock position.
3. While holding the fitting body steady with a
back-up wrench, tighten the nut 1-1/4 turns.
Watch the scribe mark, make one complete revolution and continue to the 9 o’clock position.
4. For 3/16” and 4mm or smaller tube ttings,
tighten the Swagelok nut 3/4 turns from nger-
tight.
Swagelok tubing connections can be disconnected and retightened many times. The same reliable leak­proof seal can be obtained every time the connec­tion is remade using the simple two-step procedure.

Figure 2-1

Swagelok Tube Fittings

1. Insert the tubing with pre-swaged ferrules into the fitting body until the front ferrule seats.
2. Tighten the nut by hand. Rotate the nut to the original position with a wrench. An increase in resistance will be encountered at the original position. Then tighten slightly with a wrench.
Smaller tube sizes (up to 3/16” or 4mm) take less
tightening to reach the original position than larger tube sizes.
The type of tubing and the wall thickness also has an effect on the amount of tightening required. Plastic tubing requires a minimal amount of ad­ditional tightening while heavy wall metal tubing may require somewhat more tightening. In general,
the nut only needs to be tightened about 1/8 turn
beyond finger tight where the ferrule seats in order to obtain a tight seal.
Over tightening the nut should be avoided. Over tightening the nut causes distortion (flaring) of the lip of the tube fitting where the ferrule seats. This in turn causes the threaded portion of the body to deform. It becomes difficult to tighten the nut by hand during a subsequent re-tightening when the fitting body becomes distorted in this manner.
www.parrinst.com
11
2
Installation

Figure 2-2

6200 Calorimeter Back Panel

12
Parr Instrument Company
6200
Line 3

Figure 2-3

Closed Loop Configuration with 6510
O2 Regulator
Line 2
6510 Water
Handling
System
Installation
2
Line 3
Line 1 ~27 °C

Figure 2-4

Open Loop Configuration with Tap Water
O2 Regulator
Tap Water
· <27 °C
Line 2
Line 1
Drain
Line 1 & 2 – Maximum length of 10 feet, 1/4” OD, Polyurethane (Part Number HJ0025TB035) Line 3 – Maximum length of 25 feet, 1/8” OD, Nylon (Part Number HX0012TB024)
www.parrinst.com
13
2
Installation

Figure 2-5

Open Loop Configuration with 1552
O2 Regulator
Tap Water
· >27 °C
Line 3
Line 2
Line 1 ~27 °C
Line 1 & 2 – Maximum length of 10 feet, 1/4” OD, Polyurethane (Part Number HJ0025TB035) Line 3 – Maximum length of 25 feet, 1/8” OD, Nylon (Part Number HX0012TB024)
Water
Cooler
Drain
1552
14
Parr Instrument Company
6200

chaPter 3

Instrument Description

Types of Controls

All calorimeter configurations and operations are handled by a menu-driven system operated from the bright touch screen display. The settings and controls are organized into nine main sections or pages which comprise the MAIN MENU.
Instrument Description
pad or similar screen for entering the required value. Some keys lead to multiple choices. Always clear the current value before entering a new value. Once entered the screen will revert to the previous menu and the new value will be displayed in the lower right corner of the key.
4. Data Displays. Most of these keys display values
that have been calculated by the calorimeter and are informational only. Certain ones can be overridden by the user entering a desired value through a sub-menu. The value is displayed in the lower right corner of the key.
3
Note: Keys with a “double box” in the upper left hand corner lead to sub-menus.

Menu Keys

The controls that change the data field information in the menus will be one of the following:
1. Toggles. These data elds contain ON/OFF or
YES/NO choices. Simply touching the key on the
screen toggles the choice to the other option. The current setting is displayed in the lower right corner of the key.
2. Option Selection. These data fields contain a list of options. Touching the key on the screen steps the user through the available choices. The current setting is displayed in the lower right corner of the key.
Note: Some keys will respond with an op­portunity for the user to confirm the speci­fied action to minimize accidental disrup­tions to the program and/or stored data.

Control Keys

There are five control keys which always appear in the right column of the primary displays. These keys are unavailable when they are gray instead of white.
1. Escape. This key is used to go up one level in
the menu structure.
2. Main Menu. This key is used to return to the
main menu touch screen from anywhere in the menu structure.
3. Start. This key is used to start a calorimeter test.
4. Report. This key is used to access the test
results stored in the calorimeter, to enter thermochemical corrections and to initiate report on the display, printer or attached
computer.
5. Help. This key is used to access help screens
related to the menu currently displayed on the touch screen.
6. Abort. This key appears in the start key location
while the test is running. Pressing this key will abort the test in progress.
3. Value Entry Fields. These data fields are used to enter data into the calorimeter. Touching the key on the screen brings up a sub menu with a key
7. This key appears on the main menu only and is used to prepare the calorimeter for turning off the power.
www.parrinst.com
15
Notes
16
Parr Instrument Company
6200

chaPter 4

Program Installation & Control

Software Installation

Program Installation & Control

Revising Default Settings

The default parameters of the 6200 Calorimeter can be changed to guarantee that the 6200 Calorimeter, when cold restarted, will always be in the desired configuration before beginning a series of tests.
4
The program in the 6200 Calorimeter can be exten­sively modified to tailor the unit to a wide variety of operating conditions, reporting units, laboratory techniques, available accessories and communica­tion modes.
In addition, the calculations, thermochemical cor­rections, and reporting modes can be modified to conform to a number of standard test methods and procedures.
Numerous provisions are included to permit the use of other reagent concentrations, techniques, combustion aids and short cuts appropriate for the user’s work.
Note: Changes to the program are made by use of the menu structure described in Ap­pendix A of this manual. Any of these items can be individually entered at any time to revise the operating program.

Default Settings

Users who wish to permanently revise their default settings may do so using the following procedure:
Establish the operating parameters to be stored as the user default settings.
Go to the Program Info and Control Menu, User/
Factory Settings, User Setup ID, and enter the desired User Setup ID.
Select Save User Default Settings
To re-load the user default settings, go to the Pro-
gram Info and Control Page, User/Factory Settings,
Re-load User Default Settings, and YES.
Units are pre-programmed with default settings. See pages 17 and 18 for a listing of the factory default settings.
These default settings remain in effect until changed by the user. Should the user ever wish to return to the factory default settings, go to the Program
Information and Control Menu, then to User/Factory
Settings, and then touch Reload Factory Default Settings and YES.
Non-volatile memory is provided to retain any and all operator initiated program changes; even if power is interrupted or the unit is turned off. If the unit experiences an intentional or unintentional “Cold Restart”, the controller will return to its default settings.
www.parrinst.com
17
4
Program Installation & Control

6200 Factory Default Settings

Calorimeter Operations
Operating Mode Determination
Bomb Installed/EE 1/2400.0
Heater and Pump OFF
Operating Controls
Method of Operation Dynamic
Reporting Units BTU/lb
Use Spiking Correction OFF
“OTHER” Multiplier 4.1868
Calibrate Touchscreen
LCD Backlight Timeout(s)
LCD Backlight Intensity 70%
Print Error Messages ON
Language English
1200 S

Table 4-1

Calibration Data & Controls
Calibration Run Limit
EE Max Std Deviation 0.0
Heat of Combustion of Standard 6318.4
Bomb Service Interval 500
Control Chart Parameters
Charted Value HOC Standard
Process Sigma 0.1
Temp Rise High Warning 3.3
Temp Rise Low Warning 2.0
Use Bomb 1
10
Bomb 1 Through 4
EE Value 2400.0
Protected EE Value OFF
Spike Controls
Use Spiking OFF
Heat of Combustion of Spike 6318.4
Use Fixed Spike OFF
Weight of Fixed Spike 0.0
Prompt for Spike before Weight OFF
Program Information and Control
Date & Time Settings
Volume Level Adjust 85%
Software and Hardware Info
Settings Protect OFF
User/Factory Settings
Feature Key
Bomb Type Select
User Function Setup
Cold Restart
User/Factory Settings
User Setup ID 62-1108
Reload Factory Default Settings
Reload User Default Settings
Save User Default Settings
Thermochemical Corrections Standardization
Fixed Fuse Correction ON 50
Acid Correction Fixed HNO3 10.0
Fixed Sulfur Correction ON 0.0
Determination
Fixed Fuse Correction ON 50
Acid Correction Fixed HNO3 10.0
Fixed Sulfur Correction OFF 0.0
Net Heat/Dry Factors Net Heat & Dry Disable
Fixed Hydrogen
Fixed Oxygen ON 0.0
Fixed Nitrogen ON 0.0
Calculate Net Heat of Combustion OFF
Fixed Moisture as Determined OFF 0.0
Fixed Moisture as Received OFF 0.0
Dry Calculation OFF
OFF 0.0
18
Parr Instrument Company
6200
Table 4-1 (Continued)
6200 Factory Default Settings
Calculation Factors
Nitric Acid Factor 1.58
Acid Multiplier 0.0709
Sulfur Value is Percent ON
Sulfur Multiplier 0.6238
Fuse Multiplier 1. 0
Use Offset Correction (ISO) OFF
Offset Value 0.0
Data Entry Controls
Prompt for Bomb ID ON
Weight Entry Modes Touch Screen
Acid Entry Mode Touch Screen
Net Heat Entry Modes Touch Screen
Auto Sample ID Controls ON
Sample Weight Warning above 2.0
Spike Weight Entry Mode Touch Screen
Sulfur Entry Mode Touch Screen
Moisture Entry Mode Touch Screen
Auto Preweigh Controls ON
Auto Sample ID Controls
Automatic Sample ID ON
Automatic Sample ID Increment 1
Automatic Sample ID Number 1
Auto Preweigh Controls
Automatic Preweigh ID ON
Automatic Preweigh ID Increment 1
Automatic Preweigh ID Number 1
Program Installation & Control
Communication Controls
Printer Type Parr 1758
Balance Port
Network Interface
Printer Destination Local (USB)
Bar Code Port
Network Data Devices
Balance Port Communications
Balance Type
Balance Port Device
Customize Balance Settings
Balance Port Settings
Number of Data Bits 8
Parity None
Number of Stop Bits 1
Handshaking None
Baud Rate 9600
Data Characters from Balance 8
Data Precision 4
Transfer Timeout (seconds) 10
Balance Handler Strings
Data Logger
Data Logger OFF
Data Log Interval 12s
Data Log Destination Log File and Printer
Select Data Log Items
Data Log Format Text Format
Generic
4
Reporting Controls
Report Width 40
Automatic Reporting ON
Auto Report Destination Printer
Individual Printed Reports OFF
Edit Final Reports OFF
Recalculate Final Reports OFF
Use New EE Values in Recalculation OFF
www.parrinst.com
19
Notes
20
Parr Instrument Company
6200

Operating Instructions

chaPter 5

Operating Instructions

Operating the Calorimeter

All operations required to standardize the 6200 Calorimeter, or test an unknown sample, should proceed step-wise in the following manner:
5

Operating the 1108P Oxygen Combustion Vessel

Detailed instructions for preparing the sample and charging the 1108P Oxygen Combustion Vessels are given in Operating Instructions No. 418M. Follow these instructions carefully, giving particular atten­tion to the precautions to be observed in charging and handling the bomb.

Operating the Filling Connection

To fill the bomb, connect the hose to the bomb inlet valve and push the O2 Fill button on the calorimeter control panel. The calorimeter will then fill the bomb to the preset pressure and release the residual pressure in the connecting hose at the end of the filling cycle. It will take approximately 60 seconds to fill the bomb. During this time a countdown timer on the O2 fill button will display the remaining fill time. Pushing the O2 key a second time will stop the flow of oxygen at any time. Once the display returns to its normal reading, the user can disconnect the coupling and proceed with the combustion test.
If the charging cycle should be started inadvertently, it can be stopped immediately by pushing the O2 FILL key a second time.
During extended periods of inactivity, overnight or longer, close the tank valve to prevent leakage. When changing oxygen tanks, close the tank valve and push the O2 FILL key to exhaust the system. Do not use oil or combustible lubricants on this filling system or on any devices handling oxygen under pressure. Keep all threads, fittings, and gaskets clean and in good condition. Replace the two 394HCJE O-rings in the slip connector if the connector fails to maintain a tight seal on the bomb inlet valve.
The recommended filling pressure is 450 psig (3 MPa or 30 bar). This pressure is prescribed by most of the standard bomb calorimetric test methods. Higher or lower filling pressures can be used, but the bomb must never be filled to more than 600 psig (40 atm).
1. Turn on the calorimeter and activate the pump and heater using Calorimeter Operations. Allow
at least 20 minutes for the calorimeter to warm up and the jacket temperature to stabilize. Once the jacket temperature comes within 0.5 °C of 30 °C and stays there for approximately 15 min­utes, the calorimeter is ready to begin testing.
The Start key will be available at this time. The bomb parts should be wetted and then dried in the manner used at the conclusion of a test. This serves to wet all sealing parts as well as leav­ing the bomb with the same amount of residual water which will exist in all subsequent testing.
2. Prepare the sample weighing the material to 0.1 mg and charge the oxygen bomb as described in the section entitled Operating the Filling Con- nection. Using an additional bomb and bucket can increase the throughput of the 6200 Calorim­eter. With this arrangement, the calorimeter can operate almost continuously since the operator will be able to empty a bomb and recharge it while a run is in progress. A bomb and bucket for the next run will be ready to go into the calo­rimeter as soon as it is opened. Each bomb and bucket combination will have to be standardized separately and the proper energy equivalent for each set must be used when calculating the heat of combustion.
3. Fill the calorimeter bucket by first taring the dry bucket on a solution or trip balance; then add
2000 (+/- 0.5) grams of water. Distilled water is
preferred, but demineralized or tap water con­taining less than 250 ppm of dissolved solids is satisfactory. The bucket water temperature should be approximately 3 to 5 °C below the jacket temperature. It is not necessary to use exactly 2000 grams, but the amount selected
must be duplicated within +/- 0.5 gram for each
run. Instead of weighing the bucket, it can be filled from an automatic pipet, or from any other volumetric device if the repeatability of the filling
system is within +/- 0.5 ml.
www.parrinst.com
21
5
Operating Instructions
To speed and simplify the bucket filling process, and to conserve water and energy, Parr offers a closed circuit Water Handling System (No.
6510). This provides a water supply, cooled to the starting temperature and held in an automatic pipet ready for delivery in the exact amount needed to fill the bucket. Instructions for this automatic system are given in Operating Instruction No. 454M.
4. Set the bucket in the calorimeter. Attach the lifting handle (421A) to the two holes in the side of the screw cap and partially lower the bomb into the water. Handle the bomb carefully during this operation so that the sample will not be disturbed. Push the two ignition lead wires into the terminal sockets on the bomb head. Orient the wires away from the stirrer shaft so they do not become tangled in the stirring mechanism. Lower the bomb completely into water with its feet spanning the circular boss in the bottom of the bucket. Remove the lifting handle and shake any drops of water back into the bucket and check for gas bubbles.
Note: If bubbles continue to rise from the bomb after the air in the screw cap has escaped the test must be stopped and the bomb not fired until the leak has been cor­rected.
5. Close the calorimeter cover. This lowers the stirrer and thermistor probe into the bucket. Make sure that the bucket thermistor does not touch the bucket or 1108P when the lid is lowered.
6. Select determination or standardization as appropriate on the Calorimeter Operation menu by toggling the OPERATING MODE key. After pressing the STA RT key, the calorimeter will now prompt the operator for Bomb ID number, sample ID number, sample weight and spike weight in accordance with the instructions set into the operating controls page.
the status bar. Just before it fires the bomb, it will sound a series of short beeps to warn the user to move away from the calorimeter. Once the bomb has been fired, the status bar will display POSTPERIOD. The calorimeter will check to make certain that a temperature rise occurs and will then look for the final equilibrium conditions to be met. If it fails to meet either the initial or final equilibrium conditions, or if it fails to detect a temperature rise within the allotted time, the calorimeter will terminate the test and
advise the user of the error.
8. At the conclusion of the test, the calorimeter will signal the user.
9. Open the cover and remove the bomb and bucket. Remove the bomb from the bucket and open the knurled valve knob on the bomb head to release the residual gas pressure before attempting to remove the cap. This release should proceed slowly over a period of not less than one minute to avoid entrainment losses. After all pressure has been released, unscrew the cap, lift the head out of the cylinder and examine the interior of the bomb for soot or other evidence of incomplete combustion. If such evidence is found, the test will have to be discarded. Otherwise, wash all interior surfaces of the bomb, including the head, with a jet of distilled water and collect the washings in a beaker.
10. Titrate the bomb washings with a standard sodium carbonate solution using methyl orange, red or purple indicator. A 0.0709N sodium carbonate solution is recommended for this titration to simplify the calculation. This is prepared by dissolving 3.76 grams of Na2CO3 in the water and diluting to one liter. NaOH or KOH solutions of the same normality may be used.
11. Analyze the bomb washings to determine the sulfur content of the sample if it exceeds 0.1%. Methods for determining sulfur are discussed in Operating Instructions No. 207M.
7. The calorimeter will now take over and conduct the test. During the time it is establishing the initial equilibrium, it will display PREPERIOD on
22
Parr Instrument Company
12. At the end of the testing period, turn OFF the calorimeter at the power switch.
6200
Operating Instructions
5

Samples and Sample Holders

Particle Size and Moisture Content. Solid samples burn best in an oxygen bomb when reduced to 60 mesh, or smaller, and compressed into a pellet with a 2811 Parr Pellet Press.
Large particles may not burn completely and small particles are easily swept out of the capsule by turbulent gases during rapid combustion.
Note: Particle size is important because it in­fluences the reaction rate. Compression into a pellet is recommended because the pres­sure developed during combustion can be reduced as much as 40% when compared to the combustion of the material in the powder form. In addition in giving controlled burn rates, the pelletizing of samples keeps the sample in the fuel capsule during combus­tion.
Materials, such as coal, burn well in the as-received or air-dry condition, but do not burn completely dry samples. A certain amount of moisture is desirable in order to control the burning rate. Moisture con­tent up to 20% can be tolerated in many cases, but the optimum moisture is best determined by trial combustions.
If moisture is to be added to retard the combustion rate, drop water directly into a loose sample or onto a pellet after the sample has been weighed. Then let the sample stand for awhile to obtain uniform distribution.
Also, when benzoic acid is combusted for stan­dardization runs or for combustion aid purposes, it should be in the form of a pellet to avoid possible damage to the bomb which might result from rapid combustion of the loose powder.

Oxygen Charging Pressure

The 6200 Calorimeter has been designed to operate with an oxygen filling pressure of 30 atm. Signifi­cant changes from this value are not recommended.

Combustion Capsules

Non-volatile samples to be tested in Parr oxygen bombs are weighed and burned in shallow capsules
measuring approximately 1” diameter and 7/16”
deep. These are available in stainless steel, fused silica, fused quartz, and platinum alloyed with
3-1/2% rhodium.
Stainless steel capsules 43AS are furnished with each calorimeter. When combusting samples that contain metal particles such as aluminum or mag­nesium, the non-metallic fused silica 43A3 capsule or fused quartz 43A3KQ capsule is required. When superior corrosion resistance is needed, the plati­num rhodium 43A5 capsule is required.
The stainless steel capsules will acquire a dull gray finish after repeated use in an oxygen bomb due to the formation of a hard, protective oxide film. This dull finish not only protects the capsule, but it also promotes combustion and makes it easier to burn the last traces of the sample.

Combustion Aids

Some samples may be difficult to ignite or they may burn so slowly that the particles become chilled below the ignition point before complete combus­tion is obtained. In such cases benzoic acid, white oil or any other combustible material of known purity can be mixed with the sample. Ethylene glycol, butyl alcohol or decalin may also be used for this purpose.
Note: It must be remembered, however, that a combustion aid adds to the total energy released in the bomb and the amount of sample may have to be reduced to compen­sate for the added charge.
Capsules should be monitored for wear. Do not use the capsule if the wall or base thickness is less than
0.025”.
New capsules are heated in a muffle furnace at 500 ºC for 24 hours by Parr to develop this protec­tive coating uniformly on all surfaces. This treatment should be repeated after a capsule has been pol­ished with an abrasive to remove any ash or other surface deposits. Heating in a muffle is also a good way to destroy any traces of carbon or combustible matter which might remain in the capsule from a previous test.
Note: After heating, place the capsules in a clean container and handle them only with forceps when they are removed to be weighed on an analytical balance.
www.parrinst.com
23
5
Operating Instructions

Foodstuffs and Cellulosic Materials

Fibrous and fluffy materials generally require one of three modes of controlling the burn rate. Fibrous materials do not pelletize readily and generally require either moisture content or a combustion aid such as mineral oil to retard the burn rate and avoid development of high pressures.
Partial drying may be necessary if the moisture con­tent is too high to obtain ignition, but if the sample is heat sensitive and cannot be dried, a water soluble combustion aid such as ethylene glycol can be added to promote ignition.

Coarse Samples

In most cases it may be necessary to burn coarse samples without size reduction since grinding or drying may introduce unwanted changes. There is no objection to this if the coarse sample will ignite and burn completely. Whole wheat grains and coarse charcoal chunks are typical of materials which will burn satisfactorily without grinding and without additives or a special procedure.

Corrosive Samples

The 1108P Oxygen Combustion Vessel is made of a corrosion resistant alloy designed to withstand the corrosive mixture of sulfuric and nitric acids pro­duced in normal fuel testing operations. Samples containing chlorine and particular samples contain­ing more than 20 mg of chlorine samples with high sulfur contents will greatly accelerate corrosion of the bomb. An alternate 1108PCL is available con­structed of an alloy selected to specifically resist the corrosive effects of samples with high chlorine or chloride.
While no material will offer complete corrosion resistance to these samples, the 1108PCL offers significantly enhanced corrosion resistance for this service.

Explosives and High Energy Fuels

The 1108P and 1108PCL used in the 6200 Calorimeter have been designed to provide highly automated testing of routine samples. Materials which release large volumes of gas which detonate with explosive force or burn with unusually high energy levels should not be tested with these bombs. Rather, they should be tested in a model 1104 High Pressure Oxygen Combustion Vessel designed specifically for these types of samples.

Volatile Sample Holders

Volatile samples can be handled in a Parr 43A6 platinum capsule with a spun rim, or in a Parr 43AS stainless steel capsule which has a sturdy wall with a flat top rim. These holders can be sealed with a disc of plastic adhesive tape prepared by stretching tape across the top of the cup and trimming the excess with a sharp knife. The seal obtained after pressing this disc firmly against the rim of the cup with a flat blade will be adequate for most volatile samples.
The tape used for this purpose should be free of chlorine and as low in sulfur as possible. Borden Mystic Tape, No. M-169-C or 3M Transparent Tape, No. 610, are recommended for this purpose. The 3M Transparent Tape can be ordered through Parr, Part No. 517A.
The weight of the tape disc must be determined separately and a correction applied for any elements in the tape which might interfere with the deter­mination. The approximate Heat of Combustion of
the tape is 6300 cal/g. An actual amount should be
determined by running a blank test with tape alone using a sample weighing 1.0 gram. The compensa­tion for heat of tape may be done through the spike option; see Spike Controls, Line 2 - Heat of Combus­tion of Spike.
Note: Tape should always be stored in a sealed container to minimize changes in its moisture and solvent content.
24
Parr Instrument Company
6200
Operating Instructions
5

Figure 5-7

Combustion Capsule with Adhesive Tape Seal

Use the following procedure when filling and han­dling any of these tape-sealed sample holders:
Weigh the empty cup or capsule; then cover the top with tape, trim with a knife and press the trimmed edge firmly against the metal rim. Also cut and attach a small flag to the disc (see Figure 5-7).
Puncture the tape at a point below the flag, then re-weigh the empty cup with its tape cover.
Add the sample with a hypodermic syringe; close the opening with the flag and re-weigh the filled cup.
Set the cup in the capsule holder and arrange the auxiliary fuse so that it touches the center of the tape disc.
Just before starting the test, prick the disc with a sharp needle to make a small opening which is needed to prevent collapse of the disc when pressure is applied.
Fill the bomb with the usual oxygen charging pressure.
The calorimeter will fire the bomb and complete the test in the usual manner.

Poor Combustion

Because of the difference in combustion char­acteristics of the many different materials which may be burned in an oxygen bomb, it is difficult to give specific directions which will assure complete combustions for all samples.
The following fundamental conditions should be considered when burning samples:
Some part of the sample must be heated to its ignition temperature to start the combustion and, in burning, it must liberate sufficient heat to support its own combustion regardless of the chilling effect of the adjacent metal parts.
The combustion must produce sufficient tur­bulence within the bomb to bring oxygen into the fuel cup for burning the last traces of the sample.
Loose or powdery condition of the sample which will permit unburned particles to be ejected dur­ing a violent combustion.
The use of a sample containing coarse particles which will not burn readily. Coal particles which are too large to pass a 60 mesh screen may not burn completely.
The use of a sample pellet which has been made too hard or too soft. Either condition can cause spalling and the ejection of unburned fragments.
Insufficient space between the combustion cup and the bottom of the bomb. The bottom of the cup should always be at least one-half inch above the bottom of the bomb or above the liquid level in the bomb to prevent thermal quenching.
Excessive moisture or non-combustible material in the sample. If the moisture, ash and other non combustible material in the sample amounts to approximately 20% or more of the charge, it may be difficult to obtain complete combustion. This condition can be remedied by adding a small amount of benzoic acid or other combustion aid.
Volatile samples are defined as one with an initial boiling point below 180 °C per ASTM D-2.
Low volatile samples with a high water content, such as urine or blood, can be burned in an open capsule by absorbing the liquid on filter paper pulp or by adding a combustion aid, such as ethylene glycol.
www.parrinst.com
25
Loading...
+ 61 hidden pages