Parr Instrument 3900 User Manual

Page 1
3900
Hydrogenation Apparatus
Operating Instruction Manual
271M
3910 Hydrogenation Apparatus
3920 Hydrogenation Apparatus
Page 2

Table of Contents

Table of Contents
Preface
Scope — 3
Related Instructions — 3
Safety Information — 3
Explanation of Symbols — 4
General Specifi cations — 4
Environmental Conditions — 4
Provisions for Lifting and Carrying — 4
Intended Usage — 5
The User’s Responsibility — 5
Unpack Carefully — 5
Cleaning & Maintenance — 5
Assembly Instructions — 6
Operating Procedure — 7
Standardization — 7
Pressure Limits — 8
Safety Barricades — 8
General Precautions — 8
Catalysts — 9
Bottle Heater — 9
Water Jacket — 9
Temperature Measurement and Control — 10
Semi-micro Operations — 11
Alternate Parts — 11
Maintenance Instructions
Fuse Ratings — 11
Special Instructions for the 3921 Hydrogenator
Motor and Gear Box — 12
Bottle Connectors — 12
Glass Reaction Bottles — 12
Stainless Steel Bottle — 13
Bottle Heater — 14
Temperature Measurement — 14
References — 14
Parts Lists
3910 Hydrogenation Apparatus — 16
3920 Hydrogenation Apparatus — 18
Bottle Clamp Assembly - Series 3910 — 20
Shaker Column Assembly - Series 3910 — 21
Flywheel Assembly - Series 3910 — 22
Connecting Rod Assembly - Series 3920 — 23
AA92CA Single Valve Assembly — 24
3A93CA Double Valve Assembly — 25
3910 Wiring Schematic — 26
3920 Wiring Schematic — 27
Explosion Proof Switch Assembly for 1765EEG 115V ­Series 3910 — 28
Explosion Proof Switch Assembly for 1765EEK 220V -
Series 3910 — 29
Explosion Proof Switch Assembly for A388EEG 115V
& 220V - Series 3920 — 30
Explosion Proof Switch Assembly for A388EEK 220V -
Series 3920 — 31
Thermocouple Assemblies — 32
Reaction Bottles for Parr Hydrogenators — 33
Heating Mantles — 33
Air Motor Assembly — 34
Switch Box Assembly 115V — 35
Switch Box Assembly 230V — 35
Customer Service:
Questions concerning the installation or operation of this instrument can be answered by the Parr Customer Service Department:
1-309-762-7716 • 1-800-872-7720 • Fax: 1-309-762-9453
E-mail: parr@parrinst.com • http://www.parrinst.com
Parr Instrument Company
Page 3
Product Line

Preface

Scope

These instructions describe the installation, opera­tion and maintenance of Parr 3911 or 3921 Hydro-
genators offered with bottle sizes of 250, 500, 1000, 1700, 2000 or 2250 mL. They cover the basic steps
to be followed when installing these reactors and describe the function of all standard components. They are intended to be used in conjunction with several related instruction sheets listed below. This information describes several components that are common to most Parr pressure reaction equipment, and includes safety precautions and other related information applicable to all reaction laboratories. The users should study all of these instructions carefully before starting to use these hydrogenators so that they will fully understand the capabilities and limitations of the equipment.

Related Instructions

The following Parr publication is also included to further your understanding of this instrument and its component parts:

Safety Information

To avoid electrical shock, always:
1. Use a properly grounded electrical outlet of cor­rect voltage and current handling capability.
2. Ensure that the equipment is connected to elec­trical service according to local national electri-
cal codes. Failure to properly connect may create
a fi re or shock hazard.
3. For continued protection against possible haz- ard, replace fuses with same type and rating of fuse.
4. Disconnect from the power supply before main­tenance or servicing.
To avoid personal injury:
1. Do not use in the presence of fl ammable or com-
bustible materials; fi re or explosion may result.
This device contains components which may ignite such material.
2. Refer servicing to qualifi ed personnel.
No. Description
201M Limited Warranty
311M Supplemental Instructions for the Parr
4833 Controller
448M CAL 9500P Manual
www.parrinst.com
Page 4
Hydrogenation Apparatus

Explanation of Symbols

I On Position
O Off Position
This CAUTION symbol may be present on the Product Instrumentation and literature. If present on the product, the user must consult the ap­propriate part of the accompanying product literature for more infor­mation.
This CAUTION symbol indicates that the surface may be hot.
Protective Earth (PE) terminal. Provided for connection of the protec-
tive earth (green or green/yellow) supply system conductor.
General Specifi cations
Electrical Ratings
3911 & 3921 Rated:
115 Vac, 60 Hz, 8.0 Amps or 115 Vac, 60 Hz, 3.6 Amps or 115 Vac, 60 Hz, 3.3 Amps or 230 Vac, 50 Hz, 2.1 Amps or 230 Vac, 50 Hz, 8.0 Amps
The electrical ratings are identifi ed on the data plate of the instrument.
Before connecting any Parr Hydrogenation Appara­tus to an electrical outlet, the user must be certain that the electrical outlet has an earth ground con­nection and that the line, load and other character-
istics of the installation do not exceed the following
limits:
Voltage: Fluctuations in the line voltage should not exceed 10% of the rated nominal voltage shown on
the data plate.
Frequency: Hydrogenation apparatus can be oper-
ated from a 50 or 60 Hertz power supply without
affecting the apparatus. The frequency ratings are identifi ed on the data plate of the instrument.

Environmental Conditions

This apparatus is to be used indoors.
Operating: 10 ºC to 35 ºC; maximum relative humid­ity of 80% non-condensing. Installation Category II (over voltage) in accordance with IEC 664. Pollution degree 2 in accordance with IEC 664.
Altitude Limit: 2,000 meters.
Storage: -25 ºC to 65 ºC; 10% to 85% relative humidity.
Electromagnetic Compatibility In accordance with
2004/108/EC.

Provisions for Lifting and Carrying

The 3900 Series Hydrogenation Apparatus and its components are very heavy. Before moving ensure all cables are disconnected. Use proper and safe lifting techniques when installing or moving the ap-
paratus and/or its components.
CAUTION! Do not use in hazardous atmospheres.
Current: The total current drawn should not exceed
the rating shown on the data plate by more than 10 percent.
Parr Instrument Company
Page 5
Hydrogenation Apparatus

Intended Usage

This apparatus has been designed, built, and tested to strict physical and electrical standards. However, it is the user’s responsibility to install and operate it in conformance with local pressure and electrical codes. If this equipment is used in a manner be­yond its intended usage, the protection provided by the equipment may be impaired.

The User’s Responsibility

The user must realize that it is their responsibility to keep their equipment in good condition and to use it only within the prescribed temperature and pressure limits. They must be constantly aware of the serious consequences that can result from such
things as: opening the wrong valve, mixing com­bustible vapors with air or oxidizing gases, adding
reactants too fast or failing to observe and prevent sudden increases in temperature or pressure. Quali­fi ed personnel should make frequent checks to be sure that all safety rules are being observed. In the absence of a supervised safety program the user must take time to become completely familiar with their equipment and to consider any hazards inher­ent in the reactions they intend to perform.

Unpack Carefully

Unpack the equipment carefully and check all the parts against the packing list. If shipping damage is discovered, report it immediately to the deliver-
ing carriers. Examine the components closely for
any loose parts or shipping damage and be sure to check all layers of packing materials thoroughly so as not to overlook any parts which might otherwise be discarded.

Cleaning & Maintenance

Periodic cleaning may be performed on the exterior
surfaces of the instrument with a lightly dampened cloth containing mild soap solution. All power should be disconnected when cleaning the instru­ment.
There are no user serviceable parts inside the prod­uct other than what is specifi cally called out and discussed in this manual. Advanced troubleshooting instructions beyond the scope of this manual can be obtained by calling Parr Instrument Company in order to determine which part(s) may be replaced or serviced.
CAUTION! Ensure that any hot surfaces have had adequate time to cool before cleaning or maintaining the hydrogenation apparatus and/or its components.
www.parrinst.com
Page 6
Hydrogenation Apparatus

Assembly Instructions

Rinse or blow air through the hydrogen tank to remove any dust or foreign material. Likewise, blow out the gas hose and valves to be sure that they are dust-free and clean.
Set the apparatus on a sturdy bench or table where there is convenient access to an electrical outlet with the appropriate supply voltage and current ratings in accordance with national and local electrical code
requirements. The supply voltage must not exceed
the marked nominal voltage shown on the instru-
ment by more than 10%. The supply voltage recep-
tacle must have an earth ground connection.
Attach the valves to the hydrogen tank and tighten the couplings fi rmly with a wrench. This tank is fi lled through the AA92CA valve which is attached to the right end. The gage on this valve shows the tank pressure when the valve is closed. The gage on the 3A93CA valve at the left end of the tank shows the pressure in the reaction bottle and connecting tube. The front knob on the 3A93CA valve controls the fl ow of gas from the tank to the bottle. The rear knob at the opposite end of the block is used when dis­charging gas from the bottle or when evacuating the bottle through the hose nipple.
A 6-ft hose (A118CA) is furnished for connecting the
apparatus to the pressure regulator or needle valve on a commercial hydrogen tank. Screw one end of the hose into the socket in the AA92CA valve and connect the other end to the pressure regulator or
tank valve. The 1/8” NPT pipe threads on the ends of
the pressure hose should be coated with Tefl on tape, plastic lead or other thread dope to ensure tight
seals. The 1/8” to 1/4” bushing can be removed from
the hose if it is not required but do not unscrew the fi ttings which anchor the end nipples to the hose itself.
Before using a new apparatus for the fi rst time, assemble it with an empty bottle and test for gas
leaks, as follows: Fill the tank with hydrogen to 40 psig; then close the tank fi lling valve. Open the
bottle valve and allow the full tank pressure into the bottle; then close the bottle valve. The tank and bottle gages should now read the same. Record these pressures and continue to observe them over
a period of four to six hours. The pressures should
remain constant throughout this period unless there is a signifi cant change in room temperature. If a noticeable drop is observed in either gage, fi nd and correct the leak before starting to use the apparatus. Any leaks can usually be detected by brushing the joints with a soap solution.
The 25-inch length of polypropylene tubing leading
to the reaction bottle must be fi rmly connected to the outlet fi tting on the 3A93CA valve. No special clamps or fi ttings are required to fasten this tube into the bottle stopper. Simply slide the end of the
tube through the 61CA4 washer and through the
166CA retaining ring, then push the tube through the one-hole Neoprene stopper leaving a projection of about one inch below the bottom of the stopper. Slide the safety screen over the bottle; place the bottle into the holder and tighten the thumb nuts on the ends of the two tie rods. The clamping pressure developed by the thumb nuts will hold the bottle in the shaker mechanism and it will also anchor the connecting tube in the stopper.
Parr Instrument Company
Page 7
Hydrogenation Apparatus

Operating Procedure

Samples to be treated in a Parr hydrogenator are placed in a reaction bottle with a catalyst and clamped in a shaking mechanism. A gas connection is made to the bottle from a multiple valve and all air is removed either by evacuation or by fl ushing with hydrogen. Hydrogen is then introduced from
a 4-liter reservoir while the bottle is shaken vigor-
ously to initiate the reaction. Heating or cooling can be applied, if necessary. After the reaction reaches the desired point the shaker is stopped, the bottle vented and the product and catalyst are recovered.
The individual steps in this operating procedure are listed below. These can be varied to suit each indi­vidual application.
1. Place the catalyst, solvent and sample in the reaction bottle, adding the catalyst fi rst to avoid possible vapor ignition by the dry catalyst. The
total volume of solution should not exceed two-
thirds the capacity of the bottle.
2. Attach the stopper with connecting tube and slide the bottle into the guard screen, then set the assembly in the bottle holder and tighten the knurled clamping nuts.
3. If air is removed from the bottle by evacua­tion, attach a vacuum hose to the nipple on the 3A93CA valve; close the bottle valve: open the gas release valve and evacuate until the solvent starts to boil. If a low boiling solvent is not used, evacuate to a negative pressure suffi cient to remove most of the air. Air can also be removed by alternately fi lling the bottle with hydrogen to 20 or 30 psig. and venting it at least three times. After purging the bottle, close the gas release valve and leave it closed throughout the run.
4. Starting with the hydrogen tank fi lled to 30 psig. open the bottle valve and read the bottle pressure gage after equilibrium has been estab­lished.
5. Start the shaker and follow the progress of the reaction by observing the bottle pressure gage. If complete hydrogenation is desired, continue shaking until there is no further pressure drop.
For partial or quantitative hydrogenation, contin-
ue shaking until the pressure drops to a calculat­ed value as determined by prior standardization runs.
6. At the end of the run, stop the shaker; close the bottle valve and allow the catalyst to settle. Any residual pressure in the bottle and connect­ing tube can be discharged by opening the gas release valve.
7. Open the bottle clamp and remove the bottle. Decant the solution leaving the catalyst in the bottle for a second reduction, or remove the catalyst on a fi lter.

Standardization

The apparatus can be standardized by making a preliminary run with a known amount of any com­pound that can be completely and quantitatively reduced. The pressure drop per mole of hydrogen consumed in such tests is then used as a basis for estimating the progress of a reaction with unfamil­iar materials. A procedure is described in reference (3) for standardizing the apparatus by reducing 11.6 grams (0. 1 mole) of pure maleic acid dissolved in
150 mL of 95% ethanol using 0.1 gram of catalyst.
The reaction is carried out as previously described with shaking continued until no more hydrogen is consumed. This usually takes twenty to thirty min­utes, after which the pressure drop in the tank is
recorded. Since exactly 0.1 mole of hydrogen has
been consumed in this run, this decrease in tank pressure can be used as a basis for measuring or regulating the amount of hydrogen consumed when treating other compounds.
Standardization tests are not limited to runs with
maleic acid. Fumaric acid is equally suitable for
this purpose, or any other pure compound can be used provided that it is completely or quantitatively reduced. If the intended usage for the apparatus involves the consumption of only a small amount of hydrogen, the apparatus can be standardized with the tank valve closed. The gage will then give a more signifi cant reading as gas is consumed from the connecting tube and bottle alone while these parts are isolated from the tank.
www.parrinst.com
Page 8
Hydrogenation Apparatus
Gas Tank and Valve System

Pressure Limits

The use of glass bottles in these reactors introduces certain pressure limitations and a potential hazard which the user must understand. Although each bottle is pressure tested before it is sold, the physi­cal characteristics of glass are such that it is impos­sible to guarantee these bottles against breakage
or to predict their service life. For this reason, each
apparatus is equipped with a bottle shield to restrain fl ying glass in case of breakage. The user must take whatever additional precautions he considers neces­sary to protect himself from injury in case a bottle
should unexpectedly fail. Working pressures should never exceed 60 psig when using either 250 or 500
mL bottles in the 3911 apparatus, and never more
than 40 psi for one liter bottle and 30 psi for two liter
bottle in the 3921 apparatus.

Safety Barricades

If a barricade is used it should be built of concrete, brick or steel in whatever thickness or form is con­sidered necessary to protect the operator from fl y-
ing fragments if the reaction bottle should explode.
Glass shields, either plain or reinforced with wire mesh, are not recommended. The requirements for barricades differ so widely that each should be designed and built in order to protect against the potential hazards inherent in each installation. This subject is well covered in the references listed on
page 15.

General Precautions

Pressure reactions with hydrogen are not unduly hazardous if the user maintains his hydrogenator in good condition and operates it with the realization that hydrogen is highly fl ammable and that pres­sures and reaction rates must be carefully controlled at all times.
Parr shaker type hydrogenators are usually operated in an open laboratory without additional barricades or protective screens, but the operator must realize that additional protection may be necessary if there is any possibility that a reaction might run out of
control, or if unexpected bottle breakage would pro­duce a hazardous spill of toxic or fl ammable materi­als. Potentially explosive reactions are best handled
with the apparatus located behind a suitable barri­cade or in a pressure test cell.
Parr Instrument Company
All catalysts must be handled cautiously because of their highly reactive nature. Do not add dry catalyst to a bottle containing a fl ammable solution or vapor. The vapor might ignite. Instead, add the catalyst fi rst and cover it immediately with the sample in solution. Precautions must also be taken to wash the catalyst from the thermocouple, the inlet tube and the stopper when opening the bottle. Any catalyst
left on these parts may ignite when exposed to the
Page 9
Hydrogenation Apparatus
air. If breakage or spills occur, fl ush the contami­nated area immediately with large volumes of water and keep the area wet until all traces of catalyst have been removed.
Vacuum fi ltration through a paper fi lter can be dangerous if air is drawn through the fi lter in the presence of a catalyst. To reduce this hazard, keep the paper covered with solution while it is under suction, or use a fi lter made of a non-fl ammable material.
Care must be taken to keep the apparatus free of im­purities which might poison the catalyst. Although a small amount of air trapped in the bottle will not interfere with most hydrogenations, the reaction rate can sometimes be improved by evacuating the bottle before adding hydrogen. Or the air can be removed by fi lling the bottle with hydrogen to 20 or 30 psig and venting it at least three times before starting the shaker.
If the reaction proceeds too rapidly it can usually be checked by stopping the shaker. If overheating becomes a dangerous problem, the bottle can be cooled by using the A103CA water jacket described below.
There should be no gas burners or open fl ames near a hydrogenation apparatus. The room must be well ventilated and any gas released from the apparatus
should be discharged into an explosion proof hood
or ventilating duct. Care must be taken to prevent ignition by a static charge from an insulated object.
For this reason, a good ground connection through
the power cord or directly to the base of the appara­tus must be maintained at all times.
Loss of gas is an annoying factor which can be avoided by careful maintenance and frequent test-
ing. It should not be necessary to use extreme force
to close any of the valves on this apparatus. If a tight seal cannot be secured without a hard turn on the valve handle, dismantle the valve and replace the 20VB valve seat and any other worn or damaged parts. If the valve leaks through the packing, back
the needle away from its seat and tighten the 8VB2
packing nut. If this does not stop the leak replace the
4VB3 packing rings.

Catalysts

The most active catalysts for hydrogenation reac­tions in this apparatus are made of platinum and
palladium. A hydrated platinum oxide, often called
Adams catalyst, is used in many procedures. This is prepared by drying and heating chloroplatinic
acid in air to form a brown oxide which can then be
reduced to the more active black form by shaking
with hydrogen either before or after mixing with the
sample. Instructions for preparing this and other
catalysts are given in the references on page 15.
Raney nickel is used in certain procedures but other less active nickel catalysts require pressures outside
the range of this apparatus. Platinum oxide and
other catalytic chemicals can be obtained from most laboratory supply houses.

Bottle Heater

Bottle temperatures up to 80 ° C can be developed
in the 3911 Hydrogenator by wrapping a glass fabric
heating mantle (A450E) around the 500 mL bottle.
The mantle then takes the place of the perforated metal guard screen. No special fi ttings are needed for attaching this heater. Simply wrap the mantle tightly around the bottle and run the connecting cord through the slot in the bottle clamp. Use the
plastic cable clip (453E) furnished with the heater
to anchor the heater cord to the top of the bottle clamp, then run the cord out from the shaker pivot,
arranging it for minimum fl exing when the shaker is
operating.

Water Jacket

The 500 mL reaction bottle can be
cooled while it is clamped in the shaker mechanism by installing an A103CA water jacket. This jacket fi ts into the bottle clamp in place of the
65CA perforated steel guard. A soft
rubber ring seals the neck of the bottle into the jacket. The only change required when adding a water jacket is to use the special cut-down stop-
per which is furnished, or cut 1/4 inch
from the top of the regular 62CA stopper to shorten the overall height of the combined assembly. Al­ways install the soft rubber sealing ring with the large diameter downward and the smaller diameter at the top: otherwise it will be very diffi cult to re­move the bottle from the cooling jacket.
A103CA
Water Jacket
www.parrinst.com
Page 10
Hydrogenation Apparatus

Temperature Measurement and Control

4833 Temperature Controller
Reaction temperatures can be measured and con-
trolled with a Parr 4833 Temperature Controller which operates with an A295E, Type J (iron-constan-
tan) thermocouple installed in the reaction bottle.
The thermocouple is sealed in an 1/8” diameter stainless steel sheath and held in an A159CA bottle
connector assembly which carries both the probe and the gas passage through a single opening in
the bottle stopper. A295E thermocouples are made
in two lengths to fi t different bottle sizes as listed on page 30.
To install the A159CA bottle connector, remove the existing gas inlet tube and fi ttings and install the
new fi ttings which are provided. Clamp the bottle in the shaker with the thermocouple positioned so that the tip of the probe reaches a point three-fourths of the distance from the neck of the bottle. This dis­tance can be changed by loosening the cap nut at the top of the connector and sliding the probe up or down in the tee fi tting.
Loop the thermocouple wire downward and fasten it above the tee fi tting using one of the cable ties fur­nished with the bottle connector; then run the wire along the polypropylene tube, fastening it fi rmly
to the tube at several points to prevent excessive fl exing and breakage. Bring the bottle heater cord
upward and fasten it under the clip at the top of the shaker; then run the cord parallel to the shaker piv­ots and plug it into the heater socket on the control­ler.
To activate the heater, plug the thermocouple and the heating mantle into sockets on the rear panel of the controller, and connect the controller to an electric supply that corresponds to the voltage of the heater. Turn the DISPLAY and HEATER switches ON, and set the controller to the desired operating temperature using the UP and DOWN arrow keys on the front of the controller module. The indicator light in the heater switch will illuminate when the switch is ON and current is being supplied to the heater. The set point can be changed at any time regardless of whether the heater is on or off. Copies of Parr Bulletin 311M, Supplemental Instructions for
the 4833 Controller, and the CAL-9500 User’s Manual furnished with the 4833 Controller provides instruc-
tions for tuning the Controller and using the alarm features.
The 4833 Temperature Controller has dual displays:
one for the process temperature and the other for the set point. The operating range of the Controller
covers the span from 0 ° to 100 ° C with the high limit extending somewhat beyond the temperature
limit for the 3911 and 3921 Hydrogenators. These hy­drogenators should not be operated at temperatures
above 80 ° C.
The 4833 Controller is sensitive to temperature changes of less than 0.5 ° C but temperature varia-
tions in the reaction bottle may be larger than this due to lag in the heating system and turbulence in
the bottle. In most cases temperatures in a 500 mL
bottle will not vary by more than one degree above or below the set point. Variations in larger bottles may run as much as two degrees, but this is still within permissible limits for most reactions. Bottle temperatures will tend to overshoot at low tempera-
tures in the 30° to 40° range. There will be less over-
shoot at higher settings. Overshoot can be avoided by setting the controller several degrees low during the initial period, then raising the temperature in one or two steps after cycling has been established slightly below the working level.
Fail-safe protection against a thermocouple break
is provided in the controller. This can be tested by disconnecting the thermocouple from the controller, which should turn off the heater.
10
Parr Instrument Company
Page 11
Hydrogenation Apparatus

Semi-micro Operations

Small samples can be treated in a 3911 Hydrogena-
tor by replacing the standard 500 mL bottle with a smaller 250 mL bottle (66CA2). This requires a
101CA2 spacer to compensate for the difference in
bottle heights. To install the 250 mL bottle, remove the stopper and the 61CA4 washer from the gas
inlet tube and slide the 101CA2 spacer onto the tube, then clamp the bottle and spacer in the holder in the usual manner. Since small samples may require only small amounts of hydrogen, it may be advanta­geous to replace the standard four liter hydrogen tank with an alternate one liter tank (A16CA2) for semi-micro operations. The same tank valves are used on both sizes and are easily transferred to the small tank.
Augustine (reference 1 on page 15) describes an
interesting holder for handling small samples in a 3911 Hydrogenator which he prepares by sealing a
50 mL or smaller heavy walled fl ask inside a larger
bottle.

Alternate Parts

If bottle pressure measurements are not required, the gage can be removed from the 3A93CA valve
assembly and replaced with a 94CA plug.
Tefl on tubing can be furnished in place of the stan­dard 119CA polypropylene tube for use with chemi­cals which might attack polypropylene. The same fi ttings are used to attach either Tefl on or polypropyl­ene tubing to the 3A93CA valve.
about once a month. Lubricate the fl ywheel shaft by placing a few drops of light oil in the oil cup at regular intervals. A light application of a lithium
grease such as “Lubriplate” on the shaker pivots is
also advisable. The spacing of these pivots should be adjusted so that the bottle clamp swings freely
without excessive friction.
To inspect and replace the valves, unscrew the 8VB2
packing nut and remove the needle and knob. The internal parts can then be removed with a small wire hook. These will come out in the following se-
quence: 6VB packing cover, two 4VB3 packing rings,
21VB lantern ring, and 20VB valve seat. If the plastic valve seat will not slide out of its socket, use a 1
1/2” wood screw as a removal tool. Replace these
parts in the same order; insert the valve needle and
tighten the 8VB2 packing nut fi rmly with a wrench.
Caution: Always back the valve needle away from its seat before tightening the packing nut.
It will be necessary to use a new plastic ferrule in the A102CA connector whenever a new polypropyl­ene tube is installed. A new A102CA connector is
furnished with each replacement A154CA polypro-
pylene tube so that a complete set of parts will be available when changing tubes. If the body and cap nut in the old connection are in good condition they can be used with the ferrule from the new connec­tor. But if there is any question about the old parts, discard them and install a new connector. The steps required to attach the tube to the connector are as follows: Be sure that the end of the tube has been cut squarely; then disassemble the connector and slide the cap nut and ferrule onto the tube. Insert the tube into the connector body and tighten the cap nut fi rmly.

Maintenance Instructions

Periodic cleaning may be performed on the exterior
surfaces of the instrument with a lightly dampened cloth containing mild soap solution. All power should be disconnected when cleaning the instru­ment. There are no user serviceable parts inside the product other than what is specifi cally called out and discussed in this manual. Advanced trouble­shooting instructions beyond the scope of this manual can be obtained by calling Parr Instrument Company in order to determine which part(s) may be replaced or serviced.
The connecting rod has oil-impregnated bronze bearings which do not require heavy lubrication. Place a drop or two of light oil on each bearing

Fuse Ratings

The replacement of protective fuses should be per­formed by qualifi ed personnel.
Parr No. Type Ratings
139E21 Slo-blo 5 Amps, 250 Vac 139E20 Slo-blo 4 Amps, 250 Vac
139E8 Slo-blo 2.5 Amps, 250 Vac 139E25 Slo-blo 10 Amps, 250 Vac 139E26 Slo-blo 6 Amps, 250 Vac
Note: Check the labels on Switch Box for cor­rect fuse rating.
www.parrinst.com
11
Page 12
Hydrogenation Apparatus

Special Instructions for the 3921 Hydrogenator

Customer installed location of
Blue Plastic Vent Plug
All instructions given in the preceding sections of this Manual apply equally to both the 3911 and 3921 Hydrogenators. Both models use the same hydro­gen tank and valves, but the 3921 apparatus has a larger and heavier bottle clamp and shaker mecha­nism to accommodate larger reaction bottles.

Motor and Gear Box

Before starting the motor, replace the uppermost
pipe plug on the gearbox with the blue plastic vent
plug which is furnished. The hole in the vent plug must point up. See Boston gear instruction sheets for lubricating instructions.

Bottle Connectors

Attach the valves to the gas tank and connect the polypropylene tube to the reaction bottle by slid-
ing the tube through the 82CA2 washer and 166CA
retaining ring, then push the tube through the bottle stopper leaving a projection of about one inch be­low the stopper. To complete the assembly, set the bottle on top of the rubber pad in the holder: slide the tube through the slot in the clamping screw and tighten the screw fi rmly. Always attach the steel guard screen to the front of the bottle holder before pressurizing the bottle.
An 80CA2 spacer must be placed between the
clamping screw and the stopper to compensate for the difference in bottle heights when using a 1000
mL bottle, Assemble the inlet tube with the 80CA2
spacer and 166CA retaining ring above the stopper and clamp the bottle in the shaker in the usual man-
ner. The 82CA2 washer is not required when using a
spacer.

Glass Reaction Bottles

The 1000 and 2000 mL reaction bottles (71CA and 72CA) supplied with this apparatus have been
individually pressure tested to 80 psig and 60 psig
respectively. In spite of these tests, the Parr Instru­ment Company cannot guarantee that these bottles will not break at lower pressures. The user must therefore be constantly aware of the hazards in­volved in handling large volumes of liquids in glass bottles under pressure, and he must take whatever precautions he considers necessary to protect him-
self from injury in case a bottle should unexpect-
edly fail. It is recommended that working pressures in these 1000 and 2000 mL bottles should never
exceed 40 psi and 30 psi respectively. Pressures should be held below this maximum whenever pos-
sible.
12
Parr Instrument Company
Page 13
Alternate 1000 and 2000 mL reaction bottles with a fi berglass covering (71CA2 and 72CA3) can be furnished for the 3921 Hydrogenator. These coated bottles are no stronger than the plain bottles, but the fi berglass envelope will usually retain any broken glass and prevent the loss of valuable reac-
tants in case of accidental breakage. The maximum
working pressures mentioned above applies to both plain and coated bottles alike. Both styles use the same connecting tube and fi ttings.
If higher pressures up to 60 psig are required for treating large amounts of reactants, users are urged
to purchase the special 2500 mL heavy duty bottle (72CA4) which is made specifi cally for this purpose.
This is a hand blown, borosilicate glass bottle with
an extra heavy wall which is much stronger than the
standard machine-made bottles.

Stainless Steel Bottle

Hydrogenation Apparatus
The breakage hazard which is always present when using glass bottles can be eliminated by substituting a 1700 mL stainless steel reaction bottle which can
be used at working pressures up to 60 psig maxi-
mum. The complete stainless steel bottle assembly with cover, spacer spool, connecting tube and fi t­tings can be ordered under Cat. No. A129CA3.
To install the stainless bottle, slide the 133CA2 spacer spool onto the polypropylene tube with the parts arranged so that the plain end of the spool will rest on the cover, leaving the shallow depression in the other end of the spool to engage the clamping screw. Attach the O-ring to the cover, close the bottle and slide the assembly into the holder. Tighten the clamping screw and attach the guard screen before pressurizing the bottle.
A129CA3 Stainless Steel Bottle Assembly
www.parrinst.com
13
Page 14
Hydrogenation Apparatus

Bottle Heater

The 71CA one liter bottle, 72CA two liter bottle and the A129CA3 stainless bottle assembly may be heated in the 3921 Hydrogenator by wrapping
a glass fabric heating mantle (A451E) around the
bottle. No special fi ttings are needed for attaching this heater. Simply wrap the mantle tightly around the bottle and run the connecting cord through the slot in the top housing. Use the plastic clip furnished with the heater to anchor the cord to the top edge of the bottle clamp: then run the cord out laterally on
the axis of the shaker, arranging it to avoid exces­sive fl exing when the shaker is operating.
Note: Connect cord to temperature controller to control operations of heater. See page 10 for instructions on Parr’s temperature controller and thermocouple.
The 72CA4 heavy wall bottle should not be used
with the heating mantle because the heat will de­stroy the protective polyvinyl coating on the bottle.

Temperature Measurement

Temperatures in the reaction bottle can be mea­sured by installing a stainless sheathed thermo­couple using the same fi ttings described for the 3911 Hydrogenator on page 10, but with a longer sheath on the thermocouple to match the larger bottles. The various assemblies are identifi ed in the Parts List on page 30.

References

No attempt is made here to list the hundreds of references to the Parr hydrogenator which have ap­peared in chemical literature since Dr. Roger Adams published his fi rst paper describing an apparatus of this kind in 1923. More than fi ve hundred literature references are cited in Augustine’s book on Catalytic Hydrogenation which is listed below. Additional references can be obtained from other books in this list. Among these, the books by Augustine,
Freifelder and Rylander will be particularly help-
ful to those users who want additional information regarding hydrogenation techniques, catalysts and procedures for treating specifi c functional groups. The following references are therefore highly rec­ommended:
R.L. Augustine, Catalytic Hydrogenation, Marcel
Dekker, Inc.: New York , 1965.
Morris Freifelder, Practical Catalytic Hydrogenation, Wiley-Interscience Div. of John Wiley & Sons, Inc.:
New York, 1971.
Gilman-Blatt, Organic Synthesis, Collective Volume I,
p. 65, John Wiley & Sons, Inc.: New York, 1948.
H.W. Lohse, Catalytic Chemistry, Chemical Publish-
ing Co., Inc.: New York, l945.
Paul N. Rylander, Catalytic Hydrogenation Over Platinum Metals, Academic Press: New York, 1967.
14
Paul N. Rylander, Catalytic Hydrogenation in Organic Synthesis, Academic Press: New York, 1979.
Paul N. Rylander, Hydrogenation Methods, Academ-
ic Press: New York, 1985.
Biennial conferences originally held at the New York Academy of Sciences and now in a variety of loca­tions by ORCS (Organic Reaction Catalysis Society) in even-numbered years starting in 1966 have pro-
duced excellent collections of papers under the gen-
eral title, Catalytic Hydrogenation and Analogous Pressure Reactions. The initial set is published in the Annals of The New York Academy of Sciences, Vol. 145, Art. 1, pp. 1-206, 1967. Currently titled Catalysis of Organic Reactions published by Marcel Dekker,
Inc.: New York.
Parr Instrument Company
Page 15
Notes
www.parrinst.com
15
Page 16
Hydrogenation Apparatus

Parts Lists

15
6
7
5
8
14
8
4
12
9
13
11
21
10
17
(4 PL)
16
3
(4 PL)
2
31
19
20
22
(2 PL)
45
(2 PL)
44
(4 PL)
24
(4 PL)
26
54
(4 PL)1852

3910 Hydrogenation Apparatus

39 38
51
16 35
43
48
49
32
30
41
33
42
28
34
27
(2 PL)
17
29
17
40
37
37
(2 PL)
24
(2 PL)
26
36
23
16
25B 25E
25A
25C
25D 25
Parr Instrument Company
53
56
531
Page 17
Hydrogenation Apparatus
ITEM NO.
1 221CA BASE PLATE 1
2 2CA2 SHAKER COLUMN 1
3 SB3118HX14 5/16-18 X 7/8 HHCS 8-18 SS 4
4 54CA BOTTLE CLAMP TOP 1
5 453E CLIP, 3/8 DIA TENSION CLEAR 1
6 SA1332RD04 6-32 X 1/4 RHMS SLOTTED 1
7 SW13FT #6 BOLT WASHER 18-8 SS 1
8 59CA KNURLED NUT, TIE ROD 2
9 A55CA TIE ROD SHORT ASSEMBLY 1
10 66CA REACTION BOTTLE 500ML 1
11 62CA
12 61CA4 WASHER, 1" ALUMINUM 1
13 166CA
14 101CA2 SPACER FOR 250ML BOTTLE 1
15 HL0025TB040 POLYPROPYLENE TUBE 2.08 FT
16 5CA3 PIVOT PIN 2
17 SC2520SC04 1/4-20 X 1/4 SHSS 18-8 SS 7
18 65CA BOTTLE GUARD T304, 500 mL 1
19 53CA2 BOTTLE CLAMP BASE 1
20 TD3118SC12 5/16 X 3/4 SH SHLD BOLT STEEL 1
21 56CA TIE ROD, LONG 1
22 SN3124HX 5/16-24 HEX JAM NUT 18-8 SS 1
23 1765EEF MOTOR 230/50 1/4HP 1425 RPM 1
24 SB3118HX10 5/16-18 X 5/8 HHCS 18-8 SS 6
25 X2291EB
25A 909E SWITCH, POWER 16AMP 1
25B 1471E
25C 911E FILTER, INTERFERENCE; 10AMP 1
25D 139E8
PART NUMBER
221CA2 BASE PLATE (AIR) 1
221CA3 BASE PLATE (EXP) 1
66CATC REACTION BOTTLE 500ML PED 1
66CA2 REACTION BOTTLE 250ML 1
66CA2TC REACTION BOTTLE 250ML PED
65CA2 BOTTLE GUARD T304, 250 mL
1765EEA MOTOR 115/60 1/6 HP 1725 RPM 1
1765EEG
1765EEK
X2353 AIR MOTOR PACKAGE 3910 1
DESCRIPTION QTY
STOPPER, #6 NEOPRENE 1-HOLE
RETAINING RING; TRIANGLE PUSH
MOTOR 115/60 EP 1/6HP 1725 RPM
MOTOR 230/50 EP 1/6HP 1425 RPM
SWITCH BOX PKG 115V(3910/3920)
CARRIER, FUSE 3AG 1/4 X 1-1/4
FUSE 3AG SLO-BLO 250V 2.5
AMP
ITEM NO.
25E 139E26 FUSE SLO-BLO 6 AMP 250V 1
26 SW31FT 5/16 BOLT WASHER 18-8 SS 6
27 4CA BASE, ALUM FLYWHEEL 1
28 8CA3 FLYWHEEL W/SHAFT 1
29 18CA2 COLLAR W/SET SCREWS 1
30 A58CA
31 SR3118
32 19CA3 CRANK PIN W/NUT 1
33 SN2520HX 1/4-20 HEX JAM NUT 1
34 50CA OIL HOLE COVER 1
35 A16CA HYDROGEN TANK 4L T304SS 1
36 49CA TANK SUPPORT 1
37 XC0210 TRIM, PROTECTIVE BLACK 2
38 11CA TANK CLAMP 1
39 SB3118HX24 5/16-18 X 1-1/2 HHCS 18-8 SS 1
40 34CA3
41 37CA3 VEE BELT, NOTCHED 25" 1
42 SA3118FT14 5/16-18 X 7/8FHMS 18-8 SS 3
43 A150CA BELT GUARD 1
44 SA2520RD10 1/4-20 X 5/8 RHMS 18-8 SS 2
45 SW25NL
46 SW16NL
47 SA1632RD08 8-32 X 1/2 RHMS 18-8 SS 2
48 AA92CA SINGLE VALVE ASSY W/GAGE 1
49 A118CA GAS HOSE ASSY, NYLON, 6FT 1
51 3A93CA DOUBLE VALVE W/GAGE 1
52 64PR FOOT, RUBBER; W/ 1/4-20 THRD 4
53 328E7 HOLE PLUG 1/2 2
57 504HCAA
PART NUMBER
139E21
139E20
X2291EE
A16CATC
A16CATD
DESCRIPTION QTY
FUSE 3AG SLO-BLO 250V 5.0
AMP
FUSE 3AG SLO-BLO 250V 4.0
AMP
SWITCH BOX PKG 230V (3910/3920)
CONNECTING ROD ASSY W/
BUSHINGS
5/16 DIA X 1-1/8 ROLLPIN 420SS
HYDROGEN TANK 4L T304SS
PED
HYDROGEN TANK 4L T304
(CHINESE)
MOTOR PULLEY W/SET
SCREW
1/4 LOCK WASHER, INT TOOTH 410S
#8 LOCK WASHER INT TOOTH 410SS
REDUCING BUSHING,
1/8NPTF-1/4NPTM
www.parrinst.com
17
Page 18
Hydrogenation Apparatus
(2 PL)
58
60
38
37
54
40
53
52
51
41
(4 PL)
39
(10 PL)
30
31
32
26
(4 PL)
33
34
25
35
23
(4 PL)
6
5
8
22
42
46
42
43
45
(2 PL)
48
44
47
24
(4 PL)
5
19
20
21
(4 PL)
17

3920 Hydrogenation Apparatus

26
34
33
32
31
30
25
29
36
18
10
15
14
11
7
5
16
9
59
9
4
12
13
1
2
18
52 86550 49
(4 PL)
Parr Instrument Company
(2 PL) (2 PL)
6 8
5027 28
(4 PL)
49B
49C
49D49A
Page 19
Hydrogenation Apparatus
ITEM NO.
1 73CA6 BASE PLATE 1
4 49CA TANK SUPPORT 1
5 SW31FT 5/16 BOLT WASHER 18-8 SS 12
6 TW31SL 5/16 LOCK WASHER STEEL PLATED 14
7 SB3118HX14 5/16-18 X 7/8 HHCS 18-8 SS 2
8 SN3118HX JAM NUT, 5/16-18 THD 7
9 XC0210 TRIM, PROTECTIVE BLACK 2
10 11CA TANK CLAMP 1
11 A16CA HYDROGEN TANK 4L T304SS 1
12 SA2520RD28 1/4-20 X 1-3/4 RHMS 18-8 SS 1
13 SN2520HX 1/4-20 HEX JAM NUT 18-8 SS 1
14 SW25FT 1/4 BOLT WASHER (FLAT) 18-8 SS 1
15 SN2520WG 1/4-20 WING NUT 18-8 SS 1
16 AA92CA SINGLE VALVE W/GAGE 1
17 180HC MOTOR SUPPORT 4
18 3A93CA DOUBLE VALVE W/GAGE 1
19 A387E2 GEAR BOX W/BASE 10-1 RATIO 1
20 1748HC DRIVE KEY MACH. .185 SQ X 1” LG 1
21 A388EEL MOTOR 110/220V 60 HZ 1/4 HP 1
22 220CA FLYWHEEL GUARD 1
23 SB3118HX10 5/16-18 X 5/8 HHCS 18-8 SS 8
24 SB3118HX36 5/16-18 X 2-1/4 HHCS 18-8 SS 4
25 92HC5 ROCKER COLUMN 2
26 TC2520SC04 1/4-20 X 1/4 SHSS ALLOY STEEL 4
27 TW75FT 3/4 BOLT WASHER STEEL PLATED 2
28 TB7516HX16
29 A76CA2 REAR GUARD ASSEMBLY 1
30 TB5020HX36 1/2-20 X 2-1/4 HHCS STEEL 2
PART NUMBER
X2290EA MOTOR GROUP 115/60
X2290EF MOTOR GROUP 230/50
X2290EG MOTOR GROUP 115/60EP
159X40EGX MOTOR GROUP 115/60 EPX
X2290EK MOTOR GROUP 230/50 EP
159X40EKX MOTOR GROUP 230/50 EPX
159X40EY MOTOR GROUP AIR
DESCRIPTION
3/4-16 X 1 HHCS STEEL GRD5
PLATED
QTY
2
ITEM NO.
31 427E4 BUSHING, INNER 2
32 1055HC SPHERICAL BEARING PTFE LINED 2
33 427E3 BUSHING, OUTER 2
34 SN5020HC 1/2-20 HEX CAP NUT 18-8 SS 2
35 74CA2 BOTTLE HOLDER BASE 1
36 172CA BOTTLE PAD NEOPRENE 1
37 75CA BOTTLE HOLDER TOP 1
38 79CA2 BOTTLE CLAMP SCREW 1
39 SB2520BT08 1/4-20 X 1/2” BHSCS 18-8 SS 10
40 A77CA2 FRONT GUARD ASSEMBLY 1
41 A83CA THUMB SCREW 1/4-20 X 1/2L 4
42 109CA2 SPACER 2
43 89CA2 CRANK PIN W/NUT 1
44 1065HC FLYWHEEL 1
45 SC3118SC06 5/16-18 X 3/8 SHSS CUP PT 18-8 2
46 SB3716SC16 3/8-16 X 1 SHCS 18-8 SS 1
47 SW39FT .625 X .390 X .063 WASHER 18-8 1
48 A87CA4 CONNECTING ROD ASSEMBLY 1
49 X2291 SWITCH BOX PACKAGE 1
49A 909E SWITCH, POWER 16 AMP 1
49B 1470E
49C 1471E CARRIER, FUSE 3AG 1/4 X 1-1/4 1
49D 911E FILTER, INTERFERENCE, 10AMP
50 328E7 HOLE PLUG 1/2 2
51 72CA REACTION BOTTLE 2000ML 1
52 62CA STOPPER, #6 NEOPRENE 1 HOLE 5
53 166CA
54 82CA2 WASHER 9/32” 1
58 HL0025TB040 POLYPROPYLENE TUBE 1
59 A118CA GAS HOSE ASSY NYLON 6FT 1
60 A102CA
PART NUMBER
71CA REACTION BOTTLE 1000ML
71CATC REACTION BOTTLE 1000ML PED
72CATC REACTION BOTTLE 2000ML PED
DESCRIPTION
FUSEHOLDER, INTERNATIONAL 1/4”
RETAINING RING, TRIANGLE PUSH
MALE CONN: 1/4T-1/8 NPT
BRASS
QTY
1
1
1
www.parrinst.com
19
Page 20
Hydrogenation Apparatus

Bottle Clamp Assembly - Series 3910

20
Parr Instrument Company
Page 21

Shaker Column Assembly - Series 3910

R1
DATE
SHAKER COLUMN ASSEMBLY
R1 WAS CK, REDRAWN; [EC10291] 07-02-09 JPR
APPROVED
BY
BY
SCALE
DWG NO.
DATE
DRAWN
DESCRIPTION
REVISIONS
FOR
SHEET 2
REVISED AND REDRAWN; 01-15-02 RPE
08-23-94
1:2
3910 HYDROGENATION APPARATUS
RPE
SIZE
C
3910
ANGULAR ... ±1/2°
PARR INSTRUMENT CO.
211 53rd STREET
MOLINE, ILLINOIS 61265
DO NOT SCALE DRAWING
TOLERANCES
IN INCHES
UNLESS OTHERWISE SPECIFIED
1/X .... ±1/64
.00 .... ±.010
.000 .... ±.003
08-24-94
VGM
COMPANY.
DRAWN IN 3rd ANGLE PROJECTION
OR BETTER
64
MACHINED SURFACES:
UNMARKED RADII .03"
PROPRIETARY
NEITHER THE DRAWING
NOR INFORMATION
CONTAINED HEREIN
MAY BE COPIED,
REPRODUCED, OR
OTHERWISE USED
WITHOUT WRITTEN
PERMISSION FROM
PARR INSTRUMENT
R2 REVISED 65CA [EC10603];09-14-10 JPR
Hydrogenation Apparatus
WHITE LUBE
TYP (2) PLACES
3910 SHEET 1
CRANK PIN W/NUT
19CA3
FLYWHEEL W/SHAFT
8CA3
34CA3
SC2520SC04
MOTOR PULLEY
1/4-20 X 1/4 SHSS
37CA3
W/SET SCREW
VEE BELT, NOTCHED 25"
SN2520HX
1/4-20 HEX JAM NUT
2CA2
SC2550SC04
TYP (4) PLACES
1/4-20 X 1/4 SCSS
SHAKER COLUMN
5CA3
PIVOT PIN
TYP (2) PLACES
www.parrinst.com
21
Page 22
Hydrogenation Apparatus

Flywheel Assembly - Series 3910

22
Parr Instrument Company
Page 23

Connecting Rod Assembly - Series 3920

Hydrogenation Apparatus
www.parrinst.com
23
Page 24
Hydrogenation Apparatus

AA92CA Single Valve Assembly

A150VB
KNOB
6VBBB
PACKING
CONE
8VB2
PACKING
NUT
112VB4AK
VALVE
NEEDLE
12CA GAGE
70
60
50
40
30
80
20
90
LBS. PER SQ. INCH
10
100
0
arr
P
21VBBB
LANTERN
RING
PARR
INSTRUMENT
COMPANY
4VB3
PTFE
20VB
VALVE
PACKING
SEAT
NOTE:
ALL PIPE THREADS TO BE
SEALED WITH TEFLON TAPE
24
92CA VALVE BODY
124VB NIPPLE
9VB1 NUT
Parr Instrument Company
Page 25
12CA
70
60

3A93CA Double Valve Assembly

80
90
100
Hydrogenation Apparatus
50
40
30
20
LBS. PER SQ. INCH
10
60HW Nipple
0
arr
P
A102CA
GAGE FACES
21VBBB
(2) PLACES
PARR
COMPANY
INSTRUMENT
4VB3
20VB
93CA
(2) PLACES
(4) PLACES
9VB1
A150VB
124VB
Note: AA93CA Valve Assembly as show except without gage.
6VBBB TYP
(2) PLACES
(2) PLACES
www.parrinst.com
8VB2
(2) PLACES
112VB4AK
(2) PLACES
25
Page 26
Hydrogenation Apparatus

3910 Wiring Schematic

26
Parr Instrument Company
Page 27

3920 Wiring Schematic

Hydrogenation Apparatus
www.parrinst.com
27
Page 28
Hydrogenation Apparatus

Explosion Proof Switch Assembly for 1765EEG 115V - Series 3910

28
Parr Instrument Company
Page 29
Hydrogenation Apparatus

Explosion Proof Switch Assembly for 1765EEK 220V - Series 3910

www.parrinst.com
29
Page 30
Hydrogenation Apparatus

Explosion Proof Switch Assembly for A388EEG 115V & 220V - Series 3920

30
Parr Instrument Company
Page 31
Hydrogenation Apparatus

Explosion Proof Switch Assembly for A388EEK 220V - Series 3920

www.parrinst.com
31
Page 32
Hydrogenation Apparatus

Thermocouple Assemblies

Assembly No.
A159CAPA A295E 66CA2 250 mL
A159CA2PA A295E 66CA 500
A159CA3PA A295E2 71CA 1000
A159CA4PA A295E2 72CA 2000
A159CA5PA A295E2 129CA3 1700
Equipped with Thermocouple No.
Fits Bottle
No. Size
72CA4 3000
A159CA2
Thermocouple
Assembly
32
Parr Instrument Company
Page 33

Reaction Bottles for Parr Hydrogenators

Hydrogenation Apparatus
Apparatus No.
Bottle No.
Size mL
Bottle Type
Maximum Working Pressure psig
Requires Connector No.
3911 66CA2 250 Borosilicate Glass 60 A122CA2* 3911 66CA 500 Borosilicate Glass 60 A122CA 3911 66CA3 500 Borosilicate Glass Fiberglass Covered 60 A122CA 3921 71CA 1000 Borosilicate Glass 40 A123CA2* 3921 71CA2 1000 Borosilicate Glass Fiberglass Covered 40 A123CA2* 3921 72CA 2000 Borosilicate Glass 30 A123CA 3921 72CA3 2000 Borosilicate Glass Fiberglass Covered 30 A123CA
3921 72CA4 2500 Hand Blown, Heavy Wall Borosilicate Glass 60 A123CA 3921 A129CA3 1700 Stainless Steel Bottle Assembly 60 A155CA
Tube with 133CA2 Spacer Spool
3921 A226CA Cylinder only, Stainless Steel, 1700 mL
A129CA3 Stainless
Bottle
72CA
2.0 L
Bottle
71CA
1.0 L
Bottle
66CA
500 mL
Bottle
66CA2
250 mL
Bottle
Heating Mantles
Mantle No. Watts Volts Use with Bottle No.
A450EEB 100 11 5 66CA, 66CA2 A450EEE 10 0 230 66CA, 66CA2 A451EEB 200 115 71CA, 72CA, A129CA3
or A226CA
A451EEE 200 230 71CA, 72CA, A129CA3
or A226CA
www.parrinst.com
33
Page 34
Hydrogenation Apparatus

Air Motor Assembly

34
Parr Instrument Company
Page 35

Switch Box Assembly 115V

Hydrogenation Apparatus

Switch Box Assembly 230V

Note: All components are the same as 115V except as noted.
www.parrinst.com
35
Page 36
Revision 03/14/13
Loading...