DANGER
Failure to observe any of the following may cause battery pack malfunction, overheating, explosion and fire.
Do not immerse the battery pack in liquid or get it
wet.
Do not leave the battery pack near a heat source
such as a stove.
Do not charge the battery pack on unspecified
instruments.
Do not charge the battery pack in conditions
outside the specified environment. (Over 40°C (104°F))
Do not put the battery pack into fire or heat it.
Do not short-circuit the + and terminals on the
battery pack.
Do not give strong impact to or deform the
battery pack.
• Do not disassemble or modify the battery pack.
Lithium Ion Battery SB-220V
General
The SB-220V lithium ion battery (hereafter referred to as battery pack) is for the AED-2150K/L, AED-2151K and AED-2152K/L
NOTE: When using this battery pack, also refer to the
operator’s manual for the AED and the battery charger.
for AED-2150 series
for AED-2152
• Do not charge the battery pack in a high
temperature place such as near a stove or in
sun-heated cars.
WARNING
• Replace the battery pack every 2 years.
• Do not use a damaged or deformed battery pack.
WARNING
Fully charge the battery pack at the time it is delivered, regardless of whether you will use it soon. The battery pack gradually discharges and it may become unusable.
Safety Information
DANGER
WARNING
CAUTION
Failure to observe any of the following may cause battery pack malfunction, overheating, explosion and fire.
• Do not immerse the battery pack in liquid or get it
wet.
• Do not leave the battery pack near a heat source
such as a stove.
• Do not charge the battery pack on unspecified
instruments.
• Do not charge the battery pack in conditions
outside the specified environment. (Over 40°C (104°F))
• Do not put the battery pack into fire or heat it.
• Do not short-circuit the + and – terminals on the
battery pack.
• Do not give strong impact to or deform the
battery pack.
0604-910862B
A danger is used to alert the user to a hazardous situation which will cause death or serious injury.
A warning alerts the user to the possible injury or death associated with the use or misuse of the instrument.
A caution alerts the user to possible injury or problems with the instrument associated with its use or misuse such as instrument malfunction, instrument failure, damage to the instrument, or damage to other property.
DANGER
WARNING
When the battery pack is inserted in the AED, charge the battery pack every 3 months so that it is always ready for use in an emergency. After charging, install the battery pack in the AED and confirm that the status indicator lights in green.
WARNING
When the battery pack is stored without inserting it in the AED after it is fully charged, charge it every 6 months. The battery pack gradually discharges and it may become unusable.
WARNING
If the battery pack is damaged and the substance inside the battery pack contacts the eyes, wash immediately and thoroughly with water and see a physician. Never rub your eyes, because you may lose your eyesight.
CAUTION
Do not leave the battery pack in direct sunlight or high temperatures such as in a sun-heated car. This may cause leakage of the battery pack, degrade its performance and make it unusable.
1
CAUTION
If the battery pack is damaged and the substance inside the battery pack contacts the skin or cloth, wash immediately with clear water. The skin may get irritated.
CAUTION
Store the battery pack in the specified environmental conditions. Otherwise, the battery pack may deteriorate or become unusable.
CAUTION
When recycling or disposing of the battery pack, follow your local law.
Symbol Description
The CE mark is a protected conformity mark of the
European Community. Products marked with this
symbol comply with the requirements of the Medical
Device Directive 93/42/EEC.
Products marked with this symbol require separate
waste collection according to EU battery directive
2006/66/EC.
Products marked with this symbol comply with
environmental protection use period of 5 years according to the ST/J11364 “Marking for Control
of Pollution Caused by Electronic Information Products” of the People’s Republic of China Electronic Industry Standard.
Panel Descriptions
Remaining battery check button
Press to check the remaining battery on the remaining battery indicators.
NOTE • Always have a fully charged battery pack ready
as a spare.
• When bringing the battery pack on an airplane, consult the airline beforehand.
• When transporting the battery pack on an airplane, consult the shipping company beforehand.
• Charge the battery after every use. Do not leave the battery uncharged for 2 weeks. This causes the
battery to discharge and become unusable.
Composition
Name Supply Code Qty
SB-220V lithium ion battery X233 1
This operator’s manual 1
Battery label (not used) 1
Symbols
The following symbols are used with this battery pack. The description of each symbol is shown in the table below.
Symbol Description
Connector
Latch
Remaining battery indicators
LEDs light depending on the remaining battery power.
Charging Time and Standby
Charging Time
With a new battery pack at a surrounding temperature of 20°C (68°F) using the SB-205V battery charger
Fully charged: within 3 hours
Charged to 80%: about 2 hours
NOTE: For battery operation time and number of
debrillations that can be performed with a fully
charged new battery pack, refer to the operator’s manual for the AED.
Standby
When a new fully charged battery pack is installed in the AED,
standby power lasts for about 3 months.
(Circle and slash: red)
(Circle and slash: red)
(Circle and slash: red)
(Circle and slash: red)
(Circle and slash: red)
(Background color: blue)
Recycle
Caution
Lot number
Keep away from re
Keep away from water
Avoid strong impact or dropping.
Never disassemble or modify.
Never short circuit the + and − terminals on the battery.
Follow instructions for use
Writing Down the Installation Date
When installing the battery pack
in the AED for the rst time, write the date of rst use in the Installation Date column on the label of the battery pack.
The battery pack gradually deteriorates with total hours of
use. Replace the battery pack
every 2 years regardless of the
total hours of use.
Do not use a battery pack which
is past the “Install Before” date.
Type of Battery, Voltage and Capacity
Type of battery: Lithium ion, secondary battery
Nominal voltage: 25.9 V
Rated capacity: 1600 mAh
Watt-hour rating: 41.4 Wh
2
Write the date of rst use here.
Check the due date before use.