If you have any comments or suggestions
on this manual, please contact us at:
www.nihonkohden.com
0614-904173D
Page 3
In order to use this product safely and fully understand all its functions, make sure to read this manual before using the
product.
Keep this manual near the instrument or in the reach of the operator and refer to it whenever the operation is unclear.
The contents of this manual are subject to change without notice.
Copyright Notice
The entire contents of this manual are copyrighted by Nihon Kohden. All rights are reserved. No part of this document
may be reproduced, stored, or transmitted in any form or by any means (electronic, mechanical, photocopied, recorded,
or otherwise) without the prior written permission of Nihon Kohden.
Trademark
The mark printed on the SD card that is used in this instrument is a trademark. The company name and model name are
trademarks and registered trademarks of each company.
Page 4
Contents
GENERAL HANDLING PRECAUTIONS ............................................................................. i
WARRANTY POLICY ......................................................................................................... ii
RESPONSIBILITIES – PROFESSIONAL USERS .............................................................. ii
EMC RELATED CAUTION ..................................................................................................iii
Conventions Used in this Manual and Instrument ............................................................. iv
Warnings, Cautions and Notes ................................................................................ iv
Explanations of the Symbols in this Manual and Instrument .................................... v
Caution Labels on the Analyzer ...............................................................................vii
Text Conventions in this Manual .............................................................................. ix
How to Use This Manual ..................................................................................................... x
Manual Organization................................................................................................. x
Manual Construction ................................................................................................ xi
Safety ....................................................................................................................... xi
Section 1 General ...................................................................................1.1
Appendix D Factory Default Settings ................................................... A.D.1
C.10 Operator’s Manual MEK-7300K
Page 14
GENERAL HANDLING PRECAUTIONS
This device is intended for use only by qualified healthcare personnel.
Use only Nihon Kohden approved products with this device. Use of non-approved products or
in a non-approved manner may affect the performance specifications of the device.
Please read these precautions thoroughly before attempting to operate the instrument.
1. To safely and effectively use the instrument, its operation must be fully understood.
2. When installing or storing the instrument, take the following precautions:
(1) Avoid moisture or contact with water, extreme atmospheric pressure, excessive humidity and temperatures, poorly
ventilated areas, and dust, saline or sulfuric air.
(2) Place the instrument on an even, level oor. Avoid vibration and mechanical shock, even during transport.
(3) Avoid placing in an area where chemicals are stored or where there is danger of gas leakage.
(4) The power line source to be applied to the instrument must correspond in frequency and voltage to product
specications, and have sufcient current capacity.
(5) Choose a room where proper grounding is available.
3. Before Operation
(1) Check that the instrument is in proper operating order.
(2) Check that the instrument is grounded properly.
(3) Check that all cords are connected properly.
(4) Check that battery level is acceptable and battery condition is good when using battery-operated models.
4. During Operation
(1) The instrument must receive continual, careful attention.
(2) Turn power off when trouble is found on the instrument.
5. To Shutdown After Use
(1) Turn power off with all switches and keys returned to their original positions.
(2) Remove the cords gently; do not use force to remove them.
(3) Clean the instrument together with all accessories for their next use.
6. The instrument must receive expert, professional attention for maintenance and repairs. When the instrument is
not functioning properly, it should be clearly marked to avoid operation while it is out of order.
7. The instrument must not be altered or modied in any way.
8. Maintenance and Inspection
(1) The instrument and parts must undergo regular maintenance inspection.
(2) If stored for extended periods without being used, make sure prior to operation that the instrument is in proper
operating condition.
Operator’s Manual MEK-7300K i
Page 15
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and workmanship for one year
from the date of delivery. However, consumable materials such as recording paper, ink, stylus and battery are excluded
from the warranty.
NKC or its authorized agents will repair or replace any products which prove to be defective during the warranty period,
provided these products are used as prescribed by the operating instructions given in the operator’s and service manuals.
No other party is authorized to make any warranty or assume liability for NKC’s products. NKC will not recognize any
other warranty, either implied or in writing. In addition, service, technical modication or any other product change
performed by someone other than NKC or its authorized agents without prior consent of NKC may be cause for voiding
this warranty.
Defective products or parts must be returned to NKC or its authorized agents, along with an explanation of the failure.
Shipping costs must be pre-paid.
This warranty does not apply to products that have been modied, disassembled, reinstalled or repaired without Nihon
Kohden approval or which have been subjected to neglect or accident, damage due to accident, re, lightning, vandalism,
water or other casualty, improper installation or application, or on which the original identication marks have been
removed.
In the USA and Canada other warranty policies may apply.
RESPONSIBILITIES – PROFESSIONAL USERS
This instrument must be used by a professional user with a full knowledge of operating this instrument, only for the stated
intended use and according to the instructions for use. Instructions in the operator’s manual must be followed, especially
the following points.
• Storage and stability of reagents
• Handling of reagents
• Instrument installation
• Connection of all tubes to inlets and outlets
• Connection of all tubes to reagents and waste container
• Checking the amount of reagents and waste uid
• Calibration
• Quality control
• Maintaining and servicing
If deviating from the instructions, the professional user does it at the risk and liability of the laboratory and only after
validation by the laboratory. Nihon Kohden has no responsibility over such deviations.
ii Operator’s Manual MEK-7300K
Page 16
EMC RELATED CAUTION
This equipment and/or system complies with the International Standard EN 61326-2-6 for electromagnetic
compatibility for electrical equipment and/or system for measurement, control and laboratory use. However,
an electromagnetic environment that exceeds the limits or levels stipulated in the EN 61326-2-6, can
cause harmful interference to the equipment and/or system or cause the equipment and/or system to fail
to perform its intended function or degrade its intended performance. Therefore, during the operation of the
equipment and/or system, if there is any undesired deviation from its intended operational performance,
you must avoid, identify and resolve the adverse electromagnetic effect before continuing to use the
equipment and/or system.
The following describes some common interference sources and remedial actions:
1. Strong electromagnetic interference from a nearby emitter source such as an authorized radio station
or cellular phone:
Install the equipment and/or system at another location if it is interfered with by an emitter source
such as an authorized radio station. Keep the emitter source such as cellular phone away from the
equipment and/or system.
2. Radio-frequency interference from other equipment through the AC power supply of the equipment and/
or system:
Identify the cause of this interference and if possible remove this interference source. If this is not
possible, use a different power supply.
3. Effect of direct or indirect electrostatic discharge:
Make sure all users and patients in contact with the equipment and/or system are free from direct or
indirect electrostatic energy before using it. A humid room can help lessen this problem.
4. Electromagnetic interference with any radio wave receiver such as radio or television:
If the equipment and/or system interferes with any radio wave receiver, locate the equipment and/or
system as far as possible from the radio wave receiver.
If the above suggested remedial actions do not solve the problem, consult your Nihon Kohden
representative for additional suggestions.
This equipment complies with International Standard EN 55011: 2007 Group 1, Class B. Class B EQUIPMENT is
equipment suitable for use in domestic establishments and in establishments directly connected to a low voltage
power supply network which supplies buildings used for domestic purposes.
The CE mark is a protected conformity mark of the European Community. Products with the CE mark comply
with the requirements of the In vitro Diagnostic Medical Device Directive 98/79/EC.
NOTE about Waste Electrical and Electronic Equipment (WEEE) directive 2002/96/EC
For the member states of the European Union only:
The purpose of WEEE directive 2002/96/EC is, as a first priority, the prevention of waste electrical and
electronic equipment (WEEE), and in addition, the reuse, recycling and other forms of recovery of such
wastes so as to reduce the disposal of waste.
Contact your Nihon Kohden representative for disposal.
Operator’s Manual MEK-7300K iii
Page 17
Conventions Used in this Manual and Instrument
Warnings, Cautions and Notes
Warnings, cautions and notes are used in this manual to alert or signal the reader to specic information.
WARNING
A warning alerts the user to possible injury or death associated with the use or misuse of the instrument.
CAUTION
A caution alerts the user to possible injury or problems with the instrument associated with its use or
misuse such as instrument malfunction, instrument failure, damage to the instrument, or damage to other
property.
NOTE
A note provides specific information, in the form of recommendations, prerequirements, alternative methods or
supplemental information.
iv Operator’s Manual MEK-7300K
Page 18
Explanations of the Symbols in this Manual and Instrument
The following symbols found in this manual/instrument bear the respective descriptions as given.
The CE mark is a protected
conformity mark of the European
Community. Products marked
with this symbol comply with
the requirements of the In vitro
Diagnostic Medical Device Directive
98/79/EC.
IVD
Date of manufacture
Laser on
IN VITRO DIAGNOSTIC
MEDICAL DEVICE
Attention, consult operator’s manual
Products marked with this symbol
comply with the European WEEE
directive 2002/96/EC and require
separate waste collection. For Nihon
Kohden products marked with this
symbol, contact your Nihon Kohden
representative for disposal.
Operator’s Manual MEK-7300K v
Page 19
On screen and recorded data
SymbolDescription
When displayed beside NE, NE%, LY, LY%: Room temperature high
When displayed beside NE, NE%, LY, LY%, MO, MO%, EO, EO%, BA, BA%:
Optical count error
When displayed beside NE, NE%, LY, LY%, MO, MO%: Room temperature low
When displayed beside WBC or RBC measured value: Sample error
When displayed beside HGB measured value: Dirty measurement baths
When displayed beside WBC measured value: Poor hemolyzation
When displayed beside HGB measured value: HGB voltage adjustment error
When displayed beside MCHC measured value: Abnormal MCHC
When displayed beside WBC measured value: Small nucleated cell
When displayed beside NE, NE%, MO, MO%, BA, BA%: Immature granulocytes
When displayed beside NE, NE%, EO, EO%: Ne-Eo interference
When displayed beside NE, NE%, LY, LY%, MO, MO%, EO, EO%, BA, BA%:
Blasts
When displayed beside NE, NE%, EO, EO%: Left shift
When displayed beside LY, LY%, MO, MO%: Atypical lymphocytes
When displayed beside LY, LY%, MO, MO%: Ly-Mo interference
When displayed beside HGB measured value: HGB circuit error/WBC measured
value is OVER
When displayed beside RBC and PLT measured values: PLT-RBC interference
When displayed beside PLT: PLT low value (below 50,000/μL)
C
When displayed beside WBC or PLT measured value: Platelet coagulation
Reagent level graph
SD card is inserted
Open the alarm window
vi Operator’s Manual MEK-7300K
Page 20
Caution Labels on the Analyzer
Refer to “Connecting Tubes” in Section 2.
Refer to warnings and cautions in
“Connecting the Power Cord and
Grounding the Analyzer” in Section 2.
Operator’s Manual MEK-7300K vii
Page 21
NOTE
• Replace the filters periodically.
• When attaching the filter joint assembly, be careful not to bend or
damage the filter packing at the bottom of the measurement bath.
• When there is a leakage, check that there is no scratch or damage to
the circumference of the filter.
viii Operator’s Manual MEK-7300K
Page 22
Text Conventions in this Manual
In this manual, procedural instructions are explained in logical groups, using numbered steps. Illustrations and drawings
appear where they are useful to the explanation. Text conventions are as follows:
Screen Name
The screen name is printed in mixed-case, regular letters; for example, Ready screen.
Touch Screen Keys
The screen has some touch screen keys which are pressure-sensitive. Pressing one of these touch screen keys initiates the
action specied by a screen label. Screen labels are shown in mixed-case, regular letters; for example, Quality control key.
Data Entry Field Names
Fields that accept data entered by the Operator have their names shown in regular, mixed-case font enclosed within carats < >.
Hard Keys (Keys on the Panels)
The keys on the panels are shown in regular, mixed-case font enclosed in brackets; for example, [Clean].
Screen Messages
Screen messages or other screen displays will appear in regular, mixed-case font enclosed in quotation marks, for example,
“Priming”.
InformationPresentationExamples
Screen nameRegular, Mixed-CaseMenu screen
Touch screen keysRegular, Mixed-CaseQuality control
Field namesRegular, Mixed-Case, enclosed within carats < ><Dilute mode>
Hard keys (Panel keys)Regular, Mixed-Case, enclosed within brackets[Clean]
Screen MessageRegular, Mixed-Case, enclosed within quotation marks“Priming”
Operator’s Manual MEK-7300K ix
Page 23
How to Use This Manual
Manual Organization
The major sections of the manual and their contents are as follows:
Section 1: General
This section provides an overall description of the system. It names the major
system components and describes their uses or functions.
Section 2: Preparation
This section provides detailed instructions for system setup and conguration. It
explains proper location, requirements, and steps for installation.
Section 3: Principles of Operation
This section explains the principles behind the system’s operation. It describes
what the system measures and how those measurements are made.
Section 4: Performance Characteristics and Specifications
This section contains useful details on the dimensions of the analyzer, proper
operating environment, and performance specications.
Section 5: Operating Instructions
This section explains the procedures for daily start-up and shutdown, sample
collection and handling, and routine operation of the analyzer including use of
stored data.
Section 6: Calibration Procedures
This section describes the calibration process. It discusses calibration materials,
guidelines, and methods.
Section 7: Operational Precautions and Limitations
This section contains a summary of known factors that may adversely affect the
proper operation of the analyzer or the quality of the output.
Section 8: Hazards
This section covers possible hazards arising from the operation of the analyzer,
as well as decontamination and waste handling procedures.
Section 9: Service and Maintenance
This section discusses routine maintenance and cleaning on a daily, weekly,
monthly, and as required basis. Also included are detailed instructions for
removing and cleaning certain components to ensure proper system performance.
Section 10: Messages and Troubleshooting
This section contains a troubleshooting guide to help users identify probable
causes of a system malfunction or of suspect data, and to suggest the proper
corrective action.
x Operator’s Manual MEK-7300K
Page 24
Section 11: Quality Control
This section covers the proper mixing, handling, and running of control material,
setting up QC les, and using the QC capabilities of the analyzer.
Appendices
Appendix A
This appendix lists the part numbers of components, accessories, controls,
reagents, and consumables associated with the analyzer for user convenience
when placing orders.
Appendix B
This appendix contains information on setup of the optional bar code reader.
Appendix C
This appendix contains specimen data reports.
Appendix D
This appendix contains the list of factory default settings.
Manual Construction
Safety
The physical construction of the manual supports its sectional organization.
Contents
The Contents at the beginning of this manual lists each section and its
subsections.
Section Separators
A large separator tab marks the start of each section.
Throughout the manual, signal words appear where the nature of the information
warrants special attention.
Operation, maintenance, and servicing of hematology systems may expose
individuals to potential safety and health hazards. All work must be performed as
described in the operator’s manual or as directed by Nihon Kohden. For detailed
safety information, refer to Section 8 “Hazards”.
Warnings are inserted throughout this manual to alert personnel to potential
hazards.
For detailed safety information, refer to Section 7 “Operational Precautions and
Limitations” and Section 8 “Hazards” in this manual.
Features ..............................................................................................................................................................1.4
Front Panel ................................................................................................................................................ 1.6
Right Side Panel ........................................................................................................................................ 1.7
Left Side Panel ..........................................................................................................................................1.8
Top Panel...................................................................................................................................................1.8
Screen Information ..................................................................................................................................1.10
Using Touch Screen Keys ........................................................................................................................1.11
Reagent System ................................................................................................................................................1.12
Controls and Calibrator ...........................................................................................................................1.13
The MEK-7300K Automated Hematology Analyzer provides simultaneous 23
parameter and 2 research parameter measurement. It provides quick counting and
all operations are performed automatically just by putting the sampling nozzle
into a sample container with whole blood and pressing the [ Count] switch.
The analyzer also has a cap pierce unit to measure samples in capped tubes in
closed mode. Results and data are displayed on a color LCD screen and full
reports can be printed on an optional printer.
The analyzer automatically cleans blood from the sampling nozzle, so it is safe and
there is no risk of touching the blood. The analyzer has self-check, quality control
and other programs for reliable data management system.
To achieve full performance, thoroughly read this operator’s manual before
operating the analyzer.
Measured Parameters
To use the analyzer safely and effectively and keep it in optimum condition,
follow the operating and maintenance instructions in this manual.
NOTE
Use only Nihon Kohden parts and accessories to assure maximum
performance from your instrument.
23 parameters and 2 research parameters can be measured on the MEK-7300K
hematology analyzer.
RDW-CV: Red Blood Cell Distribution Width in CV
RDW-SD: Red Blood Cell Distribution Width in SD
PLT*1: Platelet Count
PCT: Platelet Crit
MPV: Mean Platelet Volume
PDW: Platelet Distribution Width
IG%*2: Immature Granulocyte Percent
IG*2: Immature Granulocyte
*1 CBC (complete blood count) parameters
*2 Research parameters
HGB (g/dL)
× 10
RBC (× 106/µL)
HGB (g/dL)
× 100
HCT (%)
1
Operator’s Manual MEK-7300K 1.3
Page 28
1. GENERAL
Features
• Simultaneous 23 parameter measurement
The analyzer simultaneously measures 23 parameters (CBC and WBC 5 part
differential). WBC, RBC and PLT are measured by the electrical resistance
detection method. WBC is differentiated into neutrophil, lymphocyte,
monocyte, eosinophil and basophil by the light scatter technique - ow
cytometry with laser.
• Infection prevention
To help eliminate the risk of touching blood, you can perform measurement in
closed mode without removing the cap of the sample tube.
• Automatic sampling
Once the sample is aspirated through the sampling nozzle, all other operations
are performed automatically. The sample is automatically diluted and
measured. After measurement, the analyzer is cleaned and the waste uid is
automatically treated.
• Automatic sampling nozzle cleaning
The analyzer automatically cleans remaining blood from the sampling nozzle.
This also helps eliminate the risk of touching blood during measurement.
• Automatic clog removal and recounting
After each counting, the analyzer removes blood protein and dust particles
from around the aperture caps to prevent clogging. Even if a clog occurs, the
analyzer automatically removes the clog and recounts the sample.
• High accuracy and reproducibility
The analyzer reduces counting error by automatically diluting samples
and wiping the sampling nozzle. The analyzer provides high accuracy and
reproducibility with a low dilution ratio of 200:1 for WBC, 40,000:1 for RBC
and 360 µL of diluted sample. Built-in circuits automatically compensate
for cell miscount due to coincidence (simultaneous cell passage) and uid
temperature variation. The aperture is shielded from external noise and
clogging is prevented.
• Color LCD touch screen
The analyzer has a color TFT LCD with 800 × 600 pixel resolution which
clearly displays the results and various messages. The touch screen allows easy
and intuitive operation of the analyzer. You can enlarge the numerical data and
scattergram by touching them. Screen messages prompt the operator through
operation.
1.4 Operator’s Manual MEK-7300K
Page 29
1. GENERAL
• Pre-diluted blood counting
Pre-diluted blood (10 or 20 µL) can be measured. The sample needs to be
diluted before measurement. 30 µL of venous blood can also be measured.
This sample does not need to be diluted before measurement.
• Automatic self-check
When the analyzer is turned on, the analyzer automatically starts priming and
checks itself. If a problem is detected, the LCD displays an alarm message, e.g.
BUBBLE, NO DILUENT or CLOG, so you can quickly identify and x the
problem.
• Variety of quality control programs
A variety of quality control programs are provided to enable calibration for
WBC, RBC, PLT, HGB and HCT, calculation of mean value and CV value, XR and L & J quality control programs, and normal range setting on the screen.
• Data management
The analyzer can store measurement data of up to 400 samples and histograms
of up to 50 samples. Stored measurement data can be printed, deleted or
transferred to an external device.
1
• Automatic priming and cleaning
The analyzer automatically primes the uid path when the power is turned on
and cleans the uid path when the power is turned off.
• Selectable 23 parameters or 8 (CBC) parameters measurement
Measuring parameters can be selected from 8 (CBC) parameters or 23
(CBC and WBC 5 part differential) parameters. Measuring only the 8 CBC
parameters reduces reagent consumption.
• Password protected access
Measurement conditions, such as calibration coefcient, normal range and
quality control, are managed by the operator who has the password to access
these setting screens. Miscalculation caused by inappropriate measurement
condition settings can be avoided.
• High dilution measurement
When the WBC measured result is high, the sample can be recounted by
diluting the sample with higher ratio.
• Reagent management
The amount of reagent and waste uid can be monitored by the hematology
analyzer and a message can be displayed to alert the operator that a reagent is
nearly run out or the waste container is becoming full.
• Data sending and receiving with USB
The measurement data and work list can be sent and received by connecting a
personal computer to the USB connector (device) and using the QP-822V data
management software.
Operator’s Manual MEK-7300K 1.5
Page 30
1. GENERAL
Panel Description
Front Panel
8
1
2
3
6
7
5
4
15
16
No.NameDescription
1Main power lampLights when the [Main power] switch on the rear panel is turned on.
2Power lamp
3Power key
4Auto print keySwitches the printing mode between automatic and manual for the printer.
5Feed keyFeeds paper of the printer while held down.
6Print keyPrints displayed data on the printer.
Auto print mode
7
lamp
8LCD displayDisplays various messages, measured data and touch screen keys.
9Reset key
10Clean key
11Eject keyFor closed mode only. Opens the tube holder to set the sample tube.
12Tube holder
13Sampling nozzle
14Count switchFor open mode only. Aspirates the sample and starts counting.
Printer unit
15
(WA-730VK)
16Printer door
Lights when the [Main power] switch on the rear panel and [Power] key on the front
panel are turned on.
Turns the analyzer power on or off when the [Main power] switch on the rear panel
is turned on. When the power is turned on, priming and self-check are automatically
performed and the Ready screen appears.
Lights when automatic printing mode is selected.
Stops operation when pressed during operation. Returns to the Ready screen when
pressed while changing settings. Use this key only when an error occurs.
Cleans the uid path, aperture and manometer with detergent. Automatically primes
after cleaning the uid path. Press this key when clogging occurs, the manometer
becomes dirty or bubbles occur in the manometer.
For closed mode only. Holds a sealed vacuum blood collecting tube. Press the [Eject]
key to open. After measurement, the holder automatically opens.
For open mode only. Aspirates the sample. Dispenses the diluent when in the predilution blood mode.
Thermal array printer. Prints out measured data and sample ID number (optional).
For the recording paper of the WA-730VK printer unit. To open, pull the upper left
corner (optional).
9
10
11
13
14
12
1.6 Operator’s Manual MEK-7300K
Page 31
1. GENERAL
Right Side Panel
1
1234567
8
No.NameDescription
1Laser switch
ISO3
2
Diluent inlet
CLN
3
Detergent inlet
CLN3
4
Detergent inlet
HEMO3N
5
Lysing reagent inlet
HEMO5
6
Lysing reagent inlet
WASTE
7
Waste outlet
8Vent hole for fan
Turns the laser on or off with the laser key for WBC 5 part
differential measurement.
Inlet for the ISOTONAC•3 diluent.
Inlet for the CLEANAC detergent.
Inlet for the CLEANAC•3 detergent.
Inlet for the Hemolynac•3N lysing reagent.
Inlet for the Hemolynac•5 lysing reagent.
Outlet for waste such as used lyse, detergent and aspirated
samples.
Vent hole for the fan.
NOTE
Do not block the hole. It affects the measurement
capability.
Operator’s Manual MEK-7300K 1.7
Page 32
1. GENERAL
Left Side Panel
1
No.NameDescription
Top Panel
1SD card slotInsert an SD memory card.
1
No.NameDescription
1Flow cell coverFor adjusting the ow cell position.
1.8 Operator’s Manual MEK-7300K
Page 33
1. GENERAL
Rear Panel
No.NameDescription
1Main power switch
2Fuse holder
3Power socketConnects the AC power cord to supply AC power to the analyzer.
4
5USB socket (device)Connects a personal computer to send and receive data.*
6USB socket (host)Connects a barcode reader (Keyence BL-N60UB or equivalent).
7
8Serial port 1Connects to the optional WA-731V/461V card printer or PC.
9Serial port 2Connects to the optional WA-731V/461V card printer or PC.
10Option portConnects to the external instrument.
4
1
2
3
Equipotential ground
terminal
ZK-820V
Bar code reader socket
1
5
6
7
8
9
10
Supplies the power to the analyzer when it is turned on. Under
normal conditions keep this switch turned on.
Contains the time lag fuse. To replace the fuse, contact your
Nihon Kohden representative.
Connects the ground lead to the equipotential ground terminal on
the wall for earth grounding.
Connects to an optional ZK-820V hand-held bar code reader and
supplies power to the bar code reader when connected.
Power supply voltage: 5 V DC (pin 9: 5 V, pin 5: GND)
Rated current: 200 mA
* To connect to a personal computer, the QP-822V data management software is required.
Operator’s Manual MEK-7300K 1.9
Page 34
1. GENERAL
Basic Operations
Screen Information
Status bar
When the type of operator is “service”, the status bar is orange.
When the type of operator is “lab technician”, the status bar is blue.
When the type of operator is “other user”, the status bar is green.
For details, refer to “Assigning Users and Passwords” in Section 5.
Screen name
Check box
Date and time
For details, refer to
“Setting Date and Time”
in Section 5.
1.10 Operator’s Manual MEK-7300K
Page 35
1. GENERAL
Using Touch Screen Keys
Do not use a sharp object to press the touch screen. Use your finger.
Cursor (blue)
NOTE
Displays the selection list
1
Selection list
Use arrow keys to move cursorUse the numeric keys to enter a value and
press the Enter key to register the setting
Displays measure
mode selection list
Displays sample type
selection list
Displays another screen
Operator’s Manual MEK-7300K 1.11
Page 36
1. GENERAL
Reagent System
Introduction
A reagent system which is specically formulated for this analyzer provides
optimal system performance. Use of reagents other than those specied in this
manual is not recommended, as analyzer performance can be affected. Each
analyzer is tested at the factory using the specied reagents, and all performance
claims are generated using these reagents. For information on ordering reagents,
refer to Appendix A “Parts and Accessories”.
CAUTION
If reagent has been frozen, it must not be used.
Reagents
Diluent
Diluent is for doing the following:
• Act as the diluent for the WBCs, RBCs, PLTs, and HGB.
• Maintain the cell volume of each RBC and PLT during the count and sizing
portion of the measurement cycle.
• Provide a conductive medium for impedance counting and sizing of cells and
platelets.
• Rinse the sampling nozzles and ow systems.
Lysing Reagent
The analyzer uses lysing reagent as hemolysing reagents:
• Rapidly lyse the RBC and minimize the resultant cell stroma.
• Alter the WBC membrane to allow the cytoplasm to slowly diffuse and shrink
the membrane around the nucleus and any granules that may be present.
• Convert hemoglobin to a modied hemoglobin complex that is measurable at
540 nm (The quaternary ammonium lysate participates to form a chromogen
for hemoglobin measurement.).
Detergent
The analyzer can use two types of detergents: CLEANAC and CLEANAC•3:
• Provide an optically clear solution that is used to obtain the zero reference
during the HGB measurement cycle.
• Provide proper meniscus formation in both metering tubes and maintain it
during each run cycle.
• Rinse both measurement baths and sub baths, both metering tubes, and the
uid path with minimal bubble formation.
• Remove protein buildup and provide detergent cleaning within the analyzer. It
is used in scheduled and unscheduled maintenance.
1.12 Operator’s Manual MEK-7300K
Page 37
1. GENERAL
Reagent Storage
Reagents must be stored at room temperature to ensure optimal performance.
All reagents should be protected from direct sunlight, extreme heat, and freezing
during storage. Temperatures lower than 0ºC (32ºF) can cause reagent layering
that changes the tonicity and conductivity of the reagents. If any reagent has been
frozen, it must be disposed of according to federal, state, and local regulations.
Each length of reagent inlet tubing is attached to a cap that minimizes
evaporation and contamination during use. Ensure that all reagent caps are not
damaged, and are securely attached to containers during use. Reagent quality can
deteriorate with time. Therefore, use all reagents before the expiration date on
the label.
Handling Reagent
When handling reagents, pay special attention to the following:
• Wear protective gloves when handling reagents.
• Never transfer the contents of a reagent container to an unmarked container or
other reagent container.
• Thoroughly clean all spills. Remove any dried residue in and around the
reagent inlet connectors located on the right side panel of the analyzer.
• Dispose of reagents and waste uids according to federal, state, and local
regulations.
• Always wash your hands after handling reagents.
1
Controls and Calibrator
Background Count
Always run a background count after installing a fresh container of reagent.
Values reported must be within the following specications:
• WBC ≤ 0.2 (×103/μL)
• RBC ≤ 0.05 (×106/μL)
• HGB ≤ 0.1 g/dL
• PLT ≤ 10 (×103/μL)
Controls and calibrator are reference materials used to test, set, and monitor
analyzer performance. For information on ordering controls and calibrators, refer
to Appendix A “Parts and Accessories”.
Controls
Day-to-day verication of system calibration is performed using controls.
Running these stabilized reference products is recommended to test analyzer
accuracy. The following controls are used in the analyzer system:
Control: tri-level whole blood quality control materials designed to monitor CBC
and WBC values obtained on hematology systems.
Calibrator
Calibration of the directly-measured parameters can be performed using
calibrator. Calibration is discussed in Section 6 “Calibration Procedures”.
Operator’s Manual MEK-7300K 1.13
Page 38
1. GENERAL
Consumables
For information on ordering parts, accessories, reagents, controls, calibrators,
and consumables, refer to Appendix A “Parts and Accessories”.
1.14 Operator’s Manual MEK-7300K
Page 39
Section 2 Preparation
General ................................................................................................................................................................ 2.2
Standard Accessories ...............................................................................................................................2.4
Connecting an External Instrument to the Analyzer ............................................................................................ 2.7
Connecting a PC ....................................................................................................................................... 2.8
Disconnecting the USB Cable ......................................................................................................... 2.8
Connecting a Card Printer ......................................................................................................................... 2.8
Connecting a ZK-820V Hand-held Bar Code Reader (Option) .................................................................2.8
Connecting the Power Cord and Grounding the Analyzer .................................................................................2.10
Connecting the Power Cord .................................................................................................................... 2.10
Turning the Laser Switch On .............................................................................................................................2.18
Loading Recording Paper in the WA-730V Printer Unit (Option) .......................................................................2.19
Turning Power On .............................................................................................................................................. 2.20
Check Before Turning Power On ............................................................................................................. 2.20
Turning On the Power ..............................................................................................................................2.20
Checking the Date and Time Settings ..................................................................................................... 2.22
Cleaning the Analyzer ............................................................................................................................. 2.22
Check After Turning On the Power .......................................................................................................... 2.22
Safety Information about the SD Card ....................................................................................................2.30
Handling and Storage ...................................................................................................................2.30
Data Processing ............................................................................................................................ 2.30
Inserting and Removing the SD Card ..................................................................................................... 2.31
2
Operator’s Manual MEK-7300K 2.1
Page 40
2. PREPARATION
General
WARNING
Never use the analyzer in the presence of any flammable anesthetic
gas or high concentration oxygen atmosphere. Failure to follow this
warning may cause explosion or fire.
WARNING
Never use the analyzer in a hyperbaric oxygen chamber. Failure to
follow this warning may cause explosion or fire.
2.2 Operator’s Manual MEK-7300K
Page 41
Environmental Requirements
Install the analyzer outside the patient environment. If it is installed
inside the patient environment, the patient or operator may receive
electrical shock.
Use this analyzer under the following conditions.
Temperature: 15 to 30°C (59 to 86°F)
Humidity: 30 to 85% (noncondensing)
Air pressure: 70 to 106 kPa
70 kPa air pressure equals 3,000 m above sea level. Do not use the
analyzer at altitudes higher than 3,000 m above sea level.
2. PREPARATION
2
WARNING
CAUTION
15 cm
Left side
Rear
5 cm
Right side
Do not block
the vent hole
for the fan.
• Operate the analyzer in a room with a temperature range of 15 to 30°C (59 to
86°F). Keep the temperature of diluent and lysing reagent within this
temperature in order to obtain reliable data.
• No measurement can be done in dusty areas because the aperture for specimen
aspiration is very ne and can get clogged. Therefore, install the analyzer in a
dust-free area.
• Do not install the analyzer in direct sunlight.
• Do not place containers of reagent or uid on the analyzer. To prevent
electrical problems or electric shock, avoid spillage in or around the analyzer
because the uid is highly conductive.
• Select a stable, at buffering stand to set the analyzer on.
• If possible, use an independent AC outlet only for this analyzer. The analyzer
must not share an AC outlet with noise generating equipment such as a
centrifuge, constant temperature bath (thermostat), refrigerator, air conditioner
or ultrasonic cleaner.
• Make sure that there is more than 5 cm of space between the rear panel and
the wall and 15 cm of space between the left panel and the wall for adequate
ventilation.
Vent
hole
for fan
• Do not block the vent hole for the fan.
• When there is any problem in the analyzer, turn off the main power
immediately and disconnect the power cord from the AC outlet. Take the
analyzer out of service and check for damage.
Operator’s Manual MEK-7300K 2.3
Page 42
2. PREPARATION
Initial Preparation
Inventory
Standard Accessories
Conrm that the analyzer shipment contains the following:
• Celltac Es hematology analyzer
• Standard accessories
• Reagents
• Controls and calibrator
• Operator’s manual
• Printer (optional)
• ZK-820V Hand-held bar code reader (optional)
• Power cord
• Ground lead
• Fuse, 3.15 A time-lag (2)
• Filter assy (4)
• Pump tube (N) assy (2)
• Sampling nozzle
• Diluent tube, marked blue, 1.5 m
• Waste tube, marked red, 1.5 m
• Detergent tube for CLEANAC, marked green, 1.5 m
• Cleanac tube 8 for CLEANAC•3, marked white, 1.5 m
• 18 L cap (3)
• 18 L tube assy 2
• 1 L tube assy
• Hemolynac•3 cap (yellow)
• Hemolynac•3 tube (570), marked yellow
• 18 L tube assy (waste)
• Hemolynac•5 cap (black)
• Hemolynac•5 tube (570), marked black
• Waste container (10 L)
• Cleanac tube assy (8222) (2)
• Laser key (2)
• T-wrench TBS-8 (8 mm)
Visually inspect these items for damage. If there is any damage, contact your
Nihon Kohden representative.
2.4 Operator’s Manual MEK-7300K
Page 43
Unpacking
2. PREPARATION
Waste Disposal Requirements
Remove the analyzer from the shipping container and visually inspect for
damage. If there is any damage, contact your Nihon Kohden representative. Use
two people when lifting or moving the analyzer.
CAUTION
Use two people and be careful when moving the analyzer.
Otherwise, you may injure your back or be injured from dropping the
analyzer.
WARNING
Potential Biohazard. Observe all biosafety and chemical hazard
precautions for waste disposal. Operators are responsible for
disposing of waste in accordance with local, state, and federal
regulations.
2
CAUTION
Waste container must be stored at the same level or below the
analyzer, never above.
CAUTION
The waste is under pressure. Be sure that the waste tube is securely
placed in the waste container, flow of waste is unobstructed, and all
analyzer components are located away from possible waste overflow.
Operator’s Manual MEK-7300K 2.5
Page 44
2. PREPARATION
Installation Flowchart
1. Place the analyzer in the appropriate place.
2. Connect external instruments such as printer, ZK-820V hand-held bar code
reader and PC, if necessary.
3. Connect the power cord and if necessary, perform grounding. Do not turn on
the power of any instrument at this stage.
4. Connect the diluent, detergent, lysing reagent and waste container to the
analyzer.
5. Check that the pump tubes are not disconnected or damaged.
6. Turn on the power of the analyzer and connected external instruments.
7. Perform Strong cleaning on the Operation screen.
8. Check the settings on the Settings screen. Make sure that the date and time
settings are correct. Refer to Section 5.
9. Calibrate the analyzer. Refer to Section 6.
10. Check the daily accuracy.
2.6 Operator’s Manual MEK-7300K
Page 45
Connecting an External Instrument to the Analyzer
A personal computer, card printer (WA-460V/461V or equivalent) and external
printer can be connected to the serial port and a ZK-820V hand-held bar code
reader can be connected to the bar code reader socket on the rear panel.
Before connecting the external instrument to the analyzer, make sure that the
power on the instruments is turned off.
For changing the printing and communication format, refer to “Changing Output
Format” in Section 5.
CAUTION
When several medical instruments are used together, ground all
instruments to the same one-point ground. Any potential difference
between instruments may cause electrical shock to the operator.
2. PREPARATION
2
When more than one electrical instrument is used, there may be electrical
potential difference between the instruments. Potential difference between
instruments may cause current to ow to the connected to the instruments,
resulting in electrical shock.
Always perform equipotential grounding when required.
CAUTION
Connect only the specified instrument to the
analyzer and follow the specified procedure.
Failure to follow this instruction may cause
instrument malfunction.
CAUTION
The analyzer should only be connected to an
external instrument which complies with the
CISPR 11 (Edition 4: 2003), Group 1 and Class B
standard.
CAUTION
In order to avoid any safety hazard, only connect
personal computers which are approved by UL
60950.
CAUTION
Before connecting or disconnecting instruments,
make sure that each instrument is turned off and
the power cord is disconnected from the AC
socket. Otherwise, the operator may receive
electrical shock or injury.
Operator’s Manual MEK-7300K 2.7
Page 46
2. PREPARATION
Bar code socket
Connecting a PC
USB socket
(device)
Serial ports
A locally purchased PC can be connected to the serial port or USB socket
(device) on the rear panel.
NOTE
•
In order to avoid any safety hazard, only connect personal computers
which are approved by UL 60950.
•
Only use the 3-prong power cord for the PC.
• Set the correct connection settings.
• The PC can be connected to either the serial port or USB socket. When
the PC is connected to the serial port, only the numeric data can be
sent to the PC; the scattergram and histogram cannot be sent.
• To connect the PC to the USB socket (device), the optional QP-822V
data management software is required. The measurement result can be
sent to the PC after every measurement.
Disconnecting the USB Cable
Disconnect the USB cable from the PC to disconnect the USB communication.
Connect the USB cable again to restart the communication. You cannot
disconnect the communication by clicking the icon on the task bar of the PC.
Connecting a Card Printer
The WA-460V/461V card printer can be connected to the serial port on the
analyzer. On the WA-460V/461V card printer, only the numeric data is printed.
Histograms cannot be printed.
Connect the card printer cable to the card printer and the serial port on the
Serial ports
analyzer. For changing the printing and communication format, refer to
“Changing Output Format” in Section 5.
Connecting a ZK-820V Hand-held Bar Code Reader (Option)
The optional ZK-820V hand-held bar code reader can be used for reading the bar
code label (up to 13 characters) on the sample tube. For details about the handheld bar code reader, refer to the bar code reader manual.
The bar code reader can read the following codes:
• Industrial 2 of 5
• ITF
• JAN/EAN/UPC
• CODABAR (NW-7)
• CODE 39
• CODE 93
• CODE 128
2.8 Operator’s Manual MEK-7300K
Page 47
2. PREPARATION
Before turning on the analyzer power by pressing the [Power] key on the front
panel, connect the bar code reader cable to the bar code socket on the rear panel.
When the bar code reader is connected, the power is supplied through pin 9 of
the bar code reader socket.
NOTE
Do not connect instruments other than ZK-820V hand-held bar code
reader to the bar code reader socket.
To set the various settings for the hand-held bar code reader, read the bar code
attached to the bar code reader manual with the hand-held bar code reader
within 15 seconds after turning the bar code reader power on. After changing the
settings, turn the bar code reader power off.
You can also use the bar codes in “Bar Codes for Using the ZK-820V Hand-held
Bar Code Reader” in Appendix B.
2
Operator’s Manual MEK-7300K 2.9
Page 48
2. PREPARATION
Equipotential ground terminal
Connecting the Power Cord and Grounding the Analyzer
Connecting the Power Cord
AC source socket
Equipotential Grounding
CAUTION
Only use the provided power cord. Using other power cords may
result in electrical shock or injury to the operator.
Connect the provided power cord to the AC SOURCE socket on the rear panel
and plug the cord into a 3-prong AC outlet.
CAUTION
When several medical instruments are used together, ground all
instruments to the same one-point ground. Any potential difference
between instruments may cause electrical shock to the operator.
When more than one electrical instrument is used, there may be electrical
potential difference between the instruments. The potential difference between
the instruments may cause current to ow to the patient connected to the
instruments, resulting in electrical shock.
Always perform equipotential grounding when required. When equipotential
grounding is required, connect the equipotential ground terminal on the
rear panel of the analyzer to the equipotential ground terminal on the wall
(equipotential grounding system) with the equipotential grounding lead (potential
equalization conductor).
2.10 Operator’s Manual MEK-7300K
Page 49
Connecting Tubes and Installing Reagents
In order for the analyzer to operate correctly, you must install all reagents and the
waste tube before the power is turned ON.
Materials Required
• Powder-free gloves, lab coat, safety glasses
• ISOTONAC•3 diluent
• CLEANAC detergent
• CLEANAC•3 detergent
• Hemolynac•3N lysing reagent
• Hemolynac•5 lysing reagent
• Reagent inlet tubes and waste outlet tube
• Waste container (or appropriate drain)
• 2 L container
• Lint-free cloth
2. PREPARATION
2
CAUTION
Only use Nihon Kohden recommended reagents and consumables.
Otherwise the measurement result cannot be guaranteed and
incorrect reagent concentration can cause equipment damage.
CAUTION
If reagent has been frozen, it must not be used.
Diluent (ISOTONAC•3)
NOTE
• If the diluent contacts the skin or eyes or is swallowed, wash
immediately and thoroughly with water.
• Use the diluent at room temperature 15 to 30ºC (59 to 86ºF).
• When the temperature of the diluent decreases to less than 15ºC
(59ºF), it may influence WBC differential parameters (NE, LY, MO, EO,
BA) but not influence the WBC count.
• Do not put new diluent in old diluent bottles.
Operator’s Manual MEK-7300K 2.11
Page 50
2. PREPARATION
Detergent
CLEANAC•3
WARNING
Do not allow CLEANAC•3 detergent to come into
contact with acid. Contact with acids can cause
the release of poisonous chlorine gas.
CAUTION
If the CLEANAC•3 detergent contacts the skin,
eyes or mouth, wash thoroughly and immediately
with water and see a physician.
Store the detergent at room temperature 15 to 30ºC (59 to 86ºF).
CLEANAC
Do not swallow the CLEANAC detergent. If swallowed, see a physician
immediately.
WARNING
Do not swallow the CLEANAC•3 detergent. If
swallowed, see a physician immediately.
NOTE
WARNING
Store the detergent at room temperature 15 to 30ºC (59 to 86ºF).
Lysing Reagent
Hemolynac•3N
CAUTION
Do not swallow the Hemolynac•3N lysing reagent.
If swallowed, see a physician immediately.
CAUTION
If the Hemolynac•3N lysing reagent contacts the
skin, eyes or mouth, wash thoroughly and
immediately with water and see a physician.
Hemolynac•5
NOTE
CAUTION
Avoid reagent contact with the skin. If it contacts
the skin or eyes, wash thoroughly with water and
see a physician immediately.
CAUTION
Do not swallow the Hemolynac•5 lysing reagent.
If swallowed, see a physician immediately.
2.12 Operator’s Manual MEK-7300K
If the Hemolynac•5 lysing reagent contacts the
skin, eyes or mouth, wash thoroughly and
immediately with water and see a physician.
CAUTION
Page 51
Connecting Tubes
2. PREPARATION
NOTE
• Use the reagent before the expiration date on the package or label and
within the specified period after opening.
• Use only the specified cap for the hemolysing reagent. Otherwise the
reagent concentration changes and it affects the measurement.
• Do not use the reagent outside of the designated laboratory.
• Completely seal the cap of the reagent when storing it.
• Read the instructions thoroughly to use the hemolysing reagent.
• Do not put new lysing reagent in old lysing reagent bottles.
NOTE
• When connecting tubes or replacing reagents, do not let dust or
bacteria enter the port or bottle. The analyzer may get damaged.
• Do not squeeze or bend the tubes. The tube may get accidentally
disconnected or the analyzer may get damaged.
• Install the diluent and detergent container at the same level.
• Install the reagents so they do not block the vent hole for the fan.
• If necessary, cut the diluent tube and reagent tube to an appropriate
length (2.0 m or less) when the length of the tube does not fit. For the
lysing reagent and detergent, use only the specified tubes.
• Follow the instructions on each package for handling the diluent and
detergent.
• For handling the lysing reagent, refer to the manual of the reagent.
• After installing the reagents, do not block the vent hole on the bottle
cap.
2
Lysing reagent
(Hemolynac•3N)
Lysing reagent
(Hemolynac•5)
Detergent
(CLEANAC)
Waste container
Diluent
Detergent
(CLEANAC•3)
Operator’s Manual MEK-7300K 2.13
Page 52
2. PREPARATION
Diluent Tube
NOTE
Try to keep the diluent container at the same level as the analyzer.
1. Connect the diluent tube (marked blue) to the ISO3
inlet on the right side panel.
2. Pass the diluent tube through the 18 L container cap.
18 L container cap
Diluent tube
(marked blue)
Diluent (18 L)
18 L tube assy 2
Detergent Tube
Only use the specified detergent tubes for the detergent.
For performing strong clean, use CLEANAC•3 detergent. For other purposes,
use CLEANAC detergent.
CLEANAC Detergent
Try to keep the CLEANAC detergent container at the same level as the
analyzer.
3. Connect the end of the diluent tube to the 18 L tube
assy 2.
4. Put the 18 L tube assy 2 into the diluent container
and tighten the 18 L container cap.
NOTE
NOTE
Detergent tube
(marked green)
CLEANAC
detergent (5 L)
18 L container cap
Cleanac tube
assy
1. Connect the detergent tube (marked green) to the
CLN inlet on the right side panel.
2. Pass the detergent tube through the 18 L container
cap.
3. Connect the end of the detergent tube to the cleanac
tube assy.
4. Put the cleanac tube assy into the CLEANAC
container and tighten the 18 L container cap.
2.14 Operator’s Manual MEK-7300K
Page 53
2. PREPARATION
CLEANAC•3 Detergent
NOTE
Place the CLEANAC•3 detergent container at the same level as the
analyzer.
1. Connect the cleanac tube 8 (marked white) to the
CLN3 inlet on the side panel.
2. Pass the cleanac tube 8 through the 18 L cap.
2
Cleanac tube 8
(marked white)
CLEANAC•3 detergent
Detergent inlet
(for CLEANAC•3)
18 L cap
Cleanac tube
assy (8222)
Detergent tube
(marked white)
1 L cap
1 L tube assy
3. Connect the end of the cleanac tube 8 to the cleanac
tube assy (8222).
4. Put the cleanac tube assy (8222) into the
CLEANAC•3 container and tighten the 18 L cap.
When the CLEANAC•3 (1 L) is used
1. Connect the detergent tube (marked white) to the
CLN3 inlet on the side panel.
2. Pass the end of the detergent tube into the 1 L cap
and attach the 1 L tube assy.
3. Put the 1 L tube assy into the tank and tighten the
cap.
Operator’s Manual MEK-7300K 2.15
Page 54
2. PREPARATION
When the YZ-0066 reagent bottle set is used
1. Pour the detergent (CLEANAC•3) into the 2 L tank.
Detergent inlet
(for CLEANAC•3)
2 L tank
Detergent tube
(marked white)
MEK cap
Tube assy
Lysing Reagent Tube
• Place the lysing reagent container at the same level as the analyzer.
• Only use the specified lysing reagent tubes for the lysing reagent.
Hemolynac•3N
2. Connect the detergent tube (marked white) to the
CLN3 inlet on the right side panel.
3. Pass the end of the detergent tube into the MEK cap
and attach the tube assy.
4. Put the tube assy into the tank and tighten the cap.
NOTE
1. Replace the lysing reagent cap with the
Hemolynac•3 cap (yellow) and tighten the cap.
Lysing reagent inlet
(for Hemolynac•3N)
Hemolynac•3 cap (yellow)
Place the tip of the tube at
the bottom of the bottle.
Hemolynac•3
tube (570)
(marked yellow)
Attach the tape
and cap tightly.
2. Connect the Hemolynac•3 tube (570) (marked
yellow) to the HEMO3N inlet on the right side
panel.
3. Put the other end of the tube into the lysing reagent
container through the Hemolynac•3 cap (yellow).
Attach the tape and cap tightly and place the tip of
the tube at the bottom of the bottle.
4. When using the reagent tray, place the lysing reagent
container on the upper shelf of the reagent tray.
2.16 Operator’s Manual MEK-7300K
Page 55
Hemolynac•5
2. PREPARATION
1. Replace the hemolysing reagent cap with the
Hemolynac•5 cap (black) and tighten the cap.
2
Lysing reagent inlet
(for Hemolynac•5)
Hemolynac•5 cap (black)
Place the tip of the tube at
the bottom of the bottle.
Hemolynac•5
tube (570)
(marked black)
Attach the tape
and cap tightly.
Waste Fluid Tube
Try to keep the waste container at the same level as the analyzer.
2. Connect the Hemolynac•5 tube (570) (marked black)
to the HEMO5 inlet on the side panel.
3. Put the other end of the tube into the hemolysing
reagent container through the Hemolynac•5 cap
(black). Attach the tape and cap tightly and place the
tip of the tube at the bottom of the bottle.
NOTE
1. Connect the waste tube (marked red) to the waste
outlet on the right side panel.
Waste tube
(marked red)
Waste container
MEK cap
Waste tube assy
2. Pass the waste tube through the MEK cap.
3. Connect the waste tube to the waste tube assy.
4. Insert the waste tube assy into the waste container
and tighten the MEK cap.
Operator’s Manual MEK-7300K 2.17
Page 56
2. PREPARATION
Turning the Laser Switch On
1. Insert the laser key into the laser switch hole on the
right side panel.
2. Turn the laser switch to ON.
NOTE
• When the measurement is performed with the
laser switch turned off, CBC can be measured but
an alarm is displayed and there is no WBC 5 part
differential scattergram and data.
• Store the key in a safe place.
2.18 Operator’s Manual MEK-7300K
Page 57
2. PREPARATION
Loading Recording Paper in the WA-730V Printer Unit (Option)
NOTE
• The WA-730V printer unit is optional (built-in type).
• Only use the specified recording paper.
1. Open the printer door.
2. Set the recording paper in the paper tray in the direction as shown.
2
3. Insert the recording paper into the slot.
4. Press the [ Feed] key on the front panel until the recording paper comes
out from the printer unit.
5. Put the paper through the opening of the door and close the door.
6. Cut the extra paper.
Operator’s Manual MEK-7300K 2.19
Page 58
2. PREPARATION
Main power switch
Main power lamp
Turning Power On
Check Before Turning Power On
Check the following items before turning on the power.
Accessories and
consumables
Connection and settings
AppearanceNo scratches, dirt or leakage (especially in the
Quality controlRun quality control by measuring a hematology
Use after long term
storage
ItemCheck
Sufcient diluent, detergent and lysing reagent.
Power cord is connected properly.
Grounding lead is connected properly when
equipotential grounding is required.
Tubes are connected properly.
Diluent, detergent and lysing reagent containers are
connected and have no dust in them.
Waste container is in place and empty.
Enough recording paper in the optional printer.
External instruments (e.g. PC and printer) are
properly connected.
measurement baths, sub baths and pump tube).
No key or switch is broken.
No damage to the power cord.
Waist container is not full.
Analyzer is not wet.
control.
Aperture caps are clean.
Pump tube and pinch valve tube are not broken and
not disconnected.
Turning On the Power
2.20 Operator’s Manual MEK-7300K
1. Conrm that the analyzer, printer (optional), and ZK-820V hand-held bar
code reader (optional) power cords are connected to grounded power outlets.
2. Set the printer power switch ON.
3. Press the [Main power] switch on the rear panel to ON. The [Main power
lamp] on the front panel lights.
Always leave the main power ON except for storage and transport of the
analyzer.
4. Press the [Power] key on the front panel ON. The [Power lamp] lights
and the screen illuminates within 15 to 30 seconds. Cleaning of the uid
path, priming and circuit self-check are automatically performed, and the
“Checking detergent” message appears. When the laser switch on the right
side panel is turned on, the laser lamp also lights.
Power keyPower lamp
Page 59
2. PREPARATION
If there is an error, “Fail” appears on the screen.
After priming operation is completed, the Ready screen appears.
2
NOTE
If an error message appears, refer to “Alarm Messages” in Section 10.
5. Perform strong cleaning by pressing the Operation key → Strong clean key.
6. After strong cleaning, press the OK key.
7. Press the Ready key to display the Ready screen.
Operator’s Manual MEK-7300K 2.21
Page 60
2. PREPARATION
Checking the Date and Time Settings
After turning the analyzer ON, date and time must be checked. To correct the
date and time on the upper right corner of the screen, refer to “Setting Date and
Time” in Section 5.
Cleaning the Analyzer
After installing the analyzer, turn the power on to rell reagents into the analyzer.
After installing the analyzer or long term storage, perform strong cleaning. Refer
to “Using the Analyzer after Storage” in section 9.
Check After Turning On the Power
Check the following items after turning on the power to start operating safely and
properly. If any problem is detected, take the proper countermeasure according to
Section 10 “Messages and Troubleshooting”.
ItemCheck
Turning on the power
Basic operation
After long term storagePerform Strong cleaning on the Operation screen.
At the start of the day, check that the date and time settings are correct.
There is no re, smoke or smell.
The analyzer is not too hot.
There is no electric shock.
The main power and power lamps light.
No alarm message is displayed on the screen.
The messages are displayed properly.
Keys and switches operate properly.
The touch screen keys function properly.
The lamps and LED indication are correct.
The measured background noise values are proper.
The measured hematology control values are proper.
The printer works properly.
The date and time are correct.
No alarm message is displayed on the screen during
operation.
NOTE
Optical Adjustment
After installing the analyzer, perform optical adjustment to measure samples
correctly. Refer to “Performing Optical Adjustment” in section 5.
2.22 Operator’s Manual MEK-7300K
Page 61
Checking Daily Accuracy
Measuring Background Noise
2. PREPARATION
To assure measurement reliability, check the analyzer daily before measurement.
For details on the measurement reliability, refer to Section 11 “Quality Control”.
• Count the diluent to measure background noise.
• Count the MEK-5D hematology control to check accuracy.
Count the diluent to measure background noise. When using the analyzer
in closed mode, measure background noise in closed mode. When using the
analyzer in open mode, measure background noise in open mode. Background
noise increases in the following cases.
ProblemCountermeasure
Old diluent. Germs begin to breed in the
diluent 60 days after opening.
Dust in the diluent container.Replace diluent.
Extremely high or low diluent
temperature (normal range is 15 to 30°C
(59 to 86°F)).
The sampling nozzle is clogged and
bubbles occur in the sub bath.
Replace diluent.
Adjust diluent temperature to 15 to
30°C (59 to 86°F).
Replace the sampling nozzle with a
new one.
2
There are two ways to measure background noise. One is by pressing the
[ Count] switch. The other is using the Background screen of the Other screen.
When measured on the Background screen, “Fail” appears beside the parameter
which is over the acceptable value on the result screen.
Measuring Background Noise in Closed Mode
Measuring by Pressing the [ Count] Switch
1. Select “Closed” for sampling mode on the Ready screen.
2. Press the [ Eject] key on the front panel. The tube holder opens.
3. Set an empty tube in the tube holder.
4. Close the tube holder. Diluent is measured. If the tube holder is closed
without setting a tube, measurement is not performed.
5. The tube holder opens automatically after nishing diluent measurement and
the result is displayed.
Operator’s Manual MEK-7300K 2.23
Page 62
2. PREPARATION
Measuring on the Background Screen
1. Check that “Closed” is selected for sampling mode and press the Operation
key on the Ready screen.
2. Press the Background key on the Operation screen.
The “Measure background noise?” message appears on the screen.
3. Press the Yes key to measure background noise. The result is displayed after
the measurement is complete.
2.24 Operator’s Manual MEK-7300K
Page 63
Measuring Background Noise in Open Mode
Measuring by Pressing the [ Count] Switch
1. Select “Open” for sampling mode on the Ready screen.
2. PREPARATION
2
2. Press the [ Count] switch to count the diluent. There is no need to aspirate
the diluent from the sampling nozzle. The result is displayed after the
measurement is complete.
Measuring on the Background Screen
1. Check that “Open” is selected for sampling mode and press the Operation
key on the Ready screen.
Operator’s Manual MEK-7300K 2.25
Page 64
2. PREPARATION
2. Press the Background key on the Operation screen.
The “Measure background noise?” message appears on the screen.
3. Press the Yes key to measure background noise. The result is displayed after
the measurement is complete.
Results
The result is displayed on the screen after measurement.
Make sure that the values are less than or equal to the following values.
WBC: 0.2 (×103/µL)
RBC: 0.05 (×106/µL)
HGB: 0.1 (g/dL)
PLT: 10 (×103/µL)
When measured on the Background screen, “Fail” appears beside any parameter
which is over the acceptable value.
2.26 Operator’s Manual MEK-7300K
Page 65
2. PREPARATION
Disregard the other parameter values because noise does not affect them.
Measuring Hematology Control
If the values are greater than the values listed above, check the following items,
press the [ Clean] key on the front panel to clean the uid path, and recount the
diluent.
• The diluent is clean.
• No bubbles in the diluent.
• The aperture caps are clean.
• The aperture caps are rmly attached.
• The measurement baths and sub baths are clean.
To check the measurement accuracy of the analyzer, measure a hematology
control.
NOTE
The hematology control is for controlling the accuracy of the analyzer. It is
made from healthy human blood. Because living cells are used, handle and
store the control correctly according to the explanation of the assay sheet.
2
1. On the Ready screen, set the measurement settings as follows.
• Sample type: Control
On the control select window, select the control to use and
press the OK key. When the Cancel is pressed, the Sample
mode returns to Normal.
• Measure mode: Normal
• Parameter: CBC + Diff
Operator’s Manual MEK-7300K 2.27
Page 66
2. PREPARATION
Sampling nozzle
Count
switch
Set the control type as follows.
Hematology CntrolSetting
MEK-5DNNormal
MEK-5DLLow
MEK-5DHHigh
2. Measure a hematology control.
NOTE
• Before measurement, make sure that the control setting matches
the control.
• After measurement, the Sample type returns to Normal.
Open mode
i) Remove the cap of the hematology control and put the sampling nozzle
into the control as shown in the illustration.
NOTE
Do not let the sampling nozzle touch the bottom of the
hematology control. This may prevent aspiration of the control.
ii) Press the count switch to start measurement. When the switch is pressed,
the sample is aspirated, diluted in the analyzer and measured. During
measurement, the “Measuring” message is displayed.
Put the nozzle to
this position.
NOTE
Do not lower the sample while it is being aspirated.
2.28 Operator’s Manual MEK-7300K
Page 67
Eject key
Closed mode
Set the hematology control to the tube holder.
i) Press the eject [ ] key to open the tube holder.
ii) Set the hematology control into the tube holder.
2. PREPARATION
2
Tube holder
iii) Close the tube holder. The measurement automatically starts. During
measurement, the “Measuring” message is displayed.
3. Check that the measurement value is within the range of the expected value
on the assay sheet.
Operator’s Manual MEK-7300K 2.29
Page 68
2. PREPARATION
Preparing SD Card
Use a QM-001D or specied SD memory card for this analyzer.
• Only use the specified SD card. Otherwise, data cannot be saved or the
analyzer may get damaged.
• The operation of a mini or micro SD card with conversion adapter
cannot be guaranteed.
• An SDHC or SDXC memory card cannot be used.
Safety Information about the SD Card
For handling and safety information about SD card, refer to the SD card
manual together with this manual.
NOTE
NOTE
Handling and Storage
• Keep the SD card slot clean. If dust gets into the slot, the SD card will
not function.
• Do not handle the SD card while eating or drinking.
• Do not get the SD card wet.
• Do not give impact to the SD card by dropping or bending.
• Do not expose the SD card to direct sunlight or leave it in a high
temperature place.
Environmental Conditions
• Storage environment
Temperature: –20 to +65°C (–4 to +149°F)
Humidity: less than 95% (noncondensing)
• Operating environment: same as the analyzer
Data Processing
• Initialize the SD card on this hematology analyzer. An SD card that was
initialized on other instruments cannot be used. Refer to “Formatting SD
Card” in Section 5.
• Do not use an unspecified instrument to save data to the SD card. The
card may get damaged.
• Do not remove the SD card or turn off the power during data
processing. The card may get damaged.
2.30 Operator’s Manual MEK-7300K
Page 69
Inserting and Removing the SD Card
2. PREPARATION
NOTE
• When inserting or removing a SD card, turn off the power of the
hematology analyzer. The data in the card may get damaged or the
hematology analyzer might not operate correctly.
• Do not bend the SD card or insert it at an angle.
Insert the SD card into the SD card slot so that the label on the SD card is toward
front.
When removing the SD card, push the card once and remove it.
2
Operator’s Manual MEK-7300K 2.31
Page 70
Section 3 Principles of Operation
Operation Theory ................................................................................................................................................. 3.2
Electric Cell Counting ................................................................................................................................3.2
The hematology analyzer uses the volumetric impedance method of cell counting.
In this method, an electrolytic solution (diluent) containing suspended blood cells
is aspirated through the aperture. Two electrodes, an internal electrode and an
external electrode, are located close to the aperture and a constant current ows
between them. When a blood cell passes through the aperture, the resistance
between the electrodes momentarily increases and a very small voltage change
occurs corresponding to the resistance. The voltage signal is amplied and is sent
to the electronic circuit.
A threshold circuit eliminates signals caused by electrical noise, dust, debris and
particles which are smaller or larger than blood cells.
Detection circuit
Constant
current circuit
Electrodes
Electrolytic solution
Aperture
To nd the peak values, the signals are sent to the A/D converter. The acquired
data is stored in memory for each individual peak value. The data is corrected by
the CPU and displayed on the screen.
CPU
AmplifierThreshold circuit
A/D converter
Memory
The number of signals for each size cell is stored in memory as a histogram.
Counted cells of RBC and PLT can have overlap sizes so the CPU can
discriminate the count for each type of cell. See the “Red Blood Cell and Platelet
Counting” section.
Sometimes two or more cells pass through the aperture at the same time. This is
called coincidence. When the sample solution is sufciently diluted and mixed,
this can be statistically predicted to a high degree of accuracy. The software
contains a coincidence correction table to compensate for this.
3.2 Operator’s Manual MEK-7300K
Page 72
Count
3. PRINCIPLES OF OPERATION
Red Blood Cell and Platelet Counting
Data of the RBC and PLT pulses are stored in the memory as a histogram.
Normal blood shows a clear separation between the PLT volume range and RBC
volume range (Fig. A) so an accurate PLT count is easily acquired. However,
when abnormal blood such as microcytic blood is counted (Figs. B and C),
the separation is unclear. In these cases, for accurate PLT counting, the CPU
determines the PLT and RBC distribution pattern, and sets the upper threshold
(PLT HI) to the lowest count.
Count
3
PLT LO*
Count
volume range
PLT LO*
Normal blood
PLT HI*
PLT
PLT
Macro platelet
PLT
PLT
volume range
Fig. A
RBC
volume range
PLT HI*
Fig. C
RBC
RBC
volume range
amplitude
(volume)
RBC
amplitude
(volume)
Microcytic blood (Microcythemia)
PLT LO*
volume range
PLT HI*
PLT
PLT
volume range
Fig. B
PLT LO*: Lower threshold
PLT HI*: Upper threshold
RBC
RBC
amplitude
(volume)
Principle of Hydraulic Operation
For accurate counting of the blood cells in a diluent solution, a constant volume
of solution must be aspirated through the aperture. The manometer controls this
volume by measuring diluent level by an optical sensor and aspirating the diluent
by the rotary pump.
Aspiration and dispensing of the sample are performed alternately by an
electromagnetic valve. The waste uid is drained from the hematology analyzer
by the rotary pump.
The priming of diluent and cleaning with detergent are also performed by the
electromagnetic valves.
The CPU controls the electromagnetic valves, pump rotations and rotation
direction. If an air bubble enters the manometer, the CPU generates an alarm
sound and displays an error message on the screen.
Operator’s Manual MEK-7300K 3.3
Page 73
3. PRINCIPLES OF OPERATION
Hemoglobin Measurement
Chemical Processing
A hemolysing reagent is added to the diluted blood sample to break the red blood
cell membrane and release the hemoglobin.
Spectrophotometric Measurement Method
The hemoglobin is measured by spectrophotometry. This method measures the
optical density of the sample solution. The optical density is proportional to the
amount of hemoglobin in the sample solution.
Spectrophotometric measurement is based on the principle that different
materials absorb different amounts of different wavelengths of light. When light
strikes a material, some of the light is absorbed by the material and some passes
through. The amount of absorbed light at each frequency forms a unique optical
“ngerprint” for that material.
In this hematology analyzer, hemoglobin is measured in the measurement
baths. An LED shines one wavelength of light through the sample solution. A
photodiode receives the light which is not absorbed by the solution. The amount
of received light is converted to an electric signal which is amplied by the
preamplier. The amplied signal is sent to the A/D converter.
The amplied signals from the sample and diluent are required for measuring
the hemoglobin concentration. The sample data is sent to the CPU. The CPU
converts the ratio of these data to logarithmic data, multiplies it by the calibration
coefcient and displays the result on the LCD.
After measurement, the sample and diluent are drained from the hematology
analyzer. The sample is a highly concentrated protein solution. If the sample is
left in the measurement baths for a long time, the measurement baths gradually
become dirty. To prevent this problem, the measurement baths are automatically
cleaned by dispensing the diluent after each measurement.
Hemoglobin absorption characteristics
Absorbance
(0.D)
0.5
Cyanmethemoglobin
Non-cyan
surfactant
Oxidized hemoglobin
520500
540560580
Wavelength (nm)
3.4 Operator’s Manual MEK-7300K
Page 74
Principle of WBC Differential Operation
The hematology analyzer uses the light scatter technique to differentiate WBC
into neutrophil, lymphocyte, monocyte, eosinophil and basophil counts.
3. PRINCIPLES OF OPERATION
The diluted blood sample is injected into the ow cell. Blood cells pass through
the sensing zone in a single le. A laser beam through the sensing zone is
scattered by the passing cells and the scattered light is detected. The angle and
intensity of scattered light depend on the volume and characteristics of the cell.
From this, WBC can be differentiated into 5 parts.
For details about the scattergrams, refer to “Measurement Results” later in this
section.
Laser
Granularity
Complexity
Size
3
Sample
DiluentDiluent
Operator’s Manual MEK-7300K 3.5
Page 75
3. PRINCIPLES OF OPERATION
Measurement Results
WBC Scattergrams
There are 3 WBC scattergrams. You can compare 3 scattergrams at a time. You
can also enlarge one scattergram to examine it.
The main scattergram is divided into 3 areas.
Monocyte and
basophil count
Lymphocyte count
The lymphocyte count is acquired from the lymphocyte area.
The NE-EO scattergram is the neutrophil, eosinophil and ghost 2 area expanded
with the granularity as the horizontal axis. The neutrophil count and eosinophil
count can be acquired from this scattergram.
Neutrophil count
Neutrophil and
eosinophil count
Eosinophil count
The MO-BA scattergram is the monocyte and basophil area expanded with the
granularity as the horizontal axis. The monocyte count and basophil count can be
acquired from this scattergram.
3.6 Operator’s Manual MEK-7300K
Page 76
3. PRINCIPLES OF OPERATION
Histograms
Monocyte count
Basophil count
Lymphocyte count
On the Result screen, you can view three histograms. You can examine the
histograms when anisocytosis occurs or hemolyzation is poor.
Numeric result
When touched,
it enlarges to
full screen.
3
Vertical Axis Type for the Histogram
Counted value (COUNTED): The vertical axis on the graph represents
numerical count values.
Percentage (%): The vertical axis on the graph is xed at
100% full scale regardless of the total counted
value. Therefore, the trend of particle volume
distribution can be easily judged even though
the total counted value is not shown.
(Count)
Counted Value
(%)
100%
(fL)
Percentage
The horizontal axis represents the blood cell volume.
(fL = 1 × 10
-15
L)
(fL)
Operator’s Manual MEK-7300K 3.7
Page 77
3. PRINCIPLES OF OPERATION
RBC (Red Blood Cell) Distribution Histogram
The RDW indicates the deviation ratio of the red blood cell volumes on the
histogram.
RDW-CV (%)=
MCV
Standard deviation of red blood cell volumes (SD)
Mean cell volume (MCV)
SD
SD
RDW-SD (fL) = Standard deviation of red blood cell volumes (SD) × 4
(= 2 SD × 2)
2SD 2SD
(fL)
RDW-SD
PLT (Platelet) Distribution Histogram
The following parameters are automatically calculated from the PLT distribution
histogram.
• Platelet Crit (PCT)
The PCT indicates the ratio of the platelet total volume on the histogram to the
aspirated whole blood volume.
• Mean Platelet Volume (MPV)
The MPV indicates the mean of the platelets volume on the histogram.
• Platelet Distribution Width (PDW)
The PDW indicates the deviation ratio of the platelet volumes on the
histogram.
MPV
SD
SD
10
20
3.8 Operator’s Manual MEK-7300K
Page 78
Flags
Flags are displayed when the measurement result is above or below the threshold
for each item. The following list is the factory default settings. You can change
the threshold on the Flags screen of the Settings screen. Refer to “Flag Settings”
in Section 5.
You can enlarge ags by pressing the ag on the Result screen.
Close the enlarged
ag window.
3. PRINCIPLES OF OPERATION
3
Change the page.
Flag Appears when
WBC
• Leukocytosis WBC is above 18 × 103/µL
• Leukopenia WBC is below 2.5 × 103/µL
• Neutrophilia NE is above 11 × 103/µL
• Neutropenia NE is below 1.0 × 103/µL
• Lymphocytosis LY is above 4.0 × 103/µL
• Lymphopenia LY is below 0.8 × 103/µL
• Monocytosis MO is above 1.0 × 103/µL
• Eosinophilia EO is above 0.7 × 103/µL
• Basophilia BA is above 0.2 × 103/µL
• Blasts Presence of blasts is suspected
• Immature Gr Presence of immature granulocytes is suspected
• Left Shift Left shifted cells
• Atypical Ly Presence of atypical lymphocytes is suspected
• Small Nucleated Cell Presence of small nucleated cells is suspected
• Ly-Mo Interference Overlap of population of lymphocytes and monocytes
is suspected
• Ne-Eo Interference Overlap of population of neutrophils and eosinophils
is suspected
RBC
• Erythrocytosis RBC is above 6.5 × 106/µL
• Anemia HGB is below 10.0 g/dL
• Anisocytosis RDW-CV is above 20.0%
• Microcytosis MCV is below 70 fL
• Macrocytosis MCV is above 110 fL
• Hypochromia MCHC is below 29.0 g/dL
• Abnormal MCHC MCHC is below 28.0 g/dL or above 38.0 g/dL
Operator’s Manual MEK-7300K 3.9
Page 79
3. PRINCIPLES OF OPERATION
PLT
• Thrombocytosis PLT is above 600 × 103/µL
• Thrombocytopenia PLT is below 60 × 103/µL
• PLT Clumps Presence of PLT clumps is suspected
• PLT-RBC Interference Overlap of population of PLT and RBC is suspected
• If “C” appears on the right side of the WBC or PLT data, the PLT may be
coagulated.
If “*” appears on the right side of the measurement value, the measurement
result might not be reliable. Check the sample with the manual differential
counts.
Dilution Ratio .............................................................................................................................................4.3
Counting Time ...........................................................................................................................................4.3
Data Storage ............................................................................................................................................. 4.3
Power Requirements ................................................................................................................................. 4.4
Dimensions and Weight ............................................................................................................................. 4.4
Bar Code Specifications ............................................................................................................................4.5
Transfer Format PC (V03-01) .............................................................................................................................. 4.6
Transfer Example to a PC..........................................................................................................................4.6
Transfer Format PC (V03-02) .............................................................................................................................. 4.8
Transfer Example to a PC..........................................................................................................................4.8
Transfer Format PC (V05-01) ............................................................................................................................ 4.10
Transfer Example to a PC........................................................................................................................4.10
ASTM Communication ....................................................................................................................................... 4.13
4
Operator’s Manual MEK-7300K 4.1
Page 81
4. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
Specifications
Measured Parameters, Ranges and Reproducibility to Specimen
Specications except WBC population were determined using hematology control blood (MEK-5DN), counted 10 times
consecutively.
Measured ParametersMeasuring Range
WBC: White blood cell count0 to 299 × 103/µLwithin 2.0%CV
NE%: Neutrophil percent0 to 99.9%within 5.0%CV
LY%: Lymphocyte percent0 to 99.9%within 5.0%CV
MO%: Monocyte percent0 to 99.9%within 12.0%CV
EO%: Eosinophil percent0 to 99.9%within 20.0%CV
BA%: Basophil percent0 to 99.9%
NE: Neutrophil count0 to 299 × 103/µL
LY: Lymphocyte count0 to 299 × 103/µL
MO: Monocyte count0 to 299 × 103/µL
EO: Eosinophil count0 to 299 × 103/µL
BA: Basophil count0 to 299 × 103/µL
RBC: Red blood cell count0 to 14.9 × 106/µLwithin 1.5%CV
HGB: Hemoglobin concentration0 to 29.9 g/dLwithin 1.5%CV
HCT: Hematocrit0 to 99.9%—
MCV: Mean cell volume20 to 199 fLwithin 1.0%CV
MCH: Mean cell hemoglobin10 to 50 pg
RDW-CV: Red blood cell distribution width0 to 50%
PLT: Platelet count0 to 1490 × 103/µLwithin 4.0%CV
PCT: Platelet crit0 to 2.9%
PDW: Platelet distribution width0 to 50%
within CV30.0% (>2%) or average value ±1% (0
to 2%)
Reproducibility to Specimen
(CV: Coefficient of Variation)
—
—MCHC: Mean cell hemoglobin concentration10 to 50 g/dL
Open mode: 63 s/sample (from measurement start to data display)
Closed mode: 75 s/sample (from measurement start to data display)
Display
Display: 10.4 inch, LCD with backlight and touch screen keys
Resolution: 800 × 600 dots
Screen size: approx. 211.2 × 158.4 mm
Display contents: Numerical data, scattergrams, histograms, measuring conditions, alarm message and other messages,
touch screen keys
Data Storage
Numerical data for all counted parameters for up to 400 samples and histograms and scattergrams for up to 50 samples
Environmental Conditions
Storage temperature: –20 to +60ºC (–4 to +140°F)
Operating temperature: 15 to 30ºC (59 to 86°F)
Storage humidity: 10 to 95%
Operating humidity: 30 to 85% (noncondensing)
Storage atmospheric pressure: 70 to 106 kPa
Operating atmospheric pressure: 70 to 106 kPa
Operating altitude: less than 3000 m
Operator’s Manual MEK-7300K 4.3
Page 83
4. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
Power Requirements
Power requirements: 110 to 240 V ± 10% AC, 50/60 Hz
Power consumption: 250 VA
Dimensions and Weight
Dimensions: 382 W × 465 D × 532 H (mm)
Net weight: approx. 35 kg
Electromagnetic Compatibility
IEC 61326-1: 2005
IEC 61326-2-6: 2005
EN 61326-1: 2006
EN 61326-2-6: 2006
CISPR11: Edition 4: 2003, Group 1, Class B
EN 55011: 2007, Group 1, Class B
The power supply short interruption test is performed through a transformer which has at least three times the power
capacity of the instrument.
Type of protection against electrical shock: CLASS I EQUIPMENT
Degree of protection against harmful ingress of water: IPX0 (non-protected)
Degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH
OXYGEN OR NITROUS OXIDE: EQUIPMENT not suitable for use in the presence of
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR
WITH OXYGEN OR NITROUS OXIDE
Mode of operation: CONTINUOUS OPERATION
EQUIPMENT types (classication): Indoor stationary EQUIPMENT
Pollution Degree: 2 EQUIPMENT
Requirements for marking of IN VITRO DIAGNOSTIC instruments: EN1658: 1996
4.4 Operator’s Manual MEK-7300K
Page 84
4. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
Bar Code Specifications
Bar Code Format
The following formats with or without check digits are acceptable:
• Industrial 2 of 5
• ITF
• JAN/EAN/UPC
• NW-7
• CODE 39
• CODE 93
• CODE 128
Bar Code Label Specications
Refer to Appendix Bfor complete information on bar code label formats, check digits and specications.
4
Operator’s Manual MEK-7300K 4.5
Page 85
4. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
Transfer Format PC (V03-01)
Transfer Example to a PC
Common data block “_” in the Transferred Data in Characters column indicates a blank (0x20).
No. of
Transfer ItemsExample
<Start> Common data block
Start of Text0x02
Hematology analyzer information
Type“MEK-7222”
Parameter no“22”
Send data bytes“01024”
Sampling mode“MANUAL”
Parameter“CBC + Diff”
Sample code“01”
Extended data block “_” in the Transferred Data in Characters column indicates a blank (0x20).
Transfer ItemsExample
<Start> Extended data block
Start of Text0x021STX02
Hematology analyzer information
Identifier“EXP”4E X P CR45 58 50 0D
Send data bytes“00512”60 0 5 1 2 CR30 30 35 31 32 0D
Type“MEK-7222”11M E K - 7 2 2 2CR4D 45 4B 2D 37 32 32 32 20 20 0D
Unit no“1”31 CR20 31 0D
Work list data
Name“DAVID”27D A V I D...…CR44 41 56 49 44 … … 20 20 20 20 … … 0D
Sex“MALE”7M A L ECR4D 41 4C 45 20 20 0D
Date of birth“19800219”111 9 8 0 CR 0 2 CR 1 9 CR31 39 38 30 0D 30 32 0D 31 39 0D
Age“22”42 2 CR20 32 32 0D
Department“INTERNAL”14I N T E R N A LCR49 4E 54 45 52 4E 41 4C 20 20 20 20 20 0D
Physician“WATSON”27W A T S O N......CR57 57 41 54 53 4F … … 20 20 20 … … 0D
Operator name“”9CR20 20 20 20 20 20 20 20 0D
Comments“No problem.”129N o p r o b l e m . ...4E 4E 6F 20 70 72 6F 62 6C 65 6D
Normal range table no“0”20 CR30 0D
Work list flag“1”21 CR31 0D
Control mode flag“”2CR20 0D
Reserve data for work list data32SP × 31 Byte + CR × 1 Byte20 × 31 Byte + 0D × 1 Byte
Normal range setting data
WBC-LOW“4.0”54 . 0 CR20 34 2E 30 0D
WBC-HIGH“9.0”59 . 0 CR20 39 2E 30 0D
NE%-LOW“42.0”54 2 . 0 CR34 32 2E 30 0D
NE%-HIGH“85.0”58 5 . 0 CR38 35 2E 30 0D
LY%-LOW“11.0”51 1 . 0 CR31 31 2E 30 0D
LY%-HIGH“49.0”54 9 . 0 CR34 39 2E 30 0D
MO%-LOW“0.0”50 . 0 CR20 30 2E 30 0D
MO%-HIGH“9.0”59 . 0 CR20 39 2E 30 0D
EO%-LOW“0.0”50 . 0 CR20 30 2E 30 0D
EO%-HIGH“3.0”53 . 0 CR20 33 2E 30 0D
BA%-LOW“0.0”50 . 0 CR20 30 2E 30 0D
BA%-HIGH“2.0”52 . 0 CR20 32 2E 30 0D
NE-LOW“1.7”51 . 7 CR20 31 2E 37 0D
NE-HIGH“7.7”57 . 7 CR20 37 2E 37 0D
LY-LOW“0.4”50 . 4 CR20 30 2E 34 0D
LY-HIGH“4.4”54 . 4 CR20 34 2E 34 0D
MO-LOW“0.0”50 . 0 CR20 30 2E 30 0D
MO-HIGH“0.8”50 . 8 CR20 30 2E 38 0D
EO-LOW“0.0”50 . 0 CR20 30 2E 30 0D
EO-HIGH“0.3”50 . 3 CR20 30 2E 33 0D
BA-LOW“0.0”50 . 0 CR20 30 2E 30 0D
BA-HIGH“0.2”50 . 2 CR20 30 2E 32 0D
RBC-LOW“3.80”53 . 8 0 CR33 2E 38 30 0D
RBC-HIGH“5.30”55 . 3 0 CR35 2E 33 30 0D
HGB-LOW“11.0”51 1 . 0 CR31 31 2E 30 0D
HGB-HIGH“17.0”51 7 . 0 CR31 37 2E 30 0D
HCT-LOW“36.0”53 6 . 0 CR33 36 2E 30 0D
HCT-HIGH“56.0”55 6 . 0 CR35 36 2E 30 0D
MCV-LOW“80.0”58 0 . 0 CR38 30 2E 30 0D
MCV-HIGH“100”51 0 0 CR20 31 30 30 0D
MCH-LOW“28.0”52 8 . 0 CR32 38 2E 30 0D
MCH-HIGH“36.0”53 6 . 0 CR33 36 2E 30 0D
MCHC-LOW“31.0”53 1 . 0 CR33 31 2E 30 0D
MCHC-HIGH“37.0”53 7 . 0 CR33 37 2E 30 0D
RDW-LOW“11.5”51 1 . 5 CR31 31 2E 35 0D
RDW-HIGH“16.5”51 6 . 5 CR31 36 2E 35 0D
PLT-LOW“120”51 2 0 CR20 31 32 30 0D
PLT-HIGH“380”53 8 0 CR20 33 38 30 0D
PCT-LOW“0.10”50 . 1 0 CR30 2E 31 30 0D
PCT-HIGH“1.00”51 . 0 0 CR31 2E 30 30 0D
MPV-LOW“5.0”55 . 0 CR20 35 2E 30 0D
MPV-HIGH“10.0”51 0 . 0 CR31 30 2E 30 0D
PDW-LOW“12.0”51 2 . 0 CR31 32 2E 30 0D
PDW-HIGH“18.0”51 8 . 0 CR31 38 2E 30 0D
<End> Extended data block
End of Text0x031ETX03
No. of
Transferred Data in CharactersTransferred Data in Hexadecimal
Byte
... CR … … 20 20 20 20 20 20 … … 0D
4
• “0x0D” is at the end of each item as separating character.
• Each item is entered by ASCII code. Only “Start of Text” and “ End of Text” are entered by control code.
• Each item is initialized at space (0x20). Unused items are also initialized at space and have “0x0D” in the end.
Operator’s Manual MEK-7300K 4.7
Page 87
4. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
Transfer Format PC (V03-02)
Transfer Example to a PC
Common data block “_” in the Transferred Data in Characters column indicates a blank (0x20).
No. of
Transfer ItemsExample
<Start> Common data block
Start of Text0x02
Hematology analyzer information
Type“MEK-7222”
Parameter no“22”
Send data bytes“01024”
Sampling mode“MANUAL”
Parameter“CBC + Diff”
Sample code“01”
Sample label“BLOOD”
Rack location“MMM”
Seq#“0000001”
Software version“V02-01”
Analysis program version
Format version“V03-01”
Total data bytes“01536”
Data block pattern “1”
Reserve data
Measurement data
Extended data block “_” in the Transferred Data in Characters column indicates a blank (0x20).
Transfer ItemsExample
<Start> Extended data block
Start of Text0x021STX02
Hematology analyzer information
Identifier“EXP”4E X P CR45 58 50 0D
Send data bytes“00512”60 0 5 1 2 CR30 30 35 31 32 0D
Type“MEK-7222”11M E K - 7 2 2 2CR4D 45 4B 2D 37 32 32 32 20 20 0D
Unit no“1”31 CR20 31 0D
Work list data
Name“DAVID”27D A V I D...…CR44 41 56 49 44 … … 20 20 20 20 … … 0D
Sex“MALE”7M A L ECR4D 41 4C 45 20 20 0D
Date of birth“19800219”111 9 8 0 CR 0 2 CR 1 9 CR31 39 38 30 0D 30 32 0D 31 39 0D
Age“22”42 2 CR20 32 32 0D
Department“INTERNAL”14I N T E R N A LCR49 4E 54 45 52 4E 41 4C 20 20 20 20 20 0D
Physician“WATSON”27W A T S O N......CR57 57 41 54 53 4F … … 20 20 20 … … 0D
Operator name“”9CR20 20 20 20 20 20 20 20 0D
Comments“No problem.”129N o p r o b l e m . ...4E 4E 6F 20 70 72 6F 62 6C 65 6D
Normal range table no“0”20 CR30 0D
Work list flag“1”21 CR31 0D
Control mode flag“”2CR20 0D
Reserve data for work list data32SP × 31 Byte + CR × 1 Byte20 × 31 Byte + 0D × 1 Byte
Normal range setting data
WBC-LOW“4.0”54 . 0 CR20 34 2E 30 0D
WBC-HIGH“9.0”59 . 0 CR20 39 2E 30 0D
NE%-LOW“42.0”54 2 . 0 CR34 32 2E 30 0D
NE%-HIGH“85.0”58 5 . 0 CR38 35 2E 30 0D
LY%-LOW“11.0”51 1 . 0 CR31 31 2E 30 0D
LY%-HIGH“49.0”54 9 . 0 CR34 39 2E 30 0D
MO%-LOW“0.0”50 . 0 CR20 30 2E 30 0D
MO%-HIGH“9.0”59 . 0 CR20 39 2E 30 0D
EO%-LOW“0.0”50 . 0 CR20 30 2E 30 0D
EO%-HIGH“3.0”53 . 0 CR20 33 2E 30 0D
BA%-LOW“0.0”50 . 0 CR20 30 2E 30 0D
BA%-HIGH“2.0”52 . 0 CR20 32 2E 30 0D
NE-LOW“1.7”51 . 7 CR20 31 2E 37 0D
NE-HIGH“7.7”57 . 7 CR20 37 2E 37 0D
LY-LOW“0.4”50 . 4 CR20 30 2E 34 0D
LY-HIGH“4.4”54 . 4 CR20 34 2E 34 0D
MO-LOW“0.0”50 . 0 CR20 30 2E 30 0D
MO-HIGH“0.8”50 . 8 CR20 30 2E 38 0D
EO-LOW“0.0”50 . 0 CR20 30 2E 30 0D
EO-HIGH“0.3”50 . 3 CR20 30 2E 33 0D
BA-LOW“0.0”50 . 0 CR20 30 2E 30 0D
BA-HIGH“0.2”50 . 2 CR20 30 2E 32 0D
RBC-LOW“3.80”53 . 8 0 CR33 2E 38 30 0D
RBC-HIGH“5.30”55 . 3 0 CR35 2E 33 30 0D
HGB-LOW“11.0”51 1 . 0 CR31 31 2E 30 0D
HGB-HIGH“17.0”51 7 . 0 CR31 37 2E 30 0D
HCT-LOW“36.0”53 6 . 0 CR33 36 2E 30 0D
HCT-HIGH“56.0”55 6 . 0 CR35 36 2E 30 0D
MCV-LOW“80.0”58 0 . 0 CR38 30 2E 30 0D
MCV-HIGH“100”51 0 0 CR20 31 30 30 0D
MCH-LOW“28.0”52 8 . 0 CR32 38 2E 30 0D
MCH-HIGH“36.0”53 6 . 0 CR33 36 2E 30 0D
MCHC-LOW“31.0”53 1 . 0 CR33 31 2E 30 0D
MCHC-HIGH“37.0”53 7 . 0 CR33 37 2E 30 0D
RDW-LOW“11.5”51 1 . 5 CR31 31 2E 35 0D
RDW-HIGH“16.5”51 6 . 5 CR31 36 2E 35 0D
PLT-LOW“120”51 2 0 CR20 31 32 30 0D
PLT-HIGH“380”53 8 0 CR20 33 38 30 0D
PCT-LOW“0.10”50 . 1 0 CR30 2E 31 30 0D
PCT-HIGH“1.00”51 . 0 0 CR31 2E 30 30 0D
MPV-LOW“5.0”55 . 0 CR20 35 2E 30 0D
MPV-HIGH“10.0”51 0 . 0 CR31 30 2E 30 0D
PDW-LOW“12.0”51 2 . 0 CR31 32 2E 30 0D
PDW-HIGH“18.0”51 8 . 0 CR31 38 2E 30 0D
<End> Extended data block
End of Text0x031ETX03
No. of
Transferred Data in CharactersTransferred Data in Hexadecimal
Byte
... CR … … 20 20 20 20 20 20 … … 0D
4
• “0x0D” is at the end of each item as a separating character but there is a comma “,” at the end of primary type, primary
format version, RDW-SD, IG and IG%.
• Each item is entered by ASCII code. Only “Start of Text” and “ End of Text” are entered by control code.
• Each item is initialized at space (0x20). Unused items are also initialized at space and have “0x0D” in the end.
Operator’s Manual MEK-7300K 4.9
Page 89
4. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
Transfer Format PC (V05-01)
Transfer Example to a PC
Common data block “_” in the Transferred Data in Characters column indicates a blank (0x20).
No. of
Transfer ItemsExample
<Start> Common data block
Start of Text0x02
Hematology analyzer information
Type“MEK-7300”
Parameter no“23”
Send data bytes“01024”
Sampling mode“MANUAL”
Parameter“CBC + Diff”
Sample code“01”
Extended data block “_” in the Transferred Data in Characters column indicates a blank (0x20).
Transfer ItemsExample
<Start> Extended data block
Start of Text0x021STX02
Hematology analyzer information
Identifier“EXI”4E X I CR45 58 50 0D
Send data bytes“00512”60 0 5 1 2 CR30 30 35 31 32 0D
Type“MEK-7300”11M E K - 7 3 0 0CR4D 45 4B 2D 37 32 32 32 20 20 0D
Unit no“0”30 CR20 31 0D
Work list data
Name“DAVID”27D A V I D...…CR44 41 56 49 44 … … 20 20 20 20 … … 0D
Sex“MALE”7M A L ECR4D 41 4C 45 20 20 0D
Date of birth“19800219”111 9 8 0 CR 0 2 CR 1 9 CR31 39 38 30 0D 30 32 0D 31 39 0D
Age“22”42 2 CR20 32 32 0D
Department“INTERNAL”14I N T E R N A LCR49 4E 54 45 52 4E 41 4C 20 20 20 20 20 0D
Physician“WATSON”27W A T S O N......CR57 57 41 54 53 4F … … 20 20 20 … … 0D
Operator name“”9CR20 20 20 20 20 20 20 20 0D
Comments“No problem.”129N o p r o b l e m . ...4E 4E 6F 20 70 72 6F 62 6C 65 6D
Normal range table no“0”20 CR30 0D
Work list flag“1”21 CR31 0D
Control mode flag“”2CR20 0D
Reserve data for work list data32SP × 31 Byte + CR × 1 Byte20 × 31 Byte + 0D × 1 Byte
Normal range setting data
WBC-LOW“4.0”54 . 0 CR20 34 2E 30 0D
WBC-HIGH“9.0”59 . 0 CR20 39 2E 30 0D
NE%-LOW“42.0”54 2 . 0 CR34 32 2E 30 0D
NE%-HIGH“85.0”58 5 . 0 CR38 35 2E 30 0D
LY%-LOW“11.0”51 1 . 0 CR31 31 2E 30 0D
LY%-HIGH“49.0”54 9 . 0 CR34 39 2E 30 0D
MO%-LOW“0.0”50 . 0 CR20 30 2E 30 0D
MO%-HIGH“9.0”59 . 0 CR20 39 2E 30 0D
EO%-LOW“0.0”50 . 0 CR20 30 2E 30 0D
EO%-HIGH“3.0”53 . 0 CR20 33 2E 30 0D
BA%-LOW“0.0”50 . 0 CR20 30 2E 30 0D
BA%-HIGH“2.0”52 . 0 CR20 32 2E 30 0D
NE-LOW“1.7”51 . 7 CR20 31 2E 37 0D
NE-HIGH“7.7”57 . 7 CR20 37 2E 37 0D
LY-LOW“0.4”50 . 4 CR20 30 2E 34 0D
LY-HIGH“4.4”54 . 4 CR20 34 2E 34 0D
MO-LOW“0.0”50 . 0 CR20 30 2E 30 0D
MO-HIGH“0.8”50 . 8 CR20 30 2E 38 0D
EO-LOW“0.0”50 . 0 CR20 30 2E 30 0D
EO-HIGH“0.3”50 . 3 CR20 30 2E 33 0D
BA-LOW“0.0”50 . 0 CR20 30 2E 30 0D
BA-HIGH“0.2”50 . 2 CR20 30 2E 32 0D
RBC-LOW“3.80”53 . 8 0 CR33 2E 38 30 0D
RBC-HIGH“5.30”55 . 3 0 CR35 2E 33 30 0D
HGB-LOW“11.0”51 1 . 0 CR31 31 2E 30 0D
HGB-HIGH“17.0”51 7 . 0 CR31 37 2E 30 0D
HCT-LOW“36.0”53 6 . 0 CR33 36 2E 30 0D
HCT-HIGH“56.0”55 6 . 0 CR35 36 2E 30 0D
MCV-LOW“80.0”58 0 . 0 CR38 30 2E 30 0D
MCV-HIGH“100”51 0 0 CR20 31 30 30 0D
MCH-LOW“28.0”52 8 . 0 CR32 38 2E 30 0D
MCH-HIGH“36.0”53 6 . 0 CR33 36 2E 30 0D
MCHC-LOW“31.0”53 1 . 0 CR33 31 2E 30 0D
MCHC-HIGH“37.0”53 7 . 0 CR33 37 2E 30 0D
RDW-CV-LOW“11.5”51 1 . 5 CR31 31 2E 35 0D
RDW-CV-HIGH“16.5”51 6 . 5 CR31 36 2E 35 0D
PLT-LOW“120”51 2 0 CR20 31 32 30 0D
PLT-HIGH“380”53 8 0 CR20 33 38 30 0D
PCT-LOW“0.10”50 . 1 0 CR30 2E 31 30 0D
PCT-HIGH“1.00”51 . 0 0 CR31 2E 30 30 0D
MPV-LOW“5.0”55 . 0 CR20 35 2E 30 0D
MPV-HIGH“10.0”51 0 . 0 CR31 30 2E 30 0D
PDW-LOW“12.0”51 2 . 0 CR31 32 2E 30 0D
PDW-HIGH“18.0”51 8 . 0 CR31 38 2E 30 0D
<End> Extended data block
End of Text0x031ETX03
No. of
Transferred Data in CharactersTransferred Data in Hexadecimal
Byte
... CR … … 20 20 20 20 20 20 … … 0D
4
Operator’s Manual MEK-7300K 4.11
Page 91
4. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
Extended data block 2
Transfer ItemsExample
MEK-7300 extended data
<Start> Extended data block
Start of Text0x021STX02
Hematology analyzer information
Identifier“EX2”4E X 2 CR45 58 32 0D
Send data bytes“00512”60 0 5 1 2 CR30 30 35 31 32 0D
Type“MEK-7300”11M E K - 7 3 0 0CR4D 45 4B 2D 37 33 30 30 20 20 0D
Serial no“1234567”81 2 3 4 5 6 7 CR31 32 33 34 35 36 37 0D
Reserve data49 × 48 Byte + CR × 1 Byte20 × 48 Byte + 0D × 1 Byte
Measurement data
• “0x0D” is at the end of each item as separating character.
• Each item is entered by ASCII code. Only “Start of Text” and “ End of Text” are entered by control code.
• Each item is initialized at space (0x20). Unused items are also initialized at space and have “0x0D” in the end.
• For the communication, consult your Nihon Kohden representative.
4.12 Operator’s Manual MEK-7300K
Page 92
ASTM Communication
The communication specications of the analyzer are based on the OSI
Reference Model (ISO7498) and comply with the following ASTM standards.
Layer OSI Reference ModelApplicable Standard
For connection between the analyzer and a receiver, use the serial port and an
RS-232C cable. In addition, set the <External device> setting on the Serial Port
Setting to “ASTM”. For the setting, refer to “Changing Serial Port Settings” in
Section 5.
For details on the ASTM communication specications, contact your Nihon
Kohden representative.
4. PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
6Presentation layer
2Data link layerASTM E1381-02
1Physical layerASTM E1381-02
Flowchart of Screens ................................................................................................................................ 5.4
Work List Screen .......................................................................................................................................5.6
Data Screen ..............................................................................................................................................5.7
Quality Control Screen .............................................................................................................................. 5.8
System Screen ..........................................................................................................................................5.9
General .................................................................................................................................................... 5.10
System Screen ........................................................................................................................................5.10
Assigning Users and Passwords ...................................................................................................5.10
General .................................................................................................................................................... 5.78
Open and Closed Sampling Modes ..............................................................................................5.78
Auto Recount ......................................................................................................................................... 5.100
Recount on Alarm .......................................................................................................................5.100
Recount on PLT Count ................................................................................................................5.100
Work Lists ........................................................................................................................................................ 5.101
General .................................................................................................................................................. 5.101
Entering New Work List Data ................................................................................................................ 5.101
Receiving Work List Data ......................................................................................................................5.104
Editing Work List Data ........................................................................................................................... 5.105
Searching for Work List Data ................................................................................................................. 5.106
Deleting Work List Data .........................................................................................................................5.108
To Delete One Data .....................................................................................................................5.108
To Delete Several Consecutive Data ........................................................................................... 5.108
Measuring a Sample Using the Work List ............................................................................................. 5.110
Description of the Results Screen ................................................................................................................... 5.111
Printing and Sending Results ................................................................................................................5.114
Auto Printing/Sending after Measurement .................................................................................. 5.114
Printing by Pressing [Print] Key ................................................................................................... 5.114
Sending Data to PC ....................................................................................................................5.114
Handling Data .................................................................................................................................................. 5.115
General .................................................................................................................................................. 5.115
Displaying Saved Data ..........................................................................................................................5.115
Editing the ID of Saved Data .................................................................................................................5.117
Searching for Saved Data ..................................................................................................................... 5.118
Printing and Sending Saved Data ......................................................................................................... 5.124
Automatic Printing and Sending Data after Measurement ..........................................................5.124
Printing and Sending Stored Data............................................................................................... 5.124
RS-232C Data Transfer .........................................................................................................................5.128
Deleting One Data ....................................................................................................................... 5.129
Deleting Multiple Data .................................................................................................................5.130
Deleting All Stored Data .............................................................................................................. 5.131
Turning the Power Off ...................................................................................................................................... 5.132
Draining the Measurement Baths and Sub Baths ...........................................................................................5.136
5
Operator’s Manual MEK-7300K 5.3
Page 96
5. OPERATING INSTRUCTIONS
Ready
Screen Configuration
Flowchart of Screens
Power ON
Checking
Ready
Work list
Memory
Search
Data
Memory
SD Card
Daily
Search
QC (QUALITY CONTROL)
X-R
L&J
Measurement
Search
ID
Date
Sample type
Flag
Measure
History
Graph
Vial mgnt
Measure
History
Graph
Vial mgnt
Results
Output
XB
Data list
Graph
Setting
Calibration
Setting
History
Data list
5.4 Operator’s Manual MEK-7300K
Page 97
Operations
Refill all
Replace dil
Replace lys.
Replace det
Strong clean
Clean flowcell
Clean baths
Remove clog
Drain all
Drain baths
Circuit Check
Background
System
Users
User Maint.
Error log
Adj. Flowcell
Settings
5. OPERATING INSTRUCTIONS
5
Sample type
Normal range
Units
Operation
Flags
QC Settings
Reagent mgmt
Auto Clean
Date & time
Output
Formatting SD card
Sound/LCD
Initialize
Service (This screen is for service personnel when performing maintenance.)
Operator’s Manual MEK-7300K 5.5
Page 98
5. OPERATING INSTRUCTIONS
Ready Screen
The Ready screen is displayed when the analyzer is powered on and after
cleaning and priming.
On the Ready screen, you can immediately start measurement after checking
measuring accuracy and setting various necessary settings. You can set the
sample ID and select open mode or closed mode.
Work List Screen
On this screen, you can manage work lists or start measurement from a work list.
By registering work list, the analyzer can set measurement parameters for each
sample or attach patient information to the measurement data. Up to 50 work
lists can be registered. Refer to “Using Work Lists” in this section.
5.6 Operator’s Manual MEK-7300K
Page 99
Data Screen
5. OPERATING INSTRUCTIONS
On the Data screen, you can display measured results. Histograms of a measured
sample can be displayed by pressing the desired data and pressing the Detail key.
You can delete, print or transfer data to a personal computer. Refer to “Handling
Data” in this section.
5
Data-List screen
Data-Details screen
Operator’s Manual MEK-7300K 5.7
Page 100
5. OPERATING INSTRUCTIONS
Quality Control Screen
Quality control refers to the precision, accuracy and reproducibility in the
system. The analyzer provides four programs for quality control. Refer to Section
11 “Quality Control”.
Calibration Screen
The calibration coefcient is used in converting the data to the measured values
to make the measured values as close as possible to the real values. Calibration
can be performed in venous blood and pre-diluted blood individually. Refer to
Section 6 “Calibration Procedures”.
5.8 Operator’s Manual MEK-7300K
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.