LIFE SCOPE®
G7 BEDSIDE MONITOR SPECIFICATIONS
DISPLAY
Display Size/Type
CSM-1702: 19” color direct bond TFT LCD
Resolution
CSM-1702: 1680 × 1050
Characteristics
True Flat, tempered glass medical certied, capacitive touch screen with up to 20 function soft keys
and 3 quick recall screen congurations
Number of Traces
Up to 17 traces (up to 49 on three displays) moving or xed method
Waveforms
ECG (up to 12), respiration, IBP (up to 8), SpO2 pulse wave, CO2, BIS-EEG*, EEG* (up to 2 traces),
vent PAW, vent Flow, and CO Thermodilution curve.When gas is monitored: O
2
concentration curve,
CO
2
concentration curve, anesthetic agent concentration (Halothane, Isourane, Enurane,
Sevourane, Desurane*) Analog input.
Sweep Speed
Normal sweep speed: 25 mm/s, 50 mm/s
Slow sweep speed (respiration): 1.56 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s
Numeric
Data Display
Heart rate, VPC rate, ST level, RR respiration rate, NIBP (systolic, diastolic, map), IBP (systolic, dia-
stolic, mean), SpO
2
, SpO2-2, delta SpO2, pulse rate, temperature, CO, CI, Ti (injectate temperature),
Tb (blood temperature), O
2
concentration, EtCO2, BIS, inspired/ expired N2O concentration, inspired/
expired CO
2
, inspired/ expired O2 concentration, inspired/ expired anesthetic agent concentration
(Halothane, Isourane, Enurane, Sevourane, Desurane), MAC (minimum alveolar concentration),
Ppeak (peak airway pressure), PEEP (positive end expiratory pressure), Pmean (mean airway
pres- sure), MV (minute volume), TVi (inspiratory tidal volume), TVe (expiratory tidal volume),
C (compliance), R (airway resistance), Ri (inspiratory airway resistance), Re (expiratory airway
resistance), I:E (inspiration expiration ratio), SEF (90 or 95% spectral edge frequency), MDF (median
frequency), power ratio of frequency, PPF (peak power frequency), TP (total power), power of
frequency, CSA, DSA, TOF cnt, TOF rat, Tw 1 to Tw4 (Twitch height), DBS (double burst stimulation),
TET (tetanic stimulation) PTC (post tetanic count stimulation), PCCO, PCCI, tcPO
2
, tcPCO2, PPV, SPV.
With INVOS monitor: rSO
2
, SSI indicator, BL (baseline), area under the curve (AUC), change rate.
With CCO monitor: CCO, CCI, SvO
2
, ScvO2, SV, SVI, SVV, SVR,SVRI, RVEF, EDV, EDVI, ESV, ESVI,
DO
2
, VO2, O2EI, SaO22 HRV, CF, QTc and QRSd
ALARMS
Alarm Items
Vital sign, arrhythmia, technical, operational, interbed
Alarm Levels
Crisis: red blinking;
Warning: yellow blinking;
Advisory: yellow or blue light
Alarm Indication
Alarm indicator (360° visibility) highlighted message, alarm sound
Alarm Suspend
1, 2, or 3 min
Alarm Master
Adult and Pediatric up to 4, Neonatal up to 6
STORED PATIENT DATA
Trend graph
Trend parameters: up to 9 for each trend graph (up to three);
Trend display time: Up to 72 hours (short trend for the last 30 minutes on main screen)
Vital Signs List
Three lists of up to 15 parameters each for up to 72 hours; Periodic: up to 1 per minute for 72 hours
NIBP
Number of entries: 1,024 les
HEMO List
Number of entries: 1,024 les
Full Disclosure
Storage time: up to 72 hours;
Number of waveforms stored: 5 (max)
ST Recall
Number of les: 4,320 les (1 per minute for 72 hours) for all monitoring leads
History
16,384 les (Alarm & Arrythmia recall)
12-Lead
Interpretive Recall
Number of les: 18 les
Storage Capacity
72 Hours (OCRG/Hemodynamics/Trend/aEEG)
RECORDER (OPTIONAL)
Recording Method
Thermal array recording
Number of Channels
3 traces (max)
PARAMETERS
Leads
3,6 or 10-lead ECG cable for I, II, III, aVR, aVL, aVF, V1 to V6
ECG
Number of ECG waveforms channels: up to 12; Frequency response: diagnosis mode - 0.05 to
150 Hz, ST mode - 0.05 to 18 Hz, monitor mode - 0.3 to 40 Hz, maximum lter mode - 1 to 18 Hz;
Heart Rate Counting range: 0, 15 to 300 beats/min; Arrhythmia analysis method: multitemplate
software algorithm; VPC counting rate: 0 to 99 VPCs/min; Arrhythmia alarms: ASYSTOLE, VF, VT,
V RHYTHM, V BRADY, EXT TACHY, EXT BRADY, A-Fib, End A-Fib, VPC RUN, COUPLET, EARLY
VPC, BIGEMINY, TRIGEMINY, FREQ VPC, PROLONGED RR, SV TACHY, TACHYCARDIA, BRADY-
CARDIA, VPC, MULTIFORM, IRREGULAR RR, NO PACER PULSE, PACER NON-CAPTURE, PAUSE
ST Level Measurement
Number of measurement channels: Up to 12; Measuring range: ±2.5 mV
Respiration
Measuring range: 0 to 150 breaths/min (Impedance)
SpO
2
Measuring Technology: Nihon Kohden, Masimo or Nellcor; Measuring Display Range: 0 to
100% (70 to 100% at specied accuracy); Pulse rate from SpO
2
Range: 20 to 300 beats/min
(varies by SpO
2
technology)
Non-invasive Blood
Pressure, NIBP
Measuring method: Oscillometric Cuff;
Pressure display range: 0 to 300 mmHg
Invasive Blood
Pressure, IBP
Measuring range: –50 to 300 mmHg;
Pulse rate display range from IBP range: 0, 30 to 300 beats/min
Temperature
Measuring range: 0 to 45°C; Number of channels: 4 (max)
Cardiac Output
Measuring method: Thermodilution method; Measuring range: Injectate temperature (Ti): 0°C
to 27°C; Blood temperature (Tb): 15°C to 45°C; Thermodilution curve (delta Tb): 0°C to 2.5°C;
Cardiac output (CO): 0.5 to 20 L/min
CO
2
CO2 measuring range: 0 to 150 mmHg; Respiration rate counting range: 3 to 150 breaths/min
BIS
Input channels: 2; Measuring parameters: Bispectral Index (BIS), 95% Spectral Edge
Frequency (SEF90, SEF95), Suppression Ratio (SR), EMG, Signal Quality Index (SQI)
CONNECTIVIT Y
Standard: Ethernet (LS-Net), USB, third party interface (6), HDMI remote video out, recorder and RS-232 Serial out. Optional third
party interface (4), Nurse call, independent interactive remote display port, DVI independent remote video out, Ethernet (HIS),
ground terminal and AC power
OPERATING ENVIRONMENT
Temperature
41 to 104°F (5 to 40°C)
Humidity
30 to 85% RH (non-condensing)
Atmospheric Pressure
700 to 1060 hPa
Degree of Protection
Against harmful ingress of water: IPX1
POWER REQUIREMENT
AC
100 to 240 V ±10%
DC (SB-950P)
10.8 V
Line Frequency
50 or 60 Hz
Battery Emergency
Backup Time
3 minutes
Power Input
AC 150, Battery 130 VA
Noise
<48 dBA
DIMENSIONS & WEIGHT
Dimensions
CSM-1702: 18.7” W × 12.8” H × 4.5” D (475 W × 326 H × 115 D mm)
JA-170P: 7.1” W × 7.1” H × 5.3” D (182 W × 182 H × 136 D mm), (excl. cable)
BSM-1700: 5.8” W × 7.6” H × 3.7” D (147 W × 194 H × 94 D mm)
WS-171P recorder unit: 3.9” × 3.2” × 5.6” (99 W × 82 H × 142 D mm), (option)
AA-174P multi amp unit: 6.1” × 2.4” × 7.5” (156 W × 63 H × 190 D mm), (option, excluding cable)
Weight
CSM-1702: 18.3 lbs (8.3 kg); JA-170P: 2.4 lbs (1.1kg), (excluding cable) BSM-1700: 3.5 lbs
(1.57 kg without battery pack); WS-171P recorder unit: 1.4 lbs (0.62 kg) (option);
AA-174P multi amp unit: 1.8 lbs (0.82 kg) (option)
* With optional modules
WORKS
I N HAR MON Y,
SO YOU CAN TOO
G5
G7
Imagine one continuous patient record, from admit to discharge. Available to
everyone on the caregiving team, seamlessly integrating every device and decision
across the spectrum of care.
Created by clinicians, for clinicians—Nihon Kohden Patient monitoring gives you
more data, more insights, and more eciency. Combining fully optimized, adaptable
monitors with an enterprise-level server and software applications customized to
your needs. All in a consistent user interface that’s easy to use and streamlines
workow. These advanced tools were designed to help accomplish one critical goal:
to allow clinicians to spend more time with their patients and less with technology.
G5 MAX

LIFE SCOPE®
G5 BEDSIDE MONITOR SPECIFICATIONS
LIFE SCOPE®
G7 BEDSIDE MONITOR SPECIFICATIONS
DISPLAY
Display Size/Type
Resolution
Characteristics
Number of Traces
Waveforms
Sweep Speed
Numeric
Data Display
CSM-1501 (CU-151R): 12.1” color TFT LCD; CSM-1502 (CU-152R): 15.6” color direct bond TFT LCD
CSM-1501: 1280 × 800; CSM-1502: 1366 × 768
True Flat, tempered glass medical certied, capacitive touch screen (G5 Max) with up to 15 function soft keys and 3 quick recall screen congurations (CSM-1501; resistive lm touch screen)
Up to 15 traces (30 traces on two displays) moving or xed method
Up to 15 traces (30 traces on two displays) moving or xed method ECG (up to 12), respiration, IBP (up to 8), SpO2
pulse wave, CO2, BIS-EEG, EEG* (up to 2 traces), vent PAW, vent Flow, and CO Thermodilution curve. When gas is
monitored: O2 concentration curve, CO2 concentration curve, anesthetic agent concentration (Halothane, Isourane,
Enurane, Sevourane, Desurane*) Analog input.
Normal sweep speed: 25 mm/s, 50 mm/s
Slow sweep speed (respiration):
Heart rate, VPC rate, ST level, RR respiration rate, NIBP (systolic, diastolic, map), IBP (systolic, diastolic,
mean), SpO
temperature), O
inspired/expired O
Enurane, Sevourane, Desurane), MAC (minimum alveolar concentration), Ppeak (peak airway pres-
sure), PEEP (positive end expiratory pressure), Pmean (mean airway pressure), MV (minute volume),
TVi (inspiratory tidal volume), TVe (expiratory tidal volume), C (compliance), R (airway resistance), Ri
(inspiratory airway resistance), Re (expiratory airway resistance), I:E (inspiration expiration ratio), SEF (90
or 95% spectral edge frequency), MDF (median frequency), PPF (peak power frequency), TP (total power),
power of frequency, CSA, power ratio of frequency, DSA, TOF cnt, TOF rat, Tw 1 to Tw4 (Twitch height),
DBS (double burst stimulation), TET (tetanic stimulation) PTC (post tetanic count stimulation), PCCO, PCCI,
, tcPCO2, PPV, SPV. With INVOS monitor: rSO2, SSI indicator, BL (baseline), AUC (area under the
tcPO
2
curve), change rate. With CCO monitor: CCO, CCI, SvO
EDVI, ESV, ESVI, DO
ALARMS
Alarm Items
Alarm Levels
Alarm Indication
Alarm Suspend
Alarm Master
Vital sign, arrhythmia, technical, operational, interbed
Crisis: red blinking
Warning: yellow blinking
Advisory: yellow or blue light
Alarm indicator (360° visibility) highlighted message, alarm sound
1, 2, or 3 min
Adult and Pediatric up to 4, Neonatal up to 6
STORED PATIENT DATA
Trendgraph
Vital Signs List
NIBP
HEMO List
Full Disclosure
ST Recall
History
12-Lead
Interpretive Recall
Storage Capacity
Trend parameters: up to 9 for each trend graph (up to three); Trend display time: Up to 72 hours
(short trend for the last 30 minutes on main screen)
Three lists of up to 15 parameters each for up to 72 hours; Periodic: up to 1 per minute for 72 hours
Number of entries: 1,024 les
Number of entries: 1,024 les
Storage time: up to 72 hours;
Number of waveforms stored: 5 (max)
Number of les: 4,320 les (1 per minute for 72 hours) for all monitoring leads
16,384 les (Alarm & Arrythmia recall)
Number of les: 18 les
72 hours (OCRG/Hemodynamics/Trend/aEEG)
RECORDER (OPTIONAL)
Recording Method
Number of Channels
* With optional modules
Thermal array recording
3 traces (max)
1.56 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s
, SpO2-2, delta SpO2, pulse rate, temperature, CO, CI, Ti (injectate tempera- ture), Tb (blood
2
concentration, EtCO2, BIS, inspired/ expired N2O concentration, inspired/expired CO2,
2
concentration, inspired/expired anesthetic agent concentration (Halothane, Isourane,
2
, ScvO2, SV, SVI, SVV, SVR, SVRI, RVEF, EDV,
, VO2, O2EI, SaO2, HRV, CF, QTc and QRSd
2
2
PARAMETERS
Leads
ECG
ST Leve l
Measurement
Respiration
SpO
2
Non-invasive Blood
Pressure, NIBP
Invasive Blood
Pressure, IBP
Temperature
Cardiac Output
CO
2
BIS
3,6 or 10-lead ECG cable for I, II, III, aVR, aVL, aVF, V1 to V6
Number of ECG waveforms channels: up to 12; Frequency response: diagnosis mode - 0.05
to 150 Hz, ST mode -0.05 to 18 Hz, monitor mode -0.3 to 40 Hz, maximum lter mode -1 to 18 Hz;
Heart Rate Counting range: 0, 15 to 300 beats/min; Arrhythmia analysis method: multi-template
software algorithm; VPC counting rate: 0 to 99 VPCs/min; Arrhythmia alarms: ASYSTOLE, VF, VT,
V RHYTHM, V BRADY, EXT TACHY, EXT BRADY, A-Fib, End A-Fib, VPC RUN, COUPLET, EARLY
VPC, BIGEMINY, TRIGEMINY, FREQ VPC, PROLONGED RR, SV TACHY, TACHYCARDIA, BRADYCARDIA, VPC, MULTIFORM, IRREGULAR RR, NO PACER PULSE, PACER NON-CAPTURE, PAUSE
Number of measurement channels: Up to 12;
Measuring range: ±2.5 mV
Measuring range: 0 to 150 breaths/min (Impedance)
Measuring Technology: Nihon Kohden, Masimo or Nellcor; Measuring Display Range: 0 to 100%
(70 to 100% at specied accuracy); Pulse rate from SpO
technology)
Measuring method: Oscillometric Cuff;
Pressure display range: 0 to 300 mmHg
Measuring range: –50 to 300 mmHg;
Pulse rate display range from IBP range: 0, 30 to 300 beats/min
Measuring range: 0 to 45°C; Number of channels: 4 (max)
Measuring method: Thermodilution method; Measuring range: Injectate temperature (Ti): 0°C
to 27°C; Blood temperature (Tb): 15°C to 45°C; Thermodilution curve (delta Tb): 0°C to 2.5°C;
Cardiac output (CO): 0.5 to 20 L/min
CO2 measuring range: 0 to 150 mmHg; Respiration rate counting range: 3 to 150 breaths/min
Input channels: 2; Measuring parameters: Bispectral Index (BIS), 95% Spectral Edge Frequency
(SEF90, SEF95), Suppression Ratio (SR), EMG, Signal Quality Index (SQI)
Range: 20 to 300, (varies by SpO2
2
CONNECTIVIT Y
Standard: Ethernet (LS-Net), USB, third party interface (2), HDMI remote video out, recorder and RS-232 Serial out. Optional
third party interface (4), Nurse call, independent interactive remote display port (G5 Max), DVI independent remote video out
(G5 Max), Ethernet (HIS), ground terminal and AC power
OPERATING ENVIRONMENT
Temperature
Humidity
Atmospheric Pressure
Degree of Protection
41 to 104°F (5 to 40°C)
30 to 85% RH (non-condensing)
700 to 1060 hPa
Against harmful ingress of water: IPX1
POWER REQUIREMENT
AC
DC (SB-950P)
Line Frequency
Battery Operation Time
Power Input
Noise
100 to 240 V ±10%
10.8 V
50 or 60 Hz
CSM-1501: Up to 180 minutes;
CSM-1502: Up to 120 minutes
AC 120, Battery 100 VA
<48 dBA
DIMENSIONS & WEIGHT
CSM-1501: 13.4” W x 11.6” H x 8” D (341 W × 294 H × 204 D mm)
CSM-1502: 15.9” W × 11.9” H × 8” D (403 W × 302 H × 204 D mm)
Dimensions
Weight
BSM-1700: 5.8” W × 7.6” H × 3.7” D (147 W × 194 H × 94 D mm)
WS-151P recorder unit: (built in option)
AA-174P multi amp unit: 6.1” × 2.4” × 7.5” (156 W × 63 H × 190 D mm) (option, excluding cable)
CSM-1501: 13.2 lbs (6 kg) CSM-1502: 15.4 lbs (7 kg); BSM-1700: 3.5 lbs (1.57kg without battery pack);
WS-151P recorder unit: 0.77 lbs (0.35 Kg) (option) AA-174P multi amp unit: 1.8 lbs (0.82 kg) (option)
DISPLAY
Display Size/Type
Resolution
Characteristics
Number of Traces
Waveforms
Sweep Speed
Numeric
Data Display
CSM-1702: 19” color direct bond TFT LCD
CSM-1702: 1680 × 1050
True Flat, tempered glass medical certied, capacitive touch screen with up to 20 function soft keys
and 3 quick recall screen congurations
Up to 17 traces (up to 49 on three displays) moving or xed method
ECG (up to 12), respiration, IBP (up to 8), SpO2 pulse wave, CO2, BIS-EEG*, EEG* (up to 2 traces),
vent PAW, vent Flow, and CO Thermodilution curve.When gas is monitored: O
concentration curve, anesthetic agent concentration (Halothane, Isourane, Enurane,
CO
2
Sevourane, Desurane*) Analog input.
Normal sweep speed: 25 mm/s, 50 mm/s
Slow sweep speed (respiration): 1.56 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s
Heart rate, VPC rate, ST level, RR respiration rate, NIBP (systolic, diastolic, map), IBP (systolic, diastolic, mean), SpO
Tb (blood temperature), O
expired CO
(Halothane, Isourane, Enurane, Sevourane, Desurane), MAC (minimum alveolar concentration),
Ppeak (peak airway pressure), PEEP (positive end expiratory pressure), Pmean (mean airway
pres- sure), MV (minute volume), TVi (inspiratory tidal volume), TVe (expiratory tidal volume),
C (compliance), R (airway resistance), Ri (inspiratory airway resistance), Re (expiratory airway
resistance), I:E (inspiration expiration ratio), SEF (90 or 95% spectral edge frequency), MDF (median
frequency), power ratio of frequency, PPF (peak power frequency), TP (total power), power of
frequency, CSA, DSA, TOF cnt, TOF rat, Tw 1 to Tw4 (Twitch height), DBS (double burst stimulation),
TET (tetanic stimulation) PTC (post tetanic count stimulation), PCCO, PCCI, tcPO
With INVOS monitor: rSO
With CCO monitor: CCO, CCI, SvO
, VO2, O2EI, SaO22 HRV, CF, QTc and QRSd
DO
2
ALARMS
Alarm Items
Alarm Levels
Alarm Indication
Alarm Suspend
Alarm Master
Vital sign, arrhythmia, technical, operational, interbed
Crisis: red blinking;
Warning: yellow blinking;
Advisory: yellow or blue light
Alarm indicator (360° visibility) highlighted message, alarm sound
1, 2, or 3 min
Adult and Pediatric up to 4, Neonatal up to 6
STORED PATIENT DATA
Trend graph
Vital Signs List
NIBP
HEMO List
Full Disclosure
ST Recall
History
12-Lead
Interpretive Recall
Storage Capacity
Trend parameters: up to 9 for each trend graph (up to three);
Trend display time: Up to 72 hours (short trend for the last 30 minutes on main screen)
Three lists of up to 15 parameters each for up to 72 hours; Periodic: up to 1 per minute for 72 hours
Number of entries: 1,024 les
Number of entries: 1,024 les
Storage time: up to 72 hours;
Number of waveforms stored: 5 (max)
Number of les: 4,320 les (1 per minute for 72 hours) for all monitoring leads
16,384 les (Alarm & Arrythmia recall)
Number of les: 18 les
72 Hours (OCRG/Hemodynamics/Trend/aEEG)
RECORDER (OPTIONAL)
Recording Method
Number of Channels
* With optional modules
Thermal array recording
3 traces (max)
concentration curve,
2
, SpO2-2, delta SpO2, pulse rate, temperature, CO, CI, Ti (injectate temperature),
2
concentration, EtCO2, BIS, inspired/ expired N2O concentration, inspired/
, inspired/ expired O2 concentration, inspired/ expired anesthetic agent concentration
2
2
, tcPCO2, PPV, SPV.
, SSI indicator, BL (baseline), area under the curve (AUC), change rate.
2
, ScvO2, SV, SVI, SVV, SVR,SVRI, RVEF, EDV, EDVI, ESV, ESVI,
2
2
PARAMETERS
Leads
ECG
ST Level Measurement
Respiration
SpO
2
Non-invasive Blood
Pressure, NIBP
Invasive Blood
Pressure, IBP
Temperature
Cardiac Output
CO
2
BIS
3,6 or 10-lead ECG cable for I, II, III, aVR, aVL, aVF, V1 to V6
Number of ECG waveforms channels: up to 12; Frequency response: diagnosis mode - 0.05 to
150 Hz, ST mode - 0.05 to 18 Hz, monitor mode - 0.3 to 40 Hz, maximum lter mode - 1 to 18 Hz;
Heart Rate Counting range: 0, 15 to 300 beats/min; Arrhythmia analysis method: multitemplate
software algorithm; VPC counting rate: 0 to 99 VPCs/min; Arrhythmia alarms: ASYSTOLE, VF, VT,
V RHYTHM, V BRADY, EXT TACHY, EXT BRADY, A-Fib, End A-Fib, VPC RUN, COUPLET, EARLY
VPC, BIGEMINY, TRIGEMINY, FREQ VPC, PROLONGED RR, SV TACHY, TACHYCARDIA, BRADYCARDIA, VPC, MULTIFORM, IRREGULAR RR, NO PACER PULSE, PACER NON-CAPTURE, PAUSE
Number of measurement channels: Up to 12; Measuring range: ±2.5 mV
Measuring range: 0 to 150 breaths/min (Impedance)
Measuring Technology: Nihon Kohden, Masimo or Nellcor; Measuring Display Range: 0 to
100% (70 to 100% at specied accuracy); Pulse rate from SpO
(varies by SpO
Measuring method: Oscillometric Cuff;
Pressure display range: 0 to 300 mmHg
Measuring range: –50 to 300 mmHg;
Pulse rate display range from IBP range: 0, 30 to 300 beats/min
Measuring range: 0 to 45°C; Number of channels: 4 (max)
Measuring method: Thermodilution method; Measuring range: Injectate temperature (Ti): 0°C
to 27°C; Blood temperature (Tb): 15°C to 45°C; Thermodilution curve (delta Tb): 0°C to 2.5°C;
Cardiac output (CO): 0.5 to 20 L/min
CO2 measuring range: 0 to 150 mmHg; Respiration rate counting range: 3 to 150 breaths/min
Input channels: 2; Measuring parameters: Bispectral Index (BIS), 95% Spectral Edge
Frequency (SEF90, SEF95), Suppression Ratio (SR), EMG, Signal Quality Index (SQI)
technology)
2
Range: 20 to 300 beats/min
2
CONNECTIVIT Y
Standard: Ethernet (LS-Net), USB, third party interface (6), HDMI remote video out, recorder and RS-232 Serial out. Optional third
party interface (4), Nurse call, independent interactive remote display port, DVI independent remote video out, Ethernet (HIS),
ground terminal and AC power
OPERATING ENVIRONMENT
Temperature
Humidity
Atmospheric Pressure
Degree of Protection
41 to 104°F (5 to 40°C)
30 to 85% RH (non-condensing)
700 to 1060 hPa
Against harmful ingress of water: IPX1
POWER REQUIREMENT
AC
DC (SB-950P)
Line Frequency
Battery Emergency
Backup Time
Power Input
Noise
100 to 240 V ±10%
10.8 V
50 or 60 Hz
3 minutes
AC 150, Battery 130 VA
<48 dBA
DIMENSIONS & WEIGHT
CSM-1702: 18.7” W × 12.8” H × 4.5” D (475 W × 326 H × 115 D mm)
JA-170P: 7.1” W × 7.1” H × 5.3” D (182 W × 182 H × 136 D mm), (excl. cable)
Dimensions
Weight
BSM-1700: 5.8” W × 7.6” H × 3.7” D (147 W × 194 H × 94 D mm)
WS-171P recorder unit: 3.9” × 3.2” × 5.6” (99 W × 82 H × 142 D mm), (option)
AA-174P multi amp unit: 6.1” × 2.4” × 7.5” (156 W × 63 H × 190 D mm), (option, excluding cable)
CSM-1702: 18.3 lbs (8.3 kg); JA-170P: 2.4 lbs (1.1kg), (excluding cable) BSM-1700: 3.5 lbs
(1.57 kg without battery pack); WS-171P recorder unit: 1.4 lbs (0.62 kg) (option);
AA-174P multi amp unit: 1.8 lbs (0.82 kg) (option)