LIFE SCOPE
G5 & G5 MAX BEDSIDE MONITOR
NEXT GENERATION MONITORING SYSTEM, ALL-IN-ONE MONITOR
G5
G5 Max
The Life Scope G5 is a bedside monitor available with 15.6" or 12.1" display with features to support high and mid acuity clinical
settings, where clinicians need an all-in-one system with quick access to multiple configurations and trend analysis.
M O R E DATA I N T EG RAT I O N
01. Offers a complete data record, including multiwaveform, multi-parameter full disclosure
02. Includes 12-lead ECG monitoring with printing capabilities and visualization that may help indicate changes to the heart
03. Combines vital signs patient monitoring and 8-channel neurology monitoring
04. Integrates with third party devices such as ventilators, anesthesia systems, and continuous cardiac output devices
MORE EFFICIENCY
01. Seamless patient monitoring work ow with one admission and discharge across the entire hospital stay
02. Employs intuitive user interface and quick access keys to data and functions
03. Customizable escalation latching and delayed alarm functionality to help reduce alarm fatigue
04. Adaptable system solutions for specialized care needs such as isolation room
MORE INSIGHTFUL
01. Advanced features like drug, pulmonary calculations, and hemodynamic graph to aid clinicians in closer monitoring for better patient outcome
02. Offers advanced EGG analysis with advanced atrial brillation algorithm, enhanced review capability ST analysis, event recall and QTc/QRSd
03. Offers an overview of patient health status, including a comprehensive histogram, advanced interbed, HiQ View, and car seat report
MMLB 343 [B]-CO-3951
LIFE SCOPE®
G5 BEDSIDE MONITOR SPECIFICATIONS
DISPLAY
Display Size/Type
Resolution
Characteristics
Number of Traces
Waveforms
Sweep Speed
Numeric
Data Display
CSM-1501 (CU-151R): 12.1” color TFT LCD
CSM-1502 (CU-152R): 15.6” color direct bond TFT LCD
CSM-1501: 1280 × 800
CSM-1502: 1366 × 768
True Flat, tempered glass medical certied, capacitive
touch screen (G5 Max) with up to 15 function soft keys and 3 quick
recall screen congurations (CSM-1501; resistive lm touch screen)
Up to 15 traces (30 traces on two displays) moving or xed method
Up to 15 traces (30 traces on two displays) moving or xed method ECG (up to 12),
respiration, IBP (up to 8), SpO
PAW, vent Flow, and CO Thermodilution curve. When gas is monitored: O2 concentration
curve, CO2 concentration curve, anesthetic agent concentration (Halothane, Isourane,
Enurane, Sevourane, Desurane*) Analog input.
Normal sweep speed: 25 mm/s, 50 mm/s
Slow sweep speed (respiration):
Heart rate, VPC rate, ST level, RR respiration rate, NIBP(systolic, diastolic,
map), IBP (systolic, diastolic, mean), SpO
temperature, CO, CI, Ti (injectate tempera- ture), Tb (blood temperature),
concentration, EtCO2, BIS, inspired/ expired N2O concentration, inspired/
O
2
expired CO
agent concentration (Halothane, Isourane, Enurane, Sevourane,
Desurane), MAC (minimum alveolar concentration), Ppeak (peak airway
pressure), PEEP (positive end expiratory pressure), Pmean (mean airway
pressure), MV (minute volume), TVi (inspiratory tidal volume), TVe (expiratory
tidal volume), C (compliance), R (airway resistance), Ri (inspiratory airway
resistance), Re (expiratory airway resistance), I:E (inspiration expiration ratio),
SEF (90 or 95% spectral edge frequency), MDF (median frequency), PPF (peak
power frequency), TP (total power), power of frequency, CSA, power ratio of
frequency, DSA,TOF cnt, TOF rat, Tw 1 to Tw4 (Twitch height), DBS (double
burst stimulation), TET (tetanic stimulation) PTC (post tetanic count stimulation),
PCCO, PCCI, tcPO
indicator, BL (baseline), AUC (area under the curve), change rate. With CCO
monitor: CCO, CCI, SvO
ESV, ESVI, DO2, VO
ALARMS
Alarm Items
Alarm Levels
Alarm Indication
Alarm Suspend
Alarm Master
Vital sign, arrhythmia, technical, operational, interbed
Crisis: red blinking
Warning: yellow blinking
Advisory: yellow or blue light
Alarm indicator (360° visibility) highlighted message, alarm sound
1, 2, or 3 min
Adult and Pediatric up to 4, Neonatal up to 6
STORED PATIENT DATA
Trend parameters: up to 9 for each trend graph (up to three);
Trend graph
Vital Signs List
NIBP
HEMO List
Full Disclosure
ST Recall
History
12-Lead
Interpretive Recall
Storage Capacity
Trend display time: Up to 72 hours (short trend for the last 30 minutes
on main screen)
Three lists of up to 15 parameters each for up to 72 hours; Periodic: up to
1 per minute for 72 hours
Number of entries: 1,024 les
Number of entries: 1,024 les
Storage time: up to 72 hours;
Number of waveforms stored: 5 (max)
Number of les: 4,320 les (1 per minute for 72 hours) for all monitoring leads
16,384 les (Alarm & Arrythmia recall)
Number of les: 18 les
72 Hours (OCRG/Hemodynamics/Trend/aEEG)
RECORDER (OPTIONAL)
Recording Method
Number of Channels
* With optional modules
Thermal array recording
3 traces (max)
PARAMETERS
3,6 or 10-lead ECG cable for I, II, III, aVR, aVL, aVF, V1 to V6
Number of ECG waveforms channels: up to 12; Frequency response:
diagnosis mode - 0.05 to 150 Hz, ST mode -0.05 to 18 Hz, monitor mode
-0.3 to 40 Hz, maximum lter mode -1 to 18 Hz; Heart Rate Counting
range: 0, 15 to 300 beats/min; Arrhythmia analysis method: multi-template
software algorithm; VPC counting rate: 0 to 99 VPCs/min; Arrhythmia
alarms: ASYSTOLE, VF, VT, V RHYTHM, V BRADY, EXT TACHY, EXT
BRADY, A-Fib, End A-Fib, VPC RUN, COUPLET, EARLY VPC, BIGEMINY,
TRIGEMINY, FREQ VPC, PROLONGED RR, SV TACHY, TACHYCARDIA,
BRADYCARDIA, VPC, MULTIFORM, IRREGULAR RR, NO PACER PULSE,
PACER NON-CAPTURE, PAUSE
Number of measurement channels: Up to 12;
Measuring range: ±2.5 mV
Measuring range: 0 to 150 breaths/min (Impedance)
Measuring Technology: Nihon Kohden, Masimo or Nellcor;
Measuring Display Range: 0 to 100% (70 to 100% at specied accuracy);
Pulse rate from SpO
Measuring method: Oscillometric Cuff;
Pressure display range: 0 to 300 mmHg
Measuring range: –50 to 300 mmHg;
Pulse rate display range from IBP range: 0, 30 to 300 beats/min
Measuring range: 0 to 45°C; Number of channels: 4 (max)
Measuring method: Thermodilution method;
Measuring range: Injectate temperature (Ti): 0°C to 27°C;
Blood temperature (Tb): 15°C to 45°C; Thermodilution curve (delta Tb):
0°C to 2.5°C; Cardiac output (CO): 0.5 to 20 L/min
CO2 measuring range: 0 to 150 mmHg;
Respiration rate counting range: 3 to 150 breaths/min
Input channels: 2; Measuring parameters: Bispectral Index (BIS), 95%
Spectral Edge Frequency (SEF90, SEF95), Suppression Ratio (SR), EMG,
Signal Quality Index (SQI)
pulse wave, CO2, BIS-EEG, EEG* (up to 2 traces), vent
2
1.56 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s
, SpO2-2, delta SpO2, pulse rate,
2
, inspired/expired O2 concentration, inspired/expired anesthetic
2
, tcPCO2, PPV, SPV. With INVOS monitor: rSO2, SSI
2
, ScvO2, SV, SVI, SVV, SVR, SVRI, RVEF, EDV, EDVI,
2
, O2EI, SaO2, HRV, CF, QTc and QRSd.
2
Leads
ECG
ST Leve l
Measurement
Respiration
SpO
2
Non-invasive Blood
Pressure, NIBP
Invasive Blood
Pressure, IBP
Temperature
Cardiac Output
CO
2
BIS
CONNECTIVIT Y
Standard: Ethernet (LS-Net), USB, third party interface (2), HDMI remote video out, recorder and
RS-232 Serial out. Optional third party interface (4), Nurse call, independent interactive remote
display port (G5 Max), DVI independent remote video out (G5 Max), Ethernet (HIS), ground terminal
and AC power
OPERATING ENVIRONMENT
Temperature
Humidity
Atmospheric
Press ure
Degree of
Protection
41 to 104°F (5 to 40°C)
30 to 85% RH (non-condensing)
700 to 1060 hPa
Against harmful ingress of water: IPX1
POWER REQUIREMENT
AC
DC (SB-950P)
Line Frequency
Battery Operation
Time
Power In put
Noise
100 to 240 V ±10%
10.8 V
50 or 60 Hz
CSM-1501: Up to 180 minutes
CSM-1502: Up to 120 minutes
AC 120, Battery 100 VA
<48 dBA
DIMENSIONS & WEIGHT
CSM-1501: 13.4” W x 11.6” H x 8” D (341 W × 294 H × 204 D mm)
CSM-1502: 15.9” W × 11.9” H × 8” D (403 W × 302 H × 204 D mm)
Dimensions
Weight
FOR MORE INFORMATION, PLEASE CONTACT US AT 1-800-325-0283 OR VISIT US.NIHONKOHDEN.COM
Life Scope is a registered trademark of Nihon Kohden Corporation.
MMLB 343 [B]-CO-3951
BSM-1700: 5.8” W × 7.6” H × 3.7” D (147 W ×194 H × 94 D mm)
WS-151P recorder unit: (built in option) AA-174P multi amp unit: 6.1” ×
2.4” × 7.5” (156 W × 63 H × 190 D mm) (option, excluding cable)
CSM-1501: 13.2 lbs (6 kg) CSM-1502: 15.4 lbs (7 kg) BSM-1700: 3.5 lbs (1.57
kg without battery pack) WS-151P recorder unit: 0.77 lbs (0.35 Kg) (option)
AA-174P multi amp unit: 1.8 lbs (0.82 kg) (option)
Range: 20 to 300 (varies by SpO2 technology)
2