Nihon Kohden BSM-6000 Reference Guide

Life Scope® BSM-6000 Series Bedside Monitor and BSM-1700 Series Transport Monitor
Clinical Reference Guide
Version 08-12+
© 2019 All Rights Reserved
Table of Contents
Introduction 1
Life Scope
BSM-1700 Series Transport Monitor 4
Basic Operating Concepts 6
Admission Procedure 7
Discharge Procedure 8
ECG Monitoring 9
ECG Signal Acquisition – Electrode Placement 9
ECG Signal Acquisition – Skin Preparation 11
ECG Parameter Operations 12
Arrhythmia Analysis 13
ECG Monitoring – Troubleshooting 15
Non-Invasive Blood Pressure (NIBP) Data Acquisition 16
NIBP Parameter Operations 17
NIBP Monitoring - Troubleshooting 18
®
BSM-6000 Series Bedside Monitor 2
Pulse Oximetry (SpO
SpO
SpO
Parameter Operations 21
2
Monitoring - Troubleshooting 24
2
) Data Acquisition 20
2
Respiration (RESP) Data Acquisition 25
RESP (RR) Parameter Operations 26
RESP Monitoring - Troubleshooting 27
End Tidal CO
End Tidal CO cap-ONE
Data Acquisition Using the TG-920P CO2 Sensor 28
2
Data Acquisition Using the TG-980P/TG-970P
2
®
Sensor 31
CO Parameter Operations for TG-980-P/TG-970-P 32
CO Parameter Operations 33
CO Monitoring - Troubleshooting 34
Temperature (TEMP) 35
Monitoring Screen Options 36
Vital Signs, Parameter and Arrhythmia Alarms 37
Interbed Alarms 39
Noise Detection 40
ST Alarms 41
Arrhythmia Alarms 41
Atrial Fibrillation Alarm 42
Arrhythmia Troubleshooting - Interventions 43
12-Lead Electrocardiogram/ST Analysis 44
12-Lead ECG Procedure 45
ST Analysis 46
ST Alarms 46
To Change the ST Measurement Points 47
To View Minute to Minute Stored ST Data 47
Invasive Blood Pressure (IBP) 48
IBP Parameter Operations 48
Set Up/Zero IBP (Invasive Blood Pressure) 48
Change IBP Label (On Screen and in Cable) 50
Calculation Method – STANDARD, PEAK 50
IBP Monitoring - Troubleshooting 52
PA Catheter Insertion 53
Perform Hemodynamic Readings 53
CO Monitoring - Troubleshooting 55
Patient Data Review 56
Trend Window 57
Trend Graphs 57
Tables 57
NIBP Trend 58
HEMO Calculations 58
LUNG Trend 59
Arrhythmia Recall Window 60
Alarm History Window 61
Full Disclosure Window 62
Zoom In Window 62
ST Window 63
12-Lead Window 63
OCRG Window (Oxy-cardio-respiratory-gram) 64
Drug Calculations 65
Drug Calculations Operations 66
LUNG Calculations 67
LUNG FUNCTION Operations 68
Anesthesia Agent Monitoring 69
Recording 71
Transport Function 72
BSM-1700 Series Monitor 73
BSM-1700 Home Screen Display: Standard Mode/Transport Mode 74
Transport Mode 75
Transport Function using the BSM-1700 75
Connecting to the BSM-6000 Series Monitor 76
Menu Window 77
Using ZM-View Function 78
Starting and Ending ZM-View Function 79
System Settings 80
Cleaning and Disinfecting 81
Introduction
4
Introduction
This Clinical Reference Guide is designed to assist you to learn the monitors basic operations during the pre-implementation training class or as a self-study tool as you use the system on your clinical unit. It is also designed as a reference tool when you need a refresher for infrequently performed procedures.
The procedures in this guide reflect Life Scope TR Bedside Monitor (BSM) 6000 Series and the BSM- 1700 series monitor. Both are based on the use of the monitor in an adult or pediatric environment. The BSM- 1700 series monitor operates similarly to the BSM-6000 series and any differences between the two will be explained at the end of this guide. Otherwise the basic operation is nearly identical to each other and this guide will serve to reflect both BSM’s.
The Clinical Reference Guide is an adjunct to the Operators Manual and does not replace it. Please refer to that manual for critical technical and other specific information, and for additional information as directed in this guide.
Whenever you see this icon in this guide, a one touch shortcut” is available for the particular function. That shortcut
This Clinical Reference Guide is designed to assist you to learn the monitors’ basic operations during the pre-implementation training class or as a self-study tool as you use the system on your clinical unit. It is also designed as a reference tool when you need a refresher for infrequently performed procedures.
1
The procedures in this guide reflect Life Scope
®
BSM-6000 series bedside monitor and the BSM-1700 series transport monitor. Both are based on the use of the monitor in an adult or pediatric environment. The BSM-1700 series monitor operates similarly to the BSM-6000 series and any differences between the two will be explained at the end of this guide. Otherwise, the basic operations are nearly identical to each other and this guide will serve to reflect both BSM’s.
The Clinical Reference Guide is an adjunct to the Operator’s Manual and does not replace it. Please refer to that manual for critical, technical, and other specific information, and for additional information as directed in this guide.
Whenever you see this icon in this guide, a one touch “shortcut” is available for the particular function. That shortcut is described with the icon.
For additional assistance, please contact your hospital biomedical services representative. For technical or clinical product assistance, a toll­free customer support service is provided at 1-800-325-0283.
Life Scope® BSM-6000 Series Bedside Monitor
5
Life Scope TR 6000 Series
Bedside Monitor
2
3
6
The monitor operates on AC power with optional batteries for backup. When the monitor is operating on battery, the screen brightness can be dimmed to conserve power.
The batteries are recharged when the monitor is connected to AC power, and they should be changed every year. Refer to the Operator’s Manual for additional information.
BSM-1700 Series Transport Monitor
7
BSM-1700 Series Monitor
4
5
8
The BSM-1700 series monitor can operate on either battery power or DC power supplied by the optional cradle. Battery life is approximately 4-5 hours. When not in use, it is recommended to keep the BSM-1700 docked
in the cradle and plugged in to charge.
Basic Operating Concepts
9
Basic Operating Concepts
Normally, the monitoring screen is displayed on the bedside monitor. Key operations are performed on the touch screen display using a gentle touch with the fingertip or touch pen, the infrared remote controller, or a computer mouse that is connected to the USB port: 1) Touch (or click) the PARAMETER VALUE to enter the parameter menus, 2) Touch (or click) the PATIENT NAME area to enter the PATIENT INFO menu, 3) Touch a FUNCTION KEY in the upper left corner to access frequently used functions.
The BSM-6000 and 1700 Series bedside monitors have an optional infrared remote controller:
Normally, the monitoring screen is displayed on the bedside monitor. Key operations are performed on the touch screen display using a gentle touch with the fingertip or touch pen, the infrared remote controller, or a computer mouse that is connected to the USB port: 1) Touch (or click) the PARAMETER VALUE to enter the parameter menus; 2) Touch (or click) the PATIENT NAME area to enter the PATIENT INFO menu; 3) Touch a FUNCTION KEY in the upper left corner to access frequently used functions.
The BSM-6000 and -1700 Series bedside monitors have an optional infrared remote controller:
6
The MENU window can be displayed by pressing the MENU key on the front panel or on the remote controller. When you access a function on the monitor, the menu is displayed in a tabbed “file folder” format. Touch or click on the tab to access the individual menu. The monitoring screen can be displayed at any time by pressing the HOME key on the front panel or on the remote controller.
Admission Procedure
The Life Scope BSM 6000 Series bedside monitor is ready for data collection once the power is turned ON.
1. Press the power switch and release — the monitoring screen appears after monitor warms up.
2. Verify the alarm settings and input unit (see VITAL SIGNS, PARAMETER, and ARRHYTHMIA ALARMS section in this guide).
3. Connect the patient.
4. Enter patient demographic information (optional).
a. Press the MENU key
b. Touch the ADMIT button in the PATIENT section on the menu
OR Touch the patient name field at the top of the screen
c. Select the patient type (Adult, Pediatric, or Neonate), if necessary.
   i. Selecting each patient type will set the vital sign alarm limits, arrhythmia alarm settings, NIBP initial cuff inflation pressure, invasive pressure scales, QRS detection type, and trend graph scales, based on the patient type.
7
d. Touch the NAME data button and/or touch the PATIENT ID
e. Enter the patient name and/or number using the on-screen keyboard.
f. DEL clears all characters to the right of the cursor.
g. BS erases to the left of the cursor, one character at a time.
h. Touch the ENTER button to enter the patient name or number.
i. If demographics are entered after patient monitoring has commenced, do not press the ADMIT key. Data loss could occur if this action is performed post demographic entry.
Press the HOME key to return to the monitoring screen. If the bedside monitor is a part of a network with a central nursing station (CNS), the information can be entered at the central station using the ADMIT/DISCHARGE menu.
Discharge Procedure
4
Introduction
This Clinical Reference Guide is designed to assist you to learn the monitors basic operations during the pre-implementation training class or as a self-study tool as you use the system on your clinical unit. It is also designed as a reference tool when you need a refresher for infrequently performed procedures.
The procedures in this guide reflect Life Scope TR Bedside Monitor (BSM) 6000 Series and the BSM- 1700 series monitor. Both are based on the use of the monitor in an adult or pediatric environment. The BSM- 1700 series monitor operates similarly to the BSM-6000 series and any differences between the two will be explained at the end of this guide. Otherwise the basic operation is nearly identical to each other and this guide will serve to reflect both BSM’s.
The Clinical Reference Guide is an adjunct to the Operators Manual and does not replace it. Please refer to that manual for critical technical and other specific information, and for additional information as directed in this guide.
Whenever you see this icon in this guide, a one touch shortcut” is available for the particular function. That shortcut
4
Introduction
This Clinical Reference Guide is designed to assist you to learn the monitors basic operations during the pre-implementation training class or as a self-study tool as you use the system on your clinical unit. It is also designed as a reference tool when you need a refresher for infrequently performed procedures.
The procedures in this guide reflect Life Scope TR Bedside Monitor (BSM) 6000 Series and the BSM- 1700 series monitor. Both are based on the use of the monitor in an adult or pediatric environment. The BSM- 1700 series monitor operates similarly to the BSM-6000 series and any differences between the two will be explained at the end of this guide. Otherwise the basic operation is nearly identical to each other and this guide will serve to reflect both BSM’s.
The Clinical Reference Guide is an adjunct to the Operators Manual and does not replace it. Please refer to that manual for critical technical and other specific information, and for additional information as directed in this guide.
Whenever you see this icon in this guide, a one touch shortcut” is available for the particular function. That shortcut
The discharge procedure is performed between patients to delete information from the bedside monitor and the CNS and to reset default alarm settings on both monitors. When this procedure is not performed, the most current 24 hours of data is stored in memory. The amount of time that data is stored after power is turned off on the bedside monitor depends on the setup functions. Once the information is deleted, it CANNOT be retrieved.
To discharge a patient:
1. Press the MENU key
2. Touch the ADMIT button in the PATIENT section in the menu
OR Touch the patient name field at the top of the screen
3. Touch the DISCHARGE tab
4. Touch YES to confirm. This procedure readies the monitor for the next patient’s admittance.
5. Turn the monitor power off, if needed (press and hold the power switch for 3 seconds until the display blanks).
8
A “COMMUNICATION LOSS” message displays on the central station when the bedside monitor is turned off.
Suspend Monitoring
The SUSPEND MONITORING function is used to suspend data processing and alarms during temporary interruptions in monitoring. No patient information is deleted with this procedure, but the monitoring screen remains active and alarms are suspended while the patient is disconnected.
1. Disconnect the patient from the monitor.
2. Press the MENU key.
3. Touch the SUSPEND MONITORING button in the ALARM section.
The menu closes and the “SUSPEND MONITORING ALL ALARMS OFF” message appears. Monitoring resumes automatically once the patient is reattached, depending on the configured time interval.
A SUSPEND MONITORING function button may be present on the screen and/or the remote for a one-touch function.
ECG Monitoring
12
12
ECG Signal Acquisition – Electrode Placement
The Life Scope G9 bedside monitor is equipped to use 3, 6, or 10 electrodes to monitor heart rate, rhythm, and ST segments, as well as impedance respiration.
9
Chest electrodes may be placed in any chest lead position, depending on the monitoring requirements. V1 and V3 are shown. V1 is useful for arrhythmia monitoring. V3 is the most sensitive lead for anterior wall ischemia. Other V­leads are useful for ST segment and arrhythmia monitoring as well.
*Default V leads
3 Electrode Placement
Chest Lead Position
R/RA (white) Midclavicular line 2 fingers below R clavicle
L/LA (black) Midclavicular line 2 fingers below L clavicle
N/RL (green) R anterior-axillary line at edge or lower rib
F/LL (red) L anterior-axillary line at edge of lower rib
*V1 (blue) 4th inter-costal space(ICS), right sternal border
V2 4th ICS, left sternal border
*V3 (orange) Halfway between V2 and V4
V4 5th ICS, left mid-clavicular line
V5 Anterior axillary line at same level as V4
V6 Mid-axillary line at same level as V4
6 Electrode Placement
The 10-electrode ECG cable is used to perform diagnostic-quality
13
12-lead electrocardiograms, as with an ECG cart. Electrode placement for the procedure on the bedside monitor should be the same as the cart procedure to provide consistency for serial comparison between ECGs. Electrode placement is suggested here but should be decided by individual hospital protocols. Tab or patch electrodes may be used.
10
10 Electrode Placements
Limb electrodes should be placed on the actual limbs if the hospital protocols require this placement for diagnostic 12-lead ECG samples.
Continuous 12-lead monitoring is difficult with the limb leads on the extremities because of the motion artifact, so use the torso limb placement for continuous monitoring, when possible.
See the 12-lead ECG section in this guide for the 12-lead procedure.
ECG Signal Acquisition – Skin Preparation
14
The ECG waveform is monitored for heart rate, arrhythmias, ST segment measurements, and pacemaker activity. Electrical impulses are conducted through the gel on the ECG electrodes, so moist gel is required for proper detection. Inadequate detection can cause electrical noise on the ECG tracing and inaccurate interpretation by the monitor. We recommend that electrodes be changed at least daily to promote adequate signal detection but follow the recommendations of your electrode manufacturer and/or hospital policy.
Proper skin preparation is also important for respiration monitoring, as the respiration parameter uses the same electrodes. It is most important to remove hair and dead skin cells for accurate detection, and to remove skin oils so the electrodes will adhere to the patient’s skin.
1. Select the electrode site according to the diagram above.
2. Trim excess hair according to hospital protocols.
3. Gently abrade skin areas with dry gauze to remove dead skin cells.
4. If skin is oily, clean the site with soap/water and friction and then wipe off with dry gauze. You may use an alcohol prep according to your policy, if needed.
11
5. Attach the lead wire to the electrode.
6. Attach the electrode to the patient, pressing the circumference of electrode to secure it.
7. Position the ECG cable to prevent tangling.
Stress Loop: The lead wires can transmit electrical interference to the monitor; to minimize this interference with patient movement, fasten the individual lead wire to the skin with surgical tape.
ECG Parameter Operations
4
Introduction
This Clinical Reference Guide is designed to assist you to learn the monitors basic operations during the pre-implementation training class or as a self-study tool as you use the system on your clinical unit. It is also designed as a reference tool when you need a refresher for infrequently performed procedures.
The procedures in this guide reflect Life Scope TR Bedside Monitor (BSM) 6000 Series and the BSM- 1700 series monitor. Both are based on the use of the monitor in an adult or pediatric environment. The BSM- 1700 series monitor operates similarly to the BSM-6000 series and any differences between the two will be explained at the end of this guide. Otherwise the basic operation is nearly identical to each other and this guide will serve to reflect both BSM’s.
The Clinical Reference Guide is an adjunct to the Operators Manual and does not replace it. Please refer to that manual for critical technical and other specific information, and for additional information as directed in this guide.
Whenever you see this icon in this guide, a one touch shortcut” is available for the particular function. That shortcut
The heart rate is calculated by using a moving average or by counting each individual beat. The AVERAGE mode uses an average of 12 beats and updates them with each new beat to display the new average every 3 seconds. The INSTANT mode updates the heart rate with each beat and displays the new rate every 3 seconds.
The ECG menu allows adjustments while monitoring ECG. To access the ECG parameter menu, touch the heart rate value.
MAIN Tab Options:
Sensitivity – Increases or decreases the size of the ECG tracings.
 1. Touch x1/4, x1/2, Auto, x1, x2, or x4. Ensure correct electrode
a. Auto allows the monitor to set a sensitivity that is required for detection based on the previous 16 beats.
Lead – The top lead is analyzed for arrhythmias when SINGLE arrhythmia analysis is selected; the top two leads are analyzed with MULTIPLE. Ensure correct electrode placement and then select the option to display the lead with QRS complexes that are > 5 mm tall.
placement and then select the option to display the QRS at > 5 mm tall.
12
Touch the desired lead to display in the top position — unavailable leads are grayed out.
Alarms – HEART RATE/VPC/ST Increase, decrease, or turn OFF alarm limits
1. Touch limit and then
2. Touch limit and then
Change alarm limit by touching and dragging the slider.
LEARN Initiates a “relearn” of the patient’s dominant QRS for arrhythmia processing and the “LEARNING” message is displayed on screen. Use this function whenever the patient’s rhythm has changed, incorrect detection occurs, or to store a new dominant template in the MAIN or ARRHYTHMIA ANALYSIS tabs. ST ALARMS tab: allows for setting alarms on all monitored leads. Provides alarm settings for individual or ALL leads. Individual settings set alarms relative to iso point; ALL sets relative to current ST measurements.
s
to increase
t
to decrease OR
Arrhythmia Analysis
Arrhythmia alarms occur when ARRHYTHMIA ANALYSIS is turned on in the ARRHYTH ANALYSIS tab, the individual arrhythmia alarm is turned to ON, and the monitor detects an arrhythmia event. Single (top) or multiple (top two) ECG leads are selected for analysis in the ARRHYTH ANALYSIS tab.
To view or change these settings:
1. Touch the heart rate value.
2. Touch the ARRHYTH tab.
3. Touch the ON button for the arrhythmia events you want to alarm.
OTHER Tab Options:
Filters – DIAG, MONITOR, MAXIMUM Filters reduce AC interference on the ECG waveform. MONITOR should be chosen for routine patient monitoring.
HUM Filter – ON/OFF HUM filters reduce interference on the ECG waveform during battery operation. ON should be chosen for routine patient monitoring with the monitor plugged into AC power.
Number of Electrodes Signals the monitor to the type and numbers of ECG lead wires. Choose STANDARD for Nihon Kohden lead blocks. Use INDIV 6 or INDIV 10 when using individual or disposable lead wires.
13
ECG Measurement – ON/OFF This option is available if SpO
and pulse rate monitoring are needed and
2
can only be turned off if no ECG signal is detected. Turn OFF to view only SpO
and SpO2 PR when patient is connected to SpO2 probe and monitor.
2
Sync Pitch – FIXED, SpO Provides a variable change in the audible tone as the SpO
, PRESS
2
value
2
changes. FIXED does not change as the value changes.
Sync Source – ECG, SpO
, PRESS
2
Provides the sync sound from the selected pulse source. If the QRS SYNC function is enabled, a green light on top of the monitor also flashes with each pulse.
When SpO
or PRESS is selected, the associated pulse rate is displayed in
2
the corresponding color when the PR is ON. This is helpful in OR when the cautery is in use. SYNC SOUND VOLUME is controlled in MENU > VOLUME.
HR Display Mode – AVERAGE, INSTANT
Two modes that can be used to calculate heart rate:
AVERAGE – Monitor detects 12 consecutive beats (including VPCs), averages the R-R intervals, and uses this average to calculate the current HR. When a new beat is detected, the HR is recalculated using the latest 12 beats. The HR display is updated every 3 seconds.
INSTANT – HR is calculated based on the last 2 beats and is updated every 3 seconds.
Pacing
Detection – ON/OFF Turn this to ON when the patient has a pacemaker. This tells the monitor to detect the pacer spike and look for capture following that spike. False arrhythmia alarms could occur if detection is left OFF with an internal pacemaker present.
Mark – ON/OFF Turn this ON to enhance the pacer spike ON SCREEN. The enhanced spike is not printed.
14
Auto Lead Change – ON/OFF This option is available with the 6 or 10 electrode cable. This allows for the monitored lead to automatically change between lead I, II, and III if lead wires or electrodes become disconnected for more than 5 seconds.
ECG 2/3 Waves Tab: This allows you to select the numbers of ECG leads to display (1, 2, or 3), the lead to display in each position, and to change the sensitivity on the second and third leads. Choose the primary (top) ECG lead in the ECG MAIN tab.
V-Leads Tab: This allows you to assign the V-leads for the Va and Vb chest leads when using the 6 electrode cable. Once the leads are assigned, they become available on the ECG 2/3 Waves tab.
ST Point Tab: This allows you to adjust the measurement points from the default settings and the markers show the iso, r-wave, j-point, and ST point. Use the < > keys for changes.
ECG Monitoring – Troubleshooting (see operator manual)
Problem Possible Cause Action
Inaccurate HR QRS is too small Change sensitivity so QRS is >
5 mm tall
Arrhythmia alarm occurs frequently when HR is normal
ECG waveform does not appear on screen when electrodes are properly attached
AC interference on the waveform
Baseline wandering
QRS is not detected correctly
Dominant QRS is not appropriate for arrhythmia monitoring
Patient moved or EMG noise is superimposed
NUMBER OF ELECTRODES setting in ECG MENU is not correct
Electrode is dry Change electrode(s)
ECG FILTER is set to DIAGNOSTIC
Baseline is not stable due to respiration or movement
The electrode is dry Change electrodes
Change to different lead with a tall QRS and small T wave (View all leads of available ECG in the 12-Lead Analysis tab)
Learn ECG or change the ECG lead
Move electrodes to positions with less muscle. Use stress loops.
Set correct number of electrodes for attached ECG cable. Use STANDARD unless individual lead wires are used.
Set ECG FILTER to MONITOR or MAXIMUM
Move electrodes to positions with less muscle. Consider stress loops.
15
Inaccurate pacemaker detection or false VPC calls for paced beats
Contact resistance between skin and electrode is high
The ECG filter is set to DIAGNOSTIC
Pacemaker impulse is not detected and/ or paced beats are labeled as ventricular
Abrade skin gently with dry gauze before applying electrodes
Set the ECG filter to MONITOR or MAXIMUM
Ensure PACING DETECTION is ON. Display a different ECG lead, such as V1. Try alternate electrode placement: Move LL up and RA down.
Non-Invasive Blood Pressure (NIBP)
19
Data Acquisition
The Nihon Kohden NIBP monitor uses the occlusive-oscillometric method to measure systolic, diastolic, and mean arterial non- invasive blood pressure. This method may yield dierent results from direct, invasive
arterial pressure.
The initial adult inflation volume is 180mmHg, pediatric is 140mmHg, and then subsequent inflations are 30mmHg above the last systolic reading. ADULT/CHILD is displayed in the NIBP numeric area when that hose/cuff is connected.
The monitor automatically detects the neonatal hose when connected and displays a “NEONATAL” message in the NIBP numeric area. The initial neonatal inflation volume is 70-100mmHg (based on system configuration), and then subsequent inflations are 30mmHg above the last systolic reading.
The NIBP is measured from the change in amplitude pattern of pulsatile oscillation in cuff pressure as the cuff pressure is reduced from above systolic to below diastolic pressure. The systolic pressure is the pressure at which the pulsatile oscillation suddenly increases, and the diastolic pressure is the pressure at which the pulsatile oscillation suddenly decreases. The mean arterial pressure is the point where maximum pulsatile oscillation occurs.
16
INACCURATE READINGS MAY OCCUR DURING PATIENT MOVEMENT.
Cuff selection should be based on the size of the patient’s arm. The American Heart Association recommends that the cuff width be 40% of the circumference of the upper arm. Inappropriate cu size can result in
inaccurate NIBP results.
NIBP Parameter Operations
4
Introduction
This Clinical Reference Guide is designed to assist you to learn the monitors basic operations during the pre-implementation training class or as a self-study tool as you use the system on your clinical unit. It is also designed as a reference tool when you need a refresher for infrequently performed procedures.
The procedures in this guide reflect Life Scope TR Bedside Monitor (BSM) 6000 Series and the BSM- 1700 series monitor. Both are based on the use of the monitor in an adult or pediatric environment. The BSM- 1700 series monitor operates similarly to the BSM-6000 series and any differences between the two will be explained at the end of this guide. Otherwise the basic operation is nearly identical to each other and this guide will serve to reflect both BSM’s.
The Clinical Reference Guide is an adjunct to the Operators Manual and does not replace it. Please refer to that manual for critical technical and other specific information, and for additional information as directed in this guide.
Whenever you see this icon in this guide, a one touch shortcut” is available for the particular function. That shortcut
4
Introduction
This Clinical Reference Guide is designed to assist you to learn the monitors basic operations during the pre-implementation training class or as a self-study tool as you use the system on your clinical unit. It is also designed as a reference tool when you need a refresher for infrequently performed procedures.
The procedures in this guide reflect Life Scope TR Bedside Monitor (BSM) 6000 Series and the BSM- 1700 series monitor. Both are based on the use of the monitor in an adult or pediatric environment. The BSM- 1700 series monitor operates similarly to the BSM-6000 series and any differences between the two will be explained at the end of this guide. Otherwise the basic operation is nearly identical to each other and this guide will serve to reflect both BSM’s.
The Clinical Reference Guide is an adjunct to the Operators Manual and does not replace it. Please refer to that manual for critical technical and other specific information, and for additional information as directed in this guide.
Whenever you see this icon in this guide, a one touch shortcut” is available for the particular function. That shortcut
1. Ensure the appropriate cuff size for the patient and position extremity at heart level, as readings may vary with a cuff higher or lower than this position. Monitor automatically identies hose connected
(ADULT/CHILD, NEONATE) and displays that on screen.
2. Press the NIBP INTERVAL key to display the automatic interval options and select the desired interval.
3. Press the NIBP START/STOP key to initiate reading.
Single NIBP readings can be started at any time using the NIBP START/STOP key
To access the NIBP parameter menu, touch the NIBP value.
MAIN Tab Options:
Initial Cu Pressure Type:
This allows you to choose ADULT or CHILD to adjust the initial inflation pressure. Subsequent inflations are 30mmHg over the last systolic reading.
Measurement Interval – Manual to 8-hour
Touch the desired interval and then press the NIBP START/STOP key to initiate the automatic readings. Inflation may start automatically if configured to do so.
17
STAT mode takes continuous readings for 10 minutes as set in
the setup menu (view in NIBP Mode tab).
SIM mode is available in the OR mode and allows for staged
readings as set in the setup menu (view in NIBP Mode tab).
NIBP intervals may be set using the NIBP INTERVAL key.
Alarms – SYSTOLIC, DIASTOLIC, MAP Increase, decrease, or turn OFF alarm limits
1. Touch limit then
2. Touch limit then
Set value to ABOVE MAXIMUM or BELOW MINIMUM to turn alarm OFF.
PWTT Tab: ON/OFF – choose to set the trigger time in minutes. Available with NK SpO when SpO
s
to increase
t
to decrease
monitors and allows for automatic blood pressure readings
2
indicates a change in blood pressure.
2
OTHER Tab Options:
Target Cu Pressure – Adjusts inflation volume. Use this for patient comfort before the cuff inflates.
Display Oscillation Graph – Graph shows the detected pulse amplitude on a pressure scale and can be used to determine if the reading was affected by patient movement or other interference. Can be turned ON/ OFF. (Graph only displayed on home screen if there is enough room.)
NIBP Monitoring – Troubleshooting (see operator manual)
Problem Possible Cause Action
Cuff inflation pressure is < 10mmhg or NIBP data disappears for a few seconds
Cuff does not inflate when NIBP START/ STOP key is pressed
Abnormal measurement results are displayed
Auto measurement does not start even when time interval has passed
Cuff suddenly inflates
Cuff hose is not connected to monitor socket properly
Hose is not connected to socket
Cuff size is incorrect Select cuff size appropriate to
Cuff is not wrapped correctly
Patient movement Take readings without patient
Measurement on wrong site
Time interval is set incorrectly or NIBP START/STOP key was not pressed to set interval
Measurement mode is set to AUTO
NIBP measurement is triggered by PWTT
Connect hose to monitor properly
Connect the hose to socket firmly
patient’s size
Position cuff snuggly on patient’s arm
movement
Position cuff over artery
Check interval and press NIBP START/STOP key
Check time interval
PWTT TRIGGERED NIBP is set to a threshold on the SYSTEM SETUP screen
18
Problem Possible Cause Action
“MEASURE TIME OUT” message
“SYSTOLIC OVER” message is displayed
Cannot measure NIBP
Patient movement or irregular pulse
Blood pressure is higher than the NEONATAL mode can detect
Noise interfered with calculation of NIBP
Pulse wave is unstable due to arrhythmias
Cu or hose has leak Change cu or hose
Reposition, retry, or use another method
Change to the ADULT/CHILD hose and use the appropriately sized cuff for the infant.
Remove cause
Check blood pressure using alternative methods
19
Blood congestion occurs
Thrombus occurs
NIBP data on the screen is dark
“CHECK CUFF AND HOSE” message is displayed
“PATIENT MOVEMENT REMEASURING” message is displayed
Measuring over long periods of time at intervals less than2.5 minutes
Sickle cell patients NIBP is contraindicated on sickle
10 minutes elapsed from last reading
Cu or hose has leak or is hose is bent or obstructed
Patient moved during measurement
Patient moved during measurement or pulse is too small to detect.
Cu is not appropriate for patient or is attached incorrectly
Increase measuring interval
Use alternate site for NIBP measurements
cell anemia patients
NIBP data returns to normal brightness with next reading
Check cu and hose for leaks or straighten hoses
Wait for patient to stop moving and measure again
Wait for patient to stop moving and measure again
Check that cu is appropriate and positioned correctly
Pulse Oximetry (SpO2) Data Acquisition
23
2
cable, which accepts
The Nihon Kohden SpO2 monitor uses a SpO2 cable, which accepts reusable or disposable probes. The monitor comes in three options for pulse oximetry: Nihon Kohden, Nellcor OxiMax, and Masimo. The menu options vary slightly depending on the algorithm used.
Nihon Kohden probes are obtained from Nihon Kohden, but Nellcor and Masimo probes are obtained from those companies directly. There are many options for probes from each company, ranging from neonate to adult, and it is important to use the appropriate probe for the size and age of the patient. There are options for probes that are used on the ear and forehead, in addition to the traditional foot or hand and finger probes. See the operator manual for additional information.
The probe is positioned with the photo emitter (cable side) on top so light is passed through the measurement site to the photo detector directly opposite the emitter. Results are best if used on fingers without nail polish or acrylic nails. Ambient light may affect readings, so placing the extremity under cover may improve detection.
20
When measuring SpO
for long periods of time, it is recommended that
2
the site be alternated to relieve pressure on the measuring digit.
Change site every four hours for reusable and every four to eight hours for disposable probes. Discontinue use of probe if skin irritation occurs or signs of circulatory compromise.
The SpO
monitor is able to “read through”
2
motion and alerts you to this condition by placing an “M” on the screen.
The Masimo version uses this symbol to indicate a “low Signal IQ” that is discussed below.
SpO2 Parameter Operations
To access the SpO2 parameter menu, touch the SpO2 value.
MAIN Tab Options:
Sensitivity – X1/8 to X8 – AUTO finds the appropriate size and adjusts the size of the SpO Touch the setting and observe the waveform in the SpO
waveform on the screen and on the recording paper.
2
window.
2
21
Alarms – SpO
Alarm Limits
2
Increase, decrease, or turn OFF alarm limits.
1. Touch limit then
2. Touch limit then
s
to increase
t
to decrease
Set value to ABOVE MAXIMUM or BELOW MINIMUM to turn alarm OFF.
OTHER Tab Options:
Sync Source – ECG, SpO2, PRESS Provides the sync sound from the selected pulse source. If the QRS SYNC function is enabled, a green light on top of the monitor also flashes with each pulse. WHEN SpO
ASSOCIATED PULSE RATE IS DISPLAYED IN THE CORRESPONDING COLOR WHEN THE PR IS ON. This is helpful in OR when the cautery is in use.
SYNC SOUND VOLUME is controlled in MENU>VOLUME.
Sync Pitch – FIXED, SpO
, PRESS
2
Provides a variable change in the audible tone, as the SpO changes and can be configured in one of two ways. FIXED does not change as the value changes.
OR PRESS IS SELECTED,THE
2
value
2
SPO2 VALUE Sync Pitch of Sync Sound
100 to 81% High to low pitch, in 1% steps
100 to 40% High to low pitch, in 1% steps
Less than 81% or less than 40%
Low Pitch
Nihon Kohden Monitors Only
Response - FAST, NORMAL, SLOW
There are three response modes in the Nihon Kohden version of the BSM. Each uses a different time to ensure accurate measurements according to patient conditions. When measurement conditions are unstable, the response becomes slower in all modes.
FAST: Select this mode for special applications that require a fast response time, such as with short apnea episodes.
NORMAL: Select this for normal monitoring.
22
SLOW: Select this when you need to suppress a rapid change in SpO
.
2
Masimo Monitors Only
Perfusion Index
Indicates the percentage of pulsatile (arterial) to non-pulsatile (venous) signals. The range of Perfusion Index is 0.02% (very weak pulse strength) to 20% (very strong pulse strength). Placing the sensor at the site with the highest Perfusion Index number improves the performance during motion.
Perfusion Index Display
To display the Perfusion Index, press the SpO Press ON/OFF tab for PI DISPLAY. The PI will be displayed next to the associated SpO
value.
2
Signal IQ
The signal quality indicator is used to identify the occurrence of the pulse and the associated signal quality of the measurement. The Signal IQ, shown as a vertical line, coincides with the peak of the arterial pulsation on the waveform. The height of the line indicates the quality of the measured signal. When the signal quality is very low, the height of the line becomes low and a “M” is displayed along with the SpO
tab, then NUMERIC DISPLAY.
2
value on the monitor.
2
Loading...
+ 61 hidden pages