Metrohm 930 User Manual

930 Compact IC Flex
930 Compact IC Flex Oven/Deg
Manual – Short Instructions
8.930.8112EN
Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
930 Compact IC Flex
930 Compact IC Flex Oven/Deg
2.930.2160
8.930.8112EN
Manual – Short Instructions
11.2013 zst
Teachware Metrohm AG CH-9100 Herisau teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
Documentation in additional languages can be found on
http://documents.metrohm.com.
■■■■■■■■■■■■■■■■■■■■■■

Table of contents

1 About these Short Instructions 1
2 Introduction 2
2.1 Instrument description ......................................................... 2
2.2 Intended use ......................................................................... 4
2.3 Safety instructions ................................................................ 4
2.3.1 General notes on safety ........................................................... 4
2.3.2 Electrical safety ........................................................................ 4
2.3.3 Tubing and capillary connections ............................................. 5
2.3.4 Flammable solvents and chemicals ........................................... 5
2.3.5 Recycling and disposal ............................................................. 6
2.4 Symbols and conventions .................................................... 6
3 Overview of the instrument 8
Table of contents
3.1 Front ...................................................................................... 8
3.2 Rear ........................................................................................ 9
4 Installation 11
4.1 Setting up the instrument .................................................. 11
4.1.1 Packaging .............................................................................. 11
4.1.2 Checks .................................................................................. 11
4.1.3 Location ................................................................................ 11
4.2 Removing transport locking screws ................................. 11
4.3 Connecting the drainage tubing and leak sensor ............ 13
4.3.1 Installing the drainage tubing ................................................ 13
4.3.2 Connecting the leak sensor .................................................... 14
4.4 Column oven ....................................................................... 15
4.5 Connecting the eluent bottle ............................................. 15
4.6 Connecting the eluent degasser ........................................ 17
4.7 Installing the high-pressure pump .................................... 17
4.8 Installing an inline filter ..................................................... 17
4.9 Installing the pulsation absorber ...................................... 17
4.10 Injection valve ..................................................................... 17
4.11 Installing the detector ........................................................ 17
4.12 Connecting the sample degasser (optional) ..................... 18
4.13 Connecting the instrument to a computer ....................... 19
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
III
Table of contents
■■■■■■■■■■■■■■■■■■■■■■
4.14 Connecting the instrument to the power supply ............. 20
4.15 Initial start-up ..................................................................... 21
4.16 Connecting and rinsing the guard column ....................... 23
4.17 Connecting the separation column ................................... 25
4.18 Conditioning ........................................................................ 27
5 Operation 29
6 Warranty (guarantee) 30
7 Accessories 32
Index 34
■■■■■■■■
IV
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■

Table of figures

Figure 1 Front ................................................................................................. 8
Figure 2 Rear .................................................................................................. 9
Figure 3 Removing the transport locking screws ............................................ 12
Table of figures
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
V
■■■■■■■■■■■■■■■■■■■■■■
1 About these Short Instructions
This short instruction manual contains important chapters from the com­prehensive manual. In addition to an introduction, safety instructions and an overview of the instrument, you will also find information about instal­ling and operating the 930 Compact IC Flex Oven/Deg as well as informa­tion regarding the warranty. You can download the detailed manual as a PDF file from the Internet.
Downloading the manual
You can find the detailed manual on the Internet under http://docu-
ments.metrohm.com.
1. Click on Instruments Ion Chromatography.
2. Select the line of instruments, e.g. 930 Compact IC Flex.
3. Select the instrument.
All available manuals for the instrument will be displayed.

1 About these Short Instructions

930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
1

2.1 Instrument description

2 Introduction

2.1 Instrument description
The 930 Compact IC Flex is a professional ion chromatograph. It is distin­guished by:
Its intelligence: All of the functions are monitored, optimized and
documented in an FDA-compatible manner. Intelligent components, such as iColumns, save important data onto a chip.
Its compact design: It has a small footprint. Its transparency: All components are easily accessible and located for
simple visibility and can be monitored during operation through a large window.
Its safety: The wet end and electronics are physically separated, thereby
preventing fluids from coming into contact with the electronics to a large extent. A leak sensor is integrated into the wet end.
Its environmental compatibility. Its low noise emissions. The intelligent MagIC Net™ software
■■■■■■■■■■■■■■■■■■■■■■
The 930 Compact IC Flex is operated solely using the MagIC Net software. A USB cable is used to connect the instrument to a computer with MagIC Net installed. The intelligent software detects the instrument auto­matically and checks its functionality. The software controls and monitors the instrument, evaluates the measured data and manages it in a data­base.
The 930 Compact IC Flex Oven/Deg consists of the following modules:
Leak sensor
The leak sensor detects leaking liquid that collects in the instrument's base tray. Liquid that leaks in the instrument is routed to the base tray using drainage tubing and detected there.
Column oven
The column oven tempers the separation column and the eluent and thereby provides stable measuring conditions. A column holder with a chip reader is located in the column oven.
Eluent degasser
The eluent degasser removes gas bubbles and dissolved gases from the eluent.
■■■■■■■■
2
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
2 Introduction
High-pressure pump
The intelligent and low-pulsation high-pressure pump pumps the eluent through the system. It is equipped with a chip where its technical specifi­cations and "life history" (operating hours, service data, etc.) are saved.
Inline filter
Inline filters protect the separation column reliably from potential contami­nation from the eluent. The small filter pads with 2 µm pore size can be replaced quickly and easily. They remove particles from the solutions, such as bacteria and algae.
Pulsation absorber
The pulsation absorber protects the separation column from damage caused by pressure fluctuations when switching the injection valve, and reduces interfering pulsations during highly sensitive measurements.
Injection valve
The injection valve connects the eluent path to the sample path. By a quick and precise switching of the valve a quantity of sample solution defined by the size of the sample loop is injected and flushed to the sepa­ration column with the eluent.
Detector
Metrohm offers a series of different detectors for various analysis tasks. A suitable detector type must be ordered as a separate accessory.
Sample degasser
The sample degasser removes gas bubbles and dissolved gases from the sample.
Separation column
The intelligent separation column separates different components accord­ing to their interactions with the column. Metrohm separation columns are equipped with a chip where their technical specifications and history (start-up, operating hours, etc) are stored.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
3

2.2 Intended use

2.2 Intended use
The 930 Compact IC Flex Oven/Deg is used for the determination of cati­ons and anions without suppression using ion chromatography.
The present instrument is suitable for processing chemicals and flammable samples. Usage of the 930 Compact IC Flex therefore requires the user to have basic knowledge and experience in handling toxic and caustic sub­stances. Knowledge with respect to the application of the fire prevention measures prescribed for laboratories is also mandatory.

2.3 Safety instructions

2.3.1 General notes on safety

WARNING
■■■■■■■■■■■■■■■■■■■■■■
This instrument may only be operated in accordance with the specifica­tions in this documentation.
This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.

2.3.2 Electrical safety

The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.
Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.
Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.
WARNING
WARNING
■■■■■■■■
4
There are no parts inside the housing which can be serviced or replaced by the user.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
2 Introduction
Mains voltage
WARNING
An incorrect mains voltage can damage the instrument.
Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
Protection against electrostatic charges
WARNING
Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.
Do not fail to pull the mains cable out of the mains connection socket before you set up or disconnect electrical plug connections at the rear of the instrument.

2.3.3 Tubing and capillary connections

CAUTION
Leaks in tubing and capillary connections are a safety risk. Tighten all connections well by hand. Avoid applying excessive force to tubing connections. Damaged tubing ends lead to leakage. Appropriate tools can be used to loosen connections.
Check the connections regularly for leakage. If the instrument is used mainly in unattended operation, then weekly inspections are manda­tory.

2.3.4 Flammable solvents and chemicals

WARNING
All relevant safety measures are to be observed when working with flammable solvents and chemicals.
Set up the instrument in a well-ventilated location (e.g. fume cup-
board).
Keep all sources of flame far from the workplace. Clean up spilled liquids and solids immediately. Follow the safety instructions of the chemical manufacturer.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
5

2.4 Symbols and conventions

2.3.5 Recycling and disposal

This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative effects on the environment and public health.
More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.
2.4 Symbols and conventions
The following symbols and formatting may appear in this documentation:
Cross-reference to figure legend
The first number refers to the figure number, the sec­ond to the instrument part in the figure.
■■■■■■■■■■■■■■■■■■■■■■
Instruction step
Carry out these steps in the sequence shown.
Method Dialog text, parameter in the software
File New Menu or menu item
[Next] Button or key
WARNING
This symbol draws attention to a possible life-threat­ening hazard or risk of injury.
WARNING
This symbol draws attention to a possible hazard due to electrical current.
WARNING
This symbol draws attention to a possible hazard due to heat or hot instrument parts.
WARNING
This symbol draws attention to a possible biological hazard.
■■■■■■■■
6
CAUTION
This symbol draws attention to possible damage to instruments or instrument parts.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
2 Introduction
NOTE
This symbol highlights additional information and tips.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
7

3.1 Front

3 Overview of the instrument

3.1 Front
■■■■■■■■■■■■■■■■■■■■■■
1
3
5
7
■■■■■■■■
8
Figure 1 Front
Bottle holder
Offers space for the eluent bottle and addi­tional accessories.
Column oven
Inline filter
Pulsation absorber
Column holder
2
For hanging the separation column (iCol­umn). With column recognition.
Sample degasser
4
Injection valve
6
Base tray
8
With leak sensor.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
3 Overview of the instrument
Purge valve
9
For deaerating the high-pressure pump. With rotary knob in the center and pressure sensor.
Eluent degasser
11

3.2 Rear

High-pressure pump
10
Pumps the eluent through the IC system.
Detector chamber
12
Offers space for an embedded detector and additional accessories.
Figure 2
Bottle holder
1
930 Compact IC Flex Oven/Deg (2.930.2160)
Rear
Drainage tubing connection
2
For connecting the drainage tubing, which guides escaped fluids away from the bottle holder.
■■■■■■■■
9
3.2 Rear
■■■■■■■■■■■■■■■■■■■■■■
Knurled screws
3
For fastening the removable back panel.
Drainage tubing connection
5
For connecting the drainage tubing, which guides escaped fluids away from the detec­tor chamber.
Exhaust opening
7
Labeled Exhaust. For extracting the air from the vacuum chamber.
Service connection socket
9
For Metrohm service only.
Drainage tubing connections
11
For connecting two drainage tubings that guide the escaped fluid to the leak sensor and from there to the waste container.
Type plate
13
Leak sensor connection socket
15
Labeled Leak Sensor. For connecting the leak sensor connection cable.
Back panel
4
Removable. Enables access to the detector chamber.
Transport locking screws
6
For securing the vacuum pump when trans­porting the instrument.
Auxiliary connection socket
8
For connecting an 891 Professional Analog Out (2.891.0010).
Transport locking screws
10
For securing the high-pressure pump when transporting the instrument.
Base tray
12
With leak sensor.
Serial number
14
Detector connection socket
16
Labeled Detector. For connecting Metrohm detectors.
MSB connection sockets
17
Labeled MSB 1 and MSB 2. For connecting MSB devices.
PC connection socket
19
For connecting the instrument to the com­puter with the USB cable (6.2151.020).
Power socket
21
Power socket for connecting the power cable and power switch for switching the instrument on and off.
USB connection sockets
18
Labeled USB 1 and USB 2. For connecting USB devices.
Vacuum connection
20
Plugged with a stopper.
■■■■■■■■
10
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
4 Installation

4.1 Setting up the instrument

4.1.1 Packaging

The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.

4.1.2 Checks

Immediately after receipt, check whether the shipment has arrived com­plete and without damage by comparing it with the delivery note.

4.1.3 Location

The instrument has been developed for operation indoors and may not be used in explosive environments.

4 Installation

Place the instrument in a location of the laboratory which is suitable for operation, free of vibrations, protected from corrosive atmosphere, and contamination by chemicals.
The instrument should be protected against excessive temperature fluctu­ations and direct sunlight.

4.2 Removing transport locking screws

To avoid damage to the drives for the high-pressure pump and the vac­uum pump during transport, the pumps are secured with transport lock­ing screws. These are located at the rear of the instrument and labeled with Transport security screws.
Remove these transport locking screws before the initial start-up.
Accessories
For this step you need:
Hex key 4 mm (6.2621.030)
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
11
4.2 Removing transport locking screws
Remove the transport locking screws
■■■■■■■■■■■■■■■■■■■■■■
Figure 3 Removing the transport locking screws
Transport locking screws
1
For the vacuum pump. Only present if the instrument has a degasser.
Store the transport locking screws in a safe place. Reinsert the transport locking screws each time you transport the instrument a significant dis­tance.
The pumps may be damaged if you transport the instrument without inserting the transport locking screws.
CAUTION
Transport locking screws
2
For the high-pressure pump.
■■■■■■■■
12
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
4 Installation

4.3 Connecting the drainage tubing and leak sensor

The leak sensor detects leaking liquid that collects in the instrument's base tray. Liquid that leaks in the bottle holder or in the detector chamber is routed to the base tray using drainage tubing and is detected there.
If the leak sensor detects a leak in the IC system, the IC instrument is switched off and a warning is output in the software.
The leak sensor functions properly only if the following preconditions are met:
The drainage tubing is connected. The leak sensor connection cable is inserted into the leak sensor con-
nection socket.
The 930 Compact IC Flex is switched on. The leak sensor is switched to active in the software.

4.3.1 Installing the drainage tubing

Liquid that leaks in the bottle holder or detector chamber flows to the rear of the instrument and is drained through openings on the bottle holder and in the detector chamber. The drainage tubing has to be mounted at these openings. This drainage tubing guides the leaking liquid to the base tray where the leak sensor is located.
Accessories
For this step you need the following parts from the accessory kit: Vario/ Flex Basic (6.5000.000):
2 × silicone tubing (6.1816.020) Y connector (6.1807.010)
You also need scissors.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
13
4.3 Connecting the drainage tubing and leak sensor
Connecting the drainage tubing
■■■■■■■■■■■■■■■■■■■■■■

4.3.2 Connecting the leak sensor

Plugging in the leak sensor connection cable
The leak sensor connection cable is coiled up in the base tray.
■■■■■■■■
14
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■

4.4 Column oven

The column oven is completely connected. No installation work is required.

4.5 Connecting the eluent bottle

The eluent is aspirated out of the eluent bottle via the eluent aspiration tubing . The eluent aspiration tubing is installed on the input for the elu­ent degasser.
4 Installation
Accessories
For this step you need the following accessories:
These parts are part of the accessory kit Vario/Flex ONE (6.5000.010).
Eluent bottle (6.1608.070) The eluent bottle cap GL 45 accessory set (6.1602.160)
This accessory set contains the bottle cap, an M6 tubing nipple, an M8 tubing nipple, two O-rings and an M6 and M8 threaded stopper.
The tubing adapter for aspiration filter accessory set (6.2744.210)
This accessory set contains a filter holder, a clamping screw and tubing weighting.
An aspiration filter (6.2821.090)
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
15
4.5 Connecting the eluent bottle
■■■■■■■■■■■■■■■■■■■■■■
Connecting the eluent aspiration tubing
■■■■■■■■
16
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■

4.6 Connecting the eluent degasser

The eluent degasser is completely connected. No installation work is required.

4.7 Installing the high-pressure pump

The high-pressure pump is completely connected. No installation work is required.

4.8 Installing an inline filter

The inline filter is completely connected. No installation work is required.

4.9 Installing the pulsation absorber

4 Installation
The pulsation absorber is installed between the high-pressure pump and the injection valve. It protects the separation column from damage caused by pressure fluctuations such as those when switching the injection valve, and reduces interfering pulsations during highly sensitive measurements.
The pulsation absorber is completely connected. No installation work is required.

4.10 Injection valve

The injection valve is completely connected. No installation work is required.

4.11 Installing the detector

The 930 Compact IC Flex provides enough space for two detectors and additional accessories in the detector chamber. The detectors are available as accessories and are supplied with separate manuals.
Placing the detector in the instrument
Follow the instructions in the chapter Inserting the detector in the manual for the detector.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
17

4.12 Connecting the sample degasser (optional)

Connecting the detector to the eluent path
NOTE
The separation column is not inserted into the instrument until it is being started up for the first time. Until then, the detector input capil­lary has to be connected directly to the column input capillary using a coupling (6.2744.040).
Accessories For this step you need the following accessories:
Coupling (6.2744.040) 2 × pressure screw (6.2744.010)
■■■■■■■■■■■■■■■■■■■■■■
4.12 Connecting the sample degasser (optional)
Gas bubbles in the sample lead to poor reproducibility, as the quantity of sample in the sample loop is not always the same. Therefore, we recom­mend degassing samples that contain gas before injection.
NOTE
The sample degasser does not have to be connected. We recommend only using the sample degasser if the application requires it.
The rinsing time increases by at least 2 minutes when the sample degasser is connected.
Accessories
For this step you need the following accessories:
2 × pressure screw, long (6.2744.090) PTFE capillary (6.1803.040)
■■■■■■■■
18
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
Connecting the sample degasser
CAUTION
If the sample degasser is not used, the input and output must be sealed with threaded stoppers (6.2744.220).

4.13 Connecting the instrument to a computer

NOTE
The instrument must be switched off when being connected to a com­puter.
Accessories
For this step you need the following accessories:
USB connecting cable (6.2151.020) USB connection cable (6.2151.020) from the accessory kit: Vario/Flex
Basic (6.5000.000)
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
19

4.14 Connecting the instrument to the power supply

Connecting the USB cable
Insert the USB cable into the computer connection socket on the rear
1
of the instrument.
Insert the other end into a USB port on the computer.
2
■■■■■■■■■■■■■■■■■■■■■■
4.14 Connecting the instrument to the power supply
WARNING
The power supply unit must not get wet. Protect it from liquids.
The power supply cable is three-core and provided with a plug with grounding. If another plug has to be mounted, the yellow/green conduc­tor (IEC standard) must be connected to the protective ground (protection class I).
Accessories
For this step you need the following accessories:
For Switzerland, …: Power supply cable with IEC 60320 line socket,
type C13, with SEV 1011 plug, type 12 (6.2122.020), 1.5 m
For Germany, …: Power supply cable with IEC 60320 line socket, type
C13, with CEE 7 plug, type VII (6.2122.040), 1.5 m
For the USA, …: Power supply cable with IEC 60320 line socket, type
C13, with NEMA 5-15 plug, type 498 (6.2122.070), 1.5 m
■■■■■■■■
20
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
Connecting the power supply cable
1
Inserting the power supply cable
Insert the power supply cable into the instrument's power socket. Connect the power supply cable to the power supply.
2
Switching on the instrument
Switch on the instrument using the power switch.
4 Installation
After being switched on, the LED on the front of the instrument flashes. The instrument conducts a system test and establishes a con­nection to the software. Once the system test is complete and the connection to the software has been established, the LED lights up continuously.

4.15 Initial start-up

Even before the guard column and separation column are installed, the entire system has to be completely rinsed with eluent for the first time.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
21
4.15 Initial start-up
■■■■■■■■■■■■■■■■■■■■■■
Rinsing the IC system
CAUTION
The guard column and separation column must not be installed for the initial start-up.
Make sure that a coupling (6.2744.040) is being used instead of the columns.
1
Preparing the software
Start the MagIC Net computer program. Open the Equilibration tab in MagIC Net: Work-
place Run Equilibration.
Select (or create) a suitable method.
Also see: MagIC Net tutorial and online help.
2
Preparing the instrument
Ensure that the eluent aspiration tubing is immersed in the eluent
and there is enough eluent in the eluent bottle.
Switch on the instrument.
MagIC Net detects the instrument and all of its modules.
3
Starting equilibration
Start the equilibration in MagIC Net: Workplace Run Equi-
libration Start HW.
■■■■■■■■
22
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
4
Deaerating the high-pressure pump
4 Installation
Use the syringe to aspirate eluent until there are no more air
bubbles in the eluent.
5
Rinsing the instrument without columns
Rinse the instrument (without columns) with eluent for 5 minutes.

4.16 Connecting and rinsing the guard column

CAUTION
New guard columns are filled with solution and sealed with stoppers or caps on both sides.
Before inserting the guard column, ensure that this solution can be mixed with the eluent being used (follow information from the manu­facturer).
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
23
4.16 Connecting and rinsing the guard column
NOTE
The guard column may not be connected until after the instrument has already been put into operation once (see Chapter 4.15, page 21). The guard column and separation column have to be replaced by a coupling (6.2744.040) until then.
Accessories For this step you need the following accessories:
Guard column (suitable for the separation column)
Connecting the guard column
■■■■■■■■■■■■■■■■■■■■■■
Rinsing the guard column
1
Rinsing the guard column
Place a beaker under the guard column's output. Start manual control in MagIC Net and select the high-pressure
pump: Manual Manual control Pump
Flow: in accordance with column leafletOn
Rinse the guard column with eluent for approx. 5 minutes. Use manual control in MagIC Net to stop the high-pressure pump
again: Off.
■■■■■■■■
24
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■

4.17 Connecting the separation column

CAUTION
New separation columns are filled with solution and sealed with stop­pers on both sides. Before inserting the column, ensure that this solu­tion can be mixed with the eluent being used (follow information from the manufacturer).
NOTE
The separation column may not be connected until after the instrument has already been put into operation once (see Chapter 4.15, page 21). The guard column and separation column have to be replaced by a coupling (6.2744.040) until then.
4 Installation
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
25
4.17 Connecting the separation column
■■■■■■■■■■■■■■■■■■■■■■
■■■■■■■■
26
Rinse the separation column as follows.
Connecting the separation column
1
Rinsing the separation column
Place a beaker under the separation column's outlet. Start manual control in MagIC Net and select the high-pressure
pump: Manual Manual control Pump
Flow: in accordance with column leafletOn
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
Rinse the separation column with eluent for approximately 10
Use manual control in MagIC Net to stop the high-pressure pump

4.18 Conditioning

After installation, after each time the instrument is switched on and after each time the eluent is changed, the system has to be conditioned with eluent long enough that a stable baseline is attained.
The conditioning time can lengthen considerably after changing the eluent.
Conditioning the system
4 Installation
minutes.
again: Off.
NOTE
1
Preparing the software
CAUTION
Ensure that the configured flow is not higher than the flow permit­ted for the corresponding column (refer to the column leaflet and the chip data record).
Start the MagIC Net computer program. Open the Equilibration tab in MagIC Net: Work-
place Run Equilibration.
Select (or create) a suitable method.
Also see: MagIC Net tutorial and online help.
2
Preparing the instrument
Ensure that the column is inserted correctly in relation to the flow
direction marked on the sticker (arrow has to point in the direc­tion of flow).
Ensure that the eluent aspiration tubing is immersed in the eluent
and there is enough eluent in the eluent bottle.
3
Starting equilibration
Start the equilibration in MagIC Net: Workplace Run Equi-
libration Start HW.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
27
4.18 Conditioning
■■■■■■■■■■■■■■■■■■■■■■
Visually inspect whether all capillaries and their connections from
the high-pressure pump to the detector are leak-tight. If eluent is leaking out anywhere, tighten the corresponding pressure screw further, or loosen the pressure screw, check the end of the capil­lary and shorten it using the capillary cutter if necessary and retighten the pressure screw.
4
Conditioning the system
Rinse the system with eluent until the required stability of the base­line is attained (normally 30 minutes).
The instrument is now ready for measuring samples.
■■■■■■■■
28
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
5 Operation

5 Operation

The 930 Compact IC Flex Oven/Deg is operated solely using the MagIC Net software. You can find information on operating the software in the tutorial for MagIC Net or in the online help.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
29

6 Warranty (guarantee)

Metrohm guarantees that the deliveries and services it provides are free of defects in materials, design or manufacturing.
The general warranty period is 36 months (exclusions below) from the date of delivery, or 18 months in the event of continuous operation. The warranty remains valid on the condition that the servicing is provided by a service organization authorized by Metrohm at defined intervals and with a defined scope.
The warranty period for anion suppressors of the type "MSM" is 120 months from the date of delivery or 60 months in the case of continuous operation.
The warranty period for IC separation columns is 90 days after start-up.
For third-party components that are recognizable as such, the manufac­turer's warranty regulations apply.
■■■■■■■■■■■■■■■■■■■■■■
For instruments sold under the Metrohm NIRSystems brand, a full 16­month warranty is applicable. In the event of continuous operation, the warranty period is reduced by half.
Consumables and materials with limited storage life and glass breakage in the case of electrodes or other glass parts are excluded from the warranty.
Warranty claims cannot be asserted if the ordering party has failed to meet its payment obligations according to schedule.
During the warranty period, Metrohm undertakes either to replace free of charge or to credit the purchaser for any modules or components that can be shown to be faulty. Any transport or customs fees that may apply are the responsibility of the ordering party.
The precondition for this is that the ordering party has to specify the arti­cle number, the article designation, an adequate error description, the delivery date and (if applicable) the serial number or chip data in the Sup­port Tracker. Metrohm then decides whether a replacement or a credit note is to be issued or whether the faulty part has to be returned using the Return Material Authorization (RMA). If a replacement or credit note is issued, the ordering party undertakes to store the faulty part for at least 24 months in accordance with the current storage directives (in compli­ance with ESD guidelines) and to hold it in readiness for onsite inspection or for return shipment to Metrohm. Metrohm reserves the right to invoice the ordering party for these articles, including retroactively, in the event of noncompliance with these preconditions.
■■■■■■■■
30
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
6 Warranty (guarantee)
The same warranty periods that are specified for a corresponding new part apply to parts that are replaced or repaired within the above-men­tioned warranty periods. However, replacement or repair of a part does not extend the warranty period of the entire system.
Deficiencies arising from circumstances that are not the responsibility of Metrohm, such as improper storage or improper use, etc., are expressly excluded from the warranty.
Metrohm also offers a 120-month spare parts availability guarantee and a 60-month PC software support warranty, calculated from the date on which the product is withdrawn from the market. The content of this war­ranty is the ability of the customer to obtain functioning spare parts or appropriate software support at market prices during the time of the war­ranty period. This does not apply for software products sold under the Metrohm NIRSystems brand.
If Metrohm AG is unable to meet this obligation due to circumstances beyond the control of Metrohm AG, then the ordering party shall be offered alternative solutions at preferential conditions.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
31

7 Accessories

Up-to-date information on the scope of delivery and optional accessories for your instrument can be found on the Internet.
When you receive your new instrument, we recommend downloading the accessories list from the Internet, printing it out and keeping it together with the manual for reference purposes.
Instruments currently sold
If you do not know the article number of your instrument, proceed as fol­lows:
Downloading the accessories list
■■■■■■■■■■■■■■■■■■■■■■
NOTE
Go to the Metrohm website http://www.metrohm.com/com.
1
2
Click on .
The Search webpage will be displayed.
Enter a search term relating to the instrument into the search field
3
and click on Find.
The search results will be displayed.
In the search results, select the Devices tab (if it is not already
4
selected) and then click on the Metrohm article number of the required instrument (e.g. 2.852.0050).
The page with information pertaining to the searched article is dis­played.
Select the Parts tab.
5
The complete list of accessories with the scope of delivery and the optional accessories will be displayed.
6
Click on .
■■■■■■■■
32
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■■■■■■■■■■■■■■■
7 Accessories
The Partslists webpage will be displayed.
Select the desired output language.
7
With the article number entered, click on the command Generate
8
PDF.
The PDF file with the accessories data will be created in the language selected.
Direct access for all instruments
If you are unable to find your instrument using the search as described above, this may be due to the instrument not being sold anymore. Using the article number, you can download accessories lists for all instruments as follows:
Downloading the accessories list
Type http://partslists.metrohm.com into your Internet browser.
1
The Partslists webpage will be displayed.
Select the desired output language.
2
Enter the article number and click on the Generate PDF command.
3
The PDF file with the accessories data will be created in the language selected.
930 Compact IC Flex Oven/Deg (2.930.2160)
■■■■■■■■
33

Index

Index
■■■■■■■■■■■■■■■■■■■■■■
A
Aspiration tubing for eluent ...... 16
B
Baseline
Conditioning ...................... 28
C
Column
See also "Separation column"
........................................... 25
Computer connection ............... 19
Conditioning ............................ 28
Connect
To computer ...................... 19
To power supply ................. 20
D
Degasser
Sample degasser ................. 18
Drainage tubing
Installation ......................... 13
E
Electrostatic charge .................... 5
Eluent
Aspirate .............................. 16
Eluent bottle
Installation ......................... 15
Equilibration ....................... 22, 27
F
Filter
See also "Inline filter" .......... 17
G
Guarantee ................................ 30
Guard column
Installation ......................... 23
Rinse .................................. 24
H
High-pressure pump
Protection .......................... 12
I
IC column
See also "Separation column"
........................................... 25
Injection valve ............................ 3
Installation ......................... 17
Inline filter ................................ 17
Installation
Drainage tubing ................. 13
Eluent bottle ....................... 15
Guard column .................... 23
Injection valve .................... 17
Leak sensor ........................ 14
Pulsation absorber .............. 17
Sample degasser ................. 18
Separation column ............. 25
L
Leak sensor
Installation ......................... 14
Leak-tightness .......................... 23
M
Mains voltage ............................. 5
P
Power connection .............. 20, 21
Power supply cable ................... 20
Pulsation absorber
Installation ......................... 17
R
Rinse
Guard column .................... 24
Separation column ............. 26
S
Safety instructions ...................... 4
Sample degasser
Installation ......................... 18
Separation column
Installation ......................... 25
Protection ...................... 3, 17
Rinse .................................. 26
Service ....................................... 4
Switch on ................................. 21
T
Transport locking screws .......... 11
V
Vacuum pump
Protection .......................... 12
Valve
See also "Injection valve" .... 17
W
Warranty .................................. 30
■■■■■■■■
34
930 Compact IC Flex Oven/Deg (2.930.2160)
Loading...