Metrohm 761 Compact IC User Manual

761 Compact IC
METROHM Ltd. CH-9101 Herisau
Switzerland Phone ++41 71 353 85 85 Fax ++41 71 353 89 01
8.761.1063
Instructions for Use
CH-9101 Herisau/Switzerland Phone ++41 71 353 85 85 Fax ++41 71 353 89 01
Internet www.metrohm.ch E-Mail info@metrohm.ch
761 Compact IC
Program «761 PC Software 1.1»
8.761.1063 Instructions for Use
07.07.2004 / chs
Table of contents
761 Compact IC
I
Table of contents
1 Introduction............................................................................................................1
1.1 Instrument description .......................................................................................... 1
1.2 Parts and controls .................................................................................................. 3
1.3 Information on the Instructions for Use ...........................................................10
1.3.1 Organization ......................................................................................... 10
1.3.2 Notation and pictograms...................................................................... 11
1.4 Safety notes........................................................................................................... 12
1.4.1 Electrical safety..................................................................................... 12
1.4.2 General precautionary rules ................................................................. 12
2 Installation .......................................................................................................... 13
2.1 Overview ................................................................................................................13
2.1.1 Flow chart ............................................................................................. 13
2.1.2 Connections in the 761 Compact IC.................................................... 13
2.2 Setting up the instrument.................................................................................... 15
2.2.1 Packaging............................................................................................. 15
2.2.2 Check.................................................................................................... 15
2.2.3 Location ................................................................................................ 15
2.3 Attaching the accessories ..................................................................................15
2.3.1 Connection of detector block............................................................... 15
2.3.2 Connection of syringe and aspirating tubing ....................................... 16
2.3.3 Connection of the drain tube for the inner compartment..................... 16
2.3.4 Connection of the drain tube for bottle rack ........................................ 17
2.3.5 Connection of PEEK capillaries............................................................ 17
2.3.6 Filter unit PEEK ..................................................................................... 18
2.4 Mains connection ................................................................................................. 19
2.4.1 Setting the mains voltage..................................................................... 19
2.4.2 Fuses .................................................................................................... 20
2.4.3 Mains cable and mains connection ..................................................... 20
2.4.4 On/off switching of the instrument ....................................................... 20
2.5 Connection to the PC........................................................................................... 21
2.5.1 Connecting cable ................................................................................. 21
2.5.2 Software installation.............................................................................. 21
2.5.3 Basic settings ....................................................................................... 22
2.6 High-pressure pump ............................................................................................ 25
2.6.1 Removing the transport security screws .............................................. 25
2.6.2 Installing the pulsation dampener ........................................................ 25
2.6.3 Connecting the eluent bottle ................................................................ 27
2.6.4 Deaerating the pump and rinsing the pulsation dampener................. 29
2.7 Precolumns and separating columns ............................................................... 31
2.7.1 General information on precolumns..................................................... 31
2.7.2 Precolumns with cartridge head........................................................... 31
2.7.3 Precolumn glass cartridges with cartridge holder ............................... 33
2.7.4 IC anion precolumn SUPERSEP .......................................................... 34
2.7.5 General information on separating columns........................................ 35
2.7.6 Selection of the sample loop................................................................ 35
2.7.7 Installation of the separating column without suppressor ...................36
2.7.8 Installation of the separating column with suppressor ........................ 37
Table of contents
761 Compact IC
II
2.8 Suppressor module ............................................................................................. 38
2.8.1 General information on suppressor module........................................ 38
2.8.2 Preparation of the peristaltic pump...................................................... 38
2.8.3 Connection of supply bottles ............................................................... 41
2.8.4 Connection of the suppressor module ................................................ 43
2.9 Putting into operation.......................................................................................... 45
2.9.1 Putting into operation without suppressor ........................................... 45
2.9.2 Putting into operation with suppressor ................................................ 46
2.10 Connection of external devices ......................................................................... 49
2.10.1 Connection of the 750 Autosampler .................................................... 49
2.10.2 Connection of the 766 IC Sample Processor ...................................... 50
2.10.3 Connection of other devices ................................................................ 52
3 Operating tutorial ....................................................................................... 53
3.1 Requirements........................................................................................................ 53
3.2 Preparations.......................................................................................................... 54
3.3 Calibration ............................................................................................................. 55
3.4 Sample determination ......................................................................................... 63
4 Operation............................................................................................................... 67
4.1 Fundamentals of the operation.......................................................................... 67
4.1.1 Starting/closing the program ............................................................... 67
4.1.2 Glossary................................................................................................ 68
4.1.3 Overview of program windows............................................................. 69
4.1.4 Main window elements......................................................................... 70
4.1.5 Icons of the main window..................................................................... 70
4.1.6 Overview of file types............................................................................ 71
4.1.7 Context sensitive menus ...................................................................... 72
4.1.8 Keyboard and mouse functions........................................................... 72
4.1.9 Help ...................................................................................................... 73
4.2 Instrument and software settings...................................................................... 74
4.2.1 Fonts..................................................................................................... 74
4.2.2 Security system .................................................................................... 74
4.2.3 Global settings ..................................................................................... 75
4.2.4 COM port.............................................................................................. 77
4.3 Systems ................................................................................................................. 78
4.3.1 System window .................................................................................... 78
4.3.2 System file handling ............................................................................. 78
4.3.3 System functions .................................................................................. 79
Connect and disconnect system ................................................. 79
Start/stop hardware and record baseline .................................... 79
Start/stop determinations............................................................. 80
Options for determinations........................................................... 80
4.3.4 System settings .................................................................................... 82
Modify system window ................................................................. 82
Watch window display.................................................................. 82
Set start mode .............................................................................. 82
Print system parameters .............................................................. 83
4.3.5 PC icon ................................................................................................. 83
Table of contents
761 Compact IC
III
Menu options for PC icon............................................................. 83
Select processing method and data source................................ 83
4.3.6 Watch window....................................................................................... 84
Menu options for screen icon....................................................... 84
4.3.7 Instrument icon ..................................................................................... 85
Menu options for instrument icon................................................. 85
System parameters for disconnected system ............................. 85
Instrument control for connected system .................................... 86
Time program ...............................................................................88
Configuration ................................................................................90
Hardware settings ........................................................................ 91
4.3.8 Timer ..................................................................................................... 94
4.3.9 System state window............................................................................ 95
Status messages.......................................................................... 95
Error messages ............................................................................ 96
4.4 Methods .................................................................................................................97
4.4.1 Method file handling ............................................................................. 97
4.4.2 Passport................................................................................................ 97
General ......................................................................................... 98
Sample.......................................................................................... 99
Column ....................................................................................... 100
Eluent .......................................................................................... 101
Comment .................................................................................... 101
Method Log ................................................................................ 102
Data Log ..................................................................................... 103
4.4.3 Method setup...................................................................................... 104
General ....................................................................................... 104
Measure ...................................................................................... 104
Filters .......................................................................................... 105
Processing.................................................................................. 106
Math ............................................................................................ 107
4.4.4 Integration........................................................................................... 109
Setup........................................................................................... 110
Events .........................................................................................112
4.4.5 Calibration and quantification............................................................. 117
General information .................................................................... 117
Notations..................................................................................... 118
External standard calibration...................................................... 119
Component table........................................................................ 119
Peak identification ...................................................................... 121
Concentration table .................................................................... 123
Calibration curve......................................................................... 125
Update calibration ...................................................................... 127
Load and save calibration data.................................................. 128
Import and export calibration data ............................................. 128
Put out calibration curves ........................................................... 128
4.4.6 Report output...................................................................................... 129
Report options ............................................................................ 129
4.5 Chromatograms..................................................................................................141
4.5.1 Chromatogram window ...................................................................... 141
4.5.2 Chromatogram file handling............................................................... 142
Open chromatogram.................................................................. 142
Save chromatogram................................................................... 143
Close chromatogram.................................................................. 143
Delete chromatogram................................................................. 143
Export chromatogram................................................................. 143
Import chromatogram ................................................................ 144
4.5.3 Graphical representation.................................................................... 145
Table of contents
761 Compact IC
IV
Appearance................................................................................ 145
Other graphical functions........................................................... 149
4.5.4 Peak editor.......................................................................................... 150
Switching on/off the peak editor ................................................ 150
Peak editor functions.................................................................. 150
Moving the cursor....................................................................... 151
4.5.5 Printing................................................................................................ 152
Page layout for printing .............................................................. 152
Printer settings............................................................................ 153
Print preview ............................................................................... 153
Printing........................................................................................ 153
4.5.6 Miscellaneous functions..................................................................... 153
Reintegration .............................................................................. 153
Recalibration............................................................................... 153
Subtraction of a chromatogram................................................. 154
Data compression ...................................................................... 154
Invert chromatogram.................................................................. 154
Autodatabase options................................................................ 155
Indicate active Autodatabase program ..................................... 155
4.6 Sample queue ..................................................................................................... 156
4.6.1 Sample queue file handling................................................................ 156
Open sample queue................................................................... 156
Save sample queue.................................................................... 156
Delete sample queue ................................................................. 156
4.6.2 Sample queue control ........................................................................ 157
Sample queue overview table .................................................... 157
Start sample queue .................................................................... 158
Pause sample queue ................................................................. 159
Cancel last run............................................................................ 159
Reset sample queue .................................................................. 159
4.6.3 Sample queue editor.......................................................................... 159
Open queue editor window........................................................ 159
Sample queue editor functions .................................................. 161
Print sample queue .................................................................... 161
Close sample queue editor ........................................................ 161
4.7 Batch reprocessing............................................................................................ 162
4.7.1 Batch reprocessing queue file handling ............................................ 162
Open batch reprocessing queue ............................................... 162
Create new batch reprocessing queue...................................... 162
Save batch reprocessing queue ................................................ 162
4.7.2 Perform batch reprocessing............................................................... 163
Reprocess options window........................................................ 163
Merge chromatograms .............................................................. 166
4.7.3 Batch reprocessing queue editor....................................................... 167
Open batch reprocessing queue editor window ....................... 167
Batch reprocessing queue editor functions............................... 168
Print batch reprocessing queue................................................. 168
Close batch reprocessing queue editor .................................... 168
Table of contents
761 Compact IC
V
5 Notes – Maintenance – Faults.................................................... 169
5.1 Practical notes on ion chromatography .........................................................169
5.1.1 Separating columns............................................................................ 169
5.1.2 High-pressure pump .......................................................................... 170
5.1.3 Eluents ................................................................................................ 171
5.1.4 Peristaltic pump.................................................................................. 171
5.1.5 Suppressor module ............................................................................ 172
5.1.6 Connections........................................................................................ 172
5.2 Maintenance and servicing...............................................................................173
5.2.1 General information ............................................................................ 173
5.2.2 Passivation.......................................................................................... 173
5.2.3 Recycling ............................................................................................ 174
5.2.4 Shutdown............................................................................................ 174
5.2.5 Changing separating columns........................................................... 174
5.2.6 Maintenance work at the pump head ................................................176
5.2.7 Regeneration of the suppressor module ...........................................180
5.2.8 Cleaning the suppressor .................................................................... 181
5.2.9 Replacing the suppressor .................................................................. 183
5.2.10 Exchanging the pump tubing ............................................................. 185
5.3 Faults and malfunctions....................................................................................186
5.3.1 Error messages ..................................................................................186
5.3.2 Malfunctions and their rectification..................................................... 186
5.4 Diagnostic tests / Validation / GLP..................................................................188
6 Appendix.............................................................................................................. 189
6.1 Technical data.....................................................................................................189
6.1.1 Conductivity measurement................................................................. 189
6.1.2 Conductivity detector.......................................................................... 189
6.1.3 Injection valve .....................................................................................190
6.1.4 High-pressure pump .......................................................................... 190
6.1.5 Peristaltic pump.................................................................................. 191
6.1.6 Suppressor module ............................................................................ 191
6.1.7 Leak detector...................................................................................... 191
6.1.8 RS232 interface .................................................................................. 191
6.1.9 Remote interface................................................................................. 192
6.1.10 Analog output .....................................................................................192
6.1.11 Mains connection ............................................................................... 192
6.1.12 Safety specifications........................................................................... 192
6.1.13 Electromagnetic compatibility (EMC)................................................. 193
6.1.14 Ambient temperature.......................................................................... 193
6.1.15 Housing............................................................................................... 193
6.2 Standard equipment...........................................................................................194
6.3 Optional accessories .........................................................................................198
6.3.1 Accessories for capillary connections................................................ 198
6.3.2 High-pressure pump .......................................................................... 199
6.3.3 Separating columns and precolumns................................................ 200
6.3.4 Additional devices and cables ...........................................................205
6.4 Warranty and conformity...................................................................................206
6.4.1 Warranty.............................................................................................. 206
6.4.2 EU Declaration of conformity.............................................................. 207
6.4.3 Certificate of conformity and system validation .................................208
6.5 Index .....................................................................................................................209
Table of contents
761 Compact IC
VI
List of figures
Fig. 1: Front of the 761 Compact IC .......................................................................... 3
Fig. 2
: Rear of the 761 Compact IC........................................................................... 4
Fig. 3
: Interior of the 2.761.0010 Compact IC ...........................................................6
Fig. 4
: Interior of the 2.761.0020 Compact IC ...........................................................8
Fig. 5
: Connecting diagram for 2.761.0010 Compact IC without suppressor........ 14
Fig. 6
: Connecting diagram for 2.761.0020 Compact IC with suppressor............. 14
Fig. 7
: Connectors for capillaries............................................................................. 17
Fig. 8
: 6.2821.000 Filter unit PEEK ..........................................................................18
Fig. 9
: Setting the mains voltage .............................................................................20
Fig. 10
: Connection of the pulsation dampener........................................................ 26
Fig. 11
: Connection of eluent bottle........................................................................... 28
Fig. 12
: Installation of precolumn cartridges with cartridge head............................. 32
Fig. 13
: Installation of precolumn glass cartridges with cartridge holder ................. 34
Fig. 14
: Installation of column without suppressor.................................................... 37
Fig. 15
: Installing pump tubings ................................................................................39
Fig. 16
: Connection of the separating column with suppressor ...............................40
Fig. 17
: Connection of supply bottles........................................................................ 42
Fig. 18
: Connections at suppressor module ............................................................. 43
Fig. 19
: Components of the pump head .................................................................177
Fig. 20
: Replacement of the piston seal 112 ..........................................................177
Fig. 21
: Components of inlet valve 113 and outlet valve 114................................. 179
Fig. 22
: Assembling the suppressor........................................................................ 183
Table of contents
761 Compact IC
VII
List of numbered parts and controls
1 Door to interior .................................3,7,9
2 Connection for syringe...........................3
3 Feedthrough for aspirating tubing ......... 3
4 Aspirating tubing..............................3,7,9
5 Feedthrough for capillaries .................... 3
6 Connection for drain tube .....................3
7 Feedthrough for capillaries .................... 3
8 Connection to purge valve ....................3
9 Mains pilot lamp.....................................3
10 Bottle rack ........................................... 3,5
11 Opening for detector cable....................5
12 Opening for inlet capillaries ...................5
13 Opening for outlet capillaries.................5
14 Connection for drain tube ...................... 5
15 Knurled screw ........................................5
16 Rear panel opening ...............................5
17 Detachable rear panel ...........................5
18 Transport security screws ...................... 5
19 Mains switch .....................................5,20
20 Mains connection plug .....................5,20
21 Fuse holder ....................................... 5,20
22 RS232 interface...................................... 5
23 Connection for detector block ...............5
24 Remote interface.................................... 5
25 Serial number.........................................5
26 Inlet capillary for injector ................7,9,26
27 Mounting rail .............................7,9,37,40
28 Column connection
capillary ...........................7,9,32,34,37,40
29 Sample loop........................................7,9
30 Connection capillary to syringe ..........7,9
31 Rotary nipple for aspirating tube ........7,9
32 Injection valve.......................7,9,26,37,40
33 PEEK coupling ............................... 7,9,40
34 Leak detector ......................................7,9
35 Connection capillary .....................7,9,26
36 Filter unit PEEK ....................7,9,18,26,40
37 Connection capillary ......................7,9,26
38 Connection capillary ...........................7,9
39 Purge valve.....................................7,9,26
40 Aspirating capillary..............................7,9
41 Fastening screws ................................7,9
42 Pump head...................................7,9,177
43 Connection capillary............................7,9
44 Connection capillary............................7,9
45 Inlet capillary for
detector block .......................... 7,9,37,40
46 Detector block .......................... 7,9,37,40
47 Suppressor module...........................9,40
48 Tubing cartridge ...........................9,39,41
49 Contact pressure lever .................9,39,41
50 Holding clamp...................................9,41
51 Snap-action lever .........................9,39,41
52 Pump drive ........................................9,41
53 Mounting pin .....................................9,41
54 Compression fitting .... 17,18,32,34,39,42
55 Compression fitting ..............................17
56 Capillary..................................... 17,18,34
57 Filter-Screw of Filter Unit .................18,39
58 Filter.................................................18,39
59 Filter-Housing of Filter Unit ..................18
60 Pulsation dampener .............................26
61 Connection to injection valve ...............26
62 Connection to purge valve ...................26
63 Aspirating tubing ..................................28
64 Tubing nipple........................................28
65 Threaded stopper.................................28
66 Bottle attachment .................................28
67 Eluent bottle..........................................28
68 Aspirating filter......................................28
69 CO
2
absorber .......................................28
70 Cotton wool ..........................................28
71 SGJ clip ................................................28
72 Absorber tube.......................................28
73 Manufit housing....................................32
74 Steel connector ....................................32
75 PTFE gasket .........................................32
76 2 Steel meshes.....................................32
77 Precolumn cartridge.............................32
78 Steel spacer .........................................32
Table of contents
761 Compact IC
VIII
79 4 Steel meshes ....................................32
80 Manufit pressure screw ....................... 32
81 Separating column .................... 32,37,40
82 End fitting............................................. 34
83 Screw cap for precolumn .................... 34
84 Sleeve for precolumn cartridge ...........34
85 Precolumn cartridge ............................34
86 Connection piece................................. 34
87 Screw cap for column.......................... 34
88 Column holder ................................37,40
89 Aspirating tubing for H
2
O........... 39,40,42
90 Aspirating tubing for H
2SO4
....... 39,40,42
91 Coupling ......................................... 39,40
92 Pump tubing for H
2SO4
................... 39,40
93 Pump tubing for H
2
O ......................39,40
94 Stopper ................................................ 39
95 PTFE tubing ....................................39,40
96 Suppressor inlet capillary
for eluent ......................................... 41,43
97 Suppressor outlet capillary
for eluent ......................................... 41,43
98 Suppressor inlet capillary for H
2
O ............................................41,43
99 Suppressor inlet capillary for H
2SO4
.................................... 39,41,43
100 Suppressor outlet capillary for H
2
O ............................................41,43
101 Suppressor outlet capillary for H
2SO4
......................................... 41,43
102 Bottle attachment.................................42
103 Supply bottle ........................................ 42
104 Screw..................................................177
105 Zircon piston ......................................177
106 Spring retainer.................................... 177
107 Spring.................................................177
108 Piston cartridge..................................177
109 Piston guide sleeve............................177
110 Sapphire supporting ring ................... 177
111 Piston guide sleeve............................177
112 Piston seal.......................................... 177
113 Inlet valve ....................................177,179
114 Outlet valve..................................177,179
115 Screw holder ......................................177
116 Special tool ........................................ 177
117 Special tool ........................................ 177
118 Valve housing..................................... 179
119 Sealing ring ........................................179
120 Sleeve................................................. 179
121 Sapphire sleeve .................................179
122 Sapphire sphere.................................179
123 Ceramic holder...................................179
124 Seal ....................................................179
125 Screw nut ...........................................183
126 Connection piece...............................183
127 Suppressor rotor ................................ 183
128 Suppressor holder .............................183
1.1 Instrument description
761 Compact IC
1
1 Introduction
1.1 Instrument description
The 761 Compact IC is a PC-controlled system for ion chroma­tographic analyses. The two following versions are available:
2.761.0010 Compact IC without
suppressor module
2.761.0020 Compact IC with
suppressor module
The extremely compact housing of the 761 Compact IC contains every­thing needed to carry out ion chromatography at the highest quality level:
Injection valve – for individual injections or for use with a sample
changer such as the Metrohm 766 IC Sample Processor
High-pressure pump – extremely low-pulsation double piston pump with a flow range from 0.2 2.5 mL/min and a maximum pressure of 25 MPa (250 bar)
Pulsation dampener – even with low-level pressure variations the pulsation dampener reliably protects the column against damage
Column chamber – the perfect insulation of the housing provides not only thermally stable conditions for the separation column but also shields the system against electromagnetic interference
Columns – whether anion columns with or without suppression, separation columns for cations or organic acids – the 761 Compact IC can accommodate them all
Suppressor – the Metrohm Suppressor Module (MSM) is already integrated in the 2.761.0020 Compact IC – pressure-resistant, with fully automatic regeneration, highest performance and optimal re­producibility
Peristaltic pump – integrated two-channel peristaltic pump with a flow rate of 0.5 0.6 mL/min for regeneration and rinsing of the suppressor module built into the 2.761.0020 Compact IC
Detector – conductivity detector with outstanding temperature stability. The detector temperature varies by less than 0.01°C and can be optimally adapted to the ambient conditions.
All components which come into contact with eluent and sample are metal-free.
1 Introduction
761 Compact IC
2
The operation of the 761 Compact IC takes place via a PC connected to the RS232 interface with the help of the control and evaluation pro­gram «761 Compact IC». This PC program can be used to create sys­tems for recording and evaluating chromatograms. Time programs can also be created in which a large number of instrument functions can be triggered for each program step. It is also possible to use programma­ble signals to control external instruments such as the 750 Autosampler or the 766 IC Sample Processor via the remote interface.
About 80 prepared system configurations for more than 300 applica­tions are already permanently stored and additional new applications can be downloaded at any time from the Internet under «www.metrohm.ch».
The operating software for the 761 Compact IC meets all the require­ments you could place today on a modern integration software: single or multi-point calibration, internal or external standard, selectable algo­rithms for non-linear calibration, various integration modes with integra­tion parameters and integration events, different methods for peak rec­ognition, peak editor, free scaling, superimposing several chromatograms, use of sample tables and batch reprocessing; a pow­erful and GLP-conform report generator with output interfaces for moni­tor, printer and external databases.
The independent «Autodatabase» PC programm supplied can be used to save and handle results and chromatograms produced by the «761 Compact IC» program in a database. With «Autodatabase», data can be sorted, filtered and searched with the help of different criteria. In ad­dition, data and curves can be printed out according to user-defined report templates.
1.2 Parts and controls
761 Compact IC
3
1.2 Parts and controls
POWER
761 Compact IC
89 76 5
2
3
4
1
10
Fig. 1
: Front of the 761 Compact IC
1
Door to interior 6 Connection for drain tube
for discharge of spilled liquid from the interior
2 Connection for 6.2816.020 Syringe
for aspiration of the sample
7 Feedthrough for capillaries
3 Feedthrough for aspirating tubing 8 Connection to purge valve
4 Aspirating tubing
for sample
9 Mains pilot lamp
lit up when instrument switched on
5 Feedthrough for capillaries 10 Bottle rack
for holding supply bottles with eluent, regeneration solution, and rinsing solution
1 Introduction
761 Compact IC
4
WARNING - Fire Hazard -
with the same type and rating of fuse
For continued protection replace only
Type 1.761.
Fuse 100-120V: 220-240V:
1,0A(T)
0,5A(T)
Made by Metrohm Herisau Switzerland
RS 232
Detector Block
Remote
Waste B
Waste A
Transport security scre ws
17
18
19
20
21
11
13 14 13 10
23a
22
15
25
24
12 12 11
15
16
15
15
f = 50-60 Hz
S = 100 VA
Analog Output
23
Fig. 2
: Rear of the 761 Compact IC
1.2 Parts and controls
761 Compact IC
5
10 Bottle rack
for holding supply bottles with eluent, regeneration solution, and rinsing solution
18 Transport security screws
to secure the pump head when the instrument is transported
11
Opening for detector cable 19 Mains switch
to switch instrument on and off: I = ON 0 = OFF
12 Opening for inlet capillaries
for supply of eluent, regeneration solution, and rinsing solution into the inner compartment
20 Mains connection plug
mains connection, see section 2.4
13 Opening for outlet capillaries
for discharge of eluent, regeneration solution, and rinsing solution from the inner compartment
21 Fuse holder
changing the fuses, see section 2.4
14 Connection for drain tube
for discharge of spilled liquid from the bottle rack
22 RS232 interface
connection of the PC
15 Knurled screw
for fastening the rear panel 17
23
23a
Connection for detector block
Analog output for measuring signal
16 Rear panel opening
(closed with plastic stopper) for additional supply and discharge lines to and from the inner compartment
24 Remote interface
remote I/O lines for connection of external devices
17 Detachable rear panel
access to upper part of inner compartment
25 Serial number
1 Introduction
761 Compact IC
6
37
44
1
26
27
27
28
29
30
31
32
43
42
41
40
39
38
36 35 34
4
33
45
46
Fig. 3
: Interior of the 2.761.0010 Compact IC
(with permanently attached accessories and
1.733.0110 Detector block)
1.2 Parts and controls
761 Compact IC
7
1 Door to interior 36 Filter unit PEEK (6.2821.120)
4
Aspirating tubing
for aspirating the sample;
6.1803.020 PTFE tubing, length L = 52 cm
37 Connection capillary
6.1831.010 PEEK capillary, length L = 13 cm
26 Inlet capillary for injector
6.1831.010 PEEK capillary, length L = 24 cm
38 Connection capillary
6.1831.010 PEEK capillary, length L = 15 cm
27 Mounting rail
for 6.2027.0X0 column holder
39 Purge valve
28 Column connection capillary
6.1831.010 PEEK capillary, length L = 30 cm
40 Aspirating capillary
Connection for 6.1834.010 aspirating tubing
29 Sample loop 10 µL
6.1825.230 PEEK sample loop
41 Fastening screws
for pump head 42
30 Connection capillary to syringe
6.1803.020 PTFE tubing, length L = 30 cm
42 Pump head (6.2824.100)
31 Rotary nipple for aspirating tube
for fixing the aspirating tube
43 Connection capillary
Connection pump head – purge valve, fixed mounting
32 Injection valve
44 Connection capillary
in pump head, fixed mounting
33 PEEK coupling (6.2744.040)
45 Inlet capillary for detector block
PEEK capillary, fixed mounting
34 Leak detector 46 Detector block (1.732.0110)
35 Connection capillary
6.1831.010 PEEK capillary, length L = 13 cm
1 Introduction
761 Compact IC
8
37
44
1
26
27
27
28
29
30
31
32
43
42
41
40
39
38
36 35 34
4
33
45
46
47
48
49
53 52
5051
Fig. 4
: Interior of the 2.761.0020 Compact IC
(with permanently attached accessories and
1.733.0110 Detector block)
1.2 Parts and controls
761 Compact IC
9
1 Door to interior 39 Purge valve
4
Aspirating tubing
for aspirating the sample;
6.1803.020 PTFE tubing, length L = 52 cm
40 Aspirating capillary
Connection for 6.1834.010 aspirating tubing
26 Inlet capillary for injector
6.1831.010 PEEK capillary, length L = 24 cm
41 Fastening screws
for pump head 42
27 Mounting rail
for 6.2027.0X0 column holder
42 Pump head (6.2824.100)
28 Column connection capillary
6.1831.010 PEEK capillary, length L = 30 cm
43 Connection capillary
Connection pump head – purge valve, fixed mounting
29 Sample loop 20 µL
6.1825.210 PEEK sample loop
44 Connection capillary
in pump head, fixed mounting
30 Connection capillary to syringe
6.1803.020 PTFE tubing, length L = 30 cm
45 Inlet capillary for detector block
PEEK capillary, fixed mounting
31 Rotary nipple for aspirating tube
for fixing the aspirating tube
46 Detector block (1.732.0110)
32 Injection valve
47 Suppressor module
(inlet and outlet capillaries are not shown)
33 PEEK coupling (6.2744.040)
48 Tubing cartridge (6.2755.000)
for 6.1826.060 pump tubing
34 Leak detector 49 Contact pressure lever
for adjusting the contact pressure
35 Connection capillary
6.1831.010 PEEK capillary, length L = 13 cm
50 Holding clamp
for locking the tubing cartridge into place
36 Filter unit PEEK (6.2821.120) 51 Snap-action lever
for releasing the tubing cartridge
37 Connection capillary
6.1831.010 PEEK capillary, length L = 13 cm
52 Pump drive
roller head with contact rollers
38 Connection capillary
6.1831.030 PEEK capillary, length L = 20 cm
53 Mounting pin
for attaching the tubing cartridges
1 Introduction
761 Compact IC
10
1.3 Information on the Instructions for Use
Please read through these Instructions for Use carefully before you put the 761 Compact IC into operation. The Instructions for Use contain information and warnings to which the user must pay attention in order to assure safe operation of the instrument.
1.3.1 Organization
These 8.761.1063 Instructions for Use for the 761 Compact IC pro- vide a comprehensive overview of the installation, startup procedure, operation, fault rectification and technical specifications of this instru­ment. The Instructions for Use are organized as follows:
Section 1 Introduction
General description of instrument, parts and controls and safety notes
Section 2 Installation
Installation of instrument, accessories, and external devices
Section 3 Operating tutorial
Introduction to the operation using an example
Section 4 Operation
Detailed description of the operation
Section 5 Notes – Maintenance – Faults
Notes on ion chromatography, maintenance, fault rectifi­cation, diagnostic tests, validation
Section 6 Appendix
Technical data, standard equipment, options, warranty, declarations of conformity, index
To find the required information on the instruments, you will find it an advantage to use either the Table of contents or the Index at the back.
As a supplement to the Instructions for Use, the Metrohm Monograph
8.732.2003 "Ion chromatography" is also supplied. This provides an introduction to the theoretical fundamentals and general information on separating columns and sample pretreatment.
The 8.732.2013 IC Applications Collection is also supplied; this con­tains all the Application Notes on the subject of ion chromatography. Each of these applications can be carried out directly with the 761 Compact IC by loading the system file with the same name. The Appli­cations Collection can be updated at any time by downloading the lat­est applications from the Internet under «www.metrohm.ch».
You will find detailed information on the separating columns available from Metrohm and on special IC applications in the relevant "Applica- tion Bulletins", which are available on request free of charge from your Metrohm agency.
1.3 Information on the Instructions for Use
761 Compact IC
11
1.3.2 Notation and pictograms
The following notations and pictograms (symbols) are used in these In­structions for Use:
Range Menu item, parameter or entry
value
SYSTEM STATE Program window
<OK> Button
[ Ctrl ] Key
35 Part or control of 761
12 Part or control of 750
26 Part or control of 766
Hazard This symbol draws attention to a possible danger to life or of injury if the associated directions are not followed correctly.
Warning This symbol draws attention to possible damage to instruments or instrument parts if the associated directions are not followed correct­ly.
Caution This symbol marks important information. First read the associ­ated directions before you conti­nue.
Comment This symbol marks additional information and tips.
1 Introduction
761 Compact IC
12
1.4 Safety notes
1.4.1 Electrical safety
While electrical safety in the handling of the 761 Compact IC is assured in the context of the specifications IEC 1010-1 (protection class 1, de­gree of protection IP20), the following points should be noted:
Mains connection
Setting of the mains voltage, checking the mains fuse and the mains connection must be effected in accordance with the instruc-
tions in section 2.4.
Opening the 761 Compact IC
If the 761 Compact IC is connected to the power supply, the instru­ment must not be opened nor must parts be removed from it, other­wise there is a danger of coming into contact with components which are live. Hence, always disconnect the instrument from all voltage sources before you open it and ensure that the mains cable is disconnected from mains connection 20 !
Protection against static charges
Electronic components are sensitive to static charging and can be destroyed by discharges. Before you touch any of the components inside the 761 Compact IC, you should earth yourself and any tools you are using by touching an earthed object (e.g. housing of the instrument or a radiator) to eliminate any static charges which exist.
1.4.2 General precautionary rules
Handling of solvents
Check all lines of the IC system periodically for possible leaks. Follow the relevant instructions regarding the handling of flammable and/or toxic solvents and their disposal.
2.1 Overview
761 Compact IC
13
2 Installation
2.1 Overview
2.1.1 Flow chart
The following flow chart provides an overview of all installation work. You will find more detailed information in the relevant sections.
Setting up sect. 2.2
Mains connection sect. 2.4
Connecting PC sect. 2.5
Installing accessories sect. 2.3
Suppressor Connect ing suppressor module sect. 2.8
Installing sample loop sect. 2.7.6
Conditioning sect. 2.9
Connecting external devices sect. 2.10
Precolumn
Precolumn with cartridge head sect. 2.7.2 Precolumn with cartridge holder sect. 2.7.3 IC anion precolumn SUPERSEP sect. 2.7.4
Yes
No
Connecting high pr. pump sect 2.6
Connecting column sect. 2.7.7/8
Yes
No
2.1.2 Connections in the 761 Compact IC
The two following illustrations show the internal connections in the 761 Compact IC in schematic form. The meanings of the various numbered components are given in the detailed illustrations and descriptions in sections 2.2 – 2.10.
2 Installation
761 Compact IC
14
63
40
42
44 43
28 81 45 46
36 35
37
39
66
67
68
38
60
26 29 32
30
4
Fig. 5
: Connecting diagram for 2.761.0010 Compact IC without suppressor
63
40
42
44 43
28 81 96
47
97 33 45 46
36 35
37
39
66
67
68
38
60
26 29 32
99
100
98
4
30
36
92
103
102
89
90
103
102
9191 93 95
101
Fig. 6
: Connecting diagram for 2.761.0020 Compact IC with suppressor
Sample
Eluent
Injection valve
Column
Detector
Waste
Eluent
Column
Detector
Sample
H
2
OH2SO
4
Waste
2.2 Setting up the instrument
761 Compact IC
15
2.2 Setting up the instrument
2.2.1 Packaging
The 761 Compact IC is supplied together with the separately packed accessories in special packagings containing shock-absorbing foam linings designed to provide excellent protection. The instrument itself is packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all these special packagings as only they assure transport of the instrument free from damage.
2.2.2 Check
After receipt, immediately check whether the shipment is complete and has arrived without damage (compare with delivery note and list of accessories in section 6.2). In the case of transport damage, see instructions in section 6.4.1 "Warranty".
2.2.3 Location
Position the instrument in the laboratory at a location convenient for op­eration, free from vibrations and protected against a corrosive atmos­phere and contamination by chemicals.
To avoid disturbing temperature influences on the insulated column compartment, the instrument must be protected against direct sunlight.
2.3 Attaching the accessories
2.3.1 Connection of detector block
The metal-free 1.732.0110 Detector block belongs to the scope of supply of the 761 Compact IC; it must be inserted in the instrument and connected up. Proceed as follows:
1 Note the cell constant
The cell constant c = XX,X /cm is printed on the rear of the
detector block. Note this value; it must subsequently be en­tered in the software in order to ensure that an exact display of the conductivity is obtained (see section 2.5.3).
2 Install detector block
Unscrew the four knurled screws 15 from the top rear panel
17 of the 761 Compact IC and remove rear panel (see Fig. 2).
Position detector block 46 from the back in the space pro-
vided in the 761 Compact IC and push fully to the front (see Fig. 3 and Fig. 4).
2 Installation
761 Compact IC
16
Insert the cable permanently attached to the detector block
46 in one of the openings 11 and the outlet capillary in one of the openings 13 of the rear panel 17.
Replace rear panel 17 and screw to the 761 Compact IC
using the four knurled screws 15.
3 Connect detector block
Plug the gray connecting cable permanently attached to the
detector block 46 into connection 23 "Detector Block" of the 761 Compact IC and fasten to the instrument by tightening the screws in the cable connector (see Fig. 2).
4 Connect waste container
Lead the outlet capillary of the detector block 46 to a suffi-
ciently large waste container and fix in place.
2.3.2 Connection of syringe and aspirating tubing
For manual filling of the sample loop 29 mounted on the injection valve, the 6.2816.020 Syringe and the PTFE aspirating tubing 4 already screwed to the valve are needed. These accessories are mounted or adjusted as follows:
1 Connect syringe
Push 6.2816.020 Syringe (without needle) as far as it will go
into connection socket 2 (see Fig. 1).
2 Adjust aspirating tubing
Loosen the rotary nipple 31 screwed onto the interior side of
feedthrough 3 (see Fig. 3 and Fig. 4).
Pull PTFE aspirating tubing 4 (see Fig. 3 and Fig. 4) by hand
out of feedthrough 3 as far as desired.
Retighten rotary nipple 31 on the interior side of feedthrough
3 to fix the aspirating tubing 4 in place.
2.3.3 Connection of the drain tube for the inner compartment
The 761 Compact IC has a connection at the front to which a drain tube for discharged liquids in the inner compartment can be attached. Pro­ceed as follows:
1 Connect drain tube
Mount 6.1816.020 Silicone tubing on connection nipple 6
(see Fig. 1).
2 Lead drain tube to collecting vessel
Lead the other end of the drain tube to a suitable collecting
vessel and fix in place.
2.3 Attaching the accessories
761 Compact IC
17
2.3.4 Connection of the drain tube for bottle rack
The 761 Compact IC has a connection at the rear to which a drain tube for discharged liquids in the bottle rack can be attached. Proceed as follows:
1 Connect drain tube
Mount 6.1816.020 Silicone tubing on connection nipple 14
(see Fig. 2).
2 Lead drain tube to collecting vessel
Lead the other end of the drain tube to a suitable collecting
vessel and fix in place.
2.3.5 Connection of PEEK capillaries
For the connections between high-pressure pump and detector block
6.1831.010 PEEK capillaries (i.d. = 0.25 mm, e.d. =
1
/16") are used
which are connected using either 6.2744.010 PEEK compression fit- tings (long) or 6.2744.070 PEEK compression fittings (short). These PEEK connectors can also be used to connect 6.1822.010 PTFE microcapillaries (i.d. = 0.3 mm). Proceed as follows:
Capillaries fitted with new connectors must have a perfectly flat cut surface. To cut PEEK or PTFE capillaries it is best to use the
6.2621.080 Capillary tubing cutter.
1 Mount compression fitting
Slide a compression fitting 54 (6.2744.010) or a compression fitting 55 (6.2744.070) over the end of the capillary 56 to be fastened as shown in Fig. 7.
2 Insert capillary in connection
Push capillary end in the corresponding connection as far as it will go (to avoid dead volume).
3 Tighten compression fitting
Tighten compression fitting 54 or 55 by hand (never use tools).
54 Compression fitting (6.2744.010)
55 Compression fitting (6.2744.070)
56 Capillary
6.1831.010 PEEK capillary or
6.1822.010 PTFE microcapillary
54 56 55
Fig. 7
: Connectors for capillaries
2 Installation
761 Compact IC
18
2.3.6 Filter unit PEEK
One 6.2821.120 Filter unit PEEK (see Fig. 8) is already installed be- tween the high-pressure pump and the injection valve at the 761 Com­pact IC. This filter unit serves to avoid contamination by abrasive parti­cles of the piston seals.
The two other filter units PEEK supplied with the 2.761.0020 Compact IC (with suppressor) are installed between the pump tubings of the peristaltic pump and the inlet capillaries for regeneration and rinsing so­lution (see section 2.8.2). These filter units serve to protect the suppres­sor module from foreign particles and bacterial growth.
The Filter unit PEEK 36 consists of the filter-housing 59, the filter-screw 57 and the 6.2821.130 filter 58. For the connection of capillaries 56 PEEK compression fittings 54 (6.2744.010 or 6.2744.070) must be used. New filters 58 are available as an option with the ordering number
6.2821.130 (set of 10).
For the connection of the filter unit, please note the flow direction arrow printed on the housing.
56 54
57
58
59 54 56
36
Fig. 8
: 6.2821.120 Filter unit PEEK
36 Filter unit PEEK (6.2821.120) 57 Filter-Screw of Filter Unit
Part of 6.2821.120 Filter unit
54 Compression fitting (6.2744.010) 58 Filter 6.2821.130
Part of 6.2821.120 Filter unit
56 Capillary
6.1831.010 PEEK capillary or
6.1822.010 PTFE microcapillary
59 Filter-Housing of Filter Unit
Part of 6.2821.120 Filter unit
2.4 Mains connection
761 Compact IC
19
2.4 Mains connection
Follow the instructions below for connecting to the power supply. If the instrument is operated with a mains voltage set wrongly and/or wrong mains fuse, there is a danger of fire!
2.4.1 Setting the mains voltage
Before switching on the 761 Compact IC for the first time, check that the mains voltage set on the instrument (see Fig. 9) matches the local mains voltage. If this is not
the case, you must reset the mains voltage
on the instrument as follows:
1 Disconnect mains cable
Disconnect mains cable from mains connection plug 20 of the 761 Compact IC.
2 Remove fuse holder
Using a screwdriver, loosen fuse holder 21 below the mains connection plug 20 and take out completely.
3 Check and change fuse if necessary
Carefully take the fuse installed for the desired mains voltage out of fuse holder 21 and check its specifications (the position of the fuse in the fuse holder is marked by the white arrow imprinted next to the mains voltage range):
100120 V 1.0 A (slow-blow) Metrohm No. U.600.0016 220240 V 0.5 A (slow-blow) Metrohm No. U.600.0013
4 Insert fuse
Change fuse if necessary and reinsert in fuse holder 21.
5 Install fuse holder
Depending on the desired mains voltage, insert fuse holder 21 in the 761 Compact IC so that the corresponding mains voltage range can be read normally and the adjacent white arrow points to the white bar imprinted below the fuse holder (see Fig. 9).
2 Installation
761 Compact IC
20
19
220 - 240 V
100 - 120 V
20
21
220 - 240 V
100 - 120 V
Fig. 9
: Setting the mains voltage
2.4.2 Fuses
One of the two fuses 1 A/slow-blow for 100120 V or 0.5 A/slow-blow for 220240 V is installed in fuse holder 21 of the 761 Compact IC as standard.
Ensure that the instrument is never put into operation with fuses of another type, otherwise there is danger of fire!
For checking or changing fuses, process as described in section 2.4.1.
2.4.3 Mains cable and mains connection
Mains cable
The instrument is supplied with one of three mains cables
6.2122.020 with plug SEV 12 (Switzerland, …)
6.2122.040 with plug CEE(7), VII (Germany, …)
6.2133.070 with plug NEMA 5-15 (USA, …)
which are three-cored and fitted with a plug with an earthing pin. If a dif­ferent plug has to be fitted, the yellow/green lead (IEC standard) must be connected to protective earth (protection class 1).
Any break in the earthing inside or outside the instrument can make it a hazard!
Mains connection
Plug the mains cable into mains connection plug 20 of the 761 Com­pact IC (see Fig. 9).
2.4.4 On/off switching of the instrument
The 761 Compact IC is switched on and off using mains switch 19. When the instrument is switched on, the mains pilot lamp 9 lights up.
100 – 120 V 220 – 240 V
19 Mains switch
20 Mains connec-
tion plug
21 Fuse holder
2.5 Connection to the PC
761 Compact IC
21
2.5 Connection to the PC
2.5.1 Connecting cable
Always switch off 761 Compact IC and PC before you connect the two instruments with the 6.2134.100 Cable.
Connect the RS232 interface 22 at the 761 Compact IC to one of the serial COM ports at the PC using the 6.2134.100 Cable (9 pin/9 pin). If only a 25-pin COM interface is available on the PC then the 6.2125.110 Adapter cable or a commercially available adapter must be used.
2.5.2 Software installation
These Instructions for Use describe the operation of a single 761 Compact IC connected to a PC. If several instruments should be operated simultaneously with one PC, the «IC Net 2.0» PC program must be installed for this purpose (details see Instructions for Use for «IC Net»).
The PC program «761 Compact IC 1.1» is required for the operation of the 761 Compact IC; this is contained on the 6.6030.013 CD included in the accessories. This program runs under Windows 95, Windows 98 and Windows NT operating systems and is installed as follows:
1 Install program
Insert 6.6030.013 Installation CD into CD drive.
Select
<Start> and Run. Browse for the setup.exe file on the
installation CD and click on
<OK>. Follow the instructions
given in the setup program.
The software package will be installed in the desired direc-
tory. Icons are created in the program folder and in the startup folder. In addition to the program files, the following folders are installed:
Data Folder for storage of chromatogram files
(
*.chw) and batch reprocessing files (*.bar) Devices Folder for storage of device files (*.dev) Methods Folder for storage of method files (*.mtw) Reports Folder for storage of report files (*.txt)
and graphic files (
*.wmf)
Systems Folder with subfolders with system files
(
*.smt) and sample queue files (*.que).
2 Registration
Please send us your 8.761.8007 Registration card as soon as
possible. Only registered users will get updated program ver­sions at a special price.
2 Installation
761 Compact IC
22
The installed files (incl. system and method files) are generally not write-protected. To prevent these files from being deleted by mistake, switch on the write-protection or make a backup copy in another directory.
2.5.3 Basic settings
When the program is started for the first time several basic settings must be made for the 761 Compact IC. Proceed as follows:
1 Start program
Double-click the software icon to start the program. The
program window with the opening picture is opened and the Log In window appears on the screen:
Do not
enter any password here, just click on <Log In>. The
following window appears:
Enter company, division and name and click on
<OK>. This
window appears only one time after software installation.
2 COM port settings
This step must only be carried out if a different COM interface from COM1 is used for connection to the 761 Compact IC.
2.5 Connection to the PC
761 Compact IC
23
Click on Options / 761 Compact IC:COM1 to open the Links
window:
Click on
COM1 using the right mouse button and select the
Change menu item to open the following window listing all
available COM ports at the PC:
Select the desired COM port to which the 761 Compact IC
has been connected and click on
<OK>. The window is
closed.
Close the
Links window by clicking on <OK>.
3 Open a system
Click on File / Open / System in the main window. Open either
the subfolder
Non-suppressed (for 2.761.0010 Compact IC
without
suppressor) or the subfolder Suppressed (for
2.761.0020 Compact IC with
suppressor). Select the system
file
n-01.smt or s-01.smt and click on <Open>. The correspond-
ing system window is opened:
2 Installation
761 Compact IC
24
Select the Connect to workplace menu item of the Control menu
in this window.
If the connection between PC and Compact IC is working the
message
Hardware settings file for 761 unit with serial number
'#####' not found! Create?
will appear. Click <Yes> in order to
create the configuration file
#####.761 for this instrument.
If the connection between PC and Compact IC does not work
then the message
Detection of hardware failed[761 Compact IC
[ COM# ]]
appears in the SYSTEM STATE window. In this case check whether the instrument has been switched on, whether the connection cable is connected up properly and whether the COM interface has been set correctly (see point 2). Then repeat point 3.
4 Hardware settings
Of the general hardware settings only the input of the cell constant is described. Standard settings can normally be used for all other parameters.
Click the 761 icon using the right mouse button and select
the
Hardware item. The Hardware settings window is opened:
In the
Cell constant field enter the cell constant which is printed
on the 1.732.0110 Detector block (see section 2.3.1).
Click on
<OK> to close the window and save the settings.
2.6 High-pressure pump
761 Compact IC
25
2.6 High-pressure pump
In order to avoid damage to the pump it must never be operated dry. Each time that the pump is switched on always first check that the eluent supply has been connected up correctly and that sufficient eluent is present in the eluent bottle.
2.6.1 Removing the transport security screws
In order to prevent the pump drive from being damaged during trans­port the pump head is fitted with three transport security screws 18 (see Fig. 2). These transport security screws must be removed before the high-pressure pump is started up. Also remove the red sticker attached to the pump head.
In order to avoid damage to the pump head these three security screws should be attached to the pump head each time that it is to be transported.
2.6.2 Installing the pulsation dampener
To protect the column material against pressure drops caused by the injector, the 6.2620.150 Pulsation dampener MF has to be installed between the high-pressure pump and the injection valve of the 761 Compact IC. Proceed as follows (see Fig. 10):
1 Install pulsation dampener
Position the pulsation dampener 60 in the interior of the 761
Compact IC on the base.
2 Connection to the pump
Unscrew PEEK capillary 35 of coupling 33 and attach it to
connection 62 of the pulsation dampener 60.
3 Connection to injection valve
Unscrew PEEK capillary 26 of coupling 33 and attach it to
connection 61 of the pulsation dampener 60.
The pulsation dampener is filled with isopropanol and must be rinsed with eluent before connection to a separating column (see section
2.6.4).
The 6.2620.150 Pulsation dampener can be operated in both direc­tions.
2 Installation
761 Compact IC
26
37
39
36 35 62
32
26
6061
Fig. 10
: Connection of the pulsation dampener
26
Inlet capillary for injector
6.1831.010 PEEK capillary, length L = 24 cm
39 Purge valve
32 Injection valve 60 Pulsation dampener (6.2620.150)
35 Connection capillary
6.1831.010 PEEK capillary, length L = 13 cm
61 Connection to injection valve
36 Filter unit PEEK (6.2821.120) 62 Connection to purge valve
37 Connection capillary
6.1831.010 PEEK capillary, length L = 13 cm
2.6 High-pressure pump
761 Compact IC
27
2.6.3 Connecting the eluent bottle
The eluent supply line from the storage bottle to the high-pressure pump is connected as follows (see Fig. 11):
Only degassed (with N
2
, He or vacuum) and microfiltered (0.45 µm
filter) eluents should be used!
The 6.1608.070 Eluent bottle (2 L) supplied is not suitable for vacuum degassing. Use a pressure-resistant container for this.
Care must be taken that the eluent used is freely miscible with any solvent remaining in the pump head (the pump head leaves the factory filled with either isopropanol or methanol/water). If this is not the case then the pump must first be rinsed with a solvent which is miscible with both the previous eluent and the following eluent (e.g. acetone).
1 Prepare eluent bottle
Prepare, microfilter (0.45 µm microfilter) and degas (with N2,
He, or vacuum) the suitable eluent for the required application and separating column.
Fill eluent into eluent vessel 67 (clear glass, 2 L).
Place eluent bottle 67 at the front in bottle holder 10 on the
761 Compact IC (see Fig. 1).
2 Install bottle attachment
Firmly screw threaded stopper 65 (6.1446.040; part of
6.1602.160) into the smaller threaded opening (M6) of bottle attachment 66 (6.1602.105; part of 6.1602.160).
Firmly screw aspirating filter 68 onto aspirating tubing 63.
Pull the other end of aspirating tubing 63 through the larger
threaded opening (M8) of bottle attachment 66 from below.
Push O-ring (E.301.0021; part of 6.1602.160) over the free
end of aspirating tubing 63 and move it towards bottle at­tachment 66.
Push tubing nipple 64 (4.420.4300; part of 6.1602.160) over
the free end of aspirating tubing 63, move it as far as re­quired towards bottle attachment 66 and screw it loosely in the larger opening of bottle attachment 66.
Insert aspirating tubing 63 with screwed-on aspiration filter 68
into eluent bottle 67 and screw bottle attachment 66 onto eluent bottle 67.
Pull aspirating tubing 63 so far through the opening of tubing
nipple 64 that aspirating filter 68 is touching the bottom of eluent bottle 67.
Fix aspirating tubing 63 in place by screwing shut tubing
nipple 64.
2 Installation
761 Compact IC
28
3 Mount CO2 absorber tube
First place a piece of cotton wool 70 followed by CO2 ab-
sorber 69 (e.g. Merck soda-lime pellets with indicator, no.
6839.1000) in the large opening of absorber tube 72 and then close it with the plastic lid.
Fasten absorber tube 72 to bottle attachment 66 with the aid
of SGJ clip 71.
4 Connect aspirating tubing to pump
Insert the free end of aspirating tubing 63 into one of the
openings 12 in the interior of the 761 Compact IC (see Fig. 2).
Pull aspirating tubing 63 sufficiently far into the interior of the
761 Compact IC, cut off to the required length and push at least 5 mm of it onto aspirating capillary 40 (see Fig. 3 and Fig. 4) of the high-pressure pump (it may be necessary to use emery paper).
If necessary, fix aspirating tubing 63 in the required position
in the interior with the aid of the Y.107.0150 self-adhesive strap.
Instead of threaded stopper
65
, the 4.420.0311 Tubing nipple (M6) which belongs to the 6.1602.160 Bottle attachment accessories can be used together with the included second E.301.0021 O-ring for recycling operation (see section 5.2.3), in which the outlet capillary from the detector block is led back into eluent bottle
67
.
63 64
65
66
72
68
71
70
69
67
Fig. 11
: Connection of eluent bottle
63 Aspirating tubing (6.1834.010)
64 Tubing nipple (4.420.4300; M8)
with E.301.0021 O-ring
65 Threaded stopper (6.1446.040; M6)
66 Bottle attachment (6.1602.105)
67 Eluent bottle (6.1608.070)
68 Aspirating filter (6.2821.090)
69 CO2 absorber
70 Cotton wool
71 SGJ clip (6.2023.020)
72 Absorber tube (6.1609.000)
2.6 High-pressure pump
761 Compact IC
29
2.6.4 Deaerating the pump and rinsing the pulsation dampener
The first time that it is started up the high-pressure pump must be deaerated. Proceed as follows:
1 Prepare for deaeration
Open the rotary knob on purge valve 39 by approx. ½ turn in
the counterclockwise direction (see Fig. 3 and Fig. 4).
Remove the plastic stopper from connection 8 on the front
panel of 761 Compact IC (see Fig. 1).
Push 6.2816.020 Syringe (without needle) into connection 8
until the stop is reached.
2 Open and connect system
Start the «761 Compact IC» PC program, if it has not been
already been started (see section 2.5.3).
Select
File / Open / System in the main window. Open either the
subfolder
Non-suppressed (for 2.761.0010 Compact IC without
suppressor) or the subfolder
Suppressed (for 2.761.0020
Compact IC with
suppressor). Select the system file n-01.smt
or
s-01.smt and click on <Open>. The corresponding system
window is opened:
Select the
Connect to workplace item of the Control menu in this
window.
3 Set flow rate to 2 mL/min
Double-click the 761 icon in the system window to open the
window for manual control of the 761 Compact IC (see be­low).
Set the flow rate to
2 mL/min in the Flow field.
Click to
<Send to unit> to send this value to the 761 Compact
IC.
2 Installation
761 Compact IC
30
4 Deaerate pump
Make sure that the aspirating tubing 63 for the high-pressure
pump has been immersed into the eluent.
Click the
<On> button for IC pump to switch on the high-
pressure pump.
Use the syringe inserted into connection 8 to aspirate air until
eluent flows into the syringe.
Click the
<Off> button for IC pump to switch off the high-
pressure pump.
Close the rotary knob on purge valve 39 by turning it in a
clockwise direction (see Fig. 3 and Fig. 4).
Remove the syringe from connection 8.
5 Rinse pulsation dampener
Place a beaker beneath the column connection capillary 28.
Click the
<On> button for IC pump to switch on the high-
pressure pump and rinse the pulsation dampener 60 filled with isopropanol for ca. 10 min with eluent.
Click the
<Off> button for IC pump to switch off the high-
pressure pump.
6 Reduce flow rate
Reset the original flow rate under Flow (e.g. 0.5 mL/min).
Click
<Send to unit> to send this value to the 761 Compact IC.
2.7 Precolumns and separating columns
761 Compact IC
31
2.7 Precolumns and separating columns
2.7.1 General information on precolumns
The use of easily exchangeable precolumns protects the separating columns and prolongs their lifetime. The precolumns available from Metrohm (see section 6.3.2) are either real precolumns or precolumn cartridges, which are used together with the 6.2821.040 Cartridge head or the 6.2828.010 Precolumn cartridge holder.
New IC precolumns are normally filled with solution and sealed at both ends. Before the precolumn is installed in the system, it must be ensured that this solution is freely miscible with the eluent used (check manufacturer's specifications).
2.7.2 Precolumns with cartridge head
The 6.1005.020, 6.1005.040, 6.1005.050, 6.1007.010 and 6.1010.010 Precolumn cartridges are installed in the 6.2821.040 Cartridge head as follows (see Fig. 12):
1 Prepare separating column
Remove end caps from separating column 81.
Unscrew fastening screw from column inlet.
Take steel connector 74 for ferrule out of fastening screw.
2 Install cartridge
Remove end caps for precolumn cartridge 77 (the steel mesh
76 and gaskets 75 are already installed in the cartridge).
Mount steel spacer 78 on Manufit pressure screw 80 (the
steel mesh 79 and gaskets 75 are already installed in the pressure screw).
Mount precolumn cartridge 77 on the steel spacer 78 (com-
ply with flow direction if specified on the precolumn).
Screw Manufit pressure screw 80 firmly to separating column
81.
Mount steel connector 74 for ferrule on the inlet side of the
precolumn cartridge 77.
Screw on Manufit housing 73 with Manufit pressure screw 80.
2 Installation
761 Compact IC
32
7473 75 76 77 76 75 78 755428 79 80 81
Fig. 12
: Installation of precolumn cartridges with cartridge head
28 Column connection capillary from
injector
77
Precolumn cartridge
54 Compression fitting (6.2744.010) 78 Steel spacer (6.2821.080)
73 Manufit housing 79 4 Steel meshes (6.2821.020)
74 Steel connector
(of IC separating column)
80 Manufit pressure screw
75 PTFE gasket (6.2821.010) 81 IC separating column
76 2 Steel meshes (6.2821.020)
3 Connect precolumn
Fit PEEK compression fitting 54 to the column connection
capillary 28 mounted on the injection valve (see section
2.3.5).
Screw column connection capillary 28 to Manufit housing 73.
The precolumn cartridge built into the cartridge head can only be rinsed together with the separating column (see section 2.7.7/8). This increases the rinsing time by approx. 20 min.
2.7 Precolumns and separating columns
761 Compact IC
33
2.7.3 Precolumn glass cartridges with cartridge holder
The precolumn glass cartridge METROSEP Anion Dual 1 (6.1006.030) is inserted in the 6.2828.010 Precolumn cartridge holder as follows (see Fig. 13):
1 Insert cartridge
Remove end fittings 82 from screw caps 83.
Insert precolumn cartridge 85 in sleeve 84 (mark flow direc-
tion of the column cartridge on sleeve, the column cartridge should always be operated in the same flow direction).
Screw both screw caps 83 loosely to sleeve 84 by hand.
Screw both end fittings 82 into screw caps 83 so that their
capillary ends are seated in the PTFE seals of the cartridge.
Tighten both screw caps 83 by hand.
2 Connect precolumn
Provide the column connection capillary 28 mounted on the
injection valve with a PEEK compression fitting 54 (see sec- tion 2.3.5) and screw it tightly onto end fitting 82 on the inlet side of the precolumn.
Cut off as short a piece as possible from PEEK capillary 56
and fit this with two PEEK compression fittings 54 (see sec- tion 2.3.5).
Screw prepared capillary 56 to end fitting 82 on the other end
of the precolumn.
3 Rinse precolumn
Place a beaker beneath the outlet capillary of the precolumn.
Open software window for manual system control.
If necessary, modify
Flow rate to the value suited for the
inserted separating column and click on
<Send to unit> to
send this value to the 761 Compact IC.
Switch on high-pressure pump (
IC pump) by clicking <On>
and rinse precolumn with eluent for ca. 10 min.
Switch off high-pressure pump by clicking
<Off>.
The rinsed precolumn can be connected directly to the separating column instead of by means of a PEEK capillary. Proceed as follows:
Screw off PEEK capillary from end fitting 82 of the precolumn.
Screw off end fitting 82 from screw cap 83 of the precolumn.
Screw connection piece 86 onto screw cap 83.
Screw screw cap 87 of separating column 81 onto connection
piece 86.
2 Installation
761 Compact IC
34
8584 83 86 875428 8382 82 54 56
Fig. 13
: Installation of precolumn glass cartridges with cartridge holder
28 Column connection capillary from
injector
84
Sleeve for precolumn cartridge
54 Compression fitting (6.2744.010) 85 Precolumn cartridge (6.1006.030)
56 PEEK capillary (6.1831.010) 86 Connection piece
for connection precolumn – column
82 End fitting 87 Screw cap for column
83 Screw cap for precolumn
2.7.4 IC anion precolumn SUPERSEP
The 6.1009.010 IC Anion Precolumn SUPERSEP has two connections for PEEK capillaries and is installed as follows:
1 Connect precolumn
Remove end caps from the precolumn.
Fit PEEK compression fitting 54 to column connection capil-
lary 28 mounted on the injection valve (see section 2.3.5).
Screw precolumn to column connection capillary 28.
Cut a piece from the 6.1831.010 PEEK capillary as short as
possible and fit with PEEK compression fittings 54 (see sec­tion 2.3.5).
Fasten the prepared capillary to the other end of the precol-
umn.
2 Rinse precolumn
Place a beaker beneath the outlet capillary of the precolumn.
Open software window for manual system control.
If necessary, modify
Flow rate to the value suited for the
inserted separating column and click on
<Send to unit> to
send this value to the 761 Compact IC.
Switch on high-pressure pump (
IC pump) by clicking <On>
and rinse precolumn with eluent for ca. 10 min.
Switch off high-pressure pump by clicking
<Off>.
2.7 Precolumns and separating columns
761 Compact IC
35
2.7.5 General information on separating columns
New IC separating columns are normally filled with solution and sealed at both ends. Before the column is installed in the system, it must be ensured that this solution is freely miscible with the eluent used (check manufacturer's specifications).
The IC separating columns and precolumns currently available from Metrohm are listed in section 6.3.2. A test chromatogram and an infor­mation leaflet is provided with each column. You will find additional in­formation concerning these columns in the 8.732.2003 Metrohm Mono­graph «Ion chromatography» and in special "Application Bulletins", which are available on request free of charge from your local Metrohm agency.
When you install the column, always ensure that this is inserted correctly in accordance with the flow direction shown (arrow must point upwards
).
2.7.6 Selection of the sample loop
Selection of the sample loop depends on the separating column used. Normally, the following sample loops are used:
Cation columns 10 µL Anion columns with suppressor 20 µL Anion columns without suppressor 10 µL
Depending on the instrument version, the following sample loops are installed in the 761 Compact IC 761:
Version Sample loop Volume
2.761.0010 6.1825.230 (PEEK) 10 µL
2.761.0020 6.1825.210 (PEEK) 20 µL
If desired, the built-in sample loop can be replaced by one of the sam­ple loops available as an option (see section 6.3.1).
2 Installation
761 Compact IC
36
2.7.7 Installation of the separating column without suppressor
With the 2.761.0010 Compact IC without suppressor module, the IC separating column is installed as follows (see Fig. 14):
1 Connect column to injector
Remove end caps from column 81.
without precolumn:
Screw inlet end of separating column 81 (note flow direction) to column connection capillary 28 mounted on the injector.
with precolumn in cartridge head:
Install separating column 81 (note flow direction) in the car­tridge head (see Fig. 12) as described in section 2.7.2.
with precolumn in cartridge holder:
Screw inlet end of separating column 81 (note flow direction) to the precolumn fixed in the cartridge holder as described in section 2.7.3 (see Fig. 13).
with IC anion precolumn SUPERSEP:
Screw inlet end of separating column 81 (note flow direction) to the precolumn installed as described in section 2.7.4.
2 Rinse column
Place a beaker beneath the column outlet.
Open software window for manual system control.
If necessary, modify
Flow rate to the value suited for the
inserted separating column and click on
<Send to unit> to
send this value to the 761 Compact IC.
Switch on high-pressure pump (
IC pump) by clicking <On>
and rinse column with eluent for ca. 10 min.
Switch off high-pressure pump by clicking
<Off>.
3 Connect column to detector block
Screw outlet end of separating column 81 to the inlet capillary
45 permanently mounted on the detector block.
4 Fix column
Insert one or two column holders 88 (6.2027.030, 6.2027.040
or 6.2027.050) in the mounting rails 27 and fasten separating column 81 in the column holder 88.
2.7 Precolumns and separating columns
761 Compact IC
37
88
27
27
45
46
32
81
88
28
Fig. 14: Installation of column without suppressor
2.7.8 Installation of the separating column with suppressor
With the 2.761.0020 Compact IC with suppressor module, the IC sepa­rating column is first connected to the injector or precolumn. The con­nection to the suppressor module and the detector block is described in section 2.8.
1 Connect column to injector
Remove end caps from column 81.
Screw inlet end of separating column 81 (note flow direction)
to column connection capillary 28 or to the already installed precolumn (procedure see section 2.7.7).
2 Rinse column
Place a beaker beneath the column outlet.
Open software window for manual system control.
If necessary, modify
Flow rate to the value suited for the
inserted separating column and click on
<Send to unit> to
send this value to the 761 Compact IC.
Switch on high-pressure pump (
IC pump) by clicking <On>
and rinse column with eluent for ca. 10 min.
Switch off high-pressure pump by clicking
<Off>.
3 Fix column
Insert one or two column holders 88 (6.2027.030, 6.2027.040
or 6.2027.050) in the mounting rails 27 and fasten separating column 81 in the column holder 88.
27 Mounting rail
for column holder 88
28 Column connection
capillary
Connection injection valve 32 – separating column 81
32 Injection valve
45 Inlet capillary for
detector block
PEEK capillary, fixed mounting
46 Detector block
(1.733.0110)
81 Separating column
88 Column holder
(6.2027.0X0)
2 Installation
761 Compact IC
38
2.8 Suppressor module
2.8.1 General information on suppressor module
The Metrohm Suppressor Module MSM for chemical suppression installed in the 2.761.0020 Compact IC comprises a total of 3 suppres­sor units which are in turn used for suppression, regenerated with sulfu­ric acid and rinsed with water. To record every new chromatogram un­der comparable conditions, work is normally carried out with freshly regenerated suppressor. Switching is either automatic together with the valve switching or manual.
The suppressor units must never be regenerated with H
2SO4
in the same flow direction used for the eluent. You should thus always install the inlet and outlet capillaries as described in section 2.8.4 according to the scheme shown in Fig. 18.
For operation of the suppressor module, the two-channel peristaltic pump built into the 2.761.0020 Compact IC is used which conveys the regeneration solution (normally 20 mmol/L H
2SO4
) and the rinsing so-
lution (normally dist. H
2
O) to the suppressor units (flow rate of 0.5
mL/min).
The three inlets and outlets numbered 1...3 on the suppressor module each have 2 permanently mounted PTFE capillaries, which must be connected as described in section 2.8.4 (see Fig. 16 and Fig. 18).
To avoid contamination of the suppressor module by foreign particles or bacterial growth, the two 6.2821.120 Filter units PEEK (see sec- tion 2.3.6) supplied with the 2.761.0020 Compact IC must be installed between the peristaltic pump and the inlet capillaries of the suppressor module.
The suppressor module must never
be switched in the dry state as there is a danger of blocking. Before every switching operation of the suppressor module, the three suppressor units must have been rinsed for at least ½ h with eluent, regeneration and rinsing solution.
2.8.2 Preparation of the peristaltic pump
Before start-up the accessories for the 2-channel peristaltic pump built into the 761 Compact IC must be mounted according to Fig. 15. Pro­ceed as follows:
1 Attach pump tubings
Loosen both tubing cartridges 48 mounted above pump drive
52 from the holding clamp 50 by pressing down snap-action lever 51 and remove from mounting pin 53 (see Fig. 16).
2.8 Suppressor module
761 Compact IC
39
Press contact pressure lever 49 on both tubing cartridges
down as far as it will go.
Insert the pump tubings 92 and 93 (6.1826.060) into each of
the tubing cartridges as shown in Fig. 15. The white-yellow stopper 94 must click into the corresponding holder on the left-hand side of the tubing cartridge.
Place the tubing cartridges on mounting pin 53 and press
down on the right-hand side until snap-action lever 51 clicks into position on holding clamp 50. Take care that no kinks are formed in the pump tubing.
2 Install Filter units PEEK
Mount a coupling 91 (6.2744.030) to the outlet end of the two
Pump tubings 92 and 93.
Attach a piece of PTFE tubing 95 (6.1803.020) cut to the
required length (normally ca. 10 cm) using a compression fitting 54 (6.2744.010) to the other end of this coupling.
Attach the PTFE tubing 95 using a compression fitting 54
(6.2744.010) to the filter-screw 57 of the filter unit PEEK (see section 2.3.6).
89/90 91 92/93 94 48 49 94 91 98/9951 5454 95 54 59 57 54
Fig. 15
: Installing pump tubings
48
Tubing cartridge 91 Coupling (6.2744.030)
49 Contact pressure lever 92 Pump tubing (6.1826.060) for H2SO4
51 Snap-action lever 93 Pump tubing (6.1826.060) for H2O
54 PEEK compression fitting
(6.2744.010)
94 Stopper (white-yellow)
57 Filter-Screw of Filter Unit
Part of 6.2821.120 Filter unit
95 PTFE tubing (6.1803.020)
59 Filter-Housing of Filter Unit
Part of 6.2821.120 Filter unit
98 Suppressor inlet capillary for H2O
89 Aspirating tubing for H2O 99 Suppressor inlet capillary for H2SO4
90 Aspirating tubing for H2SO4
2 Installation
761 Compact IC
40
47
95
101
50
91 52
9392
32
28
46
45
33
49
91
953636
91
91
100
90
89
97
48
98
99
51
53
27
96
88
81
27 88
Fig. 16: Connection of the separating column with suppressor
27 Mounting rail
for column holder 88
81 Separating column
28 Column connection capillary
PEEK capillary (6.1831.010; 30 cm)
88 Column holder (6.2027.0X0)
32 Injection valve 89 Aspirating tubing for H2O
6.1803.020 PTFE tubing
33 PEEK coupling (6.2744.040) 90 Aspirating tubing for H2SO4
6.1803.020 PTFE tubing
36 Filter unit PEEK (6.2821.120) 91 Coupling (6.2744.030)
45 Inlet capillary to detector block
(fixed mounting)
92 Pump tubing (6.1826.060) for H2SO4
46 Detector block (1.733.0110) 93 Pump tubing (6.1826.060) for H2O
2.8 Suppressor module
761 Compact IC
41
47 Suppressor module 95 PTFE tubing (6.1803.020)
48 Tubing cartridge (6.2755.000)
for pump tubings 92/93
96 Suppressor inlet capillary for
eluent
49 Contact pressure lever
for adjusting the contact pressure
97 Suppressor outlet capillary for
eluent
50 Holding clamp
for locking the tubing cartridge into place
98 Suppressor inlet capillary for H2O
51 Snap-action lever
for releasing the tubing cartridge
99 Suppressor inlet capillary for
H
2SO4
52 Pump drive
Roller head with contact rollers
100 Suppressor outlet capillary for H2O
53 Mounting pin
for attaching the tubing cartridge
101 Suppressor outlet capillary for
H
2SO4
2.8.3 Connection of supply bottles
The supply lines for the regeneration and rinsing solution between the storage bottles and the peristaltic pump are installed as follows (see Fig. 17):
1 Prepare supply bottle for H2SO4
Prepare regeneration solution suited for the desired applica-
tion and separating column (normally 20 mmol/L H
2SO4
).
Fill regeneration solution into supply bottle 103 (amber glass,
1 L) and label the bottle.
Screw bottle attachment 102 on to supply bottle 103.
Place supply bottle 103 at one of the rear positions of the
bottle rack 10 on the 761 Compact IC (see Fig. 1).
2 Connect aspirating tubing for H2SO4
Prepare aspirating tubing 90: Cut a piece of the 6.1803.020
PTFE tubing to the required length (normally ca. 120 cm).
Pull one end of aspirating tubing 90 through a PEEK com-
pression fitting 54 (6.2744.010) so that approx. 30 cm of the tubing projects.
Screw PEEK compression fitting 54 with tubing into one
opening of bottle attachment 102 attached to the regenera­tion solution storage bottle and tighten it so that the tubing is firmly held.
Insert the free end of aspirating tubing 90 into one of the
openings in 12 of the 761 Compact IC (see Fig. 2) from above and pull it sufficiently far into the interior.
Mount a coupling 91 (6.2744.030) to the inlet end of the rear
2 Installation
761 Compact IC
42
pump tubing 92.
Mount a 6.2744.010 Compression fitting at the end of the
aspirating tubing 90 and screw this compression fitting on to the coupling 91 (see Fig. 16).
3 Prepare supply bottle for H2O
Prepare rinsing solution suited for the desired application and
separating column (normally dist. H
2
O).
Fill rinsing solution into supply bottle 103 (amber glass, 1 L)
and label the bottle.
Screw bottle attachment 102 on to supply bottle 103.
Place supply bottle 103 beside the other supply bottle on the
bottle rack 10 on the 761 Compact IC (see Fig. 1).
4 Connect aspirating tubing for H2O
Prepare aspirating tubing 89: Cut a piece of the 6.1803.020
PTFE tubing to the required length (normally ca. 120 cm).
Pull one end of aspirating tubing 89 through a PEEK com-
pression fitting 54 (6.2744.010) so that approx. 30 cm of the tubing projects.
Screw PEEK compression fitting 54 with tubing into one
opening of bottle attachment 102 attached to the rinsing so­lution storage bottle and tighten it so that the tubing is firmly held.
Insert the free end of aspirating tubing 89 into one of the
openings in 12 of the 761 Compact IC (see Fig. 2) from above and pull it sufficiently far into the interior.
Mount a coupling 91 (6.2744.030) to the inlet end of the front
pump tubing 93.
Mount a 6.2744.010 Compression fitting at the end of the
aspirating tubing 89 and screw this compression fitting on to the coupling 91 (see Fig. 16).
89/90
54
102
103
Fig. 17
: Connection of supply bottles
54 PEEK compression fitting
(6.2744.010)
89 Aspirating tubing for H2O
6.1803.020 PTFE tubing
90 Aspirating tubing for H2SO
4
6.1803.020 PTFE tubing
102 Bottle attachment (6.1602.150)
103 Supply bottle (6.1608.023)
2.8 Suppressor module
761 Compact IC
43
2.8.4 Connection of the suppressor module
The three inlets and outlets numbered 1...3 on the suppressor module 47 each have 2 permanently mounted PTFE capillaries, which must be connected as described as follows (see Fig. 16 and Fig. 18).
1 Inlet capillary for eluent
Screw inlet capillary 96 marked with "Eluent" of suppressor
module 47 to outlet end of separating column 81 using a
6.2744.010 Compression fitting.
2 Outlet capillary for eluent
Screw outlet capillary 97 marked with "Detector" of suppres-
sor module 47 to coupling 33 using a 6.2744.010 Compres­sion fitting.
Screw inlet capillary 45 of detector block 46 to other end of
coupling 33.
3 Inlet capillary for H2SO4
Attach inlet capillary 99 marked with "H2SO4" of suppressor
module 47 using a 6.2744.010 Compression fitting to the filter unit PEEK 36 connected to the rear pump tubing 92.
4 Outlet capillary for H2SO4
Pull outlet capillary 101 marked with "Waste" of the suppres-
sor module 47 from below through one of the openings 13 out of the inner compartment of the 761 Compact IC.
Lead outlet capillary 101 to a sufficiently large waste con-
tainer and fix it in place.
1
3
2
96
100
98
9799
101
Fig. 18
: Connections at suppressor module
96
Suppressor inlet
capillary for eluent
97 Suppressor outlet
capillary for eluent
98 Suppressor inlet
capillary for H
2
O
99 Suppressor inlet
capillary for H
2SO4
100 Suppressor outlet
capillary for H
2
O
101 Suppressor outlet
capillary for H
2SO4
Eluent
H2O
Detector
Waste
Waste
H2SO4
2 Installation
761 Compact IC
44
5 Inlet capillary for H2O
Attach inlet capillary 98 marked with "H2O" of suppressor
module 47 using a 6.2744.010 Compression fitting to the filter unit PEEK 36 connected to the front pump tubing 93.
6 Outlet capillary for H2O
Pull outlet capillary 100 marked with "Waste" of the suppres-
sor module 47 from below through one of the openings 13 out of the inner compartment of the 761 Compact IC.
Lead outlet capillary 100 to a sufficiently large waste con-
tainer and fix it in place.
7 Fasten capillaries to the side walls
If necessary, the two aspirating tubings 89 and 90 can be
fixed in the required position in the interior with the help of a Y.107.0150 self-adhesive strap.
If necessary, the two outlet capillaries 100 and 101 can be
fixed in the required position in the interior with the help of a Y.107.0150 self-adhesive strap.
2.9 Putting into operation
761 Compact IC
45
2.9 Putting into operation
2.9.1 Putting into operation without suppressor
Before sample solutions can be injected at the 2.761.0010 Compact IC (without
suppressor), the entire system must be tested for leaks and then conditioned with eluent until the baseline is stable. Proceed as fol­lows:
1 Open and connect system
Start the «761 Compact IC» PC program, if it has not already
been started (see section 2.5.3).
Select
File / Open / System in the main window. Select the
system file suited for the inserted separating column (e.g.
n-
01.smt
) and click on <Open>.
Select the
Connect to workplace item from the Control menu of
the system window.
2 Open control window
Double-click the 761 icon in the system window. The control
window for manual control of the 761 Compact IC appears, which indicates conductivity, pressure and current system parameters.
2 Installation
761 Compact IC
46
3 Start system
Make sure, that the aspirating tubing 63 for the high-pressure
pump is immersed in the eluent.
Select
Startup hardware (Measure baseline) from the Control
menu in the system window. The high-pressure pump is started, at the same time, a chromatogram window is opened where the baseline is recorded continuously.
4 Check for leaks
Check all capillaries and their connections between the high-
pressure pump and the detector block for escaping liquid. If eluent escapes anywhere, the appropriate compression fitting must be tightened further or changed.
5 Condition system
Rinse the system with eluent until the desired stability of the baseline is reached (normally 3060 min; if the eluent is changed, the establishment of the ion exchanger equilibrium on the separating column can take longer).
The instrument is now ready for sample determinations using the selected system.
2.9.2 Putting into operation with suppressor
Before sample solutions can be injected at the 2.761.0020 Compact IC (with
suppressor), the entire system must be tested for leaks and then conditioned with eluent until the baseline is stable. At the same time, the suppressor module must be conditioned. Proceed as follows:
1 Open and connect system "Prep-MSM*.smt"
Start the «761 Compact IC» PC program, if it has not already
been started (see section 2.5.3).
Select
File / Open / System in the main window. Open the folder
Suppressed and select the system file Prep-MSM1.smt (for an-
ion column Dual 1) or
Prep-MSM2.smt (for anion column Dual
2) suited for the inserted separating column. Click on
<Open>.
Select the
Connect to workplace item from the Control menu of
the system window.
2 Open control window
Double-click the 761 icon in the system window. The control
window for manual control of the 761 Compact IC appears, which indicates conductivity, pressure and current system
2.9 Putting into operation
761 Compact IC
47
parameters.
3 Start system
Make sure, that the aspirating tubing 63 for the high-pressure
pump is immersed in the eluent.
Select
Startup determination from the Control menu in the
system window. The high-pressure pump and the peristaltic pump are started, at the same time, a chromatogram window is opened where the baseline is recorded continuously. The suppressor module is automatically switched to the next posi­tion every 20 min and conditioned in this way.
4 Set contact pressure for pump tubings
Press contact pressure lever 49 on both tubing cartridges 48
upwards until regeneration and rinsing solution just start to be drawn in.
Then press contact pressure lever 49 upwards until it clicks
once more to obtain optimal contact pressure.
5 Check for leaks
Check all capillaries and their connections between the high-
pressure pump and the detector block for escaping liquid. If eluent escapes anywhere, the appropriate compression fitting must be tightened further or changed.
2 Installation
761 Compact IC
48
6 Condition system
Rinse the system with eluent until the desired stability of the baseline is reached (normally 3060 min; if the eluent is changed, the establishment of the ion exchanger equilibrium on the separating column can take longer). After this time the suppressor module is sufficiently conditioned too.
The instrument is now ready for sample determinations using the selected system.
Pump tubings are consumable material with a lifetime which depends on the contact pressure. This is why the tubing cartridges should be raised completely by loosening snap-action lever on the right-hand side if the pump is to remain switched off for a considerable length of time (the set contact pressure remains unchanged).
2.10 Connection of external devices
761 Compact IC
49
2.10 Connection of external devices
2.10.1 Connection of the 750 Autosampler
The 750 Autosampler available from Metrohm as an option is an auto­matic sampler for ion chromatography. It accommodates max. 128 samples each of 730 µL, which are automatically transferred to the sample loop attached to the injection valve of the 761 Compact IC. The samples on the 750 Autosampler are changed by a remote signal which is outputted by the 761 Compact IC (761 Compact IC as "Mas­ter").
In order to connect the 750 Autosampler the 6.2128.160 Cable which is available as an option is additionally required. The 750 Autosampler is connected to the 761 Compact IC as follows:
1 Electrical connection 761 – 750
Connect 6.2140.010 Connector plug to contact closure strip 9
of the 750 Autosampler (see 750 Instructions for Use).
Connect one end of the 6.2128.160 Cable to the connections
13 "EXTERNAL INJECT INPUT" and 14 "GND" on the connec­tor plug. The two cable ends are appropriately inscribed with "750" (white) and "GND" (brown).
Connect the other end of the 6.2128.160 Cable to the remote
interface 24 of the 761 Compact IC (see Fig. 2).
2 Tubing connection 750 – injection valve
Loosen rotary nipple 31 which is screwed onto the inner side
of connection 3 of the 761 Compact IC (see Fig. 3 and Fig. 4).
Pull aspiration tubing 4 completely out from connection 3 and
screw it off from connection "1" of injection valve 32.
Cut transfer tubing 8
(see 750 Instructions for Use) to the
shortest possible length between the needle of the 750 Auto­sampler and injection valve 32 of the 761 Compact IC.
Pull the free end of transfer tubing 8
through the opening of connection 3 on the 761 Compact IC and screw it onto con­nection "1" of injection valve 32 using a 6.2744.010 PEEK compression fitting.
Retighten rotary nipple on the interior side of connection 3 to
fix the transfer tubing.
3 Tubing connection injection valve – waste
Insert 6.2744.020 Coupling (from 761 accessories) into
connection 3 of the 761 Compact IC.
Screw aspirating tubing 4 onto the 6.2744.020 Coupling and
lead it into the waste container.
2 Installation
761 Compact IC
50
4 Settings at the 750 Autosampler
Set parameter "Run time" to 0.0 to activate the external start
of the 750 Autosampler (see section 3.5 of the 750 Instruc- tions for Use). Filling the sample loop of the injection valve only takes place when the impulse produced by the 761 Compact IC at remote output 3 (see point 5) makes contact at the "EXTERNAL INJECT INPUT" of the 750 Autosampler. The run time will then be determined externally.
5 Settings in the «761 Compact IC» program
A time program must be drawn up for the selected system
which first switches the injection valve to the "Fill" position, then produces an impulse at remote lead 3 to start sample addition by the 750 Autosampler and finally switches the in­jection valve to the "Inject" position. The following example shows such a program:
2.10.2 Connection of the 766 IC Sample Processor
The 766 IC Sample Processor available from Metrohm as an option is an automatic sampler for ion chromatography. It accommodates max. 127 samples each of 2.5 mL or 11 mL, which are automatically trans­ferred to the sample loop attached to the injection valve of the 761 Compact IC. Sample changing and filling the sample loop are started by a remote signal produced by the 761 Compact IC (761 Compact IC as "Master").
The 6.2141.110 Cable is required for connecting the 766 IC Sample Processor. The 766 IC Sample Processor is connected to the 761 Compact IC as follows:
1 Electrical connection 761 – 766
Connect the end of the 6.2141.110 Cable marked with "766"
to the remote interface 22
of the 766 IC Sample Processor
(see 766 Instructions for Use).
2.10 Connection of external devices
761 Compact IC
51
Connect the end of the 6.2141.110 Cable marked with "732/1"
to the remote interface 24 of the 761 Compact IC (see Fig. 2).
2 Tubing connection 766 – injection valve
At the 761 Compact IC, loosen the rotary nipple 31 screwed
onto the interior side of connection 3 (see Fig. 3 and Fig. 4).
Take aspirating tubing 4 completely out of connection 3 and
unscrew from connection "1" of injection valve 32.
Cut PEEK capillary 18
installed at the 766 IC Sample Proces-
sor (see 766 Instructions for Use) to the desired length.
Pull the free end of PEEK capillary tubing 18
through the opening of connection 3 of the 761 Compact IC and screw it onto connection "1" of injection valve 32 with the help of a
6.2744.010 PEEK compression fitting.
Retighten rotary nipple 31 on the interior side of connection 3
to fix the capillary.
3 Tubing connection injection valve – waste
Insert 6.2744.020 Coupling (from 761 accessories) into
connection 3 of the 761 Compact IC.
Screw aspirating tubing 4 onto the 6.2744.020 Coupling and
lead it into the waste container.
4 Settings at the 766 IC Sample Processor
For operation with the 761 Compact IC it is recommended
that the following program "761" is used with the 766 IC Sample Processor:
parameters method 761 number of samples: Rack
>start sequence 1 CTL:Rm: INIT >sample sequence 1 SCN:Rm : Wait1
2 MOVE 1 : sample 3 LIFT: 1 : work mm
4 SCN:Rm : Wait1
5 PUMP 1.1 : 120 s
>final sequence >changer settings rack number 0 lift rate 1 12 mm/s shift rate 20 >manual stop CTL Rmt: ************** CTL RS232:
------------
Method name
Number of samples to be processed
(entire sample rack)
Initialize remote interface
Waiting until 761 Compact IC sends
signal on remote line 3
Move needle to sample position
Place lift with needle to working
position
Waiting until 761 Compact IC sends
signal on remote line 3
Fill sample loop with sample during
120 s
Settings for changer functions
Reaction to manual stop
2 Installation
761 Compact IC
52
5 Settings in the «761 Compact IC» program
A time program must be drawn up for the selected system
which first produces an impulse at remote lead 3 to start sample changing on the 766 IC Sample Processor and then switches the injection valve to the "Fill" position. A further im­pulse is then produced to start filling the sample loop within 120 s. At the end the injection valve is switched to the "Inject" position, which also starts data acquisition.
6 Run conditions
In order that the joint operation of 761 – 766 functions properly the following conditions must be fulfilled:
The program must be started first on the 766 IC Sample
Processor, then the "Sample Queue" on 761 Compact IC.
Remote lead 3 on the Compact IC must be set to 0 at the
start of each determination (Set
System startup values:
Remote line 3 = 0).
2.10.3 Connection of other devices
Any external devices can be connected to the 25-pin remote interface 24 (see Fig. 2). The 8 output lines can be used to control these external devices.
Before an external device is connected to the remote interface 24, the 761 Compact IC must always be switched off using mains switch 19 !
The pin assignment of the remote interface, its functions and the elec­trical requirements and the conditions are described in section 6.1.
3.1 Requirements
761 Compact IC
53
3 Operating tutorial
This section introduces you to the operation of the 761 Compact IC by means of a brief operating tutorial which describes the basic operat­ing steps needed for the recording of an ion chromatogram using one of the system files supplied.
The determination of the anionic content of a drinking water sample with the METROSEP Anion Dual 2 column with chemical suppression is used an illustrative example. Please note that the steps and pa­rameter settings described apply only to this example. If you use a different separating column and system file, the procedures described in the tutorial must be modified appropriately.
For further explanations of the operation, please refer to section 4.
3.1 Requirements
For the determination of anions in drinking water described in this tuto­rial, the following instruments, accessories and solutions are required:
2.761.0020 Compact IC
with suppressor module
6.1825.210 Sample loop (20 µL, PEEK)
already integrated in the 2.761.0020 Compact IC
6.1006.100 IC anion column METROSEP Anion Dual 2
Eluent
2 mmol/L NaHCO
3
/ 1.3 mmol/L Na2CO3 in dist. H2O
flow: 0.8 mL/min
Standard
Standard solution with 0.5 mg/L F
, 5 mg/L Cl– and
10 mg/L each of NO
3
and SO
4
2–
(in dist. H2O)
3 Operating tutorial
761 Compact IC
54
3.2 Preparations
Before you start this brief tutorial, the entire IC system must be correctly installed as described in section 2. In what follows, the most important points for the installation are described once again (for details, see the sections mentioned).
1 Install 761 Compact IC
Setting up instrument section 2.2Installing and connecting detector block section 2.3.1Mounting syringe and aspirating tubing section 2.3.2Mounting drain tubes section 2.3.3/4Mains connection section 2.4Connection to PC section 2.5
2 Prepare eluent
Preparing eluent:
2 mmol/L NaHCO
3
/ 1.3 mmol/L Na2CO3 in dist. H2O
Microfiltering and degassing eluent section 5.1.3
3 Install high-pressure pump
⇒ Removing transport security screws section 2.6.1 ⇒ Mounting pulsation dampener section 2.6.2 ⇒ Installing eluent supply section 2.6.3 ⇒ Degassing pump section 2.6.4
4 Connecting separating column and suppressor
⇒ Connecting IC anion column section 2.7.8 ⇒ Connecting peristaltic pump section 2.8.2 ⇒ Connecting supply bottles section 2.8.3 ⇒ Connecting suppressor module section 2.8.4 ⇒ Conditioning system section 2.9.2
3.3 Calibration
761 Compact IC
55
3.3 Calibration
After the complete IC system has been installed as described in sec­tion 3.2, the first calibration can be started. A standard solution is re-
quired for this; it should contain the substances to be determined in approximately the same concentrations in which they can be expected in the sample.
In our example of drinking water determination using the METROSEP Anion Dual 2 IC-anion column a 20 µL sample loop is used; this is filled with the following standard solution:
0.5 mg/L F
5 mg/L Cl
10 mg/L NO
3
10 mg/L SO
4
2–
(as Na+- or K+ salts in dist. water)
Please note that all program displays refer to the condition in which the system
s-03.smt is loaded for the first time. If you want to work through
this tutorial at a later date and the system
s-03.smt has been altered in
the meantime then differences with respect to the program display and the parameter values may occur.
In the description of the calibration procedure it is assumed that the PC and 761 Compact IC are not in operation and that the system must first be conditioned again. If this is not the case (e.g. if you start the tutorial immediately after conditioning) then you can skip steps 1 to 6.
1 Switch on 761 Compact IC
Switch on 761 Compact IC with mains switch 19 on the rear
of the instrument. After the instrument has been switched on the mains pilot lamp 9 lights up.
2 Switch on PC
Switch on PC and start «761 Compact IC» program.
3 Open and connect system "PrepMSM2.smt"
Start PC program «761 Compact IC», if it is not already
running (see section 2.5.3).
Select
File / Open / System in the main window. Select the
system file
PrepMSM2.smt in the folder Suppressed and click on
<Open>.
Select
Connect to workplace of the Control menu in the system
window.
4 Start system "PrepMSM2.smt"
Select Start determination of the Control menu in the system
window. High-pressure pump and peristaltic pump are started, at the same time, a chromatogram window opens where the baseline is recorded continuously. The suppressor
3 Operating tutorial
761 Compact IC
56
module is switched to the next position every 20 min and conditioned in this way.
5 Condition system "PrepMSM2.smt"
Rinse IC system with eluent until the desired stability of the
baseline is achieved and the suppressor module is suffi­ciently conditioned (at least 1 h).
6 Stop system "PrepMSM2.smt"
Select Stop determination of the Control menu in the system
window. Data recording and time program are stopped, high­pressure pump and peristaltic pump keep running on.
7 Open and connect system "s-03.smt"
Select Change of the System menu in the system window.
Select the system file
s-03.smt suited for the METROSEP An-
ion Dual 2 column in the folder
Suppressed and click on <Open>. The system PrepMSM2.smt is closed, the system s-03.smt is opened and automatically connected.
8 Start system "s-03.smt"
Select Start determination of the Control menu in the system
window. High-pressure pump and peristaltic pump are started, at the same time, a chromatogram window opens where the baseline is recorded continuously. The status bar of this window shows the message
Measure(Baseline), beside
this running time, analysis time, absolute conductivity and number of measuring points per second are displayed. The messages
Running [761 Compact IC] and Waiting for INJECT
[Suppressed]
appear in the SYSTEM STATE window
Let the system run for several minutes. If the system has
already been conditioned then a stable conductivity value of approx. 14 µS/cm will be obtained after a few minutes.
Double-click the chromatogram or select
View all of the View
menu. The sensitivity is set automatically so that all measur­ing points are visible.
This function is only available if the integration has been started. In the present example the integration only starts after a delay period
Delay = 2.6 min.
Select the desired sensitivity using the cursor keys <> or
<>. In the chromatogram window the following display is seen, for example:
3.3 Calibration
761 Compact IC
57
9 Enter information for determination
Click on or select Passport of the Method menu. The
passport opens with the
General page, which contains the
general information about the determination.
Enter a title for the calibration as
Ident (e.g. Standard).
Change the analysis time under
Duration if required.
Change other parameters on the passport tabs to character-
ize sample and acquirement conditions if required.
10 Switch injection valve to "FILL" position
Click on the button of the 761 icon in the system
window to switch the injection valve to the "FILL" position. At the same time, the suppressor module is switched to the next position.
3 Operating tutorial
761 Compact IC
58
11 Fill sample loop
Immerse the aspirating tubing 4 attached to connection 3 in
the standard solution.
Using the syringe fixed to connection 2 siphon in ca. 1 mL
standard solution.
12 Switch injection valve to "INJECT" position
Click on the button of the 761 icon in the system
window to switch the injection valve to the "INJECT" position. At the same time, the data recording is started automatically. The status bar of the chromatogram window shows the mes­sage
Measure, the SYSTEM STATE window shows the mes-
sages
Running [761 Compact IC] and INJECT done [Suppressed].
When the analysis time of 24 min set in the method has
expired the recorded chromatogram is automatically evalu­ated and stored. The chromatogram window then closes.
13 Open chromatogram
Click on or File / Open / Chromatogram in the main win-
dow. Select the chromatogram file
*.chw just recorded and
click on
<OK>. The chromatogram window is opened where
the found peaks are numbered and the baselines are drawn in.
14 Modify integration parameters
Click on or select Integration from the Method menu in
the main window to open the integration parameters window.
If required (e.g. if the fluoride peak is not detected), change
the time delay
Delay before starting peak integration.
3.3 Calibration
761 Compact IC
59
Click on <Apply>. The Integration parameters window is
opened and the chromatogram is reintegrated.
Repeat this procedure for all other integration parameters
until the result satisfies your expectations. Click on
<Apply>
for each parameter change.
Close the
Integration parameters window with <OK>.
15 Modify peak allocation
Click on or select Calibration / Components from the
Method menu. The prepared table for the components fluo-
ride, chloride, nitrate, and sulfate is displayed below the chromatogram. For all peaks which could be allocated un­ambiguously to a component this table contains the peak number in the
Peak column and the retention time in the Time
column. In the chromatogram itself this retention time is indi­cated by the cursor (vertical line).
If the
Peak column contains a 0 the peak must be manually
allocated to a component. In this case, insert the peak num­ber from the chromatogram into the
Peak column and click
the
Time field on this row with the mouse. The corresponding
retention time is then inserted automatically as the new time for this component.
For each correctly allocated peak click on the retention time
in the
Time column. This time is automatically shown in the
chromatogram by the cursor. If the retention time is not lo­cated in the area of the middle of the peak then the value in the
Time column should be adapted accordingly.
3 Operating tutorial
761 Compact IC
60
This is done by clicking on the chromatogram and moving
the cursor with the aid of the cursor keys <> and <> to the middle of the peak and reading off the corresponding re­tention time in the status line. Enter this value (rounded off if necessary) into the
Time column.
The optimized table could, for example, then appear as
follows:
16 Start calibration
Click the <Concentrations> button in the component window.
Click on
<Add>. The Add level window appears.
3.3 Calibration
761 Compact IC
61
Confirm
Calibration level 1 in this window with <OK>. A new
column
Level 1 is added in the Concentrations window.
Enter the concentrations of all components of the standard
solution used in this column:
Click on
<Calibrate>. The following window appears:
Confirm
Level 1 with <OK>.
Click on
<OK> in the Concentrations window.
3 Operating tutorial
761 Compact IC
62
17 Display calibration curves
Click on <Graphs> in the Components window. Select the desired component in the
Component field for which
the calibration curve should be displayed (e.g. for
chloride).
Close the
Component - chloride window with <OK>.
Close the
Components window below the chromatogram by
clicking on
<OK>.
18 Save chromatogram and method
Close the chromatogram window. A window appears with the
message
Changes in *.chw. Modified: Calibration. Save changes?.
Click on
<Yes>. A window appears with the message *.chw
already exists. Overwrite?
.
Click on
<Yes>. A window appears with the message Method
s-03.mtw was modified. Save changes?
.
Click on
<Yes>. The File Save As window appears.
Save the method under the previously given name
s-03.mtw.
If you want to store the method under a different name then this must be newly selected for the system under
PC icon /
Setup / Processing method
.
3.4 Sample determination
761 Compact IC
63
3.4 Sample determination
Following the calibration of the IC system as described in section 3.3 the first sample solution can be injected.
1 Enable "Verify sample"
Switch on the Verify sample option of the Control menu in the
system window. This causes the
Edit sample description win-
dow to open at the start of each determination; the most im­portant data for the sample can then be entered in this win­dow.
2 Start system "s-03.smt"
Select Start determination of the Control menu in the system
window. An empty chromatogram window opens where the baseline is recorded continuously. The messages
Ready [761
Compact IC]
and Initialisation [Suppressed] appear in the SYS-
TEM STATE
window. Because of the Verify sample option
switched on the
Edit sample description window is opened
automatically.
3 Enter information for determination
Enter the desired information concerning the sample in the
Edit sample description window and confirm with <OK>.
4 Switch injection valve to "FILL" position
Click on the button of the 761 icon in the system
window to switch the injection valve to the "FILL" position. At the same time, the suppressor module is switched to the next position.
3 Operating tutorial
761 Compact IC
64
5 Fill sample loop
Immerse the aspirating tubing 4 attached to connection 3 in
the vessel containing the drinking water sample.
Using the syringe fixed to connection 2 siphon in ca. 1 mL
drinking water.
6 Switch injection valve to "INJECT" position
Click on the button of the 761 icon in the system
window to switch the injection valve to the "INJECT" position. At the same time, the data recording is started automatically. The status bar of the chromatogram window shows the mes­sage
Measure, the SYSTEM STATE window shows the mes-
sages
Running [761 Compact IC] and INJECT done [Suppressed].
When the analysis time of 24 min set in the method has
expired data recording is automatically stopped and the chromatogram is integrated and evaluated. In the chroma­togram window the peaks which have been found are num­bered and the baseline drawn. In the
SYSTEM STATE window
the message
Finished [Suppressed] appears. It is now possible
to record further samples with the same system.
At the end of the determination, the chromatogram window is
closed and the chromatogram is saved automatically. The file name is generated automatically and contains date and time in a coded form:
1
st
numeral: Alphabetical code for year
(e.g. j = 1999, k = 2000, l = 2001, etc.)
2
nd
numeral: Code for month
(1...9 = Jan....Sept., a...c = Oct....Dec.)
3
rd+4th
numeral: Day (01 ... 31)
5
th
- 8th numeral: Time (hh:mm)
3.4 Sample determination
761 Compact IC
65
7 Open chromatogram
Click on or File / Open / Chromatogram in the main win-
dow. Select the chromatogram file
*.chw just recorded and
click on
<OK>. The chromatogram window is opened where
the found peaks are numbered and the baselines are drawn in.
8 Report output
Click on or select Make report of the Process menu in the
main window. The
Report Options window appears.
Under
Report destination, switch on the desired target option
for report output.
Under
Items to report and More items to report, click all desired
elements for the report output.
Click on <Report> to start the report output to the selected
target.
3 Operating tutorial
761 Compact IC
66
9 Print report
Click on or select Print of the File menu in the main
window. The standard printing window is opened where printer, printing range and number of copies can be selected. After confirmation with
<OK> the results including the chro-
matogram are printed out.
4.1 Fundamentals of the operation
761 Compact IC
67
4 Operation
This section describes the most important points concerning the operation of the 761 Compact IC. For further details please refer to the on-line help in the PC program which can provide you with the required information rapidly and conveniently from any place in the program.
4.1 Fundamentals of the operation
4.1.1 Starting/closing the program
Start the «761 Compact IC» program
Start the program Double-click this icon or the
Metro761.exe file to start the «761
Compact IC 1.1» program. The Login window appears:
Enter your password and click on
<Log In>.
After software installation, the program can be started without entering a
Password. For the defini-
tion of users, see section 4.2.2.
Close the «761 Compact IC» program
761 COMPACT IC / File / Exit
Exit the «761 Compact IC» program.
The program is also quit by clicking on
in the upper right part
of the main window
761 COMPACT IC.
4 Operation
761 Compact IC
68
4.1.2 Glossary
System
The term “system“ is used to describe the combination of instrument settings, time program and process method which have been opti-
mized for the specific separating column and the determination to be carried out with it. A system is used to start single determinations or de­terminations with the help of a sample queue.
Systems are stored as system files (
*.smt) in the Systems directory.
Method
A method contains all information necessary for data acquisition, in­tegration, peak evaluation and quantification. It can be considered
as the chromatogram template, i.e. chromatogram without raw data.
Methods are stored as method files (
*.mtw) in the Methods directory.
Each system is linked to a method. This method is called processing method and is opened automatically at the start of a new determina-
tion.
Chromatogram
A chromatogram is a graphic plot of the elution curve (signal vs. time) recorded following a chromatographic separation on a separating col­umn.
Chromatograms are stored as chromatogram files (
*.chw) in the Data
directory. As well as the measuring data the chromatogram files also contain the method parameters and system settings which have been used for data recording, data processing and remote control.
Determination
In order to carry out a determination a suitable system must be se­lected for the separating problem. The result of the determination is a chromatogram, in which the measuring data and results of the deter­mination are stored.
Calibration
Calibration is used to describe the method of determining the relation­ship between the peak height or peak area found for one component and its concentration in the sample. The result of the calibration is a calibration function (calibration curve), which shows the relationship between the amount of sample and the evaluated quantity.
The determination of the calibration function with reference solutions can be carried out as a one-point or as a multiple-point calibration. The calibration method which is mainly used in ion chromatography is the external standard calibration (absolute calibration); calibration with an internal standard (relative calibration) or tabulated calibra- tion are also possible.
4.1 Fundamentals of the operation
761 Compact IC
69
Integration
Integration is to be understood as being the method for determining the peak area and peak height with the aid of approximate baselines. The integration algorithm included in the program is influenced by the inte-
gration parameters and the optionally programmable integration events which are defined in the method. In addition, the integration can be manually corrected later with the aid of the peak editor.
Sample queue
A sample queue is used for the automated processing of series of samples, particularly in combination with a sample changer.
Batch reprocessing
Batch reprocessing is understood to be the subsequent reprocessing of a series of chromatograms which have been loaded in a batch re­processing queue. During reprocessing with a selected method the set­tings for calibration, integration, passport, appearance and report can be altered at will.
4.1.3 Overview of program windows
The «761 Compact IC» program consists of different windows whose functionality is linked together. The different windows are:
761 COMPACT IC Main program window for file administration,
printing, opening of systems, methods and chromatograms, login and user rights, optional settings and window handling.
CHROMATOGRAM Window for graphic plot of running or recorded
chromatograms.
SYSTEM Window for loaded system with possibility for
manual control of the 761 Compact IC.
SYSTEM STATE Window for display of status messages for the
connected system.
WATCH WINDOW
Window for display of conductivity and pressure.
QUEUE EDITOR Window for edition of sample queues and batch
reprocessing queues.
4 Operation
761 Compact IC
70
4.1.4 Main window elements
The elements of the main window 761 COMPACT IC are the menu bar, the tool bar and the status bar, indicating prompts and logged-in user.
4.1.5 Icons of the main window
The following icons are displayed in the 761 COMPACT IC main window:
Open chromatogram
Save chromatogram
Open last batch reprocessing file
Report settings
Print preview
Send chromatograms to «Autodatabase»
Print report
Passport
General method settings
Integration parameters
Component table
Cascade all opened chromatogram windows
Vertical tiling of open chromatogram windows
Horizontal tiling of open chromatogram windows
Appearance of the chromatogram window
Menu bar
Tool bar
Status bar
Logged-in user
Prompts, information
4.1 Fundamentals of the operation
761 Compact IC
71
Enable/disable peak editor mode
View whole chromatogram
Help
Lock system
Connected system
4.1.6 Overview of file types
The following file types are produced by the «761 Compact IC» soft­ware:
*.bar Batch reprocessing file
This binary file contains data of the batch reprocessing queue. The
*.bar file is stored automatically in the Data
folder.
*.cal Calibration file
This binary file contains calibration data, which can be ex­ported with
761 COMPACT IC / Method / Calibration / Export
calibration
. The *.cal file is stored automatically in the Meth-
ods
folder.
*.chw Chromatogram file
This binary file contains chromatogram, system and method data of a determination. The
*.chw file is stored
automatically in the
Data folder.
*.mtw Method
This binary file contains the data acquisition method, which can be linked to a system. The
*.mtw file is stored automati-
cally in the
Methods folder.
*.que Sample queue file
This binary file contains sample table queue. The
*.que file
is stored automatically in the
Methods folder.
*.rtt Report template
This ASCII file contains a report template. The
*.rtt file is
stored in the program folder.
*.smt System file
This ASCII file contains the system settings. The
*.stm file is
stored automatically in the
Systems folder.
*.dev Device file
This ASCII file contains drivers for devices. The
*.dev file is
stored in the
Devices folder.
4 Operation
761 Compact IC
72
4.1.7 Context sensitive menus
Some of the menu functions of the program windows are also accessi­ble by clicking on the desired window or item and pressing the right mouse button. The pop up windows have different contents and func­tions depending on the selected active window or item type.
4.1.8 Keyboard and mouse functions
The mouse can be used to carry out the normal program operating functions such as the selection of menu items and fields. It can addi­tionally be used for magnifying a section of a chromatogram (zoom- ing). To zoom a portion of the plot it is necessary to place the mouse cursor to the upper left corner of the square to zoom, press the left mouse button and drag the cursor to the lower right corner of the rec­tangle. After releasing of the left mouse button the selected region will be zoomed full-screen. If the cursor is active in the peak editor mode then it can be moved by pressing down the right-hand mouse key.
The keyboard can also be used to scale a chromatogram in the win­dow, as described below.
Keyboard quick reference
Cursor is inactive:
[ up ] Increases sensitivity on the Y axis.
[ down ] Reduces sensitivity on the Y axis.
[ right ] Expands a chromatogram on the X axis.
[ left ] Shrinks a chromatogram on the X axis.
[ Ctrl ] + [ Home ] Autoscale procedure on the X axis (shows all
on X).
[ Ctrl ] + [ End ] Autoscale procedure on the Y axis (shows all
on Y).
[ PageUp ] Shifts a chromatogram on
1
/10 part of a screen
upwards.
[ PageDown ] Shifts a chromatogram on
1
/10 part of a screen
downwards.
[ Shift ] + [ up ] Increases a distance between channels of a
chromatogram.
[ Shift ] + [ down ] Reduces a distance between channels of a
chromatogram.
[ 0 (Zero) ] Adjusts a zero on the last point of a chroma-
togram (running chromatogram) or its lowest level (finished run).
Only part of the chromatogram is on screen:
4.1 Fundamentals of the operation
761 Compact IC
73
[ Ctrl ] + [ right ] Moves one window right (without change of
scale on X and Y axes).
[ Ctrl ] + [ left ] Moves one window left (without change of
scale on X and Y axes).
[ Home ] Shows the beginning of a chromatogram (with-
out change of scale on X and Y).
[ End ] Shows the end of a chromatogram (without
change of scale on X and Y).
[ 0 (Zero) ] Adjusts a zero on the lowest level in the win-
dow.
Cursor is active:
[ 0 (Zero) ] Adjust a zero in site of the cursor.
[ right ] Moves cursor left to right.
[ Shift ] + [ right ] Quickly moves cursor left to right.
[ left ] Moves cursor to the left.
[ Shift ] + [ left ] Quickly moves cursor to the left.
[ Home ] Moves cursor to beginning of a window.
[ End ] Moves the cursor to end of a window.
[ Shift ] + [ End ] Sets the beginning of a window in site of the
cursor.
[ Shift ] + [ Home ] Sets the end of a window in site of the cursor.
4.1.9 Help
By clicking on , by clicking on , by selecting the Help /
Contents menu item, or by pressing the [ F1 ] key you can get on-line
help on the current topic anywhere in the program.
Green texts can be clicked to jump to a different Help topic.
Violet texts identify the dialog item, parameter or button in
the corresponding window.
Blue texts identify important information.
4 Operation
761 Compact IC
74
4.2 Instrument and software settings
4.2.1 Fonts
761 COMPACT IC / Options / Fonts
This option allows the selection of fonts used by the system.
Font for dialog... Selection of font used for dialog boxes.
Default setting:
MS Sans Serif / Standard / 8 pt.
Font for reports... Selection of font used for report output to the
screen or printer. Default setting:
Courier New / Standard / 10 pt.
Font for tables... Selection of font used for data presentation in
tables on the screen. Default setting:
MS Sans Serif / Bold / 8 pt.
Font for plots... Selection of font used for labels on chroma-
togram plots and calibration curves. Default setting:
Times New Roman / Bold / 10 pt.
Save fonts configuration Save chosen font configuration.
4.2.2 Security system
The «761 Compact IC» program has a security system based on the list of users. Every user has his unique password and one of the following access levels:
Novice Restricted access to program functions. Allows
only start and stop of determinations using ex­isting system and method files and manual control of the 761 Compact IC. Modifications of system, method and data files are not allowed.
Master Access to all program functions with few excep-
tions: the user cannot set
Global preferences,
Hardware settings and security system.
Administrator Access to all program functions.
It is recommended to make user lists and enter passwords as a first ac­tion after system installation. So, select the menu item
761 COMPACT IC /
Options / Security
and click on <Log In> in the Log In window without en-
tering a password. The
Security options window appears:
4.2 Instrument and software settings
761 Compact IC
75
Enter user name, password and access level for all users. Don't forget to make one of the user an
Administrator, otherwise this window will not
open again. At the end, click on
<OK>.
After configuration of the security system the program prompts for the password every time the system starts. This user name stamps all methods, chromatograms and reports. It is possible at any time to change the user with the menu item
761 COMPACT IC / Options / Lock sys-
tem
.
4.2.3 Global settings
761 COMPACT IC / Options / Global preferences
This window is used for global program settings.
This window is only accessible for users with
Administrator
access level.
GLP On If this option is enabled, the following parame-
ters are automatically set:
4 Operation
761 Compact IC
76
Data file overwrite = Never If method changed = Don't ask to save If disk method is newer = Ask to over­write
Data file overwrite Overwriting of chromatogram files:
Never Chromatogram files cannot be overwritten. A
modified chromatogram is saved as a new file with the file name number raised by 1.
Ask The user is asked if the chromatogram should
be overwritten.
Always Chromatogram files are always overwritten
without confirmation.
If method changed Saving of method files:
Don't ask to save The method file is not saved automatically. It
can be saved only with
File / Save / Method.
Ask to save The user is asked if the method should be
saved.
Overwrite without asking
Method files are overwritten without confirma­tion.
If disk method is newer Overwriting of method files:
Ask to overwrite The user is asked if the method should be
overwritten.
Overwrite without asking
Method files are overwritten without confirma­tion.
Opening chromatograms Opening of chromatograms:
Ignore last data directory
If this option is enabled, the default data direc­tory
Data is opened.
OK, Apply buttons mean
Save object to file If this option is enabled, the <Apply> button in
system settings windows is replaced by the
<Save> button. The system settings are saved
if the
<Save> or <OK> button is clicked.
with file name confirmation
If this option is enabled, the
<Apply> button in
system settings windows is replaced by the
<Save as> button. The system settings can be
saved in a new file if the
<Save as> or <OK>
button is clicked.
4.2 Instrument and software settings
761 Compact IC
77
Chromatogram units Units for chromatograms:
Flow Unit for flow rate: µL/min, mL/min Pressure Unit for pressure: MPa, psi, bar, atm
Print via print spooler
Switch on/off printing via print spooler. Switch
off this option if you use a GDI printer.
Measuring priority Setting the priority of program execution. real-
time means that the «761 Compact IC» program
has the highest priority,
normal means that all
active programs have the same priority.
<Default colors> Default colors for chromatographic windows
(details see section 4.5.3).
4.2.4 COM port
761 COMPACT IC / Options / 761 Compact IC:COM1
This menu item opens the Links window for COM port (serial RS232 in­terface) selection and settings.
If
COM1 is clicked with the right mouse button, the following menu items
appear:
Put on desktop Possibility for setting COM port parameters for
recording data transfer (details see on-line help).
Change Possibility for changing the COM port (default
setting:
COM1). The following window is
opened, where the interface can be changed by clicking on the desired COM port.
4 Operation
761 Compact IC
78
4.3 Systems
The term “System“ describes the combination of instrument settings, time program and process method which has been optimized for a
specific separating column and the determination to be carried out with it. A system is used to start single determinations or determinations with the help of a sample queue. Systems are stored as system files (
*.smt)
in the
Systems directory.
4.3.1 System window
A system window is opened with 761 COMPACT IC / File / Open / System and the selection of the desired system file. It contains icons for PC, Watch window (screen) and 761 Compact IC.
4.3.2 System file handling
The following menu items are used for opening, changing, saving and closing of systems:
761 COMPACT IC / File / Open / System
Load an existing system file (*.smt) from the Systems directory and open the corresponding
SYSTEM window
The directory and name of the opened system file are displayed in the title bar of the
SYSTEM window. A star ( * ) at the end of the name indi-
cates that the system settings have been changed since the last sav­ing.
SYSTEM / System / Open other
Load an existing system file (*.smt) from the Systems directory and open a new
SYSTEM window. The system parameters can be changed for this
system, but it can not be connected without disconnecting the actual system first.
SYSTEM / System / Change
Disconnect the current system, open the selected new system and connect it automatically.
Watch window
PC
761 Compact IC
4.3 Systems
761 Compact IC
79
SYSTEM / System / Save
Save the current system settings in a system file (*.smt) in the Systems directory.
SYSTEM / System / Close
Disconnect the selected system (if it is connected) and close the sys­tem window.
4.3.3 System functions
Connect and disconnect system
To make possible manual control of the instrument and start of deter­minations the selected system must be connected to the PC. Only one system can be connected at the same time. Systems are connected and disconnected as follows:
SYSTEM / Control / Connect to workplace
Connect selected system to the COM port of the PC. If the system is connected, the two buttons
<Inject> and <Fill> appear on the 761 icon
in the system window for manual control of the injection valve (see sec­tion 4.3.7).
At the same time, the
icon appears on the tool bar. If this icon is
clicked, the
SYSTEM window is always displayed in front of all other win-
dows.
SYSTEM / Control / Disconnect from workplace
Disconnect selected system from the COM port of the PC. If the system is disconnected, the two buttons
<Inject> and <Fill> disappear on the
761 icon in the system window. Manual control is not available for this system any more, but all other system settings can be modified and saved.
Start/stop hardware and record baseline
SYSTEM / Control / Startup hardware (Measure Baseline)
Starting the hardware at the 761 Compact IC includes sending of
Sys-
tem startup values
, starting of the high-pressure pump and (if present)
starting of the peristaltic pump.
At the same time, the recording of the measurement signal using the method of the connected system is started. Independently of the set chromatogram
Duration, the measurement signal is recorded until the
data acquisition is stopped with
SYSTEM / Control / Stop data acquisition or
a new determination is started. Alternatively the baseline recording can be stopped by clicking the
icon of the chromatogram window. In this case the user is asked if the recorded baseline should be saved or not.
4 Operation
761 Compact IC
80
SYSTEM / Control / Shutdown hardware
High-pressure pump and (if present) peristaltic pump at the 761 Com­pact IC are immediately stopped. A running determination and an ac­tive sample queue are also stopped and the remote output lines are set to the values defined under
Hardware / Remote lines after power on.
SYSTEM / Control / Stop data acquisition
Stop recording of the baseline.
Start/stop determinations
SYSTEM / Control / Start determination
Start determination using the settings of the selected system. At this
start command, the
System startup values are set at the 761 Compact IC.
The high-pressure pump and the peristaltic pump are started if they are not already running. The time program and the data recording are started either immediately (
Start with determination) or after switching the
injection valve to the "Inject" position (
Start with inject) as set in the Start
mode
window.
SYSTEM / Control / Stop determination
Stop running determination. Data acquisition and time program are
terminated immediately. The recorded chromatogram is saved auto­matically if the
Save chromatogram after the run option on the Passport /
Processing
tab is enabled.
Alternatively the determination can be stopped by clicking the
icon of the chromatogram window. In this case the user is asked always if the determination should be saved or not.
SYSTEM / Control / Stop data acquisition
Stop data acquisition of the running determination immediately and
save the recorded chromatogram automatically if the
Save chromatogram
after the run
option on the Passport / Processing tab is enabled. The time
program of the running determination is continued normally.
Options for determinations
SYSTEM / Control / Auto restart
If this option is enabled, a new determination is started automatically using the current system after the preceding determination has been finished normally or stopped manually.
The
Auto restart option is disabled if determinations are made with an
active sample queue.
SYSTEM / Control / Verify sample
If this option is enabled, the Edit sample description window is opened automatically at the start of each determination for entry of the following sample information:
4.3 Systems
761 Compact IC
81
Ident User defined identifier (title) for the chroma-
togram to be displayed in the title bar of the chromatogram window and in the
Chromatogram
open window.
Calibration level Calibration level (0 = sample; 1n = calibra-
tion solutions).
Info 1 / Info 2 Sample description.
Volume Injected volume in µL.
Dilution Dilution of the sample.
Vial number Autosampler vial position to take sample from.
Amount Sample amount. If this value is different for the
calibration run (c) and the sample run (s), the component concentrations of the sample are calculated as follows:
C
s
= Cc Amounts / Amountc
Internal standard amount Concentration of the internal standard compo-
nent for relative concentration calculations.
Date/time when... Date and time of sample collection (the default
values are equal to the date and time when the chromatogram starts).
The
Verify sample option is disabled if determinations are made with an
active sample queue.
4 Operation
761 Compact IC
82
4.3.4 System settings
Modify system window
SYSTEM / Setup / Drag icons
If this option is enabled, the system icons can be resized and moved in the
SYSTEM window and the SYSTEM window itself can be resized.
To resize an icon or the window, move the cursor to the desired object until
appears. Press the left mouse button and resize the object to
the desired size.
To move an icon, move the cursor to the desired object until
ap­pears. Press the left mouse button and move the object to the desired place.
Watch window display
SYSTEM / Setup / Watch window
If this option is enabled, the
WATCH WINDOW with conductivity and
pressure display is automatically opened if a determination or measure baseline is started (see section 4.3.6).
Set start mode
SYSTEM / Setup / Start mode
This menu item opens the
Start mode window for definition of start mode
for time program and data acquisition.
This window contains the following two fields:
Start with determination The loaded object is started at the moment the
determination is started.
Start with inject The loaded object is started at the moment the
sample is injected.
The two objects available are:
RECORDER Data acquisition defined in the method. 761 Compact IC Time program defined on the Program tab of
the system control window.
These two objects can be moved from one field to the other using the
or buttons.
4.3 Systems
761 Compact IC
83
Print system parameters
SYSTEM / Setup / Parameters / Print
A report of the system parameters including the time program is cre­ated and opened using the «Wordpad» program. The
*.txt file opened
can be printed, saved and exported into other programs. The system report includes the name of the method linked to the system, the con­figuration settings for the 761 Compact IC, the system startup values and the time program (incl.
Remote configuration), if such a program ex-
ists and is switched on (
ENABLED).
4.3.5 PC icon
Menu options for PC icon
The PC icon is one of the three components of the
SYSTEM window. If
the system is connected and this icon is clicked with the right mouse button, the following menu appears:
Open Load the processing method linked to the system
and open an empty chromatogram window.
Setup Open PC setup window for selection of processing
method and data source.
Select processing method and data source
The
PC / Setup menu item opens the following window:
This window contains the following two fields:
Processing method Directory and name of the method file (*.mtw)
linked to the system. The method file can be selected with
<choose>
and opened with
<show>.
4 Operation
761 Compact IC
84
Connected data source Data source for main measurement channel. By
clicking on
<choose>, the Data source window is
opened for selection of the data source:
This window contains the following two fields:
Connected Connected data source.
Default setting:
Cond[ 761 Compact IC]
Available Available data sources of the 761 Compact IC:
# Cond Conductivity # Press Pressure
The data sources can be moved from one field to the other using the
or buttons (only one data source can be connected).
4.3.6 Watch window
Menu options for screen icon
The watch window icon is one of the three components of the
SYSTEM
window. If the system is connected and this icon is clicked with the right mouse button, the following menu item appears:
Open Open the WATCH WINDOW for live display of conduc-
tivity and pressure:
The colors of the watch window fields can be changed by clicking the fields with the right mouse button and selecting the menu item
Choose color / ... .
4.3 Systems
761 Compact IC
85
4.3.7 Instrument icon
Menu options for instrument icon
System
disconnected
System connected
Injection valve in
"INJECT" position
System connected
Injection valve in
"FILL" position
The instrument icon for the 761 Compact IC is one of the three compo­nents of the
SYSTEM window. If the system is connected, the 761 icon
contains two buttons for manual control of the injection valve:
<Inject> Switch injection valve to "INJECT" position.
<Fill> Switch injection valve to "FILL" position.
If the system is connected and the 761 icon is clicked with the right mouse button, the following menu appears:
Open Open the system settings window.
Hardware Open the hardware settings window.
Diagnostics Open the diagnostics window.
System parameters for disconnected system
A double-click on the instrument icon or selection of the
Open menu
item using the right mouse button opens the system settings window. For a disconnected system, the
Initial tab is displayed.
4 Operation
761 Compact IC
86
Flow, mL/min Startup value for flow rate of the high-pressure
pump. Entry range:
0.20 ... 2.50 mL/min
Pressure max, MPa Startup value for maximum pressure limit for
high-pressure pump. Entry range:
0.0 ... 25.0 MPa
Pressure min, MPa Startup value for minimum pressure limit for
high-pressure pump. Entry range:
0.0 ... 25.0 MPa
Full scale, µS/cm
Startup value for full scale range. Selection:
50, 250, 1000 µS/cm
Remote lines Startup value for remote line settings 18.
Selection:
0, 1
Instrument control for connected system
A double-click on the instrument icon or selection of the
Open menu
item using the right mouse button opens the system settings window. For a connected system, the
Control tab is displayed. It allows manual
control of the 761 Compact IC functions and setting of startup values to be sent to the instrument. This tab shows the current measurement val­ues for conductivity and pressure.
Conductivity, µS/cm Live display of measured conductivity.
Pressure, MPa Live display of measured pressure.
The color settings for this two fields can be changed by clicking the fields with the right
4.3 Systems
761 Compact IC
87
mouse button and selecting the menu item
Choose color / ... .
Actual values Display of actual values
Flow, mL/min Display of flow rate of the high-pressure pump. Pressure max, MPa Display of maximum pressure limit for high-
pressure pump.
Pressure min, MPa Display of minimum pressure limit for high-
pressure pump.
Full scale, µS/cm Display of selected full scale range. Remote lines Display of current remote line settings.
Valve Injection valve
<Inject> Switch injection valve to "INJECT" position. <Fill> Switch injection valve to "FILL" position
Suppressor Suppressor module
<Step> Switch the suppressor module to the next posi-
tion. The time since the last switching of the suppressor module is displayed in the field be­side the
<Step> button.
IC pump High-pressure pump
<On> Start high-pressure pump. <Off> Stop high-pressure pump.
Peristaltic pump Peristaltic pump
<On> Start peristaltic pump. <Off> Stop peristaltic pump.
System startup values The system startup values are sent and applied
to the 761 Compact IC each time the system is connected, a determination is started, or the values are sent manually with
<Send to unit>.
Flow, mL/min Startup value for flow rate of the high-pressure
pump. Entry range:
0.20 ... 2.50 mL/min
Pressure max, MPa
Startup value for maximum pressure limit for high-pressure pump. This limit is controlled even without connection to the PC. Entry range:
0.0 ... 25.0 MPa
Pressure min, MPa Startup value for minimum pressure limit for
high-pressure pump. This limit is controlled even without connection to the PC. Entry range:
0.0 ... 25.0 MPa
Full scale, µS/cm Startup value for full scale range.
Selection:
50, 250, 1000 µS/cm
Remote lines
Startup value for remote line settings 1…8. Selection:
0, 1
4 Operation
761 Compact IC
88
Time program
On the
Program tab of the system settings window a user-defined time
program for instrument control can be entered. This program is started automatically as defined in the
Start mode window (see section 4.3.3) ei-
ther at the moment the determination is started (
Start with determination)
or at the moment the sample is injected (
Start with inject).
The
Program tab contains the two following subpages: Program Main time program with all program steps. Remote configuration Possibility for creation of user-defined remote
commands.
Program
On the
Program subpage, program steps including time, program in-
struction and parameter can be entered.
Time (1st column) Time at which program instruction is applied.
Entry range: 0.0 ... 999.9 min If no time is entered, the program instruction is
applied together with the last instruction with time entry.
Command (2nd column) Program instruction (see List of program in-
structions).
In addition to these predefined instructions, user-defined remote commands can be en­tered if activated on the
Remote configuration
tab.
Parameter (3rd column) Parameter for program instruction (see List of
program instructions).
ENABLED Enable program start (a disabled program is
not started).
<Add> Add new program instruction. <Delete> Delete selected program instruction. <Verify> Test the time program (error messages are dis-
played if program is wrong).
4.3 Systems
761 Compact IC
89
List of program instructions
The following program instructions can be added to the time program on the
Program subpage:
Instruction Parameter entry Meaning
Valve Inject, Fill
Switch injection valve to "INJECT" or "FILL" position.
FullScale
50, 250, 1000 µS/cm
Set full scale range to the selected value.
ICPump on, off
Switch on or off the high-pressure
pump.
Flow 0.2 ... 2.5 mL/min
Set flow rate of the high-pressure pump to the desired value.
Pmax 0.0 ... 25.0 MPa
Set maximum pressure limit for the high-pressure pump to the desired value.
Pmin 0.0 ... 25.0 MPa
Set minimum pressure limit for the high-pressure pump to the desired value.
Remote 0, 1, *, p
Set remote output lines 1...8 to the desired values. For entry of the first value, enter
1, 0, p or *. For entry of the
other values, move the cursor in front of the value to be changed and enter
1, 0, p or *.
Program END, RESET
The
END flag can be used to end a
program, especially if the program time should be longer than the chromatogram duration. Additional steps after this flag are not allowed. The
RESET flag is used to reset the
parameters to the system startup values.
Suppressor
Switch suppressor module to the
next position.
Peristaltic on, off
Switch on or off the peristaltic pump.
Remote configuration
On the
Remote configuration subtab user-defined remote commands can
be defined, which can be inserted into a time program.
4 Operation
761 Compact IC
90
Name (1st column) User-definable name of the remote command
(e.g.
Start_766).
Remote command (2nd column)
Setting the remote output lines 1…8. Selection:
0 (line off, inactive, open) 1 (line on, active, 0 V) p (pulse) * (leave line in current status)
For entry of the first value, enter
1, 0, p or *. For
entry of the other values, move the cursor in front of the value to be changed and enter
1, 0,
p or *.
<Add> Add new remote command. <Delete> Delete selected remote command. <Activate> Activate the defined remote commands for in-
sertion into the time program.
Configuration
The
Configuration tab in the system settings window contains configura-
tion settings for the 761 Compact IC.
761 unit version Selection of instrument version:
761.0010 761 Compact IC without suppressor
761.0020 761 Compact IC with suppressor
Loading...