LumaSense Technologies 3434i User Manual

BE6028-14
Instruction
Manual
3434i SF6 Leak Detector
Index ________________________________________________________________________

Index

Index ................................................................................................................................... 2
3434i SF6 Leak Detector ........................................................................................................ 4
Safety Considerations. ...................................................................................................... 5
Applying Power ................................................................................................................ 7
Chapter 1 Using this Manual ................................................................................................. 8
1.1 Introduction ............................................................................................................ 9
1.2 Online Monitoring .................................................................................................... 9
1.3 PC Use BZ7002 Calibration Software .......................................................................... 9
1.4 PC Use BZ7003 Offline Software ................................................................................ 9
1.5 Stand-alone Use ...................................................................................................... 9
Chapter 2 Preliminary Tasks ............................................................................................... 12
2.1 Connecting the Monitor to the PC ............................................................................ 13
2.2 Checking /changing the Fuses in the Monitor. ........................................................... 14
2.3 Setting the USB Communication Parameters ............................................................. 17
2.4 Setting the RS-232 Communication Parameters ........................................................... 18
2.5 Setting the Ethernet Communication Parameters ....................................................... 19
2.6 Setting-up user Levels ........................................................................................... 23
2.7 Restoring of calibration data ................................................................................... 25
Chapter 3 Philosophy of Operation ....................................................................................... 26
3.1 PC Use – Online .................................................................................................... 27
3.2 Stand-alone Use .................................................................................................... 28
Chapter 4 Introduction to and Familiarization with the Monitor ............................................... 30
4.1 Special Terminology ............................................................................................... 31
4.2 Attaching the Sampling Tube and External Filter ........................................................ 34
4.3 Familiarization with the Monitor using the Front Panel ................................................ 35
Chapter 5 The Monitor ....................................................................................................... 56
5.1 Introduction .......................................................................................................... 57
5.2 Front Panel ........................................................................................................... 57
5.3 Starting-up or Restarting the Monitor ....................................................................... 61
5.4 Configuring the Monitor .......................................................................................... 64
5.5 Alarm Relay .......................................................................................................... 67
5.6 Continuous Warm System ...................................................................................... 69
Chapter 6 Keying-in Parameters .......................................................................................... 70
6.1 Types of Parameters .............................................................................................. 71
Chapter 7 Setting-up and Starting a Monitoring Task ............................................................ 74
7.1 The Parameters’ Units ............................................................................................ 75
7.2 The Measurement Parameters ................................................................................. 77
7.3 Setting-up the Monitor’s Alarm Levels ...................................................................... 85
7.4 Starting a Monitoring Task ...................................................................................... 86
Chapter 8 Viewing Measurement Data – Using the Monitor .................................................... 90
8.1 Displayed Monitoring Task Data............................................................................... 91
8.2 Printed Monitoring Task Data .................................................................................. 96
8.3 Storage Space in the Display Memory ...................................................................... 99
Chapter 9 Storage of Data in the Monitor’s Memory .............................................................. 100
9.1 Types of Memory in the Monitor ............................................................................. 101
9.2 Storage of Measurement Results ............................................................................ 109
9.3 Management of Data in Background Memory ........................................................... 112
Chapter 10 Using a Printer with the Monitor ......................................................................... 116
10.1 Introduction ......................................................................................................... 117
10.2 Printer Cables ...................................................................................................... 117
10.3 Setting-up Communication Parameters ................................................................... 118
10.4 Additional Parameters Required .............................................................................. 123
10.5 Data Logging Function .......................................................................................... 124
10.6 Error Logging Function .......................................................................................... 124
10.7 Printing-out Data from the Monitor ......................................................................... 125
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 2 of 191
Index ________________________________________________________________________
Chapter 11 Maintenance of the 3434i .................................................................................. 126
11.1 Changing the Fine Air-filters .................................................................................. 127
11.2 Cleaning the Filter in the Ventilation Unit ................................................................. 132
11.3 Cleaning the Instrument ........................................................................................ 133
Chapter 12 Calibration of the Optical Filters ......................................................................... 134
12.1 About the Installed Optical Filters ........................................................................... 135
12.2 What is Calibration and Why is it Necessary? ........................................................... 136
12.3 Tasks Before Starting Any Calibration ..................................................................... 138
12.4 The Basic Calibration Set-up .................................................................................. 144
12.5 General Settings – Stand-alone Use ........................................................................ 147
12.6 Zero-point Calibration ........................................................................................... 151
12.7 Humidity-interference Calibrations – Stand-alone Use ............................................... 155
12.8 Span Calibration ................................................................................................... 159
12.9 Calibration of the Water-vapour Filter – Stand-alone Use ........................................... 167
12.10 Setting-up a Combined Zero-point and Humidity-interference Calibration – Stand-
alone Use ................................................................................................................ 173
12.11 Setting-up any Combined Calibration – Stand-alone Use ...................................... 175
Chapter 13 Messages Related to Calibration ......................................................................... 177
13.1 “Successful” Calibration Messages .......................................................................... 178
13.2 “Successful *” Calibration Messages ........................................................................ 178
13.3 “Invalid” Calibration Messages ............................................................................... 179
13.4 Error Messages Displayed when the INFO button is used ........................................... 179
Chapter 14 Warning and Error Messages ............................................................................. 185
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 3 of 191
Safety Considerations ________________________________________________________________________

3434i SF6 Leak Detector

July 2014
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 4 of 191
Safety Considerations ______________________________________________________________________

Safety Considerations.

Throughout this manual Monitor is used for: The 3434i SF6 Leak Detector
The Monitor complies with:
EN/IEC 61010-1, 2
equipment for measurement, control and laboratory use.
Can/CSA-C22.2 No. 61010-1-04 - Safety Requirements for
Electrical Equipment for Measurement, Control, and Laboratory Use.
UL Std. No. 61010A-1 (2
Electrical Equipment for Measurement, Control, and Laboratory Use.
To ensure safe operation and retain the Monitor in safe condition, note the following:
TO AVOID THE POSSIBILITY OF AN EXPLOSION; MONITORING
OF FLAMMABLE GASES IN EXPLOSIVE CONCENTRATIONS
nd
Edition: Safety requirements for electrical
nd
Edition) - Safety Requirements for
EXPLOSION HAZARD!
MUST NEVER BE ATTEMPTED.
Never operate the 3434i SF6 Leak Detector in potentially explosive environments.
When monitoring potentially flammable or toxic gases it is essential that:
The instrument itself is placed in a well-ventilated area outside the
potentially hazardous zone.
A sufficiently long tube is connected to the air-outlet on the back
panel so that the sampled gas is carried away to the open air or to an extraction and/or filtration unit.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 5 of 191
Safety Considerations ______________________________________________________________________
Environmental Conditions for transport and storage:
Temperature: -25 to +55ºC Relative Humidity: 0 to 80% RH Atmospheric Pressure: 800 to 1060 hPa
Warnings!
Avoid water condensation in the instrument.
Switch off all equipment before connecting or disconnecting their
digital interface. Failure to do so could damage the equipment.
Whenever it is likely that correct function or operating safety of
the apparatus has been impaired, the apparatus must be made inoperative and secured against unintended operation.
Any adjustment, maintenance and repair of the open apparatus
under voltage must be avoided as far as possible and, if unavoidable, must be carried out only by trained personnel.
If a fault is reported by the monitor that indicates correct function
of the instrument may be impaired, consult your local LumaSense Technologies representative. Under no circumstances should repair be attempted by persons not qualified in service of electronic instrumentation.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 6 of 191
Safety Considerations ______________________________________________________________________

Applying Power

Before using the 3434i check that the available mains voltage match the specified voltage and frequency for the instrument.
SAFETY SYMBOLS
The apparatus will be marked with this symbol when it is important that the user refers to the associated warning statement given in the User Guide.
Frame or Chassis Dangerous voltage Protective earth
Copyright © 2014 LumaSense Technologies A/S
All rights reserved. No part of this publication may be reproduced or distributed in any form or by any means without prior consent in writing from LumaSense Technologies A/S, Denmark
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 7 of 191
Chapter 1 ______________________________________________________________________
Chapter 1
Using this Manual
July 2014
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 8 of 191
Chapter 1 ______________________________________________________________________

1.1 Introduction

This manual can be used in several ways. The first time users can work their way through the examples in order to get to know this monitoring system. The more experienced users can jump directly to the relevant chapters in order to gain assistance, and experts can use this manual as a reference book by using the index.
The monitor can be operated in different ways; Online Monitoring using LumaSoft Gas Single point 7810, Online Calibration and back­up using the Calibration Software BZ7002, Offline using the Offline Software BZ7003 , and as a Stand-alone instrument. This manual focuses on the Stand-alone use. For instruction in the use of BZ7002 and BZ7003 please refer to the User Manuals BE6034 and BE6035 respectively.

1.2 Online Monitoring

LumaSoft Gas Single Point 7810 is supplied with the instrument. This is to be used for Online monitoring (PC use) please follow the instructions in the BE6027 User Manual for online monitoring and set-up.

1.3 PC Use BZ7002 Calibration Software

The BZ7002 Calibration Software is used for calibration of the gas monitor from a PC. Please refer to the BZ7002 User Manual BE6034 and Chapter 12 of this Manual.

1.4 PC Use BZ7003 Offline Software

Data from the display memory can be copied to the background memory after a measurement task is performed. The BZ7003 Offline Software is used for upload of these measurement data from the monitor to the PC in either excel format or text format files. Please refer to the User Manual BE6035 and Chapter 9 of this manual for further instructions.

1.5 Stand-alone Use

In some situations, it may be necessary to set up without connecting it to a PC. The procedure to do this are covered in this manual under the Stand-alone use headings. The way these instructions are presented is shown below:
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 9 of 191
Chapter 1
SELECT SET-UP BRANCH
S1 S2 S3
______________________________________________________________________
MEASUREMENT FORMAT CONFIGURATION
The display above is used continuously through-out this manual. It assists you displaying the text on screen and by indicating which of the push-buttons can and should be pressed.
In general, the push-buttons shown above are used to navigate through the various modes possible within the 3434i but are described in more detail below:
These push-buttons are illustrated as S1, S2 and S3. On the 3434i they correspond to the key and their position on the instrument. These select push-buttons enable you to select one of the options displayed.
This group of 5 push-buttons are referred to in this manual collectively as the direction keys. The symbols below are used to simplify the instruction in this manual.
& enable you to increase & decrease numbers, respectively, or to go to the Previous & Next Displays, respectively.
& enable you to move across number fields or go to the Previous & Next gases, respectively.
, depending on the situations, acts as an “Enter” or “Go To Head”
key.
Memory Function push-buttons are always represented with the name
of the push-button enclosed in a box which is shaded-in.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 10 of 191
Chapter 1 ______________________________________________________________________
SYSTEM Text that appears on the display screen is shown in an open
shaded box, using UPPER case letters.
FORMAT System General Clock When referring to any part of the Set-up “tree”, the text is
shown in open shaded boxes with the same typeface as that used in the “tree”.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 11 of 191
Chapter 2 ______________________________________________________________________
Chapter 2
Preliminary Tasks
July 2014
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 12 of 191
Chapter 2 ______________________________________________________________________
When taking delivery of the 3434i SF6 Leak Detector, five very important and preliminary tasks must be completed before starting to operate it:
Install the BZ7002 Calibration Software to enable operation Back-up of the calibration file supplied with the instrument if calibrations are performed by LumaSense Technologies A/S. Connect the Monitor to a PC (see section 2.1). Check/Change the fuses in the monitor (see section 2.2). Set the Communication parameters. USB (see section 2.3), RS232 (see section 2.4), Ethernet (see section 2.5). Set-up the password protected user levels (see section 2.6).

2.1 Connecting the Monitor to the PC

The monitor comes complete with a USB interface cable. The maximum length of an USB cable is 5 meters. USB extension kits can be used to extend the USB connection to 30 meters.
The gas monitor can also be connected to a PC using a 9-pin to 9-pin null modem RS-232 interface cable (optional accessory WL950-003).
Finally the monitor can be connected to a PC using the Ethernet network connection.

2.1.1 Fitting the RS-232 Cable.

Ensure that both the monitor and the PC are switched off at the mains. Failure to do so may result in your equipment being damaged. Locate the serial port at the back of the PC; refer to your PC manual if in doubt. Push the connector on the RS-232 cable on to the serial port socket, and secure it firmly using the securing screws. Locate the output labelled “RS-232” at the back of the Monitor. Push the connector at the other end of the RS-232 cable on to this socket, and secure it firmly using the securing screws. Turn on the PC. Wait for Windows to start up.
The instruments can be turned on at the mains now.

2.1.2 Fitting the USB Cable.

The gas monitor can be connected to an USB port on the PC in order to make measurements using the LumaSoft Gas Single Point 7810 PC application, which communicates using the USB interface communication protocol.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 13 of 191
Chapter 2 ______________________________________________________________________
The USB interface cable can be connected while the gas monitor is switched on.

2.2.3 Fitting the Ethernet Cable.

The gas monitor can be connected to a local Ethernet network in order to make measurements using the LumaSoft Gas Single Point 7810 PC application, which communicates using the TCP/IP network interface protocol or to view the homepage of the gas monitor in a standard PC internet browser.
The Ethernet network cable can be connected while the gas monitor is switched on.
Connect a standard Ethernet network cable to the Ethernet socket connector on the back of the gas monitor. Connect the other end of the Ethernet network cable to your local network. The yellow LED on the Ethernet socket connector will lit, if a local network connection is detected.

2.2 Checking /changing the Fuses in the Monitor.

Unplug the Monitor from the mains power supply and pull out the
plug in the AC Mains socket of the Monitor If the fuses in the Monitor are blown it can indicate a serious fault in the instrument, therefore it is advisable to contact your LumaSense service representative before changing the fuse!

2.2.1 Checking/changing the Fuses in the 3434i

1. Turn the 3434i so that it stands on its back panel and remove
the rubber “shoes” mounted on the four “feet” of its base­plate.
2. Using Pozidrive screwdriver No. 1, undo and completely
remove the screws located under the rubber “shoes” which have just been removed. Note: these screws hold the upper and lower covers of the 3434i together. If the screws are not completely undone you will not be able to remove the upper cover as explained in step
5.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 14 of 191
Chapter 2 ______________________________________________________________________
3. Undo the upper four screws holding the frame over the back-
panel of the 3434i using Pozidrive screwdriver No. 0.
4. Using the Pozidrive screwdriver No.0 undo the upper four
screws which hold the frame over the front panel of the 3434i.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 15 of 191
Chapter 2 ______________________________________________________________________
5. Gently lift off the upper cover of the 3434i by pulling it firmly
upward.
6. Use a screwdriver to remove the fuse-holder:
7. Remove the old fuse from the holder and replace it with a new.
Check that the fuse has the correct rating.
The fuse must be:
Slow-blow (T) fuse with a rating of 2.5 A, 250 VAC, 5x20 mm, UL approved.
LumaSense order number: VF0102A
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 16 of 191
Chapter 2 ______________________________________________________________________
8. Install the fuse-holder and re-assemble the 3434i again.
The figure below shows the inside of the top cover of the 3434i. Note that the front side of the cover has square flanges. Place the top cover back in its correct position on top of the 3434i and screw in the all screws.

2.3 Setting the USB Communication Parameters

The LumaSoft Gas 7810 application with USB communication is able to communicate with the 3434i gas monitor through a USB cable.
BZ7002 and BZ7003 applications with USB communication are able to communicate with the 3434i gas monitor through a USB cable .
No setup on the 3434i gas monitor is necessary in order to be able to communicate with the 3434i gas monitor through an USB connection.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 17 of 191
Chapter 2
Baud rate
9600
Stop bits
1
Data bits
7
Parity
Even
Hardwire mode
Leased line
Handshake type
Hardwire
Text Line Terminator
CR-LF
______________________________________________________________________

2.4 Setting the RS-232 Communication Parameters

2.4.1 Default RS-232 Communication Parameters.

The Default RS-232 Communication parameters are shown in the table below.
Performing a full reset will set all parameters to the default settings.

2.4.2 Checking/Changing the RS-232 Communication Parameters

The communication parameters for the serial interface must be set using the push-buttons on the front of the monitor.
1. Press SET-UP S3 S1 S3 S1 .The screen display now shows the
following text.
If the baud rate displayed is incorrect press ↵ and use and to display the correct value. Press ↵ again to store the selection.
If the baud rate displayed is correct, then press to continue to the next parameter. Press S1 to select 1 STOP BIT. Press S1 to select 7 DATA BITS Press S2 to select EVEN PARITY Press S3 to select LEASED-LINE Press S3 to select HARD-WIRED HANDSHAKE. Press SET-UP to exit the set-up mode Press RESET and S1 in order that the new settings are enabled.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 18 of 191
Chapter 2
SELECT ADDRESSABLE SET-UP BRANCH
______________________________________________________________________

2.4.3 Checking/Changing the Text line Terminator.

The Text line Terminator must be set using the push-buttons on the front of the monitor.
Press SET-UP S2 S3
Press SET-UP to exit the set-up mode

2.4.4 Checking/Changing the Data Log and Error Log Options.

The Data Log and Error Log must be set using the push-buttons on the front of the monitor.
Press SET-UP S3 S1 S1 S3 S1 S1
Press SET-UP to exit the set-up mode

2.5 Setting the Ethernet Communication Parameters

The LumaSoft Gas 7810 application with Ethernet communication is able to communicate with the 3434i gas monitor through a local Ethernet network.
The homepage of the 3434i gas monitor can also be displayed in a web browser, when the 3434i gas monitor is connected to a local Ethernet network.

2.5.1 Checking/Changing the Ethernet Communication Parameters

The communication parameters for the Ethernet interface can be set by using the push-buttons on the front of the monitor.
1. Press SET-UP S3 S1 S3 S2. The screen display now shows the
following text.
Ethernet IEEE 488
2. Press S1 to enter the Ethernet setup. During this setup you can
also use the andkeys to go back and forth between the
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 19 of 191
Chapter 2
IS 3434i'S IP ADDRESS SET BY DHCP ?
3434i 'S IP ADDRESS 192.168.000.200
PRESS ENTER TO CHANGE VALUE
3434i'S IP ADDRESS 192.168.000.200
CANCEL DEFAULT
3434i'S IP PORT NUMBER: 23
______________________________________________________________________
available Ethernet settings. The screen display now shows the following text.
NO YES
3. Here you can select whether the IP address of the 3434i gas
monitor is to be assigned by a DHCP server in the local network. If you want the 3434i's IP address to be assigned by the DHCP server press the S3 (Yes) key, where after you will proceed to the below step 6 as the IP address is set by the DHCP server. If you want yourself to set the IP address of the 3434i press the S1 (No) key. The screen display now shows the following text.
4. Here you can change the IP address of the 3434i gas monitor in
case the IP address is not set by a DHCP server in the local network. If you want to change the IP address, press the ↵ key. If
the IP address does not need to be changed, press thekey and you will proceed to the below step 6. If you selected to change the IP address, the screen display now shows the following text.
5. Here you can change the 4 individual numbers in the IP address
by using the andkeys. Each of the 4 individual numbers can be set in the range from 000 to 255. You can step between the 4 individual numbers in the IP address by using the and keys. In case you want to return to the start value of the IP address press the S1 (CANCEL) key. In case you want to return to the default stored value of the IP address press the S3 (DEFAULT) key.
When you have set the IP address, press the ↵ key. The screen display now shows the following text.
PRESS ENTER TO CHANGE VALUE
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 20 of 191
Chapter 2
PROTECTED IP ADDRESS 000.000.000.000
3434i'S IP PORT NUMBER: 23
PROTECTED IP ADDRESS 000.000.000.000
CANCEL DEFAULT
______________________________________________________________________
6. Here you can set the IP port number for the Ethernet
communication. This normally does not need to be changed. If you want to change the IP port number press the ↵ key. If the IP
port numbers does not need to be changed, press thekey and you will proceed to the below step 8. If you selected to change the IP port number the screen display now shows the following text.
CANCEL DEFAULT
7. Here you can change the IP port number by using the and
keys. The port number can be set in the range from 0 to 32767. In case you want to return to the start value of the IP port number press the S1 (CANCEL) key. In case you want to return to
the default stored value of the IP port number press the S3 (DEFAULT) key.
When you have set the IP port number, press the ↵ key. The screen display now shows the following text.
PRESS ENTER TO CHANGE VALUE
8. In case you want the 3434i gas monitor to be able to
communicate only with a certain PC on the local network, you can set the IP address of this PC as the protected IP address. This can be desired due to security reasons so no other PC is able to communicate with the 3434i gas monitor. If no protection is desired the protected IP address should be set to a value of 000.000.000.000.
If you want to change the protected IP address, press the ↵ key. If the protected IP address does not need to be changed, press
thekey and you will proceed to the below step 10. If you selected to change the protected IP address the screen display now shows the following text.
9. Here you can change the 4 individual numbers in the protected IP
address by using the andkeys. Each of the 4 individual numbers can be set in the range from 000 to 255. You can step between the 4 individual numbers in the protected IP address by using the and keys. In case you want to return to the start values of the protected IP address press the S1 (CANCEL) key. In case you want to return to
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 21 of 191
Chapter 2
CURRENT DHCP ADDRESS 192.168.000.062
______________________________________________________________________
the default stored value of the protected IP address press the S3 (DEFAULT) key.
When you have set the protected IP address, press the ↵ key. If you previously selected that the IP address was NOT to be set by the DHCP server you will proceed to the below step 11.
If you instead selected that the IP address to be set by the DHCP server, the screen display now shows the following text.
ACCEPT
10. Here the IP address of the 3434i, which has been assigned by the
DHCP server, is displayed. This IP address can be used to
manually set Ethernet communication in a PC application, which
accesses the 3434i gas monitor.
Press the S2 (ACCEPT) key to accept.
11. This concludes the setup of the Ethernet parameters of the 3434i
gas monitor.
Press the SET-UP key to leave the setup.

2.5.2 Display the Homepage of the Monitor

The homepage of the 3434i gas monitor can be displayed if the gas monitor is connected to a Ethernet network.
In section 2.5.1 it is explained how to set-up the Ethernet parameters in the 3434i gas monitor.
Any web browser can be used to display the homepage of the gas monitor. The IP address obtained from the Ethernet set-up of the 3434i gas monitor, can be entered in the address field of the web browser in order to display the homepage.
The last measured gas values are displayed together with information about the gas monitor, like the device serial number, if water- and cross-compensation is performed. Error and warning messages (see Chapter 14 for further details) will also be displayed on the homepage. The grey "Update Measurements" link can be used to update the homepage with the last measurement results.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 22 of 191
Chapter 2 ______________________________________________________________________
Fig.2.1 The 3434i gas monitor homepage

2.6 Setting-up user Levels

Press the Lock function push button on the front of the Monitor. This enables setting up 3 different user levels, Level 1, Level 2 and Level 3. These user levels are password protected.
Level 1 is the Operator level, which enables start/stop and reading of
measurement results. Partial Reset is possible.
Level 2 is the Expert level, which in addition to Level 1 access
enables set up of parameters. Full Reset is possible.
Level 3 is the Super-User level, which allows all possible operations
including calibration.

2.6.1 Default settings for User Level and Password.

Before dispatch from the factory the User Level and the Password are set to the following default setting.
User Level 3
And the default setting for the Password is:
Password: ***** (5 times *)
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 23 of 191
Chapter 2
CURRENT USER LEVEL IS X
CHANGE USER LEVEL CHANGE PASSWORD
ENTER PASSWORD "*****"
PASSWORD IS "*****"
PRESS ENTER TO CHANGE VALUE
ENTER PASSWORD "*****"
______________________________________________________________________
These settings are to be used to change user level or password the first time.

2.6.2 Changing the Password

In order to change the Password Press the Lock function push button on the front of the Monitor.
The following text is displayed.
Press S3 “CHANGE PASSWORD”
The following text appears.
PRESS ENTER TO CHANGE VALUE
1. Press to change value.
The following text appears.
CANCEL DEFAULT
2. Enter the old Password using the andkeys.
3. Press ↵.
The following text appears.
4. Type in the new Password using the andkeys.
5. Press .

2.6.3 Changing the User Level

In order to change the User Level Press the Lock function push button on the front of the Monitor
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 24 of 191
Chapter 2
ENTER PASSWORD "*****" CURRENT USER LEVEL IS X
CHANGE USER LEVEL CHANGE PASSWORD
CURRENT USER LEVEL IS X
______________________________________________________________________
The following text is displayed.
1. Press S1 “CHANGE USER LEVEL”
The following text appears.
PRESS ENTER TO CHANGE VALUE
2. Press ↵.
3. Enter the current Password using the andkeys.
4. Press ↵.
The following text appears.
PRESS ENTER TO CHANGE VALUE
5. Press ↵.
6. Use the ▲ and ▼ keys to change the User level.
7. Press ↵.
8. Press the Lock function push button to exit.

2.7 Restoring of calibration data

If you have ordered a calibration from our calibration laboratory, you will receive a CD with a backup of the calibration data. It is very important that you store these data in your Calibration Software BZ7002. Please store the data before using the Gas Monitor. Please refer to the User Manual BE6034 regarding this procedure.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 25 of 191
Chapter 3 ______________________________________________________________________
Chapter 3
Philosophy of Operation
July 2014
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 26 of 191
Chapter 3 ______________________________________________________________________
The monitor has several essential operational features, which enable you to operate it without undue reference to this Instruction Manual. The main features of the monitor which ensure its simple and easy operation, are listed below:
The fully, comprehensive Software BZ7002 and BZ7003.
The LumaSoft Gas Single Point 7810 application software.
The short, self-explanatory texts, which appear in the monitor
display when the monitor is being used, give operational guidance
to the user.
The small light-emitting diodes at some of the push-buttons allow
the user to identify how the monitor is operating at any particular
time.
Audible warnings “beep” if any push-buttons are used in the
wrong way.

3.1 PC Use – Online

A monitoring system, where you use the PC environment to set-up the monitor prior to measuring, display measurement data on the PC screen while monitoring and store the measurement results directly on the PC’s hard disc. The LumaSoft Gas Single Point 7810 application software is used for making online measurements. The measurement results can be used by other programs that fetch data from the SQL Server Express database server featured by the 7810 application. A manual BE6027 for the 7810 application is available for reference.
Using the online mode (7810 Application Software) provides you with a variety of Software Alarm options. Not only are the alarm trigger levels definable, but the action resulting from an alarm being triggered can also be defined here. This enables you to connect external warning systems to the monitoring system.
The philosophy behind the 7810 Application Software is common to almost all Windows programs. Access to the various levels of the program is gained by use of the MENU BAR, PULL-DOWN MENUS, NEW WINDOW options and INDEX-CARD separators. These familiar procedures, together with the simple software structure, make the instructions in the On-line sections of this manual understandable and easy to implement.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 27 of 191
Chapter 3
CURRENT TIME IS 2011-06-13 17:12
SELECT SET-UP BRANCH
SELECT SET-UP BRANCH
______________________________________________________________________

3.2 Stand-alone Use

This method of operation is available to those of you who wish to use the monitor without connection it to a PC. The monitor is set-up and the results are displayed via the monitor’s front panel.
The structure of the monitor set-up can be viewed like an inverted tree. The main branches divide and separate into minor branches. These in turn divide again and again.
To move through the tree, the branches are selected using the “select” push-buttons (see section 1.5). Use the Set-up Tree, supplied with the monitor and the example steps below to see how the instructions progress. Starting from the initial screen display.
MONITOR IS READY FOR USE
1. Press SET-UP, and the text changes to that displayed below.
MEASUREMENT FORMAT CONFIGURATION
2. To select the MEASUREMENT branch, press S1. (Pressing either
of the other “select” push-buttons will take you along another
branch of the set-up tree. The following text is displayed.
MONITORING TASK ENVIRONMENT
To select the ENVIRONMENT branch, press S3. The text is displayed.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 28 of 191
Chapter 3
SELECT FLUSHING MODE
LENGTH OF SAMPLING TUBE 00.00 m
______________________________________________________________________
AUTO FIXED TIME
3. To select the AUTO flushing mode, press S1. The following text is
displayed.
PRESS ENTER TO CHANGE VALUE
This branch does not divide again. At this stage you are able to define the parameters in this branch of the tree using the direction push-buttons.
If the branch had divided again, it is a simply process of using the correct “select” push-button to continue down the desired branch.
Parameters that are already selected are displayed either with a cursor under the current selection or the value is displayed on the top line.
When the correct value or parameter is already selected, you can continue on through the tree using the direction push-button.
To exit SET-UP mode press SET-UP
Defining parameters is covered in section 6.1.
The procedures in the example above are common to all the situations where the “select” push-buttons can be used to access the different levels in the monitor.
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 29 of 191
Chapter 4 ______________________________________________________________________
Chapter 4
Introduction to and Familiarization
with the Monitor
July 2014
_____________________________________________________________________________ BE6028-14 3434i SF6 Leak Detector LumaSense Technologies A/S Page 30 of 191
Loading...
+ 161 hidden pages