All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without
the prior written authorisation of KROHNE Messtechnik GmbH.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 7
OPTIWAVE 5200 C/F
1.1 Software history
"Firmware revision" agrees with NAMUR NE 53. It is a series of numbers used to record the
revision status of embedded software (firmware) in electronic equipment assemblies. It gives
data on the type of changes made and the effect that changes have on compatibility.
SAFETY INSTRUCTIONS
1
Data about software revisions is shown in menu 1.1.0 IDENT. For more data, refer to
description
of the device (given on the device nameplate) and speak to the supplier.
Release
date
2013-01-11Converter1.00.054000342401k—HB OPTIWAVE
2016-06-27Converter1.00.114000342401pHART commands + auto tests HB OPTIWAVE
Printed circuit
assembly
Sensor1.00.064001025501h
HMI (LCD display option) 1.00.034000487601n
Sensor1.00.164001025501oAuto tests
HMI (LCD display option) 1.00.064000487601rConversion units
on page 76. If it is not possible to refer to the device menu, record the serial number
Firmware
revision
Hardware
revision
Changes and compatibility Documentation
Function
5200 R01 + R02
5200 R03
1.2 Intended use
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and
corrosion resistance of the used materials against the measured fluid lies solely with the
operator.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
This radar level transmitter measures distance, level, mass, volume, flow rate (in open
channels) and reflectivity of liquids, pastes and slurries. It does not touch the measured product.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
7
Page 8
1
SAFETY INSTRUCTIONS
1.3 Certification
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
In accordance with the commitment to customer service and safety, the device
described in this document meets the following safety requirements:
• Electromagnetic Compatibility (EMC) Directive 2014/30/EU. Thus, the device agrees with the
EMC requirements of “Electrical equipment for measurement, control and laboratory use” in
European Standard EN 61326-1.
• European Directive 2014/53/EU (Radio Equipment Directive – RED). The device agrees with
European Standard ETSI EN 302 372. For more data, refer to
• Low-Voltage Directive 2014/35/EU. The device agrees with European Standard EN 61010-1.
• SIL-approved devices agree with European Standard EN 61508.
OPTIWAVE 5200 C/F
European Union (EU)
on page 9.
All devices are based on the CE marking and meet the requirements of NAMUR Guidelines
NE 21, NE 43, NE 53 and NE 107.
8
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 9
OPTIWAVE 5200 C/F
1.4 Radio approvals
1.4.1 European Union (EU)
LEGAL NOTICE!
This level transmitter is intended for installation in closed tanks. It meets the requirements of
the RED (Radio Equipment Directive) 2014/53/EU for use in the member countries of the EU.
An industry agreement includes approval for use of the frequency band (8.5...10.6 GHz) in
industrial environments.
Refer to EN 302 372 for installation conditions.
SAFETY INSTRUCTIONS
1
Figure 1-1: Radio approval information on the nameplate
1 CE sign
According to ETSI EN 302 372-2 (2011), the radiated power outside a metallic tank is less than
-41.3 dBm.
CAUTION!
Obey the regulations in Annex B of ETSI EN 302 372-1 to prevent the emission of radar signals
and radar signal interference.
The device and the tank agree with Annex B of ETSI EN 302 372-1 if you obey the
regulations that follow:
• Install the device in a permanent position on a closed metallic tank, a reinforced concrete
tank or a structure with a closed space made of an equivalent material. Make sure that the
device is correctly attached. We recommend that you install the device on the top of the tank
and the antenna points down.
• Seal flanges and attachments with EMI/RFI shielding gaskets. This gasket is supplied as an
accessory. For the order code, refer to
• Sight glasses must have a microwave-proof layer on them.
• Close all openings and process connections.
Accessories
on page 158.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
9
Page 10
1
SAFETY INSTRUCTIONS
• Only approved personnel can install the device.
For data about how to install EMI/RFI shielding gaskets, refer to the instructions supplied with
this accessory.
The radio approval certificate is given on the DVD-ROM supplied with the device.
1.4.2 U.S.A.
LEGAL NOTICE!
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference which may cause un-
desired operation.
Changes or modifications made to this equipment not expressly approved by KROHNE may void
the FCC authorization to operate this equipment.
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and
used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his own
expense.
OPTIWAVE 5200 C/F
This legal information is shown on the device nameplate.
Figure 1-2: FCC ID number label
1 FCC ID
Devices made in France: Q6BFMCW10G52
Devices made in the USA: JH5FMCW10G52
The radio approval certificate is given on the DVD-ROM supplied with the device.
10
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 11
OPTIWAVE 5200 C/F
1.4.3 Canada
LEGAL NOTICE!
This device complies with Industry Canada licence-exempt RSS 210. Operation is subject to the
following two conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference which may cause un-
desired operation.
Changes or modifications made to this equipment not expressly approved by KROHNE may void
the IC authorization to operate this equipment.
This legal information is shown on the device nameplate.
SAFETY INSTRUCTIONS
1
Figure 1-3: IC number label
1 IC number
Devices made in France: 1991D-FMCW10G52
Devices made in the USA: 1991A-FMCW10G52
The radio approval certificate is given on the DVD-ROM supplied with the device.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
11
Page 12
1
SAFETY INSTRUCTIONS
1.5 Safety instructions from the manufacturer
1.5.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
OPTIWAVE 5200 C/F
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.5.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
12
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 13
OPTIWAVE 5200 C/F
1.5.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation or operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.5.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of icons as shown below.
SAFETY INSTRUCTIONS
1
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
13
Page 14
1
SAFETY INSTRUCTIONS
1.5.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This warning refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
OPTIWAVE 5200 C/F
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.6 Safety instructions for the operator
14
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 15
OPTIWAVE 5200 C/F
2.1 Scope of delivery
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
DEVICE DESCRIPTION
2
Figure 2-1: Scope of delivery
1 Signal converter and antenna (compact version)
2 Quick Start
3 Strap wrench
4 DVD-ROM (including Handbook, Quick Start, Technical Data Sheet and related software)
INFORMATION!
METALLIC HORN AND WAVE GUIDE ANTENNAS
If the device has an antenna extension option, this part is attached to the device if the antenna
extension length, L
≤ 300 mm / 11.8¨. If L
ext
not attached to the device. Obey the assembly procedure on page 41
> 300 mm / 11.8¨, then the antenna extension is
ext
.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
15
Page 16
2
DEVICE DESCRIPTION
2.2 Device description
The FMCW radar level transmitter is designed to measure the distance, level, mass, volume,
flow rate and reflection of liquids, pastes and slurries. Radar level transmitters use an antenna
to guide a signal to the surface of the measured product. Radar is a non-contact technology. It is
particularly suitable for the level measurement of corrosive and viscous products. For more data
about the measuring principle, refer to
The signal converter of the device has 2 versions: Compact and Remote (Field). The converter
can also be ordered with horizontal or vertical housing options for easy access to the device
terminals and the optional display.
Compact version
Compact version
Compact versionCompact version
The signal converter is attached directly to the process connection and the antenna (vertical
housing option shown).
Measuring principle
OPTIWAVE 5200 C/F
on page 108.
Figure 2-2: Compact version
Remote (Field) version
Remote (Field) version
Remote (Field) versionRemote (Field) version
The signal converter is installed away from the process connection (for example, at the bottom
of the tank). The antenna is installed at the top of the tank. The signal cable between the signal
converter and the antenna has a maximum length of 100 m / 328 ft.
Figure 2-3: Remote version
The illustration that follows shows the types of antenna. Antenna extensions are available for
PTFE Wave Horn and Metallic Horn antenna options for long nozzles.
16
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 17
OPTIWAVE 5200 C/F
Figure 2-4: Types of antenna
1 PP Wave Horn antenna (G 1½...2 or 1½...2 NPT threaded connection only)
2 PTFE Wave Horn antenna (DN50...200 / 2¨...8¨ flange connections)
3 Metallic Horn antenna (available antenna sizes: DN80 (3¨), DN100 (4¨), DN150 (6¨) and DN200(8¨))
4 Wave Guide antenna (Length options: 1...6 m / 3.28...19.68 ft, in increments of 0.5 m / 1.64 ft)
DEVICE DESCRIPTION
2
If the process connection temperature is more than +150°C / +302°F, the device has a high-
temperature (HT) extension. This is only available for Metallic Horn and Wave Guide antenna
options.
Figure 2-5: Signal converter with the HT extension
For more data about device options, refer to
Technical data
on page 108.
You can use the LCD display option or PACTware™ software installed on a computer or a
workstation for device configuration. The device configuration menu has a quick set-up wizard.
You usually will not need this document to install, set up and operate the device.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
17
Page 18
2
DEVICE DESCRIPTION
2.3 Visual Check
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
OPTIWAVE 5200 C/F
Figure 2-6: Visual check
1 Device nameplate (for more data, refer to
2 Process connection data (size and pressure rating, material reference and heat number)
3 Gasket material data – refer to the illustration that follows
Figure 2-7: Symbols for the supplied gasket material (on the side of the process connection)
1 EPDM
2 Kalrez® 6375
3 PFA
Nameplate (examples)
on page 19)
If the device is supplied with an FKM/FPM gasket, there is no symbol on the side of the process
connection.
18
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 19
OPTIWAVE 5200 C/F
2.4 Nameplates
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
2.4.1 Nameplate (examples)
DEVICE DESCRIPTION
2
Figure 2-8: Compact (C) and remote (F) versions: Non-Ex nameplate attached to the housing
1 Cable entry size
2 PROFIBUS Device ID (Profile ID) – only available with the PROFIBUS PA output option
3 Hardware revision / Software revision (according to NAMUR NE 53)
4 Signal output (analog, HART®, fieldbus, etc.), input voltage and maximum current (fieldbus options: basic current)
5 Degree of ingress protection (according to EN 60529 / IEC 60529)
6 Customer tag number
7 Date of manufacture
8 Serial number
9 Type code (defined in order)
10 Model name and number. The last letter "X" is either:
C = compact version or
F = remote (field) version
11 Company logo, name and postal address
Country of manufacture / Company web address
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
19
Page 20
2
DEVICE DESCRIPTION
Figure 2-9: Remote (F) version: Non-Ex nameplate attached to the antenna assembly
1 Company web address
2 Country of manufacture
3 Degree of ingress protection (according to EN 60529 / IEC 60529)
4 Date of manufacture
5 Serial number
6 Type code (defined in order)
7 Model name and number. The last letter = "F" remote (field) version
8 Company logo, name and postal address
OPTIWAVE 5200 C/F
20
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 21
OPTIWAVE 5200 C/F
3.1 General notes on installation
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Storage
WARNING!
Do not keep the device in a vertical position. This will damage the antenna and the device will not
measure correctly.
INSTALLATION
3
Figure 3-1: Storage conditions
1 When you put the device into storage, do not keep it in a vertical position
2 Put the device on its side. We recommend that you use the packaging in which it was delivered.
3 Storage temperature range: -50…+85°C/ -58…+185°F (min. -40°C/ -40°F for devices with the integrated LCD display
option)
• Store the device in a dry and dust-free location.
• Store the device in its original packing.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
21
Page 22
3
INSTALLATION
3.3 How to assemble the PTFE Wave Horn antenna
Figure 3-2: Equipment needed to assemble the device
1 Signal converter and process connection (compact version) or antenna housing and process connection (remote ver-
sion)
2 PTFE Wave Horn antenna in cardboard packaging. The cardboard packaging is attached with 4 screws and wingnuts.
Part 1: Remove the protective cover from the process connection
OPTIWAVE 5200 C/F
Figure 3-3: Part 1: Remove the protective cover from the process connection
• Remove the plastic cover from the process connection.
Part2: Remove the cardboard packaging from the PTFE Wave Horn antenna
Figure 3-4: Part2: Remove the cardboard packaging from the PTFE Wave Horn antenna
• Remove the 4 wingnuts and screws that attach cardboard packaging.
• Remove the top and the bottom of the cardboard packaging from the PTFE Wave Horn
antenna.
Part3A: Attach the PTFE Wave Horn antenna to the process connection
22
Figure 3-5: Part3A: Attach the PTFE Wave Horn antenna to the process connection
CAUTION!
Do not use tools to attach the antenna to the process connection.
• Attach the Wave Horn antenna to the process connection. Put the end of the antenna that has
a thread in the fitting on the process connection.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 23
OPTIWAVE 5200 C/F
Part3B: Attach the PTFE Wave Horn antenna to the process connection
Figure 3-6: Part3B: Attach the PTFE Wave Horn antenna to the process connection
CAUTION!
Do not use tools to attach the antenna to the process connection.
• Turn the PTFE Wave Horn antenna. Tighten with your fingers.
• Make sure that the thread is engaged fully.
iEnd of the procedure.
INSTALLATION
3
3.4 Transport
WARNING!
•
Depending on the version, the device will weight approx. 5...30 kg / 11...66 lb. To carry, use
both hands to lift the device carefully by the converter housing. If necessary, lift the device
with a hoist.
•
When handling the device, avoid hard blows, jolts, impact, etc. to prevent damage.
3.5 Pre-installation requirements
INFORMATION!
Obey the precautions that follow to make sure that the device is correctly installed.
• Make sure that there is sufficient space on all sides.
• Protect the signal converter from direct sunlight. If necessary, install the weather protection
accessory.
• Do not subject the signal converter to heavy vibrations.
3.6 Installation
3.6.1 Pressure and temperature ranges
DANGER!
°
If the ambient temperature is more than +70
device. Use a protective cover or metallic grid to prevent injury.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
C / +158°F, there is a risk of injury if you touch the
23
Page 24
3
INSTALLATION
Figure 3-7: Pressure and temperature ranges
1 Temperature at the process connection
Non-Ex devices: The temperature range depends on the type of antenna, process connection and the seal material.
Refer to the table that follows.
Devices with Hazardous Location approvals: see supplementary instructions
2 Ambient temperature for operation of the display
-20...+60°C / -4...+140°F
If the ambient temperature is not between these limits, the display screen switches off automatically. The device continues to operate.
3 Ambient temperature
Non-Ex devices: refer to the ambient temperature / flange temperature graphs in this section
Devices with Hazardous Location approvals: see supplementary instructions
4 Process pressure
Depends on the type of antenna and process connection. Refer to the table that follows.
OPTIWAVE 5200 C/F
WARNING!
The process connection temperature range must agree with the temperature limits of the
gasket material.
Antenna typeProcess
connection
SealProcess connection
temperature
Process pressure
[°C][°F][barg][psig]
PP Wave HornG1½...2;
1½...2 NPT
PTFE Wave HornFlange with
PTFE plate
Metallic Horn
FlangeMetaglas®
Wave Guide
1 Higher temperature on request
2 Higher pressure on request
—-20...+100-4...+212-1...16-14.5...232
—-50...+150-58...+302-1...40-14.5...580
with
-40...+200 1-40...+392
1
-1...40 2-14.5...580
FKM/FPM
Metaglas®
with
-20...+250
1
-4...+482
1
-1...40 2-14.5...580
Kalrez®
6375
Metaglas®
with PFA
Metaglas®
with EPDM
-60...+130 1-76...+266
1
-50...+130 1-58...+266
1
-1...40 2-14.5...580
-1...40 2-14.5...580
2
2
2
2
24
For more data on pressure ratings, refer to
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Pressure ratings
on page 119
Page 25
OPTIWAVE 5200 C/F
PP Wave Horn: compact and remote versions
Ambient temperature / flange temperature, flange and threaded connection, in °C
Figure 3-8: Ambient temperature / flange temperature, flange and threaded connection, in °C
Ambient temperature / flange temperature, flange and threaded connection, in °F
INSTALLATION
3
Figure 3-9: Ambient temperature / flange temperature, flange and threaded connection, in °F
1 Maximum ambient temperature, °C
2 Maximum flange temperature, °C
3 Maximum ambient temperature, °F
4 Maximum flange temperature, °F
5 Compact and remote versions of the device with the PP Wave Horn antenna
6 Remote version of the device with the PP Wave Horn antenna
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
25
Page 26
3
INSTALLATION
PTFE Wave Horn: compact and remote versions
Ambient temperature / flange temperature, flange and threaded connection, in °C
Figure 3-10: Ambient temperature / flange temperature, flange and threaded connection, in °C
Ambient temperature / flange temperature, flange and threaded connection, in °F
OPTIWAVE 5200 C/F
Figure 3-11: Ambient temperature / flange temperature, flange and threaded connection, in °F
1 Maximum ambient temperature, °C
2 Maximum flange temperature, °C
3 Maximum ambient temperature, °F
4 Maximum flange temperature, °F
5 Compact and remote versions of the device with the PTFE Wave Horn antenna
6 Compact version of the device with the PTFE Wave Horn antenna
INFORMATION!
°
When the process temperature is -50
temperature. The minimum ambient temperature is -36
C/ -58°F, there is a de-rating for the ambient
°
C / -32.8°F.
26
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 27
OPTIWAVE 5200 C/F
Metallic Horn and Wave Guide: compact and remote versions
Ambient temperature / flange temperature, flange and threaded connection, in °C
Figure 3-12: Ambient temperature / flange temperature, flange and threaded connection, in °C
Ambient temperature / flange temperature, flange and threaded connection, in °F
INSTALLATION
3
Figure 3-13: Ambient temperature / flange temperature, flange and threaded connection, in °F
1 Maximum ambient temperature, °C
2 Maximum flange temperature, °C
3 Maximum ambient temperature, °F
4 Maximum flange temperature, °F
5 Compact and remote versions of the device with the Metallic Horn or Wave Guide antenna
6 Compact version of the device with the Metallic Horn or Wave Guide antenna
INFORMATION!
°
When the process temperature is -50
C/ -58°F and the device has an EPDM gasket, there is a
de-rating for the ambient temperature. The minimum ambient temperature is -35
°
C/ -31°F.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
27
Page 28
3
INSTALLATION
Metallic Horn and Wave Guide: compact and remote versions with the hightemperature (HT) extension
Ambient temperature / flange temperature, flange and threaded connection, in °C
Figure 3-14: Ambient temperature / flange temperature, flange and threaded connection, in °C
Ambient temperature / flange temperature, flange and threaded connection, in °F
OPTIWAVE 5200 C/F
Figure 3-15: Ambient temperature / flange temperature, flange and threaded connection, in °F
1 Maximum ambient temperature, °C
2 Maximum flange temperature, °C
3 Maximum ambient temperature, °F
4 Maximum flange temperature, °F
5 Compact and remote versions of the device with the Metallic Horn or Wave Guide antenna, the HT extension and EPDM
and PFA gaskets
6 Compact and remote versions of the device with the Metallic Horn or Wave Guide antenna, the HT extension and a PFA
gasket
INFORMATION!
°
When the process temperature is -50
de-rating for the ambient temperature. The minimum ambient temperature is -39
When the process temperature is -60
rating for the ambient temperature. The minimum ambient temperature is -37
C/ -58°F and the device has an EPDM gasket, there is a
°
C / -38.2°F.
°
C/ -76°F and the device has a PFA gasket, there is a de-
°
C / -34.6°F.
28
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 29
OPTIWAVE 5200 C/F
3.6.2 Recommended mounting position
CAUTION!
Follow these recommendations to make sure that the device measures correctly. They have an
effect on the performance of the device.
INSTALLATION
3
Figure 3-16: Recommended mounting position for liquids, pastes and slurries
1 Sockets for the PP Wave Horn antenna
2 Nozzles for the PTFE Wave Horn antenna
3 Nozzles for DN150 or DN200 Metallic Horn antennas
4 Nozzles for Wave Guide antennas
5 Tank diameter
6 Minimum distance of the nozzle or socket from the tank wall (depends on the antenna type and size - refer to items
1, 2 , 3 and 4 in this list):
- PP/PTFE Wave Horn (1 and 2): 1/7 × tank height
- Metallic Horn (3): 1/10 × tank height
- Wave Guide (4): There is no minimum distance from the Wave Guide antenna to metallic walls and other metal ob-
jects
Maximum distance of nozzle from the tank wall (depends on the antenna type and size - refer to items 1, 2 and 3 in
this list):
- PP/PTFE Wave Horn (1 and 2): 1/3 × tank diameter
- Metallic Horn (3): 1/3 × tank diameter
- Wave Guide (4): There is no maximum distance from the Wave Guide antenna to metallic walls and other metal ob-
jects
7 Tank height
INFORMATION!
If there is a nozzle on the tank before installation, the nozzle must be a minimum of 200 mm /
¨
from the tank wall. The tank wall must be flat and there must not be obstacles adjacent to
7.9
the nozzle or on the tank wall.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
29
Page 30
3
INSTALLATION
Point the device in the correct direction
Figure 3-17: Point the device in the correct direction to get the best performance
Point the tag hole on the housing in the direction of the nearest tank wall.
1 Tag hole
2 Nearest tank wall
OPTIWAVE 5200 C/F
Figure 3-18: A maximum of 4 FMCW radar level meters can be operated in a tank
Figure 3-19: Tanks with dish-shaped or conical bottoms
Dish-shaped or conical bottoms have an effect on the measuring range. The device cannot measure to the bottom of the
tank.
1 Axis of radar beam
2 Minimum level reading
30
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 31
OPTIWAVE 5200 C/F
3.6.3 Mounting restrictions
CAUTION!
Follow these recommendations to make sure that the device measures correctly. They have an
effect on the performance of the device.
We recommend that you prepare the installation when the tank is empty.
Mounting restrictions: General data
INSTALLATION
3
Figure 3-20: Mounting restrictions: General data
1 Do not tilt the device more than 2°
2 We recommend that you do an empty spectrum recording if there are obstacles in the radar beam (refer to Operation
If necessary, install a bypass chamber or stilling well or use an S-bend antenna extension or an L-bend antenna extension (the device must be installed on the side of the tank) to move the device away from obstacles.
3 2.5 mm / 0.1¨ max. for high-dielectric constant liquids
4 Beam radius (DN80 (3¨) Metallic Horn antenna): increments of 290 mm/m or 3.4¨/ft (16°)
Beam radius (DN100 (4¨) Metallic Horn antenna): increments of 210 mm/m or 2.6¨/ft (12°)
Beam radius (DN150 (6¨) Metallic Horn antenna): increments of 140 mm/m or 1.7¨/ft (8°)
Beam radius (DN200 (8¨) Metallic Horn antenna): increments of 100 mm/m or 1.3¨/ft (6°)
Beam radius (PP Wave Horn and PTFE Wave Horn antenna): increments of 176 mm/m or 2.1¨/ft (10°)
Operation).
OperationOperation
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
31
Page 32
3
INSTALLATION
Obstacles in the tank
Figure 3-21: Obstacles in the tank
Do not put the device directly above obstacles (agitator, support beams, heating tubes etc.). Parasitic signals from obstacles will cause the device to measure incorrectly.
1 Solution 1: Put the device on another process connection away from obstacles
2 Solution 2: Use the same process connection, but also use an S-bend extension
3 Solution 3: Attach the device to the side of the tank and use an L-bend (right angle) extension
OPTIWAVE 5200 C/F
CAUTION!
Do not put the device near to the product inlet. If the product that enters the tank touches the
antenna, the device will measure incorrectly. If the product fills the tank directly below the
antenna, the device will also measure incorrectly.
Figure 3-22: Product inlets
1 The device is in the correct position.
2 The device is too near to the product inlet.
32
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 33
OPTIWAVE 5200 C/F
Devices with Metallic Horn antenna
Figure 3-23: Devices with Metallic Horn antenna
1 If the roof is not flat, the antenna must project out of the nozzle
2 Short tank nozzle
3 Long tank nozzle (device with an antenna extension)
4 If the roof is flat and the tank fitting is symmetrical, it is not necessary for the antenna to project out of the nozzle.
Thus, the device can have a larger measuring range.
INSTALLATION
3
The antenna must project out of the nozzle. If necessary, use an antenna extension. But if the
tank roof is flat and the tank fitting is symmetrical, it is not necessary for the antenna to project
out of the nozzle. Thus, the device can have a larger measuring range.
Devices with PTFE or PP Wave Horn antenna
Figure 3-24: Devices with PTFE or PP Wave Horn antenna
Recommended height of tank process connection, a = 44...200 mm / 1.7...7.87¨1 Device with a PTFE Wave Horn antenna and a flange connection. There are antenna extension options available
(100 mm / 4¨, 200 mm / 8¨ and 300 mm / 12¨) for long nozzles.
2 Device with a PP Wave Horn antenna and a thread connection
CAUTION!
Do not install devices with Wave Horn antennas on plastic nozzles that have a diameter of
¨
50 mm / 2
.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
33
Page 34
3
INSTALLATION
CAUTION!
If there are parasitic signals, the device will not measure correctly. Parasitic signals are caused
by:
•
Objects in the tank.
•
Sharp corners that are perpendicular to the path of the radar beam.
•
Sudden changes in tank diameter in the path of the radar beam.
OPTIWAVE 5200 C/F
Do an Empty Spectrum recording (refer to Operation
Operation) to remove parasitic signals with a filter.
OperationOperation
Requirements for flange connections
Figure 3-25: Flange connection
Equipment needed:
• Device
• Flange gasket (not supplied)
• Wrench (not supplied)
• Make sure the flange on the nozzle is level.
• Make sure that you use the applicable gasket for the flange dimensions and the process.
• Align the gasket correctly on the flange facing of the nozzle.
• Lower the antenna carefully into the tank.
• Make sure that you point the device in the correct direction. Refer to "Point the device in the
correct direction" in this section.
• Tighten the flange bolts.
iRefer to local rules and regulations for the correct torque to apply to the bolts.
34
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 35
OPTIWAVE 5200 C/F
Requirements for threaded connections
Figure 3-26: Threaded connection
Equipment needed:
• Device
• Gasket for G 1½A or G 2A connection (not supplied)
• Thread seal tape (PTFE) for 1½ NPT or 2 NPT connection (not supplied)
• 50 mm / 2¨ wrench (not supplied)
INSTALLATION
3
WARNING!
·
Do not tighten the connection to a torque more than 40 N
m / 29.5 lb·ft. If the connection is too
tight, this will damage the thread.
To prevent damage to the antenna, make sure that the minimum diameter of the hole for a
½
NPT or 2 NPT thread connection is not less than 43.4 mm / 1.71¨.
1
• Make sure the tank connection is level.
• ISO 228-1 (G) connection:
ISO 228-1 (G) connection: Make sure that you use the applicable gasket for the connection
ISO 228-1 (G) connection:ISO 228-1 (G) connection:
dimensions and the process.
• ISO 228-1 (G) connection:
ISO 228-1 (G) connection: Align the gasket correctly.
ISO 228-1 (G) connection:ISO 228-1 (G) connection:
• NPT connection:
NPT connection: Wind the thread seal tape around the process connection in agreement with
NPT connection:NPT connection:
good engineering practice.
• Lower the antenna carefully into the tank.
• Turn the threaded connection on the antenna to attach the device to the process connection.
• Make sure that you point the device in the correct direction. Refer to "Point the device in the
correct direction" in this section.
• Tighten the connection to the correct torque (not more than 40 Nm).
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
35
Page 36
3
INSTALLATION
3.6.4 Standpipes (stilling wells and bypass chambers)
Use a standpipe if:
• There is highly conductive foam in the tank.
• The liquid is very turbulent or agitated.
• There are too many other objects in the tank.
• The device is measuring a liquid (petro-chemicals) in a tank with a floating roof.
• The device is installed in a horizontal cylindrical tank (refer to the end of this section).
OPTIWAVE 5200 C/F
Figure 3-27: Installation recommendations for standpipes (stilling wells and bypass chambers)
1 A stilling well solution
2 A bypass chamber solution
3 Air circulation hole
4 Level of the liquid
CAUTION!
•
The standpipe must be electrically conductive.
•
The inside diameter of the standpipe must not be more than 5 mm / 0.2¨ over the diameter of
the antenna (for a high-dielectric constant liquid).
•
The standpipe must be straight. There must be no sudden changes in internal diameter
¨
greater than 1 mm / 0.04
•
The standpipe must be vertical.
•
Recommended surface roughness: <±0.1 mm / 0.004¨.
•
Make sure that there are no deposits at the bottom of the standpipe.
•
Make sure that there is liquid in the standpipe.
.
36
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 37
OPTIWAVE 5200 C/F
Stilling wells – general notes
CAUTION!
You can drill an air circulation hole if the stilling well is in a closed space (a metallic container).
This condition is necessary to obey TLPR standards. For more data, refer to Radio approvals on
.
page 9
Installation in tanks containing one liquid and foam
• Drill an air circulation hole (max. Ø10 mm / 0.4¨) in the stilling well above the maximum level.
• Remove the burr from the hole.
Installation in tanks containing one liquid or more without foam
• Drill an air circulation hole (max. Ø10 mm / 0.4¨) in the stilling well above the maximum level.
• Drill 1 or more liquid circulation holes in the stilling well (if there is more than 1 liquid in the
tank).
iThese holes help the liquid to move freely between the stilling well and the tank.
• Remove the burr from the hole.
Stilling wells – reference point
We recommend that you attach a flat cruciform target at the bottom of the stilling well. It must
have a width of 1/3 of the internal diameter of the stilling well tube. This target is the limit of the
measuring range in the stilling well.
INSTALLATION
3
Figure 3-28: Reference point in the stilling well
a ≥ 1/3 × ØD
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
37
Page 38
3
INSTALLATION
Stilling wells: floating roofs
If the device must be installed on a tank with a floating roof, install it in a stilling well.
OPTIWAVE 5200 C/F
Figure 3-29: Floating roofs
1 Sediment
2 Support fixtures
3 Stilling well
4 Floating roof
5 Product
6 Tank
38
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 39
OPTIWAVE 5200 C/F
Stilling wells: horizontal cylindrical tanks
We recommend that you install the device in a stilling well if the device:
• is for a horizontal cylindrical tank,
• is in a metallic tank,
• measures a product with a high dielectric constant and
• is on the centerline of the tank.
INSTALLATION
3
Figure 3-30: Horizontal cylindrical tanks
1 The device is installed without a stilling well. There are multiple reflections. Refer to the CAUTION! that follows.
2 The device is installed in a stilling well and measures correctly.
CAUTION!
If the device is installed in horizontal cylindrical tank that contains a high dielectric constant
liquid without a stilling well, do not put it on the tank centerline. This will cause multiple
reflections and the device will not measure accurately. Use the 2.3.12 Multiple Reflections
function in Supervisor > Basic Parameters
Supervisor > Basic Parameters to keep the effects of multiple reflections to a
Bypass chambers
Installation next to tanks containing one liquid and foam
• The top process connection of the bypass chamber must be above the maximum level of
liquid.
• The bottom process connection of the bypass chamber must be below the lowest measured
level of liquid.
Installation next to tanks containing more than one liquid
• The top process connection of the bypass chamber must be above the maximum level of
liquid.
• The bottom process connection of the bypass chamber must be below the lowest measured
level of liquid.
• Additional process connections are necessary for the liquids to circulate freely along the
length of the bypass chamber.
3.6.5 Wall support for the remote version
OPTIWAVE 5200 C/F
Figure 3-31: Wall support for the remote version (attached to the remote converter)
1 Use marks on the wall to help you put the wall support in the correct position. For more data,
refer to
2 Use equipment and tools that agree with health and safety regulations and good engineering
practice.
3 Make sure the wall support is correctly attached to the wall.
Dimensions and weights
on page 126.
40
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 41
OPTIWAVE 5200 C/F
INSTALLATION
3.6.6 How to attach an antenna extension (Metallic Horn or Wave Guide antennas)
CAUTION!
IF THE ANTENNA EXTENSION IS SUPPLIED WITH THE DEVICE AS AN OPTION:
If the antenna extension is attached to the device, no more work is necessary.
If the antenna extension is not attached to the device, refer to Procedure 1 in this section. It is not
necessary to change the device settings. The manufacturer sets the related menu items to the
correct values in the factory.
IF THE ANTENNA EXTENSION IS SUPPLIED AFTER DELIVERY OF THE DEVICE (SPARE PART):
You must attach the antenna extension to the device and change the device settings in the
SUPERVISOR menu. If you have a straight antenna extension, refer to Procedures 1 and 2A. If you
have an S-bend or L-bend (right-angle) antenna extension, refer to Procedures 1 and 2B.
DANGER!
If you installed the device on the tank before the procedure, make the device safe (de-energize
the circuit, clean the device etc.) before you continue the work.
Equipment needed
3
Figure 3-32: Equipment needed
1 Device (with a Metallic Horn or Wave Guide antenna option only)
2 Antenna extension. From left to right: straight, S-bend and L-bend (right-angle) extension.
3 10 mm open-end or box-end wrench (not supplied)
4 3 hex head screws M6×25-A4-70 (supplied in a plastic sachet)
Obey the assembly procedure that follows:
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
41
Page 42
3
INSTALLATION
OPTIWAVE 5200 C/F
Figure 3-33: Procedure: How to attach an antenna extension
WARNING!
Make sure that you do not damage the PTFE cone when you remove or install the antenna.
Procedure 1: How to attach an antenna extension
1 Remove the 3 hex head screws from the antenna with the 10 mm wrench.
2 Remove the antenna. Make sure that you do not damage the PTFE cone when you remove the
antenna.
3 Attach the antenna extension below the flange. Make sure that the antenna extension is fully
engaged. Make sure that you do not damage the PTFE cone when you attach the antenna extension.
4 Attach 3 hex head screws to the antenna extension with a 10 mm wrench. Tighten the screws
to a torque of 8 Nm.
5 Attach the antenna below the antenna extension.
6 Attach 3 hex head screws to the antenna with a 10 mm wrench. Tighten the screws to a torque
of 8 Nm.
iIf the antenna extension is supplied with the device, it is not necessary to change the device
settings. End of the procedure.
7 If the antenna extension is supplied after delivery of the device, it is necessary to change the
device settings. Make a selection from one of the procedures (2A or 2B) that follow: straight
42
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Procedure 2A: Device settings for a device with a straight antenna extension
• Enter the SUPERVISOR menu (2.0.0).
• Push [>>>>], 2 × [], [>>>>] and 6 × [] to go to menu item ANTENNA EXTENSION (2.3.7).
• Push [>>>>] to change the value. Push [>>>>] to change the position of the cursor. Push [] to
decrease the value or [] to increase the value.
iIf the antenna extension has a length of 500 mm, enter the value "500" (if units for this
menu item are in mm).
• Push 3 × [^^^^] to go back to the "STORE" screen.
• Push [] or [] to set the screen to STORE YES
iEnd of the procedure.
Procedure 2B: Device settings for a device with an S-bend or L-bend (right angle)
antenna extension
• Enter the SUPERVISOR menu (2.0.0).
• Push [>>>>], 2 × [], [>>>>] and 6 × [] to go to menu item ANTENNA EXTENSION (2.3.7).
• Push [>>>>] to change the value. Push [>>>>] to change the position of the cursor. Push [] to
decrease the value or [] to increase the value.
iIf units are in mm, enter the value "221" (for an S-bend extension) or "236" (for an L-bend
extension).
• Push [^^^^] to go back to the menu. Push 2 × [] to go to menu item DIST.PIECE (2.3.9).
• Push [>>>>] to change the value. Push [>>>>] to change the position of the cursor. Push [] to
decrease the value or [] to increase the value.
iIf units are in mm, enter the value "243" (for an S-bend extension) or "236" (for an L-bend
extension).
• Push 3 × [^^^^] to go back to the "STORE" screen.
• Push [] or [] to set the screen to STORE YES
iEnd of the procedure.
STORE YES and push [^^^^].
STORE YESSTORE YES
STORE YES and push [^^^^].
STORE YESSTORE YES
INSTALLATION
3
Settings for devices with antenna extensions in mm
Antenna extension typeDevice settings
ANTENNA EXTENSION (2.3.7)DIST.PIECE (2.3.9)
Straight
S-bend221243
L-bend (right angle)236236
1 This value depends on the length of the antenna extension. Enter the length of the antenna extension: 100, 200, 300,
400, 500 or 1000 mm
2 If the device has a high temperature extension, add 120 mm to this value
For more data about the dimensions of devices with antenna extensionsFor more data about the
dimensions of devices with antenna extensions, refer to
1
Dimensions and weights
0
2
2
2
on page 126.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
43
Page 44
3
INSTALLATION
3.6.7 How to turn or remove the signal converter
The converter turns 360°, but we recommend that the tag hole on the housing points to the
nearest tank wall. For more data, refer to
converter can be removed from the process connection assembly under process conditions.
Recommended mounting position
OPTIWAVE 5200 C/F
on page 29. The
Figure 3-34: How to turn or remove the signal converter
1 Tool: 5 mm Allen wrench (not supplied) for the lock screw on the signal converter
2 Cover for the wave guide hole on top of the process connection assembly (not supplied)
CAUTION!
Do not loosen the 4 socket head screws on the process connection assembly.
If you remove the housing, put a cover on the wave guide hole on top of the process connection
assembly. Make sure that the wave guide hole is clean and dry.
When the housing is attached to the process connection assembly, tighten the lock screw with
the 5 mm Allen wrench 1.
44
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 45
OPTIWAVE 5200 C/F
3.6.8 How to attach the weather protection to the device
The device and the weather protection option are supplied disassembled in the same box. The
weather protection can also be supplied as an accessory. You must attach the weather
protection when you install the device.
INSTALLATION
3
Figure 3-35: Equipment needed
1 Weather protection cover (with an R-clip to hold the cover on the clamp)
2 Device
3 Weather protection clamp (2 parts)
4 10 mm socket wrench (not supplied)
5 2 locking nuts
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
45
Page 46
3
INSTALLATION
OPTIWAVE 5200 C/F
46
Figure 3-36: Installation of the weather protection on a vertical signal converter
• Put the weather protection clamp around the top of the device.
• Attach the two locking nuts to the threads on the weather protection clamp. Tighten the
locking nuts with a 10 mm socket wrench.
• Lower the weather protection cover onto weather protection clamp until the hole for the lock
is in the slot at the front of the cover.
• Put the R-clip into the hole at the front of the weather protection cover.
• End of the procedure.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 47
OPTIWAVE 5200 C/F
INSTALLATION
3
Figure 3-37: Installation of the weather protection on a horizontal signal converter
• Put the weather protection clamp around the front of the device (the end of the device that is
nearest to the cable entry).
• Attach the two locking nuts to the threads on the weather protection clamp. Tighten the
locking nuts with a 10 mm socket wrench.
• Lower the weather protection cover onto weather protection clamp until the hole for the lock
is in the slot at the front of the cover.
• Put the R-clip into the hole at the front of the weather protection cover.
• End of the procedure.
The overall dimensions of the weather protection are on page 126.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
47
Page 48
3
INSTALLATION
3.6.9 How to open the weather protection
OPTIWAVE 5200 C/F
Figure 3-38: How to open the weather protection
1 Remove the R-clip from the hole at the front of the weather protection cover.
2 Remove the weather protection cover.
iLift the display screen cover.
48
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 49
OPTIWAVE 5200 C/F
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
ELECTRICAL CONNECTIONS
4
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
4.2 General notes
This chapter includes electrical connection data about devices with the 4...20mA output and
HART® communication options.
INFORMATION!
Devices with the FOUNDATION
Devices with the FOUNDATION
Devices with the FOUNDATIONDevices with the FOUNDATION
For electrical connection data, refer to the "Description of FOUNDATION
supplementary instructions.
INFORMATION!
Devices with the PROFIBUS PA output option:
Devices with the PROFIBUS PA output option:
Devices with the PROFIBUS PA output option:Devices with the PROFIBUS PA output option:
For electrical connection data, refer to the "Description of PROFIBUS PA interface"
supplementary instructions.
™
fieldbus output option:
fieldbus output option:
fieldbus output option: fieldbus output option:
™
fieldbus interface"
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
49
Page 50
4
ELECTRICAL CONNECTIONS
4.3 Electrical installation: 2-wire, loop-powered
4.3.1 Compact version
Terminals for electrical installation
Figure 4-1: Terminals for electrical installation
1 Grounding terminal in the housing (if the electrical cable is shielded)
2 Current output -
3 Current output +
4 Location of the external grounding terminal (at the bottom of the converter)
OPTIWAVE 5200 C/F
INFORMATION!
Electrical power to the output terminal energizes the device. The output terminal is also used for
®
HART
communication.
CAUTION!
•
Use the applicable electrical cables with the cable glands.
•
Make sure that the current is not more than 5 A or that there is 5 A-rated fuse in the
electrical circuit that energizes the device.
•
Make sure that the polarity of the power supply is correct. If the polarity is incorrect, you will
not cause damage to the device but the device will not operate.
50
Figure 4-2: How to open the terminal compartment cover
1 Loosen the lock screw with a 2.5 mm Allen wrench.
2 Turn the cover counterclockwise with a strap wrench.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 51
OPTIWAVE 5200 C/F
3 Remove the cover.
Figure 4-3: Procedure for electrical installation
Equipment needed:
• Small slotted tip screwdriver (not supplied)
ELECTRICAL CONNECTIONS
4
Procedure:
1 Do not disconnect the safety cord from the terminal compartment cover. Put the terminal
compartment cover adjacent to the housing.
2 Remove the connector from the circuit board.
3 Connect the electrical wires to the connector. Attach the connector to the circuit board. Tight-
en the cable entry glands.
Figure 4-4: How to close the terminal compartment cover
1 Put the cover on the housing and push it down.
2 Turn the cover clockwise until it is fully engaged.
3 Tighten the lock screw.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
51
Page 52
4
ELECTRICAL CONNECTIONS
4.3.2 Remote version
Terminals for electrical installation
Figure 4-5: Terminals for electrical installation
1 Grounding terminal in the housing (if the electrical cable is shielded)
2 Current output -
3 Current output +
4 Location of the external grounding terminal (on the wall support)
OPTIWAVE 5200 C/F
INFORMATION!
Electrical power to the output terminal energizes the device. The output terminal is also used for
®
HART
communication.
CAUTION!
•
Use the applicable electrical cables with the cable glands.
•
Make sure that the current is not more than 5 A or that there is 5 A-rated fuse in the
electrical circuit that energizes the device.
•
Make sure that the polarity of the power supply is correct. If the polarity is incorrect, you will
not cause damage to the device but the device will not operate.
For more electrical installation data, refer to
Compact version
on page 50.
52
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 53
OPTIWAVE 5200 C/F
ELECTRICAL CONNECTIONS
4.4 Remote device data
4.4.1 Requirements for signal cables supplied by the customer
DANGER!
An Ex-approved signal cable is supplied by the manufacturer with devices for hazardous
locations. Use of this signal cable is mandatory.
Non-Ex devices only:
Non-Ex devices only: The signal cable is an option for non-Ex devices. If the signal cable is not
Non-Ex devices only: Non-Ex devices only:
supplied by the device manufacturer, the cable must have properties that follow:
Basic properties
• Twisted cable 2 by 2, shielded or screened. For example, multicore cable — reference
MCD 5123 — from Cabletec ICS/JP Electronics.
Maximum length of the signal cable
• 100 m / 328 ft
4
Temperature
• Use electrical cable with the applicable temperature rating for the operating conditions.
• Ambient temperature range: -40...+80°C / -40...+176°F
• We recommend that the cable agrees with UL 94V-0.
Dimensions of the insulated conductors
• Min.-max. cross-sectional area of the conductors: 4×0.326...4×2.5 mm² (22....14 AWG),
shielded cable
• Use the applicable cable for the cable glands (Ø6....10 mm / 0.24...0.39¨).
• Use the applicable cable glands for the cable entry openings in the housing.
• The cable must agree with EN 60811 (Low Voltage Directive) or equivalent national
regulations.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
53
Page 54
4
ELECTRICAL CONNECTIONS
4.4.2 How to prepare a signal cable supplied by the customer
OPTIWAVE 5200 C/F
Figure 4-6: Equipment needed to prepare the signal cable
1 Signal cable (supplied on request)
2 2 heat-shrinkable sleeves for the PVC jacket (not supplied)
3 8 ferrules for the end of the conductors (not supplied)
4 2 Faston connectors for the shield wires
5 Shield wire insulation, 2 sleeves
6 Wire stripper (not supplied)
7 Crimping pliers (not supplied)
INFORMATION!
•
The Faston connector for the stranded drain wire must agree with DIN 46 228: E 1.5-8
•
The wire end ferrules for the twisted pair of conductors must agree with DIN 46 228: E 0.5-8
54
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 55
OPTIWAVE 5200 C/F
Figure 4-7: How to prepare the signal cable
ELECTRICAL CONNECTIONS
4
1 Remove the PVC jacket from the wire to dimension "a". a = 50 mm / 2¨.
2 Remove the insulation from the wire. Obey national regulations for electrical wiring.
3 Crimp the wire end ferrules on the conductors.
4 Install shield wire insulation on the 2 ends of the shield wire.
5 Crimp the Faston connectors on the 2 ends of the shield wire.
6 Install a heat-shrinkable sleeve on the PVC jacket.
4.4.3 How to connect the signal cable to the device
DANGER!
Cables may only be connected when the power is switched off.
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
CAUTION!
Do not wind the signal cable. This configuration will prevent interference from electromagnetic
fields.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
55
Page 56
4
ELECTRICAL CONNECTIONS
Equipment needed
OPTIWAVE 5200 C/F
Figure 4-8: Equipment needed to prepare the signal cable
1 Remote converter
2 Antenna housing
3 Signal cable (supplied on request for non-Ex devices) – for more data refer to, refer to
supplied by the customer
4 Small slotted-tip screwdriver (not supplied)
on page 54
How to prepare a signal cable
Connections between the remote converter and the antenna housing
Figure 4-9: Connections between the remote converter and the antenna housing
1 Remote converter
2 Antenna housing
3 Power supply: voltage in -
4 Power supply: voltage in +
5 Signal cable B
6 Signal cable A
7 Shielding wire (attached to Faston connectors in the housings of the remote converter and the antenna housing)
56
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 57
OPTIWAVE 5200 C/F
How to connect the signal cable to the remote converter
ELECTRICAL CONNECTIONS
4
Figure 4-10: How to connect the signal cable to the remote converter
CAUTION!
Bending radius of the signal cable: ≥ 50 mm / 2
1 Remove the terminal compartment cover.
2 Remove the 4-pin connector.
3 Put the signal cable into the opening of the cable gland.
4 Put the electrical wires in the connector terminals. Tighten the terminal screws with a small
slotted-tip screwdriver. Make sure that the electrical wires agree with the terminals. For
more data, refer to the electrical schema in this section.
5 Put the connector into the 4-pin socket.
6 Attach the Faston connector (drain wire).
7 Attach the terminal compartment cover.
8 Tighten the cable gland. Make sure that the remote converter is correctly sealed.
¨
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
57
Page 58
4
ELECTRICAL CONNECTIONS
How to connect the signal cable to the antenna housing
OPTIWAVE 5200 C/F
Figure 4-11: How to connect the signal cable to the antenna housing
CAUTION!
Bending radius of the signal cable: ≥ 50 mm / 2
1 Remove the terminal compartment cover.
2 Remove the 4-pin connector.
3 Put the signal cable into the opening of the cable gland.
4 Put the electrical wires in the connector terminals. Tighten the terminal screws with a small
slotted-tip screwdriver. Make sure that the electrical wires agree with the terminals. For
more data, refer to the electrical schema in this section.
5 Put the connector into the 4-pin socket. Attach the Faston connector (drain wire).
6 Attach the terminal compartment cover.
7 Tighten the cable gland. Make sure that the probe housing is correctly sealed.
¨
58
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 59
OPTIWAVE 5200 C/F
ELECTRICAL CONNECTIONS
4.5 Electrical connection for current output
4.5.1 Non-Ex devices
Figure 4-12: Electrical connections for non-Ex devices
1 Power supply
2 Resistor for HART® communication
3 Optional connection to the grounding terminal
4 Output: 11.5...30 VDC for an output of 22 mA at the terminal
5 Device
4
4.5.2 Devices for hazardous locations
DANGER!
For electrical data for device operation in hazardous locations, refer to the related certificates of
compliance and supplementary instructions (ATEX, IECEx, etc.). You can find this documentation
on the DVD-ROM delivered with the device or it can be downloaded free of charge from the
website (Download Center).
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
59
Page 60
4
ELECTRICAL CONNECTIONS
4.6 Protection category
INFORMATION!
The device fulfils all requirements per protection category IP66 / IP67. It also fulfils all
requirements per NEMA type 4X (housing) and type 6P (antenna).
DANGER!
Make sure that the cable gland is watertight.
OPTIWAVE 5200 C/F
Figure 4-13: How to make the installation agree with protection category IP67
• Make sure that the gaskets are not damaged.
• Make sure that the electrical cables are not damaged.
• Make sure that the electrical cables agree with the national electrical code.
• The cables are in a loop in front of the device 1 so water does not go into the housing.
• Tighten the cable feedthroughs 2.
• Close unused cable feedthroughs with dummy plugs 3.
Refer to the table that follows for the diameter of the outer sheath of the electrical cable:
Min. / Max. diameter of the electrical cable
Type of electrical cableApprovalMin. / Max. diameter of the electrical cable
[mm][inches]
Power supply / outputnon-Ex / Ex i6...7.50.24...0.3
Power supply / outputExd6...100.24...0.39
Signal cable (for the remote
version)
1 This electrical cable is connected between the remote converter and the antenna housing
1
non-Ex / Ex i / Ex d6...100.24...0.39
60
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 61
OPTIWAVE 5200 C/F
4.7 Networks
4.7.1 General information
The device uses the HART® communication protocol. This protocol agrees with the HART®
Communication Foundation standard. The device can be connected point-to-point. It can also
have a polling address of 1 to 63 in a multi-drop network.
The device output is factory-set to communicate point-to-point. To change the communication
mode from point-to-point
4.7.2 Point-to-point connection
point-to-point to multi-drop
point-to-pointpoint-to-point
ELECTRICAL CONNECTIONS
multi-drop, refer to
multi-dropmulti-drop
HART
® network configuration on page 88.
4
Figure 4-14: Point-to-point connection (non-Ex)
1 Address of the device (0 for point-to-point connection)
2 4...20 mA + HART®
3 Resistor for HART® communication
4 Power supply
5 HART® converter
6 HART® communication software
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
61
Page 62
4
ELECTRICAL CONNECTIONS
4.7.3 Multi-drop networks
OPTIWAVE 5200 C/F
Figure 4-15: Multi-drop network (non-Ex)
1 Address of the device (each device must have a different address in multidrop networks)
2 4mA + HART®
3 Resistor for HART® communication
4 Power supply
5 HART® converter
6 HART® communication software
62
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 63
OPTIWAVE 5200 C/F
4.7.4 Fieldbus networks
For more data, refer to the supplementary instructions for FOUNDATION™ fieldbus and
1 Field device
2 Junction box
3 H1 network
4 H1/HSE converter
5 High Speed Ethernet (HSE)
6 Workstation
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
63
Page 64
4
ELECTRICAL CONNECTIONS
PROFIBUS PA/DP network (non-Ex)
OPTIWAVE 5200 C/F
Figure 4-17: PROFIBUS PA/DP network (non-Ex)
1 Field device
2 Bus termination
3 PROFIBUS PA bus segment
4 Segment coupler (PA/DP link)
5 PROFIBUS DP bus line
6 Control system (PLC / Class 1 master device)
7 Engineering or operator workstation (Control tool / Class 2 master device)
64
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 65
OPTIWAVE 5200 C/F
5.1 How to start the device
5.1.1 Start-up checklist
Check these points before you energize the device:
• Are all the wetted components (antenna, flange and gaskets) chemically resistant to the
product in the tank?
• Does the information on the signal converter nameplate agree with the operating data?
• Did you correctly install the device on the tank?
• Do the electrical connections agree with the national electrical codes? Use the applicable
electrical cables with the cable glands.
DANGER!
Before you energize the device, make sure that the supply voltage and polarity are correct.
DANGER!
If the device has a Hazardous Locations approval, make sure that the device and the installation
agrees with the requirements of the certificate of compliance.
START-UP
5
5.1.2 How to start the device
• Connect the converter to the power supply.
• Energize the converter.
iDevices with the LCD display option only:
Devices with the LCD display option only: After 10 seconds the screen will display "Starting
Devices with the LCD display option only:Devices with the LCD display option only:
up". After 20 seconds the screen will display the software version numbers. After 30
seconds the default screen will appear.
• The device will display readings.
INFORMATION!
This chapter and the start of the chapter that follows tell you what data is given on the device
display in normal mode and how to change device settings in configuration mode. If you know
about how this device operates, you can ignore this data. Continue with the quick setup
procedure. For more data about this procedure, refer to Quick Setup (Parameters) on page 83
5.2 Operating concept
You can read measurements and configure the device with:
• A digital display screen (optional).
• A connection to a system or PC with PACTware™. You can download the Device Type
Manager (DTM) file from the website. It is also supplied on the DVD-ROM delivered with the
device.
• A connection to a system or PC with AMS™. You can download the Device Description (DD)
file from the website. It is also supplied on the DVD-ROM delivered with the device.
• A connection to a HART® Field Communicator. You can download the Device Description (DD)
file from the website. It is also supplied on the DVD-ROM delivered with the device.
.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
65
Page 66
5
START-UP
5.3 Digital display screen
5.3.1 Local display screen layout
OPTIWAVE 5200 C/F
Figure 5-1: Local display screen layout in Normal mode
1 Current output percentage (bar graph and text — only shown if the current output function is the same as the mea-
surement on the screen in normal mode)
2 Measurement type (in this example, distance)
3 Device status (NE 107 symbols)
4 Device tag name
5 Updated measurement data symbol (the symbol flashes each time the measurement data is updated)
6 Measurement value and units
7 Device status (markers)
8 Keypad buttons (refer to the table in the section that follows)
The current output percentage is only shown if the measurement type (refer to item 2 in the
illustration) is the same as the output function. The parameter is set in menu item 2.4.1 (OUTPUT
FUNC.). For example, if the output function is set to "Level" and the device shows "Level"
measurements in normal mode, the bar graph and value is shown (refer to item 1 in the
illustration).
66
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 67
OPTIWAVE 5200 C/F
Figure 5-2: Local display screen layout in configuration mode
1 Function name
2 Configuration mode symbol
3 Menu number
START-UP
5
5.3.2 Functions of keypad buttons
Keypad buttonFunction
Normal mode:
Normal mode: Enter Information menu (Enter Configuration mode)
[Right]
[Return / Escape]
[Down]
[Up]
1 If you have made a strapping table in menu item 2.8.1 INPUT TABLE for volume or mass measurement, "Conversion"
and "Ullage Conv." will be shown in the list of measurement types
For data on keypad functions, refer to
Normal mode:Normal mode:
Configuration mode:
Configuration mode: Move cursor to the right
Configuration mode:Configuration mode:
Normal mode:
Normal mode: Change units (m, cm, mm, in, ft)
Normal mode:Normal mode:
Configuration mode:
Configuration mode: Exit
Configuration mode:Configuration mode:
Normal mode:
Normal mode: Change measurement type (distance, level , output (%),
Normal mode:Normal mode:
output (mA), conversion, ullage conversion, reflection)
Configuration mode:
Configuration mode: Decrease value or change parameter
Configuration mode:Configuration mode:
Normal mode:
Normal mode: Change measurement type (distance, level , output (%),
Normal mode:Normal mode:
output (mA), conversion, ullage conversion, reflection)
Configuration mode:
Configuration mode: Increase value or change parameter
Configuration mode:Configuration mode:
Normal mode
5.4 Remote communication with PACTware™
PACTware™ displays measurement information clearly and lets you configure the device from a
remote location. It is an Open Source, open configuration software for all field devices. It uses
Field Device Tool (FDT) technology. FDT is a communication standard for sending information
between the system and the field device. This standard agrees with IEC 62453. Field devices are
easily integrated. Installation is supported by a user-friendly Wizard.
1
1
on page 69.
Install these software programs and equipment:
• Microsoft® .NET Framework version 1.1 or later.
• PACTware.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
67
Page 68
5
START-UP
• HART® converter (USB, RS232...).
• The Device Type Manager (DTM) for the device.
INFORMATION!
The DTM for this device agrees with the FDT1.2 specification. For more data, refer to the related
certificate in the Product Catalog on the FDT Group website (http://www.fdtgroup.org/productcatalog/certified-dtms).
The software and installation instructions are given on the DVD-ROM supplied with the device.
You can also download the latest version of PACTware™ and the DTM from our website.
Refer also to the PACTware™ consortium site at http://www.pactware.com.
OPTIWAVE 5200 C/F
Figure 5-3: Screen from the PACTware™ user interface
1 DTM menu
2 Information for device identification
3 Configuration summary
5.5 Remote communication with the AMS™ Device Manager
The AMS™ Device Manager is an industrial Plant Asset Management (PAM) software tool. Its
role is to:
• Store configuration information for each device.
• Support HART® and FOUNDATION™ fieldbus devices.
• Store and read process data.
• Store and read diagnostic status information.
• Help plan preventive maintenance to reduce a plant's downtime to a minimum.
The DD file is given on the DVD-ROM supplied with the device. You can also download it from our
website.
68
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 69
OPTIWAVE 5200 C/F
6.1 User modes
Normal mode
Normal modeThis mode displays measurement data. For more data, refer to
Normal modeNormal mode
Configuration mode
Configuration modeUse this mode to view parameters, commission the device, create tables
Configuration modeConfiguration mode
6.2 Normal mode
This mode shows measurement data. Use the table that follows:
• for the selection of the measurement type (level, distance, percentage, conversion and
reflection) and
• for the selection of the measurement units
OPERATION
on page 69.
mode
for volume or mass measurement, change critical values to measure in
difficult process conditions. To get access to supervisor menu, refer to
Protection of the device settings
refer to
Function description
on page 88. For more data on menu items,
on page 76.
Normal
6
Some measurement types will only be available if the device has the correct parameters entered
in the configuration mode.
Keypad functions
ButtonDescriptionFunction"Hot key" function
RightEnter configuration mode.—
Return /
Escape
DownChange the measurement type.—
UpChange the measurement type.The display language will change
Change the measurement units.The device will show the
firmware version numbers in
menu item 1.1.0 IDENT..
when you push this button for 2
seconds. Push the button again
and it will go back to the original
language.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
69
Page 70
6
OPERATION
Measurement type definitions
Measurement typeDescriptionAvailable units
OPTIWAVE 5200 C/F
LEVELThis is a display and an output function option.
DISTANCEThis is a display and an output function option.
CONVERSIONThis is a display and an output function option.
ULLAGE CONV.This is a display and an output function option.
It is the height from the bottom of the tank to
the surface of the liquid (Tank height Distance).
It is the distance from the face of the flange (or
thread stop) to the surface of the liquid.
It gives the volume or mass of the tank
contents. This data is available if you prepare a
volume or mass table in configuration mode.
For data on how to prepare the conversion
table, refer to
measure volume or mass
device is used with PACTware™ software, it
can also show flow rate in a channel. 6 flow
profiles are available in the menu: Parshall
(ISO 9826), Venturi Rectangular (ISO 4359),
Venturi Trapezoidal (ISO 4359), Venturi U
(ISO 4359), V-Notch (ISO 1438) or Rectangular
Notch (ISO 1438).
It gives the empty volume or remaining mass
that can be put in the tank. This data is
available if you prepare a volume or mass table
in configuration mode. For data on how to
prepare the conversion table, refer to
How to configure the device to
on page 90. If the
How to
m, cm, mm, in (inches), ft
(feet)
m, cm, mm, in (inches), ft
(feet)
m3, L, gal (US gallons), ImpG
(Imperial gallons), ft3, bbl (oil
barrel), kg, t, Ston, Lton, m,
cm, mm, in, ft, m3/h, ft3/h
m3, L, gal (US gallons), ImpG
(Imperial gallons), ft3, bbl (oil
barrel), kg, t, Ston, Lton, m,
cm, mm, in, ft, m3/h, ft3/h
configure the device to measure volume or
mass
on page 90.
REFLECTIONThis is a display and an output function option.
It is the percentage of the emitted radar signal
which makes a reflection on the surface of the
liquid and is received by the device.
OUTPUT I (mA)The current output of the device.mA
OUTPUT I (%)The percentage of the current output.
0% = 4 mA. 100% = 20 mA.
%
%
6.3 Configuration mode
6.3.1 General notes
Change the settings of your device in Configuration
76. You can:
• Use the 1.0.0 INFORMATION
records. For more data about the Information menu, refer to Table 1: Info.
• Use the 2.0.0 SUPERVISOR
conversion table for volume, mass or flow rate measurement, change critical parameters for
difficult process conditions, reset the device and change basic parameters (tank height etc.),
output settings, HART Address etc. For more data about the Supervisor menu, refer to Table
2: Supervisor.
CAUTION!
The commissioning (parameters) procedure is mandatory.
70
1.0.0 INFORMATION menu to read settings, device software versions and error
1.0.0 INFORMATION1.0.0 INFORMATION
2.0.0 SUPERVISOR menu to commission the device, to run diagnostic tests, set up a
2.0.0 SUPERVISOR2.0.0 SUPERVISOR
Configuration mode. Data about the menus is given on page
ConfigurationConfiguration
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 71
OPTIWAVE 5200 C/F
CAUTION!
SIL-approved devices:
SIL-approved devices: For data about critical device parameters, refer to the Safety Manual.
SIL-approved devices:SIL-approved devices:
INFORMATION!
It is not possible to enter the 3.0.0 SERVICE and 4.0.0 MASTER menus. These menus are for
factory calibration and approved personnel only.
6.3.2 How to get access to the commissioning menu
The commissioning menu contains the menu items that are necessary for most configurations of
the device. The menu items are divided into 2 groups: "Parameters" and "Empty Spectrum
Recording". The "Parameters" group lets the supervisor set the tank height, tank type (process,
storage etc.), output function, output current range, 4 mA output setting, 20 mA output setting,
error delay and tag name. "Empty Spectrum Recording" is a procedure that finds interference
signals in the tank and uses a filter to remove them from the measurement data.
Do the steps that follow:
• Push the [>>>>] button.
iThis shows the Information
password security.
• Push the [] button one time to scroll up to the Supervisor
iThe screen shows the text "2.0.0 SUPERVISOR".
Information menu. The Information
InformationInformation
Information menu is read only and does not have
InformationInformation
Supervisor menu.
SupervisorSupervisor
OPERATION
6
• Push the [>>>>] button one time.
iThe screen shows a line. You must enter the password. Push the buttons under the display
screen 6 times (in total and in a given order) to get access to Configuration mode.
• Type in the password. The factory-set password is [>>>>], [^^^^], [], [], [>>>>] and [^^^^].
iThe device shows the text "2.1.0 COMMISSION.".
• Push [>>>>]. Do the basic configuration of the device in the "'Parameters" menu. For more data on
the procedure, refer to
of the procedure to continue to the next step.
• Push [] to go to menu item 2.1.2 EMP.SPEC.REC. Push [>>>>] to start the empty spectrum
recording procedure. For more data, refer to
Quick Setup (Parameters)
on page 83. Push [^^^^] at the end of each step
Empty spectrum recording
on page 85.
CAUTION!
SIL-approved devices: For data about critical device parameters for SIL approval, refer to the
Safety Manual (SIL approval).
INFORMATION!
HOW TO SET THE SUPERVISOR PASSWORD TO "ON" OR "OFF"
The supervisor password is set to "on" by default. If it is necessary to set this function to "off",
refer to Function description on page 76
(2.7.4).
, Table 2: Supervisor menu, menu item PSWD YES/NO
INFORMATION!
HOW TO CHANGE THE SUPERVISOR PASSWORD
You can change the password for the supervisor menu. For more data, refer to Function
description on page 76
, Table 2: Supervisor menu, menu item PASSWORD (2.7.5).
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
71
Page 72
6
OPERATION
6.3.3 Menu overview
1.0.0 Info. (Information)
1.1.0Ident. (Identification)
1.2.0Output
1.3.0History
2.0.0 Supervisor
2.1.0Commission. (Commissioning)
2.2.0Tests
2.3.0Basic Param. (Basic Parameters)
2.4.0Output I
2.5.0Application
2.6.0Communicat. (Communication)
2.7.0Display
2.8.0Conv. Table (Conversion Table)
2.9.0Config/Reset
OPTIWAVE 5200 C/F
3.0.0 Service
n/aPassword locked. Menus for factory calibration and qualified
service personnel only.
4.0.0 Master
n/aPassword locked. Menus for factory calibration and qualified
service personnel only.
72
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 73
OPTIWAVE 5200 C/F
6.3.4 Keypad functions
Figure 6-1: Local display screen layout in configuration mode
1 Function name
2 Configuration mode symbol
3 Menu number
OPERATION
6
This is what you see when you are in Configuration mode. The functions of the buttons are given
in the table that follows:
Functions of buttons for menu navigation
ButtonDescriptionFunction
Right
Enter /
Esc (Escape)
Down
Up
•
Go down to the sub-menu level (for example, from menu 1.0.0 to
sub-menu 1.1.0).
•
Enter the menu item
•
Go up to the menu level (for example, from sub-menu 1.1.0 to
menu 1.0.0).
•
Go to Normal mode. If you changed settings in Configuration mode,
you must save or cancel your new settings. For more data, refer to
the end of this section.
•
Scroll down the menu list (for example, from menu 2.0.0 to menu
1.0.0).
•
Scroll down the sub-menu list (for example, from sub-menu 2.2.0
to sub-menu 2.1.0).
•
Scroll up the menu list (for example, from menu 1.0.0 to menu
2.0.0).
•
Scroll up the sub-menu list (for example, from sub-menu 2.1.0 to
sub-menu 2.2.0).
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
73
Page 74
6
OPERATION
Lists of parameters in menu items
Figure 6-2: Lists of parameters in menu items
1 Parameter
2 Menu name
OPTIWAVE 5200 C/F
This is what you see when you select a menu item that has a list of parameters. The functions of
the buttons are given in the table that follows:
Function of buttons in menu items that have a list of parameters
ButtonDescriptionFunction
Rightn/a
Enter /
Esc (Escape)
DownMove down the list
UpMove up the list
Select the parameter and go back to the menu
74
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 75
OPTIWAVE 5200 C/F
Values in menu items
Figure 6-3: Values in menu items
1 Menu item with values stored at this time (first screen)
2 Push [>>>>] again to change the values. A cursor shows on the first digit.
3 Menu item name
4 Cursor on the selected digit
OPERATION
6
This is what you see when you select a menu item that has a value. The functions of the buttons
are given in the table that follows:
Function of buttons in menu items that have values
ButtonDescriptionFunction
Right
Enter /
Esc (Escape)
DownDecrease the digit value.
UpIncrease the digit value.
•
Enter the menu item and see the value stored at this time.
•
Enter the menu item configuration level to change the value.
•
Move the cursor to the next digit on the right. If the cursor is on the
last digit, push [>>>>] again to go back to the first digit.
Accept the value and go back to the sub-menu.
How to save settings changed in the supervisor menu (menu 2.0.0)
• When you have changed parameters in all the necessary menu items, push [^^^^] to accept the
new parameter.
• Push [^^^^] to go back to the "STORE" screen.
• The device will ask you to save or cancel your settings. Push [] or [] to select STORE YES
STORE NO
STORE NO. Push [^^^^] to accept or reject the new settings.
STORE NOSTORE NO
STORE YES or
STORE YESSTORE YES
iThe display goes back to Normal mode.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
75
Page 76
6
OPERATION
6.3.5 Function description
1.0.0 Information (Info.) menu
OPTIWAVE 5200 C/F
Menu
No.
FunctionFunction descriptionSelection list or range of
values
1.1.0 IDENT.
1.1.1SERIAL NUM.The device serial number.Read only.
1.2.1SUMMARY IPush [>>>>] to read the setting at this time
for the output function (OUTPUT
FUNC.). Push [>>>>] again to read the
settings for the output range (RANGE
I), 4 mA setting (SCALE 4mA), 20 mA
setting (SCALE 20mA), and error delay
(ERROR DELAY).
Read only.
1.3.0 HISTORY
1.3.1ERROR RECORDA log of device errors. Push [>>>>] to read
the errors. Push [] or [] to scroll up
or down the list. Each error is identified
by a code. Push [>>>>] again to show the
number of incidents and the time since
the last incident in days, hours,
minutes and seconds. For more data
about errors, refer to
messages
on page 93.
Status and error
Read only.
Default
2.0.0 Supervisor menu
Menu
No.
FunctionFunction descriptionSelection list or range
2.1.0 COMMISSION.
2.1.1PARAMETERSThis starts a quick set-up procedure
applicable to most applications. The
supervisor can give the tank height (TANK
HEIGHT), type of tank (TANK TYPE), output
function (OUTPUT FUNC.), current output
range (RANGE I), 4 mA setting (SCALE 4mA),
20 mA setting (SCALE 20mA), error delay
(ERROR DELAY) and tag name (TAG NAME).
CAUTION!
CAUTION!
CAUTION!CAUTION!
Make sure that you do this procedure before
you use the device. The settings in this
procedure have an effect on the
performance of the device.
Default
of values
76
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 77
OPTIWAVE 5200 C/F
OPERATION
6
Menu
FunctionFunction descriptionSelection list or range
No.
2.1.2EMP.SPEC.REC. Fixed and moving objects in the tank cause
interference signals. Put them through this
filter to correctly measure the tank
contents. This menu item starts a quick setup procedure. We recommend that the tank
is empty or only filled to the minimum level
before you do the procedure. We also
recommend that if you installed the device
on a tank that has equipment with parts that
move (e.g. agitators), start the equipment.
Push "Yes" at the end the procedure and set
the STORE screen to "STORE YES" to use
the data. For more data, refer to
spectrum recording
to "How to make a filter to remove radar
signal interference" on page 92.
on page 85. Refer also
Empty
2.2.0 TESTS
2.2.1SET OUTPUTAfter the selection of a parameter in this
2.2.2DIAGNOSTICThis starts the hardware test. Push [>>>>] many
menu item, this sets the current output to a
test value [mA]. The output will change to
the given value, independent of the
measured value. The current output will go
back to the measured value when the
display goes back to the menu level.
times to show:
•
D1, the time of operation
•
T1, temperature of the electronic
converter board
•
I1, loop current (internally measured
value)
•
I2, load current (NOTE: this data is not
available at this time)
•
V1, voltage 5.6 V. If the voltage is not
5.0<V1<5.7, speak to the supplier.
•
V2, voltage on capacitors. If the voltage is
not 3.2<V2<3.4, speak to the supplier.
•
V3, voltage 3.3 V. If the voltage is not
3.2<V3<3.4, speak to the supplier.
•
C1, reset counter (watchdog timer). If
C1>1, replace the signal converter.
If an NE 107 symbol and a status marker are
shown, refer to
page 93.
If you push [>>>>] again, the display goes back
to the menu level.
Device status (markers)
on
of values
3.5, 4, 6, 8, 10, 12, 14, 16,
18, 20 or 22 mA
Default
3.5 mA
2.3.0 BASIC PARAM.
2.3.1TANK HEIGHTThe distance from the flange face / thread
stop of the tank connection down to the tank
bottom. If the tank has a dish-shaped or
conical bottom, the tank height is measured
to a point on the tank bottom directly below
the antenna. For more data about level
measurement, refer to
on page 90. For more data about distance
measurement, refer to
measurement
on page 89.
Level measurement
Distance
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
min: 0.5 m / 1.6 ft or 2.3.2
BLOCK. DIST. or 2.3.7
ANTENNA EXT.
max: 30 m / 98.4 ft
1
77
Page 78
6
OPERATION
OPTIWAVE 5200 C/F
Menu
FunctionFunction descriptionSelection list or range
No.
2.3.2BLOCK. DIST.Blocking distance. The distance from the
2.3.3TIME CONST.Using this function, the device processes
2.3.4STILLWELL EN.If the device is installed in a stilling well or
2.3.5STILLWELL D.The inner diameter of the stilling well. If the
2.3.6STILLWELL H.The height of the stilling well. If the device is
2.3.7ANTENNA.EXTOptional antenna extensions. These are
2.3.8ANTENNA.TYPType of antenna. For more data, refer to
2.3.9DIST.PIECEOptional distance piece between the
2.3.10TAG NAMEThe device has an identification code (tag
flange to the top limit of the measuring
range (a zone given by the user where it is
not possible to measure). We recommend a
blocking distance of 500 mm / 19.7¨ below
the process connection. We also
recommend that the limit of the measuring
range is not less than the blocking distance.
If it is possible that the tank contents will go
above the blocking distance, set 2.5.5
OVERF. ENAB. to "Yes". If the distance is
less than the blocking distance, the device
continues to display the blocking distance.
several measurement readings to filter out
disturbances. Increasing the time constant
will smoothen the integrated readings,
decreasing will roughen the readings.
s=seconds.
has a Wave-Guide antenna, set this menu
item to "Yes".
device is installed in a stilling well or has a
Wave-Guide antenna, enter the diameter.
The device will use this data if you set 2.3.4
STILLWELL EN. to "Yes". This menu item is
not shown if you set 2.3.4 STILLWELL EN. to
"No".
installed in a stilling well or has a WaveGuide antenna, enter the height. The device
will use this data if you set 2.3.4 STILLWELL
EN. to "Yes". This menu item is not shown if
you set 2.3.4 STILLWELL EN. to "No".
attached between the flange and the
antenna.
Technical data
converter and the process connection. This
is for high-temperature applications more
than +150°C / +302°F. The distance piece is
120 mm / 4.7¨ long. For more data, refer to
Technical data
name). The supervisor can enter a
maximum of 8 numbers or letters.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 79
OPTIWAVE 5200 C/F
OPERATION
6
Menu
No.
FunctionFunction descriptionSelection list or range
of values
2.4.0 OUTPUT I
2.4.1OUTPUT FUNC.The output function. Select an output
2.4.2RANGE IThis menu item sets the limits of the output
2.4.3SCALE 4mAThis gives a measurement value to 4 mA.min.-max:
2.4.4SCALE 20mAThis gives a measurement value to 20 mA.min.-max:
2.4.5ERROR DELAYThe time after which the current output
function to scale the current values in
relation to a given point (usually the device
process connection or the tank bottom). The
output current value is shown on a bar
graph in normal mode if the measurement
type is the same as the output function.
Conversion parameters are shown
(Conversion, Ullage conversion) if there is
level-volume, level-mass or level-flow rate
data in 2.8.1 INPUT TABLE.
current range to 1 of the 2 available options:
standard limits (4...20 mA) or NAMUR
NE 43-compliant limits (3.8...20.5 mA). It
also tells the device what to do if an error
occurs. If you set RANGE I to 4-20/22E
an error occurs (e.g. the tank is too full,
etc.), the device output current will change
to an error value of 22 mA. If you set RANGE
I to 4-20
4-20 and the device senses a
4-204-20
measurement error, the value will stop at
the last correct measurement.
changes to an error value. The error value
shows that there is a measurement error.
mn=minutes and s=seconds.
2.5.1TANK TYPEThe conditions in which the device is used. If
the surface of the product is flat, select
"Storage". If the surface of the product is
disturbed, select "Process". If the surface of
the product is agitated with vortexes and
foam, select "Agitator".
2.5.2TRACKING VEL.Maximum tracking velocity. This value must
agree with the maximum rate of change of
the level of the liquid in the tank.
2.5.3Er PRODUCTFor measurement in TBF mode only. The
device automatically calculates the level
based on the product εr. If you select "TBF
Partial" or "TBF Full" in menu item 2.5.4
MEASUR.MODE, you can change the εr value
manually to adjust readings.
Storage, Process,
Agitator
min-max:
0.01…10.0 m/min
min-max: 1.100 to 10.0002.000
Process
0.5 m/min
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
79
Page 80
6
OPERATION
OPTIWAVE 5200 C/F
Menu
FunctionFunction descriptionSelection list or range
No.
2.5.4MEASUR.MODEThe device uses the dielectric constant (εr)
of the tank contents to monitor level. In
direct mode (if the dielectric constant is
high), the level signal is a reflection on the
surface of the tank contents. If the dielectric
constant is low, the device uses TBF mode.
The device in TBF mode uses the radar
reflection on the bottom of the tank (the
signal goes through the tank contents). The
tank must have a flat bottom for the device
to operate correctly in TBF mode. This
menu item is set by default to "Direct" for
tank contents with an ε
(<1.6), use "TBF Full". If ε
1.5...1.7), use "TBF Partial" mode. "TBF
Partial" is an automatic mode that lets the
device make a selection between "Direct
mode and "TBF" mode. If you use "TBF Full"
or "TBF Partial", enter the dielectric
constant in menu item 2.5.3 Er PRODUCT.
Refer also to "Measuring principle" on page
108.
2.5.5OVERF. ENAB.If this function is in operation, the device will
monitor the level even if it is in the blocking
distance. The reading shown on the display
stays fixed at the blocking distance if the
level is higher than the blocking distance.
The output will show that the tank is full. An
error is recorded by default.
If this function is not in operation, the
device looks for the largest signal between
the process connection and the tank bottom.
2.5.6MULT.REF.EN.Multiple reflections will cause the device to
display smaller level readings. Installation
of the device on a manhole or at the centre
of a dome roof, and high dielectric products
(εr> 5) can cause multiple reflections. A
very calm surface or a tank with a small
convex or flat roof can also cause multiple
reflections.
>1.8. If εr is very low
r
is low (εr=
r
Default
of values
Direct, TBF Partial,
TBF Full
YES, NONO
YES, NONO
Direct
If this function is in operation, the device
looks for the first signal peak below the
process connection. This signal peak is then
used to measure the level of the tank
contents. If this function is not in operation,
the device looks for the largest signal below
the process connection.
2.5.7EMP.SPEC.EN.The supervisor can start or stop the empty
spectrum filter.
2.6.0 COMMUNICAT.
2.6.1ADDRESSAny HART® address greater than 0 will
80
activate HART® multidrop mode. The
current output stays constant at 4 mA.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
YES, NONO
min.-max: 0…630
Page 81
OPTIWAVE 5200 C/F
OPERATION
6
Menu
No.
FunctionFunction descriptionSelection list or range
of values
2.7.0 DISPLAY
2.7.1LANGUAGEData can be shown in any of the languages
2.7.2LENGTH UNITThe length unit shown in normal mode.m, cm, mm, in (inches), ft
2.7.3CONV UNITConversion unit. The length, volume, mass
2.7.4PSWD YES/NOIf it is necessary to protect your settings in
2.7.5PASSWORDThis changes the password for the
2.7.6CONTRASTThe contrast control for the display screen.
stored in the device.
or flow rate conversion unit for the
conversion table and shown in normal
mode.
the supervisor menu with a password, set
this menu item to YES
supervisor menu. Push the buttons up to 6
times in any order. This will be the new
password. To confirm the change, enter the
new password a second time. For more
data, refer to
settings
You can select a shade of grey between light
grey (level 20) and black (level 54).
on page 88.
YES.
YESYES
Protection of the device
9 languages are available
in 3 packs: (1) English,
French, German and
Italian; (2) English,
French, Spanish and
Portuguese; (3) English,
Chinese (simplified),
Japanese and Russian
(feet)
m3, L, gal (US gallons),
ImpG (Imperial gallons),
ft3, bbl (oil barrel), kg, t,
Ston, Lton, m, cm, mm,
in, ft, m3/h, ft3/h
YES, NOYES
min.-max: 20…5436
Default
4
m
L
[>>>>], [^^^^], [],
[], [>>>>] and [^^^^]
2.8.0 CONV. TABLE
2.8.1INPUT TABLEThe device uses a conversion table
2.8.2DELETE TABLEThis menu item erases the data in the
(strapping table) to convert measurements
to volume, mass and flow rate readings. The
readings are shown in normal mode. Give
the number of entries on the table (min. 2;
max. 30). Select an entry (01...30) and enter
the level and the related volume / mass /
flow rate value for that entry. Push [^^^^] to
confirm the entry values. Continue the
procedure until the device has data for all
the entries. For more data, refer to
configure the device to measure volume or
on page 90.
mass
conversion table.
How to
min. 2 entries
max. 30 entries
(level / volume, mass or
flow rate)
YES, NONO
0 entries
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
81
Page 82
6
OPERATION
OPTIWAVE 5200 C/F
Menu
No.
FunctionFunction descriptionSelection list or range
of values
Default
2.9.0 CONFIG/RESET
2.9.1SAVEThis menu item is not available.YES, NONO
2.9.2RECALLThis menu item is not available.YES, NONO
2.9.3RESTARTThis menu item starts the device again.YES, NONO
2.9.4RESET FACT.If you set this menu item to "YES", the
device goes back to its initial settings (set by
the manufacturer in the factory).
1 This value is given in the customer order data
2 If the device has a SIL 2 qualification or a WHG approval, refer to the related technical document (safety manual (SIL2) or the WHG sup-
plementary instructions
3 Units and range depend on the output function, length unit and volume unit selected
4 If the device has the LCD display option, this depends on data given in the customer order
YES, NONO
3. Service menu
Menu
No.
3.0.0SERVICEAdvanced settings. The settings in this
FunctionFunction descriptionSelection listDefault
menu are protected with a password.
Only approved personnel can change
the parameters in this menu. For more
data, speak or write to your local sales
office.
4. Master menu
Menu
No.
4.0.0MASTERFactory settings. The settings in this
FunctionFunction descriptionSelection listDefault
menu are protected with a password.
Only approved personnel can change
the parameters in this menu. For more
data, speak or write to your local sales
office.
82
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 83
OPTIWAVE 5200 C/F
6.4 Further information on device configuration
6.4.1 Quick Setup (Parameters)
Use this procedure to change the tank height, tank type, output function, output range and give
the top and bottom measuring limits. Values and parameters that can be changed are shown
between the « ... » marks in the illustrations that follow. Push the keypad buttons in the correct
sequence:
CAUTION!
Make sure that you do this procedure before you use the device. The settings in this procedure
have an effect on the performance of the device.
Procedure
ScreenStepsDescription
•
[>>>>], [] and [>>>>].Default screen.
OPERATION
Enter configuration mode (2.0.0
SUPERVISOR).
6
•
[>>>>], [^^^^], [], [], [>>>>] and [^^^^].Enter the password (the default password
•
2× [>>>>]Push this button to start the quick set-up
•
[>>>>] to change the tank height (H).
•
[>>>>] to change the position of the cursor.
•
[] to decrease the value or [] to
increase the value.
•
[^^^^] to confirm.
is shown). If it is necessary to change the
password, refer to
page 76, menu item 2.7.5 PASSWORD.
procedure.
The distance from the flange face / thread
stop of the tank connection down to the
tank bottom. If the tank has a dish-shaped
or conical bottom, the tank height is
measured to a point on the tank bottom
directly below the antenna.
Function description
on
•
[] or [] for the selection of the
conditions in which the device is used
(Storage, Process, Agitator).
•
[^^^^] to confirm.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
If the surface of the product is flat, select
"Storage". If the surface of the product is
disturbed, select "Process". If the surface
of the product is agitated with vortexes and
foam, select "Agitator".
83
Page 84
6
OPERATION
ScreenStepsDescription
•
[] or [] for the selection of the
measurement name (Distance, Level,
Conversion, Ullage Conv. or Reflection).
•
[^^^^] to confirm.
•
[] or [] for the selection of the current
output range (4-20 mA/3.6E, 4-20, 3.8-
20.5/3.6E, etc.).
•
[^^^^] to confirm.
•
[>>>>] to change Scale 4 mA.
•
[>>>>] to change the position of the cursor.
•
[] to decrease the value or [] to
increase the value.
•
[^^^^] to confirm.
OPTIWAVE 5200 C/F
The manufacturer sets the output function
to “Level” before delivery.
If it is necessary to measure volume, ullage
volume, mass or ullage mass (Conversion
or Ullage Conv.), refer to
the device to measure volume or mass
page 90.
Use this step to give the 4 mA output
setting (0% limit) in the tank. Refer to the
illustrations that follow. Illustration 1
shows the settings for level. Illustration 2
shows the settings for distance.
How to configure
on
•
[>>>>] to change Scale 20 mA.
•
[>>>>] to change the position of the cursor.
•
[] to decrease the value or [] to
increase the value.
•
[^^^^] to confirm.
•
[] or [] for the selection of the error
delay (0 s, 10 s, 20 s, 30 s, 1 mn, 2 mn,
5mn or 15mn).
•
[^^^^] to confirm.
Use this step to give the 20 mA output
setting (100% limit) in the tank. Refer to the
illustrations that follow. Illustration 1
shows the settings for level. Illustration 2
shows the settings for distance.
The time after which the current output
changes to an error value. The error value
shows that there is a measurement error.
84
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 85
OPTIWAVE 5200 C/F
ScreenStepsDescription
•
[>>>>] to change the tag name.
•
[>>>>] to change the position of the cursor.
•
[] to decrease the alphanumeric value
(A, B, ..., 1, 2, ...) or [] to increase the
alphanumeric value.
•
[^^^^] to confirm.
•
2× [^^^^] to confirm.
•
[] or [] for the selection of the save
option (STORE NO or STORE YES).
•
[^^^^] to confirm.
6.4.2 Empty spectrum recording
The empty spectrum recording procedure is important for the performance of the device. We
recommend that the tank is empty or only filled to the minimum level before you do the
procedure.
OPERATION
The device has an identification code (tag
name). The supervisor can enter a
maximum of 8 numbers or letters.
Set to STORE YES to save and use the data.
Set to STORE NO to cancel the changes to
the device settings.
6
Procedure
ScreenStepsDescription
Use this procedure (menu item 2.1.2 EMP.SPEC.REC.) if there are fixed and moving objects in the
tank that can cause parasitic signals. The device does a scan for objects that do not change their
vertical positions in the tank (heating tubes, agitators, fuel assemblies etc.) and records the
data. The device can then use this data to put the measurement signal through a filter (empty
spectrum).
When the empty spectrum filter is on (when menu item 2.3.12 EMP.SPEC.EN. is set to "on"), it
will ignore the parasitic signals. Because the device records the data from the procedure, it is
also not necessary to do the procedure again if you de-energize the device.
CAUTION!
Make sure that the tank is empty or only filled to the minimum level.
Before you do the empty spectrum recording procedure, install the device on the tank. For more
data about how to install the device, refer to
Installation
on page 21.
Values and parameters that can be changed are shown between the « ... » marks in the
illustrations that follow. Push the keypad buttons in the correct sequence:
•
[>>>>], [] and [>>>>].Default screen.
Enter configuration mode (2.0.0
SUPERVISOR).
•
[>>>>], [^^^^], [], [], [>>>>] and [^^^^].Enter the password (the default password
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
is shown). If it is necessary to change the
password refer to
page 76, menu item 2.7.5 PASSWORD.
Function description
on
85
Page 86
6
OPERATION
ScreenStepsDescription
OPTIWAVE 5200 C/F
•
[>>>>], [] and [>>>>]Push these buttons to start the empty
•
[>>>>] for the selection of YES or [] for the
selection of NO.
•
Push [>>>>] after you start equipment in the
tank with parts that move.
•
[>>>>] for the selection of "Partially filled" or
[] for the selection of "Empty".
•
[>>>>] to change the measuring distance.
•
[>>>>] to change the position of the cursor.
•
[] to decrease the value or [] to
increase the value.
•
[^^^^] to confirm.
spectrum recording procedure.
Have you completely filled the tank?
Have you completely filled the tank?
Have you completely filled the tank?Have you completely filled the tank?
If the tank is full, it is not possible to
complete this procedure. The tank must be
partially filled or empty. Push [>>>>] to go back
to the start of the empty spectrum
recording procedure.
Please activate moving parts!
Please activate moving parts!
Please activate moving parts!Please activate moving parts!
If you installed the device on a tank that has
equipment with parts that move (e.g.
agitators), start the equipment.
CAUTION!
CAUTION!
CAUTION!CAUTION!
Make sure that you start all of the
equipment in the tank with parts that move,
before you continue to the subsequent step.
Is your tank partially filled or empty?
Is your tank partially filled or empty?
Is your tank partially filled or empty?Is your tank partially filled or empty?
If the tank is partially filled, the device will
scan for the first reflection in the tank.
Continue to the subsequent step. NOTE: If
you set this step to "Partially filled", but the
tank is empty, the device will show the
error message "Failure! Pulse Lost". Push
one of the keypad buttons to go back to the
start of the Snapshot procedure.
If the tank is empty, the scan will start
immediately. Ignore the subsequent step.
Measuring Distance
If the tank is partially filled, type in a
distance shorter than that between the
flange and the tank contents.
•
[] or [] to change the parameter
("Maximum" or "Average".
•
[^^^^] to confirm.
•
Push [>>>>] to make a selection from the
available spectra.
•
[^^^^] to confirm.
•
[>>>>] to set to "Yes". [] to set to "No".Do you want to save the spectrum?
Type of empty spectrum
Type of empty spectrum
Type of empty spectrumType of empty spectrum
Use the average value for tanks which
contain fixed objects only. Use the
maximum value for tanks which contain
many objects or moving objects.
Recording in progress
Recording in progress
Recording in progressRecording in progress
Reading in progress
Reading in progress
Reading in progressReading in progress
The device does a scan for objects that do
not change their vertical positions in the
tank (heating tubes, agitators, fuel
assemblies etc.) and records the data.
This shows a set of spectra (signals found
in the tank). It also gives the amplitude and
location of each signal.
Do you want to save the spectrum?
Do you want to save the spectrum?Do you want to save the spectrum?
Set to "Yes" to use the data. Set to "No" to
erase the data.
86
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 87
OPTIWAVE 5200 C/F
6.4.3 Test
Use this procedure to test the loop current. Values and parameters that can be changed are
shown between the « ... » marks in the illustrations that follow. Push the keypad buttons in the
correct sequence:
Procedure
ScreenStepDescription
•
[>>>>], [] and [>>>>].
•
Enter the password: [>>>>], [^^^^], [], [], [>>>>]
and [^^^^].
•
[^^^^]
Default screen.
OPERATION
6
•
[].
•
[>>>>].
•
[>>>>].
•
[] to decrease the value or [] to
increase the value.
•
[^^^^] to confirm.
•
[^^^^] 3 times to go back to the default
screen.
This step sets the loop current value. Make
a selection from 3.5, 4, 6, 8, 10, 12, 14, 16,
18, 20 or 22 mA.
The loop current goes back to initial value.
Default screen.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
87
Page 88
6
OPERATION
6.4.4 Protection of the device settings
The menu item PASSWORD (2.7.5) lets you change the supervisor menu password.
How to change the supervisor menu password
• After you enter the supervisor menu, push 6 × [], [>>>>] and 4 × [] to go to the menu item
PASSWORD (2.7.5).
• Enter the new 6-character password (push the 4 buttons in any sequence).
• Enter the new 6-character password again.
iIf the second entry is the same as the first, the device will go back to the sub-menu list
(2.7).If the second entry is not the same as the first, the device will not go back to the submenu list. Push [^^^^] to start the password sequence again and enter the new 6-character
password 2 times.
• Push [^^^^] to go back to the "STORE" screen.
• Push [] or [] to set the screen to STORE YES
iThe device will save the new password and go back to normal mode.
INFORMATION!
Make a note of the password and keep it in a safe place. If you lose the password, please speak
or write to your supplier.
STORE YES and push [^^^^].
STORE YESSTORE YES
OPTIWAVE 5200 C/F
How to set the supervisor password to "on" or "off"
The supervisor password is set to "on" by default. If it is necessary to set this function to "off",
refer to
(2.7.4).
Function description
on page 76, Table 2: Supervisor menu, menu item PSWD YES/NO
6.4.5 HART® network configuration
INFORMATION!
For more data, refer to Networks on page 61
The device uses HART® communication to send information to HART®-compatible equipment. It
can operate in either point-to-point or multidrop mode. The device will communicate in
multidrop mode if you change the address.
CAUTION!
Make sure that the address for this device is different from others in the multidrop network.
How to change from point-to-point to multidrop mode
• Enter the supervisor menu.
• Push [>>>>], 5 × [] and [>>>>] to go to menu item ADDRESS (2.6.1).
• Push [>>>>] to change the value. Enter a value between 1 and 63 and push [^^^^] to confirm (refer to
the caution before this procedure).
• Push [^^^^] to go back to the "STORE" screen.
• Push [] or [] to set the screen to STORE YES
iThe output is set to multidrop mode. The current output is set to 4 mA. This value does not
change in multidrop mode.
.
STORE YES and push [^^^^].
STORE YESSTORE YES
88
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 89
OPTIWAVE 5200 C/F
How to change from multidrop to point-to-point mode
• Enter the supervisor menu.
• Push [>>>>], 5 × [] and [>>>>] to go to menu item ADDRESS (2.6.1).
• Push [>>>>] to change the value. Enter the value 0 and push [^^^^] to confirm.
• Push [^^^^] to go back to the "STORE" screen.
• Push [] or [] to set the screen to STORE YES
iThe output is set to point-to-point mode. The current output changes to a range of 4...20 mA
or 3.8...20.5 mA (this range is set in menu item RANGE I (2.4.2)).
6.4.6 Distance measurement
The device displays distance measurements when the output function is set to "Distance".
Menu items related to distance measurement are:
• Output Function (2.4.1)
• Tank Height (2.3.1)
• Blocking Distance (2.3.2)
STORE YES and push [^^^^].
STORE YESSTORE YES
OPERATION
6
CAUTION!
If the distance for 4 mA is set in the blocking distance, the device will not be able to use the full
current output range.
Figure 6-4: Distance measurement
1 Tank Height (2.3.1)
2 Blocking Distance (2.3.2)
3 4mA Setting (2.4.3)
4 20 mA Setting (2.4.4)
5 Maximum effective measuring range
6 Non-measurement zone
For more data about the menu items, refer to
Supervisor menu.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Function description
on page 76, table 2.
89
Page 90
6
OPERATION
6.4.7 Level measurement
The device displays level measurements when the output function is set to "Level".
Menu items related to level measurement are:
• Output Function (2.4.1)
• Tank Height (2.3.1)
• Blocking Distance (2.3.2)
CAUTION!
If the level for the 20 mA is set in the blocking distance, the device will not be able to use the full
current output range.
OPTIWAVE 5200 C/F
Figure 6-5: Level measurement
1 Tank Height (2.3.1)
2 Blocking Distance (2.3.2)
3 Maximum effective measuring range
4 20 mA Setting (2.4.4)
5 4mA Setting (2.4.3)
6 Non-measurement zone
For more data about the menu items, refer to
Function description
Supervisor menu.
6.4.8 How to configure the device to measure volume or mass
The device can be configured to measure volume or mass. You can set up a strapping table in the
conversion table (2.8.0 CONV. TAB) sub-menu. Each entry is a pair of data (level – volume, level –
mass or level – flow rate). The strapping table must have a minimum of 2 entries and a
maximum of 30. The reference point for the table is the bottom of the tank (as given in menu item
2.3.1 TANK HEIGHT).
CAUTION!
Enter the data in numerical sequence (strapping table entry number 01, 02 etc.).
on page 76, table 2:
90
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 91
OPTIWAVE 5200 C/F
How to prepare a strapping table (conversion table)
• Enter the supervisor menu.
• Push [>>>>], 6 × [], [>>>>] and [] to go to 2.7.2 LENGTH UNIT.
• Push [] and [] to find the length unit that you will use in the table.
• Push [^^^^] to go to the sub-menu level.
• Push [] to go to 2.7.3 CONV UNIT (conversion unit)
• Push [] and [] to find the conversion unit that you will use in the table.
• Push [^^^^] to go to the sub-menu level and then [] and [>>>>] to go to the menu item 2.8.1 INPUT
TAB
• Push [>>>>] to make the strapping table. Enter the table entry number (01).
• Enter the length value and push [^^^^].
• Enter the conversion value and push [^^^^].
• Push [>>>>] to enter the subsequent table entry number
• Repeat the last 3 steps to complete the table.
• Push [^^^^] to go back to the "STORE" screen.
• Push [] or [] to set the screen to STORE YES
iThe device will store the data for the strapping table and go back to normal mode.
The device will give more accurate volume readings if you give more conversion data in these
areas:
(02,
03, ..., 30).
STORE YES and push [^^^^].
STORE YESSTORE YES
OPERATION
6
• Surfaces with curves.
• Sudden changes in the cross section.
Refer also to the illustration that follows:
Figure 6-6: A plot of points for a volume or mass table
1 Tank with reference points
2 Tank model with plotted points
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
91
Page 92
6
OPERATION
How to delete a volume or mass table
• Enter the supervisor menu.
• Push 7 × [], [>>>>], and [] to go to 2.8.2 DELETE TABLE.
• Push [>>>>] and [] to set the parameter to YES
YES.
YESYES
• Push [^^^^] to go back to the "STORE" screen.
• Push [] or [] to set the screen to STORE YES
STORE YES and push [^^^^].
STORE YESSTORE YES
iThe device will delete the data for the strapping table and go back to normal mode. The
"CONVERSION" and "ULLAGE CONV." data are not available in normal mode.
6.4.9 How to make a filter to remove radar signal interference
If the device measures level in a tank that contains obstructions (agitator, supports, heating
pipes etc.), these objects can cause radar signal interference (parasitic signals). You can use the
empty spectrum function (menu item 2.1.2) in the Commissioning menu to make a filter to
remove radar signal interference.
INFORMATION!
We recommend that you do an empty spectrum scan when the tank is empty and all the moving
parts (agitators etc.) are in operation..
OPTIWAVE 5200 C/F
92
Figure 6-7: How to make a filter to remove radar signal interference
1 Empty tank before the device uses the empty spectrum scan (with a graph of reflections shown)
2 Partially filled tank before the device uses the empty spectrum scan (with a graph of reflections shown)
3 Partially filled tank after the device uses the empty spectrum scan (with a graph of reflections shown)
4 Agitator blades location
5 Tank bottom signal
6 Agitator blades signals (interference signals) before the device does the empty spectrum scan
7 Bad quality (mixed) signals of the liquid and the agitator blades before the device does the empty spectrum scan
8 Reflected signal if the device uses the data from the empty spectrum scan. The device only uses the reflection on the
surface of the liquid to measure distance.
• After you enter the supervisor menu, push 2 × [>>>>] and [] to go to the menu item
EMP.SPEC.REC. (2.1.2).
• Do you have a completely filled tank? Select Yes
iIf you select Yes
Yes, the device will not do the empty spectrum scan. Empty the tank and repeat
YesYes
Yes or No
YesYes
No and then push [^^^^].
NoNo
the procedure.
• Are all the moving parts switched on? Select Yes
• Is your tank partially filled or completely empty? Select Partially filled
Yes or No
No and then push [^^^^].
YesYes
NoNo
Partially filled or Empty
Partially filledPartially filled
Empty and then push
EmptyEmpty
[^^^^].
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 93
OPTIWAVE 5200 C/F
• Measured distance. If the tank is partially filled, enter the distance to the product surface -
300 mm / 12¨. The device must include the tank contents when it filters the signal
• Do you want to use the average value or the maximum value? Select Average
then push [^^^^].
iUse the maximum for tanks that have moving parts. Use the average value for tanks that do
not have moving parts. The device will do an empty spectrum scan and then show the
results on the signal screen.
• Make a selection from the available spectra to identify the correct level signal. Push [^^^^].
• Do you want to save the spectrum? Select Yes
iIf you select Yes
Yes, the device will use the empty spectrum scan results to make a filter to
YesYes
remove radar signal interference.
INFORMATION!
For more data on empty spectrum scans, refer to Function description on page 76
Supervisor (menu item 2.1.2).
6.5 Status and error messages
6.5.1 Device status (markers)
Yes or No
YesYes
No and then push [^^^^].
NoNo
OPERATION
Average or Maximum
AverageAverage
Maximum and
MaximumMaximum
–
table 2:
6
If the device senses a change in device status, the display screen will show 1 or more status
markers at the bottom right side of the display screen. The display screen will also show a
symbol that agrees with NAMUR Recommendation NE 107 (Self-Monitoring and Diagnosis of
Field Devices) and VDI/VDE 2650. This is shown at the top left side of the display screen. More
data is given if you use PACTware™ software with the appropriate DTM on a PC. Error codes and
data are shown on the device display screen and in the DTM.
Menu item 2.2.2 DIAGNOSTIC (Configuration mode / Supervisor menu) supplies more data. This
includes internal voltages, the loop current and the reset counter (watchdog timer). You can see
this data on the device display screen and in the DTM.
Figure 6-8: Status markers
1 Device status (NAMUR NE 107 symbols)
2 Symbol: Failure
3 Symbol: Function check
4 Symbol: Out of specification
5 Symbol: Maintenance
6 Status marker line (marker 3 is shown)
7 When the status marker is on, a number is shown
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
93
Page 94
6
OPERATION
Types of error message
OPTIWAVE 5200 C/F
NE 107 statusType of
Description
error
FailureErrorIf an error message is shown in ERROR RECORD (menu item 1.3.1), the current
Out of specification WarningIf a warning message is shown, there is no effect on the current output value.
Maintenance
NE 107
symbol
NE 107
Status
shown
FailureThe device does not operate
Function
check
Out of
specification
MaintenanceThe device does not operate
correctly. The fault message
stays on. The user cannot
remove the "Failure" message
from the Normal mode screen.
The device operates correctly,
but the measured value is
incorrect. This fault message is
only temporary. This symbol is
shown when the user configures
the device with the DTM or a
HART® Communicator.
It is possible that the measured
value is unstable if the operating
conditions do not agree with the
device specification.
correctly because of bad
environmental conditions (e.g.
build-up on the antenna). The
measured value is correct, but
maintenance is necessary a
short time after this symbol is
shown.
output goes to the error signal value set in menu item RANGE I (menu item 2.4.2)
after the time set in ERROR DELAY (menu item 2.4.5). For more data about menu
items, refer to
If an "Out of specification" or a "Maintenance" status symbol is shown, refer to menu item 2.2.2
DIAGNOSTIC (Configuration mode / Supervisor menu) for more data.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 95
OPTIWAVE 5200 C/F
OPERATION
6
For data on errors, error records and error codes, refer to
6.5.2 Error handling
Figure 6-9: Error record data
1 Error code for the error
2 Number of times the error occurred
3 Time since the last error record (2 days, 18 hours, 16 minutes and 43 seconds shown in this example)
Error handling
on page 95.
How to find an error record
• Push [>>>>] to enter configuration mode from normal mode.
• Push [>>>>], 2 × [] and [>>>>] to go to menu item 1.3.1 ERROR RECORD.
• Push 2 × [>>>>] to look at the error list. Push [] or [] for the selection of an error.
iThe error record gives the number of times the error occurred and the time since the last
error message.
INFORMATION!
The time since the error occurred is measured in Days (D), Hours (H), Minutes(') and Seconds (").
It only includes the time when the device is energized. The error is saved in the memory of the
device when it is de-energized. The counter continues when the device is energized again.
Description of errors and corrective actions
Error
code
Error MessageStatus marker
shown
CauseCorrective action
Failure (NE 107 status signal)
ERR 100Device reset1The device detected an internal
ERR 101Current Output Drift1The current output is not
1Hardware error.Replace the device.
error (watchdog timer issue).
calibrated.
Record the data that is in menu
item 2.2.2 DIAGNOSTIC
(Configuration mode /
Supervisor menu). Speak to the
supplier.
Speak to the supplier to get the
calibration procedure.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
95
Page 96
6
OPERATION
OPTIWAVE 5200 C/F
Error
code
ERR 102Temperature out of
ERR 103Converter memory
ERR 104Converter Voltage
ERR 200Internal
ERR 201Overfill2The level is in the blocking
ERR 203Peak Lost (Level
ERR 204Sensor no Signal3The device's hardware is
ERR 206Sensor Microwave
ERR 207Sensor Memory
Error MessageStatus marker
shown
3The ambient temperature is
range
1The device's hardware is
failure
1The device's hardware is
failure
1The device's hardware or
Communication
2The signal peak is not found
Lost)
1The device's hardware is
failure
1The device's hardware is
failure
CauseCorrective action
outside the given range. This
can cause loss or corruption of
data.
defective.
defective.
software is defective. The
converter cannot transmit
signals to or receive signals
from the antenna electronics.
distance. There is a risk that the
product will overflow and/or
cover the antenna.
within the measuring window
that filters the signals received
by the antenna. The
measurement is not correct.
The device will automatically
increase this window to find the
correct signal.
defective.
defective.
defective.
Measure the ambient
temperature. De-energize the
device until the ambient
temperature is back in the given
range. If the temperature does
not stay in the correct range,
make sure that there is
insulation around the signal
converter. If this error occurs 2
times, replace the device.
Replace the signal converter.
For more data, refer to
How to
turn or remove the signal
converter
Replace the signal converter.
For more data, refer to
on page 44.
How to
turn or remove the signal
converter
De-energize the device. Make
sure that the signal cable
engages in the terminal and the
screw connection is tight.
Energize the device. If the
problem continues, replace the
signal converter. For more data,
refer to
the signal converter
Use a different procedure to
measure the level in the tank.
Remove some of the product
until the level is below the
blocking distance.
If a viscous product touched the
antenna:
•
•
Do a check of the device, tank
and the process. Reconfigure
the device and record a new
empty spectrum. Follow the
instructions on page 92. If
necessary, speak to the
supplier.
Replace the signal converter.
For more data, refer to
on page 44.
How to turn or remove
on page 44.
remove the device and clean
the antenna, or
If the device has a purging
option, use the purging
system to clean the antenna.
How to
turn or remove the signal
converter
Replace the signal converter.
For more data, refer to
on page 44.
How to
turn or remove the signal
converter
Replace the signal converter.
For more data, refer to
on page 44.
How to
turn or remove the signal
converter
on page 44.
96
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 97
OPTIWAVE 5200 C/F
OPERATION
6
Error
code
ERR 208Sensor Voltage
ERR 210Sensor Not
Error MessageStatus marker
shown
3The device's hardware is
failure
1The software version of the
compatible
1Defective wiring.
CauseCorrective action
defective.
sensor is not compatible with
the software version of the
signal converter.
Out of specification (NE 107 status signal)
ERR 102Temperature Out of
Range
ERR 201Overfill2The level is in the blocking
ERR 203Peak Lost2The signal peak is not found
3The ambient temperature is
outside the given range. This
can cause loss or corruption of
data.
distance. There is a risk that the
product will overflow and/or
cover the device.
within the measuring window
that filters the signals received
by the antenna. The
measurement is not correct.
The device will automatically
increase this window to find the
correct signal.
Do a check of the power supply
at the device terminals. Make
sure that voltage values are in
the specified limits in menu
item 2.2.2 DIAGNOSTIC
(Configuration mode /
Supervisor menu). If the voltage
is correct, replace the signal
converter. For more data on
how to replace the signal
converter, refer to
How to turn
or remove the signal converter
on page 44.
Go to menu 1.1.0 IDENT. in
Configuration mode. Record the
version numbers of the device
software given in menu items
1.1.2, 1.1.3 and 1.1.4. Give this
data to the supplier.
Measure the ambient
temperature. De-energize the
device until the ambient
temperature is back in the given
range. If the temperature does
not stay in the correct range,
make sure that there is
insulation around the signal
converter. If this error occurs 2
times, replace the device.
Remove some of the product
until the level is below the
blocking distance.
Do a check of the device, tank
and the process and make sure
that the data agrees with the
device configuration. If
necessary, record a new empty
spectrum. Follow the
instructions on page 92. If
necessary, speak to the
supplier.
Maintenance (NE 107 status signal)
—Empty Spectrum
Invalid
—Signal Weak4The signal amplitude is less
5The empty spectrum stored in
the device does not agree with
the installation. If you change
the device configuration (tank
height etc.), this message will
be shown. The recorded empty
spectrum will not be used by
the device while this error
message is shown.
than the average value. This can
occur if the liquid is agitated or
if there is foam in the tank. If
this error occurs frequently, the
device will possibly show the
"Peak Lost (Level Lost)" error
message.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
1
1
Record a new empty spectrum.
Follow the instructions on page
92.
If this error occurs frequently, it
is possible that you must install
the device in a stilling well or
use a different antenna type.
97
Page 98
6
OPERATION
OPTIWAVE 5200 C/F
Error
code
—Signal Strong4This error can occur if there is a
—Bad Measurement
—Temperature
—Temperature
1 This error message does not have an effect on the current output signal
Error MessageStatus marker
shown
4The measurement is incorrect
Quality
3The process connection or the
<-35°C/ -31°F
3The ambient temperature is
>+75°C / +167°F
CauseCorrective action
large change in signal
amplitude.
and continues to be incorrect
after more than 10 s. This error
can occur if the tank contents
are immediately below the
antenna.
ambient temperature is less
than -35°C/ -31°F. This
temperature is near to the
minimum limit for device
operation.
more than +75°C / +167°F. This
temperature is near to the
maximum limit for device
operation.
1
1
1
1
No corrective action is
necessary.
Record the process with the
PACTware™ software tool to
find the cause of the problem. If
necessary, speak to the
supplier.
Measure the ambient
temperature. De-energize the
device until the ambient
temperature is back in the given
range. If the temperature does
not stay in the correct range,
make sure that there is
insulation around the signal
converter.
Measure the ambient
temperature. De-energize the
device until the ambient
temperature is back in the given
range. If the temperature does
not stay in the correct range,
make sure that there is
insulation around the signal
converter.
INFORMATION!
In 4.0.0 MASTER menu, the type of error shown for error codes 102, 201 and 203 can be changed
from "Error" to "Warning" (the NE 107 status signal changes from "Failure" to "Out of
specification"). The 4.0.0 MASTER menu is password locked. For more data, speak or write to the
supplier.
98
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Page 99
OPTIWAVE 5200 C/F
7.1 Periodic maintenance
No maintenance is necessary. If it is possible that there will be build-up on the antenna, clean
the antenna regularly. If your device has the purge option, refer to
under process conditions
on page 99.
How to clean horn antennas
INFORMATION!
For more data about regular inspections and maintenance procedures for devices with Ex and
other approvals, refer to the related supplementary instructions.
WARNING!
Do not use bleach to clean the signal converter.
7.2 How to clean horn antennas under process conditions
If it is possible that there will be build-up or condensation, purging options are available for
DN150 / 6¨ and DN200 / 8¨ Metallic Horn antennas. Purge the antenna at intervals to make sure
that the inner surface of the antenna stays clean and the device measures accurately. For more
data about dimensions, refer to
heating/cooling system options).
Dimensions and weights
on page 126 (Purging and
SERVICE
7
Purging system
Figure 7-1: Purging system
1 Purging system
2 Flange connection
3 Metallic Horn antenna
4 G ¼ plug for the purging system
5 Seal
6 Warning label - see the illustration that follows
Warning label
Figure 7-2: Warning label (on the top face of the flange connection)
1 Text: Attention! Purge connection. Do not open during operation.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
99
Page 100
7
SERVICE
OPTIWAVE 5200 C/F
WARNING!
Purge the antenna with a dry gas or liquid that is applicable to the process. Make sure that you
use the correct device option (a purging system that uses gas or a purging system that uses
liquid). For more data about device options, refer to Order code on page 146
For more data about how to use the purge, refer to the table that follows:
.
How to use the purge
Process conditionsHow to use the purge
Purging with gas (device option)
Purging with gas (device option)
Purging with gas (device option)Purging with gas (device option)
The device is colder than other
elements in the process. It is
possible that there will be
condensation in the antenna.
Purging with liquid (device option)
Purging with liquid (device option)
Purging with liquid (device option)Purging with liquid (device option)
There is a build-up or risk of
build-up in the antenna
Continuous use. Use gas (compressed air, nitrogen or another gas
that is applicable to the process) to remove condensation from the
antenna.
Purge at intervals. Use a liquid (hot water, solvent or another liquid
that is applicable to the process) to melt the liquid that has
crystallized on the antenna.
7.3 Heating or cooling horn antennas under process conditions
A heating / cooling system option is available for DN150 / 6¨ and DN200 / 8¨ Metallic Horn
antennas. The outer surface of the antenna is sealed in a metal heating / cooling jacket. There is
an inlet and an outlet for the heating / cooling jacket on the top of the flange. For more data
about dimensions, refer to
system options).
Dimensions and weights
on page 126 (Purging and heating/cooling
WARNING!
Make sure that the pressure in the heating / cooling system is not more than 6 bar / 87 psi.
Make sure that the flange temperature is not more than the maximum limit. For more data, refer
to Technical data on page 110
It is also possible to use a purging system with the heating / cooling system. For more data, refer
How to clean horn antennas under process conditions
to
.
on page 99.
7.4 How to replace device components
7.4.1 Service warranty
WARNING!
Only approved personnel can do an inspection of the device and repairs. If you find a problem,
send the device back to the supplier for inspection and/or repairs.
100
INFORMATION!
The converter housing (compact or remote version) can be detached from the process
connection assembly under process conditions. For more data, refer to How to turn or remove
the signal converter on page 44
.
www.krohne.com07/2016 - 4001904903 - MA OPTIWAVE 5200 R03 en
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.