KROHNE OPTISWITCH-5200C User Manual

Operating Instructions
OPTISWITCH 5200 C, 5250 C
with two-wire output

Contents

Contents
1 About this document
1.1 Function .............................
1.2 Target group ..........................
1.3 Symbolism used .......................
2.1 Authorised personnel....................
2.2 Appropriate use........................
2.3 Warning about misuse ...................
2.4 CE conformity .........................
2.5 SIL conformity .........................
2.6 Safety instructions for Ex areas ............
3 Product description
3.1 Conguration..........................
3.2 Principle of operation ....................
3.3 Operation ............................
3.4 Storage and transport ...................
4 Mounting
4.1 General instructions.....................
4.2 Mounting instructions....................
4 4 4
5 5 5 5 6 6
7 7 9 9
10 13
5 Connecting to voltage supply
5.1 Preparing the connection .................
5.2 Connection steps.......................
5.3 Wiring plan, single chamber housing.........
6 Set up
6.1 General..............................
6.2 Adjustment elements ....................
6.3 Function chart .........................
6.4 Recurring test and function test ............
7 Maintenance and fault rectication
7.1 Maintenance ..........................
7.2 Rectify faults ..........................
7.3 Instrument repair .......................
8 Dismounting
8.1 Dismounting procedure ..................
8.2 Disposal .............................
2 OPTISWITCH 5200 C, 5250 C - with two-wire output
16 16 17
20 21 21 22
25 25 26
27 27
30435-EN-060831
9 Supplement
9.1 Technical data.........................
9.2 Dimensions ...........................
Contents
28 32
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 3
About this document

1 About this document

1.1 Function

This operating instructions manual has all the information you need for quick setup and safe operation. Please read this manual before you start setup.

1.2 Target group

This operating instructions manual is directed to trained, qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
4 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the operator. For safety and warranty reasons, any internal work on the instruments must be carried out only by personnel authorised by the manufacturer.

2.2 Appropriate use

OPTISWITCH 5200 C, 5250 C is a sensor for level detection.
Detailed information on the application range of OPTISWITCH 5200 C, 5250 C is available in chapter "Product description".

2.3 Warning about misuse

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

OPTISWITCH 5200 C, 5250 C is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country-specic instal­lation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and accident prevention rules.

2.5 CE conformity

OPTISWITCH 5200 C, 5250 C is in CE conformity with EMC (89/336/EWG), fulls NAMUR recommendation NE 21 and is
in CE conformity with LVD (73/23/EWG).
Conformity has been judged according to the following
standards:
l EMC:
- Emission EN 61326/A1: 1998 (class B)
- Susceptibility EN 61326: 1997/A1:1998
l LVD: EN 61010-1: 1993
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 5
For your safety

2.6 SIL conformity

OPTISWITCH 5200 C, 5250 C fullls the requirements of functional safety according to IEC 61508/IEC 61511. You can nd further information in the supplementary instructions manual "Safety Manual - Functional safety (SIL) OPTISWITCH
5XXX".

2.7 Safety instructions for Exareas

Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex­approved instruments.
6 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831

3 Product description

3.1 Conguration
Product description
Scope of delivery
Components
The scope of delivery encompasses:
l OPTISWITCH 5200 C, 5250 C level sensor l Documentation
- this operating instructions manual
- Ex specic safety instructions (with Ex versions), if
necessary further certicates
OPTISWITCH 5200 C, 5250 C consists of the following components:
l Housing cover l Housing with electronics l process tting with tuning fork
1
2
3
Fig. 1: OPTISWITCH 5200 C, 5250 C with plastic housing 1 Housing cover 2 Housing with electronics 3 Process tting

3.2 Principle of operation

Area of application
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 7
OPTISWITCH 5200 C, 5250 C is a level sensor with tuning
fork for level detection.
Product description
It is designed for industrial use in all areas of process
technology and can be used in liquids.
Typical applications are overll and dry run protection. With a tuning fork of only 40 mm length, OPTISWITCH 5200 C, 5250 C can be also mounted, e.g. in pipelines from DN 25. The small tuning fork allows use in vessels, tanks and pipes.
Thanks to its simple and robust measuring system, OPTI­SWITCH 5200 C, 5250 C is virtually unaected by the
chemical and physical properties of the liquid.
It functions even under dicult conditions such as turbulence, air bubbles, foam generation, buildup, strong external vibration or changing products.
Fault monitoring
The electronics module of OPTISWITCH 5200 C, 5250 C continuously monitors via frequency evaluation the following criteria:
l Strong corrosion or damage on the tuning fork l loss of vibration l Line break to the piezo drive
If one of these faults is detected, the electronics signals it via a dened current to the signal conditioning instrument. The connection cable to the sensor is also monitored on line break and short-circuit.
Physical principle
The tuning fork is piezoelectrically energised and vibrates at its mechanical resonance frequency of approx. 1200 Hz. The piezos are fastened mechanically and are hence not subject to temperature shock limitations. When the tuning fork is submerged in the product, the frequency changes. This change is detected by the integrated oscillator, transferred to the signal conditioning instrument as current value and converted into a switching command.
Power supply
Depending on your requirements, OPTISWITCH 5200 C, 5250 C with two-wire electronics can be connected to dierent
signal conditioning instruments. Compatible signal condition­ing instruments are listed under "Technical data".
The exact range of the power supply is stated in the "Technical data" in the "Supplement".
8 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Product description

3.3 Operation

The switching condition of OPTISWITCH 5200 C, 5250 C with plastic housing can be checked when the housing is closed (signal lamp). With the basic setting, products with a density >0.7 g/cm³ (>0.025 lbs/in³) can be detected. The instrument can be adapted if products with lower density should be measured.
On the electronics module you will nd the following indicating and adjustment elements:
l signal lamp for indication of the switching condition (green/
red)
l DIL switch for adaptation to the product density

3.4 Storage and transport

Packaging
Storage and transport tem-
perature
Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN 55439.
The packaging of standard instruments consists of environ-
ment-friendly, recyclable cardboard. In addition, the sensor is provided with a protective cover of ABS. For special versions PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
l Storage and transport temperature see "Supplement -
Technical data - Ambient conditions"
l Relative humidity 20 85 %
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 9
Mounting

4 Mounting

4.1 General instructions

Switching point
In general, OPTISWITCH 5200 C, 5250 C can be mounted in any position. The instrument must be mounted in such a way that the tuning fork is at the height of the requested switching point.
The tuning fork has lateral markings (notches), marking the switching point with vertical mounting. The switching point refers to water with the basic setting of the sensitivity switch
>=0.7 g/cm³ (>=0.025 lbs/in³). When mounting OPTISWITCH 5200 C, 5250 C, make sure that this marking is at the height of
the requested switching point. Keep in mind that the switching point of the instrument is shifted if the medium has a density other than water - water =1 g/cm³ (=0.036 lbs/in³). For products <0.7 g/cm³ (<0.025 lbs/in³)
and >0.5 g/cm³
(>0.018 lbs/in³) the density switch must be set to >=0.5 g/cm³.
2
1
4
3
Fig. 2: Vertical mounting 1 Switching point approx. 13 mm (approx. 0.51 in) 2 Switching point with lower density 3 Switching point with higher density 4 Switching point approx. 27 mm (approx. 1.06 in)
10 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Mounting
1
Fig. 3: Horizontal mounting 1 Switching point
2
1
Fig. 4: Horizontal mounting (recommended installation location - particularly for adhesive products)
1 Switching point 2 Marking with screwed version on top - with ange versions directed to the
ange holes
With ange versions, the fork is directed to the ange holes in the following way.
Fig. 5: Fork position with ange versions
Moisture
Use the recommended cables (see chapter "Connecting to power supply") and tighten the cable gland.
You can give your OPTISWITCH 5200 C, 5250 C additional protection against moisture penetration by leading the con­nection cable downward in front of the cable entry. Rain and
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 11
Mounting
condensation water can thus drain off. This applies mainly to mounting outdoors, in areas where moisture is expected (e.g. by cleaning processes) or on cooled or heated vessels.
Fig. 6: Measures against moisture penetration
Transport
Pressure/Vacuum
Handling
Do not hold OPTISWITCH 5200 C, 5250 C on the tuning fork. Particularly with ange or tube versions, the tuning fork can be
damaged by the instrument weight. Transport coated instru­ments very carefully and avoid touching the tuning fork.
Remove the protective cover just before mounting.
The process tting must be sealed if there is gauge or low
pressure in the vessel. Before use, check if the seal material is resistant against the measured product and the process temperature.
The max. permissible pressure is stated in the "Technical data" in the "Supplement" or on the type label of the sensor.
The vibrating level switch is a measuring instrument and must be treated accordingly. Bending the vibrating element will destroy the instrument.
Warning:
The housing must not be used to screw in the instrument! Applying tightening force on the housing can damage its
internal mechanical parts.
To screw in, use the hexagon above the thread.
30435-EN-060831
12 OPTISWITCH 5200 C, 5250 C - with two-wire output

4.2 Mounting instructions

Mounting
Welded socket
OPTISWITCH 5200 C, 5250 C has a dened thread starting point. This means that every OPTISWITCH 5200 C, 5250 C is in the same fork position after being screwed in. Remove therefore the supplied seal from the thread of OPTISWITCH 5200 C, 5250 C. This seal is not required when using a welded socket with O-ring in front.
Keep in mind that this welded socket is not suitable for coated instrument versions.
Screw OPTISWITCH 5200 C, 5250 C completely into the welded socket. The later position can be determined already before welding. Mark the appropriate position of the welded socket. Before welding, unscrew OPTISWITCH 5200 C, 5250 C and remove the rubber ring from the welded socket. The welded socket has a marking (notch). Weld the socket with the notch facing upward, or in case of pipelines, aligned with the direction of ow.
1
Fig. 7: Marking on the welded socket 1 Marking
Adhesive products
In case of horizontal mounting in adhesive and viscous products, the surfaces of the tuning fork should be vertical in order to reduce buildup on the tuning fork. On the screwed version you will nd a marking on the hexagon. With this, you can check the position of the tuning fork when screwing it in. When the hexagon touches the seal, the thread can still be turned by approx. half a turn. This is sucient to reach the recommended installation position.
With ange versions, the fork is directed to the ange holes.
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 13
Mounting
When used in adhesive and viscous products, the tuning fork
should protrude into the vessel to avoid buildup. For that reason, sockets for anges and mounting bosses should be avoided when mounting horizontally.
Inowing medium
Flows
If OPTISWITCH 5200 C, 5250 C is mounted in the lling stream, unwanted switching signals may be generated. Mount OPTISWITCH 5200 C, 5250 C at a location in the vessel where no disturbing inuence from e.g. lling openings, agitators etc. can occur.
This applies particularly to instrument types with long extension tube.
Fig. 8: Inowing medium
To minimise ow resistance caused by the tuning fork, OPTISWITCH 5200 C, 5250 C should be mounted in such a
way that the surfaces of the blades are parallel to the product movement.
Agitators
Due to agitators, vibrations or similar, the level switch can be subjected to strong lateral forces. For this reason, do not use an overly long extension tube for OPTISWITCH 5200 C, 5250
C, but check if you can mount an OPTISWITCH 5100 C, 5150 C level switch on the side of the vessel in horizontal
position.
Extreme vibration caused by the process or the equipment, e. g. agitators or turbulence in the vessel can cause the extension tube of OPTISWITCH 5200 C, 5250 C to vibrate in resonance. This leads to increased stress on the upper weld joint. Should
14 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Mounting
a longer tube version be necessary, you can provide a suitable support or guy directly above the tuning fork to secure the extension tube.
This measure applies particularly to applications in Ex areas category 1G or WHG. Make sure that the tube is not bent by this measure.
Fig. 9: Lateral straining of OPTISWITCH 5200 C, 5250 C
Enamel coating
Instruments with enamel coating should be treated very carefully and shocks should be avoided. Unpack OPTI-
SWITCH 5200 C, 5250 C directly before installation. Insert OPTISWITCH 5200 C, 5250 C carefully into the vessel
opening and avoid touching any sharp vessel parts.
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 15
Connecting to voltage supply

5 Connecting to voltage supply

5.1 Preparing the connection

Note safety instructions
Take note of safety
instructions for Ex applications
Select power supply
Selecting the connection ca-
ble
Generally not the following safety instructions:
l Connect only in the complete absence of line voltage
In hazardous areas you should take note of the appropriate regulations, conformity and type approval certicates of the sensors and power supply units.
Connect the power supply according to the following diagrams. Take note of the general installation regulations. As a rule,
connect OPTISWITCH 5200 C, 5250 C to vessel ground (PA), or in case of plastic vessels, to the next ground potential. On the side of the instrument housing there is a ground terminal between the cable entries. This connection serves to drain o electrostatic charges. In Ex applications, the installation regulations for hazardous areas must be given priority.
The data for power supply are stated in the "Technical data" in the "Supplement".
OPTISWITCH 5200 C, 5250 C is connected with standard cable with round cross-section. An outer cable diameter of 5 9 mm (0.2 0.35 in) ensures the seal eect of the cable entry.
If cable with a dierent diameter or wire cross section is used, exchange the seal or use an appropriate cable connection.
In hazardous areas, only use approved cable connections for OPTISWITCH 5200 C, 5250 C.
Select connection cable for Ex applica­tions
Take note of the corresponding installation regulations for Ex applications.

5.2 Connection steps

With Ex instruments, the housing cover may only be opened if there is no explosive atmosphere present.
Proceed as follows:
1 Unscrew the housing cover 2 Loosen compression nut of the cable entry
16 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Housing overview
Connecting to voltage supply
3 Remove approx. 10 cm (4 in) of the cable mantle, strip
approx. 1 cm (0.4 in) insulation from the ends of the individual wires
4 Insert the cable into the sensor through the cable entry 5 Open the terminals with a screwdriver 6 Insert the wire ends into the open terminals according to
the wiring plan
7 Tighten the terminals with a screwdriver 8 Check the hold of the wires in the terminals by lightly
pulling on them
9 Tighten the compression nut of the cable entry, the seal
ring must completely encircle the cable
10 Screw the housing cover back on
The electrical connection is hence nished.

5.3 Wiring plan, single chamber housing

The following illustrations apply to the non-Ex as well as to the EEx d version.
44
1
Fig. 10: Material versions, single chamber housing 1 Plastic (not with EExd) 2 Aluminium 3 Stainless steel (not with EExd) 4 Filter element for pressure compensation (not with EExd)
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 17
2 3
4
Connecting to voltage supply
Electronics and connection compartment
SW E60Z
1
11-36V DC
+
12
0,5 g / cm
0,7 g / cm
3
3
4
Fig. 11: Electronics and connection compartment, single chamber housing 1 Control lamp 2 DIL switch for sensitivity adjustment 3 Ground terminal 4 Screwed terminals
2
3
Wiring plan
We recommend connecting OPTISWITCH 5200 C, 5250 C in such a way that the switching circuit is open when there is a level signal, line break or failure (safe condition).
For connection to a signal conditioning instrument dto. Ex, WHG. The sensor is powered by the connected signal conditioning instrument. Further information is available in the Technical data in the Supplement, Ex-technical data are
available in the supplied safety information manual.
The wiring example is applicable for all suitable signal conditioning instruments.
The control lamp on OPTISWITCH 5200 C, 5250 C lights in general
l red - with covered tuning fork l green - with uncovered tuning fork
Take note of the operating instructions manual of the signal conditioning instrument. Suitable signal conditioning instru­ments are listed in the Technical data.
If OPTISWITCH 5200 C, 5250 C is used in Ex areas as part of an overll protection system acc. to WHG, take note for the regulations in the safety instructions and conformity certifi-
18 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Connecting to voltage supply
cates. If OPTISWITCH with oscillator SW E60ZEX, SW E60Z
EX E1 should be operated directly on the analogue input of a PLC, a suitable safety barrier should be connected.
+-
1 2
4321
Fig. 12: Wiring plan, single chamber housing
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 19
Set up

6 Set up

6.1 General

The numbers in brackets refer to the following illustrations.
Function/Conguration
The switching status of the electronics can be checked when the cover of the plastic housing is closed (signal lamp). In the basic setting, products with density > 0.7 g/cm³ (0.025 lbs/in³) can be detected. For products with lower density, the switch must be set to > 0.5 g/cm³ (0.018 lbs/in³).
On the electronics module you will nd the following indicating and adjustment elements:
l Signal lamp (1) l DIL switch for sensitivity adjustment (2)
Mode switch (A/B)
On the signal conditioning instrument via the signal con­ditioning instrument (only permissible in mode A when used acc. to WHG).
The switching condition can be changed with the A/B switch. You can set the required mode acc. to the "Function chart" (A -
max. detection or overll protection, B - min. detection or dry run protection). The integration time can also be modified on the signal conditioning instrument.
Note:
For test purposes, immerse the tuning fork of OPTISWITCH 5200 C, 5250 C always in liquids. Do not test the function of OPTISWITCH 5200 C, 5250 C with the hand. This can
damage the sensor.
20 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831

6.2 Adjustment elements

2
1
Fig. 13: Oscillator SW E60Z - two-wire output 1 Signal lamp (LED) 2 DIL switch for sensitivity adjustment
Set up
1
2
Signal lamp (1)
Sensitivity adjustment (2)
Control lamp (LED) for indication of the switching condition
l green = tuning fork uncovered l red = tuning fork covered l o= failure
With this DIL switch (2) you can set the switching point to liquids having a density between 0.5 and 0.7 g/cm³ (0.018 and
0.025 lbs/in³). With the basic setting, liquids with a density of >0.7 g/cm³ (>0.025 lbs/in³) can be detected. In liquids with lower density, you must set the switch to >0.5 g/cm³ (>0.018 lbs/in³). The specications for the position of the switching point relate to water - density value 1.0 g/cm³
(0.
036 lbs/in³). In products with a dierent density, the
switching point will shift in the direction of the housing or tuning fork end depending on the density and type of installation.

6.3 Function chart

The following chart provides an overview of the switching conditions depending on the adjusted mode and level.
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 21
Set up
Mode on the signal condition­ing instru­ment
Mode A Overll
protection
Mode A Overll
protection
Mode B Dry run
protection
Mode B Dry run
protection
Fault signal (mode A/B)
Level Signal
current ­Sensor
approx.
8 mA
approx.
16 mA
approx.
16 mA
approx.
8 mA
any approx.
1.8 mA
Analogue input con-
1)
trol
>3.6 mA <11 mA
>13 mA <21 mA
>13 mA <21 mA
>3.6 mA <11 mA
<=3.6 mA >=21.6 mA
Signal lamp ­sensor
Green
red
red
Green
Signal lamp ­signal condi­tioning instru­ment

6.4 Recurring test and function test

Acc. to WHG and IEC 61508. The implementation of the recurring test acc. to WHG is stipulated in the general type approval, item 8.
Take note of the overwhelming approvals, if OPTISWITCH 5200 C, 5250 CEx is used as part of an overll protection system according to WHG.
OPTISWITCH 5200 C, 5250 CEx in conjunction with an SU 501 (Ex) is qualied in mode A (overll protection) for use in measuring chains of SIL2 acc. to IEC 61508 (redundant, SIL3).
1)
Electronics directly evaluated via the analogue input of a control system
(without signal conditioning instrument)
22 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Set up
WHG
SIL
Recurring test and function
test
The following instrument combinations meet the requirements acc. to WHG:
OPTISWITCH 5200 C, 5250 CEXwith
l Oscillator SWE60ZEX l signal conditioning instrument SU 501 (Ex) or PLC
The following instrument combinations meet the requirements according to SIL:
OPTISWITCH 5200 C, 5250 CEX
l Oscillator SW E60Z (EX) l signal conditioning instrument SU 501 (Ex)
The recurring test according to WHG and the function test according to IEC 61508 can be carried out by pushing the test key on the signal conditioning instrument or by briey (> 2 seconds) interrupting the cable to the sensor. The correct sequence of the switching conditions must be monitored via the two signal lamps on the signal conditioning instrument as well as the connected systems. The sensor must neither be removed nor must a response be triggered by lling the vessel. This applies to OPTISWITCH 5200 C, 5250 C with two-wire oscillator SW E60ZEX.
If you are subject to the WHG regulations, you can carry out the function test with the outputted current values also directly via a PLC or a process control system.
The implementation and switching sequence of the function test is described also in the operating instructions manual of the appropriate signal conditioning instrument.
Test key on the signal con­ditioning instrument
A function test can be carried out for measuring systems in conjunction with the two-wire oscillator SW E60ZEX. For this the integration time must be set to 0.5 s. The signal conditioning instrument has an integrated test key. The test key is recessed in the front plate of the signal conditioning instrument. Push the test key for > 2 seconds with a suitable object (screwdriver
, pen, etc.).
For WHG the following is also applicable: If OPTISWITCH 5200 C, 5250 C is connected to a processing system or a PLC, you have to interrupt the connection cable to the sensor for > 2 seconds.
After releasing the test key or interrupting the connection cable to the sensor, the complete measuring system can be checked
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 23
Set up
on correct function. The following operating conditions are simulated during the test:
l Fault signal l Empty signal l Full signal
Check if all three switching conditions occur in the correct sequence and the stated time period. If this is not the case, there is a fault in the measuring system (see also the operating instructions manual of the signal conditioning instrument). Keep in mind that connected instruments are activated during the function test. By doing this, you can check the correct function of the measuring system.
Test procedure
1. Fault signal (approx. 3 s)
2. Empty signal (approx. 1.5 s)
3. Full signal (approx. 1.5 s)
4. Return to the actual operating condition
(after releasing the test key)
Sensor current ­Sensor
<2.0 mA currentless currentless
approx. 8mAenergized currentless energized
approx. 16mAcurrentless energized
Level re­lay A - overll protection
Level re­lay B - dry run pro­tection
Signal lamp A - Overll protection
Signal lamp B - Dry run protection
Fail safe relay
currentless
energized
Note:
Mode B is not permissible when used as part of an overll protection system acc. to WHG and use in measuring chains acc. to IEC 61508.
Control lamp
You can carry out the function test with the outputted current values also directly with a PLC or a process control system
(WHG).
24 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Maintenance and fault rectication
7 Maintenance and fault rectication

7.1 Maintenance

When used as directed in normal operation, OPTISWITCH 5200 C, 5250 C is completely maintenance free.

7.2 Rectify faults

Causes of malfunction
Fault rectication
Checking the switching signal
OPTISWITCH 5200 C, 5250 C oers maximum reliability. Nevertheless faults can occur during operation. These may be
caused by the following, e.g.:
l Sensor l Process l Power supply l Signal processing
The rst measure to be taken is to check the output signal. In many cases, the causes can be determined this way and the faults rectied.
? OPTISWITCH 5200 C, 5250 C signals "covered" when the
vibrating element is not submerged (overll protection)
? OPTISWITCH 5200 C, 5250 C signals "uncovered" when
the vibrating element is submerged (dry run protection)
l Supply voltage too low
à Check the power supply
l Electronics defective
à Press the mode switch (min./max.). If the instrument
then changes the mode, the instrument may be mechanically damaged. Should the switching function in the correct mode still be faulty, return the instrument for repair.
à Push the mode switch. If the instrument then does not
change the mode, the oscillator may be defective. Exchange the oscillator.
à Check if there is buildup on the vibrating element, and if
so, remove it.
l Unfavourable installation location
à Mount the instrument at a location in the vessel where
no dead zones or air bubbles can form.
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 25
Maintenance and fault rectication
l Wrong mode selected
à Set the correct mode on the mode switch (max.: overll
protection; min.: dry run protection). Wiring should be carried out according to the quiescent current principle.
? Signal lamp ashes red
l Electronics has detected a failure
à Exchange instrument or return instrument for repair
? The signal lamp ashes alternately red and green
l instrument defective
à Exchange instrument or return instrument for repair

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form from our website krohne-mar.com/leadmin/media-lounge/PDF-Download/ Specimen_e.pdf.
By doing this you help us carry out the repair quickly and
without having to call back for needed information.
l Print and ll out one form per instrument l Clean the instrument and pack it damage-proof l Attach the completed form and possibly also a safety data
sheet to the instrument
http://www.
26 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Dismounting

8 Dismounting

8.1 Dismounting procedure

Warning:
Before dismounting, be aware of dangerous process con­ditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.
With Ex instruments, the housing cover may only be opened if there is no explosive atmosphere present.

8.2 Disposal

The instrument consists of materials which can be recycled by specialised recycling companies. We use recyclable materials and have designed the electronic modules to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/ EG and the respective national laws (in Germany, e.g. ElektroG). Pass the instrument directly on to a specialised
recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
Correct disposal avoids negative eects to persons and environment and ensures recycling of useful raw materials.
Materials: see "Technical data"
If you cannot dispose of the instrument properly, please
contact us about disposal methods or return.
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 27
Supplement

9 Supplement

9.1 Technical data

General data
Material 316L corresponds to 1.4404 or 1.4435
Materials, wetted parts
- Process tting - thread 316L, Hastelloy C4 (2.4602)
- Process tting - ange 316L, 316L with Hastelloy C4 plated, enam-
elled steel, 316L with ECTFE coated, 316L with PFA coated
- Process seal Klingersil C-4400
- Tuning fork 316L, Hastelloy C4 (2.4610)
- Extension tube ø 21.
Materials, non-wetted parts
- Housing Plastic PBT (Polyester), Alu die-casting pow-
- Seal ring between housing and
housing cover
- waveguide in the housing cover PMMA (Makrolon)
- Ground terminal 316L
- Temperature adapter (optional
- Gas-tight leadthrough (optional) 316L/glass
3 mm (0.84 in) 316L, Hastelloy C4 (2.4610), Hastelloy C4
(2.4610) enamelled, 316L with ECTFE coating,
316L with PFA coating
der-coated, 316L
NBR (stainless steel housing), silicone (Alu/ plastic housing)
) 316L
Lengths
- 316L, Hastelloy C4 (2.4610) 80 … 6000 mm (3 … 236 in)
- Hastelloy C4 (2.4610) enamelled 80 … 1500 mm (3 … 59 in)
- 316L, ECTFE coated 80 … 3000 mm (3 … 118 in)
- 316L, PFA coated 80 … 3000 mm (3 … 118 in)
Weights
- Plastic housing 760 g (
- Aluminium housing 1170 g (41 oz)
- Stainless steel housing 1530 g (54 oz)
- Tube extension approx. 110 g/m (1.2 oz/ft)
28 OPTISWITCH 5200 C, 5250 C - with two-wire output
27 oz)
30435-EN-060831
Supplement
Layer thickness
- Enamel approx. 0.8 mm (0.03 in)
- ECTFE approx. 0.5 mm (approx. 0.02 in)
- PFA approx. 0.5 mm (approx. 0.02 in)
Surface quality
- Standard R
- Hygienic version 3A (OPTISWITCH
approx. 3 µm (approx. 1.18-4in)
a
<0.8 µm (<3.15-5in)
R
a
5250 C)
Process ttings
- Thread G¾ A, ¾ NPT, G1 A, 1 NPT
- Flanges DIN from DN 25, ANSI from 1"
- hygienic ttings Bolting DN 40 PN 40, Tri-Clamp 1", Tri-Clamp
1½" PN 10, conus DN 25 PN 40, Tuchenhagen Varivent DN 50 PN 10
High voltage test (enamel) max. 5 KV
Gas-tight leadthrough (optional)
-6
- leakage rate <10
mbar l/s
- pressure resistance PN 64
- hygienic ttings Bolting DN 40 PN 40, Tri-Clamp 1", Tri-Clamp
1½" PN 10, conus DN 25 PN 40, Tuchenhagen Varivent DN 50 PN 10
Output variable
Output Two-wire output
Output signal
- empty (uncovered) approx. 8 mA
- full (covered) approx. 16 mA
- Fault signal <2.0 mA
Possible signal conditioning instruments SU 501 (Ex)
Modes (adjustable via the signal conditioning instrument
)
- AMax. detection or overow/overll protection
- BMin. detection or dry run protection
Measuring accuracy
Hysteresis approx. 2 mm (0.08 in) with vertical installation Integration time approx. 500 ms Frequency approx. 1200 Hz
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 29
Supplement
Ambient conditions
Ambient temperature on the housing -40 +70 °C (-40 +158 °F) Storage and transport temperature -40 +80 °C (-40 +176 °F)
Process conditions
Parameter Limit level of liquids Process pressure -1 64 bar (-14.5 938 psi) depending on
the process tting, e.g. ange
OPTISWITCH 5200 C, 5250 C of 316L/
-50 +150 °C (-58 +302 °F
)
Hastelloy C4 (2.4610)
Process temperature (thread or ange
temperature) with temperature adapter (option)
- OPTISWITCH 5200 C, 5250 C of
-50 +250 °C (-58 +482 °F)
316L/Hastelloy C4
- OPTISWITCH 5200 C, 5250 C
-50 +200 °C (-58 +392 °F)
enamelled
- OPTISWITCH 5200 C, 5250 C with
-50 +150 °C (-58 +302 °F)
ECTFE coating
- OPTISWITCH 5200 C, 5250 C with
-50 +150 °C (-58 +302 °F)
PFA coating
2
70˚C
(158˚F)
40˚C
(104˚F)
0˚C
(32˚F)
-50˚C
(-58˚F)
-40˚C
(-40˚F)
Fig. 14: Ambient temperature - Product temperature 1 Product temperature 2 Ambient temperature 3 Temperature range with temperature adapter
50˚C
(122˚F)
100˚C
(212˚F)
150˚C
(302˚F)
3
200˚C
(392˚F)
250˚C
(482˚F)
1
Viscosity - dynamic 0.1 10,000 mPas(requirement: with density
1)
Density 0.7 2.5 g/cm³ (0.025 0.09 lbs/in³);
30 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Supplement
0.5 2.5 g/cm³ (0.018 0.09 lbs/in³) by switching over
Electromechanical data
Cable entry/plug
2)
- Single chamber housing l 1x cable entry M20x1.5 (cable-ø 5 9 mm), 1x blind stopper M20x1.5
or:
l 1x closing cap ½ NPT, 1x blind plug ½ NPT
or:
l 1x plug (depending on the version), 1x blind
plug M20x1.5
Screw terminals for wire cross-section up to 1.5 mm²
(0.0023 in²)
Adjustment elements
Sensitivity switch
- 0.50.5 2.5 g/cm³ (0.018 0.9 oz/in³)
- 0.70.7 2.5 g/cm³ (0.025 0.9 oz/in³)
Voltage supply
Supply voltage 11 36 VDC(via the signal conditioning
instrument)
E
lectrical protective measures
Protection IP 66/IP 67
Overvoltage category III
Protection class II
Approvals
3)
ATEX II 1G, 1/2G, 2GEExiaIIC T6
ATEX II 1/2G, 2GEExd d IIC T6
4)
Overll protection acc. to WHG
Ship approvals
2)
Depending on the version M12x1, according to DIN 43650, Harting, Am­phenol-Tuchel, 7/8" FF.
3)
Deviating data in Ex applications: see separate safety instructions.
4)
Only with Aluminium housing.
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 31
/
/
/
M20x1,5
½
/
½
/
½
Supplement

9.2 Dimensions

Housing
~ 69mm
23
(2
/32")
Fig. 15: Housing versions 1 Plastic housing 2 Stainless steel housing 3 Aluminium housing
ø 77mm
1
(3
/32")
M20x1,5
~ 116mm
9
/16")
(4
ø 84mm (3
5
/16")
4
1
4
~ 69mm
23
/32")
(2
2
4
M20x1,5
ø 77mm
1
(3
/32")
4
4
32 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
OPTISWITCH 5200 C, 5250 C
")
64
/
47
18,5mm
(
32 (G3/4A, 3/4"NPT)
41 (G1A, 1"NPT) G3/4A, 3/4"NPT G1A, 1"NPT
")
64
/
27
36mm (1
")
4
/
1
57mm
(2
Supplement
1 2 3
L
")
4
/
3
19mm
ø33,7mm
21
/64")
(1
(
")
32
/
31
50mm
(1
L
4 5
L
Fig. 16: OPTISWITCH 5200 C, 5250 C, threaded version 1 Thread 2 Tri-Clamp 3 Cone DN 25 4 Bolting DN 40 5 Flange 6 Gas-tight leadthrough 7 Temperature adapter
L = Sensor length, see "Technical data"
L
L
")
32
/
11
34mm (1
")
64
/
57
175mm (6
7
6
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 33
Supplement
34 OPTISWITCH 5200 C, 5250 C - with two-wire output
30435-EN-060831
Supplement
30435-EN-060831
OPTISWITCH 5200 C, 5250 C - with two-wire output 35
Subject to change without notice
30435-EN-060831
Loading...