KROHNE OPTISONIC-3400CSA User Manual

Supplementary instructions
Supplementary instructions
OPTISONIC 3400
OPTISONIC 3400
OPTISONIC 3400OPTISONIC 3400
Supplementary instructions Supplementary instructions
Ultrasonic flow meter
Addendum for hazardous area. These additional instructions are an extension to the OPTISONIC 3400 manual and quick start.
© KROHNE 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
1 Introduction 4
1.1 Safety instructions from the manufacturer ..................................................................... 4
1.1.1 Copyright and data protection ................................................................................................ 4
1.1.2 Disclaimer ............................................................................................................................... 4
1.1.3 Product liability and warranty ................................................................................................ 4
1.1.4 Information concerning the documentation........................................................................... 5
1.1.5 Warnings and symbols used................................................................................................... 5
1.1.6 Manufacturer .......................................................................................................................... 6
1.2 Safety instructions for the operator................................................................................. 6
1.3 Approval............................................................................................................................ 6
1.4 OPTISONIC 3400 C(/i)-DIV1 .............................................................................................. 7
1.5 OPTISONIC 3000 F/...-DIV1............................................................................................... 8
1.6 UFC 400 F(/i)-DIV1............................................................................................................ 8
1.7 Marking labels (examples) ............................................................................................... 9
1.7.1 CSA labels ............................................................................................................................... 9
1.7.2 I/O sticker............................................................................................................................. 11
1.7.3 Nameplate for the measuring sensor (field version)........................................................... 11
2 Temperature limits 13
2.1 General ........................................................................................................................... 13
2.2 UFC 400 F(/i)-DIV1.......................................................................................................... 13
2.3 OPTISONIC 3400 C(/i)-DIV1 ............................................................................................ 13
2.4 OPTISONIC 3000 F/...-DIV1............................................................................................. 14
3 Connection of separate systems 15
3.1 General ........................................................................................................................... 15
3.2 Cable marking ................................................................................................................ 15
3.3 Cable parameters...........................................................................................................15
3.4 Equipotential bonding..................................................................................................... 16
3.4.1 Signal converter.................................................................................................................... 16
3.5 Signal cable connections (field versions)....................................................................... 16
4 Electrical connections 18
4.1 General ........................................................................................................................... 18
4.2 Cable glands ................................................................................................................... 20
4.3 Field Wiring..................................................................................................................... 20
4.4 Non-I.S. I/O connections ................................................................................................ 21
4.5 I.S. I/O connections......................................................................................................... 23
5 Maintenance and service 25
5.1 Maintenance ................................................................................................................... 25
5.2 Before and after opening................................................................................................ 25
5.3 Replacement of mains fuse............................................................................................ 26
5.4 Exchange of electronics unit .......................................................................................... 27
2
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
5.4.1 Field version.......................................................................................................................... 28
5.5 Service / repair information ........................................................................................... 30
5.6 Form (for copying) to accompany a returned device ..................................................... 31
5.7 Disposal .......................................................................................................................... 31
6 Notes 32
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
3
1 INTRODUCTION
1.1 Safety instructions from the manufacturer
1.1.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.
OPTISONIC 3400
We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.
1.1.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.
1.1.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.
4
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
1.1.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of underneath icons.
1.1.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
INTRODUCTION 1
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems. There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
5
1 INTRODUCTION
HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the specified sequence.
i RESULT
RESULT
RESULTRESULT This symbol refers to all important consequences of the previous actions.
1.1.6 Manufacturer
The instrument is developed and manufactured by: KROHNE Altometer Kerkeplaat 12 3313 LC Dordrecht The Netherlands
For information, maintenance or service please contact your nearest local KROHNE representative.
OPTISONIC 3400
1.2 Safety instructions for the operator
WARNING!
Do not change the device. Unauthorized changes affect the explosion safety of the devices.
The prescriptions and regulations as well as the electrical data described in the EC type examination certificate must be obeyed.
Beside the instructions for electrical installations in non-hazardous locations according to the applicable national standard , especially the regulations in ANSI/ISA RP 12.06.01 "Installation of intrinsically safe systems for hazardous (classified) locations" and the National Electrical Code (ANSI/NFPA 70), sections 504 and 505, as well as CEC section 18­090, must be strictly followed.
Installation, establishment, utilization and maintenance are only allowed to be executed by personnel with an education in explosion safety!
These additional instructions are an extension to the installation and operating instructions and only apply to the DIV1 versions of the OPTISONIC 3400 C, OPTISONIC 3000 F/... and UFC 400 F(/i) ultrasonic flowmeters. All technical information as described in the Installation and Operating Instructions is applicable, when not specifically excluded, completed or replaced by the instructions in these additional instructions.
1.3 Approval
The OPTISONIC 3400 ultrasonic flowmeters are manufactured according to the relevant Canadian and U.S.A. regulations. The flowmeters are approved for installation and use in hazardous classified locations of Class I, II and III, Division 1 or Division 2.
Approved by the Canadian Standards Association (CSA) under the following approval number:
cCSA
CSAus : 2593926
CSACSA
6
2593926
25939262593926
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
The instruments are in accordance with the following North-American standards:
United States of America:
United States of America:
United States of America:United States of America:
FM Class 3600:2011:Electrical equipment used in hazardous (classified) locations. General requirements.
FM Class 3610:2010:Intrinsically safe apparatus and associated apparatus for use in Class I, II and III, Division 1 hazardous (classified) locations.
FM Class 3615:2006:Explosionproof Electrical Equipment. General requirements.
FM Class 3810:2005:Electrical and electronic test, measuring and process control equipment.
Canada:
Canada:
Canada:Canada:
INTRODUCTION 1
C22.2 No. 0.4-M1982 (1993):Bonding and grounding of electrical equipment (protective grounding), Canadian Electrical Code, Part II.
C22.2 No. 30-M1986 (2003):Explosion-proof enclosures for use in Class I hazardous locations.
C22.2 No. 157-92 (1997):Intrinsically safe and non-incendive equipment for use in hazardous locations.
INFORMATION!
Additional drawings for cCSAus and FM (ao. examples for nameplates, schematic diagrams, I/O connections) are part of scope of the delivery. For more info see the applicable sections in this manual.
1.4 OPTISONIC 3400 C(/i)-DIV1
The OPTISONIC 3400 C(/i)-DIV1 is the compact configuration of the UFC 400 ultrasonic signal converter and the OPTISONIC 3000 ultrasonic flow sensor. It provides the ultrasonic transducers in the flow sensor with intrinsically safe signals, which are only internal circuits. The flow converter is provided with non-incendive or intrinsically safe in-/outputs, indicated by an "i" in the type name and listed as OPTISONIC 3400 C/i- DIV1.
The in-/outputs and mains supply connections are located in the terminal compartment,which can either be configured as non-incendive (installation in a Division 2 location) or explosion­proof (installation in a Division 1).
The hazardous locations marking is as follows:
USA
USA Class I, DIV1, Gps ABCD, Class II, DIV1, Gps EFG, Class III, DIV1
USAUSA
Canada
Canada Class I, DIV1, Gps CD, Class II, DIV1, Gps EFG, Class III, DIV1 b
CanadaCanada
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
7
1 INTRODUCTION
1.5 OPTISONIC 3000 F/...-DIV1
The OPTISONIC 3000 F/...-DIV1 is the remote (field) configuration of the ultrasonic flow sensor and has intrinsically safe transducer circuits. It is available in 4 versions as follows,
Cryogenic: low temperature (LT) versions, suitable for process temperatures of -200...+180°C / -328...+356°F
Standard versions: for process temperature of -45...+180°C / -58...+356°F
Extra extended versions: (XXT) for process temperatures of -45...+250°C / -58...+482°F
High Viscosity versions (HV): for process temperatures of -45...+180°C / -58...+356°F
The normal and XXT-versions are also available as flow sensors, equiped with a closed steel circuit for hot medium (e.g. steam or hot oil), which is called heating jacket (HJ). The maximum temperature of the heating medium never exceeds the maximum permissible process
temperature of 180°C (356°F) for the standard versions and 250°C (482°F) for the XXT version.
The flow sensor equipped with a heating jacket, ensures stable process temperatures.
Marking for the different flow sensor types, are listed below.
OPTISONIC 3400
OPTISONIC 3000 F/...-DIV1
OPTISONIC 3000 F/...-DIV1
OPTISONIC 3000 F/...-DIV1OPTISONIC 3000 F/...-DIV1
USA & Canada: Class I, DIV1, Gps ABCD, Class II, DIV1, Gps EFG, Class III, DIV1
The intrinsically safe transducer circuits of the OPTISONIC 3000 F/...-DIV1 are connected to an associated device and have the following maximum values:
V
) = 13,1 V I
max(Ui
INFORMATION!
When thermally insulating the ultrasonic flow sensor, make sure that the temperature of the connection box does not exceed 90
1.6 UFC 400 F(/i)-DIV1
The UFC 400 F(/i)-DIV1 is the remote (field) configuration of the ultrasonic signal converter and has intrinsically safe connections to the ultrasonic flow sensor in remote design. The ultrasonic signal converter is either provided with non-incendive or intrinsically safe signal in-/outputs indicated by an "i" in the type name as UFC 400 F/i-DIV1. The in-/outputs and mains supply connections are located in the terminal compartment, which can either be configured as non­incendive (installation in a Division 2 location) or explosion-proof (installation in a Division 1)..
The hazardous locations marking is as follows:
) = 250 mA Ci = 11,6 nF L
max(Ii
°
C (194°F).
= 500 µH
i
USA
USA Class I, DIV1, Gps ABCD, Class II, DIV1, Gps EFG, Class III, DIV1
USAUSA
Canada
Canada Class I, DIV1, Gps CD, Class II, DIV1, Gps EFG, Class III, DIV1 b
CanadaCanada
The intrinsically safe transducer output connections have the following values:
Voc(Uo) = 8,2 V Ioc(Io) = 210 mA Po = 435 mW Ca(Co) =528 nF La(Lo) = 800 μH
8
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
1.7 Marking labels (examples)
1.7.1 CSA labels
See the following marking labels (i.e. data stickers) of respectively OPTISONIC 3400 F/ -DIV1 ultrasonic compact flowmeter, UFC 400 F(/i)-DIV1 ultrasonic signal converter in field design.
INTRODUCTION 1
Figure 1-1: Example of OPTISONIC 3400 ..DIV1 sticker
1. Name and address of the manufacturer
2. Type designation of the flowmeter, serial number and year of manufacturing
3. Process temperature range and meter size
4. Mains supply data
5. NEMA rating, meter constant GK and Tag number
6. cCSA
CSAus approval mark and hazardous location marking
CSACSA
7. Ambient temperature range & code, output data reference
8. Warning message English
9. Warning message French
10. Reference to control drawing
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
9
1 INTRODUCTION
Figure 1-2: Example of UFC 400 ..DIV1 sticker
1. Name and address of the manufacturer
2. Type designation of the flowmeter, serial number and year of manufacturing
3. Reserved
4. Mains supply data
5. NEMA rating, meter constant GK and Tag number
6. cCSA
CSAus approval mark and hazardous location marking
CSACSA
7. Ambient temperature range & code, output data reference
8. Warning message English
9. Warning message French
10. Reference to control drawing
OPTISONIC 3400
10
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
1.7.2 I/O sticker
Figure 1-3: I/O sticker on cover of terminal compartment
Example of I/O sticker on the inside of the terminal compartment for modules;
INTRODUCTION 1
Fieldbus IO PA on IO board 1.
Intrinsic safety option with current out active
Pulse/Status Out/Control in on IO board 2
Other combinations are possible. See the approval description of the UFC 400 electronics unit
INFORMATION!
Detailed information of I/O label ( and/or Nameplate / stickers) is available on the KROHNE Altometer control drawings nr; 8.30736.11 D - 8.30736.12 B - 8.30736.13 D - 8.3073629 B -
8.30736.40
1.7.3 Nameplate for the measuring sensor (field version)
Examples for measuring sensor versions in CSA / DIV1 version.
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
11
1 INTRODUCTION
OPTISONIC 3400
12
1. cCSA
CSAus approval mark and hazardous location marking
CSACSA
2. Hazardous location marking
3. Warning message in English & French
4. Ambient and process temperatures, temperature code & max. pressure
5. Reference to control drawing
6. Meter constant GK
7. Year of manufacturing, meter size, NEMA rating & text "DUAL SEAL"
8. Type designation of the flowmeter, serial & tag number
9. Name and address of the manufacturer
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
2.1 General
Due to the influence of the process medium temperature (and heating temperature for HJ­versions), ultrasonic flow sensors in field design with type designation OPTISONIC 3000 F/..DIV1 and ultrasonic compact flowmeters with type designation OPTISONIC 3400 C(/i)- DIV1 are not allocated to any fixed temperature class. See the temperature classification tables on the following page for more details. The temperature limits apply under the following conditions:
The device is installed and operated in accordance with the installation directions given in the manual.
The device is not heated up by any additional heat radiation (direct solar radiation, heat from adjacent plant parts) so causing it to operate above the permissible ambient temperature range.
Insulation is not hindering free ventilation of the ultrasonic signal converter housing.
For Class II, group E and F the maximum allowed surface temperature is limited to 200°C and
for Class II, group G and Class III it is limited to 165°C.
2.2 UFC 400 F(/i)-DIV1
TEMPERATURE LIMITS 2
The UFC 400 F/-DIV1 flow converter in remote design is suitable for ambient temperatures in the range -40 to +60°C. The surface temperature always remains below 80°C.
The flow converter inremote design is not influenced by the process temperature, because it is installed on a distance of the remote flow sensor. It therefore has a temperature rating of T6.
2.3 OPTISONIC 3400 C(/i)-DIV1
The OPTISONIC 3400 C/-DIV1 compact flowmeter is suitable for ambient temperatures in the range of -40...60°C.
The following temperature limitation / classification should be maintained:
OPTISONIC 3400 C-.. with aluminum MH 300 housing and basic I/O.
Temperature class
Temperature class Maximum process temperature [
Temperature classTemperature class
T6 80 80 80 80
T5 95 95 95 95
T4 130 130 130 95
T3 180 180 145 95
Maximum process temperature [ºC] at ...
Maximum process temperature [Maximum process temperature [
Ta
Ta 40
40ºC
Ta Ta
C Ta
40 40
C C
Ta 50
50ºC
Ta Ta
50 50
C] at ...
C] at ...C] at ...
C Ta
CC
Ta 55
Ta Ta
55ºC
C Ta
55 55
CC
Ta 60
60ºC
Ta Ta
60 60
C
C C
OPTISONIC 3400 C/i-.. with aluminum MH 300 housing and all other I/O configurations.
Temperature class
Temperature class Maximum process temperature [
Temperature classTemperature class
T6 80 80 75 25
T5 95 95 75 25
T4 130 130 75 25
T3 180 130 75 25
Maximum process temperature [ºC] at ...
Maximum process temperature [Maximum process temperature [
Ta
Ta 45
45ºC
Ta Ta
C Ta
45 45
C C
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
Ta 50
50ºC
Ta Ta
50 50
C] at ...
C] at ...C] at ...
C Ta
CC
Ta 55
55ºC
Ta Ta
55 55
C Ta
CC
Ta 60
Ta Ta
60ºC
C
60 60
C C
13
2 TEMPERATURE LIMITS
OPTISONIC 3400 C-.. with stainless steel MH 300 housing and basic I/O.
Temperatur
Temperatur
TemperaturTemperatur e class
e class
e classe class
T6 80 80 80 80 65
T5 95 95 95 95 65
T4 130 130 130 105 65
T3 180 175 140 105 65
Maximum process temperature [
Maximum process temperature [ºC] at ...
Maximum process temperature [Maximum process temperature [
Ta
Ta 40
40ºC
C Ta
Ta Ta
4040
C C
Ta 45
45ºC
Ta Ta
45 45
C Ta
Ta 50
CC
Ta Ta
C] at ...
C] at ...C] at ...
50ºC
C Ta
50 50
CC
Ta 55
55ºC
Ta Ta
55 55
C Ta
Ta 60
CC
Ta Ta
OPTISONIC 3400
60ºC
C
60 60
C C
OPTISONIC 3400 C/i
Temperatur
Temperatur
TemperaturTemperatur e class
e class
e classe class
T6 80 80 80 45 10
T5 95 95 80 45 10
T4 130 120 80 45 10
T3 155 120 80 45 10
Maximum process temperature [
Maximum process temperature [ºC] at ...
Maximum process temperature [Maximum process temperature [
Ta
Ta 40
Ta Ta
-..with stainless steel MH 300 housing and all other I/O configurations.
40ºC
C Ta
4040
C C
Ta 45
Ta Ta
2.4 OPTISONIC 3000 F/...-DIV1
All the OPTISONIC 3000 F /...-DIV1 ultrasonic flow sensors in field version, are designed for an
ambient temperature range of -40..+70°C.
The following tables show the temperature limitation / classification.
OPTISONIC 3000 F/ (default & LT / XXT).
Temperature class
Temperature class Maximum process temperature [
Temperature classTemperature class
T6 80 80
T5 95 95
T4 130 130
T3 180 195
T2 n.a. 250
C] at ...
C] at ...C] at ...
45ºC
C Ta
Ta 50
50ºC
C Ta
Ta 55
55ºC
45 45
CC
Ta Ta
50 50
CC
Ta Ta
Maximum process temperature [ºC] at T
Maximum process temperature [Maximum process temperature [
OPTISONIC 3000 F-... &
OPTISONIC 3000 F-... &
OPTISONIC 3000 F-... & OPTISONIC 3000 F-... & Cryogenic versions (LT)
Cryogenic versions (LT)
Cryogenic versions (LT)Cryogenic versions (LT)
C Ta
55 55
CC
C] at T
C] at TC] at T
OPTISONIC 3000 F & Extended
OPTISONIC 3000 F & Extended
OPTISONIC 3000 F & Extended OPTISONIC 3000 F & Extended temperature versions (XXT)
temperature versions (XXT)
temperature versions (XXT)temperature versions (XXT)
Ta 60
60ºC
Ta Ta
60 60
= 70ºC
aaaa
C
C C
14
OPTISONIC 3000 F//HJ-... (with heating jacket).
Temperature class
Temperature class Maximum process (& heating for HJ) temperature [
Temperature classTemperature class
T6 80 80
T5 95 95
T4 130 130
T3 180 195
T2 n.a. 250
Maximum process (& heating for HJ) temperature [ºC] at T
Maximum process (& heating for HJ) temperature [Maximum process (& heating for HJ) temperature [
OPTISONIC 3000 F/HJ-...
OPTISONIC 3000 F/HJ-... OPTISONIC 3000 F/XXT/HJ-...
OPTISONIC 3000 F/HJ-...OPTISONIC 3000 F/HJ-...
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3000 F/XXT/HJ-...
OPTISONIC 3000 F/XXT/HJ-...OPTISONIC 3000 F/XXT/HJ-...
C] at T
C] at TC] at T
= 70ºC
aaaa
OPTISONIC 3400
3.1 General
In the case of separate systems, the electrical connection between OPTISONIC 3000 F/...-DIV1 ultrasonic flow sensor and UFC 400 F...-DIV1 ultrasonic signal converter is established via a signal cable of type MR06-RG316. This cable consists of six coaxial cables surrounded by an additional screen and outer insulation layer. The ends of the coaxial cables are provided with SMB plugs. The signal cable is included with the field system by KROHNE.
3.2 Cable marking
The SMB plugs of the coaxial cables are marked by yellow tubing with a black number of subsequently 1.1, 1.2, 2.1, 2.2, 3.1 and 3.2. The first number represents the path 1, 2 or 3 and the second the transducer of that path (1 or 2). The marking of the SMB counter-plugs in the connection box of the signal converter is respectively 1.1, 1.2, 2.1, 2.2, 3.1 and 3.2. Also refer to
Signal cable connections (field versions)
box types.
3.3 Cable parameters
The maximum permitted total capacitance and inductance for the connecting cable is:
CONNECTION OF SEPARATE SYSTEMS 3
on page 16 for the connection of the several connection
= Ca(Co) -Ci = 528 - 11.6 = 516.4 nF
C
L
= La (Lo) - Li = 800 - 500 = 300 μH
L
L
The cable supplied by the manufacturer has the following parameters: distributed capacitance C
distributed inductance L
(core/screen) = 94 pF/m
C
(core/screen) = 0.24 μH/m
C
INFORMATION!
The standard length of the signal cable is 5 m / 16.4 ft. In case a longer length is required, please contact your local representative for detailed information.
INFORMATION!
More than 30 m signal cable length is not recommendable, it will have a negative effect on the measuring accuracy.
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
15
3 CONNECTION OF SEPARATE SYSTEMS
3.4 Equipotential bonding
3.4.1 Signal converter
OPTISONIC 3400
The UFC 400 F...-DIV1 ultrasonic signal converter must always equipotential bonding system of the installation in the hazardous classified location. For this purpose it must be connected to the external U-clamp screw terminal (size M5) of the UFC 400 F converter.
The separate bonding conductor must be at least 4 mm it is mechanically protected (see regulations ANSI/ISA RP 12.06.01 - ANSI/NFPA 70). Make sure that the core of the bonding wire is properly mounted under the U-clamp of the PE-terminal and that the screw is tightly fixed.
The intrinsically safe transducer circuits of the UFC 400 F signal converter are galvanically isolated from earth, therefore an equipotential bonding conductor between the flow sensor (grounded intrinsically safe transducer circuits) and the signal converter (floating-intrinsically safe transducer circuits) does not have to be connected, but can be used if desired.
CAUTION!
°
When the ambient temperatures exceed 60
C / 140°F, the cables and cable gland used (for
connection of the UFC 400F/...) must be suitable for of at least 75
3.5 Signal cable connections (field versions)
See the pictures below for details.
must always be incorporated within the
must alwaysmust always
2
(11 AWG) or 2.5 mm2 (14 AWG), in case
°
C / 167°F.
Electrical connection - Standard version
Figure 3-1: Connect the cables in the connection box of the flow sensor
16
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
Connection of flow sensor type Cryogenic and XXT
Figure 3-2: Connect the cables in the connection box of the flow sensor
CONNECTION OF SEPARATE SYSTEMS 3
Figure 3-3: Construction of field version
1 Signal converter 2 Open connection box 3 Tool for releasing connectors 4 Marking on cable 5 Insert cable(s) into terminal compartment
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
17
4 ELECTRICAL CONNECTIONS
4.1 General
The OPTISONIC 3400 C(/i) -DIV1 compact flowmeter and UFC 400 F(/i) -DIV1 signal converter (separate version) must be incorporated in the equipotential bonding system of the installation. This can be established internally by connection of the protective earth (PE) conductor of the mains supply system to the internal PE clamp, or externally, by connecting a separate equipotential bonding conductor to the external U-clamp terminal (size M5) at respectively the flange of the mounting support (in case of compact instruments) or at the mounting device (for signal converters in separate version). A separate bonding conductor must have a cross-
2
sectional area of at least 4 mm
The display cover seals the electronics compartment of the converter housing and provides type
of protection explosion-proof enclosure. The terminal compartment is default in type of
protection "non-incendive" and can optionally be performed as explosion-proof enclosure . The threaded joints formed by the covers and housing are a tight fit due to the requirements for
type of protection explosion-proof enclosure. Screw the covers on and off with care and never
use excessive force !
Keep the screw-threads free of dirt and well-greased (e.g. with PTFE grease). The grease will help to prevent the threads from locking due to corrosion. To unscrew the covers, first release the interlocking devices (one at each cover). Therefore unscrew the M4 head screw with internal hexagon socket set using a No. 3 Allen key until the interlocking device can be turned. After the covers are screwed back onto the housing, make sure that the interlocking devices are properly refitted.
WARNING!
Allow the electronics to de-energize before opening the electronics compartment of the flow converter housing. Wait at least 35 minutes for T6 and 10 minutes for T5 before opening.
.
OPTISONIC 3400
INFORMATION!
For more detailed information regarding electrical connections for USA / Canada (FM / CSA) DIV1 systems, see the KROHNE Altometer control drawings nr; 8.30736.32 A & 8.30736.30 B
Figure 4-1: Electrical connections
1 Unscrew interlocking head screw 2 Turn cover counterclockwise and remove 3 Open / close safety lid of mains supply section 4 Mains supply & signal / data terminals
18
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
Figure 4-2: Electrical connections
1 Unscrew interlocking head screw 2 Turn cover counterclockwise and remove 3 Open / close safety lid of mains supply section 4 Mains supply & signal / data terminals
ELECTRICAL CONNECTIONS 4
Terminals
Terminals Function, electrical data
TerminalsTerminals
L, N
L+, L-
A, A-, A+ B, B­C, C­D, D-
Function, electrical data
Function, electrical dataFunction, electrical data
Connections for mains supply, always non- I.S.
100...230 V AC, +10%/-15%, 22 VA
12...24 V DC, +30%/-25%, 12 W 24 V AC, +10%/-15%, 22 VA 24 V DC, +30%/-25%, 12 W Um = 253 V
Connections for signal I/Os (PELV circuits), non-I.S. or I.S., are dependent on the specific version of the UFC 400 converter ordered. Consult the tables with CG35 numbers for details.
The exact I/O-configuration for circuits A, B, C and D is order-specific and can be determined by the CG35 number shown on the I/O sticker inside the terminal compartment. Therefore check the data on the back of the UFC 400 electronics unit. The CG35 number contains 10 characters of which the last three characters (XYZ) determine the configuration of the I/O circuits:
CG35
CG35 **** **** **** XXXX YYYY ZZZZ
CG35CG35
Pos 1...4 5 6 7 8 9 10
determine I/O circuits
For schematic overviews of the CG35 numbers, refer to refer to
I.S. I/O connections
on page 23. These overviews do not show all details. The exact
Non-I.S. I/O connections
on page 21 and
connection diagram of a specific UFC 400 signal converter can be found on the sticker inside the terminal compartment.
INFORMATION!
For more detailed information regarding I/O connections for USA / Canada (FM / CSA) DIV1 systems, see the KROHNE Altometer control drawing nr; 8.30736.31 A
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
19
4 ELECTRICAL CONNECTIONS
4.2 Cable glands
The three cable entry holes in the MH-300-... housing have a M20x1.5 screw thread. Make sure that the custom selected cable glands and/or stopping plugs have the same screw thread and are approved for installation and use in North-America.
For use in gaseous hazardous areas:
For use in gaseous hazardous areas: the chosen cable glands and/or stopping plugs, must have
For use in gaseous hazardous areas:For use in gaseous hazardous areas: the appropriate type of protection, namely non-incendive when installed in a Division 2 and explosion-proof in a Division 1 hazardous location. They MUST be suitable for the conditions of use and correctly installed.
Default the MH 300 housing is provided with two for Division 1 approved cable glands and one approved stopping plug. Unused openings must always be sealed with Division 1 certified stopping plugs.
4.3 Field Wiring
OPTISONIC 3400
The wiring of instruments has to be in accordance with the requirements as specified in the relevant national or international standard for electrical installations in hazardous areas. Make
sure to use wires that are rated at least 18°C / 64°F above ambient temperature, for the field
wiring connection of the UFC 400...DIV1 flow converter (compact or remote design). A sticker with warning messsage is located in the terminal compartment of the MH 300.
INFORMATION!
For more detailed information regarding the warning sticker, see control drawing 8.30736.40
20
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
4.4 Non-I.S. I/O connections
The following non-intrinsically safe I/O (inputs/outputs) are available:
I/O PCB
I/O PCB Input/output functions;
I/O PCBI/O PCB
Basic I/O Current Output, active or passive, with HART
Modular I/O Current Output, active or passive, with HART
Modular carrier with 1 or 2 I/O modules Each module: 1 out of following 3 in-/output
Fieldbus I/O Foundation Fieldbus
RS 485 Modbus Modbus with or without termination
ELECTRICAL CONNECTIONS 4
Input/output functions;
Input/output functions;Input/output functions; Vn < 32 V DC, In < 100 mA, Vm = 253 V
Status Output / Control Input Status Output Pulse / Status Output
Pulse / Status Output, active or passive, highC or Namur
functions: Current Output, active or passive Pulse / Status Output, active or passive, highC or Namur Control Input, active or passive, highC or Namur
Profibus-PA
Notes:
The options separated by / are software selectable (can be changed by the user).
The options separated by or are hardware versions (must be ordered as such).
All outputs are passive unless otherwise indicated.
HighC means High Current input/output, Namur means that the in-/outputs are according to
the NAMUR NE43 standard.
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
21
4 ELECTRICAL CONNECTIONS
Overview of the possible combinations, defined by characters XYZ of the CG35 number
Overview of the possible combinations, defined by characters XYZ of the CG35 number
Overview of the possible combinations, defined by characters XYZ of the CG35 numberOverview of the possible combinations, defined by characters XYZ of the CG35 number
Characters XYZ
Characters XYZ Name of I/O
Characters XYZCharacters XYZ
100 Basic I/O CO
488 to 4LL 588 to 5LL 688 to 6LL 788 to 7LL 888 to 8LL A88 to ALL B88 to BLL C88 to CLL
D88 Fieldbus I/O
D8A to DLL Fieldbus I/O
E88 Fieldbus I/O
E8A to ELL Fieldbus I/O
G00 to GLL RS485 Modbus Many combinations possible DP(a) DP(a)
H00 to HLL Modbus with 1
Used abbreviations for in-/output functions: CO = Current Output, PO = Pulse Output, SO = Status Output, CI = Control Input, PA = Profibus PA, FF = Foundation Fieldbus, RS485 = RS485 Modbus, n.c. = not connected. All in-/outputs are passive unless otherwise noted as active with extension (a).
Name of I/O
Name of I/O Name of I/O circuits
circuits
circuitscircuits
Modular I/O
or
Modular Carrier with 1 or 2 I/O modules
Profibus PA
Profibus PA with Module Carrier with 1 or 2 I/O modules
Foundation Fieldbus
Foundation Fieldbus with Module Carrier with 1 or 2 I/O modules
or 2 I/O modules
Terminals
Terminals
TerminalsTerminals A, A-, A+
A, A-, A+
A, A-, A+A, A-, A+
CO(a) over A+
Many combinations possible
n.c. n.c. PA PA
Many combinations possible PA PA
n.c. n.c. FF FF
Many combinations possible FF FF
OPTISONIC 3400
Terminals
Terminals
TerminalsTerminals B, B-
B, B-
B, B-B, B-
SO/CI SO PO/SO
Terminals
Terminals
TerminalsTerminals C, C-
C, C-
C, C-C, C-
RS485 RS485
Terminals
Terminals
TerminalsTerminals D, D-
D, D-
D, D-D, D-
22
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
4.5 I.S. I/O connections
The following intrinsically safe I/O connections are available:
I/O PCB
I/O PCB I/O functions
I/O PCBI/O PCB
Ex i I/O Current Output, passive, +
Ex i Option Current Output, passive V
Fieldbus I/O Profibus-PA
ELECTRICAL CONNECTIONS 4
I/O functions
I/O functionsI/O functions
V
) = 30 V, I
HART communication
Pulse / Status Output V
Current Output, active + HART communication
Pulse / Status Output or Control Input
Current Output, active Voc(Uo) = 21 V, Isc(Io) = 77.5 mA, Po = 407 mW
Foundation Fieldbus
max(Ui
= 5 nF, Li= 5 μH
C
i
) = 30 V, I
max(Ui
Ci = 2.2 nF, Li= 5 μH
Voc(Uo) = 21 V, Isc(Io) = 77.5 mA, Po = 407 mW liniair characteristics, C La(Lo)= 6.0 mH
) = 30 V, I
max(Ui
= 5 nF, Li= 5 μH
C
i
V
) = 30 V, I
max(Ui
Ci = 2.2 nF, Li= 5 μH
liniair characteristics, C La(Lo)= 6.0 mH
V
) = 24 V, I
max(Ui
= 5 nF, Li= 10 μH
C
i
Suitable for connection to an intrinsically safe fieldbus in accordance with the FISCO­model
) = 100 mA, Pi = 1,0 W,
max(Ii
) = 100 mA, Pi = 1,0 W,
max(Ii
(Co) = 188 nF,
a
) = 100 mA, Pi = 1,0 W,
max(Ii
) = 100 mA, Pi = 1,0 W,
max(Ii
(Co) = 188 nF,
a
) = 250 mA, Pi = 1,2 W,
max(Ii
The I/O circuits titled Ex i I/O and Ex i Option are always provided with type of protection Intrinsic Safety . The I/O-circuits Fieldbus I/O Profibus-PA as well as Fieldbus I/O Foundation Fieldbus can be provided with type of protection Intrinsic Safety.
Up to a maximum of 4 intrinsically safe in-/outputs are possible. All intrinsically safe circuits are galvanically isolated with respect to earth and each other. To avoid summation of voltages and current, the wiring of these circuits must be sufficiently separated.
The intrinsically safein-/outputs may only be connected to other intrinsically safe approved devices (e.g. intrinsically safe isolation amplifiers), even if such devices are installed in a non­hazardous location !
Connection to non-I.S. devices, cancels the intrinsically safeproperties of the flowmeter.
Terminals L and N (or L+ and L-) for connection of the mains supply are not available with type of
protection intrinsic safety. To achieve the necessary separation distances between the non-I.S.
and I.S.circuits, the mains terminals are provided with a semi-circular protection cover with a
snap-in lock. This cover MUST be closed before establishing the power supply to the converter.
INFORMATION!
For flow converters installed in a Division 2, the non-incendive terminal compartment can be opened in an energized state for short periods of time, to access the intrinsically safe terminals for possible checks. However, the semi-circular insulation cover over the non-intrinsically safe mains supply terminals L and N (or L+ and L-) MUST be kept closed.
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
23
4 ELECTRICAL CONNECTIONS
Overview of possible intrinsically safe in-/outputs, defined by characters XYZ of the CG 35 numbers
Overview of possible intrinsically safe in-/outputs, defined by characters XYZ of the CG 35 numbers
Overview of possible intrinsically safe in-/outputs, defined by characters XYZ of the CG 35 numbersOverview of possible intrinsically safe in-/outputs, defined by characters XYZ of the CG 35 numbers
Characters XYZ
Characters XYZ Name of I/O
Characters XYZCharacters XYZ
200 Ex i I/O n.c. n.c. CO(a) PO/SO
300 n.c. n.c. CO PO/SO
210 Ex i I/O with Ex
220 CO PO/SO/CI CO(a) PO/SO
310 CO(a) PO/SO/CI CO PO/SO
320 CO PO/SO/CI CO PO/SO
D00 Fieldbus I/O
D10 Fieldbus I/O
D20 CO PO/SO/CI PA PA
E00 Fieldbus I/O
E10 Fieldbus I/O
E20 CO PO/SO/CI FF FF
Used abbreviations for in-/output functions: CO = Current Output, PO = Pulse Output, SO = Status Output, CI = Control Input, PA = Profibus PA, FF = Foundation Fieldbus, n.c. = not connected All in-outputs are passive unless otherwise noted as active with extension (a).
Name of I/O
Name of I/O Name of I/O circuits
circuits
circuitscircuits
i Option
Profibus PA
Profibus PA with Ex i Option
Foundation Fieldbus
Foundation Fieldbus with Ex i Option
Terminals
Terminals
TerminalsTerminals A, A-, A+
A, A-, A+
A, A-, A+A, A-, A+
CO(a) PO/SO/CI CO(a) PO/SO
n.c. n.c. PA PA
CO(a) PO/SO/CI PA PA
n.c. n.c. FF FF
CO(a) PO/SO/CI FF FF
Terminals
Terminals
TerminalsTerminals B, B-
B, B-
B, B-B, B-
Terminals
Terminals
TerminalsTerminals C, C-
C, C-
C, C-C, C-
OPTISONIC 3400
Terminals
Terminals
TerminalsTerminals D, D-
D, D-
D, D-D, D-
24
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
5.1 Maintenance
The flowmeters are maintenance free with respect to the flowmetering properties. Within the scope of periodic inspections required for electrical equipment installed in hazardous areas it is recommended to check the flameproof converter housing and covers for signs of damage or corrosion.
For information about the explosion-proof joints please contact your KROHNE representative.
The four M6 bolts to screw the MH300-Ex housing have a strength of 700N/mm2.
5.2 Before and after opening
WARNING!
the following instructions must always be carefully followed, if the housing of the signal converter has to be opened respectively closed again.
MAINTENANCE AND SERVICE 5
Before opening:
Make absolutely sure that there is no explosion hazard!
Make sure that all connecting cables are safely isolated from all external sources!
Allow the electronics to de-energize before opening the electronics compartment of the
converter housing. Wait at least 35 minutes for T6 and 10 minutes for T5 before opening.
When the instructions above are strictly followed, the display cover (includes glass window) of the electronics compartment may be removed. First unscrew the head screw with internal hexagon socket set (size M4) of the interlocking device by a No. 3 Allen key, until the cover can rotate freely.
After opening:
Before the cover is screwed back onto the housing, the screw-thread must be clean and well­greased with an acid and resin-free grease, e.g. PTFE grease.
Screw the cover as tight as possible onto the housing by hand, until it cannot be opened by hand anymore. Fixate the screw of the interlocking device tight with the No. 3 Allen key.
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
25
5 MAINTENANCE AND SERVICE
5.3 Replacement of mains fuse
WARNING!
Before commencing the work, refer to Before and after opening on page 25
follows:
Pull the display unit of the mounting frame and turn display unit carefully aside.
Unscrew the two screws size M4 that hold the mounting frame with the electronics unit.
Carefully pull the mounting frame with electronics unit out of the housing, until the small
printed circuit board with the six soldered coaxial cables can be pulled off of the sensor driver PC-board. Now carefully remove the unit from the housing, while keeping the small printed circuit board with coaxial cables down, close to the housing wall.
The mains fuse is located in a fuse holder at the back-end of the electronics unit on the top
printed circuit board (power supply PCB). The specifications must be as follows:
Fuse type: 5 x 20 mm (H) CSA/UL certified
Fuse type: 5 x 20 mm (H) CSA/UL certified
Fuse type: 5 x 20 mm (H) CSA/UL certifiedFuse type: 5 x 20 mm (H) CSA/UL certified
Power supply
Power supply Electrical data
Power supplyPower supply
12...24 V DC 250 V / 2 A 5060200000
24 V AC/DC 250 V / 2 A 5060200000
100...230 V AC 250 V / 0,8 A 5080850000
Electrical data KROHNE part number
Electrical dataElectrical data
OPTISONIC 3400
, then continue as
KROHNE part number
KROHNE part numberKROHNE part number
WARNING!
Before reassembling the unit, refer to Before and after opening on page 25
Reassemble the unit in reverse order.
, then:
26
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
5.4 Exchange of electronics unit
Important customer specific data should be documented before replacing the UFC 400 electronics unit. Under normal circumstances, the menu settings are stored on the backplane PC-board that is fixed inside the signal converter housing. When replacing the electronics unit, this information is automatically loaded into the new unit.
Contact KROHNE if:
the unit to be replaced is damaged in such a way that the (customer) settings are lost.
the new electronics unit contains a different (newer) software version that results in data
error messages during start-up.
WARNING!
Before commencing the work, refer to Before and after opening on page 25
follows:
Pull the display unit of the mounting frame and turn the display unit carefully aside.
Unscrew the two screws size M4 that hold the mounting frame with the electronics unit.
Carefully pull the mounting frame with electronics unit out of the converter housing, until the
small printed circuit board with the four (default) or six (optional) soldered coaxial cables can be pulled off of the sensor driver PC-board (take care not to apply a large force on the circuit board while removing the connection board from the sensor driver board). Then remove the complete electronics unit from the housing.
Check if the new electronics unit is undamaged and meets the same mains voltage
specifications and in-/output properties.
Carefully insert the new electronics unit until the small PC-board with six soldered coaxial
cables can be clicked upon the sensor driver PC-board. Install the new unit completely into the housing and screw the two M4 screws. Put the display unit back onto the front of the mounting frame.
MAINTENANCE AND SERVICE 5
, then continue as
WARNING!
Before screwing the display cover back onto the housing, please refer to Before and after opening on page 25
.
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
27
5 MAINTENANCE AND SERVICE
5.4.1 Field version
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
OPTISONIC 3400
Figure 5-1: Unscrew the cover and remove the display
Figure 5-2: Pull off printed circuit board
Perform the following procedures:
Unscrew the display cover of the electronics compartment by hand, by turning it counter
clockwise 1.
Remove the display by using two screwdrivers 2.
Unscrew the two M4 screws 3 at the electronics unit 4.
Pull the two metal pullers 5 at the left and right of the display, using a screwdriver or similar
tool and partially pull out the electronics unit.
28
CAUTION!
Please pay attention that the same amount of force is applied on both pullers, otherwise the connector at the backside can be damaged.
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
Figure 5-3: Small printed circuit board and electronics unit
MAINTENANCE AND SERVICE 5
DANGER!
Electrostatic discharge (ESD) can damage electronic parts. Make sure to discharge yourself by wearing a wrist strap. If no wrist strap is available, ground yourself by touching a metal surface that is grounded.
Remove the printed circuit board 6 from the electronics unit 4.
Check compatibility between the removed and new electronics unit 4, by checking the power
voltage.
Slide the new electronics unit 4 partially back into the housing.
Mount the small printed circuit board back onto the electronics unit 4.
Push the metal pullers 5 back to their original position.
Don't use excessive force, otherwise the connector at the backside can be damaged!
Screw the electronics unit back to the housing.
Re-install the display and make sure not to kink the display's flat ribbon cable.
Replace cover and tighten by hand.
Connect power.
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
29
5 MAINTENANCE AND SERVICE
5.5 Service / repair information
This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:
Due to statutory regulations on environmental protection and safeguarding the health and safety of our personnel, the manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.
This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle.
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products, you are kindly requested:
to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from such dangerous substances,
to enclose a certificate with the device confirming that is safe to handle and stating the product used.
OPTISONIC 3400
30
www.krohne.com 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
OPTISONIC 3400
MAINTENANCE AND SERVICE 5
5.6 Form (for copying) to accompany a returned device
Company: Address:
Department: Name:
Tel. no.: Fax no.:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
This medium is: water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such substances.
We have flushed out and neutralized all cavities in the device.
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned.
Date: Signature:
Stamp:
5.7 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
www.krohne.com10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01
31
KROHNE product overview
Electromagnetic flowmeters
Variable area flowmeters
Ultrasonic flowmeters
Mass flowmeters
Vortex flowmeters
Flow controllers
Level meters
Temperature meters
Pressure meters
Analysis products
Products and systems for the oil & gas industry
Measuring systems for the marine industry
Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 47058 Duisburg (Germany) Tel.:+49 (0)203 301 0 Fax:+49 (0)203 301 10389 info@krohne.de
The current list of all KROHNE contacts and addresses can be found at:
© KROHNE 10/2013 - 7313222100 - AD CSA OPTISONIC 3400 -en- R01 - Subject to change without notice.
www.krohne.com
Loading...