4.3 Standard flow sensor ..................................................................................................... 28
4.4 Variant flow sensor; XXT - High Viscosity and Cryogenic (SS) versions........................ 34
2
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
CONTENTS
4.5 Signal converter housing ............................................................................................... 40
5 Notes41
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
3
1
SAFETY INSTRUCTIONS
Warnings and symbols used
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
OPTISONIC 3400
HANDLING
• This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
Safety instructions for the operator
CAUTION!
Installation, assembly, start-up and maintenance may only be performed by appropriately
trained personnel. The regional occupational health and safety directives must always be
observed.
LEGAL NOTICE!
The responsibility as to the suitability and intended use of this device rests solely with the user.
The supplier assumes no responsibility in the event of improper use by the customer. Improper
installation and operation may lead to loss of warranty. In addition, the "Terms and Conditions of
Sale" apply which form the basis of the purchase contract.
INFORMATION!
•
Further information can be found on the supplied CD-ROM in the manual, on the data sheet,
in special manuals, certificates and on the manufacturer's website.
•
If you need to return the device to the manufacturer or supplier, please fill out the form
contained on the CD-ROM and send it with the device. Unfortunately, the manufacturer
cannot repair or inspect the device without the completed form.
4
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
2.1 Scope of delivery
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
The field device will arrive in two cartons. One carton contains the converter and one carton
contains the sensor.
INFORMATION!
Make sure to combine the sensor and the converter correctly, so they match by the devices
serial number.
INSTALLATION
2
Figure 2-1: Scope of delivery - compact version
1 Ordered flowmeter
2 Product documentation
3 Factory calibration certificate
4 CD-ROM with product documentation in available languages
5 Signal cable (remote versions only)
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
5
2
INSTALLATION
2.2 Device description
This ultrasonic flowmeter is designed for the continuous measurement of actual volume flow,
mass flow, flow speed, velocity of sound, gain, SNR and diagnosis value.
Exclusively for measuring conductive and / or non-conductive fluids in closed, completely filled
pipeline circuits.
Your measuring device is supplied ready for operation. The factory settings for the operating
data have been made in accordance with your order specifications.
The following versions are available:
• Compact version (the signal converter is mounted directly on the measuring sensor)
• Remote version (electrical connection to the measuring sensor via signal cable)
OPTISONIC 3400
1 Compact version
2 Remote version
6
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
2.3 Nameplates
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
2.3.1 Example of nameplate for the compact version
INSTALLATION
2
Figure 2-2: Example of nameplate for the compact version
1 Ambient temperature
2 Protection class
3 Tag number
4 PED data, type I / II / II or SEP
5 Mains supply data
6 Electronic revision number
7 Calibration data
8 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies
9 Name and address of the manufacturer
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
7
2
INSTALLATION
2.3.2 Nameplate for the measuring sensor (field version)
Examples for measuring sensor versions in Standard version.
1. Ambient temperature
2. Protection class
3. Tag number
4. PED data, type I / II / II or SEP
5. Calibration data
6. Type designation of the flowmeter and CE sign with number(s) of notified body / bodies
7. Name and address of the manufacturer
OPTISONIC 3400
2.3.3 Examples of nameplates on the signal converter (field version)
Figure 2-3: Examples of nameplates on the signal converter (field version)
1 Ambient temperature
2 Protection class
3 Tag number
4 PED data, type I / II / II or SEP
5 Mains supply data
6 Electronics revision numbers
7 Calibration data
8 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies
9 Name and address of the manufacturer
8
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
Electrical connection data of inputs/outputs (example of basic version)
Electrical connection data of inputs/outputs (example of basic version)
Electrical connection data of inputs/outputs (example of basic version)Electrical connection data of inputs/outputs (example of basic version)
1 Power supply (AC: L and N, DC: L+ and L-, PE for ≥ 24V AC, FE for ≤ 24 VAC and DC)
2 Connection data of connection terminal D/D-
3 Connection data of connection terminal C/C-
4 Connection data of connection terminal B/B-
5 Connection data of connection terminal A/A-, A+ only operable in basic version
INSTALLATION
2
• A = active mode; the signal converter supplies the power for connection of the subsequent
devices
• P = passive mode; external power supply required for operation of the subsequent devices
• N/C = connection terminals not connected
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
9
2
INSTALLATION
2.4 Storage
• Store the device in a dry, dust-free location.
• Avoid continuous direct sunlight.
• Store the device in its original packaging.
• Storage temperature: -50...+70°C / -58...+158°F
2.5 Transport
Signal converter
• Do not lift the signal converter by the cable glands.
Measuring sensor
• Do not lift the measuring sensor by the connection box.
• Use hoisting belts only.
• To transport flange devices, use lifting straps. Wrap these around both process connections.
OPTISONIC 3400
Figure 2-4: Transport
10
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
2.6 Pre-installation requirements
INFORMATION!
To assure a quick, safe and uncomplicated installation, we kindly request you to make provisions
as stated below.
Make sure that you have all necessary tools available:
• Allen key (4 mm)
• Small screwdriver
• Wrench for cable glands
• Wrench for pipe mounting bracket (remote version only) see; on page 17
• Torque wrench for installing flowmeter in pipeline
2.7 General requirements
INFORMATION!
The following precautions must be taken to ensure reliable installation.
•
Make sure that there is adequate space to the sides.
•
Protect the signal converter from direct sunlight and install a sun shade if necessary.
•
Signal converters installed in control cabinets require adequate cooling, e.g. by fan or heat
exchanger.
•
Do not expose the signal converter to intense vibration. The flowmeters are tested for a
vibration level in accordance with IEC 68-2-6.
INSTALLATION
2
2.7.1 Vibration
Figure 2-5: Avoid vibrations
INFORMATION!
In case of expected vibrations, please install a field version.
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
11
2
INSTALLATION
2.8 Installation conditions
2.8.1 Inlet and outlet
Figure 2-6: Recommended inlet and oulet
1 Refer to chapter "Bends in 2 or 3 dimensions"
2 ≥ 3 DN
2.8.2 Bends in 2 or 3 dimensions
OPTISONIC 3400
Figure 2-7: 2 and 3 dimensional bends, in front of flowmeter
1 Bends in 2 dimensions: ≥ 5 DN; bends in 3 dimensions: ≥ 10 DN
2.8.3 T-section
Figure 2-8: Distance behind a T-section
1 ≥ 5 DN
12
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
2.9 Bends
Figure 2-9: Installation in bending pipes
INSTALLATION
2
Figure 2-10: Installation in bending pipes
2.10 Open feed or discharge
Figure 2-11: Open discharge
Install meter on a lowered section of the pipe to ensure a full pipe condition through the meter.
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
13
2
INSTALLATION
2.11 Position of pump
CAUTION!
Never install flowmeter at a pump suction side in order to avoid cavitation or flashing in the
flowmeter.
Figure 2-12: Position of pump
1 ≥ 15 DN
OPTISONIC 3400
2.12 Control valve
Figure 2-13: Installation in front of a control valve
1 ≥ 20 DN
14
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
2.13 Down going pipeline over 5 m /16 ft length
Install air vent downstream of the flowmeter to prevent vacuum. Although this will not harm the
meter, it may cause gases to come out of solution (cavitate) and interfere with proper
measurements.
INSTALLATION
2
Figure 2-14: Down going pipeline over 5 m / 16 ft length
1 ≥ 5 m / 16 ft
2 Install air vent
2.14 Insulation
Figure 2-15: Insulation
1 Connection box
2 Insulation area
WARNING!
The flow sensor can be insulated completely, except for the connection box.
(Ex: maximum temperature, refer to Ex supplement)
For devices used in hazardous area, additional maximum temperature and insulation
precautions apply. Please refer to the Ex documentation!
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
15
2
INSTALLATION
2.15 Mounting
2.16 Flange deviation
CAUTION!
Max. permissible misalignment of pipe flange faces: M
accoording ASME B16.5 Individual flanges. See Appendix 12 ; Flange face aligment of the
General Piping Requirements DEP 31.38.01.11-GEN
0.5 degree,
max
OPTISONIC 3400
Figure 2-16: Flange deviation
1 M
max
2.17 Mounting position
Figure 2-17: Horizontal and vertical mounting
16
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
2.18 Mounting the field housing, remote version
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
2.18.1 Pipe mounting
Figure 2-18: Pipe mounting of the field housing
INSTALLATION
2
1 Fix the signal converter to the pipe.
2 Fasten the signal converter using standard U-bolts and washers.
3 Tighten the nuts.
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
17
2
INSTALLATION
2.18.2 Turning the display of the field housing version
OPTISONIC 3400
Figure 2-19: Turning the display of the field housing version
The display of the field housing version can be turned in 90° increments.
1 Unscrew the cover from the display and operation control unit.
2 Using a suitable tool, pull out the two metal puller devices to the left and right of the display.
3 Pull out the display between the two metal puller devices and rotate it to the required position.
4 Slide the display and then the metal puller devices back into the housing.
5 Re-fit the cover and tighten it by hand.
CAUTION!
The ribbon cable of the display must not be folded or twisted repeatedly.
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resinfree and acid-free grease.
Ensure that the housing gasket is properly fitted, clean and undamaged.
18
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
3.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
ELECTRICAL CONNECTIONS
3
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Signal cable (remote versions only)
The flow sensor is connected to the signal converter via one signal cable, with 6 (labeled) inner
coax cables for the connection of three acoustic paths.
Figure 3-1: Construction of field version
1 Signal converter
2 Open connection box
3 Tool for releasing connectors
4 Marking on cable
5 Insert cable(s) into terminal compartment
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
19
3
ELECTRICAL CONNECTIONS
CAUTION!
To ensure smooth functioning, always use the signal cable(s) included in the delivery.
OPTISONIC 3400
Figure 3-2: Clamp the cables on the shielding bush
1 Cables
2 Cable glands
3 Grounding clamps
4 Cable with metal shielding bush
Electrical connection - Standard version
Figure 3-3: Connect the cables in the connection box of the flow sensor
20
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
Connection of flow sensor type Cryogenic and XXT
Figure 3-4: Connect the cables in the connection box of the flow sensor
INFORMATION!
Connect the cable on connector with similar numeral marking
ELECTRICAL CONNECTIONS
3
3.3 Power supply
WARNING!
When this device is intended for permanent connection to the mains.
It is required (for example for service) to mount an external switch or circuit breaker near the
device for disconnection from the mains. It shall be easily reachable by the operator and marked
as the disconnecting the device for this equipment.
The switch or circuit breaker and wiring has to be suitable for the application and shall also be in
accordance with the local (safety) requirements of the (building) installation
(e.g. IEC 60947-1 / -3)
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
INFORMATION!
The power terminals in the terminal compartments are equipped with additional hinged lids to
prevent accidental contact.
1 100...230 VAC (-15% / +10%), 22 VA
2 24 VDC (-55% / +30%), 12 W
3 24 VAC/DC (AC: -15% / +10%; DC: -25% / +30%), 22 VA or 12 W
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
21
3
ELECTRICAL CONNECTIONS
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.
100...230 VAC (tolerance range: -15% / +10%)
• Note the power supply voltage and frequency (50...60 Hz) on the nameplate.
• The protective ground terminal PE
clamp terminal in the terminal compartment of the signal converter
• For measurement process reasons, a functional ground FE
separate U-clamp terminal in the terminal compartment of the signal converter.
• When connecting to functional extra-low voltages, provide a facility for protective separation
(PELV) (acc. to VDE 0100 / VDE 0106 and/or IEC 364 / IEC 536 or relevant national
regulations).
OPTISONIC 3400
PE of the power supply must be connected to the separate U-
PEPE
FE must be connected to the
FEFE
INFORMATION!
For 24 VDC, 12 VDC-10% is included in the tolerance range.
3.4 Laying electrical cables correctly
Figure 3-5: Protect housing from dust and water
1 Lay the cable in a loop just before the housing.
2 Tighten the screw connection of the cable entry securely.
3 Never mount the housing with the cable entries facing upwards.
4 Seal cable entries that are not needed with a plug.
22
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
3.5 Inputs and outputs, overview
3.5.1 Combinations of the inputs/outputs (I/Os)
This signal converter is available with various input/output combinations.
Basic version
• Has 1 current output, 1 pulse output and 2 status outputs / limit switches.
• The pulse output can be set as status output/limit switch and one of the status outputs as a
control input.
Ex i version
• Depending on the task, the device can be configured with various output modules.
• Current outputs can be active or passive.
• Optionally available also with Foundation Fieldbus and Profibus PA
Modular version
• Depending on the task, the device can be configured with various output modules.
ELECTRICAL CONNECTIONS
3
Bus systems
• The device allows intrinsically safe and non intrinsically safe bus interfaces in combination
with additional modules.
• For connection and operation of bus systems, please note the separate documentation.
Ex option
• For hazardous areas, all of the input/output variants for the housing designs C and F with
terminal compartment in the Ex d (pressure-resistant casing) or Ex e (increased safety)
versions can be delivered.
• Please refer to the separate instructions for connection and operation of the Ex-devices.
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
23
3
ELECTRICAL CONNECTIONS
3.5.2 Description of the CG number
Figure 3-6: Marking (CG number) of the electronics module and input/output variants
1 ID number:5
2 ID number: 0 = standard
3 Power supply option
4 Display (language versions)
5 Input/output version (I/O)
6 1st optional module for connection terminal A
7 2nd optional module for connection terminal B
The last 3 digits of the CG number (5, 6 and 7) indicate the assignment of the terminal
connections. Please refer to the following examples.
Examples for CG number
OPTISONIC 3400
CG 350 x1 100100...230 VAC & standard display; basic I/O: Ia or Ip & Sp/Cp & Sp & Pp/S
p
CG 350 x1 7FK100...230 VAC & standard display; modular I/O: Ia & PN/SN and optional module PN/SN & C
CG 350 x1 4EB24 VDC & standard display; modular I/O: Ia & Pa/Sa and optional module Pp/Sp & I
p
Description of abbreviations and CG identifier for possible optional modules
on terminals A and B
AbbreviationIdentifier for CG No.Description
I
a
I
p
Pa / S
a
Pp / S
p
PN / S
N
C
a
C
p
C
N
-8No additional module installed
-0No further module possible
AActive current output
BPassive current output
CActive pulse output, frequency output, status output or limit switch
(changeable)
EPassive pulse output, frequency output, status output or limit switch
(changeable)
FPassive pulse output, frequency output, status output or limit switch acc.
to NAMUR (changeable)
GActive control input
KPassive control input
HActive control input to NAMUR
Signal converter monitors cable breaks and short circuits acc. to
EN 60947-5-6. Errors indicated on LC display. Error messages possible
via status output.
N
24
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
3.5.3 Fixed, non-alterable input/output versions
This signal converter is available with various input/output combinations.
• The grey boxes in the tables denote unassigned or unused connection terminals.
• In the table, only the final digits of the CG no. are depicted.
• Connection terminal A+ is only operable in the basic input/output version.
CG-No.Connection terminals
A+AA-BB-CC-DD-
Basic in-/output (I/O) (Standard)
1 0 0
Ip + HART® passive
Sp / Cp passive
1
ELECTRICAL CONNECTIONS
2
Sp passivePp / Sp passive
2
3
Ia + HART® active
1
Ex-i in-/outputs (Option)
2 0 0
3 0 0
2 1 0Ia activePN / SNNAMUR
passive
C
p
3 1 0Ia activePN / SNNAMUR
passive
C
p
2 2 0Ip passivePN / SNNAMUR
Cp passive
3 2 0Ip passivePN / SNNAMUR
passive
C
p
1 Function changed by reconnecting
2 Changeable
2
2
2
2
Ia + HART® active
Ip + HART® passive
Ia + HART® active
Ip + HART® passive
Ia + HART® active
Ip + HART® passive
PN / SN NAMUR
PN / SN NAMUR
PN / SN NAMUR
PN / SN NAMUR
PN / SNNAMUR
PN / SNNAMUR
2
2
2
2
2
2
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
25
3
ELECTRICAL CONNECTIONS
3.5.4 Alterable input/output versions
This signal converter is available with various input/output combinations.
• The grey boxes in the tables denote unassigned or unused connection terminals.
• In the table, only the final digits of the CG no. are depicted.
• Term. = (connection) terminal
CG no.Connection terminals
A+AA-BB-CC-DD-
Modular IOs (option)
4 _ _max. 2 optional modules for term. A + B
8 _ _max. 2 optional modules for term. A + B
6 _ _max. 2 optional modules for term. A + B
B _ _max. 2 optional modules for term. A + B
7 _ _max. 2 optional modules for term. A + B
C _ _max. 2 optional modules for term. A + B
Ia + HART® active
Ip + HART® passive
Ia + HART® active
Ip + HART® passive
Ia + HART® active
Ip + HART® passive
OPTISONIC 3400
Pa / Sa active
Pa / Sa active
Pp / Sp passive
Pp / Sp passive
PN / SN NAMUR
PN / SN NAMUR
1
1
1
1
1
1
PROFIBUS PA/DP
D _ _max. 2 optional modules for term. A + BPA+ (2)PA- (2)PA+ (1)PA- (1)
F _ _max. 2 optional modules for term. A + BPA+ (2)PA- (2)PA+ (1)PA- (1)
FOUNDATION Fieldbus (option)
E _ _max. 2 optional modules for term. A + BV/D+ (2)V/D- (2)V/D+ (1)V/D- (1)
Modbus (option)
G _ _
2
1 changeable
2 not activated bus terminator
max. 2 optional modules for term. A + BCommonSign. B
(D1)
Sign. A
(D0)
26
www.krohne.com02/2015 - QS OPTISONIC 3400 - 4002741204-R04 en
OPTISONIC 3400
4.1 Dimensions and weights
Remote version
Remote versiona = 88 mm / 3.5"
Remote versionRemote version
Compact version
Compact versiona = 155 mm / 6.1"
Compact versionCompact version
TECHNICAL DATA
b = 139 mm / 5.5"
c = 106 mm / 4.2"
Total height = H + a
b = 230 mm / 9.1"
c = 260 mm / 10.2"
Total height = H + a
1
2
1
2
4
1 The value may vary depending on the used cable glands.
2 The value depends on version
4.2 Variants
Standard version and
Standard version and
Standard version and Standard version and
Extended temperature -
Extended temperature -
Extended temperature - Extended temperature High Viscosity -
High Viscosity -
High Viscosity - High Viscosity -
Cryogenic versions;