All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without
the prior written authorisation of KROHNE Messtechnik GmbH.
For all GDC devices, the "Electronic Revision" (ER) is consulted to document the revision status
of the electronics according to NE 53. It is easy to see from the ER whether any fault repairs or
major changes to the electronic equipment have taken place and what effect they have had on
compatibility.
Changes and effect on compatibility
1Downwards compatible changes and fault repair with no effect on operation (e.g. spelling
mistakes on display)
2-_Downwards compatible hardware and/or software change of interfaces:
H
HART® Version 7
PPROFIBUS (pending)
FFoundation Fieldbus
MModbus
Xall interfaces
3-_Downwards compatible hardware and/or software change of inputs and outputs:
ICurrent output
F, PFrequency / pulse output
SStatus output
CControl input
Xall inputs and outputs
4Downwards compatible changes with new functions
5Incompatible changes, i.e. electronic equipment must be changed.
SAFETY INSTRUCTIONS 1
INFORMATION!
In the table below, "x" is a placeholder for possible multi-digit alphanumeric combinations,
depending on the available version.
Responsibility for the use of the measuring devices with regard to suitability, intended use and
corrosion resistance of the used materials against the measured fluid lies solely with the
operator.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
The OPTISONIC 3400
OPTISONIC 3400 is designed exclusively for measurements on conductive and / or non-
OPTISONIC 3400OPTISONIC 3400
conductive fluids, in closed completely filled pipeline circuits. Excess of contaminations (gas,
particles, 2 phases) disturb the acoustic signal and thus must be avoided.
OPTISONIC 3400
The overall functionality of the OPTISONIC 3400
actual volume flow, mass flow, flow speed, velocity of sound, gain, SNR, totalized flow mass and
diagnosis values.
1.3 Certification
CE marking
The device fulfils the statutory requirements of the following EC directives:
• EMC Directive 2004/108/EC in conjunction with EN 61326-1: 2006
• Low Voltage Directive 2006/95/EC in conjunction with EN 61010-1: 2001
• NAMUR NE 21/04
The manufacturer certifies successful testing of the product by applying the CE marking.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
OPTISONIC 3400 flowmeter, is the continuous measurement of
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
SAFETY INSTRUCTIONS 1
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.4.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation and operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.4.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of underneath icons.
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
SAFETY INSTRUCTIONS 1
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.5 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
The device will arrive in two cartons. One carton contains the converter and one carton contains
the sensor.
OPTISONIC 3400
Figure 2-1: Scope of delivery - compact version
1 Ordered flowmeter
2 Product documentation
3 Factory calibration certificate
4 CD-ROM with product documentation in available languages
5 Signal cable (remote versions only)
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
This ultrasonic flowmeter is designed for the continuous measurement of actual volume flow,
mass flow, flow speed, velocity of sound, gain, SNR and diagnosis value.
Exclusively for measuring conductive and / or non-conductive fluids in closed, completely filled
pipeline circuits.
Your measuring device is supplied ready for operation. The factory settings for the operating
data have been made in accordance with your order specifications.
The following version is available:
• Compact version (the signal converter is mounted directly on the measuring sensor)
• Remote version (electrical connection to the measuring sensor via signal cable)
1 Cover for electronics and display
2 Cover for power supply and inputs/outputs terminal compartment
3 Cover for measuring sensor terminal compartment
4 Use cable entry 4 or 5 for measuring sensor signal cable
5 (see 4)
6 Cable entry for power supply
7 Cable entry for inputs and outputs
8 Mounting plate for pipe and wall mounting
OPTISONIC 3400
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased.
Use only resin-free and acid-free grease.
Ensure that the housing gasket is properly fitted, clean and undamaged.
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
2.3.1 Example of nameplate for the compact version
DEVICE DESCRIPTION 2
Figure 2-3: Example of nameplate for the compact version
1 Ambient temperature
2 Protection class
3 Tag number
4 PED data, type I / II / II or SEP
5 Mains supply data
6 Electronic revision number
7 Calibration data
8 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies
9 Name and address of the manufacturer
2.3.2 Nameplate for the measuring sensor (field version)
Examples for measuring sensor versions in Standard version.
1. Ambient temperature
2. Protection class
3. Tag number
4. PED data, type I / II / II or SEP
5. Calibration data
6. Type designation of the flowmeter and CE sign with number(s) of notified body / bodies
7. Name and address of the manufacturer
OPTISONIC 3400
2.3.3 Examples of nameplates on the signal converter (field version)
Figure 2-4: Examples of nameplates on the signal converter (field version)
1 Ambient temperature
2 Protection class
3 Tag number
4 PED data, type I / II / II or SEP
5 Mains supply data
6 Electronics revision numbers
7 Calibration data
8 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies
9 Name and address of the manufacturer
Electrical connection data of inputs/outputs (example of basic version)
Electrical connection data of inputs/outputs (example of basic version)
Electrical connection data of inputs/outputs (example of basic version)Electrical connection data of inputs/outputs (example of basic version)
1 Power supply (AC: L and N, DC: L+ and L-, PE for ≥ 24V AC, FE for ≤ 24 VAC and DC)
2 Connection data of connection terminal D/D-
3 Connection data of connection terminal C/C-
4 Connection data of connection terminal B/B-
5 Connection data of connection terminal A/A-, A+ only operable in basic version
DEVICE DESCRIPTION 2
• A = active mode; the signal converter supplies the power for connection of the subsequent
devices
• P = passive mode; external power supply required for operation of the subsequent devices
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Storage
• Store the device in a dry, dust-free location.
• Avoid continuous direct sunlight.
• Store the device in its original packaging.
• Storage temperature: -50...+70°C / -58...+158°F
OPTISONIC 3400
3.3 Transport
Signal converter
• Do not lift the signal converter by the cable glands.
Measuring sensor
• Do not lift the measuring sensor by the connection box.
• Use hoisting belts only.
• To transport flange devices, use lifting straps. Wrap these around both process connections.
Install air vent downstream of the flowmeter to prevent vacuum. Although this will not harm the
meter, it may cause gases to come out of solution (cavitate) and interfere with proper
measurements.
INSTALLATION 3
Figure 3-11: Down going pipeline over 5 m / 16 ft length
1 ≥ 5 m / 16 ft
2 Install air vent
3.6.9 Insulation
Figure 3-12: Insulation
1 Connection box
2 Insulation area
WARNING!
The flow sensor can be insulated completely, except for the connection box.
(Ex: maximum temperature, refer to Ex supplement)
For devices used in hazardous area, additional maximum temperature and insulation
precautions apply. Please refer to the Ex documentation!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
3.8.1 Pipe mounting
Figure 3-15: Pipe mounting of the field housing
INSTALLATION 3
1 Fix the signal converter to the pipe.
2 Fasten the signal converter using standard U-bolts and washers.
3 Tighten the nuts.
3.8.2 Turning the display of the field housing version
OPTISONIC 3400
Figure 3-16: Turning the display of the field housing version
The display of the field housing version can be turned in 90° increments.
1 Unscrew the cover from the display and operation control unit.
2 Using a suitable tool, pull out the two metal puller devices to the left and right of the display.
3 Pull out the display between the two metal puller devices and rotate it to the required position.
4 Slide the display and then the metal puller devices back into the housing.
5 Re-fit the cover and tighten it by hand.
CAUTION!
The ribbon cable of the display must not be folded or twisted repeatedly.
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resinfree and acid-free grease.
Ensure that the housing gasket is properly fitted, clean and undamaged.
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
ELECTRICAL CONNECTIONS 4
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
4.2 Signal cable (remote versions only)
The flow sensor is connected to the signal converter via one signal cable, with 6 (labeled) inner
coax cables for the connection of three acoustic paths.
Figure 4-1: Construction of field version
1 Signal converter
2 Open connection box
3 Tool for releasing connectors
4 Marking on cable
5 Insert cable(s) into terminal compartment
Figure 4-4: Connect the cables in the connection box of the flow sensor
INFORMATION!
Connect the cable on connector with similar numeral marking
ELECTRICAL CONNECTIONS 4
4.3 Power supply
WARNING!
When this device is intended for permanent connection to the mains.
It is required (for example for service) to mount an external switch or circuit breaker near the
device for disconnection from the mains. It shall be easily reachable by the operator and marked
as the disconnecting the device for this equipment.
The switch or circuit breaker and wiring has to be suitable for the application and shall also be in
accordance with the local (safety) requirements of the (building) installation
(e.g. IEC 60947-1 / -3)
INFORMATION!
The power terminals in the terminal compartments are equipped with additional hinged lids to
prevent accidental contact.
1 100...230 VAC (-15% / +10%), 22 VA
2 24 VAC/DC (AC: -15% / +10%; DC: -25% / +30%), 22 VA or 12 W
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.
• Connect the protective ground conductor PE of the mains power supply to the separate
terminal in the terminal compartment of the signal converter.
• Connect the live conductor to the L terminal and the neutral conductor to the N terminal.
24 VAC/DC
• Connect a functional ground FE to the separate U-clamp terminal in the terminal
compartment of the signal converter.
• When connecting to functional extra-low voltages, provide a facility for protective separation
(PELV) (VDE 0100 / VDE 0106 and/or IEC 364 / IEC 536 or relevant national regulations).
4.4 Laying electrical cables correctly
OPTISONIC 3400
Figure 4-5: Protect housing from dust and water
1 Lay the cable in a loop just before the housing.
2 Tighten the screw connection of the cable entry securely.
3 Never mount the housing with the cable entries facing upwards.
4 Seal cable entries that are not needed with a plug.
Figure 4-6: Marking (CG number) of the electronics module and input/output variants
1 ID number:5
2 ID number: 0 = standard
3 Power supply option
4 Display (language versions)
5 Input/output version (I/O)
6 1st optional module for connection terminal A
7 2nd optional module for connection terminal B
The last 3 digits of the CG number (5, 6 and 7) indicate the assignment of the terminal
connections. Please refer to the following examples.
Examples for CG number
OPTISONIC 3400
CG 350 11 100100...230 VAC & standard display; basic I/O: Ia or Ip & Sp/Cp & Sp & Pp/S
p
CG 350 11 7FK100...230 VAC & standard display; modular I/O: Ia & PN/SN and optional module PN/SN & C
CG 350 81 4EB24 VDC & standard display; modular I/O: Ia & Pa/Sa and optional module Pp/Sp & I
p
Description of abbreviations and CG identifier for possible optional modules
on terminals A and B
AbbreviationIdentifier for CG No.Description
I
a
I
p
Pa / S
a
Pp / S
p
PN / S
N
C
a
C
p
C
N
-8No additional module installed
-0No further module possible
AActive current output
BPassive current output
CActive pulse output, frequency output, status output or limit switch
(changeable)
EPassive pulse output, frequency output, status output or limit switch
(changeable)
FPassive pulse output, frequency output, status output or limit switch acc.
to NAMUR (changeable)
GActive control input
KPassive control input
HActive control input to NAMUR
Signal converter monitors cable breaks and short circuits acc. to
EN 60947-5-6. Errors indicated on LC display. Error messages possible
via status output.
Depending on the version, the control inputs must be connected passively or actively or
according to NAMUR EN 60947-5-6! Which I/O version and inputs/outputs are installed in your
signal converter are indicated on the sticker in the cover of the terminal compartment.
• All control inputs are electrically isolated from each other and from all other circuits.
• All operating data and functions can be adjusted.
• Passive mode: external power supply required:
≤ 32 VDC
U
ext
• Active mode: use of the internal power supply:
=24VDC
U
nom
• NAMUR mode: in accordance with EN 60947-5-6
(Active control input to NAMUR EN 60947-5-6: signal converter monitors cable breaks and
short circuits acc. to EN 60947-5-6. Errors indicated on LC display. Error messages possible
via status output.
• For information on the adjustable operating states refer to
ELECTRICAL CONNECTIONS 4
Function tables
on page 63
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
The current outputs must be connected depending on the version! Which I/O versions and
inputs/outputs are installed in your signal converter are indicated on the sticker in the cover of
the terminal compartment.
• All outputs are electrically isolated from each other and from all other circuits.
• All operating data and functions can be adjusted.
• Passive mode: external power U
• Active mode: load impedance R
≤ 450 Ω at I ≤ 22 mA for Ex i outputs
R
L
• Self-monitoring: interruption or load impedance too high in the current output loop
• Error message possible via status output, error indication on LC display.
• Current value error detection can be adjusted.
• Automatic range conversion via threshold or control input. The setting range for the
threshold is between 5 and 80% of Q
smaller to larger range of 1:20 to 1:1.25).
Signaling of the active range possible via a status output (adjustable).
• Forward / reverse flow measurement (F/R mode) is possible.
ext
L
≤ 32 VDC at I ≤ 22 mA
≤ 1kΩ at I ≤ 22 mA;
, ± 0...5% hysteresis (corresponding ratio from
100%
OPTISONIC 3400
INFORMATION!
For further information refer to Connection diagrams of inputs and outputs on page 39
.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
Depending on the version, the pulse and frequency outputs must be connected passively or
actively or according to NAMUR EN 60947-5-6! Which I/O version and inputs/outputs are
installed in your signal converter are indicated on the sticker in the cover of the terminal
compartment.
• All outputs are electrically isolated from each other and from all other circuits.
• All operating data and functions can be adjusted.
• Passive mode:
External power supply required: U
I ≤ 20 mA at f ≤ 10 kHz (override up to f
I ≤ 100 mA at f ≤ 100 Hz
• Active mode:
Use of the internal power supply: U
I ≤ 20 mA at f ≤ 10 kHz (over range up to f
I ≤ 20 mA at f ≤ 100 Hz
• NAMUR mode: passive in accordance with EN 60947-5-6, f ≤ 10 kHz,
over range up to f
• Scaling:
Frequency output: in pulses per time unit (e.g. 1000 pulses/s at Q
Pulse output: quantity per pulse.
• Pulse width:
symmetric (pulse duty factor 1:1, independent of output frequency)
automatic (with fixed pulse width, duty factor approx. 1:1 at Q
fixed (pulse width adjustable as required from 0.05 ms...2 s)
• Forward / reverse flow measurement (F/R mode) is possible.
• All pulse and frequency outputs can also be used as a status output / limit switch.
max
≤ 12 kHz
≤ 32 VDC
ext
max
=24VDC
nom
ELECTRICAL CONNECTIONS 4
≤ 12 kHz)
≤ 12 kHz)
max
);
100%
) or
100%
INFORMATION!
For further information refer to Connection diagrams of inputs and outputs on page 39
.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
Depending on the version, the status outputs and limit switches must be connected passively or
actively or according to NAMUR EN 60947-5-6! Which I/O version and inputs/outputs are
installed in your signal converter are indicated on the sticker in the cover of the terminal
compartment.
• The status outputs / limit switches are electrically isolated from each other and from all
other circuits.
• The output stages of the status outputs/limit switches during simple active or passive
operation behave like relay contacts and can be connected with any polarity.
• All operating data and functions can be adjusted.
• Passive mode: external power supply required:
≤ 32 VDC; I ≤ 100 mA
U
ext
• Active mode: use of the internal power supply:
=24VDC; I ≤ 20 mA
U
nom
• NAMUR mode: passive in accordance with EN 60947-5-6
• For information on the adjustable operating states refer to
Function tables
OPTISONIC 3400
on page 63.
INFORMATION!
For further information refer to Connection diagrams of inputs and outputs on page 39
.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
Depending on the version, the inputs/outputs must be connected passively or actively or acc. to
NAMUR EN 60947-5-6! Which I/O version and inputs/outputs are installed in your signal
converter are indicated on the sticker in the cover of the terminal compartment.
• All groups are electrically isolated from each other and from all other input and output
circuits.
• Passive mode: An external power supply is necessary to operate (activation) the subsequent
devices (U
• Active mode: The signal converter supplies the power for operation (activation) of the
subsequent devices, observe max. operating data.
• Terminals that are not used should not have any conductive connection to other electrically
conductive parts.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
ext
).
Description of the used abbreviations
I
I
a
P
a
P
N
S
a
S
N
C
a
C
N
Current output active or passive
p
P
Pulse/frequency output active or passive
p
Pulse/frequency output passive acc. to NAMUR EN 60947-5-6
S
Status output/limit switch active or passive
p
Status output/limit switch passive acc. to NAMUR EN 60947-5-6
C
Control input active or passive
p
Control input active acc. to NAMUR EN 60947-5-6:
Signal converter monitors cable breaks and short circuits acc. to EN 60947-5-6. Errors
indicated on LC display. Error messages possible via status output.
0...20 mA or 4...20 mA and other
RL is the internal resistance of the measuring point including the cable
resistance
DC voltage source (U
OPTISONIC 3400
), external power supply, any connection polarity
ext
Table 4-1: Description of symbols
DC voltage source (U
connection diagrams
Internal DC voltage source
Controlled internal power source in the device
Electronic or electromagnetic counter
At frequencies above 100 Hz, shielded cables must be used to connect the
counters.
Ri Internal resistance of the counter
The measuring device, consisting of the measuring sensor and the signal converter, is supplied
ready for operation. All operating data have been set at the factory in accordance with your order
specifications.
When the power is switched on, a self test is carried out. After that the device immediately
begins measuring, and the current values are displayed.
Figure 5-1: Displays in measuring mode (examples for 2 or 3 measured values)
x, y and z denote the units of the measured values displayed
START-UP 5
It is possible to change between the two measured value windows, the trend display and the list
with the status messages by pressing the keys ↑ and ↓. For possible status messages, their
meaning and cause refer to
5.2 Switching on the power
Before connecting to power, please check that the system has been correctly installed.
This includes:
• The device must be mechanically safe and mounted in compliance with the regulations.
• The power connections must have been made in compliance with the regulations.
• The electrical terminal compartments must be secured and the covers have been screwed
on.
• Check that the electrical operating data of the power supply are correct.
Figure 6-1: Display and operating elements (Example: flow indication with 2 measuring values)
1 Indicates a possible status message in the status list
2 Tag number (is only indicated if this number was entered previously by the operator)
3 Indicates when a key has been pressed
4 1st measured variable in large representation
5 Bargraph indication
6 Operating keys, optical and mechanical (see table below for function and representation in text)
7 Interface to the GDC bus (not present in all signal converter versions)
8 Infrared sensor (not present in all signal converter versions)
OPTISONIC 3400
CAUTION!
The use of a jumper is only permitted for custody transfer devices to lock the access to custody
transfer relevant parameters. For non custody transfer devices (i.e. process instruments) this
jumper must not be used!
INFORMATION!
•
The switching point for the 4 optical keys is located directly in front of the glass. It is
recommended to activate the keys at right angles to the front. Touching them from the side
can cause incorrect operation.
•
After 5 minutes of inactivity, there is an automatic return to measuring mode. Previously
changed data is not saved.
•
Step over from optical- to push buttons is directly possible. After using push buttons, wait
several minutes before the optical buttons become active again.
mode to menu mode;
press key for 2.5 s,
"Quick Start" menu is
then displayed
^Reset display; "Quick
Access" function
↓ or ↑Switch between display
pages: measured
value 1 + 2, trend page
and status page
Esc (> + ↑)--Return to menu mode
Access to displayed
menu, then 1st
submenu is displayed
Return to measuring
mode but prompt
whether the data should
be saved
Select menuSelect submenu or
Access to displayed
submenu or function
Press 1 to 3 times,
return to menu mode,
data saved
function
without acceptance of
data
6.1.1 Display in measuring mode with 2 or 3 measured values
Parameter and data
mode
For numerical values,
move cursor
(highlighted in blue) one
position to the right
Return to submenu or
function, data saved
Use cursor highlighted
in blue to change
number, unit, property
and to move the decimal
point
Return to submenu or
function without
acceptance of data
Figure 6-2: Example for display in measuring mode with 2 or 3 measured values
1 Indicates a possible status message in the status list
2 Tag number (is only indicated if this number was entered previously by the operator)
3 1st measured variable in large representation
4 Bargraph indication
5 Depiction with 3 measured values
6.1.2 Display for selection of submenu and functions, 3 lines
Figure 6-3: Display for selection of submenu and functions, 3 lines
1 Indicates a possible status message in the status list
2 Menu, submenu or function name
3 Number relating to 6
4 Indicates position within menu, submenu or function list
5 Next menu, submenu or function
(_ _ _ signalise in this line the end of the list)
6 Current menu, submenu or function
7 Previous menu, submenu or function
(_ _ _ signalise in this line the beginning of the list)
OPTISONIC 3400
6.1.3 Display when setting parameters, 4 lines
Figure 6-4: Display when setting parameters, 4 lines
1 Current menu, submenu or function
2 Number relating to 7
3 Denotes factory setting
4 Denotes permissible value range
5 Permissible value range for numeric values
6 Currently set value, unit or function (when selected, appears with white text, blue background)
This is where the data is changed.
7 Current parameter
8 Factory setting of parameter
Figure 6-5: Display when previewing parameters, 4 lines
1 Current menu(s), submenu or function
2 Number relating to 6
3 Denotes a changed parameter (simple check of changed data when browsing through lists)
4 Next parameter
5 Currently set data from 6
6 Current parameter (for selection press key >; then see previous chapter)
7 Factory setting of parameter
OPERATION 6
6.1.5 Using an IR interface (option)
The optical IR interface serves as an adapter for PC-based communication with the signal
converter without opening the housing.
INFORMATION!
•
This device is not part of the scope of delivery.
•
For more information about activation with the functions A6 or C6.6.7.
Figure 6-6: IR interface
1 Glass panel in front of the control and display panel
2 IR interface
3 LED lights up when IR interface is activated.
4 Suction cups
Timeout function
Following activation of the IR interface in Fct. A6 or C6.6.7 the interface must be properly
positioned and attached to the housing with the suction cups within 60 seconds. If this does not
happen within the specified time period, the device can be operated using the optical keys again.
Upon activation, the LED 3 lights up and the optical keys no longer function.
The following tables describe the functions of the standard device with HART® connection.
The functions for Modbus, Foundation Fieldbus and Profibus are described in detail in the
corresponding supplementary instructions.
•
Depending on the device version, not all functions are available.
6.3.1 Menu A, Quick Setup
No.FunctionSetting / Description
A1 Language
A1LanguageLanguage selection depends on the device version.
A2 Tag
A2Tag
Measuring point identifier (Tag no.) (also for HART® operation) appears in
the LCD header (max. 8 digits).
OPERATION 6
A3 Reset
A3Reset
A3.1Reset ErrorsReset Errors? Select: no/yes
A3.2Totaliser 1Reset Totaliser? Select: No / Yes
A3.3Totaliser 2Reset Totaliser? Select: No / Yes
A3.4Totaliser 3Reset Totaliser? Select: No / Yes
A4 Analogue Outputs (only for HART®)
A4Analogue OutputsApplicable to all current outputs (terminals A, B and C), frequency outputs
A4.1Measurement1) Select measurement: Volume Flow / Velocity of Sound/ / Mass Flow /
A4.2UnitSelection of the unit from a list, depending on the measurement.
A4.3Range1) Setting for main current output (range: 0...100%)
A4.4Low Flow Cutoff1) Setting for main current output (sets output value to "0")
A4.5Time Constant1) Setting for main current output (applicable to all flow measurements)
(terminals A, B and D), limit switch (terminals A, B, C, and / or D) and the 1st
display page / line 1.
2) Use for all outputs? (also use this setting for Fct. A4.2...A4.5!)
Setting: no (applies only to the main current output ) / yes (applies to all
analogue outputs)
Setting: 0…x.xx (format and unit, depending on measurement, see A4.1 and
A4.2 above)
2) Use for all outputs? Make setting, see Fct. A4.1 above!
Setting: x.xxx ± x.xxx% (Range: 0.0…20%)
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value ≤ 1st value
2) Use for all outputs? Make setting, see Fct. A4.1 above!
Setting: xxx.x s (range: 000.1…100 s)
2) Use for all outputs? Make setting, see Fct. A4.1 above!
A5Digital OutputsValid for all pulse outputs (terminals A, B and/or D) and totaliser 1.
A5.1Measurement1) Select measurement: Volume Flow / Mass Flow
2) Use for all outputs? (also use this setting for Fct. A5.2...A5.5!)
Setting: No (only for pulse output D) / Yes (for all digital outputs)
A5.2Pulse Value UnitSelection of the unit from a list, depending on the measurement.
A5.3Value p. Pulse1) Setting for pulse output D (volume or mass value per pulse)
A5.4Low Flow Cutoff1) Setting for pulse output D (sets output value to "0")
Setting: xxx.xxx in l/s or kg/s
2) Use for all outputs? Make setting, see Fct. A5.1 above!
Setting: x.xxx ± x.xxx% (range: 0.0…20%)
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value ≤ 1st value
2) Use for all outputs? Make setting, see Fct. A5.1 above!
A6 GDC IR interface
A6GDC IR InterfaceAfter this function has been activated an optical GDC adapter can be
connected to the LC display. After 60 seconds pass without a connection
being established or after the adapter is removed, the function is exited and
the optical keys are active again.
Break (exit function without connection)
Activate (the IR interface adapter and interrupt the optical keys)
B3.2Status DetailsPresent errors and warnings in NE107 groups
B3.3C NumberDisplays C number of the installed electronics
B3.4Process inputDisplays information of the sensor electronics PCB
B3.4.1 Sensor CPUDisplays information of the sensor CPU software
B3.4.2 Sensor DSPDisplays information of the sensor DSP software
B3.4.3 Sensor DriverDisplays information of the sensor driver hardware
B3.5SW. REV. MSDisplays information of Main Software
B3.6SW. REV. UISDisplays information of User Interface
B3.7"Bus interface"Only appears with Profibus, Modbus and FF
B3.7.0 ProfibusDisplays information of the Profibus interface
B3.7.0 Foundation Field
Bus
B3.7.0 ModbusDisplays information of the Modbus interface
B3.8Electronic RevisionDisplays information of the Electronics Revision
B3.9Change LogThe last changes of parameters are listed in this menu point, together with
Displays information of the Foundation Fieldbus interface
date and time. As a reference a CRC (checksum) over all parameters is used.
This reference can be used by the customer for their documentation. The
preview shows the actual CRC.
6.3.3 Menu C; setup
No.FunctionSettings / descriptions
C1 Setup
C1.1 Process Input
C1.1Meter sizeSets the pipe diameter
C1.2 CalibrationZero Offset
C1.2.1 Zero Calibration Direct setting of zero offset
C1.2.2 GKSets the meter factor
C1.3Filters
C1.3.1 LimitationLimits for the flow speed
C1.3.2 Flow DirectionSet polarity of flow direction
C1.3.3 Time constantSets timing constant of the sensor
C1.3.4 Low flow cutoffSets the low flow cutoff
C1.4PlausibilityError filtering
1.4.1 error limitsets the error limit in percentage of the measured value: exceeding values
1.4.2 counter decrease Sets the decrease of the plausibility counter when the measurement is
1.4.3 counter limitSets the limit for the plausibility counter for which measurements will not
C1.5SimulationSimulation
will be discarded and plausibility counter increased
Select measurement:
Volume Flow / Mass Flow / Velocity of sound / flow speed / Gain / SNR /
Diagnostics 1 / Diagnostics 2 /
x.xx…xx.xx _ _ _ (format and unit depend on the measurement, see above)
Select:
Both Polarities (plus and minus values are displayed) /
Positive Polarity (display for negative values = 0) /
Negative Polarity (display for positive values = 0) /
Absolute Value (always displays positive, with both negative and positive
values)
±xxx …±xxx%; range: -150…+150%
x.xxx ± x.xxx%; range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value ≤ 1st value
C2.2.11 Special FunctionsThis function is only available at the terminal B frequency output. At the
C2.2.12 InformationSerial no. of the I/O board, software version no. and production date of the
C2.2.13 SimulationSequence see B1._ Frequency Output X
Off (activated output: switch closed) /
On (activated output: switch open)
same time, 2 frequency outputs must be available:
1st output at terminal A or D / 2nd output at terminal B
The B output is operated as a slave output, controlled and set using master
output A or D
Select:
Off (no phase shift) /
Phase Shift w.r.t. D or A (slave output is B and master output is D or A)
circuit board
C2._ Pulse Output A
C2._Pulse Output XX stands for one of the connection terminals A, B or D
C2._.1Pulse ShapeSpecify the pulse shape
Select:
Symmetric (about 50% on and 50% off) /
Automatic (constant pulse with about 50% on and 50% off at 100% pulse
rate) /
Fixed (fixed pulse rate, setting see below Fct. C2._.3 100% pulse rate)
C2._.2Pulse WidthOnly available if set to "Fixed" in Fct. C2._.1.
Range: 0.05…2000 ms
Note: max. setting value Tp [ms] ≤ 500 / max. pulse rate [1/s], gives the pulse
width = time where the output is activated
C2._.3Max. Pulse RatePulse rate for 100% of the measuring range
Range: 0.0…10000 1/s
Limitation 100% pulse rate ≤ 100/s: I
Limitation 100% pulse rate > 100/s: I
C2._.4MeasurementMeasurements for activating the output
Select: Volume Flow / Mass Flow
C2._.5Pulse Value UnitSelection of the unit from a list, depending on the measurement
C2._.6Value p. PulseSet value for volume or mass per pulse.
xxx.xxx, measured value in [l] or [kg] depending on setting in C3._.6
At max. pulse rate see above 2._.3 Pulse Output
C2._.7PolaritySet polarity, please note Flow Direction
Select:
Both Polarities (plus and minus values are displayed) /
Positive Polarity (display for negative values = 0) /
Negative Polarity (display for positive values = 0) /
Absolute Value (always displays positive, with both negative and positive
values)
C2._.8Low Flow CutoffSets the measurement to "0" for low values
x.xxx ± x.xxx%; range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value ≤ 1st value
C2._.11Phase Shift w.r.t. BOnly available when configuring the A or D terminal and only if output B is a
C2.3.11Special FunctionsThis function is only available at the pulse output of terminal B. At the same
C2._.12InformationSerial no. of the I/O board, software version no. and production date of the
C2._.13SimulationSequence see B1._ Pulse Output X
Off (activated output: switch closed) /
On (activated output: switch open)
pulse or frequency output. If setting in Fct. C2.2.7 is "Both Polarities", the
phase shift is prefixed by a symbol, e.g. -90° and +90°
Select:
Off (no phase shift) /
0° Phase Shift (between outputs A or D and B, inversion possible) /
90° Phase Shift (between outputs A or D and B, inversion possible) /
180° Phase Shift (between outputs A or D and B, inversion possible)
time, 2 pulse outputs must be available:
1st output at terminal A or D / 2nd output at terminal B
The B output is operated as a slave output, controlled and set using master
output A or D
Select:
Off (no phase shift) /
Phase Shift w.r.t. D or A (slave output is B and master output is D or A)
circuit board
C2._ Status Output X
C2._Status Output XX (Y) stands for one of the connection terminals A, B, C or D
C2._.1ModeThe output shows the following measuring conditions:
Out Of Specification (output set, signals status of category "Error in Device"
or "Application Failure" or "Out Of refer to
information
Application Failure (output set, signals status of category "Error in Device"
or "Application Failure" refer to
information
Flow Polarity (polarity of the current flow)
Flow Over Range (over range of the flow)
Totaliser 1 Preset (activates when totaliser X preset value is reached) /
Totaliser 2 Preset (activates when totaliser X preset value is reached) /
Totaliser 3 Preset (activates when totaliser X preset value is reached) /
Output A (activated by the status of output Y, additional output data see
below) /
Output B (activated by the status of output Y, additional output data see
below) /
Output C (activated by the status of output Y, additional output data see
below) /
Output D (activated by the status of output Y, additional output data see
below) /
Off (switched off) /
Empty Pipe (when pipe empty, output activated) /
Error in Device (output set, signals status of category "Error in Device" refer
to
Status messages and diagnostic information
C2._.2Current Output YOnly appears if output A...C is set under "Mode" (see above), and this output
C2._.2Frequency Output Y and
Pulse Output Y
is a "Current Output".
Polarity (is signalled)
Over Range (is signalled)
Automatic Range signals lower range
Only appears if output A, B or D is set under "Mode" (see above), and this
output is a "Frequency/Pulse Output".
C2._.2Status Output YOnly appears if output A...D is set under "Mode" (see above), and this output
C2._.2Limit Switch Y and Control
Input Y
C2._.2Output YOnly appears if output A...D is set under "Mode" (see above) and this output
C2._.3Invert SignalSelect:
C2._.4InformationSerial no. of the I/O board, software version no. and production date of the
C2._.5SimulationSequence see B1._ Status Output X
is a "Status Output".
Same Signal (like other connected status output, signal can be inverted, see
below)
Only appears if output A...D / input A or B is set under "Mode" (see above),
and this output / input is a "Limit Switch / Control Input".
Status Off (is always selected here if Status Output X is connected with a
Limit Switch / Control Input Y).
is switched off.
Off (activated output: switch closed) /
On (activated output: switch open)
circuit board
C2._ Limit Switch X
C2._Limit Switch XX stands for one of the connection terminals A, B, C or D
C2._.1MeasurementSelect:
C2._.2ThresholdSwitching level, set threshold with hysteresis
C2._.3PolaritySet polarity, please note Flow Direction
C2._.4Time ConstantRange: 000.1…100 s
C2._.5Invert SignalSelect:
C2._.6InformationSerial no. of the I/O board, software version no. and production date of the
C2._.7SimulationSequence see B1._ Limit Switch X
Volume Flow / Mass Flow / Flow Speed / Velocity of Sound / Gain / SNR /
Diagnostics 1 / Diagnostics 2
xxx.x ±x.xxx (format and unit depend on the measurement, see above)
(1st value = threshold / 2nd value = hysteresis),
condition: 2nd value ≤ 1st value
Select:
Both Polarities (plus and minus values are displayed) /
Positive Polarity (display for negative values = 0) /
Negative Polarity (display for positive values = 0) /
Absolute Value (always displays positive, with both negative and positive
values)
Off (activated output: switch closed) /
On (activated output: switch open)
Off (control input switched off) /
Hold All Outputs (hold current values, not display and totalisers) /
Output Y (hold current values) /
All Outputs To Zero (current values = 0%, not display and totalisers) /
Output Y To Zero (current value = 0%) /
All Totalisers (reset all totalisers to "0") /
Totaliser "Z" Reset (set totaliser 1, (2 or 3) to "0") /
Stop All Totalisers /
Stop Totaliser "Z" (stops totaliser 1, (2 or 3) /
Zero Outp.+Stop Tot. (all outputs 0%, stop all totalisers, not the display) /
External Range Y (control input for external range of current output Y) - also
make this setting on current output Y (no check if current output Y is
available) /
Error Reset (all resettable errors are deleted)
Zero Calibration
C2._.2Invert SignalSelect:
C2._.3InformationSerial no. of the I/O board, software version no. and production date of the
C2._.4SimulationSequence see B1._ Control Input X
Off (activated output: switch closed) /
On (activated output: switch open)
circuit board
No.FunctionSettings / descriptions
C3 I/O Totalisers
C3.1Totaliser 1Set function of totaliser.
C3.2Totaliser 2
C3.3Totaliser 3
C3._.1Totaliser FunctionSelect:
C3._.2MeasurementSelection of the measurement for Totaliser _
C3._.3Low Flow CutoffSets the measurement to "0" for low values
C3._.4Time ConstantRange: 000.1…100 s
C3._.5Preset ValueIf this value is reached, positive or negative, a signal is generated that can be
C3._.6Reset TotaliserSequence see Fct. A3.2, A3.3 and A3.4
_ stands for 1, 2, 3 (= Totaliser 1, 2, 3)
The basic version (standard) has only 2 totaliser!
Absolute Total (counts positive + negative values) /
+Totaliser (counts only the positive values) /
-Totaliser (counts only the negative values) /
Off (Totaliser is switched off)
Select:
Volume Flow / Mass Flow
Range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value ≤ 1st value
used for a status output at which "Preset Totaliser X" has to be set.
Preset value (max. 8 digits) x.xxxxx in selected unit, see C6.7.10 + 13
C3._.7Set TotaliserSet Totaliser _ to the desired value
Select:
Break (exit function) / Set Value (opens the editor to make the entry)
Query: Set Totaliser?
Select:
Select: No (exit function without setting the value) / Yes (sets the totaliser
and exits the function)
C3._.8Stop TotaliserTotaliser _ stops and holds the current value.
Select:
No (exits the function without stopping the totaliser) / Yes (stops the totaliser
and exits the function)
C3._.9Start TotaliserStart Totaliser _ after that totaliser is stopped
Select:
No (exits the function without starting the totaliser) / Yes (starts the totaliser
and exits the function)
C3._.10InformationSerial no. of the I/O board, software version no. and production date of the
circuit board
No.FunctionSettings / descriptions
C4 I/O HART
C4I/O HART
C4.1PV isCurrent output (primary variable)
C4.2SV is(secondary variable)
C4.3TV is(tertiary variable)
C4.44V is(4th variable)
C4.5HART UnitsChanges units of DVs (dynamic variables) in the display
C4._.1Current Output XShows the current analog measured value of the linked current output. The
C4._.1Frequency Output XShows the current analog measured value of the linked frequency output, if
C4._.1HART Dynamic Var.
Selection or display of the 4 dynamic variables (DV) for HART
The HART® current output (terminal A basic I/Os or terminal C modular I/Os)
always has a fixed link to the primary variables (PV). Fixed links of the other
DVs (1-3) are only possible if additional analogue outputs (current and
frequency output) are available; if not, the measurement can be freely
selected from the following list: in Fct. A4.1 "Measurement"
_ stands for 1, 2, 3 or 4
X stands for connection terminals A...D
Break: return with ^ key
HART® display: copies the settings for the display units to the settings for
DVs
C5.2.1LanguageLanguage selection depends on the device version.
C5.2.2ContrastAdjust display contrast for extreme temperatures. Setting: -9…0…+9
This change takes place immediately, not just when setting mode is exited!
C5.2.3Default DisplaySpecification of the default display page that is returned to after a short
C5.2.4Optical KeysActivate or deactivate the optical keys
delay period.
Select: None (the current page is always active) / 1st Meas. Page (show this
page) / 2nd Meas. Page (show this page) / Status Page (show only status
messages) / Graphic Page (trend display of the 1st measurement)
Select: On / Off
C5.3 and C5.4 1st Meas. Page and 2nd Meas. Page
C5.31st Meas. Page_ stands for 3 = 1st Meas. Page and 4 = 2nd Meas. Page
C5.42nd Meas. Page
C5._.1FunctionSpecify number of measured value lines (font size)
Select: One Line / Two Lines / Three Lines
C5._.21st Line VariableSpecify 1st Line Variable
Select measurement:Volume Flow / Mass Flow / Flow Speed / Velocity of
Sound / Gain / SNR / Diagnostics1 / Diagnostics2
C5._.3Range0…100% of the measurement set in Fct. C5._.2
x.xx…xx.xx _ _ _ (format and unit depend on the measurement)
C5._.4LimitationLimitation before applying the time constant
±xxx ... ±xxx%; range: -150…+150%
C5._.5Low Flow CutoffSets low flow values to "0"
x.xxx ± x.xxx %; Range: 0.0…20 %
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value ≤ 1st value
C5._.6Time ConstantRange: 000.1…100 s
C5._.71st Line FormatSpecify decimal places.
Select: Automatic (adaptation is automatic) /
X(=none)…X.XXXXXXXX (max. 8 digits) depends on size of font
C5._.82nd Line VariableSpecify 2nd Line Variable (only available if this 2nd line is activated)
Select: Bar Graph (for measurement selected in the 1st line ) Volume Flow /
Mass Flow / Flow Speed / Velocity of Sound / Gain / SNR / Diagnostics1 /
Diagnostics2
Totalisers / Operating hours
C5._.92nd Line FormatSpecify decimal places
Select: Automatic (adaptation is automatic) /
X(=none)…X.XXXXXXXX (max. 8 digits) depends on size of font
C5._.103rd Line VariableSpecify 3rd Line Variable (only available if this 3rd line is activated)
Select: Volume Flow / Mass Flow / Flow Speed / Velocity of Sound / Gain /
SNR / Diagnostics1 / Diagnostics2 /Totalisers / Operating hours
C5._.113rd Line FormatSpecify decimal places.
Select: Automatic (adaptation is automatic) /
X(=none)…X.XXXXXXXX (max. 8 digits) depends on size of font
C5.5.1Select RangeGraphic page always shows trend curve of the measurement of the 1st page
C5.5.2RangeSet the scaling for the Y axis. Only available if "Manual" is set in C5.5.1.
C5.5.3Time ScaleSet the time scaling for the X axis, trend curve
/ 1st line, see Fct. C6.3.2
Select: Manual (set range in Fct. C5.5.2) ; Automatic (automatic depiction
based on the measured values)
Reset only after parameter change or after switching off and on.
±xxx ±xxx%; range: -100…+100%
(1st value = lower limit / 2nd value = upper limit),
condition: 1st value ≤ 2nd value
xxx min; range: 0…100 min
C5.6 Special Functions
C5.6Special Functions-
C5.6.1Reset ErrorsReset Errors?
Select: No / Yes
C5.6.2Save SettingsSave current settings.
C5.6.3Load SettingsLoad saved settings
C5.6.4Password Quick SetPassword required to change data in the quick setup menu.
C5.6.5Password SetupPassword required to change data in the setup menu
C5.6.6Date and TimeSet real time
C5.6.7Quick AccessSet Quick Access function;
C5.6.8GDC IR InterfaceAfter this function has been activated an optical GDC adapter can be
Select: Break (exit function without saving) / Backup 1 (save in storage
location 1) / Backup 2 (save in storage location 2)
Query: Continue To Copy? (cannot be done afterwards)
Select: No (exit function without saving) / Yes (copy current settings to
storage backup 1 or backup 2)
Select: Break (exit function without loading) / factory settings (restore
factory settings) / backup 1 (load data from storage location 1) / backup 2
(load data from storage location 2)
Query: Continue To Copy? (cannot be done afterwards)
Select: No (exit the function without saving) / Yes (load data from the
selected storage location)
0000 (= to quick setup menu without password)
xxxx (password required); range 4 digits: 0001…9999
0000 (= to quick setup menu without password)
xxxx (password required); range 4 digits: 0001…9999
Select: Off (deactivated) / Reset Totaliser 1, 2, 3 or All Totalisers
connected to the LC display. If approximately 60 seconds pass without a
connection being established or after the adapter is removed, then the
function is exited and the optical keys are active once again.
Break (exit function without connection)
Activate (the IR interface adapter and interrupt the optical keys)
If approximately 60 seconds pass without a connection being established,
then the function is exited and the optical keys are active once again.
Select: 00 (Point-to-Point operation, current output has normal function,
current = 4…20 mA) / 01...15 (Multi-Drop operation, current output has a
constant setting of 4 mA)
A…Z ; a…z ; 0…9 ; / - + , . *
A…Z ; a…z ; 0…9 ; / - + , . *
®
activated) possible current range for current output 4…20 mA /
®
not activated) possible current range for current output
OPTISONIC 3400
C5.9 Quick Setup
C5.9Quick SetupActivate quick access in Quick Setup menu:
Select: Yes (switched on) / No (switched off)
C5.9.1Reset Totaliser 1Reset Totaliser 1 in Quick Setup menu?
Select: Yes (activated) / No (switched off)
C5.9.2Reset Totaliser 2Reset Totaliser 2 in Quick Setup menu?
Select: Yes (activated) / No (switched off)
C5.9.3Reset Totaliser 3Reset Totaliser 3 in Quick Setup menu?
Select: Yes (activated) / No (switched off)
6.3.4 Set free units
Free unitsSequences to set texts and factors
Texts
Texts
TextsTexts
Volume flow, mass flow and density:3 digits before and after the slash
These settings make it possible to change the status signal of the respective diagnosis message
(status group).
6.4.4 Optical keys
This function can deactivate the optical keys. In the display, the switched off state of the optical
keys is represented by a lock 1.
OPTISONIC 3400
In this case, the device may only be operated using the push buttons.
6.4.5 Graphic page
With this converter, the trend of the main measurement can be graphically displayed. The first
measurement on display page 1 is always defined as the main measurement.
• Menu C5.5.1 defines the range for the trend indicator (manual or automatic).
• Menu C5.5.2 defines the range for manual setting.
• Menu C5.5.3 defines the time span for the trend indicator.
6.4.6 Save settings
This function allows all settings to be stored in a memory.
• Backup 1: Saves settings in backup memory area 1
• Backup 2: Saves settings in backup memory area 2
6.4.7 Load settings
This function allows the complete stored settings to be loaded again.
• Backup 1: Loading from backup - memory 1
• Backup 2: Loading from backup - memory 2
• Factory: Uploading the original factory settings
To create a password for the Quick Set menu or Setup menu, you must enter a 4-digit code into
the menu. You are then prompted for this password every time changes are to be made to the
corresponding menus. There is a hierarchy. The Setup password can also be used to perform
changes in the Quick Setup menu. Enter 0000 in each menu to disable the password.
6.4.9 Date and time
The signal converter has a real time clock which is used for all of the log functions in the device.
This function can be used to set the date and time of the real time clock.
6.4.10 Quick Access
In measuring mode, pressing the ^ key for 2.5 seconds carries out the "Quick Access" function.
Totalisers 1, 2, 3 and All Totalisers can be reset.
6.4.11 Low flow cutoff
The low flow cutoff can be individually set for each output and each display line. If the low flow
cutoff has been activated, the respective output or display is set to zero when the flow is below
the low flow cutoff value entered.
OPERATION 6
The value can either be entered as a percentage of the nominal flow of the sensor or, in the case
of a pulse output, as a discrete flow value.
Two values must be entered. The first is for the operating point of the sensor and the second is
for hysterisis. Condition: 1st value > 2nd value
1 Flow
2 Time
3 Currently indicated flow
4 Display set to zero
5 Currently indicated flow
6 Positive hysteresis
7 Operating point
8 Negative hysteresis
To better process widely fluctuating measured values in the device, the measured values are
digitally filtered to stabilise the output. The time constant can be individually set for each output,
the first line of the display and the density measurement. However, keep in mind that the degree
of filtration affects the response time of the device in the event of rapid changes.
Short time constantFast response times
Long time constantSlow response time
The time constant corresponds to the elapsed time until 67% of the end value has been reached
according to a step function.
6.4.13 Dual phase pulse output
A dual phase pulse or frequency output is possible. This operating mode requires 2 terminal
pairs. Terminal pairs A and B or D and B can be used.
OPTISONIC 3400
Fluctuating reading
Stable reading
In this case, perform the following settings:
• C2.3.11: Phase shift to D or shift to A
• All functions for output B are set using output D or output A.
• C2.5.11: Setting phase shift from output B relative to D, if terminal pair D was selected in
C2.3.11. 0°, 90° or 180° are offered as options.
6.4.14 Timeouts in programming mode
Normal menu function:
Normal menu function: If no key has been pressed for 5 minutes in a normal menu function, the
Normal menu function:Normal menu function:
display automatically switches to measuring mode. All changes are lost.
Test function:
Test function: In test mode, the test function is finished after 60 minutes.
Test function:Test function:
GDC IR Interface:
GDC IR Interface: If the GDC-IR connection is activated, it is cancelled after 60 seconds if no
GDC IR Interface:GDC IR Interface:
connection is established. If the connection is interrupted, the display can be operated again
after 60 seconds using the optical keys.
6.4.15 Output hardware
Depending on the hardware modules used (see CG number), it may be possible to change the
output options on terminals A, B, C or D in the menus C2.1.x. For example: A pulse output to a
frequency output or a status output to a control input.
82
The available options are determined by the hardware module used. It is not possible to change
the type of output, e.g. from active to passive or to NAMUR.
The diagnostic messages are displayed in accordance with NAMUR standard NE 107. NE 107
states that there are up to 32 status groups which can have different status signals. NE 107 was
implemented with 16 status groups with fixed status signals and 8 groups with variable status
signals. To facilitate identification of the problem source, the status groups are then divided into
the groups: Sensor, Electronics, Configuration and Process.
OPERATION 6
The variable status signal can be changed in menu Mapping; C1.10.3 ...8
Mapping; C1.10.3 ...8. Changing the status
Mapping; C1.10.3 ...8Mapping; C1.10.3 ...8
signal to "Information", switches off the message.
INFORMATION!
As status message always the name of the relevant status group and the status signal (F/S/M/C)
is displayed.
Each status message (= status signal) has a specific symbol, determined by NAMUR, which is
displayed with the message. The length of each message is limited to one line.
Symbol LetterStatus signalDescription and consequence
FFFFFailureNo measurement possible.
SOut of specificationMeasurements are available but are
MMaintenance requiredMeasurements are still accurate but
CFunction checkA test function is active; The displayed
IInformationNo direct influence on the
no longer sufficiently accurate and
should be checked
this could soon change
or transferred measured value does
not correspond to the actual measured
value.
measurements
All status messages are saved in the status log (menu B3.1). Use the ↑ and ↓ keys to navigate
through this list. Use the ^ key to exit the list.
The status screen shows the status groups of all errors that have occurred since the last time
the status screen was opened. All errors that are not current disappear after 2 seconds. They
are shown in the list in brackets.
Mass Flow Out of RangeThe flow is out of range. The
Vol. Flow Out of Range
Velocity Out of Range
Backplane Data FaultyThe backplane data record is
Factory Data FaultyFactory settings are invalid.Contact manufacturer.
Backplane DifferenceThe data on the backplane
PROFIBUS BaudrateThe PROFIBUS searches for
Backup 1 Data FaultyError when checking the
Backup 2 Data FaultyError when checking the
by a filter.
actual flow is higher than
the displayed value.
faulty.
differ from the data in the
device.
the current Baudrate.
data record for Backup 1.
data record for Backup 2.
Check the range setting of
the output.
Check process conditions.
Check the correct
installation of the signal
converter electronics. After
changing one parameter,
the message should
disappear within one
minute. If not, contact the
manufacturer.
After changing one
parameter, the message
should disappear within one
minute. If not, contact the
manufacturer.
Use "Setup > Device >
Special Functions > Save
Settings" to save the data
record. If the message
continues to appear, contact
the manufacturer.
The manufacturer adheres to the basic principle that functionally adequate spare parts for each
device or each important accessory part will be kept available for a period of 3 years after
delivery of the last production run for the device.
This regulation only applies to spare parts which are subject to wear and tear under normal
operating conditions.
7.2 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the
warranty. These include repair, maintenance, technical support and training.
INFORMATION!
For more precise information, please contact your local sales office.
7.3 Returning the device to the manufacturer
OPTISONIC 3400
7.3.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance
with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict
attention to the following points:
•
Due to statutory regulations on environmental protection and safeguarding the health and
safety of our personnel, manufacturer may only handle, test and repair returned devices that
have been in contact with products without risk to personnel and environment.
•
This means that the manufacturer can only service this device if it is accompanied by the
following certificate (see next section) confirming that the device is safe to handle.
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products,
you are kindly requested:
•
to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from
such dangerous substances,
•
to enclose a certificate with the device confirming that is safe to handle and stating the
product used.
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle Ultrasonic transit time
Application range Flow measurement of (non) conductive fluids
Ambient temperature Depending on the version and combination of outputs
Protect inside electronics against self-heating (increasing of the electronics temperature will lead to a
reduction of the corresponding service life by a factor 2 for each step of 10°C / 50°F increase).
Protect the signal converter from external heat sources such as direct sunlight, as higher temperatures
reduce the life cycle of all electronic components.
Storage temperature -50…+70°C / -58…+158°F
Pressure
Pressure
PressurePressure
Atmospheric
EN 1092-1DN25…50: PN 40
ASME B16.51…24": 150 lb RF
JISDN25...40: 20K
Properties of medium
Properties of medium
Properties of mediumProperties of medium
Physical condition Liquid, single phase (well mixed, rather clean)
Permissible gas content≤ 2% (volume)
Permissible solid content≤ 5% (volume)
ViscosityStandard:
TECHNICAL DATA 8
Compact version: -45...+140°C / -49...+284°F
Compact version:Compact version:
(for stainless steel housing at ambient temperature ≤
45°C / +113°F)
Remote version:
Remote version: -45...+180°C / -49...+356°F
Remote version:Remote version:
Extended temperature version:
Extended temperature version: -45...+250°C / -49...+482°F
Extended temperature version:Extended temperature version:
(only remote version)