KROHNE OPTIFLUX 1000 Specifications

Technical Datasheet
Technical Datasheet
OPTIFLUX 1000
OPTIFLUX 1000
OPTIFLUX 1000OPTIFLUX 1000
Technical DatasheetTechnical Datasheet
Electromagnetic flow sensor in sandwich design
Lightweight and compact
Excellent price performance ratio
The documentation is only complete when used in combination with the relevant documentation for the signal converter.
© KROHNE 07/2017 - 4000690405 - TD OPTIFLUX 1000 R05 en
CONTENTS
OPTIFLUX 1000
1 Product features 3
1.1 Cost efficient and reliable flow sensor ............................................................................ 3
1.2 Options.............................................................................................................................. 5
1.3 Measuring principle.......................................................................................................... 6
2 Technical data 7
2.1 Technical data................................................................................................................... 7
2.2 Measuring accuracy ....................................................................................................... 11
2.3 Dimensions and weights ................................................................................................ 12
3 Installation 15
3.1 Intended use ................................................................................................................... 15
3.2 General notes on installation ......................................................................................... 15
3.2.1 Vibration ................................................................................................................................ 15
3.2.2 Magnetic field........................................................................................................................ 15
3.3 Installation conditions ....................................................................................................16
3.3.1 Inlet and outlet...................................................................................................................... 16
3.3.2 Bends in 2 or 3 dimensions................................................................................................... 16
3.3.3 T-section ............................................................................................................................... 16
3.3.4 Bends .................................................................................................................................... 17
3.3.5 Open feed or discharge......................................................................................................... 17
3.3.6 Flange deviation.................................................................................................................... 18
3.3.7 Pump ..................................................................................................................................... 18
3.3.8 Control valve ......................................................................................................................... 18
3.3.9 Air venting and vacuum forces ............................................................................................. 19
3.3.10 Mounting position................................................................................................................ 19
4 Electrical connections 20
4.1 Safety instructions.......................................................................................................... 20
4.2 Grounding ....................................................................................................................... 20
4.3 Virtual reference for IFC 300.......................................................................................... 21
5 Notes 22
2
www.krohne.com 07/2017 - 4000690405 - TD OPTIFLUX 1000 R05 en
OPTIFLUX 1000
1.1 Cost efficient and reliable flow sensor
PRODUCT FEATURES
1
The flangeless OPTIFLUX 1000 design is robust with the highly resistant, reinforced PFA liner and Hastelloy
OPTIFLUX 1000 electromagnetic flow sensor is compact and lightweight. The
OPTIFLUX 1000OPTIFLUX 1000
®
electrodes. This
offers an excellent chemical resistance.
The device is a cost-efficient and reliable solution for a wide range of applications. For industries varying from water and wastewater, agriculture, utilities and from fire-fighting to machine building.
1 Sandwich design 2 PFA liner
3 Hastelloy
®
electrodes
www.krohne.com07/2017 - 4000690405 - TD OPTIFLUX 1000 R05 en
3
1
PRODUCT FEATURES
Highlights
Sandwich (wafer) design
Lightweight and compact for easy handling and space saving installation
Affordable price
Excellent chemical resistance
Bi-directional measurements
No pressure loss
Insensitive to vibrations
No internal moving parts, no maintenance
Industries
Machine building
Energy, HVAC
Water & wastewater
Agriculture
Process industries
OPTIFLUX 1000
Applications
Mixing, batching and dosing systems, filtration systems, pump control
Water flow monitoring
Water circulation and treatment systems
Fire-fighting systems, foam mixing, control of sprinkler systems
Heat transfer and cooling systems
Water including; raw water, process water, wastewater, salt water, heated and cooled water
Mud, slurry, sludge, manure
4
www.krohne.com 07/2017 - 4000690405 - TD OPTIFLUX 1000 R05 en
OPTIFLUX 1000
1.2 Options
PRODUCT FEATURES
The OPTIFLUX 1000
OPTIFLUX 1000 flow sensor is available in a
OPTIFLUX 1000OPTIFLUX 1000 diameter range from DN10 up to DN150 (3/8...6"). The compact flangeless flow sensor meets all applicable process connections: EN 1092, DIN, ANSI and JIS.
Signal converters
Signal converters
Signal convertersSignal converters The OPTIFLUX 1000
OPTIFLUX 1000 flow sensor is compatible with
OPTIFLUX 1000OPTIFLUX 1000
the IFC 050, IFC 100 and IFC 300 signal converter.
1
The flangeless flowmeter is suitable for compact and remote (field) mounting.
www.krohne.com07/2017 - 4000690405 - TD OPTIFLUX 1000 R05 en
5
1
PRODUCT FEATURES
1.3 Measuring principle
An electrically conductive fluid flows inside an electrically insulated pipe through a magnetic field. This magnetic field is generated by a current, flowing through a pair of field coils. Inside of the fluid, a voltage U is generated: U = v * k * B * D
U = v * k * B * D
U = v * k * B * DU = v * k * B * D
in which: v = mean flow velocity k = factor correcting for geometry B = magnetic field strength D = inner diameter of flowmeter
The signal voltage U is picked off by electrodes and is proportional to the mean flow velocity v and thus the flow rate Q. A signal converter is used to amplify the signal voltage, filter it and convert it into signals for totalizing, recording and output processing.
OPTIFLUX 1000
Figure 1-1: Measuring principle
1 Field coils 2 Magnetic field 3 Electrodes 4 Induced voltage (proportional to flow velocity)
6
www.krohne.com 07/2017 - 4000690405 - TD OPTIFLUX 1000 R05 en
OPTIFLUX 1000
2.1 Technical data
The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local sales office.
Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Downloadcenter).
Measuring system
Measuring principle Faraday's law of induction
Application range Electrically conductive fluids
Measured value
Measured value
Measured valueMeasured value
Primary measured value Flow velocity
Secondary measured value Volume flow
Design
Features Sandwich design
TECHNICAL DATA
2
PFA liner and Hastelloy® electrodes
Light weight and compact
Modular construction The measurement system consists of a flow sensor and a signal converter.
Compact version With signal converter IFC 050: OPTIFLUX 1050 C
Remote version In wall (W) mount version with signal converter IFC 050: OPTIFLUX 1050 W
Nominal diameter DN10...150 / 3/8...6"
It is available as compact and as separate version. Additional information can be found in the documentation of the signal converter.
With signal converter IFC 100: OPTIFLUX 1100 C
With signal converter IFC 300: OPTIFLUX 1300 C
In wall (W) mount version with signal converter IFC 100: OPTIFLUX 1100 W
In field (F), wall (W) or rack ( R) mount version with signal converter IFC 300: OPTIFLUX 1300 F, W or R
Measuring accuracy
Maximum measuring error IFC 050: down to 0.5% of the measured valuesPPV
Repeatability ±0.1% of MV, minimum 1 mm/s
Calibration 2 point calibration by direct volume comparison
IFC 100: down to 0.4% of the measured value ± 1 mm/s
IFC 300: down to 0.3% of the measured value ± 2 mm/s
The maximum measuring error depends on the installation conditions.
For detailed information refer to
Optional:
Optional: special calibration on request.
Optional:Optional:
Measuring accuracy
on page 11.
www.krohne.com07/2017 - 4000690405 - TD OPTIFLUX 1000 R05 en
7
2
TECHNICAL DATA
Operating conditions
Temperature
Temperature
TemperatureTemperature
Process temperature -25...+120°C / -13...+248°F
Ambient temperature -25…+65°C / -13+149°F
Protect electronics against self-heating at ambient temperatures above +55°C / +131°F
Storage temperature -50+70°C / -58+158°F
Measuring range
Measuring range -12...+12 m/s / -40...+40 ft/s
Measuring rangeMeasuring range
Pressure
Pressure
PressurePressure
Ambient pressure Atmospheric
Operating pressure Up to 16 bar / 230 psi
Vacuum load 0 mbar / psi absolute
Pressure loss Negligible
Pressure ranges for secondary containment
Chemical properties
Chemical properties
Chemical propertiesChemical properties
Physical condition Electrically conductive liquids
Electrical conductivity Standard: t 5 PS/cm
Permissible gas content (volume)
Permissible solid content (volume)
OPTIFLUX 1000
Pressure resistant up to 40 bar / 580 psi
Burst pressure up to approx. 160 bar / 2320 psi
Demineralised water: t 20 PS/cm
IFC 050: d 3%
IFC 100: d 3%
IFC 300: d 5%
IFC 050: d 10%
IFC 100: d 10%
IFC 300: d 70%
Installation conditions
Installation Assure that the flow sensor is always fully filled.
For detailed information refer to
Flow direction Forward and reverse
Arrow on flow sensor indicates positive flow direction.
Inlet run t 5 DN
Outlet run t 2 DN
Dimensions and weights For detailed information refer to
Installation
Dimensions and weights
on page 15.
on page 12.
8
www.krohne.com 07/2017 - 4000690405 - TD OPTIFLUX 1000 R05 en