Kern FKB 15K0.5A, FKB 15K1A, FKB 30K1A, FKB 30K2A, FKB 60K2A User guide

...
0 (0)

KERN & Sohn GmbH

Ziegelei 1

Tel: +49-[0]7433- 9933-0

D-72336 Balingen

Fax: +49-[0]7433-9933-149

E-Mail: info@kern-sohn.com

Internet: www.kern-sohn.com

Operating instruction

Compact balance

KERN FKB_A

Version 1.1 07/2007

GB

FKB_A-BA-e-0711

GB

KERN FKB_A

Version 1.1 07/2007

 

Operating instruction

 

Compact balance

 

 

 

 

Table of Contents

1Technical data __________________________________________________ 4

2Declaration of conformity ________________________________________ 6

3Basic Information (General)_______________________________________ 7

3.1Proper use _________________________________________________ 7

3.2Improper Use _______________________________________________ 7

3.3Warranty ___________________________________________________ 7

3.4Monitoring of Test Resources _________________________________ 8

4Basic Safety Precautions_________________________________________ 8

4.1Pay attention to the instructions in the Operation Manual __________ 8

4.2Personnel training ___________________________________________ 8

5Transport and storage ___________________________________________ 8

5.1Testing upon acceptance _____________________________________ 8

5.2Packaging__________________________________________________ 8

6Unpacking, Setup and Commissioning _____________________________ 9

6.1Installation Site, Location of Use _______________________________ 9

6.2Unpacking _________________________________________________ 9

6.2.1Setup ____________________________________________________ 9

6.2.2Scope of delivery __________________________________________ 10

6.2.3Basic structure ____________________________________________ 10

6.3Mains connection __________________________________________ 10

6.4Operation using a (rechargeable) battery (optional) ______________ 10

6.5Connection of peripheral devices _____________________________ 11

6.6Initial Commissioning _______________________________________ 11

6.7Adjustment ________________________________________________ 11

6.8Adjustment ________________________________________________ 12

6.9Underfloor weighing ________________________________________ 13

7Operation _____________________________________________________ 14

7.1Overview of display_________________________________________ 14

7.2Weighing__________________________________________________ 14

2

FKB_A-BA-e-0711

7.3Taring ____________________________________________________ 15

7.4PRE-Tare function __________________________________________ 15

7.5Plus/minus weighings_______________________________________ 15

7.6Parts counting _____________________________________________ 16

7.7Net-total weighings _________________________________________ 17

7.8Percent weighings__________________________________________ 17

7.9Weighing units (Unit)________________________________________ 18

7.10Display background illumination ______________________________ 20

7.11Animal weighing function____________________________________ 21

7.12Default setting _____________________________________________ 22

8 Settings ______________________________________________________ 23

8.1Menu structure: ____________________________________________ 23

8.2Leave menu structure _______________________________________ 24

8.3Dosing and Zero-tracking ____________________________________ 24

8.4Selection of the adjustment weight ____________________________ 25

8.5Interface RS232C ___________________________________________ 26

8.5.1Data transfer mode ________________________________________ 26

8.5.2Baud rate ________________________________________________ 26

8.6Selection printed edition_____________________________________ 27

9 Data output RS 232 C ___________________________________________ 28

9.1Technical data _____________________________________________ 28

9.2Pin allocation of the balance output socket (front view), __________ 28

9.3Explanation of the data transfer_______________________________ 28

9.3.1Pr PC ___________________________________________________ 28

9.3.2AU Pr___________________________________________________ 28

9.3.3AU PC __________________________________________________ 29

9.3.4rE Cr ___________________________________________________ 29

9.4Output on bar code printer ___________________________________ 30

10 Service, maintenance, disposal_________________________________ 31

10.1Cleaning __________________________________________________ 31

10.2Service, maintenance _______________________________________ 31

10.3Disposal __________________________________________________ 31

11 Instant help _________________________________________________ 32

FKB_A-BA-e-0711

3

1 Technical data

KERN

FKB 15K0.5A

 

FKB 15K1A

 

FKB 30K1A

 

 

 

 

 

 

Readability (d)

0.5 g

 

1 g

 

1 g

 

 

 

 

 

 

Weighing range (max) (d)

15 kg

 

15 kg

 

30 kg

 

 

 

 

 

 

Taring range (subtractive)

15 kg

 

15 kg

 

30 kg

 

 

 

 

 

 

Reproducibility

0.5 g

 

1 g

 

1 g

 

 

 

 

 

 

Linearity

1.5 g

 

3 g

 

3 g

 

 

 

 

 

 

Minimum unit weight at

1 g

 

2 g

 

2 g

piece counting

 

 

 

 

 

 

 

 

 

 

 

 

 

Warm-up time

2 hours

 

30 minutes

 

2 hours

 

 

 

 

 

 

Reference quantities at

 

5, 10, 25, 50

 

 

piece counting

 

 

 

 

 

 

 

 

 

 

 

 

Weighing Units

Details „Weighing units“ chapter 7.9

 

 

 

 

 

 

Recommended adjustment weight,

 

 

 

 

 

not added (class)

15 kg

 

15 kg

 

30 kg

 

 

 

Details for „Selection of the

(F2)

 

( M1 )

 

(F2)

Adjustment weight“ in chapter

 

 

 

 

 

 

 

8.4

 

 

 

 

 

 

 

 

 

 

 

Stabilization time (typical)

 

 

3 sec.

 

 

 

 

 

 

 

Operating temperature

 

+5°C….. + 35° C

 

 

 

 

Humidity of air

max. 80 % (not condensing)

 

 

 

 

 

Housing (B x D x H) mm

 

350 x 390 x 120

 

 

 

 

 

 

 

 

Weighing plate mm

 

 

340 x 240

 

 

 

 

 

 

 

Weight kg (net)

 

6.5

 

 

 

 

 

 

 

 

4

FKB_A-BA-e-0711

KERN

FKB 30K2A

FKB 60K2A

 

FKB 60K5A

 

 

 

 

 

Readability (d)

2 g

2 g

 

5 g

 

 

 

 

 

Weighing range (max) (d)

30 kg

60 kg

 

60 kg

 

 

 

 

 

Taring range (subtractive)

30 kg

60 kg

 

60 kg

 

 

 

 

 

Reproducibility

2 g

2 g

 

5 g

 

 

 

 

 

Linearity

6 g

6 g

 

15 g

 

 

 

 

 

Minimum unit weight at piece

4 g

4 g

 

10 g

counting

 

 

 

 

 

 

 

 

 

 

Warm-up time

30 minutes

2 hours

 

30 minutes

 

 

 

 

 

Reference quantities at

 

5, 10, 25, 50

 

 

piece counting

 

 

 

 

 

 

 

 

 

Weighing Units

Details „Weighing units“ chapter 7.9

 

 

 

 

 

Recommended adjustment weight,

 

 

 

 

not added (class)

30 kg

60 kg

 

60 kg

 

 

Details for „Selection of the

( M1 )

(F2)

 

( M1 )

Adjustment weight“ in chapter

 

 

 

 

 

8.4

 

 

 

 

 

 

 

 

 

Stabilization time (typical)

 

3 sec.

 

 

 

 

 

Operating temperature

 

+5°C... + 35° C

 

 

 

Humidity of air

max. 80 % (not condensing)

 

 

 

 

Housing (B x D x H) mm

 

350 x 390 x 120

 

 

 

 

 

Weighing plate mm

 

340 x 240

 

 

 

 

 

 

Weight kg (net)

 

6.5

 

 

 

 

 

 

 

FKB_A-BA-e-0711

5

Kern FKB 15K0.5A, FKB 15K1A, FKB 30K1A, FKB 30K2A, FKB 60K2A User guide

2 Declaration of conformity

Declaration of conformity

The electronic balances

Type :

KERN FKB 15K0.5A

 

KERN FKB 15K1A

 

KERN FKB 30K1A

 

KERN FKB 30K2A

 

KERN FKB 60K2A

 

KERN FKB 60K5A

 

 

fulfill the requirements of the following CE guideline:

CE-EMC guideline (directive

about electromagnetic in the version 89/336/EWG compatibility)

Applied harmonized standards, in particular

EN 55022 : 1998

EN 61000-3-2: 1995/A1: 1998/A2 : 1998/A14 :2000

EN 61000-3-3 : 1995

EN 55024 :1998

Any modifications of the appliances described above that are carried out without prior permission by KERN will render this declaration invalid.

Date: 19.12. 2006

Signature:

Gottl. KERN & Sohn GmbH

Management

Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72322 Balingen-Frommern, Tel. +49-07433/9933-0, Fax +49-07433/9933-149

6

FKB_A-BA-e-0711

3 Basic Information (General)

3.1Proper use

The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.

3.2Improper Use

Do not use balance for dynamic weighings. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (Example: Slowly draining fluids from a container on the balance).

Do not leave permanent load on the weighing plate. This may damage the measuring system.

Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this.

Never operate balance in explosive environment. The serial version is not explosion protected.

The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance.

The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.

3.3Warranty

Warranty claims shall be voided in case

Our conditions in the operation manual are ignored

The appliance is used outside the described uses

The appliance is modified or opened

mechanical damage and damage caused by media, liquids, natural wear and tear

The appliance is improperly set up or incorrectly electrically connected

The measuring system is overloaded

FKB_A-BA-e-0711

7

3.4Monitoring of Test Resources

In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.

4 Basic Safety Precautions

4.1Pay attention to the instructions in the Operation Manual

Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.

4.2Personnel training

The appliance may only be operated and maintained by trained personnel.

5 Transport and storage

5.1Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.

5.2Packaging

Keep all parts of the original packaging in case you need to return the appliance. Only use original packaging for returning.

Before sending, disconnect all connected cables and loose/movable parts.

Attach possibly existing transport safeguards. Secure all parts, e.g. weighing plate, power unit etc., to prevent slipping and damage.

8

FKB_A-BA-e-0711

6 Unpacking, Setup and Commissioning

6.1Installation Site, Location of Use

The balances are designed in a way that reliable weighing results are achieved in common conditions of use.

You will work accurately and fast, if you select the right location for your balance.

Therefore, observe the following for the installation site:

Place the balance on a firm, level surface;

Avoid extreme heat as well as temperature fluctuation caused by installing next to a radiator or in the direct sunlight;

Protect the balance against direct draughts due to open windows and doors;

Avoid jarring during weighing;

Protect the balance against high humidity, vapours and dust;

Do not expose the device to extreme dampness for longer periods of time. Nonpermitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.

Avoid static charging of the material to be weighed, weighing container and windshield.

Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference.

6.2Unpacking

Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.

6.2.1 Setup

The balance must be installed in a way that the weighing plate is exactly in horizontal position.

FKB_A-BA-e-0711

9

6.2.2 Scope of delivery

Serial accessories:

Balance

Weighing plate

2 x transport security

Mains power supply

Protective cover

Operating Manual

6.2.3 Basic structure

Place the balance on a horizontal and solid base.

Remove the transport security on the 4-point support.

Pull off the protection foil from the weighing plate if existing.

Attach the weighing plate.

6.3Mains connection

Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage.

Only use original KERN mains adapters. Using other makes requires consent by

KERN.

6.4Operation using a (rechargeable) battery (optional)

Lift-off the battery cover on the lower side of the balance. Connect 9 V compound battery.

Reinsert the battery cover.

For battery operation the balance has an automatic switch-off function which can be activated or deactivated in the menu (chapter 8.1). Proceed as follows:

Switch-on the balance using the key and wait for the „0“ display.

Press the key and keep it pressed until „UNIT“ appears on the display.

Press the key four times, in the display appears „AF“.

Confirm by pressing the key.

Use the key to choose between the two following settings:

1)

„AF

on“:

To save the battery, the balance switches off automatically

 

 

 

3 minutes after having finished the weighing procedure.

2)

„AF

off“:

Switch-off function deactivated.

Use the key to confirm your selected setting.

10

FKB_A-BA-e-0711

Loading...