KERN DE15K5N, DE6K2N, DE24K2N, DE12K1N, DE30K1N Operating Manual

...
0 (0)
DE-BA-defsi-0640
3
7
GB
Version 4.0 12/2006
Operating Manual Electronic Platform Balance
Table of Contents
1 Technical data ................................................................................................. 39
2 Declaration of conformity............................................................................... 46
3 Basic Information (General) ........................................................................... 47
3.1 Proper use ................................................................................................ 47
3.2 Improper Use............................................................................................ 47
3.3 Warranty.................................................................................................... 47
3.4 Monitoring of Test Resources................................................................. 48
4 Basic Safety Precautions................................................................................ 48
4.1 Pay attention to the instructions in the Operation Manual................... 48
4.2 Personnel training.................................................................................... 48
5 Transport and storage .................................................................................... 48
5.1 Testing upon acceptance ........................................................................ 48
5.2 Packaging ................................................................................................. 48
6 Unpacking, Setup and Commissioning......................................................... 49
6.1 Installation Site, Location of Use............................................................ 49
6.2 Unpacking................................................................................................. 49
6.2.1 Setup...................................................................................................... 49
6.2.2 Scope of delivery.................................................................................... 49
6.2.3 Grundaufbau........................................................................................... 50
6.2.3.1 Assembly instructions for the use of the tripod (option)................... 50
6.3 Mains connection..................................................................................... 51
6.4 Operation using a (rechargeable) battery (optional)............................. 51
6.5 Connection of peripheral devices........................................................... 51
6.6 Initial Commissioning.............................................................................. 52
6.7 Adjustment ............................................................................................... 52
6.8 Adjustment ............................................................................................... 52
7 Operation ......................................................................................................... 53
7.1 Overview of display.................................................................................. 53
7.2 Weighing................................................................................................... 53
DE-BA-defsi-0640
38
7.3 Taring........................................................................................................ 54
7.4 PRE-Tare function.................................................................................... 54
7.5 Plus/minus weighings.............................................................................. 54
7.6 Parts counting.......................................................................................... 55
7.7 Net-total weighings.................................................................................. 55
7.8 Percent weighings.................................................................................... 56
7.9 Weighing units (Unit)............................................................................... 57
7.10 Display background illumination............................................................ 59
7.11 Animal weighing function........................................................................ 60
8 Settings............................................................................................................ 61
8.1 Call-up menu structure............................................................................ 61
8.2 Leave menu structure.............................................................................. 62
8.3 Dosing and Zero-tracking........................................................................ 62
8.4 Selection of the adjustment weight........................................................ 63
8.5 Interface RS232C...................................................................................... 64
8.5.1 Data transfer mode................................................................................. 64
8.5.2 Baud rate................................................................................................ 64
8.6 Selection printed edition ......................................................................... 65
8.7 Reset to factory setting ........................................................................... 66
9 Data output RS 232 C...................................................................................... 67
9.1 Technical data .......................................................................................... 67
9.2 Pin allocation of the balance output socket (front view)...................... 67
9.3 Explanation of the data transfer.............................................................. 67
9.3.1 Pr PC...................................................................................................... 67
9.3.2 AU Pr..................................................................................................... 67
9.3.3 AU PC.................................................................................................... 68
9.3.4 rE Cr...................................................................................................... 68
9.4 Output on bar code printer...................................................................... 69
10 Service, maintenance, disposal.................................................................. 70
10.1 Cleaning.................................................................................................... 70
10.2 Service, maintenance............................................................................... 70
10.3 Disposal .................................................................................................... 70
11 Instant help................................................................................................... 71
DE-BA-defsi-0640
39
1 Technical data
KERN DE3K1N DE6K0.5N DE6K2N DE12K1N
Readability (d) 1 g 0.5 g 2 g 1 g Weighing range (max) (d) 3 kg 6 kg 6 kg 12 kg Taring range (subtractive) 3 kg 6 kg 6 kg 12 kg Minimum piece weight 2 g 1 g 4 g 2 g Reproducibility 1 g 0.5 g 2 g 1 g Linearity ± 3 g ± 1.5 g ± 6 g ± 3 g Warm-up time 10 minutes 30 minutes 10 minutes 30 minutes Reference unit weights at
piece count
5, 10, 25, 50
Weighing Units
Details „Weighing units“ chapter 7.9
Recommended adjustment weight, not added (class)
Details for „Selection of the Adjustment weight“ in chapter 8.4
3 kg
( M2 )
6 kg
( M1 )
6 kg
( M2 )
12 kg
( M1 )
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing) Terminal (B x D x H) mm 200 x 100 x 55 Platform (B x D x H ) mm 315 x 305 x 75 Weight kg (net) 4 4 4 4
DE-BA-defsi-0640
40
KERN DE15K0.5N DE15K5N DE24K2N DE30K1N
Readability (d) 0.5 g 5 g 2 g 1 g Weighing range (max) 15 kg 15 kg 24 kg 30.5 kg Taring range (subtractive) 15 kg 15 kg 24 kg 30.5 kg Minimum piece weight 1 g 10 g 4 g 2 g Reproducibility 0.5 g 5 g 2 g 1 g Linearity ± 1.5 g ± 15 g ± 6 g ± 3 g Warm-up time 2 hours 10 minutes 30 minutes 2 hours Reference unit weights at
piece count
5, 10, 25, 50
Weighing Units
Details „Weighing units“ chapter 7.9
Recommended adjustment weight, not added (class)
Details for „Selection of the Adjustment weight“ in chapter 8.4
15 kg
(F2)
15 kg
( M2 )
20 kg
( M1 )
30 kg
(F2)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing) Terminal (B x D x H) mm 200 x 100 x 55 Platform (B x D x H ) mm
315 x 305 x 82 315 x 305 x 75 315 x 305 x 75 315 x 305 x 82
Weight kg (net) 5 4 4 5
DE-BA-defsi-0640
41
KERN DE36K10N DE36K10NL DE60K2N DE60K5N
Readability (d) 10 g 10 g 2 g 5 g Weighing range (max) 36 kg 36 kg 60 kg 60 kg Taring range (subtractive) 36 kg 36 kg 60 kg 60 kg Minimum piece weight 20 g 20 g 4 g 10 g Reproducibility 10 g 10 g 2 g 5 g Linearity ± 30 g ± 30 g ± 6 g ± 15 g Warm-up time 10 minutes 10 minutes 2 hours 30 minutes Reference unit weights at
piece count
5, 10, 25, 50
Weighing Units
Details „Weighing units“ chapter 7.9
Recommended adjustment weight, not added (class)
Details for „Selection of the Adjustment weight“ in chapter 8.4
30 kg
( M2 )
30 kg
( M2 )
60 kg
(F2)
60 kg
( M1 )
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing) Terminal (B x D x H) mm 200 x 100 x 55 Platform (B x D x H )mm
315 x 305 x 75 522 x 403 x
100
315 x 305 x 82 315 x 305 x 75
Weight kg (net) 5 16 5 5
DE-BA-defsi-0640
42
KERN DE60K5NL DE60K20N DE60K20NL
Readability (d) 5 g 20 g 20 g Weighing range (max) 60 kg 60 kg 60 kg Taring range (subtractive) 60 kg 60 kg 60 kg Minimum piece weight 10 g 40 g 40 g Reproducibility 5 g 20 g 20 g Linearity ± 15 g ± 60 g ± 60 g Warm-up time 30 minutes 10 minutes 10 minutes Reference unit weights at
piece count
5, 10, 25, 50
Weighing Units
Details „Weighing units“ chapter 7.9
Recommended adjustment weight, not added (class)
Details for „Selection of the Adjustment weight“ in chapter 8.4
60 kg
( M1 )
60 kg
( M2 )
60 kg
( M2 )
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing) Terminal (B x D x H) mm 200 x 100 x 55 Platform (B x D x H ) mm
522 x 403 x 100 315 x 305 x 75 522 x 403 x 85
Weight kg (net) 16 5 16
DE-BA-defsi-0640
43
KERN DE120K10N DE120K10NL DE150K5N Readability (d) 10 g 10 g 5 g Weighing range (max) 120 kg 120 kg 150 kg Taring range (subtractive) 120 kg 120 kg 150 kg Minimum piece weight 20 g 20 g 10 g Reproducibility 10 g 10 g 5 g Linearity ± 30 g ± 30 g ± 15 g Warm-up time 30 minutes 30 minutes 2 hours Reference unit weights at
piece count
5, 10, 25, 50
Weighing Units
Details „Weighing units“ chapter 7.9
Recommended adjustment weight, not added (class)
Details for „Selection of
the Adjustment weight“
in chapter 8.4
120 kg
( M1 )
120 kg ( M1 )
150 kg
(F2)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing) Terminal (B x D x H) mm 200 x 100 x 55 Platform (B x D x H ) mm 315 x 305 x 75 522 x 403 x 100 315 x 305 x 82 Weight kg (net) 5 16 5
DE-BA-defsi-0640
44
KERN DE150K50N DE150K50NL DE150K50NXL
Readability (d) 50 g 50 g 50 g Weighing range (max) 150 kg 150 kg 150 kg Taring range (subtractive) 150 kg 150 kg 150 kg Minimum piece weight 100 g 100 g 100 g Reproducibility 50 g 50 g 50 g Linearity ± 150 g ± 150 g ± 150 g Warm-up time 10 minutes 10 minutes 10 minutes Reference unit weights at
piece count
5, 10, 25, 50
Weighing Units
Details „Weighing units“ chapter 7.9
Recommended adjustment weight, not added (class)
Details for „Selection of
the Adjustment weight“
in chapter 8.4
150 kg
( M2 )
150 kg
( M2 )
150 kg
( M2 )
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing) Terminal (B x D x H) mm 200 x 100 x 55 Platform (B x D x H ) mm 315 x 305 x 75 522 x 403 x 100
650 x 500 x 105
Weight kg (net) 5 16 28
DE-BA-defsi-0640
45
KERN DE240K20NL DE300K100N DE300K100NL
Readability (d) 20 g 100 g 100 g Weighing range (max) 240 kg 300 kg 300 kg Taring range (subtractive) 240 kg 300 kg 300 kg Minimum piece weight 40 g 200 g 200 g Reproducibility 20 g 100 g 100 g Linearity ± 60 g ± 300 g ± 300 g Warm-up time 30 minutes 10 minutes 10 minutes Reference unit weights at
piece count
5, 10, 25, 50
Weighing Units
Details „Weighing units“ chapter 7.9
Recommended adjustment weight, not added (class)
Details for „Selection of the Adjustment weight“ in chapter 8.4
200 kg ( M1 )
300 kg
( M2 )
300 kg
( M2 )
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing) Terminal (B x D x H) mm 200 x 100 x 55 Platform (B x D x H ) mm 522 x 403 x 100 522 x 403 x 85 650 x 500 x 105 Weight kg (net) 16 16 28
DE-BA-defsi-0640
46
2 Declaration of conformity
Declaration of conformity
The electronic balances
Type : KERN DE3K1N
KERN DE6K0.5N KERN DE6K2N KERN DE12K1N KERN DE15K0.5N KERN DE15K5N KERN DE24K2N
KERN DE30K1N KERN DE36K10N KERN DE36K10NL KERN DE60K2N KERN DE60K5N KERN DE60K5NL KERN DE60K20N KERN DE60K20NL KERN DE120K10N
KERN DE120K10NL KERN DE150K5N KERN DE150K50N KERN DE150K50NL KERN DE150K50NXL KERN DE240K20NL KERN DE300K100N KERN DE300K100NL
fulfill the requirements of the following CE guideline: CE-EMC guideline (directive about
electromagnetic compatibility)
in the version 89/336/EWG
Applied harmonized standards, in particular: EN 55022 : 1998
EN 61000-3-2: 1995/A1: 1998/A2 : 1998/A14 :2000
EN 61000-3-3 : 1995
EN 55024 :1998
Any modifications of the appliances described above that are carried out without prior
permission by KERN will render this declaration invalid.
Date: 19.12.2006 Signature: Gottl. KERN & Sohn GmbH Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72322 Balingen, Tel. +49-07433/9933-0,Fax +49-074433/9933-149
DE-BA-defsi-0640
47
3 Basic Information (General)
3.1 Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use balance for dynamic weighings. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage and damage caused by media, liquids
Natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
Loading...