INFICON PSG500, PSG500-S, PSG502-S, PSG510-S, PSG512-S Operating Manua

Pirani Standard Gauge
pp
PSG500/-S, PSG502-S, PSG510-S, PSG512-S
Operating Manual
In all communications with INFICON, please specify the in­formation on the product nameplate. For convenient refe­rence copy that information into the space provided below.
INFICON AG, LI-9496 Balzers
Model: PN: SN: V W
Validity
This document applies to products with the following part numbers:
PSG500 PSG500-S ( 350-060 350-080
350-062 350-082 350-061 350-081 350-064 350-084 350-065 350-085 350-063 350-083 350-066 350-086
350-067 350-087 350-068 350-088
PSG502-S (Ni filament) 350-140
350-142 350-141 350-144 350-145 350-143 350-146
350-147 350-148
(DN 16 ISO-KF) (DN 16 CF-R) (1/8" NPT) (8 VCR®) (4 VCR®) (½"-Rohr) (7/16-20 UNF)
(DN 16 ISO-KF long tube) (DN 16 CF-R long tube)
PSG510-S (W filament) PSG512-S (Ni filament) 350-200
The part number (PN) can be taken from the product name­plate.
If not indicated otherwise in the legends, the illustrations in this document correspond to the gauge with part number 350 060. They apply to gauges with other part numbers by analogy.
We reserve the right to make technical changes without prior notice.
All dimensions in mm.
(DN 16 ISO-KF) 350-300 (DN 16 ISO-KF)
Intended Use
The Pirani Standard Gauges PSG500/-S, PSG502-S, PSG510-S, PSG512-S have been designed for vacuum measurement of gases in the pressure range of
-4
5×10
… 1000 mbar.
Incl. EU Declaration of Conformity
tina44e1-i (2015-10)
W filament)
(DN 16 ISO-KF) (DN 16 CF-R) (1/8" NPT) (8 VCR®) (4 VCR®) (½"-Rohr) (7/16-20 UNF)
(DN 16 ISO-KF long tube) (DN 16 CF-R long tube)
They must not be used for measuring flammable or combus­tible gases in mixtures containing oxidants (e.g. atmospheric oxygen) within the explosion range.
They can be operated in connection with an INFICON con­troller or with another controller.
Trademark
VCR® Swagelok Marketing Co.
Safety
Symbols Used
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and envi­ronmental damage.
Caution
Information on correct handling or use. Disregard can lead to malfunctions or minor equipment damage.
Personnel Qualifications
Skilled personnel
All work described in this document may only be carried out by persons who have suitable technical training and the necessary experience or who have been instructed by the end-user of the product.
General Safety Instructions
Adhere to the applicable regulations and take the nec­essary precautions for the process media used.
Consider possible reactions between the materials and the process media.
Consider possible reactions (e.g. explosion) of the process media due to the heat generated by the product.
Adhere to the applicable regulations and take the neces­sary precautions for all work you are going to do and con­sider the safety instructions in this document.
Before beginning to work, find out whether any vacuum components are contaminated. Adhere to the relevant re­gulations and take the necessary precautions when hand­ling contaminated parts.
Communicate the safety instructions to all other users.
Liability and Warranty
INFICON assumes no liability and the warranty becomes null and void if the end-user or third parties
disregard the information in this document
use the product in a non-conforming manner
make any kind of interventions (modifications, alterations
etc.) on the product
use the product with accessories not listed in the product documentation.
The end-user assumes the responsibility in conjunction with the process media used.
Gauge failures due to contamination or wear and tear, as well as expendable parts (e.g. filament), are not covered by the warranty.
Technical Data
Measurement principle thermal conductance according
Measurement range (air, O
, CO, N2)
2
Accuracy (N
)
2
-3
1×10
… 100 mbar ±15% of reading
-4
5×10
… 1×10-3 mbar ±50% of reading
100 … 1000 mbar ±50% of reading Resolution 1% of reading Repeatability
-3
1×10
… 100 mbar 2% of reading
Output signal (measure­ment signal)
Voltage range V (dc) 0 … +10.3
Measurement range V (dc) +1.9 … +10.0 Voltage vs. pressure logarithmic
Error signal
Filament rupture V V
Output impedance
Minimum loaded imped­ance
Response time ms 80
Gauge identification
Adjustment one tactile switch for ATM and
Switching functions SP1, SP2
Threshold value indi-
cation and setting
Setting range 2×10
Hysteresis 10% above lower threshold
Relay contact
closed open
Supply
The gauge may only be connected to power sup­plies, instruments or control devices that conform to the requirements of a grounded extra-low vol­tage (PELV). The connection to the gauge has to be fused
Supply voltage
At gauge V (dc) +14 … +30
Ripple V Current consumption mA <500
Power consumption W ≤1 Fuse required
1)
AT
Electrical connection FCC 68 / RJ45 appliance
Sensor cable 8 poles plus shielding Cable length 100 m (8×0.14 mm
Grounding concept
Vacuum connection to
signal common
Supply common to
signal common
Materials exposed to vacuum
PSG500/-S, PSG502-S
PSG510-S,
PSG512-S
Filament
PSG500/-S, PSG510-S
PSG502-S, PSG512-S
1)
INFICON controllers fulfill these requirements.
to Pirani
-4
… 1000 mbar
5×10
1.286 V/decade 0 … +0.5
+0.1
2×4.7
Ω
10, short-circuit proof
kΩ
27.0 kΩ, referenced to supply common (voltage at pin 4 5 V)
HV adjustment
one tactile switch at measure-
ment value output. Press briefly for threshold indication. Keep pressing or press repeatedly for threshold setting.
-3
… 500 mbar
30 V, 0.5 A (dc), floating
at low pressure (lamp is lit) at high pressure, error, missing supply
DANGER
1)
.
1
(max. starting current)
1
(slow)
connector, 8 poles, male
"Electrical Connection"
connected via 1 MΩ (voltage difference <15 V)
conducted separately, for
differential measurement
DIN 1.4301, DIN 1.4305, DIN 1.4435, glass, Ni, NiFe
Al
(ceramics), Ni,
2O3
DIN 1.4435, DIN 1.4305 DIN 1.3981
W Ni
Internal volume
DN 16 ISO-KF DN 16 CF-R 1/8" NPT
®
8 VCR
®
4 VCR ½"-Rohr 7/16-20 UNF
DN 16 ISO-KF long tube DN 16 CF-R long tube
Admissible pressure bar
Admissible temperatures
Operation °C Vacuum connection
DN 16 ISO-KF DN 16 CF-R 1/8" NPT
®
8 VCR
®
4 VCR ½"-Rohr 7/16-20 UNF
°C °C °C °C °C °C °C
3
cm
3
cm
3
cm
3
cm
3
cm
3
cm
3
cm
3
cm
3
cm
(abs.)
+5 … +60
2)
80
2)
80 80 80 80 80 80
1.5 1.5 2 2 2 2 1.5
10 10
10, limited to inert gases
in horizontal mounting ori­entation
Filament °C 110 Storage °C –20 … +65
Relative humidity % 80 at temperatures up to
Use indoors only, altitude up to
+31 °C, decreasing to 50 at +40 °C
2000 m NN Mounting orientation any Protection category IP40
Dimensions mm
65
50
20.9
Measurement Signal vs. Pressure
Pressure p
1E+6
1E+5
1E+4
1E+3
1E+2
1E+1
1E+0
1E–1
1E–2
1E–3
1E–4
valid in the range 5×10
U p c
[V] [mbar] 6.143 [V] [µbar] 2.287 [V] [Torr] 6.304 [V] [mTorr] 2.448
where p pressure
[mbar]
Underrange
Sensor error
0.0 0.5 1.0 1.5 2.5 3.5 4.5 5.5 6.5 7.5 8.5 9.5
p = 10
2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0
Measurement signal U[V]
((U-c)/1.286)
U measurement signal c constant (depending on pressure unit)
3.75×10 5×10
U = c + 1.286 × log10 p
-4
mbar <p< 1000 mbar
-4
Torr <p< 750 Torr
-2
Pa <p< 1×105 Pa
U p c
[V] [micron] 2.448 [V] [Pa] 3.572 [V] [kPa] 7.429
Pa
mbar
Torr
Gas Type Dependence
Pressure reading (gauge adjusted for air)
p (mbar)
2
10
8 6
4
2
1
10
8 6
4
2
0
10
8 6
4
2
–1
10
8 6
4
2
Water vapor
–2
10
8 6
4
2
–3
10
–3
10
24
–2
10
6
24
–1
10
6
24
Calibration factors for the pressure range below 1 mbar
115
p
= C × pressure reading
eff
)
15*
DN 16 ISO-KF *) PSG51x-S: 20 mm
31.5
8 VCR
32.8
14
DN 16 CF-R
41.5
4 VCR
29
1/8" NPT
24
7/16-20 UNF
70.3
19
tube ½"
116
Gas
2
)
Weight
DN 16 ISO-KF DN 16 CF-R 1/8" NPT
®
8 VCR
®
4 VCR ½"-Rohr 7/16-20 UNF
DN 16 ISO-KF long tube DN 16 CF-R long tube g g
2)
250 °C with long tube.
DN 16 ISO-KF
g
80
g
100
g
70
g
130
g
100
g
70
g
80 130
140
DN 16 CF-R
type
He Ne
Ar Kr
Xe
Calibration
factor C
0.8
1.4
1.7
2.4
3.0
Gas type Calibration
H
0
10
6
24
2
, CO, N2
air, O
2
CO
water vapor
freon 12
H2He Ne
10
6
2
1
24
(mbar)
p
eff
Air O
2
CO N
2
CO
2
Ar
Freon 12
Kr
Xe
10
6
factor C
0.5
1.0
0.9
0.5
0.7
2
Installation
Vacuum Connection
DANGER
DANGER: overpressure in the vacuum system >1 bar
Overrange
10.3
Remove the protective lid and install the product to the vac­uum system.
Original: German tina44d1-i (2015-10)
Injury caused by released parts and harm caused by escaping process gases can result if clamps are opened while the vacuum system is pressurized.
Do not open any clamps while the vacuum sys­tem is pressurized. Use the type of clamps which are suited to overpressure.
DANGER
DANGER: overpressure in the vacuum system >2.5 bar
KF connections with elastomer seals (e.g. O-rings) cannot withstand such pressures. Pro­cess media can thus leak and possibly damage your health.
Use O-rings provided with an outer centering ring.
DANGER
DANGER: protective ground Incorrectly grounded products can be extremely
hazardous in the event of a fault. The gauge must be electrically connected to the
grounded vacuum chamber. This connection must conform to the requirements of a protective connection according to EN 61010:
CF, NPT, VCR
this requirement.
For gauges with a KF connection, use a con-
ductive metallic clamping ring.
If a ½" tube is used, take appropriate meas-
ures for this requirement to be fulfilled.
®
and UNF connections fulfill
Caution
Caution: vacuum component Dirt and damages impair the function of the vac-
uum component. When handling vacuum components, take ap-
propriate measures to ensure cleanliness and prevent damages.
Caution
Caution: dirt sensitive area Touching the product or parts thereof with bare
hands increases the desorption rate. Always wear clean, lint-free gloves and use
clean tools when working in this area.
The gauge may be mounted in any orientation. To
keep condensates and particles from getting into the measuring chamber preferably choose a horizontal to upright position and possibly use a seal with a centering ring and filter. If adjustment should be pos­sible after the gauge has been installed, be sure to install it so that the button can be accessed with a
"Adjusting the Gauge").
pin (
Keep the protective lid.
t i na44e1- i
(2015-10)
Electrical Connection
Make sure the vacuum connection is properly made
"Vacuum Connection").
(
If no sensor cable is available, make one according to
the following diagram.
Gauge with switching functions
<SP1> <SP2>
SP1
Measurement
signal
1M
SP2
Ω
SP2
SP1
4.7
4.7
27k
Ω
Ω
Ω
+
6 8 7 3 5 4 1 2
+ –
Ident
+ –
1
Electrical connection Pin 1 Supply Pin 2 Supply common, GND Pin 3 Measurement signal or thresholds SP1/2 Pin 4 Gauge identification Pin 5 Signal common Pin 6, 8 Relay SP2, closing contact Pin 7, 8 Relay SP1, closing contact
8
8-pole
FCC-68
connector
Connect the sensor cable to the gauge and the control-
ler.
Operation
When the supply voltage is applied, the measurement signal is available between pins 3 and 5 (relationship between measurement signal and pressure
Allow a stabilization period of at least 10 minutes. It is advis­able to operate the gauge continuously, irrespective of the pressure.
Gas Type Dependence
The measurement value is gas dependent. The pressure reading applies to dry air, O has to be corrected (
If the gauge is operated with an INFICON controller, a cali­bration factor for correction of the actual reading can be ap­plied (
of the corresponding controller).
"Technical Data").
Adjusting the Gauge
The gauge is factory calibrated. Due to long time operation or contamination, a zero drift could occur. Periodically check the zero and adjust it if necessary.
For adjusting the zero, operate the gauge under the same ambient conditions and in the same mounting orientation as normally.
The gauge is adjusted to default values. However, it can also be adjusted to other pressure values, if the exact pressure value is known (reference measurement).
"Technical Data").
, CO and N2. For other gases, it
2
If you are using a seal with centering ring and filter,
check that they are clean or replace them if necessary (
"Deinstallation").
Activate the gauge and operate it at atmospheric pres-
sure for at least 10 minutes.
Press the button with a pin (max. ø1.1 mm) and the
ATM adjustment is carried out: The gauge is adjusted to 1000 mbar (10 V (dc)) by default. By pressing the button >5 s the pressure value is increased towards 1200 mbar (or, by pressing it again, decreased towards 500 mbar) until the button is released or the limit is reached.
ATM
HV
max. ø1.1 mm
Evacuate to p << 10
pressure in the range of 10 least 2 minutes.
-4
mbar (recommended) or to a
-4
… 10-2 mbar and wait at
Press the button with a pin and the HV adjustment is
carried out: The gauge is adjusted to 1.2×10 (1.1 V (dc)) by default. By pressing the button >5 s the pressure value is increased toward 1×10 the button is released or the limit is reached.
Switching Functions
(PSG500-S and PSG502-S only)
The setpoints are adjustable within a pressure range of
-3
2×10
… 500 mbar (voltage range of 2.67 … 9.61 V (dc)). Each switching function provides a floating relay contact (
"Electrical Connection").
Lamps
Measurement signal (Pressure p)
Setpoint
Off
Adjusting the Setpoints
On
DANGER
DANGER: malfunction If processes are controlled via the signal output,
keep in mind that by pressing a button <SP> the measururement signal is suppressed and that the corresponding threshold value is output instead. This can cause malfunctions.
Press a button <SP> only if you are sure that no damages can arise from a malfunction.
The status of the relay and lamp is not affected by
pressing the button.
The status of the switching func­tion is indicated by a lamp.
Status Lamp Relay
off
dark
on
lit
e
r
u
s
a
e
M
Hysteresis (10% of threshold value)
Threshold value
Off
Press the button <SP1> with a pin (max. ø1.1 mm):
The gauge changes to the switching function mode and outputs the current lower threshold value at the measurement value output for about 5 s. When the button is kept depressed for more than 5 s, the threshold setting is modified until the button is released or until the limit of the setting range is reached.
max. ø1.1 mm
Keep the button
depressed
W hen the button is pressed again within 5 s the thresh-
old setting is adjusted in the reverse direction.
-4
mbar
-2
mbar until
deenergized
energized
e
u
l
a
v
t
n
e
m
Time t
The upper thresh­old is 10% above the lower one (hysteresis).
Release the button. The gauge resumes operation after
5 s and the connected controller displays the current measurement value.
The adjustment procedure for <SP2> is the same as de­scribed for <SP1>.
Deinstallation
DANGER
DANGER: contaminated parts Contaminated parts can be detrimental to health
and environment. Before beginning to work, find out whether any
parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts.
Caution
Caution: vacuum component Dirt and damages impair the function of the vac-
uum component. When handling vacuum components, take ap-
propriate measures to ensure cleanliness and prevent damages.
Caution
Caution: dirt sensitive area Touching the product or parts thereof with bare
hands increases the desorption rate. Always wear clean, lint-free gloves and use
clean tools when working in this area.
Vent the vacuum system.
Put the gauge out of operation.
Unplug the sensor cable.
Remove the gauge from the vacuum system and install
the protective lid.
Maintenance, Repair
In case of severe contamination or a malfunction, the sensor can be replaced.
Gauge failures due to contamination or wear and
tear, as well as expendable parts (e.g. filament), are not covered by the warranty.
INFICON assumes no liability and the warranty becomes null and void if any repair work is carried out by the end-user or third parties.
Spare Parts
When ordering spare parts, always indicate:
all information on the product nameplate
description and ordering number according to the spare
parts list
Sensor for gauge Ordering number
350-060, 350-080 350-062, 350-082 350-061, 350-081 350-064, 350-084 350-065, 350-085 350-063, 350-083 350-066, 350-086
350-067, 350-087 350-068, 350-088
350-200
350-140 350-142 350-141 350-144 350-145 350-143 350-146
350-147 350-148
350-300
350-920 350-922 350-921 350-924 350-926 350-923 350-925
350-927 350-928
350 930
350-900 350-902 350-901 350-904 350-906 350-903 350-905
350-907 350-908
350-940
Returning the Product
WARNING
WARNING: forwarding contaminated products Contaminated products (e.g. radioactive, toxic,
caustic or microbiological hazard) can be detri­mental to health and environment.
Products returned to INFICON should preferably be free of harmful substances. Adhere to the forwarding regulations of all involved countries and forwarding companies and enclose a duly completed declaration of contamination
*)
Form under www.inficon.com
Products that are not clearly declared as "free of harmful sub­stances" are decontaminated at the expense of the customer.
Products not accompanied by a duly completed declaration of contamination are returned to the sender at his own expense.
*)
.
Disposal
DANGER
DANGER: contaminated parts Contaminated parts can be detrimental to health
and environment. Before beginning to work, find out whether any
parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts.
WARNING
WARNING: substances detrimental to the en­vironment
Products or parts thereof (mechanical and elec­tric components, operating fluids etc.) can be detrimental to the environment.
Dispose of such substances in accordance with the relevant local regulations.
Separating the components
After disassembling the product, separate its components according to the following criteria:
Contaminated components
Contaminated components (radioactive, toxic, caustic, or biological hazard etc.) must be decontaminated in accor­dance with the relevant national regulations, separated according to their materials, and disposed of.
Other components
Such components must be separated according to their materials and recycled.
EU Declaration of Conformity
Products
We, INFICON, hereby declare that the equip­ment mentioned below complies with the pro­visions of the Directive relating to electro­magnetic compatibility 2014/30/EU and the Directive on the restriction of the use of certain hazardous substances in electrical and electronic equipment 2011/65/EU.
Pirani Standard Gauge
PSG500/-S, PSG502-S, PSG510-S, PSG512-S
Standards
Harmonized and international/national standards and specifi­cations:
EN 61000-6-2:2005 (EMC: generic emission standard)
EN 61000-6-3:2007 + A1:2011 (EMC: generic immunity
standard)
EN 61010-1:2010 (Safety requirements for electrical equip-
EN 61326-1:2013 (EMC requirements for electrical equip-
Manufacturer / Signatures
INFICON AG, Alte Landstraße 6, LI-9496 Balzers
19 October 2015 19 October 2015
Dr. Urs Wälchli Managing Director
ment for measurement, control and laboratory use)
ment for measurement, control and laboratory use)
Marco Kern Product Manager
LI–9496 Balzers Liechtenstein Tel +423 / 388 3111 Fax +423 / 388 3700 reachus@inficon.com
www.inficon.com
Loading...