Eppendorf Repeater plus User Manual

4981 900.095
Multipette® plus
Repeater® plus

Operating manual

Copyright© 2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner.
Trademarks
Eppendorf Combitip, Biopur are registered trademarks of Eppendorf AG,
and the Eppendorf logo, Multipette and Repeater,
Hamburg, Germany.
Registered trademarks are not marked in all cases with ™ or manual.
in this
U.S. Patent No.
5,591 408
5,620,660
5,620,661
U.S. Design Patent No. Combitip plus
384, 163
384, 162
387, 426
4981 900.095-08/012013
2
Multipette® plus · Repeater® plus — Operating manual
Operating manual — Multipette® plus · Repeater® plusseeon p.Fig.Tab.p.
Operating manual

Table of contents

1Table of contents
1 User instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1 Using this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.1 Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
3.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
4.1 Choose Combitip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4.2 Insert Combitip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.3 Select volume. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.3.1 Filling Combitip and reverse stroke . . . . . . . . . . . . . . . . . . . . 8
4.3.2 Discard first dispensing step (reverse stroke). . . . . . . . . . . . .9
4.4 Dispense . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
4.5 Release Combitip. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.6 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
4.7 Empty Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.1 General errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
6.1 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
6.2 Change battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
6.3 Decontamination before shipment . . . . . . . . . . . . . . . . . . . . 12
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
8 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
9 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
10 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
10.1 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16

1 User instructions

1User instructions

1.1 Using this manual

Please read this operating manual completely before using the
device for the first time.
Please view this operating manual as part of the product and keep
it somewhere easily accessible.
When passing the device on to third parties, be sure to include this
operating manual.

2 Product description

2Product description

2.1 Features

In combination with Eppendorf Combitips advanced, the Multipette plus/the Repeater plus forms a dispensing system for repeat dispensing operations.
3
Multipette® plus · Repeater® plus — Operating manual
Fig. 1: Multipette plus/Repeater plus
1 Volume selection dial 2 Position indicator
3 Battery compartment 4 Digital display
5 Dispensing lever 6 Filling lever
7 Ejection keys 8 Serial number

2.2 Combitips advanced

A supply of liquid is aspirated into the Combitip advanced and then dispensed step by step.
The built-in electronics automatically recognize
the Combitip advanced inserted and
the position of the volume selection dial,
and the current dispensing volume appears in the display. Nine Combitips advanced with a variety of filling volumes are
available for single use (see Choose Combitip on p. 7). The Combitips advanced 25 mL and 50 mL can only be used in combination with an adapter. The adapter is a wear part.
For each Combitip advanced, you can use the volume selection dial to set 20 different volumes.
The smallest volume can be dispensed up to a max. of 100 times and the largest volume a max. of 5 times.
^
1=100µL
1
2
US-PAT.No. 5.620,660
34
5 mL
5
67
Fig. 2: Combitip advanced
1 Dispensing volume with the
manual dispenser at selection dial position 1
2 Graduation
3 US patent numbers 4 Maximum volume
5 Color code 6 Coding
7 Piston

2.3 Combitips plus

Multipette plus/Repeater plus can also be used with Combitips plus (predecessor model).
4
Multipette® plus · Repeater® plus — Operating manual

3Safety

3Safety

3.1 Intended use

The Multipette plus/Repeater plus is a lab device and, in combination with a Combitip plus/Combitip advanced, it is intended for dispensing aqueous solutions in the volume range between 1 μL and 10.00 mL. In vivo applications (applications in or on the human body) are not permitted.
The Multipette plus/Repeater plus may only be operated by trained specialist staff. All users must have read the operating manual carefully and familiarized themselves with the device's mode of operation.
CAUTION! Danger to persons from careless use.
Never point the opening of a Multipette plus / Repeater
CAUTION!
CAUTION!
NOTICE!
plus equipped with a Combitip at yourself or anyone else.
Only perform dispensing operations when you do not
endanger yourself or anyone else.
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may impair the safety, func­tion and precision of the device. Eppendorf cannot be held liable or accept any liability for damage resulting from the use of incorrect or non-recommended accessories and spare parts or from the improper use of such equipment.
Only use accessories and original spare parts
recommended by Eppendorf.
NOTICE! Damage to device from opening the housing.
Do not remove the housing of the Multipette plus/
Repeater plus.
Do not allow the housing to be opened by anyone
except service staff authorized by Eppendorf.
NOTICE! Damage to sensor from damaged or incorrectly inserted Adapter plus.
When using 25 mL and 50 mL Combitips, make sure
NOTICE!
that the adapter is in a perfect condition. The adapter is a wear part.
Do not use the Adapter if it is damaged or visibly worn. Connect the Adapter and the Combitip always outside
of the Multipette plus/Repeater plus.
Do not turn the inserted Combitip. Always hold the Multipette plus/Repeater plus by the
housing.
NOTICE! Damage to device from penetration of liquids.
Do not allow any liquids to penetrate the inside of the
NOTICE!
housing.
If liquid has penetrated the inside of the housing, have
the inner parts of the Multipette plus / Repeater plus repaired only by service partners of Eppendorf AG. Contact your responsible sales office before returning a product.
5
Multipette® plus · Repeater® plus — Operating manual
NOTICE! Material damage from incorrect use.
Only use the product for its intended purpose as
NOTICE!
described in the operating manual.
Ensure adequate material resistance when using
chemical substances.
Do not use the Multipette plus/Repeater plus with
liquids which attack the materials ABS, ASA, PBT, PC, POM and PPS.
Do not use the Combitip with liquids which attack the
materials PP, PE or LCP.
In case of doubt, contact your local Eppendorf partner.