HPLC adapter — Instructions for use
This document supplements the instructions for use of the Centrifuge, but does not replace them.
Therefore, also read the instructions for use. The current version can be found online at
www.eppendorf.com
Instructions for use
.
Product description
The HPLC adapters are designed for centrifugation of HPLC vials with max. 1.5 mL filling volume
in the F-45-18-17-Cryo rotor for Centrifuge 5430/5430 R and Centrifuge 5702/5702 R/5702 RH.
Safety notes
Prior to using HPLC vials, observe manufacturer specifications on resistance to
centrifugation (max. g-force).
Do not exceed the maximum tube length of 33 mm (without cap) so the tubes can be
optimally supported by the adapters.
Do not exceed a g-force of 3,500 x g at a max. load of 9.5 g per rotor bore (adapter + tube +
contents). At a load of up to 8.7 g per rotor bore, the adapters can be centrifuged at max.
g-force (rcf)/rotational speed (rpm).
Load the rotor symmetrically. When doing so, follow the loading instructions included in the
instructions for use of the centrifuge.
Cleaning and Disinfection
Use a mild cleaning agent for cleaning.
Use alcohol (ethanol, isopropanol) or disinfectants containing alcohol for the disinfection.
The HPLC adapter can be autoclaved (121°C, 20 min).
Hint!
Technical data
For the Centrifuge 5430/5430 R, ensure that the g-force displayed during switch-over is
standardized to the corresponding rotor without adapter and, for the Centrifuge 5702/5702 R/
5702 RH, standardized to the A-4-38 rotor. You can achieve the following maximum g-forces (rcf)
at maximum speed (rpm) when adapters are used:
Rotor, centrifuge Adapters Max. centrifugation
F-45-18-17-Cryo,
Centrifuge 5430/5430 R
F-45-18-17-Cryo,
Centrifuge 5702/5702 R/5702 RH
without adapter
HPLC adapter
without adapter
HPLC adapter
radius (cm)
9.4
8.3
9.4
8,3
Max. g-force (rcf)
/ speed (rpm)
8,320 / 8,900
7,350 / 8,900
2,030 / 4,400
1,800 / 4,400
1
HPLC adapter — Instructions for use
Max. allowed tube lengths without cap 33 mm
Adapter bore (Ø) 11.8 mm
Material Polypropylene
You can calculate the g-force (rcf) for a special adapter using the following formula according to
DIN 58 970:
rcf = 1.118 * 10
Example: The max. centrifugation radius of the HPCL adapter is 8.3 cm in the F-45-18-17-Cryo
rotor. A max. g-force of 3,500 x g is reached at 6142 rpm.
Ordering information
-5
* n2 * r
, n: rotational speed (rpm), r
max
: max. centrifugation radius (cm)
max
Order No.
(International)
Order No.
(North America)
Description
HPLC adapter
5427 708.006 5427708006
18 pcs.
5427 708.960-00/0210
Your local distributor: www.eppendorf.com/worldwide
Eppendorf AG · 22331 Hamburg · Germany · Tel: +49 40 538 01-0 · Fax: +49 40 538 01-556
Eppendorf North America, Inc. · 102 Motor Parkway, Suite 410 · Hauppauge, NY 11788-5178 · USA
Tel: +1 516 334 7500 · Toll free phone: +1 800 645 3050 · Fax: +1 516 334 7506 · E-Mail: info@eppendorf.com
Europe, International: Tel: +49 1803 666 789 · E-Mail: support@eppendorf.com
North America: Tel: +1 800 645 3050 ext. 2258 · E-Mail: support_na@eppendorf.com
Asia Pacific: Tel: +60 3 8023 6869 · E-Mail: support_asiapacific@eppendorf.com
E-Mail: eppendorf@eppendorf.com
Application Support
2