6 Technical data ............................................................................................................. 88
7 Ordering information .................................................................................................. 89
7aOrdering information for North America ................................................................... 92
51
Table of contents
8 Index ............................................................................................................................. 95
Appendix A
Operation via a computer ........................................................................................... 97
A.1 Description of interface ................................................................................................. 97
A.2 Commands and queries ............................................................................................... 97
Table of contents
52
1 Safety precautions and application
•
Observe operating instructions!
•
The device is only designed for use in the research lab!
•
The FemtoJet express is a lab device. It may only be operated by appropriately qualified lab
personnel.
•
Only use the device for microinjection under observance of all relevant legal regulations!
•
Before plugging in the device, please compare your voltage requirements with the
specifications on the identification plate!
•
The unit may not be connected to a socket that is not grounded! Use the appropriate mains
cable.
•
Electrical connections may be effected only when the device is switched off!
•
The unit may be opened only by authorized service personnel!
Potentially lethal voltage inside the unit.
•
Do not use the unit if damaged. This applies in particular when the mains cable is damaged.
•
The connection of the pressure tube to the high-pressure flasks, compressors or building
pressure systems is not part of these operating instructions! The connection may only be
carried out by trained personnel or tradesmen, taking into account the local regulations.
•
The device may only be operated with dry, non-oiled, pressurized air or purified nitrogen
gas with 4000 hPa to a maximum of 8000 hPa (4 bar – 8 bar, 58 PSI – 116 PSI)!
Other gases are not permitted!
•
Loud hissing may occur when coupling or uncoupling the pressure tube! Don't be alarmed!
When coupling, hold the ears at an adequate distance from the coupling!
•
The pressure tube of the external pressure supply is not permanently UV light resistant.
When using, for example, under a sterile bench with UV sterilization, this should be taken
into account, and the pressure tube should, where necessary, be removed prior to UV
sterilization.
•
This pressure tube may not be bent. The smallest bending radius is 50 mm.
Hold the pressure tube tightly when uncoupling! The brief pressure surge when uncoupling
•
can cause the pressure tube to fly off! Following the uncoupling, the pressure tube remains
under pressure when the pressure is not released using the Standby function!
If the unit is not used in accordance with purpose for which it is intended, its protective
•
function may be impaired.
•
•
Caution!
Take care when mounting glass capillaries or tools!
Mounting must always take place without the use of pressure!
Capillaries which have been mounted incorrectly may break loose under pressure.
Never point mounted capillaries at persons.
Caution!
Take care when using glass capillaries! During mounting or when placed upon a
hard surface, the glass capillaries may break and glass splinters may cause injury.
It is therefore essential to wear eye protection!
1
Safety precautions and application
53
1 Safety precautions and application
•
In the case of independently-manufactured capillaries, ensure that the outer diameter of the
capillary corresponds to the specifications of the capillary grip head.
•
Over longer idle periods the device should be separated from the pressurized air supply and
then depressurized using the key and stored switched off!
1
•
Use the unit only for the described purpose! The unit and in particular the high pressure of
the Clean function may not be used for purposes other than microinjection!
•
Transfer
If the device is passed on to someone else, please include the instruction manual.
•
Disposal
In case the product is to be disposed of, the relevant legal regulations are to be observed.
•
Information on the disposal of electrical and electronic devices in the European Community
The disposal of electrical devices is regulated within the European Community by national
regulations based on EU Directive 2002/96/EC on waste electrical and electronic equipment
(WEEE).
According to these regulations, any devices supplied after 13.08.05 in the business-tobusiness sphere, to which this product is assigned, may no longer be disposed of in
municipal or domestic waste. They are marked with the following symbol to indicate this.
•
Safety precautions and application
•
•
•
As disposal regulations within the EU may vary from country to country, please
contact your supplier if necessary.
Use only original Eppendorf accessories. Only original accessory parts may be used.
The only exceptions are auxiliary pieces of equipment recommended by the manufacturer.
Otherwise all claims under guarantee and liability become void immediately. Use other
accessories only if you have a certificate from Eppendorf AG confirming technical safety in
use.
The responsibility for the correct functioning of the device passes onto the owner or user if
the device is maintained or serviced improperly by persons not belonging to the service
team authorized by Eppendorf or if used for a purpose other than that intended. Guarantee
and conditions of liability of the terms of sale and delivery of Eppendorf AG are not
expanded by the above notes.
The Eppendorf AG reserves the right to make technical alterations to this device!
54
2 Device description
The FemtoJet® express enables minute quantities of liquid to be injected into cells. The relevant
parameters (injection pressure pi and injection time ti) are set using the variable regulators.
The key or connected hand or foot control or the INJECT key of the connected
micromanipulator are used to trigger these two parameters. If no injections are being
performed, compensation pressure pc is applied permanently.
The connection of an external pressure supply enables faster injections, even with higher
pressure.
2
Device description
55
2 Device description
2.1 Overview
2
Device description
1
2
3
4
1 Display
–Displays relevant parameters during injection or setting of parameters.
–Guides operator through functions.
2 Keypad
–To trigger actions.
–To call up the function menu.
–To switch to and from Standby mode.
3 Variable regulators
–To set parameters injection pressure p
–To select functions and parameters in Function mode.
4 Tube connection
–Connect the injection tube for the capillary holder to the device via the bayonet connection.
56
, injection time ti and compensation pressure pc.
i
2 Device description
Standard accessories:
1 Pressure tube, 2,5 m, G 1/4-inch for external pressure supply incl. couplings
1 Injection tube, 2 m
1 Universal capillary holder with grip head 0 for capillary diameters
of 1.0 mm and adapter for Femtotips
1 Hand control for triggering injection and the Clean function
1 Femtotips®, Femtotips® II and Microloader
1 Mains cable
1 Operating manual
Optional accessories:
1 Foot control for triggering the Inject function
1 Connecting cable (Interface Cable 5171) to the Micromanipulator 5171 or InjectMan
®
2
Device description
®
57
2 Device description
2.2 Short description
Display
Display of all relevant parameters and information.
2
Function keys
Device description
Clean
To apply the maximum pressure for cleaning the capillary to the tube outlet.
Auto / Manu
To switch between manual and automatic time-controlled injection.
Count
To reset to zero the counter for injections which have been actuated.
58
Inject
To execute automatic or manual injection.
Menu
To switch the device to menu control; selected functions may be executed.
Standby
When the key is pressed briefly:
To switch the device between operating mode and standby mode or vice
versa.
When the key is held down:
As above, but also deairs the valve system.
2 Device description
Variable regulator
In the normal operating mode, the parameters for injection are set by the three variable
regulators. The arrangement of the variable regulators corresponds to the arrangement of the
parameters in the two bottom lines of the display:
From left to right: injection pressure pi, injection time ti and compensation pressure pc.
In other operating modes, the variable regulators have either the significance indicated in the
display or no significance at all.
Tube connection
The injection tube for the capillary holder is connected to the device via the bayonet connection.
2
Device description
59
2 Device description
2.3 Startup
2
Device description
Fig. 2: Diagram of rear of device
–Electrical connection is effected via the mains socket with integrated mains switch (1).
–The connection to the external pressure supply (2) takes place via the delivered pressure
tube, 2.5 m.
–Connection to an Eppendorf manipulator designed for injection is effected via the 15-pin
interface (3).
–Optional switches (e.g. foot control, hand control) or a computer can be connected to the
9-pin interfaces (4).
4
4
3
2
1
60
2 Device description
Connection of the external pressure supply
Dry, non-oiled, compressed air or purified nitrogen gas is to be used!
Pressure range from 4000 hPa to a maximum of 8000 hPa (4 bar – 8 bar, 58 PSI – 116 PSI),
constant pressure.
The connection for the external pressure supply is found on the rear side of the device.
Remove the black plastic cap.
The delivered pressure tube can be connected directly to the device.
2
Device description
1
Fig. 3: pressure tube, 2.5 m
1 Quick-lock coupling for mounting on the compressor, building pressure system
(e.g. when no connection system is available in the lab)
2 Plug sleeve, for coupling 1
3 Quick-lock coupling for connection to the FemtoJet
4 Plug sleeve to the FemtoJet® express, for coupling 3
The connection to the compressor, compressed air flask or building system can take place with
the supplied lock coupling (G1/4-inch threaded) or must be adapted by building services to the
existing connections.
The connection of the pressure tube to the high-pressure flasks, compressors or building
pressure systems is not part of these operating instructions!
Switching on the device
Connect the device to the external pressure supply. The injection tube may not be connected
when switching on and during the initialization phase.
Set the mains switch on the rear of the device to "I".
After being switched on, the unit runs through an initialization phase with self-test and valve
adjustment. Duration of approx. 1 minute.
Switching off the device
If it is not going to be used for an extended period, the device is to be separated from the
external pressure supply and to be depressurized via the switch-off routine using the
key. Then switch off the device at the rear.
2
3
®
express
4
61
3 Mode of operation
3.1 Display, variable regulators and keypad
Before operating the device for the first time, users should read this section carefully. This
section describes operation of the device only. Handling of the injection capillaries is described
in Section 3.3.
3
AUTOn= 0
pi= 150 hPa
Supply pressure 5261
pi[hPa]ti[s]pc[hPa]
1500.2 50
AUTOn= 0
Mode of operation
pi= 2.15 PSI
Supply pressure 76.3
pi[PSI]ti[s]pc[PSI]
2.150.2 0.72
3.1.1 Setting the injection parameters
The injection parameters have to be adapted to the injection capillaries used and the cell type
and liquid which are to be injected.
The following injection parameters can be set using the variable regulators:
Injection pressurep
Injection time t
Compensation pressurep
The settings are in increments of 1 hPa/0.1 s. The display can show hPa or psi, as required.
The setting and internal pressure conversion are always performed in hPa. Rounding for the
display in psi can result in increments of 0.01 to 0.02 psi.
The display of pressure in psi varies according to the size of the value (e.g.: 9.99 may also
appear in the display as 10.0).
Note:
An adjustment of pi or pc above the internally adjacent system pressure ps doesn't make sense.
The values for pi, pc should lie at least 500 hPa (7.2 PSI) below the system pressure in order to
ensure secure valve regulation.
Slight deviations between the nominal value (specified value) and displayed compensation
pressure pc (actual value) may occur for technical reasons. The relevant value is the actual value
measured.
Display
The display shows the following:
1st line:
injection mode AUTO or MANU,
counter n for number of injections performed
2nd line (bold):
The most-recently set parameter or relevant parameter
Middle line:
In the device, adjacent maximum system pressure p
Bottom lines:
Parameters which affect injection
Compensation pressure p
the normal mode is displayed as an actual measured value.
The arrangement of the three parameters shown, pi, ti, pc,
corresponds to the arrangement of the variable regulators.
0; 5 – 6,000 hPa (0.00; 0.07 – 87.0 psi)
i
0.1 – 99.9 s
i
0; 5 – 6,000 hPa (0,00; 0.07 – 87.0 psi)
c
which is applied permanently in
c
.
s
62
3 Mode of operation
The following settings are used as a basis for initial
injection experiments using Femtotip® / Femtotip® II:
Display:
p
left-hand variable regulator: 150 hPa
i
ti central variable regulator: 0.5 s
pcright-hand variable regulator: 50 hPa
Variable regulators:
3.1.2 Variable regulators
The variable regulators are set so that when they are turned slowly, the values are changed by
1 hPa / 0.1 second per notch (0.01 – 0.02 psi).
If the regulators are turned faster, the values are altered in larger increments.
3.1.3 Keypad
Inject
AUTOn= 0
pi= 150 hPa
pi[hPa]ti[s]pc[hPa]
1500.5 50
AUTO - INJECTn= 0
3
Mode of operation
Performs automatic or manual injection. Automatic
injection is controlled via the injection time ti, and the
injection pressure pi is used. In the second line, injection
time ti is counted down from the nominal value.
Manual injection is performed for as long as the key is held down. Injection pressure p
used. In the second line, injection time ti is counted upwards from 0.0.
If the FemtoJet® express is connected to the Eppendorf Micromanipulator InjectMan® NI 2,
5171 or Inject Man® the injection movement of the micromanipulator is also started via the
key. Both units then combine as a system to perform semi-automatic microinjection.
Alternatively, the INJECT key of the connected micromanipulator or the hand / foot control can
be used.
If synchronization is set to "limit" (SYNC = LIMIT) in the INJECT function of the micromanipulator, semi-automatic injection can only be triggered via the micromanipulator.
Supply pressure 5261
pi[hPa]ti[s]pc[hPa]
0.4 s
1500.6 50
i
is
63
3 Mode of operation
Clean
CLEANn= 0
The maximum available pressure for cleaning is applied
at the tube outlet for as long as the key is held down.
3
Once the key is released, the pressure drops to
compensation pressure pc again. The pressure curve is
shown in the form of a bar.
Auto / Manu
To switch between manual and automatic,
time-controlled injection.
In the case of manual injection, time is counted upwards
from 0.0 seconds. The parameter ti is not displayed in the
Mode of operation
bottom two lines in order to emphasize that the time
cannot be set via the variable regulators in this case.
Count
To reset to zero the counter for injections which have
been actuated.
Menu
To switch the device to menu control; selected functions
may be executed.
In the first line, FUNC indicates that the device is in the
function menu and shows which function number is
active.
The second line contains the function description
(e.g. "Select pressure unit").
The third line requests an action (e.g. press the ,
key to change the pressure unit).
Current status: [hPa]
The two bottom lines show information about which
action is assigned to which variable regulator
(e.g. other functions can be selected using left-hand
variable regulator pi).
≡≡≡≡≡≡≡≡≡≡≡≡≡≡≡_____
Supply pressure 5261
pi[hPa]ti[s]pc[hPa]
1500.2 50
MANU - INJECTn= 0
0.4 s
Supply pressure 5261
pi[hPa]pc[hPa]
150 50
FUNC 1
Push Inject to select
Pressure Unit hPa
Turn pi to
select function
PSI
64
3 Mode of operation
Standby
in the case of a brief interruption
When key is pressed briefly:
The device switches to the Standby mode without
deairing. The illumination of the display switches off.
"STANDBY" appears in the display with the suffix
"Supply pressure = rxyz" as a code for system pressure
ps being maintained.
Application:
For temporary interruptions or use of as a Reset
function.
When key is pressed again:
The device is reactivated and runs through a warm-up
routine. This lasts approximately one minute.
Standby
with deairing of pressure accumulator
Hold down the key (approx. 2 seconds; the display
moves to "STANDBY"):
The device switches to the Standby mode. The valve
system is completely deaired, the illumination of the
display switches off.
In addition to WAIT, the display also instructs that the
compressed air supply is to be separated from the
device. At the same time, some of the system pressure is
released in order to support this instruction.
A noise can be heard at this stage.
"STANDBY" appears in the display with the note
"Supply pressure = x" (x ~ 0).
Application:
For correct shutdown, with deairing of the entire pressure
system in the event of longer periods of non-use or
standstill (weekend or storage) with subsequent
switch-off using the mains switch.
Press the key again:
The device is reactivated and runs through the warm-up
routine.
STANDBY
3
Supply pressure 6789
Mode of operation
Hold standby
to exhaust
WAIT
Close supply pressure
Supply pressure 5261
STANDBY
Supply pressure 0
Reset
To reset the device to the initial mode by pressing the
key briefly or via the mains switch.
65
Loading...
+ 39 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.