Eppendorf FemtoJet User Manual

Page 1
FemtoJet
®

Operating Manual

Your local distributor: www.eppendorf.com/worldwide
Eppendorf AG 22331 Hamburg · Germany Tel. +49 40 538 01-0 Fax +49 40 538 01-556 E-Mail: eppendorf@eppendorf.com
Eppendorf North America, Inc. One Cantiague Road · P.O. Box 1019 Westbury, N.Y. 11590-0207 · USA Tel. +1 516 334 7500 Toll free phone 800 645 3050 Fax +1 516 334 7506 E-Mail: info@eppendorf.com
Application Support
Europe, International: Tel. +49 1803 666 789 E-Mail: support@eppendorf.com
North America: Tel. 800 645 3050 ext. 2258 E-Mail: support_NA@eppendorf.com
Asia, Pacific: Tel. +603 8023 2769 E-Mail: support_AsiaPacific@eppendorf.com
eppendorf® is a registered trademarkPrinted in Germany
Page 2
Eppendorf Offices
AUSTRALIA / NEW ZEALAND
Eppendorf South Pacific Pty. Ltd. Tel. +61 2 98 89 50 00 Fax +61 2 98 89 51 11 E-mail: Info@eppendorf.com.au Internet: www.eppendorf.com.au
AUSTRIA
Eppendorf Austria Tel. +43 1 2901756-0 Fax +43 1 2901756-20 E-Mail: office@eppendorf.at Internet: www.eppendorf.at
BRAZIL
Eppendorf do Brasil Ltda. Tel. +55 11 3095 9344 Fax +55 11 3095 9340 E-Mail: eppendorf@eppendorf.com.br Internet: www.eppendorf.com.br
CANADA
Eppendorf Canada Ltd. Tel. +1 905 826 5525 Fax +1 905 826 5424 E-Mail: canada@eppendorf.com Internet: www.eppendorfna.com
CHINA
Eppendorf China Ltd. Tel. +86 21 68760880 Fax +86 21 50815371
E-Mail: market.info@eppendorf.cn
Internet: www.eppendorf.cn
GERMANY
Eppendorf Vertrieb Deutschland GmbH Tel. +49 2232 418-0 Fax +49 2232 418-155 E-Mail: vertrieb@eppendorf.de Internet: www.eppendorf.de
INDIA
Eppendorf India Limited Tel. +91 44 42 11 13 14 Fax +91 44 42 18 74 05 E-Mail: info@eppendorf.co.in Internet: www.eppendorf.co.in
ITALY
Eppendorf s.r.l. Tel. +390 2 55 404 1 Fax +390 2 58 013 438 E-Mail: eppendorf@eppendorf.it Internet: www.eppendorf.it
JAPAN
Eppendorf Co. Ltd. Tel. +81 3 5825 2363 Fax +81 3 5825 2365 E-Mail: info@eppendorf.jp Internet: www.eppendorf.jp
NORDIC
Eppendorf Nordic Aps Tel. +45 70 22 29 70 Fax +45 45 76 73 70 E-Mail: nordic@eppendorf.dk Internet: www.eppendorf.dk
SPAIN
Eppendorf Ibérica S.L.U. Tel. +34 91 651 76 94 Fax +34 91 651 81 44 E-Mail: iberica@eppendorf.es Internet: www.eppendorf.es
SWITZERLAND
Vaudaux-Eppendorf AG Tel. +41 61 482 1414 Fax +41 61 482 1419 E-Mail: vaudaux@vaudaux.ch Internet: www.eppendorf.ch
UNITED KINGDOM
Eppendorf UK Limited Tel. +44 1223 200 440 Fax +44 1223 200 441 E-Mail: sales@eppendorf.co.uk Internet: www.eppendorf.co.uk
USA
Eppendorf North America Tel. +1 516 334 7500 Fax +1 516 334 7506 E-Mail: info@eppendorf.com Internet: www.eppendorfna.com
OTHER COUNTRIES
Internet: www.eppendorf.com/worldwide
FRANCE
EPPENDORF FRANCE S.A.R.L. Tel. +33 1 30 15 67 40 Fax +33 1 30 15 67 45 E-Mail: eppendorf@eppendorf.fr Internet: www.eppendorf.fr
SOUTH & SOUTHEAST ASIA
Eppendorf Asia Pacific Sdn. Bhd. Tel. +60 3 8023 2769 Fax +60 3 8023 3720 E-Mail:
asiapacifichq@eppendorf.com.my
Internet: www.eppendorf.com.my
Page 3

Contents

1 Safety precautions and application
2 Device description
2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Short description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Mode of operation
3.1 Display, variable regulators and keypad . . . . . . . . . . . . . . . . 12
3.1.1 Setting the injection parameters . . . . . . . . . . . . . . . . . . . . . . 12
3.1.2 Variable regulators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1.3 Keypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2 Hand control / foot control . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3 Connecting the capillary holder . . . . . . . . . . . . . . . . . . . . . . . 18
3.4 Information on working practices. . . . . . . . . . . . . . . . . . . . . . 19
3.4.1 Compensation pressure p
3.4.2 Injection pressure pi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.4.3 Injection time ti . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.4.4 Clean function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.4.5 Manual or automatic injection . . . . . . . . . . . . . . . . . . . . . . . . 23
3.4.6 Constant pressure/constant working pressure . . . . . . . . . . . 23
3.4.7 General note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.5 The first injection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6 Connecting an Eppendorf Micromanipulator . . . . . . . . . . . . . 25
3.7 Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.7.1 Changing the capillaries . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.7.2 Setting the pressure unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.7.3 Switching on/off the loudspeaker. . . . . . . . . . . . . . . . . . . . . . 27
3.7.4 Switching on/off the background illumination of the display . 27
3.7.5 Constant pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.7.6 Dewatering the pressure accumulator. . . . . . . . . . . . . . . . . . 28
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
. . . . . . . . . . . . . . . . . . . . . . . . . . 21
c
. . . . . . . . . . . . . . . . . . . 3
4 Cleaning and Service
4.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.2 Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.3 Decontamination prior to shipping . . . . . . . . . . . . . . . . . . . . . 29
4.3 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
No part of this publication may be reproduced without the prior permission of the copyright owner.
Copyright© 2006 by Eppendorf AG, Hamburg
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
B 5247 900.011-12/0108
Page 4
Contents
5 Troubleshooting
5.1 Potential errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.1 Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.2 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.3 Error messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6 Technical data
7 Ordering information
7a Ordering information for North America
8 Index
Appendix A
A.1 Description of interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
A.2 Commands and queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
EC Conformity Declaration
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Operation via a computer
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
. . . . . . . . . . . . . . .44
. . . . . . . . . . . . . . . . . . . . . . . . . . 49
2
Page 5

1 Safety precautions and application

Observe operating instructions!
The device is only designed for use in the research lab!
The FemtoJet is a lab device. It may only be operated by appropriately qualified lab personnel.
Only use the device for microinjection under observance of all relevant legal regulations!
Before plugging in the device, please compare your voltage
requirements with the specifications on the identification plate!
The unit may not be connected to a socket that is not grounded!
Use the appropriate mains cable.
Electrical connections may be effected only when the device is
switched off!
The unit may be opened only by authorized service personnel! Potentially lethal voltage inside the unit.
Do not use the unit if damaged. This applies in particular when the
mains cable is damaged.
If the unit is not used in accordance with purpose for which it is
intended, its protective function may be impaired.
In the case of independently-manufactured capillaries, ensure that the
outer diameter of the capillary corresponds to the specifications of the capillary grip head.
In the case of extended periods of non-use, the unit should be depressurized using the key and switched off before being stored!
Caution!
Take care when mounting glass capillaries or tools! Mounting must always take place without the use of pressure! Capillaries which have been mounted incorrectly may break loose under pressure. Never point mounted capillaries at persons.
Caution!
Take care when using glass capillaries! During mounting or when placed upon a hard surface, the glass capillaries may break and glass splinters may cause injury. It is therefore essential to wear eye protection!
3
Page 6
1 Safety precautions and application
Use the unit only for the described purpose! The unit and in particular
the high pressure of the Clean function may not be used for purposes other than microinjection!
Transfer
If the device is passed on to someone else, please include the instruction manual.
Disposal
In case the product is to be disposed of, the relevant legal regulations are to be observed.
Information on the disposal of electrical and electronic devices in the European Community
The disposal of electrical devices is regulated within the European Community by national regulations based on EU Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). According to these regulations, any devices supplied after 13.08.05 in the business-to-business sphere, to which this product is assigned, may no longer be disposed of in municipal or domestic waste. They are marked with the following symbol to indicate this.
As disposal regulations within the EU may vary from country to country, please contact your supplier if necessary.
Use only original Eppendorf accessories. Only original accessory parts may be used. The only exceptions are auxiliary pieces of equipment recommended by the manufacturer. Otherwise all claims under guarantee and liability become void immediately. Use other accessories only if you have a certificate from Eppendorf AG confirming technical safety in use.
The responsibility for the correct functioning of the device passes
onto the owner or user if the device is maintained or serviced improperly by persons not belonging to the Eppendorf service team or if used for a purpose other than that intended. Eppendorf AG is not liable for damage occurring from non-observance of these notes. Guarantee and conditions of liability of the terms of sale and delivery of Eppendorf AG are not expanded by the above notes.
The Eppendorf AG reserves the right to make technical alterations to this device!
4
Page 7

2 Device description

2 Device description
The FemtoJet enables minute quantities of liquid to be injected into cells. The relevant parameters (injection pressure pi and injection time ti)
are set using the variable regulators. The key or connected hand or foot control or the INJECT key of the connected micromanipulator are
used to trigger these two parameters. If no injections are being performed, compensation pressure pc is applied permanently.
5
Page 8
2 Device description

2.1 Overview

1
2
3
4
1
Display
– Displays relevant parameters during injection or setting
of parameters.
– Guides operator through functions.
2
Keypad
– To trigger actions. – To call up the function menu. – To switch to and from Standby mode.
3
Variable regulators
– To set parameters injection pressure pi, injection time ti and
compensation pressure pc.
– To select functions and parameters in Function mode.
4
Tube connection
– To connect the injection tube to the device via the bayonet joint.
6
Page 9
2 Device description
Standard accessories:
1 Injection tube, 2 m 1 Universal capillary holder with grip head for capillary diameters of 1.0 mm and adapter for Femtotips 1 Hand control for triggering injection and the Clean function 1 Femtotips, Femtotips II and Microloader 1 Mains cable 1 Operating manual
Optional accessories:
1 Foot control for triggering the Inject function 1 Connecting cable (Interface Cable 5171) to the Micromanipulator 5171 or InjectMan
7
Page 10
2 Device description

2.2 Short description

Display
Display of all relevant parameters and information.
Function keys
Clean
To apply the maximum pressure for cleaning the capillary to the tube outlet.
Auto / Manu
To switch between manual and automatic time-controlled injection.
Count
To reset to zero the counter for injections which have been actuated.
Inject
To execute automatic or manual injection.
Menu
To switch the device to menu control; selected functions may be executed.
Standby
When the key is pressed briefly: To switch the device between operating mode and standby mode. When the key is held down: As above, but also deairs the pressure accumulator.
8
Page 11
2 Device description
Variable regulator
In the normal operating mode, the parameters for injection are set by the three variable regulators. The arrangement of the variable regulators corresponds to the arrangement of the parameters in the two bottom lines of the display: From left to right: injection pressure pi, injection time ti and compensation pressure p In other operating modes, the variable regulators have either the significance indicated in the display or no significance at all.
Tube connection
The injection tube is connected to the device via a bayonet joint.
.
c
9
Page 12
2 Device description

2.3 Startup

Diagram of rear of device
– Electrical connection is effected via the mains socket with integrated
mains switch (1).
3 3
2
1
– Connection to an Eppendorf manipulator designed for injection is
effected via the 15-pin interface (2).
– Optional switches (e.g. foot control, hand control) or a computer can
be connected to the 9-pin interfaces.
Switching on the device
Set the mains switch on the rear of the device to "I". The unit is switched on, the display lights up, the unit runs through a warm-up routine and is then ready for operation. The injection mode is shown in the display.
The injection tube may not be connected when switching on and during the initialization phase.
After being switched on, the unit runs through an initialization phase and supply pressure is built up. This takes approximately three minutes.
10
Page 13
2 Device description
The drainage valve of the pressure cylinder is switched several times at the beginning of the initialization phase. The pressure cylinder is also drai­ned every 3 hours when switched on. This results in a certain noise level following a series-connected reminder signal.
Switching off the device
If it is not going to be used for an extended period, the device should
be depressurized via the switch-off routine using the key. This guarantees a long service life of the device (see Section 3.1.3).
The switch-off routine ensures that the system is dewatered and deaired correctly. Do not disconnect the device from the mains supply until dewatering and deairing have taken place. If only the mains switch is actuated, the existing pressure remains in the system.
11
Page 14

3 Mode of operation

3 Mode of operation

3.1 Display, variable regulators and keypad

Before operating the device for the first time, users should read this section carefully. This section describes operation of the device only. Handling of the injection capillaries is described in Section 3.3.
AUTO n= 0
pi= 150 hPa
pi[hPa] ti[s] pc[hPa]
150 0.2 50
AUTO n= 0
pi= 215 PSI
pi[PSI] ti[s] pc[PSI]
2.15 0.2 0.72
Display
The display shows the following:
1st line: injection mode AUTO or MANU, counter n for number of injections performed
2nd line (bold): The most-recently set parameter or relevant parameter
Bottom lines: Parameters which affect injection
Compensation pressure pc which is applied permanently in the normal mode is displayed as an actual measured value.
The arrangement of the three parameters shown, p corresponds to the arrangement of the variable regulators.
, ti, pc,
i
3.1.1 Setting the injection parameters
The injection parameters have to be adapted to the injection capillaries used and the cell type and liquid which are to be injected.
The following injection parameters can be set using the variable regulators:
Injection pressure pi 0; 5 – 6,000 hPa (0.00; 0.07 – 87.0 psi) Injection time ti0.1 – 99.9 s Compensation pressure pc0; 5 – 6,000 hPa (0,00; 0.07 – 87.0 psi)
The settings are in increments of 1 hPa/0.1 s. The display can show hPa or psi, as required. The setting and internal pressure conversion are always performed in hPa. Rounding for the display in psi can result in increments of 0.01 to 0.02 psi. The display of pressure in psi varies according to the size of the value (e.g.: 9.99 may also appear in the display as 10.0).
12
Page 15
3 Mode of operation
Slight deviations between the nominal value (specified value) and displayed compensation pressure pc (actual value) may occur for technical reasons. The relevant value is the actual value measured.
The following settings are used as a basis for initial injection experiments using Femtotip
pileft-hand variable regulator: 150 hPa ti central variable regulator: 0.5 s pcright-hand variable regulator: 50 hPa
/ Femtotip II:
Variable regulators:
Display:
AUTO n= 0
pc= 50 hPa
pi[hPa] ti[s] pc[hPa]
150 0.5 50
3.1.2 Variable regulators
The variable regulators are set so that when they are turned slowly, the values are changed by 1 hPa / 0.1 second per notch (0.01 – 0.02 psi). If the regulators are turned faster, the values are altered in larger increments.
13
Page 16
3 Mode of operation
3.1.3 Keypad
Inject
Performs automatic or manual injection. Automatic injection is controlled via the injection time ti, and the injection pressure pi is used. In the second line, injection time ti is counted down from the nominal value.
Manual injection is performed for as long as the key is held down. Injection pressure pi is used. In the second line, injection time ti is counted upwards from 0.0.
If the FemtoJet is connected to the Eppendorf Micromanipulator InjectMan NI 2, 5171 or Inject Man, the injection movement of the
micromanipulator is also started via the key. Both units then combine as a system to perform semi-automatic microinjection. Alternatively, the INJECT key of the connected micromanipulator or the
hand/foot control can be used. If synchronization is set to "limit" (SYNC = LIMIT) in the INJECT function
of the micromanipulator, semi-automatic injection can only be triggered via the micromanipulator.
Clean
The maximum available pressure for cleaning is applied at the tube outlet for as long as the key is held down. Once the key is released, the pressure drops to compensation pressure p pressure curve is shown in the form of a bar.
again. The
c
AUTO - INJECT n= 0
0.4 s
pi[hPa] ti[s] pc[hPa]
150 0.6 50
CLEAN n= 0
≡≡≡≡≡≡≡≡≡≡≡≡≡≡≡_____
pi[hPa] ti[s] pc[hPa]
150 0.2 50
14
Page 17
3 Mode of operation
Auto / Manu
To switch between manual and automatic, time-controlled injection.
In the case of manual injection, time is counted upwards from 0.0 seconds. The parameter t bottom two lines in order to emphasize that the time cannot be set via the variable regulators in this case.
Count
To reset to zero the counter for injections which have been actuated.
Menu
To switch the device to menu control; selected functions may be executed.
is not displayed in the
i
MANU - INJECT n= 0
0.4 s
pi[hPa] pc[hPa]
150 50
In the first line, FUNC indicates that the device is in the function menu and shows which function number is active. The second line contains the function description (e.g. "Select pressure unit").
The third line requests an action (e.g. press the key to change the pressure unit). Current status: [hPa]
The two bottom lines show information about which action is assigned to which variable regulator (e.g. other functions can be selected using left-hand variable regulator p
).
i
FUNC 1 Select pressure unit INJECT = PSI [hPa]
Tu r n pi to select function
15
Page 18
3 Mode of operation
Standby
in the case of a brief interruption
When key is pressed briefly: The device switches to the Standby mode
without
of the display switches off. "STANDBY" appears in the display with the suffix "Supply pressure = rxyz" as a code for system pressure ps being maintained.
Application: For temporary interruptions or use of
as a Reset function.
When key is pressed again: The device is reactivated and runs through a short warm-up routine. This lasts approximately one minute.
Standby
with deairing of pressure accumulator
Hold down the key (approx. 2 seconds; the display moves to "STANDBY"):
The device switches to the Standby mode. The pressure accumulator is completely deaired, the illumination of the display switches off. "STANDBY" appears in the display with the note "Supply pressure = x" (x ~ 0).
A noise can be heard at this stage.
Application: For correct shutdown, with deairing of the entire pressure system in the event of longer periods of non-use or standstill (weekend or storage) with subsequent switch-off using the mains switch.
Press the key again: The device is reactivated and runs through the warm-up routine. This takes approximately three minutes.
deairing. The illumination
STANDBY
Supply pressure 6789
Hold standby
to exhaust
STANDBY
Supply pressure 0
16
Page 19
3 Mode of operation
Reset
To reset the device to the initial mode by pressing the key briefly or via the mains switch.
The pressure accumulator is not deaired.

3.2 Hand control / foot control

The functions of the hand/foot control correspond to the functions of the
and keys on the
FemtoJet
.
The function of the foot control corresponds to the key.
17
Page 20
3 Mode of operation

3.3 Connecting the capillary holder

Universal capillary holder
The universal capillary holder is used to locate the injection capillaries. Prefabricated Eppendorf Femtotips / Femtotips II (6) are screwed into the universal capillary holder using the adapter (5). To do so, the adapter for Femtotips (5) is first screwed into the front rotatable knurled screw (4) and then the Femtotip / Femtotip II is attached. If self-pulled capillaries are to be attached, the grip head 0 (7) needs to be screwed loosely into the front rotatable knurled screw (4). The self-pulled capillary is pushed into the front opening of the grip head through the two O-rings and fixed in position by tightening the grip head.
12 3 4 5 6
7
Universal capillary holder
1Tube connection 2 Rear knurled screw 3Tension piece 4Front knurled screw, rotatable 5 Adapter for Femtotips 6Femtotip / Femtotip II 7Grip head 0, plastic, with two O-rings and sealing washer
(microcapillaries for microinjection)
Connecting the universal capillary holder
The two-meter-long injection tube is connected to the switched on FemtoJet (1) via the bayonet joint (2). The injection tube (3) is screwed to the rear knurled screw (4) of the universal capillary holder. Femtotips Femtotips II are fitted at the front end of the universal capillary holder (5) or independently-manufactured capillaries are clamped with the aid of the appropriate grip head.
12 3 4
18
/
5
Page 21
3 Mode of operation
Other accessories
Grip head 0, incl. 2 sets of O-rings and sealing washer.
Using grip head 0 supplied, capillaries with an outer diameter of 1.0 mm to 1.1 mm can be used. In the case of outer diameters which differ from these measurements, the relevant grip heads can be orde­red as non-standard accessories for the universal capillary holder (see Section 7, Ordering informa­tion). The transparent grip head allows the user to visually monitor the position of the capillary.

3.4 Information on working practices

The following pressures can be used:
Name Pressure range Comments
Compensation pressure pc
Injection pressure p
Clean pressure maximum
Constant working pressure pw
System pressure ps [supply pressure]
0; 5 – 6,000 hPa
0.00; 0.07 – 87.0 psi
0; 5 – 6,000 hPa
i
0.00; 0.07 – 87.0 psi
up to 7,000 hPa maximum up to 102 psi
0; 5 – 6,000 hPa
0.00; 0.07 – 87.0 psi
maximum up to 7,000 hPa maximum up to 102 psi
Select Function 4 via the key.
This is used directly for
.
In certain phases of initialization, the system pressure may temporarily increase to a maximum of 8,000 hPa (117 psi).
19
Page 22
3 Mode of operation
The usual values for the permanent compensation pressure Pc are bet­ween 30 – 50 hPa (0,44 – 0,73 PSI) for Femtotips and similar capillaries. The FemtoJet shows good, constant values within this range Immediately after the device is switched on and maintains the balance between undesired absorption of medium on the one hand and excessive loss of the liquid to be injected on the other hand.
Quite low P (e.g. capillaries with an interior diameter of 5 µm and greater)! In order to be able to also work precisely and efficiently here, the FemtoJet should be switched on app. 60 to 90 minutes before beginning the experiment to guarantee the correct thermal settings. A pressure setting below 5 hPa is not possible. You may merely also set 0 hPa. An internal self-calibration of 3 seconds for more precise adjustment of the pressure takes place every 10 minutes. No injection is carried out during this time. Injection can continue as usual following the self-calibration. An already activated injection will be carried out by the FemtoJet. The self-calibration takes place afterward. Message during the self-calibration on the display of the FemtoJet:
values are sometimes advisable for special applications
c
WAIT
self calibration
If the FemtoJet is connected with the InjectMan NI 2, the following message appears on the display of the micromanipulator:
Warning 001 Check Injector and cable Press joystick button
Self-calibration intervals:
– In the first hour after the machine is switched on: every 10 minutes.
– Afterwards: Every 30 min at pc < 30 hPa or every 15 min at pc > 30 hPa
following the last activation.
20
A brief signal sounds at the beginning and end of the calibration.
Press the joystick button again to repeat the injection.
Page 23
3 Mode of operation
3.4.1 Compensation pressure pc
The compensation pressure pc ensures that no medium flows into the capillary. Capillary forces would make liquid flow out of the cell culture dish into the injection capillary and thus dilute the injection material. To prevent this, a permanent compensation pressure p be selected so that there is a permanent slight flow-out of liquid from the injection capillary. The individual pressure level can be determined in a preliminary test. For example, a fluorescent dye can be added to the capillary to enable outflow to be monitored directly. The pressure required is heavily dependent on the surface tension and viscosity of the injection material. Experience has shown that compensation pressures of between 30 hPa and 300 hPa (0.44 – 4.4 psi) are appropriate values for ready-to­use Eppendorf Femtotips
injection capillaries.
is set. This should
c
Capillary suction
Compensation pressure p
Hydrostatic pressure P
hyd
c
Pressure in the capillaries
3.4.2 Injection pressure p
i
Injection pressure pi is applied during the injection process. This pressure is generally higher than the compensation pressure. In many cases, a pressure between 50 hPa and 500 hPa (0.73 – 7.2 psi) is appropriate for Femtotips. The volume injected is dependent on the injection pressure and the injection time. The injection pressure must be greater than the internal pressure of the cell into which injection should take place. However, if the cell or the cell nucleus swell up during injection, the injection pressure is too high.
If the compressor starts up occasionally during experiments, this does not impair the function of the FemtoJet
. Pressure is always available up to the maximum value of 6,000 hPa (approx. 87.0 psi), even when the compressor is topping up the supply pressure.
21
Page 24
3 Mode of operation
3.4.3 Injection time t
Injection time ti gives the period for which injection pressure pi is maintained. t
1.5 s. The relevant factor for the injected volume is the product of the injection pressure and the injection time.
Depending on the setting on the Eppendorf Micromanipulator which is connected, the injection time ti is measured from the point at which the
is already present during the injection movement and during penetration of the cell. Once the end of injection time has been reached, the device switches to compensation pressure pc. At the same time, the micromani­pulator is moved back into the working level. When "Injection at Z limit" is set, measurement of the injection time ti does not begin until the Z limit has been reached. Once ti has elapsed, the FemtoJet switches back to pc and the micromanipulator moves back into its starting position in the working level.
is usually selected from a range between approx. 0.3 s and
i
key is pressed. The injection pressure is built up immediately, so it
i
3.4.4 Clean function
The Clean function is used to flush out blocked capillaries. The maximum available pressure is applied to the outlet with the key. It is applied for as long as the key is pressed. The maximum pressure of the FemtoJet
changed. The pressure curve up to maximum pressure appears in the display in the
form of a bar chart. By pressing the key briefly, a lower pressure can also be obtained.
is in the range of 7,000 hPa (102 psi). The value cannot be
22
Page 25
3 Mode of operation
3.4.5 Manual or automatic injection
In the Manu operating mode, injection is performed for as long as the
key or hand / foot control is held down. The duration (ti) of
injection is measured and shown in the display. In the case of combination with an Eppendorf Micromanipulator, the injection movement is triggered.
In the Auto operating mode, the built-in timer facilitates a reproducible injection time. The injection runs automatically for the duration of set time
t
. It is triggered by the key or the hand / foot control being pressed
i
briefly. During injection, the Inject key is inactive, i.e. pressing it again has no effect. In combination with an Eppendorf Micromanipulator, the auto­matic injection procedure can also be triggered from the operating panel of the micromanipulator.
3.4.6 Constant pressure/constant working pressure
Using the Constant Pressure function, it is possible to select a higher pressure permanently if required. This function is helpful when special injection techniques require either a high pressure up to a maximum of 6,000 hPa (87 psi) or a permanent flow of injection solution (see Section 3.7.5). Control accuracy drops in the pressure range from 2,000 – 6,000 hPa (29 – 87 psi).
3.4.7 General note
FemtoJet is fitted with an integrated compressor. For this reason, the compressor may start up or excess pressure may be let off briefly before, during and after injections or during periods when the FemtoJet being used. This is not a malfunction of the device and does not impair injection.
In the case of injections under high pressure at very short intervals or in the case of the Clean function being carried out several times in succession, the pressure in the system accumulator may drop significantly. This is indicated by a warning or an error message. In this case, wait a short time until the supply pressure has stabilized again. It is then possible to continue working immediately.
is not
23
Page 26
3 Mode of operation
3.5 The first injection
The following are available:
Adherent cells in a Petri dish (possibly on CELLocates to observe
specific cells after injection),
Injection liquid (purified, e.g. by centrifugation),
Femtotips
Microloader,
0.5 – 10 µl pipette to take the Microloader,
A universal capillary holder on the micromanipulator which has been
preadjusted with a Femtotip
/ Femtotips II; other capillaries as an alternative
.
Filling the Femtotips
with the Microloader:
– Select function 0 "Capillary exchange" by pressing . Connect
the injection tube.
– Aspirate approximately 0.5 – 5.0 µl of centrifuged injection fluid with
the Microloader without touching the base of the vessel. Using the Microloader, fill the Femtotip tip approximately up to the centre of the Femtotips
through the rear opening, inserting the
and releasing the
fluid there.
– Hold the Femtotip
vertically, loosen the cap by turning it and allow it
to drop vertically.
– Screw the Femtotip
into the capillary holder.
– Complete Function 0 by pressing . Compensation pressure is
applied.
– Move the capillary into the cell medium (Petri dish) with the micro-
manipulator.
– Focus the cells with the microscope, position the capillary slightly
above the cells but do not allow it to come into contact with the base of the Petri dish. Focus the capillary.
– Using the key, check that the capillary is not blocked. Either
the fluid can be seen flowing out or it is rendered visible by the fluorescent substance it entrains. As a rule, any air bubbles present are blown out.
– Perform a sample injection using the preset parameters
~ 150 hPa, ti ~ 0.5 s, pc ~ 50 hPa, corresponding to
(e.g. p
i
p
~ 2.18 psi, ti ~ 0.5 s, pc ~ 0.73 psi). Injection can be monitored
i
visually when the cell size changes by approx. 5 % to 10 % or via fluorescence. If required, adapt parameters p
and/or ti.
i
24
Page 27
3 Mode of operation

3.6 Connecting an Eppendorf Micromanipulator

Connection to an Eppendorf Micromanipulator designed for automated injection is effected with the 15-pin interface.
1
2
FemtoJet
1 Foot control (optional) with Inject function 2 Hand control with Inject and Clean functions 3FemtoJet 4 Micromanipulator InjectMan NI 2 5 Connection socket of micromanipulator for Interface cable FemtoJet 6 Interface cable FemtoJet (accessories: InjectMan 7 Interface of FemtoJet
(rear) with all connection options, with InjectMan NI 2
3
to micromanipulator
67
45
NI 2)
25
Page 28
3 Mode of operation
2
1
345
8
67
FemtoJet (rear) with all connection options, with power unit 5171/ InjectMan 5179
1 Foot control (optional) with Inject function 2 Hand control with Inject and Clean functions 3FemtoJet 4 Micromanipulator 5171 or InjectMan
5 Connection socket of micromanipulator for
Transmission cable 5246 ("Y cable")
6 Transmission cable 5246
(accessory for Micromanipulator 5171 and InjectMan
7 Interface cable 5171 (optional)
(Connection from FemtoJet
8 Interface of FemtoJet
to Transmission cable 5246 "Y cable")
to micromanipulator
)

3.7 Functions

The Function menu is called up using the key. One of seven functions is selected using left-hand variable regulator p functionality is assigned to the variable regulators in the bottom line of the display.
Exit the Function menu by pressing the key again.
In cases of error, pressure control is switched off from within the Function menu. Then exit the Menu using the key.
Function 6, self-test, is described in Section 5, "Troubleshooting".
26
. The relevant
i
Page 29
3 Mode of operation
3.7.1 Changing the capillaries
FUNC 0 Change capillary
Capillary may be
changed now Tu r n pi to select function
FUNC 0
When this function is called up, the injection tube is deaired and any pressure present in the system is let out. Once the capillary has been changed,
exit the function using the key or variable regulator p
restart.
3.7.2 Setting the pressure unit
FUNC 1 Push Inject to select Pressure Unit hPa PSI
Tu r n pi to select function
FUNC 1
The other pressure unit is selected using the key and all pressure values are converted. The pressure unit
currently selected (in this case: hPa) is underlined.
3.7.3 Switching on/off the loudspeaker
FUNC 2 Push Inject to select Beeper ON
Tu r n pi to select function
OFF
FUNC 2
Use the key to switch the loudspeaker on or off. The currently selected setting (in this case: ON) is underlined. Acoustic error messages are not emitted if the setting is OFF.
; pressure control will
i
3.7.4 Switching on/off the background illumination of the display
FUNC 3 Push Inject to select Illumination ON OFF
Tu r n pi to select function
FUNC 3
Use the key to switch the illumination on or off. The currently selected setting (in this case: ON) is underlined.
In the normal operating mode, illumination switches off after three minutes if no operating control has been actuated. When an operating control is actuated, the illumination lights up again.
27
Page 30
3 Mode of operation
3.7.5 Constant pressure
FUNC 4 Continuous flow INJECT = start
Tu r n pi to select function
FUNC 4 Continuous flow
pw= 50 hPa
Tu r n pi to pw[hPa] select function 50
FUNC 4
This function can be used to apply a constant working pressure p range from 5 to 6,000 hPa (87.0 psi) at the pressure outlet. The pressure unit psi can also be used.
The working pressure p the right-hand variable regulator (p the specified value is shown in large characters. The actual value is shown in the bottom right of the display.
The Clean function can be activated. The pressure of the Clean function is shown in the bottom right of the display in the form of p
.
w
3.7.6 Dewatering the pressure accumulator
FUNC 5 Drain pressure supply INJECT = start
Tu r n pi to select function
FUNC 5
– Briefly opens the dewatering valve
of the pressure accumulator; any condensed humidity present is blown away.
– When this function is performed, the
noise of air being blown away can be heard.
This function is automatically performed every time the unit is switched on. It can be performed manually as described if condensation is suspected to be present.
in the
w
is set with
w
),
c
28
Page 31

4 Cleaning and Service

4.1 Cleaning

Before cleaning the FemtoJet, disconnect the plug.
Ensure that no fluids enter into the inside of the injector in order to avoid short-circuits in the electrical installation as well as corrosion.
Wipe painted parts and aluminum surfaces using a cloth and mild detergent and then with a dry cloth.
Warning:
or polishes.

4.2 Disinfection

Clean the FemtoJet before disinfecting.
Wipe every part of the FemtoJet, including accessories and connector cables, with disinfectant like alcohol (ethanol, isopropanol) or alcoholic disinfection media.
Spray disinfection is not advisable, as disinfectants can enter into the microinjector.
The disinfection method used must comply with current disinfection regulations and guidelines.
Do not use any corrosive, solvent or abrasive detergents

4.3 Decontamination prior to shipping

If the device is to be inspected, repaired or calibrated by the
service partners of the Eppendorf AG, it must be free of hazardous substances and clean!
A decontamination certificate must accompany the FemtoJet.
A printed form "Decontamination certificate for return of goods" and general instructions for decontamination are available at our www.eppendorf.com homepage.
The serial number of the device must be entered in the decontamination certificate.

4.4 Service

Warranty or service is taken over by the supplier or the service partner of the Eppendorf AG.
29
Page 32

5 Troubleshooting

5 Troubleshooting

5.1 Potential errors

– Pressure loss due to capillary not fitted with capillary holder
connected. Solution: Connect the capillary or select Function 0, Change capillary,
– Capillary is blocked.
Solution: Press the key for long enough.
– Compressor runs for a long time.
Solution: Take a short break.
–Valve rattles
Solution: Connect tube to filled capillary. Pressure control is designed
– No communication with the micromanipulator.
Solution: Check cables and connections.
via the key and variable regulator p
Change capillary. Purify injection fluid (e.g. by means of centrifugation).
Do not use the key too often within a short period. If possible, set a lower compensation pressure p
Perform self-test (Section 5.1.1).
for the tube connected.
.
i
c.
30
5.1.1 Self-test
FUNC 6 Self-Test INJECT = Complete CLEAN = Leak only
Tu r n pi to select function
FUNC 6
The FemtoJet is checked for leaks in the self-test.
This self-test can also be used if a leak is suspected.
The test runs automatically.
The injection tube must be connected to a closed capillary (e.g. melted closed or filled) for the pressure control to work properly.
Page 33
5 Troubleshooting
FUNC 6 Self-Test Compressor _____ s Ps leak _____ _____ hPa Pw leak _____ _____ hPa Injection _____ _____ hPa
ps=7000 pw= 18 hPa
FUNC 6 Self-Test
Ps leak _____ _____ hPa Pw leak _____ _____ hPa
ps=7000 pw= 18 hPa
key: Complete self-test
– Checks compressor run time up to
the point when a specific pressure is reached and thus for leaks in the area of the compressor.
– In the case of two different
pressures, checks for leaks in the pressure accumulator (p the internal tightness of the valves.
) and for
s
– Checks for leaks in the area of
working pressure (pw) in the case of two different pressures.
– Checks behavior during injection
at different pressures (short continuous test).
If certain values are outside the specified range, an error message is emitted. If this recurs, inform Service.
key: Self-test to search for leaks
– Checks for leaks in the pressure
accumulator (p different pressures, as well as for
) in the case of two
s
inner tightness of the valves.
– Checks for leaks in the area of wor-
king pressure (pw) in the case of two different pressures.
If certain values are outside the specified range, an error message is emitted. If this recurs, inform Service.
31
Page 34
5 Troubleshooting

5.2 Warnings

As a rule, warnings are emitted if certain operating conditions are not (yet) reached or if an action is not possible.
Acknowledge the warning with the key and the unit will return to normal status.
WARNING 001
Please wait for
supply pressure
INJECT = Continue
WARNING 002
Check capillary
e.g. broken/missing ?
INJECT = Continue
Warning 1
Supply pressure has not yet been reached.
The minimum supply pressure required for an injection or the Clean function was not present when the function was triggered.
Possible cause: Multiple triggering of Clean or very rapid injection at a high injection pressure.
Solution: Wait until the supply pressure has built up and trigger again.
Warning 2
Compensation pressure cannot be maintained.
Possible cause: Capillary or capillary holder not attached or only connected loosely.
Solution: Check all connections and tighten. Activate Function 0 to change the capillary and when the capillary holder is open.
32
Page 35
5 Troubleshooting
WARNING 003
Switch on manipulator
or disconnect cable
INJECT = Continue
Warning 3
Micromanipulator does not react
A micromanipulator is connected but does not react to a start pulse from the FemtoJet.
Possible cause: Micromanipulator is switched off. INJECT function on manipulator is not released. No limit is set on the micromanipulator. In the INJECT function on the micro­manipulator, synchronization is set to limit (SYNC = LIMIT).
Solution: Switch on the micromanipulator and make it operational. Disconnect the cable to micro­manipulator. With SYNC = LIMIT, trigger semi-auto­matic injection via the micromanipulator only.

5.3 Error messages

Error messages are emitted if certain operating conditions are not reached or the device records a deviation from specified conditions.
Acknowledge the error message with the key and the unit will return to normal status. The measures described under “Solution” should be carried out. If the error message recurs after acknowledgement with the key, inform Service.
33
Page 36
5 Troubleshooting
ERROR 010
Supply pressure
cannot be reached
INJECT = Continue
ERROR 011
Supply pressure is
too high
INJECT = Continue
Error 10
Supply pressure cannot be reached
The compressor runs without the maximum supply pressure being reached.
Possible cause: High compensation pressure when the capillary holder is not attached or is poorly attached.
Technical cause: Compressor does not start up. Leak in compressor or pressure accumulator.
Solution: Check the capillary holder and tube for leaks or tighten up the tube connections. Service.
Error 11
Supply pressure is too high
Supply pressure has exceeded the maximum permissible supply pressure.
Possible cause: supply pressure is not relieved by the relevant standby function. Technical defect.
Solution: Switch off using the key (hold down for approximately two seconds). Service.
34
ERROR 012
Supply pressure is
does not fall
INJECT = Continue
Error 12
Supply pressure does not drop far enough
Possible cause: Technical defect
Solution: Service
Page 37
5 Troubleshooting
ERROR 013
Supply pressure is
does not raise
INJECT = Continue
ERROR 014
Supply pressure
leak is too high
INJECT = Continue
ERROR 015
Supply pressure
leak is too high
INJECT = Continue
ERROR 016
Valve 1 pressure
leak is too high
INJECT = Continue
ERROR 020
Working pressure
cannot be reached
INJECT = Continue
Error 13
Supply pressure does not rise
Possible cause: Technical defect
Solution: Service
Error 14
Massive leak in pressure accumulator
Possible cause: Technical defect
Solution: Service
Error 15
Leak in pressure accumulator
Possible cause: Technical defect
Solution: Service
Error 16
Leak through Valve V1 is too large
Possible cause: Technical defect
Solution: Service
Error 20
Working pressure has not been reached
Compressor runs without supply pressure being reached within a specific time.
Possible cause: Capillary holder not attached or poorly attached. (Broken) capillary with large opening. Technical defect.
Solution: Check the capillary. Check the capillary holder and tube for leaks or tighten up the tube connections. Service.
35
Page 38
5 Troubleshooting
ERROR 021
Inject pressure
cannot be reached
INJECT = Continue
ERROR 022
Compensation pressure
cannot be reached
INJECT = Continue
Error 21
Injection pressure has not been reached
Possible cause: Capillary holder not attached or poorly attached. (Broken) capillary with large opening. Technical defect.
Solution: Check capillary. Check capillary holder and tube for leaks or tighten up tube connections. Service.
Error 22
Compensation pressure has not been reached
Possible cause: Capillary holder not attached or poorly attached. (Broken) capillary with large opening. Technical defect.
Solution: Check capillary. Check capillary holder and tube for leaks or tighten up tube connections. Service.
36
ERROR 023
Inject n pressure
overflow too large
INJECT = Continue
ERROR 024
Working pressure
does not rise
INJECT = Continue
Error 23
Increase in working pressure is smaller than required
Possible cause: Technical defect
Solution: Service
Error 24
No increase in working pressure
Possible cause: Technical defect
Solution: Service
Page 39
5 Troubleshooting
ERROR 025
Working pressure
does not fall
INJECT = Continue
ERROR 026
Working pressure
leak is too high
INJECT = Continue
ERROR 027
Working pressure
leak is too high
INJECT = Continue
Error 25
No drop in working pressure
Possible cause: Technical defect
Solution: Service
Error 26
Working pressure leak rate is too high
Possible cause: Capillary holder not attached or poorly attached. (Broken) capillary with large opening. Technical defect.
Solution: Check capillary. Check capillary holder and tube for leaks or tighten up tube connections. Service.
Error 27
Working pressure leak rate is too high
Possible cause: Capillary holder not attached or poorly attached. (Broken) capillary with large opening. Technical defect.
Solution: Check capillary. Check capillary holder and tube for leaks or tighten up tube connections. Service.
37
Page 40
5 Troubleshooting
ERROR 028
Working pressure
leak is too high
INJECT = Continue
ERROR 030
Sensoroffset supply
is too low
INJECT = Continue
ERROR 031
Sensoroffset supply
is too high
INJECT = Continue
ERROR 032
Sensoroffset work
is too low
INJECT = Continue
Error 28
Working pressure leak rate is too high
Possible cause: Capillary holder not attached or poorly attached. (Broken) capillary with large opening. Technical defect.
Solution: Check capillary. Check capillary holder and tube for leaks or tighten up tube connections. Service.
Error 30
Sensor offset supply pressure is too low
Possible cause: Technical defect
Solution: Service
Error 31
Sensor offset supply pressure is too high
Possible cause: Technical defect
Solution: Service
Error 32
Sensor offset working pressure is too low
Possible cause: Technical defect
Solution: Service
38
ERROR 033
Sensoroffset work
is too high
INJECT = Continue
Error 33
Sensor offset working pressure is too high
Possible cause: Technical defect
Solution: Service
Page 41
5 Troubleshooting
ERROR 034
Val ve 1 offset
is too low
INJECT = Continue
ERROR 035
Val ve 1 offset
is too high
INJECT = Continue
ERROR 036
Val ve 2 offset
is too low
INJECT = Continue
ERROR 037
Val ve 2 offset
is too high
INJECT = Continue
ERROR 040
Self-Calibration
is missing
INJECT = Continue
Error 34
Valve 1 offset is too low
Possible cause: Technical defect
Solution: Service
Error 35
Valve 1 offset is too high
Possible cause: Technical defect
Solution: Service
Error 36
Valve 2 offset is too low
Possible cause: Technical defect
Solution: Service
Error 37
Valve 2 offset is too high
Possible cause: Technical defect
Solution: Service
Error 40
Self-calibration is missing
Possible cause: Technical defect
Solution: Service
ERROR 041
Fullscale Calibration
is missing
INJECT = Continue
Error 41
Full-scale calibration is missing
Possible cause: Technical defect
Solution: Service
39
Page 42

6 Technical data

Voltage/Frequency: 100 V – 240 V ± 10 %, 50 – 60 Hz Set voltage: adapts automatically Fuse for 100 – 240 V: T1.0 A, 250 V (2 pcs.) Power consumption: 50 W Max. current consumption: < 0.5 A Protection class: Overvoltage category: Pollution degree: 2 (IEC 664) Ambient temperature: 15 °C – 35 °C Ambient relative humidity: max. 70 % Weight: 6.2 kg Dimensions: Width: 220 mm
Injection time t
Injection pressure p
Compensation pressure p
Regulating accuracy: 1 % of value set in each case,
Clean rinsing pressure: max. approx. 7,000 hPa (102 psi) RS 232 computer Baud rate: 9,600
interface: Start bit: 1
: 0.0 – 99.9 s
i
:0; 5 – 6,000 hPa (87 psi),
i
I
II
(IEC 61010-1)
Depth: 170 mm Height: 280 mm
in 0.1 s increments
in 1 hPa (0.01 – 0.02 psi) increments
: 0; 5 – 6,000 hPa (87 psi),
c
in 1 hPa (0.01 – 0.02 psi) increments
at least ± 1 hPa
Data bits: 8 Parity: non Stop bit: 2
The device is approved with the sign.
U.S. Pat.No. 4,836,244
Technical specifications subject to change!
40
Page 43

7 Ordering information

7 Ordering information
Order no.
5247 000.013 FemtoJet
100 V – 240 V ± 10 %, 50 – 60 Hz Fully programmable microinjector with integrated pressure supply. For injection of smallest amounts of liquid into living cells. Delivery package: 1 FemtoJet
1 Power cable 1 Universal capillary holder 1 Injection tube, 2 m
5247 617.002 1 Hand control for remote control
10 Femtotips
/ Femtotips II (injection capillaries) 10 Microloader 1 Operating Manual
5247 000.013 Version Europe 5247 000.021 Version USA and Canada 5247 000.030 Version UK/Hongkong 5247 000.048 Version Japan 5247 000.056 Version South America 5247 000.072 Version Australia 5247 000.080 Version China
Accessories (not standard)
5247 623.002 Foot control FemtoJet
5325 620.007 Interface cable PC
for control and for program updates
5181 070.015 Interface cable FemtoJet
for connection to the
InjectMan NI 2
5247 622.006 Interface cable 5171
for connection to the micromanipulator 5171 or InjectMan
5242 952.008 20 Femtotips
5179
5242 957.000 20 Femtotips II 5242 009.012 Filling stand for Femtotips
5242 956.003 200 Microloader 5246 164.004 1 Injection tube, 2 m, for universal capillary holder
41
Page 44
7 Ordering information
Order no.
5176 190.002 Universal capillary holder
(incl. grip head 0, adapter for Femtotips
and spare O-rings) 4910 000.018 Eppendorf Reference 3111 000.122 Eppendorf Research
pipette, variable, 0.5 to 10 µl
pipette, variable, 0.5 to 10 µl
5176 210.003 Grip head 0 for universal capillary holder;
for microcapillaries with an outer diameter
of 1.0 to 1.1 mm 5176 212.006 Grip head 1 for universal capillary holder;
for microcapillaries with an outer diameter
of 1.2 to 1.3 mm 5176 214.009 Grip head 2 for universal capillary holder;
for microcapillaries with an outer diameter
of 1.4 to 1.5 mm 5176 207.002 Grip head 3 for universal capillary holder
for microcapillaries with an outer diameter
of 0.7 to 0.9 mm 5176 196.000 O-ring set
for grip heads 0 to 3 with extraction tool 5176 195.004 Service kit
for grip head and capillary holder,
consisting of O-ring set, adapter for Femtotip
and clamping piece (for 5171, 5179)
Micromanipulation
InjectMan NI 2
Micromanipulator for automatic injection in adherent
cells in combination with FemtoJet
5181 000.017 Version Europe 5181 000.025 Version USA and Canada 5181 000.033 Version UK/Hongkong 5181 000.041 Version Japan 5181 000.076 Version Australia 5181 000.084 Version China
42
Page 45
7 Ordering information
Order no.
Tr ansferMan NK 2 Proportional micromanipulator for microinjection
in suspension cells 5188 000.012 Version Europe 5188 000.020 Version USA and Canada 5188 000.039 Version UK/Hongkong 5188 000.047 Version Japan 5188 000.071 Version Australia 5188 000.080 Version China
PatchMan NP 2
Micromanipulator for electrophysiology or other
applications 5183 000.014 Version Europe 5183 000.022 Version USA and Canada 5183 000.030 Version UK/Hongkong 5183 000.049 Version Japan 5183 000.073 Version Australia 5183 000.081 Version China
5246 621.006 Transmission cable 5246
"Y cable" for connection with the
interface cable 5171 5211 058.007 Foot control for microinjection,
connectable to transmission cable 5246 "Y cable"
43
Page 46

7a Ordering information for North America

Order no.
FemtoJet
100 V – 240 V ± 10 %, 50 – 60 Hz Fully programmable microinjector with integrated pressure supply. For injection of smallest amounts of liquid into living cells. Delivery package: 1 FemtoJet
1 Power cable 1 Universal capillary holder 1 Injection tube, 2 m
920010512 1 Hand control for remote control
10 Femtotips
/ Femtotips II (injection capillaries) 10 Microloader 1 Operating Manual
920010504 Version USA and Canada 5247 000.013 Version Europe 5247 000.030 Version UK 5247 000.048 Version Japan 5247 000.056 Version South America 5247 000.072 Version Australia 5247 000.080 Version China
Accessories (not standard)
920005098 Foot control FemtoJet
015160501 Interface cable PC
for control and for program updates
920005845 Interface cable FemtoJet
for connection to the
InjectMan NI 2
5247 622.006 Interface cable 5171
for connection to the micromanipulator 5171 or InjectMan
930000035 20 Femtotips
5179
930000043 20 Femtotips II 930000051 Filling stand for Femtotips
930001007 200 Microloader 920007431 1 Injection tube, 2 m, for universal capillary holder
44
Page 47
7a Ordering information for North America
Order no.
920007392 Universal capillary holder
920007414 Grip head 0 for universal capillary holder;
910012509 Grip head 1 for universal capillary holder;
910012517 Grip head 2 for universal capillary holder;
920007406 Grip head 3 for universal capillary holder
920005870 O-ring set
920005888 Service kit
(incl. grip head 0, adapter for Femtotips and spare O-rings)
for microcapillaries with an outer diameter of 1.0 to 1.1 mm
for microcapillaries with an outer diameter of 1.2 to 1.3 mm
for microcapillaries with an outer diameter of 1.4 to 1.5 mm
for microcapillaries with an outer diameter of 0.7 to 0.9 mm
for grip heads 0 to 3 with extraction tool
for grip head and capillary holder, consisting of O-ring set, adapter for Femtotip and clamping piece (for 5171, 5179)
Micromanipulation
InjectMan NI 2 Micromanipulator for automatic injection in adherent
cells in combination with FemtoJet 920000029 Version USA and Canada 5181 000.017 Version Europe 5181 000.033 Version UK/Hongkong 5181 000.041 Version Japan 5181 000.076 Version Australia 5181 000.084 Version China
45
Page 48
7a Ordering information for North America
Order no.
Tr ansferMan NK 2 Proportional micromanipulator for microinjection
in suspension cells 920000011 Version Europe 5188 000.020 Version USA and Canada 5188 000.039 Version UK/Hongkong 5188 000.047 Version Japan 5188 000.071 Version Australia 5188 000.080 Version China
PatchMan NP 2
Micromanipulator for electrophysiology or other
applications 920000037 Version USA and Canada 5183 000.014 Version Europe 5183 000.030 Version UK/Hongkong 5183 000.049 Version Japan 5183 000.073 Version Australia 5183 000.081 Version China
920012043 Transmission cable 5246
920005047 Foot control for microinjection,
46
"Y cable" for connection with the
interface cable 5171
connectable to transmission cable 5246 "Y cable"
Page 49
8

Index

A
Adapter for Femtotips 18 Application 3, 4
C
Capillary change 27 Clean 8
pressure 20
function 22 Cleaning 29 Commands 46 Compressor 21, 23 Compensation pressure pc 5, 20,
21 Connection
capillary holder 18
micromanipulator 25
universal capillary holder 18 Constant
pressure 23, 27
working pressure 23
working pressure p
20
w
D
Disinfection 29 Display 6, 8, 12
in hPa 12
in psi 12
switch on/off 27
E
Error
messages 33
potential 30
F
Foot control 7, 17 Functions 26
keys 8
menu 15
G
Grip head 18, 19
H
Hand control 17
I
Information on working practices
20
Injection
automatic 14, 23 first 24 manual 14, 23 movement 14, 23 parameters 12 pressure pi 5, 14, 20, 21 time t
5, 14, 22
i
trial 24 tube 9, 18
Interface 10, 25
description 45
K
Key
Auto/Manu 8, 14 Clean 8, 14, 22 Count 8, 15 Inject 5, 8, 14, 32 Menu 8, 26 Standby 8, 16
Keypad 6, 14
L
Loudspeaker, switch on/off 27
M
Mains switch 10 Menu 15 Mode of operation 12
O
Operation 12
via computer 44, 46 Ordering information 41 Overview 6
47
Page 50
8 Index
P
Pressure
– dewatering pressure accumulator 28 – set pressure unit 26
Q
Query 45
R
Reset 9, 17
function 16
S
Safety precautions 3, 4 Self-test 30
complete 31
to find leaks 31 Servicing 29 Short description 8 Startup 10 Standby without deairing 16 Switch
on 10
off 11 System pressure ps 20
T
Technical data 40 Tension piece 19 Troubleshooting 30 Tube connection 6, 9
U
Unit
description 5
rear of 10 Universal capillary holder 18
V
Variable regulator 5, 6, 9, 13, 26
W
Warnings 32
48
Page 51

Appendix A

Operation via a computer
Certain commands and queries (Section 6.2) can be transmitted to the
unit using the optional cable/PC interface.

A.1 Description of interface

A computer can be connected to the remote control of the FemtoJet via
one of the two RS 232 interfaces. Parameters can be entered or read out
via the computer.
The computer is connected by the special cable/PC interface cable.
The computer to be connected must comply with EN 60950/UL 1950
standards.
The serial interface of the computer should be set to the following
transmission parameters:
Baud rate: 9,600
Start bit: 1
Dat bits: 8
Parity: non
Stop bit: 2
Only ASCII symbols (incl. CR/LF) which can be represented are
transmitted.Therefore a terminal program, for example, is suitable for
remote control.

A.2 Commands and queries

The commands to be sent are defined with the following structure:
Commands to enter parameters: "Cxxx=ssssssssss"
Commands to execute and to read out parameters: "Cxxx"
C: Command
xxx: Code for the executing command
=: Separation for the following input parameters
ssssssssss: Input parameters
In response to any command, the device provides a symbol string of a
similar structure:
"Axxx" for answers without output parameter
"Axxx=rrrrrrrrrr" for responses with output parameter
A: Answer
49
Page 52
Appendix A
If the command is faulty, an error message is returned instead of the answer:
"ERR=eee" ERR: Error =: Separation eee: Error number
Lower and upper case letters can be used.
Command to enter parameters
Short acoustic signal 003 1 – 254 Number of signals
Long acoustic signal 004 1 – 254 Number of signals
Automatic injection (inject at t
Set injection counter n to 0
Perform Clean 013 1 – 100 Time in 1/10 seconds
Execute Standby 014 0 or 1 0 = short
Switch acoustic signal on/off
Set pressure unit 041 0 or 1 Unit:
Set injection time ti 042 1 – 999 Injection time
)
i
Code
Input parameter
[C]
[xxx]
[ssssssssss]
010 0; 1 – 999 Injection time
012 – – – No parameter input
040 0 or 1 Status:
Meaning of input parameters
in 1/10 seconds 0: injection time set on the FemtoJet
1 = long
0 = off / 1 = on
0 = hPa / 1 = psi
in 1/10 seconds
50
Page 53
Appendix A
Command to enter parameters
Set compensation pressure pc
Set injection pressure pi 044 0; 5 – 6,000 /
Set background illumination
Code
Input parameter
[C]
[xxx]
[ssssssssss]
043 0; 5 – 6,000 /
87
87
045 0 or 1 0 = off
Meaning of input parameter
Pressure in hPa/psi
Pressure in hPa/psi
1 = on
Dewater 046 – – – No parameter input
Command to read out parameters
[C]
Code
Significance of output parameter
[xxx]
[A]
Output parameter [rrrrrrrrrr]
Query program version 101 Versions number x.xx
Query bleeper 150 Status:
0 or 1
0 = off / 1 = on
Query pressure unit 151 Unit:
0 or 1
0 = hPa / 1 = psi
Query injection time t
152 Injection time
i
in 1/10 seconds
1 – 999
Query compensation pressure p
c
Query injection pressure p
i
Query background illumination
153 Pressure in hPa/psi 0 – 6,000
154 Pressure in hPa/psi 0 – 6,000
155 Status:
0 or 1
0 = off / 1 = on
/ 87
/ 87
51
Page 54
Appendix A
Error no. for incorrect operating construction ERR = [eee]
060 The first character is not
061 The 2nd to 4th character
062 The 5th character is not =. Check entry parameters.
063 The parameter is not
064 The command is not
Cause Solution
a c or C.
is not a digit.
within the permitted range.
known.
Check entry parameters. Change entry.
Check entry parameters. Change entry.
Change entry.
Check entry parameters. Change entry.
Check entry parameters. Change entry.
52
Page 55
Appendix A
Error no.
Cause Solution during execution ERR = [eee]
070 FemtoJet is not in routine
mode.
Check FemtoJet. Bring FemtoJet
into routine operation. Eliminate error.
071 FemtoJet
is not ready. Check FemtoJet or
micromanipulator. As necessary, wait or eliminate error.
072 The command is not finished
within the scheduled time.
Check FemtoJet micromanipulator.
or
Eliminate error.
073 This command cannot be
executed at the moment,
Check FemtoJet Eliminate error.
.
because of insufficient storage pressure. See: Warning 001
074 The connected micro-
manipulator did not respond. See: Warning 003
Check micro­manipulator. Eliminate error.
Other error numbers as well as the error cause and solution are described in Section 5, "Troubleshooting". During remote control, no error texts appear in the FemtoJet
display.
53
Page 56
Appendix A
54
Page 57
Loading...