Eppendorf CellTram vario User Manual

Page 1
Operating manual
CellTram® Air · Oil · varioseeon p.Fig.Tab.p
O
l
.
perating manua
5176 900.018
CellTram® Air CellTram® Oil
CellTram® vario

Operating manual

Page 2
Copyright© 2012 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior permission of the copyright owner.
CellTram, Eppendorf and TransferTip are registered trademarks of Eppendorf AG, Hamburg, Germany.
Registered trademarks are not marked in all cases with ™ or ® in this manual.
5176 900.018-16/102012
Page 3
CellTram® Air · Oil · vario — Operating manual

Table of contents

1 Table of contents
1 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Using this manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Danger symbols and danger levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.1 Danger symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2.2 Danger levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3 Symbols used. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1 Main illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1.1 CellTramAir . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1.2 CellTramOil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.3 CellTramvario . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.4 Universal capillary holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1.5 Grip head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1.6 Adapter for Femtotips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.1 CellTramAir . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2 CellTramOil/CellTramvario . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 Delivery package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3.1 CellTram Air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3.2 CellTram Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3.3 CellTram vario . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.1 User profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.2 Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3 Warnings for intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4 Safety instructions on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.1 Selecting the location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.2 Mounting the CellTram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.3 Filling oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3.1 Heating the syringe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.3.2 Connecting the filling tube. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.3.3 Filling the piston chamber and universal capillary holder with oil . . . . . . . . . . 16
4.3.4 Sealing the filling bore. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3.5 Removing air bubbles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.4 Mounting the capillary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.5 Setting the angle of the CellTram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3
Page 4
CellTram® Air · Oil · vario — Operating manual
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.1 Holding and transferringcells. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.2 Suctioning liquid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.1 Replacing O-rings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.3 Disinfection/decontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.1 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.1.1 CellTram Air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.1.2 CellTram Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.1.3 CellTram vario . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.2 Weight/dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.2.1 CellTram Air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.2.2 CellTram Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.2.3 CellTram vario . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.2.4 Pressure tubing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.3 Application parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.3.1 CellTram Air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.3.2 CellTram Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.3.3 CellTram vario . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
8 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
8.1 CellTram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
8.2 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
8.3 Capillaries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
9 Transport, storage and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
9.1 Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
9.2 Decontamination before shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4
Page 5
CellTram® Air · Oil · vario — Operating manual

1 Operating instructions

1 Operating instructions

1.1 Using this manual

Read this operating manual completely before using the device for the first time. Please also
note the operating instructions for the accessories, if applicable.
This operating manual is part of the product. Please keep it in a place that is easily
accessible.
Enclose this operating manual when transferring the device to third parties.If this manual is lost, please request another one. You will find the current version on our
webpage www.eppendorf.com/worldwide.

1.2 Danger symbols and danger levels

The safety precautions in these instructions have the following danger symbols and danger levels:

1.2.1 Danger symbols

Hazard point Cuts
Material damage

1.2.2 Danger levels

DANGER Will lead to severe injuries or death.
WARNING Can lead to severe injuries or death.
CAUTION Can lead to light to moderate injuries.
ATTENTION May lead to material damage.

1.3 Symbols used

Symbol Meaning
Handling
1.
2.
Actions in the specified order
List
Text Name of fields in the software
Useful information
5
Page 6
CellTram® Air · Oil · vario — Operating manual

2 Product description

2 Product description

2.1 Main illustration

2.1.1 CellTramAir
Abb. 1:Side view
1 2 3 4 5 6 7 8
Fig. 1: Side view
1 Rotary knob 2 Scale (from 0 to 9)
3 Tilting joint 4 Piston window
5 Foot 6 Ventilation screw
7 Barb screw 8 Pressure tubing
Abb. 2:Top view
1 2
15
10
0
5
Fig. 2: Top view
1 Position indicator 2 Piston window with scale range
6
Page 7
CellTram® Air · Oil · vario — Operating manual
2.1.2 CellTramOil
Abb. 3:Side view
1
2 3 4 5 6 7 8
9
10
Fig. 3: Side view
1 Pressure tubing 2 Barb screw
3 Cylinder body (transparent) 4 Filling bore with filling screw
5 Piston (in initial position) 6 Housing
7 Scale 8 Rotary knob
9 Tilting joint 10 Foot
2.1.3 CellTramvario
Abb. 4:Side view
1
2 3 4 5 6 7 8 9
10
11
Fig. 4: Side view
1 Pressure tubing 2 Barb screw
3 Cylinder body (transparent) 4 Filling bore with filling screw
5 Piston (in initial position) 6 Housing
7 Rotary knob, coarse setting 8 Scale
9 Rotary knob, fine setting 10 Tilting joint
11 Foot
7
Page 8
CellTram® Air · Oil · vario — Operating manual

2.1.4 Universal capillary holder

Abb. 5:Universal capillary holder
1 2 3 4 5
Fig. 5: Universal capillary holder
1 Barb 2 Rear knurled screw
3 Tension piece 4 Front knurled screw
5 Grip head0 for capillaries with an outer
diameter of 1.0 to 1.1mm

2.1.5 Grip head

Abb. 6:Grip head 0 with O-rings and sealing washer
Fig. 6: Grip head 0 with O-rings and sealing washer
The transparent grip head allows the user to visually monitor the position of the capillary. The capillary is fixed in the universal capillary holder via the grip head.

2.1.6 Adapter for Femtotips

Abb. 7:Adapter for Femtotips with spare O-rings
Fig. 7: Adapter for Femtotips with spare O-rings
8
Page 9
CellTram® Air · Oil · vario — Operating manual

2.2 Features

The CellTramAir, CellTramOil and CellTramvario are manual piston pumps. Using a movable piston, the CellTram generates differences in pressure which are transferred directly to a capillary via a pressure tube. The piston is moved in the CellTram using the dial. A clockwise rotation, i.e. to the right, results in an increase in pressure, the capillary contents are pressed out. A counterclockwise rotation, i.e. to the left, results in a decrease in pressure, aspiration into the capillary takes place. To assist fine pressure adjustment, the dial has a scale.
The capillary is inserted into a capillary holder which in turn provides the connection to the pressure tube. The universal capillary holder enables permanent mounting of various capillaries in the laboratory.
The CellTramAir, CellTramOil and CellTramvario have a very ergonomic design which enables easy handling, particularly in conjunction with all control units of Eppendorf micromanipulators.
2.2.1 CellTramAir
The CellTramAir uses air for pressure transmission. This guarantees clean and simple handling of the piston pump.
A complete revolution of the dial corresponds to a change in volume of approx. 88L. The position of the piston is displayed in the piston window on a scale from 0 to 15mm
corresponding to the strokes.
2.2.2 CellTramOil/CellTramvario
The CellTramOil and CellTramvario use oil for pressure transmission which, unlike air, can hardly be compressed. This ensures its operating cycles are very controlled and constant. The simple filling technique reduces the filling time and simplifies oil filling. To ensure controlled, constant pressure transmission, the cylinder, pressure tube and part of the capillary are filled with oil. A complete revolution of the dial (coarse drive) corresponds to a change in volume of approx.
9.6L. The CellTramvario has an additional fine drive dial. The transmission ratio between fine drive
and coarse drive is 10:1. A revolution of the fine drive corresponds to a change in volume of approx. 0.96L. This enables ultra-fine cell movements.
The position of the piston can be observed in the transparent cylinder. The filling bore with the plug screw is at the top of the cylinder. The pressure tube is connected to the CellTramOil/ CellTramvario with the aid of the tube connection screw. The pressure tube has a joint at both ends and thus connects the CellTramOil/CellTramvario and the universal capillary holder.
9
Page 10
CellTram® Air · Oil · vario — Operating manual

2.3 Delivery package

2.3.1 CellTram Air

QuantityOrder no.
(international)
1 5176 000.017 CellTram Air
1 5176 114.004
1 5176 190.002
1- Allen key (4mm)
1 5176 900.018 Operating Manual CellTram Air/oil/vario

2.3.2 CellTram Oil

QuantityOrder no.
(international)
1 5176 000.025 CellTram Oil
1 5176 114.004
1 5176 190.002
1-
1-
1- Allen key (4mm)
1 5176 900.018 Operating Manual CellTram Air/oil/vario

2.3.3 CellTram vario

QuantityOrder no.
(international)
1 5176 000.033 CellTram vario
1 5176 114.004
1 5176 190.002
Description
Pressure tube
1m
Universal capillary holder
For holding Femtotips, Femtotips II or grip heads Including grip head 0, adapter for Femtotips and spare O-rings
Description
Pressure tube
1m
Universal capillary holder
For holding Femtotips, Femtotips II or grip heads Including grip head 0, adapter for Femtotips and spare O-rings
Quick valve system
For simple filling of the CellTram Consists of a premounted filling screw, disposable syringe and filling tube with a Luer Lock connector
Bottle with mineral oil
25mL
Description
Pressure tube
1m
Universal capillary holder
For holding Femtotips, Femtotips II or grip heads Including grip head 0, adapter for Femtotips and spare O-rings
10
Page 11
CellTram® Air · Oil · vario — Operating manual
QuantityOrder no.
(international)
1-
1-
1- Allen key (4mm)
1 5176 900.018 Operating Manual CellTram Air/oil/vario
Description
Quick valve system
For simple filling of the CellTram Consists of a premounted filling screw, disposable syringe and filling tube with a Luer Lock connector
Bottle with mineral oil
25mL
11
Page 12
CellTram® Air · Oil · vario — Operating manual

3 Safety

3 Safety

3.1 User profile

The device and accessories may only be operated by trained and skilled personnel. Before using the device, read the operating manual carefully and familiarize yourself with the
device's mode of operation.

3.2 Intended use

The CellTram has been designed and manufactured for use in the field of Intracytoplasmic sperm injection (ICSI) in conjunction with Eppendorf VacuTips and Transfertips. The CellTram is thus considered a medical device within the scope of Directive 93/42/EEC.
The CellTram is intended exclusively for indoor use, and for operation by qualified staff.

3.3 Warnings for intended use

WARNING! Risk of injury due to flying capillaries and glass splinters.
If exposed to high pressures, capillaries can detach themselves from the grip heads and
WARNING!
become projectiles. Capillaries can crack as a result of incorrect handling.
Wear protective goggles.Never aim capillaries at people.Use capillaries with an outer diameter that matches the grip head specifications.Always mount/dismount capillaries when they are depressurized.Mount the capillary correctly in the grip head.Do not touch the capillary with the Petri dish or other objects.
12
WARNING! Damage to health from infectious liquids and pathogenic germs.
When handling infectious liquids and pathogenic germs, observe the national
WARNING!
regulations, the biological security level of your laboratory, the material safety data sheets, and the manufacturer's application notes.
Wear personal protective equipment.Comprehensive regulations on handling germs or biological material in risk group II
or higher can be found in the "Laboratory Biosafety Manual" (source: World Health Organization, Laboratory Biosafety Manual, in the currently valid version).
WARNING! Damage to health due to toxic, radioactive or aggressive chemicals.
Wear personal protective equipment.
WARNING!
Observe the national regulations for handling these substances. Observe the material safety data sheets and manufacturer's application notes.
Page 13
CellTram® Air · Oil · vario — Operating manual
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf
CAUTION!
may impair the safety, functioning and precision of the device. Eppendorf cannot be held liable or accept any liability for damage resulting from the use of incorrect or non-recommended accessories and spare parts, or from the improper use of such equipment.
Only use accessories and original spare parts recommended by Eppendorf.

3.4 Safety instructions on the device

Depiction Meaning
WARNING
General hazard point. Follow the operating manual.
13
Page 14
CellTram® Air · Oil · vario — Operating manual

4 Installation

4 Installation

4.1 Selecting the location

Please select the location according to the following criteria:
Stable
Level
Horizontal
Vibration-damped

4.2 Mounting the CellTram

Valid for:
CellTram Air
CellTram Oil
CellTram vario
Abb. 8:ExampleCellTram Air
1 2 3
Fig. 8: ExampleCellTram Air
1 Pressure tubing 2 Universal capillary holder
For holding Eppendorf capillaries (e.g., TransferTips, VacuTips)
3 Grip head0
For capillaries with an outer diameter of
1.0mm to 1.1mm
1. Screw the pressure tubing on to the CellTram.
2. Screw the pressure tubing on to the universal capillary holder.
14
Page 15
CellTram® Air · Oil · vario — Operating manual

4.3 Filling oil

Valid for:
CellTram Oil
CellTram vario
Work with an air cushion in the visible area of the capillary in order to avoid contact
Hint!
between the oil and the medium.
Observe the legal regulations for purity of the oil. The supplied oil is generally used in
Hint!
the ICSI area to coat drops containing embryos/ egg cells or sperm. The supplied oil is not sterile. Additional instructions can be found in the safety data sheet of the manufacturer.
Aspirating oil with the syringe may lead to the formation of tiny air bubbles. This should
Hint!
be avoided as these air bubbles may be transferred to the CellTram, which may impair or slow down the filling process. If the syringe contains air bubbles, these should be collected by long-term storage. Large air bubbles can be pressed out during filling.
Abb. 9:CellTram OilandCellTram vario
1 2 3 4 5 6
Fig. 9: CellTram OilandCellTram vario
1 Initial position of piston 2 Filling bore with plug screw
3 Final position of piston 4 Barb screw
5 Pressure tubing connection to the rear
6 Grip head 0 for capillary
knurled screw
15
Page 16
CellTram® Air · Oil · vario — Operating manual

4.3.1 Heating the syringe

1. Screw the filling hose on to the tip.
2. Aspirate at least 3mL of oil into the syringe.

4.3.2 Connecting the filling tube

1. Turn the rotary knob to move the piston to the initial position.
2. Unscrew the blind screw.
3. Screw the filling tube in the filling bore.

4.3.3 Filling the piston chamber and universal capillary holder with oil

Prerequisites
The piston is in the starting position.
1. Hold the CellTram vertically.
2. Push some oil into the piston chamber. During the first filling, air bubbles may be
present between the piston and the cylinder.
16
3. Allow these bubbles to rise by moving the rotary knob or coarse drive back and forth.
Page 17
CellTram® Air · Oil · vario — Operating manual
4. Completely fill the piston chamber with oil. Make sure that there are no air bubbles in the tip of the piston chamber.
5. Completely fill the universal capillary holder with oil.
6. Visually check that no air bubbles remain in the universal capillary holder.

4.3.4 Sealing the filling bore

1. Position the CellTram horizontally.
2. Unscrew the filling tube.
3. Fill the filling bore with oil.
17
Page 18
CellTram® Air · Oil · vario — Operating manual
4. Screw on the blind screw. When doing so, make sure that no new air bubbles enter the bore.
5. Soak up any excess oil using a soft cellulose cloth.

4.3.5 Removing air bubbles

1. If there are still air bubbles in the piston chamber: slant the CellTram to allow air bubbles to escape to the filling bore. Remove the air bubbles.

4.4 Mounting the capillary

WARNING! Risk of injury due to flying capillaries and glass splinters.
If exposed to high pressures, capillaries can detach themselves from the grip heads and
WARNING!
become projectiles. Capillaries can crack as a result of incorrect handling.
Wear protective goggles.Never aim capillaries at people.Use capillaries with an outer diameter that matches the grip head specifications.Always mount/dismount capillaries when they are depressurized.Mount the capillary correctly in the grip head.Do not touch the capillary with the Petri dish or other objects.
18
CAUTION! Risk of cuts when unpacking the capillaries
Capillaries can break as a result of incorrect unpacking.
CAUTION!
Do not reach into the capillary transport protection.
Grip head 0 is only for capillaries with an outer diameter of 1.0mm and 1.1mm. Other
Hint!
capillaries may only be used with the corresponding grip head (see Ordering information) (see Accessories on p. 27).
Page 19
CellTram® Air · Oil · vario — Operating manual
1. Slightly screw the grip head into the universal capillary holder.
2. Slide the capillary through the inner O-rings up to the stop into the universal capillary holder.
No stop is felt with capillaries with an inner diameter <0.8mm.
Hint!
3. Tighten the grip head.
4. In the case of CellTram Oil and CellTram vario, carefully wipe any excess oil off of the universal capillary holder and the capillary using a soft cellulose cloth if necessary.
When using Eppendorf Femtotips
1. Screw the Femtotip adapter into the universal capillary holder.
2. Screw the Femtotip into the Femtotip adapter.

4.5 Setting the angle of the CellTram

1. Loosen the fastening screw in the tilting joint with an Allen key (4mm).
2. Set the CellTram via the tilting joint.
3. Tighten the fastening screw in the tilting joint with an Allen key (4mm).
19
Page 20
CellTram® Air · Oil · vario — Operating manual

5 Operation

5 Operation
5.1 Holding and transferringcells
CellTramAir:
Hint!
An overpressure or vacuum can be built up in the capillary using the volume of the piston chamber (cylinder) and the pressure tube. Therefore, the pressure behavior varies depending on the position of the piston. If the piston is located, e.g., in position 10to 15, the differences in pressure are greater than, e.g., in position 0 to 5. By carefully selecting the position of the piston, the sensitivity of the CellTramAir piston pump can be adapted to your requirements.
1. Fix the universal capillary holder into the tool holder of the Eppendorfmicromanipulator.
2. Determine the correct length of the capillary by moving the universal capillary holder in the tool holder.
3. Determine the correct rotation of the capillary by turning the front knurled screw.
4. Move the micromanipulator upwards (see Operating Manual of the micromanipulator).
5. Position the petri dish.
6. Move the capillary into the medium and ensure its correct positioning by observing it under a microscope.
Establishing pressure equilibrium in the capillary
7. You have three options:
Wait until equilibrium has been established.
Turn the piston clockwise (i.e. to the right) to establish a counter pressure which
compensates the active capillary force.
Only CellTramAir: loosen the ventilation screw. Tighten the screw again after pressure
compensation.
8. Move the capillary near to the desired cell.
9. Generate a vacuum by turning the dial counterclockwise (to the left).
In hold mode:
Hint!
Excessive constant pressure can damage the cell. After the cell has reached the capillary tip, an optimal constant pressure should be set.
In transfer mode:
Hint!
Ensure that pressure is equilibrated sensitively and evenly during cell aspiration into the capillary.
10. While the cell moves towards the capillary: it may be necessary to turn the dial clockwise (to the right) to slightly equilibrate the vacuum generated in order that the cell is not fixed too firmly to the capillary tip.
20
Page 21
CellTram® Air · Oil · vario — Operating manual

5.2 Suctioning liquid

Valid for:
CellTramAir
1. Dip the capillary into the liquid that is to be aspirated.
2. Turn the rotary knob counterclockwise to the left (by approx. 2revolutions) to aspirate the liquid by creating a low pressure.
3. Wait for a short period and then loosen the ventilation screw.
4. Retighten the ventilation screw after pressure compensation.
21
Page 22
CellTram® Air · Oil · vario — Operating manual

6 Maintenance

6 Maintenance

6.1 Replacing O-rings

If the O-rings on the universal capillary holder or on the grip head are damaged, they can be replaced.
1. Remove the O-rings and the sealing washer with the extraction tool.
2. Insert the 2 new O-rings and a new sealing washer in the correct order.
3. Make sure that the 2 sealing washers and 2 matching O-rings are correctly arranged in the grip head.

6.2 Cleaning

NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as
NOTICE!
Hint!
strong and weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.
If the device becomes contaminated with aggressive chemicals, clean it
immediately using a mild cleaning agent.
Choose the disinfection method which corresponds to the legal regulations and guidelines in place for your range of application.
If you have any questions on cleaning, disinfection, decontamination, or the cleaning agents to be used, please contact Eppendorf AG Application Support. The relevant contact details are provided on the back of this operating manual.
1. Clean the CellTram, pressure tube and universal capillary holder with water or a mild lab detergent.
2. To remove residual liquid, blow out the pressure tube and universal capillary holder with compressed air after cleaning.
There must not be any residual liquid remaining in the pressure tube or in the universal
Hint!
capillary holder after cleaning as otherwise steady pressure behavior will be compromised.

6.3 Disinfection/decontamination

NOTICE! Damage from UV and other high-energy radiation.
Do not use UV, beta, gamma, or any other high-energy radiation for disinfecting.
NOTICE!
Avoid storage in areas with strong UV radiation
NOTICE! Damage to the device due to autoclaving.
Do not autoclave the CellTram.
NOTICE!
22
Page 23
CellTram® Air · Oil · vario — Operating manual
Required equipment
Alcohol (ethanol, isopropanol) or disinfectants containing alcohol
Proceed as follows:
1. Choose the disinfection method which corresponds to the legal regulations and guidelines in place for your range of application.
2. Wipe all surfaces of the CellTram and accessories using the disinfectant.
3. Flush the pressure tube and universal capillary holder with the disinfectant if required.
4. To remove residual liquid, blow out the pressure tube and universal capillary holder with compressed air after cleaning.
There must not be any residual liquid remaining in the pressure tube or in the universal
Hint!
capillary holder after cleaning as otherwise steady pressure behavior will be compromised.
23
Page 24
CellTram® Air · Oil · vario — Operating manual

7 Technical data

7 Technical data

7.1 Ambient conditions

7.1.1 CellTram Air

Ambience For indoor use only
Ambient temperature max. 40°C
Max. relative humidity Max. 80%
Maximum pressure 2900hPa

7.1.2 CellTram Oil

Ambience For indoor use only
Ambient temperature max. 40°C
Max. relative humidity Max. 80%
Maximum pressure Approx. 20000hPa

7.1.3 CellTram vario

Ambience For indoor use only
Ambient temperature max. 40°C
Max. relative humidity Max. 80%
Maximum pressure Approx. 20000hPa
24
Page 25
CellTram® Air · Oil · vario — Operating manual

7.2 Weight/dimensions

7.2.1 CellTram Air

Footprint (w/o pressure tubing) 175mm×60mm
Weight Approx. 1.4kg
Rotary knob diameter 40mm

7.2.2 CellTram Oil

Footprint (w/o pressure tubing) 275mm×60mm
Weight Approx. 1.7kg
Rotary knob diameter 40mm

7.2.3 CellTram vario

Footprint (w/o pressure tubing) 29.5x6cm
Weight Approx. 1.7kg
Rotary knob diameter (coarse) 42mm
Rotary knob diameter (fine) 34mm

7.2.4 Pressure tubing

Material Polyethylene
Length 1m
Inner diameter 1mm
25
Page 26
CellTram® Air · Oil · vario — Operating manual

7.3 Application parameters

7.3.1 CellTram Air

Change in volume (per rotary knob revolution):
Minimum adjustment volume < 0.2L
Adjustable total volume 2.640L
Entire stroke of the piston 15mm (corresponds to 30 revolutions)
Generation of pressure Using piston and cylinder

7.3.2 CellTram Oil

Change in volume (per rotary knob revolution):
Minimum adjustment volume: <0.02L
Adjustable total volume 960L
Total filling volume (incl. 1m tubing and universal
capillary holder)
Entire stroke of the piston 50mm (corresponds to 100 revolutions)
Generation of pressure Using piston and cylinder

7.3.3 CellTram vario

Change in volume (per rotary knob revolution)
Minimum adjustment volume Coarse: <0.02L
Adjustable total volume 960L
Total filling volume (incl. 1m tubing and universal
capillary holder)
Entire stroke of the piston 50mm (corresponds to 100 revolutions)
88L
9,6L
<2mL
Coarse: 9.6L Fine: 0.96L
Fine: <0.002L
<2mL
Generation of pressure Using piston and cylinder
26
Page 27
CellTram® Air · Oil · vario — Operating manual

8 Ordering information

8 Ordering information

8.1 CellTram

Order no. (international)
5176 000.017 CellTram Air
5176 000.025 CellTram Oil
5176 000.033 CellTram vario
Description

8.2 Accessories

Order no. (international)
5176 114.004
5176 190.002
5176 210.003
5176 212.006
5176 214.009
5176 207.002
5176 196.000
5176 195.004
5176 120.004
5176 220.009
Description
Pressure tube
1m
Universal capillary holder
For holding Femtotips, Femtotips II or grip heads Including grip head 0, adapter for Femtotips and spare O-rings
Grip head 0
For microcapillaries with an outer diameter of 1.0mm to 1.1mm, setof2
Grip head 1
For microcapillaries with an outer diameter of 1.2mm to 1.3mm, setof2
Grip head 2
For microcapillaries with an outer diameter of 1.4mm to 1.5mm, setof2
Grip head 3
For microcapillaries with an outer diameter of 0.7mm to 0.9mm, setof2
O-ring set
For grip heads 0-3 with removal tool
Service kit
For grip heads and capillary holder Consisting of O-ring set, adapter for Femtotips
Filling hose for Quick Valve-system
With a Luer Lock connector (female), 12cm tube coupling for pressure tube for extending/connecting 2pressure tubes
Tube coupling
For extending or connecting pressure tubes
27
Page 28
CellTram® Air · Oil · vario — Operating manual

8.3 Capillaries

Various capillaries for sperm injection and the containment of cells are available. VacuTips have been optimized for holding suspension cells (e.g., oocytes or blastocysts). TransferTips have been especially developed for transferring sperm in ICSI applications.
Microcapillary Technical data
VacuTip
Microcapillary for holding suspension cells.
TransferTip RP (ICSI)
Microcapillary for injecting human sperm (ICSI).
TransferTip F (ICSI)
Microcapillary for injecting human sperm (ICSI).
TransferTip R (ICSI)
Microcapillary for injecting human sperm (ICSI).
Sterile
15m inner diameter
100m outer diameter
35° capillary angle
1mm flange
Sterile
4m inner diameter
7m outer diameter
35° capillary angle
Rigid 0.5mm parallel flange
Sterile
4m inner diameter
7m outer diameter
35° capillary angle
Flexible 0.4mm flange
Sterile
4m inner diameter
7m outer diameter
35° capillary angle
Rigid 1mm flange
Order no. (international)
5175 108.000
5175 114.000
5175 106.008
5175 113.004
28
Description
VacuTip
25pieces, sterile
TransferTip-RP (ICSI)
25pieces, sterile
TransferTip-F (ICSI)
25pieces, sterile
TransferTip-R (ICSI)
25pieces, sterile
Page 29
CellTram® Air · Oil · vario — Operating manual

9 Transport, storage and disposal

9 Transport, storage and disposal

9.1 Storage

Air temperature Relative humidity Atmospheric
In transport packaging -25°C– 55°C 10%– 95%70kPa– 106kPa
Without transport packaging
-5°C– 45°C 10%– 95%70kPa– 106kPa
pressure

9.2 Decontamination before shipment

If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer for disposal please note the following:
WARNING! Risk to health from contaminated device
WARNING!
1. Please observe the information in the decontamination certificate, which is available on our webpage as a PDF (www.eppendorf.com/decontamination).
2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.
29
Page 30
Page 31
Eppendorf AG · 22331 Hamburg · Germany · Tel: +49 40 53801-0 · Fax: +49 40 538 01-556 · E-mail: eppendorf@eppendorf.com
Application Support Europe: Tel: +49 1803 666 789 (Preis je nach Tarif im Ausland; 9 ct/min aus dem dt. Festnetz; Mobilfunkhöchstpreis 42 ct/min)
Eppendorf North America, Inc. · 102 Motor Parkway · Hauppauge, N.Y. 11788-5178 · USA
Tel: +1 516 334 7500 · Toll free phone: +1 800-645-3050 · Fax: +1 516 334 7506 · E-mail: info@eppendorf.com
Your local distributor: www.eppendor f.com/worldwide
North America: Tel: +1 800 645 3050 · E-mail: techserv@eppendorf.com
Asia Pacific: Tel: +60 3 8023 6869 · E-mail: support_asiapacific@eppendorf.com
support@eppendorf.com
Loading...