Druck DPI 812 Operating Manual

Page 1
GE Infrastructure
Sensing
Druck DPI 811/812
RTD calibrator and RTD loop calibrator
User manual - K345
Page 2
A1
B1
A2
10
1
A2
10
9
A3
8
2
3
4
7
6
5
11
B1
12
K345 Issue 2
1513
14
A3
17182619 2716
20
21
22 23
24
25
Page 3

Table of Contents

Introduction ......................................................................... 1
Safety ..................................................................................... 1
Safety - Marks and symbols on the instrument ............... 2
To start .................................................................................. 2
To start - Location of items ........................................................ 2
To start - Items on the display .................................................. 2
To start - Prepare the instrument ........................................... 2
To start - Power on or off ............................................................ 3
To start - Set up the basic operation ..................................... 3
To start - Select a task (Measure and/or simulate) ......... 3
To start - Set up the settings ..................................................... 4
To start - Edit functions ................................................................ 4
Operation .............................................................................. 6
Operation - RTD connections .................................................... 6
Operation - Communications port connections .............. 6
Operation - Measure RTD values ............................................. 6
Operation - Change the output values ................................. 6
Operation - Simulate RTD values ............................................ 7
Operation - Transmitter calibration ....................................... 7
Operation - mA measurements ............................................... 8
Operation - Switch test ................................................................ 8
Operation - UPM Pressure measurements ......................... 9
Operation - Error indications ..................................................... 9
Maintenance ..................................................................... 10
Maintenance – Clean the unit ................................................ 10
Maintenance – Replace the batteries ................................ 10
Calibration ......................................................................... 10
Calibration - Before you start ................................................. 10
Calibration - Procedures (Ohms input) ............................... 10
Calibration - Procedures (Ohms output) ........................... 11
Calibration - Procedures (mA input) .................................... 11
Calibration - Procedures (IDOS UMM) ................................. 11
Specification data ............................................................ 12
Specification - General .............................................................. 12
Specification - Temperature ranges ................................... 12
Specification - Resistance ranges ........................................ 12
Specification - Electrical connectors (A2) ......................... 12
Customer service ............................................... Back cover
© 2005 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.

Introduction

The DPI 811 RTD (Resistance Temperature Detector) Calibrator and DPI 812 RTD Loop Calibrator are part of the Druck DPI 800 series of hand held instruments. The DPI 800 series uses Intelligent Digital Output Sensor (IDOS) technology to give instant plug and play functionality with a range of Universal Measurement Modules (UMM). Example: the Universal Pressure Module (UPM). The DPI 811/812 include these functions:
Function DPI 811 DPI 812
Measure/simulate RTD temperature or resistance
Step/Ramp functions Automatic/Manual Communications port IDOS or RS232 Language selection Yes Measure pressure/Leak test ** External IDOS UPM ** Snapshot Up to 1000 displays with a
Measure mA No 0 - 55 mA HART® resistor No Yes V dc output No 24 V Switch test No Yes Other functions Hold, Maximum/Minimum/Average,
Filter, Tare, Scaled values, Backlight, Alarm
* Refer to “Specification data”. ** Optional item
* Yes
date/time stamp

Safety

Before you use the instrument, make sure that you read and understand all the related data. This includes: all local safety procedures, the instructions for the UMM (if applicable), and this publication.
WARNING
• It is dangerous to ignore the specified limits for the instrument or to use the instrument when it is not in its normal condition. Use the applicable protection and obey all safety precautions.
• Do not use the instrument in loca tions with explosive gas, vapor or dust. There is a risk of an explosion.
• To prevent electrical shocks or damage to the instrument, do not connect more than 30V between the terminals, or between the terminals and the ground (earth).
Continued
[EN] English - 1K345 Issue 2
Page 4
Safety (Continued)
• UPM only. To prevent a dangerous release of pressure, isolate and bleed the system before you disconnect a pressure connection.
Before you start an operation or procedure in this publication, make sure that you have the necessary skills (if necessary, with qualifications from an approved training establishment). Follow good engineering practice at all times.

Safety - Marks and symbols on the instrument

Complies with European Union directives
Read the manual Battery
Ground (Earth)
Warning - refer to the manual
ON/OFF

To start

To start - Location of items

Item Description
1. On or off button.
2.
3.
4.
5.
6.
7.
8. Display. Refer to A3
9.
10. RTD connectors: To measure, use the 2 Wire, 3 Wire,
11. Connection point for some of the optional
12. Battery compartment. Refer to B1.
13., 14., 15. DPI 812 only. Terminals to measure current, to supply
Left-hand soft-key. Selects the function above it on
■ ■
the display (Item 25). Example: Edit Moves back one menu level.
ESC
Leaves a menu option. Cancels the changes to a value. Increases or decreases a value.
▲ ▼
Highlights a different item. Holds the data on the display. To continue, press the
HOLD
HOLD button ag ain. Shows the Select Task menu.
MENU
Selects or accepts an item or value.
OK
Selects [] or cancels [ ] a selection. Right-hand soft-key. Selects the function above it on
■ ■
the display (Item 25). Example: Settings
Communications port. Use to connect a Universal
SENSOR
/ PC
Measurement Module (UMM) or a RS232 cable.
or 4 Wire “INPUT” connectors. To simulate, use only the two “OUT” connectors. Refer to “Operation”.
accessories. Refer to the datasheet.
24V source, and to do switch tests.
A1 A2

To start - Items on the display

Item Description
16. DPI 812 only. Task indication for the switch test.
17. DPI 812 only. The loop power supply is on.
18. The measured value satisfies one of the alarm
19. The data on the display is on hold. To continue, press
20.
21. Identifies the type of data and the measurement
22. ... 24. The settings applied to the input or output.
22. °C The units or a specified scale (x:y) - (Table 4/5).
Pt... The RTD type (Pt50, ...) - (Table 4/5).
23.
24. = Filter
25. A soft-key function. To select an available function,
26. The measured value or values applicable to the task
27. The Edit display to set up text labels ( 6 characters):
= switch closed
UPM only. Task indication for the leak test.
Refer to: Select Task (Table 2/3)
Refer to: Select Task (Table 2/3)
conditions. Refer to: Settings (Table 4)
the HOLD button again.
Shows the battery level: 0 ... 100%.
range.
= Input = Output = IDOS input
Refer to: Select Task (Table 2/3)
RTD input conn ections: 2, 3, or 4 (Figure 1)
, ... ,
= Tare
press the soft-key below it. Example:
= Move left = Move right
selection.
x:y Scaling (Table 4).
OK = Accept the new text label
Shift = Change the keys: 1 23ABC or -_+abc
= Add a space
BS = Back space (Delete character)
A3
= switch open
= Output operation (Table 5)
= Maximum = Average = Minimum
(Table 4)

To start - Prepare the instrument

Before you use the instrument for the first time:
• Make sure that there is no damage to the instrument, and that there are no missing items.
• Remove the plastic film that protects the display. Use the tag (◗ ) in the top right-hand corner.
• Install the batteries (refer to B1). Then re-attach the cover.
2 - [EN] English K345 Issue 2
Page 5

To start - Power on or off

To turn the instrument on or off, press ❍ (A 1 - item [1]). The instrument does a self test and then shows the applicable data. When the pow er is off, the last set of con figuration options stays in memory. Refer to “Maintenance”.

To start - Set up the basic operation

Use the Set Up menu to set up the basic operation of the instrument.
Menu:
1
Select Task
(Table 2) (Table 1) []/[ ]
23
▲ ▼
Menu: Set Up
45
▲ ▼
If there is additional data for a menu option, select Settings (■■
) to see the values that are set up. If
necessary, adjust the values.
Table 1: Menu options - Set Up
Options
(If applicable)
Descript ion
DPI 812 only. To add a series resistor into the mA circuit. You can then use this instrument together with a HAR T® communic ator to set up and calibrate HART® devices.
... Scale To select the applicable international temperature
scale: IPTS 68 or ITS 90. To select and set up the backlight facility + timer.
Additional data: Select Settings ( ) To select and set up the power off facility + timer. Additional data: Select Settings ( ) To show the battery level (%).
To set the display contrast (%). Increases %, decreases %
To set the time + date. The calibration facility uses the date to give service and calibration messages. To set the language option.
To calibrate the instrument.
Additional data: Refer to “Calibration”.
To select and show the applicable status data. (Software Build, Calibration Due date, Serial Number, IDOS Information).

To start - Select a task (Measure and/or simulate)

When the instrument is set up (Table 1), use the Select Task menu to select the applicable task.
1
Menu:
Select Tas k
(Table 2/3)
23
▲ ▼
Display:
RTD output
Sk1 = Edit Sk2 = Settings
In Table 2/3, IDOS is a Universal Measurement Module (UMM). If you attach a UMM to the communications port (A1 - item [9]), the Select Task menu shows the applicable IDOS options.
Table 2: Menu options - Select Task
Options
(If applicable)
IDOS UMM only. An IDOS measurement task.
UPM only. A leak test.
Description
RTD or
An input measurement task:
Ohms
RTD - Measure RTD temperature OR Ohms - Measure RTD resistance.
RTD or
An output task:
Ohms
RTD - Simulate RTD temperature OR
Ohms - Simulate RTD resistance. mA DPI 812 only. A mA measurement task. mA(24V) DPI 812 only. A mA measurement task + the loop
power supply is on .
DPI 812 only. A switch test.
To set up the way the instrument works.
Additional data: Refer to: Set Up (Table 1) .
Table 3 shows all the one and two function operations that are available. If you attach a UMM, you can only use the options that include IDOS.
Table 3: Permitted 1 and 2 function operations
Function
mA (1) x (2) x (2) (2) mA(24V) (1) x (2) x (2) (2)
IDOS (1) (2) (2) (2) (2) x
RTD RTD Ohms Ohms IDOS (1) (1) (1) (1) (1)
x x (2) x (2) (2) xxxxx(2)
= DPI 812 only
[EN] English - 3K345 Issue 2
Page 6

To start - Set up the settings

When the task is set up (Table 2/3), use the Settings menu to adjust the input and/or output operation
Display:
RTD + mA
Sk1 = Start/Stop Sk2 = Settings
Menu:
Settings
(Table 4) []/[ ] Sk1 = Edit
Menu:
Settings
(Table 5) []/ [ ]
Settings selection
1
■ ■
4A 5A 6A
▲ ▼
4B 5B
▲ ▼
(If applicable)
■ ■
.
23
▲ ▼
Settings
x:y
If there is additional data for a menu option, select Settings (■■
) to see the values that are set up. If
necessary, adjust the values. Refer to “Edit functions”.
Table 4: (Part of table) Menu options - Settings (Input)
Options
(If applicable)
Description
... Units To select the temperature units (°C or °F).
UPM only = “Pressure Units” if you select an IDOS task (Table 2/3). Select one of the fixed units of measurement (psi, mbar ... ).
... type To select an applicable RTD type (Pt50, Pt100 ... )
To include maximum, minimum and average values for the measurement task. To select and set up a tare value for the measurement task (a specified value or the reading on the display). The instrument subtracts a positive tare value, and adds a negative tare value.
Additional data: Select Settings ( )
Table 4: (Part of table) Menu options - Settings (Input)
Options
(If applicable)
Description
To select and set up a scale of values: One local scale for each measurement task (Maximum: 5).
Additional data (Example 1/2): Select Settings ( ) To select and set up the filter values to give a smoother output for the measurement task:
Band as a % of full scale (FS). The filter
compares each ne w value with the previous value. If the new value is outside the band, it is not filtered.
Low pass filter time constant in seconds.
Increase the value to increase da mping factor. Additional data: Select Settings ( )
To select and set up the alarm values for the measurement task (maximum and minimum).
Additional data: Select Settings ( ) UPM only. Gage sensors or sensors with differential
operation. A zero correction that makes the instrument read zero at local pressure.
Leak Test only. To set an applicable period for the leak test (Hours:Minutes:Seconds).

To start - Edit functions

Example 1) Set up a label for x:y Scaling = %.
Settings
x:y
Sk1 = Edit A3 - item [27]
1 Edit 2 Edit 3
■ ■
Edit ... Edit 45
A3 - item [27]
Shift
■ ■
OK
%_
■ ■
4 - [EN] English K345 Issue 2
Page 7
Example 2) Set up values for x:y Scaling = 0 to 100%.
Settings
x:y
Sk1 = Edit
Settings
1
▲ ▼
= Flow scaling (mA, pressure only)
2 Edit 34
x:y
■ ■
▲ ▼
■ ■
Table 5: (Part of table) Menu options - Settings (Output)
Options
(If applicable)
Description
... Units To select the temperature units (°C or °F ).
... type To select a n applicable RTD type (Pt50, Pt100 ... )
To select and set up a v alue for the “Nudge” output. Example: 10°C increments. Additional data: Select Settings ( )
To select and set up values for the “Span Check”
output. Example outp ut cycle:
This cycle repeats automatically.
Additional data (Table 6): Select Settings ( ) To select and set up values for the “% Step” output.
% Step
Example output cycle:
Auto Repeat - Opt ional
Table 5: (Part of table) Menu options - Settings (Output)
Options
(If applicable) To select and set up values for the “Ramp” output.
Description
Example output cycle:
Auto Repeat - Optional
Additional data (Table 6): Select Settings ( )
Table 6: Additional data for Settings (Output):
Item Value
Span Check
Low (0%) Set the 0% value. High (100%) Set the 100% value. Dwell (d) Set the period (Hours:M inutes:Seconds) between
% Step Low (0%), High (100%), Dwell (d): As above. Step Size (s)
... % Defined Step Low (0%), High (100%), Dwell (d): As above. Step Size (s) Set the change in value for each step as a
Ramp Low (0%), High (100%), Dwell (d): As above. Travel (t) Se t the period (Hours:Minutes:Seconds) to go from
Auto Repeat If applicable, select this item to repeat a cycle
each change in value.
Set the change in value for each step as a percentage of the full-scale range (High - Low).
temperature or resistance valu e.
the Low (0%) value to the High (100%) value.
continuously.
Additional data (Table 6): Select Settings ( ) To select and set up values for the “Defined Step”
... Step
output. Example outp ut cycle:
Additional data (Table 6): Select Settings ( )
Auto Repeat - Opt ional
[EN] English - 5K345 Issue 2
Page 8

Operation

This section gives examples of how to connect and use the instrument. Before you start:
• Read and understand the “Safety” section.
• Do not use a damaged instrument.

Operation - RTD connections

To prevent instrument errors, make sure that the RTD connections (A1-item [10]) are correct. In the examples that follow 2W, 3W, and 4W identify the 2, 3, and 4 wire connections.

Operation - Communications port connections

Use the communications port (A1 - item [9]) to attach an IDOS Universal Measurement Module (UMM). When you attach the cable from a UMM (Figure 7/8), the instrument automatically changes the menus to give you all the applicable options (Table 2/3).

Operation - Measure RTD values

To measure the temperature or resistance values of an RTD:
1. Connect the instrument (Figure 1) and, if necessary, adjust the Set Up (Table 1).
2. Select the RTD or Ohms input task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4).
Pt100
The display shows the number of RTD connections.
= Four-wire RTD attached.
If this symbol does not agree with the number of RTD connections:
• Make sure that the RTD connections are correct.
• Make sure that the wires and the sensor are serviceable.

Operation - Change the output values

When the output operation is set up (Table 5), use one of these procedures to change the output values:
Table 7: Procedures to change the output
Output Procedure
,
Select Edit (■ ■) and/or use the ▲ ▼ buttons. See the example below.
Select Start/Stop (■ ■) or use the ▲ ▼ buttons to make the step changes manually. Select Start/Stop (■ ■).
Example procedure (“Nudge” output):
Display:
RTD output
Sk1 = Edit Sk2 = Settings
Display:
RTD output
1 Edit 23 Edit 4
■ ■ ■ ■
56
+100.00
°C
SettingsSettings
a) Temperature b) Resistance
▲ ▼
101.00
Increment = 1 (Table 5)
102.00
Figure 1: Example configuration - To measure the
temperature or resistance of an RTD
6 - [EN] English K345 Issue 2
Page 9

Operation - Simulate RTD values

To simulate the temperature or resistance values of an RTD:
1. Connect the instrument (Figure 2) and, if necessary, adjust the Set Up (Table 1).
2. Select the RTD or Ohms output task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 5).
3. Supply the output values to the system (Table 7).
Maximum: 30V
Edit Settings Edit Settings
a) Temperature b) Resistance
Figure 2: Example configuration - To simulate the
temperature or resistance of an RTD

Operation - Transmitter calibration

DPI 812 only. To calibrate a transmitter:
1. Connect the instrument (Figure 3/4) and, if necessary, adjust the Set Up (Table 1).
2. Select the applicable calibration task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4/5).
3. Supply the output values to the system (Table 7).
a) Temperature b) Resistance
Start SettingsEdit Settings
Figure 3: Example configuration - Transmitter calibration
with external loop power
Edit Settings Start Settings
a) Temperature b) Resistance
Figure 4: Example configuration - Transmitter calibration
with internal loop power
[EN] English - 7K345 Issue 2
Page 10

Operation - mA measurements

DPI 812 only. To measure a current:
1. Connect the instrument (Figure 5) and, if necessary, adjust the Set Up (Table 1).
2. Select the applicable mA input task from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 4).
Maximum: 30V
Settings Settings
a) with external loop
power
b) with internal loop
power
Figure 5: Example configuration - mA measurement

Operation - Switch test

DPI 812 only. To do tests on an RTD switch:
1. Connect the instrument (Figure 6) and, if necessary, adjust the Set Up (Table 1).
2. Select the applicable switch test from Select Task (Table 2/3) and, if necessary, adjust the Settings (Table 5). The display shows the switch condition (open or closed) in the top right-hand corner.
3. Supply the output values to the system (Table 7).
• Example - “Nudge” output. a. Use Edit (
) to set a value less than the switch
value.
b. Use the ▲ ▼ buttons to change the value in small
increments.
• Example - “Ramp” output. a. Set “High” and “Low” values that are applicable to
the switch value (Table 6). Then, to get an accurate switch value, set a long “Travel” period.
b. Use Start/Stop (
) to start and stop the “Ramp”
cycle.
4. If necessary, supply the output values in the opposite direction until the switch changes condition again.
The display shows the applicable values to open and close the switch.
5. To do the test again, press ESC to reset the values.
Start SettingsEdit Settings
a) Temperature b) Resistance
Figure 6: Example configuration - Switch test
8 - [EN] English K345 Issue 2
Page 11

Operation - UPM Pressure measurements

Read all the instructions supplied with the UPM and then use the specified procedures to connect it (Figure 7/8).
Settings
a) Pressur e b) Leak test
Start Settings
Figure 7: Example configuration - Pressure measurement
with a UPM
When the connections are complete, make the necessary IDOS selections (Table 2/3). If you re-attach a UPM, the instrument uses the same measurement units that you used before. The instrument keeps a record for the last 10 modules.

UPM - Measure the pressure

To measure the pressure (Figure 7):
1. Select the applicable pressure task from Select Task (Table 2/3) and, if necessary, adjust the Set Up (Table 1), and the Settings (Table 4/5).
2. If necessary, do a zero correction (Table 4).
To measure pressure with another operation (Figure 8), use the same procedure.
Settings
Figure 8: Example configuration - To measure pressure
and temperature

UPM - Leak test

To do a leak test on a pressure system (Figure 7):
1. Select an applicable leak test from Select Task (Table 2/3) and, if necessary, adjust the Set Up (Table 1), and the Settings (Table 4).
2. Set the period for the leak test (Table 4).
3. If necessary, do a zero correction (Table 4).
4. To start the leak test, select Start (
). When the test is
finished, the instrument calculates the leak rate in the applicable units/minute.

Operation - Error indications

If the display shows <<<< or >>>> :
• Make sure that the range is correct.
• Make sure that all the related equipment and connections are serviceable.
[EN] English - 9K345 Issue 2
Page 12

Maintenance

This section gives procedures to maintain the unit in a good condition. Return the instrument to the supplier for all repairs.

Maintenance – Clean the unit

Clean the case with a moist, lint-free cloth and a weak detergent. Do not use solvents or abrasive materials.

Maintenance – Replace the batteries

B1
To replace the batteries, refer to B1. Then re-attach the cover. Make sure that the time and date are correct. The calibration facility uses the date to give service and calibration messages. All the other configuration options stay in memory.

Calibration

Note: GE can provide a calibration service that is traceable to international standards.
We recommend that you return the instrument to the manufacturer or an approved service agent for calibration. If you use an alternative calibration facility, make sure that it uses these standards.

Calibration - Before you start

To do an accurate calibration, you must have:
• the calibration equipment specified in Table 8.
• a stable temperature environment: 70 ± 2°F (21 ± 1°C)
Function Calibration equipment
Pressure UPM only. Refer to the user manual for the IDOS
mA mA calibrator.
* Or an equivalent resistance simulator
Before you start the calibration, make sure that the time and date on the instrument are correct (Table 1).
Table 8: Calibration equipment
(ppm = parts per million)
Ohms - Stan dard 0 resistor
- *Standard resistor (): 100, 200, 300 Tolerance: 50 ppm + 0.6 ppm/°C + 5 ppm/year
- *Standard resistor (): 400, 1k, 2k, 4k Tolerance: 10 ppm + 0.6 ppm/°C + 5 ppm/year
Ohms An ohmmeter or an RTD measurement system
with these limits for the excitation current:
range (): Excitation (mA)
0 ... 400: 0.50 to 3.0 mA
400 ... 1500Ω: 0.05 to 0.8 mA 1500 ... 3200: 0.05 to 0.4 mA 3200 ... 4000: 0.05 to 0.3 mA
UPM.
Accuracy: Refer to Table 12.
Selection sequence:
Select Task (Table 2) ➤ Set Up (Table 1) ➤ Calibration ➤.
Display: Enter
1
Calibration PIN
Factory PIN = 4321 (Table 9)
23
▲ ▼
+
Menu:
Select Channel
45
▲ ▼
Table 9: Calibration options
Options Description
Ohms To calibrate the input resistance. Ohms To calibrate the output resistance
IDOS ... UMM only. To calibrate the specified IDOS UMM.
Refer to the user manual for the IDOS UMM.
mA DPI 812 only. To calibrate the mA input.
Calibration Due: To set the date of the next calibration for the instrument. After the specified calibration date, there is a warning message. There is a selection box to stop the warning. To change the calibration PIN (Personal Identification Number).
When you select a channel, the display shows the applicable instructions to complete the calibration. When the calibration is complete, select Calibration Due and set the new calibration date for the instrument.

Calibration - Procedures (Ohms input)

1. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
2. Use the calibration menu (Table 9) to do a two-point calibration for each range.
• Range: 0-399.9 a. Nominal zero ohms: Make a 4 wire connection to
the 0 resistor (Figure 1).
b. Nominal positive full-scale ohms: Make a 4 wire
connection to the 400 resistor (Figure 1).
• Range: 400Ω-4kΩ a. Nominal zero ohms: Make a 4 wire connection to
the 400 resistor (Figure 1).
b. Nominal positive full-scale ohms: Make a 4 wire
connection to the 4k resistor (Figure 1).
The display shows the applicable instructions to calibrate each range.
10 - [EN] English K345 Issue 2
Page 13
3. To make sure that the calibration is correct, select the applicable ohms input task (Table 2).
4. Make a 4 wire connection to the applicable standard resistor (Table 10) and measure the value (Figure 1).
5. Make sure that the error is in the specified limits (Table 10).
Table 10: Ohms input error limits
Standard Resistor* ()
0 (Short circuit) - 0.05 100 0.008 0.05 200 0.013 0.05 300 0.018 0.05 400 0.007 0.05 1k 0.042 0.25 2k 0.052 0.25 4k 0.072 0.50
* Or an equivalent resistance simulator
Resistor error ()
Permitted DPI 811/812 error ()

Calibration - Procedures (Ohms output)

1. Connect the instrument to the calibration equipment (Figure 2).
2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a two-point calibration for each range.
• Range: 0-399.9
• Range: 400Ω-1999.9Ω
• Range: 2kΩ-4kΩ The display shows the applicable instructions to calibrate each range.
4. To make sure that the calibration is correct, select the applicable ohms output task (Table 2).
5. Supply the specified values (Table 11). Make sure that the error is in the specified limits.
Table 11: Ohms output error limits
Ohms
()
0 0.003 0.05 100 0 .004 0.06 200 0 .005 0.06 300 0 .007 0.07 400 0 .008 0.07 1000 0.015 0.30 2000 0.026 0.40 4000 0.049 0.80
Calibrator error ()
Permitted DPI 811/812 error ()

Calibration - Procedures (mA input)

1. DPI 812 only. Connect the instrument to the calibration equipment (Figure 5).
2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a three-point calibration (-FS, Zero and +FS). The display shows the applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the applicable mA input task (Table 2) and apply these values:
• mA: -55, -40, -24, -18, -12, -6, 0 (short circuit) Then mA: 0, 6, 12, 18, 24, 40, 55.
5. Make sure that the error is in the specified limits (Table 12).
Table 12: mA input error limits
Applied mA
±55 0.0022 0.005 ±40 0.0018 0.004 ±24 0.0014 0.003 ±18 0.0004 0.003 ±12 0.0003 0.002 ±6 0.0002 0.002 0 (Short circuit) - 0.001
Calibrator error (mA)
Permitted DPI 811/812 error (mA)

Calibration - Procedures (IDOS UMM)

Refer to the user manual for the IDOS UMM. When the calibration is complete, the instrument automatically sets a new calibration date in the UMM.
[EN] English - 11K345 Issue 2
Page 14

Specification data

All accuracy statements are for one year.

Specification - General

Languages English [Default] Operating
temperature Storage temperature Humidity 0 to 90% without condensation
Shock/Vibration BS EN 61010:2001; Def Stan 66-31, 8.4 cat III EMC BS EN 61326-1:1998 + A2:2001 Safety Electrical - BS EN 61010:2001; CE Marked Size (L: W: H) 7.1 x 3.3 x 2.0 in (180 x 85 x 50 mm) Weight 14 oz (400 g) Power supply 3 x AA alkaline batteries Duration (Measure)
Duration (Supply)

Specification - Temperature ranges

RTD type Standard Range °F Range °C Accuracy °F * Accuracy °C *
Pt50 (385) IEC 751 -328 ... 1562 -200 ... 850 0.90 0.50 Pt100 (385) IEC 751 -328 ... 1562 -200 ... 850 0.45 0.25 Pt200 (385) IEC 751 -328 ... 1562 -200 ... 850 1.08 0.60 Pt500 (385) IEC 751 -328 ... 1562 -200 ... 850 0.72 0.40 Pt1000 (385) IEC 751 -328 ... 752 -200 ... 400 0.36 0.20 D 100 (392) JIS 1604-1989 -328 ... 1202 -200 ... 650 0.45 0.25 Ni 100 DIN 43760 -76 ... 482 -60 ... 250 0.36 0.20 Ni 120 MINCO 7-120 -112 ... 500 -80 ... 260 0.36 0.20 *Temperature coefficient: 14 … 50°F, 86 … 122°F = 0.0028% FS / °F (-10 … 10°C, 30 … 50°C = 0.005% FS / °C)

Specification - Resistance ranges Specification - Electrical connectors (A2)

Range () Excitation (mA) Accuracy (Ω)*
0 ... 400 0.15 ... 0.5 0.15 0 ... 400 0.50 ... 3.0 0.10
400 ... 1500 0.05 ... 0.8 0.50 1500 ... 3200 0.05 ... 0.4 1.00 3200 ... 4000 0.05 ... 0.3 1.30
*Temperature coefficient: 14 … 50°F, 86 … 122 °F = 0.0028% FS / °F (-10 … 10°C, 30 … 50°C = 0.005% FS / °C)
14 … 122°F (-10 … 50°C)
-4 … 158°F (-20 … 70°C)
(Def Stan 66-31, 8.6 cat III)
RTD, Ω: ≈ 70 hours mA: 35 hours mA: 10 hours (24 V Source at 12 mA) RTD, Ω: ≈ 65 hours
Range (Measure) 0 to ±55 mA Accuracy 0.02% of reading + 3 counts Temperature coefficient 14 … 50°F, 86 … 122°F (-10 … 10°C, 30 … 50°C) Switch detection Open a nd closed. 2 mA current. Loop power output 24 V ± 10%
HART® resistor Connectors (A2) Three 0.16 in (4 mm) sockets
0.0011% FS / °F (0.002% FS / °C)
250 (menu select ion)
12 - [EN] English K345 Issue 2
Page 15

Customer Service

Visit our web site: www.gesensing.com
Loading...