Druck DPI 812 Operating Manual

GE Infrastructure
Sensing
Druck DPI 811/812
RTD calibrator and RTD loop calibrator
User manual - K345
A1
B1
A2
10
1
A2
10
9
A3
8
2
3
4
7
6
5
11
B1
12
K345 Issue 2
1513
14
A3
17182619 2716
20
21
22 23
24
25

Table of Contents

Introduction ......................................................................... 1
Safety ..................................................................................... 1
Safety - Marks and symbols on the instrument ............... 2
To start .................................................................................. 2
To start - Location of items ........................................................ 2
To start - Items on the display .................................................. 2
To start - Prepare the instrument ........................................... 2
To start - Power on or off ............................................................ 3
To start - Set up the basic operation ..................................... 3
To start - Select a task (Measure and/or simulate) ......... 3
To start - Set up the settings ..................................................... 4
To start - Edit functions ................................................................ 4
Operation .............................................................................. 6
Operation - RTD connections .................................................... 6
Operation - Communications port connections .............. 6
Operation - Measure RTD values ............................................. 6
Operation - Change the output values ................................. 6
Operation - Simulate RTD values ............................................ 7
Operation - Transmitter calibration ....................................... 7
Operation - mA measurements ............................................... 8
Operation - Switch test ................................................................ 8
Operation - UPM Pressure measurements ......................... 9
Operation - Error indications ..................................................... 9
Maintenance ..................................................................... 10
Maintenance – Clean the unit ................................................ 10
Maintenance – Replace the batteries ................................ 10
Calibration ......................................................................... 10
Calibration - Before you start ................................................. 10
Calibration - Procedures (Ohms input) ............................... 10
Calibration - Procedures (Ohms output) ........................... 11
Calibration - Procedures (mA input) .................................... 11
Calibration - Procedures (IDOS UMM) ................................. 11
Specification data ............................................................ 12
Specification - General .............................................................. 12
Specification - Temperature ranges ................................... 12
Specification - Resistance ranges ........................................ 12
Specification - Electrical connectors (A2) ......................... 12
Customer service ............................................... Back cover
© 2005 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.

Introduction

The DPI 811 RTD (Resistance Temperature Detector) Calibrator and DPI 812 RTD Loop Calibrator are part of the Druck DPI 800 series of hand held instruments. The DPI 800 series uses Intelligent Digital Output Sensor (IDOS) technology to give instant plug and play functionality with a range of Universal Measurement Modules (UMM). Example: the Universal Pressure Module (UPM). The DPI 811/812 include these functions:
Function DPI 811 DPI 812
Measure/simulate RTD temperature or resistance
Step/Ramp functions Automatic/Manual Communications port IDOS or RS232 Language selection Yes Measure pressure/Leak test ** External IDOS UPM ** Snapshot Up to 1000 displays with a
Measure mA No 0 - 55 mA HART® resistor No Yes V dc output No 24 V Switch test No Yes Other functions Hold, Maximum/Minimum/Average,
Filter, Tare, Scaled values, Backlight, Alarm
* Refer to “Specification data”. ** Optional item
* Yes
date/time stamp

Safety

Before you use the instrument, make sure that you read and understand all the related data. This includes: all local safety procedures, the instructions for the UMM (if applicable), and this publication.
WARNING
• It is dangerous to ignore the specified limits for the instrument or to use the instrument when it is not in its normal condition. Use the applicable protection and obey all safety precautions.
• Do not use the instrument in loca tions with explosive gas, vapor or dust. There is a risk of an explosion.
• To prevent electrical shocks or damage to the instrument, do not connect more than 30V between the terminals, or between the terminals and the ground (earth).
Continued
[EN] English - 1K345 Issue 2
Safety (Continued)
• UPM only. To prevent a dangerous release of pressure, isolate and bleed the system before you disconnect a pressure connection.
Before you start an operation or procedure in this publication, make sure that you have the necessary skills (if necessary, with qualifications from an approved training establishment). Follow good engineering practice at all times.

Safety - Marks and symbols on the instrument

Complies with European Union directives
Read the manual Battery
Ground (Earth)
Warning - refer to the manual
ON/OFF

To start

To start - Location of items

Item Description
1. On or off button.
2.
3.
4.
5.
6.
7.
8. Display. Refer to A3
9.
10. RTD connectors: To measure, use the 2 Wire, 3 Wire,
11. Connection point for some of the optional
12. Battery compartment. Refer to B1.
13., 14., 15. DPI 812 only. Terminals to measure current, to supply
Left-hand soft-key. Selects the function above it on
■ ■
the display (Item 25). Example: Edit Moves back one menu level.
ESC
Leaves a menu option. Cancels the changes to a value. Increases or decreases a value.
▲ ▼
Highlights a different item. Holds the data on the display. To continue, press the
HOLD
HOLD button ag ain. Shows the Select Task menu.
MENU
Selects or accepts an item or value.
OK
Selects [] or cancels [ ] a selection. Right-hand soft-key. Selects the function above it on
■ ■
the display (Item 25). Example: Settings
Communications port. Use to connect a Universal
SENSOR
/ PC
Measurement Module (UMM) or a RS232 cable.
or 4 Wire “INPUT” connectors. To simulate, use only the two “OUT” connectors. Refer to “Operation”.
accessories. Refer to the datasheet.
24V source, and to do switch tests.
A1 A2

To start - Items on the display

Item Description
16. DPI 812 only. Task indication for the switch test.
17. DPI 812 only. The loop power supply is on.
18. The measured value satisfies one of the alarm
19. The data on the display is on hold. To continue, press
20.
21. Identifies the type of data and the measurement
22. ... 24. The settings applied to the input or output.
22. °C The units or a specified scale (x:y) - (Table 4/5).
Pt... The RTD type (Pt50, ...) - (Table 4/5).
23.
24. = Filter
25. A soft-key function. To select an available function,
26. The measured value or values applicable to the task
27. The Edit display to set up text labels ( 6 characters):
= switch closed
UPM only. Task indication for the leak test.
Refer to: Select Task (Table 2/3)
Refer to: Select Task (Table 2/3)
conditions. Refer to: Settings (Table 4)
the HOLD button again.
Shows the battery level: 0 ... 100%.
range.
= Input = Output = IDOS input
Refer to: Select Task (Table 2/3)
RTD input conn ections: 2, 3, or 4 (Figure 1)
, ... ,
= Tare
press the soft-key below it. Example:
= Move left = Move right
selection.
x:y Scaling (Table 4).
OK = Accept the new text label
Shift = Change the keys: 1 23ABC or -_+abc
= Add a space
BS = Back space (Delete character)
A3
= switch open
= Output operation (Table 5)
= Maximum = Average = Minimum
(Table 4)

To start - Prepare the instrument

Before you use the instrument for the first time:
• Make sure that there is no damage to the instrument, and that there are no missing items.
• Remove the plastic film that protects the display. Use the tag (◗ ) in the top right-hand corner.
• Install the batteries (refer to B1). Then re-attach the cover.
2 - [EN] English K345 Issue 2

To start - Power on or off

To turn the instrument on or off, press ❍ (A 1 - item [1]). The instrument does a self test and then shows the applicable data. When the pow er is off, the last set of con figuration options stays in memory. Refer to “Maintenance”.

To start - Set up the basic operation

Use the Set Up menu to set up the basic operation of the instrument.
Menu:
1
Select Task
(Table 2) (Table 1) []/[ ]
23
▲ ▼
Menu: Set Up
45
▲ ▼
If there is additional data for a menu option, select Settings (■■
) to see the values that are set up. If
necessary, adjust the values.
Table 1: Menu options - Set Up
Options
(If applicable)
Descript ion
DPI 812 only. To add a series resistor into the mA circuit. You can then use this instrument together with a HAR T® communic ator to set up and calibrate HART® devices.
... Scale To select the applicable international temperature
scale: IPTS 68 or ITS 90. To select and set up the backlight facility + timer.
Additional data: Select Settings ( ) To select and set up the power off facility + timer. Additional data: Select Settings ( ) To show the battery level (%).
To set the display contrast (%). Increases %, decreases %
To set the time + date. The calibration facility uses the date to give service and calibration messages. To set the language option.
To calibrate the instrument.
Additional data: Refer to “Calibration”.
To select and show the applicable status data. (Software Build, Calibration Due date, Serial Number, IDOS Information).

To start - Select a task (Measure and/or simulate)

When the instrument is set up (Table 1), use the Select Task menu to select the applicable task.
1
Menu:
Select Tas k
(Table 2/3)
23
▲ ▼
Display:
RTD output
Sk1 = Edit Sk2 = Settings
In Table 2/3, IDOS is a Universal Measurement Module (UMM). If you attach a UMM to the communications port (A1 - item [9]), the Select Task menu shows the applicable IDOS options.
Table 2: Menu options - Select Task
Options
(If applicable)
IDOS UMM only. An IDOS measurement task.
UPM only. A leak test.
Description
RTD or
An input measurement task:
Ohms
RTD - Measure RTD temperature OR Ohms - Measure RTD resistance.
RTD or
An output task:
Ohms
RTD - Simulate RTD temperature OR
Ohms - Simulate RTD resistance. mA DPI 812 only. A mA measurement task. mA(24V) DPI 812 only. A mA measurement task + the loop
power supply is on .
DPI 812 only. A switch test.
To set up the way the instrument works.
Additional data: Refer to: Set Up (Table 1) .
Table 3 shows all the one and two function operations that are available. If you attach a UMM, you can only use the options that include IDOS.
Table 3: Permitted 1 and 2 function operations
Function
mA (1) x (2) x (2) (2) mA(24V) (1) x (2) x (2) (2)
IDOS (1) (2) (2) (2) (2) x
RTD RTD Ohms Ohms IDOS (1) (1) (1) (1) (1)
x x (2) x (2) (2) xxxxx(2)
= DPI 812 only
[EN] English - 3K345 Issue 2
Loading...
+ 10 hidden pages