Druck DPI 705E Specifications

DPI 705E Series
Druck Handheld Pressure Indicators
The Druck DPI 705E Series of handheld pressure and optional temperature indicators combine tough and rugged design with accurate and reliable measurements.
Compact and robust, the DPI 705E Series is designed for single handed operation and provides many essential features required for routine maintenance and system troubleshooting.
Features
48 pressure ranges from ±25 mbar to 1,400 bar (±1.69 psi to 20,000 psi / 2.5 KPa to 140 MPa)
Total 1 year uncertainty down to 0.05% full scale (FS) over temperature range of -10°C to +50°C
Integral calibration record with calibration due
Safe Area
Pressure Indicator
Hazardous Area
Pressure Indicator
count-down display
Rugged, handheld design with backlit high-contrast display
Leak test, tare, maximum/minimum and filter
Hazardous Area (Intrinsically safe) version available
Optional remote plug and play pressure and Resistance Temperature Detector (RTD) sensors
Optional pneumatic and hydraulic hand pumps
Druck.com
DPI 705E Improves Test Efficiency
Power on and use. No warm up time required.
19 pressure measurement units (See below)
1, 3 or 5 minute leak test
Impact resistant, sealed to IP54
Large digit high-contrast LCD with backlight
Optional carrying case with belt loop
Optional magnetic/loop hanging strap
Long battery life
Integral desk stand and hanger
Zero
Zero pressure correction (gauge/differential sensors)
Tare
0 to 100% FS temporary zero offset capability by subtracting the current reading from subsequent measurements.
Filter
Enables a filtered pressure reading by showing a rolling average of the last 10 measurements. Providing a more stable reading of a noisy measurement.
Special Features
Pressure Units
mbar, bar, Pa, hPa, Pa, MPa, psi , lb/ft2, kgf/cm2, kgf/m2, mmHg(0°C), mHg(0°C), inHg 0, mmH2O, cmH2O, mH2O, inH2O (4°C, 20°C), ftH2O (4°C, 20°C)
Calibration Certificates
Supplied as standard with bar, psi and kPa
Optional UKAS approved calibrations available
Leak Test
Used to determine if there is a leak in the system by recording the pressure change over of time.
Leak test can also be used with the RTD sensor to record a temperature change over time.
User defined 1, 3 or 5 minute leak test.
Alarm
User adjustable; high and low pressure alarms with visual (Bell Icon, Pressure Reading and Backlight flash) and audible warning for 60 seconds.
Calibration
Used in conjunction with the optional pneumatic or hydraulic hand pumps, the DPI 705E Series provides a simple, low-cost calibration solution.
Battery
4 x AA Alkaline/NiCd/NiMH cells
Power efficient design, can be used for 8 Hours per day, 6 days per week for 1 Yearusing a single set of batteries.
Maximum/Minimum
Captures both maximum and minimum readings in peak-hold mode.
Display
16mm LCD with ±99999 readout
Hazardous Area Approvals
ATEX, IECEx, NEPSI, ECASEx (Ordering code ‘H1‘)
INMETRO (Ordering code ‘H2‘)
IS Class I, Groups A, B, C & D, T4 Ex ia; Class I, Zone 0, AEx/Ex ia IIC T4 Ga (-10°C ≤ Ta ≤ +50°C) to CSA Std. 157-92, UL 913 (7th Ed.), CAN/CSA-C22.2 No. 60079­0:2018, CAN/CSA-C22.2 No. 60079-11:2012, ANSI/UL 60079-0:2018 and ANSI/UL 60079-11:2012.
Ex ia IIC T4 Ga (-10°C ≤ Ta ≤ +50°C) to ATEX IEC 60079-0:2017 and EN 60079-11:2011
Pressure Ranges
Internal and remote sensors are fitted with pressure connectors as detailed below:
25 mbar – 200 bar (10 inH2O - 3000 psi / 2.5 kPa - 20 MPa): G1/8 BSP female
Differential reference port: Legris 4 mm tubing adaptor
≥350 bar (5000 psi / 35 MPa): Autoclave 9/16 x 18 UNF Male
DPI 705E Internal Pressure Sensors
The DPI705E includes an internal Absolute, Gauge or Differential sensor.
Available ranges as per table:
Internal Pressure Sensor Availability
Pressure Range Abs Gauge Differential Standard
25 mbar / 10 inH2O / 2.5 kPa ­70 mbar / 1 psi / 7 kPa ­200 mbar / 3 psi / 20 kPa ­350 mbar / 5 psi / 35 kPa 700 mbar / 10 psi / 70 kPa 1 bar / 15 psi / 100 kPa 2 bar / 30 psi / 200 kPa
3.5 bar / 50 psi / 350 kPa 7 bar / 100 psi / 700 kPa 10 bar / 150 psi / 1000 kPa 20 bar / 300 psi / 2 MPa 35 bar / 500 psi / 3.5 MPa 70 bar / 1000 psi / 7 MPa 100 bar / 1500 psi / 10 MPa 135 bar / 2000 psi / 13.5 MPa 200 bar / 3000 psi / 20 MPa 350 bar / 5000 psi / 35 MPa 700 bar / 10000 psi / 70MPa
ü ü ü ü ü ü
ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü
- -
ü
- -
Accuracy
-
-
-
-
-
-
-
-
-
Note: Do not apply pressure greater than 1 bar (16 psi) absolute to the negative port of differential pressure indicators.
High Accuracy
-
-
-
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
PM 700E External Remote Pressure Sensors
Any numberof remote sensors can be used with a single DPI705E as all sensors hold their own calibration data and are supplied with a 2.9m (9.5 feet) cable
Available ranges as per table:
External Remote Pressure Sensor Availability
Pressure Range Abs Gauge Differential Standard
25 mbar / 10 inH2O / 2.5 kPa ­70 mbar / 1 psi / 7 kPa ­200 mbar / 3 psi / 20 kPa -
350 mbar / 5 psi / 35 kPa 700 mbar / 10 psi / 70 kPa 1 bar / 15 psi / 100 kPa 2 bar / 30 psi / 200 kPa
3.5 bar / 50 psi / 350 kPa 7 bar / 100 psi / 700 kPa 10 bar / 150 psi / 1000 kPa 20 bar / 300 psi / 2 MPa 35 bar / 500 psi / 3.5 MPa 70 bar / 1000 psi / 7 MPa 100 bar / 1500 psi / 10 MPa 135 bar / 2000 psi / 13.5 MPa 200 bar / 3000 psi / 20 MPa 350 bar / 5000 psi / 35 MPa 700 bar / 10000 psi / 70 MPa 1000 bar/ 15000 psi/ 100 MPa 1400 bar/20000 psi/140 MPa
Accuracy
ü ü ü ü ü ü
ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü
ü
ü
ü
- -
- -
- -
- -
-
-
-
-
-
-
-
-
-
High Accuracy
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
-
-
-
PM700E (Gauge, Absolute)
PM700E (Differential)
Accuracy Levels
Total Uncertainty includes 1-year drift for both Standard and High accuracy levels.
1-Standard ±0.1% FS NLH&R over temperature, including non-linearity, hysteresis and repeatability. 2-High ±0.05% FS NLH&R over temperature, including non-linearity, hysteresis and repeatability.
Loading...