DPI 705E Series
Druck Handheld Pressure Indicators
The Druck DPI 705E Series of handheld pressure
and optional temperature indicators combine
tough and rugged design with accurate and
reliable measurements.
Compact and robust, the DPI 705E Series is
designed for single handed operation and
provides many essential features required
for routine maintenance and system
troubleshooting.
Features
• 48 pressure ranges from ±25 mbar to 1,400 bar
(±1.69 psi to 20,000 psi / 2.5 KPa to 140 MPa)
• Total 1 year uncertainty down to 0.05% full scale (FS)
over temperature range of -10°C to +50°C
• Integral calibration record with calibration due
Safe Area
Pressure Indicator
Hazardous Area
Pressure Indicator
count-down display
• Rugged, handheld design with backlit high-contrast
display
• Leak test, tare, maximum/minimum and filter
• Hazardous Area (Intrinsically safe) version available
• Optional remote plug and play pressure and
Resistance Temperature Detector (RTD) sensors
• Optional pneumatic and hydraulic hand pumps
Druck.com
DPI 705E Improves Test Efficiency
• Power on and use. No warm up time required.
• 19 pressure measurement units (See below)
• 1, 3 or 5 minute leak test
• Impact resistant, sealed to IP54
• Large digit high-contrast LCD with backlight
• Optional carrying case with belt loop
• Optional magnetic/loop hanging strap
• Long battery life
• Integral desk stand and hanger
Zero
Zero pressure correction (gauge/differential sensors)
Tare
0 to 100% FS temporary zero offset capability by
subtracting the current reading from subsequent
measurements.
Filter
Enables a filtered pressure reading by showing a rolling
average of the last 10 measurements. Providing a more
stable reading of a noisy measurement.
Special Features
Pressure Units
mbar, bar, Pa, hPa, Pa, MPa, psi , lb/ft2, kgf/cm2, kgf/m2,
mmHg(0°C), mHg(0°C), inHg 0, mmH2O, cmH2O, mH2O,
inH2O (4°C, 20°C), ftH2O (4°C, 20°C)
Calibration Certificates
• Supplied as standard with bar, psi and kPa
• Optional UKAS approved calibrations available
Leak Test
Used to determine if there is a leak in the system by
recording the pressure change over of time.
Leak test can also be used with the RTD sensor to record
a temperature change over time.
User defined 1, 3 or 5 minute leak test.
Alarm
User adjustable; high and low pressure alarms with
visual (Bell Icon, Pressure Reading and Backlight flash)
and audible warning for 60 seconds.
Calibration
Used in conjunction with the optional pneumatic or
hydraulic hand pumps, the DPI 705E Series provides a
simple, low-cost calibration solution.
Battery
• 4 x AA Alkaline/NiCd/NiMH cells
• Power efficient design, can be used for 8 Hours per
day, 6 days per week for 1 Yearusing a single set of
batteries.
Maximum/Minimum
Captures both maximum and minimum readings in
peak-hold mode.
Display
16mm LCD with ±99999 readout
Hazardous Area Approvals
• ATEX, IECEx, NEPSI, ECASEx (Ordering code ‘H1‘)
• INMETRO (Ordering code ‘H2‘)
• IS Class I, Groups A, B, C & D, T4 Ex ia; Class I, Zone 0,
AEx/Ex ia IIC T4 Ga (-10°C ≤ Ta ≤ +50°C) to CSA Std.
157-92, UL 913 (7th Ed.), CAN/CSA-C22.2 No. 600790:2018, CAN/CSA-C22.2 No. 60079-11:2012, ANSI/UL
60079-0:2018 and ANSI/UL 60079-11:2012.
• Ex ia IIC T4 Ga (-10°C ≤ Ta ≤ +50°C) to ATEX IEC
60079-0:2017 and EN 60079-11:2011
Pressure Ranges
Internal and remote sensors are fitted with pressure connectors as detailed below:
• 25 mbar – 200 bar (10 inH2O - 3000 psi / 2.5 kPa - 20 MPa): G1/8 BSP female
• Differential reference port: Legris 4 mm tubing adaptor
• ≥350 bar (5000 psi / 35 MPa): Autoclave 9/16 x 18 UNF Male
DPI 705E Internal Pressure Sensors
The DPI705E includes an internal Absolute, Gauge or
Differential sensor.
Available ranges as per table:
Internal Pressure Sensor Availability
Pressure Range Abs Gauge Differential Standard
25 mbar / 10 inH2O / 2.5 kPa 70 mbar / 1 psi / 7 kPa 200 mbar / 3 psi / 20 kPa 350 mbar / 5 psi / 35 kPa
700 mbar / 10 psi / 70 kPa
1 bar / 15 psi / 100 kPa
2 bar / 30 psi / 200 kPa
3.5 bar / 50 psi / 350 kPa
7 bar / 100 psi / 700 kPa
10 bar / 150 psi / 1000 kPa
20 bar / 300 psi / 2 MPa
35 bar / 500 psi / 3.5 MPa
70 bar / 1000 psi / 7 MPa
100 bar / 1500 psi / 10 MPa
135 bar / 2000 psi / 13.5 MPa
200 bar / 3000 psi / 20 MPa
350 bar / 5000 psi / 35 MPa
700 bar / 10000 psi / 70MPa
ü ü ü
ü ü ü
ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü
- -
ü
- -
Accuracy
-
-
-
-
-
-
-
-
-
Note: Do not apply pressure greater than 1 bar (16 psi) absolute to
the negative port of differential pressure indicators.
High
Accuracy
-
-
-
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
PM 700E External Remote Pressure Sensors
Any numberof remote sensors can be used with a
single DPI705E as all sensors hold their own calibration
data and are supplied with a 2.9m (9.5 feet) cable
Available ranges as per table:
External Remote Pressure Sensor Availability
Pressure Range Abs Gauge Differential Standard
25 mbar / 10 inH2O / 2.5 kPa 70 mbar / 1 psi / 7 kPa 200 mbar / 3 psi / 20 kPa -
350 mbar / 5 psi / 35 kPa
700 mbar / 10 psi / 70 kPa
1 bar / 15 psi / 100 kPa
2 bar / 30 psi / 200 kPa
3.5 bar / 50 psi / 350 kPa
7 bar / 100 psi / 700 kPa
10 bar / 150 psi / 1000 kPa
20 bar / 300 psi / 2 MPa
35 bar / 500 psi / 3.5 MPa
70 bar / 1000 psi / 7 MPa
100 bar / 1500 psi / 10 MPa
135 bar / 2000 psi / 13.5 MPa
200 bar / 3000 psi / 20 MPa
350 bar / 5000 psi / 35 MPa
700 bar / 10000 psi / 70 MPa
1000 bar/ 15000 psi/ 100 MPa
1400 bar/20000 psi/140 MPa
Accuracy
ü ü ü
ü ü ü
ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü ü ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü
ü
ü
ü
- -
- -
- -
- -
-
-
-
-
-
-
-
-
-
High
Accuracy
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
ü ü
-
-
-
PM700E (Gauge, Absolute)
PM700E (Differential)
Accuracy Levels
Total Uncertainty includes 1-year drift for both Standard and High accuracy levels.
1-Standard ±0.1% FS NLH&R over temperature, including non-linearity, hysteresis and repeatability.
2-High ±0.05% FS NLH&R over temperature, including non-linearity, hysteresis and repeatability.