Druck DPI 620 Data sheet

GE
Sensing & Inspection Technologies
DPI 620 Series
Advanced Modular Calibration and HART Communication System
Combines an advanced multi-function calibrator and HART communicator with world-class pressure measurement and generation.
Modularity brings a whole new perspective to multi-function test instruments, -calibrators and HART communicators
• Modular concept provides greater fl exibility
• Re-range during use
• Expand over time
• Adapt by application
• Allows signifi cant inventory reductions
• Simplifi es user training
• Reduces cost of ownership
• A Complete HART communicator and multifunction calibrator in one device
The system includes:
A combined Multifunction calibrator…
• Accuracy from 0.0025%rdg + 0.002%FS
• Measure, source and simulate mA, mV, V, ohms, frequency, RTDs and thermocouples
• Easy to use with photo quality touch screen
• Robust, weatherproof and extremely compact
…and a HART digital communicator…
• Device Description Library for HART devices
• Free of charge upgrades and new device descriptions
• Integral 24V loop power supply to energise HART devices and loops
• Menu selectable 250 ohm loop resistor
• Measures mA and sources the primary variable for device trims and calibration
…with advanced connectivity and documenting features
• PDA version with Windows CE – view loop diagrams, datasheets, safety procedures, spreadsheets and create text documents.
• IEEE 802.11g wireless connection with Internet access
Pressure measurement modules
• 25 mbar to 1000 bar (10 inH2O to 15000 psi)
• Accuracy from 0.005%FS
• Fully interchangeable modules with no need for set-up, calibration or tools
Pressure generation stations
• Advanced pressure generation
- 95% vacuum to 20 bar pneumatic
- 95% vacuum to 100 bar pneumatic
- 0 to 1000 bar hydraulic
• Stand-alone pressure stations can replace hand pumps and can be used as comparators
Applications
• Instrumentation installation, commissioning, maintenance and calibration
• HART test, confi guration and calibration
• System measurement and monitoring
• Indicator, recorder and controller testing
• Process loop set up and diagnostics
• Switch, trip and safety system testing
For …
• Plant and process engineers
• Service companies and site contractors
• Installation and commissioning engineers
• Laboratory technicians
• Automate calibration procedures and document As Found and As Left results
• Store a complete plant database including tags, devices and procedures
• Graphical views of data log fi les, calibration results and historical trends
2
A fl exible modular system
The DPI 620 Series - Advanced Modular Calibration and HART Communication System uses three basic components to provide the multi-functionality to perform duties formerly requiring a wide range of diff erent instruments.
Features
• Multi-function capabilities: electrical, frequency, temperature and pressure
• Complete HART communicator included
• Modular re-rangeable and expandable concept
• Individual components can be used as stand-alone instruments
• Allows signifi cant inventory reductions
• Simplifi es training and improves operator safety
• Reduces cost of ownership
Its basic component is an ultra-compact, electrical, frequency and temperature calibrator and HART communicator, which provides simultaneous measurement and source capabilities with a fully featured HART digital interface.
Pressure measurement is provided by interchangeable pressure modules that can be attached to the calibrator by a pressure module carrier.
If pressure generation is required the calibrator and pressure module can be attached to one of three pressure generation stations, to form a fully-integrated pressure calibrator, of unparalleled performance.
The simple yet sophisticated design and concept combines, for the fi rst time, an advanced electrical calibrator with state-of-the-art pressure measurement and generation, so that there is no longer any need to compromise one measurand capability in favour of another.
Measure and source mA, mV, V, ohms, frequency, RTD’s and thermocouples.
Re-rangeable pressure measurement and generation from 25 mbar (10 inH2O) to 1000 bar (15000 psi)
Re-rangeable dual channel pressure measurement from 25 mbar (10 inH
O) to 1000 bar (15000 psi)
2
3
The DPI 620 Series ­Advanced Modular Calibrator and HART Communicator
Features
• Measure, source and simulate multiple parameters
• Complete HART communicator included
• Easy-to-use, with photo quality colour touch screen, which can be operated by a gloved hand
• Weather-proof to IP 65
• Can display up to 6 active readings
• Available with a range of add-on modules, pressure measurement modules and generation stations, Intecal Field Calibration documenting module, 300Vac measurement probe
• Digital interface for pressure modules and future options
• USB master and slave for connecting to a PC and peripherals
• Available in Windows CE version
• Long-life Lithium-Polymer battery pack
The high resolution, colour touch screen of the calibrator shows only those buttons necessary for the immediate, selected tasks. The buttons have been sized so that they can be operated by a gloved hand, with no need for a stylus. The display can be confi gured to show up to six reading windows and can show electrical and temperature inputs and outputs, two pressure module readings, an IDOS pressure measurement, and a USB-connected device, as well as a HART summary window or full HART communicator with active mA measurement and actual primary variable readings.
A Compact and Powerful Electrical Calibrator and HART Communicator
The DPI 620 electrical calibrator can measure and source mA, mV, V, Ohms, frequency and a variety of RTDs and T/Cs. It provides an isolated 24 V loop power supply to energise devices and control loops and a stabilized dc voltage supply for ratio metric transducers.
The reading windows also act as active keys and, when pressed, they expand to show more detailed information or to provide advance functions such as step and ramp outputs. Colour is used to emphasize important information, for example, red for critical or failed values and green for pass or within tolerance.
The DPI 620 includes comprehensive documenting features for automating instrument calibration when used in conjunction with Intecal Calibration Management software.
The compact and lightweight calibrator is designed for rugged fi eld application, with a robust casing and weather-proofi ng to IP 65. It features a high-speed digital interface for connecting to pressure modules and pressure generating stations. This interface will also allow expansion to meet future requirements when new modules and stations become available. It also features USB connectivity and a microSD card slot with 2 Gb of memory installed as standard.
4
High performance HART communicator
The DPI 620 multifunction calibrator now incorporates a fully featured HART communicators in functionality, ease of use and upgrade support such that a single hand held device can now confi gure the full range of HART devices and perform mA trims and sensor calibrations without the need for secondary equipment such as ammeters, calibrators, power supplies and loop resistors. This can signifi cantly reduce equipment inventory, cost of ownership and greatly simplify maintenance activities The DPI 620 contains the latest application software and a complete library of registered HART device descriptions. Upgrades can be simply downloaded free of charge from our website.
®
communicator which surpasses market leading
The DPI 620 CE Advanced Modular Calibrator and HART Communicator
A Multifunction Electrical Calibrator and HART Communicator with Windows CE
The Windows CE-driven version of the electrical calibrator provides all the computing power of a conventional hand-held PC or PDA. It provides standard Windows fi le management and off ers popular Windows features to allow users to create text documents and view common fi le types including Excel, Power Point, Word, pdf and images. This means that operators can consult user manuals, training presentations, data sheets, installation drawings etc while in the fi eld or on the plant.
5
The DPI 620 CE with WiFi Advanced Modular
Calibrator and HART Communicator
A wirless IEEE 802.11g enabled Multifunction Electrical Calibrator and HART Communicator
The DPI 620 CE version can be further enhance with wireless IEEE
802.11g communications. For the fi rst time, in a calibrator of this
type, it is possible to link to the Internet and remote networks in order to access information and to transfer data. This powerful feature will benefi t service technicians who spend extended periods away from head offi ce and for those who need instant access to data, safety information, system drawings, product datasheets etc., while on the move. It will also provide an interface to future system modules when a physical connection is a hindrance.
Technical Specifi cations
General Specifi cations
Display Size: 110 mm (4.3 in) diagonal; 480 x 272 pixels
Languages English {Default} Operating temperature -10° to 50°C (14° to 122°F) Storage temperature -20° to 70°C (-4° to 158°F) Ingress Protection IP65 (Dust-tight, jets of water) Humidity 0 to 90% RH Non condensing Shock / Vibration BS EN 61010:2001; Def Stan 66-31, 8.4 cat III, 1 m Drop Tested EMC Electromagnetic compatibility: BS EN 61326-1:2006 Electrical safety Electrical – BS EN 61010 : 2001 Pressure safety Pressure Equipment Directive - Class: Sound Engineering Practice (SEP) Approved CE Marked Size (L: W: H) DPI 620 only: 183 x114 x 42 mm (7.2 x 4.5 x 1.7 in) + MC 620: 265 x 114 x 64 mm ( 10.4 x 4.5 x 2.5 in)
Weight DPI 620 only: 575 g (1.3 lb) – battery included. MC 620 only: 640 g (1.4 lb). PM 620 only: 100 g (0.2 lb). Power supply
Duration Measure functions (CH1): 12 hours continuous. Dual Function, mA measure (CH2): 7 hours (24 V Source at 12 mA)
LCD: Colour display with touch-screen
+ PM 620: 265 x 114 x 93 mm (10.4 x 4.5 x 3.7 in)
Lithium-Polymer battery (GE Part number : IO620-Battery); Capacity: 5040 mAh (minimum), 5280 mAh (typical); Nominal voltage:
3.7 V.
Charge temperature: 0° to 40°C (32° to 104°F) Discharge temperature: -20° to 60°C (-4° to 140°F).
Note: For best battery performance, keep the temperature less than 60°C (140°F). Charge/discharge cycles: > 500 > 70% capacity.
6