GE
Sensing & Inspection Technologies
Druck DPI 620
advanced modular calibrator
safety and quick reference guide - K0454
A1.1 DPI 620: Channel 1 (CH1)
Measure (M) / Source (S) / Power (P)
a ±30 V (M)
b 0 to 12 V (S)
±2000 mV (M)
c
0 to 2000 mV (S)
d 0 to 4000 Ω (M/S)
e 0 to 50 kHz (M/S)
Switch (M)
f
g ±55 mA (M)
h 0 to 24 mA (S)
i 11 RTDs (M/S) Figure C3 (2, 3, 4-wire)
j 12 TCs (M/S) Figure C4
k ±300 Vac: AC probe Figure C5 (Part: IO620-AC)
A1.2 DPI 620: Channel 2 (CH2)
l ±30 V (M)
m ±2000 mV (M)
n ±55 mA (M)
o 0 to 24 mA (S)
p 24 V loop (P)
Switch (M)
q
Figure C1
Note: = See figure or
table.
Figure C2
Figure D1
A1.3 DPI 620 + MC 620 + PM 620
Pressure* (M) Figure E1
Gauge: 25 mbar to 200 bar (0.36 to 3000 psi)
Absolute: 350 mbar to 1000 bar (5 to 15000 psi)
Note: Maximum pneumatic pressure: 500 bar (7250 psi)
*Caution: To prevent damage to the PM 620 module, only
use it within the specified pressure limit on the label.
© 2008 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.
A2.1
A1.1 ( a, b, c, d, e, f )
V, mV, Ω, Hz,
A1.2 ( l, m, n, o, p, q )
V, mV, mA, 24 Vo,
DPI 620
A2.2
DPI 620
C1
DPI 620: CH1
C2
DPI 620: CH1
A3
DPI 620
B1
MC 620 + PM 620
C3
C5
DPI 620: CH1
DPI 620: CH1
C4
E1
DPI 620: CH1
DPI 620 + MC 620 + PM 620
C
4
B2
DPI 620 + MC 620
+ PM 620
D1
DPI 620: CH2
K0454 Issue 1
Quick Reference
TAP
Channel Settings
1
3
2
4
Task Settings
6 8
Task Settings
5
Select Function Channel Settings
7
WARNING: Before you use this instrument, read and understand the “Safety”
section. It is dangerous to ignore the specified warnings.
Start operations (S1 to S4)
S1: Install the battery. Refer to Section 5.
S2: Power on/off sequence
ON
➀
Press and hold Press
First display Normal output
S3: Example change of function (Voltage to Current)
OFF
➀
[EN] English - K0454 Issue 1 Quick Reference 1
S4: Touch-screen operations (maximise, minimise, set voltage)
TAP
Set voltageMaximise
1
2
1
3
4
Minimise
*
*
* Alternative options for step ➀
Menu sequence
Task*
Settings
CH1
CH2
2 Quick Reference K0454 Issue 1 - [EN] English
* A channel can only have one function at a time.
Channel
Settings
Function
Units
Utility
Caption
Select
“Function”
None
Current
Voltage