To select the manual in an available language go to:
//www.gesensing.com/toolsupport/manuals.htm
Français
Pour choisir le manuel dans une langue disponible, accédez à :
//www.gesensing.com/toolsupport/manuals.htm
Deutsch
Um das Handbuch in einer vorhandenen Sprache auszuwählen, gehen Sie zu:
//www.gesensing.com/toolsupport/manuals.htm
Italiano
Per scaricare il manuale in una delle lingue disponibili consultare la pagina:
//www.gesensing.com/toolsupport/manuals.htm
Español
Para seleccionar el manual en uno de los idiomas disponibles vaya a:
//www.gesensing.com/toolsupport/manuals.htm
Português
Para selecionar o manual em uma língua disponível vá para:
//www.gesensing.com/toolsupport/manuals.htm
WARNING
Before operating this intrinsically safe instrument, read the safety instructions and
the special conditions stated on the ATEX certificate (Appendix 2)
Safety
The manufacturer has designed this equipment to be safe when operated using the
procedures detailed in this manual. Do not use this equipment for any other purpose
than that stated.
This publication contains operating and safety instructions that must be followed to
ensure safe operation and to maintain the equipment in a safe condition. The safety
instructions are either warnings or cautions issued to protect the user and the equipment
from injury or damage.
Use suitably qualified * technicians and good engineering practice for all procedures in
this publication.
Pressure
Do not apply pressures greater than the safe working pressure to this equipment.
Maintenance
The equipment must be maintained using the procedures in this publication. Further
manufacturer’s procedures should be carried out by authorized service agents or the
manufacturer’s service departments.
www.gesensinginspection.com
For technical advice contact the manufacturer.
*A qualified technician must have the necessary technical knowledge,
documentation, special test equipment and tools to carry out the required work on
this equipment.
Symbols
This equipment meets the requirements of all relevant European safety
directives. The equipment carries the CE mark.
This symbol, on the instrument, indicates that the user should refer to
the user manual. This symbol, in this manual, indicates a hazardous
operation.
This symbol, on the instrument, indicates do not throw-away in
domestic bin, hazardous material, dispose correctly in accordance with
local regulations.
iK0430 Issue No. 2
K0430 Issue No. 2ii iiiK0430 Issue No. 2
ATEX Approved Models
Introduction
These instructions detail the requirements for using the DPI 610 IS and DPI 615 IS
intrinsically safe pressure calibrator in a hazardous area. Read the whole publication
before starting.
Markings
II 1 G.................................................................... Equipment group & category
Ex ia IIC T4 Ga..................................................................Hazardous location markings
BAS02ATEX1174X................................................................................ Certificate number
1180..................................................................................................................................CE Mark
DPI 61X IS ......................................................................................Specific apparatus type
(Pressure Range in mbar or psi........................................Full-scale pressure rating
Druck LTD. Groby LE6 OFH, UK.................... Manufacturer's name and address
SN *******/YY-MM.................................. Serial number and date of manufacture,
Year-Month.
Requirements and Conditions
Batteries
WARNING: Only replace batteries in a safe area.
•Power supply use only 6 x LR14 (C): Duracell MN1400-LR14, Procell lndustrial
MN1400-LR14, Energizer E93.KR14.C.AM2, Energizer Industrial EN93 or Varta
4014 LR14.C.AM2.
Special Conditions for Safe Use
1. The DPI 61X IS Series Pressure Calibrator is not capable of withstanding the 500V
r.m.s. electric field strength test between the external connectors and frame of the
apparatus as required by Clause 6.4.12 of EN 50020 and this must be taken into
account when using the apparatus for input measurements in a system.
2. The outer enclosure may contain light metals in the form of aluminium, magnesium,
titanium or zirconium. Therefore, the apparatus must be installed in such a manner
as to prevent the possibility of it being subject to impacts or friction. An optional
carrying case is case is available for transporting the pressure calibrator to and from
the location of use. When the carrying case is used, special condition for safe use
item #2 does not apply.
Electrical Parameters
Maximum Output Parameters at the External Measurement Connectors:
lin (SK1) Vin (SK2) SwitchIn (SK3)
Uo= 1.1V d.c.
= 0.16 mA d.c.
I
o
= 0.15 mW
P
o
= 0.05 µF
C
i
L
= 0
i
U
= 1.1V d.c.
o
= 0.11 µA d.c.
I
o
= 0.03 µW
P
o
= 0
C
i
L
= 0
i
U
= 1.1V d.c.
o
= 12 mA d.c.
I
o
= 11 mW
P
o
= 0.05 µF
C
i
L
= 0
i
lout (SK6) RS232External Transducer
Uo= 7.9V d.c.
= 0
C
i
L
= 0.1 mH
i
= 7.6V d.c.
U
o
= 82 mA d.c.
I
o
P
= 162 mW
o
C
= 0
i
= 0
L
i
U
= 250V
m
= 7.6V d.c.
U
o
= 155 mA d.c.
I
o
P
= 0.43 W
o
C
= 0.15 µF
i
= 0.9 mH
L
i
C
= 8.6 µF
o
= 0.3 mH
L
o
The output parameters at sockets SK1, SK2 and SK3 do not exceed the values specified
in Clause 5.4, Simple Apparatus, of EN 50020.
Maximum Safe Input Parameters:
Sockets SK1, SK2, SK3 and SK6
Ui= 30V
li= 100 mA
P
= 1.0W
i
Installation
WARNING: Do not use tools on the pressure calibrator that might cause
incendive sparks - this can cause an explosion.
•Provide additional protection for equipment that may be damaged in service.
•Installation should be carried out by qualified plant installation technicians in
compliance with the latest issue of EN 60079-14.
Declaration Requirements
The DPI 610 IS and DPI 615 IS pressure calibrators are designed and manufactured to
meet the essential health and safety requirements not covered by the EC Type
Examination Certificate BAS02ATEX1174X when installed as detailed above. The
intrinsically safe pressure calibrators are designed and manufactured to protect against
other hazards as defined in paragraph 1.2.7 of Annex 11 of the ATEX Directive 94/9/EC.
K0430 Issue No. 2iv
Specification
Safe working pressure
20 bar range (300 psi)1.75 x full-scale
350 bar range (5000 psi)1.2 x full-scale
400 bar range (6000 psi)1.5 x full-scale
All other ranges2 x full-scale
Accuracy
Combined non-linearity, hysteresis and repeatability
±70 mbar range (2 inHg)0.05% F.S.
up to ±150 mbar (4.4 inHg)0.05% span
200 mbar to 20 bar (3 psi to 300 psi) [Calibrator]):0.025% F.S.
35 bar to 700 bar (500 psi to 10000 psi) [Indicator]0.025% F.S.
70 bar to 400 bar (1000 psi to 6000 psi) [Hydraulic]0.025% F.S.
Pressure Ranges
Refer to the pressure range matrix in the data sheet .
Temperature Effects
±0.004% of reading/°C (averaged over -10° to +40°C w.r.t. 20°C)
±0.002% of reading/°F (averaged over +14° to 104°F w.r.t. 68°F)
Power supply
Batteries6 x 1.5 V C cells, alkaline (up to 60 hours nominal use at 20°C)
Voltage Inputs
Range:±30V
Accuracy±0.05% rdg, ±0.004% F.S.
Resolution100µV max
Current Inputs
Range:±55mA
Accuracy±0.05% rdg, ±0.004% F.S.
Resolution1µA max
Current sink
Range:24mA
Accuracy±0.05% rdg, ±0.01% F.S.
Resolution1µA max
Display
Size:60 x 60 mm (2.36” x 2.36”) LCD Graphics
Reading±99999, update rate 2 readings/sec
Environment
Operating Temperature:-10°C to 50°C (+14°F to 122°F)
Calibrated Temperature:-10°C to 40°C (+14°F to 104°F)
Storage Temperature:-20°C to 60°C (-4°F to 140°F)
Calibration Temperature: 21°C ±2°C (70°F ±4°F)
Sealing
Sealed to IP54 (NEMA 4)
Physical
Size:300 x 170 x 140 mm (11.8” x 6.7” x 5.5”)
Weight:3 kg (6.6lb)
Fitting Batteries 7
Switching On 7
Change Pressure Units8
Voltage and Current Measurement 8
Typical Calibration Set-up (Pressure to Voltage)9
Zero Display Reading9
Task Selection
Task Key10
Using Task Functions10
Set Units 10
Cal Mode (DPI 615 instruments only) 11
Basic Mode (Task BASIC)11
Taking Measurements
Pressure Transmitter (P-I) Task12
Voltage Output Pressure Transmitter (P-V) Task12
Pressure Converter (P-P) Task13
Current to Pressure Converter (I-P) Task14
Pressure Switch Test (P-Switch) Task14
Pressure to Display (P-Display) Task15
Leak Test (Leak Test) Task16
Transmitter Simulator (TX SIM) Task17
Relief Valve Test (REL VALVE) Task18
K0430 Issue No. 2vi
Advanced Task
General19
Select Input19
Ambient Temperature Measurement19
Process Functions20
Tare Process Function21
Min/Max Process Function22
Filter Process Function22
Flow Function23
% Span23
Select Output24
Electrical Outputs (Loop Power)24
mA Step25
mA Ramp26
mA Value27
Define New Task28
Clear Task28
Memory Operations
Saving Display or Data Log29
Store Operations (Screen Snapshots)29
Recalling Stored Data (Screen Snapshots)29
Datalog Operations 30
Auto Log (Timer)30
Manual Logging30
Recall Data Log Files31
Uploading Data Log Files32
Delete Data Log and Procedure Files 32
Downloading Procedure Files (DPI 615 instruments only) 33
Running Procedure Files (DPI 615 instruments only) 34
Recalling Data Files (DPI 615 instruments only)35
Using Set-up
General36
Store Mode36
Contrast36
Settings - Select Set-up Option37
Units37
Define Special Units37
Language38
RS23238
Powerdown39
Calibration39
viiK0430 Issue No. 2
Date and Time (Real Time Clock)40
Date Format40
Set Date 40
Set Time 40
Calibration
General41
Calibration Check41
Calibration Adjustment41
Guide to Calibration Procedures41
Test Equipment42
Using the Calibration Menu43
Change PIN43
Calibrate Internal Ranges43
Internal Pressure Range44
Voltage Input Range (5 Volts)45
Voltage Input Range (30 Volts)47
Current Input Range (55 mA)49
Current Output Range (24 mA)51
Ambient Temperature Channel54
Calibrate External Sensors55
Add External Sensor56
Hydraulic Calibrator Versions
Introduction59
Safety Instructions60
Preparation for Use60
Bleeding the System61
Operation62
Draining the Hydraulic Fluid62
Flushing, Replenishing or Changing the Hydraulic Fluid63
The DPI 610 IS and DPI 615 IS intrinsically safe instruments measure and display
pneumatic and hydraulic pressure applied to the test port. Pressure measurement can
be absolute, gauge and sealed gauge and in ranges from 2.5 mbar to 700 bar (1.0 inH
to 10000 psi).
Calibrator versions of this instrument contain pneumatic or hydraulic pressure
generation components to produce pneumatic pressure ranges between -1 to 20 bar (-
14.5 psi to 300 psi) and hydraulic pressure ranges up to 400 bar (6000 psi)
Using external electrical connections, the DPI 610 IS and DPI 615 IS intrinsically safe
instruments measure ±30 volts d.c. and ±55 mA. An integral sensor provides
measurement of ambient temperature. Additional sensors (option B1) connect to an
external connector and extend the pressure measurement range and include differential
pressure measurement. The DPI 615 instrument has an RS232 connector to enable
downloading of test data to a compatible documenting system. Six alkaline C size
batteries, IEC Type LR14, power the instrument.
Important Notice
Zinc-carbon and zinc-chloride cells must NOT be used in this instrument.
Use only the battery types as shown in the table on page 7.
Description of Procedures
The procedures apply to both the DPI 610 IS and the DPI 615 IS instruments unless
otherwise stated. In the procedures in this manual, hard (fixed function) and soft
(variable function) key operations are shown in bold type : TASK and F1. These
statements mean press the TAS K key and press the F1 key. Soft key operations can be
assigned to both the F1 and F2 keys. Where a specific soft function is referred to it is
written in bold italics (e.g.) PROCESS.
This instrument has a number of operating modes that are described in simplified form in
the following sections. Diagrams accompanying the procedures give typical selection
sequences and shaded controls indicate that this control key should be pressed in the
appropriate sequence. Diagrams should be read from left to right, top to bottom where
appropriate. A shaded display soft box indicates that the function key immediately
below that soft box should be pressed (either F1 for the left hand soft box or F2 for the
right).
2
O
In the above diagram the following key sequence is indicated.
(a)Press the F2 key (the key immediately below the UNITS soft box).
(b)Use the Up and Down cursor keys (only) to select the required option. (If all keys
shaded, use all these keys to select or enter data).
(c)Press the ENTER key.
1K0430 Issue No. 2
INTRODUCTIONSummary of Functions
Using This Guide
The following key symbols are used in the procedure diagrams:
Shaded cursor keys indicate that a combination of these four keys, Up,
Down, Left and Right should be used to (e.g.) enter an alpha numeric
value or to select a function.
Indicates the ENTER key. Used to confirm an operation or a selection.
Shading indicates key operation.
Exit key, used to clear current menu selection and return to next menu
level above current level. Used as an escape key from current operation.
Shading indicates key operation.
The following table shows the maximum measurement input ratings of the instrument
which should not be exceeded.
PRESSURE120% FULL SCALE
VOLTAGE30 V d.c.
CURRENT55 mA d.c.
Note 1: The display flashes if the input pressure, voltage or current overrange.
Note 2: Max applied voltage for external loop supply = 30V dc (see page 8).
K0430 Issue No. 22
PRESSURE INT
bar
VOLTAGE
V
F2F1
2
3
4
5
7
6
1
CURRENT
PRESSURE
UNITS
max 30V
TASK: BASIC
DPI 615 IS
PRESSURE INT
bar
VOLTAGE
V
CURRENT
UNITS
PRESSURE
TASK: BASIC
+
-
Input display
Output display
Soft boxes
Status display
INTRODUCTIONSummary of Functions
OPERATOR CONTROLS (Figure 1 and 2)
These divide into two groups, the operator/display controls (Figure 1) and the pressure/
vacuum generation components (Figure 2). The operator controls and a typical display,
common to all instrument versions, is shown below.
The display and key-pad of the instrument divides into four distinct sections. The two
main sections of the display are used to show an input and an output. The remaining
two sections show the status display area and define the soft key functions. A typical
display is shown below:
3K0430 Issue No. 2
INTRODUCTIONSummary of Functions
HARD KEY FUNCTIONS (Fig. 1)
KeyFunction
I/O
SETUP*
ZERO
INPUT*
TAS K
OUTPUT*
STORE*
RECALL*
This key selects the instrument ON and OFF.7
The
SETUP
configuration parameters that are set-up to certain default
parameters on delivery.
The
the display reading is within 5% of zero. Attempts to zero a larger
offset result in an error message,
The
The
different types of external device calibration. There are twenty task
configurations available, eleven pre-programmed configurations and
nine user def ined configurations
The
Depending upon how the instrument’s
is used either to store up to 20 display screens (in
or to manually log a screen in
This key recalls a previously stored screen to the display. Depending
on the
snapshot of a previously stored screen or data log file. In
mode, selection displays the last screen stored. By using the cursor
keys, the operator can scroll either forward or back through memory
locations.
key provides access to the instrument’s general
ZERO
key zeroes either the selected input or output display, only if
INPUT
key selects the input parameter to be displayed.18, 19
TAS K
key rapidly configures the instrument for a number of
OUTPUT
key selects the output parameter to be displayed.24-27
DATALOG
STORE
mode set-up, operation of this key recalls either the
Zero too large.
STORE
mode is set-up, this key
mode.
SNAPSHOT
mode),
STORE
Page
reference
36
9
10
29, 36
28, 31, 35
The
ENTER
ENTER
EXIT
conjunction with the soft keys, accepts a given selection.
The
keys to exit from the current screen or menu level, to the level
immediately preceding it. To quit completely from any menu level,
press
key either enters data (accept entered data), or, in
EXIT
key operates in conjunction with all the other hard and soft
EXIT
until the
MEASURE/SOURCE
screen is displayed.
* These key functions are not available in BASIC mode
K0430 Issue No. 24
2
2
7
8
9
5
6
4
3
10
11
1
2
INTRODUCTIONSummary of Functions
SOFT KEYS (Fig. 1)
Three soft keys, designated F1, EXIT and F2, are situated immediately below the display
as shown below. These keys have their function allocated by the instrument software
which is indicated in the bottom of the display (Voltage for F1 and Units for F2 in this
example). They are used to select menu (program) options and are fully described under
the appropriate section headings.
CURSOR KEYS (Fig. 1)
The cursor keys consist of a block of four keys, designated up , down , left , and
right . In programs where options need to be selected from a list, (e.g.) the TAS K
selection program, the up and down cursor keys are used to highlight one of the
options, from which it can be selected by the ENTER key. In TA SK mode, where more
than one page of options are provided, the left , and right cursor keys will switch
between pages.
1Test port, connect to unit under test2Hard keys
3Cover (external interfaces)4Electrical inputs
5Cursor keys6Function (soft) keys
7Release valve (releases pressure through 8)8Vent port
9Select positive or negative pressure10 Pump
11 Fine pressure adjuster
Figure 2 - DPI 610/615 IS Calibrator Controls
5K0430 Issue No. 2
1
2
3
USERS232
INSAFE
AREAONLY
4
PRESSURE INT
bar
VOLTAGE
V
F2F1
CURRENT
UNITS
INPUTS
OUTPUTS
4
3
1
2
PRESSURE
max 30V
CAT II
TASK: BASIC
DPI 615 IS
INTRODUCTIONSummary of Functions
ELECTRICAL CONNECTIONS
1Cover, closed when
not using connectors
2External transducer
3RS232 connector
4Temperature sensor
Figure 3 - Electrical System Connections
Measurement inputs and Source outputs are made via the control panel sockets as
shown below:
WARNING: BATTERIES MUST ONLY BE FITTED IN A SAFE AREA.
USE ONLY THE BATTERIES SPECIFIED IN THE TABLE.
Caution: Old batteries can leak and cause corrosion. Never leave discharged batteries in
Switching On
Press the I/O switch on the front panel and proceed as follows:
The first time that the instrument is powered up, it will power-up in BASIC mode with the
main screen displaying voltage in the input display area and pressure in the source
display area. To switch to Current as input, press F1 as shown. Similarly, F1 to return to
Voltage.
Note: No other keys are active in this mode and the instrument can only be reconfigured
by pressing the TASK key and selecting another mode.
the instrument. Old batteries should be treated as hazardous waste and
disposed of accordingly.
7K0430 Issue No. 2
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
V
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
Current
Maximum input current = 55mA dc (at 30 V dc)
Voltage
Maximum applied voltage = 30V dc.
Getting Started
Change Pressure Units
To change the pressure units proceed as follows. If the four units displayed are not the
units required, press TAS K and select any task, other than BASIC, press SETUP and
proceed as detailed on page 36. To return to BASIC mode, press TAS K and select BASIC.
In BASIC mode, the unit is configured to carry out basic Pressure to Voltage (P to V) or
Pressure to Current (P to I) tests, a typical test procedure follows:
Voltage and Current Measurements
Connect the electrical input sockets as follows for voltage and current measurements.
Use the test leads provided and DO NOT push bare wires into the sockets.
K0430 Issue No. 28
P
V
+
A
B
PRESSURE INT
bar
VOLTAGE
F2F1
EXIT
V
CURRENT
PRESSURE
UNITS
Max30V
CATII
TASK:BASIC
DPI 615 IS
CD
3
--
+
+
-
-
C
D
E
F
D
Getting Started
Typical Calibration Set-up (Pressure to Voltage)
Connect a device under test to the instrument as shown below:
A - External pressure source (indicator instruments only) B - Pressure regulator
C - Pressure/voltage device D - Barrier E - Excitation 10V F - Safe area
•General Procedure
Use the hand-pump to pressurize the system to the required level as indicated on
the display. Allow the display to settle and screw the volume adjuster in or out as
a fine adjustment to the required pressure. Record the input: Voltage, reading at
each applied pressure.
Zero Display Reading
Both the input and output readings can be set to zero by using the ZERO key and if the
displayed reading is within 5% of zero. To zero either the INPUT or OUTPUT displays,
proceed as follows:
9K0430 Issue No. 2
Task Selection
Task Key
The TASK key is used to set-up the instrument for a number of specific types of test.
There are two modes BASIC and ADVANCED and nine other specific types of test which
automatically configure the instrument on selection from the TAS K menu. The tasks
available under the TASK menu are held on three pages shown below. To select a task
from the menu, press the TASK key, position the cursor over the required task and press
the ENTER key as shown below. Use the right/left cursor keys to switch between pages.
Using TASK Functions
Specific tasks are selected as shown above. The following diagrams show how to
connect the unit under test (UUT) for each task selectable under the TA SK menu.
Input and output units, where applicable, can be selected by pressing either the INPUT or
OUTPUT keys as shown below.
Set Units
Note: If the four units displayed are not the units required, press SETUP, select SETTINGS
and refer to select regular units on page 37.
K0430 Issue No. 210
0.00
2000.00
OUTPUT
MAX. ERROR
ERROR TYPE
CHANGE
VALUE
psi
=
=
=
=
=
mA
4.00
20.00
LINEAR
0.05%
%span
PI
SELECT VALUE
psi
mA
TASK : P-I
DATALOG
24V OFF
..........
.%span FAIL
TURN OFF
CAL MODE
TURN ON
CAL MODE
mA
TASK : P-I
DATALOG
24V OFF
..........
psi
Task Selection
Cal Mode (DPI 615 versions only)
Cal mode, which is available in tasks P-I, P-P, P-V, P-P, P-DISPLAY and P-SWITCH, provides
a method of setting up test parameters manually. Downloaded test procedures can also
automatically set up and turn on the Cal Mode function. The method of turning on and
setting up Cal Mode is shown below for a P-I task. A similar method can be used for all
the other tasks applicable to the Cal Mode function.
Pressing the F1 key (TURN ON CAL MODE), provides the set-up screen for the CAL mode.
Initially, the cursor is placed in the UUT SPAN field to allow the required span range to be
entered. The corresponding values for the UUT output parameter (current) are then set,
followed by the maximum error value and error type (%rdg or % span). When all test
parameters have been set-up, the screen changes to display the input and output and
the test results. The test result can only be displayed to within a range of ±9.99%. If the
test result is outside this range, either the left pointing (-ve error) or right pointing (+ve
error) chevrons are displayed. Within this error band, the actual tolerance value is
displayed. Test results can either be stored as snapshots or logged as data log files,
depending on how the instrument has been set-up.
Basic Mode (Task BASIC)
This instrument will power-up in this mode the first time that it is used. To select BASIC
from any other task, press the TA SK key and select BASIC and press the ENTER key.
BASIC mode is fully described in the Getting Started, section (see page 7).
11K0430 Issue No. 2
A
B
PRESSUREINT
bar
CURRENT
mA
TASK:P-I
SNAPSHOTMODE
DPI 615 IS
F2F1
EXIT
Max30V
CATII
3
P
I
+
-
+
-
C
D
E
D
F
P
V
+
A
B
PRESSURE INT
bar
VOLTAGE
F2F1
EXIT
V
CURRENT
PRESSURE
UNITS
Max30V
CATII
TASK:BASIC
DPI 615 IS
CD
3
--
+
+
-
-
C
D
E
F
D
Taking Measurements
Pressure Transmitter (P-I) Task
Select the P-I task from the task menu and connect the Unit Under Test (UUT) to the
calibrator as shown below:
A - External pressure source (indicator instruments only) B - Pressure regulator
C - Pressure to current device D - Barrier E - External supply F - Safe Area
•If required, select the output units as described on page 10.
•If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
Voltage Output Pressure Transmitter (P-V) Task
Select the P-V task from the task menu and connect the Unit Under Test (UUT) to the
instrument as shown below:
A - External pressure source (indicator instruments only) B - Pressure regulator
C - Pressure to voltage device D - Barrier E - External supply F - Safe Area
•If required, select the output units as described on page 10.
•If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
K0430 Issue No. 212
USERS232
INSAFE
AREAONLY
3
A
B
C
D
E
Taking Measurements
Pressure Converter (Pressure to Pressure) Task
Select the P-P task from the task menu and connect the Unit Under Test (UUT) to the
calibrator as shown below. Testing a converter requires one pressure to be applied to
the unit under test (UUT) and another (converter output) to be measured. The additional
measurement is provided by the external transducer option.
Method
•Connect the UUT to the instrument as shown below. Plug the external transducer
into the instrument as shown below:
A - External pressure source (indicator instruments only) B - Pressure regulator
C - External pressure sourceD - Pressure to pressure device E - External transducer
•Press the TASK key and select the P-P task. Providing the external transducer has
been calibrated and its parameters stored in the instrument, the display will
show External pressure in the input window and calibrator Output pressure in
the output window. If an error message “NO SENSOR OR CAL INVALID” is
displayed, this indicates that the external transducer has not been entered and/
or calibrated with the instrument. Refer to page 56 for details of adding an
external transducer. If an external transducer change is made, switch the
calibrator off and then on to load new transducer data.
•If required, select the input and output units as described on page 10.
•If applicable, turn on Cal Mode and set-up test parameters as detailed on page
11.
13K0430 Issue No. 2
P
I
-
PRESSURE INT
bar
OUTPUT
NEW VALUE
CURRENT
mA
OK
TASK: I-P
SNAPSHOT MODE
F2F1
EXIT
Max30V
CATII
DPI 615 IS
CD
3
+
-
+
-
C
C
E
D
A
B
PRESSURE INT
bar
F2F1
RUN
CONTACTSTATE
max30V
TASK: P-SWITCH
SNAPSHOT MODE
DPI 615 IS
A
B
3
C
Taking Measurements
Current to Pressure Converter (I-P) Task
A - External pressure source B - Pressure to current device (24V)C - Barrier
D - Safe areaE - External supply
•Use the up and down cursor keys to adjust the loop current to the required
value. Alternatively, press ENTER and use cursor keys to enter a finite value.
Cursor keys can then be used to nudge the output either up or down.
If required, change pressure units with INPUT key. A flashing CHECK LOOP
message indicates either an open circuit supply loop (or no external supply).
Pressure Switch Test (P-SWITCH) Task
A - External pressure sourceB - Pressure regulatorC - Pressure switch under test
•Contact state will be shown on display. When contacts close, buzzer sounds.
•To run switch test, close vent valve and press the RUN (F1) key.
•Using the hand-pump, increase the applied pressure to just below the switch
operating point. Screw the volume adjuster in until the switch operates (the
operating pressure of the switch is then written to the display).
•Reduce pressure until the switch releases (indicated by the switch symbol). The
release pressure is then written to the display and the hysteresis displayed.
K0430 Issue No. 214
A
B
PRESSURE INTbar
F2F1
DISPLAY
CHANGE
VALUE
TASK : P-DISPLAY
SNAPSHOT MODE
max30V
DPI 615 IS
C
Taking Measurements
Pressure to Display (P-DISPLAY) Task
P-Display is a special application of Data Log. To use this mode, select Data Log from the
Store Mode menu as detailed on page 36. Connect the UUT to the instrument as shown
below and, if required, turn on and set-up Cal Mode (see page 11).
A - External pressure sourceB - Pressure regulatorC - Dial gauge under test
•Press TASK and select P-DISPLAY. If required, use OUTPUT key to change
pressure units.
•Set-up a data log file as detailed on page 30.
Note: TRIGGER field, automatically set to KEYPRESS, cannot be changed.
•Apply a series of test pressures to the device under test. Enter displayed
reading at each pressure and log each point:
•After logging final test point, terminate as follows:
15K0430 Issue No. 2
A
B
PRESSURE INT
bar
F2F1
RUN
CHANGE
VALUE
WAIT
DURATION
STARTPRESS
STOP PRESS
PRESS CHANGE
LEAK RATE
secs
secs
bar
bar
bar
bar/m
60
60
max30V
TASK: LEAK TEST
SNAPSHOT MODE
DPI 615 IS
C
Taking Measurements
Leak Test (LEAK TEST) Task
A - External pressure sourceB - Pressure regulatorC - Device/system under test
•If required, use the INPUT key to change pressure units.
•Set-up the leak test WAIT and DURATION times to the required values as shown
below. A minimum wait period of 3 minutes is recommended.
•Close the vent valve and pressurize the device/system to the required LEAK TEST
pressure.
•Press the RUN (F2) key to start the leak test. When completed, the beeper sounds
and the leak test results are written to the display.
K0430 Issue No. 216
Taking Measurements
Transmitter Simulator (TX SIM) Task
When used with an external voltage source (see page 24), provides a current output
proportional to the calibrator’s measured output pressure (indicated pressure on
indicator only version). Select task TX SIM. Press EXIT to skip set-up screen if parameters
are correct.
On completion of Tx SIM set-up, the display is configured as follows :.
Connect an external power source to the output loop as detailed on page 24.
To subsequently change any of the Tx SIM scaling parameters, press CHANGE VALUE key
(F1) to obtain the TX Simulation set-up display.
To change the pressure units, press INPUT and select the required scale units. If the
required scale units are not listed, press SETUP, select SETTINGS and proceed as detailed
on page 37.
17K0430 Issue No. 2
PRESSURE INT
bar
bar
RESET
MAXIMUM
MINIMUM
MAX/MIN
F2F1
A
B
C
PRESSURE INT
max30V
TASK: RELVALVE
SNAPSHOT MODE
DPI 615 IS
3
Taking Measurements
Relief Valve Test (REL VALVE) Task
To carry out a relief valve test, press TAS K and select REL VALVE. Connect the output
pressure port of the instrument to an external system as shown below:
A - External pressure source (indicator only) B - Pressure regulatorC - Relief valve under test
•To change the pressure units, if required, press INPUT and select the required
units using the cursor keys.
•Close the vent valve and, using the hand-pump or external pressure supply, apply
pressure to the relief valve under test.
•When the relief valve operates, the maximum recorded pressure indicates the
operating point of the valve.
•Record the test results.
Note: The STORE key can be used for this purpose. Use right cursor key initially,
followed by up/down keys to enter Snapshot text.
•Open vent valve to release test pressure.
Note: If using external pressure supply, isolate supply before opening the vent
valve.
K0430 Issue No. 218
SELECT UNITS
SELECT AIR
TEMPERATURE
INPUT
SELECT INPUT
SELECT INPUT
PROCESSUNITS
UNITS
CELSIUS
FAHRENHEIT
AMBIENT TEMP
NO INPUT
Advanced Task Select Input
General
Advanced task allows the user to configure the instrument to monitor one of a number of
different input measurements and outputs (sources). Additionally, five process functions,
Tare, Max/Min, Filter, Flow and % Span can be applied to the input functions.
Select Input
To select an input channel, select ADVANCED task from the task menu. The display
shows the list of the input selections and, if available, the PROCESS soft box (F1) and the
UNITS soft box (F2).
The following procedure shows the method of input channel selection and the method of
changing units:
Note: Left/right arrow keys function as page up/down keys.
Refer to pages 20 to 23 for details of process functions.
Ambient Temperature Measurement
To set-up the instrument to read ambient temperature, proceed as follows:
Note: Ensure that the temperature reading has stabilised.
19K0430 Issue No. 2
Loading...
+ 67 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.