These fully self-contained pressure test and calibration systems have interchangeable pressure
modules for exible ranging, generate pneumatic or hydraulic pressures, measure process
signals and supply loop power. They provide all the convenience of the Druck DPI 610/615 series
with signicantly improved generation capabilities, higher accuracy and simplied touch screen
operation.
Features
• The complete pressure calibration
tool: pump, measure, analyse
• Best in class pressure accuracy
• Fast and dependable from vacuum
to 1000 bar/15000 psi
• Easy to use touch screen
• No in-the-eld leaks with quick-to-t
pressure adaptors & hoses
Page 2
Druck DPI 612 Flex Series
Precision engineering
Range Flexible Pressure Calibrators
The Druck DPI 612 Flex series is the fth
generation in the DPI 600 family, which was
rst introduced in 1984. The DPI 600 family
revolutionized test and calibration by providing
all the tools for pressure generation and signal
measurement in self-contained portable
packages. The DPI 600 soon became the industry
workhorse and today it is simply known as the
“Druck.”
Building on the technical legacy and more than
three decades of experience in pressure sensor
development and calibration, the DPI 612 Flex
series provides all the convenience and reliability
of a true “Druck,” yet oers interchangeable
pressure modules, higher accuracy and
signicantly improved pressure generation.
Dependable and ecient pressure
generation
Performance is a function of precision engineering.
The innovative design of the DPI 612 Flex series
pressure systems achieves ecient generation and
precise control through the use of carefully selected
materials with high tolerance machining and perfect
nishes.
The choice of case materials and precision
moulding ensures that DPI 612 Flex are rugged and
weatherproof.
State-of-the-art analogue and digital
microelectronics contribute both accuracy and
processing power to provide a class leading pressure
and electrical capability with unique pressure re-
ranging and simple-to-use interface.
Three model choices to generate from 95%
vacuum to 1000 bar/ 15000 psi
• 20 bar/300 psi in 30 seconds
• 100 bar/1500 psi without gas bottles and
regulators
• 1000 bar/15000 psi hydraulic pressure
Flexible pressure ranging – swap
the module, not the instrument
• Re-range for dierent applications and expand
the system over time
• 31 interchangeable pressure modules from 25
mbar/10 inH20 to 1000 bar/15000 psi
• Accuracy from 0.005% FS
• “Hot swap” without tools, seals or cables
For convenience and to protect your budget,
the pressure modules are fully interchangeable
between the DPI 612 Flex series and the DPI 620
Genii multi-function calibrator / communicator
series.
Leak free, eld serviceable
pressure connections
• “Quick-to-t” pressure adaptor/hose system
requires no tools or sealing
Simple touch screen operation
• Application DASHBOARD, quick TASK selection
and FAVOURITES storage
• Fast, three touch set-up for any application
• Calculates PASS/FAIL errors, documents results
and interfaces with calibration software
Page 3
PERFORMANCE IS A FUNCTION
PERFORMANCE IS A FUNCTION
OF PRECISION ENGINEERING
OF PRECISION ENGINEERING
Page 4
Pressure generation
Quick, eortless and dependable
There are three models in the Druck DPI 612 Flex series to
generate pressures from 95% vacuum to 1000 bar/ 15000
psi. Their dependable, high performance pressure systems
provide quick and easy generation with great accuracy.
• DPI 612 pFlex: 95% vacuum to 20 bar/300 psi
• DPI 612 pFlexPro: 95% vacuum to 100 bar/1500 psi
• DPI 612 hFlexPro: 0 to 1000 bar/15000 psi
The pFlex pneumatic versions have a simple selector to let
you convert from vacuum to pressure and with the high
eciency hand pumps and volume adjusters you can
generate accurate pressure quickly and with ease.
Pressure accuracy
Druck technology throughout
Using advanced “Druck” silicon
technology from the agship DPI 620
Genii series, the PM 620 interchangeable
pressure modules achieve up to 0.005%
FS accuracy.
Taking stability and temperature
errors into account over the one year
calibration period, the DPI 612 Flex
is more than twice as accurate as its
predecessor. This is expressed as a
total uncertainty to give you complete condence in the
measurement accuracy between annual calibrations.
The hFlex hydraulic version has an internal reservoir for
hydraulic oil or water, a priming pump to expel air from the
connected system and an intensier to quickly and easily
generate pressure.
• They’re fast, 20 bar/300 psi pneumatic or 1000 bar/15000
psi hydraulic in less than 30 seconds
• Generate 100 bar/1500 psi without potentially dangerous
bottled gas
• Ergonomic design and a detachable hand/shoulder strap
make the DPI 612 Flex equally at home in the eld or on a
calibration bench
• Proven mechanical designs provide simple-to-use,
dependable systems with long service life
Pressure range exibility
The closest thing to a pressure multi-
meter
The DPI 612 Flex utilises PM 620 interchangeable pressure
modules allowing in-eld re-ranging to the highest accuracy
and most suitable range for any application
Electrical capability
Just as important as pressure accuracy
The DPI 612 Flex series combines the comprehensive electrical
measurement and sourcing capability of the DPI 610 series,
with higher accuracy and simplied connections.
• Field interchangeable, automatic detection
• Simple screw t: no tools, sealing or cables required
• 31 ranges: 25 mbar/10 inH2O to 1000 bar/15000 psi
• Accuracy from 0.005% FS e.g. 0.001 bar/0.015 psi error @
20 bar/300 psi
The PM 620 is the latest development in digital output sensor
technology and incorporates key innovations to allow
pressure re-ranging without the need for tools, sealing, cables
or user set-up. The modules cost a fraction of xed range
instruments and can be calibrated independently, so you
can benet from reduced inventory, shared resources, lower
cost of ownership and less down-time for annual calibration
checks.
P1P2 IDOSmAVmV10 Vdc24 VSwitch
Measure
Source
P2 IDOS is an optional external pressure sensor.
• The DPI 612 Flex electrical measurement accuracy includes
one year stability, temperature errors and calibration
uncertainty, making it three times more accurate than the
DPI 610 series.
• For ease of use the electrical connectors have been
rationalised to four 4mm sockets.
Page 5
Quick-to-t pressure connections
No more leaks, service downtime or frustration
Making leak-tight, pressure connections is inevitably frustrating. The
DPI 612 Flex series have quick-to-t adaptor and hose systems with
time-saving advantages over conventional methods:
• All adaptors, hoses and accessories, including the dirt and moisture
traps, are quick and simple to t. No tools or sealing are required
and connections are leak free every time.
• A damaged thread on an instrument pressure port often
necessitates a costly repair and signicant downtime. With the
DPI 612 quick-to-t system damaged adaptors are very simply
replaced in-the-eld.
• Making leak tight connections wastes time and for several joints,
it can take longer than a calibration. The DPI 612 Flex system is
proven to signicantly reduce set-up time.
Simplied touch screen
How many key presses to congure your
calibrator for the next job?
DPI 612 Flex calibrators have the same user interface design as
the unique and award-winning DPI 620 Genii (Measures magazine
innovative product of 2014).
• The DASHBOARD allows quick application selection without
menus or special keys - just tap the app.
• The TASK menu provides a library of popular congurations, so
from the calibrator screen three simple gestures will completely
recongure the DPI 612 Flex for the next job.
• From the FAVOURITES menu it’s even quicker to access regularly
used and customised TASKS.
• The touch screen only shows function keys when they’re
required, making it quicker and simpler to use than complex
keypads with special function keys and key combinations.
Typical TASK
Output mA
Measure pressure
Check box to add this TASK to
FAVOURITES
Connections for mA lit in green
Easy as 1, 2, 3
1) Swipe from right to left 2) Touch to make a selection 3) Touch to select the TASK
Page 6
Fully documenting
Let the calibrator do the work
The DPI 612 Flex is a simple-to-use “everyday” tool for
maintaining and calibrating pressure instruments and
due to its processing power and advance software,
the automation of procedures reduces even the most
complex task to a few simple user actions.
• Automated calibration procedures
• PASS/FAIL error analysis
• Multi-channel data logging
• 8Gb (nominal) user memory
• Connectivity with leading calibration and
maintenance software including 4Sight from GE
Save time by automating your
calibration procedures
Procedures generated by calibration management
software can be downloaded to the DPI 612 Flex.
These are presented as a list of work orders and when
selected, each one will congure the DPI 612 Flex to
calibrate a specic device and report a PASS/FAIL
condition. The data is recorded digitally ready to be
uploaded to the management software.
Intecal calibration management software
Intecal is a simple to use single PC application which provides
an asset data base with dened procedures that can be
downloaded to the DPI 612 calibrator. It will automate your
calibration process and provide accurate audit ready data.
• User accounts and access levels
• Asset database
• Automated scheduling
• Dened procedures
• Supports portable eld and xed workshop calibrators
• Data analysis, certicate generation, archiving
• Extensive connectivity including the DPI 611, DPI 612, DPI
620 and PACE series of instruments
Please visit https://www.gemeasurement.com/test-calibration/
software/intecal-v10-calibration-management-software for
more information and a free 30 day trial.
Using the DPI 612 Flex with automated procedures
signicantly reduces calibration time, from typically
40 minutes to less than 10 minutes including the time
to set-up. Further time is saved when evaluating the
data and creating calibration reports because these
operations are automated within the software.
Avoid surprises and re-work with
PASS/FAIL error analysis
The error of a device being tested is displayed live with
a PASS or FAIL indication. Zero and span adjustments
can be evaluated in the eld as they are made.
Diagnosing faults? Data log multiple
channels
Data from three channels can be logged
simultaneously by manually touching a record button
or automatically at a user set interval. Data can be
reviewed on screen or the data le can be transferred
to a PC for further analysis.
Calibration software
Connect to leading calibration and
maintenance software
The DPI 612 Flex integrates with leading calibration
and maintenance software including Intecal and
4Sight from GE. Typically such applications provide an
automated and paperless solution to calibration and
realise signicant benets including reduced operating
costs, regulatory compliance and improved process
eciency.
4Sight calibration and maintenance
software
4Sight is the new state-of-the-art, integrated, web-based
Software as a Service (SaaS).
• Maintains compliance with industry standards
• Provides a full-time and date stamped audit trail
• Signicantly reduces your operating costs
• Provides automated paperless solutions
• Ensures that you are always ready for an audit
• Optional web hosting means no IT overhead
4Sight is a scalable solution from a single user to multiple
users across multiple sites to give you total control for all your
calibration and maintenance tasks.
4Sight will help you comply with regulations, reduce running
costs and improve process eciency. As your calibration
manager, its automated workow, robust data and
complete traceability will signicantly reduce calibration and
maintenance costs.
Please visit www.gemeasurement.com/test-calibration/
software/4sight-calibration-and- maintenance-software for
more information.
Page 7
Advanced features
Step and ramp mA output: Simply congured for
simulating transmitter outputs into control loops,
testing valve positioners and checking safety systems.
The function has programmable end points, manual
or automatic sequencing and the following options for
quick set-up:
• % step: The step size is dened as a percentage. For
example 25% provides ve test points of 4, 8, 12, 16
and 20 mA.
• Dened step: The step size is dened as a value in mA.
• Span check: Toggles between two end points, for
example, 4 and 20 mA for checking zero and FS.
• Ramp: A linear ramp between two end points with
programmable travel and dwell times is perfect for
dynamically testing switches.
Pressure leak test: This automated procedure for
detecting leaks and determining leak rates has
programmable settle and run times. The start and
end pressures are reported along with the pressure
change and leak rate.
Max/min: Captures maximum and minimum values
and calculates the mean.
25% step manual advanceRAMP automatic cycle
Nudge: Simply used to make a small incremental change
to a mA output using up/down keys. This is great for
determining trip values.
Switch test: Automates the capture of pressure switch
actuation and de-actuation values and calculates the
hysteresis.
Relief valve: Captures the venting pressure of a
relief valve.
Scaling: Scales the measured value into a process
value. For example, mA expressed as a %. Flow
correction is available for scaling dierential ow
transmitter outputs.
Pressure Resolution: Adjustable from four to seven
digits; this matches the displayed value to that of
the test device for easy comparison.
Help: The DPI 612 is provided with a multi-lingual
quick user guide to get you up and running
without delay. For convenience, the full manual is
stored digitally within the instrument and can be
transferred to a PC for viewing or printing.
In the help application you can also view wiring
connection diagrams.
Page 8
Quick-to-t pressure
adaptor system (no tools
required or seals)
Interchangeable pressure
module
Electrical connections for
measuring mA, mV, V and
detecting switch contacts.
Full colour touch screen
display
On/O/Sleep
Anti-slip feet (front and rear)
Pressure/vacuum selector
Pump
Pressure/vacuum
selector
Pump
DPI 612 pFlex
Pressure release valve
Volume adjuster for precise
pressure setting
Page 9
DPI 612 pFlexPro
Pressure release and
non-return valve
10V DC regulated supply
and 24V loop power
Hand strap attachment
USB master and
slave sockets
DPI 612 hFlexPro
Protective cover
Hand strap attachment
Combined pressure
intensier and ne
adjustment
Reservoir rell valve
Pressure release valve
Priming pump
Non-return valve
Internal uid reservoir
Combined pressure
intensier and ne
adjustment
Page 10
Druck DPI 612 Flex Series Specications
DPI 612 pFlex low to medium pressure pneumatic calibrator with range exibility
This model is ideal in low to medium pressure applications from a few mbar/psi to 20 bar/300 psi.
Ordering Information
Product Name Model Range
DPI 612
PFX
The following standard DPI 612 pFlex kits include a PM 620 pressure module and accessories as indicated
2A0 to 2 abs0 to 30 abs
7A0 to 7 abs0 to 100 abs
20A0 to 20 g/abs0 to 300 g/abs
1G-1 to 1 g-14.5 to 15 g
2G-1 to 2 g-14.5 to 30 g
7G-1 to 7 g-14.5 to 100 g
10G-1 to 10 g-14.5 to 150 g
20G-1 to 20 g-14.5 to 300 g
For other pressure range and accessory combinations please use the “00” pressure range and order PM 620 pressure
modules and accessories as separate items.
00Pressure module not included
Each DPI 612 pFlex includes battery charger ( IO620-PSU), carrying strap (IO612-STRAP), test leads (IO61X-LEADS), user guide
and calibration certicate.
Ranges shown as g/abs can be used in sealed gauge or absolute mode via a software feature.
Please refer to the PM 620 pressure module specication table for pressure accuracies
Pressure range
( g = gauge, a b s = a b s olut e )
bar psi
Rechargeable
battery
CC3800GE
Hose
IOHOSE-NP1
Carry case
IO620-CASE-3
Dirt moisture
trap
IO620-IDT621
DPI 612 pFlex general specication
Maximum working pressure20 bar / 30 psi pneumatic
Maximum vacuum-95% of atmospheric pressure
Pressure mediaNon-corrosive gases
Pressure connectionQuick-to-t connection system. G1/8 female and 1/8 NPT female adaptors are provided
Size and weight350 mm x 170 mm x 145 mm ( 13.7 in x 6.7 in x 5.7 in) excluding accessories
3.4 kg (7.5 lb) including battery and pressure module
Page 11
Druck DPI 612 pFlexPro low to high pressure pneumatic calibrator with range exibility
This model excels in a wide range of applications from a few mbar/psi through to 100 bar/1500 psi. It retains the low to medium
pressure capability of the pFlex, but incorporates an advanced generation system extending its range ve times. The pressure
intensier makes generating pressures from a few mbar/psi through to 40 bar/600 psi much quicker and easier and on the
occasions when high pressure is required from a clean gas source, the pFlexPro is a safe and easy-to-transport alternative to
bottled gas.
Ordering information Pressure
( g = g aug e , ab s = a b sol u te)
bar psi
Product nameModelRange
DPI 612
PFP
The following standard DPI 612 pFlexPro kits include a PM 620 pressure module and accessories as indicated
20A0 to 20 g/abs0 to 300 g/abs
20G-1 to 20 g-14.5 to 300 g
35G0 to 35 g0 to 500 g
70G0 to 70 g0 to 1000 g
100G0 to 100 g0 to 1500 g
For other pressure range and accessory combinations please use the “00” pressure range and order PM 620 pressure modules and accessories as
separate items.
00Pressure module not included
Each DPI 612 pFlexPro includes battery charger ( IO620-PSU), carrying strap (IO612-STRAP), test leads (IO61X-LEADS), user guide and calibration certicate.
Ranges shown as g/abs can be used in sealed gauge or absolute mode via a software feature.
Please refer to the PM 620 pressure module specication table for pressure accuracies.
Rechargeable
battery
CC3800GE
Hose
IO620-HOSE-P1
Carry case
IO620-CASE-3
Dirt moisture
trap
IO620-IDT622
Pressure relief
valve
IO620-PRV-PX
-P3
-P3
-P4
-P5
-P5
DPI 612 pFlexPro general specication
Maximum working pressure100 bar / 1500 psi pneumatic
Maximum vacuum-95% of atmospheric pressure
Pressure mediaNon-corrosive gases
Pressure connectionQuick-to-t connection system. G1/8 female and 1/8 NPT female adaptors are provided
Size and weight350 mm x 170 mm x 160 mm ( 13.7 in x 6.7 in x 6.3 in) excluding accessories
4.1 kg (9 lb) including battery and pressure module
Page 12
Druck DPI 612 hFlexPro very high pressure hydraulic calibrator with range exibility
This model generates hydraulic pressure to 1000 bar/15000 psi using water or mineral oil from its internal reservoir.
Ordering information Pressure
Product nameModelRange
DPI 612
HFP
The following standard DPI 612 hFlexPro kits include a PM 620 pressure module and accessories as indicated
200G0 to 200 g0 to 3000 g
3500 to 350 g/abs0 to 5000 g/
7000 to 700 g/abs0 to 10000 g/
10000 to 1000 g/
For other pressure range and accessory combinations please use the “00” pressure range and order PM 620 pressure
modules and accessories as separate items.
00Pressure module not included
Each DPI 612 hFlexPro includes battery charger ( IO620-PSU), carrying strap (IO612-STRAP), test leads (IO61X-LEADS), uid
bottle, user guide and calibration certicate.
Ranges shown as g/abs can be used in sealed gauge or absolute mode via a software feature
Please refer to the PM 620 pressure module specication table for pressure accuracies.
DPI 612 hFlexPro general specications
Maximum working pressure1000 bar / 15000 psi hydraulic
Pressure mediaDe-mineralized water or mineral oil (ISO viscosity grade < 22). 100cc internal reservoir
Pressure connectionQuick-to-t connection system with G1/8 female and 1/8 NPT female adaptors
Size and weight350 mm x 170 mm x 170 mm ( 13.7 in x 6.7 in x 6.7 in) excluding accessories
4.7 kg (10.4 lb) including battery and pressure module
0 to 24 mA (internal loop power)0.0180.0060.0010.001
FS = full scale. Rdg = reading
Total uncertainty includes reference standard uncertainty, temperature errors, NLH&R and typical long term stability for one
year (K=2).
Additional error
-10° to 10°C
(14° to 50°F)
& 30° to 50°C
(86° to 122°F)
%FS/°C
Resolution
General Specications
DisplaySize: 110 mm (4.3in) diagonal. 480 x 272 pixels. LCD colour display with touch screen
Internal memory8Gb (nominal) user memory for automated procedures, calibration data and data log les
Battery eliminator / charger (GE part number: IO620-PSU) supplied as standard.
Alternatively 8 AA alkaline cells (not supplied)
For AA alkaline cells 18 to 26 hours depending on functions
External pressure modules
IDOS (intelligent digital output sensor) pressure modules can be connected via a USB converter P/N IO620-USB-IDOS to
provide a second measurement channel or to isolate the pressure media from the generation systems.
Multiple parameter display capability
The display can be congured to show a maximum of three simultaneous reading windows as follows: pressure,
electrical measurement or electrical source, IDOS external pressure module
Page 14
PM 620 Interchangeable Pressure Module
The PM 620 is the latest development in digital output sensor technology incorporating
a number of key innovations to allow pressure re-ranging of compatible equipment. A
simple screw t makes both the pressure and electrical connections without the need for
tools, sealing tape, cables or plugs and digital characterisation allows interchangeability
without set-up or calibration.
PM 620 pressure module specication and recommended use
ModelPressure rangePrecision
(NLHR)
%FS
Ordering informationDPI 612 pFlex
PM 620barpsigaugeabsolutegauge
+/- 0.025 g+/-10 inH20 g0.090.1
+/- 0.07 g+/-1 g0.0250.047
+/- 0.2 g+/-3 g0.020.045
+/- 0.35 g+/-5 g0.020.044
+/- 0.7 g+/-10 g0.0150.041
+/- 1 g-14.5 to 15 g0.0150.041
-1 to 2 g-14.5 to 30 g0.0150.025
-1 to 3.5 g-14.5 to 50 g0.010.025
-1 to 7 g-14.5 to 100 g0.010.025
-1 to 10 g-14.5 to 150 g0.0050.025
-1 to 20 g-14.5 to 300 g0.0050.025
0 to 35 g0 to 500 g0.0050.025
0 to 70 g0 to 1000 g0.0050.025
0 to 100 g0 to 1500 g0.0050.025
0 to 135 g0 to 2000 g0.0050.025
0 to 200 g0 to 3000 g0.0050.025
0 to 0.35 a0 to 5 a0.03
0 to 1.2 a0 to 35 inHg a0.02
0 to 2 a0 to 30 a0.015
0 to 3.5 a0 to 50 a0.015
0 to 7 a0 to 100 a0.015
0 to 10 a-14.5 to 150 a0.0050.015
0 to 20 a-14.5 to 300 a0.0050.015
0 to 35 a0 to 500 a0.0050.015
0 to 70 a0 to 1000 a0.0050.015
0 to 100 a0 to 1500 a0.0050.015
0 to 135 a0 to 2000 a0.0050.015
0 to 200 a0 to 3000 a0.0050.015
0 to 350 a0 to 5000 a0.0050.015
0 to 700 a0 to 10000 a0.0050.015
0 to 1000 a0 to 15000 a0.0050.015
FS = full scale
NLH&R = Non-linearity, hysteresis and repeatability
Media 1 compatible with non-corrosive gas/uid
Media 2 Compatible with stainless steel
* The reading can be referenced to ambient air pressure, allowing the same module to be switched between absolute and sealed gauge measurement
DPI 612 Flex pressure resolution is adjustable from 4 to 7 digits.
Total uncertainty includes reference standard uncertainty, temperature errors, NLH&R and typical long term stability for one year (K=2).
PM 620 General Specication (where dierent from DPI 612 Flex general specication)
Maximum intermittent pressure2 x full scale pressure
Maximum working pressure110% of full scale pressure
Ingress protectionIP 65 (protected against dust and jets of water)
Size and weightL. 56 mm. Dia. 44 mm. 106 g
Total uncertainty
0° to 50°C
(32° to 122°F)
for one year
%FS
MediaRecommended PM 620 pressure module and IO620-PRV-XX pressure relief valve
Pneumatic Hydraulic
(20 bar/300 psi max)
absolute
Gauge ranges (referenced to atmosphere)
PM 620PRVPM 620PRVPM 620PRV
1
1
1
2
2
2
2
2
2
2
2
2
2
2
2
2
Absolute ranges (referenced to vacuum)
0.082
0.072
0.0522
0.052
0.052
Dual use sealed gauge and absolute*
0.0472
0.0472
P1P1
P1P1
P1P1
P6P6
P7P7
P7P7
P3P3
0.0472
0.0472
0.0462
0.0462
0.0462
0.0492
0.0492
0.0492
DPI 612 pFlexPro
(100 bar/1500 psi max)
P1
P1
P1
P1
P6
P2
P7
P7
P3
P4
P5H2
P5H2
(PRV)
DPI 612 hFlexPro
(1000 bar/15000 psi max)
P1
P1
P1
P1
P6
P2
P7
P7
P3
P4
P5
P5
H2
H2
H3
H4
H5
H2
H2
H2
H2
Page 15
Ordering Information
Please provide the ordering information highlighted in the
specication tables. For example, to order a standard DPI 620
pFlexPro kit with a 300 psi gauge pressure module order DPI
612, PFP, 20G. This will include the accessories indicated in the
DPI 612 pFlexPro specication table.
To order a DPI 612 Flex with a dierent pressure module to
those indicated in the specication table, for example, 0.35 bar
gauge, use DPI 612, PFX, 00 and then order the pressure module
PM 620, 0.35 bar g and any accessories as a separate items.
Accessories
Note – accessories identied by * are supplied as standard
with applicable DPI 612 models except DPI 612-XXX–00. Please
refer to the specication tables.
*Carrying case (P/N IO620-CASE-3)
A protective carrying case with
shoulder strap and large pocket for
accessories.
USB to RS 232 cable (P/N IO620-USB-RS232)
Connects the DPI 612 to an RS 232 interface
*Dirt and moisture trap
Prevents contamination of the DPI 612
pFlex pneumatic systems and cross
contamination from one device under
test to another. The trap connects
directly to the DPI 612 pFlex pressure
port and replicates the quick t
connection for compatibility with the
hose and adaptor kits. Not for use with
DPI 612 hFlex models.
*P/N IO620-IDT621: Maximum working
pressure 20 bar / 300 psi.
*P/N IO620-IDT622: Maximum working pressure 100 bar
/ 1500 psi.
Quick-to-t hose and adaptor system
These high pressure hoses and adaptors are compatible
with the pressure ports of the DPI 612 Flex, DPI 611
and DPI 620 series and are quick-to-t without tools or
Spare/replacement lithium ion battery for
the DPI611 and DPI 612 Flex series.
Spare/replacement mains adaptor
(P/N IO620-PSU)
A universal input mains adaptor
connects to the DPI 612 to charge the
battery and power the instrument. Input
voltage 100 to 240 VAC 50/60 Hz. Mains
socket adaptors are provided.
Battery charging station (P/N
CX6100GE)
This external battery charging
station and universal input mains
adaptor allows a spare battery to be
charged independently of the DPI
612 for minimum instrument down
time. A complete charge cycle takes
approximately 6.5 hours.
(CX6100GE Not available in Korea).
*IOHOSE-NP1: Pneumatic hose 1m/3.28ft. Maximum
pressure 20 bar/300 psi.
*IOHOSE-NP2: Pneumatic hose 2m/6.56ft. Maximum
pressure 20 bar/300 psi.
*P/N IO620-HOSE-P1:
Pneumatic hose 1m/3.28ft.
Maximum pressure 400
bar/5800 psi.
P/N IO620-HOSE-P2:
Pneumatic hose 2m/6.56ft.
Maximum pressure 400 bar/5800 psi.
*P/N IO620-HOSE-H1: Hydraulic hose 1 metre/3.28 ft.
Maximum pressure 1000 bar/15000 psi.
P/N IO620-HOSE-H2: Hydraulic hose 2 metre/6.56 ft.
Maximum pressure 1000 bar/15000 psi.
P/N IO620-BSP: Pressure adaptors G1/8 male and G1/4
male, G1/4 female, G3/8 female and G1/2 female.
P/N IO620-NPT: Pressure
adaptors 1/8” male and
1/4”male, 1/4” female,
3/8” female, and 1/2”
female.
P/N IO620-MET: 14 mm female and 20 mm female.
USB cable (P/N IO620-USB-PC)
Connects the DPI 612 to a PC.
IDOS to USB converter (P/N IO620-IDOSUSB)
Allows connection of an IDOS universal
pressure module to the DPI 612. P/N IO620USB-PC is also required to connect the
converter to the DPI 612 USB port.
Page 16
P/N IO620-COMP Comparator adaptor
For greater eciency, two test devices
can be connected at the same time.
The adaptor connects to the pressure
port of the DPI 612 and provides two
outlet ports.
*Pressure relief valve
Protects the PM 620 pressure module
and the device under test from overpressure.
Simple to use, single PC installation of Intecal v10
supporting portable eld calibrators including the
DPI 612 and x workshop/laboratory calibration
systems including PACE pressure controllers.
Compatible with Windows XP, 7 and 8 with 32
or 64 bit operating systems. Please refer to the
Intecal datasheet for more information